Hioki Hioki PQ3100-01/600 KIT Quickstart

PQ3100
POWER QUALITY
ANALYZER
Measurement Guide
5 Commonwealth Ave Woburn, MA 01801 Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
July 2017 Revised edition 1 Printed in Japan PQ3100A971-01 17-07H
4. Connections with the Instrument
Connect the voltage cords to the voltage
1
Colored clips
input jacks. Connect the current sensors to the current
2
input jacks.
The current sensors will be automatically identied.
Quick Set for Setting
Try to check for the power supply malfunction in a 3-phase 4-wire 230 V line.
Setting Items Setting Example
Wiring: 3P4W (3-phase/4-wire) Declared input voltage: 230 V Measurement frequency: 50 Hz Current sensor: CT7045 Current range: 50 A Easy settings course: Voltage events Recording start method: Interval Recording stop method: Manual
1. Preparations
Install the
1
Battery Pack.
Z1003
Easy
Colored clips for current sensors
Wind the clips of the same
2 3 4
color as the channels around the leads.
(CH1) (CH2)
Red
Yellow
CT7045
You will need
Model PQ3100
Model Z1002 AC Adapter
Red (CH1) Yellow (CH2)
Blue (CH3)
Model Z1003 Battery Pack
Insert the SD memory card.
(CH3)
Blue
Right side of the instrument
Model L1000-05 Voltage Cord
Model CT7045 AC Flexible Current Sensor (optional)
Model Z4001 SD Memory Card 2GB
(optional)
Connect the AC adapter.
Refer to Chapter 2 on the Instruction Manual.
(Wiring image)
Left side of the instrument
Check that the SD memory card is inserted.
Connect the cords to each of the same color jacks.
Refer to Sections 4.3 through 4.5 on the Instruction Manual.
3
Without connecting the voltage cords and
4
current sensors to the measuring lines, press the [F2] (Next) key.
Zero adjustment will be automatically performed.
5. Wiring Voltage Cords to the Measuring Object
Refer to the wiring diagram to check the locations
1
to which the voltage cords have to be connected. Attach the voltage cords to the secondary side of
2
the breaker. Check the vectors and measured values.
3
Check the declared input voltage.
4
In completion of the wiring, values will be set automatically. If the values are different from the actual values, change the
5
6
Refer to Section 4.6 on the Instruction Manual.
values.
Check the wiring judgment.
If all the items are judged to be
(green):
(You can proceed to the next step even with
Press the [F2] (Next) key.
If (red) or (yellow) is displayed:
1. Move the cursor to the (red) or (yellow) items.
(red) or (yellow).)
2. Press the [ENTER] key.
3. Refer to the key points shown in the dialog to correct the wiring.
Align the arrow with the concave portion of the terminal to insert the connector.
Current input jack
A
B
C
N
For a bus bar, pinch the metal part.
Red Yellow
Blue
Black
2. Starting Quick Set
Turn on the instrument.
1
(Language, clock and measuring frequency are
required to be set only during the rst setting.
Refer to the Instruction Manual.)
Press the
2
key.
Press the key.
3
Refer to Chapter 3 on the Instruction Manual.
To exit in the middle of settings
Press any of the other screen switching keys. All the settings before the exit are saved.
3.Basic Settings
Settings for the wiring.
1
3P4W
Press the (Next)
2
key.
Refer to Section 1.6 on the Instruction Manual.
1. Move cursor
2. Display drop­down list
3. Select wiring
4. Settle setting
6. Wiring Current Sensors to Measuring Object
Refer to the wiring diagram to check the
1
locations to which the current sensors have to be connected.
Attach the current sensors around the wires
2
connected to the secondary side of the breaker. Verify that the measured values are displayed.
3
Set the current range.
4
Press the [F2] (Next) key.
5
Tip
Set the current range based on the maximum load current expected to ow during the measurement period.
(Consult the operating status, load rating, breaker rating, and other data to make this determination.) If the range is too low, the instrument will experience an overrange event during measurement. The error component increases if the range is too high. Current cannot be measured accurately in any of the above cases.
Refer to Sections 4.7 and 4.8 on the Instruction Manual.
Current direction mark
Attach the sensor around only one of the conductor.
Source
Load
7. Wiring Check
Check the measured values
1
and vectors.
Check the wiring in the following cases.
• Measured values of the channels are low, or active power Psum shows a negative value.
• Displacement power factor DPFsum is below 0.5.
• The vector position is outside the PASS range. Refer to Section 4.9 on the Instruction Manual.
8. Event Settings
Check the declared input
1
voltage. Select the easy settings course.
2
Voltage events
Check the recording interval.
3
1 min
Changes can made in “Step 9. Recording Settings”.
Press the [F2] (Next) key.
4
Events that can be measured with the selected menu are displayed. (Events are displayed with light color cannot be measured.)
Easy settings course
Threshold values for events and recording interval will be automatically congured.
To make any change to the event settings, press the [SETUP] key after completion of Quick Set to display the Event Settings screen.
Check the wiring judgment.
2
If (red) or (yellow) is displayed:
1. Move the cursor to the item in (red) or (yellow) item.
2. Press the [ENTER] key.
3. Refer to the key points shown in the dialog to correct the wiring.
If all the items are judged to be (green):
The color was (yellow) but the wiring check did not indicate any problems:
Press the [F2] (Next) key.
3
Voltage events
This is used to investigate the cause of power supply abnormalities such as equipment malfunction. Voltage components (swell, dip, interruption) and frequency are monitored. The recording interval will be set to 1 minute.
Inrush current
This is used to measure the inrush current. Event thresholds for inrush current is set to 200% of current RMS and the recording interval to 1 minute.
Trend record only
This is used to record measured values over an extended period of time. All the event settings (effective only for manual events, recording start events, and recording stop events) are set to OFF and recording interval is set to 10 minutes.
EN50160
This is used to measure in conformance to the European Norm EN50160. The recording interval is set to 10 minutes.
(The recording interval is xed to 10
minutes. Cannot be changed.)
Refer to Section 5.3 on the Instruction Manual.
10. Checking Settings and Recording
Check the settings.
1
To make any changes to the settings, press the [F1] (Previous) key to return to applicable screen.
START/STOP LED
Recording start
Press the key.
2
The instrument enters the standby state. (START/STOP LED: Blinking)
The recording will start at the time set by the interval*.
The instrument enters the recording state. (START/ STOP LED: On)
To start recording after setting the items that are not listed in Quick Set.
Press the [F5] (End) key.
The settings congured up to this point will be
saved.
Recording stop
Press the key.
3
The recording stop dialog will be displayed.
Press the key.
4
Recording will be stopped. (START/STOP LED: Off)
Fluctuations in measured values during recording can be monitored.
Press the [TREND] key to display the
TREND screen.
The measured items in the form of a time series graph can be observed.
Event occurrence status during recording can be monitored.
Press the [EVENT] key to display the
EVENT screen.
Event occurrence status can be checked.
Refer to Chapter 7 on the Instruction Manual.
*: Interval
In case of Recording interval:
5 min
Recording
12
11
10
9
8
7
6
Recorded
Example 1: Example 2:
Recorded data can be post­analyzed with a computer.
Data after completion of recording can be analyzed with a computer using the supplied PC application software.
12:43 12:45
start
1
Recorded
2
Recorded
3
4
Recorded
5
Recorded
4:02 4:05
9. Recording Settings
1
2
Congure the Recording start and Recording stop.
Interval: Recording will start at a well-dened time in accordance with the Recording interval.
Interval
Manual
Tip
If the Save time is less than the measurement period, the following methods can be used to increase the save time:
Recording interval: Lengthened
• SD memory card: Delete unnecessary data, and format it. (Exit the Quick Set and use the FILE screen.)
Press the [F2] (Next) key.
Refer to Section 5.2 on the Instruction Manual.
Refer to “8. Verifying the Trends (Fluctuations) in Measured Values” on the Instruction Manual for details.
Refer to “9. Checking Events” on the Instruction Manual for details.
Functions:
• Observing time series data, event data, and event waveform
• Observing statistics data
• Creating reports
Refer to “11. Analysis (with Computer)” on the Instruction Manual for details.
Loading...