After using the instrument, remove the test sample and turn OFF the power supply.
Calibration/compensation execution timing
• Before measurements
• After the length of measurement cable is changed
• After the type of measurement sample is changed
• After the xture is changed
HIOKI IM7585A981-05
Introduction
HIOKI IM7585A981-05
Introduction
Thank you for purchasing the HIOKI IM7580A, IM7581, IM7583, IM7585, IM7587 Impedance Analyzer.
To obtain maximum performance from the instrument, please read this manual rst, and keep it handy for
future reference.
The screen display has been explained using Model IM7585 as an example.
Model Information
ModelMeasurement frequencyCable length
1
2
Rear
(Example: IM7585)
IM7580A-1
1 MHz to 300 MHz
IM7580A-22 m
IM7581-01
100 kHz to 300 MHz
IM7581-022 m
IM7583-01
1 MHz to 600 MHz
IM7583-022 m
IM7585-01
1 MHz to 1.3 GHz
IM7585-022 m
IM7587-01
1 MHz to 1.3 GHz
IM7587-022 m
1 m
1 m
1 m
1 m
1 m
Verifying Package Contents
• When you receive the instrument, inspect it carefully to ensure that no damage occurred during
shipping.
In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to
operate according to the specications, contact your authorized Hioki distributor or reseller.
• Store the packaging in which the instrument was delivered, as you will need it when transporting the
instrument.
3
4
5
6
7
Conrm that the following items have been provided:
• Measurement cable ×1
(IM7580A-1: 1 m, IM7580A-2: 2 m,
IM7581-01: 1 m, IM7581-02: 2 m,
IM7583-01: 1 m, IM7583-02: 2 m,
IM7585-01: 1 m, IM7585-02: 2 m,
IM7587-01: 1 m, IM7587-02: 2 m)
IM7580A, IM7581IM7583, IM7585, IM7587
Instruction Manual ×1Power cord ×1
Impedance Analyzer Application Disc ×1
(Communications user manual [PDF],
communications commands manual,
USB driver, sample application, and initial
settings table)
• The latest version can be downloaded
from our website.
8
9
10
Appx.Ind.
1
Options (Sold Separately)
HIOKI IM7585A981-05
Options (Sold Separately)
Contact your authorized Hioki distributor or reseller when ordering.
The options are subject to change. Visit our website for updated information.
Test xtures
IM9200Test Fixture Stand
IM9201SMD Test Fixture (for SMDs)
IM9202Test Fixture (for through-hole components and SMDs)
IM9906Adapter (3.5mm/7mm)
IM9905Calibration Kit
Interfaces
Z3000GP-IB Interface
Z3001RS-232C Interface
Connection cables
9151-02GP-IB Connector Cable (2 m)
9637RS-232C Cable (9pin-9pin/1.8m)
Safety Information
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested
for safety prior to shipment. However, using the instrument in a way not described in this manual may
negate the provided safety features.
Before using the instrument, be certain to carefully read the following safety notes.
DANGER
Mishandling during use could result in injury or death, as well as damage to the
instrument. Be certain that you understand the instructions and precautions in
the manual before use.
2
WARNING
With regard to the electricity supply, there are risks of electric shock, heat
generation, re, and arc discharge due to short circuits. If persons unfamiliar
with electricity measuring instrument are to use the instrument, another person
familiar with such instruments must supervise operations.
Notation
HIOKI IM7585A981-05
In this manual, the risk seriousness and the hazard levels are classied as follows.
Safety Information
DANGER
WARNING
CAUTION
IMPORTANT
*
Symbols on the instrument
Indicates an imminently hazardous situation that will result in death or serious injury to
the operator.
Indicates a potentially hazardous situation that may result in death or serious injury to
the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury
to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance tasks
with which the operators must be fully familiar.
Indicates a high voltage hazard.
If a particular safety check is not performed or the instrument is mishandled, this may
give rise to a hazardous situation; the operator may receive an electric shock, may get
burnt or may even be fatally injured.
Indicates prohibited actions.
Indicates the action which must be performed.
Additional information is presented below.
1
2
3
4
5
Indicates cautions and hazards. When the symbol is printed on the instrument, refer to a
corresponding topic in the Instruction Manual.
Indicates AC (Alternating Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Symbols for standards
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU
member states.
Indicates that the product conforms to regulations set out by the EU Directive.
6
7
8
9
10
Appx.Ind.
3
Safety Information
HIOKI IM7585A981-05
Accuracy
We dene measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values,
with the following meanings:
(maximum display value or scale length)
f.s.
The maximum displayable value or scale length. This is usually the name of the currently
selected range.
rdg.
dgt.
(reading or displayed value)
The value currently being measured and indicated on the measuring instrument.
(resolution)
The smallest displayable unit on a digital measuring instrument, i.e., the input value that
causes the digital display to show a “1” as the least-signicant digit.
Measurement categories
To ensure safe operation of measurement instruments, IEC 61010 establishes safety standards
for various electrical environments, categorized as CAT II to CAT IV, and called measurement
categories.
DANGER
• Using a measuring instrument in an environment designated with a higher-
numbered category than that for which the instrument is rated could result in a
severe accident, and must be carefully avoided.
• Using a measuring instrument without categories in an environment designated
with the CAT II to CAT IV category could result in a severe accident, and must
be carefully avoided.
CAT II: When directly measuring the electrical outlet receptacles of primary electrical circuits
in equipment connected to an AC electrical outlet by a power cord (portable tools,
household appliances, etc.).
CAT III: When measuring the primary electrical circuits of heavy equipment (xed installations)
connected directly to the distribution panel, and feeders from the distribution panel to
outlets.
CAT IV: When measuring the circuit from the service drop to the service entrance, and to the
power meter and primary overcurrent protection device (distribution panel).
4
Service entrance
Service drop
CAT IV
Power meter
Distribution panel
Internal wiring
CAT III
Fixed installations
CAT II
T
Outlet
Operating Precautions
HIOKI IM7585A981-05
Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benets of the various functions.
WARNING
If the measurement cable or the instrument is damaged, there is a risk of electric
shock. Perform the following inspection before using the instrument:
• Before using the instrument, make sure that the insulation on the measurement
cables are undamaged and that no bare conductors are improperly exposed. If
the instrument is damaged, contact your authorized Hioki distributor or reseller.
• Before using the instrument for the rst time, verify that it operates normally
to ensure that no damage occurred during storage or shipping. If you nd any
damage, contact your authorized Hioki distributor or reseller.
CAUTION
Avoid using an uninterruptible power supply (UPS) or DC/AC inverter with rectangular
wave or pseudo-sine-wave output to power the instrument. Doing so may damage the
instrument.
Instrument installation
For details on the operating temperature and humidity, refer to the specications (p. 275).
1
2
3
4
5
Installing the instrument in inappropriate locations may cause a malfunction of
instrument or may give rise to an accident. Avoid the following locations.
• Exposed to direct sunlight or high temperature
• Exposed to corrosive or combustible gases
• Exposed to a strong electromagnetic eld or electrostatic charge
• Near induction heating systems (such as high-frequency induction heating
systems and IH cooking equipment)
• Susceptible to vibration
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
Installation method
Example: IM7585
50 mm or more
WARNING
To prevent overheating, be sure to leave the specied
clearances around the instrument.
• Install the instrument with the bottom facing down.
• Vents must not be obstructed.
• A distance of 50 mm or more must be maintained between the
rear and the surroundings.
6
7
8
9
10
Appx.Ind.
5
Operating Precautions
HIOKI IM7585A981-05
Warranty
Hioki disclaims responsibility for any direct or indirect damages that may occur if this instrument has been
combined with other devices by a systems integrator prior to sale or during resale. Please note.
Handling the instrument
DANGER
• To avoid electric shock, do not remove the instrument’s case. The internal
components of the instrument carry high voltages and may become very hot
during operation.
• Do not allow the instrument to get wet, and do not take measurements with wet
hands. This may cause an electric shock.
CAUTION
• If the instrument exhibits abnormal operation or display during use, review the
information given in “13.3 Troubleshooting” (p. 305) and “13.4 Error Display” (p. 310)
before contacting your authorized Hioki distributor or reseller. Do not connect charged
capacitors, input voltage or current to the measuring terminals. The instrument will get
damaged.
• To avoid damage to the instrument, protect it from physical shock when transporting
and handling. Be especially careful to avoid physical shock from dropping.
• This instrument is not designed to be entirely water- or dust-proof. Do not use it in an
especially dusty environment, nor where it might be splashed with liquid. This may
cause damage.
• Do not use excessive force on the touch panel, and do not use hard or sharp objects
that could damage the touch screen.
• Do not apply heavy downward pressure with the stand extended. The stand could be
damaged.
• To avoid damage to the instrument, do not short-circuit the terminal/output terminal
and do not input voltage to the terminal/output terminal.
• After use, always turn OFF the power.
IMPORTANT
This instrument may cause interference if used in residential areas. Such use must be avoided
unless the user takes special measures to reduce electromagnetic emissions to prevent
interference to the reception of radio and television broadcasts.
Before turning ON the power
• Before turning the instrument on, make sure the supply voltage matches that
indicated on its power connector. Connection to an improper supply voltage
may damage the instrument and present an electrical hazard. Be careful to
avoid connecting the supply voltage improperly. Doing so may damage the
instrument’s internal circuitry. To avoid electrical accidents and to maintain the
safety specications of this instrument, connect the power cord provided only
to an outlet.
6
WARNING
Handling of cords and xtures
HIOKI IM7585A981-05
Use only the designated power cord with this instrument. Using other power cords may
cause re.
Operating Precautions
WARNING
1
• To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it
• To avoid breaking the cables or probes, do not bend or pull them.
• Bare conductors may get exposed if the insulation melts. Keep the cables well away
• Keep in mind that, in some cases, conductors to be measured may be hot.
• Use only the specied measurement cables. Using a non-specied measurement cable
may result in incorrect measurements due to poor connection or other reasons.
• Read the instruction manual supplied with the product to be used before using a xture.
CD precautions
• Exercise care to keep the recorded side of discs free of dirt and scratches. When writing
text on a disc’s label, use a pen or marker with a soft tip.
• Keep discs inside a protective case and do not expose to direct sunlight, high
temperature, or high humidity.
• Hioki is not liable for any issues your computer system experiences in the course of
using this disc.
CAUTION
2
from the power outlet.
3
from heat sources.
4
5
6
7
Before connecting to the EXT I/O terminals
WARNING
To avoid electric shock or damage to the equipment, always observe the following
precautions when connecting to the EXT I/O connector.
• Always turn OFF the power to the instrument and to any devices to be
connected before making connections.
• Be careful to avoid exceeding the ratings of EXT I/O connector (p. 216).
• During operation, a wire becoming dislocated and contacting another
conductive object can be a serious hazard. Use screws to secure the EXT I/O
connectors.
• Ensure that devices and systems to be connected to the EXT I/O terminals are
properly isolated.
• The ISO_5V pin of the EXT I/O connector is a 5V power output. Do not apply
external power to this pin.
8
9
10
Appx.Ind.
7
Operating Precautions
HIOKI IM7585A981-05
USB ash drives
• Hioki cannot recover or analyze data from damaged or faulty storage media. We
cannot compensate for such data loss, irrespective of the contents or cause for the
failure or damage. We recommend you to make a backup of all important data in a
computer or other devices.
• Avoid inserting the USB ash drive with the wrong orientation. This can damage the
USB ash drive or instrument.
• When a USB ash drive is being accessed, the color of the USB icon changes from
blue to red. Do not turn off the power of the instrument while the USB ash drive is
being accessed. Also, do not remove the USB ash drive from the instrument while it
is being accessed. This may result in the loss of data stored in the USB ash drive.
• Do not transport the instrument while a USB ash drive is connected. Damage could
result.
• Some USB ash drive are susceptible to static electricity. Exercise care when using
such products because static electricity could damage the USB ash drive or cause
malfunction of the instrument.
• With some USB ash drives, the instrument may not start up if power is turned on
while the USB ash drive is inserted. In such a case, turn the power on rst, and then
insert the USB ash drive. We recommend that various operations such as copy and
save be carried out with the USB ash drive before using it for actual measurements.
CAUTION
USB ash drives have a limited usable lifetime. Data reading and writing will fail after long-term
use. Replace the USB ash drive in this case.
Input modules (optional)
Always turn both devices OFF when connecting and disconnecting an interface
connector. This may cause an electric shock.
• To connect or disconnect optional interfaces, hold the metal part. Touching the PCB
with bare hands could damage the instrument due to static electricity. (Antistatic wrist
strap is recommended when disconnecting the interface.)
• To avoid equipment failure, do not disconnect the communications cable while
communications arein progress.
• Use a common ground for both the instrument and the computer. Using different
ground circuits will resultin a potential difference between the instrument’s ground and
the computer’sground. If the communications cable is connectedwhile such a potential
difference exists, it may result inequipment malfunction or failure.
• Before connecting or disconnecting any communications cable, always turn off the
instrument and the computer. Failure to do so could result in equipmentmalfunction or
damage.
• After connecting the communications cable, tightenthe screws on the connector
securely. Failure to secure theconnector could result in equipment malfunction or
damage.
WARNING
CAUTION
8
1
HIOKI IM7585A981-05
Overview
1.1 Overview and Features
The HIOKI IM7580A, IM7581, IM7583, IM7585, IM7587 Impedance Analyzer is an impedance measuring
instrument that has achieved high speed and high accuracy.
The IM7585 combines the functionality of two devices: an impedance analyzer that can perform
measurement while sweeping the measurement frequency and signal level, and an LCR meter that can
simultaneously display up to four parameters under a single set of measurement conditions.
A wide range of measurement conditions can be set, and the instrument can be used for a wide range of
applications such as measurement of high frequency inductors.
ModelMeasurement frequencySignal level
Wide range of
measurement
conditions
IM7580A-1
IM7580A-2
IM7581-01
IM7581-02
IM7583-01
IM7583-02
IM7585-01
IM7585-02
1 MHz to 300 MHz
100 kHz to 300 MHz
1 MHz to 600 MHz
1 MHz to 1.3 GHz
-40.0 dBm to +7.0 dBm
-40.0 dBm to +1.0 dBm
1
Overview
IM7587-01
IM7587-02
Fast measurementThe measurement speed is 0.6 ms (typical value) at the fastest.
The measurement frequency and sweep function of measurement level
Graph display
Equivalent circuit
analysis
CONTINUOUS
measurement mode
measures and displays the frequency characteristics and level characteristics
as a graph on the color LCD display of the instrument.
A Cole-Cole plot and admittance pie graph can also be easily displayed.
It provides ve types of equivalent circuit models for circuit element
components.
It is capable of continuous measurements using measurement conditions
stored in the instrument memory. This function enables making pass/fail
judgments with different measurement conditions.
(Example: Performs C-D measurement with 1 MHz and Ls measurement with
100 MHz in succession.)
1 MHz to 3 GHz
Various interfaces are
supported
It supports EXT I/O (handler interface), which is the most suitable for
production lines, USB, GP-IB, RS-232C, and LAN.
* GP-IB and RS-232C are optional.
9
Overview and Features
HIOKI IM7585A981-05
Comparator function
• LCR mode: (p. 46) It is capable of making pass/fail judgements by
determining whether measurement values qualify as higher, within a range,
or lower (hereafter referred to as HI, IN, and LO, respectively) regarding four
parameters.
• ANALYZER mode: (p. 108) It is capable of making pass/fail judgments for
sweep measurement results.
BIN function
LCR mode can divide the rank up to 10 classications based on the
measurement values.
10
1.2 Names and Functions of Parts
HIOKI IM7585A981-05
Front panel of the instrument
Names and Functions of Parts
Example: IM7585
Coaxial connectors (p.17)
•CONTROL
•RF OUT
•PORT 1
•PORT 2
IM7580A and IM7581 only
Connector for USB
ash drive (p.243)
Connect a USB ash
drive storage device.
Coaxial connectors (p. 17)
• RF OUT
• PORT 1
• PORT 2
LCD display
This is a touch panel display.
Press the keys displayed on the screen to
operate the instrument.
HOME button
• Returns to the measurement screen.
• Use this button for the full reset operation
(p. 309).
Start button (p. 24)
(The main power switch is located at the rear.)
Lamp statusInstrument status
Green
Red
The start button turns the state from inactive to
active. Accurate measurement requires at least
60 minutes of warm-up.
Measurement lamp
Lamp statusInstrument status
GreenDuring measurement
Red
Power supply ON
Active
Power supply OFF
Inactive
Full reset under
preparation
1
Overview
Bottom panel of the instrument
Example: IM7585
This instrument can be rack mounted.
Refer to “Appx. 6 Rack Mounting” (p. A7).
Parts removed from this instrument should be stored in a safe place to
enable future reuse.
11
Names and Functions of Parts
HIOKI IM7585A981-05
Rear of the instrument
Example: IM7585
Vents (p. 5)
Install so that the vents
are not covered.
Main power switch
(p. 24)
Power inlet (p. 20)
Connect the supplied power
cord.
(p. 6)
Serial number and model name
The 9-digit serial number indicates the year (rst two digits) and the
month of manufacture (next two digits). Do not remove this sticker as
the number is important.
Communication interface (option) (p. 235)
Optional interfaces can be installed.
Refer to the Communication Instruction Manual (Impedance Analyzer
Application Disc).
• Model Z3000 GP-IB Interface
• Model Z3001 RS-232C Interface
MAC address
EXT I/O Connector
(p. 199)
(p. 7)
USB cable connector
Refer to the
Communication
Instruction Manual
(Impedance Analyzer
Application Disc).
LAN connector
Refer to the
Communication
Instruction Manual
(Impedance Analyzer
Application Disc).
IM7580A, IM7581
In Models IM7580A and IM7581, the following are the
differences with Model IM7585.
• The position and shape of vents
• The position of the MAC address
• The position of the serial number and model name
Right side panel of the instrument
Handle
Used for carrying the instrument.
Example: IM7585
When setting up the stands
Open till the end without stopping in between.
Ensure that both stands are straight.
12
When closing the stands
Close till the end without stopping in between.
Stands
Enables the instrument to be tilted.
Side of the test head
HIOKI IM7585A981-05
IM7580A, IM7581
IM7583, IM7585, IM7587
Meauring terminal
(p. 6)
Measuring terminal
(p. 6)
Coaxial connectors
RF OUT
PORT1
PORT2
Coaxial connectors
Names and Functions of Parts
1
Overview
CONTROL
RF OUT
PORT1
PORT2
Bottom of the test head
IM7580A, IM7581
2×M4
Effective Depth 5
Threaded holes for xing the IM9200 Test Fixture Stand
These holes can also be used when xing a test head to an automated machine.
The depth of the threaded holes is 5 mm.
IM7583, IM7585, IM7587
Do not use screws of length exceeding M4 × 5 mm. Doing so may damage the device.
13
Screen Operations
HIOKI IM7585A981-05
1.3 Screen Operations
This instrument allows you to use a touch panel to set and change all measurement conditions.
Gently touch the key on the screen to select the item or numerical value set for that key.
In this manual, gently touching the screen is referred to as “press”.
CAUTION
Do not use excessive force on the touch panel, and do not use hard or sharp objects
that could damage the touch screen.
Example: Setting the measurement frequency in LCR mode
2
3
1
On LCR mode
1
screen, press
[SETUP] of the
menu key.
Press the [BASIC] tab.
2
Press [FREQ].
3
4
14
5
4
[DIGIT] (input with /), or [10-KEY] (input with the
numeric keypad).
6
5
Set the measurement
4
frequency with
the numeric keypad.
Press any unit key to
5
conrm the setting.
Press [EXIT] to close
6
the setting screen.
/,
or
Example: Moving the window
HIOKI IM7585A981-05
Screen Operations
You can move the window
by moving the top of the
window (green bar) while
pressing it.
1
Overview
15
Screen Operations
HIOKI IM7585A981-05
16
2
HIOKI IM7585A981-05
Read “Operating Precautions” (p. 5) before installing and connecting this instrument.
Refer to “Rack Mounting” (p. A7) for rack mounting.
Measurement Preparations
2.1 Connecting the Test Head
Connect the test head.
Refer to “5 Calibration and Compensation” (p. 141).
CAUTION
• If the instrument and the connectors of the test head are not correctly connected, the
instrument may get damaged or accurate measurements may not be possible.
• Tighten the connector with a torque of 0.56 Nm (recommended value). Tightening
the connector with a torque other than the recommended value may damage the
instrument or accurate measurements may not be possible.
2
Measurement Preparations
IMPORTANT
• Check that there are no problems with the connector before connecting the cable. If
there is a problem with the connector, you cannot perform accurate measurement due
to large measurement errors.
Refer to “Appx. 5 Maintenance of Coaxial Connector” (p. A6).
• The instrument, test head and measuring cables have been adjusted as a set before
shipment.
Connect the test head having the same serial number as the instrument with the
measuring cables supplied.
Example: IM7585
Rear of the instrumentRear of the test head
Serial No.
17
Connecting the Test Head
HIOKI IM7585A981-05
1
Check that the power switch of the instrument is turned off.
2
Connect CONTROL, RF OUT, PORT1, and PORT2 of the instrument to CONTROL, RF OUT,
PORT1, and PORT2 of the test head with the supplied measurement cable.
Example: IM7585
0.56 Nm
(recommended value)
If the specied torque is applied to the
torque wrench, it will rotate to the position
shown in the gure.
Do not rotate the cable when connecting the SMA connector of the cable to the
instrument and the test head. If the cable is rotated while connecting the connector, the
core wires of the connector or cable may get damaged. Rotate the nut of the connector
and connect.
18
Pre-Operation Inspection
HIOKI IM7585A981-05
2.2 Pre-Operation Inspection
Be sure to read “Operating Precautions” (p. 5) before use.
Before using the instrument for the rst time, verify that it operates normally to ensure that no
damage occurred during storage or shipping.
If you nd any damage, contact your authorized Hioki distributor or reseller.
Inspection of accessories and options
Inspection itemSolution
Is the power cord insulation torn, or is any metal
exposed?
Is the insulation on the measurement cable torn, or is
any metal exposed?
Do not use the instrument if it is found to be damaged, as it
can result in electric shocks or short-circuit accidents.
Contact your authorized Hioki distributor or reseller.
If there is any damage, measurement values may not be
stable and measurement errors may occur.
We recommend using a new cable without any damage.
Instrument inspection
Inspection itemSolution
Is the instrument damaged?If the instrument is damaged, request repairs.
Does the splash screen appear (Model no., version
no.) when the power is turned on?
Screen when the power is turned on
(Example: IM7585)
If the splash screen does not appear, the power cord may
be damaged, or the instrument may be damaged internally.
Request repairs.
2
Measurement Preparations
Is there an error displayed instead of the splash
screen?
Error display screen
If there is an error displayed, the instrument may be
damaged internally. Request repairs.
Refer to “Troubleshooting” (p. 305) and “Error Display”
(p. 310).
Example: An interface board that cannot be used is installed
(LAN board).
19
Connecting the Power Cord
HIOKI IM7585A981-05
2.3 Connecting the Power Cord
Please read “Before turning ON the power” (p. 6) and “Handling of cords and xtures” (p. 7)
before connecting the measurement cable or xture.
Connect the power cord to the power inlet on the instrument, and plug into an outlet.
Turn OFF the main power switch before disconnecting the power cord.
1
2
Check that the main power switch of the instrument is turned off.
Connect a power cord that is compatible with the line voltage to the power inlet on the
instrument (100 V to 240 V AC).
Rear (Example: IM7585)
3
Plug the other end of the power cord into an outlet.
20
Connecting a Measurement Cable/Fixture
HIOKI IM7585A981-05
2.4 Connecting a Measurement Cable/Fixture
Please read “Before turning ON the power” (p. 6) and “Handling of cords and xtures” (p. 7)
before connecting the measurement cable or xture.
Connect the measurement cables, or optional Hioki test xture to the measurement terminals.
Refer to “Options (Sold Separately)” (p. 2) for options.
Refer to the Instruction Manual of the xture for the operating details.
2
Note the following items when extending the distance between the test sample and
measuring terminals.
• Use 50 Ω coaxial cable for the measurement cable.
• Make the length of the cable as short as possible.
• Perform open/short/load calibration using the connecting terminal of the test sample.
Use the specied probes and xtures. When you make your own probe, it may not satisfy the
specications of this instrument.
Refer to “Options (Sold Separately)” (p. 2).
You will need:
• Test head × 1
• Model IM9200 Test Fixture Stand × 1
• Model IM9906 Adapter × 1
• 3.5 mm Connector torque wrench × 1 (This is not provided with the device.)
1
Metal ttings
Test head
Pull the metal ttings to the model IM9200 Test
1
Fixture Stand side.
Place the test head on the stand and tighten the knob.
2
Measurement Preparations
4
2
Model IM9906
3
Measuring terminal
Notch
3.5 mm connector
5
Model IM9200
nut
Place the notch of the model IM9906 Adapter parallel
3
to the metal ttings, then install onto the measurement
terminal of the test head.
At this time it is stopped temporarily. Position the notch of the
model IM9906 Adapter where the metal ttings can slide.
Fix the notch of model IM9906 adapter by sliding the
4
metal ttings.
Tighten the nut of the 3.5 mm connector by using a
5
torque wrench.
Recommended torque: 0.9 N·m
Tightening the nut until the handle of the wrench bends slightly is
sufcient. Do not over tighten.
21
Connecting an Interface
HIOKI IM7585A981-05
2.5 Connecting an Interface
Rear (Example: IM7585)
To prevent instrument damage or electric shock, use only the screws (M3 × 6 mm)
shipped with the instrument for installing the interface.
If you have lost any screws or nd that any screws are damaged, please contact
your Hioki distributor for a replacement.
• Please read “Input modules (optional)” (p. 8) before connecting the
Interface.
• Read the Instruction Manual of the optional interface before installing or
replacing an optional interface or to use the instrument after removing
the interface.
• You can check the information of the interface installed in the
instrument on the screen.
Refer to “Setting the Interface” (p. 235) and “Checking the Instrument
Version” (p. 236).
WARNING
CAUTION
To connect or disconnect optional interfaces, hold the metal part. Touching the PCB with
bare hands could damage the instrument due to static electricity. (An antistatic wrist
strap is recommended when disconnecting the interface.)
22