Hioki IM9201 Instruction Manual

IM9201
Instruction Manual
SMD TEST FIXTURE
July 2016 Edition 1 IM9201A961-00 16-07H
EN

Introduction

Thank you for purchasing the Hioki IM9021 SMD Test Fixture. To obtain maximum performance
from the device, please read this manual rst, and keep it handy for future reference.
This device is used by connecting to the impedance analyzer. Before using the impedance analyzer, read the instruction manual of the impedance analyzer carefully.
1

Contents

Verifying Package Contents Safety Information Operating Precautions Product Overview Names and Functions of Parts Measuring Method
■ Measurement Procedure
■ Inspection Before Use
■ Attaching the Test Head and the Adapter
■ Calibration
■ Attaching the Device
■ Electrical Length Correction
■ Short Correction
■ Open Correction
p.2 p.3 p.4 p.5 p.6 p.7 p.7 p.8 p.9 p.10 p.13 p.14 p.15 p.18
2 3 4 5 6
■ Starting the Measurement
Specications
Maintenance and Service
p.20
7
p.21 p.22
8 9
Appx. Index
IM9201A961-00
1

Verifying Package Contents

Verifying Package Contents
When you receive the device, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories. If damage is evident, or if it fails to operate according
to the specications, contact your authorized Hioki distributor or reseller.
Conrm that the following items are provided.
(1)
(3) (4)
Model IM9201 SMD Test Fixture × 1 ..............................................................................(1)
Short plate × 5 ................................................................................................................
GND plate × 1 Device guide × 1
Hexagon socket head bolts (spare, M1.4 × 4 mm) × 4...................................................
Knob (for xing tip pin) × 1
Tip pin × 2
Brush for cleaning × 1 ....................................................................................................(4)
Hex driver × 1 .................................................................................................................(5)
Carrying case × 1 ...........................................................................................................(6)
Instruction Manual × 1
(6)
(5) (11)(2)
(9)(8)(10)(7)
(2)
(3)
GND plate and device guide for the sizes 1005 (0402), 1608 (0603), 2012 (0805), 3216 (1206), 3225 (1210) are installed in the device. Refer to "Tip pins, device guides, and plates for each DUT size" (p.5).
Options (sold separately)
Model IM9906 Adapter (3.5 mm/7 mm) ..........................................................................(7)
Model IM9905 Calibration Kit (LOAD) ............................................................................(8)
Model IM9905 Calibration Kit (OPEN) ............................................................................(9)
Model IM9905 Calibration Kit (SHORT) .........................................................................(10)
Magnifying glass (Model IM9200 Test Fixture Stand accessory) ...................................(11)
Model IM9200 Test Fixture Stand
Options (sold separately) can be stored inside (7) to (11) of the carrying case. Please store them after removing the packed sponge at the time of unpacking.
2

Safety Information

Before using the device, be certain to carefully read the following safety notes.
CAUTION
• Mishandling during use could result in damage to the device. Be certain that you understand the instructions and precautions in the manual before use.
• Individuals using an electrical measuring device for the rst time should be supervised
by a technician who has experience in electrical measurement.
Safety Information
1 2
Notation
In this document, the risk seriousness and the hazard levels are classied as follows.
WARNING
CAUTION
IMPORTANT
Symbols for standards
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Indicates a potentially hazardous situation that may result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the device or malfunction of the device.
Indicates information related to the operation of the device or
maintenance tasks with which the operators must be fully familiar.
Indicates prohibited action.
Indicates the action which must be performed.
3 4 5 6 7
Indicates that the product conforms to regulations set out by the EC Directive.
8 9
Appx. Index
3

Operating Precautions

Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benets of the various
functions.
Installation Environment
WARNING
Installing the device in inappropriate locations may cause malfunctions or may give rise to accidents. Avoid the following locations:
• Exposed to direct sunlight or high temperature
• Exposed to corrosive or combustible gases
Exposed to strong electromagnetic elds or electrostatic charges
• Near induction heating systems (such as high-frequency induction heating systems and IH cooking equipment)
• Susceptible to mechanical vibration
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity and condensation
• Exposed to high quantities of dust particles
Handling the Device
• Do not press the tip pin in the electrode when the DUT is not attached.
• While pressing the tip pin into the DUT, the DUT may pop out. To avoid injuries, lower
the pusher slowly and press the tip pin into the DUT.
• To avoid damage to the device, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
• To prevent damage to the device, use the bolts included in the factory shipment for
securing the pusher and plate. If you have lost any bolts or nd any damaged bolts,
please contact your authorized Hioki distributor or reseller for a replacement.
Measurements
Do not input a voltage in excess of the maximum applied voltage listed in the
specications. It may cause damage to the device or connecting equipments.
To avoid damage, do not connect the charged capacitor to the device.
CAUTION
CAUTION
Precautions during shipment
During shipment of the device, handle it carefully so that it is not damaged due to a vibration or shock.
4

Product Overview

0603
3225 3216
1005 1608
2012
3225 3216
1005 1608
2012
3225
3216
1005 1608
2012
3225 3216
1005 1608
2012
Product Overview
This device is the test xture for SMD (Surface Mount Device) to be used by connecting to the
impedance analyzer of Hioki. It measures the 2 terminals of SMD.
Measurable DUT size
1
Nominal size [mm]
JIS (EIA)
0603 (0201) (0.6±0.03) × (0.3±0.03) H ≥ 0.25 1005 (0402) (1.0±0.1) × (0.5±0.1) H ≥ 0.4 1608 (0603) (1.6±0.15) × (0.8±0.15) H ≥ 0.4 2012 (0805) (2.0±0.15) × (1.2±0.15) H ≥ 0.4 3216 (1206) (3.2±0.15) × (1.6±0.15) H ≥ 0.4 3225 (1210) (3.2±0.15) × (2.5±0.15) H ≥ 0.4
L × W [mm] H [mm]
L
H
W
Tip pins, device guides, and plates for each DUT size
Dimensions
2 3 4 5 6
DUT size
JIS (EIA)
Accessories
Tip pin
Device guide
0603 (0201) 1005 (0402) 1608 (0603) 2012 (0805)
(Common)
3216 (1206) 3225 (1210)
7 8
(Common)
9
Short plate (At the time of
short correction)
GND plate (At the time of open
correction and measurement)
0603
2012
3225 3216
1005 1608
(Common)
Appx. Index
5

Names and Functions of Parts

Names and Functions of Parts
(9)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(10)
(11)
(12)
(11)
(13)
(1) Pusher Presses and xes the DUT. (2) Tip pin The tip of the pusher. Replaced according to the DUT to be measured. (3) Shock absorber Absorbs the shock when the pusher comes down. (4) Device guide Placed on the top of the GND plate or short plate for use. (5) Short plate or
GND plate
(6) Electrode section The measurement unit of the device. Measures by placing the GND plate
(7) Model IM9906 Adapter
(3.5 mm/7 mm) (8) Test head The test head is provided with the impedance analyzer. (9) Loupe The magnifying glass provided with the model IM9200 Test Fixture Stand
(10) Knob The knob is loosened to move the tip pin to the DUT location. Then the
(11) Stand xing knob The knob xes the device to the model IM9200 Test Fixture Stand. (12) Stopper Locks the pusher so that it does not come down. (13) Model IM9200 Test
Fixture Stand
Short plate is used at the time of short correction. GND plate is used at the time of open correction and measurement.
and DUT on top of the electrode. The exclusive adapter for connecting the device to the test head of the
impedance analyzer side.
(option). It is used by installing to the model IM9200.
knob is tightened to x it.
The exclusive stand for xing the test head of the impedance analyzer and
the device.
6

Measuring Method

Measurement Procedure

You will need (besides the device and
• Model IM7580 series Impedance Analyzer × 1
• Model IM9200 Test Fixture Stand × 1
• Model IM9906 Adapter × 1
• 3.5 mm Connector torque wrench × 1 (This is not provided with the device.)
• Tweezers × 1 (This is not provided with the device. Tweezers which are thinner than the DUT size, clean, and not magnetized are recommended.)
accessories
Measuring Method
1
):
2
Connect the test head to the
1
impedance analyzer.
See the instruction manual of the impedance analyzer.
Attach the test head and adapter
2
to the stand.
Execute the calibration.
3
Attach the device.
4
(p.9)
(p.10)
(p.13)
Connect the cable of the test head to the impedance analyzer.
Test head
Attach the test head and model IM9906 Adapter to the model IM9200 Test Fixture Stand.
Model IM9906 Model IM9200
Execute the open calibration, short calibration, and load calibration.
Attach the device to the model IM9200 Test Fixture Stand.
Device (model IM9201)
3 4 5 6 7 8
Execute the correction.
5
Start the measurement.
6
(From p.14)
(p.20)
Execute the electrical length correction, short correction, and open correction.
Connect the DUT to the measurement location of the device guide and start the measurement by the impedance analyzer.
9