This instruction manual provides details on the communication interfaces of the IM7580A series, IM7581
series, IM7583 series, IM7585 series and IM7587 series Impedance Analyzer.
In this document, the “instrument” means the IM7580A series, IM7581 series, IM7583 series , IM7585
series and IM7587 series.
1
Safety Information
This manual contains information and warnings essential for safe operation of the instrument and for
maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety
precautions.
Notation
In this manual
WARNING
CAUTION
IMPORTANT
the risk seriousness and the hazard levels are classied as follows.
,
Indicates a potentially hazardous situation that may result in death or serious
injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate
injury to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance
tasks with which the operators must be fully familiar.
Indicates prohibited actions.
Indicates the action which must be performed.
*
[ ]
Indicates the action which must be performed.
Menus, commands, dialogs, buttons in a dialog, and other names on the screen
and the keys are indicated in brackets.
2
3
4
5
6
7
CURSOR
( Bold character)
Windows
DialogueDialogue box represents a Windows dialog box.
Mouse Operation
Click:Press and quickly release the left button of the mouse.
Right-click:Press and quickly release the right button of the mouse.
Double click:Quickly click the left button of the mouse twice.
Drag:
Bold characters within the text indicate operating key labels.
Unless otherwise specied
Windows 8 or Windows 10.
While holding down the left button of the mouse, move the mouse and then release the
left button to deposit the chosen item in the desired position.
“Windows” represents Windows Vista, Windows 7,
,
8
9
10
付録索引
1
Safety Information
2
1
Specications
1.1 RS-232C Specications
Transmission MethodCommunication method: Full duplex
• When the data in the receive buffer exceeds 75% of the
buffer, XOFF(13H) is sent and the controller is notied
that there is not much space remaining in the buffer.
• Processing of the data in the buffer continues, and then
XON(11H) is sent and the controller is notied that there
is sufcient remaining space in the buffer when the
amount of data becomes less than 25%.
Control during Sending
XON/XOFF control:
• When XOFF is received, the sending of data is halted. When XON is received, the sending of data is
resumed.
Input Buffer
Usage
Amount
85%
75%
25%
Buffer Space
Send XOFF
Send XON
3
GP-IB Specications
1.2 GP-IB Specications
SH1Supports all source handshake functions.
AH1Supports all acceptor handshake functions.
T6Supports standard talker functions.
Supports serial poll functions.
Talk only mode is not supported.
Supports the talker cancel function by MLA (My Listen Address).
L4Supports standard listener functions.
Listener only mode is not supported.
Supports the listener cancel function by MTA (My Talk Address).
SR1Supports all service request functions.
RL1Supports all remote/local functions.
PP0Parallel poll functions are not supported.
DC1Supports all device clear functions.
DT1Supports all device trigger functions.
C0Controller functions are not supported.
Code used: ASCII code
1.3 USB Specications
ConnectorSeries B receptacle
Compliance tandardUSB2.0 (Full Speed/High Speed)
No. of ports1
ClassCommunication class
Supported OSWindows Vista, Windows 7, Windows 8, Windows 10
1.4 LAN Specications
ConnectorRJ-45 connector × 1
Compliance
standard
IEEE 802.3-compliant Ethernet
Transfer system10BASE-T/ 100BASE-TX/ 1000BASE-T Auto detected
ProtocolTCP/IP
FunctionCommand control
4
2
Connection and Setting
2.1 Overview of Communication
You can control the instrument with communication commands from a computer via the USB, GP-IB, RS232C and LAN interfaces.
There are the following four communication methods. To enable communication, the communication
conditions need to be set on the instrument.
USB communication (p. 7)
The instrument is communication class compatible.
LAN communication (p. 9)
2
Command control using the TCP/IP protocol is possible.
GP-IB communication (when the Z3000 is connected) (p. 14)
• Commands common to IEEE-488-2 1987 (requirement) can be used.
• The instrument has been designed with reference to the following standard.
(Reference standard: IEEE-488.2 1987)
RS-232C communication (when the Z3001 is connected) (p. 16)
WARNING
• Always turn both devices OFF when connecting and disconnecting an interface
connector. Otherwise, an electric shock accident may occur.
• To avoid damage to the instrument, do not short-circuit the terminal and do not
input voltage to the terminal.
• Failure to fasten the connectors properly may result is sub-specication
performance or damage to the equipment.