Hioki Hioki PQ3198 Manual

PQ3198
POWER QUALITY ANALYZER
Measurement Guide
Mar. 2019 Edition1 PQ3198A981-00 19-03H
Read First
Thank you for purchasing the Hioki PQ3198 Power Quality Analyzer.
This guide introduces the basic measurement procedure to first­time users.
Before using the instrument, be sure to read the instruction manual carefully.
EN
Preparation before measurement
1
Getting
For more information, see the
instruction manual.
Measurement
4
1
Perform zero adjustment. Configure the connection and current sensor settings.
After purchasing the instrument (first time only)
Attach the Z1003 Battery Pack to the instrument.
Turn on the instrument and set its clock.
-1 Perform the pre-measurement inspection.
-2 Insert an SD memory card into the instrument.
(Be sure to close the cover.)
-3 Connect the AC adapter.
-4 Connect the voltage cords and current sensors.
-5 Turn on the instrument.
Push (Zero adjustment will be performed.)
Push
2
Push to display
2
Initial
Turn on the instrument.
Perform zero adjustment after 30-min. warm-up.
Configure initial
settings.
Load settings.
(Load settings file from the SD
memory card. For more information,
see the instruction manual.)
Connect the instrument to the measurement line.
Verify connections.
Perform simple
configuration.
Configure settings as
desired.
Verify settings and connections.
Start recording.
Stop recording.
Verify that data has been saved and analyze.
Turn off the instrument. (The instrument will be
unavailable for checking or analyzing data.)
Analyze data on a computer.
PQ3198
mode
[Setting]
[Record
ing]
[Analyz-
ing]
3
4 6 5
7
8
(Number indicates reference step no.)
2
4
2
3
5
(Shown: Connection selected)
Key functions
A. Menu keys
SYSTEM:
Configure various settings and event thresholds.
VIEW:
Display instantaneous values and waveforms.
TIMEPLOT:
Display measurement data as a time series graph.
EVENT:
Display measured events as a list.
B. DF keys
Select detailed screen display from each screen.
C. Cursor key, ENTER key
Select and accept settings.
D. ESC key
Cancel selections and changes.
E. DATA RESET key
Delete displayed measurement data. (Data stored on the SD memory card will not be deleted.)
F. START/STOP key
Start and stop recording.
G. START/STOP LED
Recording standby : Flashing
green
Recording : Steady green
H. MANU EVENT key
Generate events.
I. COPY key
Record data on screen currently being displayed.
J. F keys
Select and change display content and settings.
K. POWER LED
When using AC adapter: Steady
green
When using battery : Steady
red
A
C
B
D
E F
G
H
I
K
J
Display pull-down menu
Select
Accept
Select a setting
3
Procedure
ready settings
1
3
Verify connections to the measurement line.
Connect the instrument to the measurement line, referring to the connection diagram shown on the screen.
Attach voltage cords to measurement lines
Securely clip the leads to metal parts
such as terminal screw terminals or bus bars. (Example: Secondary side of breaker)
Current flow direction arrow
Make certain that the current flow direction arrow points toward the load side.
Verify that the connections are correct, referring to the vector diagrams and measured values on the [Wiring] screen.
If you discover an error, verify the connections and return to step (2) to reconfigure the initial settings.
Vector diagrams
Measured values
Simple
4
Settings such as the current range, nominal input voltage, measurement frequency, and event thresholds will be automatically configured based on the selected connection mode.
(You will need to set the measurement line type, current sensor type, and external VT/CT ratio.) Select from the five available patterns according to your objective. To investigate the cause of a power supply issue, select the abnormal voltage detection pattern. To investigate power supply qual­ity (i.e., to monitor a power system), select the basic power supply quality measurement pattern.
Example: Configuring settings for the abnormal voltage detection pattern
1
2
3
Push (The following screen will be displayed.)
Push
Push to display
5
Push (The following screen will be displayed.)
Display pull-down menu
Select
[Voltage event detection]
Accept
Select
[Easy setting course]
Easy setting course patterns
Setting Contents Description
Voltage event
detection
Monitors voltage factors (dips, swells, inter­ruptions, etc.) and frequency to detect events. This pattern is used to investigate the cause of equipment malfunctions. The TIMEPLOT interval will be set to 1 minute.
Standard Power
Quality
Monitors voltage factors (dips, swells, inter­ruptions, etc.), frequency, current, voltage and current harmonics, and other characteristics to detect events. This pattern is primarily used to monitor systems. The TIMEPLOT interval will be set to 10 minutes.
Inrush Current
Measures rush current. The TIMEPLOT inter­val will be set to 1 minute, and the rush cur­rent threshold will be set to 200% of the RMS current (reference value) set during simple configuration.
Record measured
value
Records measured values over an extended period of time using a TIMEPLOT interval of 10 minutes. All event detection functionality other than manual, start, and stop events is turned off.
EN50160
Performs EN50160-compliant measurement. Standard-compliant evaluation and analysis can be performed by analyzing data using the application software PQ ONE, which is sup­plied with the instrument.
For more information about settings, see the
instruction manual.
4
Applying clamp sensors to lines to be measured
Always clamp the instrument around only one conductor. Clamping the instrument around two or more of conductors in a bundle prevents the instrument from measuring any current regardless of whether the measurement target is a single-phase or three-phase circuit.
OK
Source side
Line
Load side
configuration
6
7
Ver ify [Declared input voltage] and [Frequency]
These values will be set automatically. Change the values if they are incorrect.
Push (A message informing you that the simple configuration process is starting will be displayed. Select [Yes] to continue with the process.)
Event indicator
Counting starts when recording starts. Reaches capacity at 9999 events.
5
Verifying settings and connections
-2. Are too many events occurring?
(Is the icon shown continuously?)
If too many events are occurring, check which events are being generated on the event list on the [EVENT] screen (see step 7) after recording some data
and then change the thresholds for
the problematic events. Alternately, wiring (connections) to the measurement line may
be incorrect. Verify the connections.
-3. Are the voltage and current phase relationships shown properly in the vector diagrams?
If not, the wiring (connections) to the measurement line or cur­rent sensor orientations may be incorrect. Verify the connections and current sensor orientations.
1
2
Example: To change the voltage
RMS (Upper limit)
Push
Push to display
Select the level setting
Change the threshold
Accept
Select
[(Upper limit)]
5
4
Push
6
User settings
You can change event thresholds, the recording start date and time, recorded items, and other set­tings as desired.
1
Example: To change the [Time Start] start date
and time (example: set to 12:00)
Push
2
Push to display
Accept
The time can only be changed when
the [Time Start] setting is [Time].
-1. Are measured values or crest factors out of range?
If you see any warning indicators, the clamp sensor, range, or connection settings may be incorrect. Verify the connections and return to step (2) to reconfigure the initial settings.
Warning indicators: Values shown in red as below.
-4. Are waveforms and measured values shown properly?
If waveforms and measured values are not shown properly, the clamp sensor, range, or connection settings may be incorrect. Return to step 2 to reconfigure the initial settings. Alternately, wir­ing (connections) to the measurement line or current sensor ori­entations may be incorrect.
(Voltage and crest factor out of range)
(Current and crest factor out of range)
The vectors should resemble a fan shape.
Event icon
Normally shown in white but turns orange when too many events occur.
No. of events recorded
Up to 9999 events can be recorded. (when [Max. recordable
events]
is set to [9999].) Measurement data continues to be recorded even when the number of events exceeds 9999.
3
Push
Push to display
Push
Push to display
Push to display
(changing settings)
4
Select
Select the value so it can be changed
Set to
[12]
Accept
3
Push
Creating reports
Output information shown on the screen as-is in the form of a report. Create reports that suit your needs without the need to configure complex settings.
Viewing data
Use event statistics functionality to analyze measurement data at a high level of detail. By reviewing event occurrence by date or time, you can discover events that are likely to occur at a specific time or on a specific day of the week.
Operating mode and key operation
[Setting]
[Recording] [Analyzing]
Start recording Stop recording
Revert to setting mode
Change settings
Monitoring fluctuations in measured values
You can view measured items in the form of a time series graph. You can also display flicker values as a graph or list.
Press the TIMEPLOT key to display the [TIMEPLOT] screen. You can change the screen display with the DF keys.
7
Starting recording
Press the START/STOP key to start recording. Data will be automatically saved to the SD memory card.
• When recording, the operating mode indi­cator on the top right of the screen will show [Recording].
• To change a setting, set the operating mode to [Setting]. (Settings cannot be changed while recording or analyzing.)
8
Analysis
After recording has stopped, data can be analyzed on a computer using the application software PQ ONE, which is supplied with the instrument.
Display file data in list form
Drag and drop folders containing measurement data to display a list of settings information and event occurrence data for all the data contained in the folder.
Converting data
Convert event data and time plot data (binary data) to the CSV format so that it can be opened with a spreadsheet application on a computer.
Easily create graphs to suit your needs
Organize time plot graphs by convenient times or group three phases of data together on a single graph.
To hold waveforms and values
Push
To active the key lock
For more information, see the PQ ONE
i
nstruction manual (on the included
CD-ROM).
Press and hold for at least 3 seconds. To cancel the key lock, press and hold for at least 3 seconds again.
Calculating demand and integral power
Display a demand graph, or display the maximum and integrated power values for a specified interval.
To check instantaneous values
Push
(Switch screens with the DF key.)
Monitoring event generation
You can check whether events are occurring and the num­ber of events occurring with the event list.
Press the EVENT key to display the [EVENT] screen.
Event list
Select an event and press the ENTER key to view waveforms and measured values at the time of its generation on the [VIEW] screen.
Event list details
When multiple events occurred at the same time, “1/N” (where N indicates the total number of events) is displayed at the top of the list.
Waveforms at event occurrence
Select an event
Select an event from the event list details
1
2
Hold function is valid on the
[VIEW] screen only.
Loading...