Hioki RM3542, RM3542-01 Instruction Manual

Instruction Manual
RM3542
RM3542-01
RESISTANCE HiTESTER
September 2013 Revised edition 3 RM3542A981-03 13-09H
Contents
i
1
Task-Oriented Reference...............................1
Measurement Flow.........................................2
Introduction.....................................................3
Verifying Package Contents...........................3
Safety Information..........................................5
Operating Precautions....................................7
Chapter 1 Overview 11
1.1 Product Overview and Features ........11
Block Diagram ..........................................14
1.2 Names and Functions of Parts ...........15
1.3 Screen Organization ..........................17
Chapter 2 Measurement Preparations 21
2.1 Connecting the Power Cord............... 22
2.2 Connecting Measurement Cables and
Test Fixtures ......................................23
Making Your Own Measurement Cables ..24
2.3 Turning the Power On and Off ...........25
Chapter 3 Measurement Settings (Basic Measurements)
27
3.1 Pre-Operation Inspection ...................27
3.2 Measurement Object Types ...............28
3.3 Setting the Measurement Speed .......29
3.4 Setting Measurement Start Conditions
(Trigger Source) .................................30
3.5 Selecting the Measurement Range ....31
3.6 Zero Adjustment .................................32
3.7 Judging Measured Values
(Comparator Function) .......................34
Enabling and Disabling the Comparator
Function .................................... ................35
Decide According to Reference Value and
Tolerance (REF% Mode) ..........................35
Decide According to Upper/Lower Thresholds
(ABS Mode) ............................................. 37
3.8 Confirming Faulty Measurements ......38
Chapter 4 Customizing Measurement Settings 41
4.1 Making Range-Specific Measurement Settings 41
4.2 Setting Pre-Measurement Delay ........42
4.3 Setting the Measurement Integration Time Option 44
4.4 Checking for Poor or Improper Contact
(Contact Check Function) ...................46
4.5 Improving Probe Contact
(Contact Improver Function) ...............47
4.6 Detecting Measurement Voltage Faults
(Voltage Level Monitor Function) .......49
4.7 Applying Current Only When Measuring
(Current Mode Setting) .......................50
4.8 Test for Short-Circuited Probe (Probe Short-Circuit Detection Function)
............................................................51
4.9 Comparing the Measurement Settings of Two Instruments
(Settings Monitor Function)................ 53
4.10 Retrying Measurement After a Fault
(Retry Function) ..................................56
4.11 Maintaining Measurement Precision
(Self-Calibration) ................................57
4.12 Compensating for Thermal EMF Offset (Offset Voltage Compensation - OVC) 57
Chapter 5 System Settings 59
5.1 Disabling and Enabling Key Operations 59
Disabling Key Operations
(Key-Lock Function) ................................. 59
Re-Enabling Key Operations
(Key-Lock Cancel) ................................... 60
5.2 Setting the Comparator Decision and
Key Beepers .......................................61
Enabling or Disabling the Key Beeper ..... 61
2 3 4
5
ii
Contents
Setting the Comparator Decision (“JUDGE”)
Beeper ..................................................... 62
5.3 Power Line Frequency Manual Setting 63
5.4 Setting the Clock ................................64
5.5 Adjusting Screen Contrast ..................65
5.6 Adjusting the Backlight .......................66
5.7 Initializing (Reset) ...............................67
Default Settings ........................................ 68
Chapter 6 Storing and Exporting Data 69
6.1 Storing Data at Specific Times
(Data Memory Function) .....................70
6.2 Store as soon as Measurement is
Stable (Auto-Memory Function) ........71
6.3 Performing Statistical Calculations on
Measured Values ...............................74
Using Statistical Calculations ................... 75
Confirming, Printing, and Erasing
Calculation Results .................................. 76
6.4 Auto-Exporting Measured Values (at End of Measurement)
(Data Output Function) .......................77
Chapter 7 Printing 79
7.1 Connecting the Printer ........................79
Connecting the Printer to the Instrument . 80
7.2 Instrument Settings ............................81
7.3 Printing ...............................................82
Printing Measured Values and
Comparator Decisions 82
Printing Statistical Calculation Results ..... 82
Chapter 8 External Control 85
8.1 External Input/Output Connector
and Signals......................................... 85
Connector Type and Signal Pinouts ........ 86
Signal Descriptions .................................. 87
8.2 Timing Chart .......................................88
8.3 Internal Circuitry .................................90
Electrical Specifications ........................... 91
Connection Examples ............................ .. 92
8.4 External I/O Settings .......................... 93
Setting End-of-Measurement Signal Output
Signal Setting) ...............................93
(EOM
Setting the Trigger (TRIG) Signal Logic ... 94
8.5 External Control Q&A ........................ 95
8.6 Supplied Connector Assembly ...........96
Chapter 9 Communications (RS-232C/ GP-IB Interface) 97
9.1 Overview and Features ......................97
9.2 Specifications .....................................98
9.3 Connecting .........................................99
Using the RS-232C Interface ................... 99
Using the GP-IB Interface
(RM3542-01 only) ........ ... ... ... .... ... ... ... ... . 100
9.4 Configuring the Communications
Protocol ............................................101
Configuring RS-232C Interface
Communications ....................................101
Configuring GP-IB Interface Communications
(RM3542-01 only) 102
Select the Measurement Data Transmission
Format (both RS-232C and GP-IB) 103
9.5 Communication Methods .................104
To cancel the Remote state
(enter the local state) .............................104
Message Format .................................... 105
Output Queue and Input Buffer .............. 108
Status Byte Register ..............................109
Event Registers ...................................... 111
Initialization Items ...................................114
Command Execution Time ..................... 115
Errors During Communications .............. 115
9.6 Message List ....................................116
Standard Commands ............................. 116
Device-Specific Commands ...................117
9.7 Message Reference .........................122
Message Reference Interpretation ......... 122
Standard Commands ............................. 123
Device-Specific Commands ...................127
9.8 Data Exporting Methods ..................148
9.9 Sample Programs ............................150
Using Visual Basic 5.0 or 6.0 ................. 150
Creation Procedure (Visual Basic 2005) 160Sample Programs (Visual Basic 2005) .. 162
9.10 Device Compliance Statement ........ 164
Chapter 10Specifications 167
Chapter 11 Maintenance and Service 179
11.1 Troubleshooting ...............................179
Inspection and Repair .............................179
11.2 Cleaning ........................................... 180
11.3 Error Displays and Remedies ..........181
11.4 Disposing of the Instrument .............183
Removing the Lithium Battery .................183
iii
Contents
Appendix A 1
Appendix 1Four-Terminal (Voltage-Drop)
Method......................................A 1
Appendix 2Effect of Thermal emf ...............A 2
Appendix 3Unstable Measurement Values.A 3
Appendix 4Rack Mounting..........................A 8
Appendix 5Dimensional Diagram .............A 10
Appendix 6Calibration...............................A 11
Appendix 7Adjustment Procedure ............A 13
Appendix 8Table of Commands Compatible
with the ADEX AX-162D.........A 14
Appendix 9Zero Adjustment .....................A 16
Index Index 1
6 7 8 9 10 11 12
Appendix
Index
iv
Contents

Task-Oriented Reference

Setting the measurement speed (p. 29)
1
Task-Oriented Reference
To minimize measurement error
To judge measurement results
To correct faulty measurements
To automatically store measured values
Setting the measurement speed integration time option (p. 44)
Zero-adjustment (p. 32)
Judge measured values (comparator function) (p. 34) Compare the measurement settings of two instruments
(Settings Monitor function) (p. 53)
Confirm faulty measurement (p. 38)
Improve probe contact (Contact Improver function) (p. 47) Test for short-circuited probe (probe short-circuit detection
function) (p. 51) Compare the measurement settings of two instruments
(Settings Monitor function) (p. 53)
Store as soon as measurement is stable (Auto Memory function) (p. 71)
To print measurement results Printing (p. 79)
To measure by PLC connection
(PLC: Programmable Logic Controller)
To connect to a computer Communications (RS-232C/GP-IB interface) (p. 97)
To automatically send measure­ment data to a computer
(RS-232C interface only)
To check operation
External control (p. 85)
Communications (RS-232C/GP-IB interface) (p. 97)
Export measurement values automatically (when finished measuring) (p. 77)
Setting Measurement Start Conditions (Trigger Source) (p. 30) Internal trigger (INT)
Calibration (p. A11)
2
Computer communications (p. 97)
Printing (p. 79)
External control (p. 85)
EXT I/O
RS-232C
GP-IB
Compare the settings of two instruments (p. 53)
SET MONITOR
Connecting the Power Cord (p. 22)
Turn the power on (p. 25)
Connect (p. 21)
Install (p. 7)
Connect the measurement cables (p. 23)
Set measurement
conditions (as needed)
Confirm the screen confi guration (p. 17)
Confirm the initial setup (p. 68)
Basic Settings (p. 27)
Configure settings for your
specific conditions (p. 41)
System-related settings
(p. 59)
When changing settings
Change basic settings such as measurement speed
Change to detailed settings (measurement conditions and system-related)
Set decision criteria
(p. 34)
Data transmission (p. 77) Printing (p. 79)
Computer communications
(p. 97)
External control (p. 85)
Statistical calculations
(p. 74)
Instrument interface settings must be configured before printing or using communications or remote control.
Turn the power off (p. 25)

Measurement Flow

Measurement Flow
Be sure to read the "Operating Precautions " (p. 7) b efore us e.
Installing, Connecting and Turning On
Settings
Calculation, Printing, Communication, and External Control Settings
When Finished

Introduction

When you receive the instrument, inspect it carefully to ensure that no damage occurred during ship­ping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Confirm that these contents are provide d.
Model RM3542 or
RM3542-01 (with GP-IB included)...............1
Power Cord (2-line + ground)(p. 22)EXT I/O Male Connector (p. 96)
Instruction Manual (This document)............ 1
Operation Guide.......................................... 1
Thank you for purchasing the HIOKI Model RM3542/ RM3542-01 Resistance HiTester. To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference.
Model RM3542-01 is the same as the RM3542, but with GP-IB included.
Registered trademarks
Windows and Visual Basic are registered trademark of Microsoft Corporation in the United States and/or other countries.

Verifying Package Contents

Inspection
3
Introduction
Content confirmation
4
Measurement Probes and Fixtures
(connect to measurement jacks)
Model 9140 4-terminal Probe
Model 9262 Test Fixture
Model 9263 SMD Test Fixture
Alligator-clip-type measurement probes. These general-purpose dual-electrode clips fit a wide range of conductor thicknesses.
This fixture is for measuring lead components. (less than 10 m residual resistance after zero adjustment)
This fixture is for measuring chip components. (less than 10 m residual resistance after zero adjustment)
Interface Cables
Model 9637 RS-232C Cable
(9-pin to 9-pin/ crossover cable)
Model 9638 RS-232C Cable
(9-pin to 25-pin/ crossover cable)
Model 9151-02 GP-IB Connector Cable (2m)
Verifying Package Contents
Options
Contact your dealer or Hioki representative for details.
5

Safety Information

Safety Information
This instrument is designed to comply with IEC 61010 Safety S t andards , and has b een thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. Using the instru­ment in a way not described in this manual may negate the provided safety features. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from instrument defects.
This manual contains information and warnings essential for safe operation of the instrument and for main­taining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions.
Safety Symbols
In the manual, the symbol indicates particularly important information that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with the symbol) before usi ng the relevant function.
Indicates AC (Alternating Current).
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents a significant hazard th at could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or dama ge to the instru­ment.
Indicates advisory items related to performance or correct operation of the instru ment.
Symbols for Various Standards
This symbol indicates that the product conforms to regulations set out by the EC Directive.
WEEE marking: This symbol indicates that the electrical and electronic appliance is put on the EU market after August 13, 2005, and producers of the Member States are required to display it on the appliance under Arti­cle 11.2 of Directive 2002/96/EC (WEEE).
6
Safety Information
Other Symbols
Indicates the prohibited action.
(p. )
*
[ ] SET
(Bold characters)
Unless otherwise specified, “Windows” represents Windows 95, 98, Me, Widows NT4.0, Windows 2000, Windows XP, or Windows Vista.
Click: Press and quickly release the left button of the mouse. Double click: Quickly click the left button of the mouse twice.
Indicates the location of reference information. Indicates that descriptive information is provided below. Square brackets indicate instrument display labels (such as setting item names).
Bold characters within the text indicate operating key labels.
Accuracy
We define measurement tolera nces in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values, with the following mean­ings.
f.s. (maximum display value)
The maximum displayable value. This is usually the name of the currently selected range.
rdg. (reading or displayed value)
The value currently being measured and indicated on the measuring instrument.
dgt. (resolution)
The smallest displayable unit on a digital measuring instrument, i.e., the input value that causes the digital display to show a “1” as the least-significant digit.

Operating Precautions

The instrument can be used with the stand (p. 16). It can also be rack-mounted. (p. A8).
50 mm or more
10 mm or more
Rear
50 mm or more
Unplugging the power cord kills power to the instrument. Be sure to provide enough unobstructed space to unplug the power cord immediately in an emergency.
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Preliminary Checks
• Before using the instrument for the first time, verify that it operates normally to ensure that no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
• Before using the instrument make sure that the insulation on the powe r co rd is unda mag ed and that no bare conductors are improp erly exposed. Using the instru ment in such condi­tions could cause an electric shock, so contact your dealer or Hioki representative for repair.
• Before using the instrument, make sure that the insulation on the measurement cables is undamaged and that no bare metal is improperly exposed. If there is any damage, mea­sured values may be unstable and measurement errors may occur.
7
Operating Precautions
Instrument Installation
Storage temperature and humidity: -10°C to 50°C at 80% RH or less (non-condensating) Operating temperature and humidity: 0 to 40°C at 80% RH or less (non-condensating)
Avoid the following locations that could cause an accident or damage to the instrument.
Installation Precautions
Exposed to direct sunlight Exposed to high tempera­ture
Exposed to liquids Exposed to high humidity or condensation
Exposed to high levels of particulate dust
Subject to vibration
In the presence of corrosive or ex­plosive gases
Exposed to strong electromag­netic fields Near electromagnetic radiators
Near induction heating systems (e.g., high-frequency induction heating systems and IH cooking utensils)
• The instrument should be operated only with the bottom downwards.
• Do not place the instrument on an unstable or slanted surfa ce.
8
Operating Precautions
Handling the Instrument
Do not allow the instrument to get wet, and do not take measurements with wet
hands. This may cause an electric shock.
Do not attempt to modify, disassemble or repair the instrument; as fire, electric
shock and injury could result.
To avoid damag e to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
This instrument may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to redu ce electromagnetic emissions to p re­vent interference to the reception of radio and television broadcasts.
Handling the Cords and Probes
• Avoid stepping on or pinching cables, which could damage the cable insulation.
• To avoid breaking the cables, do not bend or pull them.
• To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it from the power outlet.
• When disconnecting the BNC connector, be sure to release the lock before pulling off the connector. Forcibly pulling the connector without releasing the lock, or pulling on the cable, can damage the connector (p. 23).
• Use only the specified connection cables. Using a non-specified cable may result in incor­rect measurements due to poor connection or other reasons.
• Before using a test fixture, read the instructions provided with it.
Before Turning Power On
Before turning the instrument on, make sure the supply voltage matches that indi­cated on its power connecto r. Connection to an improper supply voltage may damag e the instrument and present an electrical hazard.
Before Connecting EXT I/O
To avoid electric shock or damage to the equipment, always observe the following precautions when connecting to the EXT I/O connector.
Always turn off the power to the instrument and to any devices to be connected
before making connections.
Be careful to avoid exceeding the ratings of external terminals (p. 91).
During operation, a wire becoming dislocated and contacting another conductive
object can be serious hazard. Make sure that connections are secure and use screws to secure the external connectors.
Properly insulate any devices and mechanisms to be connected to the EXT I/O con-
nector.
The ISO_5V pin of the EXT I/O connector is a 5V power output. Do not apply ext ernal
power to this pin.
9
Operating Precautions
Before Measuring
• Do not apply any voltage to the measurement jacks. Doing so could damage the instru­ment.
• Never attempt to measure at a point wher e voltage is pr esent. In particular, do not measure a transformer or motor immediately after a temperature increase test or withstand-voltage test, as the instrument could be damaged by induced voltage or residual charge.
• Battery internal resistance cannot be measured with this instrument. It will sustain damage. To measur e battery internal resistance, we recommend the HIO KI 3554, 3555, BT3562, BT3563, and 3561 Battery HiTesters.
• To obtain the guaranteed measurement accuracy, allow at least 30 minutes warm-up.
• The instrument internally retains all settings (but not measured values), such as measure­ment range and comparator settings. However, measurement settings made through the RS-232C or GP-IB interface are not memorized.
• In the 100 and higher ranges (LOW POWER set to OFF), thermal emf can cause mea­surement errors.
• The DC resistance of a power transformer cannot be measured. When measuring objects with a large L, such as choke coils and other inductors, measured values may be unst able. In such cases, contact your dealer or Hioki rep r esentative.
• Carefully insulate all H cannot be performed and an error will occur if core and shield wires touch.
CUR
, H
POT
, L
POT
, and L
wiring. Proper 4-terminal measurements
CUR
10
Operating Precautions

1.1 Product Overview and Features

The Hioki RM3542 Resistance HiTester employs the 4-terminal method to quickly and accurately measure the DC resistance of components such as resistors and ferrite beads. It includes advanced contact-check, comparator, and data output functions. The intuitive user interface and high noise immunity are ideal for use with taping machines and separators.
Resistance Measurement
The factory defaults (initial settings) are optimized for chip-component resistance measurements. The RM3542 can also measure devices that are otherwise difficult to mea­sure with high current, such as fer­rite-bead and small multilayer inductors (low-power resistance measurement, p. 28).
Interface Communications
Connect the instrument to a controller via the RS-232C or GP-IB interface to control measurement data acquisi­tion (p. 97).
Send Measurement Data and Calculation Results to a Printer
Use a commercially available print­er with a serial interface to print measured values and calculation results (p. 79).
Save and Output Measured Values
Measured values can be stored in internal memory (p. 69) Statistical calculations can be per­formed on the stored data, which can be transferred to a computer in batch form (however, stored data cannot be confirmed internally).
Judge Measured Values
Measured values are compared with a pre-specified referenc e value or thresholds, and the result is out­put externally and indicated by the COMP indicators (comparator func­tion, p. 34).
Connect a PLC or I/O Board
To control from a PLC, con­nect to the EXT I/O connec­tor. In addition to comparator results, various measurement anomaly signals can be out­put (p. 85).
Upper limit
Lower limit
Compare Two Instrument’s Setting Conditions
When measuring with two inter­connected instruments, settings are compared, and an alarm is output and measurement is inhib­ited if the settings differ (Settings Monitor function, p. 53).
Optional Hioki probes and fixtures are available to connect to the mea­surement jacks (BNC jacks, p. 4). Alternatively, commercially available cables such as 1.5D-2V coax can be used (p. 24).
Overview Chapter 1
1.1 Product Overview and Features
11
1
12
Ultra Fast and Accurate Measurements Increase Productivity
The factory default settings are optimized for chip­component resistance measurements. Enhanced contact-to-measurement and contact-check-to-deci­sion times are only 1 ms. The offset-voltage compen­sation (OVC) function minimizes the effects of thermal emf when using low-power resistance measurement and the 100 m
to 10 ranges (p. 57). Measurement
results are judged pass/fail with 10 ppm resolution, ideal for high-speed Class B resistor testing.
High-Speed Dat a Output and Ample Memory
The Data Output function transfers measured data at 5 ms/sample, even via RS-232C. Up to 30,000 mea­surements can be stored, and all data can be exported at the end of measuring each reel. This function is ide­al for system setup, debugging and process manage­ment.
Multiple Interfaces
EXT I/O is isolated from the measurement and control circuits to provide noise immunity (p. 85). All data can be acquired in real time using the built-in
38.4-kbps high-speed RS-232C interface. Connect the commercially available printer with a serial inter­face to print measured values and statistical calcula­tion results (p. 79). The GP-IB interface is available for Model RM3542-01 (specified when ordering, p. 97).
Low-Power Function (p. 28)
For ranges from 1000 m to 1000 , low-power re­sistance measurement is provided to minimize mea­surement current. Safely measure devices that are otherwise difficult to measure with high current, such as ferrite-bead and multilayer inductors.
Clearly Visible Display and Intuitive Opera­tion
High-contrast LCD provides clear visibility, helping avoid setting mistakes. The optimum range is selected automatically when comparator thresholds are en­tered.
Auto Memory Function Convenient for Sam­pling Tests (p. 71)
The auto memory function is convenient for sampling tests after screen-printing. Measured values are auto­matically acquired as soon as they stabilize, and statis­tical calculations proceed until the specified quantity is obtained, upon which an alert notification (alarm) oc­curs. Selecting [PRINT] (screen display) prints mea­sured values and statistical calculation results (p. 82).
Fixtures for Component Measurements (p. 4)
The BNC-type measurement jacks exhibit good noise immunity. Ready availability and easy assembly en­sure smooth system setup. Various test fixtures are available for Hioki LCR HiTesters.
Features
1.1 Product Overview and Features
Reliable Contact Checkin g (p. 46)
Contact checking (that was previously performed be­fore and after measuring) is now performed during mea­surement, so probe bounce and contact resistance fluctuations can be detected. Contact checking time can be shortened, improving tact times.
Features
Measurement Circuit Strongly Immune to Contact Resistance Fluctuations
The effects of contact resistance fluctuations are re­duced even when scattering occurs near the end of probe life. Such effects are minimized by the fast re­sponse of the measurement circuit.
Contact Improver Function (p. 47)
The Contact Improver function improves bad contacts between probes and test samples. Contacts errors are reduced by penetrating oxidation and impurities be­tween probes and samples. Reducing contact errors can increase productivity and quality. The intensity of the contact improver function can be adjusted accord­ing to probe type.
Reject Faulty Data – Voltage Level Monitor Function (p. 49)
When the contact resistance of the H
CUR
and L
CUR
leads fluctuates, the measurement current changes momentarily. Such momentary changes are not detect­able by typical contact checking. The Voltage Level Monitor function detects a contact error if the detection voltage changes significantly, which can increase the reliability of measured values.
Minimize Human Error and Risk – the Settings Monitor Function (p. 53)
If the settings of two instruments are different, trigger­ing is inhibited and an alarm notification is generated to avoid setting mistakes due to human error.
Reliable Four-Terminal Measurement – Probe Short-Circuit Detection Function (p. 51)
Four-terminal measurements are inhibited when a conductive foreign object is present between the POT and CUR probe tips. Short-circuit probe anoma­lies are detected by checking the resistance between these tips when not measuring.
Strong Electrical Noise Immunity
The specified measurement accuracy is achieved even with ±1.5 kV mixed pulse noise. The floating measurement section design is highly impervious to electrical noise, minimizing the effect on measured values even when turning large-induction motors on and off. The free-range power supply input (90 to 264 V) is practically unaffected by voltage fluctuations, so stable measurements are possible even in under poor power conditions.
Previously
Model RM3542
Contact Condition
Contact Check
Measuring
Probe Bounce
Contact Condition
Detection Voltage
Error
Good Contact
Poor Contact
Excessive detection voltage fluctuation error
Good Contact
ERROR
Measurement Status
Contact Check
Contact Condition Contact Improver
Function
Contact Improvement
Measuring
Checking
ONON
DUT
Foreign object
POT
CUR
DUT electrode
13
1.1 Product Overview and Features
1
14
AB
C
D
E
F
G
H
I
E
D
1.1 Product Overview and Features

Block Diagram

Constant current (determined by the measurement range) is applied between the H
age is measured between the H voltage (B) by the constant current flow (A).
POT
and L
terminals. The resistance value is obtained by dividing the measured
POT
CUR
and L
terminals while volt-
CUR
The effects of large offset voltage such as from thermal emf are reduced by current flowing in the positive and negative
directions (A).
The constant current source (A) and voltmeter (B) circuit designs are largely unaffected by contact resistance.
Faulty measurement values caused by unstable or chattering contact conditions can be eliminated by monitoring (C)
the detection voltage (B) waveform (Volta ge Le ve l Mo ni to r fu nction).
Stable measurements are ensured by providing sufficient integration time (the default setting is 0.3 ms). (The integra-
tion time can be reduced to 0.1 ms to support even higher speed (B).)
Before measuring, the Contact Improver circuit (D) optimizes contact when the probes touch the DUT.
By also performing contact checking (E), short circuits between CUR and POT terminals caused by a clogged probe tip
can be detected (probe short-circuit detection function).
When measurement starts, the contact check circuit (E) and constant current monitor (F) are activated to monitor for
fault conditions while measuring. The dual-CPU (C and G) design provides ultra-high-speed measurements and fast system response.
Immunity from electrical noise is provided by isolation between the Measurement and Control blocks (H).
The auto-ranging 100-to-240 V switching power supply (I) can provide stable measurements even in poor power quality
environments.
15
Display Screen
Monochrome graphical LCD There are three general screen types: Measurement, Basic Set­tings and Detailed Settings.
Screen Configurations (p. 17)
F1 to F4ÉLÅ[
(FÉLÅ[: ëçèÃ)
âÊñþâEë§Ç…ï\é¶ÇŠÇÍÇÈçÄ ñÇšëIëšÇµÇ‹Ç²ÅB
Judging measured values (p. 34)
(Comparator function) Set a reference value and range for judging measurements, which can be confirmed by the COMP indicators.
Select this to judge measured val­ues relative to a reference value and tolerance (%). The REF% set­ting display appears.
Press to judge measured values relative to upper/lower thresholds. The ABS setting display appears.
Setting
POWER Button
Turns the instrument on and off (p. 25).
Unlit: power off
(when no power supplied)
Red light: power off
(while power is supplied)
Green light: power on
Power On/Off
Entering numerical values
Enter a numerical value. (we call these the “tenkeys”)
Switches the sign of a numer­ical value.
Selects the unit of measure. Sets the tolerance values.
Deletes the value in the se­lected field.
Accepts the displayed com­parator threshold values.
Aborts comparator setting and returns to the previous display.
Measurement Terminals
Connect measurement cables or a fix­ture (p. 23).
H
CUR
jack: Current source terminal
H
POT
jack: Detected voltage high
terminal
L
POT
jack: Detected voltage low ter-
minal
L
CUR
jack: Measurement current
detected terminal
GUARD jack: Shield (measurement
ground) terminal
Connecting Probes
COMP indicator LEDs
Indicate the decision re­sult of the measured value (p. 34).
Hi Measured value is
above upper limit
IN Pass (meets crite-
ria)
Lo Measured value is
below lower limit
Viewing
Comparator
Results
Viewing Measured
Values and Settings
Audible Alarm (beeper)
Selecting the setting contents
(“the F keys”) F1 to F4 keys Selects the corresponding item on the right side of the display.
Cursor keys Move among the displayed setting items. The cursor location is indicat­ed by reverse characters.

1.2 Names and Functions of Parts

1.2 Names and Functions of Part s
Front Panel
1
16
Manufacturer's Serial Number
Shows the serial number. Do not remove this label, as it is required for product support.
Connect the supplied power cord (p. 22).
Connecting the Power Cord
RS-232C Connector
The RS-232C interface can be used to connect to a PLC or computer (p. 97). It is also used by the commercially available printer with a serial interface (p. 79).
RS-232C Communications
Printer Output
GP-IB Connector (RM3542-01 only)
The GP-IB interface can be used to connect to a computer (p. 97).
GP-IB Communications
SET MONITOR Connector
Connect another RM3542 here to compare instru­ment settings (p. 53).
Compare Two Instruments’
Settings
EXT I/O Connector
Connect to a PLC or I/O board to control measurement start, and to acquire com­parator results (p. 85).
External Control
Bottom Panel
Stand
When using the stand
Extend the stand until it clicks into place. Make sure to extend both legs of the stand.
Collapsing the stand
Fold in the stand until it clicks into place.
This instrument can be rack mounted.
See: "Appendix 4 Rack Mounting" (p. A8)
Parts removed from this instrument should be stored in a safe place to enable future reuse.
Do not apply heavy downward pressure with the stand extended. The stand could be damaged.
1.2 Names and Functions of Parts
Rear Panel
17
Detailed Settings Screens (p. 19)
Measurement Settings Screens [MEAS SETTINGS]
Data Settings Screens [DATA SETTINGS]
System Screens [SYSTEM]
Measurement Screen (p. 18)
Basic Settings Screen (p. 18)
Return to Previous Screen
Indicates a Continued Screen
Return to Previous Screen

1.3 Screen Organization

1.3 Screen Organization
The instrument has three general display screen types: Measurement, Basic Settings and Detailed Settings. Refer to "11.3 Error Displays and Re medies" (p. 181) for error displays.
The screen examples in this guide appear reversed (black on white) for best visibility on the printed page. However, the instrument screens can actually be displayed only as white characters on black background.
1
18
Measurement Screen
This screen normally appears while measuring. View cur­rently measured values and measurement conditions. Some parts of the display depend on the comparator mode and other settings.
Measured Value
Settings Menu (corresponding to F keys)
Displayed contents depend on the curre nt function set­tings. Parentheses ( ) indicate the corresponding F-key
MENU (F1) Displays the Basic Settings screen PRINT (F2) Print (p. 81)
Appears only when the interface is set for the printer
ST AT (F3) Statistical calculation results (p. 74)
Appears only when statistical calcu­lation is enabled
NUMBER (F4) Set Auto-Memory number (quantity)
(p. 71) Displays the number of stored data
items and the number of passed and failed products at the lower left.
UNDO (F3) Deletes the previously stored mea-
surement and calculation result (only one can be deleted) (p. 76) Appears only when calculation re­sults are displayed
ALLCLR (F4) Clears all memory and calculation
(p. 76) Appears only when calculation re­sults are displayed
LOCAL (F1) Enables local controll (p. 104) UNLOCK (F1) Cancels the key-lock state (hold 1s)
(p. 60)
Criteria Setting Values (p. 34) Displayed contents depend on the selected comparator mode. (Ex.: REF% mode)
To display the Basic Settings screen
Measurement Conditions
Shows current setting contents. Displayed contents de­pend on the current settings.
INT/ EXT Trigger source type (p. 30) Measurement range (p. 31) FAST/ MED/ SLOW Measurement speed (p. 29) 0ADJ/
OFF (not shown)
Appears only when zero-adjust is enabled (p. 32)
OVC/
OFF (not shown)
(OVC: Offset Voltage Compensa­tion) Appears only when OVC is en­abled (p. 57)
LP/
OFF (not shown)
Appears only when the Low- Power Resistance function is enabled (p. 28)
NUM Appears only when tenkey input is
enabled
RMT Remote control (p. 104) M.LOCK Disables all operations except
comparator settings (p. 59)
F.LOCK Disables all operations including
comparator settings (p. 59)
Basic Settings Screen
Make basic measurement condition settings on this screen. Measurement speed and range can be changed while viewing measured values. (Trigger source: INT)
TRG
Change trigger source (measurement start control method) (p. 30) When EXT triggering is enabled, [MANU] ap-
pears beside the F4 key (press to trigger measurement manually).
RANGE Change range (p. 31) SPEED Change measurement speed (p. 29) 0ADJ Turn zero-adjust on/off (p. 32) COMP Turn comparator on/off (p. 34) LOCK Enable/disable key lock (p. 59)
Select measurement condition settings Move with cursor keys.
Measurement condition setting selections
Return to previous screen
To display the [MEAS SETTINGS] screen To display the [DATA SETTINGS] screen To display the [SYSTEM] screen
MISC To display the Detailed Settings screen
1.3 Screen Organization
Measurement Settings Screen [MEAS SETTINGS]
Set instrument system-related settings on this screen.
SET MONITOR Turn two-instrument setting comparison
on/off (p. 53)
PROBE CHECK Turn probe short-circuit detection on/off
(p. 51)
RETRY Retry function setting (p. 56) TRIG EDGE (External I/O) Set trigger rising/falling
edge (p. 94)
EOM (External I/O) Set EOM
(end-of-mea-
surement) signal (p. 93)
INTERFACE Communications interface settings
(p. 101)
LOW POWER Low-power resistance component mea-
surement on/off (p. 28)
JUDGE BEEP Comparator beeper settings (p. 62) KEY BEEP Key beeper on/off (p. 61) CLOCK (Y-M-D) Set internal clock (p. 64) LINE FREQ Set power source frequency (p. 63) CONTRAST Adjust screen contrast (p. 65) BACKLIGHT Adjust screen backlight (p. 66) RESET Initialize (p. 67) ADJUST Instrument Adjustment (p. A13)
System Screen [SYSTEM]
Shows detailed settings for measurements. Adjust the measurement speed, stability and measurement fault detection functions.
DELAY1 Adjust the delay from probing to trigger
input (p. 42)
DELAY2 Adjust measurement object electrical re-
sponse (p. 42)
INT (FAST/ MED/ SLOW)
Make fine adjustment to integration time (p. 44)
CONT CHECK Contact check threshold setting (p. 46) CONT IMP Contact Improver function setting (p. 47) VOL T MONITOR Voltage level monitor function setting
(p. 49)
CURRENT MODE Current mode setting (p. 50)
Data Settings Screen (Save, Analysis, and Output) [DATA SETTINGS]
These are settings for memory and statistical calculation func­tions.
AUTO MEMORY Turn Auto-Memory on/off (p. 71) STATISTICS Statistical calculation on/off (p. 74) DATA OUT Automatically output measured values
(communications) (p. 77)
Detailed Settings Screens
19
1.3 Screen Organization
1
20
1.3 Screen Organization
21
Install this instrument (p. 7)
1
Rear PaneFront Panel
Connect measurement cables (p. 23)
Turn the power on (p. 25)
5
Connect to the test sample When finished measuring, turn the
power off (p. 25).
Connect the external interface
(as needed)
4
• Using the printer (p. 79)
• Using the RS-232C or GP-IB interface (p. 97)
• Connecting to a PLC or I/O board (p. 85)
• Automatically comparing the settings of two instruments (Settings Monitor function) (p. 53)
Connect the power cord (p. 22)
2 3
2
3
4
5
Make instrument settings (p. 27)
6
Measurement
Preparations Chapter 2
Be sure to read the "Operating Precautions" (p.7) before installing and connecting this instrument. Refer to "Appendix 4 Rack Mounting" (p. A8) for rack mounting.
2
22
Rear Panel
1 Confirm that the mains supply voltage matches the
instrument, and connect the power cord to the power inlet on the instrument.
2 Plug the power cord into the mains outlet.
The POWER button on the front panel lights red. In event of a power outage, operation resumes with the same
settings when power is restored (breaker reset, etc.).
Power inlet

2.1 Connecting the Power Cord

2.1 Connecting the Power Cord
Before turning the instrument on, make sure the supply voltage matches that indi-
cated on its power connector. Connection to an improper supply voltage may dam­age the instrument and present an electrical hazard.
To avoid electrical accidents and to maintain the safety specifications of this instru-
ment, connect the power cord only to a 3-contact (two-conductor + ground) outlet.
Before using the instrument, make sure that the insulation on the power cord is
undamaged and that no bare conductors are improperly exposed. Any damage could cause electric shock, so cont act your dealer or Hioki representative.
To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it from the power outlet.

2.2 Connecting Measurement Cables and Test Fixtures

Example of defeated guard measurement
Connecting measurement cables
Connecting a fixture
Connect directly to the measurement jacks with the label side up, and affix with the levers on the left and right.
Connect the red plugs to the H
CUR
and H
POT
jacks, and the
black plugs to the L
CUR
and
L
POT
jacks.
Black plugs
Red plugs
BNC Jack Guide Pins (on the instrument)
Lock
BNC plug slots
2
1
Align the slots in the BNC plug with the guide pins on the jack on the instrument, then push and twist the plug clockwise until it locks. Disconnecting BNC connectors Push the BNC plug, twist it counterclock­wise, and pull it out.
Making your own probes and extenders (p. 24)
Connection Methods
2.2 Connecting Measurement Cables and Test Fixtures
23
Connect your measurement cables, optional Hioki probes or test fixture to the measurement jacks. Refer to "Options" (p.4) for details. See the instructions provided with the fixture for operating details.
• Do not apply a voltage to the measurement terminals. Doing so may damage the unit.
• When disconnecting the BNC connector, be sure to release the lock before pulling off the connector. Forcibly pulling the connector without releasing the lock, or pulling on the cable, can damage the connector.
• We recommend using optional Hioki fix­tures.
• Use the GUARD jack only for Faraday shield, and avoid more than 10 mA cur­rent flow. This jack is not for guarding network resistance measurements.
2
24
Wiring Diagram
Before Wiring
• Twist together the H
POT
and L
POT
wires, and the H
CUR
and L
CUR
wires. If not twisted together, mea­sured values may be unstable and errors occur whe n measuring with low-power resistan ce, or low resis­tance values.
• Refer to the block diagram (p. 14) for internal circuit details.
• Probes and measurement objects should be shielded at BNC or GUARD jack potential.
• Measurement cable length should not exceed 2 m (with conductor resistance 500 m/m or less). Long cables are more susceptible to noise, and measured values may be unstable.
• Extensions should maintain the four-terminal structure. If con verted to a two-te rminal circuit in the wiring, correct measurement may not be possible due to the effects of wiring and contact resis­tance.
• Cables and measurement objects should be shielded.
• After extending measurement cables, confirm operation and accuracy ("Measurement Specifica­tions" (p.176)).
• If cutting the ends off of optional measurement cables, make sure that the shield does not touch the center conductor of the H
CUR
, H
POT
, L
POT
and L
CUR
cables. Correct measurem ent is n ot pos -
sible with a shorted cable.
Extending Measurement Cables
Observe the following when extending measurement cables:
2.2 Connecting Measurement Cables and Test Fixtures

Making Your Own Measurement Cables

Recommended Measurement Cable Specifications
Conductor resistance Capacitance 150 pF/m or less
Cable dielectric material Polyethylene (PE), Teflon* (TFE), polyethylene foam (PEF)
Connector insulating material Teflon* (TFE ), polybutylene terephtalate (PBT)
Length 2 m or less
500 m
/m or less
Insulation resistance at least 10 G
Insulation resistance at least 10 G
Example: JIS standard 3C-2V, 1.5D-2V MIL standard RG-58A/U * Teflon is a registered trademark of E. I. du Pont de Nemours and Company.
Loading...
+ 188 hidden pages