Hioki Hioki LR8431-20 PRO KIT Manual

Page 1
LR8431-20
Instruction Manual
MEMORY HiLOGGER
June 2018 Revised edition 4 LR8431B980-04 18-06H
5 Commonwealth Ave Woburn, MA 01801 Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Page 2
Page 3

Contents

Introduction ..............................................................................1
Confirming Package Contents................................................2
Safety Information....................................................................3
Operating Precautions.............................................................6
Chapter 1 Overview ____________________________________13
1.1 Product Overview and Features .............................13
1.2 Measurement Flow .............. .. ..................................14
1.3 Names and Functions of Parts ...............................16
1.4 Screen Configurations ............................................18
1.5 Basic Operation .......................................................21
Screen Operations ...........................................................21
Starting and Stopping Recording .....................................22
Disabling Key Operations (Key-Lock Function) ...............23
Saving Data ......................................................................23
Verifying the Input Level (Level Monitor) ..........................24
i
Contents
1 2
Chapter 2 Measurement Preparations_________________25
2.1 Using the Battery Pack (Option) ............................26
Install the Battery Pack ....................................................29
Charge the Battery Pack ..................................................29
2.2 Connecting the AC Adapter ............................... ... ..30
2.3 Connecting Measurement Cables to the Memory
HiLogger ...................................................................31
Connecting Measurement Leads or Thermocouples
(for voltage or temperature, respectively) ........................32
Connecting the Optional Model 9641 Connection Cable
(for Pulse Signal Input) .....................................................33
2.4 Turning the Power On and Off ...............................34
2.5 Zero Adjustment ......................................................35
LR8431B980-04
Page 4
ii
Contents
Chapter 3 Setting before Measurement_______________ 37
3.1 Pre-Operation Inspection ....................................... 37
3.2 Operating Flow Overview ..................... .................. 38
3.3 Setting Configuration 1 – Recording Settings ..... 42
Specify the Recording Interval .........................................42
Setting the Display Time Base (horizontal axis
magnification,set as needed) ...........................................43
Selecting the Recording Start/stop Method .....................44
Set the Recording Length for Repeated or One-time
Recording ..................................................................... ....45
Enable Digital Filtering (noise suppression) (as needed) 45
Automatic Saving .............................................................46
Replacing Removable Storage During Real-Time Saving 48
3.4 Setting Configuration 2 – Input Channel Settings 49
Voltage Measurement Settings ........................................50
Temperature Measurement Settings ...............................51
Integration (Count) Measurement Settings ......................53
Revolution Measurement Settings ...................................54
3.5 Display Settings ...................................................... 56
Selecting Waveform Display Color ..................................56
Specifying Vertical Display Range by Magnification and
Zero Position (vertical axis expansion/compression) .......56
Specifying the Vertical Display Range by Upper and
Lower Limits (expansion/compression) ............................57
Converting Units (Scaling function) ............. ......... ............58
3.6 Entering Comments ................................................ 60
3.7 Viewing All Channel Settings in a List .................. 62
Batch Copying Channel Settings .....................................63
Batch Setting Waveform Display/Hide and Waveform
Color Settings for All Channels ........................................63
Initializing Settings (to factory defaults) ...........................64
Aligning Zero Positions on the Grid .................................64
Display Celsius (°C) temperature values as
Fahrenheit (°F) .................................................................65
Page 5
Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer) ______________67
4.1 Setting Criteria to Start and Stop Recording ........68
Setting the Input Signal Trigger Threshold (Level Trigger) 69Setting Lower and Upper Trigger Thresholds
(Window Trigger) ..............................................................70
Selecting Trigger Sources and Combinations ..................71
Setting Criteria for Pre-Trigger Recording (Pre-Trig) .......72
4.2 Alarm Output ............................................................73
4.3 Confirming Trigger and Alarm Criteria Settings
(List) ......................... .......................... ....................... 75
4.4 Starting and Stopping Recording by Timer ..........76
4.5 About Recording Operation ...................................78
iii
Contents
3 4
5
Chapter 5 Waveform Analysis__________________________81
5.1 Viewing Waveforms .................................................82
Scrolling Waveforms ........................................................82
Verifying Waveform Display Position ...............................82
Magnifying and Compressing Horizontally .......................83
Viewing Any Waveform Location (Jump Function) ..........83
5.2 Selecting Display Options ......................................84
Displaying Gauges ...........................................................84
Displaying Cursor Values .................................................84
Specifying a Waveform Time Span ..................................86
Viewing Input Signals as Numerical Values .....................87
5.3 Inserting Event Marks (Search Function) ..............88
Inserting Event Marks While Viewing Waveforms ............88
Inserting Event Marks Using External Input Signals ........89
Searching Event Marks ....................................................90
Chapter 6 Saving & Loading Data _____________________91
6.1 About Saving and Loading Data ............................91
6.2 Using a CF Card/USB flash drive ...........................93
CF Card Insertion & Removal ..........................................94
6
Page 6
iv
Contents
USB flash drive Insertion & Removal ...............................94
Formatting a CF Card/USB flash drive ............................95
6.3 Saving Data ............................................................. 96
Automatic Saving .............................................................97
Selecting the Manual Saving Method
[Quick Save]/[Select & Save] ...........................................98
Saving Waveform Data (with the SAVE Key) ..................99
Capturing a Screen Image (With the SAVE Key) ...........100
Saving Numerical Calculation Results
(With the SAVE Key) ......................................................101
Saving Setting Configurations ........................................102
6.4 Loading Data on the Memory HiLogger .............. 103
Loading a Setting Configuration .....................................103
Loading Waveform Data and Screen Images ................104
6.5 Data Management ................................................. 105
Switching removable storage .........................................105
Viewing Folder Contents and the Parent Folder ............106
Deleting Data .................................................................106
Renaming Files and Folders ..........................................107
Copying Data .................................................................108
Sorting Files ...................................................................109
6.6 Transferring Data to a Computer
(USB Drive Mode) ................................................. 110
Select the USB Drive Mode ...........................................110
Connecting the USB Cable ............................................111
Chapter 7 Numerical Calculations ___________________113
7.1 Calculation Methods .......................................... ... 113
Auto Calculation .............................................................114
Manual Calculation ........................................................115
Apply Calculations to a Specific Time Span
(Manual Calculation Only) ..............................................116
7.2 Numerical Value Calculation Expressions ......... 117
Page 7
Chapter 8 System Environment Settings ____________ 119
8.1 Screen and Key Operation Settings ....................120
Using the Auto-Resume Function
(Resume After Power Restoration) ................................120
Adjust Backlight Brightness ............................................120
Enabling and Disabling the Backlight Saver ..................121
Selecting Black or White Screen Background ................121
Enabling or Disabling the Beeper ...................................121
Selecting the Horizontal (Time) Axis Display .................122
Selecting the Display Language .....................................122
8.2 CSV File Saving Settings ......................................123
CSV File Data Decimal and Separator Characters ........123
8.3 Making System Settings ............... ........................124
Setting the Date and Time .............................................124
Initializing the Memory HiLogger (System Reset) ..........125
Switching the USB mode ...............................................126
Self-Test ......................................................................... 12 7
Chapter 9 External Control __________________________ 129
9.1 Connecting to the External Control Terminals ...129
9.2 External Trigger Input ........... ... ... ..........................131
9.3 External Signal Output (Trigger Output) .............132
9.4 Alarm Signal Output (Alarm Output) ...................134
9.5 Synchronous Measurements with Multiple
Instruments ....................... ..................................... 135
v
Contents
6 7 8 9
10
Chapter 10 Specifications_____________________________ 137
Chapter 11 Maintenance and Service_________________ 147
11.1 Troubleshooting ....................................................147
11.2 Cleaning .................................................................149
11.3 Disposing of the Instrument .................................150
11
Appendix
Index
Page 8
vi
Contents
Appendix____________________________________ A1
Appendix 1 Error Messages and Remedial Actions .........A1
Appendix 2 File Naming ......................................................A7
Appendix 3 Text File (CSV) Internal Format............... .. .....A8
Appendix 4 Binary File Size Calculation............................A9
Appendix 5 List of Default Settings..................................A10
Appendix 6 Maximum Recordable Time ..........................A11
Appendix 7 Reference .......................................................A12
Appendix 8 Concerning Noise Countermeasures ..........A13
Appendix 9 Using the Application Program....................A19
Installing Logger Utility ....................................................A20
Installing the USB Driver .................................................A23
Connecting the Memory HiLogger to the Computer with
a USB Cable ....................... ........ ......... ........ ................. ..A26
Starting and Ending Logger Utility ..................................A29
Uninstalling the Logger Utility .............. ...........................A29
Appendix 10Frequently Asked Questions.......................A30
In sta llation and Settings .................................................. A30
Trigge ring .................................... .................................... A32
Me asuring .......................................................................A33
Data Saving ....................................................................A34
Logger Utility ...................................................................A38
Appendix 11Pulse Input Circuit Diagram.........................A39
Index___________________________________Index 1
Page 9

Introduction

Thank you for purchasing the HIOKI "Model LR8431-20 Memory HiLogger." To obtain maximum performance from the instrument, please read this man
Trademarks
• Microsoft, Windows, Excel, are either registered trademarks or trademarks of Microsoft Corporation in the United States and other countries.
• CompactFlash is a registered trademark of Sandisk Corporation (USA).
1
Introduction
ual carefully, and keep it handy for future reference.
Page 10
2
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Confirm that these contents are provided.
Model LR8431-20 ......................... 1
Memory HiLogger
Model Z1005 AC Adapter............1
with supplied power cord
(p. 30)
Measurement Guide....................1
USB Cable....................................1
CD .................................................1
• Instruction Manual (PDF) (This document)
• Logger Utility Instruction Manual (PDF)
• Logger Utility (Data acquisition application program)
(p. A26)
The latest version can be downloaded from our web site.

Confirming Package Contents

Confirming Package Contents
About options:
Contact your dealer or Hioki representative for details.
Model 9780 Battery Pack Model Z1005 AC Adapter Model 9641 Connection Cable (for pulse inputs) Model 9782 Carrying Case Model 9812 Soft Case
Model 9727 PC Card (256MB) Model 9728 PC Card (512MB) Model 9729 PC Card (1GB) Model 9830 PC Card (2GB) Model 9809 Protection Sheet
Page 11
3

Safety Information

Safety Information
This instrument is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. However, using the instrument in a way not desc manual may negate the provided safety features. Be certain that you understand the instructions and precau­tions in the manual before use. ity for accidents or injuries not resulting directly from instrument defects.
This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before u sing it, be sure to carefully read the following safety precautions.
We disclaim any responsibil-
Safety Symbols
In the manual, the symbol indicates particularly important information that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with the symbol) before using the relevant function.
Indicates DC (Direct Current). Indicates AC (Alternating Current).
ribed in this
Indicates the ON side of the power switch. Indicates the OFF side of the power switch.
The following symbols in this manual indicate and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument.
Indicates advisory items related to performance or correct operation of the instrument.
the relative importance of cautions
Page 12
4
Ni-MH
Safety Information
Symbols for Various Standards
Indicates that the product conforms to regulations set out by the EU Directive.
This is a recycle mark established under the Resource Recycling Promotion Law
(only for Japan).
WEEE marking: This symbol indicates that the electrical and electronic appliance is put on the EU mar
ket after August 13, 2005, and producers of the Member States are required to display it on the appliance under Article 11.2 of Directive 2002/96/ EC (WEEE).
Other Symbols
Indicates the prohibited action.
(p. #)
[ ]
SET
(Bold characters)
Unless otherwise specified, "Windows" represents Windows XP, Windows Vista, Win­dows 7, Windows 8 or Windows 10.
Click: Press and quickly release the left button of the mouse. Double click: Quickly click the left button of the mouse twice.
Indicates the location of reference information. Indicates that descriptive information is provided below. The names of setting objects and buttons on the screen are indicated
by square brackets [ ]. Bold characters within the text indicate operating key labels.
Accuracy
We define measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values, with the following meanings:
f.s. (maximum display value or scale length)
The maximum displayable value or scale length. Th rently selected range.
Example: For the 1 V range, f.s. = 1 V
rdg. (reading or displayed value)
The value currently being measured and indicated on the measuring instrument.
dgt. (resolution)
The smallest displayable unit on a digital measurin that causes the digital display to show a "1" as the least-significant digit.
is is usually the name of the cur-
g instrument, i.e., the input value
Page 13
Safety Information
Measurement categories
To ensure safe operation of measurement instruments, IEC 61010 establishes safety stan­dards for various electrical environments, categorized as CAT II to CAT IV, and called mea­surement categories.
Primary electrical circuits in equipment connected to an AC electrical outlet
CAT II
y a power cord (portable tools, household appliances, etc.)
b CAT II covers directly measuring electrical outlet receptacles.
5
CAT III
CAT IV
Using a measurement instrument in an environment d esignated with a higher-numbered ca t­egory than that for which the instrument is rated could result in a severe accident, and must
e carefully avoided.
b Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV measurement applications could result in a severe accident, and must be carefully avoided.
Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets.
The circuit from the service drop to the service entrance, and to the power m
eter and primary overcurrent protection device (distribution panel).
Difference between "Measurement" and "Recording"
The measurement and recording processes are distinguished as follows for the pur poses of these instructions.
Measurement:
Recording:
Measured data (data acquired in internal memory) is erased when ever a new measurement st
arts. To retain data, always record (save) it.
The acquisition of input values into internal HiLogger memory or to a PC via com
munications.
Storing measurement data on a CF card, USB flash drive or on a PC via dat
a communication.
Page 14
6

Operating Precautions

Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Before Use
• Before using the instrument for the first time
, verify that it operates normally to ensure that no damage occurred during storage or ship ping. If you find any dam­age, contact your dealer or Hioki representative.
• Before using the instrument, make sure that the insulation on the cables is undam­aged and that no bare conductors are improperly e
xposed. Using the instrument in such conditions could cause an electric shock, so contact your dealer or Hioki rep­resentative for replacements.
Instrument Installation
• Operating temperature and humidity: 0 to 40 sating)
• Temperature and humidity range for guarante less
Avoid the following locations that could cause an accident or damage to the instrument.
Exposed to direct sunlight Exposed to high tempera­ture
Exposed to water, oil, other chemica Exposed to high humidity or conden
Exposed to high levels of particulate dust
Subject to vibration
ls, or solvents
sation
°C at 80% RH or less (non-conden-
ed accuracy: 23±5°C, 80%RH or
In the presence of corrosive or explosive gases
Exposed to strong electro­magnetic fields Near electromagnetic radia­tors
Near induction heating sys­tems (e.g., high-frequency induc­tion heating systems and IH cooking
utensils)
The maximum operating (ambient) temperature for the LR8431-20 is 40°C. Do not attempt to use in higher temperature environ­ments.
Page 15
• Correct measurement may be impossible in the presence of
Horizontal placement Upright placement
strong magnetic fields, such as near transformers and high-cur rent conductors, or in the presen fields such as near radio transmitters.
• If liquid en ters the enclo sure through an air vent or othe r opening, it ma
y damage the instrument's internal circuitry. Exercise caution con-
cerning the surrounding environment when inst alling the instrument .
Installation Precautions
• If the instrument is used in any state other than the following, the measurement accuracy may not satisfy the device specifications.
• Leave sufficient space around the ventilation holes and install the instrumen
• Avoid temperature changes around the avoid directed airflow such as from an electric fan or air conditioner vent.Thermocouple inputs are prone to measurement errors.
• When the HiLogger is moved to a location with significantly dif­ferent ambient temperature, allow a equalization before measuring.
7
Operating Precautions
-
ce of strong electromagnetic
t with the holes unobstructed.
terminal block. Especially
t least 30 minutes for thermal
Handling the Instrument
• Do not allow the instrument to get wet, and do not take mea­surements with wet hands. This may cause an el ectric shock.
• Do not attempt to modify, disasse ment; as fire, electric shock and injury could result.
To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
This instrument may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interfer ence to the reception of radio and television broadcasts.
mble or repair the instru-
-
Page 16
8
Operating Precautions
Handling the Cords and Cables
• Avoid stepping on or pinching cables, which could damage the cable insulation.
• To avoid breaking the cables and pro
• To avoid damaging the power cord, grasp the plug, not the cord, wh
en unplugging it from the power outlet.
Before Turning Power On
Using the Battery Pack
• For battery operation, use only the HIOKI Model 9780 Battery Pack. We do not take any responsibility for accidents or damage related to the use of any other batteries.
See: 2.1, "Using the Battery Pack (Option)" (p. 26)
Using the AC Adapter
• Use only the supplied Model Z1005 AC Adapter. AC adapter input voltage range is 100 to 240 VAC (with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this range.
• Turn the instrument off before connecting the AC the instrument and to AC power.
• To avoid electrical accidents and to maint ain the safety spec­ifications of this instrument, connec vided only to a 3-contact (two-conductor + ground) outlet.
• Use only the designated power co Use of other power cords may cause fire.
• Before turning the instrument on, make sure the age matches that indicated on it tion to an improper supply voltage may damage the instrument and present an electrical hazard.
bes, do not bend or pull them.
adapter to
t the power cord pro-
rd with this instrument.
supply volt-
s power connector. Connec-
When the power is turned off, do not apply voltage or current to the connectors. Doing so may damage the instrument.
• After use, always turn OFF the power.
• Brief power interruptions of 40 ms or less will not cause this instrument to malfunction. However, Longer interruptions may cause the Memory HiLogger to shut itself off, so consider local power conditions before installing, as appropriate.
• To ensure that recording is not interrupted by power outages, you can use th
e Z1005 AC Adapter and 9780 Battery Pack together.
Page 17
About Inputs and Measurement
• The maximum input voltage (and the maximum rated volt­age to earth) for the analog input terminal V DC). If these limits are exceeded, the instrument may be damaged and personal injury or death could occur, so do not attempt measurement.
• Do not leave the Memory HiLogger connected to test objects i
n environments where a voltage surge might exceed the dielectric withstand voltage. Doing so could result in damage to the Memory HiLogger, bodily injury or fat al accident.
• Channels are insulated by semiconduc voltage beyond the specification is applied between the channels, the semiconductor relay may short circuit. Please ensure that a voltage beyond specification, especially a surge such as a lightning, is never applied. When an abnor­mal measurement value is observed, please contact your dealer or Hioki representative for inspection.
The waveform for an open channel may sometimes appear to be influenced by the signals of the other channels being measured. If you do not like this, please set the waveform display of the open channel to OFF or short-circuit the input terminals of the open channel by connecting the positive and negative terminal.
9
Operating Precautions
s is 30 Vrms (or 60
tor relays. When a
Page 18
10
Operating Precautions
CD Handling
• Always hold the disc by the edges, so as not to make fingerprints on the disc or scratch the printing.
• Never touch the recorde d side of the disc. Do not place the disc dire
• Do not wet the disc with volatile alcohol or water, as there is a possibility of th
• To write on the disc label surface, use a spirit-based felt pen. Do not
use a ball-point pen or hard-tipped pen, because there is a danger of scratching the surface and corrupting the data. Do not use adhesive labels.
• Do not expose the disc directly to the sun's rays, or keep it in conditions of high temp of warping, with consequent loss of data.
• To remove dirt, dust, or fingerprints from the disc, wipe with a dry clo
th, or use a CD cleaner. Always wipe from the inside to the outside, and do no wipe with circular movements. Never use abrasives or solvent cleaners.
• Hioki shall not be held liable for any problems with a computer syste related to the purchase of a Hioki product.
ctly on anything hard.
e label printing disappearing.
erature or humidity, as there is a danger
m that arises from the use of this CD, or for any proble m
Page 19
Using a CF Card/USB flash drive
USB method of use
Connector used
[System] screen USB
mode setting Use a USB flash drive. Type A USB Memory (Default) Communicate with the LR8431-20
and initiate measurement using the Logger Utility software from a com­puter (using a USB cable).
Type B USB Communication
Read files on a CF card that is con­nected to the LR8431-20 from a computer (using a USB cable).
Type B USB Drive
• Inserting a CF card/USB flash drive upside down, backwards or in the wrong direction may damage the CF card, USB flash drive, or HiLogger.
• Never eject a CF card /USB flash drive while measuring or when the HiLo
gger is or accessing the card. Data on the CF card/USB flash drive may be destroyed. (The CF icon/USB flash drive icon at the lower right is red while the card is being accessed.)
• Do not transport the HiLogger while a USB flash drive is connected. Dam
age could result.
• As th e CF card/ USB flash drive is sen sitive t o static elec tricit y, damage to the CF card/USB flash driv e or wrong operations by the HiLogger may occur due to static electricity. Please be careful when handling it.
• With some USB flash drives, the HiLogger may not start up if power is tu
rned on while the USB flash drive is inserted. In such a case, turn power on first, and then insert the USB flash drive. It is recommended to try out operation with a USB flash drive before starting to use it for actual measurements.
• The Flash memory in a CF card/USB flash drive has a limited operat­ing life. After long-term usage, data storage and retrieval become diffi­cult. In this case, replace the CF card/USB flash drive with a new one.
• We cannot provide compensation for data loss in a CF card/USB flash d
rive, regardless of content or cause of the damage. Data is also cleared from memory if a long time passes after measuring. Always maintain a backup of important data stored on a CF card/USB flash drive.
• Although real-time saving to USB flash drive is supported, a CF card is r
ecommended for data preserva tion. Performanc e cannot be guaran teed
when using storage media othe r tha n a H iok i-spe cifie d CF ca rd o ption.
• Use a USB flash drive whose continuous current consumption does not e
xceed 300 mA (peak 500 mA). (The peak value is displayed as "Max
Power" under the USB flash drive sel f-tes t on the [Sy ste m] sc reen .)
• Depending on how USB is used, the USB connector and instrument set­tings may vary as shown in the chart belo w.
• The three USB methods of use described in the chart below involve e
xclusive settings and cannot be us ed sim ult an eousl y.
11
Operating Precautions
Page 20
12
Operating Precautions
Page 21
13
Record anomalies
Abnormal events can be ana­lyzed by recording with the trigger function (p. 67) Use the alarm function to out­put signals when anomalies occur (p. 73).
Measurement data and setting configurations are stored, read and managed as data files
Measurement data can be stored to and read from a CF card or USB flash drive (p. 91).
Waveform analysis
mV A
Cursor Measurements (p. 84)
A
B
Using the A/B cursors, values at cursor lo­cations on waveforms and the time at trig­ger positions can be displayed.
Performs six calculation types, including maximum and average values.
Numerical Calculations (p. 113)
123.4V
Acquire pulse signals
Measure integrated pulses or revolutions from sensors and pulse output devices such as watt-hour meters.
Gauge display (p. 84)
Zoom in/out (p. 83)
Numerical display (p. 87)
Analysis on a PC
Monitor and analyze measurement data with the supplied application program (p. A19).
Event search (p. 88)
T
1 2
Display in converted units
Using the scaling function, input val­ues can be displayed in units of the actual physical quantity being mea­sured, such as current or flow vol­ume (p. 58).
Observe voltage fluctua­tions and temperature changes
Just connect measurement leads or thermocouples to mea­sure as needed.

1.1 Product Overview and Features

Overview Chapter 1
1.1 Product Overview and Features
The Hioki LR8431-20 Memory HiLogger is a compact, lightweight, easy-to­operate waveform recorder. It can run on batteries, and can be quickly deployed when a power anomaly occurs. Measurement data can be monitored, subjec on a personal computer using the supplied application program.
ted to calculations and analyzed
1

Chapter 1 Overview

3
Page 22
14
To save data, insert a CF card or USB flash drive. (p. 94)
AC adapter connection (p. 30)
Install the Battery Pack (option) (p. 26)
Connect to the Model LR8431-20 (p. 25), and Set Up (p. 7)
Turn the Power On (p. 34)
Measurement cables connections (p. 31)
1
2
Configure Recording Settings (p. 38)
Using a previously saved
setting configuration
Reload a previously stored setting con­figuration from Model LR8431-20 memory or a CF card or USB flash drive, and measure (p. 103)
Set the data acquisition (recording) inter­val, and recording length (time span). Select the type of measurement data to save when automatically saving to a CF card or USB flash drive. Remaining data stored in the instrument can be saved when finished measuring.
Move to a setting item
Open the setting options
Select Apply
1
2

1.2 Measurement Flow

1.2 Measurement Flow
Installing, Connecting and Turning On
Settings
Page 23
1.2 Measurement Flow
Configure Input Channels (p. 38)
Select input channels, and set input types and measurement ranges. Make other settings as desired.
Move to a setting item
Open the setting options
Select Apply
Press the START/STOP Key (p. 22)
Record once, and stop. [Repeat]: Off
Stop
Measurement
Start
Measurement
When recording using the trigger function, recording occurs only when the input waveform satisfies specified trigger criteria.
Record repeatedly. [Repeat]: On
(default setting)
View Measurement Data (p. 81)
Waveforms can be zoomed and numerical values confirmed.
Save Data (p. 91)
Measurement data, waveform data, screen images and numerical calculation results can be saved.
Calculate (p. 1 1 3)
Numerical measurement data can be applied to calculations.
View on a Computer (p. A19)
Analyze recorded data using the supplied application program.
Turn the Memory HiLogger Off (p. 34)
Starting and Stopping Measurement
15
1
Chapter 1 Overview
3
Data Analysis
When Finished
Page 24
16
Front Panel
Operating Keys
Rear Panel
Battery Compartment (p. 26)
The optional Model 9780 Battery Pack is installed here.
Manufacturer's Serial No.
Shows the serial number.The serial number consists of 9 digits. The first two (from the left) indicate the year of manufacture, and the next two indicate the month of manufacture. Do not remove this label, as it is required for product support.
Display Screen (p. 81)
4.3-inch TFT color LCD Screen Configurations (p. 18)
WAVE/DATA
Selects among waveform screen displays (p. 18).
SET
Displays the Settings screens, and switches among the screen tabs with each press (p. 20).
FILE
Displays file information (p. 19).
CH/
Select channels.
ESC
Cancels changes to settings.
Cursor Keys
Moves the position of the cursor (blinking selection) on the screen.
ENTER
Accepts displayed settings.
KEY LOCK
Disables keypad op­erations. Press and hold the left and right cursor keys simultaneously for three seconds to lock and un­lock the keys.
(Zero Adjust)
Performs zero adjust­ment. Press the up and down keys simul­taneously to execute (p. 35).
Press to save data manu­ally (p. 96).
Press the center key to select waveform scrolling or A/B cursor movement, then press the left and right cursor keys to scroll or move (p. 21).
Setup and display
Saving operations
Scroll waveforms and
read cursor values
Choose a screen
Start and stop measure­ment. The LED at the left lights green while mea­suring (p. 22).
Start and stop
measurement
Operating Keys

1.3 Names and Functions of Parts

1.3 Names and Functions of Parts
Page 25
17
Top Panel
AC Adapter Socket
Connect the supplied Model Z1005 AC Adapter (p. 30).
CF Card Slot
Insert an optional CF card. (p. 93)
Analog Input Terminals
Connect measurement leads for voltage mea­surement, and thermocouples for temperature (p. 31).
CHARGE LED
This LED lights when the battery is charging (p. 29).
Pulse Input Connector
Connect an optional 9641 cable (p. 33).
External Control Terminals
Control signals can be received from and output to external devices (p. 129).
Right Side
POWER Switch
Turns the instrument on and off (p. 34)
Power Off
Power On
USB Port
(USB 2.0 mini-B receptacle)
• Using the included Logger Utility applica­tion software, you can monitor measure­ment data on the computer and transfer settings to the instrument. (p. A26) (To use, set the USB mode on the [System] screen to [USB Communication].)
• You can transfer data on the instrument’s CF card to a computer. (p. 110) (To use, set the USB mode on the [System] screen to [USB Drive].)
USB flash drive Slot (Type A USB 2.0 receptacle)
Used to save data to a USB flash drive. (p. 94) (To use, set the USB mode on the [System] screen to [USB Memory].)
1.3 Names and Functions of Parts
1
Chapter 1 Overview
3
The [USB Memory], [USB Communication], and [USB Drive] USB mode settings are exclusive and cannot be used simultaneously.
Page 26
18
Selects between seven display types.
The screen switches each time you press the key.
Operational infor­mation is displayed along the bottom of the screen.
[Wave] Screen
Measurement data is displayed as waveforms (p. 81).
[Value+Cmnt] Screen
Measurement data is displayed as numerical val ues with com­ments (p. 87).
[Wave+Calc] Screen
Measurement data is displayed as waveforms with calculation results (p. 115).
[Gauge+Wave] Screen
Measurement data is displayed as waveforms with gauges (p.
84).
[Wave+Value] Screen
Measurement data is displayed as waveforms and numerical values (p. 87).
[Value] Screen
Measurement data is displayed as numerical values (p. 87).
[Wave+Crsr] Screen
Measurement data is displayed as waveforms with cursor val­ues (p. 84).
Waveform/Numerical Screens
Selection is also available from the name of the current screen displayed near the bottom right
[Gauge+Wave]
[Wave] [Wave+Value] [Value+Cmnt]
[Value]
[Wave+Calc] [Wave+Crsr]

1.4 Screen Configurations

1.4 Screen Configurations
Page 27
19
Waveform/Numerical Screens (About the Icons)
Clock
"Setting the Date and Time" (p. 124)
Power source indicator
Indicates the Memory HiLogger's power source.
AC adapter operation Battery pack operation
(
Fully charged battery pack) Battery pack operation Battery pack operation
(Low battery indicator)
CF card
Displayed when a CF card is inserted. The icon appears red when accessing the card.
USB source indicator
Indicates the USB status.
Operating in USB Memory mode.
Operating in USB Communication mode.
File Screen
View and manage files on the CF card or USB flash drive (p. 91).
File Screen
Operational infor­mation is displayed along the bottom of the screen.
1.4 Screen Configurations
1
Chapter 1 Overview
3
Page 28
20
[CH] Screen
Make input channel settings while viewing the monitor dis­play (p. 49).
[Scaling] Screen
Make these settings to convert measured values to arbitrary units for display (p. 58).
[Comment] Screen
Enter channel comments (p. 60)
[Setting] Screen
[Range] Screen
Make settings while viewing all channel settings (p. 62).
[Trig & Alm] Screen
Recording criteria (triggering) and warning sounds can be set for each channel (p. 67).
[System] Screen
Configure the system environ­ment (p. 119).
Settings Screens
Press the left/right cursor keys to se­lect between the Settings screens.
Selects between seven display types.
The screen switch­es each time you press the key.
Operational infor­mation is displayed along the bottom of the screen.
Make settings for recording (p.
42). Set numerical calculation, auto-saving and timers.
1.4 Screen Configurations
Page 29
21
Select the item to change.
Show available setting options.
Select the desired setting.
Changing screen contents
Setting Screen
Waveform Screen
Apply the new setting, or cancel it.
Scrolling a waveform
Earlier
Later
See: "Scrolling Waveforms" (p. 82)
Reading values at cursors
1
Display the A/B cursors.
2
Move the A/B cursors.
Values at cursor positions can be read numerically.
You can select the cursor display type and which cur­sor(s) to move (p. 84).

1.5 Basic Operation

1.5 Basic Operation

Screen Operations

1
Chapter 1 Overview
3
Page 30
22
Press the START/STOP key. The green LED lights. When using the trigger function, the times when measurement
starts and when recording (data acquisition) starts are not the same.
A confirmation message appears when you press the
START/
STOP
key to start recording. Save important data to a CF card or USB flash drive before starting, because previous measure­ment data will be overwritten when recording starts.
Start Measurement
When [Repeat] (record multiple instances of the specified recording length) or
[Cont] (record continuously) is
enabled:
Pressing the
START/STOP key stops recording.
When
[Repeat] and [Cont] are disabled (Off):
Recording stops after the specified recording length. Pressing the
START/STOP key during recording displays a confirmation
message. Select
[Yes] to stop recording.
For details about repeating and continuous recording:
See: "Selecting the Recording Start/stop Method" (p. 44)
"Set the Recording Length for Repeated or One-time Record­ing" (p. 45)
Finish Measurement
1.5 Basic Operation

Starting and Stopping Recording

Acquire measurement data on the Memory HiLogger, and start recording. When recording stops depends on data recording settings (repeat recording, t
riggering, timers, etc.). (p. 78)
When you want measurement data to be automatically saved, before starting recording, check that au is correctly inserted (p. 94), and that sufficient storage capacity is available (p. A11).
to-save is enabled (p. 46), that a CF card or USB flash drive
Page 31
23
Press and hold the right and left cursor keys for three seconds to lock the other keys, and repeat to unlock.
Insert a CF card or USB flash drive.
Press the
SAVE key.
A dialog appears.
Select the save destination and item to save, and press ENTER key. Select
[Yes] in the confirma-
tion dialog, and press ENTER again to save the data to the CF card or USB flash drive.
1.5 Basic Operation

Disabling Key Operations (Key-Lock Function)

Keys can be disabled to avoid inadvertent operations.

Saving Data

Measurement data, settings, screen images can be saved. For details of the saving methods, see Chapter 6, "Saving & Loadin g Data" (p.
91).
(Default settings) Saving Method: [Select & Save] (Press the SA data.)
and numerical calculation results
VE key to select and save
1
Chapter 1 Overview
3
• When using a USB flash drive, set the USB mode on the [Sys ­tem] screen to [USB Memory]. USB flash drives cannot be used while
the USB mode setting is [USB Communication] or [USB
Drive].
• The [USB Memory], [USB Communication], and [USB Drive] USB mo neously.
de settings are exclusive and cannot be used simulta-
Page 32
24
Current Input Signal Level and Channel Number
Level Monitor
Display Channel and Gauge
Shows the gauge for the current­ly selected channel. The gauge display is linked to the measure­ment and display range settings. Press the
CH / keys to dis-
play the gauge for another chan­nel.
Changing the Monitor Display
Select whether to display all channels, or only se­lected channels. [All Ch]/[SlctCh]/[Value]
[All Ch] The channels displayed in the level monitor are those with a check mark ( ) displayed next to
the channel number in the settings.
[SlctCh] (Selected Channels) The level monitor displays the channel selected on this screen.
[Value] The input values of all channels are displayed numerically.
Displayable Waveform Range
This is determined by the waveform display range setting.
Input Waveform Amplitude
Indicates real-time input ampli­tude. Using the Scaling function (p. 58), measurement units can be converted as desired for display.
1.5 Basic Operation

Verifying the Input Level (Level Monitor)

You can verify the input status and display range while making settings on the Setting screen. This is not available while measuring.
Page 33
Measurement
Rear View
Install the 9780 Battery Pack (option) (p. 26)
1
3
(To save data) Insert a CF card or USB flash
drive (p. 93)
4
2
Connect measurement cables (p. 31)
Pulse signals: Connect the Hioki 9641 connection cable
Voltage measurement: Connect the measurement leads.
Temperature measurement: Connect thermocouples.
External control signal inputs
External control signal outputs (p. 129)
Turn the power on (p. 34) Execute zero adjustment (p. 35)
5
Connect the AC adapter (p. 30)
25
Preparations
Chapter 2
2

Chapter 2 Measurement Preparations

3
Page 34
26

2.1 Using the Battery Pack (Option)

2.1 Using the Battery Pack (Option)
If commercial power is not available when the AC adapter is connected, the LR8431-20 Memory HiLogger can operate from the Model 9780 battery pack, so when using commercial power, the battery pack serves as a backup supply during power outages. Charge the battery pack fully before using it the first time.
Be sure to observe the following precautions. Incorrect han­dling may result in liquid leaks, h bursting and other hazards.
• Do not use a battery pack other than the 9780 Battery Pack. We cannot accept responsibility related to the use of any other batteries.
• To avoid the possibility of explos disassemble or incinerate battery pack.
• When storing the instrument, make could short-circuit the connectors are placed near them.
• The battery pack contains lye, whi comes into contact with the eyes. Should battery liquid get into your eyes, do not rub your eyes but rinse them in plenty of water and seek the immediate attention of a physician.
eat generation, ignition,
for accidents or damage
ion, do not short circuit,
sure no objects that
ch can cause blindness if
Installing and replacing the battery pack
• To avoid electric shock, turn of connect the cables before replac
• After installing or replacing the battery pack, replace the battery comp
• Handle and dispose of batteries regulations.
artment cover and screw.
f the power switch and dis-
ing the battery pack.
in accordance with local
Page 35
27
2.1 Using the Battery Pack (Option)
Observe the following to avoid damage to the instrument.
• Use the battery pack in an ambient temperature range of 0 to 40°C a
nd charge it in an ambient temperature range of 5 to 30°C
to be on the safe side.
• If the battery packs fails to complete charging within the stipu­lated time, disconnect the AC adapter from the pack to stop
arging.
ch
• Consult your dealer or nearest se leaks, strange odor, heat, discoloration, deformation and other abnormal conditions occur during use, charging or storage. Should these conditions occur during use or charging, turn off and disconnect the instrument immediately.
• Do not expose the instrument to water and do not use it in exces­sively humid locations or locations exposed to rain.
• Do not expose the instrument to strong impacts and do not throw
nd.
it arou
• The battery pack is subject to self-discharge. Be sure to charge the battery pack before initial use.
• When recharging a discharged battery pack in the LR8431-20, allow it to charge for at least ten minutes with the LR8431-20 turned off to maximize battery longevity.
• The battery pack is a consumable. If the battery capacity remains very low end. It can be recharged up to about 500 times.
• To prevent battery pack deterioration when the battery will not be u
sed for 1 month or longer, remove it and store it in a dry location with an ambient temperature range of between -10 to 30°C. Be sure to discharge and charge it every two months. Long-term storage when capacity has decreased will make charging impos­sible and reduce performance.
• Remove the battery pack from instrument once a month and che
• As long as battery charge remains, waveform data is retained in inte battery continues to gradually discharge. When the battery charge is depleted, stored waveform data is lost.
after correct recharging, the useful battery life is at an
ck that appearance is normal.
rnal memory while the Memory HiLogger is turned off, so the
rvice station should liquid
2
Chapter 2 Measurement Preparations
3
Page 36
28
When to charge
When powering the instrument from the battery pack without the Model Z1005 AC Adapter, the low battery indicator ( ) is displayed when the battery charge is depleted, indicating that the battery pack requires charging. Note that if the Memory HiLogger is turned off under this situation, waveform data may fail to be backed up, and could be lost.
Approximate charging time:
About 200 minutes at 23°C when recharged at the first appearance of the low battery indicator.
Approximate continuous battery-only operating time
(When saving data to a CF card at a temperature of 23°C) * When using a USB flash drive, the continuous operating time may decrease
depending on the particular USB flash drive in use.
• After a full charge, with the LCD backlight always on (default setting): approximately 2.5 hours
• After a full charge, with the LCD backlight off (five minutes after last key­press): approximately 3 hours
See: "Enabling and Disabling the Backlight Saver" (p. 121)
2.1 Using the Battery Pack (Option)
Page 37

Install the Battery Pack

Required tool:
One Phillips screwdriver (No.2)
Important
Disconnect the AC adapter when installing the battery pack.
Battery Compartment Cover
1
Place the Memory HiLogger face down, and remove the battery com­partment cover screw and cover.
2
Connect the plugs of the battery pack to the connectors and insert the battery pack with the label side up.
Replace the bat­tery compartment cover and screw.
3
Pull the battery pack plug straight out to remove the battery pack.
Make sure that the cables are not pinched.
2
Be sure to properly seat the connectors.
1
Connect the power cord to the inlet socket on the AC adapter.
1
2
Connect the output plug of the AC adapter to the instrument.
Plug the power cord into the mains outlet.
3
Refer to 2.2, "Connecting the AC Adapter" (p. 30) for details about the AC adapter.
The CHARGE LED lights red when charging begins. Charging is finished when the CHARGE LED turns off.
29
2.1 Using the Battery Pack (Option)
2
Chapter 2 Measurement Preparations
3

Charge the Battery Pack

Regardless of whether the Memory HiLogger is on or off, the battery pack recharges whenever the Z1005 AC Adapter is plugged in to a power source. Therefore, charging is provided by merely keeping the battery installed in the Memory HiLogger .
Page 38
30
Connect the power cord to the inlet socket on the AC adapter.
Rated supply voltage is 100 to 240 VAC, and rated supply fre­quency is 50 or 60 Hz.
1
2
Connect the output plug of the AC adapter to the instrument.
Plug the power cord into the mains outlet.
3

2.2 Connecting the AC Adapter

2.2 Connecting the AC Adapter
Connect the power cord and the instrument to the supplied Model Z1005 AC Adapter, then plug the power cord into an outlet. When used with the battery pack installed, the battery serves as an operating backup supply in case of power failure, and the AC adapter otherwise has priority.
Before Connecting
• Use only the supplied Model Z1005 AC Adapter. AC ada pter input voltage range is 100 to 240 VAC (with ±10% st ability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this range.
• Turn the instrument off before connecting the AC adapter to the instrument a
• To avoid electrical accidents and to maintain the safety specifications of this instr only to a 3-contact (two-conductor + ground) outlet.
To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it from the power outlet.
nd to AC power.
ument, connect the power cord
Page 39

2.3 Connecting Measurement Cables to the Memory HiLogger

6 mm max.
6 mm max.
2.3 Connecting Measurement Cables to the Memory HiLogger
31
Connect the appropriate cables for the intended recording application.
• When recording voltage fluctuations or temperature changes (p. 32)
• When recording pulse signals (p. 33)
Be sure to read the "Operating Precautions" (p. 6) before connecting the cabl es.
Connecting to the Analog Input Terminals
• The maximum input voltage (and the maximum rated volt­age to earth) for the analog input terminal V DC). If the voltage is exceeded, this device will be dam­aged and personal injury will form measurement in this case.
• Do not leave the Memory HiLogger connected to test objects in environments where a voltage surge might exceed the dielectric withstand voltage. Doing so could result in damage to the Memory HiLogger, bodily injury or fat al accident.
• Do not connect a cable to the instrument while it is con­nected to the object to be meas shock accident may occur.
• To avoid short-circuit accidents, make certain that connec­tions are secure.
• Channels are insulated by semiconductor relays. When a voltage beyond the specification i the channels, the semiconductor relay may short circuit. Please ensure that a voltage beyond specification, espe­cially a surge such as a lightning, is nev abnormal measurement value is observed, please contact your dealer or Hioki representative for inspection.
result. Therefore, do not per-
ured. Otherwise, an electric
s is 30 Vrms (or 60
s applied between
er applied. When an
2
Chapter 2 Measurement Preparations
3
• Measurements may be affected by noise or other electromag­netic ingress if input leads are longer than about three meters.
• Route measurement leads and cables as far away as possible from power and earth lines.
• Measurement values may be scattered if the leads are routed in
arallel with other devices. If they must be parallel, check opera-
p tion before taking measurements.
• Avoid temperature changes around the terminal blo directed airflow such as from an electric fan or air-conditioner vent.
• If connecting crimped terminals to the analog input terminals, use insu
lated terminals for M3 screws with the size shown below .
ck. Especially avoid
Page 40
32
Connect to the measurement object.
2
Using the screwdriver, loosen the terminal block screws, insert the bare ends of the leads, and tighten the screws.
Analog Input Terminals
1
When applying 4 - 20 mA input current
Connect a 250 shunt resistor as shown.
Tools required: Phillips screwdriver (for M3 screws)
(For details on instrumentation control equipment measure­ments, see p. 50 .)
The insulation colors of wires, which are to be connected to the plus and minus terminals, differ depending on countries or manufactures. Check with the manu­facturer on the colors and connect them according to instructions.
2.3 Connecting Measurement Cables to the Memory HiLogger

Connecting Measurement Leads or Thermocouples (for voltage or temperature, respectively)

Connect the voltage and temperature measurement (thermocouple) leads to the analog input terminals.
K,E thermocouples experience a physicalphenomenon known as short range ordering, which can prevent accurate measurement within the range of 250°C to 600°C. Please choose a sensor after checking with the manufacturer of the thermocouple you intend to use.
Page 41
2.3 Connecting Measurement Cables to the Memory HiLogger
Measurement Object
Memory HiLogger
Connection Cable
P1 White (H) P1 Red (L)
P2 Yellow (H) P2 Black (L)
P3 Green (H) P3 Blue (L)
P4 Gray (H) P4 Brown (L)
Ends of the 9641 Connection Cable
(connects to Memory HiLogger common ground)
Shield
Connect the other end of the cable to the pulse source(s) to be mea­sured.
Pulse Input Connector
1
2
Connecting to the Pulse Input Connector
To avoid electric shock and short circuit accident s or damage to the instrument, pay attention to the following:
• The ground terminal for the optional 9641 Connection C
able is not isolated from the Memory HiLogger's ground (chassis ground). Therefore, if the measure­ment object connects to A
C, it should have a grounded, polarized plug, and be con­nected to the same power outlet as the Memor AC adapter. If the Memory HiLogger and me nected to different mains circuit power cord is used, the potential difference between the dif­ferent grounding paths may result in current flow through a c
onnection cable that could damage the measurement
object or Memory HiLogger.
• The input voltage range limits are 0 are exceeded, the instrument could be damaged and per­sonal injury or death could occur, so do not attempt mea­surement.
33
y HiLogger’s
asurement object are con-
s, or if an ungrounded
to +10 V. If these limits
2
Chapter 2 Measurement Preparations
3
Connecting the Optional Model 9641 Connection Cable (for Pulse Signal Input)
Connect the optional 9461 connection cable to
For safety, we recommend using only Hioki’s connection cable (measurement lead cable).
the pulse input connector.
Page 42
34
Right Side
Turn the power switch on ( ).
If the " " indicator is not displayed, the instrument is not being powered from the AC line. In this state, the battery charge will be depleted if measuring for a long time, and measurement may be interrupted. Verify that the AC adapter is securely connected to an AC power source and to the instrument.
Turning Power On
Turn the power switch off ( ).
The installed battery pack is charged whenever the AC adapt­er is plugged into a power outlet, even when the Memory HiLogger is off. In addition, providing there is some remaining battery charge, waveform data and the setting configuration are memorized whenever the Memory HiLogger is turned off, so that when the Memory HiLogger is turned back on, the same operating state is displayed.
Verify that the instrument and peripheral devices are correctly connected.
Turning Power Off
Execute zero-adjust 30 minutes after turning power on (p. 35).
Initial screen
The power supply icon appears near the bottom right corner of the screen (p. 19).

2.4 Turning the Power On and Off

2.4 Turning the Power On and Off
Using the AC Adapter
Before turning the instrument on, make sure the supply volt­age matches that indicated on its power connector. Connec­tion to an improper supply volt instrument and present an electrical hazard. Rated supply voltage is 100 to 240 VAC (Voltage fluctuations of ±10% from the rated supply voltage ar
e taken into account.)
Rated supply frequency is 50 or 60 Hz.
age may damage the
Page 43
35
Turn on the power and wait for 30 minutes to stabilize the internal temperature of the Memory HiLogger.
With the Waveform or [CH] screen displayed, press and hold these keys simultaneously for one second.
The Zero Adjust dialog appears.
From the Setting screen
Select [Zero Adjust].
Apply
1
2
Zero adjustment is executed.
(simulta­neously)

2.5 Zero Adjustment

2.5 Zero Adjustment
Zero adjustment corrects for volt age offset at the input terminals, so that Mem­ory HiLogger measurements are relative to zero volts. Perform the zero adjustment procedure when tion where the ambient temperatu
re is markedly different.
using the instrument in a loca-
2
Chapter 2 Measurement Preparations
3
Any zero-adjust setting is cleared upon system reset (p. 125).
Page 44
36
2.5 Zero Adjustment
Page 45
Setting before
Is the insulation of the measure­ment cable to be used damaged, or is bare metal exposed?
1
2
Peripheral Device Inspection
Before using the instrument the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Metal Exposed
No Metal Exposed
Do not use if damage is present as you could re­ceive an electric shock. Replace the damaged items.
To
If damage is evident, request repairs.
Yes
Is damage to the instrument evident?
Instrument Inspection
When turning power on
Does the HIOKI logo appear on the screen?
No
2
The power cord may be damaged, or the instru­ment may be damaged internally. Request repairs.
Is the Waveform or Setting screen displayed? (When power is turned on the first time after purchase, the Set­ting screen should appear.)
No
Yes
The instrument may be damaged internally. Request repairs.
Nothing appears, or the display is abnormal
Yes
Inspection complete
37

3.1 Pre-Operation Inspection

Measurement
Chapter 3
3.1 Pre-Operation Inspection
3

Chapter 3 Setting before Measurement

3
Page 46
38
Record
Display the Setting screen
Configure measurement settings
Connect
Configure
Connect measurement cables (p. 31).
Configure input channels (p. 49)
Analyze
Analyze recorded data on the Waveform screen (p. 81). Calculations can be applied to measured data (p. 113).
Save
Select and save data (p. 91).
When configuration is finished, start recording (p. 22).
Reload stored
settings (p. 103)
[Setting] screen
[Setting] screen
[CH] screen
Configure data
saving
[Setting] screen
(as needed)
Configure data recording settings (p. 42)

3.2 Operating Flow Overview

3.2 Operating Flow Overview
Page 47
39
. Must be set before measuring.
Make other settings as needed.
Save automatically
while measuring (p. 46)
Reloading Settings
(p. 103)
Saving Settings
(p. 102)
Enter a title for mea-
surements (p. 60)
Filter out noise
(p. 45)
Perform numerical
calculations (p. 113)
Record by timer
(p. 76)
Configure recording settings: [Setting] screen
Data recording (acquisition) interval (p. 42)*Display time base (time per horizontal division) (p. 43)Recording start/stop method (p. 44)*Recording length (time span) for repeated or one-time r e-
cording (p. 45)*
Horizontal Axis (Time Axis) Settings
3.2 Operating Flow Overview
Measurement Configuration (Setting Screen)
Before measuring, configure the measurement settings on the [Setting] and
[CH] screens. A brief operational description of the currently selected (high-
lighted) setting item appears a
t the bottom of the screen.
Three recording methods are available. T between specified start and stop times, can be use d in combination with these methods (p. 76).
Recording Method Description Normal Recording
Continuous Recording
Repeat Recording
Records for the specified recording length beginning when
measurement starts (Cont: Off).
Records continuously beginning when measurement
rts (Cont: On).
sta Records repeatedly for the specified recording length
(time spa
n) (Repeat: On).
imed recording, which record
3
Chapter 3 Setting before Measurement
3
Page 48
40
Make other settings as needed.
Entering channel comments (p. 60)Selecting channel display colors (p. 56)Setting the display range
Specifying the display position (p. 56) Specifying upper and lower display limits (p. 57)
Recording specific events (triggering) (p. 68)Setting alarm output (alarm function) (p. 73)Setting Unit Conversion (Scaling) (p. 58) Switching the monitor display (p. 24)
Configure input channel settings: [CH] screen
Voltage measurement (p. 50)Temperature measurement
(p. 51)
Integration (pulse count) (p. 53)Revolution measurement
(p. 54)
Input Channel Settings
3.2 Operating Flow Overview
Page 49
41
Specify the recording interval (data acquisition
interval) (p. 42)
Specify the display time base (time per hori-
zontal division) (p. 43)
Enable digital filtering (noise suppression)
(p. 45)
Waveform Screen
Horizontal Axis Settings
Voltage measurement (p. 50)Temperature measurement (p. 51)Integration (pulse count) (p. 53)Revolution measurement (p. 54)Alarm settings (p. 73)
Setting the display range
Specifying the display position (p. 56) Specifying upper and lower display limits (p. 57)
Select the alarm display type (p. 84)Eject (replace the CF card/USB flash drive
during real-time save operation)(p. 48)
Displaying any point in time (Jump function)
(p. 83)
Waveform Display Settings
Input Channel Settings
About Channel Settings
Pressing the CH / keys selects channel setting and gauge display items. However, while measurement data is displayed, such changes affect only those channels selected (by a check mark [ ]).
3.2 Operating Flow Overview
Measurement Configuration (Waveform Screen)
Certain setting items are available on the W aveform screen. Setting det ails are the same as for those on the Setting screen.
3
Chapter 3 Setting before Measurement
3
Page 50
42
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
Select the [Setting] screen.
Select from the [Interval] setting options.
Setting options:
10ms, 20ms, 50ms, 100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s, 1min, 2min, 5min, 10min, 20min, 30min, 1h ( default setting)
Some waveform peaks may not be displayable with certain interval settings. For example, when the [Interval] is 1 s and the [Time/ DIV] (time base) is 5 s.
5s/div
1s
Peak
Broken wire detection is automatically disabled when the interval is set to 10, 20 or 50 ms (p. 51).
3.3 Setting Configuration 1 – Recording Settings
3.3 Setting Configuration 1 – Recording Settings
Configure the horizontal axis and recording settings on the [Setting] screen. To save data while measuring, auto save mus
t be enabled beforehand (p. 46).

Specify the Recording Interval

Select the data acquisition interval to suit your measurement objectives. Note
horter recording intervals restrict the maximum recording time.
that s
Page 51
3.3 Setting Configuration 1 – Recording Settings
1 div
Select from the [Time/DIV] setting options.
Setting options:
100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s, 1min, 2min, 5min, 10min, 20min, 30min, 1h, 2h, 5h, 10h, 12h, 1day [/DIV] ( default setting)
When auto save is enabled for waveforms, 100, 200 and 500 ms settings are automatically changed to 1 s when measurement starts, and the shorter values cannot be se­lected while measuring.

Setting the Display Time Base (horizontal axis magnification, set as needed)

Displayed waveforms can be expanded or compressed along the horizontal axis. Shorter setting values corre­spond to greater magnification. This changed both during and after measurement without affecting acquired data (p. 83).
Available display time base settings depend on the selected recording interval. Horizontal axis display options are affected when the recording interval is changed af
ter measuring, because the options for expanding (or compressing) the displayed
axis depend on the recording interval used when measuring.
Hori-
zontal
axi
s
10 20 50 100 200 500 1 2 5 10 20 30 1 2 5 10 20 30 1
100ms 200ms 500ms 1s 2s 5s 10s 20s 30s 1min 2min 5min 10min 20min 30min 1h 2h 5h 10h 12h 1day
                                                                                                             
                                                                          
[ms] [s] [min] [h]
: Selectable, ×: Not selectable, : Not selectable while measuring
Recording interval
× × × × × × × × × × × × × × × ×
× × × × × × × × × × × × × × × × × × × × × × × × × × × × × ×
× × × × × × × × × × × × ×
× × × × × × × × × × × × × × × × × × × × × × × ×
43
setting can be
3
Chapter 3 Setting before Measurement
3
× × × × × × × × × ×
× × × × × × × × × × × × × × × × × ×
× × × × × × ×
× × × × × × × × × × × ×
× × × ×
× × × × × ×
×
Page 52
44
Continuous Recording
Repeat Re­cording (speci­fied lengths)
Start Measurement
Stop Measurement
To record continuously, Enable continuous recording (set
[Cont] to
[On]). (default setting)
Recording continues until you press the START/STOP key, or until a “stop trigger” event occurs. When auto save is enabled for measurement data, the saving method depends on whether [Deleting] or [Split Save] (segmented save) is selected (p. 46).
To record for a specified length (time span),
1 Set [Cont] to [Off]. 2 Specify the duration of one recording length
(time span).
Setting options: d (days) 0 to 499, h
(hours) 0 to 23,
min (minutes) 0 to 59, s (
seconds) 0 to 59
1 2
Recording proceeds for the specified length. Record­ing stops early if you press the
START/STOP key be-
fore the recording length has elapsed, or if a “stop trigger” event occurs.
Change 1’s digit Change 10’s digit
Maximum recording length and data samples
Indicates the maximum record­ing time to internal memory, and the maximum number of data samples. These values are limit­ed by the specified recording length and the number of mea­surement channels used.
See: Appendix 6, "Maximum
Recordable Time" (p. A11)
3.3 Setting Configuration 1 – Recording Settings

Selecting the Recording Start/stop Method

Specify how you want recording to start and stop. The following methods are available.
• Record continuously: Record continuously between starting and stop­ping measurement.
• Specified length (time span): Record only for a specified time span (can be set to
automatically repeat) (p. 45)
Page 53
3.3 Setting Configuration 1 – Recording Settings
Repeated recording
On
Off
Select from the [Repeat] setting options.
Setting options:( default setting)
Off
Recording stop after one recording length (time span).
On Recording restarts after the specified recording
length has elapsed or after recording is stopped by a “stop trigger” event. To manually stop re­cording, press the START/STOP key.
Dead time
When repeating recording is enabled ([Repeat] set to [On]) After a recording length has elapsed, some “dead time” is needed for internal processing before recording can re­sume. Because recording is not available during the dead time, enable continuous recording (set [Cont] to [On]) if you need to record without data loss.
Recording
Select from the [Filter] setting options.
Setting options:60Hz, 50Hz, Off ( default setting)
fc: cut-off frequency
When the 60 Hz digital filter is selected, optimum line noise suppression occurs when the recording length is between two seconds and one hour. Noise is suppressed with either 50 Hz or 60 Hz filter selec­tions, although the measurement timing difference between channels may be increased, depending on the recording length. Disabling the filter (set to Off) minimizes such timing differences (to less than 1 ms between channels).
When disabled (set to Off), broken wire detection is also disabled (p. 51).

Set the Recording Length for Repeated or One-time Recording

Select whether recording should be resumed after the specified recording length has
sed or after recording is stopped by a
elap “stop trigger” event.
45
3
Chapter 3 Setting before Measurement
3

Enable Digital Filtering (noise suppression) (as needed)

Electrical noise mixed w be removed by digital filtering. Noise suppres­sion is most effective for longer recordings,
re it can provide higher measurement
whe accuracy and less scattering. Set as needed to suit your measurement environment.
See:"Cut-off frequency" (p. 141)
ith input signals can
Page 54
46
3.3 Setting Configuration 1 – Recording Settings

Automatic Saving

Measurement data can be automatically saved to a CF card or USB flash drive while measuring. The available types and formats of automatically saved data are as follows.
Saved Content
Waveform
ata only
d
Calcula­tion results on
ly *1
Waveform d
ata and calcula­tion results *1
Setting Waveform
(realtime)
CSV (realtime)*2
Calc (post meas.)
Waveform + Calc
CSV + Calc*2
File Extension
.MEM
.CSV
.CSV
.MEM, .CSV
.CSV, .CSV
Description Waveform data is saved to the CF card while re-
cording. Waveform data is saved in CSV text format while
r
ecording. Saved data cannot be reloaded in the
HiLogger. After recording, calculation results are saved to
the
CF card. When [Re sults for each recording are included in the record­ed data file.
Waveform data is saved while recording, and cal­culation results are saved when recording finishes.
Waveform is saved in CSV text format while re­cording, and calculation results are saved when r
ecording finishes.
Saved data cannot be reloaded in the HiLogger.
peat] is enabled ([On]), calculation re-
*1.Numerical calculation settings must be made before starting recording (p. 113). *2.The [CSV (real time)] and [CSV + Calc] settings are not available when the
interval is 10ms or 20ms.
Also, when opening a CSV file in a spreadsheet program,
record
note that the number of rows that can be loaded at one time is limited (p. A7).
Page 55
47
1 Select which content to save from the [Auto
Save] settings.
Setting options: Off*, Waveform(realtime), CSV(re-
altime), Calc
(post meas.), W aveform + Calc, CSV
+ Calc ( default setting)
When [Waveform(realtime)], [CSV(realtime)],
[Waveform + Calc], [CSV + Calc] is selected
2 (As needed) You can enter a file name for the
waveform files.
See: 3.6, "Entering Comments" (p. 60)
If left blank, file names are created automatically.
See: Appendix 2, "File Naming" (p. A7)
3 Select the removable storage priority (CF card
or USB flash drive)
4 Select the [Deleting] method for saving data when
the CF card or USB flash drive becomes full.
Setting options:( default setting)
5 Select whether to create a new file for each
recording length, or to create new files at specified intervals (using
[Split Save]).
Setting options:( default setting)
Off* Saving stops when the CF card or USB flash
dr
ive becomes full.
On When space on the CF card or USB flash drive
de
clines to a certain level when starting or dur­ing automatic saving, the oldest file is deleted. If the
file cannot be deleted, saving stops. Only waveform files can be deleted in this way. Appendix 2, "File Naming" (p. A7)
Off* One file is created for each recording
length.
On Specify the length (time span) of saved
s
egments. A measurement data file is cre-
ated for each segment.
Ref TimeSpecify a reference time and segment
le
ngth (time span). A file is saved after each segment time span beginning at the reference time.
1 2
3 4
[Split Save: On]
[Split Save: Ref Time]
Ref Time (reference time): 0:0 to 23:59 Length (Segment length): 1, 2, 5, 10, 15, 20, 30 minutes, 1, 2, 3, 4, 6, 8, 12 hours, 1day
Segment length: Days (0 to
30), Hours (0 to 23), Minutes (0 to 59)
When starting measurement, the message “Sync fixed time” is displayed until the re­cording interval synchronizes with specified reference time.
Horizontal axis settings are in­hibited during automatic sav­ing of waveforms (p. 43).
5
3.3 Setting Configuration 1 – Recording Settings
3
Chapter 3 Setting before Measurement
3
Regardless of the split save setting, the file will be automatically split when the file size nears 2 GB.
Page 56
48
Move the cursor to [EJECT].
Execute
Insert a formatted removable storage.
Measurement data acquired while the de­vice was removed is automatically written.
See: Formatting: (p. 95)
Insertion: (p. 94)
Remove the USB flash drive (p. 94).
Right bottom of the screen
Move the cursor to
[Unmount USB memory].
Execute
Move the cursor to [Yes].
Execute
Bottom of the screen
When both a CF card and USB flash drive are in­stalled and either is removed, the save destina­tion automatically switches to the remaining media.
Execute
The media icon display will change to the eject display.
3.3 Setting Configuration 1 – Recording Settings

Replacing Removable Storage During Real-Time Saving

During real-time saving, removable storage can be replaced without interrupting mea­surement. This procedure describes replacing a USB flash drive.
• If a storage device is not replaced within two minutes after selecting "Yes" in the "Do you want to remove the USB memory/CF card?" con firmation dialog, data may overflow internal memory and be lost.
• If a storage device is replaced during real-time saving, data is saved to a new file.
• If measurement ends while the removable storage is ejected, even if a
removal storage is then inserted, the remaining data is lost. If th i s o c cu r s, th e m e as u re me n t d a ta r em a in in g on t he instrument can be retrieved with the manual save function.
See: 6.3, "Saving Data" (p. 96).
-
Page 57
49
Select the [CH] screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed op­tions.
1 2
Level monitor (p. 24)
Use the CH / keys to change the selected channel.
3.4 Setting Configuration 2 – Input Channel Settings
3.4 Setting Configuration 2 – Input Channel Settings
Configure the input channel settings on the [CH] screen. You can make these settings while viewing input signal amplitude on the level monitor. There are two general types of input signals, analog channels CH1 – CH10 a
nd pulse channels P1 to P4. The analog channels can be used to measure voltage and temperature (with thermocouples), and the pulse channels can be used for pulse counting (integration) and revolution measurement.
Analog channels Pulse channels
"Voltage Measurement Settings" (p. 50)
"Temperature Measurement Settings" (p. 51) "Revolution Measurement Settings" (p. 54)
Settings can be changed while viewing all channel settings on the [Range], [Sc
aling], [Trig & Alm], and [Comment] screens. In this example, settings are made on the [CH] screen while viewing the level mo
nitor. You can specify recording criteria with the input setting or to output alarm signals. See Chapter 4, "Specifying Criteria for Me ments (Triggering, Alarm and Timer)" (p. 67) for details.
"Integration (Count) Measurement Settings" (p. 53)
s to control recording,
asure-
3
Chapter 3 Setting before Measurement
3
Page 58
50
1 Select a channel (CH 1 to CH 10) for the
[Channel] setting contents.
2 To enable measurement on the channel,
select On to display the check mark ( ).
3 Select [Voltage] for the [Input] (input type)
setting.
4 Select the appropriate measurement ampli-
tude range for the
[Range] setting.
Setting options:
100mV, 1V, 10V, 20V, 100V, 1-5V
( default setting)
1
2 3 4
: On : Off
Instrumentation and Control Device Measurements
For instrumentation control signal measurements, select [1-5 V]. This range corre­sponds to the 4-20 mA output standard for instrumentation control. This setting automat­ically sets 1- and 5-volt lower and upper limits, respectively, in the 10 V range. Measurement accuracy is ±10 mV ±5 mV. To measure 4-20 m A current, connect a 250 shunt resistor between + and – analog input terminals (p. 32).
Use the scaling function to scale measurements to any arbitrary unit value for display (p.
58).
3.4 Setting Configuration 2 – Input Channel Settings

Voltage Measurement Settings

Configure these settings for each voltage measurement channel.
Page 59
3.4 Setting Configuration 2 – Input Channel Settings
1 Select a channel (CH 1 to CH 10) for the
[Channel] setting contents.
2 To enable measurement on the channel,
select On to display the check mark ( ).
3 Select [Tc] for the [Input] (input type) set-
ting.
4 Select the type of thermocouple to be
used.
Setting options:
K, J, E, T, N, R, S, B ( default setting)
5 Enable or disable [Burn Out].
Setting options:( default setting)
6 Select the reference junction compensa-
tion (RJC) method for the
[RJC] setting.
Setting options:( default setting)
Off
Do not detect broken (disconnected) wires. Normally when the thermocouple is dis­connected, the value will fluctuate, etc. a
nd will not be stable.
On Detect broken wires.
(This setting is available only when the re­cording interval is at least 100 ms, and the 5
0 Hz or 60 Hz digital filter is enabled.) When the wires are disconnected, “OVE
R” will be displayed as the measure-
ment value.
Int
Use the Memory HiLogger’s internal com­pensation. Measurement accuracy is the su
m of the temperature measurement ac-
curacy and the RJC accuracy.
Ext Use an external RJC method instead of
th
e Memory HiLogger’s internal compen­sation. Memory HiLogger measurement a
ccuracy is equivalent to the temperature
measurement accuracy.
12
: On : Off
Temperature Measurement Range (Depends on thermocouple type)
JIS C 1602-1995
With type B, the display shows 0 to 400°C, but accuracy is not guaranteed.
K
-200°C to 1350°C
J
-200°C to 1200°C
E
-200°C to 1000°C
N
-200°C to 1300°C
T
-200°C to 400°C
R
0°C to 1700°C
S
0°C to 1700°C
B
400°C to 1800°C
34 5 6

Temperature Measurement Settings

Configure these settings for each temperature measurement channel.
51
3
Chapter 3 Setting before Measurement
3
Page 60
52
[C]
1000
400
0
-200
-1000
If Thermocouple Measurement Range is Exceeded
Example: Using a type T thermocouple with range -200 to +400°C When measurements are outside of the range of the thermocou­ple, the waveform saturates and clings to the top of the screen as shown at the right. The numerical measurement values are displayed as “OVER” and the cursor measurement value, calculation values and saved data are handled as full-scale output values of the 16-bit A/D converter.
See: "LR8431-20 Measurement Values" (p. A12)
About Burn Out Detection
Broken (disconnected) thermocouple wires are detected using a tiny (300 nA) sensing current. The thermal equivalent of the thermoelectromotive force (thermocouple resis­tance () times the 300 nA sensing current) is added to the measured value. However, measurement errors can occur if the thermocouple wire is long or if the wire itself has a large resistance, in which cases broken wire detection should be disabled (set to Off).
3.4 Setting Configuration 2 – Input Channel Settings
Page 61
53
1 Select a channel (P 1 to P 4) for the [Chan-
nel]
setting contents.
2 To enable measurement on the channel,
select On to display the check mark ( ).
3 Select [Count] for the [Input] (input type)
setting.
4 Select the measurement method from the
[Count Mode] settings.
Setting options:( default setting)
5 Select the basis of measurement from the
[Slope] settings.
Setting options:( default setting)
Add* Counts the number of pulses acquired
from the start of measurement.
Inst Counts the number of pulses acquired dur-
ing each recording interval. The pulse coun
t is reset at the beginning of each re-
cording interval.
Counts the number of L - H transitions. Counts the number of H - L transitions.
Totalized
[Add]
Instantaneous
[Inst]
0 V
L
H
12
: On : Off
345
3.4 Setting Configuration 2 – Input Channel Settings

Integration (Count) Measurement Settings

Configure these settings for each pulse channel receiving input from a pulse output device such as a watt-hour or flow meter. Use the scaling function to display integra ing physical units (such as Wh or VA) of the measurement parameter (p. 58).
ted pulse counts as the correspond-
3
Chapter 3 Setting before Measurement
3
• When using the scaling function, the integrated pulse count can be displayed after conversion to the measurement target’s prop erty (Wh, VA, etc.). (p. 58)
• The count will be saturated when the pulse count exceeds
3,741,823. If there is a possibility that this number may be
1,07 exceeded, it is recommended to set the integration mode to instantaneous and tabulate the results later with Excel ilar application.
®
or a sim-
-
Page 62
54
Change 1’s digit Change 10’s digit
0 V
L
H
1 Select a channel (P 1 to P 4) for the [Chan-
nel]
setting contents.
2 To enable measurement on the channel,
select On to display the check mark ( ).
3 Select [Revolve] for the [Input] (input type)
setting.
4 Set the [Pulse/Rev] setting to the number of
pulses per revolution.
Setting options: 1 to 1000 ( default setting) The range is automatically determined by this set­ting.
5 Select the basis of measurement from the
[Slope] setting.
Setting options:( default setting)Counts the number of revolutions based on
L - H transitions.
Counts the number of revolutions based on
H - L transitions.
5000 r/s (n=1)
12
: On : Off
3
45
3.4 Setting Configuration 2 – Input Channel Settings

Revolution Measurement Settings

Configure these settings for each pulse channel on which you will be counting pulses corresponding to revolutions, such as output from a rotary encoder or tachometer. The revolution measurement is obtained by counting the number of pulses input per second. The scaling function can be used to display the results of converting measured values to physical properties of the measure­ment target. (p. 58).
Page 63
3.4 Setting Configuration 2 – Input Channel Settings
0 10 ms 20 ms 1 s 2 s
Pulse Count
P200P100P2P10
Pulses per revolution
Pulse count at t
s
minus pulse count at (t -1)
s
r [rps] =
(rps: Revolutions per second)
r
t=2
=
4
2000 1000
= 250 [rps]
Pulses per revolution
Pulse count at t
s
minus pulse count at (t -0.01)
s
r [rps] = x 100
Revolution Measurement Theory
Pulse count is measured using the Memory HiLogger’s internal 10 ms sampling interval.
The revolution rate (r) per second during time ts is obtained by dividing the number of pulses from (t-1) to ber of pulses per revolution.
Example: Pulses per revolution = 4
When the number of pulses counted in one second is
P
= 1000 c,
100
and the number of pulses counted in two seconds is P = 2000 c, then the revolution rate (rps) at t=
When time ts is less than one second
The pulse count at (t –1) between (t -0.01) rate. As a result, revolution measurements when t is less tha
s
and ts serves and the pulse count for one second to obtain the revolution
s
2s (r
(whole seconds) cannot measured, so 100 times the pulse count
ts by the num-
) can be obtained.
t=2
200
n one second are scattered.
55
3
Chapter 3 Setting before Measurement
3
Page 64
56
Select from the waveform display color set­ting options.
Setting options:Off ( x ), 24-color
Zero Position: 50%
25%
0 V
100%
0%
50%
0 V
25%
Normal Display
Magnified
Display (x2)
1 Select [Position] for the [Type] setting options. 2 Select from the [Zoom] setting options.
Setting options:
x100, x50, x20, x10, x5, x2, x1, x1/2 ( default setting)
3 Set the desired zero position [0 pos] as a
percentage of screen height.
Setting options:
-50 to 150% (
with x1 magnification)
1 2
3
Change 1’s digit The expansion/com-
pression range de­pends on the selected magnification.

3.5 Display Settings

3.5 Display Settings
Configure how each channel is displayed, as needed. These settings can be changed on the Numerical/Waveform screen after measurement.

Selecting Waveform Display Color

Different display colors can be selected for each input channel wav eform.

Specifying Vertical Display Range by Magnification and Zero Position (vertical axis expansion/compression)

The vertical display range and zero position can be specified for each channel. The magnification setting determines the display or compressed vertically relative to the cen­ter of the screen. Magnification and zero-pos ignored when the display range is defined by upper and lower limits. In the 1-5V range of a voltage measure­ment and in integration measurement,
[Position] cannot be specified.
ed range. Waveforms are expanded
ition settings are
Page 65
57
50%
100%
0%
50%
100%
50%
100%
0%
0%
100%
50%
0V
Zero Position: 0%
Display range
(Zoom: x1)
Measurement range
Default Setting
Input type Zoom
Zero
position
Input type Zoom
Zero
position
Voltage
x1 50
Integration (Count)

Temperature (Tc)
x20 0
Revolution (Revolve)
x1 0
10
-10
0
5
-5
0
1 Select [Upper/Lower] for the [Type] setting
item.
2 Specify the [Upper] value for the top of the
display.
3 In the same way, specify the [Lower] value
for the bottom of the display.
The display range and scaling settings are in­terlinked.
The range is not automatically optimized when the upper and lower limits are set on the wave­form screen.
1 2
3
Default Setting
Input type Range
Voltage
-50m to 50m
Temperature (Tc)
0 to 100
Integration (Count)
0 to 5000
Revolution (Revolve)
0 to 5000
After making the settings, press [OK].
3.5 Display Settings
Specifying the Vertical Display Range by Upper and Lower Lim­its (expansion/compression)
The vertical display range can be defined by upper and lower limits. Using this method, the voltage range does
t need to be selected because the opti-
no mum range is selected automatically for the disp
lay. However, upper and lower limit set­tings are ignored when magnification and ze
ro position are enabled to define the dis­play range. As the upper and lower limits in the 1-5V range
are fixed, they cannot be changed.
3
Chapter 3 Setting before Measurement
3
Page 66
58
Normal Display
(Off)
Scaling Enabled
A
mV
1 For the [Cond](conditi on) sett i ng, select th e
desired numerical notation.
Setting options:( default setting)
2 Move to the next item to the right, and
select the scaling method.
Setting options:( default setting)
3 Enter the [Unit] name on the comment entry
screen.
See: 3.6, "Entering Comments" (p. 60)
4 Ratio: Enter the conversion ratio and offset.
2-pt: Enter the values at two points before
and after conversion.
Dec Displays converted values using fixed-point
notation
Exp Displays converted values using floating-
point notation
Off* Scaling is disabled.
Ratio* Convert by specifying the ratio of physical
units per volt of input signal (conversion ra­tio), offset, and the name of the units.
2-pt Convert by specifying the input signal volt-
age at two points, the physical unit values at those points, and the name of the units.
12 3
Conversion Ratio Setting
Conversion Ratio Setting
2-Point Setting
2-Point Setting
[Cond: Exp]
[Cond: Dec]
4
A
H
A
L
VLVH[V]
Actual measurement values (input values)
Converted unit values
Conversion Ratio Setting Method
Units (eu)
Offset
Slope (Conversion ratio: eu/v)
2-Point Setting Method
Units: [A] Param 1:V
H
value AH value
Param 2:V
L
value AL value
Example
V
H
:High value
A
H
:Converted
high value
V
L
:Low value
A
L
:Converted
low value
Combining scaling and display positioning (upper/lower limits)
To use both scaling and upper/lower display limits, set the scaling first, then upper/lower limit val­ues. With scaling enabled, the upper and lower limit setting ranges are constrained as follows:
-9.9999E+19 Setting value-1.0000E-19, Setting value = 0, +1.0000E-19 Setting value +9.9999E+19
3.5 Display Settings

Converting Units (Scaling function)

Use the scaling function to convert input voltage to the physical units of the measure­ment parameter for display, such to convert voltage input for display as electrical curren t. Converted values can be displayed in fixed­or floating-point not
Setting Analog Channel or Pulse Channel (Revolut
ation.
ion) Scaling
Page 67
59
1 For the [Cond] (condition) setting, select the
desired numerical notation.
Setting options:( default setting)
2 Enter the [Unit] name on the comment entry
screen.
See: Entering Comments (p. 60)
3 Specify the physical units per pulse or the
number of pulses per fundamental unit (for example, if 1 c = 1 pulse, select [1 pulse]).
The settings for physical units per pulse and pu
lses per fundamental unit are interlinked.
Dec Displays converted values using fixed-point
notation
Exp Displays converted values using floating-
point n
otation
Off* Scaling is disabled.
1
3
2
Using fixed-point notation
Using floating-point notation
Scaling Setting Example
For integration with a 50,000 pulse/kWh watt-hour meter: Cond: Dec Units: kWh 1kwh=50000 pulse
For integration with a 10 liter/pulse flow meter: Cond: Dec Units: L 1 pulse=10L
3.5 Display Settings
Setting Pulse Channel (Integration) Scaling
Integrated pulse values can be converted for display as physical units of the mea ­surement parameter (such as Wh or VA). Pulse output devices determine the amplitude per pulse in physica ber of pulses per fundamental unit (for example, p
er kWh, liter or m3).
l units, or the num-
3
Chapter 3 Setting before Measurement
3
Page 68
60
Select the [CH] screen.
Select the field to enter the comment.
1 2
Apply
The character entry screen appears.
3
Select the characters to enter
To enter a space, select a blank character.
Apply
Characters appear in the entry field as you select them.
Entry field
Select [OK]
Accepts your entry and returns to the [Comment] screen.
Navigate the entry field
Finished with entry

3.6 Entering Comments

3.6 Entering Comments
Comments of up to 40 characters can be entered as a title for the measure­ment data, and as a label for each input waveform. Titles and comments are displayed on the waveform screen (Comments ap
pear only when [Value+Cmnt] display is selected. Titles can be entered on the [Setting] an can be entered on the [CH] and [Comment] screens. This description enters a comment on the [CH] screen.
d [Comment] screens, and comments
The following characters are converted as follows when saved to a text file.
Entered character Converted plain
text equivalent
^2 ^3 ^u ~o ~c ~e
Page 69
Operating Panel
Only the keys displayed on the operating panel are en­abled.
List (WAVE/DATA key) Measurement related terms are registered in the list beforehand.
Hist (History, SET key) You can select previously entered terms from the pick list.
Clear (FILE key) Clears all entered characters.
Char Input (CH
key)
Switches between the pick list and character selection frames.
/ (Cursor up/down key) Selects from the pick list frame.
Select (ENTER key) Accepts the item selected in the pick list.
Cancel (ESC key) Cancels entry.
<</ >> (SCROLL/CURSOR left/right key) Move input position.
BS (SCROLL/CURSOR middle key) Deletes one character.
OK (START/STOP key) Accepts entry.
The operating panel depicts the functions of Memory HiLogger's operating keys.
3.6 Entering Comments
61
3
Chapter 3 Setting before Measurement
3
Page 70
62
[Range] screen
[Scaling] screen
Input Types
( : Voltage, : Temperature, : Integration, : Revolution, : Alarm)
[Trig & Alm] screen
Input Types
( : Voltage, : Temperature, : Integration, : Revolution, : Alarm)
[Comment] screen

3.7 Viewing All Channel Settings in a List

3.7 Viewing All Channel Settings in a List
All channel settings can be viewed and changed in the list. You can copy batches of settings between channels.
Page 71
3.7 Viewing All Channel Settings in a List
Select the screen showing the channel to copy. [Range], [Scaling], [Trig & Alm]
Select [Batch Proc].
1 2
Apply
The selected items are displayed.
3
Select the desired action.
Apply
[Copy settings of CH1 to all
other analog channels] :
Copy the settings from channel 1 to channels 2 through 10.
[Copy settings of P1 to all
other pulse channels] :
Copy the settings from channel P1 to channels P2 through P4.
Except for the following, the settings dis­played on the screen are copied.
• Waveform display color
• Input type (although this can also be copied from the [Range] screen, p. 62 )
This setting is only available on the [Range] screen.
Select [Batch Proc].
1
2
Apply
The selected items are displayed.
Select the desired action.
Apply
[Set all analog channels color off]: Sets all channel waveforms to be hidden.
[Set all analog channels color default]: Sets all channel waveform colors to their defaults.

Batch Copying Channel Settings

The CH1 and P1 channel settings can be copied to other channels.
63
3
Chapter 3 Setting before Measurement
3
Batch Setting Waveform Display/Hide and Waveform Color Set­tings for All Channels
The display of all waveforms can be hidden, and all waveform display colors
an be set to defaults.
c
Page 72
64
Select which screen you want to initialize, from the [Range], [Scaling], [Trig & Alm] screens.
Select [Batch Proc].
1 2
Apply
The selected items are displayed.
3
Select [Reset all settings in
this page].
Apply
This setting is only available on the [Range] screen.
Select [Batch Proc].
1
2
Apply
The selected items are displayed.
Select the desired action.
Apply
3.7 Viewing All Channel Settings in a List

Initializing Settings (to factory defaults)

All settings in the current window are reset to their factory default values.

Aligning Zero Positions on the Grid

Analog channel waveforms can be positioned evenly along the vertical grid, with CH1 at the to This functionality is valid only for channels with the s and for which [Display range] has been set to [Setting: Position]. CH1 to CH10 waveforms can be displayed 10% of the screen height.
p. ([Range] screen only)
ame input time as CH1
together with each allotted 5% or
Page 73
65
CH10 0 V
100%
0%
50%
CH1 0 V
This setting is only available on the [Scaling] screen.
Select [Batch Proc].
1
2
Apply
The selected items are displayed.
Select the desired action.
Apply
3.7 Viewing All Channel Settings in a List
Alignment Example (with CH1 zero-position aligned at 95% and occupying 10% of vertical screen height)
Channel Zero-position (%) CH1 95 CH2 85 CH3 75 CH4 65 CH5 55 CH6 45 CH7 35 CH8 25 CH9 15 CH10 5
Display Celsius (°C) temperature values as Fahrenheit (°F)
All measured temperature values can be converted at once from Celsius to Fahrenheit units. The conversion is simultaneously reflected appropriately on all sc To revert from Fahrenheit units to Celsius, disable (set to Of tion (p. 58).
aled values.
f) the scaling func-
3
Chapter 3 Setting before Measurement
3
Page 74
66
3.7 Viewing All Channel Settings in a List
Page 75
Specifying Criteria for
Trigger criteria met
T
T
Pre-trigger
Start (Start trigger)
Stop (Stop trigger)
Alarm
Timer
T T
Measurements (Triggering,
67
Alarm and T
You can set recording to start and stop under specific criteria (start/stop triggers), and to output alarm signals. You can also set specific times to start and stop record­ing, using the Timer function. Triggering is the process of controlling the start and stop of recording b recording is started or stopped by a specific signal, we say the trigger is "applied" or "triggering occurs". In this manual, indicates a "trigger point", as the time at which a trigger is applied.
Criteria can be specified for the fo
Trigger Type Description
Start Trigger
Stop Trigger
Pre-trigger Trigger
Sour External
Trigger Alarm Timer Records at a specified date and time.
y specific signals or conditions (criteria). When
Recording starts when the specified Start Trigger crite­ria are satisfied.
Recording stops when the specified Stop Trigger crite­ria are satisfied. When [Repeat] is enabled (set to [On])
, recording restarts automatically.
Records data acquired during a specified period prior to ea
ch trigger event.
ce
Various trigger criteria can be combined with logical AND/OR operations for each channel.
Triggering can be applied by an external signal at the EXT.TRIG input
An alarm signal can be output when any specified cri­teria are met.
imer)
llowing event trigger types.
terminal.
Chapter 4
Setting
Scr
een
[CH]/ [Trig & Alm]
[CH]/ [Trig & Alm]
[Trig & Alm]
[Trig & Alm]
[Trig & Alm] [CH]/
[Trig & Alm] [Setting]
Ref.
(p. 68)
(p. 68)
(p. 72)
(p. 71)
(p. 71)
(p. 73) (p. 76)
3
4

Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer)

Set trigger criteria for individual channels on th ger and alarm settings for all channels on the [Trig & Alm] screen (p. 75).
e Channel [CH] screen, and set trig-
Page 76
68
[ ]
[ ]
T T
Trigger Level
Input
Trigger Slope
T
T
Upper
Threshold
Lower
Threshold
T
T
Upper
Threshold
Lower
Threshold
Enable the Trigger Function
Select the trigger to use, and press the
ENTER key. Press it again when the
confirmation message appears, and you are ready to set trigger criteria (at the same time, the [Trigger] setting on the [Trig & Alm] screen turns on).
Set trigger criteria
Select the trigger criteria to be applied.
The [Timming] setting on the [Trig & Alm] screen determines the effect of triggering, as follows:
• Start recording upon trigger: [Start]
• Stop recording upon trigger: [Stop]
• Alternately start and stop upon trigger: [Start&Stop]
Recording
T
Start Stop Start/Stop
T T T

4.1 Setting Criteria to Start and Stop Recording

4.1 Setting Criteria to Start and Stop Record­ing
There are two ways to set the criteria to start and stop recording according to waveform slope. Select whether to trigger at the rising
Type of Analog Trigger Trigger Example Description
or falling edge of the input signal.
Level Trigger (p. 69)
[
Level]
Window Trigger (p. 70)
[In]
[Out]
Recording starts (or stops) when the input signal crosses the specified trigger level (voltage).
Recording starts (or stops) when the input signal enters a range defined by upper and lower thresholds.
Recording starts (or stops) when the input signal exits a range defined by upper and low­er thresholds.
Page 77
4.1 Setting Criteria to Start and Stop Recording
Select the [CH] Screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
T
T
1 Select [Level] as the [Cond] setting option
for the start or stop trigger.
2 Select the input signal rising () or falling ()
edge setting option.
Setting options: ( default setting)
3 Set the [Level] setting to the input signal level
at which to start or stop recording.
 Recording starts (or stops) when the signal cross-
es the threshold on the upslope (rising edge).
Recording starts (or stops) when the signal cross-
es the threshold on the downslope (falling edge ).
1 2
3
When scaling is enabled, the scaled value is also dis­played.
Default setting: 0

Setting the Input Signal Trigger Threshold (Level Trigger)

Enter the signal level threshold at which to start or stop recording, and whether triggering occurs on the up
slope or downslope of the input signal. Recording starts or stops when the signal crosses the specified threshold. The level value is an instanta­neous (not rms) value. Levels can be confirmed on the level monitor or on the W
aveform screen.
69
3
4
Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer)
Level Setting Ranges and Resolution
Channels Input type Range Resolution Channels Input type Range Resolution CH1 -
CH
10
Voltage
Tc
100mV 25 1V 250 10V 2.5 mV 20V 5 mV 100V 25 mV 1-5V 2.5 mV

V
V Revolve
0.5C
P1 - P4
Count
 
1 c 1 r/s
Page 78
70
T
T
T
In Out
T
1 Select [Window] as the [Cond] setting option
for the start or stop trigger.
2 Select inside [In] or outside [Out].
Setting options: ( default setting)
3 Enter the upper threshold value for the
[Upper] setting.
4 Enter the lower threshold value for the
[Lower] setting.
In Trigger when the input signal level enters the
window defined by the upper/lower thresholds.
Out Trigger when the input signal level exits the
wind
ow defined by the upper/lower thresholds.
1 2 3
4
When scaling is enabled, the scaled value is also dis­played.
4.1 Setting Criteria to Start and Stop Recording

Setting Lower and Upper Trigger Thresholds (Window Trigger)

An input signal level “window” within which recording will (or will not) occur can be defined by upper and
er threshold levels. You can select whether record-
low ing starts or stops when the input signal level enters
or exits (Out) of this window.
(In) Upper and lower thresholds can be verified in the level
r or on the Waveform screen. When scaling is
monito enabled, the scaled threshold values are displayed.
Upper/Lower Threshold Ranges and Resolution
Channels Input Range Resolution
100mV 25 V 2 mV -2 mV 1V 250
CH1 - CH10
Voltage
P1 - P4
Tc Count Revolve
10V 2.5 mV 0.2 V -0.2 V 20V 5 mV 0.4 V -0.4 V 100V 25 mV 2 V -2 V 1-5V 2.5 mV 0.2 V -0.2 V
  
V 20 mV -20 mV
0.5C 40C -40C 1 c 80 0 1 r/s 80 0
Default setting Upper Lower
Page 79
4.1 Setting Criteria to Start and Stop Recording
1 Set the [Trigger] to [On]. 2 Select the [Timming] setting option to deter-
mine how triggering will affect recording.
Setting options: ( default setting)
(Start and stop triggers can be set for each channel in­dependently.)
3 Select how trigger sources are logically com-
bined using the
[Trig Source] - [Start] and [Stop]
setting options.
Setting options:( default setting)
4 (When an external trigger is applied)
Set the
[Ext trig] - [Start] or [Stop] setting to
Start* Start recording according to Start Trigger
criteria.
Stop Stop recording according to Stop Trigger
criteria.
Start&StopStart and stop recording according to the
Start and Stop Trigger criteria, respec­tively.
OR* Triggering occurs when any one of the spec-
ified trigger source criteria is met. (includes external trigger signal)
AND Triggering occurs only when all of the speci-
fied trigger source criteria are met. (includes exte
rnal trigger signal)
1 2
3 4
[AND]
One waveform has crossed above 0V as the other crosses on the upslope
[OR]
Either waveform crosses 0V on the upslope
Start Measurement
Trigger.......Level
Level.............0 V
Slope
To apply a trigger when the upslope () of the waveform crosses zero volts:
Start Measurement
T T T T
When the trigger combining logic is set to [AND] If trigger criteria are already met when you press the
START/STOP key, no trig-
gering occurs. Triggering occurs only after all trigger sources have ceased to meet the criteria at once, and are sub­sequently met again.

Selecting Trigger Sources and Combinations

Enable the trigger function (set to On), select recording start/stop timing, and set trigger criteria. Triggering is affected only by those channels for which trig­ger criteria have been set. Make these settings on the [Trig & Alm] screen. If trigger settings have been made on the [CH] scree
n, skip steps 1 and 2.
The settings in steps 1 and 2 are interlinked with those on the [CH] screen.
71
3
4
Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer)
Page 80
72
T
For the [Pre-Trig] setting, specify the time (or number of days) to record before a trigger event.
To record waveforms after a trigger event, simply spec­ify a recording time that is longer than the pre-trigger recording time.
Difference between [Waiting for pre-trigger] and [Wait­ing for trigger]
When measurement is started, the specified pre-trigger length is recorded. This period is indicated as the [Waiting for pre-trigger]. After the specified pre-trigger length has been recorded, the period indicated as [Waiting for trigger] continues until a trigger occurs. During the [Waiting for pre-trigger] period, trigger events are not recognized even when the trigger criteria are met.
The maximum settable time span is displayed.
Change 1’s digit
Change 10’s digit
T
Pre-Tirg: 15 minutes
Recording length (10 minutes)
Recording ends at trigger event
Recording Time shorter than Pre-Trigger Time
Recording Time longer than
Pre-Trigger Time
T
Pre-Tirg: 5 minutes
Recording length (15 minutes)
Records both before
and after trigger event
Recording ends when recording length has passed
Pre-Trigger vs. Recording Time
4.1 Setting Criteria to Start and Stop Recording

Setting Criteria for Pre-Trigger Recording (Pre-Trig)

Select the [Start] or [Start&Stop] trigger timing setting to record waveforms both before and after trigger even
ts. However, pre-trigg er recording is disabled when [Stop] trigger timing Make the setting on the [Trig & Alm] screen.
is selected.
Page 81
73
Alarm Criteria
Channel alarm status (CH1 to CH10 and P1 to P4)
• Yellow*: alarm criteria are satisfied
• Green: alarm criteria are not satisfied *. When [Hold Alarm] is enabled (set to [On]), the indicator remains yellow because the criteria remain satisfied (until measurement is finished). The [Buzzer] alarm sound can be set to on or off. When [All display] is set to [On], the de­tected waveforms for CH1 to 10 and P1 to 4 set to the alarm source will also be displayed.
Alarm output status (ALM)
• Red: alarm signal output active
• Green: alarm signal output inactive
1 Select the [Channel] setting option to [ALM]. 2 Set a check mark [ ] as shown to enable
alarm output.
3 Set the [Hold Alarm] as required for your
alarm output destination device.
Setting options:( default setting)
4 Enable or disable the alarm beeper by the
[Buzzer] setting.
Setting options:( default setting)
Off
Disable the alarm output signal when alarm cr
iteria are no longer met.
On Hold the alarm output active state until mea-
surement is finished.
Off
The Memory HiLogger will not sound an au­dible beep during alarm signal output.
On The Memory HiLogger will sound an audible
beep during alarm signal output.
12
: On : Off
34
Configuring Alarm Output
Select the [CH] Screen.
To stop the alarm beeper while measuring, set the beeper to [Off] in the ALM channel settings on the Waveform screen.

4.2 Alarm Output

4.2 Alarm Output
You can enable beep tones and an alarm output signal (for external use) for each input channel by setting its alarm criteria. Y ou can also view alarm output status on screens that display numerical values ([Wave+Value], [V alue+Cmnt] and [Value]).
3
4
Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer)
External alarm output requires connection to th See Chapter 9, "External Control" (p. 129) for details. Select alarm input channels
on the [CH] screen, and specify alarm criteria on
the [Trig & Alm] screen.
e external control terminals.
Page 82
74
1 For the [Channel], select a channel for which you
want to set alarm criteria.
2 Select [Alarm on.] to enable alarm criteria settings. 3 Select the alarm criteria type from the [Cond]
setting options.
Setting options:( default setting)
Alarm criteria settings are not available unless a check mar
k [ ] is displayed next to the channel number.
4 Set the alarm criteria.
If you have selected
[Level].
Setting options:
If you have selected [Window].
Setting options:
Off* Disable alarms for this channel. Level Trigger an alarm when the input reaches a
specified le
vel.
WindowTrigger an alarm when the input transitions
through upper and lower limits (window).
Alarm output occurs when the input signal rises
above the specified level threshold.
Alarm output occurs when the input signal falls be-
low the specified level threshold.
In Alarm output occurs when the input signal enters
the r
ange defined by the thresholds.
Out Alarm output occurs when the input signal exits
the r
ange defined by the thresholds.
1
Setting Alarm Criteria for Each Channel
ALM appears when no channel output is checked ( ). In this case, select [Alarm on.] to enable alarm output settings, and then set alarm criteria.
Select how you want to combine alarm output crite­ria from the
[Alarm] setting options.
Setting options:( default setting)
OR* Alarm output occurs when the alarm criteria specified
for
any alarm-enabled channel are satisfied.
AND Alarm output occurs only when the alarm criteria for
eve
ry alarm-enabled channel are satisfied.
Selecting Combined Alarm Output Criteria
Select the [Trig & Alm] Screen.
If set to [Off], change it to [On].
4.2 Alarm Output
Page 83
75
Channel-specific set­tings (start, stop and alarm triggers) are dis­played to the right of the list for the channel selected (highlighted) by the cursor.
Select [Batch Proc] to initialize settings and to copy trigger and alarm settings from CH1 or P1 to the other channels. Copyable settings include level, window thresholds and slope.
See: "Batch Copying Channel Settings" (p. 63)
Input type
: Voltage : Temperature : Integration : Revolution : Alarm

4.3 Confirming Trigger and Alarm Criteria Settings (List)

4.3 Confirming Trigger and Alarm Criteria Set­tings (List)
You can view and change trigger and alarm criteria settings for all channels. However, detailed criteria settings such as level, threshold and slope can only b
e changed on the [CH] screen (p. 68).
3
4
Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer)
Page 84
76
Select the [Setting] Screen.
Move to a setting item.
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
Change 1’s digit
Change 10’s digit
1 Set the [Timer] setting to [On]. 2 Enable (set to [On]) the desired timer set-
ting items.
Setting options:
3 Set the year, month, day, hour, minute and
second for the start or stop time.
To enter the current time as the start or stop time, select the clock mark ( ) and press the
ENTER
key.
4 For the [Interval] settings, set the amount of
time (days, hours, minutes and seconds) between recording repetitions.
5 Press START.
[Wait Timer] will be displayed at the bottom of the
waveform screen.
Start Set the recording start time. Stop Set the recording stop time. Interval Set the recording repetition interval.
This setting is enabled on when [Repeat] is set to
[On].
Stop
measurement
Start measurement
Recording
Recording time
Recording Interval
When recording times are specified, recording operation depends on the repetition and recording time settings(p. 78).

4.4 Starting and Stopping Recording by Timer

4.4 Starting and Stopping Recording by Timer
Make these settings to record at a specific time. Recording can be set to repeat at specific intervals between the set start and stop times. Before setting, confirm that the Memory time. If not, reset the clock on the [System] screen (p. 124).
HiLogger clock is set to the correct
Page 85
77
Example: To record daily from 9:00 to 17:00 for one month beginning 1/1/2012, Repeat Recording: On, Timer: On
Note: When all channels are measured and the recording interval is slower than 200ms.
(YY-M-D) (H:M:S)
(From) Start
On 12-1-1 9: 0: 0
(Until) Stop
On 12-1-31 17: 0: 0
(Time from one record­ing to the next)
Interval
On 1 0: 0: 0
(recording length) Cont: Off
8: 0: 0
9:00 to 17:00 is 8 hours
12-1-1 9:00
12-1-1 17:00
8 hours
1 day
12-1-2 9:00
4.4 Starting and Stopping Recording by Timer
3
4
Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer)
Page 86
78
Press START/STOP
Timer start/stop time Trigger criteria match
T
Stop recordingStart recording
Recording
Recording time
Stop recordingStart recording
Recording
Recording
time
Dead time
Stop recordingStart recording
Recording
Recording time
Stop recording
Start recording
Recording
Dead time
Recording
time
Stop recordingStart recording
Recording
Recording time
Stop recordingStart recording
Recording
Dead time
Recording
time
Stop recordingStart recording
Recording
Recording time
Stop recordingStart recording
Recording
Dead time
Recording
time
Stop recordingStart recording
Recording
Recording time
Stop recordingStart recording
Recording
Dead time
Recording
time
Stop recordingStart recording
Recording
Recording time

4.5 About Recording Operation

4.5 About Recording Operation
Recording operation depends on the combination of trigger, timer and repeat recording (On/Off) settings.
Trig-
Cont.
Off Off Off
Off Off Off
Recording is stopp one recording length has elapsed
Off Off Start
Timer
ger
ed before
Repeat: Off Repeat: On
Off Off Stop
The stop time
tting is after the
se recording length has elapsed
Off Off Stop
The stop time
tting is within
se the recording length
Off Off
The stop time
tting is within
se the recording length
Start/
op
St
(Repeat Recording: same as Off)
Page 87
ig-
Press START/STOP
Timer start/stop time Trigger criteria match
T
Stop recording
Recording
T
Recording time
Start recording
Recording
time
Stop recordingStart recording
Recording
T T
Stop recordingStart recording
Recording
T
Recording time
Recording
time
Stop recordingStart
recording
Recording
Dead time
T T
Stop recordingStart recording
Recording
T T
Recording time
Stop recordingStart recording
Recording
Dead time
T T T T
Recording
time
Recording
time
Stop recordingStart recording
Recording
Stop recordingStart recording
Recording
T
Stop recordingStart recording
Recording
Dead time
T T
Tr
Cont.
Off Start Off
Off Stop Off
A stop trigger oc­curs before one recordi has elapsed
Off
A start trigger oc­curs after the tim­er start time, and a s curs before the
mer stop time
ti
Timer
ger
ng length
Start/
Start/
op
Stop
St
top trigger oc-
79
4.5 About Recording Operation
Repeat: Off Repeat: On
3
4
Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer)
On Off Off (Repeat Recording: same as Off)
On Stop Off
Page 88
80
4.5 About Recording Operation
Page 89
Waveform
Eject
Allows you to change the CF card or USB flash drive while sav­ing data in real time. (p. 48)
Analog and pulse waveforms
Displays acquired data as waveforms.
Time Values
Scroll Bar
Shows the range and position of the displayed waveform.
Viewing the Scroll bar (p. 82) Information such as measure­ment start time and trigger times is displayed below the scroll bar.
Measurement Configuration
You can change these settings (p. 41) You can switch between channel settings and event search settings.
“Event Search” (p. 88)
Jump Function (p. 83)
Trigger Mark
Indicates a trigger event
You can select from seven screen configurations (p. 18)
Gauges can be displayed for each channel (p. 84)Numerical values at waveform cursor positions can be
viewed (p. 84)
Numerical calculation results can be displayed (p. 113)Measurement data can be displayed as waveforms and
numerical values, as numerical values with comments, or as numerical values only (p. 87).
Move to a setting item
Apply
Open the setting options for the item to be set.
Select from the listed options.
1 2
Select the Waveform Screen.
81
Analysis
About the Waveform and Numerical Value Screens
Settings can be changed on the W
Chapter 5
aveform/Numerical screen.
3
5

Chapter 5 Waveform Analysis

Page 90
82
Screen Display
Earlier
Later
If internal memory becomes full when measuring with [Cont] set to [On], recording continues by overwriting the oldest acquired data. Therefore, even when scrolling earlier parts of waveforms, the overwritten data is lost.
Internal Memory (max. recording length): scrollable viewing time span
Erased Data
Screen Display Range
Whole Waveform

5.1 Viewing Waveforms

5.1 Viewing Waveforms

Scrolling Waveforms

When measuring or displaying an existing waveform, use the You can scroll to any part of a waveform while mea­suring. To return to the current waveform position, move of the screen, and press the ENTER key. To jump to the start or end of a waveform, hold the
ESC key while scrolling.
SCROLL/CURSOR keys to scroll.
the selection cursor to [Trace] at the lower right

Verifying Waveform Display Position

From the scroll bar you can verif y the re lative pos ition and size of the displayed portion of a waveform within the
overall recorded waveform. The width of the currently displayed time span indica­tor within the scroll bar depends on the time base (time per horizontal division) setting, even when the recording length is unchanged.
Page 91
5.1 Viewing Waveforms
Select from the [Time/DIV] setting options.
Setting options:
100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s, 1min, 2min, 5min, 10min, 20min, 30min, 1h, 2h, 5h, 10h, 12h, 1day
[/div] ( default setting)
Indicates the center point for horizontal magnification. When the A/B cursors are displayed, the position of cur­sor A becomes the reference point.
Select [Jump].
Apply
1 2
The location of the currently displayed portion of the over­all waveform is indicated by the mark on the scroll bar.
3
Move the mark to dis­play a different portion of the waveform.
The portion of the waveform at the mark is displayed.
Apply
Select the Waveform Screen.
4

Magnifying and Compressing Horizontally

Waveforms can be magnified (expanded or com­pressed) along the time axis b horizontal division. Detailed waveform data is usually best observed with
e time axis expanded, while longer-term trends are
th best observed with it compressed. Horizontal magnification can be changed while mea­suring. However, when measuring with auto saving
led, the 100, 200 and 500 ms settings are not
enab selectable.
y changing the time per

Viewing Any Waveform Location (Jump Function)

83
3
5
Chapter 5 Waveform Analysis
When the recording length of a waveform is long, you can specify the portion to be displayed immediately.
Page 92
84
1 2
Select the Waveform Screen.
Gauge
Select [Gauge+Wave].
Apply
A
B
A
B
B-A
B-A
AB
B-A
A
B
B-A

5.2 Selecting Display Options

5.2 Selecting Display Options

Displaying Gauges

A gauge corresponding to the measurement range of each channel can be displayed at the left side of the sc
reen, for confirming measurement values. The color of the gauge matches the waveform display color of its input channel. Press the channels. However, note that when measurement data is display On ) for measurement can be selected.
CH or key to display the gauges for other
ed, only those channels that are enabled (set to

Displaying Cursor Values

Time difference and potential difference (and when scaling is enabled, scaling values) can be read as numerical val­ues using the A/B cursors.
About Cursor Values
Cursor Type Example Cursor Value
Displays the time and measurement values at the A/B cur-
Trace Cursors
Vertical Cursors
Horizontal Cursors
sors, or the time and measurement differences between the Displays the intersections (trace points) of cursors and wave lected channels)
Displays the time and frequency values at the A/B cursors, or sors.
Displays the measurement values at the A and B cursors for cursor values. A/B cursors can be enabled for any channel.
A/B cursors.
forms. (the intersections of waveform traces of se-
the time and frequency differences between the A/B cur-
the selected channel(s), or the difference between A/B
Page 93
85
1
Displays the A/B cursors and the values at the cursor positions. To hide the cursor items again, press the
SCROLL/CURSOR key.
2
Move the cursors to display numer­ical values along the waveform.
Moving the Cursors
Select from the [Move] setting options.
A Cur B Cur AB Cur
( default setting)
Changing Cursor Type
Select from the [Type] setting options.
Trace Vert (Vertical) Horz (Horizontal)
( default setting)
Cursors
Selecting which channels have their cursor values dis­played
Select from the
[AB Crsr]
setting options.
[All Ch]* Displays all channels. The cursor selected for [Move] is displayed (select [AB Cur] to display B – A difference val­ues).
[Ch Set] Specify which channels to display with A and B cursors. The cursor values for the specified channels are dis­played.
( default setting)
Cursor Values for All Channels
Cursor Value for Specified Channels
To move cursors more quickly, hold the ESC key.
5.2 Selecting Display Options
Cursor values can be displayed for all channels, or for only desired channels.
3
5
Chapter 5 Waveform Analysis
Page 94
86
1
The A/B cursors and cursor values are displayed.
3
Move the A/B cursors to specify the time span.
Select [Move].
2
Apply
Open the setting options for the item to be set.
Select which cursor(s) to move.
[Move] A Cur B Cur AB Cur
5.2 Selecting Display Options

Specifying a Waveform Time Span

Specify a waveform time span when saving a partial waveform or applying numerical calculations (Trace cursors or Vertical cursors).
Page 95
87
Select [Wave+Value].
1 2
Apply
Select the Waveform Screen.
Select [Value+Cmnt].
1 2
Apply
Select the Waveform Screen.
Select [Value].
1 2
Apply
Select the Waveform Screen.
(Channels whose combination of mea­surement value and unit based on the scaling setting is longer than 13 charac­ters are indicated by smaller display characters.)
5.2 Selecting Display Options

Viewing Input Signals as Numerical Values

Numerical values can be displayed in three ways: numerical values only, waveforms and numerical values, and numerical values and comments. Displayed numerical values are those
To display waveforms and numerical values
To display numerical values and comments
of the current input signals.
3
5
Chapter 5 Waveform Analysis
To display numerical values only
Page 96
88
1 23
Select [Event].
1
Event mark settings are displayed.
3
Apply
Select the Waveform Screen.
Move to [Make Mark].
Press to insert an event mark. Event numbers appear on the waveform displayed when you select
[Make Mark].
2
Select [Channel].
2 3

5.3 Inserting Event Marks (Search Function)

5.3 Inserting Event Marks (Search Function)
You can insert up to 100 event marks at any point while measuring, to help find them later. See "Searching Event Marks" (p. 90) Event marks can be inserted while viewing waveforms or by
applying external signals.

Inserting Event Marks While Viewing Waveforms

Use this procedure to insert event marks while viewing data on the Waveform/ Nu
merical Value screen during measurement. Event marks are numbered in
the sequence they are inserted.
Page 97
5.3 Inserting Event Marks (Search Function)
Select the [System] screen.
Select [External Trig In].
Apply
Open the setting options for the item to be set.
Select [Event].
1 2
3
Connect the external signal input (p. 129). Event marks are inserted when the external trigger signal (EXT.TRIG input
terminal 2) transitions on the falling edge from High level (3.0 to 5.0 V) to Low level (0 to 0.8 V).

Inserting Event Marks Using External Input Signals

Event marks can be inserted by applying external input signals.
See: Input signal specifications in section 9.2, "External Trigger Input" (p. 131)
89
3
5
Chapter 5 Waveform Analysis
Page 98
90
1 2
3
Apply
Select the Waveform Screen.
For the [MoveNo.] setting, select the number of the event you want to find.
Select [Move].
4
The waveform with the specified event number is displayed.
You can search by incrementing and decrementing the event number.
[PrevEvent]: Search for the previous
event mark.
[NextEvent]: Search for the next
event mark.
When [Channel] is displayed, switch to
[Event].
The event setting items are displayed.
2
3 4
5.3 Inserting Event Marks (Search Function)

Searching Event Marks

Any event mark can be found by searching.
Page 99
Saving &
91

6.1 About Saving and Loading Data

Loading Data
Chapter 6
Measurement data acquired by the Memory HiLogger can be saved to a CF card or USB flash drive. (p.93) The CF card is treated as Drive A:, while the USB flash drive is treated as D
rive B:. There are three save methods: [Auto Save], which saves data auto-
matically at the time of meas
urement; [Select & Save], which allows data to be saved after measurement; and [Quick Save]. Refer to 6.3, "Saving Data" (p. 96)
for details.
6.1 About Saving and Loading Data
When saving data, a folder named HIOKI_LR84 stored in the folder as follows.
File Type
Settings Data Waveform
Data*1 Numerical
Calculation
esults
R Captured
Screen Image
*1.Save waveform data in binary format if you intend to view it later in the HiLogger or on a
computer with the supplied Logger Utility program. Waveform data and some measure­ment-related settings data are saved. To save a waveform partially, specify a period by using A/B cursors befo re saving (p.86).
*2. BMP Format: This is a standard Windows
by many graphics programs. *3.Appendix 2, "File Naming" (p. A7) *4.Date (YY-MM-DD) folders are automatically created inside the DATA folder. *5.When opening a CSV file in a spreadsheet program, note that the number of rows that
can be loaded at one time is limited (p. A7)
When an option other than [Separator: Comma] has been selected on the [System]
screen, the file extension will be .TXT rather than .CSV.
File Format
Binary Binary Text *5 WAVE0001.CSV
Text *5
BMP *2
Folder Name
CONFIG DATA\(Date)*4
(e.g. 12-07-30)
MEASUREMENT
PICTURE
File Name *3 (Auto-numbered fr
om 1)
CONF0001.SET × WAVE0001.MEM
MEAS0001.CSV
SCR00001.BMP ×
®
graphics format. These files can be hand led
.
31 is created, and files are
O = Available, X = Not available
Save Load
Model
Man-
Auto
O O O O O O
O O
LR8431
ual
-20
O O
×
×
O O O
PC
×
O
O
3
6

Chapter 6 Saving & Loading Data

Page 100
92
HIOKI_LR8431
CONF0001.SET
WAVE0001.MEM
(Settings data folder)
CONFIG
(Waveform data folder)
DATA
MEASUREMENT
WAVE0001.CSV
PICTURE
(Numerical calculation results folder)
(Screen image folder)
12-07-01
(the folder name is the date of file creation)
MEAS0001.CSV
SCR00001.BMP
6.1 About Saving and Loading Data
Depending on file size and CF card or USB flash drive capacity, each folder can store over 1,000 files. However, the file screen display is limited to a maximum of 1,000 files. Also, as more files are created, more time is needed to start and stop recording. We recommend setting measurement criteria so that the number of saved files stays below 1,000 whenever possible.
Directory Structure for Saved Data
Loading...