Hioki IM9202 Instruction Manual

Page 1
IM9202
HIOKI IM9202A961-00
TEST FIXTURE
Instruction Manual
Apr. 2020 Edition 1 IM9202A961-00 20-04H
Be sure to read this manual before using the instrument.
When using the instrument for the
rst time
Part Names and Functions
Making Measurements
p.6 Maintenance and Service p.25
Troubleshooting
EN
Page 2
HIOKI IM9202A961-00
Page 3

Introduction

HIOKI IM9202A961-00
Thank you for choosing the Hioki IM9202 Test Fixture. To ensure your ability to get the most out of the device over the long term, please read this manual carefully and keep it available for future reference. The device is used by connecting to the impedance analyzer. Before using the impedance analyzer, read the instruction manual of the impedance analyzer carefully.
Intended audience
This manual has been written for use by individuals who use the product or provide information about how to use the product. In explaining how to use the product, it assumes electrical knowledge (equivalent of the knowledge possessed by a graduate of an electrical program at a technical high school).
Contents
Introduction
Checking Package Contents
Safety Information
Operating Precautions
Overview
Part Names and Functions
Making Measurements
■Measurement Procedure
■Inspecting the Device Before Use
■Attaching the Test Head and the Adapter
■Calibration
■Attaching the Device
■Electrical Length Correction
■Short Correction
■Open Correction
■Starting Measurement
p. 2
p. 3
p. 4
p. 5
p. 6
p. 8
p. 8
p. 9
p. 10
p. 11
p. 14
p. 15
p. 16
p. 19
p. 22
IM9202A961-00
Specications
Maintenance and Service
Warranty Certicate
p. 23
p. 25
1
Page 4

Checking Package Contents

HIOKI IM9202A961-00
Checking Package Contents
When you open the package, carefully inspect the device to ensure that everything is in good condition, and that no damage occurred during shipping. Carefully check the accessories. If
device seems to have been damaged or does not work as specied, contact your authorized Hioki
distributor or reseller.
Conrm that the following items are provided.
(9)
(1)
(2)
(13)(3)
(4) (7)
IM9202 Test Fixture ........................................................................................................(1)
H-side lead component clip* ...........................................................................................(2)
L-side lead component electrode (with clip)* ..................................................................(3)
L-side SMD electrode .....................................................................................................(4)
SMD open correction jig ................................................................................................ (5)
Short plate ..................................................................................................................... (6)
Brush for cleaning...........................................................................................................(7)
Carrying case .................................................................................................................(8)
Instruction manual
* The H-side lead component clip and the L-side lead component electrode are installed on the device.
(5)
(6)
(11)(10)
(8)
(12)
Options (sold separately)
The options listed below are available for the device. To purchase an option, please contact your authorized Hioki distributor or reseller. Options are subject to change. Please check Hioki's website for the latest information.
IM9906 Adapter (3.5 mm/7 mm) .....................................................................................(9)
IM9905 Calibration Kit (LOAD) .......................................................................................(10)
IM9905 Calibration Kit (OPEN).......................................................................................(11)
IM9905 Calibration Kit (SHORT)* ...................................................................................(12)
Magnifying glass (IM9200 Test Fixture Stand accessory) ..............................................(13)
IM9200 Test Fixture Stand
2
Options (sold separately) can be stored inside (9) to (13) of the carrying case. Please store them after removing the packed sponge at the time of unpacking. * The IM9905 Calibration Kit (SHORT) can be stowed side by side with the short plate (6).
Page 5

Safety Information

HIOKI IM9202A961-00
Carefully read the following safety notes before using the device
CAUTION
• Mishandling during use could damage to the device. Familiarize yourself with the instructions and precautions in this manual before use the device.
• If you have not used electrical measuring instrument before, you should be supervised by a technician who has experience in electrical measurement.
Notations
In this document, the severity levels of risk and hazard are classied as follows.
Safety Information
WARNING
CAUTION
IMPORTANT
Indicates a potentially hazardous situation that may result in death of or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the device or malfunction.
Indicates information or content that is particularly important from the standpoint of operating or maintaining the device.
Indicates an action that must not be performed.
Indicates an action that must be performed.
Notations related to standards compliance
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Other notations
Indicates that the product conforms to regulations required by the EU Directive.
*
(p. ) Indicates the page number to reference.
Instructs the reader to see below for additional information.
3
Page 6

Operating Precautions

HIOKI IM9202A961-00
Operating Precautions
Observe the following precautionary information to ensure that the device can be used safely and
in a manner that allows it to perform as described in its specications. Use of the device should conrm not only to its specications, but also to the specications of all
accessories, options and other equipment in use.
Installation Environment
WARNING
Installing the device in inappropriate locations could cause a malfunction of the device, an accident. Avoid locations that are:
• Exposed to direct sunlight or high temperatures
• Exposed to corrosive or combustible gases
Exposed to strong electromagnetic elds or electrostatic charges
• Near induction heating systems (such as high-frequency induction heating
systems and IH cooking equipment)
• Susceptible to vibration
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to high concentrations of dust particles
Do not place the device on an unstable or uneven surface. Doing so could cause the device to fall or turn over, causing bodily injury or damage to the device.
Handling the Device
• The DUT could pop off the device and into the air while being connected. To prevent injury, use gradual movements when handling clips and electrodes.
• To avoid damage to the device, avoid subjecting it to vibration or mechanical shock during transportation and handling. Exercise particular care to avoid subjecting the device to mechanical shock, for example by dropping.
• To prevent device damage or an electric shock, use only the screws that are originally installed for securing clips or the short plate in place. If you have lost any screws or
nd that any screws are damaged, please contact your authorized Hioki distributor or
reseller.
Measurements
• To avoid bodily injury or damage to the device, do not attempt to measure voltage exceeding 42V.
• To avoid damage to the device, do not connect the charged capacitor to the device.
• Do not apply a voltage across the device’s measurement terminals while it is connected to an impedance analyzer.
CAUTION
CAUTION
CAUTION
It may cause damage to the impedance analyzer.
Shipping Precautions
4
IMPORTANT
The accuracy of measured values obtained using the device is not guaranteed. Measured values should be treated as reference values.
During shipment of the device, handle it carefully so that it is not damaged due to a vibration or shock.
Page 7

Overview

G
HIOKI IM9202A961-00
The device is the test xture to be used by connecting to the impedance analyzer of Hioki.
It performs two-terminal measurement of lead components and SMDs.
Measurable DUT size
Lead components
Axial
Overview
B
Radial
SMD*
A
E
D
2
A Lead interval
(Component body length)
C
B Lead length 2 mm to 10 mm
C Height to lead*
Lead wire diameter 0.5 mm to 2 mm
D Lead interval 2 mm to 26 mm
E Lead length*
Lead wire diameter 0.5 mm to 2 mm
F Component length 1.6 mm to 23 mm
G Component width 0.8 mm or longer
1
1
1 mm to 25 mm
2.5 mm or shorter
2 mm or longer
H Component height 0.65 mm or longer
F
H
*1: The component’s H-side lead must not make contact with the xture’s L-side electrode. *2: The electrode on the bottom surface of the H side of the component must not make contact with the xture’s
L-side electrode.
Supported SMD size: EIA0603 (JIS1608) or larger
5
Page 8

Part Names and Functions

HIOKI IM9202A961-00
Part Names and Functions
(7)
(8)
(9)
(10)
(8)
(1)
(2)
(3)
(4)
(5)
(6)
(1) Base plate The base plate is at the same potential as the L-side electrodes.
(2) H-side lead component
clip
Connect lead components to the H-side electrode by inserting the lead wire here.
(11)
(12)
(3) H-side electrode Serves as the device’s measurement unit.
(4) IM9906 Adapter
(3.5 mm/7 mm)
(5) Test head The test head comes with the impedance analyzer.
(6) IM9200 Test Fixture Stand
(option)
(7) Loupe The magnifying glass provided with the IM9200 Test Fixture Stand
(8) Stand xing knob The knob xes the device to the IM9200 Test Fixture Stand.
(9) L-side lead component
clip
(10) L-side lead component
electrode
(11) Serial number The 9-digit serial number indicates the year (rst two digits) and the
(12) Knob This knob is used to lock the L-side lead component electrode in
This specially designed adapter is used to connect the device to the impedance analyzer’s test head.
The exclusive stand for xing the test head of the impedance analyzer
and the device.
(option). It is used by installing to the IM9200.
Connect lead components to the L-side electrode by inserting the lead wire here.
Serves as the device’s measurement unit.
month of manufacture (next two digits). Do not remove this sticker as the number is important.
place.
6
Page 9
The electrode and the clips for measurement of lead components
HIOKI IM9202A961-00
Part Names and Functions
(9)
(10)
The electrode for measurement of SMDs
(13)
(14)
(13) L-side SMD electrode Used when measuring SMDs.
(2)
(14) Lever This lever causes the L-side SMD electrode to slide back and forth.
Short plate
(15) Short plate Used when performing short correction.
(15)
7
Page 10

Making Measurements

HIOKI IM9202A961-00
Making Measurements

Measurement Procedure

You will need (besides the device and accessories):
• IM7580 series Impedance Analyzer ×1
• IM9200 Test Fixture Stand ×1
• IM9905 Calibration Kit ×1
• IM9906 Adapter ×1
• 3.5 mm connector torque wrench ×1 (not included with the device)
• Tweezers ×1 (This is not included with the device. Tweezers which are thinner than the DUT size, clean, and not magnetized are recommended.)
Connect the test head to theConnect the test head to the
1
impedance analyzer.impedance analyzer.
See the instruction manual
of the impedance analyzer.
Attach the test head and adapter Attach the test head and adapter
2
to the stand.to the stand.
Execute the calibration.Execute the calibration.
3
Connect the cable of the test head to the impedance analyzer.
Test head
Attach the test head and the IM9906 Adapter to the IM9200 Test Fixture Stand.
(p. 10)
IM9906
IM9200
Execute the open calibration, short calibration, and load calibration.
(p. 11)
4
5
6
8
Attach the device.Attach the device.
Execute the correction.Execute the correction.
Start the measurement.Start the measurement.
Attach the device to the IM9200 Test Fixture Stand.
Device (IM9202)
(p. 14)
Execute the electrical length correction, short correction, and open correction.
(From p. 15)
Connect the DUT to the measurement location and start the measurement by using the impedance analyzer.
(p. 22)
Page 11
Making Measurements
HIOKI IM9202A961-00

Inspecting the Device Before Use

To ensure that the device is properly operating, conduct an inspection and check device operation before using device to ensure that no damage has occurred during storage or transport. If there is any damage to the device, contact your authorized Hioki distributor or reseller for repair.
It is recommend to inspect the device prior to use as described below. Measure residual parameters with an impedance analyzer and verify that the measured values fall within the allowable range. See "Residual parameters (reference value)" (p. 24)
Check items Remedy
Are the electrode or the short plate worn?
Does either of the clips exhibit any damage or unusual appearance?
Is the device damaged? Send the device for repair if it exhibits any deformation in shape or
If any of the following applies, send the device for repair:
• Residual parameters’ measured values diverge from the allowable values.
• An electrode or the short plate exhibits wear.
If any of the following applies, send the device for repair:
• Either of the clips is cracked or worn.
• Either of the clips will not move.
• The clips are not able to accommodate a component.
other damage.
9
Page 12
Making Measurements
HIOKI IM9202A961-00

Attaching the Test Head and the Adapter

Attach the test head and the IM9906 Adapter to the IM9200 Test Fixture Stand.
IMPORTANT
The impedance analyzer, test head, and measurement cable are adjusted together at the time of shipment. Use the test head and measurement cable included along with the impedance
analyzer. Conrm that the serial number* of the impedance analyzer and test head is same.
* The 9-digit serial number indicates the year (rst two digits) and the month of manufacture (next
two digits). Do not remove this sticker as the number is important.
You will need:
• Test head ×1
• IM9200 Test Fixture Stand ×1
• IM9906 Adapter ×1
• 3.5 mm connector torque wrench ×1 (not included with the device)
11
Metal ttings
22
44
Test head
IM9200
IM9906
33
Measurement terminal
Pull the metal ttings to the IM9200 Test Fixture
1
Stand side.
Place the test head on the stand and tighten the
2
knob.
Place the notch of the IM9906 Adapter parallel to the
3
metal ttings, then install onto the measurement
terminal of the test head.
At this time it is stopped temporarily. Position the notch of the
IM9906 Adapter where the metal ttings can slide.
Fix the notch of the IM9906 Adapter by sliding the
4
metal ttings.
10
Notch
3.5 mm connector nut
55
Tighten the nut of the 3.5 mm connector by using a
5
torque wrench.
Recommended torque: 0.9 N·m Tightening the nut until the handle of the wrench bends slightly
is sufcient. Do not over tighten.
Page 13
Making Measurements
HIOKI IM9202A961-00

Calibration

Connect the three standard units, open, short, and load of the IM9905 Calibration Kit to the reference surface (terminal) of the IM9906 Adapter one by one, and measure the respective calibration data. This reference surface is referred to as the “calibration reference surface”. The cause of errors between the impedance analyzer itself and the calibration reference surface is eliminated.
The dened value differs according to the calibration kit to be used. For more information about how to input dened values to the impedance analyzer, see the instruction manual of the impedance analyzer and the Dened Value Setting Application Installation Procedure on the CD-R that came
with the IM9905.
CAUTION
While attaching and detaching the calibration kit, make sure that the internal conductor of the calibration kit does not rotate. If the internal conductor rotates, it may cause damage to the electrode.
When the calibration cannot be executed:
• Conrm that the calibration kit and the IM9906 Adapter are connected properly.
• Check that the DC measurement limit value of the impedance analyzer is appropriate.
• Connect the IM9906 adapter, calibration kit and the device with a 7 mm connector. Recommended torque for the 7 mm connector is 1.35 N·m.
You will need:
IM9905 Calibration Kit (OPEN, SHORT, and LOAD) ×1 of each
Set the measurement conditions like frequency,
1
measurement signal level, and measurement speed in the impedance analyzer.
Set the dened values of the calibration kit and offset
2
delay in the impedance analyzer.
The initial values of the impedance analyzer are set to the values of the following table.
OPEN G: 0 S, Cp: 82 fF 0 s
SHORT Rs: 0
LOAD Rs: 50
To use the IM9905’s LOAD dened value, use the
Value Setting Application
congure the impedance analyzer with the dened value.
Dened values Offset delay
Ls: 0 H 0 s
,
Ω
Ls: 0 H 0 s
,
Ω
Dened
that came with the IM9905 to
11
Page 14
Making Measurements
HIOKI IM9202A961-00
Screw thread
33
Nut
Open calibration kit
Nut
55
Raise the screw thread by turning the nut on the
3
IM9906 Adapter
Place the open calibration kit on the reference
4
surface.
As shown in gure, connect by turning only the nut
5
of open calibration kit while pressing the convex portion on the top part of the open calibration kit.
The convex part of the open calibration kit is connected with the internal conductor. If the calibration kit is rotated without pressing the convex part, the internal conductors of calibration kit and adapter may graze each other, and the calibration kit and adapter may get damaged.
Execute open calibration by using the impedance
6
analyzer.
33
11
44
22
Short calibration kit
Nut
Remove the open calibration kit by turning the nut of
7
the open calibration kit.
Place the short calibration kit on the reference
8
surface.
99
As shown in gure, connect by turning only the nut
9
of short calibration kit while pressing the convex portion on the top part of the short calibration kit.
The convex part of the short calibration kit is connected with the internal conductor. If the calibration kit is rotated without pressing the convex part, the internal conductors of calibration kit and adapter may graze each other, and the calibration kit and adapter may get damaged.
12
Page 15
Making Measurements
HIOKI IM9202A961-00
Execute short calibration by using the impedance
10
analyzer.
33
11
44
22
Load calibration kit
Nut
1313
Remove the short calibration kit by turning the nut of
11
the short calibration kit.
Place the load calibration kit on the reference surface.
12
As shown in gure, connect by turning only the nut
13
of load calibration kit while pressing the top part of the load calibration kit.
If the top part of the load calibration kit is rotated, the internal conductors of the calibration kit and adapter may graze each other, and the calibration kit and adapter may get damaged.
Execute load calibration by using the impedance
14
analyzer.
33
11
44
22
Remove the load calibration kit by turning the nut of
15
the load calibration kit.
13
Page 16
Making Measurements
HIOKI IM9202A961-00

Attaching the Device

Attach the device to the IM9200 Test Fixture Stand and adapter.
22
33
22
Screw thread
Nut
11
Lower the screw thread by turning the nut of the
1
IM9906 Adapter
Loosen the stand xing knob.
2
Place the device on the IM9200 Test Fixture Stand.
3
As shown in gure, slide the notch part.
55
55
44
Turn the nut of the IM9906 Adapter and connect to the
4
device.
Tighten the stand xing knob.
5
14
Page 17
Making Measurements
HIOKI IM9202A961-00

Electrical Length Correction

The error, caused by the phase shift occurring between the calibration reference surface and the measurement DUT's connection surface is compensated. Set the electrical length correction to 0 mm in the impedance analyzer.
22
11
33
44
15
Page 18
Making Measurements
HIOKI IM9202A961-00

Short Correction

Short correction is performed to eliminate the device’s residual short impedance. The short plate is used.
CAUTION
To prevent damage to the screw in the center of the short plate, orient the screw perpendicularly before insertion and tighten with an appropriate amount of force.
About short impedance
At high frequencies, a DUT’s structure and dimensions exert a signicant effect on impedance
values.
DUTs have a variety of shapes and dimensions, and it is not possible to precisely dene a short
impedance that will accommodate all of those possibilities. Although short correction is performed
to correct for the residual impedance of the test xture, the fact that short impedance is undened
means measured values will incorporate an error component.
There are two methods for performing short correction of the test xture.
• Using a short device with the same shape as the DUT The value after short correction is the difference in impedance between the DUT and the short device. The short device’s impedance will affect measured values.
Using a xed reference without regard to the size of the DUT The value after short correction is the sum of the DUT’s impedance and the impedance caused by
the effects of the xture’s electrodes. The impedance of the xture’s electrodes affects measured
values.
The device use a xed reference to perform short correction without regard to the size of the DUT. Consequently, measured values include the impedance of part of the xture’s base plate and of the L-side electrode. Since a deterioration in the xture’s residual impedance will affect measured
values, it is recommended to check the residual impedance on a daily basis.
You can also prepare your own short device with the same shape as the DUT you plan to measure and use it to perform short compensation.
However, be sure to use the same correction method when comparing measured values.
IMPORTANT
Do not scratch the electrode and short plate of the device or touch them directly with your hand. It may affect the measurement accuracy and reproducibility.
16
Page 19
You will need:
HIOKI IM9202A961-00
• Flat-head screwdriver ×1 (not included with the device)
• Short plate ×1
• Tweezers ×1 (This is not included with the device. Tweezers which are thinner than the DUT size, clean, and not magnetized are recommended.)
Making Measurements
Set the reference value of correction (COMPEN) in
1
the impedance analyzer.
By default, the impedance analyzer is congured using the
values shown in the following table. Use the device’s short
correction and open correction dened values shown in the
following table.
OPEN G: 0 S, Cp: 0 F
SHORT Rs: 0
,
Ω
Ls: 0 H
Electrodes and clips for lead component measurement
Knob
L-side lead component electrode
Electrodes for SMD measurement
L-side SMD electrode
Base plate
H-side lead component clip
Screws
Knob
Base plate
Knob
Short plate
(If electrodes and clips for lead component
2
measurement have been installed)
Remove the H-side lead component clip and L-side lead component electrode from the base plate.
Using a screwdriver, remove the H-side lead component clip. Turn the knob to remove the L-side lead component electrode.
(If the SMD measurement electrodes have been installed)
Remove the L-side SMD electrode from the base plate.
Turn the knob and remove the L-side SMD electrode.
Using a at-head screwdriver, install the short plate.
3
Tighten the short plate’s screw to the correct position. If the screw catches partway, do not force it, but rather back it out and then attempt to tighten it again after inserting it in a perpendicular orientation.
Screw
View from beside the plate
Screw in correct position
Partially tightened screw
17
Page 20
Making Measurements
HIOKI IM9202A961-00
Execute short correction by using the impedance
4
analyzer.
When the correction cannot be executed:
Conrm that the IM9906 Adapter is connected properly.
• Check that the DC measurement limit value of the impedance analyzer is appropriate.
22
11
55
33
44
Using a at-head screwdriver, remove the short
5
plate.
66
18
Page 21
Making Measurements
HIOKI IM9202A961-00

Open Correction

Open correction is performed to eliminate the device’s residual open impedance. Open correction is performed by adjusting the electrode width based on the size of the DUT.
You will need:
• Phillips screwdriver ×1
• Tweezers ×1 (This is not included with the device. Tweezers which are thinner than the DUT size, clean, and not magnetized are recommended.)
Measuring lead components
Using a phillips screwdriver, install the H-side lead
1
component clip.
Gently tighten the knob that holds the L-side lead
2
component electrode in place.
22
11
33
55
Marker
Edge
DUT width
44
H-side electrode
Slide the L-side lead component electrode all the
3
way to the H-side electrode.
Check where the edge of the H-side electrode falls on the marker located on upper right to ascertain the DUT’s width.
Slide the L-side lead component electrode to the
4
left until the distance between the edge of the moving plate and the edge of the marker is equal to the DUT width.
Lock the L-side lead component electrode in place
5
with the knob.
19
Page 22
Making Measurements
HIOKI IM9202A961-00
Start the open correction in the impedance
6
analyzer.
22
11
33
44
Measuring SMDs
11
(4) (3) (2) (1)
55
When correction cannot be executed, check that the DC measurement limit value of the impedance analyzer is appropriate.
Choose the best L-side SMD electrode lock
1
position from (1) through (4) as shown in the
gure on the left based on the SMD’s dimensions
(width).
H-side electrode
the L-side SMD electrode
66
Lock position for
(1) 1.6 mm to 7 mm
SMD dimensions
(width)
20
33
22
(2) 7 mm to 12.5 mm
(3) 12.5 mm to 18 mm
(4) 18 mm to 23 mm
Install the L-side SMD electrode while pressing
2
down on and turning the knob.
Pull the lever on the L-side SMD electrode toward
3
you to align the DUT interval with the electrode width.
Page 23
Making Measurements
HIOKI IM9202A961-00
Leading edge of the L-side SMD electrode
Protruding part
SMD open
44
Gap between the top surface of the electrode and the plate
View from the side with the jig installed
SMD open correction jig
correction jig
Protruding part
Install the SMD open correction jig and lock the
4
electrode in place.
Firmly push the protruding part of the SMD open correction jig into the gap between the top surface of the L-side SMD electrode and the plate. Catch the other protruding part on the leading edge of the L-side SMD electrode.
If you’re unable to secure electrode position
sufciently, you’ll need to replace the SMD open
correction jig. Please contact your authorized Hioki distributor or reseller
.
Start the open correction in the impedance
5
analyzer.
22
11
55
33
44
When correction cannot be executed, check that the DC measurement limit value of the impedance analyzer is appropriate.
66
Remove the SMD open correction jig.
6
21
Page 24
Making Measurements
HIOKI IM9202A961-00

Starting Measurement

Lock the DUT in place and start measurement.
You will need:
• Tweezers ×1 (This is not included with the device. Tweezers which are thinner than the DUT size, clean and not magnetized are recommended.)
• DUT ×1
Lead components
11
Open the clip on the L-side lead component electrode
1
and the H-side lead component clip and insert the
component’s lead wires into the electrodes.
Start the measurement using the impedance analyzer.
2
SMDs
IMPORTANT
To boost measurement reproducibility, use the same component position, orientation, and lead connection locations when measuring components with the same shape.
CAUTION
Do not measure components that would readily deform in shape or suffer damage when subjected to force. Doing so may cause damage the component.
Operate the lever on the L-side electrode to sandwich
1
the component between the H-side electrode and the L-side SMD electrode.
11
Verify that the edge of the L-side SMD electrode is in contact with the base plate. Proper measurement cannot be performed
if the edge of the electrode is oating relative to the base plate.
Start the measurement using the impedance analyzer.
2
22
IMPORTANT
• To boost measurement reproducibility, use the same component position and orientation
when measuring components with the same shape.
• If the L-side SMD electrode is exhibiting unstable contact, the SMD open correction jig can
be used to improve contact.
• The impedance of some DUTs varies with the amount of force exerted on the DUT. Reducing
the amount of force applied between the xture’s electrodes and the DUT’s electrodes can
lead to increased measurement error by increasing the contact resistance between the electrodes. Adjust the force and contact after assessing how the component in question will be affected by force.
Page 25
Specications
HIOKI IM9202A961-00
General Specications
Operating environment Indoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
Operating temperature and humidity range
Storage temperature and humidity range
Dimensions Approx. 135W × 82H × 37D mm (5.31″W × 3.23″H × 1.46″D)
Mass Approx. 190 g (6.7 oz.)
Product warranty period 3 years
Accessories and options p. 2
Measurement Specications
Basic Specications
According to the operating temperature and humidity, and accuracy guarantee for temperature and humidity of connectable measuring instrument.
−10°C to 50°C (14°F to 122°F), 80% RH or less (non-condensing)
Depth varies with the electrodes in use.
Moving parts and parts that make contact with DUTs are not covered by the warranty.
Specications
Connectable measuring instruments
Connecting part to the DUT
Dimensions of measurable DUT
Measurable range (Z, θ) According to the measurable range of connectable measuring instrument.
Used frequency range DC to 600 MHz
Maximum applied voltage ±42 V peak
Electrical length 0 mm
Connector 7 mm connector
IM7580, IM7580A, IM7581, IM7583, IM7585, IM7587 (as of Jan. 2020) Connectable measuring instruments are subject to change. Please check Hioki’s website for the latest information.
2-terminal structure Lead components: The lead component’s leads are pinched between the device’s electrodes and clips. SMDs: The SMD is pinched between the device’s electrodes.
Lead components
Axial
Lead interval (component length): 1 mm to 25 mm Lead length: 2 mm to 10 mm Height to lead: 2.5 mm or shorter
(The component’s H-side lead must not make contact with the xture’s L-side
electrode.) Lead wire diameter:
Radial
Lead interval: Lead length: (
The electrode on the bottom surface of the H side of the component must not make contact with the fixture’s L Lead wire diameter: 0.5 mm to 2 mm
SMD
Component length: 1.6 mm to 23 mm Component width: 0.8 mm or longer Component height: 0.65 mm or longer (The component’s H-side lead must not make contact with the fixture’s L-side electrode.)
0.5
mm to 2 mm
2
mm to 26 mm
2
mm or longer
-
side electrode.
)
23
Page 26
Specications
HIOKI IM9202A961-00
Residual parameters (reference value)
Residual parameters Short inductance (Short plate)
Residual parameters Short inductance (L-side lead component
electrode)
Residual parameters Short inductance (L-side SMD electrode)
Conditions
• Inductance value as seen from the 7 mm connector calibration surface (Electrical Length Correction 0 mm) Measurement parameter Ls
• Short plate connection
1.6 nH ±1.1 nH: 30 MHz
1.6 nH ±0.8 nH: 100 MHz
1.6 nH ±0.7 nH: 300 MHz, 600 MHz
Conditions
• Inductance value as seen from the 7 mm connector calibration surface (Electrical Length Correction 0 mm) Measurement parameter Ls
• The L-side lead component electrode is directly connected to the H-side electrode.
• The H-side clip is present.
1.2 nH ±1.1 nH: 30 MHz
1.2 nH ±0.8 nH: 100 MHz
1.2 nH ±0.7 nH: 300 MHz, 600 MHz
Conditions
• Inductance value as seen from the 7 mm connector calibration surface (Electrical Length Correction 0 mm) Measurement parameter Ls
• The L-side SMD electrode is directly connected to the H-side electrode.
1.2 nH ±1.1 nH: 30 MHz
1.2 nH ±0.8 nH: 100 MHz
1.2 nH ±0.7 nH: 300 MHz, 600 MHz
Residual parameters Open capacitance
Conditions
• Capacitance value as seen from the 7 mm connector calibration surface (Electrical Length Correction 0 mm) Measurement parameter Cp
• Neither the L-side electrode nor the H-side clip is present.
1.3 pF ± 0.3 pF (30 MHz)
1.3 pF ± 0.2 pF (100 MHz, 300 MHz, 600 MHz)
24
Page 27

Maintenance and Service

HIOKI IM9202A961-00
CAUTION
Do not attempt to modify, disassemble, or try to repair the device. Doing so may cause
a re, electric shock, or injury.
Storage
If you do not intend to use the device for an extended period, store it in the carrying case.
Cleaning
• If the device becomes dirty, wipe the device clean with a soft cloth slightly moistened with water or a neutral detergent.
• The measurement accuracy and reproducibility of the device’s electrode and short plate may decrease due to dirt or deformation. If dirt sticks to the device, remove it by using the provided cleaning brush.
• Remove the H-side lead component clip, the L-side lead component electrode, the L-side SMD electrode, and the short plate. Gently brush the H-side electrode and base plate.
Maintenance and Service
Moving parts service life
• The performance of some parts used in the device may degrade during the course or extended use. When repair is necessary, please contact your authorized Hioki distributor or reseller.
• Service life varies with the operating environment and frequency of use. Please note that performance is not guaranteed for the duration of the recommended replacement cycle.
Parts Recommended
replacement cycle
Base plate 10000 times Number of times the L-side lead component
electrode and L-side SMD electrode have slid back and forth
H-side lead component clip 10000 times Number of open/close cycles
L-side lead component electrode
L-side lead component clip 10000 times Number of times slid back and forth over base
10000 times Number of times slid back and forth over base
plate
plate
Conditions
Troubleshooting
If the device seems to be malfunctioning, contact your authorized Hioki distributor or reseller.
Transporting
Be sure to follow these precautions:
• To avoid damaging the device, store it in the carrying case with the H-side lead component clip and the L-side lead component electrode connected to the base plate and with the electrode locked in place with the knob. Remove the short plate and L-side SMD electrode.
• Keep the removed accessories in the original packing materials in which it was shipped, and store in the carrying case. Additionally, be sure to double box the device. Accidental damage suffered in transit is not covered by the warranty.
• Attach a description of the issue when sending out your device for repair.
Disposal
Dispose of the device in accordance with local regulations.
25
Page 28
Maintenance and Service
HIOKI IM9202A961-00
26
Page 29
HIOKI IM9202A961-00
Page 30
HIOKI IM9202A961-00
Page 31
HIOKI IM9202A961-00
Page 32
HIOKI IM9202A961-00