Hioki Hioki U8552 Manual

Feb. 2020 Revised edition 1 U8550A960-01 20-02H
Plug-In Modules Dedicated to
LR8450 (-01) Memory HiLogger
Instruction Manual
EN
NOOKNO
99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Model names
U8550 Voltage/Temp Unit
U8551 Universal Unit
U8552 Voltage/Temp Unit
U8553 High Speed Voltage Unit
U8554 Strain Unit
Introduction
Thank you for choosing the U8550/U8551/U8552/U8553/ U8554, built-in modules dedicated to Hioki LR8450/LR8450­01 Memory HiLogger, The built-in modules can be connected to the LR8450/LR8450-01. For information, such as operating precautions and safety, see LR8450 Quick Start Manual.
Safety notations
This manual classies seriousness of risks and hazard levels as described below.
Indicates a potentially hazardous situation
WARNING
IMPORTANT
that, if not avoided, could result in death of or serious injury to the operator.
Indicates information or content that is particularly important from the standpoint of operating or maintaining the device.
Connecting the Module to the Instrument
Connect the built-in module to the instrument (LR8450/ LR8450-01). You can connect up to four modules to one LR8450 (-01).
WARNING
Turn o󰀨 the instrument and disconnect any
cables before connecting or removing any plug­in modules.
When not using a plug-in module slot, secure the
connector cover over the instrument’s module connector with the screws.
Failure to do so could cause electric shock or damage the plug-in module and the LR8450 (-01).
You will need: A Phillip’s head screwdriver (No. 2)
Turn o󰀨 the LR8450 (-01) and disconnect the AC
1
Adapter.
Remove the two screws from the connector cover
2
of the LR8450 (-01).
If installing multiple plug-in modules on the LR8450 (-01), do so in the order indicated in the diagram below. Do not remove the connector covers for any plug-in module slots not in use.
When connecting 1 plug-in module When connecting 3 plug-in
modules
OK
Connect the module to slot
Connecting plug-in units as shown below will cause an error to be displayed on the screen.
When connecting 2 or more
plug-in modules on one side
or .
Connect the modules to slots
or slots , , and .
and
When connecting 5 or more
plug-in modules
, ,
Package Contents
U8550 Voltage/Temp Unit
U8551 Universal Unit U8552 Voltage/Temp Unit U8553 High Speed Voltage Unit U8554 Strain Unit ……………………… any one of them
Screw M3×35 mm (spare)

Connection conrmation label (U8554 only) …………… 1

Instruction Manual (this manual) ………………………… 1

………………………………… 2
Remove the connector cover.
3
Store the connector cover removed from the LR8450 (-01) for future use.
2
3
Connect the plug-in module to the connector on the
4
LR8450 (-01).
Insert and tighten the two screws removed in step 2.
5
5
NO
IMPORTANT
Plug-in modules ship with two spare screws (M3 × 35 mm).
Exercise care not to lose them.