19.4.5 System Conguration Check ................... 429
19.5 Cleaning the instrument ............ 431
19.6 Disposing of the Instrument
(Removing Lithium Battery) ...... 433
Appendix Appx.1
Appx. 1 Default Values for Major
Settings ............................. Appx.1
Appx. 2 For Reference ................... Appx.2
Appx. 3 About Options ................ Appx.13
Appx.4 FFTDenitions ............... Appx.20
16
17
18
19
Appx.Ind.
v
Contents
vi
Usage Index
Basic measurement procedure
1 Installing the instrument
(p. 25)
Installing the instrument
Installing modules
Usage Index
Performing measurement in the automatic
range setting (
Monitoring changes in input signals (p. 201)
Manually triggering the instrument (p. 216)
Entering comments (p. 156)
p. 76
)
1
2
Connecting cables
Loading the recording paper
Turning on the instrument
2 Setting the instrument
(p. 51)
Selecting a function
Selecting measurement settings
Selecting input channels
3 Measuring input signals (p. 73)
Starting measuring input signals
Freely setting the waveform display (p. 64)
Converting input values (p. 167)
Copying settings to other channels (p. 179)
Eliminating noise (Low-pass lter) (p. 70)
Plotting X-Y composite curves (p. 144)
Locking the operation keys (p. 17)
Formatting a CF Card (p. 43)
Scaling measured values obtained with current
clamp sensors (p. 169)
3
4
5
6
7
Completing the measurement
4 Analyzing (
Saving/printing results
5 Completing the measurement (p. 46)
Turning off the instrument
p. 133
Performing analysis
(optionally)
), saving (
p. 85
), and printing data (
p. 119
8
)
9
10
Appx.Ind.
1
Introduction
Introduction
Thank you for purchasing the Hioki MR8847A Memory HiCorder (MR8847-51, MR8847-52, MR8847-53).
To obtain maximum performance from the instrument, please read this manual rst, and keep it handy for
future reference.
The optional clamps (p. Appx.13) collectively mean “clamp sensors.”
The following instruction manuals are available for this instrument. Refer to the relevant manual as usage.
Instruction ManualDescription
1Measurement Guide
(booklet)
2
(This
document)
3Communication
4U8793, MR8790,
Instruction Manual
(booklet)
Command
Instruction Manual
(PDF)
MR8791 Instruction
Manual
(PDF)
Read this booklet rst.
Contains basic operating procedures for those who use this instrument
for the rst time.
Contains details and specifications regarding the functions and
operations of this instrument.
Contains a list of the communication commands and their explanations
to control the instrument with a computer.
Contains specications and explanations of functions/operations of
Models U8793 Arbitrary Waveform Generator Unit, MR8790 Waveform
Generator Unit, MR8791 Pulse Generator Unit, and SF8000 Waveform
Maker.
Trademarks
• Microsoft Windows, Excel and Internet Explorer are either registered trademarks or trademarks of
Microsoft Corporation in the United States and other countries.
• CompactFlash is a registered trademark of SanDisk Corporation (USA).
• Sun, Sun Microsystems, Java, and any logos containing Sun or Java are trademarks or registered
trademarks of Oracle Corporation in the United States and other countries.
2
Verifying Package Contents
Verifying Package Contents
When you receive your instrument, inspect it carefully to ensure that no damage occurred during shipping.
In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to
operate according to the specications, contact your authorized Hioki distributor or reseller.
Store the packaging in which the instrument was delivered, as you will need it when transporting the
instrument.
1
Instrument and accessories
Conrm that you received the following items: (One each)
Other options as specied in your order “Appx. 3.1 Options” (p. Appx.13)
*1: When one or more pieces of Model 8967 Temp Unit are installed in the instrument, two ferrite clamp-
on chokes (small) are supplied per module.
*2: The latest version can be downloaded from our website.
7
8
9
10
Appx.Ind.
3
Safety Information
Safety Information
This instrument and modules are designed to conform to IEC 61010 Safety Standards, and has been
thoroughly tested for safety prior to shipment. However, using the instrument in a way not described in
this manual may negate the provided safety features.
Before using the instrument, be certain to carefully read the following safety notes:
DANGER
Mishandling during use could result in injury or death, as well as damage to the
instrument. Be certain that you understand the instructions and precautions in
the manual before use.
WARNING
With regard to the electricity supply, there are risks of electric shock, heat
generation, re, and arc discharge due to short circuits. Individuals using an
electrical measuring instrument for the rst time should be supervised by a
technician who has experience in electrical measurement.
Protective Gear
WARNING
This instrument measures live lines. To prevent electric shock, use appropriate
protective insulation and adhere to applicable laws and regulations.
Notation
In this document, the risk seriousness and the hazard levels are classied as follows.
DANGER
WARNING
CAUTION
IMPORTANT
Indicates an imminently hazardous situation that will result in death or serious injury
to the operator.
Indicates a potentially hazardous situation that may result in death or serious injury to
the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury
to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance tasks
with which the operators must be fully familiar.
Indicates a high voltage hazard.
If a particular safety check is not performed or the instrument is mishandled, this may
give rise to a hazardous situation; the operator may receive an electric shock, may
get burnt or may even be fatally injured.
Indicates prohibited actions.
Indicates the action which must be performed.
*
Additional information is presented below.
4
Symbols Afxed to the Instrument
Indicates cautions and hazards. When the symbol is printed on the instrument, refer to the
corresponding topic in the Instruction Manual.
Safety Information
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Indicates a fuse.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates a burn hazard if touched directly.
Standards Symbols
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU
member states.
Indicates that the product conforms to regulations set out by the EU Directive.
1
2
3
4
5
6
Other Symbols
This manual uses the following symbols to indicate specic information for operating the instrument.
(p. )Indicates the location of reference information.
CURSOR
(Bold-faced)
[ ]
Names of settings, buttons, and other screen elements are written in bold blue text.
Unless otherwise specied, “Windows” represents Windows Vista, Windows 7, and Windows 8.
IE is an acronym for Internet Explorer.
Menus, commands, dialogs, buttons in a dialog, and other names on the screen and keys
are indicated in brackets.
Indicates that the memory function supports the function.
Indicates that the recorder function supports the function.
Indicates that the X-Y recorder function supports the function.
Indicates that the FFT recorder function supports the function.
Click: Press and quickly release the left button of the mouse.
Right-click: Press and quickly release the right button of the mouse.
Double-click: Quickly click the left button of the mouse twice.
7
8
9
10
Appx.Ind.
5
Safety Information
Accuracy
We dene measurement tolerances in terms of f.s. (full scale), rdg. (reading), and setting values
with the following meanings:
f.s. (maximum display value
or scale length)
rdg. (reading or displayed
value)
SettingIndicates the value set as the output voltage, current, or other quantity.
The maximum displayable value or scale length.
For this instrument, the maximum displayable value equals the numerical
number of a presently set range (unit: V/div) multiplied by the number of divisions (20) on the vertical axis.
Example: When the range is set to 1 V/div, f.s. = 20 V
The value currently being measured and indicated on the measuring
instrument.
Measurement Categories
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards
for various electrical environments, categorized as CAT II to CAT IV, and called measurement
categories.
DANGER
• Using a measuring instrument in an environment designated with a higher-
numbered category than that for which the instrument is rated could result in a
severe accident, and must be carefully avoided.
• Never use a measuring instrument that lacks category labeling in a CAT II to
CAT IV measurement environment. Doing so could result in a serious accident.
CAT II: When directly measuring the electrical outlet receptacles of the primary electrical
circuits in equipment connected to an AC electrical outlet with a power cord (portable
tools, household appliances, etc.)
CAT III: When measuring the primary electrical circuits of heavy equipment (xed installations)
connected directly to the distribution panel, and feeders from the distribution panel to
outlets
CAT IV: When measuring the circuit from the service drop to the service entrance, and to the
power meter and primary overcurrent protection device (distribution panel)
Distribution Panel
Service Entrance
Service Drop
CAT IV
Power Meter
Fixed Installation
Internal Wiring
CAT III
CAT II
T
Outlet
The applicable measurement category is determined based on the module being used.
Refer to “18.6 Specications of Modules” (p. 387).
6
Operation Precautions
Before Use
Operation Precautions
Follow these precautions to ensure safe operation and to obtain the full benets of the various
functions.
DANGER
If the connection cables or the instrument are damaged, there is a risk of an
electric shock. Perform the following inspection before using the instrument:
• Before using the instrument, check that the coatings of the connection cables
are neither ripped nor torn and that no metal parts are exposed. Using the
instrument under such conditions could result in an electric shock. Replace the
connection cables with those specied by our company.
• Verify that it operates normally to ensure that no damage occurred during
storage or shipping. If you nd any damage, contact your authorized Hioki
distributor or reseller.
Installing the instrument and modules
WARNING
Installing the instrument and modules in inappropriate locations may cause a
malfunction of the instrument or may give rise to an accident. Avoid the following
locations:
• Exposed to direct sunlight or high temperatures
• Exposed to corrosive or combustible gases
• Exposed to a strong electromagnetic eld or electrostatic charge
• Near induction heating systems (such as high-frequency induction heating
systems and IH cooking equipment)
• Susceptible to vibration
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
1
2
3
4
5
6
7
8
CAUTION
Do not place the instrument on an unstable table or an inclined place. Dropping
or knocking down the instrument can cause injury or damage to the instrument.
Installing the instrument
To prevent overheating, be sure to leave the specied clearances around the instrument.
• The instrument should be operated only with the bottom or rear side downwards.
• Vents must not be obstructed.
• Do not install the instrument at an angle.
Left side
Vents
At least 5 cm
on all sides
9
10
Appx.Ind.
7
Operation Precautions
Handling the Instrument and Modules
• Do not use the modules or the cables with circuits that exceed those ratings or
specications.
Doing so may damage the instrument or cause it to become hot, resulting in
bodily injury.
Even including any devices, such as an attenuator, in the input terminal will
never increase the maximum rated voltage to earth. Take care of the connection
not to allow any input voltage to exceed the maximum rated voltage to earth.
• To avoid an electric shock, do not remove the instrument’s cover and the
modules’ cases.
The internal components of the instrument carry high voltages and may
become very hot during operation.
• To avoid an electric shock, before removing or replacing an input module,
conrm that the instrument is turned off and that the connection cords are
disconnected.
• To avoid the danger of an electric shock, never operate the instrument with an
input module removed. To use the instrument with a module removed, install a
blank panel over the opening of the removed module.
• To prevent the instrument damage or an electric shock, use only the screws
that are originally installed for securing the module in place.
If you have lost any screws or nd that any screws are damaged, please contact
your Hioki distributor for a replacement.
DANGER
WARNING
CAUTION
• To avoid damage to the instrument, protect it from physical shock when
transporting and handling it. Be especially careful to avoid physical shock due
to dropping it.
• The mounting screws must be rmly tightened or the module may not perform
to specications, or may even fail.
• To avoid damaging modules, do not touch the connectors, installed in the
instrument, to which the modules are connected.
• Before carrying the instrument, disconnect all cables and remove the CF card, USB
ash drive, and the recording paper.
• Displayed waveforms can frequently uctuate due to induction potential even when no
voltage is applied. This, however, is not a malfunction.
• This instrument may cause interference if used in residential areas. Such use must be
avoided unless the user takes special measures to reduce electromagnetic emissions to
prevent interference to the reception of radio and television broadcasts.
Handling the printer and the recording paper
WARNING
The print head and surrounding metal parts can become hot. Be careful to avoid
touching these parts.
8
Operation Precautions
CAUTION
Be careful not to cut yourself with the paper cutter.
• Please use only the specied recording paper. Using non-specied paper may not only
result in faulty printing, but printing may become impossible.
• If the recording paper is skewed on the roller, paper jams may result.
• Always use the paper cutter to cut the printed paper. Excessive paper dust can
accumulate on the roller if the paper is cut with the print head, which may result in paper
jams or white streaks in the printing.
1
2
Storing data recordings
The recording paper is thermally sensitive. Observe the following precautions to avoid paper discoloration
and fading.
• To avoid paper discoloration, do not expose it to direct sunlight. Store the paper at no more than 40°C
and 90% RH.
• Store the paper away from dew and damp places.
• Make photocopies of recording printouts that are to be handled or stored for legal purposes.
• If the thermal paper is exposed to an organic solvent such as alcohol or ketone, it may no longer
develop properly, and recorded data may fade. Keep the printer papers away from exible PVC lms
and pressure sensitive tapes including scotch tapes because they contain organic solvents.
• Also, the thermal recording paper is ruined by contact with wet diazo copy paper.
Avoid exposure to direct
sunlight.
Do not store the paper at
more than
40°C (104°F) and 90%
RH.
Avoid exposure to volatile
organic solvents like
alcohol, ethers, and
ketones.
Avoid contact with exible
PVC lms or adhesive
tapes such as scotch
tapes.
3
4
5
6
7
Avoid stacking with wet
Diazo copy paper.
Storing recording paper
• Store thermal paper where its temperature will not exceed 40°C.
• The paper will deteriorate if exposed to light for a long time; thus, do not remove the wrapping paper
from the roll until it is ready to be used.
8
9
10
Appx.Ind.
9
Operation Precautions
Handling storage devices
CAUTION
• Do not remove the storage device while it is being accessed by the instrument (
SAVE
key is lit in blue). Data saved on the device could be lost.
• Do not turn off the instrument while it is accessing the storage device (
key is lit up in blue). Data saved on the device could be lost.
• Do not carry the instrument with a USB ash drive left connected. Damage could result.
• Exercise care when using such products because static electricity could damage the
storage device or cause a malfunction of the instrument.
•
Do not subject the SSD to extreme shock or vibration. Shock can cause it to be
damaged.
IMPORTANT
• No compensation is available for loss of data stored on the built-in drive (SSD) or a removable
storage device, regardless of the content or cause of damage or loss. Be sure to back up any
important data saved on the built-in drive (SSD) or the removable storage device.
• Use only CF cards sold by Hioki. (No adapter is required to insert a CF card into the
instrument.)
• Compatibility and performance are not guaranteed for PC cards made by other manufacturers.
You may be unable to read from or save data to such cards.
■Hioki optional CF cards (with an adapter accompanying)
Model 9728 PC Card 512M, Model 9729 PC Card 1G, Model 9830 PC Card 2G
• With some external storage device, the instrument may not start up if it is turned on while the
external storage device is inserted. In such a case, turn on the instrument rst, and then insert
the external media. Prior testing is recommended.
• The instrument does not support particular kind of USB ash drives, such as those that require
ngerprint authentication or a password.
• When saving or loading data, insert the storage device before selecting data to be saved.
When the storage device is not inserted, no devices are not displayed in the le list.
• All storage devices (built-in drive [SSD], USB ash drive, and CF card) have a limited service
life. After extensive use for a long period, saving and loading data may be disabled. In that
case, replace the device with a new one.
• The built-in drive (SSD) is a consumable part. When the saved data reaches the capacity (about
60 TB), no further data can be recorded. In such a case, the SSD should be replaced with a
new one.
• When the instrument is left powered off for a one year or more, the data saved on the built-in
drive (SSD) may be lost. Be sure to back up the data if the instrument is left powered off for a
long time.
• Devices the automatic data saving supports are the built-in drive (SSD), a USB ash drive, and
a CF card.
• Data can also be automatically saved on a USB ash drive; however, we recommend using
Hioki optional CF card instead for data protection.
while the
while the
SAVE
10
Before connecting cables
Operation Precautions
DANGER
When measuring power line voltage
• Connect the connecting cables to only the secondary side of a breaker. Even
if a short-circuit occurs on the secondary side of the breaker, the breaker will
interrupt a short-circuit current. Do not connect them to the primary side of the
breaker because an unrestricted current ow could damage the instrument and
facilities if a short circuit occurs.
• To prevent an electrical shock and a bodily injury, do not touch any input
terminals on the VT (PT), CT or the instrument when they are in operation.
• Do not leave the measurement cables connected to the instrument in an
environment where voltage surges exceeding the maximum input voltage may
occur. Subjecting the instrument to such a voltage may result in damage to the
instrument or a serious accident.
• Do not short-circuit two wires to be measured by bringing the connection
cables into contact with them. Arcs or such grave accidents are likely to occur.
• To avoid a short-circuit or an electric shock, do not touch the metal parts of the
connecting cable clips.
• To avoid electrical shock, be careful to avoid shorting live lines with the
connection cable chips.
WARNING
To avoid an electric shock and a short-circuit accident, use only the specied test
leads to connect the instrument input terminals to the circuit to be tested.
• To avoid an electric shock, do not exceed the lower of the ratings shown on the
instrument and connection cords.
1
2
3
4
5
6
To prevent an electric shock, conrm that the white or red portion (insulation
layer) inside the cable is not exposed. If a color inside the cable is exposed, do
not use the cable.
CAUTION
• The cable is hardened in the freezing temperatures. Do not bend or pull it to avoid
tearing its shield or cutting cable.
• Connecting cables to the BNC jacks on modules
Do not use any cable terminated with a metal BNC connector. If you connect a metal
BNC cable to an insulated BNC connector, the insulated BNC connector and the
instrument may be damaged.
To prevent cable damage, do not step on cables or pinch them between other objects.
Do not bend or pull on cables at their base.
IMPORTANT
• Use only the specied connection cables. Use of any cable not specied by our company does
not allow safe measurements due to poor connection or other reasons.
• For detailed precautions and instructions regarding connections, refer to the instruction
manuals for your modules, connection cables, etc.
7
8
9
10
Appx.Ind.
11
Operation Precautions
Measurement
Functional Earth
Before connecting a logic probe to the measurement object
DANGER
To avoid an electric shock, a short-circuit, and damage to the instrument, observe
the following precautions:
• The ground pin in the logic connector (plug) of Model 9320-01 Logic Probe and
Model 9327 Logic Probe are not isolated from the instrument’s ground (common
ground).
Supply power to the instrument with the provided power cord and measurement
objects from a single mains circuit.
Connecting the instrument and a measurement object to different mains circuits
from one another or using a non-grounding power cord may cause damage to
the measurement object or the instrument because of current owing through
the logic probes resulting from the potential difference between the grounds of
the different wiring systems.
To avoid that, we recommend the following connection procedure:
Connect the provided power cord
to the instrument and supply
power from the same outlet as the
measurement object.
Object
Logic probe
Memory
HiCorder
Connect the measurement object’s
ground to the GND terminal
(functional earth terminal) of the
instrument.
(Always supply power from the
single mains circuit.)
Refer to
Wire to the GND Terminal (Functional
Earth Terminal)” (p. 45).
“2.5.2 Connecting an Earthing
Before turning on the instrument
• To avoid electrical accidents and to maintain the safety specications of this
instrument, connect the power cord provided only to an outlet.
• Before turning the instrument on, make sure the supply voltage matches that
indicated on its power connector. Connection to an improper supply voltage
may damage the instrument and present an electrical hazard.
Avoid using an uninterruptible power supply (UPS), DC/AC inverter with rectangular-
wave or pseudo-sine-wave output to power the instrument. Doing so may damage the
instrument.
Measurement
Object
GND
WARNING
CAUTION
Logic probe
Memory
HiCorder
Terminals
12
Before connecting the instrument to an external device
DANGER
Operation Precautions
To avoid electrical hazards and damage to the instrument, do not apply voltage
exceeding the rated maximum to the external control terminals.
I/O terminalsMaximum input voltage
InstrumentSTART/EXT.IN1−0.5 V to 7 V DC
STOP/EXT.IN2−0.5 V to 7 V DC
PRINT/EXT.IN3−0.5 V to 7 V DC
GO/EXT.OUT150 V DC, 50 mA DC, 200 mW
NG/EXT.OUT250 V DC, 50 mA DC, 200 mW
EXT.SMPL−0.5 V to 7 V DC
TRIG OUT50 V DC, 50 mA DC, 200 mW
EXT.TRIG−0.5 V to 7 V DC
U8793
Arbitrary Waveform
Generator Unit
IN−0.5 V to 7 V DC
OUT30 V DC, 50 mA DC
WARNING
To avoid an electric shock or damage to the equipment, always observe the
following precautions when connecting the cables to external control terminals.
• Always turn off the instrument and any devices to be connected before making
connections.
• Be careful to avoid exceeding the ratings of the external control terminals and
the external connectors.
• Ensure that devices and systems to be connected to the external control
terminals are properly isolated from one another.
1
2
3
4
5
6
CAUTION
To avoid equipment failure, do not disconnect the USB cable while communications are
in progress.
• Use a common ground for both the instrument and the connection equipment. Using
different ground circuits will result in a potential difference between the instrument’s
ground and the connected equipment’s ground. If the communications cable is
connected while such a potential difference exists, it may result in equipment
malfunction or failure.
• Before connecting or disconnecting any communication cable, always turn off the
instrument and the device to be connected. Failure to do so may result in equipment
malfunction or damage.
• After connecting the communications cable, tighten the screws on the connector
securely. Failure to secure the connector could result in equipment malfunction or
damage.
7
8
9
10
Appx.Ind.
13
Operation Precautions
CD precautions
• Exercise care to keep the recorded side of discs free of dirt and scratches. When writing text on a
disc’s label, use a pen or marker with a soft tip.
• Keep discs inside a protective case and do not expose to direct sunlight, high temperatures, or
high humidity.
• Hioki is not liable for any issues your computer system experiences in the course of using this
disc.
When the instrument is not used for a long period
• To avoid straining some parts of the printer, and to prevent dirt adhering to the print head, close
the printer cover.
• Perform test prints (printer check) three or four times before using the printer that has been in
storage and has left unused for a long period.
Precautions during shipment
Store the packaging in which the instrument was delivered, as you will need it when transporting
the instrument.
14
1
Overview
1.1 Product Overview
This instrument enables you to measure and analyze various waveforms with simple methods.
You can use this instrument mainly for facility diagnosis, preventive maintenance, and troubleshooting.
Sturdy body with easy-to-grasp
handle installed
You can install this portable instrument anywhere.
1
Overview
Logic modules can measure
signals input on 64 channels
Easy loading of recording paper
High-speed printing
High-speed sampling
at 20 MS/s
Arbitrary Waveform Generator Unit
can output waveforms simulating
measured signals
You can take multiple measurements simultaneously.
You can load the recording paper through one-touch
operation.
You can conduct reliable response evaluation.
You can have the instrument output realistically simulated
waveforms.
15
Part Names and Functions
1.2 Part Names and Functions
Left side
Vents
Right side
1
2
3
Printer
(p. 9)
6
7
Front side
Handle
Screen
USB connector (Type B)
1
Connect the USB cable to
operate the instrument with a
computer.
(p. 352)
USB connector (Type A)
2
Connect a USB ash drive or
a mouse. (p. 41)
CF card slot
Operation keys (p. 17)
External control terminals
6
Input an external sampling
signal.
(p. 361)
Connect signal cables to
operate the instrument
externally.
Standard LOGIC
7
terminals
Connect optional Hioki logic
probes.
(p. 28)
4
5
16
8
9
10
100BASE-TX connector
3
Connect a LAN cable.
(p. 331)
Power switch (p. 46)
4
Flip the switch to turn on and
off the instrument.
: Power-on
: Power-off
GND terminal (Functional
5
earth terminal)
Connect a grounded
conductor.
(p. 45)
Various modules
8
(p. 26), (p. 28)
For details, refer to “8.10
Setting Details of Modules” (p.
180) or “18.6 Specications
of Modules” (p. 387).
Power inlet
9
Connect the provided power
cord. (p. 45)
Serial number
10
The serial number consists
of 9 digits. The rst two (from
the left) indicate the year of
manufacture, and the next
two indicate the month of
manufacture.
Required for production
control. Do not peel off the
label.
Operation keys
Part Names and Functions
87
1
9
10
2
3
11
12
13
4
14
5
6
15
16
DISP key
1
Displays the waveform
screen.
CH.SET key
2
Displays the channel
settings window on the
waveform screen (p. 64).
TRIG.SET key
3
Displays the trigger settings
window on the waveform
screen (p. 201).
F key
4
Selects setting items.
ESC key
5
Cancels the last action.
Closes the displayed dialog
and window.
KEY LOCK:
Press and hold the ESC
key for 3 seconds to
engage the key lock
function, which prevents
accidental operation.
Press and hold this key for
3 seconds to disengage
the key lock function.
STATUS key
8
Displays the status screen.
CHAN key
9
Displays the channel
screen.
FILE key
10
Displays the File screen.
(p. 108)
AB CSR key
11
(Lights up in red when
selected.)
Sets Cursors A and B. (p.
134)
WAVE key
12
(Lights up in red when
selected.)
Assigns waveform scrolling
to the jog dial and shuttle
ring. (p. 141)
Inner: Jog dial
13
Outer: Shuttle ring
Scrolls waveforms display.
(p. 141)
Increases and decreases a
setting value.
(p. 21)
1
Overview
STOP key
6
Stops the measurement in
progress.
Press the key once:
Stops the measurement
in progress after the
instrument records the
specied recording length
of waveforms.
Press the key twice:
Immediately stops the
measurement in progress.
(p. 328)
SYSTEM key
7
Displays the system
screen.
(p. 325)
CURSOR key
14
Moves the cursor up, down,
left, and right on the screen.
Manual trigger key
15
Manually trigger the
instrument.
(p. 216)
START key
16
Starts measurement.
(Lights up in green during
measurement.)
(p. 328)
17
Part Names and Functions
17181920212223
PRINT key
17
Prints waveforms and lists. (p. 119)
COPY key
18
Prints a screenshot. (p. 130)
FEED key
19
Feeds paper.
SAVE key (Lights up in blue while the instrument is
20
accessing a storage device.)
Saves data to a storage device. (p. 85)
The dialog box can be switched between visible and
invisible during auto-saving.
HELP key
21
Displays help information. (p. 22)
AUTO key
22
Starts measurement in the auto-range setting.
(p. 76)
TIME/DIV key
23
Sets the timebase.
18
Screens Conguration
1.3 Screens Conguration
The screens are congured as listed below. Pressing each of the keys listed below displays a
corresponding screen or window.
The waveform screen can display the trigger settings window, and the channel settings window.
Waveform screen
The display used to observe waveforms.
Congure measurement conditions using the settings window on the right.
Trigger settings window, channel settings window
The display used to congure the trigger settings
The display used to congure the settings of analog channels and logic channels
1
Overview
Status screen
The window used to congure the measurement methods and numerical calculation
settings.
Pressing the STATUS key switches the sheets to be displayed in the following order:
The screen used to view saved data les in storage devices (a CompactFlash card, the
built-in drive, a USB ash drive, the internal memory).
19
Screens Conguration
Explanation of screen contents
Waveform screen
Title comment
Shows a previously
entered title
comment.
(p. 156)
Trigger marker
Indicates the point when
the instrument triggered.
(p. 201)
Upper and lower limits
Shows upper and lower
limit values for each
channel. (p. 152)
Trigger time
Shows the date
and time when
the instrument
triggered.
(p. 201)
Storage counter
Shows the number of times the
instrument triggered. (p. 74)
Logic waveform (p. 70)
Analog waveform (p. 67)
Vertical axis display
Shows a value per division for each
channel linked to the range settings
of the vertical axis (voltage axis).
(p. 67)
Storage device icon
Displays the status of
storage devices.
(p. 41)
Current date and time
Shows the current date
and time in the manner
previously congured.
(p. 47)
Settings cursor
The present
cursor position
ashes.
Settings window
The window
used to congure
measurement
conditions.
(p. 54)
Scroll bar
The red bar indicates the waveform range
written in the memory. The blue frame
indicates the displayed waveform range.
(p. 141)
Items common to the Status, Channel, System, and File screens
Sheet tabs
Shows names of sheets
that can be selected.
Pressing each of the
MENU keys switches a
sheet to another.
Hint
Shows details about the item at the present settings cursor
position.
Messages such as “Online,” “Key Lock active.,” and error
messages also appear here.
Next Page button
Appears when more
than ve setting
items are available.
Pressing this button
switches other
groups of items to be
displayed.
20
1.4 Basic Key Operation
Press the CURSOR key and move the cursor to an item to be changed.
1
Cursor
GUI
Basic Key Operation
1
Overview
2
Check the illustrations on the GUI and press the function key (F key) to change the
settings.
The function assigned to the F key varies depending on the setting items.
To select an item to be set
Press the F key to change settings.
When there are more than six setting items, press the
F5 [Next Page] key to switch to the next page.
To increase and decrease a setting value
Increases a
numerical value at an
accelerated rate.
Decreases a numerical
value at an accelerated
rate.
Press the F key to change the setting value.
(Turning the jog dial or shuttle ring enables you
to change values.)
3
For some settings, press the CH.SET key to select [Exec], and press the TRIG.SET key
to select [Cancel].
To enter characters and numbers
See “8.1.3 Entering Alphanumeric Characters” (p. 159).
21
Basic Key Operation
1.4.1 Using the HELP Key
Pressing the HELP key displays a simple explanation of the item at the cursor position. You can also
search the help messages for the information for which you are looking.
Cursor position help
Move the cursor to the item for which you want to display a help message.
1
Press the HELP key. The [Cursor Pos Help] sheet that contains the help message of
2
the cursor position appears.
Pressing the CURSOR up and down keys or turning the Jog dial scrolls the information.
Cursor
• Pressing the CH.SET key switches the display mode of the [Cursor Pos Help] sheet between the
following modes: The full-screen display, the upper-half display, and the lower-half display. The gure
above illustrates the upper-half display mode.
• Pressing the HELP key closes the [Cursor Pos Help] sheet.
22
Basic Key Operation
1.4.2 Using Mouse to Enable Key Operation
Using a commercially available USB mouse enables you to operate the instrument in the manner similar
to that with the keys on the instrument.
• The instrument may not support some types of mouses.
• Do not connect any type of device other than a mouse or a USB ash drive to the USB connector of
the instrument.
• Operating the instrument with a mouse may cause temporary operating delay or abnormal screen
display.
• While operating the instrument with a mouse, do not change the interface on the system screen
to anything other than “LAN.” Operating the instrument with a mouse is disabled while USB
communications are in progress.
• External noise may cause the instrument to malfunction while it is operated with a mouse. Keep the
mouse and mouse cable as far away as possible from sources of noise.
The gure below illustrates the basic operation of the instrument with a mouse.
Clicking the right button (rightclick)
Displays a menu with a list of
screens.
Dimmed screen items are
unavailable.
1
Overview
Clicking the left button (click)
Selects a menu or executes the
selected menu. During measurement
with the memory division engaged,
you can change blocks to be
displayed by double-clicking the left
button.
Click the current path shown on the
File screen to move to the upper
folder in the folder hierarchy.
Clicking the center wheel
Changes an item to be
selected.
On the File screen, clicking
the center wheel changes
a le to be selected. During
measurement with the
memory division engaged,
you can change blocks to be
displayed.
Move the mouse forward/
backward/leftward/
rightward
Moves the mouse cursor
on the screen.
23
Basic Key Operation
The operation keys of the instrument and the shortcut menu relate to each other as follows.
To operate the functions assigned to the CH.SET, WAV E, and AB CSR keys and to congure those
settings, click the icons displayed while a mouse is connected to the instrument.
IconOperation key
CH.SET key
WAVE key
WAVE key
Useful functionality
Dragging the right button of the mouse (hold down the button, moving the mouse rightward, leftward, or
forward, and then release the button) performs the same function as when pressing the following keys:
Rightward: START
Leftward: STOP
Upward: ESC
24
2
Procedure
Preparing for Measurement
Install the instrument.
1
Install or remove modules.
2
(When adding or replacing modules)
Connect logic probes to the LOGIC
3
terminals.
(When measuring logic signals)
Connect connection cables to the
4
modules.
(When measuring analog signals)
Probes and cables differ depending on the type of
measurement to be performed.
Insert a storage device (CF card, USB
5
ash drive).
Load a roll of recording paper.
6
Connect the power cord.
7
Connect an earthing wire to the GND
8
terminal (functional earth terminal).
(When performing measurement in noisy
environments)
Turn on the instrument.
9
Set the clock
10
Perform zero-adjustment
11
Perform calibration
(For the instrument with Model MR8990
installed)
(p. 7)
(p. 26)
(p. 28)
(p. 28)
(p. 41)
(p. 43)
(p. 45)
(p. 45)
(p. 46)
(p. 47)
(p. 48)
(p. 49)
2
Preparing for Measurement
After preparation terminates, start
measurement. (p. 51)
To operate the instrument with a computer
Refer to “16 Connecting the Instrument to a Computer” (p. 331).
To control the instrument externally
Refer to “17 Controlling the Instrument Externally” (p. 361).
25
Installing and Removing Modules
2.1 Installing and Removing Modules
Read “Handling the Instrument and Modules” (p. 8) carefully.
Modules ordered with the instrument has already been installed in the instrument. Follow the procedures
below to add, replace, or remove modules from the instrument.
• Up to three logic units can be installed. The instrument ignores the fourth logic module and
later modules that are installed in the instrument.
• For information on the analog channel resolution when logic channels are used, refer to “8.10
Setting Details of Modules” (p. 180).
Installing a module
Right side
Removing a module
Right side
Knob
(Example: Model 8966)
Knob
Required items: Phillips-head screwdriver (No. 2)
Turn off the instrument.
1
Orient the module and insert it all the way into the instrument.
2
Make certain that the module is installed in such a way that the
characters printed on the module’s panel are right side up about
those printed on the instrument.
Tighten the two module mounting screws with a Phillips
3
screwdriver.
Required items: Phillips-head screwdriver (No. 2)
Turn off the instrument.
1
Remove all connection cables and thermocouples connected to
2
the module.
Remove the power cord.
3
Loosen the two module mounting screws with a Phillips
4
screwdriver.
Pinch the knobs and pull out the module.
5
When not installing another module after removal
Right side
Blank
panel
Place a blank panel.
1
Tighten two screws with a Phillips screwdriver.
2
If measurement is performed with the instrument without a blank
panel installed, the instrument may fail to meet specications
because of temperature instability within modules.
26
Installing and Removing Modules
Channel conguration
Modules are numbered beginning at the top, and channels are numbered beginning at the left of the
module installed at the top.
You can nd out information about the modules installed in the instrument in the System Information (p.
429).
2
Preparing for Measurement
LA
[1:4]LB[1:4]LC[1:4]LD[1:4]
CH1
CH3
CH15
CH2
CH4
CH16
LA
[1:4]LB[1:4]LC[1:4]LD[1:4]
L7A
[1:4]
L7B
[1:4]
L7C
[1:4]
L7D
[1:4]
Module 1
Module 2
Module 3
Module 4
Module 5
Module 6
Module 7
Module 8
Analog channels only
Both analog and logic modules are installed.
27
Attaching Connection Cables
2.2 Attaching Connection Cables
Read “Before connecting cables” (p. 11) carefully.
For detailed precautions and instructions regarding connections, refer to the instruction manuals for your
modules, connection cables, etc.
Measuring voltage
Applicable modules
• Model 8966 Analog Unit
• Model 8968 High
Resolution Unit
• Model 8972 DC/RMS Unit
Connect connection cables to the
BNC female connectors on modules.
The following connection cables can be connected
to the modules:
• Model L9197 Connection Cord
(Maximum input voltage: 600 V)
Large alligator clip type
• Model L9198 Connection Cord
(Maximum input voltage: 300 V)
Small alligator clip type
• Model L9217 Connection Cord
(Maximum input voltage: 300 V)
For measuring output signal
from a BNC connector
When a voltage to be measured exceeds a maximum input
rating of a module being used
• Model 9322 Differential Probe*
• Model P9000-01/-02 Differential Probe*
Example: Model P9000-02 Differential Probe
1
2
30
Measuring temperature
Attaching Connection Cables
Applicable Module
• Model 8967 Temp Unit
Connect a thermocouple to the
terminal block on the module.
The following thermocouple can be connected to the
module.
Thermocouple
(Compatible wire: from 0.4 mm
to 1.2 mm in diameter)
Connecting thermocouples to the terminal blocks
Required items:
Thermocouple, at-blade screwdriver (2.6-mm
blade)
Recommended wire:
Compatible wire: Thermocouple element
wires from 0.4 mm to 1.2 mm in diameter
Strip length: 10 mm
Outer insulation
25 mm
Inner
insulation
10 mm
1
Thermocouple
element wire
1
Terminal block
Connect thermocouple to
terminal block
2
Preparing for Measurement
Strip the insulation of the
thermocouple wires as shown on the
left.
Strip length: approx. 10 mm
Depress the button on the terminal
Connecting a thermocouple
Connection holes
3
2
block on the module with the at-
blade screwdriver.
Insert each thermocouple wire into
3
the appropriate terminal hole while
depressing the button.
Conrm proper polarity.
2
Release the button.
4
Connect the thermocouple to a
5
measurement object
• If noise inuences surrounding equipment, turn the thermocouple around the
accessory ferrite clamp-on choke several times (as seen in the right diagram).
• When connecting the thermocouple that is more than 3 meters long, the
measurement may be affected by EMC environments including external
noise.
4
The thermocouple is connected.
Connect to the measurement object.
5
To remove the thermocouple
Pull the thermocouple wire while depressing
the button.
31
Attaching Connection Cables
Measuring vibration or displacement with a strain gauge transducer
Applicable Module
• Model U8969 Strain Unit
• Model 8969 Strain Unit
Connect a strain gauge transducer to a connector on Model U8969 Strain Unit via Model L9769
Conversion Cable; Model 8969 Strain Unit via Model 9769 Conversion Cable.
The following device can be connected to the
module.
• Strain gauge transducer (Not available from Hioki)
Connect L9769 or 9769 Conversion Cable to the strain gauge
transducer.
Connecting the strain gauge transducer to a module’s connector
Example: Connecting the strain gauge transducer to Model U8969 Strain Unit via Model L9769
Conversion Cable
Required items:
Model L9769 Conversion Cable, strain gauge
transducer
Insert Model L9769 into a connector
1
of Model U8969 with the slot of
the plug aligned with the outward
indentation of the connector.
Insert the plug into the connector
2
until they are locked together.
Connect Model L9769 to the strain
3
gauge transducer.
Connector’s
indentation
U8969 Strain Unit
1
Connect the L9769.
Plug’s slot
2
Connect Model L9769 to the strain
3
gauge transducer.
Connect the strain gauge transducer to
4
a measurement object.
Connect the strain gauge transducer
4
to a measurement object.
How to disconnect Model L9769
Pull the sleeve of the plug gently,
releasing the plug, and disconnect the
cable.
The instrument describes Model U8969 as “8969.”
32
Connector pin-out
Attaching Connection Cables
Model U8969 Strain Unit
The metal shell is connected to
the GND of the instrument.
Pin markDescription
ABRIDGE+
BINPUT−
CBRIDGE−
DINPUT+
E
FSENSE+
GSENSE−
H, JN.C.
FLOATING COMMON
Model L9769 Conversion Cable
(Strain gauge transducer end)
AF
G
B
CD
The metal shell is connected to
the GND of the instrument.
Pin markDescription
ABRIDGE+, SENSE+
BINPUT−
CBRIDGE−, SENSE−
DINPUT+
E
F, GN.C.
Applied voltage:
E
bridge voltage of
FLOATING COMMON
2 V
2
Preparing for Measurement
Connection of Model L9769
• Pin F of the module end is connected with Pin A of the strain gauge transducer end.
• Pin G of the module end is connected with Pin C of the strain gauge transducer end.
33
Attaching Connection Cables
Example: Connecting the strain gauge transducer to Model 8969 Strain Unit via Model 9769
Conversion Cable
1
Connect a strain gauge transducer to
2
Connect the strain gauge transducer to
3
a measurement object.
Connector pin-out
Model 8969 Strain Unit
8969 Strain Unit
Connect the 9769.
Model 9769.
Required items:
Model 9769 Conversion Cable, strain gauge
transducer
Connect Model 9769 to a connector
1
on the module.
Insert the connector of Model 9769 with
the orange part facing upward.
Connect the Model 9769 to the strain
2
gauge transducer.
Connect the strain gauge transducer
3
to a measurement object.
Model 9769 Conversion Cable
(Strain gauge transducer end)
(Pin 1 is on the left when the module’s
top side is on the up)
1 2 3 4 5 6 7
Pin No.Description
1BRIDGE+
2SENSE+
3INPUT+
4INPUT−
5BRIDGE−
6SENSE−
7FLOATING COMMON
AF
G
B
CD
The metal shell is connected to
the GND of Model 8969.
Pin markDescription
ABRIDGE+
BINPUT−
CBRIDGE−
DINPUT+
EFLOATING COMMON
F, GN.C.
Applied voltage:
E
bridge voltage of 2V
IMPORTANT
• Preforming measurement with a strain gauge requires a bridge box. Use a strain gauge and bridge
box both of which are commercially available.
• The bridge box may be susceptible to the effect of noise. For more information about how to ground
the bridge box, refer to its instruction manual or contact the manufacturer of the bridge box.
IMPORTANT
Do not excessively bend the cable and the base between cable and connector, pull on them, nor twist
them. Doing so may cause the conversion cable to break.
34
Measuring current
Attaching Connection Cables
Applicable Module
• Model 8971 Current Unit
Connect a clump sensor to the connector
on the module via Model 9318 Conversion
Cable.
The following clamp sensors can be connected to
the module.
• Model 9272-10
Clamp On Sensor
• Model 9709, CT6862, CT6863,
CT6865
AC/DC Current Sensor
• Model CT6841, CT6843,
CT6844, CT6845, CT6846
AC/DC Current Probe
Connecting a clamp sensor to the module’s connector
Example: Connecting Model 9272-10 Clamp On Sensor
Sensor connector
Connect the 9318 Conversion
Cable and clamp sensor
Conversion cable plug
1
Conversion
cable connector
2
Required items:
Model 9318 Conversion Cable, Model 9272-10
Clamp On Sensor
Aligning the slots of and the conversion
1
cable plug with the module’s sensor
connector, insert the plug until it locks.
Align the plug of the clamp sensor to be
2
used with the slots of the conversion
cable connector and insert the plug until
it locks.
Attach the clamp sensor to a
3
measurement object.
2
Preparing for Measurement
Example: Model 9272-10 +
Model 9318
To disconnect the conversion cable
Clamp sensor
plug
9272-10
Attach the current sensor to a
3
measurement object
Pull the plug collar to release the lock and then
unplug the cable.
When measuring current with Model 9018-50 Clamp On Probe
Using Model 9018-50 enables voltage measurement modules such as Model 8966 Analog Unit to
measure current.
For more information about how to congure the instrument for use in this type of application, refer to “8.5.1
Example of Scaling Settings” (p. 169).
35
Attaching Connection Cables
Measuring logic signals
Read “Before connecting a logic probe to the measurement object” (p. 12) carefully.
For more information about logic probe specications, refer to the instruction manual of your logic probe.
Applicable Module
• Model 8973 Logic Unit
The logic channels LA through LD are
installed in the instrument.
The following logic probes can be connected to the
module.
• Model 9320-01 Logic Probe
• Model MR9321-01 Logic Probe
• Model 9327 Logic Probe
Connecting logic probes to the logic terminals
Example: Connecting Model 9327 Logic Probe
Right side
Required items:
Model 9327 Logic Probe
1
LOGIC terminals
2
1
LOGIC
terminals
Aligning the plug slots of the logic
probe with a logic terminal, insert the
logic probe plug.
Connect the logic probe to a
measurement object.
Connect the
2
logic probes to
measurement objects.
36
Attaching Connection Cables
Measuring voltage with a high degree of accuracy (digital voltmeter)
Applicable Module
• Model MR8990 Digital
Voltmeter Unit
Connect the test lead to the banana jacks
on the module.
The following lead can be connected to the module.
• Model L2200 Test Lead
(Maximum input voltage:
1000 V)
Connecting test leads to the banana jacks
Connect the test leads
Banana jacks
1
Required items: Test leads mentioned
above
Connect the test leads to the banana
1
jacks on the module.
Connect the black lead to the L jack; and
the red lead to the H jack. Make sure the
test lead plugs are fully inserted in the
jacks.
Connect the test lead to a
2
measurement object.
2
Preparing for Measurement
BlackRed
Connect the test leads to a
2
measurement object.
37
Attaching Connection Cables
Measuring high voltage
The following cables can be connected to the module.
Applicable Module
• Model U8974 High Voltage
Unit
Connect the connection cable to the
banana jacks on the module.
• Model L4940 Connection Cable Set
(Maximum input voltage: 1000 V)
Connecting connection cables to the banana jacks
Required items:
Connect the connection cable
Model L4940 Connection Cable Set
1
Banana jacks
Connect the plugs of the connection
cable to banana jacks on the module.
Connect each of the plugs to the banana
jack in the same color.
+
Model L4934
1
BlackRed
Connect the clips.
2
Accessory clips
Insert accessory clips into another
2
side of the cable.
Connect the cable clips to a
3
measurement object.
Model L4934 Small Alligator Clip Set
* Model L4932 is required when using
Model L4934.
Model L4935 Alligator Clip Set
Model 9243 Grabber Clip
Model L4936 Bus Bar Clip Set
Connect the
3
clips to a
measurement
object.
38
Model L4937 Magnetic Adapter Set
Model L4932 Test Pin Set
Outputting waveforms
Attaching Connection Cables
Applicable Modules
• Model U8793 Arbitrary
Waveform Generator Unit
• Model MR8790 Waveform
Generator Unit
Connect the connection cable to the output
terminal of a module.
The following cables can be connected to the
• Model L9795-01 Connection Cable (Electrical clips)
• Model L9795-02 Connection Cable (BNC output)
Connecting a connection cable to the output terminals
Example: Model U8793
Output terminal
Required item: Connection cable mentioned above
Insert the SMB connector of connection cable
1
in the output terminal of the module until it
clicks.
Connect the connection cable
Output terminal
1
Connect the cable clips to an object to which
2
the instrument applies the waveform.
module.
2
Preparing for Measurement
SMB connector
Connect the connection
2
cable to an object to which
the instrument applies the
waveform.
To disconnect output connectors
Firmly pinch the head of the SMB connector (other than
the cable), and pull it out.
39
Attaching Connection Cables
Outputting a pulse waveform
Applicable Module
• Model MR8791 Pulse Generator
Unit
Required items: Commercially available cable
Connecting a connection cable to the output connector
Output connector
Required items: Commercially available cable
Connect the connection cable to the output
1
connector of the module.
Connect the connection cable to the object to
2
Connect the connection cable
Output connector
1
Buttons
which the instrument applies the waveform.
To disconnect the connection cable from output
connectors
Pull the connector while depressing the buttons of the
connection cable.
2.3.1 Available Storage Devices (Inserting a CF Card and a USB
Flash Drive)
Icons indicating storage
devices
USB
ash
drive
Storage deviceHow to insert devices, notes
CF cardBuilt-in
drive
Insert a CF card
Fully insert a CF card with the top facing upward and
the mark (p) pointing toward the slot.
When the eject button sticks out, depress the button,
and then insert the CF card completely. Inserting
the CF card while the eject button is sticking out
may cause damage to the instrument. If the CF
card cannot be inserted completely, do not insert it
forcedly. Depress the eject button to make it stick out,
and then insert the CF card again.
The icons indicating the status of storage devices always appear at the
top of the screen.
: Storage devices are inserted.
: Storage devices are inserted and selected as the save
destination.
(Shown in red)
: Storage devices are not inserted; however, selected
as the save destination.
(Shown in black)
Front side
2
Preparing for Measurement
CF card slot
CF card
Built-in drive
Insert CF card
Eject button
To remove the CF card
Depress the eject button. After the button is
extended, depress it again to eject the CF card.
Model U8331 SSD Unit (storage capacity: 128 GB* 1 GB = 1,000,000,000 bytes)
* After formatting the drive, its capacity is decreased to less than 128 GB, which is an actually
available capacity.
The optional Model U8331 SSD Unit (factory option) is required.
Hioki formatted the hard disk during the production process.
Eject CF card
41
Preparing Storage Devices
Storage deviceHow to insert devices, notes
Memory (INT)
• The memory installed in the instrument is available. Only settings les can be written in the
memory.
• Automatic writing of data is not possible.
• Do not connect any devices other than USB ash drives.
• The instrument does not support all commercially available USB ash drives.
• To use a USB ash drive, you have to congure the setting on the instrument. See the
procedure described on the following page.
USB ash
drive
Inserting a USB ash drive
Orient the USB ash drive and insert it all the way
into the connector.
Removing the USB ash drive
Make sure that the instrument is not accessing the
USB ash drive (for saving or loading data, etc.), and
then remove the USB ash drive.
(No operation is required on the instrument.)
Right side
USB connector
(Type A)
Depending on the intended use of the USB ash drive, connector types and settings on the instrument
differ.
USB usageConnector
Using a USB ash driveType ALANUSB StickSee the following
Retrieving les saved on a CF card or the
built-in drive with a computer (via a USB
cable)
Type BLANMass storage HDD,
Instrument settings
InterfaceUSB setting
Mass storage CF
Reference
information
procedure.
“16.4” (p. 350)
Operating the instrument from a computer
(via a USB cable)
Type BUSBInterface“16.6.1” (p. 352)
Even when the SSD is installed as a built-in drive, it is recognized as an HDD.
Procedure
To display the screen
Press the SYSTEM key several times to display the [Interface] sheet on the system screen.
Move the cursor to the [Interface] box, and then select
1
[LAN].
Move the cursor to the [USB Set] box, and then select
2
[USB Stick].
42
Loading Recording Paper
2.3.2 Formatting Storage Devices
The instrument can format CF cards, USB ash drives, the built-in drive, and the internal memory. Once
they are formatted, the “HIOKI8847” folder is created.
Note that formatting a storage device deletes all the information on the storage device, and the
deleted information is unrecoverable.
Procedure
To display the screen
Press the FILE key to display the le screen.
Insert a storage device.
1
Select [Next Page].
2
Select [Format].
The cursor is placed on the
Select a storage device to be formatted, and then select
3
[Exec].
The specied storage device is formatted.
A conrmation window is displayed.
Select [Yes] to proceed the formatting action.
Select [No] to cancel the formatting action.
[Format Target]
box.
2.4 Loading Recording Paper
Read “Handling the printer and the recording paper” (p. 8) carefully.
2
Preparing for Measurement
Procedure
Paper roll axle
Required items: Model MR9231 Recording Paper, Paper
roll axles (accessory)
Press the button to open the printer cover.
1
Insert the paper roll axles into the core of Model 9321
2
Recording Paper.
43
Loading Recording Paper
Put the paper into the holder.
3
Insert rst the left side of the paper into the holder, and then,
until the holder clicks, put the paper into the holder while
pressing the paper leftward.
Face the print side of the paper upward. If the paper is
put in the holder without the paper roll axle installed, the
printer cover cannot be open, resulting in damage to the
printer.
Pull out the edge of the paper toward yourself, and
4
close the printer cover while applying the face of the
paper to the cover’s side surface.
The surface of the paper may become sticky due to the
residue left behind by the adhesive tape, with the result
that the printing on that area becomes unprintable.
Unroll approximately the rst 200 mm of paper before
inserting it into the printer.
How to withdraw the recording paper
As shown in the left gure, withdraw the paper while pressing
the paper roll axle leftward.
44
2.5 Supplying Power
Read “Before turning on the instrument” (p. 12) carefully.
2.5.1 Connecting the Power Cord
Procedure
Supplying Power
Right side
Connect the power cord to the power inlet on the instrument.
1
Plug the power cord into the mains outlet.
2
1
2
2.5.2 Connecting an Earthing Wire to the GND Terminal (Functional
Earth Terminal)
When performing measurement in a noisy environment, you can
improve the noise immunity by connecting an earthing wire to the GND
terminal (functional earth terminal).
Right side
2
Preparing for Measurement
When using a potential transformer (PT) for AC power line measurement
or others, connect an earthing wire to the GND terminal of the PT.
45
Supplying Power
2.5.3 Turning On and Off the Instrument
Turning on the Instrument
Right side
Power switch
: ON
Before starting measurement
To perform precise measurement, warm up the instrument about 30 minutes after turning on
the instrument to stabilize the internal temperature of the modules. After warm-up, perform zero
adjustment before performing measurement.
Turning off the instrument
Before turning off the instrument
When the instrument is turned off, data recorded in the internal memory is deleted. When you do
not want to lose recorded data, save it rst to a CF card or an external storage device.
Refer to “5 Saving/Loading Data and Managing Files” (p. 85).
Right side
Set the power switch in the on position ().
The splash screen is displayed rst, and then the
waveform screen is displayed.
Set the power switch in the off position ().
Power switch
: OFF
The instrument is turned off.
After the instrument is turned on again, the display is
displayed with the same settings as when the instrument
was turned off last time.
When the auto-setup function is enabled, settings are
read and congured automatically. (p. 107)
46
Setting the Clock
2.6 Setting the Clock
Set date and time for the built-in clock as follows.
The clock has an automatic calendar with leap year correction and 24-hour format.
The functions listed below make use of the clock. Ensure that the clock is set correctly before using these
functions.
• Performing measurement with timer-based trigger enabled
• Printing data including times of trigger events
• Saving measured data
Procedure
To display the screen
Press the SYSTEM key several times to display the [Init] sheet on the system screen.
Move the cursor to the [Clock] box.
1
2
Preparing for Measurement
Select the digits to change and set the
2
numeric values.
While the cursor is placed on one of the [Clock]
3
settings, select [Apply].
The date and time settings are accepted.
The date and time are displayed at the top right of
the screen.
Date setting
(Year, month, day)
Time setting
(Hour, minute,
second)
47
Adjusting the Zero Position (zero-adjustment)
2.7 Adjusting the Zero Position (zero-adjustment)
This operation compensates potential deection of modules and sets the reference potential of the
instrument to 0 V.
The compensation operation is performed for all channels and ranges.
Before performing zero-adjustment
• Warm up the instrument for about 30 minutes after the power-on to stabilize the internal temperature of
the modules, and then perform zero-adjustment.
• Note that zero-adjustment cannot be performed during measurement.
• Key operation is not accepted during zero-adjustment.
• The time required for zero-adjustment varies depending on types and the number of modules installed
in the instrument. (The operation may take several seconds.)
Procedure
To display the screen
Press the CHAN key several times to display the [Unit List] sheet on the channel screen.
Move the cursor to the [Zero-Adjust] box.
1
Select [Exec Zero-Adjust].
2
The zero-adjustment is executed.
Zero-adjustment has no effect on the 8969 and U8969 Strain Unit.
(Adjust the zero position using the auto-balance. (p. 183))
Perform zero-adjustment in the following cases.
• When a module has been replaced
• When the instrument has been cycled.
• When settings has been initialized (system reset)
• When measurement mode has been switched between DC and RMS on Model 8971 Current Unit, 8972
DC/RMS Unit, or U8974 High Voltage Unit
• When the ambient temperature has changed signicantly
The zero-position drift* may occur.
* Drift:
This means a spurious output caused by a shift in an operating point of an operational amplier. Drift can
occur due to changes in temperature and due to component aging over a period of use.
48
Performing Calibration (When Model MR8990 is Installed)
2.8 Performing Calibration (When Model MR8990 is
Installed)
This operation calibrates the scale of MR8990 Digital Voltmeter Unit.
The compensation operation is performed for all channels and ranges.
Before performing calibration
• Warm up the instrument for about 30 minutes after the power-on to stabilize the internal temperature of
the modules, and then perform zero-adjustment.
• Note that calibration cannot be performed during measurement. However, when the calibration setting
is set to on, calibration is performed at the start of measurement.
• Key operation is not accepted during calibration.
• The time required for calibration varies depending on types and number of modules installed in the
instrument. (The operation may take several seconds.)
Procedure
To display the screen
Press the CHAN key several times to display the [Unit List] sheet on the channel screen.
2
Preparing for Measurement
Move the cursor to the [Zero-Adjust] box.
1
Select [Exec Zero-Adjust].
2
Calibration is executed.
In the following cases, calibration should be executed again.
• When a module has been replaced
• When the instrument has been cycled.
• When settings has been initialized (system reset)
• When the ambient temperature has changed signicantly
49
Performing Calibration (When Model MR8990 is Installed)
50
3
Measurement
3.1 Measurement Procedure
1 Inspecting the instrument before measurement
Refer to “3.2 Inspecting the Instrument Before
Measurement” (p. 53).
2 Conguring the basic measurement settings
Refer to the following pages:
Select a suitable recording method for
an object to be measured
Set the sampling rate
Set the recording length
Set the waveform display format
and printing format
Advanced settings
Refer to the following pages:
“7.4 Plotting X-Y Composite Curves” (p. 144)
“8.2 Displaying Waveforms During the Writing in the Memory Simultaneously
(Roll Mode)” (p. 163)
“8.3 Overlaying New Waveforms With Past Waveforms” (p. 164)
“10 Numerical Calculation Functions” (p. 225)
“3.3.1 Measurement Functions” (p. 54)
“3.3.2 Time Axis Range and Sampling Rate”
(p. 56)
“3.3.3 Recording Length (Number of
Divisions)” (p. 60)
“3.3.4 Screen Layout” (p. 63)
3 Conguring the input channel settings
3
Measurement
Congure the analog channel settings
Congure the logic channel settings
Advanced settings
Refer to the following pages:
“8.1 Adding Comments” (p. 156)
“8.5 Converting Input Values (Scaling Function)” (p. 167)
“8.6 Setting the Waveform Position (Variable Function)” (p. 174)
“8.1 Adding Comments” (p. 156)
“8.8 Inverting the Waveform (Invert Function)” (p. 178)
4 Conguring the trigger settings
Refer to “9 Setting the Trigger” (p. 201).
Refer to the following pages:
“3.4.2 Conguring Analog Channels
Settings” (p. 67)
“3.4.3 Conguring Logic Channel Settings”
(p. 70)
51
Measurement Procedure
5 Starting measurement
Refer to the following pages:
“3.5 Starting and Stopping Measurement” (p. 73)
“5 Saving/Loading Data and Managing Files” (p. 85)
“6 Printing Data” (p. 119)
“7.1 Reading Measured Values (Using Cursors A and B)” (p. 134)
“7.3.2 Scrolling the Waveforms With the Jog Dial and Shuttle Ring” (p. 141)
“7.5 Magnifying and Demagnifying Waveforms” (p. 146)
6 Completing the measurement
Refer to “3.5 Starting and Stopping Measurement” (p.
73).
To observe the input signal
Pressing the AUTO key congures the timebase of the input waveform, vertical axis (voltage
axis) range, and zero-adjust settings automatically and then starts measurement.
Refer to “3.6 Measurement With Automatic Range Setting (Auto-Ranging Function)” (p. 76).
To load previously saved settings
Load the settings le using the le screen.
Saving the settings with respect to each measurement object and each application enables you
to perform measurement more convenient.
Refer to “5.3 Loading Data” (p. 104).
To automatically load previously saved settings at power-on
Previously saved measurement settings can be automatically loaded from a le when the
instrument is turned on. Simply insert a CF card containing an auto-settings le before turning on
the instrument.
Refer to “5.4 Automatically Loading Settings (Auto-setup Function)” (p. 107).
Make sure that the “HIOKI8847” folder of the CF card contains the le.
To initialize the instrument (restoring the basic default settings)
From the system screen, select the [Init] sheet to reset the instrument to the factory default
settings.
After initialized, the instrument is set up to perform simple measurement easily.
If the operation of the instrument seems unusual or overly complex, initialize the instrument.
Refer to “19.2 Resetting the Instrument” (p. 420).
52
Inspecting the Instrument Before Measurement
3.2 Inspecting the Instrument Before Measurement
Before using the instrument for the rst time, verify that it operates normally to ensure that no
damage occurred during storage or shipping. If you nd any damage, contact your authorized Hioki
distributor or reseller.
(1) Inspecting products used with the instrument
When using probes and connection cables
Is insulation of a probe or connection cable to
be used damaged, or is bare metal exposed?
No metal exposed
When using clamp sensors
Is a clamp cracked or damaged?
Metal exposed
Go to Step (2)
Yes
No
Do not use them if damage is present,
as you could receive an electric shock.
Request repairs.
3
Measurement
(2) Inspecting the instrument and modules
Is there any visible damage to the instrument
or modules?
No
When turning on the instrument
Does the Hioki logo appear on the screen?
Yes
Does the waveform screen appear?
Yes
Nothing appears, or
the display is abnormal.
Inspection terminates.
Go to Step (2)
Yes
No
If damage is evident, request repairs.
The power cord may be damaged,
or the instrument may be damaged.
Request repairs.
The instrument may be damaged.
Request repairs.
53
Setting Measurement Conditions
3.3 Setting Measurement Conditions
This section describes how to set measurement conditions.
The settings window displayed on the Waveform display enables you to congure basic settings
conveniently while you are observing waveforms. The basic settings can also be congured on the [Status]
sheet of the status screen
How to open the settings window
Settings
window
(P. 54)
Timebase
(Sampling
rate)
(P. 56)
Press the [DISP]
key.
Recording
length
(Number of
divisions)
(P. 60)
3.3.1 Measurement Functions
Select a function depending on a type of measurement.
Procedure
To display the screen
Press the DISP key to display the waveform screen.
Move the cursor to the function box (the uppermost
column of the settings window).
Used to measure instantaneous waveforms and transient phenomenon.
You can use the trigger and calculation functions.
Used to observe and record uctuations of phenomenon changing slowly
You can print data in real time.
Used to observe signals endlessly in X-Y curve format
You can set the X-Y composite curves plotting setting to on/off in the same manner as the
pen raising/lowering setting installed in typical X-Y pen recorders.
Refer to “4 X-Y Recorder” (p. 79).
Used to analyze frequencies
You can conduct various types of the spectrum and octave analyses.
Refer to “13 FFT Function” (p. 263).
Setting Measurement Conditions
Values obtained by the recorder function
While the recorder function is being used, the instrument obtains measured values at a previously set
sampling rate and lets both the maximum and minimum values be a sampling data value during each of
the sampling data periods. Thus, every sampling data value ranges in amplitude.
1 2 3 4........... ....100
1 div
If the input waveform uctuates widely due to noise even though usually it uctuates slightly, the
difference between the maximum and minimum values obtained with the instrument at the sampling rate
that is relatively faster can be inordinately large.
Set the sampling rate to a slower setting to prevent this phenomenon from being observed.
1 2 3 4........... ....100
One sampling data
Maximum value
Minimum value
Data acquired at the specied sampling rate
One sampling data
Maximum value
Minimum value
3
Measurement
1 div
When the sampling rate is fast
55
Setting Measurement Conditions
3.3.2 Timebase and Sampling Rate
The timebase setting denes a time length per division of the horizontal axis (time/div).
The sampling setting denes an interval at which the instrument samples waveforms.
(While the memory function is being used, the sampling rate appears, enclosed in
parentheses, under the timebase box. See the right gure. The displayed sampling
rate changes along with the timebase)
Procedure
To display the screen
Press the DISP key to display the waveform screen.
For the memory function
Move the cursor to the [Timebase] box.
1
Set the time per division on the horizontal axis
2
(time axis).
To control the sampling rate using an external signal, select [External] and input the signal to the
external sampling terminal. To set back the setting to the internal sampling, select [Internal] and
set the time.
When the external sampling is used, the settable number of sampling per division can range from
10 S/div to 1000 S/div.
Refer to “17.2.3 External Sampling (EXT.SMPL)” (p. 367).
(Select a sampling rate from among the options
that are 1/100 of the timebase or less)
56
The selectable range varies depending on the timebase setting.
The higher the sampling rate is, the better you can observe minute changes.
Setting Measurement Conditions
How to select the timebase
Refer to the table below to set the timebase.
For example, to measure a waveform with a frequency of 100 kHz, a maximum display frequency of
between 200 kHz and 800 kHz should be selected according to the following table. When you want to set
the maximum display frequency to 400 kHz, the timebase should be set to 10 µs/div.
TimebaseSampling rate
5 µs/div
10 µs/div
20 µs/div
50 µs/div
100 µs/div
200 µs/div
500 µs/div
1 ms/div
2 ms/div
5 ms/div
10 ms/div
20 ms/div
50 ms/div
100 ms/div
200 ms/div
500 ms/div
1 s/div
2 s/div
5 s/div
10 s/div
30 s/div
50 s/div
1 min/div
100 s/div
2 min/div
5 min/div
50 ns (20 MS/s)
100 ns (10 MS/s)
200 ns (5 MS/s)
500 ns (2 MS/s)
1 µs (1 MS/s)
2 µs (500 kS/s)
5 µs (200 kS/s)
10 µs (100 kS/s)
20 µs (50 kS/s)
50 µs (20 kS/s)
100 µs (10 kS/s)
200 µs (5 kS/s)
500 µs (2 kS/s)
1 ms (1 kS/s)
2 ms (500 S/s)
5 ms (200 S/s)
10 ms (100 S/s)
20 ms (50 S/s)
50 ms (20 S/s)
100 ms (10 S/s)
300 ms (3.33 S/s)
500 ms (2 S/s)
600 ms (1.67 S/s)
1 s (1 S/s)
1.2 s (0.83 S/s)
3 s (0.33 S/s)
Maximum display
frequency
800 kHz
400 kHz
200 kHz
80 kHz
40 kHz
20 kHz
8 kHz
4 kHz
2 kHz
800 Hz
400 Hz
200 Hz
80 Hz
40 Hz
20 Hz
8 Hz
4 Hz
2 Hz
0.8 Hz
0.4 Hz
0.13 Hz
0.08 Hz
0.04 Hz
0.067 Hz
0.033 Hz
0.013 Hz
The unit “S/s” stands for the
number of samplings per second.
3
Measurement
57
Setting Measurement Conditions
What is the maximum display frequency?
To plot sine waves that allow you to observe those peaks on the LCD, the
instrument needs to sample the waveforms at least 25 points per cycle.
The maximum display frequency varies depending on the timebase
setting.
If instrument plots non-existent waveforms (aliasing)
If a measured signal uctuates at a higher frequency compared to the
specied sampling speed, the instrument can probably plot non-existent
waveforms uctuating at a lower frequency than that of realistic signal
when the signal frequency reaches a certain level. This effect is called aliasing.
Actual input signal
Sampling points
Sampling rate
Observer waveform
Aliasing occurs because the sampling rate is slower than
the cycle of the input signal.
For the memory function, because the sampling rate can signicantly vary according to the timebase
setting, select the timebase setting carefully to avoid aliasing from occurring. Since the timebase setting
determines the maximum display frequency, perform measurement at as fast timebase setting as
possible.
When recording signals repeatedly, the auto-range function (P. 76) is useful.
The sampling rate is automatically set to 1/100 of the selected timebase.
When Model MR8990 Digital Voltmeter Unit is installed, the sampling rate of its channels is automatically
set to 1/50 of the selected timebase.
Example: When Model 8966 is installed as Unit 1 (CH1 and CH2) and Model MR8990 as Unit 2 (CH3
and CH4), then the timebase is set to 1 s/div
Sampling rate of Unit 1 (Model 8966): 10 ms
Sampling rate of Unit 2 (Model MR8990): 20 ms
To automatically set the timebase
Pressing the AUTO key selects a suitable timebase for an input signal and starts measurement.
(This function can be used in the memory function only.)
Refer to “3.6 Measurement With Automatic Range Setting (Auto-Ranging Function)” (p. 76).
58
Setting Measurement Conditions
• The timebase and sampling rate can be set independently. The sampling rate is selected depending on
the timebase setting.
• As mentioned below, when the former timebase settings are selected, the instrument displays
waveforms that are demagnied horizontally (in the time axis direction) at the latter scale size.
20 ms/div, ×1/2; 10 ms/div, ×1/5
To perform measurement minimizing the effect of noise
While measuring a waveform that usually uctuates only slightly, with a fast sampling rate set, a
difference between the maximum and minimum values may become large if the amplitude of the
waveform changes sharply. To prevent this effect, select a slower sampling rate or enable the
module’s low-pass lter (P. 67).
3
Note common to
The data refresh rate is not allowed to exceed the maximum sampling rate of the module.
During the period while data is being not updated, the same data is measured, with the result that stair-
like waveforms are plotted.
Even when a single signal is sampled simultaneously by two modules, the data may differ with one
another due to differences in the sampling rate, frequency bandwidth, and frequency response.
Data refresh rate for various modules
ModuleMax. timebaseMax. sampling rateReference information
Model 89665 µs/div50 ns (20 MS/s)–
Model 8967
Model 8968100 µs/div1 µs (1 MS/s)–
Model 8969
Model U8969
Model 8970Depends on settingDepends on settingRefer to “8.10.5” (p. 184).
Model 8971100 µs/div1 µs (1 MS/s)Refer to “8.10.6” (p. 187).
Model 8972
Depends on data refresh
setting
500 µs/div5 µs (200 kS/s)–
Depends on response
setting
Depends on data refresh
setting
Depends on response
setting
Refer to “8.10.3” (p. 182).
Refer to “8.10.7” (p. 187).
Measurement
Model 89735 µs/div50 ns (20 MS/s)–
Model MR8990Depends on NPLC settingDepends on NPLC settingRefer to “8.10.8” (p. 188).
Model U8974100 µs/div1 µs (1 MS/s)Refer to “8.10.9” (p. 189).
59
Setting Measurement Conditions
3.3.3 Setting the Recording Length
(Number of Divisions)
Set the recording length (number of divisions) of each data acquisition.
Procedure
To display the screen
Press the DISP key to display the waveform screen.
For the memory function
Move the cursor to the [Shot] box.
1
Select the type.
2
Fixed ShotSelect the length from among the xed values.
User ShotFreely specify a value in one division increments.
Set the recording length.
3
(Fixed Shot)
Model MR8847-51 (Memory capacity: 64 MW)
25, 50, 100, 200, 500, 1,000, 2,000, 5,000, 10,000, 20,000 div
50,000 div (in 2/4/8-ch mode)
100,000 div (in 2/4-ch mode)
200,000 div (in 2-ch mode)
Model MR8847-52 (Memory capacity: 256 MW)
25, 50, 100, 200, 500, 1,000, 2,000, 5,000, 10,000, 20,000,
50,000, 100,000 div
200,000 div (in 2/4/8-ch mode)
500,000 div (in 2/4-ch mode)
1,000,000 div (in 2-ch mode)
Model MR8847-53 (Memory capacity: 512 MW)
25, 50, 100, 200, 500, 1,000, 2,000, 5,000, 10,000, 20,000,
50,000, 200,000 div
500,000 div (in 2/4/8-ch mode)
1,000,000 div (in 2/4-ch mode)
2,000,000 div (in 2-ch mode)
(User Shot)
Model
MR8847-51
1 div to 40,000 div (in 16-ch mode)
1 div to 80,000 div (in 8-ch mode)
1 div to 160,000 div (in 4-ch mode)
1 div to 320,000 div (in 2-ch mode)
60
Model
MR8847-52
Model
MR8847-53
1 div to 160,000 div (in 16-ch mode)
1 div to 320,000 div (in 8-ch mode)
1 div to 640,000 div (in 4-ch mode)
1 div to 1,280,000 div (in 2-ch mode)
1 div to 320,000 div (in 16-ch mode)
1 div to 640,000 div (in 8-ch mode)
1 div to 1,280,000 div (in 4-ch mode)
1 div to 2,560,000 div (in 2-ch mode)
For the recorder function
Move the cursor to the [Shot] box.
1
Select the type.
2
Setting Measurement Conditions
Fixed ShotSelect the length among the prepared values.
User ShotFreely specify a value in one division increments.
When [ON] is selected, the instrument performs
measurement while continuously overwriting
the data in the memory. You can save and print
the data obtained during the period from the
measurement was stopped until the recording
reached its maximum record length.
1 div to 160,000 div
3
Measurement
Model
MR8847-52
Model
MR8847-53
1 div to 80,000 div
1 div to 160,000 div
61
Setting Measurement Conditions
Recording length and the number of samples
Every division of the recording length includes 100 samples. The total number of samples in a specied
recording length is: Set recording length (divisions) × 100 samples + 1.
However, when the Model MR8990 Digital Voltmeter Unit is installed, the number of samples is as follows:
• When only Model MR8990 is installed
Number of samples in a division: 50 samples
Total number of samples in the recording length:
Set recording length (divisions) × 50 samples + 1
• When both Model MR8990 and other modules are installed
Channels for Model MR8990
Number of samples in a division: 50 samples
Total number of samples in the recording length:
Set recording length (divisions) × 50 samples + 1
Channels for other modules
Number of samples in a division: 100 samples
Total number of samples in the recording length:
Set recording length (divisions) × 100 samples + 2
Each recording length division includes 100 samples, each of which consists of two values: the maximum
and minimum values.
Data obtained with Model MR8990 using the recorder function has a resolution of 16 bits.
When recording length is set to [Cont.]
• The instrument saves data that is obtained during the period that lasts until the measurement terminates
and has the maximum recording lengths (20,000 div for Model MR8847-51, 80,000 div for Model
MR8847-52, and 160,000 div for Model MR8847-53).
• For the range settings of from 10 ms/div to 200 ms/div, the instrument does not perform the real-
time print even when the printer setting (real-time print) is set to on. Perform printing manually after
measurement terminates (P. 124)
• When the auto-saving is set to on, the instrument does not save any data during the measurement.
When the measurement terminates, the data presently written in the internal memory is saved.
When the recording length is changed during measurement
Measured data already acquired in the memory is discarded, and measurement is started using the
newly set recording length.
62
Setting Measurement Conditions
3.3.4 Setting Screen Layout
You can lay out waveform displays of input signals on the waveform screen.
This setting also applies to the printing setting.
Selecting X-Y1 mode or X-Y4 mode plots X-Y composite curves.
(This function can be used in the memory and X-Y recorder functions.)
Refer to “7.4 Plotting X-Y Composite Curves” (p. 144).
Procedure
To display the screen
Press the STATUS key several times to display the [Status] sheet on the status screen.
For the memory function
Move the cursor to the [Format] box.
SingleDisplays and records waveforms in a graph. (Default setting)
DualDisplays and records waveforms in two graphs.
QuadDisplays and records waveforms in four graphs.
OctDisplays and records waveforms in eight graphs.
HexDisplays and records waveforms in 16 graphs.
X-Y SinglePlots and records curves that illustrate correlations between them
assigning multiple input signals to the X and Y axes in a graph.
X-Y QuadPlots and records curves that illustrate correlations between them
assigning multiple input signals to the X and Y axes in four graphs.
For the recorder function
Move the cursor to the [Format] box.
SingleDisplays and records waveforms in a graph. (Default setting)
DualDisplays and records waveforms in 2 graphs.
QuadDisplays and records waveforms in 4 graphs.
3
Measurement
OctDisplays and records waveforms in 8 graphs.
HexDisplays and records waveforms in 16 graphs.
Analog channel assignment
In Dual, Quad, Oct, and Hex mode, you can freely assign analog channels to the different graphs.
Procedure
To display the screen
Press the CHAN key several times to display the [Unit List] sheet on the channel screen.
63
Conguring Input Channels Settings
Move the cursor to the [Graph] box.
1
Select the display screen with respect to
2
<Example screen after assigning channels>
each channel.
The graphs are numbered beginning from
the top like Gr1, Gr2, Gr3.
3.4 Conguring Input Channels Settings
Conguring analog and logic channels settings.
To open the channel settings window
Refer to “Setting the variable function while the waveforms obtained
through all of the channels are being displayed” (p. 176).
Gr1
Gr2
Gr3
Gr4
Pressing CH.SET key switches
the sheets.
[Analog]
[Variable]
[Logic]
[Wave Calc] (when [Wave
Calculation] is set to on)
or
[Generator] (when a waveform
generator module is installed)
No display
If the channel setting screen is hard to be observed due to waveforms
overlay, reduce the waveform display width to separate the waveform
display screen from trigger settings screen.
Refer to “7.7.3 Switching the Waveform Display Width” (p. 152).
To reduce the number of displayed waveforms
Set the waveform display color setting on the Channel setting window to off.
Refer to “1. Waveform display color” (p. 67).
To apply a channel setting to another channel
Refer to “8.9 Copying Settings to Other Channels (Copy Function)” (p. 179).
64
Conguring Input Channels Settings
3.4.1 Channel Setting Procedure
The procedure below shows how to congure the analog channels (CH1 through CH16) settings.
1 Conguring the settings of input and screen display
Refer to the following pages:
Conguring the input coupling setting
Selecting an appropriate range
for the input signal
Scaling the input value (As needed)
Conguring the lter setting
(If performed in noisy environment)
Finely adjust the waveform amplitude
(As needed)
Magnify/demagnify the waveform vertically
(in the voltage-axis direction)
(As needed)
(P. 68)
(P. 67)
(P. 167)
(P. 70)
(P. 68)
(P. 68)
2 Conguring the trigger settings (as needed)
Refer to (P. 201).
3
Measurement
3 Conguring the settings of waveform display
colors and positions
Refer to the following pages:
Setting the waveform display color
Setting the display position and
magnication ratio (As needed)
(P. 67)
(P. 174)
4 Conguring the graph display settings
Refer to the following pages:
In Single, Dual, Quad, Oct, and Hex mode
In X-Y Single and X-Y Quad mode
(P. 63)
(P. 145)
65
Conguring Input Channels Settings
• Setting the input coupling set to GND disables the range settings because the waveforms seem to
have no amplitude.
• Filter attenuation may inuence correct range settings.
• When choosing trigger settings, set the vertical axis (voltage axis) range rst. When you change the
range after specifying the trigger, the trigger setting is changed.
• When using the variable function, set the vertical axis (voltage axis) range rst. When you change the
range setting after applying the variable function, observation of waveforms with sufcient precision
may not be possible.
• When using the variable and scaling functions together, enable the scaling function rst. When
you enable the scaling function after enabling the variable function, an unintended display may be
displayed.
This section describes the setting procedure for logic channels (standard logic terminals: LA through LD,
expansion logic terminals: L1A through L8D).
1 Conguring the display settings
Conguring the logic recording width settings
Refer to (P. 70).
2 Conguring the settings of waveform display colors
and positions
Refer to the following pages:
Setting the waveform display positions
Set waveform display colors
• The waveform display positions can be specied in one percent point increments.
• No logic waveforms are displayed in X-Y1 and X-Y4 modes
• Op to three logic modules can be installed. The fourth logic module and later modules installed in the
instrument are ignored.
(P. 70)
(P. 70)
66
Conguring Input Channels Settings
3.4.2 Conguring Analog Channels Settings
For information about specic settings for Model 8967 Temp Unit, 8969 Strain Unit, U8969 Strain Unit,
8970 Freq Unit, 8971 Current Unit, 8972 DC/RMS Unit, MR8990 Digital Voltmeter Unit, and U8974 High
Voltage Unit, see “8.10 Setting Details of Modules” (p. 180).
Procedure
To display the screen
Press the DISP key to display the waveform screen, and then press the CH.SET key to open the channel settings
window ([Analog] sheet).
Move the cursor to a box for which you
1
want to change the settings.
Select settings by pressing the F key.
2
1.2.3. 4. 5.6.7.
3
Measurement
To apply the channel settings to another channel
Refer to “8.9 Copying Settings to Other Channels (Copy Function)” (p. 179).
1. Waveform
display color
2. Vertical axis
(voltage axis)
range
Sets the color of the waveform of the selected channel. Each channel
can have a different color setting from or the same color setting as other
channels.
OffDisplays no waveform. When [Save Channel] of the auto-saving
setting is set to [Display Ch], the instrument does not save any data
of this channel.
Refer to “Select the channels to be saved.” (p. 93).
OnDisplays the waveform. To select the display color, press [↑] and [↓]
in the F keys.
ALL On-OffYou can switch the waveforms display between on and off for entire
channels.
Sets the vertical axis (voltage axis) range with respect to each channel. You
can set the vertical axis range in terms of voltages per division.
For information on the full-scale values (maximum displayable values) for
various modules, refer to the table on (P. 68).
When the variable auto-correction is set to [Off] and the variable function
is set to [On], the size of the waveform displayed on the screen is xed
regardless of the vertical axis (voltage axis) ranges.
If a waveform amplitude falls outside of the range
Changes the vertical axis (voltage axis) range to a lower sensitivity setting.
67
Conguring Input Channels Settings
3. Input coupling
4. Vertical axis
(voltage axis)
magnication
ratio
You can set the input signal coupling method. Typically, select the DC
coupling.
DC (V)Measures the input signals including both DC components and AC
components. (Default setting)
AC (
GND (
)
V
)
Measures the input signal, including AC components only. DC
components are rejected.
Measures the input signals short-circuiting the input signal to the
ground (This enables you to check the zero position of the signal).
You can set the magnication ratio of waveforms in the vertical axis (voltage
axis) direction for displaying and printing waveforms respect to each channel.
Magnifying waveforms are performed based on the zero position. Magnifying
waveforms does not inuence the measurement resolution.
Refer to “7.5.3 Magnifying/demagnifying the Waveforms Vertically (in the Voltage Axis
Direction)” (p. 149).
To display waveforms magnied in user-dened magnication ratios, enables
the variable function.
When signs of displayed waveforms are reversed, you can ip the waveforms
vertically.
Refer to “8.6 Setting the Waveform Position (Variable Function)” (p. 174) and “8.8
Inverting the Waveform (Invert Function)” (p. 178).
5. Vernier
6. Zero position
You can nely adjust input voltage amplitudes freely on the waveform screen
(This ne adjustment changes waveforms shape only seemingly).
Finely adjusting waveform amplitudes enables you to calibrate the instrument
for physical values recorded with sensors such as noise, temperature, and
acceleration sensors.
Refer to “8.7 Fine-Adjusting Input Values (Vernier Function)” (p. 177).
You can set the zero-volt level display position. If the zero-volt input level is
displaced, perform the zero-adjust.
Refer to “2.7 Adjusting the Zero Position (zero-adjustment)” (p. 48) and “2.8
Performing Calibration (When Model MR8990 is Installed)” (p. 49).
To move the zero positions of entire channels at a time, press [Preset].
Ascending
(3%→96%)
Descending
(96%→3%)
All Ch 0%Moves the zero position of entire channels to the 0% level of the
All Ch 50%Moves the zero position of entire channels to the 50% level of the
Moves the zero position of CH1 to the 10% level of the vertical
axis and those of the following channels, through CH16, to each
position that increases in 5 percentage points increments.
Moves the zero position of CH1 to the 90% level of the vertical
axis and those of the following channels, through CH16, to each
position that decreases in 5 percentage points increments.
vertical axis.
vertical axis.
68
If the zero position of the waveform obtained with Model 8969 or U8969
Strain Unit is displaced, perform the auto-balance.
Refer to “8.10.4 Setting Model 8969 and U8969 Strain Unit” (p. 183).
Conguring Input Channels Settings
• The zero position setting merely change waveform positions, not applying any offsets to inputs
signals.
• The vertical axis (voltage axis) magnication/demagnication of waveforms is performed based on
the zero position.
• The zero position adjustment and the vertical axis (voltage axis) magnication/demagnication merely
change displayable voltage ranges on the waveform screen. No actual measurement ranges are not
changed.
The zero positions are moved as shown in the illustration below. (Example:
Model 8966 Analog Unit)
Zoom factor ×1
A/D data
2047
2000 LSB
00 V
A/D data
−2047
Displayed area
(Zero position: 0%)
100%
50%
0%
Displayed area
(Zero position: 50%)
Displayed area
(Zero position: 100%)
Resolution of full-scale values (maximum displayable values) for modules in
various vertical axis magnication ratios (LSB)
Module
Model 8966
(Analog)
Model 8971
(Current)
Model 8972
(DC/RMS)
×1/10×1/5×1/2×1×2×5×10×20×50×100
20000
(4000)
10000
(4000)
4000200010004002001004020
Magnication ratio
3
Measurement
Model 8967
(Temperature)*
Model 8968
(High
resolution)
Model U8974
(High voltage)
Model 8969
Model U8969
(Strain)
Model 8970
(Power
frequency)
Model 8970
(Count)
Model 8970
(Excluding
power
frequency and
count)
Model MR8990
(DVM)
200000 100000400002000010000400020001000400200
320000
(64000)
250000
(64000)
20000100004000200010004002001004020
400000 200000800004000020000800040002000800400
100000500002000010000500020001000500200100
1200000 1200000 1200000 1000000
160000
(64000)
125000
(64000)
640003200016000640032001600640320
500002500012500500025001250500250
500000 200000 100000500002000010000
69
Conguring Input Channels Settings
The magnication ratios enclosed in the brackets represent ranges of valid data.
*: For Model 8967 Temp Unit, the valid range varies depending on thermocouples. For information about the
minimum resolution, see the specications of Model 8967 Temp Unit.
7. Low-pass
Filter
You can set the low-pass lter installed in the module. Enabling this
function can eliminate unwanted high-frequency components.
Settable lter types varies depending on modules. Select a suitable setting
according to an input signal characteristics.
3.4.3 Conguring Logic Channel Settings
The [Logic] sheet appears in Single, Dual, Quad, Oct, and Hex mode.
Procedure
To display the screen
Press the DISP key to display the waveform screen, and then press the CH.SET key to open the channel settings
window ([Logic] sheet).
Move the cursor to a box for which you
1
want to change the settings.
Select settings by pressing the F key.
2
1.
To apply the channel settings to another channel
Refer to “8.9 Copying Settings to Other Channels (Copy Function)” (p. 179).
1. Logic width
2. Waveform display
position
3. Waveform
display color
You can change the display width of the logic waveform.
Making waveforms narrower can enhance the readability of the display
containing a large number of waveforms.
WideDisplays waveforms at expanded width.
NormalDisplays waveform at the normal width.
NarrowDisplays waveform at reduced width. (Default setting)
You can set the logic waveforms positions.
You can freely move the positions of the logic waveforms within the range
of the display.
You can set the waveforms color of the selected channel. You can set the
same color as those for other channels.
For logic modules, either each module or each channel can have a
different color setting.
2.
3.
70
Conguring Input Channels Settings
OFFDisplays no waveform. When [Save Channel] of the auto-
saving setting is set to [Display Ch], the instrument does not
save any data of this channel.
Refer to “Select the channels to be saved.” (p. 93).
ONDisplays the waveform. Press the F key to set the display
colors.
Probe ON-OFFSwitches the waveform display between on and off at a time for
those obtained with a single probe.
ALL ON-OFFSwitches the entire logic waveform display between on and off
at a time.
This setting is selectable when the cursor is placed on any item
of the waveform display position box.
Setting the standard logic (LA, LB, LC, and LD) display setting to [ON] disables Model 8970 Freq Units
installed in Unit 1 and Unit 2 positions. Furthermore, it deteriorates the resolutions of Models 8967
Temp Unit, 8968 High Resolution Unit, 8969 Strain Unit, U8969 Strain Unit, and U8974 High Voltage
Unit from 16 bits to 12 bits. Installing Models MR 8990 Digital Voltmeter Unit, U8793 Arbitrary Waveform
Generator Unit, MR8790 Waveform Generator Unit, or MR8791 Pulse Generator Unit installed in Unit 1
and Unit 2 positions disables the standard logic terminals.
3.4.4 Display Sheet
The input channel settings can be set with respect to each display sheet. Up to four sheets are available.
You can assign waveforms to be displayed to each display sheets and switch a display sheet to another.
Switching sheets (1 through 4)
3
Measurement
71
Conguring Input Channels Settings
• You can set the following display-related settings only with respect to each display sheet.
Analog waveform: Display on/off, waveform color, magnication ratio, zero position, vernier,
invert, graph, variable (on/off, upper and lower limits)
Logic waveform: Display on/off, waveform color, display position, logic width
X-Y curve: Display on/off, display color, X ch, Y ch, waveform calculation (X ch, Y ch)
Common setting: Display format
• The measurement-related settings other than above are common to all displayed sheets.
When a range is changed, the ranges of entire display sheets are changed.
• Saving settings les operation saves the settings of all displayed sheets.
• Saving waveform les operation saves the waveforms in the setting of the sheet displayed at the time
of saving. When you try to load a waveform le, only the sheet displayed at the time of saving can be
loaded because other sheets were not saved.
You can configure the waveform calculation settings using the channel setting window.
See “3.4 Conguring Input Channels Settings” (page 64)
72
Starting and Stopping Measurement
3.5 Starting and Stopping Measurement
Procedure
To display the screen
Press the DISP key to display the waveform screen.
To start measurement
Press the START key to start measurement.
• When measurement is started, waveform data displayed on the screen is cleared.
• Applying a signal to the external control terminal can start measurement.
Refer to “17 Controlling the Instrument Externally” (p. 361).
To prevent starting measurement unintentionally
To prevent starting measurement by operation mistakes, you can set the START key acceptance
condition.
Refer to “Start Action” (P. 328).
To automatically save data during measurement
Refer to “5.2.2 Automatically Saving Waveforms” (p. 90).
To print waveforms
Refer to “6 Printing Data” (p. 119).
To stop the measurement in progress
Press the STOP key once to stop the measurement in progress after the specied recording length
expires.
Press the STOP key once again to stop the measurement in progress immediately.
You can change key acceptance conditions for the START and STOP keys.
Refer to “15 Setting the System Environment” (p. 325).
Pressing the STOP key may display the following GUI on screen.
3
Measurement
Stops the measurement in progress immediately.
(the same behavior as that when pressing the STOP key twice)
Cancels the stopping action and continue the measurement in progress.
When no keys are operated during this status, the measurement is terminated after the specied
recording length expires, and the display returns to the previous screen.
73
Starting and Stopping Measurement
End of measurement
[Storing Finished]
[ ]: Status bar display
RecordingStart measurement
End of measurement
[Trigger Wait]
T
Start measurement
[Storing Finished]
Recording
Stop measurement
[Trigger Wait]
[Now Storing]
Start measurement
[Trigger Wait]
Measurement and internal operations
Two options of the measurement methods available: the normal measurement (starts measuring and
recording waveforms at once) and trigger measurement (starts recording waveforms when trigger
conditions are satised). In this manual, “Start of measurement” indicates the time when you press the
START key, and “start of acquisition” indicates a start time of waveform acquisition on the waveform
screen.
To trigger the instrument once or repeatedly
Select the trigger mode. (P. 203)
To observe data acquired preceding the trigger point
Set a pre-trigger. (P. 217)
Typical
measurement
Triggered
measurement
Trigger mode:
[Single]
With the pre-trigger
not enabled
Without trigger setting
START key
Recording
starts
[Now Storing]
Recording
stops
The recording terminates after the recording length expires.
Triggering the instrument once
START key
Recording
starts
[Now Storing]
Recording
stops
A trigger event starts recording, and the recording terminates after the recording length
of the data is acquired.
Triggering the instrument repeatedly
Trigger mode:
[Repeat]
With the pre-trigger
not enabled
START keySTOP key
T
Recording
starts
Recording
Recording
stops
T
Recording
starts
A trigger event starts recording, and the recording terminates after the recording length
of the data is acquired, setting the instrument to wait for another trigger.
74
Starting and Stopping Measurement
[Pre-Trigger Wait]
The specified pre-trigger wait period is recorded before each trigger event.
Post-trigger waveform
Triggering the instrument repeatedly and recording phenomena before
each trigger event
Trigger mode:
[Repeat]
With the pre-trigger
enabled
Start measurement
START key
Recording
starts
[Trigger Wait]
After starting measurement and internally acquiring data for the
specified pre-trigger period, the Trigger Wait state is activated.
T
Recording
[Now Storing]
End of measurement
Recording
stops
[Storing Finished]
Operation is
repeated from
pre-triggering
stage.
Waveform during
specified pre-trigger
period
T
The data before a trigger event (for the pre-trigger period) is acquired.
When the trigger mode is set to [Repeat] or [Auto] (the memory function
only), the number of trigger events is displayed at the top of the screen. (storage
counter)
Refer to “Explanation of screen contents” (p. 20).
3
Measurement
75
Measurement in Automatic Range Setting (Auto-Range Function)
3.6 Measurement in Automatic Range Setting
(Auto-Range Function)
This function is enabled only for measurement with the analog modules with the memory function
enabled.
When you press the AUTO key after inputting signals to an analog module and select [Auto Range],
the horizontal axis range (timebase), vertical axis (voltage axis) range, and zero position are selected
automatically, and then measurement is started.
The timebase is set to the lowest numbered channel of the channels with those waveform display setting
set to on. In addition, it is set automatically so that waveforms with a length of from one cycle to 2.5 cycles
in 25 divisions are recorded.
Setting the auto-range function set to on changes the following items.
Module-related conditions (For all channels)
Vertical axis (voltage axis) range
Zero position
Automatically-set value
Vertical axis (voltage axis)
magnication ratio
Low-pass lterOff
Input couplingDC
Trigger conditions (For a single channel only)
Trigger modeAuto
Logical operation between trigger
sources
Pre-trigger20%
Internal trigger
Trigger type
Condition of [Status] sheet on the status screen
TimebaseAutomatically-set value (time axis magnication ratio: ×1)
×1
OR
Set to on is only the lowest numbered channel of channels with those waveform
display setting. (If the difference between maximum and minimum value is eight
divisions or less, the second lowest numbered channel is turned on.)
Level trigger: Slope, Rising
Trigger level: Automatically-set value
Filter: Off
76
Measurement in Automatic Range Setting (Auto-Range Function)
• When measurement is started in the auto-range setting, the trigger output signal is output from the
external control terminal block’s TRIG OUT terminal. Keep this in mind when performing auto-range
measurement while using the trigger output terminal.
• The auto-range function automatically selects settings for an input signal when the function is
enabled. After supplying an input signal (waveform), start auto-range measurement.
• If the input signal of the lowest numbered channel with the waveform display setting set to on
have extremely low amplitude, the timebase is set according to the input signal of the next lowest
numbered channel.
• If the range cannot be set for any channels with waveform display setting set to on, a warning
message is displayed, with the result that the measurement in progress is interrupted.
• The auto-range measurement disables the auto-saving and auto-printing functions.
• For a low-frequency signal of less than 10 Hz, the optimal ranges cannot be selected automatically.
Set the ranges manually.
• Using the following modules disables the auto-range function:
Models 8967 Temp Unit, 8969 Strain Unit, U8969 Strain Unit, 8970 Freq Unit, MR8990 Digital
Voltmeter Unit
3
Measurement
77
Measurement in Automatic Range Setting (Auto-Range Function)
78
4
• You can plot X-Y composite curves each of which presents the relationship between an input signal and
another in real time.
• Since the instrument writes the data of plotted curves in the memory, it can save the data as les as well
as print the curves.
• You can set the waveforms plotting setting to on and off in the same manner as the pen raising and
lowering setting installed in typical X-Y pen recorders.
• You can observe up to eight X-Y composite curves simultaneously.
• Since the instrument writes X-Y composite curves in the memory, it can redraw the X-Y composite
curves in other display settings after measurement.
Example of X-Y recorder waveform screen (in X-Y4 mode)
Refer to “3.4.2 Conguring Analog Channels
Settings” (p. 67).
80
4 Starting measurement
Refer to the following pages:
“4.3 Starting and Stopping Measurement” (p. 82)
“7.1 Reading Measured Values (Using Cursors A and B)”
(p. 134)
The X-Y recorder function does not
support the following operations:
• Trigger setting
• Auto-saving
• Auto-printing
5 Completing the measurement
Refer to the following pages:
“4.3 Starting and Stopping Measurement” (p. 82)
“4 Redraw X-Y composite curves” (p. 82)
“5.2.3 Saving Data Selectively (SAVE Key)” (p. 97)
“6.3 Manually Printing Data by Pressing the PRINT Key (Selection Print)” (p. 124)
Setting Measurement Conditions
4.2 Setting Measurement Conditions
To set the measurement conditions, press the STATUS key to display the status screen, and then select
the [Status] sheet. (The measurement functions and sampling rates can also be set on the waveform
screen.)
Setting items
Measurement
function
SamplingSet the sampling rate.
The setting [1 ms] can be set only when the [Dots-Line] is set to [Dots].
FormatSelect the graph format for the waveform screen display or printout of the input
Dots-Line
(interpolation)
Set the measurement function to [X-Y REC].
1 ms, 10 ms, 100 ms (Default setting)
signal. Up to eight phenomena can be observed simultaneously.
X-Y SingleThe waveforms of Graphs 1 through 8 are shown and recorded on a single
screen. (Default setting)
X-Y QuadThe waveforms of Graphs 1 through 8 are shown and recorded on four
screens.
Select whether to display and print the input waveform (sampling data) as dots
(sampling points) or perform the linear interpolation. Dot display allows faster
sampling.
DotsDisplays and prints sampling data as dots.
4
X-Y Recorder
LineDisplays and prints linearly interpolated sampling data. (Default setting)
Disp ClearWhen previously plotted X-Y composite curves remain at the start of
measurement, you can choose between leaving them and clearing them. When
you choose to leave them, the new waveforms are displayed as an overlay.
OffOverlays new waveforms with the existing waveforms.
OnClears the previously plotted waveforms. (Default setting)
Setting measurement conditions terminates.
Next, congure the analog channel settings.
For details, refer to “3.4.2 Conguring Analog Channels Settings” (p. 67).
81
Starting and Stopping Measurement
To choose channels used to plot X-Y composite curves
Refer to “7.4 Plotting X-Y Composite Curves” (p. 144).
4.3 Starting and Stopping Measurement
Press the DISP key to go to the waveform screen.
Start measurement.
1
Press the START key to start measurement.
Setting the X-Y composite curves plotting set to on and off (setting the pen setting to
2
up and down)
You can choose this setting both before and during measurement. Setting the pen setting to Down
enables the X-Y composite curve plotting function. Setting the pen setting to Up disables the X-Y
composite curve plotting function. This setting applies to all channels.
Move the cursor to the [Pen] box to change the setting.
UpDisable the X-Y composite curve
plotting function.
DownEnable X-Y composite curves to be
plotted.
Setting the pens in the Down position
The pens are shown in colors and curves are
plotted.
Stop the measurement in progress
3
Press the STOP key to stop the measurement in progress.
Redraw X-Y composite curves
4
The instrument plots the X-Y composite curves that start from any specied sampling point and consists
of 4,000,000 samples of the data recorded before the measurement terminated, playing back the X-Y
composite curve plotting in the same manner as video recorders.
This setting applies to all channels. You can specify the playback speed. (You can change it also during
playback.) X-Y composite curves are only played back on the screen, not affecting the printout data. To
print the same X-Y composite curves as shown on the screen, use the screenshot function (p. 130).
Pressing the TRIG.SET key or the Manual
Trigger key can switch the [Pen] setting
between up and down as well.
Setting the pens in the Up position
The pens are shown in white, and no X-Y
composite curves are plotted.
82
Starting and Stopping Measurement
Move the cursor to the [Player] box.
Move the cursor to the [Speed] box.
In the [Normal] setting, plays back the X-Y
composite curves at the same speed as at
which they recorded.
Inputting signals to the external control terminal can start and stop measurement and switch the [Pen]
setting between up and down as well. (p. 363)
ClearClears only the displayed X-Y composite
curves.
(The instrument does not discard the
waveform data written in the memory.)
RedrawRedraws the X-Y composite curves. The
X-Y composite curve display conditions
can also be changed for plotting the curves
again.
FirstMoves the pens to the start points of the
X-Y composite curves.
LastMoves the pens to the end points of the
X-Y composite curves.
Play/StopStarts and stops the X-Y composite curve
playback.
Very Fast / Fast / Normal (Default setting) / Slow / Very
Slow
4
X-Y Recorder
To redraw the X-Y composite curves in different X-Y composite curve display
settings
• Even after the X-Y composite curve display is cleared, the waveform data is still retained in
the memory. This makes it possible to redraw the X-Y composite curves in different settings,
which include the display format, display color, channels used for plotting the curves,
channel magnication ratio, and channel offset, from those previously set. (Select [Redraw])
Executing [Redraw] update the screen and printout data.
(Until [Redraw] is not executed, the screen and print data are not updated, even if the
settings are changed.)
• The X-Y composite curves can also be redrawn by operating the jog dial and shuttle ring.
Jog dial Plots the X-Y composite curves and plays back the plotting of the curves inversely
in one sample increments.
Shuttle ring Plots the X-Y composite curves and plays back the plotting of the curves
inversely at speed proportional to the rotation angle.
• To play back the plotting of the X-Y composite curves inversely, plot curves in the direction of earlier
data. Erasing the X-Y composite curves is not possible.
• If the measured data exceeds 4,000,000 samples, it is only possible to redraw up to 4,000,000
samples in the past, counting backward from the point when measurement terminated. The
instrument discards waveform data obtained earlier.
• Redrawing the X-Y composite curves with the jog dial and shuttle ring affects the X-Y composite
curves displayed on the screen only and does not affect the printout data. To print the same X-Y
composite curves as shown on the screen, use the screenshot function (p. 130).
83
Observing X-Y Composite Curves
4.4 Observing X-Y Composite Curves
Up to 4,000,000 samples of waveform data are written in the memory, and you can observe those
measured values using Cursors A and B. (p. 134)
The bar at the top of the screen indicates the amount of the occupied memory.
When the number of samples reaches 4 million, the sign [OVER] appears
next to the screen bar at the top.
s
Saving X-Y
composite
curves
The data saved in this way as a le can be loaded into the instrument; however, cannot be loaded into
computers.
Printing X-Y
composite
curves
Pressing the SAVE key or executing the saving command using the le screen saves
the waveform data written in the memory as a le.
Refer to “5.2.3 Saving Data Selectively (SAVE Key)” (p. 97).
Pressing the PRINT key prints out the X-Y composite curves.
Refer to “6.3 Manually Printing Data by Pressing the PRINT Key (Selection Print)”
(p. 124).
84
Saving/Loading Data and
5
Managing Files
This chapter explains how to save and load data and manage les.
Before saving data, congure the save settings on the [File Save] sheet of the system screen.
The File screen allows you to load data and manage les.
To display the [File Save] sheet
Pressing the SYSTEM key
switches the sheets to be
displayed in the following
order:
[Environment][Init]
[File Save]
[Printer][Interface]
Operations available on the [File Save] sheet
The following settings are available on the [File Save] sheet. Refer to (p. 89).
Auto-saving settings
Setting the method of the auto-saving of waveform data.
SAVE key settings
Setting the method of saving the data when pressing the SAVE key.
5
Saving/Loading Data and Managing Files
85
To display the le screen
Indicated the order
in which theles are
sorted.
: Ascending order
△
:
Descending order
▽
Press [FILE] key.
The selected le is indicated
by the ashingcursor.
Press Left/Right CURSOR
key to move between folder
levels.
Press the Up/down CURSOR
keys to select les.
Read-only les and folders are
shown in blue.
These les and folders cannot
be renamed or deleted.
Storage device information
Shows information about the selected storage device.
Count: Displays the number of les in the selected folder.
Dir Count: Displays the number of folders in the selected folder.
Count: Displays the number of les and folders you can create
at the selected folder level.
When the storage device is not displayed, and you cannot
perform any operations, choose a setting option other than
[Mass storage HDD] or [Mass storage CF] in the [USB
Set] box.
To change storage devices
Verify that a storage device is inserted.
1
Refer to “2.3 Preparing Storage Devices” (p. 41).
Select [Change Media] and select a storage device type.
2
The les saved on the selected storage device appears on the le list.
Operations available from the le screen
Formatting storage devices (p. 43)
Loading data (p. 104)
Saving and loading auto-settings les (p. 107)
Displays information of the
selected le.
Managing les (p. 89)
• Saving data (p. 109)
• Creating a new folder (p. 112)
• Copying a le (p. 116)
• Sorting les (p. 114)
• Deleting a le (p. 113)
• Renaming a le (p. 115)
• Printing a le table (p. 117)
86
Data That Can Be Saved and Loaded
5.1 Data That Can Be Saved and Loaded
File type
Settings data
*1
File
format
BinarySET
IconFile extension and description
Auto.
Settings data (measurement
conditions)
Save
–
Manual
Load
: Yes, –: No
Computer-
readable
–
Waveform data
The entire length
of the waveforms
or the portion of
waveforms within
the range between
Cursors A and B
(acquired in the
instrument)
*2
Waveform
management
*3
data
(memory division /
division save)
Captured screen
images (Display/
Waveform
screens)
Numerical
calculation results
Comments for
printing
MEM
REC
Binary
XYC
FFTFFT function data
Text
(Index
le)
*5
BMP
TextCSVText data
TextTXTText data––
CSVText data
IDXIndex data for division save
SEQ
BMPImage data–
Memory function waveform
data
Recorder function waveform
data
X-Y recorder function X-Y
curve data
Index data for memory
division (automatically
created during batch saving)
–
*4
–
*4
–
–
–
–
–
–
5
Saving/Loading Data and Managing Files
–
–
*6
Waveform
evaluation
BinaryARE
settings data
Waveform
evaluation area
Arbitrary
waveform data
User-dened
waveform data
Pulse pattern
data
Generation
program data
*1: Multiple les can be registered in the instrument and loaded selectively. Automatic loading at power-on is also
possible. (p. 107)
*2: To load the data onto the instrument
Save the data in binary format. The waveform data and a part of the measurement settings data are saved.
To load the data onto computers
Save the data in text format. (p. 89)
To save a portion of the waveform
Specify the range to be saved with Cursor A and B. (p. 134), (p. 139)
*3: To load the data, which is saved with the memory division enabled, of entire blocks at once
Save measured data in the [All blocks] setting. A directory including waveform data for each block and index
data (SEQ) is automatically created. Load this index data.
To load waveform data that is saved using division save
Load the IDX index data.
*4: Load les with Wave Viewer (Wv).
*5: This is a standard Windows
*6: You can print out text les created with a computer along with loaded waveforms. Other processing operations
are not available.
®
graphics format. Many graphics programs can open les in this format.
Files that exceed 2 GB cannot be saved.
Data that cannot be loaded onto the instrument
• Data saved by other instruments
• Image les other than waveform evaluation areas (
• Files displayed with the icon
.
)
88
5.2 Saving Data
5.2.1 Save Types and Setting Procedure
There are basically three types of save operations.
Saving Data
To save data automatically
during measurement (p. 90)
Auto-saving
Every time the instrument acquires
the recording length of measured
data, saves the data automatically.
Select the save destination and data
type before starting measurement.
(p. 90)
The data is saved to the CF card.
After installation and
connection, turn on the
instrument.
Insert a storage device.
Set the measurement
conditions.
[File Save] sheet
To save data manually by pressing the SAVE key (p. 97)
To save data immediately
Quick save
This function enables you to
save waveforms repeatedly
and to save waveforms quickly
while monitoring them. Specied
the save destination and data,
and then press the SAVE key.
(p. 97)
After installation and
connection, turn on the
instrument.
Insert a storage device.
[File Save] sheet[File Save] sheet
Set [Select at save] to [No].
To save data after
selecting items
Selection Save
(Default setting)
Pressing the SAVE key enables
you to specify the data to be
saved, and then saves the data.
You can congure save settings
not only on the [File Save] sheet
but also on any screen. (p. 97)
After installation and
connection, turn on the
instrument.
Insert a storage device.
Set [Select at save] to [Yes].
5
Saving/Loading Data and Managing Files
Set the [Auto Save] setting to
[On].
Select the save destination.
Set the items to be saved.
• Waveform data
• Numerical calculation
results*
Measuring waveforms
Automatically saves the data.
Select the save destination.
Set the items to be saved.
Data that can be saved:
• Settings data
• Waveform data
• Display screen
• Waveform screen
• Numerical calculation
results
• Waveform evaluation
setting data
• Waveform evaluation area
• Pulse pattern data
• User-dened waveform
data
• Program data
Press the SAVE key.
Saves the data.
Press the SAVE key.
Dialog box
Select the save destination.
Set the items to be saved.
Data that can be saved:
• Settings data
• Waveform data
• Display screen
• Waveform screen
• Numerical calculation
results
• Waveform evaluation
setting data
• Waveform evaluation area
• Pulse pattern data
• User-dened waveform
data
• Program data
Select [Exec].
Saves the data.
Check the following before saving data:
• Have you already inserted and initialized the storage device?
(p. 41), (p. 43)
• Is the save destination specied correctly?
• Is [Auto Save] set to [On]? (For auto-saving)
* To save numerical calculation results
automatically
Refer to “10.4 Saving the Numerical
Calculation Results” (p. 236).
89
Saving Data
5.2.2 Automatically Saving Waveforms
The instrument automatically saves the waveform data as les every time it writes the recording length
of the waveform data in the memory. Set the save destination and items to be saved before starting
measurement.
Procedure
To display the screen
Press the SYSTEM key to open the system screen, and then select the [File Save] sheet.
Enable auto-saving.
1
Move the cursor to the [Auto Save] box, and select
[On].
Default setting: Off (no automatic saving)
Set the save type.
2
Move the cursor to the [Save Type] box.
1
2
3
4
5
Wave
Binary
Wave
Text
Saves waveform data in binary format.
(Only the data saved in binary format can be
loaded onto the instrument.)
Saves waveform data in text format.
The instrument decimates the data and saves
it as a le.
(You can open the les with editors and
spreadsheet software installed in computers;
however, cannot load onto the instrument.)
90
Set the save destination.
3
Move the cursor to the [Save To] box and select
[Edit].
The folder dialog box is displayed (at the bottom right).
Move the cursor to a device to be set as the save
destination*, and select [Conrm] to conrm the
selection.
HDDAutomatically saves the waveform data on the
built-in drive.
(When Model U8331 SSD Unit is installed.)
CFAutomatically saves the waveform data on a CF
card.
USBAutomatically saves the waveform data on a
USB ash drive.
LANAutomatically saves the waveform data on a
computer connected via LAN.
Model 9333 LAN Communicator is required.
When the root directory (the topmost folder on the storage
device) is selected, the “HIOKI8847” folder is automatically
created. (When the storage device has been formatted
with the instrument, the instrument has already created the
folder.) Files are saved in this folder.
Saving Data
To select a
storage device
To open the
lower level folder
Up/down
CURSOR
Right
CURSOR
*
•
To save the waveform data on
a computer connected via LAN
The 9333 LAN Communicator must be
installed on the computer on which the
data is to be saved. For more information
about how to operate and congure the
Model 9333 LAN Communicator, as well
as install it on computers, refer to the
instruction manual of Model 9333 LAN
Communicator.
•When the save destination is
set to
The IP addresses of the instrument and
the computer connected via LAN must
be set. For more information about how
to set the IP addresses, refer to (p. 332).
[LAN]
5
Saving/Loading Data and Managing Files
To create a new folder to be set as the save
destination,
select [New Folder]. When the save destination is
set to [LAN], the [New Folder] command is ignored,
and a folder named with the current date is created.
Set the le name.
4
Move the cursor to the [Save Name] box and enter a
le name.
Refer to “Save operations (When the save destination is set
to [LAN])” (p. 96).
When the save destination is set to [LAN], the le is
named in the previously-determined format and the
name entered in the [Save Name] box is ignored.
Refer to “Save operations (When the save destination is set
to [LAN])” (p. 96).
File name
The maximum number of characters for [Save Name] is 123. The maximum path length including le
name is 255 in characters.
91
Saving Data
Specify the action when a le with the same name
5
exists in the target folder.
Move the cursor to the [Same Name] box.
AutoWhen no les with the same name exist, the
SerialUnconditionally, the instrument gives the
instrument gives the predetermined name to
a le and saves it. When another le with the
same name exists, the instrument gives the
name beginning with a 4-digit number to a
le and saves it (default setting).
When the name of the existing le begins
with a numeral, the instrument gives the
name beginning with the sequence number
that follows the existing le name’s number
to a le.
name beginning with a 4-digit number to a le
and saves it. When another le with the same
name exists, the instrument gives the name
beginning with a 4-digit number increasing
the number in one increments to a le and
saves it.
Set the save area.
6
Move the cursor to the [Save Area] box.
Whole
Wave
A-B
Wave
(When [SaveType] is set to [Wave Binary])
7
Saves the entire data written in the memory.
(Default setting)
Saves the data within the range between
Cursors A and B. When only Cursor A is set,
the instrument saves entire data within the
range after Cursor A point.
To set Cursors A and B, refer to (p. 134).
Select whether to divide les.
Move the cursor to the [Division] box.
OffSaves a le without dividing it.
16M,
32M,
64M
Divides a le into sections that have a
selected size and saves them.
(When [SaveType] is set to [Wave Text])
Set the data decimation factor.
Move the cursor to the [Save Thin] box.
92
OFFNo data is decimated.
1/2 to
1/1000
Sets the decimation factor (ratio of retained
samples to whole samples).
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.