Hioki FT6031-50 kit User Manual

FT6031-50
EARTH TESTER
Instruction Manual

July 2020 Edition 1 FT6031B961-00 20-07H
EN

Contents

Introduction .........................................................................1
Verifying Package Contents ..............................................5
Options (sold separately) ..................................................6
Safety Notes ........................................................................ 7
Usage Notes ...................................................................... 10
1 Overview 15
1.1 Overview and Features ................................. 15
Overview ................................................................... 15
Features .................................................................... 15
1.2 Part Names and Functions ........................... 17
1.3 How to Use Carrying Case ...........................24
2 Preparing for Measurement 25
2.1 Attaching/Removing Protector ....................25
2.2 Inserting/Replacing Batteries ......................26
Procedure .................................................................28
2.3 Connecting the Z3210 Wireless Adapter
(Option) ..........................................................29
Procedure .................................................................30
3 Making Measurements 31
3.1 Measurement Procedure ..............................31
3.2 Inspecting Before Use ..................................32
3.3 Setting Up Comparator (PASS/FAIL Test) ... 36
How to set up comparator ......................................... 37
How to disable comparator ....................................... 37
FT6031B961-00
i
Contents
3.4 Precise Measurement for Earth Resistance (Precise Measurement Method,
3-pole Method) .............................................. 38
Performing zero adjustment ...................................... 40
Connecting measurement cables .............................41
Measuring earth resistance ......................................45
Stowing ..................................................................... 48
Continuous measurement function ........................... 49
3.5 SimpliedMeasurementforEarth
Resistance
(SimpliedMeasurementMethod,
2-pole Method) ............................................... 51
Performing zero adjustment ...................................... 54
Connecting test leads ...............................................55
Measuring earth resistance ......................................57
Stowing ..................................................................... 57
3.6 Cautions and Tips for Measurement ...........58
Distance between earthing electrodes ......................58
Location to install auxiliary earthing rod ....................59
How to insert/pull out auxiliary earthing rod .............. 60
3.7 Auto Power Save
(Power-saving Function) ..............................62
3.8 Wireless
Communications F
unction
(GENNECT Cross) ...........................................63
3.9 Activating the LCD All-on Display ...............65
3.10 Displaying the Serial Number ......................66
4 Specications 67
4.1 GeneralSpecications .................................67
4.2 Input, Output, and Measurement
Specications ................................................ 69
ii
Contents
5 Maintenance and Service 77
5.1 Repair, Inspection, and Cleaning .................77
5.2 Troubleshooting ............................................79
5.3 Error Display .................................................. 83
6 Appendix 85
6.1 Earth Resistance ........................................... 85
6.2 Measurement Principle ................................. 87
WarrantyCerticate
iii
Contents
iv

Introduction

Introduction
Thank you for choosing the HIOKI FT6031-50 Earth Tester. To ensure your ability to get the most out of this instrument over the long term, please read this manual carefully and keep it available for future reference. Carefully read the separate document entitled “Operating Precautions” before use.
Intended audience
This manual has been written for use by individuals who use the instrument or provide information about how to use the instrument. In explaining how to use the instrument, it assumes electrical knowledge (equivalent of the knowledge possessed by a graduate of an electrical program at a technical high school).
1
Introduction
Notations
Safety notations
In this document, the severity levels of risk and hazard are classied
as follows:
Indicates an imminently hazardous situation
DANGER
WARNING
CAUTION
IMPORTANT
that will result in death of or serious injury to the operator.
Indicates a potentially hazardous situation that may result in death of or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the instrument or malfunction.
Indicates information or content that is particularly important from the standpoint of operating or maintaining the instrument.
Indicates a high-voltage hazard. Failure to verify safety or improper handling of the instrument could lead to an electric shock, burn, or death.
Indicates an action that must not be performed.
Indicates an action that must be performed.
2
Symbols afxed to the instrument
Indicates the need for caution or the presence of danger. For more information about locations where this symbol appears on instrument components, see the "Usage Notes" section
(p. 10), warning messages listed at the beginning of operating
instructions , and accompanying the document entitled "Operating Precautions."
Indicates a double-insulated device.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates the power button that switches the instrument between on and off states.
Symbols for various standards
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Indicates that the product conforms to regulations required by the EU Directive.
Other notations
Instructs the reader to see below for additional information.
*
Indicates the page number to reference.
(p.)
Operation key names are highlighted in bold.
Fn
Introduction
3
Introduction
Screen display
The instrument screen displays the alphanumeric characters as follows:
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
1 2 3 4 5 6 7 8 9 0
Accuracy
Hioki expresses accuracy as error limit values specied in terms of
percentages relative to reading and full scale, and digits.
(Displayed value)
Reading
Full scale
Digits
Refers to the displayed value of the measuring instrument. The limit values of reading errors are expressed in percent of reading (% of reading, % rdg).
(Maximum displayable value) Refers to the maximum displayable value of each measurement range. The instrument has measurement ranges whose values are equal to the maximum displayable values. The limit values of full-scale errors are expressed in percent of full scale (% of full scale, % f.s.).
(Resolution) Refers to the smallest change in the indication on the digital measuring instrument, i.e., the numeral one in the rightmost place. The limit values of digit errors are expressed in terms of digits (dgt).
4

Verifying Package Contents

Verifying Package Contents
When you open the package, carefully inspect the instrument to ensure that everything is in good condition, and that no damage occurred during shipping. Carefully check the accessories, panel keys, and connectors. If the instrument seems to have been
damaged or does not work as specied, contact your authorized
Hioki distributor or reseller.
Check the package contents as follows.
FT6031-50
L9840 Auxiliary Earthing Rod
L9841 Measurement Cable
L9842-11 Measurement Cable
L9842-22 Measurement Cable
C0106 Carrying Case ×1
Protector
AA alkaline battery (LR6) ×4 Instruction Manual Operating Precautions (0990A907)
Accessories
(2 piece set) ×1
(alligator clip, black 4 m) ×1
(yellow 10 m, equipped with winder) ×1
(red 20 m, equipped with winder) ×1
5

Options (sold separately)

Options (sold separately)
The options listed below are available for the instrument. To order an option, please contact your authorized Hioki distributor or reseller. Options are subject to change. Please check Hioki's website for the latest information.
L9787 Test Lead
L9840 Auxiliary Earthing Rod
L9841 Measurement Cable
L9842-11 Measurement Cable
(for simplied measurement method,
indoor use only, red and black 1.2 m each)
(for precision measurement method, 2 pcs in 1 set) (φ 6 mm, entire length of 270 mm, straight section 235 mm, material: stainless SUS304)
(for precision measurement method, alligator clip, black 4 m)
(for precision measurement method, yellow 10 m, equipped with winder)
L9842-22 Measurement Cable
L9843-51
L9843-52
L9844 Measurement Cable
9050 Earth Nets
C0106 Carrying Case
Z3210 Wireless Adapter
(for precision measurement method, red 20 m, equipped with winder)
Measurement Cable (for precision measurement method,
yellow 50 m, equipped with at cable winder)
Measurement Cable (for precision measurement method,
red 50 m, equipped with at cable winder)
(for earthing terminal board, alligator clip, 3 cables in 1 set, red/yellow/black 1.2 m each)
(2 sheets in 1 set, 300 mm × 300 mm)
6

Safety Notes

Safety Notes
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, using the instrument in a way not described in this manual may negate the provided safety features. Carefully read the following safety notes before using the instrument.
DANGER
Mishandling during use could result in injury or death, as well as damage to the instrument. Familiarize yourself with the instructions and precautions in this manual before use.
WARNING
Electricity can potentially cause serious events such as an electric shock, heat generation, re, and an arc ash due to a short-circuit. If you have not used any electrical measuring instruments before, you should be supervised by a technician who has experience in electrical measurement.
7
Safety Notes
Measurement categories
To ensure safe operation of measuring instruments, IEC 61010
species the measurement categories, which classies testing and
measuring circuits into three categories according to the types of mains circuits to which they are intended to be connected.
DANGER
• Do not use a measuring instrument for
measurements on a mains circuit that exceeds the range of the measurement category rated for the instrument. Failure to observe this can cause a severe accident.
Do not use a measuring instrument without a
measurement category rating for measurements on
a mains circuit. Failure to observe this can cause a severe accident.
This instrument conforms to the safety requirements for CAT II 300 V, CAT III 150 V, CAT IV 100 V measuring instruments.
Measurement category II (CAT II)
Applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low­voltage mains installation. Example: Measurements on household appliances, portable tools,
and similar equipment, and on the consumer side only
of socket-outlets in the xed installation.
8
Safety Notes
Measurement category III (CAT III)
Applicable to test and measuring circuits connected to the distribution part of the building’s low-voltage mains installation. Example: Measurements on distribution boards (including
secondary meters), photovoltaic panels, circuit breakers, wiring, including cables, bus-bars, junction boxes,
switches, socket-outlets in the xed installation, and
equipment for industrial use and some other equipment such as stationary motors with permanent connection to
the xed installation.
Measurement category IV (CAT IV)
Applicable to test and measuring circuits connected at the source of the building’s low-voltage mains installation. Example: Measurements on devices installed before the main fuse
or circuit breaker in the building installation.
Distribution Panel
Service Entrance
Service Drop
CAT IV
Power Meter
Internal Wiring
CAT III
Fixed Installation
CAT II
T
Outlet
9

Usage Notes

Usage Notes
Observe the following precautionary information to ensure that the instrument can be used safely and in a manner that allows it to
perform as described in its specications. Use of the instrument should conrm not only to its specications, but also to the specications of all accessories, options, batteries,
and other equipment in use.
Verifying before usage
DANGER
If the test lead or the instrument is damaged, there is a risk of an electric shock. Perform the following
inspection before use:
Check that the insulation of the test leads are
neither ripped nor torn and that no metal parts are exposed. Using the instrument under such conditions could result in an electric shock. Replace the test leads with those specied by Hioki.
• Check the instrument for any damage that may have
occurred during storage or shipping, and perform functional checks before use.
10
Use environment of the instrument
WARNING
Installing the instrument in inappropriate locations
may cause a malfunction of instrument or may give
rise to an accident. Avoid the following locations that
are:
Exposed to direct sunlight or high temperatures
Exposed to corrosive or combustible gases
Exposed to strong electromagnetic elds or
electrostatic charges
• Near induction heating systems (such as high-
frequency induction heating systems and IH
cooking equipment)
• Susceptible to vibration
Exposed to water, oil, chemicals, or solvents
Exposed to high humidity or condensation
Exposed to high concentrations of dust particles
CAUTION
• Do not place the instrument on an unstable or uneven surface. Doing so could cause the instrument to fall or turn over, causing bodily injury or damage to the instrument.
• Do not place the instrument on an unstable or uneven surface. Doing so could cause the instrument to fall or turn over, causing bodily injury or damage to the instrument.
Usage Notes
11
Usage Notes
The protection rating for the enclosure of the instrument (based on EN60529) is IP65/67*.
CAUTION
Although the instrument has a dust-proof, jet-proof and waterproof structure, it is not for completely shutting out the water intrusion into the inside. Please note that it can be a cause of failure.
*IP65/IP67:
This indicates the degree of protection provided by the enclosure of the device against use in hazardous locations, entry of solid foreign objects, and the ingress of water.
6: Protected against access to hazardous parts with wire
measuring 1.0 mm in diameter. Dust-proof type (Dust shall not penetrate the enclosure.)
5: The equipment inside the enclosure is protected against
the harmful effects of water projected in jets against the enclosure from any direction.
7: Watertight (Quantities of water that may harm the enclosure
when it is temporarily immersed in water shall not penetrate the enclosure.)
For details on the operating temperature and humidity range, see
the specications. (p. 67)
12
Usage Notes
Handling the cables
CAUTION
To prevent cable damage, do not step on cables or pinch them between other objects. Do not bend or pull on cables at their base.
The ends of the L9787 Test Lead and L9840 Auxiliary Earthing Rod are sharp. Be careful to avoid injury.
IMPORTANT
Use only the specied auxiliary earthing rod, cables, and test leads. Using a non-specied cable may result in incorrect
measurements due to poor connection or other reasons.
Precautions during measurement
WARNING
If the instrument is used in locations where the rating indicated on the instrument or cords is exceeded, the
instrument may be damaged resulting in personal
injury. Do not use the instrument in such locations. See “Measurement categories” (p.8)
Precautions during shipment
Be sure to follow these precautions when transporting the instrument:
• Remove accessories and optional equipment from the instrument in order to avoid damage. Additionally, use the original packing materials, and be sure to double box the instrument. Accidental damage suffered in transit is not covered by the warranty.
• Attach a description of the issue when sending out your instrument for repair.
13
Usage Notes
14

Overview

1

1.1 Overview and Features

Overview

The grounding works applied to the distribution lines and electrical
facilities are essential for preventing electric shock and re as well
as for safeguarding the equipment. The instrument is an earth tester that is fully functional for measuring earth resistance for grounding works. It is capable of providing accurate and highly reliable measurement. The instrument can provide precision measurement (3-pole method)
and simplied measurement (2-pole method).

Features

High accuracy
Auto-range and
auto-check for
auxiliary earthing
electrode
High allowable
earth resistance for
auxiliary earthing
electrode
The accuracy of the 200.0
±1.5% rdg ±4 dgt.
The measurement is performed by simply pressing the MEASURE button. There is no need for cumbersome range-switching. In addition, the earth potential and auxiliary earthing electrode are automatically checked.
The resistance that the auxiliary earthing
electrode can tolerate is now about 10 times
higher than the conventional level. As a result, it can provide measurement under adverse conditions.
range is
Ω
15
Overview and Features
Dustproof,
jet-proof and
waterproof
Supplied with
winders
Drop-proof
(when equipped with
a protector)
The instrument can withstand water at a
depth of 1 m for 30 minutes.
The instrument is supplied with useful winders so that it can be easily prepared and packed up before/after measurement.
Its robust structure can withstand a drop from 1 m onto concrete surface.
16
Part Names and Functions

1.2 Part Names and Functions

Front
Cover of the measurement terminals (p. 48)
Protector
Removed when replacing the batteries or attaching
the Z3210. (p. 25, p. 28, p. 30)
Operation buttons
(p. 19)
Measurement terminals
E Connected with the black cable. S(P) Connected with the yellow cable. H(C) Connected with the red cable.
Display
(p. 22)
IMPORTANT
Please do not press the buttons of the instrument with a sharp object. It may damage the instrument.
17
Part Names and Functions
Rear and sides
Battery cover (side view)
Waterproof seal
It needs to be replaced when degraded. Please contact your authorized Hioki distributor or reseller.
Battery cover
Removed when replacing the
batteries or attaching the Z3210. (p. 28, p. 30)
Serial number
Please do not remove the label as it is needed for product control such as product warranty etc.
(The serial number consists of 9 digits. The rst two (from the left)
indicate the year of manufacture, and the next two indicate the month of manufacture.)
18
Operation buttons
Part Names and Functions
1
2
3
4
5
6
7
19
Part Names and Functions
1
2
3
4
5
6
MEASURE button
• Start/stop for measuring earth resistance
Fn button
• Displays the resistance of each earthing electrode.
(p. 45)
Display switching (3-pole method, 2-pole method)
• DC/AC auto switching (when measuring earth potential)
• Wireless communications function on/off (Press and hold
for at least 1 s.)* (p. 63)
3P/2P button
Switches between 3-pole method and 2-pole method.
• Releases the measured value when it is hold and displays the current earth potential value.
• Continuous measurement on/off (Press and hold for at least 1 s.)
Power button
• Power on/off
Live wire warning LED
• For 3-pole method Blinks under any of the following conditions:
A voltage of 30 V or higher is applied between the S(P) and E terminals.
A voltage of 85 V or higher is applied between the H(C) and E terminals.
A voltage of 85 V or higher is applied between the H(C) and S(P) terminals.
For 2-pole method
Blinks when a voltage of 30 V or higher is applied
between the H(C) and E terminals.
• Lights up during earth resistance measurements.
COMP button
• Sets/cancels comparator setting. (The COMP indicator
lights up/goes off.) (p. 37)
• Wireless communications function on/off (Press and hold
for at least 1 s.)* (p. 63)
20
0ΩADJ
7
*: With Z3210 installed
• Sets/cancels zero adjustment setting. (Press and hold
for at least 1 s.) (p. 40, p. 53)
Part Names and Functions
button
21
Part Names and Functions
Display
8
9
10
1 2
11
7 3 4
5
6
1
2
3
4
5
6
7
8
Comparator comparison result (PASS)
Comparator comparison result (FAIL)
Appears when DC/AC auto-detect setting is enabled for measuring earth potential.
Appears when continuous measurement function is enabled.
3-pole method (p. 38)
2-pole method (p. 51)
Zero adjustment (
COMP
Appears when comparator setting is enabled. (p. 36)
Indicates that the measured value i s being retained.
Appears when earth potential exceeds allowable range.
Indicates wireless communication function status* (p. 63)
Flashes: Communicating Lights up: Wireless function on Lights off: Wireless function off
p. 40, p. 54
)
22
Part Names and Functions
Indicates the battery level. (p. 27)
9
10
11
*: Optional Z3210 Wireless Adapter required separately
See p. 83 for error display.
Appears 30 s before auto-power-save function is activated.
APS
(p. 62)
Indicates where to connect a measurement cable or test lead.
Appears when the resistance of each (auxiliary) earthing
OPEN
electrode is high or a measurement cable is not connected.
Earth resistance of earthing electrode
Earth resistance of auxiliary earthing electrode S
Earth resistance of auxiliary earthing electrode H
Live wire warning indicator (Blinks during earth resistance measurements)
23
How to Use Carrying Case

1.3 How to Use Carrying Case

Please store the instrument, winder and other accessories/options
into the C0106 Carrying Case as shown in the gure below.
• Please do not store commercially available pegs in this carrying case as those have sharp tips. It may damage the case.
• Please do not wash the carrying case.
• Winder
• Auxiliary
Earthing Rod
24
Earth Tester
Please make sure to retract the winder knob before storing it in the carrying case.
Measurement cable (Black)
Instruction Manual

Preparing for Measurement

2
2.1

Attaching/Removing Protector

Attaching the protector
Tilt the instrument and slide it into the protector, and then push the entire instrument into the protector.
Removing the protector
Hold it with both hands and push the one end of the protector down.
IMPORTANT
If the protector is removed, the drop-proofness (p. 67) will be
void.
25
Inserting/Replacing Batteries

2.2 Inserting/Replacing Batteries

Before using the instrument, insert four AA alkaline batteries (LR6). Before measurements, check that the battery level is sufcient.
When the battery charge is low, replace the batteries.
WARNING
• To avoid electric shock, turn off the instrument and
disconnect the measurement cables or test leads before replacing batteries.
• Batteries may explode if mistreated. Do not short-
circuit,disassemble the batteries, or dispose of
them in re. Do not recharge alkaline batteries.
Handle and dispose of batteries in accordance with local regulations.
After inserting the batteries, reattach the cover,
screws, and put on the protector before using the instrument.
• To prevent instrument damage or an electric shock,
use only the screws that are originally installed for securing the battery cover in place. If you have lost
any screws or nd that any screws are damaged,
please contact your authorized Hioki distributor or reseller.
26
Inserting/Replacing Batteries
CAUTION
Poor performance or damage from battery leakage could result. Observe the cautions listed below.
• Do no mix old and new batteries, or different types of batteries
• Pay attention to the polarity markings “+” and “–”, so that you do not insert the batteries the wrong way around.
• Do not use batteries after their recommended expiry date.
• Do not leave a depleted batteries inside the instrument.
Replace batteries only with the specied type.
• Remove the batteries from the instrument if it is to be stored for a long time.
• The indicator ashes when the remaining battery capacity is low. In this case, measurement is not possible. Replace the batteries.
After use, always turn off the instrument.
Although rechargeable batteries (nickel-hydride) can be used for measurement, the battery level indicator will not be accurately displayed.
• When replacing the batteries after cleaning, the replacement should be carried out after the instrument has been completely
dried.
• The operating temperature of the batteries included in the
shipment is −10°C to 45°C (14°F to 113°F). When using the
instrument outside this temperature range, use batteries that can support such a low or high temperature range. (Example: lithium
battery)
Battery warning indicator
Fully charged.
As the battery charge diminishes, black charge bars disappear, one by one, from the left of the battery indicator. The battery voltage is low. Replace the batteries as soon as possible.
(Blinks) The battery is exhausted. Replace the batteries.
27
Inserting/Replacing Batteries

Procedure

You will need:
Phillips-head screwdriver (No. 2)
AA alkaline battery (LR6) ×4
1
Battery Cover
2
Rear
3
4
5
6
Turn off the instrument and remove the measurement cables or test leads from the instrument.
Remove the protector. (p. 25)
Unscrew the screws of the battery cover with a Phillips­head screwdriver.
Remove the battery cover.
Do not remove the waterproof seal
from the battery cover. (p. 18)
When replacing the batteries, all of the old batteries should be removed.
Pay attention to the polarities of the batteries when inserting
four new batteries (LR6).
28
Reattach the battery cover and
7
tighten the screws.
Attach the protector. (p. 25)
8

Connecting the Z3210 Wireless Adapter (Option)

2.3 Connecting the Z3210 Wireless
Adapter (Option)
The wireless communications function can be used by connecting
the Z3210 Wireless Adapter (option) to the instrument. See “3.8 Wireless Communications Function (GENNECT Cross)” (p. 63).
WARNING
To avoid electric shock, turn off the instrument and disconnect the measurement cables or test leads before replacing batteries.
• After connecting the Z3210, reattach the cover,
screws, and put on the protector before using the instrument.
• To prevent instrument damage or an electric shock,
use only the screws that are originally installed for securing the battery cover in place. If you have lost
any screws or nd that any screws are damaged,
please contact your authorized Hioki distributor or reseller.
29
Connecting the Z3210 Wireless Adapter (Option)

Procedure

You will need:
Phillips-head screwdriver (No. 2)
Flat-head screwdriver
Z3210 Wireless Adapter
Turn off the instrument and
1
Rear
Use a at-head screwdriver or
other suitable tool to remove
the Z3210.
3,8
6
4,7
5
remove the measurement cables or test leads from the instrument.
Remove the protector.
2
(p. 25)
Unscrew the screws of the
3
battery cover with a Phillips­head screwdriver and Remove the battery cover.
Do not remove the waterproof seal
from the battery cover. (p. 18)
Remove the batteries from
4
the instrument
Remove the protective cap
5
with a athead screwdriver.
Exercising care to orient the
6
Z3210 correctly, install the Z3210 as far as it will go.
Exercising care to the
7
polarities of the batteries when inserting them to the instrument.
Reattach the battery cover
8
and tighten the screws.
Attach the protector.
9
(p. 25)
30

Making Measurements

3

3.1 Measurement Procedure

Before using the instrument, be sure to read “Usage Notes” (p. 10).
Preparing for measurement
Insert the batteries. (p. 26)
As necessary, have other
Perform the startup check. (p. 32)
Measurement
Turn the power on and select the measurement method.
Connect measurement cables or test leads to the measurement terminals.
Perform zero adjustment. ( 3-pole method [p. 40],
2-pole method [p. 54])
Start the measurement.
optional items available and ready.
Finishing the measurement
Turn the power off and remove measurement cables or test leads from the object under measurement.
31
Inspecting Before Use

3.2 Inspecting Before Use

Check the instrument/device for any damage that may have occurred during storage or shipping, and perform functional checks
before use. If you nd any damage to the instrument, please contact
your authorized Hioki distributor or reseller for repair.
Check items Solution
Is the battery level sufcient?
Is there any missing segment in
the display items?
Is there any damage or crack in
the instrument?
Is there any foreign material
(sand etc.) inside the measurement terminals?
Is there any damage or exposed internal white part or metal in the coating of measurement cable or
test lead?
Check the battery level indicator on the upper right corner while the power has been turned on. If battery level is low. Replace the batteries with new ones.
(p. 26)
Check by activating the LCD all-on display.
(p. 65)
If there is a missing segment, the instrument needs to be repaired.
Conduct visual checking. If damage is found, the not be used and needs to be repaired as it can cause electric shock.
Remove all foreign materials if any. If it cannot be removed, the to be repaired.
If damage is found, it should not be used and needs to be replaced as it can cause electric shock.
is blinking, the
instrument
instrument
should
needs
32
Check items Solution
Use the following method to check whether the measurement cable or test lead has a break.
1. For 3-pole method
Press the 3P/2P button to display
For 2-pole method
Press the 3P/2P button to display
2. Connect measurement
cables or test leads to the instrument their tips.
.
.
and short circuit
Inspecting Before Use
If not indicating approx. 0
• The measurement cable or test lead has not been fully inserted.
→Fully insert the cord or lead.
• The measurement cable or test lead may have a break.
→ Replace it with another lead that is
specied by Hioki. If the symptom
persists even after the measurement cable or test lead is replaced, the instrument may have a failure. The instrument needs to be repaired.
Ω
3. Press the MEASURE button
to check that approx. 0
indicated.
is
Ω
33
Inspecting Before Use
• Connect the instrument to the secondary side of
a distribution panel. If a short-circuit occurs on the secondary side of the distribution panel, the panel will interrupt the short-circuit current. Do not connect the instrument to the primary side of the distribution panel because an unrestricted current
ow can damage the instrument and facilities if a
short-circuit occurs.
• Do not cause a short-circuit between another wire
and the wire to be measured with the test leads. Arcs or such grave accidents are likely to occur.
• To prevent a short-circuit or electric shock, do not
touch the metal part of the connecting test leads tip.
To prevent an electric shock, conrm that any of the white or red portion (insulation layer) of the L9787 Test Lead and the L9844 Measurement Cable is not
exposed. If the cable is exposed, do not use the cable.
DANGER
WARNING
34
Inspecting Before Use
WARNING
To prevent an electric shock, observe the following precautions:
• Prior to measurement, please make sure that the
earthing electrode has been disconnected from the distribution system. The Measurement Cables
L9841, L9842-11, L9842-22, L9843-51 and L9843-52
are measurement cables with the maximum rated voltage of 30 V (between input terminals and the
ground) and are designed to measure the earth
resistance of an earthing electrode disconnected from the distribution system.
• Turn off all power before connecting measurement
cables and test leads.
• Connect measurement cables or test leads to
terminals securely. If a terminal is loose, the contact resistance will increase, resulting in overheating,
equipment burnout, or a re.
CAUTION
• To avoid damaging the measurement cables or test leads, unplug them by grasping the connectors, not the cables or cords.
35

Setting Up Comparator (PASS/FAIL Test)

3.3
Setting Up Comparator (PASS/FAIL Test)
The instrument has a comparator function, which can indicate PASS or FAIL with the display indicator and beep. The earth resistance can be measured without setting up the comparator.
The comparator function operates as follows:
Comparison result Display Beep
Measured value ≤Reference value (PASS)
Measured value > Reference value (FAIL)
The comparison reference value can be chosen from the reference values in the following table.
Reference value (Ω)
1 2 3 4 5
6 7 8 9
10 20 30 40 50
60 70 80 90
100* 200 300 400 500 OFF
*: Default
Intermittent
Continuous
36
Setting Up Comparator (PASS/FAIL Test)

How to set up comparator

Press the 3P/2P button to set
1
to
or (2-pole method).
Press the COMP button.
2
The indicator blinks, and the
2,3
1
The comparator setting is saved even after the power is turned off.
resistance for the comparison reference appears.
Press the COMP button again
3
and then choose a comparison reference.
If there is no operation for approx. 2 s after
the desired comparison reference has
been chosen, the comparator is enabled and the display.
(3-pole method),
indicator appears in the

How to disable comparator

Press the COMP button while the comparator has been
1
enabled.
The indicator blinks.
Press the COMP button several times until the OFF
2
indicator appears on the display.
The comparator will be disabled if no button is pressed for about 2 s after
the OFF indicator appears on the display.
37
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
3.4 Precise Measurement for Earth Resistance (Precise Measurement Method,
3-pole Method)
WARNING
The instrument can output a voltage of approx. 30 V. Although the instrument has a dustproof, jet-proof and waterproof structure, the instrument should always be dried before using it for measurement so as to avoid electric shock.
CAUTION
Do not connect the measurement cables if any foreign material remains inside the measurement terminals. It may cause failure.
There are two types of measurement method for earth resistance:
precision measurement method (3-pole method) and simplied measurement method (2-pole method), and the precision measurement method (3-pole method) is the basic measurement
method for earth resistance.
The simplied measurement method is used when measurement
cannot be performed by the precision measurement method. The precision measurement is performed by inserting two auxiliary
earthing rods into the ground as shown in the gure on p. 42.
Measurement of large-scale earthing electrodes
When measuring a large-scale earthing electrode such as a mesh earthing electrode, ring earthing electrode, or earthing electrode provided by a large building structure, it cannot be accurately
measured since the H(C) electrode and S(P) electrode come inside
38
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
the earth resistance area of E electrode.
If long cables are used to avoid the H(C) electrode and S(P)
electrode getting inside the earth resistance area, accurate
measurement cannot be carried out as it is signicantly affected by
noise. Generally, measurement of large-scale earthing electrode requires
a large measurement current of approx. 20 A. Use measuring
instrument designed for measuring large-scale earthing electrodes for this measurement. (No measuring instrument available for this
purpose from Hioki.)
39
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)

Performing zero adjustment

Always perform zero adjustment prior to measurement. The range that the instrument can provide zero adjustment for is 3 Ω or less. If the measured value exceeds 3 Ω, the message Err
0ΩADJ appears and zero adjustment will not be executed.
Remove the cover of the
1
6
5
2
7
4
How to disable zero adjustment
Press and hold the 0
1
measurement terminals.
Connect black, yellow and red
2
measurement cables respectively
to E, S(P), and H(C) terminals of
the instrument.
Short-circuit the tips of the three
3
measurement cables.
Press the Power button to turn it
4
on.
Press the 3P/2P button to display
5
Earth potential appears on the display.
Press the MEASURE button.
6
3
ADJ button for at least 1 s while
Ω
A measured value of approx. 0 Ω appears
in approx. 8 s, and
When the OPEN indicator appears
(p. 23), the measurement cable may not
have been connected to the measurement terminal, or the cable may have a break. Check the connection and perform cable check with a tester etc.
Press the 0ΩADJ button.
7
adjustment is completed.
(3-pole method).
lights up and then zero
lights up.
is on.
40
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)

Connecting measurement cables

WARNING
• Do not use the instrument to measure circuits that
exceed its ratings or specications. Damage to the
instrument can cause an electric shock.
• To prevent an electric shock, be careful to avoid
shorting live lines with the test leads tip.
If the insulation on a cord melts, the metal conductor may be exposed. Do not use any cord whose metal conductor is exposed. Doing so could result in an electric shock, burn, or other hazards.
CAUTION
• To prevent cords damage, do not step on cords or pinch
them between other objects. Do not bend or pull on cords at their base.
• The ends of the auxiliary earthing rod are sharp. Be
careful to avoid injury.
• For safety reasons, when taking measurements,
only use the measurement cables provided with the instrument.
• The cable is hardened in freezing temperatures. Do not
bend or pull it to avoid tearing its shield or causing a break.
41
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
5 3
Auxiliary Earthing Electrode H
10 m 10 m
Auxiliary Earthing Electrode S
1
Earthing Electrode E
Black
Yellow
Red
E
HS
42
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
Use the measurement cable (black)
1
to connect between earthing electrode and E terminal.
Carry two winders along to the
2
measuring location while pulling out the measurement cables.
2
• Insert the auxiliary earthing rods into a moist layer in the ground. Since the instrument can accept a large resistance of auxiliary earthing electrode, the auxiliary earthing rods do not need to be inserted unnecessarily deep into the ground
• For accurate measurement, the distance between E, S, and H
needs to be approx. 5 m. The measurement cables (yellow) and (red) should be positioned approx. 10 cm away from each other.
Make sure the cables do not tangle or overlap together.
At the location where the
3
measurement cable (yellow)
has been fully pulled out, insert the auxiliary earthing rod into the ground and connect the
measurement cable (yellow).
Carry the winder (measurement
4
cable: red) along the straight line
to a further distance between the earthing electrode E and auxiliary earthing electrode S while pulling out the measurement cable.
At the location where the
5
measurement cable (red) has been
fully pulled out, insert the auxiliary earthing rod into the ground and connect the measurement cable
(red).
43
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
NOTE
Measurement on concrete
Since concrete is conductive, auxiliary earthing electrodes can be installed on concrete. Place an auxiliary earthing rod on concrete and pour water over it, or cover the auxiliary earthing rod with a wet rag to form an auxiliary earthing electrode. If the earth resistance of the auxiliary earthing electrode is not reduced by
the above methods, place the optional 9050 Earth Nets on concrete and then
position the auxiliary earthing rod on the Earth Nets and pour water over it. Before measurement, allow some time for the water to well soak into the concrete.
• As an alternative to the Earth Nets, a metal plate or aluminum foil etc. may
be used.
• Since asphalt is insulator, it is generally not possible to install the auxiliary
earthing electrode on asphalt. However, measurement may be possible on asphalt that has water permeability.
9050 Earth Nets
44
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)

Measuring earth resistance

Press the Power button to turn it
1
on.
3
2 1
The displayed parameter can be changed after measurement. Press the Fn button to switch parameters.
When NOISE appears
If the earth potential is high (25 V rms or 35.4 V peak or higher),
and the earth potential peak value appears on the display. When it is not possible to measure earth resistance.
Since leak current is owing into the earthing electrode, rst
disconnect any electric equipment connected to the earthing electrode and then press the MEASURE button again.
• The instrument automatically detects AC/DC ( / ) of the earth potential. If AC/DC needs to be switched, it can be switched by pressing the while the current earth potential has been displayed.
is on, or when (live wire warning LED) is blinking,
Press the 3P/2P button to display
2
(3-pole method).
Earth potential appears on the display.
Press the MEASURE button.
3
Automatically executes the earth potential
check→auxiliary earth resistance check→earth resistance measurement in
sequence. The measurements will be completed in approx. 6 s, and then the measured value appears and
Check the measured value.
4
lights up.
Fn
button
45
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
• To check earth potential After measurement of earth resistance, press the Fn button while
is lit so that the earth potential will be displayed. When the 3P/2P button so that the retained value is unlocked and the current
earth potential will be displayed.
PENが表 示され るとき
is lit, press
When OPEN appears
The instrument automatically check whether or not each earth resistance is within the allowable range by measuring the earth resistance of the auxiliary earthing rod prior to measuring the resistance of the earthing electrode E. If the earth resistance of the auxiliary earthing electrode exceeds the allowable range, the OPEN indicator appears below the measurement terminal that exceeds the allowable range and the earth resistance of the earthing electrode E will not be measured.
Countermeasures when OPEN appears
Take the following actions for the auxiliary earthing rod connected with the measurement terminal for which the OPEN indicator appears:
• Pour water
• Stick in deeper
• Insert the auxiliary earthing rod into another location
• If the clip connecting point is dirty, wipe and clean the auxiliary
earthing rod with a soft cloth.
The Fn button allows to display the earth resistance and earth potential of each earthing electrode.
Earth resistance of earthing electrode E
Earth resistance of auxiliary earthing electrode S
Earth resistance of auxiliary earthing electrode H
V
Earth potential
46
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
• The auxiliary earthing electrodes’ earth resistance (RS, RH)
measured values are used to automatically check whether the earthing electrode’s earth resistance (RE) can be measured. RS and RH do not provide the same level of resolution as the earthing electrode’s earth resistance (RE). For the resolution of earth resistance of auxiliary earthing
electrode, see “Specications” (p. 67).
• When the resistance (RH) of auxiliary earthing electrode exceeds
5 kΩ, the display range becomes the range of 200 Ω without displaying the digit of 0.01 Ω even though the resistance of earthing electrode is 20 Ω or less. The following table shows
examples.
If the digit of 0.01 Ω is required, pour some water over the auxiliary earthing rod (H) to reduce the earth resistance to 5 kΩ
or less.
Resistance RH of auxiliary earthing
electrode H
0 to 5 k
Ω
to 50 k
5 k
Ω
Ω
Example of indicated
value (1)
When 9.52
9.52 Ω (Range: 20 Ω) 13.48 Ω (Range: 20 Ω)
9.5 Ω (Range: 200 Ω) 13.5 Ω (Range: 200 Ω)
Ω
measured
Example of indicated
is
value (2)
When 13.48 Ω is
measured
47
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)

Stowing

Cover of the measurement terminals
IMPORTANT
Always t the cover of the measurement terminals immediately
after use. Any foreign object entering the measurement terminal may cause a failure.
Press the Power button to turn it
1
off.
Remove the measurement
2
cables from the measurement
terminals and t the cover of the
measurement terminals.
Remove the measurement cables
3
from the auxiliary earthing rods and pull out the auxiliary earthing rods without bending them.
(See p. 61 for how to pull out.)
Rewind the measurement cables
4
(red, yellow) with the respective
winders and stow the winders in the carrying case with the auxiliary earthing rods inserted into the winder's holders.
Remove the measurement cable
5
(black) from the earthing electrode
and fold it, and then stow it into
the carrying case. (p. 24)
48
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)

Continuous measurement function

This function is convenient when you wish to insert an auxiliary earthing rod into the ground while checking measured values.
IMPORTAMT
• Measured values provided by the continuous measurement function may differ from those provided by normal earthing
resistance measurement (p. 45).
• Measured values provided by the continuous measurement function are for reference purposes only. If you require precise values, be sure to measure earthing resistance using the
measurement method described on p. 45.
Enabling the continuous measurement function
Press the Power button to turn it
1
on.
Press and hold the 3P/2P(CONT
2
3
2 1
button for at least 1 s until the
indicator appears on the
display.
Press the MEASURE button.
3
The instrument will measure earthing resistance continuously. To switch the parameter displayed during measurement, press the Fn button.
49
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
Disabling the continuous measurement function
Press the MEASURE button
1
during continuous measurement.
1
2
Continuous operating time
When using four AA alkaline batteries (LR6) (reference value at 23°C)
Approx. 3.5 hours (3-pole method, auxiliary earthing resistance of 100
• Approx. 3 hours (3-pole method, auxiliary earthing resistance of 100
measuring 10 without Z3210 installed)
measuring 10 with Z3210 installed, using wireless communications)
Continuous measurement will stop.
Press and hold the 3P/2P(CONT
2
button for at least 1 s until the
indicator is no longer
shown on the display.
using the 20 Ω range,
Ω
using the 20 Ω range,
Ω
,
Ω
,
Ω
50
Simplied Measurement for Earth Resistance
(Simplied Measurement Method, 2-pole Method)
3.5 Simplied Measurement for Earth
Resistance
(Simplied Measurement Method, 2-pole Method)
DANGER
Use the neutral side (ground side) of the commercial
power supply for this measurement. Prior to connection, use a voltage detector etc. to make sure that it is going to be connected with the neutral side and take caution for electric shock.
• The instrument can be connected to the neutral
side of an outlet with a voltage-to-ground of 300 V or less, or to the wire on the neutral side of the breaker's secondary side with a voltage-to-ground
of 150 V or less. Do not connect with anything other than those specied above. It can be hazardous.
WARNING
Use the optional L9787 Test Lead for simplied
measurement for safety. Connecting the
Measurement Cables L9841, L9842-11, L9842-22, L9843-51 and L9843-52 to a commercial power
supply may cause electric shock.
• The instrument can output a voltage of approx. 30 V.
Although the instrument has a dustproof, jet-proof and waterproof structure, the instrument should always be dried before using it for measurement so as to avoid electric shock.
51
Simplied Measurement for Earth Resistance
(Simplied Measurement Method, 2-pole Method)
WARNING
• Removable sleeves are attached to the metal pins at
the ends of the test leads. To prevent a short circuit accident, be sure to use the test leads with the sleeves attached when performing measurements in the CAT III measurement category.
See “Measurement categories” (p.8).
• If the sleeves are inadvertently removed during
measurement, stop the measurement.
CAUTION
When the instrument is set to 2-pole method, even if it is connected to the earth side of a commercial power supply, it will not trigger the earth leakage breaker since the measurement current is suppressed to 4 mA or less. However, do not use this method for measurement at a location where an earth leakage breaker or leakage
relay with the current sensitivity of less than 10 mA has
been installed, since such an earth leakage breaker or leakage relay may be triggered.
The simplied measurement function of the instrument is a function for measuring earth resistance with small inductance component. Therefore, a resister connected in series with inductance component of 3 mH or more may not be accurately measured.
• When carrying out measurements with the sleeves in place, be careful to avoid damaging the sleeves.
The L9787 is a test lead specically designed for indoor use. Do not use it outdoor.
52
Simplied Measurement for Earth Resistance
(Simplied Measurement Method, 2-pole Method)
Removing and attaching the sleeves of L9787 Test Lead
CAUTION
The tips of the metal pins are sharp and may cause injury. Do not touch the tips.
Removing the sleeves Hold the bottom of the sleeves and pull the sleeves off. Safely store the removed sleeves so as not to lose them.
Attaching the sleeves Insert the metal pins of the test leads into the holes of the
sleeves, and rmly push them all the way in.
What is simplied measurement method (2-pole method)?
Simplied measurement method (2-pole method) is a measurement
method to check the earth resistance of the equipment earthing with an earth system called TT method. In case of not being able to insert an auxiliary earthing rod, the earth resistance is obtained with use of an existing low earthing resistor as an auxiliary electrode.
In this method, the measured value is the sum (Rx+Ro) of the earth
resistances of the object under measurement and existing earthing resistor, according to the measurement principle. Therefore, the earth resistance of the existing earthing resistor used needs to be lower than that of the earthing electrode of the object under measurement.
In the simplied measurement method, the resistance of the earthing element
used is added to the measurement result.
It is very difcult to measure 10
or less in the simplied method
Ω
53
Simplied Measurement for Earth Resistance
(Simplied Measurement Method, 2-pole Method)

Performing zero adjustment

The range that the instrument can provide zero adjustment for is 3 Ω or less. If the measured value exceeds 3 Ω, the message Err
0ΩADJ appears and zero adjustment will not be executed.
Remove the cover of the
1
6
4
2
E
7
3
How to disable zero adjustment
Press and hold the 0
1
measurement terminals.
Connect the E terminal and
2
H(C) terminal of the instrument respectively with the L9787 Test Lead (black) and L9787 Test Lead (red).
Press the Power button to turn it
3
on.
Press the 3P/2P button to display
4
H(C)
A measured value (approx. 0 V) of the earth
potential appears.
Connect (short-circuit) the tips of
5
test leads.
Press the MEASURE button.
6
A measured value of approx. 0 Ω appears
5
ADJ button for at least 1 s while
Ω
in approx. 3 s, and
When the the measurement terminal, a test lead may not have been connected to the measurement terminal, or the lead may have a break. Check the connection and perform cable check with a tester etc.
Press the 0ΩADJ button.
7
adjustment is completed.
(2-pole method).
lights up and then zero
lights up.
indicator appears below
is on.
54
Simplied Measurement for Earth Resistance
(Simplied Measurement Method, 2-pole Method)

Connecting test leads

WARNING
• Do not use the instrument for measuring the voltage
of commercial power supply.
• If there is a large voltage on the neutral side, (live
wire warning LED) blinks and beep sound goes
off. Immediately remove the test leads from the commercial power supply.
• If the earth potential is high, attention needs to be
paid for the risk of electric shock.
• To prevent an electric shock, do not exceed the
every rating shown on either the instrument or each test lead, whichever is worse.
The gure below shows an example of connection in the case of using a commercial power supply that is grounded on its N (neutral)
side.
Home distribution panel
Master breaker
Earth leakage
breaker
Branch breaker
Pole-mounted transformer
Outdoor Indoor
Ro
Rx
Central earth terminal
55
Simplied Measurement for Earth Resistance
(Simplied Measurement Method, 2-pole Method)
Use a voltage detector etc. to make
1
sure that there is no voltage on
4
3 2
5
E
H(C)
the N (neutral) side of commercial
power supply.
Press the Power button to turn it
2
on.
Press the 3P/2P button to display
3
(2-pole method).
When (3-pole method) is chosen, the earth leakage breaker etc. may be triggered due to its large measurement current.
Connect the L9787 test lead (black)
4
with the earthing electrode of object under measurement.
Connect the L9787 Test Lead
5
(red) with the N (neutral) side of
commercial power supply.
Earth potential appears on the display.
• When
or higher) exists between the earth and the N (neutral) side of commercial power supply. A large leak current may be owing through the earthing
electrode of the object under measurement or through the earthing element of the commercial power supply used for measurement. Therefore, conduct an insulation resistance test or leakage current test etc. In addition, disconnect any electric equipment from the earthing electrode.
• The instrument automatically detects AC/DC of the earth potential.
• If AC/DC needs to be switched during earth potential measurement, it can
be switched by pressing the Fn button.
• After measurement of earth resistance, press the Fn button while
is lit so that the earth potential will be displayed. When press the 3P/2P button so that the retained value is unlocked and the current earth potential will be displayed.
• When
possible to execute earth resistance measurement.
appears, a large earth potential (25 V rms or 35.4 V peak
is lit,
is on, or when (live wire warning LED) is blinking, it is not
56
Simplied Measurement for Earth Resistance
(Simplied Measurement Method, 2-pole Method)

Measuring earth resistance

Press the MEASURE button so that the instrument automatically executes the following measurements in sequence. The measured
value appears in approx. 3 s and
Check earth potential.
1
Check whether or not the peak value of the earth potential is within the allowable range.
Measure earth resistance.
2
Measure the sum (Rx + Ro) of the earth resistance of earthing electrode
and the earth resistance on the neutral side of commercial power supply. To switch the parameter displayed during measurement, press the Fn button.
lights up.

Stowing

Press the Power button to turn it off for safety.
1
Remove the test leads from the measurement terminals and
2
t the cover of the measurement terminals. (p. 48)
Fold the test leads together and stow it into the carrying
3
case.
57
Cautions and Tips for Measurement

3.6 Cautions and Tips for Measurement

Distance between earthing electrodes

When the distance between E-H(C) is I m as shown in Figure (a), if
the resistance of the earthing element E is measured while changing the distance x m between the electrodes E-S(P), measurement
results such as shown in Figure (b) are obtained.
Therefore, its error becomes greater as the location of the auxiliary
earthing rod S(P) moves closer to the earthing element E or auxiliary earthing rod H(C). In addition, if the distance between the electrodes E-H(C) is short,
the measurement error is greater as the earth resistance under test
(Rx) and the earth resistance of auxiliary earthing rod (Rc) cannot
be separated from each other. If earthing is provided by a building structure etc. in a large area,
the resistance area of earth resistance (Rx) indicated in Figure (a)
becomes very wide. In order to carry out accurate measurement, the auxiliary earthing
rods (S[P] and H[C]) need to be installed at a location well away from the earthing element (Rx). As to conrm the above, carry out measurements at several locations while moving the location of the auxiliary earthing rod S(P) from the earthing element (Rx) to the auxiliary earthing rod H(C) and check whether or not there is an almost at section shown in Figure (b) in the measured resistance even though the auxiliary earthing rod S(P) has been moved. If no at section is seen, the measurement distance is insufcient.
Move the installation locations of the auxiliary earthing rods (S[P]
and H[C]) to further locations.
58
Ground
Surface
Cautions and Tips for Measurement
Measured Value
Flat Section
Resistance Area of E Electrode
The distance between E-H(C) of the instrument can be up to 50 m by using the optional L9843-51 and L9843-52. In principle, it is possible to carry out measurement even if the distance is more than 50 m. However,
the measurement result cannot be guaranteed.
S(P)
Resistance
Area of H(C)
Electrode

Location to install auxiliary earthing rod

It is ideal to install the auxiliary earthing rod S(P) at the middle point
of the straight line between the earthing element E and auxiliary
earthing rod H(C).
However, if installation is not possible due to an obstacle etc., the measurement error can be reduced by installing the auxiliary
earthing rod S(P) within 29° from the straight line between the earthing element E and auxiliary earthing rod H(C), wherein the location should be outside the radius of 5 m from the earthing element E and auxiliary earthing rod H(C) as shown in the gure.
E
5 m
S(P) H(C)
29°
5 m
59
Cautions and Tips for Measurement

How to insert/pull out auxiliary earthing rod

How to Insert
The accessory auxiliary earthing rods are suitable for providing auxiliary earthing electrodes and are designed for thickness and hardness that allow insertion into a general ground by hand. Since it is thinner than previous models, it can be inserted into a small gap.
Wear gloves and insert it perpendicular to the ground surface. If the ground is too hard for inserting by hand, use a hammer to perpendicularly drive into the ground. Hammering the auxiliary earthing rod too hard may result in bending it. In case of not going
into the ground with gentle taps, use the optional 9050 Earth Nets
for measurement.
60
How to pull out
• Hold the loop part of the auxiliary earthing rod and pull it out while turning it.
• If it does not come out by hand, put a hard metal bar etc. (other than
auxiliary earthing rod) through the
loop part of the auxiliary earthing rod and pull the auxiliary earthing rod while turning it. If pulling the auxiliary earthing rod with another auxiliary earthing rod put through the loop, it will result in bending it.
• Do not apply force to the auxiliary earthing rod from the side as it may result in bending the auxiliary earthing rod.
Cautions and Tips for Measurement
61

Auto Power Save (Power-saving Function)

3.7 Auto Power Save
(Power-saving Function)
Approx. 10 min after the last operation or the last time (live wire warning LED) turns on or blinks, the auto power save becomes
active and the screen turns off.
How to recover from auto power save
Press the Power button turn the screen on.
How to disable auto power save
Turn the power on while pressing the 0
IMPORTANT
• The settings of an disabled auto power save will not be saved when the power is turned off.
• After use, always turn off the instrument. Even when the auto power save is enabled, the instrument consumes a small amount of the battery power.
ADJ button.
Ω
62

Wireless Communications Function (GENNECT Cross)

3.8
Wireless
Communications F
unction
(GENNECT Cross)
When the wireless communications function is enabled, you can review measurement data and create measurement reports on mobile devices. For more information about this functionality, see “GENNECT Cross
Ofcial Website” (application software, free of charge).
GENNECT Cross Ofcial Website
https://gennect.net/en/cross/index
Connect the Z3210 Wireless
1
Flashes:
Communicating
Lights up:
Wireless function on
Lights off:
Wireless function off
To toggle the wireless communications function on and off, press and hold the
Fn button and the COMP
button for at least 1 s.
Adapter (option) to the instrument. (p. 29)
Install the GENNECT Cross on
2
your mobile device.
Turn on the instrument.
3
Press and hold the Fn
4
button and the COMP button simultaneously for at least 1 s to enable the wireless communications function.
When the wireless communications function is on, display.
Launch the GENNECT Cross
5
and pair it with the instrument.
Select the measurement
6
function and start measurement.
will ash on the
63
Wireless Communications Function (GENNECT Cross)
The communication distance is approx. 10 m (line of sight).The distance over which data can be sent and received varies greatly depending on whether there are any obstructions between the
paired instruments (for example, walls, metal barriers, etc.) and on the distance between the instrument and the oor (or ground).
To ensure stable communication, verify adequate signal strength.
• Although the GENNECT Cross is provided free of charge, downloading or using the application software may incur Internet connection charges. Such charges are the sole responsibility of the user.
GENNECT Cross
• The
is not guaranteed to operate on all
mobile devices.
The Z3210 uses 2.4 GHz band wireless technology. It may not be possible for the device to establish a wireless connection when used in the vicinity of other devices that use the same frequency
band, for example Wi-Fi devices (IEEE 802.11.b/g/n).
When the app is launched for the rst time (before being paired with any instrument), the instrument settings screen will be
displayed.
• While the mobile device is displaying the GENNECT Cross's instrument settings screen, simply move it close to the instrument to automatically pair it with the instrument (the app can be paired
with up to 8 instruments).
Allow about 5 s to 30 s for the instrument to pair with the app after being turned on. If the instrument fails to pair within 1 min.,
relaunch GENNECT Cross and cycle the instrument’s power.
64
Activating the LCD All-on Display

3.9 Activating the LCD All-on Display

This section describes how to check for screen issues by enabling all indicators on the LCD display at once.
Turn the power on while pressing the Fn button.
1
Press any button.
2
Its normal measurement screen will appear.
65
Displaying the Serial Number

3.10 Displaying the Serial Number

Turn the power on while pressing the 3P/2P button.
1
The rst 4 digits and the last 5 digits of the serial number (9 digits)
appear alternately.
Press any button.
2
Its normal measurement screen will appear.
First 4 digits
(Example of a serial number: 200756789)
The 9-digit serial number indicates the year (rst two digits) and the month of manufacture (next two digits).
Last 5 digits
66
Specications
4
4.1 General Specications
Operating environment
(scope of application)
Operating temperature and humidity range
Storage temperature and humidity range
Dustproof and waterproof
Drop-proofness Drop height: 1 m, impact surface: concrete (with
Standards Safety
Power supply AA alkaline battery (LR6) ×4
Indoor, outdoor (excluding farmland*), pollution degree 3, altitude up to 2000 m (6562 ft.) *: According to the requirements regarding the
regulations for open-circuit voltage in EN 61557-5
−25°C to 40°C (−13°F to 104°F): 80% RH or less (non-condensing) 40°C to 45°C (104°F to 113°F): 60% RH or less (non-condensing) 45°C to 50°C (113°F to 122°F): 50% RH or less (non-condensing)
50°C to 55°C (122°F to 131°F): 40% RH or less (non-condensing) 55°C to 60°C (131°F to 140°F):
30% RH or less (non-condensing) 60°C to 65°C (140°F to 149°F): 25% RH or less (non-condensing)
−25°C to 65°C (−13°F to 149°F): 80% RH or less (non-condensing)
IP65/IP67 (EN 60529)
protector attached)
Instrument: EN 61010 Measuring circuit: EN 61010
EMC: EN 61326
Earth tester: EN 61557
Rated supply voltage: 1.5 V DC× 4
Maximum rated power: 3 VA
Available effective battery voltage: 4.5 V ±0.19 V to
6.8 V
67
General Specications
Number of measurements before replacing batteries
Dimensions Approx. 185W × 111H × 44D mm (7.28″W × 4.37″H
Mass Approx. 570 g (20.1 oz.) (including batteries and
Product warranty period
Accessories
Options
When using four AA alkaline batteries (LR6)
(reference value at 23°C)
• 500 measurements (3-pole method, auxiliary earthing resistance of 100 using the 20 intervals, without Z3210 installed)
400 measurements (3-pole method, auxiliary earthing resistance of 100 using the 20 intervals, with Z3210 installed, using wireless communications)
× 1.73″D)
(including protector, excluding cover of the measurement terminals)
protector, excluding other accessories)
3 years
See “Verifying Package Contents” (p.5)
See “Options (sold separately)” (p.6)
range, measure on at 10-sec.
Ω
range, measure on at 10-sec.
Ω
, measuring 10 Ω
Ω
, measuring 10 Ω
Ω
68
Input, Output, and Measurement Specications
4.2 Input, Output, and Measurement Specications
Basic specications / accuracy specications
Measured parameters
A/D conversion method
Input/output terminals
Maximum rated terminal-to­ground voltage
Nominal operating ranges
• Earth resistance measurement
• Earth potential measurement
method (peak voltage measurement: sequential
∆Σ
comparison method)
H terminal: Terminal that outputs the measurement
S terminal: Terminal that detects the voltage being
E terminal: Terminal connected to the earthing electrode
100 V AC, DC (measurement category IV) 150 V AC, DC (measurement category III) 300 V AC, DC (measurement category II)
Anticipated transient overvoltage: 2500 V
• Operating temperature and humidity range:
See “Operating temperature and humidity range” in “General Specications” (p. 67).
Position: Level ±90°
• Power supply voltage: Available effective battery voltage
• Earth potential:
See “Allowable earth potential” in “Basic specications / accuracy specications”.
• Resistance of auxiliary earthing electrode: See
electrode” in “Basic specications / accuracy specications”
External magnetic eld: 400 A/m or less, DC and frequency 50 Hz/60 Hz
voltage or that detects the voltage being measured (when using the 2-pole method)
measured (when not using the 2-pole method)
being measured to receive the measurement current output from the H terminal
“Allowable resistance of auxiliary earthing
69
Input, Output, and Measurement Specications
Accuracy guarantee conditions
Accuracy guarantee period: 1 year Accuracy guarantee period after adjustment made by Hioki: 1 year Accuracy guarantee temperature and humidity range:
23°C ±5°C (73°F ±9°F), 80% RH or less
70
Input, Output, and Measurement Specications
Measurement of earth resistance
(RE: Earth resistance of object under measurement,
R
: Earth resistance of H electrode,
H
R
: Earth resistance of S electrode)
S
Conditions: auxiliary earthing electrode resistance 100 Ω ±5%,
Operation system
Measurement system
Output voltage 30 V rms or less and 42.4 V peak or less
Measurement current
Measurement current waveform
Measuring frequency
Measurement time
earth potential 0 V
Voltage application, measurement of voltage and current (Effective resistance is measured by synchronous detection.)
2-pole method / 3-pole method, switchable
3-pole method: 25 mA rms or less
2-pole method: 4 mA rms or less
Sine wave (Distortion factor of 5% or less)
128 Hz ±2 Hz
3-pole method: Within 6 s
2-pole method: Within 3 s
71
Input, Output, and Measurement Specications
Range
conguration
(auto-range)
Phase difference tolerance
Effect of positioning
E
)
(
1
Effect of supply
E
voltage (
)
2
Effect of temperature
*7
E
)
(
3
Display range
*1
Applied
conditions
(RE and RH)
Display range
Resolution
Accuracy
*3
3-pole 0.01
2-pole
*4
(specic uncertainty A)
Allowable
resistance of
auxiliary earthing
electrode
Operation
uncertainty
*5
±30% rdg (applied to 3-pole method)
R
E
R
H
R
H
R
S
*2
20
Ω
20 Ω
or less
5 kΩ
or less
0.00 Ω to
20.00
Ω
Ω
±1.5% rdg
±8 dgt
5 k
Ω
5 k
Ω
200 Ω2000
200 Ω or less
50 kΩ
or less
0.0 Ω to
200.0
0.1
Ω
1
Ω
±1.5% rdg ±4 dgt
Guaranteed range
of operation
uncertainty
*6
5.00 Ω to 2000
±3.5°
Not applicable due to digital type
Accuracy × 0.5 and within accuracy specications
Accuracy × 1.0 (−10°C to 50°C)
Accuracy × 2.0 (−25°C to −10°C, 50°C to 65°C)
Ω
50 k
50 k
Ω
200
Ω to
2 k
Ω or less
50 kΩ
or less
0 Ω to
2000
1
1
Ω
Ω
Ω
Ω
Ω
Ω
72
Input, Output, and Measurement Specications
Effect of earth
E
potential (
)
4
Allowable earth potential
Effect of the resistance of auxiliary earthing electrode (
*5
E
)
5
Effect of system frequency (
E
)
7
Effect of system
E
voltage (
)
8
Effect of external magnetic eld
Earth resistance of auxiliary earthing electrode auto-
check
(auto-check after start of measurement)
Auto-check for earth potential
*8
(auto-check after start of measurement)
Overload protection
16 2/3 Hz,
50 Hz, 60 Hz
DC
400 Hz
0 V to 3 V Accuracy × 1.0
3 V to 25 V Accuracy × 2.0
0 V to 3 V Accuracy × 1.0
3 V to 10 V Accuracy × 2.0
0 V to 3 V Accuracy × 1.0
3 V to 5 V Accuracy × 2.0
25 V rms or 35.4 V peak
Either RH or RS electrode Variable value
or less Accuracy × 1.0
10 k
Ω
Above 10 k
to 50 k
Ω
Ω
Accuracy × 2.0
Not applicable
Not applicable
Accuracy × 0.5
Display range
Maximum display
value
Resolution 10 Ω0.1 kΩ1 k
1000 Ω10 kΩ100 kΩ200 k
1000 Ω10.0 kΩ100 kΩ200 k
Ω
Accuracy Not specied
Display range 42 V peak
Maximum display
value
42.0 V peak
Accuracy ±2.3% rdg ±8 dgt
360 V AC (for 1 min., between each terminal)
360 V DC (for 1 min., between each terminal)
10 k
Ω
Ω
Ω
73
Input, Output, and Measurement Specications
Z
cosθ
Automatic measurement termination
Effect of earth capacity (2-pole method)
Zero adjustment allowable range
Negative resistance value processing
Automatic stop after measured values are nalized
Up to 10 nF Within accuracy specications
Above 10 nF
to 500 nF
Above 500 nF Not specied
Parallel capacitance is canceled from impedance and phase.
Calculated as:
(correction applied only when θ is negative)
R=
(θ: Difference between the measurement current phase
and voltage signal phase; display reads “OVERFLOW”
when θ is 90°.)
3
Displayed as absolute value
Accuracy × 2.0
or less
Ω
*1: Automatically select the minimum display range that can meet both R
*2: 3-pole method only. *3: When using the 3-pole method, the minimum display value is 0.01
*4: Applied after zero adjustment, ±0.3 Ω is added before zero adjustment
*5: Does not apply when using the 2-pole method, in which case the
*6: It shall be the range in which the operation uncertainty within ±30% is
74
values.
and R
H
range) or 0.1 Ω (200 Ω range), with the following exceptions:
(20
Ω
• When R
• When RH = 100 Ω ±20 Ω and RS = 100 Ω ±20
• When RH = 500 Ω ±20 Ω and RS = 500 Ω ±20
(when using the L9841).
auxiliary earthing electrode’s allowable resistance is used as the upper limit value.
guaranteed; applied to 3-pole method.
= 0 Ω ±20 Ω and RS = 0 Ω ±20
H
Ω
Ω Ω
E
Ω
Input, Output, and Measurement Specications
*7: Applied in the range excluding 18°C to 28°C. *8: Function to check whether or not the peak of earth potential is within
the allowable range; no AC coupling; display the peak value (peak on +
side).
75
Input, Output, and Measurement Specications
Earth potential measurement
DC/AC auto­detection range
Measurement method
Terminal for voltage measurement
Overload protection
Display refresh interval
Input resistance 4 M
Response time Within 2 s (when input voltage is changed from 0 V to
Effect of temperature
Range conguration
0.3 V DC ±0.2 V or higher is detected as DC
Manual switching function provided
Pulsating ow with superimposed AC component that periodically zero-crosses is detected as AC.
Average value rectication, RMS value indication
3-pole method:
Between S terminal (+) and E terminal (−)
2-pole method:
Between H terminal (+) and E terminal (−)
360 V AC (for 1 min., between each terminal)
360 V DC (for 1 min., between each terminal)
Within 1 s
or higher (DC / 50 Hz / 60 Hz)
Ω
30 V)
Accuracy × 1.0 ( −10°C to 50°C, applied in the range
Accuracy × 2.0 (−25°C to −10°C, 50°C to 65°C)
Display range 30 V rms
Maximum display value 30.0 V rms
Resolution 0.1 V
Accuracy
excluding 18°C to 28°C)
DC ±1.3% rdg ±4 dgt
50 Hz/60 Hz ±2.3% rdg ±8 dgt
76

Maintenance and Service

5

5.1 Repair, Inspection, and Cleaning

WARNING
Do not attempt to modify, disassemble, or try to
repair the instrument. Doing so may cause a re,
electric shock, or injury.
Calibrations
The calibration frequency varies depending on the status of the instrument or installation environment. We recommend that the calibration frequency is determined in accordance with the status of the instrument or installation environment and that you request that calibration be performed periodically.
Cleaning
IMPORTANT
Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
• If the instrument becomes dirty, slightly moisten a soft cloth moistened with water or a neutral detergent and wipe the instrument clean.
• When rinsing with water, make sure that the battery cover has been attached. Make sure that the protector has been removed and rinse with room-temperature water. Washing with hot water may compromise the instrument’s waterproofness.
• After cleaning, thoroughly wipe off water of the instrument and protector. Completely dry them before putting the protector on.
77
Repair, Inspection, and Cleaning
• Do not put water while the battery cover has been taken off. It may cause failure due to water intrusion.
• When replacing the batteries after cleaning, the replacement should be carried out after the instrument has been completely dried.
• The instrument is waterproof to a water depth of 1 m for 30 min. Do not soak and wash the instrument. Do not soak the instrument into hot water.
• Do not wash the instrument with a washing machine
• Do not use any electronic appliance such as a dryer or microwave for drying the instrument.
• Wipe the LCD gently with a soft, dry cloth.
If the instrument is not to be used for an extended period of time
IMPORTANT
To avoid corrosion and damage to the instrument from battery leakage, remove the batteries from the instrument if it is to be stored for a long time.
Disposal
Dispose of the instrument in accordance with local regulations.
Protector
A protector is available as a service part. Contact your authorized Hioki distributor or reseller.
78
Troubleshooting

5.2 Troubleshooting

If damage is suspected, read the “Before sending the instrument for repair” (p. 80) section before contacting your authorized Hioki distributor or reseller.
Be sure to follow these precautions when transporting the instrument:
• Remove batteries, accessories, and optional equipment from the instrument in order to avoid damage. Additionally, use the original packing materials, and be sure to double box the instrument. Accidental damage suffered in transit is not covered by the warranty.
• Attach a description of the issue when sending out your instrument for repair.
79
Troubleshooting
Before sending the instrument for repair
If it is not operating correctly, check the following items.
Symptom Cause, check, countermeasure
The power does not turn on. There are no batteries inside.
The instrument is
unintentionally turned off
during a measurement.
The batteries have been incorrectly installed.
→ Refer to how to install the batteries.
Have the batteries been depleted?
→ Replace the batteries with new ones
If the power still will not turn on, the instrument has a failure and needs to be
repaired.(“Inserting/Replacing Batteries”
[p. 26])
The batteries to be used may have been left for a long time.
→ Battery voltages may recover as time
proceeds, which seem to be high enough to work. However, such batteries, which provide only a little energy, cannot work. Replace the batteries with new ones.
The batteries have high internal resistance.
→ Such high-internal-resistance batteries,
which provide only a little energy even if they are new, cannot work. Use batteries produced by other manufactures.
80
Symptom Cause, check, countermeasure
The resistance of auxiliary earthing electrode does not come down (precision measurement).
appears and
measurement is not allowed (cannot measure the earth resistance due to high earth potential).
Troubleshooting
Bad connection of measurement cable, the ground is dry.
→ Make sure that the measurement cables
are connected to the earthing electrode/ auxiliary earthing rod.
→ Short-circuit the tips of the measurement
cables and then carry out the measurement. If the measured value is approx. 0 it is caused by a high earth resistance of the earthing electrode. Insert the auxiliary earthing rod deeper. Alternatively, pour water over the auxiliary earthing rod. If the resistance still will not come down, move the auxiliary earthing electrode to another location. The measurement cables (yellow) and (red) should be positioned approx. 10 cm away from each other.
The earth potential exceeds the value that can be accepted by the instrument. A large
leakage current may be owing into the
earthing electrode from equipment that has been connected to the earthing electrode. Or, the earth resistance of earthing electrode may be high and a large earth potential may have occurred by a small leakage current. Remove the equipment that has been
connected to the earthing electrode and then carry out the measurement. (“When
NOISE appears” [p. 45])
,
Ω
81
Troubleshooting
Symptom Cause, check, countermeasure
Trying to carry out measurement on a newly built residential building, but
the simplied measurement
(2-pole method) is not allowed.
The measured value becomes approx. 0 measuring with use of an existing auxiliary earthing electrode for measurement.
Small noise occurs inside the instrument during measurement.
The measured value is high. It has been set to 2-pole method.
Measurement is not allowed when trying to measure with an earth net laid on asphalt.
Auxiliary earthing rod has been bent.
Ω
Measurement is not allowed if the electricity has not been distributed from the power company.
The earthing electrode and the auxiliary
when
electrode for measurement may have been connected through concrete.
→ Drive an auxiliary earthing rod into the
ground and then carry out measurement instead of using an auxiliary electrode for measurement.
It is the noise from the oscillator circuit inside the instrument. It is not a failure.
→ Press the 3P/2P button to select 3-pole
method.
Since asphalt is insulator, measurement is not possible with an earth net.
It is recommended to purchase the L9840
Auxiliary Earthing Rod. (Commercially available pegs have sharp tips, which may damage the carrying case.)
82

5.3 Error Display

Error Display
Error
display
Err 0ΩADJ
Err1
Err2
Err4
Err5
Err8
APS →P.oFF
bAtt
→P.oFF
Description Solution
Zero adjustment is outside the allowable range, zero adjustment has been executed while it is not on HOLD.
Abnormality in rmware
Adjustment data damaged
Setting data damaged
Abnormality in measurement circuit
Z3210 communications error (connection failure; Z3210 or
hardware failure)
Instrument powered off by
APS
Instrument powered off due to supply voltage drop
Execute zero adjustment again in accordance with the zero adjustment procedure. (
Instrument failure. Contact your authorized Hioki distributor or reseller to organize repair.
Take the following actions:
Reinstall the Z3210.
Install a different Z3210.
(p. 29)
If the error persists, you are experiencing a instrument failure. Contact your authorized Hioki distributor or reseller to organize repair.
Cycle the power.
p. 40, p. 54
)
83
Error Display
84

Appendix

6

6.1 Earth Resistance

The resistance between earthing electrode and ground is usually called earth resistance. To be exact, it is the sum of the resistance of earthing conductor, the contact resistance between earthing conductor and ground, and the resistance of ground. Earth resistance is different from commonly known resistors and has the following special characteristics.
Polarizing action Since the ground has characteristics just like electrolyte, it
has the polarizing action, and if DC current ows through, an
electromotive force occurs in the opposite direction to the current, interfering with correct measurement. Therefore, rectangular wave or sine wave of several tens Hz to 1 kHz is usually used to measure the earth resistance.
Special conguration Earth resistance is a resistance between earthing electrode and ground. It is not possible to take it out from the ground and measure it. Since the resistivity of ground is relatively high, a voltage drop occurs near the electrode through which the current to be
measured ows. Therefore, each electrode (E electrode, S(P) electrode, H(C) electrode) needs to be away from each other to
approx. 10 m to accurately measure the resistance of earthing electrode.
Presence of disturbance factors There are disturbance factors such as effects from earth potential and auxiliary earthing electrode in the measurement of earth resistance. The earth potential caused by a leakage current from a device
85
Earth Resistance
that has been connected to the earthing electrode affects the measured value as it is superimposed over the signal to be detected by the earth tester. In addition, if the earth resistance of auxiliary earthing electrode is high, the measurement current is reduced, making it susceptible to noise such as earth potential. The instrument employs a system that is less susceptible to these disturbances and allows accurate measurement under adverse conditions.
86
Measurement Principle

6.2 Measurement Principle

While applying a voltage of AC power supply between H(C) electrode and E electrode, the owing AC current I is measured with an ammeter. In addition, the voltage V between S(P) electrode and E electrode that is caused by the ow of the current I is measured
by an AC voltmeter. The earth resistance RX of E electrode is derived from the measured current I and voltage V. It is not possible to accurately measure the
voltages between H(C) and E electrodes and between H(C) and S(P) electrodes.
AC Power
supply
I
I
S(P) electrode H(C) electrode
E electrode
=
R
V / I
X
Ammeter
l
Voltmeter
=
×
V
R
X
87
Measurement Principle
88
Loading...