Hioki PW3337 Service Manual

Part names

Switches the display channel (when in the shift state, changes the rectifier).
Switches display parameters.
Displays measured values and settings.
value displays.
SHIFT: Press to access functions shown in blue.
POWER switch
Warning display lamps (for each channel)
Switches the voltage range. Switches the current range.
Switches the voltage and current range channel (applies to both voltage and current ranges).
Status display lamps (for each channel)
Function settings
PW3336
PW3336-01, PW3336-02, PW3336-03
PW3337
PW3337-01, PW3337-02, PW3337-03

POWER METER

Measurement Guide

June 2013 Edition 1 Printed in Japan
PW3336A984-00 13-06H
Thank you for purchasing the HIOKI PW3336, PW3336-01, PW3336-02, PW3336-03, PW3337, PW3337-01, PW3337-02, or PW3337-03 Power Meter. This Measurement Guide introduces basic measurement methods to individuals using the instrument for the first time. Be sure to review the Instruction Manual before using the instrument.
1. Connecting the instrument
Example 1: 1P2W
Before connecting the instrument, read “Operating Precautions” in the Instruction Manual carefully.
Inverter
Power supply
Wiring diagram
CH(3)
Example 2: 3V3A
Motor
Inverter
Power supply
Wiring diagram
CH(3) CH(1)CH(2)
2. Turning on the instrument
: Power ON
21
Connect the power cord. Turn on the POWER switch.
Do not activate input to the instrument before it has been turned on.
3. Selecting the wiring mode
Display
Description
1P2W×3 (PW3336: 1P2W×2)
Allows you to measure three 1-phase/2­wire circuits. (PW336: Allows you to measure two such circuits.)
1P3W&1P2W (PW336: 1P3W only)
Measures a 1-phase/3-wire circuit with CH1 and CH2. The PW3337 can simul­taneously measure a 1-phase/2-wire circuit with CH3.
3P3W&1P2W (PW336: 3P3W only)
Measures a 3-phase/3-wire circuit with CH1 and CH2 and measures active power using the two-power-meter method. The PW3337 can simultane­ously measure a 1-phase/2-wire circuit with CH3.
Display
Description
3P3W2M
Measures a 3-phase/3-wire circuit with CH1 and CH2 and measures active power using the two-power-meter method. Calculates line voltage and phase current, which are not actually measured, using vector calculations and displays the result as the CH3 voltage and current.
3V3A
Measures a 3-phase/3-wire circuit with connections for all line voltages and phase currents and measures the active power using the two-power-meter method.
3P3W3M
Measures a 3-phase/3-wire circuit. Uses the same con­nections as the 3V3A wiring mode but converts line volt­ages into phase voltages by means of vector calculations and displays the results.
3P4W Measures a 3-phase/4-wire circuit.
Press to select the wiring mode.
Display Description
V Voltage (U) A Current (I)
W Active power (P)
VA Apparent power (S)
var Reactive power (Q)
PF Power factor (λ)
° Phase angle ()
V, Hz
or
A, Hz
Frequency (f)
Wh, + Positive direction Wh
Wh, - Negative direction Wh
Wh Total Wh
4. Selecting the display
Press to light up the blue lamp before setting functions shown in blue underneath the keys.
Display Description
Ah, + Positive direction Ah
Ah, - Negative direction Ah
Ah Total Ah
TIME Integration time
V, pk Voltage waveform peak value
A, pk Current waveform peak value η1 or η2 Efficiency (η) CF, %, V Voltage crest factor
CF, %, A Current crest factor
T.AV, W Time average active power
T.AV, A Time average current
RF, %, V Voltage ripple rate
Shift state
Select the parameters to display by pressing the parameter key.
Display
Description
CH1 Lights up when displaying CH1 measured values. CH2 Lights up when displaying CH2 measured values. CH3 Lights up when displaying CH3 measured values.
sum
Lights up when displaying measured values for the sum of the selected display parameters when using a wiring mode other than 1P2W.
Display
Description
AC+DC
AC + DC measurement; display of true RMS values for both voltage and current
AC+DC
Umn
AC + DC measurement; display of mean value rectified RMS converted values for voltage and true RMS values for current
DC DC measurement; display of the DC component only AC AC measurement; display of the AC component only
FND
Display of the fundamental wave component from harmonic measurement
Select the rectifier by pressing the CH key while holding down the SHIFT key.
Press the CH key to select the channel to display.
Display Description
RF, %, A Current ripple rate THD, %, V Total harmonic voltage distortion THD, %, A Total harmonic current distortion
FND, θ, V, °
Inter-channel voltage fundamental wave phase difference
FND, θ, A, °
Inter-channel current fundamental wave phase difference
HRM, V, LEVEL Harmonic voltage RMS value HRM, A, LEVEL Harmonic current RMS value
HRM, W, LEVEL Harmonic active power RMS value
HRM, %, V, HD% Harmonic voltage content percentage
HRM, %, A, HD% Harmonic current content percentage
HRM, %, W, HD%
Harmonic active power content percentage
5. Selecting the voltage and current ranges
To select auto-range operation, set after pressing .
The lamp for the selected range key will light up, and the display value will correspond to the range.
Voltage range Current range
Example: Voltage range: CH1, 300 V; current range: CH1, 20 A
Switches the channel for which the range is being set (applies to both voltage and current ranges).
2 Integration will start, and the RUN lamp will light up.
6. Performing integration
Integration is performed to measure the amount of power consumed in the set amount of time. In the example shown, the integration time has been set to 24 hours.
Performing integration Stopping integration
1 Press .
2 Integration will stop, and the RUN lamp will flash.
1 Press while the RUN lamp is lit up.
Troubleshooting
While integration is being performed ( the RUN lamp is lighting) or stopped (the RUN lamp is flashing), range, wiring mode or other settings cannot be changed. To change settings or perform functions, follow the procedure describ ed below. Integrated value will be reset by this operation.
Procedure
If integration is stopped (if the RUN lamp is flashing), skip to Step 2.
1 Press to stop integration.
The RUN lamp, which had been lit up, will start flashing.
2 Press to place the instrument in the shift state.
The blue lamp will light up.
3 Press .
The RUN lamp will turn off, and integrated values will be reset. You can now change settings.
* Wiring mod e, current input method, frequ ency measure ment range (zero-cross filter), timeout, inte gration time, synchronization
source, VT ratio, CT ratio, harmonic analysis order upper limit value, zero-adjustment (can be performed while integration is stopped)
If the instrument displays an error, see “Troubleshooting” in the instruction manual (for example , etc.).
If an error is being displayed
Even if the keys are pressed, no settings can be changed
( is displayed).
?
7. Turning off the instrument
: Power OFF
Once measurement is complete, turn off the instrument and disconnect connection cables and other wiring.
Turn off the POWER switch.