Hioki IM7580A-1, IM7580A-2, IM7581-01, IM7581-02, IM7583-01 Communication Instruction Manual

...
IM7580A IM7581
IM7583 IM7585 IM7587
Communication Instruction Manual
IM7580A-1 IM7580A-2
IM7583-01 IM7583-02
IMPEDANCE ANALYZER
Sept. 2016 Revised edition 2 IM7585A987-02 16-09H
EN

Contents

Introduction .............................................................................................................................1
Safety Information ..................................................................................................................1
1 Specications 3
1
1.1 RS-232CSpecications ................................................................................................3
Handshake (About Buffer Flow Control) ...............................................................................................3
1.2 GP-IBSpecications .....................................................................................................4
1.3 USBSpecications .......................................................................................................4
1.4 LANSpecications .......................................................................................................4
2 Connection and Setting 5
2.1 Overview of Communication ........................................................................................5
Screen Displayed while Setting Interfaces ........................................................................................... 6
2.2 USBSettingsandConnection .....................................................................................7
Setting USB ..........................................................................................................................................7
Connecting the USB Cable ................................................................................................................... 8
2.3 LANSettingsandConnection .....................................................................................9
LAN Settings .........................................................................................................................................9
Connecting a LAN Cable .................................................................................................................... 13
2.4 GP-IB Connection and Settings
(whentheZ3000isconnected) ..................................................................................14
Connecting the GP-IB Cable .............................................................................................................. 14
Setting GP-IB ......................................................................................................................................14
2.5 RS-232CConnectionandSettings
(whentheZ3001isconnected) ..................................................................................16
Connecting the RS-232C Cable ......................................................................................................... 16
Setting RS-232C .................................................................................................................................17
2.6 Remote Mode ...............................................................................................................18
Canceling Remote Mode .................................................................................................................... 18
2 3 4 5 6 7
Appendix A1
Appx.1 CheckingtheUSBVirtualCOMPort A1
IM7585A987-02
8 9
10
付録索引
i
Contents
ii

Introduction

Introduction
This instruction manual provides details on the communication interfaces of the IM7580A series, IM7581 series, IM7583 series, IM7585 series and IM7587 series Impedance Analyzer. In this document, the “instrument” means the IM7580A series, IM7581 series, IM7583 series , IM7585 series and IM7587 series.
1

Safety Information

This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions.
Notation
In this manual
WARNING CAUTION
IMPORTANT
the risk seriousness and the hazard levels are classied as follows.
,
Indicates a potentially hazardous situation that may result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance tasks with which the operators must be fully familiar.
Indicates prohibited actions.
Indicates the action which must be performed.
*
[ ]
Indicates the action which must be performed. Menus, commands, dialogs, buttons in a dialog, and other names on the screen
and the keys are indicated in brackets.
2 3 4 5 6 7
CURSOR
( Bold character)
Windows
Dialogue Dialogue box represents a Windows dialog box.
Mouse Operation
Click: Press and quickly release the left button of the mouse. Right-click: Press and quickly release the right button of the mouse. Double click: Quickly click the left button of the mouse twice.
Drag:
Bold characters within the text indicate operating key labels.
Unless otherwise specied
Windows 8 or Windows 10.
While holding down the left button of the mouse, move the mouse and then release the left button to deposit the chosen item in the desired position.
“Windows” represents Windows Vista, Windows 7,
,
8 9
10
付録索引
1
Safety Information
2
1
Specications
1.1 RS-232C Specications
Transmission Method Communication method: Full duplex
Synchronous method: Start-stop synchronization
Transmission Speed 9600 bps, 19200 bps, 38400 bps, 57600 bps Data Bits 8 bits Parity None Stop bit 1 bits Message terminator (delimiter) CR+LF, CR Flow control Software (XON/XOFF control)
“Handshake (About Buffer Flow Control)” (p. 3)
Electrical Specications

Handshake (About Buffer Flow Control)

Control during Receiving
Input voltage level
Output voltage level
5 to 15 V ON
-15 to -5 V OFF 5 to 9 V ON
-9 to -5 V OFF
1
XON/XOFF control:
• When the data in the receive buffer exceeds 75% of the buffer, XOFF(13H) is sent and the controller is notied that there is not much space remaining in the buffer.
• Processing of the data in the buffer continues, and then XON(11H) is sent and the controller is notied that there
is sufcient remaining space in the buffer when the
amount of data becomes less than 25%.
Control during Sending
XON/XOFF control:
• When XOFF is received, the sending of data is halted. When XON is received, the sending of data is resumed.
Input Buffer Usage Amount
85%
75%
25%
Buffer Space
Send XOFF
Send XON
3
GP-IB Specications
1.2 GP-IB Specications
SH1 Supports all source handshake functions.
AH1 Supports all acceptor handshake functions.
T6 Supports standard talker functions.
Supports serial poll functions. Talk only mode is not supported. Supports the talker cancel function by MLA (My Listen Address).
L4 Supports standard listener functions.
Listener only mode is not supported. Supports the listener cancel function by MTA (My Talk Address).
SR1 Supports all service request functions. RL1 Supports all remote/local functions. PP0 Parallel poll functions are not supported. DC1 Supports all device clear functions. DT1 Supports all device trigger functions. C0 Controller functions are not supported.
Code used: ASCII code
1.3 USB Specications
Connector Series B receptacle Compliance tandard USB2.0 (Full Speed/High Speed) No. of ports 1 Class Communication class Supported OS Windows Vista, Windows 7, Windows 8, Windows 10
1.4 LAN Specications
Connector RJ-45 connector × 1
Compliance standard
IEEE 802.3-compliant Ethernet
Transfer system 10BASE-T/ 100BASE-TX/ 1000BASE-T Auto detected Protocol TCP/IP Function Command control
4
2

Connection and Setting

2.1 Overview of Communication

You can control the instrument with communication commands from a computer via the USB, GP-IB, RS­232C and LAN interfaces. There are the following four communication methods. To enable communication, the communication conditions need to be set on the instrument.
USB communication (p. 7)
The instrument is communication class compatible.
LAN communication (p. 9)
2
Command control using the TCP/IP protocol is possible.
GP-IB communication (when the Z3000 is connected) (p. 14)
• Commands common to IEEE-488-2 1987 (requirement) can be used.
• The instrument has been designed with reference to the following standard. (Reference standard: IEEE-488.2 1987)
RS-232C communication (when the Z3001 is connected) (p. 16)
WARNING
• Always turn both devices OFF when connecting and disconnecting an interface connector. Otherwise, an electric shock accident may occur.
• To avoid damage to the instrument, do not short-circuit the terminal and do not input voltage to the terminal.
Failure to fasten the connectors properly may result is sub-specication performance or damage to the equipment.
5