Hioki P9000-01, P9000-02 Instruction Manual

P9000-01
HIOKI P9000A981-02
P9000-02
DIFFERENTIAL PROBE
Instruction Manual
EN
Apr. 2021 Revised edition 2 P9000A981-02 21-04H
Warranty
Warranty malfunctions occurring under conditions of normal use in conformity with the Instruction Manual and Product Precautionary Markings will be repaired free of charge. This warranty is valid for a period of three (3) years from the date of purchase. Please contact the distributor from which you purchased the product for further infor­mation on warranty provisions.
Introduction
Thank you for choosing the Hioki P9000-01, P9000-02 Di󰀨erential
Probe. Preserve this manual carefully and keep it handy to make full use of this product for along time.
Overview
The product is a di󰀨erential probe that is connected to a Memory
HiCorder input unit. The probe divides two high voltages input to its high and low terminals by 100 or 1000 and then outputs the di󰀨er­ence as a signal relative to ground. The P9000-02 includes AC RMS value output functionality. Power can be supplied via the optional AC adapter or a commer­cially available USB cable.
Verifying Package Contents
When you open the package, carefully inspect the product to ensure that everything is in good condition, and that no damage occurred during shipping. Carefully check the accessories and connectors. If the product seems to have been damaged or does not work as
specied, contact your authorized Hioki distributor or reseller.
Cleaning
If the product becomes dirty, slightly moisten a soft cloth moistened with water or a neutral detergent and wipe the product clean.
IMPORTANT
Never use solvents such as benzene, alcohol, acetone, ether,
ketones, thinners or gasoline, as they can deform and discolor the case.
Repair
If the product seems to be malfunctioning, conrm that the AC adapt­er and the USB cable are connected properly before contacting your
authorized Hioki distributor or reseller.
WARNING
Touching any of the high-voltage points inside the product is very dangerous. Do not attempt to mod­ify, disassemble, or try to repair the product. Doing
so may cause a re, electric shock, or injury.
Calibration
IMPORTANT
Periodic calibration is necessary in order to ensure that the in-
strument provides correct measurement results of the specied
accuracy.
The calibration period varies with the conditions and environment of use. It is recommended to determine a calibration period based on those factors and to have the product regularly calibrated by Hioki. Please contact your Hioki distributor to have your product periodi­cally calibrated.
Precautions during shipment
Be sure to follow these precautions when transporting the product:
• Remove accessories and optional equipment from the product in order to avoid damage. Additionally, be sure to double box the
product. Accidental damage su󰀨ered in transit is not covered by the
warranty.
• Attach a description of the issue when sending out your product for repair.
Disposal
Dispose of the product in accordance with local regulations.
Safety Notes
This product is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. How­ever, using the product in a way not described in this manual may negate the provided safety features. Carefully read the following safety notes before using the product.
WARNING
Mishandling product could result in bodily injury or
even death, as well as damage to the product. Fa­miliarize yourself with the instructions and precau­tions in this manual before using the product.
WARNING
• Protective Gear
Performing measurement using this product in­volves live-line work. To prevent an electric shock, use appropriate protective insulation and adhere to applicable laws and regulations.
• Electricity can potentially cause serious events
such as an electric shock, heat generation, re, and an arc ash due to a short-circuit. If you have
not used electrical measuring instrument before, you should be supervised by a technician who has experience in electrical measurement.
Notations
In this document, the severity levels of risk and hazard are classied
as follows.
Indicates an imminently hazardous situation
DANGER
WARNING
CAUTION
IMPORTANT
that will result in death of or serious injury to the operator.
Indicates a potentially hazardous situation that
may result in death of or serious injury to the operator.
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the operator or damage to the product or malfunc­tion.
Indicates information or content that is particu­larly important from the standpoint of operating or maintaining the product.
Indicates cautions and hazards. When the sym­bol is printed on the product, refer to a corresponding topic in the Instruction Manual.
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates an action that must not be performed.
Indicates action that must be performed.
Indicates that the product conforms to regula­tions required by the EU Directive.
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Measurement categories
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various electrical environments, cat-
egorized as CAT II to CAT IV, and called measurement categories.
WARNING
• Using a measuring product in an environment
designated with a higher-numbered category than that for which the product is rated could result in a severe accident, and must be carefully avoided.
• Never use a measuring product that lacks cat-
egory labeling in a CAT II to CAT IV measurement environment. Doing so could result in a serious accident.
This product conforms to the safety requirements for CAT III 1000 V, CAT IV 600 V measuring instruments.
CAT II: When directly measuring the electrical outlet recep-
tacles of the primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
CAT III: When measuring the primary electrical circuits of heavy
equipment (xed installations) connected directly to
the distribution panel, and feeders from the distribution panel to outlets
CAT IV: When measuring the circuit from the service drop to the
service entrance, and to the power meter and primary overcurrent protection device (distribution panel)
Distribution panel
Service entrance Service drop
CAT IV
Power meter
Internal wiring
CAT III
Fixed installation
CAT II
T
Outlet
Usage Notes
Observe the following precautionary information to ensure that the product can be used safely and in a manner that allows it to perform
as described in its specications.
WARNING
To prevent an electric shock, conrm that the white
or red portion (insulation layer) inside the cable is not exposed. If a color inside the cable is exposed, do not use the cable.
WARNING
Do not allow the product to get wet, and do not take measurements with wet hands. This may cause an electric shock.
CAUTION
• To avoid damage to the product, avoid subjecting it to vibration or mechanical shock during transportation and handling. Exercise particular care to avoid sub­jecting the product to mechanical shock, for example by dropping.
• Do not store or use the product where it could be ex­posed to direct sunlight, high temperatures or humid­ity, or condensation. Under such conditions, the prod­uct may be damaged and insulation may deteriorate
so that it no longer meets specications.
• This product is not designed to be entirely waterproof or dustproof. Do not use it in an especially dusty en­vironment, nor where it might be splashed with liquid. This may cause damage.
This product complies with EN 61326 Class A. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
Parts Names
HIOKI P9000A981-02
Step
3
Alligator clip
Maximum rated voltage to earth:
1000 V (CAT lll) 600 V (CAT IV)
Red lead
(HIGH)
Step
4
POWER switch
Selects the voltage division ratio (power on).
Turns o󰀨 the product.
Step
5
Output selector switch
(Model P9000-02 only)
Step
1
Power jack
Connect a model Z1008 AC adapter.
USB-microB jack
Connect a USB cable (commercially available).
Step
Range Scaling setting
6
Connected instrument (pictured: Model MR8875)
Step
7
To measurement target
(pictured: P9000-02)
Black lead
(LOW)
OVER LED
Lights up when excessively high input is received.
POWER LED
Lights up when the product is receiving power.
BNC connector
Step
2
For more information about devices that can be connect-
ed, contact your authorized
Hioki distributor or reseller.
Inspection Before Use
To ensure that the product is properly operating, conduct an inspec­tion and check product operation before using product to ensure that no damage has occurred during storage or transport. If there is any
damage to the product, contact your authorized Hioki distributor or
reseller for repair.
WARNING
Conrm that the insulation on the using the probes
are undamaged and that no bare conductors are im­properly exposed before using the product. If there is any damage to the product that leads to an elec­tric shock, contact your authorized Hioki distributor or reseller for repair.
Measurement Method
WARNING
• For your safe operation, do not connect the prod-
uct to the primary of the distribution panel.
• Do not connect the product to a voltage or current
that exceeds the rating indicated on the product labeling or the measurement range indicated in its
specications. Doing so may damage the product
or cause it to become hot, resulting in electric
shock or bodily injury.
• To avoid short circuits and electric shock, do not
short-circuit two wires to be measured by bring­ing the metal part of the test lead into contact with them. Never touch the metal end.
WARNING
When supplying power through the USB-micro B
jack, make sure that the USB ground pin of the
power source is connected to the ground. Using an ungrounded USB terminal could cause an electric shock.
CAUTION
To prevent damage to the connector, be sure to release the locking mechanism, grip the head of the connector (not the cord), and pull it out.
Select which of the following two sources you will
1
use to power the product:
Method 1 Connect a model Z1008 AC adapter
Model Z1008
To commercial power supply
Method 2 Connect a USB cable (
USB (type A male)USB (type B female))
To USB terminal on connected instrument
Connect the product’s BNC connector to the input
2
terminal on the connected instrument.
(1) Line up
Line up
commercially available,
(2) Lock
Connect the alligator clips to the input leads.
3
Purchase the tip option that is best suited to the measure­ment target. These options can be purchased via authorized Hioki distributor or reseller.
CAUTION
Precautions when using the L4932 Test Pin Set
To prevent a short circuit accident, be sure to use the test leads with the sleeves attached when performing measurements in the CAT III measurement category. (For the measurement categories, see “Measurement categories”.) If the sleeves are inadvertently removed during mea­surement, stop the measurement.
600V 1000V
600V 1000V
IMPORTANT
Use only the specied tip options. Using a non-specied tip option
may result in incorrect measurements due to poor connection or other reasons.
Select the voltage division ratio with the POWER
4
switch.
Voltage division ratio
1/1000 Divides the high signal and low signal by 1,000
1/100 Divides the high signal and low signal by 100
Select the output type (P9000-02 only).
5
Output Description
Wave Outputs the signal using the selected voltage
AC-RMS Outputs the RMS value using the selected
Set the input range scaling for the connected in-
6
Description
and then outputs the di󰀨erence between the
two signals as a signal relative to ground.
and then outputs the di󰀨erence between the
two signals as a signal relative to ground.
division ratio. Choose this setting when you wish to view the power supply waveform.
voltage division ratio. Input is AC coupled.
strument.
Connect the alligator clips to the measurement target.
7
After measurement is complete, turn o󰀨 the POW-
8
ER switch.
Specications
Product warranty duration
Accuracy guaran­tee duration
DC output accu­racy
Temperature characteristic
Frequency characteristic
Voltage division
ratio
3 years (except for breaks in cables caused by use)
1 year
±0.5% f.s.
f.s. = 1.0 V (ratio 1000:1), 3.5 V (ratio 100:1)
±0.05% f.s./C
DC to 100 kHz, -3 dB
1000:1, 100:1 (user-switchable)
Common-mode rejection ratio
Input resistance 10.5 M Input capacitance 5 pF or less (at 100 kHz) Input jacks Banana plug Output connector Insulated BNC Measuring range 1000 V AC/DC (ratio 1000:1)
Maximum input voltage
Maximum rated line­to-ground voltage
Operating temper­ature and humidity range
Storage tempera­ture and humidity range
Operating environment
Dimensions Approx. 128.0W×36.0H×22.0D mm
Weight Approx. 170 g/ 6.0 oz. Power supply (1) Model Z1008 AC adapter
Maximum rated power
E󰀨ects of radiated radio-frequency electromagnetic eld
E󰀨ects of conducted radio-frequency electromagnetic eld
Standards EMC EN61326 Class A
Accessories Instruction manual ×1
Options The following options are available for the product. To purchase an option,
please contact your authorized Hioki distributor or reseller.
Options are subject to change. Please check Hioki’s website for the latest information.
Z1008 AC Adapter
L9243 Grabber Clip (1000 V CAT II, 1 A) L4936 Bus Bar Clip Set (600 V CAT III, 5 A) L4937 Magnetic Adapter Set (1000 V CAT III, 2 A)
L4931 Extension Cable Set (For extending the length of the test leads, 1.5 m)
L4932 Test Pin Set L4934 Small Alligator Clip Set (300 V CAT III, 600 V CAT II, 3 A)
Model P9000-02 only
Measurement modes Waveform, AC-RMS (selected with a slide switch) Convertible voltage 1000 V AC (ratio 1000:1), 250 V AC (ratio 100:1) RMS value measure-
ment accuracy
RMS value measure­ment response time
f.s.: maximum display value or scale length (This is usually the maxi-
mum value of the currently selected range.)
80 dB or more (50 Hz/60 Hz shorted input)
±1% (between HIGH and LOW terminals)
Ω
250 V AC, 350 V DC (ratio 100:1) 1000 V AC/DC
1000 V AC/DC, measurement category III anticipated transient overvoltage: 8000 V
Temperature: -40C to 80C/ -40F to 176F Humidity: -40C to 40C [non-inclusive] / -40F to 104F, 80% RH or less 40C to 45C [non-inclusive] / 104F to 113F, 60% RH or less 45C to 80C/ 113F to 176F, 50% RH or less (no condensation) Temperature: -40C to 80C/ -40F to 176F
Humidity: 80% RH or less (no condensation)
Indoors, Pollution Degree II, altitude up to 2000 m (6562 ft.)
/ Approx. 5.04″W×1.42″H×0.87″D
(12 V DC, DC jack φ5.5×2.1 mm) Rated supply voltage: 100 V AC to 240 V AC (Voltage uctuations of ±10% from the rated supply
voltage are taken into account.)
Rated supply frequency: 50 Hz/60 Hz Anticipated transient overvoltage: 2500 V
(2) USB bus power (5 V DC USB-microB connector)
(3) External power supply, 2.7 V DC to 15 V DC (1) When using a model Z1008 AC Adapter:
6 VA (model P9000-01/-02 with a AC adapter)
0.9 VA (model P9000-01/-02 only)
(2) When using a USB bus power: 0.8 VA
(3)
When using a 15 V DC external power supply: 1 VA
±5% f.s. max. at 3 V/m
±5% f.s. max. at 3 V
Safety EN61010
Alligator clips×1 set (red and black/ one each) Carrying case×1
(600 V CAT IV, 1000 V CAT III, 10 A)
(600 V CAT IV, 1000 V CAT III, 1000 V CAT II, 10 A)
(L4932 is required when using L4934.)
±1% f.s. (30 Hz to 1 kHz [non-inclusive], sine wave)
±3% f.s. (1 kHz to 10 kHz, sine wave)
f.s. = 1.0 V (ratio 1000:1), 2.5 V (ratio 100:1) 300 ms
Rising 0% f.s. → 100% f.s.
Response time for input from 0% f.s. to 90% f.s. 600 ms
Falling 100% f.s. → 0% f.s.
Response time for input from 100% f.s. to 10% f.s.