Hioki IM3523 Data Sheet lists accessories and test fixture

LCR METER
Component measuring instruments
IM3523/IM3533/IM3533-01
A New Series of LCR Meters to Meet Your Applications
From Production Lines to Research and Development
New LCR METER Models IM3523, IM3533, and IM3533-01 are highly cost-effective testers that provide greater performance and better functionality than previous HIOKI models, such as a high basic accuracy of ±0.05%, a wide measurement frequency from 1 mHz (40 Hz for the IM3523) to 200 kHz, high­speed measurement of up to 2 ms, highly reliable measurement using the contact­check function, and measurement of turn ratio and mutual inductance. Select the best model according to your application, from production lines to research and development.
2
IM3533,IM3533-01
IM3523
For Production Lines
The Perfect Impedance Analyzer
Product Lineup
Model
LCR METER IM3523 LCR METER IM3533
LCR METER IM3533
-01
Usage
*1
Research and
development
✔✔
Transformer and
coil production
✔✔ ✔✔
LCR component
production
✔✔ ✔✔ ✔✔
Measurement
items
Basic measurement items
Z (impedance [Ω]) Y (admittance [Ω]) θ (phase angle [°])
Rs (equivalent series resistance = ESR [Ω])
Rp (parallel resistance [Ω])
X (reluctance [Ω])
G (conductance [ S]) B (susceptance [S])
Ls (series inductance [H])
Lp (parallel inductance [ H] ) Cs (series capacitance [ F]) Cp (parallel capacitance [F] )
Q (Q factor (Q = 1/D)) D (loss coeffi cient = tanδ)
DCR
(direct current resistance)
3 3
(with temperature compensation function)
Transformer measurement
N (turn radio) M (mutual inductance) ΔL (inductance difference)
Temperature T
3
Basic accuracy ±0.05%rdg.
Measurement frequency 40Hz to 200kHz 1mHz to 200kHz
Measurement voltage 5mV to 5V 5mV to 5V/2.5V
*2
Measurement time 2ms 2ms
Comparator 2 items: HI/IN/LO, ABS/%/Δ%
BIN measurement
Main item: 10 categories
Sub-item: 1 categor y
2 items: 10 categories
Cable length 0m /1m 0 m/1m 0m/1m/2m/4m
Contact check 4-terminal contact check (threshold change) / Hi-Z reject
Internal DC bias measurement −5V to 5V
Sweep measurement Frequency 2 to 801 points
Display Monochrome LCD Color TFT 5.7-inch LCD touch panel
Interface
EXT I/O, USB
3 3
USB fl ash drive
3
RS-232C, GP-IB, LAN
Option (select one)
*1 The check and double-check marks in the “Usage” rows indicate the recommendation level.
The double-check mark represents a highly recommended application.
Highlighted functions in bold-type in the IM3533 and IM3533
-01 section are more advanced than those of
the IM3523.
*2
2.5 V in the low impedance high accuracy mode
3
For Lead Components and Surface Mounted Devices (SMDs)
Probes & Test Fixtures
Please use the probes specified below. All probes are constructed with a 1.5D-2V coa xial cable.
Probes and Test Fixtures for Lead Components
MAX : 2
85 188
1000
FOUR-TERMINAL PROBE L2000
Cable length 1 m (3.28 ft), DC to 5 MHz, impedance char-
acteristics of 50 Ω, 4-terminal pair conguration, measur­able conductor diameter: ø5 m m (0.20 in) max.
94 160
1000
FOUR-TERMINAL PROBE 9140-10
Cable length 1 m (3.28 ft), DC to 200 kH z, impedance
characteristics of 50 Ω, 4-terminal pair conguration, measu rable conductor diameter: ø5 mm (0.20 in) max.
.....................Coming soon
Test Fixtures for SMDs
SMD type
: Measurable
3 3*: May not be measur-
able depending on the shape.
JIS CODE EIA CODE
0603 0201 100 5 0402 160 8 0603 20 12 0805 3216 12 06 3225 12 10 4532 1812 5750 2220
5
20
MAX φ5.0
5
20
MAX φ5.0
Applicable SMD size
Length L
(mm)
0.60 0.30
1.00 0.50
1.60 0.80
2.00 1.25
3.20 1.60
3.20 2.50
4.50 3.20
5.70 5.00
3
TEST FIXTURE 9262
Direct con nection type, DC to 5 MHz ,
measu rable conductor diameter: ø2 mm (0.08 in) max.
38 38
84
TEST FIXTURE 9261-10
Cable length 1 m (3.28 ft), DC to 5 MHz, impedance characteristics of
50 Ω, 4-terminal pair conguration, measurable conductor diameter:
ø1.5 mm (0.06 i n) max.
.....................Coming soon
Width W
(mm)
9263 9677 9699 914 3-10
*
3 3
*
3 3 3 3 3 3 3
3 3 3
*
3 3
* *
3 3 3 3
MAX: 1.5
3 3
SMD TEST FIXTURE 9263
Direct con nection type, DC to 5 MHz , Tes t
sample dimension s:1 mm (0.04 in) to 10 mm (0.39 in)
SMD TEST FIXTURE 9699
Direct con nection type, Electrodes on bottom for SMD,
DC to 120 MH z, Test sample di mensions: 1.0 mm (0.04 in) to 4.0 mm (0.16 in) wide, maximum 1.5 mm (0.06 in) high
MAX: 10
4
DEPTH:
0.3
SMD TEST FIXTURE 9677
Direct con nection type, Electrodes on side for SM D,
5.8
DC to 120 MH z, Test sample di mensions: 3.5 mm ±0.5 mm (0.14 in ±0.02 i n)
145
1150
10
1
4
PINCHER PROBE 9143-10
Cable length 1 m (3.28 ft), DC to 5 MHz, impedance
characteristics of 50 Ω, 4-terminal pair conguration
.....................Coming soon
0.7
MAX: 4 MIN: 0.1
1
6
5
4
DC
200kHz
1mHz
IM3533
IM3533-01
DC 40Hz 200kHz
IM3523
Measurement frequency[Hz]
Signifi cant ly hig h impe danc e is determined to be a Hi-Z error.
Hi-Z reject function
High contact resistance is de­termined to be an error. The threshold of contact resistance can be ch anged.
Four-terminal contact check
Basic Performance
IM3523 IM3533
IM3533-01
Relationship between capacitance and residual voltage
against which the LCR meter can be protected
0
400
1 10
Capacitance [
μ
F]
3522-50
(Previous
model)
IM3523
IM3533( -01)
Residual
voltage
[V]
100 1000
Features
High-Speed, High-Accuracy, and Easy-to-Use
Functions and Features for LCR Measurements on Production Lines
IM3523 IM3533
IM3533-01
Wide setting range for measurement voltage and current
In addition to normal open-loop signal generation, these models en­able voltage/current dependent measurement in constant voltage/cur­rent modes. The signal levels can be set over wide ranges from 5 mV to 5 V and
from 10 μA to 50 mA. (The setting range of measurement signal lev-
els varies depending on the frequency and measurement mode.)
Wide measurement frequency range
The measurement frequency can be freely set to DC or any value in
the 1 mHz (40 Hz for the IM3523) to 200 kHz range at high resolu­tion (five-digit resolution [1 mHz resolution for less than 100 Hz]). This makes it possible to measure the resonant frequency and perform
measurement and evaluation under conditions close to actual condi­tions.
Basic accuracy ±0.05%
The basic accuracy of Z is ±0.05%. This fi ts a wide array of applica-
tions ranging from the inspection of parts to research and develop­ment measurements.
Accuracy guaranteed at measurement cables of up to 4 meters
Four-terminal pair configuration reduces the influence of measure­ment cables and accuracy is gua ranteed at the measurement cable
lengths of up to 4 meters. This simplifies the wiring of automated machinery. With models IM3523 and IM3533, accuracy is guaranteed at measurement cable lengths of up to 4 meters with the cable length correction set to 1 meter. (The frequency range for which accuracy is
guaranteed varies depending on the cable length.)
15 parameters can be measured
The following parameters can be measured and selected parameters
can be imported to a computer: Z, Y, θ, Rs (ESR), Rp, Rdc (DC resis­tance), X, G, B, Ls, Lp, Cs, Cp, D (tanδ), and Q.
Fastest measurement time 2 ms
The fastest measurement time of 2 ms at a measurement frequency of 1 kHz and the measurement speed FAST improves the inspection
throughput used in automated machinery.
Contact check function incorporated
The contact check function for four-terminal measurement and the
Hi-Z reject function for two-terminal measurement ensure the mea­surement electrode is in contact with the measurement object during measurement.
Continuous measurement under different measurement conditions
Different measu rement items can be measured continuously under
different measurement conditions (frequency, level, and mode).
Protection against charged capacitors*
To address situations when a charged capacitor is incorrectly connect­ed to the measurement terminal, the protection function* has been improved to 10 times of the amount of residual charge of the previous
model 3522
-50.
* This function does not guarantee the measurement of charged capaci­tors. Be sure to discharge the capacitor before measuring it.
Advantage #1
35.6ms
74.2ms
739ms
46.2ms
824ms
348ms
552ms
328ms
FAST
MED
SLOW
SLOW2
Measurement
speed
Comparison of continuous measurement time of
IM3523/IM3533
(-01) and 3522-50
IM3523, IM3533
(-01)
measurement time
3522
-50
measurement time
Reference values
With continuous measurement under varying measurement conditions such as C-D + ESR measurement of capacitors, the total measurement time has been shortened signifi­cantly from the previous HIOKI model 3522
-50. In addition
to the reduction of the time required for individual mea­surements, the time required to change ranges such as a frequency range has been reduced signifi cantly.
5
General specifi cations of the IM3523
Measure-
ment items
Basic measure-
ment items
Z,Y, θ,Rs,Rp,X,G,B,Ls,Lp,Cs,Cp,Q,D
DCR
3
Transformer
measurement
Temperature T
Basic accuracy ±0.05%rdg.
Measurement frequency 40Hz to 200kHz
Measurement voltage 5mV to 5V
Measurement time 2ms
Comparator 2 items: HI/IN /LO, ABS/%/Δ%
BIN measurement 10 main classifi cations/1 sub-classifi cation
Cable le ngth 0m /1m
Contact check
4-terminal contact check (threshold change) / Hi-Z reject
Internal DC bias measurement
Sweep measurement
Display Monochrome LCD
Interface
EXT I/O, USB
3
USB fl ash drive
RS-232C, GP-IB, L AN
Option (select one)
Features of LCR Meter Model IM3523
Integration into Production Lines and Automated Machinery
Easy setup using a numeric keypad on a simple, easy-to-read monochrome LCD
A simple user interface is provided with a high-contrast graphic LCD display, function keys, and numeric keypad. For numeric value set­tings such as the comparator setting, the numeric keypad can be used to enter numbers easily and quickly.
IM3523
Compact size ideal for integration into production lines and automated machinery
The size is the same as that of compact measuring instruments for
bench use - smaller than the previous model - fi tting easily into auto-
mated machinery and production processes.
IM3523
Comparator
In LCR mode, the meter allows for Hi, IN, and Lo
judgments of two types from the measurement items. For the judg-
ment method, % setting and Δ% setting are available in addition to
absolute value setting. If continuous measurement is used, judgments which span over multiple measurement conditions and measurement items are possible.
IM3523
BIN measurement
With the IM3523, the main item can be classifi ed into
10 categories and out of range, and the sub-item into 1 category and out of range.
IM3523
Auto-range control function
When a measurement object crosses over multiple ranges, measure­ment can be tailored by controlling the moving-range of the auto-range.
Measurement can be performed by taking advantage of both the wide
measurement range of the auto-range and the reduction of the mea-
surement time achieved by completing a search only in the specifi ed
range.
Individual items of two continuous measurements can be output from EXT I/O
For two types of continuous measurement judgment items, individual
judgment results can be captured from EXT/IO. This makes it possible
to perform more detailed inspections and sorting.
Functions and Features Suitable for Measurements and Inspection on Production Lines
IM3523 IM3533
IM3533-01
Limit-linked range setting and range-linked setting function
The optimal range is automatically set according to the set reference value or range. In addition, the measurement conditions can be auto­matically set to be optimized according to the change in the range, re­ducing the preparation time.
OPEN/SHORT compensation area setting function
When the measurement frequency range is limited, OPEN/SHORT
compensation can be executed by limiting the compensation area to the actual frequency range being measured. The time required to ex-
ecute OPEN/SHORT compensation is then signifi cantly reduced com-
pared to the time needed to compensate the entire range.
Functions and Features to Reduce the Time Needed to Prepare for Measurement
IM3523 IM3533
IM3533-01
Loading...
+ 11 hidden pages