Hioki MR8870-20 Instruction Manual

MR8870-20
Instruction Manual
MEMORY HiCORDER
Aug. 2018 Revised edition 3 MR8870B982-03 18-08H
EN
Contents
Contents
1
Introduction ..............................................................................1
Confirming Package Contents................................................2
Safety Information....................................................................3
Operating Precautions.............................................................6
Chapter 1 Overview ____________________________________11
1.1 Product Overview and Features .............................11
1.2 Measurement Flow ........... .......................................12
1.3 Names and Functions of Parts ...............................15
1.4 Screen Configurations ............................................17
1.5 Basic Operation .......................................................19
Chapter 2 Measurement Preparations_________________23
2.1 Using the Battery Pack (Option) ............................24
Install the Battery Pack ....................................................27
Charge the Battery Pack ..................................................27
2.2 Connecting the AC Adapter ....................................28
2.3 Connecting Measurement Cables to the
Memory HiCorder ....................................................29
Connecting Connection Cords
(to record voltage waveforms) ..........................................31
Connecting a Clamp
(to re
cord current waveforms) ..........................................31
Connecting the Logic Probe (to record logic signals) .......32
2.4 Attaching the Carrying Strap ..................................33
2.5 Turning the Power On and Off ...............................34
2.6 Zero Adjustment ......................................................35
Sensor
2 3
MR8870B982-03
ii
Contents
Chapter 3 Setting before Measurement_______________ 37
3.1 Pre-Operation Inspection ....................................... 37
3.2 Operating Flow Overview ........................ ............... 38
3.3 Automatic Configuration of Measurement Settings
(Auto Range) .. ... .. .................................................... 42
3.4 Measurement Configuration 1
(Horizontal Axis Settings) ...................................... 44
Select a Timebase ...........................................................44
Setting Recording Length (no. of divisions) .....................45
Setting Display Zoom (as needed) ................................... 45
3.5 Measurement Configuration 2
(Setting Recording Criteria) ................................... 46
Setting the Trigger Mode .................................................46
3.6 Measurement Configuration 3
(Analog Channel Settings) ..................................... 47
Setting the Voltage Axis (Vertical Axis) Range ................47
Selecting Waveform Display Color (as needed) ..............47
Setting the Zero Position (as needed) .............................48
Selecting the Input Coupling Method (as needed) ...........48
Zooming the Vertical Axis (as needed) ............................49
Selecting a Low-Pass Filter (L.P.F.) (as needed) ............49
Converting Units (Scaling function) ...................... ........ ....50
3.7 Measurement Configuration 4
(Logic Channel Settings) ....................................... 51
Setting Logic Channel Display Colors (as needed) .........51
Setting Logic Channel Display Positions .........................51
3.8 Entering Comments ................................................ 52
Chapter 4 Making Trigger Settings____________________ 55
4.1 Setting Recording Criteria ..................................... 55
Setting Combining Logic (AND/OR)
for Multiple Trigger Sources ............................................56
Setting the Trigger Start Point (Pre-Trigger) ....................56
4.2 Triggering by Analog Signals ................................ 57
Selecting an Analog Trigger Type ....................................58
Selecting a Trigger Filter ..................................................59
Selecting Rising or Falling (Slope) of the Input Signal .....59
Setting the Signal Threshold Level (Trigger Level) ..........59
Setting Lower and Upper Trigger Thresholds ..................60
Selecting Measurement Frequency .................................60
4.3 Logic Channel Trigger Settings .............................61
Configuring Logical (AND/OR) Trigger Criteria ................61
Setting a Trigger Pattern ..................................................62
Setting a Trigger Filter ......................................................62
4.4 Triggering Manually (Manual Trigger) .... .. ... .. ........62
Chapter 5 Waveform Analysis__________________________63
5.1 Viewing Waveforms .................................................63
Scrolling Waveforms ........................................................64
Verifying Waveform Display Position ...............................64
Magnifying and Compressing Horizontally (Time Axis) ....64
Viewing Any Waveform Location (Jump Function) ..........65
5.2 Viewing Measurement Values ................................66
Displaying Gauges ...........................................................66
Displaying Cursor Values .................................................66
Specifying a Waveform Time Span ..................................68
Viewing Input Signals as Numerical Values
(Numerical Value Display Function) .................................69
iii
Contents
3 4
5 6
Chapter 6 Saving & Loading Data _____________________71
6.1 About Saving and Loading Data ............................71
6.2 Using a CF Card .......................................................74
CF Card Insertion & Removal ..........................................75
Formatting a CF Card ......................................................76
6.3 Saving Data ..... .........................................................77
Automatic Saving .............................................................78
Enabling Delete-and-save Operation ...............................79
Selecting the Manual Saving Method
[Quick Save]/[Select & Save] ...........................................80
Saving Waveform Data (with the SAVE Key) ...................81
Capturing a Screen Image (With the SAVE Key) .............82
iv
Contents
Saving Numerical Calculation Results
(With the SAVE Key) ........................................................83
Saving Setting Configurations ..........................................84
6.4 Loading Data on the Memory HiCorder ................ 85
Loading a Setting Configuration .......................................85
Load the setting configuration data automatically ............86
Loading Waveform Data and Screen Images ..................87
6.5 Data Management ................................................... 88
Viewing Folder Contents and the Parent Folder ..............88
Deleting Data ...................................................................89
Renaming Files and Folders ............................................89
Sorting Files .....................................................................90
6.6 Transferring Data to a Computer .......................... 91
Chapter 7 Numerical Calculations ____________________ 93
7.1 Calculation Methods ........ ... .................................... 93
Auto Calculation ...............................................................94
Manual Calculation ..........................................................95
Apply Calculations to a Specific Time Span
(Manual Calculation Only) ................................................96
7.2 Numerical Value Calculation Expressions ........... 97
Chapter 8 System Environment Settings______________ 99
8.1 Screen and Key Operation Settings ................. ... 100
Using the Auto-Resume Function
(Resume After Power Restoration) ................................100
Adjust backlight brightness ............................................100
Enabling and Disabling the Backlight Saver ..................101
Selecting Black or White Screen Background ...............101
Enabling or Disabling the Beeper ..................................101
8.2 Making System Settings ...................................... 102
Setting the Date and Time .............................................102
Initializing the Memory HiCorder (System Reset) ..........103
Selecting the Display Language ....................................103
Self-Test .........................................................................104
Chapter 9 External Control __________________________ 105
9.1 Connecting to the External Control
Terminals ......................... .. .....................................105
9.2 External Trigger Input ......... .. ... .............................107
9.3 External Signal Output (Trigger Output) .............108
9.4 Synchronous Measurements with Multiple
Instruments ....................... ..................................... 109
Chapter 10 Specifications_____________________________ 111
Chapter 11 Maintenance and Service_________________ 127
v
Contents
11.1 Troubleshooting ....................................................127
11.2 Cleaning .................................................................130
11.3 Disposing of the Instrument .................................131
Appendix ___________________________________ A1
Appendix 1 Error Messages and Remedial Actions......... A1
Appendix 2 File Naming ......................................................A5
Appendix 3 Wave Processor Application Program.......... A6
Appendix 4 List of Default Settings ...................................A9
Appendix 5 Supplemental Technical Information...........A11
Appendix 6 Options...........................................................A13
Appendix 7 Easy setting for RMS recorder mode ..........A15
Appendix 8 Executing zero adjustment
for Model CT969x series and Model CT6590A18
Appendix 9 To 8870-20 Users
(Firmware Ver. 1.08 or Earlier)......................A19
Appendix 10To MR8870-20 Users
(Firmware Ver. 2.03 or Earlier)......................A21
Index __________________________________ Index 1
6 7 8 9
10 11
Appendix
Index
vi
Contents

Introduction

Thank you for purchasing the HIOKI "Model MR8870-20 Memory HiCorder." To obtain maximum performance from the instrument, please read th erence.
1
Introduction
is manual carefully, and keep it handy for future ref-
The MR8870-20 offers two modes: mem RMS recorder mode. This manual primarily describes operation of the instrument in memory recorder mode. For more information about RMS recorder mode, see the included Measurement Guide (RMS Recorder). An optional clamp-on probe is required for current measurement w
ith the Memory HiCorder. In this manual, all models of these probes are collectively referred to as “clamp sensors.” For details, refer to the instruction manual of the particular clamp sensor to be used.
ory recorder mode and
Registered trademarks
• Windows is a registered trademark of Microsoft Corporation in
the United States and/or other countries.
• CompactFlash is a registered trademark of Sandisk Corporation
(USA).
2
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Confirm that these contents are provided.
Model MR8870-20
Memory HiCorder.........................1
Model Z1005 AC Adapter ............1
with supplied power cord
USB Cable ....................................1
Model 9809 Protection Sheet......1
Please stick this over the screen to protect it from any damage.
(p. 28)
(p. 91)
Instruction Manual
(This document)...........................1
Measurement Guide
(Memory Recorder)......................1
Measurement Guide
(RMS Recorder) ...........................1
Strap .............................................1
Dedicated Wave Processor
Application Program for the
8870 (CD)......................................1
(p. 33)
The latest version can be downloaded from our web site.

Confirming Package Contents

Confirming Package Contents
For information about options:(p. A13)
Contact your dealer or Hioki representative for details.
3

Safety Information

Safety Information
This instrument is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. How­ever, using the instrument in a way not described ual may negate the provided safety features. Be certain that you understand the instructions and precau­tions in the manual before use. ity for accidents or injuries not resulting directly from instrument defects.
This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions.
We disclaim any responsibil-
Safety Symbols
In the manual, the symbol indicates particularly important information that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with the symbol) before using the relevant function.
Indicates a double-insulated device. Indicates DC (Direct Current).
in this man-
Indicates a grounding terminal. Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
The following symbols in this manual indicate and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument.
Indicates advisory items related to performance or correct operation of the instrument.
the relative importance of cautions
4
Ni-MH
Safety Information
Symbols for Various Standards
Indicates that the product conforms to regulations set out by the EC Directive.
This is a recycle mark established under the Resource Recycling Promotion
(only for Japan).
Law
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Other Symbols
Indicates the prohibited action.
(p. #)
[ ]
SET
(Bold characters)
Unless otherwise specified, "Windows" represents Windows XP, Windows Vista, Win­dows 7, Windows 8, or Windows 10
Click: Press and quickly release the left button of the mouse. Double click: Quickly click the left button of the mouse twice.
Indicates the location of reference information. Indicates that descriptive information is provided below. The names of setting objects and buttons on the screen are indicated
by square brackets [ ]. Bold characters within the text indicate operating key labels.
Accuracy
We define measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values, with the following meanings:
f.s. (maximum display value or scale length)
The maximum displayable value or scale length. In this instrument, the maximum
layable value is the range (V/div) times the number of divisions (10) on the verti-
disp cal axis. Example: For the 1 V/div range, f.s. = 10 V
rdg. (reading or displayed value)
The value currently being measured and indicated on the measuring instrument.
dgt. (resolution)
The smallest displayable unit on a digital measuring instrument, i.e., the input value that causes the digital display to show a "1" as the least-significant digit.
Safety Information
Measurement categories
This instrument complies with CAT II safety requirements. To ensure safe operation of measurement instruments, IEC 61010 establishes safety stan­dards for various electrical environments, categorized as CAT II to CAT IV, and called mea­surement categories.
5
CAT II: Primary electrical circuits in equipment
CAT III: Primary electrical circuits of heavy
CAT IV: The circuit from the service drop to the service entrance, and to the power meter
Using a measurement instrument in an environment d esignated with a higher-numbered ca t­egory than that for which the instrument is rated could result in a severe accident, and must b Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV measurement a
connected to an AC electrical outlet by a powe
r cord (portable tools, household appliances, etc.) CAT II covers directly measuring electrical outlet receptacles.
eq
uipment (fixed installations) connect­ed directly to the distribution panel, and fe
eders from the distribution panel to
outlets. an
d primary overcurrent protection device (distribution panel).
e carefully avoided. pplications could result in a severe accident, and must be carefully avoided.
6

Operating Precautions

Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Before Use
• Before using the instrument the first time, verify that it operates normally to en or shipping. If you find any damage, contact your dealer or Hioki representative.
• Before using the instrument, make sure that the insulation on the pro
bes, connection cords, and clamp sensor leads is undam­aged and that no bare conductors are improperly exposed. Using the instrumen so contact your dealer or Hioki representative for replacements.
Instrument Installation
Operating temperature and humidity: 0 to 40°C at 80% RH or less
-condensating)
(non Temperature and humidity range for guaranteed accuracy: 2
3±5°C, 80%RH or less
sure that the no damage occurred during storage
t in such conditions could cause an electric shock,
Avoid the following locations that could cause an accident or damage to the instrument.
Exposed to direct sunlight Exposed to high temp
erature
Exposed to liquids Exposed to high humid
ity or con-
densation
Exposed to high levels o
f particulate
dust
In the presence of corrosive or explo­sive gases
Exposed to strong electr
omagnetic fields Near electromag­netic radiators
Subject to vibration
The maximum operating (ambient) temperature for the MR8870-20 is 40°C. Do not attempt to use in higher temperature environme nts.
Correct measurement may be impossible in the presence of strong magnetic fields, such as near transformers and high-cur rent conductors, or in the presence of strong electromagnetic fields such as near radio transmitters.
Installation Precautions
• Do not install the instrument with any side except the bottom fac­ing down.
• Leave sufficient space around the ventilation holes and install the instrumen
Handling the Instrument
• Do not allow the instrument to get wet, and do not take mea­surements with wet hands. This may cause an electric shock
.
• Do not attempt to modify, disasse ment; as fire, electric shock and injury could result.
• To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
• This i nstrument may cause interference if used in residential are
as. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent inter­ference to the reception of radio and television broadcasts.
7
Operating Precautions
-
t with the holes unobstructed.
mble or repair the instru-
Handling the Cords and Probes
• Avoid stepping on or pinching cables, which could damage the cable insulation.
• To avoid breaking the cables and prob
• To avoid damaging the power cord, grasp the plug, not the cord , whe
n unplugging it from the power outlet.
• When disconnecting the BNC connector, be sure to release the
ck before pulling off the connector. Forcibly pulling the connec-
lo tor without releasing the lock, or pulling on the cable, can dam ­age the connector.
es, do not bend or pull them.
8
Operating Precautions
• To prevent an electric shock accident, confirm that the white or red portion (insulation layer) inside the cable is not exp osed. If a color inside the cable is exposed, do not use the cable.
• Use only the specified measurement cables. Using a non-speci­fied cable may result in incorrect measurements due to poor con­nection or other reasons.
• Before using a clamp sensor or logic probe, read the instruction manual supplied with it.
Before Turning Power On
Using the Battery Pack
• For battery operation, use only the HIOKI Model 9780 Battery Pack. We do not take any responsibility for accidents or
amage related to the use of any other batteries.
d
See: "2.1 Using the Battery Pack (Option)" (p. 24)
Using the AC Adapter
• Use only the supplied Model Z1005 AC Adapter. AC adapter input voltage range is 100 to 240 VAC (with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this range.
• Turn the instrument off before connecting the AC the instrument and to AC power.
• To avoid electrical accidents and to maint ain the safety spec­ifications of this instrument, connec vided only to a 3-contact (two-conductor + ground) outlet.
• Before turning the instrument on, make sure the age matches that indicated on it tion to an improper supply voltage may damage the instrument and present an electrical hazard.
adapter to
t the power cord pro-
supply volt-
s power connector. Connec-
When the power is turned off, do not apply voltage or current to the BNC jacks, logic connector, or external control terminals. Doing so may damage the instrument.
After use, always turn OFF the power.
About Inputs and Measurement
Measure­ment cables
Maximum rated volt­age between terminals
maximum rated voltage to earth
Model L9197 600 VAC/DC
600 VAC/DC (CAT III) 300 VAC/DC (CAT IV)
Model 9197 Model L9198 300 VAC/DC
600 VAC/DC (CAT II) 300 VAC/DC (CAT III)
Model L9217 Model L9790 600 VAC/DC When using the L9790-01 Alliga-
tor clip, the 9790-03 Contact pin 600 VAC/DC (CAT II) 300 VAC/DC (CAT III) When using the L9790-01 Grab­ber clip 300 VAC/DC (CAT II) 150 VAC/DC (CAT III)
Model 9322 Differential Probe
2000 VDC, 1000 VAC When using the Grabber clip
1500 VAC/DC (CAT II) 600 VAC/DC (CAT III) When using the Alligator clip 1000 VAC/DC (CAT II) 600 VAC/DC (CAT III)
Note the following maximum rated voltage between terminals and maximum rated voltage to earth. Maximum rated voltage between terminals: 400 VDC Maximum rated voltage to earth: 300 VAC/DC (CAT II) The following tables list the maximum ra terminals and the maximum rated voltage to earth for connec­tion cords. To avoid electric shock and instru voltages in excess of these figures. The lower of the instrument’s and the imum rated voltage between terminalss applies. When these voltage levels are exceeded, the instrument will be damaged. Do not make measurements as doing so may result in bodily injury. The maximum rated voltage to earth doe you make measurements with an attenuator or other compo­nent connected to the input. Consider your connection method and do not exce maximum rated voltage to earth.
9
Operating Precautions
ted voltage between
ment damage, do not input
connection cord’s max-
s not change, even if
ed the
For example, if the instrument is used in combination with Model L9198, the maximum rated voltage between terminals is 300 VDC and the maximum rated voltage to earth is 300 VAC (CAT II).
10
Operating Precautions
Connect the clamp-on sensors or measurement cables to the instrument first, and then to the active lines to be measured. Observe the following to avoid electric shock and short cir­cuits.
• To avoid short circuits and potentia ards, never attach the clamp to a c more than the maximum rated voltage to earth, or over bare conductors.
• Do not allow the connection cord clip the same time. Never touch the edge of the metal clips.
• When the clamp sensor is opened, do not allow the metal
art of the clamp to touch any exposed metal, or to short
p between two lines, and do not use over bare conductors.
• Do not leave the Memory HiCorder connected to test object in environments where a voltage surge might exceed the dielectric withstand voltage. Doing so could result in dam­age to the Memory HiCorder, bodily injury or fatal accident.
CD Handling
• Always hold the disc by the edges, so as not to make fingerprints on the disc or scratch the printing.
• Never touch the recorde d side of the disc. Do not place the disc dire
• Do not wet the disc with volatile alcohol or water, as there is a possibility of th
• To write on the disc label surface, use a spirit-based felt pen. Do not
use a ball-point pen or hard-tipped pen, because there is a danger of scratching the surface and corrupting the data. Do not use adhesive labels.
• Do not expose the disc directly to the sun's rays, or keep it in conditions of high temp of warping, with consequent loss of data.
• To remove dirt, dust, or fingerprints from the disc, wipe with a dry
th, or use a CD cleaner. Always wipe from the inside to the
clo outside, and do no wipe with circular movements. Never use abrasives or solvent cleaners.
• Hioki shall not be held liable for any problems with a computer syste related to the purchase of a Hioki product.
lly life-threatening haz-
ircuit that operates at
s to touch two wires at
s
ctly on anything hard.
e label printing disappearing.
erature or humidity, as there is a danger
m that arises from the use of this CD, or for any prob lem
11
Waveform analysis
T
mV A
Cursor Measurements (p. 66)
A
B
Using the A/B cursors, values at cursor lo­cations on waveforms and the time and pe­riod at trigger positions can be displayed.
Up to seven types of calculations such as RMS, maximum and average values can be applied to measurement data, and waveforms and calculation results can be displayed separately on screen.
Numerical Calculations (p. 93)
123.4V
Recording logic signals
Measures using optional logic probes. Mixed ana­log and logic recording is possible.
Recording various analog signals
Measurements for a wide variety of applications are available using the appropriate measurement leads or clamp sensor options.
Gauge display (p. 66)
Zoom in/out (p. 64)
Numerical display (p. 69)
Measurement data and set­ting configurations are stored, read and managed as data files
Measurement data can be stored to and read from an optional CF card (p. 71).
Recording waveforms of abnormal events
Abnormal events can be an­alyzed by recording with the trigger function (p. 55).
Analysis on a PC
Data stored on the CF card can be transferred for analysis to a comput­er using the supplied USB cable (p. 91). Measurement data can be analyzed using the supplied ap­plication program (p. A6).
Display in converted units
Using the scaling function, input val­ues can be displayed in units of the actual physical quantity being mea­sured, such as speed or tempera­ture (p. 50).

1.1 Product Overview and Features

Overview Chapter 1
1.1 Product Overview and Features
The Hioki MR8870-20 Memory HiCorder is a compact, lightweight, easy-to­operate waveform recorder. It can run on batteries, and can be quickly deployed when a power anomaly occurs. Measurement data can be monitored, subjec on a personal computer using the supplied application program.
ted to calculations and analyzed
1

Chapter 1 Overview

3
12
AC adapter connection (p. 28)
Install the battery pack (option) (p. 24)
Connect to the Model MR8870-20 (p. 23), and set up (p. 6)
Turn the power on (p. 34)
Measurement cables connections (p. 29)
Configure measurement mode (p. 38)
Memory recorder mode
The instrument records instanta­neous signal waveforms. Trigger and numerical calculation function­ality can be used.
RMS recorder mode
The instrument records RMS waveforms for commercial power supply(50Hz,60Hz) and DC sig­nals.
Thereafter, it will be explained on the screen of Memory recorder mode.

1.2 Measurement Flow

1.2 Measurement Flow
Installing, Connecting and Turning On
13
Configure measurement settings (p. 44)
Using a previously saved setting configuration
Reload a previously stored setting configuration from Model MR8870-20 memory or a CF card, and measure (p. 85).
Select the timebase (horizontal ax­is), voltage range (vertical axis) and input channels. If the measurement ranges are un­known, they can be set automatical­ly (p. 42).
Move to a setting item
Open the setting options
Select
Apply
Configure Automatic Data Storage to a CF card (as needed) (p. 77)
To save (store) measured data auto­matically, enable Auto-Save before starting to measure. Alternatively, you can save data manually after measuring.
1.2 Measurement Flow
Settings
1
Chapter 1 Overview
3
14
Press the START/STOP key (p. 20)
Record once, and stop. Trigger Mode: [Single]
Start
Measurement
When recording using the trigger function, recording occurs only when the input waveform satisfies specified trigger criteria.
Record repeatedly. Trigger Mode: [Repeat]
Stop
Measurement
View Measurement Data (p. 63)
Waveforms can be zoomed and numerical values confirmed.
Save Data (p. 71)
Measurement data, waveform data, screen images and numerical calculation results can be saved.
Calculate (p. 93)
Numerical measurement data can be applied to calculations.
View on a Computer (p. 91)
To download data from the CF card in the Memory HiCorder, connect a computer with the supplied USB cable. Recorded data can be analyzed on the com puter with the bun­dled Wave Processor program (p. A6).
Turn the Memory HiCorder Off (p. 34)
1.2 Measurement Flow
Starting and Stopping Measurement
Data Analysis
When Finished
15
Front Panel
Operating Keys
Rear Panel
Battery Compartment (p. 24)
The optional Model 9780 Battery Pack is installed here.
Display Screen (p. 63)
4.3-inch TFT color LCD Screen Configurations (p. 17)
Manufacturer's Serial Number
Shows the serial number. Do not remove this label, as it is required for product support.
Operating Keys
WAVE/DATA
Switches between Waveform and Numerical Value screens (p. 17).
SET
Displays the Settings screens, and switches among the screen tabs with each press (p. 18).
FILE
Displays file information (p. 18).
GAUGE
Alternately displays and hides the measurement scale on the Waveform screen.
(Manual trigger)
Press to trigger manually.
ESC
Cancels changes to settings.
Cursor Keys
Moves the position of the cursor (blink­ing selection) on the screen.
ENTER
Accepts displayed settings.
KEY LOCK
To lock the keys, press and hold the left and right cursor keys for three seconds. Repeat to unlock.
Press to save data manually (p. 77).
Use the left and right keys to scroll waveforms or move the A/B cursors. Press the middle key to select which of these actions the left and right keys perform (p. 19).
Setup and display
Saving operations
Scroll waveforms and
read cursor values
Choose a screen
Start and stop measure­ment. The LED at the left lights green while mea­suring (p. 20).
Start and stop
measurement

1.3 Names and Functions of Parts

1.3 Names and Functions of Parts
1
Chapter 1 Overview
3
16
Top Panel
AC Adapter Socket
Connect the supplied Model Z1005 AC Adapter (p. 28).
CF Card Slot
Insert an optional CF Card (p. 74).
BNC Jack
Connect optional measurement cables or clamp sensors (p. 29).
CHARGE LED
This LED lights when the battery is charging (p. 27).
LOGIC Connector
Connect an optional logic probe (p. 32).
External Control Terminals
Control signals can be received from and output to external devices (p. 105).
Right Side
POWER Switch
Turns the instrument on and off (p. 34).
Power Off
Power On
USB Port
(USB 2.0 mini-B receptacle) Open the dust cap and connect the USB
cable (p. 91). Connect to a computer to transfer data.
Strap Hole (p. 33)
1.3 Names and Functions of Parts
17
Press this key to switch between the screens.
Operational infor­mation is displayed along the bottom of the screen.
Waveform Screen
Data stored in the Memory Hi­Corder is displayed as wave­forms (p. 63).
Numerical Screen
Displays the real-time numeri­cal values of the measurement signals. Switch between in­stantaneous (peak-to-peak) and RMS values, or Display Hold. (p. 69)
Waveform/Numerical Screens
About the Icons
Clock
"Setting the Date and Time" (p. 102)
Power Source Indicator
Indicates the Memory HiCorder's power source.
AC adapter operation Battery pack operation
Fully charged battery pack Battery pack operation Low battery indicator during
battery pack operation
CF Card
Displayed when a CF card is inserted.

1.4 Screen Configurations

1.4 Screen Configurations
1
Chapter 1 Overview
3
18
Press this key to switch between the screens.
Operational infor­mation is displayed along the bottom of the screen.
Setting Screen
Configure measurement set­tings (p. 44). Set trigger criteria as needed (p. 55).
Calc/Save Screen
Configure numerical calcula­tions (p. 93) and auto-saving (p. 78).
Comment Screen
Enter per-channel comments (p. 52).
System Screen
Configure the system environ­ment (p. 99).
Settings Screens
File Screen
View and manage files on the CF card (p. 71).
File Screen
Operational infor­mation is displayed along the bottom of the screen.
1.4 Screen Configurations
19
Select the item to change.
Show available set­ting options.
Select the desired setting.
Changing screen contents
Setting Screen
Apply the new setting, or cancel it.
Waveform Screen
Scrolling a waveform
Earlier
Later
See: "Scrolling Waveforms" (p. 64)
Reading values at cursors
A/B cursor values
1
Display the A/B cursors.
Cursor type
2
Move the A/B cursors.
Values at cursor positions can be read numerically.
Select the cur­sor type, and which cursor(s) to move (p. 66).

1.5 Basic Operation

1.5 Basic Operation
Screen Operations
1
Chapter 1 Overview
3
20
Start Measurement
Press the START/STOP key. The green LED lights. When using the trigger function, the times when measurement starts and when recording (data acquisition) starts are not the same.
Finish Measurement
Trigger Mode: [Repeat] (default setting) Pressing the START/STOP key stops recording.
Trigger Mode: [Single] Recording stops when the specified recording time has elapsed.
Press and hold the right and left cursor keys for three seconds to lock the other keys, and repeat to unlock.
Insert a CF card. Display the screen to be saved,
and press the SAVE key.
Select [Screen Image] in the di­alog, and press the ENTER key to display the confirmation dialog. Select [Yes] and press ENTER again. The image is saved as a bitmap image file (named date_number. BMP).
1.5 Basic Operation
Starting and Stopping Recording
Acquire measurement data on the Memory HiCorder, and start recording. Recording stops according to the trigger mode
46).
Disabling Key Operations (Key-Lock Function)
Keys can be disabled to avoid inadvertent operations.
(recording criteria) setting (p.
Saving Display Images
Screen images of measurement data and s tured and saved (p. 82).
etting configurations can be cap-
21
Displayable waveform range
The displayable range is deter­mined by the waveform magnifi­cation (vertical zoom) and zero­position settings.
Current Input Signal Level
Measurement Units of Input Waveform
Using the Scaling function (p. 50), measurement units can be converted as desired for display.
Waveform monitor
Gauge
Shows the selected volt­age range (vertical axis) for channels 1 and 2.
1.5 Basic Operation
Verifying the Input Level (Level Monitor)
You can verify the input status and display range while making settings on the Setting screen. This is not available while measuring.
1
Chapter 1 Overview
3
22
1.5 Basic Operation
Measurement
Rear View
Install the 9780 Battery Pack (option) (p. 24)
1
Connect the AC adapter (p. 28)
3
(To save data) Insert a CF card (option) (p. 74)
4
2
Connect measurement cables (p. 29)
Logic signal measurement:
connect logic probe
Voltage measurement: connect connection cords
Current measurement: connect clamp sensors
Attach the strap (p. 33)
External control signal inputs External control signal outputs
(p. 105)
Turn the power on (p. 34) Execute zero adjustment (p. 35)
5
23
Preparations
Chapter 2
2

Chapter 2 Measurement Preparations

3
24

2.1 Using the Battery Pack (Option)

2.1 Using the Battery Pack (Option)
If commercial power is not available when the AC adapter is connected, the MR8870-20 Memory HiCorder can operate from the Model 9780 battery pack, so when using commercial power, the battery pack serves as a backup supply during power outages. Charge the battery pack fully before using it the first time.
Be sure to observe the following precautions. Incorrect han­dling may result in liquid leaks, h bursting and other hazards.
• Do not use a battery pack other than the 9780 Battery Pack. We cannot accept responsibility related to the use of any other batteries.
• To avoid the possibility of explos disassemble or incinerate battery pack.
• When storing the instrument, make could short-circuit the connectors are placed near them.
• The battery pack contains lye, whi comes into contact with the eyes. Should battery liquid get into your eyes, do not rub your eyes but rinse them in plenty of water and seek the immediate attention of a physician.
eat generation, ignition,
for accidents or damage
ion, do not short circuit,
sure no objects that
ch can cause blindness if
Installing and replacing the battery pack
• To avoid electric shock, turn of connect the cables before replac
• After installing or replacing the battery pack, replace the battery comp
• Handle and dispose of batteries regulations.
artment cover and screw.
f the power switch and dis-
ing the battery pack.
in accordance with local
25
2.1 Using the Battery Pack (Option)
Observe the following to avoid damage to the instrument.
• Use the battery pack in an ambient temperature range of 0 to 40°C and to be on the safe side.
• If the battery packs fails to complete charging within the stipu­lated time, disconnect the AC adapter from the pack to stop ch
• Consult your dealer or nearest se leaks, strange odor, heat, discoloration, deformation and other abnormal conditions occur during use, charging or storage. Should these conditions occur during use or charging, turn off and disconnect the instrument immediately.
• Do not expose the instrument to water and do not use it in exces­sively humid locations or locations exposed to rain.
• Do not expose the instrument to strong impacts and do not throw it arou
• The battery pack is subject to self-discharge. Be sure to charge the battery pack before initial use.
• When recharging a discharged battery pack in the Model MR8870-20, allow it to charge for at least ten minutes with the Model MR8870-20 turned off to maximize battery longevity.
• The battery pack is a consumable. If the battery capacity remains very low end. It can be recharged up to about 500 times.
• To prevent battery pack deterioration when the battery will not be u
sed for 1 month or longer, remove it and store it in a dry location with an ambient temperature range of between -10 to 30°C. Be sure to discharge and charge it every two months. Long-term storage when capacity has decreased will make charging impos­sible and reduce performance.
• Remove the battery pack from instrument once a month and che
• As long as battery charge remains, wave form data is retained in inte battery continues to gradually discharge. When the battery charge is depleted, stored waveform data is lost.
charge it in an ambient temperature range of 5 - 30°C
arging.
rvice station should liquid
nd.
after correct recharging, the useful battery life is at an
ck that appearance is normal.
rnal memory while the Memory HiCorder is turned off, so the
2
Chapter 2 Measurement Preparations
3
26
When to charge
When powering the instrument from the battery pack without the Model Z1005 AC Adapter, the low battery indicator ( ) is displayed when the battery charge is depleted, indicating that the battery pack requires charging. Note that if the Memory HiCorder is turned off under this situation, waveform data may fail to be backed up, and could be lost.
Approximate charging time:
About 200 minutes at 23ºC when recharged at the first appearance of the low battery indicator.
Approximate continuous battery-only operating time
(operating at 23°C)
• After a full charge, with the LCD backlight always on (default setting): approximately two hours
• After a full charge, with the LCD backligh t off (five minutes after last key­press): approximately 2.5 hours
See: "Enabling and Disabling the Backlight Saver" (p. 101)
2.1 Using the Battery Pack (Option)

Install the Battery Pack

Required tool:
One Phillips screwdriver (No.2)
Important
Disconnect the AC adapter when installing the battery pack.
Battery Compartment Cover
1
Place the Memory HiCorder face down, and remove the battery compartment cover screw and cover.
2
Connect the plugs of the battery pack to the con­nectors and insert the battery pack with the label side up.
Replace the battery compartment cover and screw.
3
Pull the battery pack plug straight out to remove the battery pack.
Make sure that the cables are not pinched.
2
Be sure to properly seat the connectors.
1
Connect the power cord to the inlet socket on the AC adapter.
1
2
Connect the output plug of the AC adapter to the instrument.
Plug the power cord into the mains outlet.
3
Refer to "2.2 Connecting the AC Adapter" (p. 28) for details about the AC adapter.
The CHARGE LED lights red when charging begins. Charging is finished when the CHARGE LED turns off.
27
2.1 Using the Battery Pack (Option)
2
Chapter 2 Measurement Preparations
3

Charge the Battery Pack

Regardless of whether the Mem ory HiCorder is on or off, the battery pack recharges whenever the Z1005 AC Adapter is plugged in to a power source. Therefore, charg­ing is provided by merely keeping the battery installed in the Memory HiCorder.
28
Connect the power cord to the inlet socket on the AC adapter.
Rated supply voltage is 100 to 240 VAC, and rated supply fre­quency is 50 or 60 Hz.
1
2
Connect the output plug of the AC adapter to the instrument.
Plug the power cord into the mains outlet.
3

2.2 Connecting the AC Adapter

2.2 Connecting the AC Adapter
Connect the power cord and the instrument to the supplied Model Z1005 AC Adapter, then plug the power cord into an outlet. When used with the battery pack in
stalled, the battery serves as an operating backup supply in case of power failure,
and the AC adapter otherwise has priority.
Before Connecting
• Use only the supplied Model Z1005 AC Adapter. AC ada pter input voltage range is 100 to 240 VAC (with ±10% st ability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this range.
• Turn the instrument off before connecting the AC adapter to the instrument a
• To avoid electrical accidents and to maintain the safety specifications of this instr only to a 3-contact (two-conductor + ground) outlet.
To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it from the power outlet.
nd to AC power.
ument, connect the power cord

2.3 Connecting Measurement Cables to the Memory HiCorder

Measure­ment cables
maximum rated volt­age between terminals
Maximum rated voltage to earth
Model L9197 600 VAC/DC
600 VAC/DC (CAT III) 300 VAC/DC (CAT IV)
Model 9197 Model L9198 300 VAC/DC
600 VAC/DC (CAT II) 300 VAC/DC (CAT III)
Model L9217 Model L9790 600 VAC/DC When using the L9790-01 Alliga-
tor clip, the 9790-03 Contact pin 600 VAC/DC (CAT II) 300 VAC/DC (CAT III) When using the L9790-01 Grab­ber clip 300 VAC/DC (CAT II) 150 VAC/DC (CAT III)
Model 9322 Differential Probe
2000 VDC, 1000 VAC When using the Grabber clip
1500 VAC/DC (CAT II), 600 VAC/DC (CAT III) When using the Alligator clip 1000 VAC/DC (CAT II) 600 VAC/DC (CAT III)
2.3 Connecting Measurement Cables to the Memory HiCorder
29
Connect the appropriate cables for the intended recording application.
• To record voltage waveforms (p. 31)
• To record current waveforms (p. 31)
• To record logic signals (p. 32)
Be sure to read the "Operating Precautions" (p. 6) before connecting the cables.
Connecting to the BNC terminals
Note the following maximum rated voltage between terminals and maximum rated voltage to earth.
Maximum rated voltage betwe
Maximum rated voltage to earth: 300 VAC/DC (CAT II) The following tables list the maximum ra terminals and the maximum rated voltage to earth for connec­tion cords. To avoid electric shock and instrument damage, do not i
nput voltages in excess of these figures. The lower of the instrument’s and the imum rated voltage between terminalss applies. When these voltage levels are excee be damaged. Do not make measurements as doing so may result in bodily injury. The maximum rated voltage to earth doe you make measurements with an attenuator or other compo­nent connected to the input. Consider your connection method and do not exce maximum rated voltage to earth.
en terminals: 400 VDC
ted voltage between
connection cord’s max-
ded, the instrument will
s not change, even if
ed the
2
Chapter 2 Measurement Preparations
3
For example, if the instrument is used in combination with Model L9198, the maximum rated voltage between terminals is 300 VDC and the maximum rated voltage to earth is 300 VAC (CAT II).
30
Measurement Object
Memory HiCorder
Logic Probe
2.3 Connecting Measurement Cables to the Memory HiCorder
Do not connect a cable to the instrument while it is connected to the object to be measured. Otherwise, an electric shock accident may occur.
To prevent damage to the instrument and sensor, never connect or disconnect a sensor while the power is on, or while the sensor is clamped around a conductor.
Connecting to the logic terminals
T o avoid electric shock and short circuit accidents or damage to the instrument, pay attention to the following:
• The ground terminal for the optional logic probe is not isolated from the Memory
order's ground (chassis
HiC ground). Therefore, if the measurement object con­nect to AC, it should have a gr
ounded, polarized plug, and be connected to the same power outlet as the Memory HiCorder’s AC adapter. If the Memory HiCorder and measurement object are con­nected to different mains circuits, or if an ungrounded power c ferent grounding paths may result in current flo w through a logic Memory HiCorder.
• Do not allow the m etal tip of a between conductors on the measurement object. Never touch the metal tip of a probe.
• Maximum rated voltage between terminal follow the indications below. Do not measure if the Maximum rate minals would be exceeded, as damage the instrument or personal Model 9320-01 Logic Probe: +50 VDC Model MR9321-01 Logic Probe: 250 Vrms (HIGH range), 150 Vrms (LOW range)
ord is used, the potential difference between the dif-
probe that could damage the measurement object or
logic probe to cause a short
d voltage between ter-
injury may result.
s of logic probe
31
Lock
BNC Jack Guide Pins (on the instrument)
BNC plug slots
Connect to the measurement object
(Example: Model L9198 Connection Cord)
BNC jack
1
2
Connect the BNC plug on the Connection Cord to a BNC jack on the Memory HiCorder.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock­wise, and pull it out.
Align the slots in the BNC plug with the guide pins on the jack on the instrument, then push and twist the plug clockwise until it locks.
(Example: Model 9018-50 Clamp-On Probe)
1
2
Connect the BNC plug on the clamp sensor to a BNC jack on the Memory HiCorder.
The connection method is the same as described for other Connection Cables.
OK
Attach the clamp around only one conduc­tor. Single-phase (2-wire) or three-phase (3-wire) cables clamped together will not produce any reading.
The arrows on the clamp indicating the direction of current flow should point toward the load side.
L
O
A
D
S
O
U
R
C
E
Conductor
Current flow
direction
arrow
The scaling function needs to be set according to the clamp sensor model being used. Refer to "Converting Units (Scaling function)" (p. 50) for details.
Clamp the sensor around the measurement object.
BNC jack
2.3 Connecting Measurement Cables to the Memory HiCorder

Connecting Connection Cords (to record voltage waveforms)

Connect an optional Hioki Connection Cord. Use the Hioki Differential Probe (P9000-01/02, 9322) if the voltage of the measurement object may exceed the Memory HiCorder’s maximum rated voltage between terminals.

Connecting a Clamp Sensor (to record current waveforms)

2
Chapter 2 Measurement Preparations
3
Connect an optional Hioki clamp sensor. Refer to the instructions pro clamp sensor for usage instructions.
vided with the
32
Connect to the measurement object.
(Example: Model MR9321-01 Logic Probe)
LOGIC jack
1
2
Connect the logic probe by aligning the groves on the plug and a LOGIC jack.
When a logic probe is not connected, the corresponding logic waveform appears at HIGH level on the waveform screen.
2.3 Connecting Measurement Cables to the Memory HiCorder

Connecting the Logic Probe (to record logic signals)

Connect the optional logic probe. Refer to the instruction provided with the probe.
33
Strap Hole
2
Feed the strap through the buckle.
Put the end of the strap under the retain­ing loop.
3
Push the strap through one of the strap holes on the instrument.
Repeat the same pro­cedure with the other end of the strap and the other strap hole.
1
4
Feed the strap through the buckle so that it is tight and not twisted.

2.4 Attaching the Carrying Strap

2.4 Attaching the Carrying Strap
Use the strap to avoid dropping the instrument while carrying, or when you need to hang it on a hook.
Attach both ends of the Strap securely to the instrument. If inse­curely attached, the instrument may fall and be damag ed when carr
ying.
2
Chapter 2 Measurement Preparations
3
34
Right Side
Turn the power switch on ( ).
If the " " indicator is not displayed, the instrument is not being powered from the AC line. In this state, the battery charge will be depleted if measuring for a long time, and measurement may be interrupted. Verify that the AC adapter is securely connected to an AC power source and to the instrument.
See: "Approximate continuous battery-only operating time" (p. 26)
Turning Power On
Turn the power switch off ( ).
The installed battery pack is charged whenever the AC adapt­er is plugged into a power outlet, even when the Memory Hi­Corder is off. In addition, providing there is some remaining battery charge, waveform data and the setting configuration are memorized whenever the Memory HiCorder is turned off, so that when the Memory HiCorder is turned back on, the same operating state is displayed.
Verify that the instrument and peripheral devices are correctly connected.
Turning Power Off
Execute zero-adjust 30 minutes after turning power on. (p. 35)
Initial screen
The power supply icon appears near the bottom right corner of the screen (p. 17).

2.5 Turning the Power On and Off

2.5 Turning the Power On and Off
Using the AC Adapter
Before turning the instrument on, make sure the supply volt­age matches that indicated on its power connector. Connec­tion to an improper supply volt instrument and present an electrical hazard. Rated supply voltage is 100 to 240 VAC (Voltage fluctuations of ±10% from the rated supply voltage ar
e taken into account.)
Rated supply frequency is 50 or 60 Hz.
age may damage the
35
Turn on the power and wait for 30 minutes to stabilize the internal temperature of the Memory HiCorder.
With the Waveform or Setting screen displayed, press and hold these keys simultaneously for one second.
The Zero Adjust dialog appears.
From the Setting screen
Select [Zero Adjust].
Apply
1
2
Zero adjustment is executed.
(simulta­neously)

2.6 Zero Adjustment

2.6 Zero Adjustment
Zero adjustment corrects for volt age offset at the input terminals, so that Mem­ory HiCorder measurements are relative to zero volts. Repeat zero adjustment if the ambient temperature changes suddenly.
2
Chapter 2 Measurement Preparations
3
Any zero-adjust setting is cleared upon system reset (p. 103).
36
2.6 Zero Adjustment
Setting before
Do not use if damage is present as you could re­ceive an electric shock. Replace the damaged items.
Metal Exposed
Is the insulation of the probe or connection cord to be used dam­aged, or is bare metal exposed?
When using probes and connection cords
When using a clamp
Is the clamp cracked or damaged?
Yes
1
No Metal Exposed
2
To
No
2
Peripheral Device Inspection
Before using the instrument the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
To
2
If damage is evident, request repairs.
Yes
Is damage to the instrument evident?
Instrument Inspection
When turning power on
Does the HIOKI logo appear on the screen?
No
The power cord may be damaged, or the instru­ment may be damaged internally. Request repairs.
Is the Waveform or Setting screen displayed? (When power is turned on the firs t time after purchase, the Setting screen should appear.)
No
The instrument may be damaged internally. Request repairs.
Nothing appears, or the display is abnormal
Yes
Yes
3
To
3
Input signals Inspection
Is “OVER” being displayed on the Numerical value screen?
Yes
A signal whose magnitude ex­ceeds the maximum rated termi­nal-to-terminal voltage is being input. Halt measurement, as this condition is hazardous.
Inspection complete
No
37

3.1 Pre-Operation Inspection

Measurement
Chapter 3
3.1 Pre-Operation Inspection
3

Chapter 3 Setting before Measurement

3
38
The measurement mode can also be changed using the measurement mode area on the display, located at the top right of screens such as the Waveform screen (p. 17) and the Setting screen (p. 18).
Connect
Connect measurement cables or a clamp sensor (p. 29).
Memory recorder mode
The instrument records instantaneous signal waveforms. Trigger and numerical calculation functionality can be used.
RMS recorder mode
The instrument records RMS waveforms for commercial power sup­ply(50Hz,60Hz) and DC signals.
When memory recorder mode has been selected
measure­ment mode
Selecting the mode from the screen that is displayed at power-on
Selecting the mode from other screens
When RMS recorder mode has been selected
The rest of this manual uses the memory recorder screen to describe instrument operation. For more information about RMS recorder mode settings, see the Measurement Guide (RMS Recorder).

3.2 Operating Flow Overview

3.2 Operating Flow Overview
39
Record
Manual configuration (p. 44)
Display the Setting screen
Automatic
configuration (p. 42)
Manual configuration
Configure
Configure input channels (p. 47)
Configure data saving (p. 78)
Analyze
Analyze recorded data on the Waveform screen (p. 63). Calculations can be applied to measured data (p. 93).
Save
Select and save data (p. 71).
When configuration is finished, start recording (p. 20).
Reload stored
settings (p. 85)
Specify recording criteria (p. 55)
(to record specific phenomena)
(to save auto­matically)
3.2 Operating Flow Overview
3
Chapter 3 Setting before Measurement
3
40
Selecting the Timebase (p. 44)Setting the Recording Length
(divisions) (p. 45)
Selecting Horizontal Display
Magnification (p. 45)
Selecting Logic Channel Waveform Colors
(p. 51)
Setting Logic Channel Waveform Display
Positions (p. 51)
Trigger Settings (p. 61)
Selecting the Voltage Axis
(Vertical Axis) Range (p. 47)
Selecting Waveform Display
Colors (p. 47)
Setting the Zero Position (p. 48)Selecting the Input Coupling
Method (p. 48) 1
Selecting Vertical Axis
Magnification (p. 49)
Selecting the Low-Pass Filter
(p. 49) 1
Setting Unit Conversion (Scaling)
(p. 50) 1
1. Can be set only from the
[Setting] screen.
Logic Channel Settings
Input Channel Settings
Time Axis (Horizontal Axis) Settings
Selecting the Trigger Mode (p. 46)Trigger Settings (p. 55)
Setting Recording Criteria
Automatic Setting Configuration (p. 42)Reloading Settings (p. 85)Saving Settings (p. 84) 1
Automatic Setting Configuration,
Saving and Reloading
3.2 Operating Flow Overview
Measurement Configuration (Setting Screen)
Before starting measurement, configure settings on the [Setting] screen according to the desired measurement. To observe their effects, settings can be made while viewing input waveforms
e waveform monitor window (p. 21).
in th Previously stored setting configuratio
ns can also be reloaded (p. 85).
Information about the currently highlighted item is display of the [Setting] screen.
ed along the bottom
41
Selecting the Timebase (p. 44)Setting the Recording Length
(divisions) (p. 45)
Selecting Horizontal Display
Magnification(p. 45)
Waveform Screen
Time Axis (Horizontal Axis) Settings
Selecting the Trigger Mode (p. 46)Trigger Settings (p. 55)
Selecting the Voltage Axis (Vertical
Axis) Range (p. 47)
Selecting Waveform Display Colors
(p. 47)
Setting the Zero Position (p. 48)Selecting Vertical Axis Magnification
(p. 49)
Input Channel Settings
Setting Recording Criteria
3.2 Operating Flow Overview
Measurement Configuration (Waveform Screen)
Certain setting items are available on the W aveform screen. Setting details are the same as for those on the Setting screen. Changing the range, recording length or trigge causes any existing measurement data to be erased as measurement restarts with the new settings (Restart).
r setting while measuring
3
Chapter 3 Setting before Measurement
3
42
Select [Auto Range]
Apply
1
2
Press and hold together for one second
A dialog appears.
Select the Waveform Screen
(simulta­neously)
If the appropriate range cannot be automatically detected, "Auto Range failed" appears. In this case, configure the range settings manually from the Setting screen while watching the waveform monitor.
Select the [Setting] Screen.
Select [Setup Wizard].
Apply
Apply
1 2
A dialog appears. Follow the instructions in the dialog.
Set according to the input waveform amplitude.
If appropriate settings for the input signal cannot be determined, select [Next] to display the manu­al range selection procedure, and follow the pro­cedure to make the settings.

3.3 Automatic Configuration of Measurement Settings (Auto Range)

3.3 Automatic Configuration of Measurement Settings (Auto Range)
If the appropriate measurement range settings are unknown, Auto Range can be enabled from the Waveform or Setting screen. Setting configurations saved to Memory HiCorder memory or to CF card can be reloaded for later reuse (p. 85).
Auto Range from the Waveform Screen
an installed
Selecting Auto Range on the Setting Screen
3.3 Automatic Configuration of Measurement Settings (Auto Range)
Auto Range Settings
Setting Item Settings
Timebase [Timebase] Auto setting value
Zoom (magnification) [Time Mag] Settings for CH1 and CH2 Voltage-axis range [Rng/div] Auto setup value Zero position [Pos] Auto setup value Zoom (magnification) [Mag] x1 Low-pass filter [L.P.F] Off Input coupling [Coupling] Trigger Criteria Trigger mode [Trig Mode] Repeat Trigger source AND/OR [Source] OR Pre-Trigger [Pre-Trig] 20% Trigger [Trig] Level trigger
Slope [Slope] (Rising) Level (Trigger Level) [Level] Auto setup value Filter [Filter] Off
Automatically displays 1 to 2.5 cycles on the Wave
form screen. When both channels are enabled for waveform display, the waveform on CH1 determines the timebase setting.
x1
DC
CH1 only. However, if CH1 display is dis­abled when executing Auto Range from the Wave
form screen, this setting is applied to CH2 only. This setting is also applied to CH2 when the difference between maximum and minimum values of the input signal on CH1 is two divisions or less.
43
3
Chapter 3 Setting before Measurement
3
When Using External Control Terminals
Executing Auto Range generates a trigger signal on the trigger output external control terminal. Bear this in mind when using the trigger output terminal and the Auto Range function.
44
Select the [Setting] Screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
Waveform monitor
1 division
Select from the [Timebase] setting options.
Setting options:
100, 200, 500 s, 1, 2, 5, 10, 20, 50, 100, 200, 500 ms, 1, 2, 5, 10, 30 s, 1, 2, 5 min [/div]
( default setting)

3.4 Measurement Configuration 1 (Horizontal Axis Settings)

3.4 Measurement Configuration 1 (Horizontal Axis Settings)
Make settings while viewing the Waveform monitor at the left side of the [Set­ting] screen.

Select a Timebase

Set the amount of time to display per horizontal division. Sampling rate = timebase (seconds/div)/100 (data samples/ di
vision), so at x1 display magnification, each division con­tains 100 data samples. Decreasing the timebase increases the resolution of data a
vailable for analysis. See "Selecting a Timebase (Horizon­tal Axis)" (p. 11). When the duration of 1 div. as shown on the Waveform scre
en is 50 ms or more, the waveform will be scrolled automatically as it is displayed during measurement (the Roll Mode display function).
45
20 div
Recording length
Select from the [Shot] setting options.
Setting options:
20, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000 [div], Cont (continuous)
( default setting)
Normal Display
Magnified
Display (x2)
Select from the [Time Mag] setting options.
Setting options:
x10, x5, x2, x1, x1/2, x1/5, x1/10, x1/20, x1/50, x1/100, x1/200, x1/500, x1/1000
( default setting)
3.4 Measurement Configuration 1 (Horizontal Axis Settings)

Setting Recording Length (no. of divisions)

Set the recording length in units of display divi­sions. Recording time = timebase (s/div) x recording length
(div)
Recording length (Cont) (continuous)
• The [Cont] recording length can be used when the time axis is 50 ms/div or slower. For example, setting the recording length (Cont) when the time axis is set to 10 ms/div will automatically set the time axis to 50 ms/div.
• Time values are not indicated during measurements.
• The trigger function (p. 55) and external trigger function (p. 107) are not available.
• The trigger mode (p. 46) is fixed at [Single].
When recording exceeds the maximum recording length (20,000 div), the following will take
ace
pl
• A total of 20000 divisions worth of data remains when recording is stopped.
• Data saved on a CF card during auto save and manual save extends a total of 20000 divi­sions back from the end of recording.
• The time values indicated in the waveform screen indicates the time at end of recording
0 s, a negative value.
as
3
Chapter 3 Setting before Measurement
3

Setting Display Zoom (as needed)

Select display magnification of the horizontal axis as needed.
46
Select from the [Trig Mode] setting options.
Setting options:( default setting)
Single Records one time and stops. RepeatRecords continuously until you press the
START/STOP key.

3.5 Measurement Configuration 2 (Setting Recording Criteria)

3.5 Measurement Configuration 2 (Setting Recording Criteria)

Setting the Trigger Mode

Select whether to record one recording length per trigger event, or to record continu­ously ([Repeat]). To specify other recording criteria required.
See: "Chapter 4 Making Trigger Settings" (p. 55)
, additional trigger settings are
When the trigger mode is set to [Rep processing (auto save, waveform display processing and calculation) before going to the next trigger standby status. Therefore, events that meet the trigger criteria during this pro­cessing period are not recognized as trigger events.
Recording length [Cont], the trigger mode is fixed at [Single].
eat], triggering is disabled during the end of recording
47
Select the [Setting] Screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
1 div
Select from the [Rng/div] setting options.
Setting options:
10, 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50 V [/div]
( default setting)
Measurable effective range is up to ±10 times of the voltage axis range (For 50 V/div, the maximum rated voltage between terminals is up to 400 V DC).
Select from the waveform display color setting options.
Setting options:
OFF ( x ), Red (* CH1), Green (* CH2), Blue, Yellow, Pink, Light Blue
( default setting)

3.6 Measurement Configuration 3 (Analog Channel Settings)

3.6 Measurement Configuration 3 (Analog Channel Settings)
Configure the analog input channels (CH1 and CH2) from the [Setting] screen.

Setting the Voltage Axis (Vertical Axis) Range

Select the voltage amplitude (voltage axis range) per verti­cal division for each input channel.
3
Chapter 3 Setting before Measurement
3
If the input signal exceeds the measurable range for the selected measurement range, the measured value will be shown as [OVER].

Selecting Waveform Display Color (as needed)

Different display colors can be selected for each input channel waveform.
48
Normal
Waveform(50%)
Changed Zero
Position(25%)
0 V
100%
0%
50%
0 V
25%
Select from the [Pos] setting options.
Setting options:- 50 to 150% (in 1% steps, when the [Mag] setting is x1), (default setting: 50%),
Magnification and compression (p. 49) in the voltage axis direction is based on the zero position. Although the range of voltage that can be displayed on the Waveform screen depends on the zero position and magnification/compression of the voltage axis, the measure­ment range is unaffected.
0V
Zero Position: 0%
50%
100%
(Magnification: x1)
Measurement range
Display range
0%
50%
100%
50%
100%
0%
0%
100%
50%
The valid setting range depends on display magnification. With x10 magnification (maximum setting range), -950 to 1050%
Select from the [Coupling] setting options.
Setting options:( default setting)
DC
DC Coupling Select this to acquire both DC and AC components of an input signal.
GND The input signal is disconnected.
Zero position can be confirmed.
3.6 Measurement Configuration 3 (Analog Channel Settings)

Setting the Zero Position (as needed)

The vertical display position of each waveform can be changed. Set the waveform zero position (in this exam­ple, zero volts) for display on the vertical axis.

Selecting the Input Coupling Method (as needed)

Select the coupling method for input signals.
3.6 Measurement Configuration 3 (Analog Channel Settings)
Normal Display
Zoomed Display
(x2)
Select from the [Mag] setting options.
Setting options: x10, x5, x2, x1, x1/2, x1/5
( default setting)
Normal Display
OFF)
Cut-Off Frequency
Selected
Select from the [L.P.F] setting options.
Setting options: OFF, 5Hz, 50Hz, 500Hz, 5kHz ( default setting)

Zooming the Vertical Axis (as needed)

Display magnification of the vertical (voltage) axis can be selected for each input waveform.
ification is applied relative to the wave-
Magn form’s zero position.

Selecting a Low-Pass Filter (L.P.F.) (as needed)

To suppress high-frequency components, select a low-pass filter.
49
3
Chapter 3 Setting before Measurement
3
50
Normal Display
(OFF)
Scaling Enabled
A
mV
Select from the [Scaling] setting options.
Setting options: OFF, Model, Ratio, 2-Point ( default setting)
A dialog appears when selected.
Ratio
Select the output ratio (conversion ratio) of the clamp sensor, or the attenuating ratio of the at­tenuating probe.
2-Point
Set the voltage values of two points of the input signal, the converted unit value of these two points and the name of the converted measure­ment units, so measurement values acquired as voltage are converted to the specified units. Example:
Voltage value at 2 points
Voltage of units to convert
V
H
: High value AH: Converted high value
V
L
: Low value AL: Converted low value
Model
Select your clamp sensor and attenuating probe model from the Model list. Setting options:
3283, 3284, 3285, 9010-50, 9018-50, 9132-50, 9657-10, 9675, CT9691, CT9692, CT9693, CT9667, 9322, P9000, 9660, 9661, 9669, 9694, 9695-02, 9695-03, CT6500, CT7044/CT7045/CT7046, CT7631/CT7731, CT7636/CT7736, CT7642/CT7742
A
H
A
L
VLV
H
[A]
[V]
Conversion ratio and offset value are calcu­lated from the two points and converted
Actual measurement values
Converted unit values
Valid setting ranges for conversion ratio and offset are as follows. Note that the conversion ratio cannot be set to zero.
-9.9999E+9 (conversion ratio, offset) -1.0000E-9 (offset) = 0
+1.0000E-9 (conversion ratio, offset) +9.9999E+9 Values outside the above range cannot be set. Scaled values (and cursor values when using the A/B cursors) can be verified in the waveform screen.
Y
A
HAL
V
HVL
-------------------
X
V
HAL
VLAH
V
HVL
---------------------------------------------+=
Conversion Ratio Offset
X: Voltage value Y: Converted value
*When using Model CT969x series in combination with
Model CT6590, execute zero adjustment.(p. A18)
3.6 Measurement Configuration 3 (Analog Channel Settings)

Converting Units (Scaling function)

Input voltage values can be converted to physi­cal units for display, such as by converting volt­age to current. Preset scaling settings are
d for current measurement with the
provide optional clamp sensor models.

3.7 Measurement Configuration 4 (Logic Channel Settings)

Select the [Setting] Screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
Select from the waveform display color setting options.
Setting options:
OFF*, Red, Green, Blue, Yellow, Pink, Light Blue
( default setting)
Position1
2 3 4
Select from the [Pos] setting options.
Setting options: Pos1, Pos2, Pos3, Pos4 ( default setting)
3.7 Measurement Configuration 4 (Logic Channel Settings)
Configure the input channels when measuring with a logic probe.
To use logic signals as a trigger source:
See: "4.3 Logic Channel Trigger Settings" (p. 61)
51
3
Chapter 3 Setting before Measurement
3

Setting Logic Channel Display Colors (as needed)

You can select a display color for each logic channel waveform

Setting Logic Channel Display Positions

You can select the display position of each logic waveform. Waveform overlap on the display
n be minimized when recording simulta-
ca neously with an analog waveform.
52
Select the [Comment] Screen.
Select the field in which to enter the comment
1 2
Apply
The character entry screen appears.
3
Select the characters to enter
To enter a space, select a blank character.
Apply
Characters appear in the en­try field as you select them.
Navigate the entry field
Select [OK]
Accepts your entry and returns to the [Comment] screen.
Finished with entry
Entry field

3.8 Entering Comments

3.8 Entering Comments
Comments of up to 40 characters can be entered as a title for the me asurement data, and as a label for each input waveform. The comments can be displayed when the data is viewed in an application program on a
computer.
The following characters are converted as follows when saved to a text file.
Entered character Converted plain
text equivalent
^2 ^3 ^u ~o ~c ~e
Operating Panel
Only the keys displayed on the operating panel are enabled.
List (WAVE/DATA key) Measurement related terms are registered in the list beforehand.
Hist (History, SET key) You can select previously entered terms from the pick list.
Char Input ( key) Switches between the pick list and character selection frames.
/ (Cursor up/down key) Selects from the pick list frame.
Select (ENTER key) Accepts the item selected in the pick list.
Cancel (ESC key) Cancels entry.
<</ >> (SCROLL/CURSOR left/right key) Move input position.
BS (SCROLL/CURSOR middle key) Deletes one character.
OK (START/STOP key) Accepts entry.
The operating panel depicts the functions of Memory Hi­Corder's operating keys.
53
3.8 Entering Comments
3
Chapter 3 Setting before Measurement
3
54
3.8 Entering Comments
Making Trigger
Trigger criteria met
T
T
T
Move to a setting item.
Apply
Open the setting op­tions for the item to be set.
Select from the listed options.
1 2
Select the [Setting] Screen.
55

4.1 Setting Recording Criteria

Settings
Triggering is the process of controlling the start and stop of recording by specific signals or conditions (criteria). When rec
ording is started or stopped by a specific signal, we say the trigger is "applied" or "triggering occurs". In this manual, indicates a "trigger poi which a trigger is applied.
The trigger function cannot be used in RMS recorder mode.
Signals that can be used for triggering (trigger sources) are as follows.
Trigger Source Description Ref.
Analog Trigger
Logic Trigger
External Trigger Manual Trigger Applies a trigger by pressing . (p. 62)
A trigger can be applied by combining (AND/OR) criteria from multiple trigger sources (except manual triggering). When the r ecor din g leng th is [Con t], th e tri gger func tion is no t avai la ble.
Applies a trigger according to a signal input on an analog ch
annel. (Level, In-Window, Out-of-Window, or Voltage
Sag trigger) Applies a trigger according to signals input on logic chan-
nels (Ch A to Ch D). Applies a trigger according to an input signal at the
EXT.TRIG ter
minal (External Trigger Input)
Chapter 4
nt", as the time at
(p. 57)
(p. 61)
(p. 105)
4.1 Setting Recording Criteria
3
4

Chapter 4 Making Trigger Settings

Make these settings when selecting the type of triggering to be applied to input channel (CH1 or CH2) recording.
56
Select from the [Source] setting options.
Setting options:( default setting)
OR
Triggering occurs when any one of the speci­fied trigger source criteria is met.
AND Triggering occurs only when all of the speci-
fied trigger source criteria are met.
[AND]
One waveform has crossed above 0V as the other crosses on the upslope
[OR]
Either waveform crosses 0V on the upslope
T
Trigger.......Level
Filter..............Off
Level .............0 V
Slope  
To apply a trigger when the upslope () of the waveform crosses zero volts
When the trigger combining logic is set to [AND]
If trigger criteria are already met when you press the START/STOP key, no triggering oc­curs. Triggering occurs only after all trigger sources have ceased to meet the criteria at once, and are subsequently met again.
T T T
Start Measurement
Start Measurement
When the recording length is [Cont], the trigger source setting is disabled.
80%
20%
T
Select from the [Pre-Trig] setting options.
Setting options:
0, 5, 10, 20, 30, 40, 50, 60, 70, 80, 90, 95, 100%
( default setting)
Difference between [Waiting for pre-trigger] and [Waiting for trigger]
When measurement is started, the specified pre-trigger length is recorded. This period is indicated as the [Waiting for pre-trigger]. After the specified pre-trigger length has been recorded, the period indicated as [Waiting for trigger] continues until a trigger occurs. During the [Waiting for pre-trigger] period, trigger events are not recognized even when the trigger criteria are met.
When the recording length is [Cont], the pre-trigger setting is disabled.
4.1 Setting Recording Criteria

Setting Combining Logic (AND/OR) for Multiple Trigger Sources

Analog, logic, and external trigger criteria can be combined by AND/OR logic to define complex trigger criteria.

Setting the Trigger Start Point (Pre-Trigger)

Make this setting to specify a portion of the waveform to be recorded prior to a trigger event, as a percentage of the o
verall specified recording length.
57
Select the [Setting] Screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
Displays the Waveform screen.
Select [ ].
T
Apply
1 2
Trigger setting items are displayed. Set the trigger level, slope, trigger filter and frequency.
To return to the channel setting display , select [ ] again, and press the ENTER key.
Channel setting display
Trigger setting display

4.2 Triggering by Analog Signals

4.2 Triggering by Analog Signals
Make this setting to define trigger criteria according to input signal cha racteris­tics on CH1 or CH2.
Settings can also be made on the Waveform screen.
3
4
Chapter 4 Making Trigger Settings
58
Select from the [Trig] setting options.
Setting options: OFF, Level, In, Out, Drop ( default setting)
Trigger Level
Input
Trigger Slope
[ ]
[ ]
T T
T
T
Upper
Threshold
Lower
Threshold
Upper
Threshold
Lower
Threshold
TT
1/2 Period
Trigger Level
T
Apply a trigger when the input signal exceeds 200 mV
Trigger............... Level (Level Trigger)
Level..................200 mV
Slope (rising)
Apply a trigger when the input signal is outside of the range ±1 V
Trigger............... Out (Out-Window
Trigger)
Upper................. 1 V
Lower................. -1 V
Apply a trigger if a 50-Hz mains frequency input signal, nominally about 220 Vrms (311.1 Vpeak), drops below 198 Vrms (280.1 Vpeak)
Trigger................ Drop
Level................... 280 V
Freq.................... 50 Hz
4.2 Triggering by Analog Signals

Selecting an Analog Trigger Type

Select the appropriate trigger type. Setting options differ by trigger type.
Type of Analog Trigger
Trigger Example Description
Level Trigger
[Level]
In-Window Trigger
[In]
Out-of-Win­dow Trigger
[Out]
Voltage dip Trigger
[Drop]
Setting Example
A trigger is applied when an input signal crosses the specified trigger level (threshold voltage).
A trigger is applied when the input signal enters a range defined by upper and low­er thresholds.
A trigger is applied when the input signal exits a range defined by upper and lower thresholds.
A trigger is applied when the amplitude of the input signal (at commercial mains fre­quency) dips below the specified trigger leve
l.
4.2 Triggering by Analog Signals
Noise
Filter Width
T
Select from the [Filter] setting options.
Setting options: OFF, 10, 20, 50, 100S ( default setting)
TT
Select from the [Slope] setting options.
Setting options: ( default setting)
 A trigger occurs when the signal crosses the
threshold on the upslope (rising edge).
A trigger occurs when the signal crosses the
threshold on the downslope (falling edge ).
T
Select from the [Level] setting options.
Setting options: One twentieth of the voltage range [Rng/div] (Default setting: Zero volts for Level Trigger mode, and one division above the zero position for Voltage Sag trigger mode.)

Selecting a Trigger Filter

Sets the filter width (trigger filter) for triggering. Sup­presses triggering from noise. The filter width setting is defined according to the number of
samples* of acquired data.
* Number of samples: 1 division = 100 samples (with x1 display magnification)

Selecting Rising or Falling (Slope) of the Input Signal

Set the slope (rising () or falling () edge) of the input sig­nal on which to trigger.
59
3
4
Chapter 4 Making Trigger Settings

Setting the Signal Threshold Level (Trigger Level)

Set the signal threshold level (volt triggering is to occur. The trig instantaneous value (not RMS). The setting value can be verified in the waveform monitor or on the Waveform screen. When scaling is enabled, the scaled trigger level value can be verified in the waveform monitor.
age value) at which the
ger level setting value is an
60
Select from the [Lower]/[Upper] setting options.
Setting options: One twentieth of the voltage range [Rng/div] (Default setting: One division below the low­er limit zero position, and one division above the up­per limit zero position.)
Select from the [Freq] setting options.
Setting options: 50Hz, 60Hz ( default setting)
4.2 Triggering by Analog Signals

Setting Lower and Upper Trigger Thresholds

Set the upper and lower thresholds for a trigger window. Upper and lower thresholds can be verified in the waveform monitor or on the Waveform screen. When scaling is enabled, the scaled threshold values are displayed.

Selecting Measurement Frequency

Select the frequency of the commercial mains power to be measured.

4.3 Logic Channel Trigger Settings

Select the [Setting] Screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
Select from the [Trig] setting options.
Setting options:( default setting)
OFF
Logic triggering is disabled.
OR Triggering occurs when input signal logic
matches any setting in the trigger pattern.
AND Triggering occurs only when input signal logic
matches all settings in the trigger pattern.
Trigger pattern
1 0
X X
T
[OR]
1 0
X X
[AND]
T
When the trigger combining logic is set to [AND]
If trigger criteria are already met when start­ing measurement, no triggering occurs. Instead, triggering occurs once as soon as the criteria are no longer met, and again when the criteria are subsequently met.
4.3 Logic Channel Trigger Settings
Make these settings to utilize logic channel input signals for triggering. A trigger event occurs when the logic signal p ical AND/OR trigger criteria. You can enable a trigg mum interval (the filter width) during which trigge triggering occurs.
attern matches the specified log-
er filter to specify a mini-
r criteria must be met before
61
3
4
Chapter 4 Making Trigger Settings

Configuring Logical (AND/OR) Trigger Criteria

Specify a logical pattern of logic inputs to serve as trigger criteria.
62
1 0
HIGH LOW
Select from the [Pattern] setting options.
Setting options:( default setting)
X Ignore signal 0 Triggering occurs when the signal transitions
from HIGH to LOW.
1 Triggering occurs when the signal transitions
from LOW to HIGH.
Noise
Filter width
T
Select from the [Filter] setting options.
Setting options: OFF, 10, 20, 50, 100 S ( default setting)

4.4 Triggering Manually (Manual Trigger)

Setting a Trigger Pattern

Specify the state of each input signal required for trigger­ing.

Setting a Trigger Filter

To enable trigger filtering, just specify a filter). The filter can prevent f The filter width setting is defin of samples* of acquired data. * Number of samples: 1 division = 100 samples (w ith x1 di
splay magnification)
filter width (trigger
alse triggering due to noise.
ed according to the number
4.4 Triggering Manually (Manual Trigger)
Triggers can be applied manually. Manual triggering takes priority over all other trigger sources, regardless of settings.
At any time after starting measurement (by pressing the START/ST OP key), you can press the key to apply a manual trigger. However when the input channel trigger setting is OFF ing measurement, the [Waiting for trigger] mod be generated by pressing the
key.
. After setting trigger criteria and start-
, manual triggering is disabled
e is active, during which a trigger can

5.1 Viewing Waveforms

Logic waveforms
Analog waveforms
Displays acquired data as waveforms.
A/B Cursor Values
(when A/B cursors are displayed) Shows the values at cursor locations on the waveform (p. 66). This can be switched to show numerical calculation results (p. 93).
Time Values
Trigger Mark ( )
The trigger mark appears at the point where a trigger is applied.
T
Scroll Bar
Shows the range and position of the displayed waveform.
Viewing the Scroll bar (p. 64)
Displays a gauge (p. 66)
Measurement Configuration
You can change these settings (p. 41).
Analog Trigger Settings (p. 55)
Jump Function (p. 65) Trace Display (p. 64)
Select the Waveform Screen.
Move to a setting item
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
Waveform Analysis Chapter 5
5.1 Viewing Waveforms
Waveform Screen
63
3
5

Chapter 5 Waveform Analysis

64
Screen Display
Earlier
Later
Time Axis Magnification x1
x2
x1/2
When Recording Length is
200 div
Screen Display Range (20 div)
Scroll bar
Whole Waveform (200 div)
Even for identical recording lengths, the width of the screen display range indica­tion in the scroll bar depends on the specified magnifica­tion/compression of the time axis.
Select from the [x1](display magnification) setting options.
Setting options:
x10, x5, x2, x1, x1/2, x1/5, x1/10, x1/20, x1/50, x1/100, x1/200, x1/500, x1/1000 ( default setting)
5.1 Viewing Waveforms

Scrolling Waveforms

When measuring or displaying an existing waveform, use the SCROLL/CURSOR keys to scroll. During Roll Mode display,* manually scrolling terminates the Roll Mode, so you can freely view any part of the measured waveform. To return to Roll Mode, move the cursor to [T the bottom right corner of the display, and press the ENTER key. Alternatively, just scroll to the right end of the waveform to resume Roll Mode.
out the Roll Mode display:
* Ab When each div shown on the Waveform screen is at least 50 ms, the part of the waveform is always displayed. To jump to the start or end of a waveform (or between trig­ger events), hold the ESC key while pressing the right or lef
t SCROLL/CURSOR key.

Verifying Waveform Display Position

From the scroll bar you can verify the relative position and size of the displayed portion of a waveform within the over­all recorded waveform. Trigger A/B cursor positions (when using vertical or trace cursors) are also displayed.
screen scrolls automatically so that the newest
a waveform
race] near
time, trigger position and

Magnifying and Compressing Horizontally (Time Axis)

Data details can be observed by magnifying the waveform along the time axis. Also, by compressing the time axis,
verall waveform fluctuations can be readily seen. Time axis
o magnification can be changed while measuring, and when the A/B cursors are displayed, magnification can be set according to cursor position (but only when finished measur­ing). Memory recorder mode from the magnification, RMS
er mode Select from the settings of the time axis.
record
Select [Jump].
Apply
1
2
The location of the cur­rently displayed portion of the overall waveform is indicated by the mark on the scroll bar.
3
Move the mark to display a different por­tion of the waveform.
The portion of the wave­form at the mark is displayed.
Apply
Select the Waveform Screen.
4
65
5.1 Viewing Waveforms

Viewing Any Waveform Location (Jump Function)

When the recording length of a waveform is long, you can specify the portion to be displayed immediately.
3
5
Chapter 5 Waveform Analysis
66
A
B
Changing Cursor Type
Cursor Values
[Motion] A Cur
B Cur AB Cur
( default setting)
1
Displays the A/B cursors and the values at the cursor positions.
To hide the cursor items again, press the SCROLL/CURSOR key.
2
Move the cursors to display numer­ical values along the waveform.
Hold the ESC key to move by whole divisions.
To select which cursor(s) to move
Select from the [Motion] setting options.
To select the type of cursor values to display
Select from the [Type] setting options.
[Type] Trace
Vert (Vertical) Horz (Horizontal)
( default setting)
Cursor A Cursor B
The current [Motion] selection can be changed directly, by holding the ESC key while pressing the central SCROLL/CURSOR key.

5.2 Viewing Measurement Values

5.2 Viewing Measurement Values

Displaying Gauges

Press the GAUGE key to display gauges at the left side of the screen corresponding to the measurement range of e
ach channel. Measurement values can be confirmed on the gauges. The color of each gauge matches the color of its waveform.
Press the GAUGE key again to hide the gauges.

Displaying Cursor Values

Time difference, frequency and potential difference (and when scaling is enabled, scaling values) can be read as n
umerical values using the A/B cursors. See "About Cursor Values" (p. 67)
About Cursor Values
Trace Cursor case
A Cursor value B Cursor value B-A value
AB
B-A
B-A
AB
B-A
A
B
B-A
Cursor Type Example Cursor Value
Trace Cursors
Displays the time and measurement values at the A/B cursors, or the time and measurement differences be­tween the A/B cursors. Displays the intersections (trace points) of cursors and wa
veforms. (the intersections of waveform traces of se-
lected channels) (Time and Measurement Values)
Time Values A Cursor value, B Cursor value: Time from trigger point o
r recording start
B-A value: Time difference between A/B cursors Measurement Values
A Cursor value, B Cursor value: Measurement value B-A value: Difference between measured values at A/B cursors
67
5.2 Viewing Measurement Values
3
5
Chapter 5 Waveform Analysis
Vertical Cursors
Horizontal Cursors
Displays the time and frequency values at the A/B cur­sors, or the time and frequency differences between the A/B cursor
(Time Value and Frequency) Time Values (t): Time from trigger point or recording st
art Frequency (f): Frequency having period t B–A value: Time difference between A/B cursors
Displays the measurement values at the A and B cur­sors for the selected channel(s), or the difference be­tween A/B cursor values.
(Measurement Values) A Cursor value, B Cursor value: Measured value of ch
annel
B–A value: Difference between values at A/B cursors
s.
68
1
The A/B cursors and cursor values are displayed.
3
Move the A/B cursors to specify the time span.
Select [Motion].
2
Apply
Open the setting options for the item to be set.
Select which cursor(s) to move.
[Motion] A Cur B Cur AB Cur
5.2 Viewing Measurement Values

Specifying a Waveform Time Span

Specify a waveform time span when saving a partial waveform or applying numerical calculations (Trace cursors or Vertical cursors).
69
CH1 Numerical Value
CH2 Numerical Value
RMS display mode indicator (Indicates the current mode: RMS or instantaneous (P-P))
Display Hold
The displayed val­ue is held con­stant.
Normal Display
Scaling Indicators
Gauge
Maximum count is 5500
Instantaneous logic values (: 0, : 1)
Scaling Unit Indicators
Appears only when scaling is enabled for that channel (p. 50).
Units displayed af­ter conversion by scaling settings
These depict op­erating key func­tions
Select the Numerical Screen.
1
2
Switching between instantaneous and RMS values
Holding and releasing display values
Toggles only CH1
Press these keys to toggle the selection
Toggles both CH1 and CH2
Toggles only CH2
Press these keys to toggle the selection
3
5.2 Viewing Measurement Values

Viewing Input Signals as Numerical Values (Numerical Value Display Function)

AC voltage measured at commercial mains frequencies (50/60 Hz) or DC voltage can be displayed as a numerical value. Either instantaneous or RMS numerical value can be displayed. Instantaneous value: the input voltage RMS value: the root-mean-square of the Note that the RMS value is only displayed correctly for commercial mains frequen­cies (50/60 Hz) and DC signals. If the input voltage peak value exceeds 450 V, “OVER” will be shown.
actual input voltage (p. 97)
3
5
Chapter 5 Waveform Analysis
70
5.2 Viewing Measurement Values
Display when the input voltage peak value exceeds 450V
Saving &
71

6.1 About Saving and Loading Data

Loading Data
Chapter 6
Measurement data acquired by the Memory HiCorder can be saved to a CF card.
See: "6.2 Using a CF Card" (p. 74)
*To load data saved by the MR8870-20 in memory recorder mode with the 8870-20, upgrade the 8870-20 to firmware version 2.00 or later.
See: "Appendix 9 To 8870-20 Users (Firmware Ver. 1.08 or Earlier)" (p. A19)
Methods available for saving data consist of Auto Save, for saving data auto­matically while measuring; and Select & Save and Quick Save for saving data a
fter measurement is finished. Refer to "6.3 Saving Data" (p. 77) for details.
6.1 About Saving and Loading Data
When saving data, a folder named in the folder as follows. Up to 1,000 files can be saved in one folder.
File Type
Settings Data Binary CONFIG Settings List Binary CONFIG
Waveform Data*1
Numerical Cal­culation Results
Captured Screen Image
*1.To reload data into the instrument or load it into the Wave Processor program:
Save it in binary format. Waveforms and some of the measurement settings are saved. To save a waveform partially, specify a period by using A/B cursors (p. 68).
*2.BMP Format: This is a standard Windows graphics format. These files can be handled by
ma
ny graphics programs.
*3.When using auto-save functionality, a folder named AUTOhhmmss is created under a
folder named with the date of saving (based on the trigger time), where hhmmss is the measurement start time. (Example: The file AUTO131031 is created at the time 13:10:31)
File Format
Binary Text
Text
BMP *2
Folder Name
Date of saving (e.g. 14-03­30
)
*3,*5,*6
[HIOKI8870] is created, and files are stored
O = Available, X = Not available
File Name (Auto-numbered from 1)
CONFIG0001.SET*7 CONFIG0001.SRM*8
LIST0001.BDL X O O X WAVE0001.MEM*7
WAVE0001.RMS*8 WAVE0001.CSV
MEAS0001.CSV
SCR0001.BMP*4
Save
Manu-
Auto
al
X O O X
O O O O O O X O
O O X O
O O O O
Load
PC­Rea able
d-
3
6

Chapter 6 Saving & Loading Data

72
6.1 About Saving and Loading Data
*4.Also when auto-saving, screen images are saved in the file named [WAVE0001.BMP]. *5.When the retrieved waveform is re-saved as waveform or display image data, it is saved
*6.When delete-and-save has been selected, a folder named
*7.This filename is used when saving data in memory recorder mode. *8.This filename is used when saving data in RMS recorder mode.
See: "Appendix 2 File Naming" (p. A5)
r the date (year/month/date) folder (based on the trigger time) where the retrieved
unde waveform file is saved.
[AUTO_D] will be created in-
side the HIOKI8870 folder, and all data for which auto-save has been enabled will be saved
in that folder. All data saved in this folder is targeted by delete-and-save ope ration
(i.e., potentially subject to deletion).
Setting data (.SRM) and waveform data (.RMS) saved in RMS recorder mode cannot be loaded in memory recorder mode. To load these files, first switch the instrument to RMS recorder mode from any screen.(p. 38)
6.1 About Saving and Loading Data
HIOKI8870
WAVE0003.MEM
(Settings data folder)
CONFIG
(Folder for all data other than settings)
14-03-30 (Date)
14-04-01
Up to 1000 files
(Automatically saved folder including its creation time)
CONFIG0001.SET CONFIG0002.SET
WAVE0001.MEM
AUTO131031
WAVE0002.MEM
LIST0001.BDL LIST0002.BDL
AUTO_D
14-04-012
WAVE0004.MEM
Up to 1000 files
(Folder targeted for delete-and-save operation) Files are saved in this folder when both auto-save operation and delete-and-save operation are en­abled (excluding numerical calculation results files). Once the CF card becomes full, the files here will be deleted in the order of their creation, starting with the oldest file.
Directory Structure for Saved Data
Model MR8870-20 waveform data size list (Memory recorder mode)
73
3
6
Chapter 6 Saving & Loading Data
Recording
length(div)
20 12.8 KB 82.5 KB
50 27.4 KB 205.6 KB 100 51.8 KB 410.6 KB 200 100.7 KB 820.8 KB 500 247.1 KB 2.0 MB
1,000 491.3 KB 4.0 MB 2,000 979.6 KB 8.0 MB
5,000 2.4 MB 20.0 MB 10,000 4.8 MB 40.1 MB 20,000 9.5 MB 80.1 MB
In binary saving, all channels will be saved, regardless of whether the displa y is ON
Binary Text
or OFF. When saving as text, only the channel that is disp The save column in the chart above is an example of the case of displaying and sav­ing all channels. For example, when using the 9729 PC Card (1GB), a record length 20,000div files can be saved.
layed will be saved.
pproximately 100 binary data of
74
Hioki options PC cards (includes adapter)
• Model 9726 PC Card 128M (*End of sale)
• Model 9727 PC Card 256M (*End of sale)
• Model 9728 PC Card 512M
• Model 9729 PC Card 1G
• Model 9830 PC Card 2G
Important
Use only CF cards sold by Hioki. Compatibility and performance are not guaran­teed for CF cards made by other manufactur­ers. You may be unable to read from or save data to such cards.
This Memory HiCorder does not require a CF (CompactFlash) adapter. See "6.5 Data Management" (p. 88) for details about managing data storage in the card.

6.2 Using a CF Card

6.2 Using a CF Card
The following options are available for saving and loading measurement data with the Memory HiCorder.
• Format new CF cards before use. To format with this instrument: (p. 76)
• Inserting a CF card upside down, backwards or in the wrong dire
ction may damage the CF card or instrument.
• Never eject a CF card while it ment. Data on the CF card could be lost.
• During battery-only operation, data may fail to be stored properly if bat
tery charge is exhausted while saving. In the worst case, the CF card could be damaged, so pay appropriate attention to the discharge state of the battery pack.
• If the Eject button is in the released position, press it in first before inserting the CF card all the way in. Inserting the CF card when the Eject button is released may damage the instrument. If the CF card does not go all the way in, do not force it in. Press the Eject button once to release it, then press it ag ain and insert the CF card all the way in.
• As the CF card is sensitive to static electricity, damage to the CF
or wrong operations by the this instrument may occur due to
card static electricity. Please be careful when handling it.
is being accessed by the instru-
• The Flash memory in a CF card has a limited operating life. After long-term usage, data storage and retrieval become difficult. In this case, replace the CF card with a new one.
• We cannot provide compensation for data loss in a CF card, regardless of content or cause of the damage. Always maintain a backup of important data stored on a CF card.

CF Card Insertion & Removal

Eject Button
Inserting a CF card
1 Open the CF card slot cover. 2 Press the Eject button in if it is in the
released position.
3 Face the CF card with the arrow mark
() on top, and insert it in the direc­tion of the arrow all the way in the slot.
Removing a CF card
1 Open the CF card slot cover. 2 Press the Eject button (to release it). 3 Press the Eject button again and pull
the card out.
Cover
Top View
75
6.2 Using a CF Card
3
6
Chapter 6 Saving & Loading Data
76
Select the File Screen.
Select [Format].
Apply
Apply
1 2
The Format Confirmation dialog appears.
Select [Yes].
Open the Control Dialog box
Apply
The Format Confirmation dialog appears again.
Select [Yes].
6.2 Using a CF Card

Formatting a CF Card

CF Cards are formatted from the Memory HiCorder’s File screen.
Formatting irretrievably erases all data on the CF card. Always backup important data from the CF card before formatting.
6.3 Saving Data
Press the SAVE key to save the data previously specified on the System screen. This is convenient for saving certain types of data, or when you want to save immediately while monitoring (p. 80).
(Default setting) Press the SAVE key to se­lect and save the selected contents. No settings are needed be­fore saving this way (p. 80).
Automatically saves all ac­quired measurement data for the specified recording length. Various types of data can be saved at the same time. Select the save contents before measuring (p. 78).
To save automatically while measuring
To save immediately upon pressing the SAVE key
To save selected contents
Auto Save Quick Save Select & Save
Insert a CF card.
Set the Auto Save setting
[ON].
Select what to save.
Enable delete-and-save
operation (as needed).
Measurement
Saving proceeds
automatically
Calc/Save Screen
The following can be saved simultaneously:
• Waveform Data
• Numerical Calculation Results
• Screen Image
Insert a CF card.
Set the function of the
SAVE key to
[Quick Save].
Select what to save*.
Press the SAVE key*.
Save
System Screen
• Waveform Data
• Numerical Calculation Results
• Screen Image
Insert a CF card.
Set the function of the
SAVE key to
[Select & Save].
Press the SAVE key.
Select what to save
Press the ENTER key.
Save
System Screen
Save Dialog
• Waveform Data
• Numerical Calculation Results
• Screen Image
: To save a waveform partially, specify a period by using
A/B cursors. (Saving a waveform section is not available with Auto Save) To save a screen image, display the desired screen be­fore saving.
Basically, three methods are available for saving.
77

6.3 Saving Data

3
6
Chapter 6 Saving & Loading Data
78
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1
2
Select the [Calc/Save] Screen.
Set [Waveform] to [ON].
From the [Format] options, select the data sav­ing format.
Binary Select this format to be able to reload
the saved data into the instrument, or so that it can be loaded into the Wave Pro­cessor program.
Text Select this format to save the data as a
text file.
(When [Text] format is selected)
From the [Thinning] options, select the amount of data thinning desired.
1/2, 1/5, 1/10, 1/20, 1/50, 1/100, 1/200, 1/500, 1/1000, OFF(* default setting)
1/2: Saves every other data point
( x x x)
1/5: 5Saves one of every five data points
(xxxxxxxxxxxx)
To Save Waveform Data
Saving Screen Image s
Set [Screen Image] to [ON].
[Format]
BMP (fixed)
This is selected automatically when setting [Screen Image] to [ON].
To Save Numerical Calculation Results
Set [Calc Results] to [ON].
Numerical calculation settings are necessary (p. 93).
After making any other necessary settings, press the START/STOP key.
Data is automatically saved to the CF card when measurement finishes. "Directory Structure for Saved Data" (p. 73)
Set [Auto Save] to [ON], and select the item(s) to save.
3
6.3 Saving Data

Automatic Saving

Before measuring, configure saving on the [Calc/Save] screen. Waveform data, screen images and numerical calculation results can be saved at the sa
me time. Before starting to measure, confirm that Auto the CF card is properly installed.
Save is configured correctly, and that
When the time axis setting is 50 ms/div. or slower, data will be saved while measurement is being performed. During automatic saving, a red progress bar will be shown on the Save in Progress dialog box. When the time axis setting is faster than 50 ms/div., data will be saved once measurement is complete.
79
[ON]
Once available space on the CF card becomes limited, the file with the oldest date of creation in the folder
[HIOKI8870]-[AUTO_D] will be
deleted to make space for a new file.
[OFF]
Data will be saved as long as there is space re­maining on the CF Card. Once the card be­comes full, data will no longer be saved.
6.3 Saving Data

Enabling Delete-and-save Operation

When delete-and-save operation is enabled, the oldest data will be deleted to make space for new files once available space on the CF card becomes limited during measurement that is performed with auto-save enabled.
3
When enabling Auto Save operation in RMS recorder mode, ensure that the CF card has at least 10 MB of available space when measurement is started. An error will result if there is not adequate space. On the other hand, when enabling delete-and-save operation, 10 MB of available space obtained by deleting files in the folder will avoid an error.
6
Chapter 6 Saving & Loading Data
[HIOKI8870]-[AUTO_D]
80
Select the [System] Screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1
2
Selecting [SAVE Key Operation] setting options.
Setting options:( default setting)
Select & Save
Pressing the SAVE key displays the setting options for saving.
Quick Save Select the setting options for saving on the [Calc/Save] screen
before pressing the SAVE key.
The Save dialog appears. Select from the saving options in this dialog. Save Waveforms (p. 81) Save Screen Image (p. 82) Save Numerical Calculation Results (p. 83)
Setting items are displayed. Select your saving options.
[Select & Save]
[Quick Save]
6.3 Saving Data
Selecting the Manual Saving Method [Quick Save]/[Select & Save]
Two manual saving methods are available: [Quick Save] and [Select & Save], both of which offer the same setting options. When saving waveforms, a folder (named with the date of saving) is created for data on the CF card, and saved files are auto­matically numbered (p. 71).
81
Quick Save
(Before pressing the SAVE key, make
settings on the System screen)
(Select from the dialog that appears
when you press the SAVE key)
1
For the [Format] setting, select the data saving format.
Binary Select this format to be able to reload the saved data into the instru-
ment, or so that it can be loaded into the Wave Processor program.
Text Select this format to save the data as a text file.
2
(When [Text] format is selected)
From the [Thinning] options, select the amount of data thinning desired.
1/2, 1/5, 1/10, 1/20, 1/50, 1/100, 1/200, 1/500, 1/1000, OFF
1/2: Saves every other data point ( x x x) 1/5: Saves one of every five data points (xxxxxxxxxxxx)
3
From the [Period] setting options, select the portion of the data to save.
Whole *, A-B, Start-A, Start-B, A-End, B-End
(Whole = all acquired waveforms, A = Cursor A, B = Cursor B, Start = start of acquired waveforms, End = end of acquired waveforms)
Select [Save] and press the ENTER key.
Select [Yes] in the confirmation dialog,
and press the ENTER key.
Select [Waveform], and
press the ENTER key.
Select [Waveform] as
the [Save Type].
( default setting)
Select & Save
6.3 Saving Data

Saving Waveform Data (with the SAVE Key)

When saving waveform data, created files are automatically named WAVEnnnn.MEM, RMS (or .CSV) in a folder automatically named [HIOKI8870]­[save date], where nnnn is an automatically generated seria l number starting from 0001 and da To save a partial waveform, specify the time span to save beforehand (p. 68).
te is the saving date (p. 71).
3
6
Chapter 6 Saving & Loading Data
82
(Before pressing the SA VE key, make
settings on the System screen)
(Select from the dialog that appears
when you press the SAVE key)
Display the screen to capture
Select [Yes] in the confirmation
dialog, and press the ENTER key.
Select [Screen Image], and
press the ENTER key.
Select [Screen Image] as
the [Save Type].
Display the screen to capture, and press the SAVE key.
Quick SaveSelect & Save
6.3 Saving Data

Capturing a Screen Image (With the SAVE Key)

Captured screen image files are automatically named SCRnnnn.BMP in a folder named [HIOKI8870]-[save date] (p. 71).
6.3 Saving Data
(Before pressing the SAVE key, make
settings on the System screen)
(Select from the dialog that appears
when you press the SAVE key)
Display the numerical calculation
results to be saved.
Select [Yes] in the confirmation
dialog, and press the ENTER key.
Select [Calc Results], and
press the ENTER key (p. 93).
Select [Calc Results] as
the [Save Type].
By specifying a particular time span for numerical calculation, only the results for that time span are saved.
Quick SaveSelect & Save

Saving Numerical Calculation Results (With the SAVE Key)

Numerical calculation settings are necessary before saving results (p. 93). Numerical calculation result files folder named [HIOKI8870]-[save date] (p. 71).
are automatically named MEASnnnn.CSV in a
83
3
6
Chapter 6 Saving & Loading Data
84
Select the [Setting] Screen.
Select [Save Setting].
Apply
The Save dialog appears.
1 2
3
Select whether to save to the MR8870-20’s memory or to a CF card.
Apply
4
(When [CF Card] is selected)
Setting configuration files are automatically named CONFIGnnnn.SET (at memory re­corder mode) in a folder named [HIOKI8870]-[CONFIG] (p. 71). To reload a setting configuration, see p. 85. In RMS recorder mode, files will be saved with the extension .SRM.
(When [Memory] is selected)
Select [Save] for the con­figuration instance (No.) to save. Enter a comment as need­ed.
Apply
When [Export] is selected, all the setting configurations saved in the internal
memory can be saved as one file in the CF card.
Saved under the file name [LIST + Automatic No. (0001).BDL] inside the
[HIOKI8870]-[CONFIG] folder.
When reading the setting list file (LISTxxxx.BDL) saved in the CF card on this
instrument, all the 10 settings inside the internal memory will be overwritten.
Comments entered on the [Comment] screen are displayed here.
6.3 Saving Data

Saving Setting Configurations

Setting configurations can be saved as data files and later reloaded into the Memory HiCorder when you need to make more measurements with the same settings. Up to ten setting configurations can be saved to internal memory, and more can be store
d on a CF card.
85
Select the [Setting] Screen.
Select [Load Setting].
1 2
Apply
The loading dialog appears.
3
Select whether to load a
setting configuration from
MR8870-20’s memory or
CF card
Apply
When Loading from Memory HiCorder Memory The list of configurations stored in Memory Hi­Corder memory is displayed.
(If you selected
[Memory])
Select [Load] for the con­figuration you want to load
Apply
4

6.4 Loading Data on the Memory HiCorder

6.4 Loading Data on the Memory HiCorder
Previously stored binary waveform data, captured screen images, saved set­ting configurations and setting lis (p. 71). Data stored on a CF card can also be the supplied USB cable (p. 91).
Waveform files (.RMS) and setting files (.SRM) saved in RMS recorder mode cannot be loaded in memory recorder mode. Addi­tionally, waveform files (.MEM) and setting files (.SET) saved in me
mory recorder mode cannot be loaded in RMS recorder mode.
Change the mode as appropriate before loading files.

Loading a Setting Configuration

Setting configurations saved in the Memory HiCorder’s memory or on a CF card can be reloaded. How to save a setting configuration: p. 84
t can be reloaded into the Memory HiCorder
transferred to a computer using
3
6
Chapter 6 Saving & Loading Data
86
Apply
Select the configuration file to load.
(CONFIGnnnn.SET)
Memory HiCorder settings are reconfigured to those in the loaded configuration file.
(If you selected
[CF Card])
When Loading from CF card
Configuration files stored in the [HIOKI8870]­[CONFIG] folder on the CF card are displayed.
1
2
Method 1. Loading from instrument memory
Open the screen for loading the setting configuration. (Up to step 3 on the previous page)
Specify the number of the setting configuration you want to load automatically and input it into
[Auto Load].
Method 2. Loading from external storage media
Setting options:( default setting)
Off
Loading is not performed automatically.
[No1] to [No10]
Select the file to load automatically.
• Setting configuration data saved as a file named STARTUP.SET or STAR­TUP.SRM in the [HIOKI8870] - [CONFIG] folder of the CF card is automatically loaded at power-on.STARTUP.SET will start up in memory recorder mode, STAR­TUP.SRM will start up in RMS recorder mode.
• In both cases, you can read in favor of the measurement mode during start-up.
6.4 Loading Data on the Memory HiCorder

Load the setting configuration data automatically

Select the File Screen.
The contents of the CF card are displayed.
1
Apply
The control dialog box appears.
Select a file to load.
Displays the parent folder.
Displays the child folder. After selecting a folder by pressing the ENTER key, select [Change folder] in the displayed control dialog box.
2
The selected file content is displayed on the waveform screen.
Select [Load].
Apply
3
When binary format waveform data and screen images have been automatically saved together, you can switch the display between the two.
Load the screen image (WAVEnnnn.BMP)
Loads the waveform data cor­responding to the screen im­age file with the same name.
The loading dialog appears Select [Yes] and press the
ENTER key to display the waveform data.
Switch to the screen image files in the same folder.
87
6.4 Loading Data on the Memory HiCorder

Loading Waveform Data and Screen Images

Waveform data saved in binary format, or captured screen images, can be reloaded into the Memory HiCorder.
3
6
Chapter 6 Saving & Loading Data
88
Select the File Screen.
1
2
Opens the control dialog box.
Select the File Screen.
The contents of the card are displayed.
1
Displays the parent folder.
View child folder (view contents of the currently highlighted folder)
(press the ENTER key and select [Folder] in the displayed control dialog box)
2
Select a folder to view.

6.5 Data Management

6.5 Data Management
You can manage data stored on a CF card in the Memory HiCorder.
• Load a file (when the file is selected) (p. 85)
• Move displayed folders (when the folder is selected) (p. 88)
• Delete data (p. 89)
• Rename files and folders (p . 89)
• Sort files (p. 90)
• Format a CF card (p. 76)
The following operations can be performe d from the co pressing the ENTER key from the File screen.

Viewing Folder Contents and the Parent Folder

ntrol dialog box displayed by
89
Select the File Screen.
The contents of the CF card are displayed.
1
Apply
Select a file to delete.
2
A confirmation dialog appears. Select [Yes] and press the ENTER key to delete.
3
Select [Delete].
Apply
Select the File Screen.
The contents of the CF card are displayed.
1
Apply
Select the file or folder to rename.
2
The character entry dialog appears. Enter the new file name (the same way as when entering a comment) (p. 52) A confirmation dialog appears. Select
[Yes] and press the ENTER key to apply the new name.
3
Select [Rename].
Apply
6.5 Data Management

Deleting Data

Folder and files on the CF card can be deleted.

Renaming Files and Folders

Folders and files on a CF card can be renamed. File names may consist of up to 26 regu
lar characters.
3
6
Chapter 6 Saving & Loading Data
90
Select the File Screen.
1
Apply
Select the key on which to sort.
4
3
Select [Sort].
Apply
2
Apply
The control dialog box appears.
Select the desired sort order.
(Ascending or Descending) Files are displayed in the new sort order
5
6
Finish
6.5 Data Management

Sorting Files

Files can be sorted in ascending or descending order according to a selected so rt key.
91
Open the protective cover over the USB connector.
Orient the USB cable plug properly, and insert it into the recep­tacle.
1
2
Protective Cover
Connect the other end of the cable to a USB port on the computer.
3
The computer should recognize the Memory HiCorder as a remo vable disk when the cable is connected. Only data in the CF card inserted in this Memory HiC order can be accessed from the computer. Data not saved in the CF card cannot be accessed.
Connection Precautions
Connecting the cable while measuring or with the File screen displayed will pre­vent the computer from recognizing the Memory HiCorder. Also, the Memory HiCorder cannot be operated during the connection process.
Removable Disk

6.6 Transferring Data to a Computer

6.6 Transferring Data to a Computer
Using the supplied USB cable, data stored on a CF card in the Memory HiCorder can be transferred to a computer. To analyze the data using an appli­cation program, refer to the instructions for
See: "Appendix 3 Wave Processor Application Program" (p. A6)
Computer Requirements
A personal computer running Windows XP, Vista, 7, or 8
• Do not eject the CF card or pull out the USB cable during data transfer. Doing so would prevent proper data transfer.
• The Memory HiCorder and computer should be connected to the
me earth ground. If grounded separately, potential difference
sa between the ground points can cause malfunctions or damage when connecting the USB cable.
the program.
3
Connecting the USB Cable
USB Disconnection Procedure
Before unplugging the USB cable connected to the Memory HiCorder, select the “Safely Remove Hardware” icon in the Windows notification area, and click the mass storage device corresponding to the Memory HiCorder to remove.
6
Chapter 6 Saving & Loading Data
92
6.6 Transferring Data to a Computer
Loading...