Hioki Hioki FT3424 Manual

Page 1
FT3424
LUX METER
Instruction Manual
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
September 2014 Revised edition 1 FT3424A980-01 14-09H
English
Page 2
Page 3
Contents
Introduction .........................................................................1
Verifying Package Contents ..............................................2
Options (sold separately) ..................................................3
Safety Notes ........................................................................4
Usage Notes ........................................................................7
1 Overview 11
1.1 Overview and Features .................................11
1.2 Parts Names and Functions .........................12
1.3 LCD Display ...................................................16
2 Measurement Methods 19
2.1 Measurement Workfl ow ................................19
2.2 Inserting/Replacing Batteries ......................20
2.3 Attaching the strap .......................................22
2.4 Inspection Before Use ..................................24
2.5 Making Measurements ..................................25
2.6 Selecting the Measurement Range ..............29
3 Applied Functionality 31
3.1 Retaining the Measured Value after a Set Amount of Time
(Timer Hold Function) ...................................31
Retaining the measured value (TIMER) ...................31
3.2 Disconnecting the Display Unit and
Sensor Unit ....................................................33
Mounting the sensor unit on a tripod ........................34
3.3 Limiting Battery Consumption
(Auto Power Off Function) ...........................35
i
Page 4
Contents
3.4 Saving Measured Values
(Memory Function) ........................................36
Saving the measured value (MEM) ..........................36
Reading the measured memory data (READ) ..........37
Deleting the most recently saved measured
value (CLEAR) ..........................................................38
Clearing all the saved measurement data ................38
3.5 Logging Illuminance Data
(Output Function) ..........................................39
3.6 Communicating with PC ...............................41
3.7 Disabling the buzzer .....................................42
3.8 Turning On the Backlight ..............................43
4 Specifi cations 45
4.1 Basic Specifi cations .....................................45
4.2 Measurement Specifi cations ........................46
Accuracy ...................................................................46
Characteristic ............................................................47
4.3 Output Specifi cations ...................................48
4.4 Functional Specifi cations .............................48
4.5 General Specifi cations ................................. 49
5 Maintenance and Service 51
5.1 Repair, Inspection, and Cleaning .................51
5.2 Troubleshooting ............................................52
5.3 Error Display ..................................................54
5.4 Display Messages .........................................55
ii
Page 5
Contents
Appendix Appx.1
Appx. 1 Recommended Levels of Illumination
(Reference) ........................................Appx.1
Appx. 2 Sensor Characteristics Graphs .......Appx.3
Relative Spectral Response Characteristics in
the Visible Spectrum ..........................................Appx.3
Angled Incident Light Characteristics ................Appx.4
Appx. 3 Other Characteristics .......................Appx.5
Appx. 4 Dimensional drawing........................Appx.6
iii
Page 6
Contents
iv
Page 7
Introduction
Introduction
Thank you for purchasing the HIOKI FT3424 Lux Meter. maximum performance from the product, please read this manual rst, and keep it handy for future reference.
To obtain
1
Page 8
Verifying Package Contents
Verifying Package Contents
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, operation keys of the panel, and connectors. If damage is evident, or if it fails to operate according to the specifi cations, contact your authorized Hioki distributor or reseller.
Check the package contents as follows.
FT3424 LR6 Alkaline battery × 2
Carrying Case
(soft)
Instruction manual CD* (USB driver,
dedicated computer application software, and communications specifi cations)
Strap (for instrument)
(p. 22)
Sensor cap
(with strap)
(p. 25)
USB Cable
(Length: 0.9 m)
* The latest version can be downloaded from our web site.
2
Page 9
Options (sold separately)
Options (sold separately)
The following options are available for the instrument. Contact your authorized Hioki distributor or reseller when ordering.
Connecting cable/Output cord
Use when positioning the sensor unit and display unit separately during use.
Required when using the instrument’s output functionality.
L9820 Connection Cable
(Length: 2 m)
9094 Output Cord
(Length: 1.5 m)
Carrying Case
Handy for storing the instrument with the L9820 Connection cable, 9094 Output cord, and USB cable.
C0201 Carrying Case
(semi-hard)
L9820 Connection Cable cannot be stored.
C0202 Carrying Case
(soft)
3
Page 10
Safety Notes
Safety Notes
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, using the instrument in a way not described in this manual may negate the provided safety features. Before using the instrument, be certain to carefully read the following safety notes.
CAUTION
• Mishandling during use could lead to damage to the instrument. Be certain that you understand the instructions and precautions in the manual before use.
• If persons unfamiliar with electricity measuring instrument, another person familiar with such instruments must supervise operations.
Notation
In this manual, the risk seriousness and the hazard levels are classifi ed as follows.
WARNING
CAUTION
IMPORTANT
Indicates a potentially hazardous situation that may result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance tasks with which the operators must be fully familiar.
Indicates the prohibited action.
Indicates the action which must be performed.
4
Page 11
Safety Notes
*
Additional information is presented below.
Symbols affi xed to the instrument
Indicates cautions and hazards. When the symbol is printed on the instrument, refer to a corresponding topic in the Instruction Manual.
Indicates DC (Direct Current).
Symbols for various standards
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Indicates that the instrument conforms to regulations set out by the EC Directive.
Screen display
The screen of this instrument displays characters in the following manner.
ABCDEFGH I J KLMNOPQRSTUVWXYZ
1234567890
A different display is used in the case below.
Displays when power is shutdown (p. 18)
5
Page 12
Safety Notes
Displays when display unit and sensor unit is not connected.
Accuracy
We defi ne measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values, with the following meanings:
(Maximum display value)
f.s.
Indicates the maximum displayable value. This is usually the name of the currently selected range.
(Reading value)
rdg.
The value currently being measured and displayed on the measuring instrument.
(Resolution)
dgt.
The minimum display unit, indicating a minimum digit of 1.
6
Page 13
Usage Notes
Usage Notes
Follow these precautions to ensure safe operation and to obtain the full benefi ts of the various functions.
Before Use
Verify that the instrument operates normally to ensure that no damage occurred during storage or shipping. If you fi nd any damage, contact your authorized Hioki distributor or reseller.
Installation
For details on the operating temperature and humidity, see the specifi cations.(p. 49)
WARNING
Installing the instrument in inappropriate locations may cause a malfunction of instrument or may give rise to an accident. Avoid the following locations.
• Exposed to high temperature
• Exposed to corrosive or combustible gases
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
• Susceptible to vibration
When the instrument is not in use, store the instrument in a cool, dark place because optical components are vulnerable to heat.
7
Page 14
Usage Notes
Handling the cables and cords
CAUTION
• Before use, verify that the insulation on cables and cords is not damaged and that no metal is exposed. If you fi nd any damage, replace the cable or cord with those specifi ed by our company, as the instrument will not be able to make accurate measurements or send/ receive data otherwise.
• Avoid stepping on or pinching the cables and cords, which could damage the cable insulation.
• To avoid breaking the base of connectors and jacks, do not bend or pull them.
• Cables and cords become stiff and rigid at temperatures of 0°C and lower. Exercise caution in such environments as bending or pulling on cables and cords may damage their insulation or cause them to break.
Handling the instrument
CAUTION
• The instrument consists of a sensor unit and a display unit that can be positioned apart from one another during operation. To avoid damage, be sure to turn off the instrument before disconnecting or connecting the sensor and display units.
• To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
• To avoid damage to the instrument, do not short-circuit the D/A OUTPUT terminal and do not input voltage to the D/A OUTPUT terminal.
8
Page 15
Usage Notes
IMPORTANT
• Use only the specifi ed connection cables. Using a non- specifi ed cable may result in incorrect measurements due to poor connection or other reasons.
• When measuring illuminance underneath a standard lighting xture, the display may not stabilize. In most cases, a failure to stabilize is due to fl uctuations in the lighting fi xture’s supply voltage or to the surrounding environment (for example, a person’s shadow). Exercise care concerning these factors when performing measurement.
• The LCD display includes a backlight for use when making measurements in dim locations. The backlight will activate automatically when the measured value is retained or when the measured value data stored in the internal memory is in read mode, both in low-light environments (approx. 750 To avoid affecting measurement results, the backlight cannot be turned on during measurement.
• The instrument’s measurement reference level (REF.LEVEL) is the colored part in the drawing below
• Do not attempt to disassemble the instrument or subject it to mechanical shock.
.
measurement reference level (REF.LEVEL)
or less).
lx
9
Page 16
Usage Notes
CD disc precautions
• Exercise care to keep the recorded side of discs free of dirt and scratches. When writing text on a disc’s label, use a pen or marker with a soft tip.
• Keep discs inside a protective case and do not expose to direct sunlight, high temperature, or high humidity.
• Hioki is not liable for any issues your computer system experiences in the course of using this disc.
Precautions during shipment
Observe the following during shipment.Hioki cannot be responsible for damage that occurs during shipment.
CAUTION
• Handle the instrument carefully so that it is not damaged due to a vibration or shock.
• To avoid damage to the instrument, remove the accessories and optional equipment from the instrument before shipment.
If the instrument is not to be used for an extended period of time
IMPORTANT
To avoid corrosion and/or damage to the instrument due to battery leakage, remove the battery and store the instrument in a cool, dark place if it will not be used for an extended period of time.
10
Page 17
Overview
1
1.1 Overview and Features
The FT3424 is a multifunctional, high-precision lux meter which ensures durability.
Engineered for use in a wide range of fi elds and settings, including with lighting equipment, in lighting work, and in equipment management.
Wide-range illuminance measurement (0.00 lx to 200000 lx)
Use the display unit and sensor unit separately
L9820 Connection cable (option)
Large, easy-to-read LCD display
Backlight automatically turns on when you retain the measured value in a
low-illuminance
environment.
Retaining the measurement (HOLD)
Features the timer hold function.
Sensor unit
Features a luminosity fi lter designed to approximate the response of the human eye with the instrument’s sensitivity to wavelength.
Ability to send data to a computer and to be controlled by a
computer
Use the included dedicated computer application (after connecting the instrument with a USB cable) to download data and control the instrument.
Memorizes measured value
Internal memory can store up to 99 measured values, which can be sent together to a computer.
11
Page 18
Parts Names and Functions
3
4
6
1.2 Parts Names and Functions
Front/Right Side
Sensor unit
Illuminance
sensor
Sensor
cap
(p. 25)
Reference
level
indication
Display unit
LCD display
Operation
Lux meter could be seperated into sensor unit and display unit. (p. 33) (Use the option L9820 Connection cable)
(p. 16)
keys
(p. 12)
Strap hole (p. 22)
USB terminal
(mini B type)
D/A OUTPUT
terminal
(p. 39)
Operation keys
1 2
3
4
5
6
12
Page 19
Parts Names and Functions
1 2
3
4
Normal
for at least 1
second
Press and hold
1 2
3
4
Power ON Power OFF -
Retains the measured value or cancels retention of the measured value.
• Switches the range. (p. 29)
• Cancels read mode, which allows you to view the measured values stored in the internal memory. (p. 37)
Performs zero­adjustment.
Start timer hold function (p. 31) Automatically retain after 5 to 60 seconds (designating the time is possible)
Switches to auto range.
*1
• Allows you to delete the last saved measured value. (p. 38)
• Cancels zero adjustment when
is displayed.
[CAP]
Turn on the
instrument
while pressing
and holding
Cancels the auto power off function (APS). (p. 35)
Displays the software version of the instrument.
Places the instrument in the
*1
state that enables all measured values stored in the internal memory to be deleted.
(p. 38)
13
Page 20
Parts Names and Functions
5
6
Normal
for at least 1
second
Press and hold
5
6
*1: Read mode, which allows you to view the measured values stored in
the internal memory.
*2: The
measured values stored in the internal memory.
*3: During operation of timer hold function.
• Save measured value in internal memory. (p. 36)
• Increases the memory No.
• Increases the time remaining on the timer.
• Decreases the memory No.
• Decreases the time remaining on the timer.
and can only be operated when there are multiple
Continuously increases the memory
*1
(p. 37)
No.
*1, *2
*3
• Loads measured
*1, *2
values stored in the internal memory for viewing. (p. 37)
*3
• Continuously decreases the memory No. (p. 37)
*1
Turn on the
instrument
while pressing
and holding
All LCDs light up.
Sets the buzzer sound non­activated. (p. 42)
14
Page 21
Parts Names and Functions
Tripod mounting thread
Use when mounting the sensor unit on a tripod. The hole is located exactly behind the center point of the illuminance sensor on the front of the sensor unit. (p. 34)
Collation No.
Verify that the collation numbers on the sensor unit and display unit match before use.
Serial No.
It is necessary for production control such as product warranty. Do
Do not peel off the label.
Measurement range
The measurement range of the instrument is noted.
Battery Cover
When replacing the batteries remove the cover. Explains which battery to use.
See p. 20
(p. 20),
15
Page 22
LCD Display
1
5
8
1 2
3
4
5
1.3 LCD Display
For message displays and error displays, see “5.3 Error Display” (p. 54), and “5.4 Display Messages” (p. 55).
3
2
1
4
5
16
6
8
7
1 2
3
4
5
Holds measured value. (p. 26, p. 31) Communicating with the USB. (p. 41)
The memory function is activated. (p. 36) Instrument is in read mode, allows viewing
measured values stored in its internal memory. (p. 37), Memory No.
The timer hold function is activated. (p. 31) The time shown is the time remaining (in seconds) until the measured value is retained.
/ can be operated. (These keys are used to load measured values and set the timer remaining time.)
Battery indicator (p. 17) The auto power off function is activated. (p. 35) The buzzer sound is activated. (p. 42)
The measured value exceeded the set range’s maximum illuminance range.
Page 23
LCD Display
6 7
8
6 7
8
Auto range is activated. (p. 29) The output function has been activated. (p. 39)
Represents the unit used to measure illuminance (lux).
When the measured value exceeds the maximum value in each range
The maximum displayable value blinks, and
lights up on the LCD display.
Battery indicator
Fully charged. As the battery charge diminishes, black charge bars disappear,
one by one, from the left of the battery indicator. The battery is almost out of charge. Have a new battery handy.
(Lights up)
The battery is exhausted. Replace with new batteries immediately. (p. 20)
(Blinks)
The battery is exhausted. Replace with new batteries immediately. If you keep using the instrument, the power may shutdown. (p. 20)
During USB communications and while the instrument is connected
OFF
to USB bus power, the battery indicator turns off.
The battery charge indicator is only a reference for the continuous operation time.
17
Page 24
LCD Display
When using a manganese battery or nickel-hydride battery, the battery indicator may not operate properly.
Power shutdown
When the charge is gone, the display for 3 seconds and the power is shut down automatically.
[b. Lo]
blinks in
18
Page 25
Measurement Methods
2
2.1 Measurement Workfl ow
Before using the instrument, be sure to read “Usage Notes” (p. 7).
Installation and connection
Insert the batteries with the sensor cap on. (p. 20)
Perform the startup check. (p. 24)
As necessary, have other optional items available and ready.
Measurement
Turn on the power and perform zero-adjustment.
Remove the sensor cap and start the measurement.
(As necessary) Hold the (display of the) measured value. Save the measurement data in the internal
memory. (p. 36)
End of the measurement
Turn the power off and put the sensor cap on.
19
Page 26
Inserting/Replacing Batteries
2.2 Inserting/Replacing Batteries
Before using the instrument fi rst time, insert two LR6 alkaline batteries. Before measurements, check that the battery level is suffi cient. When the battery charge is low, replace the batteries.
Nickel-metal hydride batteries
Nickel-metal hydride batteries can be used. However, the discharge characteristic of these batteries is different from that of alkaline batteries. Be aware that the remaining battery power display does not function properly.
WARNING
• To avoid the possibility of explosion, do not short
circuit, charge, disassemble, or incinerate batteries.
• To avoid electric shock, disconnect the output cord
and USB cable from the object to be measured before replacing the batteries.
After battery replacement but before using the instrument, reattach and screw down the battery cover.
CAUTION
Poor performance or damage from battery leakage could result. Observe the cautions listed below.
• Do no mix new and old batteries, or different types of batteries.
• Be careful to observe the battery polarity during installation.
• Do not use batteries after their recommended expiry date.
• Do not allow used batteries to remain in the instrument.
To avoid corrosion from battery leakage and/or damage to the instrument, remove the batteries from the instrument if it is to be kept in storage for an extended period.
20
Page 27
Inserting/Replacing Batteries
3
2
6
4, 5
• The indicator appears when the battery is almost out of Have a new battery handy.
charge.
When the indicator lights up continuously or fl ashes, there is no battery life remaining. Replace the battery immediately.
• During USB communications and while the instrument is
connected to USB bus power, the battery indicator turns off.
• Turn off the power before replacing the batteries.
• After use, be sure to turn off the instrument.
• Handle and dispose of batteries in accordance with local regulations.
Have the following items
1
available and ready.
2
• LR6 Alkaline battery × 2
Press for at least 1 second to turn off the instrument.
Rear
4, 5
3
6
Turn off the instrument.
2
Remove the battery cover.
3
Remove all of the old batteries.
4
Insert two new batteries
5
(LR6), being careful to the battery polarity.
Reattach the battery cover.
6
21
Page 28
Attaching the strap
Although you can use an R6 Manganese battery, doing so will give the instrument a shorter continuous operating time than an alkaline battery.
2.3 Attaching the strap
You can attach the included strap (for instrument) and the strap for the sensor cap to the strap hole on the bottom of the display unit.
CAUTION
Attach the strap securely to the instrument. If insecurely attached, the instrument may fall and be damaged when carrying.
When attaching one strap
22
Page 29
When attaching both straps
3
2
1
1
Strap for the sensor cap
2
3
Attaching the strap
Strap (for instrument)
23
Page 30
Inspection Before Use
2.4 Inspection Before Use
Verify that the instrument operates normally to ensure that no damage occurred during storage or shipping. If you fi nd any damage, contact your authorized Hioki distributor or reseller.
Appearance check of the instrument
Check item Action
• The instrument is neither damaged nor cracked.
• The internal circuits are not exposed.
Check when turning on the power
Check item Action
The battery voltage is suffi cient.
No indicators are missing. Display all indicators and ensure that no
Visually check the instrument. If it is damaged, it could not be measured accurately. Do not use the instrument but send it for repair.
When the right corner of the LCD display, replace with new batteries immediately. If you keep using the instrument, the power may shutdown. (p. 20)
indicators are missing. (p. 14, p. 16) If any of the indicators are missing, send the instrument for repair.
indicator appears in the top
24
Page 31
2.5 Making Measurements
Put the sensor cap on.
Turn on the instrument with the
1
included sensor cap attached to the illuminance sensor.
A value will be displayed on the LCD display.
Making Measurements
Press to turn the instrument.
on
Press .
2
is displayed, and zero-adjustment
[ADJ]
of all ranges will be performed. When zero-adjustment is completed,
goes off.
[ADJ]
Remove the sensor cap, and
3
bring the sensor unit near the measuring location.
25
Page 32
Making Measurements
(To use a particular range to make a measurement)
Press to select the range.
4
See: “2.6 Selecting the Measurement Range” (p. 29)
Read the measured value when it
5
stabilizes.
(When retaining the measured value)
Press and read the measured value.
Pressing again will cancel retention of the measured value.
You can also retain the measured value after a set amount of time elapses. See: “3.1 Retaining the Measured Value after a Set Amount of Time (Timer Hold Function)” (p. 31)
26
Page 33
Making Measurements
Put the sensor cap on.
Press and hold for at least 1 second to turn off the instrument.
• If zero-adjustment is performed immediately after the
• While the measured value is retained, zero-adjustment can not
is displayed when the measuring range is exceeded.
instrument is turned on, several count digits may remain. In that case, perform zero-adjustment again.
be performed.
When the measurement is
6
nished, put the sensor cap on and turn off the instrument.
27
Page 34
Making Measurements
If is pressed without the sensor cap attached
If you press without the included sensor cap attached to the illuminance
sensor (when the count is equivalent to 1 or greater), LCD display.
Press sensor cap.
will be displayed on the
[CAP]
again after attaching the
Zero-adjustment will be canceled when pressing for at least 1 second while
[CAP]
is displayed.
lx
28
Page 35
Selecting the Measurement Range
2.6
Selecting the Measurement Range
Auto or Manual range can be selected.
• Auto range Sets the optimum range automatically in accordance with
• Manual range Sets the range and fi xes it specifi cally.
the actual measurement. (Disabled when the output function (OUTPUT) is in use.)
Measuring with the auto range
The auto range measurement starts when the instrument power is turned on.
lights up. (default setting)
Measuring with the manual range
Press .
The instrument will switch from the auto range to manual range, which will be fi xed to the range that was selected during auto-range
operation. (
Each time is pressed, the range is specifi ed. 200000 lx 20.00 lx 200.0
20000
lx
goes off.)
2000
lx
lx
To switch to auto range
Press and hold for at least 1 second. ( lights up.)
29
Page 36
Selecting the Measurement Range
While the measured value is retained, switching of the range can not be performed.
30
Page 37
Applied Functionality
3
3.1 Retaining the Measured Value after a Set Amount of Time (Timer Hold Function)
This section describes how to retain the measured value after a set amount of time has elapsed. The timer hold function is convenient when measuring low illuminance values, for example from emergency lighting or along an evacuation route.
Retaining the measured value (TIMER)
Press and hold for at least 1 second.
The timer hold function will be activated, and the time remaining until the measured value
is held will be displayed (counted down) at the top right of the LCD display. (
lights up.)
When the remaining time is 10 seconds or more, the instrument will beep every 5 seconds. When the remaining time is 10 seconds or less, the instrument will beep every second.
You can change the time remaining on the timer by pressing
while the timer hold function is active.
(Select from 5, 10, 15, 20, 30, 45, 60 seconds) Default setting: 5 seconds
or
31
Page 38
Retaining the Measured Value after a Set Amount of Time (Timer Hold Function)
The measured value will be retained once the set amount of time has elapsed.
and lights up, and a
( continuous beep sounds for 3 sec.)
When state is canceled, and the timer hold function
is not activated. ( and goes off.)
is pressed again, the hold
• Pressing while the timer hold function is active (while the timer is counting down) will cause the measured value
to be retained.
At this moment, the timer hold function is not
activated. ( goes off.)
• While the measured value is retained, switching of the range can not be performed.
32
Page 39
Disconnecting the Display Unit and Sensor Unit
3.2 Disconnecting the Display Unit and Sensor Unit
Lux meter could be measured by disconnecting the display unit and sensor unit.
Turn off the instrument.
1
Hold the display unit and sensor unit, and pull them apart.
2
Connect the display unit and sensor unit with L9820
3
Connection Cable (option).
Do not seperate and connect the display unit and the sensor unit while the instrument power is on.
33
Page 40
Disconnecting the Display Unit and Sensor Unit
Mounting the sensor unit on a tripod
When you need to place (fi x) the sensor unit in a location while making measurements, you can mount it to a commercially available tripod. Use the mounting thread* on the back of the sensor unit.
* Thread size: 1/4” (ISO 1222)
• Mount the sensor unit on the tripod after disconnecting it from the display unit.
• When mounting the sensor unit on the tripod, turn the thread (not the sensor unit).
• Do not lift the tripod by the sensor unit after it has been mounted on the tripod.
34
Page 41
Limiting Battery Consumption (Auto Power Off Function)
3.3 Limiting Battery Consumption (Auto Power Off Function)
This function limits the battery consumption. If the instrument has not been operated for approx. 10 minutes, the power turns off automatically. In the (default setting), the auto power off function is set to enabled. ( lights up.) When the auto power off function is enabled, the on the LCD display will fl ash 30 seconds before, along with the beeping sound 15 seconds before the instrument automatically turns off. To continuously use the instrument without turning off the power, press any key on the front panel. When the instrument has not been operated for approx. 10 minutes again since the key was pressed, the power turns off automatically.
• If the instrument will be used continuously for an extended period of time, disable the auto power off function.
• After use, be sure to turn off the instrument.
• When using the output function (OUTPUT), during USB communications, and when connected to USB bus power, the auto power off function will be disabled.
Disabling the auto power off function
If the instrument is on, turn it off.
Press while holding down
to turn on the instrument.
Auto power off function is disabled.
The auto power off function will be disabled until the instrument is turned off. Exercise care concerning battery consumption.
Check that of the LCD display is not lit up.
35
Page 42
Saving Measured Values (Memory Function)
y
yyy
y
y
y
3.4 Saving Measured Values (Memory Function)
The measurement result can be saved and read using the memory function. Up to 99 measured data can be saved. You can also delete saved measured values. (p. 38) Measured values saved in the internal memory can be downloaded to a computer using the instrument’s USB communications capability. (p. 41)
Memory function is disabled when the output function (OUTPUT) is in use.
Saving the measured value (MEM)
Memory No.
When the internal memory is full
Press (MEMORY) while measuring.
The measured value when (MEMORY) is pressed will be saved in the internal
memory, starting with the lowest memory No. At this moment, the buzzer sounds, and the memory No. and second.
If you try to save a measured value when the internal memory is full (when 99 measured values have been saved), the LCD display will show To save a new measured value to the internal memory, you must fi rst delete one or more previously saved measured values. (p. 38)
lights up for 1
.
[FULL]
36
Page 43
Saving Measured Values (Memory Function)
Reading the measured memory data (READ)
Memory No.
Press (READ) and hold for
1
at least 1 second.
( and lights up.)
Read mode of the measurement data saved in the internal memory starts.
Select the desired memory No.
2
using
or . (upper right
side of the LCD display)
Continuously increase or decrease the memory No. by holding the
for at least 1 second. The and can only be operated
when there are multiple measured values stored in the internal memory.
Backlight automatically turns on when you read the measured data of the internal memory in low-light environments (approx. 750
lx
or less).
or
Canceling the read mode
Press (ESC). ( goes off.)
When measurement data is not saved
When no measurement data is saved in the internal memory, press
appears in the LCD display for
[ - - - ]
approx. 1 second, and then the measurement
display reappears.
(READ) and
37
Page 44
Saving Measured Values (Memory Function)
y
y
y
y
Deleting the most recently saved measured value (CLEAR)
Memory No.
You cannot delete the measured value for a specifi c memory No. other than the last one.
Press (READ) and hold for
1
at least 1 second.
( and lights up.)
Read mode of the measurement data saved in the internal memory starts.
Press (CLEAR) and hold
2
for at least 1 second.
The most recently saved measured value (with the last memory No.) will be deleted.
Clearing all the saved measurement data
If the instrument is on, turn it off.
Press while holding down
1
to turn on the instrument.
Press and hold for at least
2
1 second while
All the saved measurement datas are cleared. After measurement display appears.
[CLr] is displayed.
blinks, the
[CLr]
38
Page 45
Logging Illuminance Data (Output Function)
3.5 Logging Illuminance Data (Output Function)
You can connect the instrument to a logger or other recording instrument and have it generate voltage output based on measured values. This functionality outputs a voltage of 1 mV DC for each effective count digit in the measured value. The voltage is updated at the same rate as the instrument’s LCD display.
CAUTION
To avoid damaging the output cord, unplug it by grasping the connector, not the cord.
• When using the output function (OUTPUT), the following functions are disabled.
Auto power off function Memory function Auto range
• When using the output function for an extended period of time, connect a USB cable to the instrument so that it operates on USB bus power.
• +2.5 V signal will be output temporarily when turning the power on, while the output cord is plugged in; however this is not a malfunction.
39
Page 46
Logging Illuminance Data (Output Function)
1, 4
2
3
Press and hold for at least 1 second to turn off the
1
instrument. Connect the mini jack of 9094 Output Cord (option) to the
2
D/A OUTPUT terminal of the right side of the instrument. (Set the recording instrument in advance.)
3
Connect the banana terminal of the 9094 Output Cord to the logger or other recording instrument.
Press to turn on the instrument.
4
Connection example
Logger or other
3
2
1, 4
Perform zero-adjustment as neccesary, and select the
5
output rate by pressing
Range Output rate
lx
20
200 lx1 mV DC / 0.1
2000 lx1 mV DC / 1
20000 lx1 mV DC / 10
200000 lx1 mV DC / 100
9094 Output Cord
The voltage is output from the D/A OUTPUT terminal, depending on the measurement value.
1 mV DC / 0.01
lx
( lights up.)
lx
lx
lx
lx
recording instrument
. (See table below)
When each range's full scale exceeds, the output is 2.5 V DC.
lights up on
( the LCD display.)
40
Page 47
Communicating with PC
3.6 Communicating with PC
Using the included USB cable, it is possible to transmit data to the PC or to control the instrument. For details, see the communications specifi cations which accompanies with the CD.
Install the special PC application software on the PC.
Connect to the PC.
The virtual COM ports of the PC can be used as the USB interface.
USB cable
Transmits data.
Controls the
instrument.
• When connecting a USB cable to the instrument, exercise care to orient the connector properly.
• During USB communication, lights up on the LCD display.
• During USB communication, do not disconnect the USB cable. Disconnecting the cable stops the communication. In that case, a warning is displayed by the special PC application software. Connect the USB cable again.
41
Page 48
Disabling the buzzer
3.7 Disabling the buzzer
The buzzer sound is enabled when factory default settings. Turn off the power of the instrument when changing the settings.
while holding down
Press
to turn on the instrument.
[bP oFF]
activated.
The buzzer sound is disabled until the power is turned off.
When you release , the screen will return to the measured value display. (
is displayed, and the buzzer is not
42
goes off.)
Page 49
Turning On the Backlight
3.8 Turning On the Backlight
The LCD display includes a backlight for use when making measurements in dim locations. The backlight will activate automatically when the measured value is retained or when the measured value data stored in the internal memory is in read mode,
lx
both in low-light environments (approx. 750 To avoid affecting measurement results, the backlight cannot be turned on during measurement.
Forcing the backlight to turn on
To forcibly turn on the backlight in an environment with illuminance that is greater than or equal to approximately 750 lx, attach the sensor cap to the illuminance sensor while retaining the measured value.
Activation and deactivation of the backlight is unrelated to the measured value that is being retained. The illuminance sensor continuously monitors illuminance, and the instrument determines whether to turn the backlight on or off based on a monitored illuminance level of approximately 750 lx.
or less).
43
Page 50
Disabling the buzzer
44
Page 51
Specifi cations
4
4.1 Basic Specifi cations
Classifi cations Grade JIS C 1609-1: 2006 General Class AA Display • Display LCD display 4 digits
Measurement ranges
• Effective display digits 2000 counts
• Display unit
• Display update rate 500 ms ± 20 ms
Range
20
2000
20000
200000
Measurement
lx
0.00 lx to 20.00
lx
lx
lx
lx
000 lx to 200000
0.0 lx to 200.0
00 lx to 20000
lx
range
0 lx to 2000
(lux)
lx
lx
1 count step200
lx
10 count step
lx
100 count step
lx
Display steps
Measurement range selection
Auto / Manual
45
Page 52
Measurement Specifi cations
4.2 Measurement Specifi cations
Accuracy
Linearity ±2% rdg.
Accuracy guarantee conditions
Accuracy guarantee for temperature and humidity
Accuracy warranty period
• f.s. (maximum display value): The maximum displayable value. This is usually the name of the currently selected range.
• rdg. (reading value): The value currently being measured and displayed on the measuring instrument.
• dgt. (resolution): The minimum display unit, indicating a minimum digit of 1.
(Multiply by 1.5 for display values in excess of
lx.
3,000
) (Add ±1dgt. for display values that are less than 1/3 of the range.)
Display unit and sensor unit must bear the same collation No.
21°C to 27°C (69.8°F to 80.6°F), 75% RH or less (no condensation)
2 years
46
Page 53
Characteristic
Measurement Specifi cations
Angled incident light
characteristics
Response time Auto range: 5 seconds or less
Temperature characteristic
Humidity characteristic
Relative spectral response characteristics in the visible spectrum
Response characteristics in ultraviolet and infrared spectrums
Fatigue characteristics
Systematic deviation f Deviation from cosine characteristics:
Angle
30° ±2% 60° ±7% 80° ±25%
Manual range: 2 seconds or less Deviation from value measured at 23°C (73.4°F)
between -10 to 40°C (14°F to 104°F): ±3% rdg. Deviation from value measured in an environment with
a temperature and humidity of 23°C (73.4°F), and 45% to 70% RH when the unit is left in an 85% to 95% RH environment for 3 hours and then is returned to the original environment: ±3% rdg.
Deviation from spectral luminous effi ciency (relative luminous effi ciency) f 6% or less
Response to ultraviolet and infrared radiation : 1% or less
The change in value of 1 minute and 10 minutes after light strikes sensor : ±1% rdg.
: 3% or less
2
Deviation from cosine
characteristics
' :
1
47
Page 54
Output Specifi cations
Characteristics regarding intermittent light
Deviation in value when subjected to intermittent light for 1/2 cycle at a frequency of 100 Hz or 120 Hz : ±2% rdg.
4.3 Output Specifi cations
Output method D/A output Output level 2 V/ range f.s.
Resolution 1 mV
Output update rate 500 ms ± 20 ms Output accuracy ±1% rdg.±5 mV (at display count) Output resistance 1.1 k
2.5 V is output when the range f.s. is exceeded.
Range Output rate
lx
20
1 mV DC / 0.01
200 lx1 mV DC / 0.1
2000 lx1 mV DC / 1
20000 lx1 mV DC / 10
200000 lx1 mV DC / 100
or less
Ω
lx
lx
lx
lx
lx
4.4 Functional Specifi cations
Hold function Retains the measured value. Timer hold
function
Memory function Up to 99 measured data can be saved.
Retains the measured value after the set timer time has elapsed after executing. Select and set the timer time from 5, 10, 15, 20, 30, 45, 60 seconds.
48
Page 55
General Specifi cations
Auto power off function
Power shutdown When the charge is gone,
Buzzer sound Emits a buzzer sound when keys are operated,
Backlight Activates when the measured value is retained or
Zero adjustment Performs zero-adjustment.
Turns off the instrument approx.10 minutes after the last key operation (can be canceled).
blinks in the
display for 3 seconds and the power is shut down.
when the timer hold function activates, and when it is auto power off function. (The buzzer sound can be canceled).
when the measured value data stored in the internal memory is in read mode, both while the ambient illuminance is less than 750
Zero-adjustment execution time: 3 seconds or less
[b. Lo]
lx
.
4.5 General Specifi cations
Product warranty period
Light receiving element
Interface USB 2.0 Operating
temperature and humidity
Storage temperature and humidity
Operating environment
3 years
Silicon photo diode
-10°C to 40°C (14°F to 104°F), 80% RH or less (no condensation)
-20°C to 50°C (-4°F to 122°F), 80% RH or less (no condensation)
Indoors, pollution degree 2, altitude up to 2,000 m (6,562 ft.)
49
Page 56
General Specifi cations
Power supply
Continuous operating time
Maximum rated power
Dimensions Approx. 78W × 170H × 39D mm (3.07 W × 6.69 H ×
Mass Approx. 310 g (10.9 oz.) (including the batteries) Standard
compliance Applicable
standards Dustproof and
waterproof
Accessories • Instruction Manual
Options See: “Options (sold separately)” (p. 3)
LR6 Alkaline battery × 2 R6 Manganese battery × 2
Rated power voltage 1.5 V DC × 2 (Maximum allowable voltage 3.6 V DC)
USB bus power 5 V DC Approx. 300 hours (when LR6 alkaline batteries are used)
500 mVa
1.54 D)
• JIS C 1609-1: 2006 General Class AA
• DIN 5032-7: 1985 Class B
• Safety: EN61010
• EMC: EN61326
IP40 (EN60529)
To avoid any failure, do not allow the instrument to get wet. If the instrument gets wet, have your authorized Hioki distributor or reseller inspect or repair it, if necessary.
• LR6 Alkaline battery × 2
• Sensor cap (with strap)
• Carrying Case (soft)
• Strap (for instrument)
• USB cable (0.9 m)
• CD (USB driver, dedicated computer application software, and communications specifi cations)
50
Page 57
Maintenance and Service
5
5.1 Repair, Inspection, and Cleaning
Calibrations
IMPORTANT
Periodic calibration is necessary in order to ensure that the instrument provides correct measurement results of the specifi ed accuracy.
The calibration interval for this instrument is 2 years. It is recommended to calibrate it every 2 years for accurate measurement.
Cleaning
• To clean the instrument, wipe it gently with a soft cloth moistened with water or mild detergent.
• Wipe the illuminance sensor and the LCD display gently with a soft, dry cloth.
IMPORTANT
Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
Disposal
Handle and dispose of the instrument in accordance with local regulations.
51
Page 58
Troubleshooting
5.2 Troubleshooting
• When a malfunction of the instrument is suspected, check the information in “Before sending the instrument for repair” and then, if necessary, contact your authorized Hioki distributor or reseller.
• When sending the instrument for repair, remove the batteries and pack it carefully to prevent damage during transportation. Include cushioning material so the instrument cannot move within the package. Be sure to include details of the problem. Hioki cannot be responsible for damage that occurs during transportation.
Before sending the instrument for repair
Symptom Check and/or remedy
Nothing appears in the display. Or the display disappears after a short time.
The display does not stabilize and the value fl uctuates; it is diffi cult to read the value.
The range can not be changed.
Check that the batteries are not exhausted. (p. 17) Replace with new batteries. (p. 20) When using a manganese battery or nickel-hydride battery, the battery indicator may not operate properly.
When the auto power off function is enabled and the instrument has not been operated for approx. 10 minutes, the power turns off automatically. Check the setting of the auto power off function. (p. 35)
When measuring illuminance underneath a standard lighting fi xture, the display may not stabilize. In most cases, a failure to stabilize is due to fl uctuations in the lighting fi xture’s supply voltage or to the surrounding environment (for example, a person’s shadow). Exercise care concerning these factors when performing measurement.
While the measured value is retained, the range can not be changed. Please cancel the hold function.
52
Page 59
Symptom Check and/or remedy
Turning on the power brings up the error display.
When nothing is connected, the error display appears.
Send the instrument for repair. Refer to “5.3 Error Display” (p. 54)
Frequently asked questions (FAQ)
Question Solution
Would like to perform zero-adjustment.
Can rechargeable batteries be used?
Would like to control multiple instruments with 1 PC.
The instrument cannot communicate with the PC.
Would like to know commands.
Would like to perform communication using own software.
Perform zero-adjustment. Refer to “2.5 Making Measurements” (p. 25)
Nickel-metal hydride batteries can be used. However, the discharge characteristic of these batteries is different from that of alkaline batteries. Be aware that the remaining battery power display does not function properly.
It is possible to control multiple instruments by connecting the USB cable.
• Check the communication setting between the instrument and the PC. For details, see the communications specifi cations which accompanies with the CD.
• Check that the USB cable is connected correctly. (p. 41)
• Verify that the illuminance sensor is not dirty.
To communicate with the instrument and PC, installation of the USB driver and special PC application software is necessary. For details on communication commands, see the communications specifi cations which accompanies with the CD.
Troubleshooting
53
Page 60
Error Display
5.3 Error Display
Error display Description Solution
Err 01
Err 02
Err 04
ROM error Malfunction of the program.
ROM error Malfunction of the adjustment data.
EEPROM error Malfunction of the memory data.
When the error appears in the LCD display, it is necessary to repair the instrument. Contact your authorized Hioki distributor or reseller.
54
Page 61
Display Messages
5.4 Display Messages
Display Description Reference
Performing zero-adjustment. p. 25
The battery is exhausted. Replace the batteries.
Disabling the buzzer. p. 42
Zero-adjustment cannot be performed since the sensor cap in not covered. Affi x the sensor cap.
All saved measured values will be deleted. Continue?
An internal ROM or EEPROM error has occurred. Send the instrument for repair.
The internal memory is full. Delete the measured data in the internal memory.
The display part and sensor unit is not connected. Connect the display unit and sensor unit, either directly or with a cable.
There is no measured data saved in the internal memory.
p. 20
p. 28
p. 38
p. 54
p. 38
p. 37
-
55
Page 62
Error Display
56
Page 63
Appendix
Appx. 1 Recommended Levels of
Illumination (Reference)
Suitable levels of illuminance (according to the JIS standard Z 9110).
Offi ces
Recommended illuminance [lx]
750 500 to 1000 Design rooms, offi ces, board rooms 500 300 to 750 300 200 to 500
200 150 to 300 Kettle rooms, locker rooms, restrooms
Illuminance
level [lx]
Place/work activity
Computer rooms, conference rooms, reception rooms
Reception area, dining rooms, elevator halls
Factories
Recommended illuminance [
1500 1000 to 2000
750 500 to 1000 500 300 to 750
150 100 to 200 Stairways, loading and unloading
50 30 to 75 Indoor emergency stairways
Illuminance
lx
]
level [lx]
Place/work activity
Extremely precision visual work such as producing precision mechanical equipments and electronic parts
Precision visual work such as analyzing in chemical factories
Ordinary visual work in manufacturing plants
Appx.1
Page 64
Recommended Levels of Illumination (Reference)
Schools
Recommended illuminance [
1000 750 to 1500
750 500 to 1000 Precision drawing or drafting 500 300 to 750
300 200 to 500 100 75 to 150 Corridors, connecting corridors, entrance
Illuminance
lx
]
level [lx]
Place/work activity
Precision handicraft, precision experimenting
Experiment demonstration rooms, library reading rooms, nurse’s offi ce, kitchen
Classrooms, gymnasium, offi ce rooms, cafeteria
Appx.2
Page 65
Sensor Characteristics Graphs
Appx. 2
Sensor Characteristics Graphs
Relative Spectral Response Characteristics in the Visible Spectrum
Human perception of brightness ranges from 380 nm to 780 nm in the wavelength and is the maximum at 555 nm. The International Commission on Illumination (CIE) has established comparative standards for luminosity, setting the maximum perception for 1 and indicating the amount of perception of each wavelength by the relative value, and calculating the average of many people. In this instrument, the relative spectral response characteristics are close to the comparative standards for luminosity. The deviation from the comparative standards for luminosity is determined by the f1’ value of JIS standard C 1609-1:2006.
1.2
1.0
0.8
0.6
0.4
0.2
Relative spectral responsivity
0.0 350 400 450 500 550 600 650 700 750 800 850
Wavelength
[ nm ]
Deviation from spectral luminous efficiency
FT3424
Graph illustrates typical characteristics. Characteristics exhibited by individual products may vary slightly.
Appx.3
Page 66
Sensor Characteristics Graphs
Angled Incident Light Characteristics
It is known that the luminance is proportional to the cosine of the incident angle of light (the cosine law). In this instrument, the shape of the light sensor, hook etc. is so made that it can follow the cosine law closely.
1.2
1.0
㻌㻌
0.8
0.6
0.4
0.2
0.0
Angled Incident Light Characteristics
0 102030405060708090
Incident angle
[ ° ]
Ideal Characteristics
㻌㻌㻌
(㻌
ș)
cos
FT3424
Graph illustrates typical characteristics. Characteristics exhibited by individual products may vary slightly.
Appx.4
Page 67
Other Characteristics
Appx. 3 Other Characteristics
Color correction factor for a general light source relative to standard illuminant A
Light source k
Fluorescent lamp F6 1.003 Fluorescent lamp F8 1.002 Fluorescent lamp F10 1.002 High-pressure sodium lamp 1.011 Metal halide lamp H1 1.002 Metal halide lamp H2 1.003 High-pressure mercury lamp 0.995
Table shows typical characteristics. Characteristics exhibited by individual products may vary slightly.
Range of distances in which the law of inverse squares relative to distance applies
50 cm from the measurement reference level
Incidence uniformity
The instrument is designed to be used under conditions in which the illuminance distribution on the sensor surface is roughly uniform. An error component will be introduced into measurements if the illuminance distribution on the sensor surface is non-uniform, for example when measuring a light source with strong directionality.
Appx.5
Page 68
Dimensional drawing
Appx. 4 Dimensional drawing
Unit : mm
㻌㻟㻝㻚㻟㻌
㻾㻱㻲㻚㻸㻱㼂㻱㻸
㻝㻣㻜㻌
Appx.6
㻌㻣㻤㻌
䃥㻞㻜
㻌㻟㻥㻌
Center of tripod mounting thread
Page 69
13-09
Page 70
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Loading...