Hioki FT3405, FT3406 Instruction Manual

FT3405 FT3406
TACHO HiTESTER
Instruction Manual

Sept. 2016 Revised edition 4 FT3405A981-04 16-09H
EN
i
Contents
Contents
Introduction ................................................................1
Verifying Package Contents ......................................1
Safety Information .....................................................2
Operating Precautions ...............................................5
Chapter 1 Overview 9
1.1 Product Overview........................................9
1.2 Features......................................................9
1.3 Names and Functions of Part....................10
Chapter 2 Preparing for Measurement 17
Installing or Replacing the Batteries.....17
Connecting the AC Adapter (Option)
(FT3406 only)........................................19
Connecting the Contact Adapter
(Option).................................................20
Connecting the Instrument to a Tripod .20 Connecting the Analog Output Cable
(FT3406 only)........................................21
Connecting the Pulse Output Cable
(FT3406 only)........................................23
Chapter 3 Measurement Procedures 25
Pre-Operation Inspection......................25
3.1 Turning the Power On and Off..................26
3.2 Selecting Measurement Functions............27
3.3 Measurement ............................................28
FT3405A981-04
Contents
ii
3.4 Precautions in Measurement of High-Rotation
Objects......................................................29
Chapter 4 Instrument Functionality 31
Measured Value Hold...........................31
Measured Value Overrun Display.........31
Averaging Function.............................. 31
Maximum/Minimum Value Display....... 32
Rotation Detection Display Function.... 32
Configuring the Buzzer......................... 33
Display Backlight..................................33
Auto Power Save (APS) Function........ 33
Resetting the System...........................34
Battery Warning
(Remaining Battery Life Detection)....... 34
Chapter 5 Specifications 35
 Function specifications ...................... 35
 Accuracy (Not applicable to count
measurement) ................................... 39
 General specifications ....................... 40
Chapter 6 Maintenance and Service 41
6.1 Cleaning....................................................41
6.2 Consumables............................................ 41
6.3 Discarding the Instrument.........................41
6.4 Troubleshooting........................................42
1

Introduction

Confirm that these contents are provided.
L9094 Output Cord
(FT3406 only)
FT3405 or FT3406
LR6alkaline battery×2
Instruction manual
(This document)
9211 Reflective Tape
C0202 Carrying
Case
Thank you for purchasing the HIOKI Model FT3405, FT3406 TACHO HiTESTER. To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference.
The “instrument” in this manual means FT3405 and FT3406. Refer to " Options" (p.14) for details.

Verifying Package Contents

• When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Use the original packing materials when transporting the instrument, if possible.
ment, be sure to remove the Contact Adapter, Contact Tip, Peripheral Ring, Output Cord and AC Adapter before shipping.
1 2
3 4
5
6
To avoid damage to the instru-
7 8 9
10
索引
2

Safety Information

This instrument is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instru­ment. However, using the instrument in a way not described in this manual may negate the provided safety features. Be certain that you understand the instructions and precau­tions in the manual before use. We disclaim any responsi­bility for accidents or injuries not resulting directly from instrument defects.
Safety Symbols
This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions.
In the manual, the symbol indicates particularly important information that the user should read be­fore using the instrument. The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with the symbol) before us­ing the relevant function.
Indicates DC (Direct Current). Indicates AC (Alternating Current).
Symbols for Various Standards
This symbol indicates that the product conforms to regulations set out by the EC Directive.
Indicates the Waste Electrical and Electronic Equip­ment Directive (WEEE Directive) in EU member states.
3
The following symbols in this manual indicate the relative impor­tance of cautions and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in se­rious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage
the instrument.
to Indicates advisory items related to perfor-
mance or correct operation of the
ment.
Other Symbols
( p. )
The screen of this instrument displays characters in the follow­ing manner.
Indicates the location of reference information. Indicates that descriptive information is pro-
*
vided below.
instru-
1 2
3 4
5
6 7 8 9
10
索引
4
Accuracy
We define measurement tolerances in terms of f.s. (ful l scale), rdg. (reading) and dgt. (digit) values, with the following meanings:
f.s. (maximum display value or scale length)
rdg. (reading or displayed value)
dgt. (resolution)
The maximum displayable value or scale length. This is usually the name of the currently selected range.
The value currently being measured and indicated on the measuring instrument.
The smallest displayable unit on a digital measuring instrument, i.e., the input val­ue that causes the digital display to show a “1” as the least-significant digit.
5

Operating Precautions

Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Before use
Before using the instrument the first time, verify that it operates normally to ensure that no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Before using the instrument, make sure that the insulation on the cable is undamaged and that no bare conductors are improperly exposed. Using the instrument in such conditions could cause an electric shock, so contact your dealer or Hioki representative for repair .
Setting up the instrument
• Although this instrument is dust resistant, it is not completely dust- or waterproof. To preven t possible damage, avoid using in dusty or wet environments.
• The protection rating for the enclosure of this device (based on EN60529) is *IP50.
*
IP50:
This indicates the degree of protection provided by the enclo­sure of the device against use in hazardous locations, entry of solid foreign objects, and the ingress of water.
5: Protected against access to hazardous parts with wire mea-
suring 1.0 mm in diameter. Dust-proof type (The penetration of dust cannot be prevented completely, but quantities of dust that may hinder the stated operation of equipment or safety cannot penetrate the enclosure.
0: The equipment inside the enclosure is not protected against
the harmful effects of water.
1 2
3 4
5
6 7 8 9
10
索引
6
Instrument Installation
To avoid damage to the instruments and potentially life­threatening hazards, observe the following precautions.
• Use only the specified/supplied Model Z1004 AC Adapter. AC adapter input voltage range is 100 to 240 VAC (with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage out­side of this range.
• Before turning the instrument on, make sure the supply voltage matches that indicated on the AC adapter. Con­nection to an improper supply voltage may damage the instrument or AC adapter and present an electrical haz­ard.
• If the protective functions of the instrument are damaged, either remove it from service or mark it clearly so that others do not use it inadvertently.
• To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
• To avoid breaking the cables, do not bend or pull them.
• For safety reasons, when taking measurements, only use the L9094 Output Cord provided with the instrument (FT3406 only).
7
Storage temperature and humidity:
-10°C to 50°C (14 to 122°F), 80%RH or less (non-condensating) Accuracy guarantee for temperature and humidity: 23°C±5°C (73°F±9°F), 80%RH or less (non-condensating)
Avoid the following locations that could cause an accident or damage to the instrument.
Exposed to direct sunlight Exposed to high temperature
Exposed to water, oil, other chemi­cals, or solvents Exposed to high humidity or con­densation
Subject to vibra­tion
In the presence of cor­rosive or explosive gases
Exposed to strong electromagnetic fields Near electromagnetic radiators
Near induction heating systems (e.g., high-frequency induction heating sys­tems and IH cooking utensils)
1 2
3 4
5
6 7
Measurement Precautions
Rotational speed measurement
Always wear protective eyewear as debris from the rotating object may cause injury.
• Depending on the material of the object to be measured or method of touching the contact tip, the measurement error may occur.
• Avoid scratching or dirtying the two lenses inside the detector window.
8 9
10
索引
8
Using the Contact Adapter
• Always use the screw to tighten the contact adapter to the main body. If it becomes loose, the instrument may vibrate or be dislocated and become a hazard.
• Be careful when measuring with the contact adapter because rotation and vibration of the instrument can pro­duce erratic results with either high or low rot ation spee ds. Hold the main body firmly against the rotator. Do not place the instrument on a tripod when making measurements.
• When using the contact adapter for measuring, the instrument is subject to vibration, so measurements should be made only below 19,999 r/min or 333 r/s.
• When using the Contact Adapter, do not mount the instru­ment on a tripod.
Handling contact tips
• Verify that the contact tip has been inserted sufficiently into the adapter axis before beginning measurement. Fail­ure to insert the contact tip sufficiently into the adapter axis may cause it to separate from the axis, contact the rotating object, and fly off in the vicinity of the object.
• If the contact tip is able to come into contact with the rotating object, exercise caution when winding it onto the object.
• Do not allow rubber contact tips to come into contact with hot measurement targets. Doing so may cause the con­tact tip to melt.
Analog Output Terminals and Pulse Output Terminals
To avoid electrical hazards and damage to the instrument, do not apply voltage exceeding the rated maximum to the Analog Output Terminals and Pulse Output Terminals.
9
Chapter 1 Overview
Overview Chapter 1

1.1 Product Overview

This portable, contactless tachometer uses reflected visible­spectrum light to measure the speed of a rotating object to which reflective tape has been affixed. It can also be used as a con­tact-type tachometer by attaching the Z5003 Contact Adapter.
1 2
3

1.2 Features

Contactless measurement for long detection range
The instrument’s ability to detect speed from up to 50 cm distance al­lows its use in situations where it would be hazardous to approach the object being measured.
Easy-grip design
Easy-to-grip design and light weight make the instrument easy to han ­dle.
LCD with a wide field of vie
The FT3405/3406 features a transmissive FSTN liquid crystal d isplay (LCD), ensuring its readout is easy to read.
Maximum/minimum value hold
Maximum and minimum values are updated continuously so you can verify variations in rotational speed.
Dust-proof design
The FT3405/3406’s enclosure delivers IP 50-grade protection, allow­ing it to be used with confidence and peace of mind in environments characterized by debris (dust, sand, grit, etc.) thrown off by the rotat­ing object.
Drop Proof
The FT3405/3406 can withstand being dropped from a height of 1 m, making it less likely the unit will be damaged.
Analog Output/ Pulse Output (FT3406 only)
The FT3406 can be used in trend management applications by con­necting it to a recorder.
4
5
6 7 8 9
10
索引
Chapter 1 Overview
10
Front
Base
Rear
Side
AC adapter connector
Analog/pulse output port
Model FT3405
Battery Cover
Tripod screw hole
Grip
Model FT3406
Projection surface
Model FT3406
(shown with cap)
Red light projection/ detection window
Z5003 Con­tact Adapter Hole for fix­ing screw

1.3 Names and Functions of Part

11
Chapter 1 Overview
3
1
2
5
6
4
1 2
3 4
POWER POWER key
1
MODE MODE key
2
AVG AVERAGE key
3
CLR CLEAR key
4
MAX/MIN MAX/MIN key
5
HOLD HOLD key
6
Turns the instrument on and off.
Switches the measurement mode (units).
Switches the sampling time (gate time). Disabled during count measurement.
Clears the current, maximum, minimum, and count values.
Displays the maximum and minimum values. Disabled during count measurement.
Starts and stops updating of measured values. When reflected light is detected, a buzzer sounds and an LED lamp built into the key flashes.
5
6 7 8 9
10
索引
Chapter 1 Overview
12
[FT3405 display] [FT3406 display]
(FT3406 only)
LCD Display
Display Function Notes
Lights up when the buzzer is enabled.
Lights up when the Auto Pow­er Save function is enabled.
Displays remaining battery life.
Turns on when the AC adapt­er is connected.
Lights up when the hold func­tion is active.
Lights up when the Contact Adapter is connected.
Lights up during circumferen­tial speed measurement.
Displays the maximum value obtained since the last time the instrument was reset.
Displays the minimum value obtained since the instrument was last reset.
Lights up when the averaging function is enabled.
Displays values.
At rotation detection, key operation
Turns off when the AC adapter is connected.
Four stages; measurement is stopped when only the outline of the battery re­mains. Turns off when the AC adapter is connected.
Turns off when no power is being supplied from the AC adapter.
Indicates that the range has been exceeded.
Flashes when the range is exceeded.
13
Chapter 1 Overview
(FT3406 only)
71
186
38
Unit: mm
Indicates the unit of measure­ment for rpm measurement, circumferential speed mea­surement, and period mea­surement.
Indicates the unit of measure­ment for count measurement.
Selects pulse output or ana­log output. When analog out­put is selected, you can select from ×10, ×1, ×0.1, and ×0.01.
Five options:
r/s, r/min, m/s, m/min, ms
Five options:
Pout, Aout×10, Aout×1, Aout×0.1, Aout×0.01
1 2
3

External Dimensions

4
5
6 7 8 9
10
索引
Chapter 1 Overview
14

Options

The following options are available for the FT3405, FT3406 Tacho Hitester. Ask your dealer or Hioki representative when ordering.
Model, product,
and diagram
Z5003 Contact Adapter
9032 Metal Contact Tip
9033 Rubber Contact Tip
9212 Peripheral Ring
Z1004 AC Adapter
Notes
Dimensions: Approx. 41W×107H×44D mm (1.61"W×4.21"H×1.73"D)(excluding screw) Mass: Approx. 46 g (1.6 oz) *The Z5003 Contact Adapter includes
9032×1, 9033×2, and 9212×1.
Attaches to the Z5003 Contact Adapter.
Attaches to the Z5003 Contact Adapter.
Attaches to the Z5003 Contact Adapter. Outer circumference: 10 cm
For model FT3406 Rated supply voltage:100 V to 240 V Output voltage: 5 V Output current: 1.2 A Mass: Approx. 69 g (2.4 oz) Plug not included. Dimensions: Approx. 45W×58H×34D mm
Operating temperature and humidity : 0°C to 40°C (32 to 104°F), 20 to 80%RH Storage temperature and humidity : -20°C to 80°C (-4 to 176°F), 10 to 95%RH
(1.77"W×2.28"H×1.34"D)
15
Chapter 1 Overview
UL type VDE type BS type SAA type
Press firmly

Using Z1004 AC Adapter Attachments

The Z1004 AC adapter comes with four plugs. Select and attach the plug that fits the type of outlet used in your area.
Attaching the plug:
Slide the plug into position as shown in the diagram and insert until you hear it click into place.
Removing the plug:
Press down firmly on the tab and slide the plug in the direction indicated by the arrow to remove it.
1 2
3 4
5
6 7 8 9
10
索引
Chapter 1 Overview
16
Instruction Manual
Reflective Tape
Peripheral Ring
Contact Tip x3
Tacho HiTester
Output cord
Contact Adapter
Strap
Inner cover
Fixable with the hook and
loop fastener
Accesory pocket
Rubber bands
Divider
Store the instrument as shown below.

Carrying Case inside structure

17
Chapter 2 Preparing for Measurement
Preparing for Measurement
Chapter 2
1

Installing or Replacing the Batteries

Install two alkaline batteries (LR6) before using the instrument. Verify that sufficient battery life remains before starting measure­ment. If the instrument indicates low battery life, replace the bat­teries.
• To avoid electric shock, turn off the power switch and dis­connect the AC adapter and output cord before replacing the batteries.
• Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result.
• Battery may explode if mistreated. Do not short-circuit, recharge, disassemble or dispose of in fire.
• Handle and dispose of batteries in accordance with local regulations.
• The “ ” indicator flashes when the remaining battery capacity is low. In this case, the instrument's reliability is not guaranteed. Replace the battery immediately.
• To avoid corrosion and damage to this instrument from battery leakage, remove the batteries from the instrument if it is to be stored for a long time (several months or more).
2
3 4
5
6 7 8 9
10
索引
Chapter 2 Preparing for Measurement
18
Turn off the instrument and disconnect the output cord and
1.
AC adapter. Remove the screws that secure the battery cover in place
2.
on the bottom of the instrument (use a screwdriver). Remove the battery cover.
3.
If replacing the batteries, remove the old batteries.
4.
Insert two new batteries (LR6), taking care to insert them in
5.
the proper orientation. Attach the battery cover and tighten the screws.
6.
19
Chapter 2 Preparing for Measurement

Connecting the AC Adapter (Option) (FT3406 only)

1
Use only the specified Model Z1004 AC Adapter. AC adapter input voltage range is 100 to 240 VAC (with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage out­side of this range.
• AC adapter output may decrease due to power supply noise, a momentary loss of power, or other factors. To keep the instrument operation from stopping due to these effects when acquir­ing data for an extended period, it is recom­mended to use alkaline batteries in the device, even when operating on AC adapter power.
• Make sure the power is turned off before con­necting or disconnecting the AC adapter.
• The AC adapter has priority when connected. Consequently, power is not drawn from batteries when the instrument is operating using power supplied from the AC adapter.
• The indicator is not displayed when no power is being supplied to the AC adapter or when there is a power outage (at this time, the instrument will operate using battery power).
Turn off the instrument.
1.
Remove the cap on the bottom of the grip
2.
and insert the AC adapter jack into the connector.
Connect the AC adapter plug to the
3.
power supply. Turn on the instrument.
4.
5.
The indicator will be shown on the instrument’s display while power is being supplied.
2
3 4
5
6 7 8 9
10
索引
Chapter 2 Preparing for Measurement
20
1
2
*Screw size: 1/4 inch

Connecting the Contact Adapter (Option)

To perform contact measurement, attach the Z5003 Contact Adapter.
Connect the Contact Adapter as shown
1.
in the diagram to the right. Tighten the fixing screw on the bottom of
2.
the instrument. Attach the 9032 Metal Contact Tip, 9033
3.
Rubber Contact Tip, and 9212 Periph­eral Ring as needed for your application.

Connecting the Instrument to a Tripod

The instrument can be mounted on a commercially available tri­pod to perform measurements from an installed (fixed) position. Mount the instrument on the tripod using the tripod screw hole on the bottom of the device.
• Do not attempt to rotate the instrument to tighten the screw when mounting it on a tripod. Instead, rotate the screw.
• Do not lift or carry the instrument while it is mounted on a tripod.
21
Chapter 2 Preparing for Measurement
(Unit: mV)
Output is calculated from the display value M as follows
Measurement mode
Setting r/min r/s ms counts m/min m/s
Aout×10 M×10/60 M×10
(1,000/M) ×10
r/min output
M×10/ 60×10
M×10 ×10
Aout×1M×1/60 M×1
(1,000/M) ×1
M×1/ 60×10
M×1 ×10
Aout×0.1 M×0.1
M×0.1 ×60
(1,000/M) ×0.1×60
M×0.1/ 60×10
M×0.1
×
60
×10
Aout×0.01 M×0.01
M×0.01 ×60
(1,000/M) ×0.01×60
M×0.01/ 60×10
M×0.01 ×60×10
Recorder, etc.
L9094 Output Cord
Output port

Connecting the Analog Output Cable (FT3406 only)

Use only the specified L9094 Output Cord. Using a non-specified cable may result in incorrect mea­surements due to poor connection or other rea­sons.
Electrical specifications
1.
Output resistance: 1 kΩ; full scale: 1 V; resolution: 1 mV Instrument output setting (analog output/pulse output)
2.
Turn off the instrument. Then press the [POWER] key while pressing and holding the [MODE] key to turn it back on. Hold
[MODE] key for no more than 3 seconds and select the
the desired "Aout’__" output port setting from Pout, Aout×10, Aout×1, Aout×0.1, and Aout×0.01. Refer to the following table for more information.
1 2
3 4
5
6 7 8
After 3 seconds, the setting will take effect, and the instru-
3.
ment will switch to the selected measurement mode. Connection method
4.
1. Remote the cap from the bottom of the instrument.
2. Connect the L9094 Output Cord to the Output port.
9
10
索引
Chapter 2 Preparing for Measurement
22
[For
averaging off
]
[For
averaging on
]
Example output waveform
5.
• At OVER, 1 V is output.
• During hold operation, analog output generates the current value rather than the hold value.
23
Chapter 2 Preparing for Measurement
Frequency counter, etc.
Output port
L9094 Output Cord

Connecting the Pulse Output Cable (FT3406 only)

Function
1.
Outputs pulses detected by the receiver. Electrical specifications
2.
Output resistance: 1 k Compatible jack diameter: 3.5 mm Output level: 0 to 3.3 V (Active low, low level fixed at 300 μs)
Instrument output setting (analog output/pulse output)
3.
Turn off the instrument. Then press the pressing and holding the [MODE] key to turn it back on. Hold the [MODE] key for no more than 3 seconds and select "Pout" from Pout, Aout×10, Aout×1, Aout×0.1, and Aout×0.01 as the output port setting.
After 3 seconds, the setting will take effect, and the instru-
4.
ment will switch to the selected measurement mode. Connection method
5.
1. Remote the cap from the bottom of the instrument.
2. Connect the L9094 Output Cord to the Output port.
[POWER] key while
1 2
3 4
5
6 7 8 9
10
索引
Chapter 2 Preparing for Measurement
24
Example output waveform
6.
• Output is generated under all measurement con­ditions, regardless of the set function or unit.
• To avoid damage to the instrument, do not apply voltage to output terminals. To avo id damage to the instrument, do not apply voltage to output ter­minals
25
Chapter 3 Measurement Procedures
Do not use the instrument if even one point of damage is found, as it can cause electric shock. Replace the damaged items.
Yes
• Is the insulation of the output cord to be used damaged, or is bare metal exposed?
• Is damage to the instrument evi­dent?
• Is damage to the AC adapter evi­dent? (FT3406 only
<When using batteries>
When the power is switched on, does the screen come on?
The batteries may have run out. Replace the batteries with new ones and then check again.
The power cord may be disconnected or the AC adapter or instrument in­ternal may be malfunc­tioning. Request repairs.
<When using AC adapter FT3406 only>
Yes
Yes
No
An error is displayed
When the power is switched on, does the screen come on?
No
The instrument may be damaged in­ternally. Request repairs.
Yes
Check the power supply.
The instrument may be damaged internally. Request repairs.
Inspection complete
An error is displayed
Inspection complete
Has there been a power outage or other interruption of power?
No
Measurement Procedures
• Before using, be sure to read " Safety Information" (p.2), " Operating Pre­cautions" (p.5) and Chapter 2 Preparin g fo r Meas ure ment ( p.17 ).
• Before using the Tester the first time, verify that it operates normally to ensure that no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.

Pre-Operation Inspection

Instrument inspection
Chapter 3
1 2
3 4
5
Inspection at power on
6 7 8 9
10
索引
Chapter 3 Measurement Procedures
26
[All indicators on]
[measurement]
[POWER]

3.1 Turning the Power On and Off

Press and hold the [POWER] key for at least 1 second to turn
1.
on the instrument. All the indicators on the LCD display will turn on briefly, and the instrument will enter measurement mode.
To turn off the instrument, press and hold the [POWER] key
2.
for at least 1 second. Setting and measurement data is treated as follows when
3.
the instrument is turned off:
Measured value (last value) Cleared Maximum value
Minimum value Count value Cleared Measurement unit setting Saved Averaging setting Saved Auto Power Save setting Reverts to “ON” Buzzer setting Saved Output function setting Saved
• Settings are saved when the instrument is turned off. Consequently, if power to the instrument is interrupted without pressing the after settings have been changed (for example, by removing the batteries or turning off power to the AC adapter), those settings will not be saved.
• Settings are saved when the instrument is turned off by the Auto Power Save function.
Cleared
[POWER] key
27
Chapter 3 Measurement Procedures
[MODE]
[MODE]
[MODE]
[MODE]
[MODE]
[MODE]
[MODE]
with Contact Adapter connected
Rotational speed measurement
Circumferential speed measurement
Rotational speed measurement
Period measurement
Count measurement
Circumferen­tial speed measurement

3.2 Selecting Measurement Functions

The unit of display changes each time the [MODE] key is pressed.
1 2
3 4
5
6 7 8 9
10
索引
When selecting circumferential speed measure­ment, attach the 9212 Peripheral Ring to the Z5003 Contact Adapter.
Chapter 3 Measurement Procedures
28

3.3 Measurement

Measurement Method
When performing contactless measurement, reflective
1.
tape is applied to the measurement target. Contact mea­surement is used to measure rotating objects that cannot be stopped since it is impossible to apply reflective tape in such applications. Please note that 19,999 r/min. is the maximum rotational speed that can be measured.
If the measured value is being held, press the [HOLD] key
2.
to cancel the hold function. Orient the instrument so that the red light projector is per-
3.
pendicular to the reflective tape. If performing contact measurement, place the contact tip or circumferential speed ring in contact with the measure­ment target.
Hold and read the measured value.
4.
Stopping measured value display updates Limiting measured value variation Displaying maximum and minimum values Stopping the buzzer Canceling the Auto Power Save function Resetting the system Reading the battery life indicator
• When using the Contact Adapter, slippage of the apparatus may introduce an error component into readings.
• When using the Contact Adapter, the method with which the apparatus is brought into contact with the measurement target may introduce an error component into readings.
(p.31) (p.31) (p.32) (p.33) (p.33) (p.34) (p.34)
29
Chapter 3 Measurement Procedures
*Radius is the distance between the center of the rotating ob-
ject and the center of the tape.
Rotation (r/min.)
Reflective tape
Center
The area of the reflective tape.
60
40
20
60
10
6
5
10 (×10
4
)
1
Radius
Rotating body
Radius (mm)

3.4 Precautions in Measurement of High-Rotation Objects

Detection of reflected light uses modulated light to minimize the effects of incident light. When this modulated light is input for a fixed period of time (about 0.2 ms) or longer, a single pulse is detected. For this reason, if the light pulse generated by the passing reflective tape is less than 0.2 ms detection is not possi­ble. The range that can be detected with a 12 mm square target of reflective tape is indicated below.
1 2
3 4
5
6 7 8 9
10
索引
Chapter 3 Measurement Procedures
30
Attach reflective tape Area where reflective tape cannot be at-
tached Rotating body
Rotating body Reflective portion Area where light reflects
Non-reflective portion
If the reflective tape cannot be attached within this detection range, increase the area of the reflective tape so that the gener­ated pulse is 0.2 ms or higher.
For measurement of 30,000 r/min. or higher, use the following method:
The red light from the instrument should be adjusted slightly off center as shown, not to the center of the rotating body.
31
Chapter 4 Instrument Functionality
[AVG]
Instrument Functionality

Measured Value Hold

Stops updating the measured value to allow easy reading.
1. Press
2.
[HOLD] key once to fix the
measured value, preventing updates. The indicator will
light up.
Press the [HOLD] key again to resume display updates.
Chapter 4
1 2
3 4
5

Measured Value Overrun Display

The indicator will flash when the measurement range is exceeded.

Averaging Function

The averaging function is used when measured values exhibit instability. Pressing the on and off. Activating the averaging function enables the last digit of the display.
[AVG] key toggles the averaging function
6 7 8 9
10
索引
Chapter 4 Instrument Functionality
32
[MAX/MIN]
[MAX/MIN]
[MAX/MIN]
B
u
z
z
e
r

Maximum/Minimum Value Display

1. Pressing the
Current value Maximum value Minimum value. Pressing the [CLR] key while displaying the maximum or
2.
minimum value clears that value to zero.
[MAX/MIN] key toggles the display as follows:

Rotation Detection Display Function

1. When the measurement signal (reflected light signal) is
detected, an LED indicator built into the
[HOLD] key will
flash. The buzzer will sound when the LED indi-
2.
cator flashes.
See
" Configuring the Buzzer" (p.33)
33
Chapter 4 Instrument Functionality
Buzzer disabled
Buzzer enabled
APS enabled
APS disabled

Configuring the Buzzer

The buzzer is configured when turning on the instrument.
Turning off the buzzer:
Turn on the instrument by pressing the
[POWER] key while holding down the [AVG]
key to disable the buzzer.
Re-enabling the buzzer:
Turn on the instrument by pressing the
[POWER] key while holding down the [AVG]
key again to re-enable the buzzer.

Display Backlight

The display backlight remains on at all times and cannot be turned off.
1 2
3 4
5

Auto Power Save (APS) Function

The Auto Power Save function is configured when turning on the instrument.
• The function operates automatically when the instrument is turned on.
• After five minutes of inactivity and no measurement signal detection, the instrument turns off automatically. This function is only operative when using battery power.
• "APS" starts to blink at 30 seconds before turning off.
Disabling Auto Power Save: :
Turn on the instrument by pressing the [POWER] key while holding down the [HOLD] key to disable Auto Power Save.
Re-enabling Auto Power Save:
Turn off the instrument and then turn it back on. The APS function will be enabled.
6 7 8 9
10
索引
Chapter 4 Instrument Functionality
34

Resetting the System

The system is reset when turning on the system. Turn on the instrument by pressing the
ing down the
[CLR] key to reset the system.
[POWER] key while hold-
Reset parameters
Measured value Cleared Maximum and minimum values Cleared Measurement function Rotational speed measurement Measurement unit r/min Auto Power Save function ON Averaging function OFF Buzzer ON Output function Aout×10

Battery Warning (Remaining Battery Life Detection)

The instrument’s remaining battery life is shown at the top right corner of the display.
Remaining battery
life indicator
• Use of manganese batteries will result in a dra-
• The remaining battery life indicator will not oper-
Battery status
After new alkaline batteries have been loaded
When 2/3 of the battery life remains
When 1/3 of the battery life remains (Flashing) There is no battery life remaining. When
this indicator is displayed, further measurement is not possible. Replace the batteries in the instrument.
matically shorter continuous operating time. ate properly when using nickel-metal-hydride
batteries.
35
Chapter 5 Specifications
Specifications Chapter 5

Function specifications

Measurement functions
Measurement method
Measurement mode switching
Averaging function
Rotational speed measurement:
Rotational speed measurement, period measurement, count measurement
Circumferential speed measurement:
(when using Z5003 Contact Adapter and 9212 Periph­eral Ring)
Visible-spectrum light photoelectric reflection:
Using red visible-spectrum light and reflective tape or a
reflective plate (Z5003 Contact Adapter) Connection method: Contact or contactless Contactless measurement detection range:
50 to 500 mm (when instrument is oriented perpendicular
to reflective tape) (when using 12 mm Reflection detection indicator:
Flashing red LED built into Sampling period: 62.5 ms to 2 s
Varies with contact/contactless measurement and aver-
aging setting Display refresh rate: Approx. 0.5 to 10 times/sec Processing at no input:
Displays 0 (during period meas ur e m e nt , di sp l a ys m a xi-
mum value) Processing at input overrun: Displays [OVER] Range switching: Automatic
Setting method: Repeated [MODE] key input Operation:
Unit of measurement changes with repeated key input Without Contact Adapter:
Rotational speed (r/min) Rotational speed (r/s)
Period (ms) Count (counts) With Contact Adapter:
Rotational speed (r/min) Rotational speed (r/s)
Period (ms) Cou nt (counts) → Circumferential speed
(m/min) ( lights up) Circumferential speed (m/s)
( lights up) Setting method: Repeated [AVG] key input
Operation: Off Averaging on with AVG indicator lit Function description:
Off: Max. gate time of 0.2 s
On: Max. gate time of 2.0 s Default setting: Off
*Function does not operate (i.e., is turned off) during
count measurement.
2
reflective tape)
[HOLD] key, buzzer
1 2
3 4
5
6 7 8 9
10
索引
Chapter 5 Specifications
36
Display value hold
Display of maximum/ minimum val­ues
Clearing of measured val­ues
APS (Auto Power Save)
Buzzer Operation description:
Remaining battery life de­tection
Contact Adapter detection
Power supply input port (FT3406 only)
Activation method:
Updating of measured values is stopped with input. The indicator will light up.
Deactivation method: Repeated Configuration method: Repeated [MAX/MIN] key input
Operation description:
Normal value Maximum value ( lights up) Minimum value ( lights up)
*This function is not available during count measurement
(the key is disabled).
Operation method: Operation description:
The current value, maximum value, minimum value, count measured value, and circumferential sp eed value are reset to zero.
Notes:
The measured value is also cleared when switching mea ­surement modes and changing the averaging sett ing.
Operation description:
The instrument turns off after 5 minutes of inactivity and
no measurement signal detection. Initial state: On ( lights up) Deactivation method: Power-on option
*APS is automatically disabled when the AC adapter is
connected (when a standard voltage is detected).
The buzzer sounds for 30 ms each time a rotation is de-
tected. Initial state: On ( lights up) Deactivation method: Power-on option
*The buzzer also sounds when keys are pressed (this
functionality cannot be disabled). Operation description:
At 1.9 V (±0.1 V), the remaining battery life is shown as
0, the battery outline flashes, and the measured value
display is turned off. At and below 1.8 V (±0.1 V) , t he i n-
strument turns off. Function description:
Attachment of the Contact Adapter is automatically detected. Operation description:
When switching measurement modes, the circumferen-
tial measurement setting is available. Application: Z1004 AC adapter connection
Output port: 5.5 mm diameter, center positive Max. input voltage: DC ±5 V ±0.3 V
[CLR] key input
[HOLD] key
[HOLD] key input
[HOLD] + [POWER]
[AVG] + [POWER]
37
Chapter 5 Specifications
Measurement mode
Setting r/min r/s ms
counts
m/min m/s
Aout×10 M×10/60 M×10 (1,000/M)
×10
r/min
output
M×10/ 60×10
M×10 ×10
Aout×1M×1/60 M×1 (1,000/M)
×1
M×1/ 60×10
M×1 ×10
Aout×0.1 M×0.1 M×0.1
×60
(1,000/M) ×0.1×60
M×0.1/ 60×10
M×0.1
×
60
×10
Aout×0.01 M×0.01 M×0.01
×60
(1,000/M) ×0.01×60
M×0.01/ 60×10
M×0.01×60 ×10
AC adapter detection (FT3406 only)
Pulse output (FT3406 only)
Analog output (FT3406 only)
Output information: Output is calculated from the display value M as follows
Operation description:
The indicator lights up when voltage is detected.
Notes:
The indicator turns off when the detected voltage value is 0 (for example, during a power outage), even if the adapter is connected.
Port profile: 3.5 mm diameter, earphone jack type
(Shared port; switchable between pulse output and ana-
log output) Output resistance: 1 k Output level: 0 to 3.3 V Output information: Outputs the detectio n pulse. Active-low output Low output duration: 300 μs
Output port: 3.5 mm diameter, earphone jack type (shared port; switchable between pulse output and ana­log output) Output resistance: 1 k Output level: 0 to 1 V f.s. Resolution: 1 mV Response speed: Varies with averaging setting Accuracy: ±2% f.s
Accuracy guarantee temperature and humidity range: 23°C ±5°C, 80% RH or less Outside above temperature range: Add accuracy specifications × 0.1 × (|T-23| ), where T is the operating temperature in Celsius.
.
1 2
3 4
5
6 7 8 9
10
Output port setting (FT3406 only)
Configuration method: Power-on option [MODE] + [POW-
, then [MODE] key input to cycle Aout×10 Aout×1
ER]
Aout×0.1 Aout×0.01 Pout Setting takes effect after 3 seconds of inactivity. Initial setting: Aout×10
(Unit:mV)
索引
38
Instrument
Assigned
function
Regular function Power-on option
[POWER] Power switch ---­[MODE] Measurement mode
switching
Output port setting (FT3406 only)
[AVG] Averaging function
setting
Buzzer setting
[MAX/MIN] Maximum and mini-
mum value display
----
[CLR] Value clear System reset [HOLD] Hold/cancel
measured value
APS setting
operation
Chapter 5 Specifications
Display Transmissive LCD: FSTN
Ranges and measurement ranges
Note 1: The lowermost digit is fixed at 0 at speeds of 20,000 r/min and over. Note 2: The lowermost digit is fixed at 0 when the averaging setting is off.
Backlight: Always on
(1) Rotational speed measurement
MODE Range
Rotational
speed
measure-
ment
(r/min)
Rotational
speed
measure-
ment
(r/s)
Contactless measurement Contact measurement AVG=ON AVG=OFF AVG=ON AVG=OFF
30.00 to
1
199.99
200.0 to
2
1999.9
2000 to
3
19999
20000 to
4
99990
0.5000 to
1
1.9999
2.000 to
2
19.999
20.00 to
3
199.99
200.0 to
4
1600.0
--
300.0 to
1999.0
2000 to
19990
20000 to
99990
--
5.000 to
19.990
20.00 to
199.90
200.0 to
1600.0
15.00 to
199.99
200.0 to
1999.9
2000 to
19999
-- --
0.2500 to
1.9999
2.000 to
19.999
20.00 to
199.99
200.0 to
333.0
--
150.0 to
1999.0
2000 to
19990
--
2.500 to
19.990
20.00 to
199.90
200.0 to
333.0
39
Period
AVG=ON AVG=OFF
Up to 9,999 counts ±1dgt. ±10dgt. 10,000 counts or more ±2dgt. ±20dgt. 20,000 counts or more
(r/min mode only)
±20dgt. ±100dgt.
Period measurement only
±0.5%rdg. is added to
above-mentioned accuracy.
measure-
ment (ms)
Count
measure-
ment
(count)
Chapter 5 Specifications
0.6000 to
1
1.9999
2.000 to
2
19.999
20.00 to
3
199.99
200.0 to
4
1999.9
1
0.6000 to
1.9990
2.000 to
19.990
20.00 to
199.90
--
(Input conditions: Up to rotational speed
0 to 999999
measurement upper limit)
-- --
3.000 to
19.999
20.00 to
199.99
200.0 to
3999.9
3.000 to
19.990
20.00 to
199.90
200.0 to
399.0
1 2
3
(2) Circumferential speed measurement
MODE Range
Circumferential speed
measurement
(m/min)
Circumferential speed
measurement
(m/s)
1 1.500 to 19.999 -­2 20.00 to 199.99 15.00 to 199.90 3 200.0 to 1999.9 200.0 to 1999.0 1 0.0250 to 1.9999 0.2500 to 1.9990 2 2.000 to 19.999 2.000 to 19.990 3 20.00 to 33.30 20.00 to 33.30
AVG=ON AVG=OFF

Accuracy (Not applicable to count measurement)

Contact measurement
4
5
6 7 8 9
10
索引
Chapter 5 Specifications
40

General specifications

Guaranteed accuracy period
Operating temperature
Operating humidity
Storage temperature and humidity
Operating environment
Drop proof 1 m onto concrete surface Power supply 5 VDC or LR6 alkaline battery × 2 Rated supply
voltage Maximum rated
power Continuous op-
erating time
Dimensions (max.)
Mass Approx. 230g (8.1 oz) (Including battery) Dust resistance IP50 (EN60529) Applicable
Standards
Accessory Instruction manual ...............................................................1
Option Z5003 Contact Adapter
1 year
0 to 50°C (32 to 122°F)
Up to 40°C (104°F) 80%RH or less 40°C to 45°C (104 to 113°F) 60%RH or less 45°C to 50°C (113 to 122°F) 50%RH or less (non-condensing)
-10°C to 50°C (14 to 122,°F) 80%RH or less (non-condensing)
Indoors, Pollution degree 2, altitude up to 2000 m (6562-ft.)
Battery power 1.5 V DC × 2 AC adapter 5 V DC ±0.3 V
0.5 VA
FT3405: Approx. 30 hours FT3406: Approx. 25 hours *When using alkaline batteries with buzzer off.
Approx. 71(W)×186(H) × 38(D)mm (2.80"W × 7.32"H ×1.50"D)
EMC :EN61326 Safety :EN61010
9211 Reflective Tape...........................................................1
L9094 Output Cord (FT3406 only).......................................1
C0202 Carrying Case..........................................................1
LR6 alkaline battery.............................................................2
9032 Metal Contact Tip 9033 Rubber Contact Tip 9212 Peripheral Ring Z1004 AC Adapter (FT3406 only) L9094 Output Cord (FT3406 only) 9211 Reflective Tape C0202 Carrying Case
41
Chapter 6 Maintenance and Service
Maintenance and Service
Chapter 6
1

6.1 Cleaning

• To clean the instrument, wipe it gently with a soft cloth moist­ened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gaso­line, as they can deform and discolor the case.
• Wipe the Lens and LCD gently with a soft, dry cloth.

6.2 Consumables

Replaceable Parts and Operating Lifetimes. Useful life depends on the operating environment and frequency of use. Operation cannot be guaranteed beyond the following periods For replacement parts, contact your dealer or Hioki rep-
resentative.
Part Operating lifetime
Measurement vis­ible-spectrum red LED and sensor
LCD backlight 50,000 hours
Batteries Operating lifetime varies with conditions of use.
8,000 hours (about 3 years at 8 hours of use per day) A reduction in detection range indicates the operating lifetime is almost over. The instrument should be re­paired.
If the display becomes difficult to see (dim), the instru­ment should be repaired.
When the remaining battery life indicator is displayed as , replace the batteries with new batteries.

6.3 Discarding the Instrument

• Battery may explode if mistreated. Do not short-circuit, recharge, disassemble or dispose of in fire.
• Handle and dispose of batteries in accordance with local regu­lations.
2
3 4
5
6 7 8 9
10
索引
Chapter 6 Maintenance and Service
42

6.4 Troubleshooting

If damage is suspected, check the “Before returning for repair” before contacting your dealer or Hioki representative.
Before returning for repair:
Symptom Checks and Remedy
Although you have turned on the pow­er switch, the screen display does not appear.
The instrument turns off.
is lit up. The RING setting cannot be activated without f irst at-
The detection range for contact­less measurement is low.
Have the batteries been inserted in the proper orien­tation? (p.17)
Are the batteries correctly inserted? Insert the new batteries.
Has the AC adapter been connected properly? (FT3406 only) (p.19)
Is the battery dead? Replace the battery.
Is the auto power save function being activated? Check the auto power save setting. (p.33)
AC adapter output may decrease due to power sup­ply noise, a momentary loss of power, or other fac­tors, causing the instrument to switch to battery power. If no batteries are loaded in the i nstrument, it will turn off. Keep fresh batteries in the instrument.
taching the Contact Adapter. The sensor’s characteristics vary with temperature.
As the ambient temperature increases, the detection range will shorten. This does not indicate an instru­ment malfunction.
This message indicates an internal instrument mal­function. The instrument should be repaired.
(p.17)
(p.17)
Pack the instrument so that it will not sustain damage during shipping, and include a description of existing damage. We do not take any responsibility for damage incurred during shipping.
Loading...