P9000-01
P9000-02
DIFFERENTIAL PROBE
Instruction Manual
EN
Apr. 2021 Revised edition 2
P9000A981-02 21-04H
Warranty
Warranty malfunctions occurring under conditions of normal use in
conformity with the Instruction Manual and Product Precautionary
Markings will be repaired free of charge. This warranty is valid for a
period of three (3) years from the date of purchase. Please contact
the distributor from which you purchased the product for further information on warranty provisions.
Introduction
Thank you for choosing the Hioki P9000-01, P9000-02 Dierential
Probe. Preserve this manual carefully and keep it handy to make full
use of this product for along time.
Overview
The product is a dierential probe that is connected to a Memory
HiCorder input unit. The probe divides two high voltages input to its
high and low terminals by 100 or 1000 and then outputs the dierence as a signal relative to ground. The P9000-02 includes AC RMS
value output functionality.
Power can be supplied via the optional AC adapter or a commercially available USB cable.
Verifying Package Contents
When you open the package, carefully inspect the product to ensure
that everything is in good condition, and that no damage occurred
during shipping. Carefully check the accessories and connectors.
If the product seems to have been damaged or does not work as
specied, contact your authorized Hioki distributor or reseller.
Cleaning
If the product becomes dirty, slightly moisten a soft cloth moistened
with water or a neutral detergent and wipe the product clean.
IMPORTANT
Never use solvents such as benzene, alcohol, acetone, ether,
ketones, thinners or gasoline, as they can deform and discolor the
case.
Repair
If the product seems to be malfunctioning, conrm that the AC adapter and the USB cable are connected properly before contacting your
authorized Hioki distributor or reseller.
WARNING
Touching any of the high-voltage points inside the
product is very dangerous. Do not attempt to modify, disassemble, or try to repair the product. Doing
so may cause a re, electric shock, or injury.
Calibration
IMPORTANT
Periodic calibration is necessary in order to ensure that the in-
strument provides correct measurement results of the specied
accuracy.
The calibration period varies with the conditions and environment of
use. It is recommended to determine a calibration period based on
those factors and to have the product regularly calibrated by Hioki.
Please contact your Hioki distributor to have your product periodically calibrated.
Precautions during shipment
Be sure to follow these precautions when transporting the product:
• Remove accessories and optional equipment from the product
in order to avoid damage. Additionally, be sure to double box the
product. Accidental damage suered in transit is not covered by the
warranty.
• Attach a description of the issue when sending out your product for
repair.
Disposal
Dispose of the product in accordance with local regulations.
Safety Notes
This product is designed to conform to IEC 61010 Safety Standards,
and has been thoroughly tested for safety prior to shipment. However, using the product in a way not described in this manual may
negate the provided safety features.
Carefully read the following safety notes before using the product.
WARNING
Mishandling product could result in bodily injury or
even death, as well as damage to the product. Familiarize yourself with the instructions and precautions in this manual before using the product.
WARNING
• Protective Gear
Performing measurement using this product involves live-line work. To prevent an electric shock,
use appropriate protective insulation and adhere
to applicable laws and regulations.
• Electricity can potentially cause serious events
such as an electric shock, heat generation, re,
and an arc ash due to a short-circuit. If you have
not used electrical measuring instrument before,
you should be supervised by a technician who
has experience in electrical measurement.
Notations
In this document, the severity levels of risk and hazard are classied
as follows.
Indicates an imminently hazardous situation
DANGER
WARNING
CAUTION
IMPORTANT
that will result in death of or serious injury to the
operator.
Indicates a potentially hazardous situation that
may result in death of or serious injury to the
operator.
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunction.
Indicates information or content that is particularly important from the standpoint of operating
or maintaining the product.
Indicates cautions and hazards. When the symbol is printed on
the product, refer to a corresponding topic in
the Instruction Manual.
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates an action that must not be performed.
Indicates action that must be performed.
Indicates that the product conforms to regulations required by the EU Directive.
Indicates the Waste Electrical and Electronic
Equipment Directive (WEEE Directive) in EU
member states.
Measurement categories
To ensure safe operation of measuring instruments, IEC 61010
establishes safety standards for various electrical environments, cat-
egorized as CAT II to CAT IV, and called measurement categories.
WARNING
• Using a measuring product in an environment
designated with a higher-numbered category than
that for which the product is rated could result in
a severe accident, and must be carefully avoided.
• Never use a measuring product that lacks cat-
egory labeling in a CAT II to CAT IV measurement
environment. Doing so could result in a serious
accident.
This product conforms to the safety requirements for CAT III 1000 V,
CAT IV 600 V measuring instruments.
CAT II: When directly measuring the electrical outlet recep-
tacles of the primary electrical circuits in equipment
connected to an AC electrical outlet by a power cord
(portable tools, household appliances, etc.)
CAT III: When measuring the primary electrical circuits of heavy
equipment (xed installations) connected directly to
the distribution panel, and feeders from the distribution
panel to outlets
CAT IV: When measuring the circuit from the service drop to the
service entrance, and to the power meter and primary
overcurrent protection device (distribution panel)
Distribution panel
Service entrance
Service drop
CAT IV
Power meter
Internal wiring
CAT III
Fixed installation
CAT II
T
Outlet
Usage Notes
Observe the following precautionary information to ensure that the
product can be used safely and in a manner that allows it to perform
as described in its specications.
WARNING
To prevent an electric shock, conrm that the white
or red portion (insulation layer) inside the cable is
not exposed. If a color inside the cable is exposed,
do not use the cable.
WARNING
Do not allow the product to get wet, and do not take
measurements with wet hands. This may cause an
electric shock.
CAUTION
• To avoid damage to the product, avoid subjecting it to
vibration or mechanical shock during transportation
and handling. Exercise particular care to avoid subjecting the product to mechanical shock, for example
by dropping.
• Do not store or use the product where it could be exposed to direct sunlight, high temperatures or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate
so that it no longer meets specications.
• This product is not designed to be entirely waterproof
or dustproof. Do not use it in an especially dusty environment, nor where it might be splashed with liquid.
This may cause damage.
This product complies with EN 61326 Class A.
This product may cause interference if used in residential areas.
Such use must be avoided unless the user takes special measures
to reduce electromagnetic emissions to prevent interference to the
reception of radio and television broadcasts.

Parts Names
Step
3
Alligator clip
Maximum rated voltage
to earth:
1000 V (CAT lll)
600 V (CAT IV)
Red lead
(HIGH)
Step
4
POWER switch
Selects the voltage division
ratio (power on).
Turns o the product.
Step
5
Output selector switch
(Model P9000-02 only)
Step
1
Power jack
Connect a model Z1008
AC adapter.
USB-microB jack
Connect a USB cable
(commercially available).
Step
Range Scaling setting
6
Connected instrument (pictured: Model MR8875)
Step
7
To measurement target
(pictured: P9000-02)
Black lead
(LOW)
OVER LED
Lights up when
excessively high
input is received.
POWER LED
Lights up when
the product is
receiving power.
BNC connector
Step
2
For more information about
devices that can be connect-
ed, contact your authorized
Hioki distributor or reseller.
Inspection Before Use
To ensure that the product is properly operating, conduct an inspection and check product operation before using product to ensure that
no damage has occurred during storage or transport. If there is any
damage to the product, contact your authorized Hioki distributor or
reseller for repair.
WARNING
Conrm that the insulation on the using the probes
are undamaged and that no bare conductors are improperly exposed before using the product. If there
is any damage to the product that leads to an electric shock, contact your authorized Hioki distributor
or reseller for repair.
Measurement Method
WARNING
• For your safe operation, do not connect the prod-
uct to the primary of the distribution panel.
• Do not connect the product to a voltage or current
that exceeds the rating indicated on the product
labeling or the measurement range indicated in its
specications. Doing so may damage the product
or cause it to become hot, resulting in electric
shock or bodily injury.
• To avoid short circuits and electric shock, do not
short-circuit two wires to be measured by bringing the metal part of the test lead into contact with
them. Never touch the metal end.
WARNING
When supplying power through the USB-micro B
jack, make sure that the USB ground pin of the
power source is connected to the ground. Using an
ungrounded USB terminal could cause an electric
shock.
CAUTION
To prevent damage to the connector, be sure to release
the locking mechanism, grip the head of the connector
(not the cord), and pull it out.
Select which of the following two sources you will
1
use to power the product:
Method 1 Connect a model Z1008 AC adapter
Model Z1008
To commercial
power supply
Method 2 Connect a USB cable (
USB (type A male)↔USB (type B female))
To USB terminal on
connected instrument
Connect the product’s BNC connector to the input
2
terminal on the connected instrument.
(1) Line up
Line up
commercially available,
(2) Lock
Connect the alligator clips to the input leads.
3
Purchase the tip option that is best suited to the measurement target. These options can be purchased via authorized
Hioki distributor or reseller.
CAUTION
Precautions when using the L4932 Test Pin Set
To prevent a short circuit accident, be sure to use the
test leads with the sleeves attached when performing
measurements in the CAT III measurement category.
(For the measurement categories, see “Measurement
categories”.)
If the sleeves are inadvertently removed during measurement, stop the measurement.
600V
1000V
600V
1000V
IMPORTANT
Use only the specied tip options. Using a non-specied tip option
may result in incorrect measurements due to poor connection or
other reasons.
Select the voltage division ratio with the POWER
4
switch.
Voltage
division ratio
1/1000 Divides the high signal and low signal by 1,000
1/100 Divides the high signal and low signal by 100
Select the output type (P9000-02 only).
5
Output Description
Wave Outputs the signal using the selected voltage
AC-RMS Outputs the RMS value using the selected
Set the input range scaling for the connected in-
6
Description
and then outputs the dierence between the
two signals as a signal relative to ground.
and then outputs the dierence between the
two signals as a signal relative to ground.
division ratio. Choose this setting when you
wish to view the power supply waveform.
voltage division ratio. Input is AC coupled.
strument.
Connect the alligator clips to the measurement target.
7
After measurement is complete, turn o the POW-
8
ER switch.
Specications
Product warranty
duration
Accuracy guarantee duration
DC output accuracy
Temperature
characteristic
Frequency
characteristic
Voltage division
ratio
3 years (except for breaks in cables caused by use)
1 year
±0.5% f.s.
f.s. = 1.0 V (ratio 1000:1), 3.5 V (ratio 100:1)
±0.05% f.s./C
DC to 100 kHz, -3 dB
1000:1, 100:1 (user-switchable)
Common-mode
rejection ratio
Input resistance 10.5 M
Input capacitance 5 pF or less (at 100 kHz)
Input jacks Banana plug
Output connector Insulated BNC
Measuring range 1000 V AC/DC (ratio 1000:1)
Maximum input
voltage
Maximum rated lineto-ground voltage
Operating temperature and humidity
range
Storage temperature and humidity
range
Operating
environment
Dimensions Approx. 128.0W×36.0H×22.0D mm
Weight Approx. 170 g/ 6.0 oz.
Power supply (1) Model Z1008 AC adapter
Maximum rated
power
Eects of radiated radio-frequency electromagnetic eld
Eects of conducted radio-frequency electromagnetic eld
Standards EMC EN61326 Class A
Accessories Instruction manual ×1
Options
The following options are available for the product. To purchase an option,
please contact your authorized Hioki distributor or reseller.
Options are subject to change. Please check Hioki’s website for the latest
information.
Z1008 AC Adapter
L9243 Grabber Clip (1000 V CAT II, 1 A)
L4936 Bus Bar Clip Set (600 V CAT III, 5 A)
L4937 Magnetic Adapter Set (1000 V CAT III, 2 A)
L4931 Extension Cable Set (For extending the length of the test leads, 1.5 m)
L4932 Test Pin Set
L4934 Small Alligator Clip Set (300 V CAT III, 600 V CAT II, 3 A)
Model P9000-02 only
Measurement modes Waveform, AC-RMS (selected with a slide switch)
Convertible voltage 1000 V AC (ratio 1000:1), 250 V AC (ratio 100:1)
RMS value measure-
ment accuracy
RMS value measurement response time
f.s.: maximum display value or scale length (This is usually the maxi-
mum value of the currently selected range.)
80 dB or more (50 Hz/60 Hz shorted input)
±1% (between HIGH and LOW terminals)
Ω
250 V AC, 350 V DC (ratio 100:1)
1000 V AC/DC
1000 V AC/DC, measurement category III
anticipated transient overvoltage: 8000 V
Temperature: -40C to 80C/ -40F to 176F
Humidity: -40C to 40C [non-inclusive]
/ -40F to 104F, 80% RH or less
40C to 45C [non-inclusive]
/ 104F to 113F, 60% RH or less
45C to 80C/ 113F to 176F,
50% RH or less (no condensation)
Temperature: -40C to 80C/ -40F to 176F
Humidity: 80% RH or less (no condensation)
Indoors, Pollution Degree II, altitude up to 2000 m (6562
ft.)
/ Approx. 5.04″W×1.42″H×0.87″D
(12 V DC, DC jack φ5.5×2.1 mm)
Rated supply voltage: 100 V AC to 240 V AC
(Voltage uctuations of ±10% from the rated supply
voltage are taken into account.)
Rated supply frequency: 50 Hz/60 Hz
Anticipated transient overvoltage: 2500 V
(2) USB bus power (5 V DC USB-microB connector)
(3) External power supply, 2.7 V DC to 15 V DC
(1) When using a model Z1008 AC Adapter:
6 VA (model P9000-01/-02 with a AC adapter)
0.9 VA (model P9000-01/-02 only)
(2) When using a USB bus power: 0.8 VA
(3)
When using a 15 V DC external power supply: 1 VA
±5% f.s. max. at 3 V/m
±5% f.s. max. at 3 V
Safety EN61010
Alligator clips×1 set (red and black/ one each)
Carrying case×1
(600 V CAT IV, 1000 V CAT III, 10 A)
(600 V CAT IV, 1000 V CAT III, 1000 V CAT II, 10 A)
(L4932 is required when using L4934.)
±1% f.s. (30 Hz to 1 kHz [non-inclusive], sine wave)
±3% f.s. (1 kHz to 10 kHz, sine wave)
f.s. = 1.0 V (ratio 1000:1), 2.5 V (ratio 100:1)
300 ms
Rising 0% f.s. → 100% f.s.
Response time for input from 0% f.s. to 90% f.s.
600 ms
Falling 100% f.s. → 0% f.s.
Response time for input from 100% f.s. to 10% f.s.