Thank you for purchasing the HIOKI DT4281, DT4282 Digital
Multimeter. To obtain maximum performance from the product,
please read this manual fi rst, and keep it handy for future reference.
1
Verifying Package Contents
When you receive the instrument, inspect it carefully to ensure that
no damage occurred during shipping.
In particular, check the accessories, panel switches, and
connectors. If damage is evident, or if it fails to operate according
to the specifi cations, contact your authorized Hioki distributor or
reseller.
Check the package contents as follows.
DT4281 or DT4282
DT4281DT4282
L9207-10 Test Lead (p. 28)
LR6 Alkaline battery × 4
Instruction Manual
(English)
Instruction manuals may also
be available in other languages.
Please visit our website at
http://www.hioki.com
/
2
3
4
5
6
7
Ind.Appx.
1
Page 7
Options (sold separately)
Options (sold separately)
The following options are available for the instrument. Contact your
authorized Hioki distributor or reseller when ordering.
Connecting cables
*1: CATIV 600 V/CATIII 1000 V/CATII 1000 V
*2: CATIV 600 V/CATIII 1000 V
*3: CATIII 1000 V *5: CATIII 300 V/CATII 600 V
*4: CATIII 600 V *6: 33 V AC/70 V DC
*7: CATIII 600 V/CATII 600 V
6
L4933*
Contact Pin Set
L9207-10*
1
Test Lead
2
L4930*
Connection Cable Set
(Length: 1.2 m)
2
L4931*
Extension Cable Set
(Length: 1.5 m, with the
coupling connector)
5
L4934*
Small Alligator Clip Set
2
L4935*
Alligator Clip Set
3
9243*
Grabber Clip
4
L4936*
Bus Bar Clip Set
3
L4937*
Magnetic Adapter Set
1
L4932*
Test Pin Set
7
L4938*
Test Pin Set
4
L4939*
Breaker Pin Set
2
Page 8
Options (sold separately)
For the clamp current measurement (only compatible with the DT4281)
9010-50, 9018-50, 9132-50*
Clamp-on Probe
9704 Conversion Adapter
4
1
Clamp-on probe
9010-50, 9018-50500 A rms46 mm or less
9132-501000 A rms55 mm or less, 80 × 20 mm bus-bar
Temperature measurement
C0202 Carrying Case
Z5004 Magnetic Strap (p. 31)
Rated
current
DT4910 Thermocouples (K) (p. 45)
• Temperature measuring junction: Exposed type (welding)
• Sensor length: Approx. 800 mm
• Operating temperature: -40°C to 260°C (temperature
measuring part), -15°C to 55°C (connector)
The instrument, test leads, instruction manual, and others
can be stored in the case.
Attach this strap to the instrument and secure it on the
wall surface such as a metal plate for use.
Diameter of the measurable
conductor
2
3
4
5
6
7
Ind.Appx.
3
Page 9
Options (sold separately)
DT4900-01 Communication Package (USB) (p. 72)
A communication adapter, USB cable, PC software,
and communication specifi cations are included.
The instrument data can be stored on the PC.
4
Page 10
Safety Notes
Safety Notes
This instrument is designed to conform to IEC 61010 Safety
Standards, and has been thoroughly tested for safety prior to
shipment. However, using the instrument in a way not described in
this manual may negate the provided safety features.
1
Before using the instrument, be certain to carefully read the
following safety notes.
DANGER
Mishandling during use could result in injury or death,
as well as damage to the instrument. Be certain that
you understand the instructions and precautions in
the manual before use.
WARNING
With regard to the electricity supply, there are risks
of electric shock, heat generation, fi re, and arc
discharge due to short circuits. If persons unfamiliar
with electricity measuring instruments are to use
the instrument, another person familiar with such
instruments must supervise operations.
Protective gear
WARNING
To avoid electric shock when measuring live lines,
wear appropriate protective gear, such as insulated
rubber gloves, boots and a safety helmet.
2
3
4
5
6
7
Ind.Appx.
5
Page 11
Safety Notes
Notation
In this manual, the risk seriousness and the hazard levels are
classifi ed as follows.
DANGER
WARNING
CAUTION
IMPORTANT
Indicates an imminently hazardous situation that will
result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may
result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may
result in minor or moderate injury to the operator or
damage to the instrument or malfunction.
Indicates information related to the operation of the
instrument or maintenance tasks with which the
operators must be fully familiar.
Indicates a high voltage hazard.
If a particular safety check is not performed or the
instrument is mishandled, this may give rise to a
hazardous situation; the operator may receive an
electric shock, may get burnt or may even be fatally
injured.
Indicates a strong magnetic-fi eld hazard.
The effects of the magnetic force can cause
abnormal operation of heart pacemakers and/or
medical electronics.
Indicates prohibited actions.
Indicates the action which must be performed.
*
Additional information is presented below.
6
Page 12
Symbols affi xed to the instrument
Indicates cautions and hazards. When the symbol is printed on
the instrument, refer to a corresponding topic in the Instruction
Manual.
Indicates that dangerous voltage may be present at this terminal.
Safety Notes
1
Indicates a double-insulated device.
Indicates a fuse.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates DC (Direct Current) or AC (Alternating Current).
Symbols for various standards
Indicates the Waste Electrical and Electronic Equipment
Directive (WEEE Directive) in EU member states.
Indicates that the instrument conforms to regulations set out by
the EC Directive.
2
3
4
5
6
7
Ind.Appx.
7
Page 13
Safety Notes
Screen display
This instrument uses the following screen displays.
AB CD E FGH I J K LMNOPQR S TU VWXY Z
123456 7890
A different display is used in the case below.
Appears when a broken Thermocouple (K) is detec ted. (p. 45)
Accuracy
We defi ne measurement tolerances in terms of rdg. (reading) and
dgt. (digit) values, with the following meanings:
(Reading or displayed value)
rdg.
The value currently being measured and indicated on the
measuring instrument.
(Resolution)
The smallest displayable unit on a digital measuring instrument,
dgt.
i.e., the input value that causes the digital display to show a “1”
as the least-signifi cant digit.
8
Page 14
Safety Notes
Measurement categories
To ensure safe operation of measuring instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement categories.
DANGER
• Using a measuring instrument in an environment
designated with a higher-numbered category than
that for which the instrument is rated could result in
a severe accident, and must be carefully avoided.
• Using a measuring instrument without categories in
an environment designated with the CAT II to CAT
IV category could result in a severe accident, and
must be carefully avoided.
This instrument conforms to the safety requirements for CAT III 1000 V, CAT
IV 600 V measuring instruments.
CAT II: When directly measuring the electrical outlet receptacles of
CAT III: When measuring the primary electrical circuits of heavy equipment
CAT IV: When measuring the circuit from the service drop to the service
See: “2.3 Using Test Leads” (p. 28)
the primary electrical circuits in equipment connected to an
AC electrical outlet by a power cord (portable tools, household
appliances, etc.)
(fi xed installations) connected directly to the distribution panel,
and feeders from the distribution panel to outlets
entrance, and to the power meter and primary overcurrent
protection device (distribution panel)
Distribution panel
Service entrance
Service drop
CAT IV
Power meter
Internal wiring
CAT III
Fixed installation
CAT II
T
Outlet
1
2
3
4
5
6
7
Ind.Appx.
9
Page 15
Usage Notes
Usage Notes
Follow these precautions to ensure safe operation and to obtain the
full benefi ts of the various functions.
DANGER
If the test lead or the instrument is damaged, there is
a risk of electric shock. Before using the instrument,
perform the following inspection.
• Before using the instrument, check that the coating of
the test leads are neither ripped nor torn and that no
metal parts are exposed. Using the instrument under
such conditions could result in electrocution. Replace
the test leads with those specifi ed by our company.
• Before using the instrument for the fi rst time, verify
that it operates normally to ensure that no damage
occurred during storage or shipping. If you fi nd any
damage, contact your authorized Hioki distributor or
reseller.
Installation
Installing the instrument in inappropriate locations may cause a
malfunction of instrument or may give rise to an accident. Avoid the
following locations. For details on the operating temperature and
humidity, see the specifi cations. (p. 95)
CAUTION
• Exposed to direct sunlight or high temperature
• Exposed to corrosive or combustible gases
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to a strong electromagnetic fi eld or electrostatic
charge
• Exposed to high quantities of dust particles
• Near induction heating systems (such as high-frequency
induction heating systems and IH cooking equipment)
• Susceptible to vibration
10
Page 16
Usage Notes
Handling the cables
WARNING
To prevent electric shock, when measuring the
voltage of a power line use a test lead that satisfi es
the following criteria:
• Conforms to safety standards IEC61010 or EN61010
• Of measurement category III or IV
• Its rated voltage is higher than the voltage to be
measured
All of the optional test leads for this device conform
to the safety standard EN61010. Use a test lead in
accordance with its defi ned measurement category
and rated voltage.
CAUTION
• Avoid stepping on or pinching the cable, which could
damage the cable insulation.
• To avoid damaging the cables, do not bend or pull the
leads and the probe bases.
The ends of the test leads are sharp. Be careful to avoid
injury.
For the test leads supplied with the instrument or the options to be connected
to the instrument, see the following information.
See the Instruction Manual which
accompanies the optional clamp.
“2.4 Installation in Measurement
Location” (p. 31)
1
2
3
4
5
6
7
Ind.Appx.
11
Page 17
Usage Notes
Precautions during measurement
WARNING
If the instrument is used in locations where the rating
indicated on the instrument or probes is exceeded,
the instrument may be damaged resulting in personal
injury. Do not use the instrument in such locations.
See “Measurement categories” (p. 9).
• With regard to the 10 A range, the maximum input
current is 10 A DC/10 Arms AC. Supplying a current
in excess of the maximum input may damage the
instrument and result in personal injury. Do not
supply current in excess of the specifi ed limit. (Only
the DT4282)
Observe the following to avoid electric shock and/or
short circuits.
• Hazardous voltage may be generated in a free
measurement terminal. Do not touch the free
terminal.
• Use only test leads and optional equipment
specifi ed by our company.
• Do not allow the metal part of the test lead to touch
any exposed metal, or to short between 2 lines.
Never touch the metal end.
• When connecting the clip-type test lead to the
active terminal, do not allow the lead to touch any
exposed metal, or to short between 2 lines.
• When the clamp-on probe is opened, do not allow
the metal part of the clamp to touch any exposed
metal, or to short between 2 lines, and do not
use over bare conductors. (For the clamp current
measurement, only the DT4281)
12
Page 18
Usage Notes
CAUTION
• Do not input voltage or supply current exceeding the
specifi ed measurement range. Doing so may damage
the instrument.
• During the continuity check, diode test, or measurement
of resistance, conductance, or electrostatic capacity,
measurement signals are generated in the terminals of
the instrument. Depending on the target for measurement,
the measurement signal may cause damage.
Seeing “Measurement current” and “Open circuit
voltage” in the accuracy table (p. 81), check, in advance,
that there are no adverse effects of the measurement
current and the open circuit voltage.
1
2
3
Precautions during shipment
Observe the following during shipment. Hioki cannot be responsible
for damage that occurs during shipment.
CAUTION
• During shipment of the instrument, handle it carefully so
that it is not damaged due to a vibration or shock.
• To avoid damage to the instrument, remove the
accessories and optional equipment from the instrument
before shipment.
If the instrument is not to be used for an extended
period of time
IMPORTANT
To avoid corrosion and/or damage to the instrument due to battery
leakage, remove the batteries from the instrument if it is to be
kept in storage for an extended period.
13
4
5
6
7
Ind.Appx.
Page 19
1
Overview
1.1 Overview and Features
This measuring instrument is a multi-function, high-precision digital
multimeter which ensures both safety and durability.
Main features
• Speedy display of the RMS measured value
• Environmental performance (can be used
anywhere) (Operation temperature: -15 to 55°C)
• High noise-proof performance
• Filter function effective for the inverter measurement
• Solid body which can be used for an extended
period of time (drop-proof)
• High accuracy (DCV: 0.025%), wide band (20 Hz to 100 kHz) measurement
• Speedy measurement via a fast response (0 V 100 V response 1 second*)
* Until the value falls within the accuracy specifi cation range.
If there is an excessive input,
a hazard is indicated by the
red display.
Large, easily-viewable display
Backlighting for use in dark
environments
For the DC power
supply maintenance
The peak measurement
function is used to
capture the ripple voltage
superimposed on the DC
signal.
PEAK max
The measurement test
leads and the end pins can
be selected depending on
the purpose of use.
PEAK min
100V85.9V114.1V
Problem fi nding a
suitable installation
location?
The strap with magnet allows the
device to be hung conveniently.
Useful functions during
measurement
• Noise reduction (FILTER)
• Display stability (SLOW)
• Display hold (HOLD)
• Maximum/Minimum value
display
Memorizing
measured values
For control of the UPS
battery cell voltage, etc.,
the measurement value can
be saved to the embedded
memory (up to 400 data).
The measurement value can
also be read.
Data transmission
to PC,
control
The optional
DT4900-01
Communication
Package is required.
1
2
3
4
5
6
7
Appx.Ind.
15
Page 20
Parts Names and Functions
1.2 Parts Names and Functions
Front
Some indications are different between the DT4281 and DT4282.
Display
(p. 21)
Operation
keys (p. 17)
Rotary
switch
(p. 18)
Measurement
terminals
(p. 19)
DT4281DT4282
16
Page 21
Operation keys
46
1
2
3
5
Parts Names and Functions
(*) Pressing a key for at
8
least 1 second activates
the function indicated
9
above the key
the function, change the
rotary switch setting.
7
10
. To disable
1
2
10
1
2
3
4
5
6
7
8
9
• Displays the maximum or minimum value. (p. 61)
• (*) activates the relative value display function. (p. 63)
• Selects the range (manual range). (p. 55)
• (*) changes to the auto range.
(The default is the auto range.)
Changes to the peak measurement. The maximum or
minimum instantaneous value is measured. (p. 62)
• Deletes the stored data. (p. 68)
• Clears the maximum or minimum value. (p. 61)
• Clears the peak value. (p. 62)
• Retains the displayed value. (p. 57)
• (*) activates the auto hold function.
• Reads the stored data. (p. 67)
• (*) toggles the fi lter function between on and off. (p. 60)
• Saves the measurement data. (p. 65)
• (*) toggles the display update speed between normal and
slow. ([SLOW] is used to stabilize the display.) (p. 59)
Turns on or off the display backlighting. (p. 69)
Increases/Decreases the memory No. or numeric value.
(When / are displayed on the LCD, these keys can be
operated.)
Changes the function of the rotary switch to that indicated
shown in blue.
3
4
5
6
7
Appx.Ind.
17
Page 22
Parts Names and Functions
Rotary switches and measurement descriptions
5
4
3
2
1
6
7
3
8
2
9
1
DT4281
is used to change to the measurement shown in blue.
DT4281
: DT4281 only,
10
(
1
2
3
4
5
6
7
8
9
Turns off the power to the instrument.
AC voltage measurement (p. 38) dBm measurement
dBV measurement (p. 41)
measurement (p. 51)
DC voltage measurement (p. 39)
Synthesized voltage measurement of DC and AC Voltage
measurement of AC and DC components (p. 39)
Continuity check (p. 42) Diode test (p. 43)
Resistance measurement (p. 44) Tem pe rat ur e
measurement (compatible with the Thermocouples (K)) (p. 45)
(6000 A range) DC measurement AC measurement (p. 49)
(600 mA range) DC measurement AC measurement (p. 49)
% conversion of the 4-20 mA input (The % conversion of the
0-20 mA input can be specifi ed via the power-on option.) (p. 53)
(10 A range)
DT4282
DC measurement AC measurement (p. 49)
4
DT4282
DT4281
DT4282
5
DT4282
: DT4282 only)
Clamp current
Conductance
6
7
8
9
10
18
Page 23
Parts Names and Functions
Measurement terminals
12 3 4234
DT4281DT4282
DT4282
1
Current measurement (A) terminal. The red test lead is connected.
When the rotary switch is set to the current measurement, the shutter
opens.
Current measurement (A, mA) terminal. The red test lead is
2
connected. When the rotary switch is set to the current measurement,
the shutter opens.
Commonly used for each measurement. The black test lead is
3
connected.
Used for voltage measurement, resistance measurement, continuity
4
check, diode test, temperature measurement, clamp current
measurement
referred to as “V terminal”.
The red test lead is connected.
When the rotary switch is set to any of the measurements above, the
shutter of the current measurement terminal is closed.
DT4281
, or conductance
DT4282
. Hereafter
1
2
3
4
5
6
Be sure to carefully read the precautions in the following references.
See “Precautions during
measurement” (p. 12).
See “Replacing Fuses”
(p. 106).
19
7
Appx.Ind.
Page 24
Parts Names and Functions
(p)
Rear
Serial number label
It is necessary for production
control such as product
warranty.
Do not peel off the label.
Communication port
When the communication
adapter supplied with the optional
DT4900-01 Communication
Package is connected, the data
can be transmitted to the PC.
(p. 72)
Test lead holder
The test lead can be held.
Strap hole
The optional Z5004 Magnetic
Strap can be attached. (p. 31)
Stand
The instrument can be set on the
stand. (p. 31)
Battery cover
When replacing the batteries
(p. 26) or fuse (p. 106), remove
the cover.
See p. 26.
20
Page 25
1.3 Display
For error displays, see “6.3 Error Display” (p. 105).
The auto power save function is activated. (p. 70)
The buzzer can be used. (p. 69)
5
6
7
Main display
Sub display
Display
1
2
3
4
5
6
7
Appx.Ind.
21
Page 26
Display
6
(Unit)Each unit
7
8
9
10
11
Maximum value (p. 61)
Minimum value (p. 61)
The display update (sampling) is performed at SLOW
speed. (p. 59)
The fi lter function is activated. (p. 60)
The relative value display function is activated. (p. 63)
Auto range (p. 55)
Manual range (p. 55)
Maximum value in the peak measurement (p. 62)
Minimum value in the peak measurement (p. 62)
When are displayed, / can be used to
change values.
Threshold of the continuity check (p. 42)
Threshold of the diode test (p. 43)
AC measurement
DC measurement
AC measurement + DC measurement
During the temperature measurement, the
temperature difference from the standard is displayed.
(p. 45)
Continuity check (p. 42)
Diode test (p. 43)
Clamp current measurement (p. 51)
DT4281
% conversion measurement of 4-20 mA (0-20 mA)
(p. 53)
DT4281
(
22
: DT4281 only,
DT4282
: DT4282 only)
Page 27
Alarm Display and Battery Indicator
1.4 Alarm Display and Battery Indicator
The following conditions are informed via the red backlight,
display, and buzzer.
When the maximum input range is exceeded
Voltage/Current measurement
blinks, the red backlight blinks,
the maximum value within the maximum
range blinks, and the buzzer sounds.
Corrective action:
Immediately move the test leads away
from the measurement object.
When the measured value exceeds the maximum value
in each range ( Range over)
Voltage/Current measurement
blinks, the red backlight lights up,
and the maximum value blinks.
1
2
3
4
5
Corrective action:
and the maximum value blinks.
Change the range.
Measurements other than voltage and current
6
7
Appx.Ind.
Corrective action:
Change the range, or measure the
samples in the specifi ed range.
If the same symptom still occurs, check
that the test leads are not broken.
(p.34)
23
Page 28
Alarm Display and Battery Indicator
When the Thermocouple (K) is broken (Temperature
measurement)
Corrective action:
Check that the thermocouple has been connected
correctly to the measurement terminal. If the display
does not change, replace with a new Thermocouple
(K). (p. 45)
Battery indicator
Fully charged.(Charge: 60% or more)
As the battery charge diminishes,
black charge bars disappear, one
by one, from the left of the battery
indicator.
The battery voltage is low. Replace the
batteries as soon as possible.
(Blinks) The battery is exhausted.
Replace the batteries.
The charge is only a reference for the continuous operation time. (p. 96)
(Charge: 20% or more)
(Charge: 5% or more)
(Charge: less than 5%)
Power shutdown
When the charge is 0% (less than 3.8 V ± 0.2 V),
[bAtt Lo] is displayed for 1 second and the power is
shut down.
24
Page 29
Preparation for Measurements
2
3
1
2
2.1 Measurement Workfl ow
Before using the instrument, be sure to read “Usage Notes” (p. 10).
Installation and connection
Insert the batteries. (p. 26)
Perform the startup check. (p. 33)
Measurement
Turn on the power and select the
measurement function.
Attach the test leads to the
measurement terminals. (p. 28)
(As necessary, perform zero adjustment. (p. 64))
Connect the test leads to the
measurement object.
(As necessary)
Hold the display of the measured value. (p. 57)
Save the measured value. (p. 65)
As necessary, have other
optional items available and
ready.
1
Black
2
Red
1
2
3
4
3
5
6
7
Appx.Ind.
End of the measurement
Move the test leads away from the measurement object and then turn off the power.
25
Page 30
Inserting/Replacing Batteries
2.2 Inserting/Replacing Batteries
Before using the instrument fi rst time, insert four LR6 alkaline
batteries. Before measurements, check that the battery level is
suffi cient. When the battery charge diminishes, replace the batteries.
Nickel-metal hydride batteries
Nickel-metal hydride batteries can be used. However, the discharge
characteristic of these batteries is different from that of alkaline
batteries. Be aware that the remaining battery power display does
not function properly.
WARNING
To avoid electric shock, disconnect the test leads from
the object to be measured before replacing the batteries.
To avoid the possibility of explosion, do not short
circuit, charge, disassemble, or incinerate batteries.
After battery replacement but before using the
instrument, reattach and screw down the battery cover.
CAUTION
Poor performance or damage from battery leakage
could result. Observe the cautions listed below.
• Do no mix new and old batteries, or different types of
batteries.
• Be careful to observe the battery polarity during installation.
• Do not use batteries after their recommended expiry date.
• Do not allow used batteries to remain in the instrument.
• To avoid corrosion from battery leakage and/or damage to
the instrument, remove the batteries from the instrument
if it is to be kept in storage for an extended period.
26
Page 31
Inserting/Replacing Batteries
• The indicator appears when the battery charge diminishes.
Replace the batteries as soon as possible.
• After use, be sure to turn off the instrument.
• Handle and dispose of batteries in accordance with local regulations.
Have the following items
1
available and ready.
• Phillips screwdriver
•Alkaline (LR6) battery or
manganese (R6) battery × 4
Remove the test leads from
Screws
Rear
2
the instrument.
Set the rotary switch to OFF.
3
Using a Phillips screwdriver,
4
remove the screws (2
locations) from the battery
cover on the rear of the
instrument.
Remove the battery cover.
5
When replacing the batteries,
6
remove all old batteries.
Insert 4 new batteries (LR6
7
or R6), being careful to the
battery polarity.
Reattach the battery cover.
8
Secure the cover with the
9
screws.
After the battery cover is removed,
the fuse can be seen. When
replacing the fuse, see “6.4
Replacing Fuses” (p. 106).
1
2
3
4
5
6
7
Appx.Ind.
27
Page 32
Using Test Leads
2.3 Using Test Leads
The L9207-10 Test Lead supplied with the instrument are used for
measurements.
Depending on measurement locations, use our optional
measurement cables. For details on the optional items, see “Options
(sold separately)” (p. 2).
WARNING
• To prevent a short circuit accident, be sure to
use the test leads with the sleeves attached when
performing measurements in the CAT III and CAT
IV measurement categories. (For the measurement
categories, see “Measurement categories” (p. 9).)
• If the sleeves are inadvertently removed during
measurement, stop the measurement.
CAUTION
• To ensure safe operation, use only test leads specifi ed
by our company.
• When carrying out measurements with the sleeves in
place, be careful to avoid damaging the sleeves.
• The tips of the metal pins are sharp and may cause
injury. Do not touch the tips.
28
Page 33
L9207-10 Test lead
Black
Red
Barriers
Using Test Leads
Sleeves
1
Plugs
Metal pinConnect to the object to be measured.
SleeveAttach to the metal pins to prevent short circuit accidents.
BarrierRepresents the safe handling distance from the metal pins.
PlugConnect to the measurement terminals on this instrument.
CableDouble sheathed cables (Length: approx. 900 mm, Diameter:
Cables
The plugs of the test leads are covered
with the safety caps.
Before use, remove the caps.
Safety cap
4 mm or less (sleeve attached)
19 mm or less (sleeve removed)
Diameter approx. 2 mm
During measurement, do not touch the area between
the barrier and the tip of the sleeve.
approx. 3.6 mm)
When the white portion inside the cable is exposed,
replace with a new L9207-10 Test Lead.
Metal pins
29
2
3
4
5
6
7
Appx.Ind.
Page 34
Using Test Leads
1
2
Removing and attaching the sleeves
Removing the sleevesAttaching the sleeves
Gently hold the bottom of the
sleeves and pull the sleeves off.
Safely store the removed sleeves
so as not to lose them.
Connecting to the instrument
Turn the rotary switch to the desired
1
measurement function.
Connect the test leads to the relevant
2
measurement terminals.
• Besides the current measurement
(excluding the clamp)
1
COM terminalConnect the black test lead.
V terminalConnect the red test lead.
• Current measurement
2
COM terminalConnect the black test lead.μA/mA terminal Connect the red test lead.
A terminal (Only the DT4282)
30
Insert the metal pins of the test
leads into the holes of the sleeves,
and fi rmly push them all the way
in.
Page 35
Installation in Measurement Location
2.4 Installation in Measurement Location
Using the instrument with the stand
Position the instrument with the stand at the rear.
CAUTION
• Do not position the instrument on an
unstable table or inclined surface.
• When the instrument is set on the stand,
do not apply a strong force above.
Doing so may damage the stand.
Hanging the instrument with the strap
Attach the optional Z5004 Magnetic Strap to the instrument and
attach the magnet to the wall surface (with metal plate affi xed).
Strap holes
Magnet
Attach it to the wall surface
(with metal plate affi xed).
1
2
3
4
5
6
7
Appx.Ind.
31
Page 36
Installation in Measurement Location
DANGER
Those with medical electronics such as pacemakers
should not use the Z5004 Magnetic Strap. Nor should
such persons approach the Z5004. It is extremely
dangerous. The electronics may not operate properly
and the life of the operator may be put at great risk.
CAUTION
• Do not use the Z5004 in locations where it may
be exposed to rainwater, dust, or condensation. In
those conditions, the Z5004 may be decomposed or
deteriorated. The magnet adhesion may be diminished.
In such case, the instrument may not be hung in place
and may fall.
• Do not bring the Z5004 near magnetic media such
as fl oppy disks, magnetic cards, pre-paid cards, or
magnetized tickets. Doing so may corrupt and may
render them unusable. Furthermore, if the Z5004 is
brought near precision electronic equipment such as
PCs, TV screens, or electronic wrist watches, they may
fail.
32
Page 37
Performing Measurements
3
3.1 Inspection Before Use
Before using the instrument the fi rst time, verify that it operates
normally to ensure that the no damage occurred during storage
or shipping. If you fi nd any damage, contact your authorized Hioki
distributor or reseller.
Appearance check of the instrument and test leads
Check itemAction
The instrument is neither
damaged nor cracked.
The internal circuits are not
exposed.
The terminals are not
contaminated with debris.
The coating of the test leads is
neither broken nor frayed, or
the white portion or metal part
within the lead is exposed.
Check when turning on the power
(Set the rotary switch to any position other than OFF.)
Check itemAction
The battery voltage is suffi cient.
Visually check the instrument.
If it is damaged, there is a risk of electric
shock. Do not use the instrument but send
it for repair.
Remove contamination with a cotton
swab.
If the test lead is damaged, there is a
risk of electric shock. Do not use the
instrument but send it for repair.
When the
top right corner of the display, the battery
voltage is low. Replace the batteries as
soon as possible.
indicator appears in the
1
2
3
4
5
6
7
Appx.Ind.
33
Page 38
Inspection Before Use
Check itemAction
No indicators are missing.Display all indicators and ensure that no
indicators are missing. (p. 74) If any of the
indicators are missing, send the instrument
for repair.
Operation check
This section introduces some of the operation checks. Periodical
calibration is necessary in order to ensure that this instrument
operates according to its specifi cations.
Check that the test leads are not broken.
1
Check methodAction
Regarding the continuity check,
deliberately short circuit the test
leads and then check the display.
Black
Red
Normal:
The value stabilizes at around 0.1
.
to 0.2
Abnormal:
A numeric value other than the
above appears.
Corrective action:
The test leads may be broken.
Replace with those specifi ed by our
company.
If the same phenomena persist
even after the test leads are
replaced, a malfunction may occur.
Halt inspection and then send the
instrument for repair.
34
Page 39
Inspection Before Use
Measure samples (such as battery, commercial power
2
supply, and resistor) of which values have already been
known, and check that the appropriate values appear.
Check methodAction
Example:
Perform the AC voltage
measurement to measure the
commercial power supply, and then
check the display.
Normal:
An already-known value appears.
(In this example, the commercial
voltage level should appear.)
Abnormal:
The measured value does not
appear.
The malfunction may occur.
Stop the inspection and do not use
the instrument.
1
2
3
4
Black
Check that the fuse is not broken.
3
Check methodAction
1. Remove the fuse from the
instrument. (p. 106)
2. Reattach the battery cover.
3. In the resistance measurement,
check the resistance of the
fuse. (Resistance measurement
(p. 44))
Red
Normal:
Fuse ratingResistance
630 mAApprox. 1.2
11 A0.1 or less
Abnormal:
If the value above is not obtained (the
value higher than that is displayed),
replace the fuse. (p. 106)
35
5
6
7
Appx.Ind.
Page 40
Inspection Before Use
Before measurements
Observe the following to avoid short circuit accidents.
• Always verify the appropriate setting of the rotary
switch before connecting the test leads.
• Disconnect the test leads from the measurement
object before switching the rotary switch.
• Operate or connect the instrument by following
the procedure of each measurement example (or
procedure steps).
WARNING
36
Page 41
Inspection Before Use
Auto power save function
• Before shipping (In the default setting), the auto power save
function is set to enabled. If the instrument has not been operated
for approx. 15 minutes, it enters the sleep mode. (When the
instrument is in the sleep mode, press any key or turn the rotary
switch to recover from the sleep mode.)
When the sleep mode continues for approx. 45 minutes, the
power turns off automatically.
• To recover from the power off state, set the rotary switch to OFF
and then turn on the power again.
• During current measurement, before setting the rotary switch to
OFF, disconnect the test leads. If the rotary switch is forcibly turned,
the shutter may be damaged.
• If the instrument will be used for an extended period of time, disable
the auto power save function. (p. 70)
• After use, set the rotary switch to OFF. The auto power save function
consumes a small amount of current.
1
2
3
4
Numerical value display with no input
When the measurement terminal is open during DC voltage (DCV)
measurement or AC voltage (ACV) measurement in the 60 mV range or
600 mV, a random value is displayed. This does not indicate a malfunction
of the instrument. When the probe is connected to the measurement target,
a normal numerical value is displayed. A high-input impedance voltmeter
is used in the instrument for highly sensitive measurement. Consequently,
external noise, such as inductive noise, appears as a numerical value.
37
5
6
7
Appx.Ind.
Page 42
Measuring Voltage
1
2
3
3.2 Measuring Voltage
AC voltage, DC voltage, synthesized voltage of DC and AC, and
voltage of AC and DC components can be measured. Furthermore,
the maximum, minimum, and peak (instantaneous) values of
measured values can be checked. (p. 61)
Before measurements
WARNING
If the instrument is used in locations where the rating
indicated on the instrument or probes is exceeded,
the instrument may be damaged resulting in personal
injury. Do not use the instrument in such locations.
See “Measurement categories” (p. 9).
The autoranging function of this instrument automatically selects the
optimum measurement range. To change the range arbitrarily, use
the manual range. (p. 55)
Measuring AC voltage
Measure the AC voltage.
1
Measure the frequency simultaneously.
The measured value is a true RMS.
(p. Appx.1)
3
38
Black
2
Red
Page 43
Measuring DC voltage
2
3
1
2
3
1
Measuring Voltage
Measure the DC voltage.
1
3
Black
2
Red
Measuring synthesized voltage of DC and AC
1
3
Black
2
Red
Measure the synthesized voltage of
DC and AC.
The measured value is a true RMS.
(p. Appx.1)
1
2
3
4
5
6
7
Appx.Ind.
It is also possible to check the AC or DC components individually. (p. 40)
39
Page 44
Measuring Frequencies
3
4
1
2
Measuring voltage of DC and AC components
2
1
Synthesized
voltage
4
Black
Red
3
DC
component
AC
component
Voltage of the DC
component
Voltage of the AC
component
3.3 Measuring Frequencies
During voltage/current measurement of AC, the frequency can
be checked in the sub display. It is not possible to change the
frequency range.
Frequency
• If signals out of the range of frequency measurement are measured,
[-----] appears. Be aware of it.
• The sensitivity of the frequency measurement is regulated by range.
(Minimum sensitivity voltage (p. 92), Minimum sensitivity current (p. 92))
When the value is less than the minimum sensitivity voltage (current),
the indicated value may fl uctuate. When the voltage (current) range is
lowered, the value stabilizes. This does not apply to cases where the value
fl uctuates due to noise.
• During the measurement of low frequency, if the auto range does not
stabilize and frequency cannot be measured, fi x the range and measure
again.
40
Page 45
Decibel Conversion (dBm/dBV)
3
4
1
2
3.4 Decibel Conversion ( dBm/dBV)
The result of the AC voltage measurement is decibel-converted
for the standard and then displayed. For details on the conversion
formula, see “Decibel conversion measurement” (p. 93).
dBmFor the voltage measurement, the “power ratio” for power 1 mW by
the standard resistance is decibel-converted and then displayed.
(Cable loss)
dBVFor the voltage measurement, the “voltage ratio” for the standard
voltage 1 V is decibel-converted and then displayed. (V
2
1
Black
3
Example: Measuring the [dBm] cable loss
Electric
telecommunication line
Amplifi er
Transmitter
Red
4
([dBm] or [dBV] lights up)
In the case of [dBm]
In the case of [dBV]
oltage gain)
Frequency
Frequency
1
2
3
4
5
6
Changing the standard impedance of the dBm conversion
Turn on the power while pressing
1
.
/ (Select a desired value.)
2
Standard impedance
selection range (p. 93)
3
The regular display reappears.
Even after the power is turned off, the setting
is retained.
(Confi rm the value.)
7
Appx.Ind.
41
Page 46
Checking Continuity
2
3
1
3.5 Checking Continuity
The input short circuit is detected and informed via a buzzer and red
backlight.
WARNING
Before measuring, be sure to turn off the power to the
measurement circuit. Otherwise, electric shock may
occur or the instrument may be damaged.
1
3
Black
Red
2
Changing thresholds
Use
/ to change a threshold. Even after the power is
turned off, the setting is retained.
ThresholdMeasurement result
Detection
Open
detection
Short
circuit
detection
20
(default)
220
or more
20
or less
50
250
or more
50
or less
100 500 Buzzer
300
or more
100
or less
600
or more
500
or less
Resistance
Threshold
Does not
sound
SoundsTurns on
Red
backlight
Turns off
42
Page 47
Measuring Diode
3
4
1
2
3.6 Measuring Diode
The forward voltage of the diode is measured. If the forward voltage
of the diode is the threshold or less, it is informed via a buzzer and
red backlight.
1
WARNING
Before measuring, be sure to turn off the power to the
measurement circuit. Otherwise, electric shock may
occur or the instrument may be damaged.
Forward voltage
2
1
Black
3
Changing thresholds
Use
Threshold: 0.15 V/0.5 V (default)/1 V/1.5 V/2 V/2.5 V/3.0 V
Even after the power is turned off, the setting is retained.
/ to change a threshold.
AnodeCathode
Red
4
Threshold
In the case of the opposite
connection
2
3
4
5
6
7
Appx.Ind.
43
Page 48
Measuring Resistance
2
3
1
3.7 Measuring Resistance
Resistance is measured.
WARNING
Before measuring, be sure to turn off the power to the
measurement circuit. Otherwise, electric shock may
occur or the instrument may be damaged.
1
3
Black
The open terminal voltage is approx. 2.5 V or less. The
measurement current (DC) varies depending on the range. (p. 86)
To avoid damage to the measurement object, check the
specifi cations before use.
Red
2
44
Page 49
Measuring Temperatures
1
2
3
3.8 Measuring Temperatures
Using our optional DT4910 Thermocouples (K), temperatures can
be measured.
CAUTION
To avoid damage to the instrument, do not input any
voltage or supply current to the thermocouple.
1
2
1
When a breaking state of
the Thermocouples (K) is
detected
3
4
2
3
Be careful to
observe the polarity
when connecting
the thermocouple.
DT4910
Checking the temperature change
Check the following values.
• Relative value display (p. 63)
• Maximum/Minimum value display
(p. 61)
Changing the temperature units
Celsius and Fahrenheit can be
switched. (p. 75)
5
6
7
Appx.Ind.
45
Page 50
Measuring Temperatures
When measuring temperatures with the thermocouple
applied to the surface of the measurement object
Clean the surface so that the thermocouple can make contact
with the object securely.
If no numeric value is displayed after the thermocouple is
attached ([OPEn] is displayed):
The instrument or thermocouple may be malfunctioning.
Check this with the following procedure.
Short-circuit the V and COM terminals of the instrument
1
using the test leads.
The ambient temperature
is displayed.
The ambient temperature
is not displayed.
Connect the thermocouple in the correct direction.
2
[OPEn] remains
displayed.
To step 2
The instrument is malfunctioning. Send
it for repair.
The thermocouple may be
malfunctioning (blown).
Replace the thermocouple with a new one.
46
Page 51
Measuring Electrostatic Capacities
2
3
1
3.9 Measuring Electrostatic Capacities
The capacity of the capacitor is measured.
WARNING
Before measuring, be sure to turn off the power to the
measurement circuit. Otherwise, electric shock may
occur or the instrument may be damaged.
Do not measure the capacitor which has been charged.
1
3
1
2
3
4
5
Black
2
• When measuring the polar capacitor
Connect the V terminal (red test lead) to the + terminal of
the capacitor and the COM terminal (black test lead) to the terminal.
• For components on a circuit board, measurement may not be
possible due to the effect of the peripheral circuit.
• If the capacity changes when the measurement range is
changed (p. Appx.3)
Red
6
7
Appx.Ind.
47
Page 52
Measuring Conductances (DT4282)
3
4
1
2
3.10 Measuring Conductances
The resistance is measured and the inverse number is displayed
(Unit: nS, nano-siemen). When the resistance is excessively large,
it is used.
Example: When the resistance is 50 M
(M = 106, n = 10-9)
When the terminal is open, 0 nS appears.
, 1/50 M = 20 nS.
(DT4282)
WARNING
Before measuring, be sure to turn off the power to the
measurement circuit. Otherwise, electric shock may
occur or the instrument may be damaged.
2
1
4
Black
Red
3
The open terminal voltage is approx. 2.5 V or less. The
measurement current (DC) is approx. 96 nA. (p. 87)
To avoid damage to the measurement object, check the
specifi cations before use.
48
Page 53
Measuring Current
3.11 Measuring Current
DC/AC is measured.
DANGER
• Do not input any voltage to the current
measurement terminals.
Doing so may result in short circuit accidents.
• To avoid electrical accidents, turn off the power to
the circuit before measuring and then connect the
test leads.
Measuring DC/AC
Function
μA Selected to measure 6000 A
mA Selected to measure 600 mA or less.
ASelected to measure 10 A or less.
(DT4282)
* The 0-20 mA input can be selected via the power-on option. (p. 77)
or less.
is used to toggle
between DC and AC.
→
*
→
1
2
3
4
5
6
When measuring an unknown current
Set to the high range (mA for the DT4281, A for the DT4282).
7
Appx.Ind.
49
Page 54
Measuring Current
3
4
1
2
3
4
1
2
DT4281, DT4282
1
2
Control board
4
Red
3
Example: Measuring the current of the burner fl ame (μA)
The measured current value of the burner fl ame varies with the
input impedance of the instrument.
The A input impedance of this instrument is approx. 100 .
DT4282
1
Power
supply
Black
3
Black
2
Load
4
Red
50
Page 55
Measuring AC Using Clamp-on Probe (DT4281)
1
2
5
3
9
Clamp-
e
C
r
accordance w
measuremen
o
c
3.12 Measuring AC Using Clamp-on
Probe (DT4281)
The current is measured using our optional clamp-on probe (901050, 9018-50, 9132-50). To connect to this instrument, the 9704
Conversion Adapter is required. Before using the clamp-on probe,
be sure to read the Instruction Manual which accompanies the
optional clamp.
1
3
Set the clamp-on probe and the instrument to the same range.
5
Current
range
2
9704 Conversion Adapter
704 Conversion Adapter
Clamp-on Probe
on Prob
4
Change the range in
hange the
accordance with the actual
measurement.
If the range of the clamp-
If the range
on probe is changed during
on probe is
measurement, change the
range of the instrument as
well.
1
2
3
4
5
6
7
Appx.Ind.
51
Page 56
Measuring AC Using Clamp-on Probe (DT4281)
When clamping a cable
Attach the clamp around only one conductor.
Single-phase (2-wire) or three-phase (3-wire) cables clamped
together will not produce any reading.
OK
When blinks
The measured value exceeds maximum display counts. Increase
the range.
NONO
52
Page 57
4-20 mA (0-20 mA) % Conversion
3
4
1
2
3.13 4-20 mA (0-20 mA) % Conversion
The 4-20 mA (or 0-20 mA) signal of the instrumentation system can
be converted to 0% to 100% and checked.
• 4 mA - 20 mA 0% - 100%
(An input exceeding 20 mA is displayed up to a maximum of 350%.)
• 0 mA - 20 mA 0% - 100%
(An input exceeding 20 mA is displayed up to a maximum of 300%.)
2
1
3
Changing to the 0-20 mA signal
Sensor
Black
Two-wire
transmitter
Red
4-20 mA
24 V
4
Distributor
Turn on the power while
1
pressing .
Output signal
1
2
3
4
5
6
/ (Select a desired
2
value.)
3
The regular display reappears.
Even after the power is turned off,
the setting is retained.
(Confi rm the value.)
7
Appx.Ind.
53
Page 58
Using Instrument
4
Conveniently
4.1 Selecting the Measurement Range
Auto or Manual range can be selected. In the case of measurement
where the desired range can be selected, [RANGE:] lights up in the
bottom left of the display.
• Auto rangeSets the optimum range automatically in
accordance with the actual measurement.
• Manual rangeSets the range and fi xes it specifi cally.
Measuring with the auto range
When the manual range is set, pressing
for at least 1 second changes to the auto range.
The instrument automatically selects the
optimum measurement range.
Measuring with the manual range
Press .
Each time the key is pressed, a higher range is
specifi ed.
When the key is pressed at the highest range,
the lowest range is specifi ed once again.
Example: When measuring the AC voltage
60 mV 600 mV 600 V 1000 V
When the measurement function is switched using the rotary switch or
SHIFT, or when the peak measurement is canceled, the auto range is
enabled.
1
2
3
4
5
6
7
Appx.Ind.
55
Page 59
Selecting the Measurement Range
Range display list
ACV, DCV60 mV, 600 mV, 6 V, 60 V, 600 V, 1000 V
DC+ACV6 V, 60 V, 600 V, 1000 V
Ω
(Electrostatic
capacity)
DCA, ACA600 A, 6000 A, 60 mA, 600 mA, 6 A, 10 A
AC clamp10 A, 20 A, 50 A, 100 A, 200 A, 500 A, 1000 A
PEAK (DCV)6 V, 60 V, 600 V, 1000 V
PEAK (ACV)18 V, 180 V, 1500 V
PEAK (DC+ACV)18 V, 180 V, 1500 V
PEAK (DCA/ACA) 1200 A, 12000 A, 120 mA, 1200 mA, 12 A, 15 A
PEAK (AC clamp) 30 A, 60 A, 150 A, 300 A, 600 A, 1500 A, 3000 A
The measured value is retained manually or automatically.
• ManuallyWhen HOLD is pressed, the measured value is
retained.
Automatically When HOLD is pressed and held for at least 1
•
second, auto mode starts. When the measured
value stabilizes, it is retained.
1
2
Retaining the measured value manually ( HOLD)
To retain the measured value,
press .
measurement value is retained.)
To cancel the hold state, press it again.
(
( lights up and the
goes off.)
Automatically retaining the measured value when
the value stabilizes ( AUTO HOLD)
Press for at least 1 second.
( lights up.)
When the measured value stabilizes, a beeping
sound is generated and the value is retained.
lights up.)
(
If HOLD is pressed again, or the input signal
exceeds the dead zone threshold again (see
table on the next page), or when the range is
switched internally and the measured value
is stabilized once again, the hold state is
canceled. (
To disable the auto hold function, press it for at
least 1 second again. (
goes off.)
goes off.)
57
3
4
5
6
7
Appx.Ind.
Page 61
Retaining the Measured Value
• If the input signal is too small for the relevant range (dead zone
threshold p. 58), the measured value cannot be automatically
retained.
• If the difference between the current and previous measured
values remains less than the counts in the stable fi eld of the
table below, the instrument determines the current value has
stabilized.
Dead zone threshold
FunctionStable (count)
AC voltage
DC voltage
DC voltage + AC voltage
Continuity check1005900
Diode test803520
Resistance
Conductance12001200
A12001200
DC
AC
AC (clamp input)5% of the range5% of the range
* No function is available for the mV range.
mA12001200
A
A12001200
mA12001200
A
1200, 200
(1000 V range)
1200, 200
(1000 V range)
1200, 200
(1000 V range)
1200, 120
(60/600M
1200, 200
(10 A range)
1200, 200
(10 A range)
range)
Dead zone threshold
(count)
1200, 200
(1000 V range) *
1200, 200
(1000 V range) *
1200, 200
(1000 V range)
58800, 5880
(60/600M range)
1200, 200
(10 A range)
1200, 200
(10 A range)
58
Page 62
When the Measured Value Fluctuates (SLOW)
4.3 When the Measured Value
Fluctuates ( SLOW)
When the measured value fl uctuates and cannot be read, it is
possible to stabilize the value. (When is lit)
• When
• When
is off (normal): refer to the display update rate
(p.80) (Default setting)
is lit (slow):fi ve times normal (average of 5 times)
1
2
Press for at least 1 second.
( lights up.)
To cancel the state, press it for at least 1
second again.
goes off.)
(
When the measurement function is switched using the rotary
switch or SHIFT, the SLOW setting is disabled.
3
4
5
6
7
Appx.Ind.
59
Page 63
Removing the Harmonic Components of the Inverter (FILTER)
4.4 Removing the Harmonic
Components of the Inverter
(FILTER)
WARNING
To prevent electric shock, before setting the fi lter
function, confi rm if there is a voltage or not with the fi lter function disabled (OFF).
Remove the harmonic components when measuring the inverter
output.
This function can be used when measuring the AC voltage or
AC+DC voltage. The cut-off frequency is 630 Hz.
Press for at least 1 second.
( lights up.)
To cancel the state, press it for at least 1
second again.
goes off.)
(
Only 600V/1000V range.
When the measurement function is switched using the rotary
switch or SHIFT, the fi lter function is canceled.
When measuring voltage with a 400 Hz
fundamental frequency (such as that on an
aircraft), be aware that the FILTER function
may cause attenuation and the indicated
voltage may be 20% lower than the actual
voltage.
The characteristics of the FILTER function
(low pass fi lter) are shown in the graph.
60
0dB
-10dB
-20dB
-30dB
-40dB
10Hz100Hz 1000Hz 10000Hz
Page 64
Checking the Maximum/Minimum Value (MAX/MIN)
4.5 Checking the Maximum/Minimum
Value ( MAX/MIN)
The maximum/minimum value after start of the measurement can
be checked.
1
Press .
Each time the key is pressed, the display is
changed in the order of the maximum value
(MAX), minimum value (MIN), and current
value.
A buzzer sounds when the maximum value is
updated with MAX displayed or the minimum
value is updated with MIN displayed.
The maximum and minimum values are for the displayed value;
they do not relate to peak values such as AC signals.
The main and sub displays update their maximum and minimum
values individually.
Clearing the existing maximum/minimum value
When MAX or MIN appears, press
Both maximum and minimum values are cleared.
After this, the maximum and minimum measurement values after clearing the
previous ones are retained.
When saving to memory, only the value displayed on the screen (current
value, maximum value, or minimum value) can be saved.
.
2
3
4
5
6
7
Appx.Ind.
61
Page 65
Checking the Peak Value (V • A PEAK)
4.6 Checking the Peak Value (V • A
PEAK)
After starting the voltage/current peak measurement, the maximum/
minimum instantaneous value can be checked.
The peak measurement can be performed only with the manual
range. Press RANGE to select the appropriate range before
starting the measurement. (For the range confi guration and
display range, see page 84)
Maximum
instantaneous
value
Minimum
instantaneous
value
A buzzer sounds when PEAKmax or PEAKmin is updated.
Press .
The maximum/minimum
instantaneous value can be
retained, from the point at which
is pressed.
To cancel the display of the peak
value, press it again.
Clearing the maximum/minimum instantaneous value
Press .
The maximum/minimum instantaneous value can be displayed, from the point
at which
is pressed.
62
Page 66
Checking the Relative Value/Performing Zero Adjustment
4.7 Checking the Relative Value/
Performing Zero Adjustment
The relative value comparing to the standard value can be checked
(relative function).
It can also be used as the zero adjustment function.
Zero adjustment eliminates the infl uences of the wiring resistance
(resistance measurement) and the wiring capacity (capacitor
measurement).
When the following measurement function is selected, this function
is disabled.
Peak measurement, Continuity, Diode, dBm/dBV, 4-20 mA
The measurement range cannot be changed when [REL] is lit. To
change the range, press REL for at least 1 second to reset the
REL function.
1
2
3
4
Checking the relative value ( REL)
Example 1: DC voltage measurement
When the standard value is measured,
press
lights up)
(
To cancel the state, press it for at least 1
second again. (
Example 2: Temperature measurement
When measuring temperature, [REF] lights up
and a standard temperature appears in the
sub display.
Deviation from the standard temperature
Standard temperature
for at least 1 second
.
goes off.)
63
5
6
7
Appx.Ind.
Page 67
Checking the Relative Value/Performing Zero Adjustment
1
2
3
1
2
3
Performing zero adjustment
When performing zero adjustment, the condition of the test leads
varies depending on the measurement function.
Perform zero adjustment, referring to the table below.
Measurement functionV, A,
Condition of the test leads Short circuitOpen
1
4
3
2
RedBlack
1
4
2
Black
Red
3
Example 1: Resistance
measurement
Select the measurement
1
function.
Connect the test leads to
2
the measurement terminals.
Allow the test leads to
3
short circuit.
Press for at least
4
1 second.
(After zero adjustment: 0.000 )
Measure the resistance.
5
Example 2: Capacitor
measurement
Select the measurement
1
function.
Connect the test leads to
2
the measurement terminals.
Allow the test leads to
3
open.
Press for at least
4
1 second.
(After zero adjustment: 0.000 nF)
Measure the capacitor.
5
, nS
64
Page 68
Using the Memory Function
4.8 Using the Memory Function
The measurement result can be saved and read using the memory
function. Up to 400 data can be saved. Data can also be deleted
arbitrarily.
The saved data can be imported to the PC using the DT4900-01
Communication Package (optional).
Saving the measured value (MEM)
Press ( lights up).
1
Memory mode starts.
Memory No.
This function applies until the power is turned
off.
The memory No. appears (only for the
measurement which can be saved).
1
2
3
4
Select the memory No. to be
2
saved using / .
When the key is pressed and held,
the memory No. can be incremented/
decremented faster.
Press to save the
3
measured value.
When the measured value is saved, the
memory No. for which the value is to be
saved next appears.
Even when the power is turned off, the
memory data are retained.
5
6
7
Appx.Ind.
65
Page 69
Using the Memory Function
• If the memory No. for which the data have already been saved is
selected, [
• When MEM is pressed, the data are overwritten.
When the value is saved with memory No. “400” displayed, a buzzer
•
sounds and “FULL” appears in the display, and then “001” reappears.
• When [
value.
• The current value, maximum value (MAX), minimum value (MIN),
instantaneous maximum value (Peak max), and instantaneous minimum
value (Peak min) cannot be saved simultaneously. Only the displayed
numeric value can be saved.
• The continuity check data and the diode test data cannot be saved.
] appears.
] (data reading) is displayed, it is not possible to save the
When both hands cannot be used during measurement
Enable the auto hold function. After the measured value is retained
automatically, press MEM to save the data.
Items to be saved per data
• Function
• Measurement range
• Measured value in the main
display
• Measured value in the sub
display
• Whether or not the REL function
is executed
• Filter ON/OFF
66
Page 70
Reading the memory data (READ)
Using the Memory Function
Memory No.
Press ( lights up).
1
Read mode starts.
Select the desired memory No.
2
using / .
Only the memory Nos. corresponding to saved
data are displayed.
Canceling the read mode
Press READ again or turn the rotary switch.
goes off.)
(
When no memory data exist
The relevant message appears and then the
regular measurement display reappears.
1
2
3
4
5
6
7
Appx.Ind.
67
Page 71
Using the Memory Function
Clearing the memory data (CLEAR)
Memory No.
Press ( lights up).
1
Read mode starts.
Select the No. to be cleared using
2
/ .
Press .
3
Clearing all memory data
Turn on the power while pressing
1
.
The relevant message appears in the display.
Press again within 3
2
seconds.
All memory data are now cleared.
68
When 3 seconds elapse without pressing
CLEAR, all memory data are not cleared
and the regular display reappears.
Page 72
Muting the Buzzer
4.9 Muting the Buzzer
The buzzer sound can be disabled.
Note, however, that the buzzer cannot be muted in the following cases.
Continuity check, diode test, and overload warning (only for the
maximum value)
1
Turn on the power while pressing
When is released, the regular display appears ( goes off).
Canceling the buzzer mute function
Turn on the power again while pressing
When is released, the regular display appears ( lights up).
Even after the power is turned off, the buzzer setting is retained.
.
.
4.10 Turning On the Backlight
The backlight can be turned on/off by pressing .
When PC communication is started, the backlight is turned off forcibly.
Enabling/disabling the automatic deactivation of the backlight
Turn on the power while pressing .
Even after the power is turned off, the automatic backlight deactivation setting
is retained.
Automatic deactivation
Enable
Disable
• When the automatic deactivation setting
is enabled (on), the backlight goes off
automatically, approximately 40 seconds after
the backlight comes on.
• When the automatic deactivation setting is
cancelled (oFF) and the auto power save
function is disabled, if the instrument has not
been operated for 3 minutes, the backlight
automatically goes off.
2
3
4
5
6
7
Appx.Ind.
69
Page 73
Using the Auto Power Save (APS)
4.11 Using the Auto Power Save (APS)
The auto power save function saves on power consumption. If the
instrument has not been operated for approx. 15 minutes, it enters
the sleep mode. (When the instrument is in the sleep mode, press
any key or turn the rotary switch to recover from the sleep mode.)
When the sleep mode continues for approx. 45 minutes, the power
turns off automatically.
Before shipping (as one of the default settings), the auto power
save function is set to enabled. (
It is also possible to disable the auto power save function.
At 15 seconds before the power is turned off, the APS blinks and
an intermittent buzzer sounds. To continuously use the instrument,
press any key or turn the rotary switch.
After use, set the rotary switch to OFF. The auto power save
function consumes a small amount of current.
Recovering from the power off state
Set the rotary switch to OFF and turn on the power again.
(When the current measurement is used, disconnect the measurement
cable and then turn on the power again.)
Disabling the auto power save function
Turn on the power while pressing
lights up)
70
.
When is released, the regular display
appears.
(APS goes off.)
This function is disabled until the power is turned
off.
Page 74
Using Plus/Minus Judgment Function for Measurement Value
4.12 Using Plus/Minus Judgment
Function for Measurement Value
If the measured DC voltage falls below the following standard value,
a buzzer sounds and the backlight lights up red.
This function is useful for checking for any incorrect connection of
the DC power.
Standard value: -10 V or less
Measurement function: DCV
“Enabling/disabling the Plus/Minus judgment function” (p. 78)
1
2
3
4
5
6
71
7
Appx.Ind.
Page 75
Communicating with PC
4.13 Communicating with PC
Using the optional DT4900-01 Communication Package, it is
possible to transmit data to the PC or to control the instrument.
For details, see the Instruction Manual which accompanies the
communication package.
Install the special software on the PC.
(See the Instruction Manual which accompanies with the
communication package.)
Dedicated Software (p.Appx.4)
Attaching the USB cable to the instrument (p. 73)
Connect to the PC.
The virtual COM ports of the PC can be used as the USB interface.
• Communication method: Start-stop system, half-duplex transmission
• Baud rate: 19,200 bps fi xed • Parity:None
• Data bit length: 8 bits • Flow control: None
• Stop bit:1 bit • Delimiter:CR+LF
72
Transmits data.
Controls the
instrument.
Page 76
Communicating with PC
3
4, 5
6
Attaching the communication adapter to the instrument
Using a Phillips
USB cable
6
Communication
3
adapter
Communication
port
1, 2
4, 5
1
screwdriver, remove
the screws (2
locations) from the
battery cover.
Remove the battery
2
cover.
Attach the
3
communication
adapter.
Reattach the battery
4
cover.
Secure the battery
5
cover with the screws.
Connect the
6
USB cable to the
communication
adapter.
1
2
3
4
5
• Connect the cables, being careful to orient each cable correctly.
• During communication, appears in the display.
• When is lit, the operation keys of the instrument is
disabled.
• During communication, do not disconnect the USB cable.
Disconnecting the cable stops the communication. In that case,
a warning is displayed by the PC software. Connect the cable
again.
• It is possible to use the instrument while the communication
adapter is attached, however, the communication adapter is
excluded from the drop-proof.
6
7
Appx.Ind.
73
Page 77
Setting and Checking the System
4.14 Setting and Checking the System
Checking that all indicators are displayed
Check that there is no missing indicator.
If any indicator is missing, stop using the instrument and send it for
repair.
Turn on the power while pressing
.
When is released, the regular display
appears.
Checking the software version of the instrument
For repair or calibration, check the version if necessary.
Example: Ver 1.00
Turn on the power while pressing
.
74
When is released, the regular display
appears.
Page 78
Setting and Checking the System
Changing the temperature display unit
The units of temperature (°C or °F) can be changed.
Turn on the power while pressing
1
and simultaneously.
Press for at least 3
2
(Display: tEMP)
The setting of the temperature unit is retained even after the
power is turned off.
seconds.
Change the unit of the temperature
3
using
Press to determine the
4
setting.
After a few moments, the regular display
reappears.
.
1
2
3
4
5
6
7
Appx.Ind.
75
Page 79
Resetting the System
4.15 Resetting the System
The system is returned to the state before shipping. The power-on
option setting is also reset.
The memory data are not reset. To clear the memory data, clear
the data individually for each memory No. (p. 68) or clear all data
(p. 68).
The temperature unit is not reset, either.
Turn on the power while pressing
1
and .
Within 3 seconds, press
2
again.
After [CLr] blinks, the system returns to the
state before shipping.
If the instrument is not operated for 3 seconds,
the system is not reset but the regular display
reappears.
Table of default settings
Filter functionDisabled
Relative functionDisabled
Display updateNormal
Display holdDisabled
RangeAUTO[
Auto holdDisabled
Maximum/Minimum valueDisabled
Peak valueDisabled
Auto power save functionEnabled[
Automatic backlight deactivationOff
BuzzerEnabled[
% conversion4-20 mA
dBm standard impedance600
Temperature display unit°C
Plus/Minus Judgment Function
for Measurement Value
76
Setting itemDefault settingDisplay
] is lit.
] is lit.
] is lit.
Disabled
Page 80
Power-on Option Table
4.16 Power-on Option Table
The settings in the instrument can be changed or checked.
After changing the setting, the regular display then reappears.
+
Canceling the
auto power
save function
(APS)
Buzzer
(ON/OFF)
Automatic
backlight
deactivation
setting
Turn on the power while pressing the operation key.
Enabling/
disabling the
Plus/Minus
judgment
function
MethodRef.
Select the
desired setting.
Select the
desired
setting.
When is released, the regular display appears.
Confi rm the
change.
Serial No. display (9 digits)
1st to 4th digits: Main display
5th to 9th digits: Sub display
Confi rm the
change.
3 seconds
or more
p. 53
p. 74
p. 74
p. 75
p.71
78
Page 82
Specifi cations
5
5.1 Electrical Characteristics
AC measurement
system
Noise
suppression
Response time
True RMS measurement
• NMRR DCV, -60 dB or less (50 Hz/60 Hz)
• CMRR DCV, -120 dB or less
• CMRR ACV, -60 dB or less
• Power ON time: within 2 seconds
• DCV Within 1
• ACV Within 3
•
•
*1: Until the value comes in the range of the accuracy
*2: Until the displayed value stabilizes (within ±2
(DC/50 Hz/60 Hz, 1 k
(DC/50 Hz/60 Hz, 1 k
(When the range does not move until the measured
value is displayed in the screen)
*1*2
second
seconds
Within 2
seconds
Within 2
seconds
specifi cation
counts)
Within 1.5
seconds
Within 8
seconds
Within 3
seconds
Within 3
seconds
unbalance)
unbalance)
(0 V 100 V auto
range operation)
(0 V 100 V auto
range operation)
(Infi nity 0 auto
range operation)
(0 30 M auto
range operation)
1
2
3
4
5
6
7
Appx.Ind.
79
Page 83
Electrical Characteristics
Display update
rate
• V, Continuity,
Frequency, A, dBm, dBV: 5 times/s
• Electrostatic
capacity
nS, Diodo, Clamp,
,
0.05 to 5 times/s (varies
depending on the measured
value)
• Temperature1 time/s (including disconnection
check)
• Peak measurement Up to 155 times/s when peak
value is updated
• DC+ACV2.5 times/s
Input impedance See “Accuracy Table” (p. 81).
Overload protection
FunctionOverload protection
• ACV, dBm, dBV, Hz,
AC clamp (DT4281)
• DCV
• DC+ACV, Hz
• Continuity check,
diode test
, temperature (K)
•
• Electrostatic capacity
• nS (DT4282)
1100 V DC/AC (50/60 Hz) or 2 × 10
whichever is the lower value
• Voltage applied for 1 minute
• Transient overvoltage 8000 V
1000 V DC/AC or 2 × 10
the lower value
• Voltage applied for 1 minute
• Current steady state under overload 15 mA
or less
• Current transient state under overload 0.8 A
or less
•DCA, ACA, DCmA,
ACmA, (4-20mA)%, Hz
• DCA (DT4282),
ACA (DT4282), Hz
630 mA/1000 V fuse
Breaking capacity 50 kA AC/30 kA DC
11 A/1000 V fuse
Breaking capacity 50 kA AC/30 kA DC
7
V • Hz,
7
V • Hz, whichever is
80
Page 84
5.2 Accuracy Table
Accuracy Table
Guaranteed accuracy period1 year
Regulated power supply range Until the power shutdown (3.8 V ± 0.2 V)
Temperature and humidity for
guaranteed accuracy
Temperature characteristicAdds “Measurement accuracy × 0.1/°C”
Comply with the temperature characteristic in the applicable accuracy table if
individually specifi ed.
Other conditions: For information related to the L4931 extension cable set
(coupled 2 cables, 3 m), see the accuracy table.
• rdg. (reading or displayed value): The value currently being measured and
displayed on the measuring instrument.
• dgt. (resolution): The smallest displayable unit, i.e., the input value that
causes the digital display to show a “1”.
AC voltage (V AC, mV AC)
1
Range
60.000 mV
600.00 mV
6.0000 V
60.000 V
600.00 V
1000.0 V
20 to
below 45
[Hz]
±1.3% rdg.
±60 dgt.
±1.3% rdg.
±60 dgt.
±1% rdg.
±60 dgt.
Not
specifi ed
Not
specifi ed
Not
specifi ed
45 to 65
[Hz]
±0.4% rdg.
±40 dgt.
±0.4% rdg.
±40 dgt.
±0.2% rdg.
±25 dgt.
±0.2% rdg.
±25 dgt.
±0.2% rdg.
±25 dgt.
±0.2% rdg.
±25 dgt.
23°C ± 5°C (73°F ± 9°F), 80%RH or less
(non-condensating)
(except 23°C ± 5°C (73°F ± 9°F))
Accuracy
Over 65
to 1 k
±0.6% rdg.
±40 dgt.
±0.6% rdg.
±40 dgt.
±0.3% rdg.
±25 dgt.
±0.3% rdg.
±25 dgt.
±0.3% rdg.
±25 dgt.
±0.3% rdg.
±25 dgt.
[Hz]
Over 1 k
to 10 k
[Hz]
±0.9% rdg.
±40 dgt.
±0.9% rdg.
±40 dgt.
±0.4% rdg.
±25 dgt.
±0.4% rdg.
±25 dgt.
±0.4% rdg.
±25 dgt.
±0.4% rdg.
±25 dgt.
Over 10 k
to 20 k
[Hz]
±1.5% rdg.
±40 dgt.
±1.5% rdg.
±40 dgt.
±0.7% rdg.
±40 dgt.
±0.7% rdg.
±40 dgt.
Not
specifi ed
Not
specifi ed
Over 20 k
to 100 k
[Hz]
±20% rdg.
±80 dgt.
±8% rdg.
±80 dgt.
±3.5% rdg.
±40 dgt.
±3.5% rdg.
±40 dgt.
Not
specifi ed
Not
specifi ed
81
1
2
3
4
5
6
7
Appx.Ind.
Page 85
Accuracy Table
Input impedance1 M ±4%, 100 pF or less
Crest factor3 or less
Accuracy
specifi cation range
DC superimposition • For the superimposition on 200 V DC or less, see
DC voltage (V DC, mV DC)
2
RangeAccuracyInput impedance
60.000 mV±0.2% rdg. ±25 dgt.
600.00 mV±0.025% rdg. ±5 dgt.
6.0000 V±0.025% rdg. ±2 dgt.11.0M
60.000 V±0.025% rdg. ±2 dgt.10.3M ±2%: 100 pF or less
600.00 V±0.03% rdg. ±2 dgt.10.2M
1000.0 V±0.03% rdg. ±2 dgt.10.2M
*1: After leaving the instrument in an environment where the ambient air
temperature has been stable for at least 30 minutes, allow the input to short
circuit and activate the relative value display function (REL). The accuracy is
then specifi ed. After activating the REL, the temperature fl uctuates within ±5°C.
(For the 60.000 mV/ 600.00 mV/ 1000.0 V range, 1.5
or less for 100% of the input of the range
3 or less for 50% of the input of the range)
5% or more of each range (less than 5% of each
range is outside of the guaranteed accuracy)
With the fi lter ON, the accuracy is specifi ed in 100
Hz or less. Furthermore, 2% rdg. is added.
Frequency range: 20 Hz to 100 kHz (A measured
value outside of the accuracy guarantee range for
frequency is also displayed.)
the accuracy table.
• For the superimposition on over 200 V DC to 500
V DC or less, 2% rdg. is added to the accuracy
specifi cation in 1 kHz or less.
• For the superimposition on over 500 V DC, 10%
rdg. is added to the accuracy specifi cation in 1
kHz or less.
*1
1G or more: 100 pF or less
*1
1G or more: 100 pF or less
±2%: 100 pF or less
±2%: 100 pF or less
±2%: 100 pF or less
82
Page 86
DC+ACV measurement
3
Range
6.0000 V
60.000 V
600.00 V
1000.0 V
20 to
below 45
[Hz]
±1.2% rdg.
±65 dgt.
Not
specifi ed
Not
specifi ed
Not
specifi ed
45 to 65
[Hz]
±0.3% rdg.
±30 dgt.
±0.3% rdg.
±30 dgt.
±0.3% rdg.
±30 dgt.
±0.3% rdg.
±30 dgt.
Accuracy
Over 65
to 1 k
[Hz]
±0.4% rdg.
±30 dgt.
±0.4% rdg.
±30 dgt.
±0.4% rdg.
±30 dgt.
±0.4% rdg.
±30 dgt.
Over 1 k
to 10 k
[Hz]
±0.4% rdg.
±30 dgt.
±0.4% rdg.
±30 dgt.
±0.4% rdg.
±30 dgt.
±0.4% rdg.
±45 dgt.
Accuracy Table
Over 10 k
to 20 k
[Hz]
±1.5% rdg.
±45 dgt.
±1.5% rdg.
±45 dgt.
Not
specifi ed
Not
specifi ed
Over 20 k
to 100 k
[Hz]
±3.5% rdg.
±125 dgt.
±3.5% rdg.
±125 dgt.
Not
specifi ed
Not
specifi ed
1
2
3
Input impedance1 M ±4%, 100 pF or less
Crest factor3 or less
Auto rangeThe range moves depending on the DC+AC
Accuracy
specifi cation range
Accuracy of DCV (main display) and ACV (sub display) when SHIFT is
pressed during DC+ACV measurement
DCV (main display): the accuracy of 45 Hz to 65 Hz in the accuracy
1.
table is applied. Note however that 2%rdg. should be added when there
is AC voltage superimposition less than 45 Hz.
2. ACV (sub display): according to the accuracy table
(For the 1000.0 V range, 1.5 or less for 100% of the
input of the range, 3 or less for 50% of the input of
the range)
calculation result.
5% or more of each range (less than 5% of each
range is outside of the guaranteed accuracy)
With the fi lter ON, the accuracy is specifi ed in 100
Hz or less. Furthermore, 2% rdg. is added.
Frequency range: 20 Hz to 100 kHz (A measured
values outside of the accuracy guarantee range for
frequency is also displayed.)
4
5
6
7
Appx.Ind.
83
Page 87
Accuracy Table
Peak measurement
4
(At the time of ACV, DCV, DC+ACV, Clamp, DCA, DCmA, DCA, ACA,
ACmA, ACA)
Main
measurement
DCV4 ms or more (single)±2.0% rdg. ±40 dgt.
Other than
DCV
Signal widthAccuracy
1 ms or more (repeated)±2.0% rdg. ±100 dgt.
1 ms or more (single)±2.0% rdg. ±40 dgt.
250 s or more (repeated) ±2.0% rdg. ±100 dgt.
*1
*2
*3, *4
*4, *5
*1: The accuracy is specifi ed via the 5 V/4 ms single signal.
*2: The accuracy is specifi ed at the peak value of 40000 counts/25 Hz sine
wave.
*3: The accuracy is specifi ed via the 5 V/1 ms single signal. (ACV, DC+ACV)
*4: The accuracy is specifi ed at the peak value of the Max. counts/100 Hz
sine wave in the main measurement range. (AC measurement)
*5: The accuracy is specifi ed at the peak value of 40000 counts/100 Hz sine
wave in the main measurement range. (DC measurement, 7 A/100 Hz for
the 10 A range)
Maximum input range
Voltage
measurement
range
Peak measurement range
DCVACVDC+ACV
6.0000 V6.000 V18.000 V18.000 V
60.000 V60.00 V180.00 V180.00 V
*1
600.00 V600.0 V1500.0 V
1500.0 V
1000.0 V1000 VNoneNone
*1: Up to 1000 V for the RMS
84
*1
Page 88
Accuracy Table
Current
measurement
range
600.00 A1200.0 A
6000.0 A12000 A
60.000 mA120.00 mA
600.00 mA1200.0 mA
6.0000 A (DT4282) 12.000 A
10.000 A (DT4282) 15.00 A
AC clamp
measurement
range (DT4281)
10.00 A30.00 A
20.00 A60.00 A
50.00 A150.0 A
100.0 A300.0 A
200.0 A600.0 A
500.0 A1500 A
1000 A3000 A
Continuity check
5
RangeAccuracyMeasurement current
600.0
Open circuit voltage2.5 V DC or less
Continuity threshold20
Threshold setting tolerance ±1% setting±0.5
Response timeOpen circuit or short circuit is detected for at
±0.5% rdg. ±5 dgt.640 A ±10%
Peak measurement range
DCA/ACA
*2
*3
*3
Peak measurement range
(default) / 50 / 100 / 500
least 10 ms.
*2: Up to 600 mA for the
RMS
*3: Up to 10 A for the
RMS
1
2
3
4
5
6
7
Appx.Ind.
85
Page 89
Accuracy Table
Diode test
6
RangeAccuracyMeasurement current
3.600 V±0.1% rdg. ±5 dgt.1.2 mA or less
Open circuit voltage4.5 V DC or less
If the reading is lower than the threshold during the forward connection, a
buzzer sounds and red backlight comes on.
When the reading exceeds the threshold by 0.01 V or more during the
forward connection, the buzzer stops and the red backlight turns off.
*1: Allow the input to short circuit and activate the relative value display
function (REL). The accuracy is then specifi ed.
*2: The accuracy is specifi ed for humidity 60%RH or less.
*3: The accuracy is specifi ed for 200.00 M
*4: The accuracy is specifi ed for more than 200.00 M
or less.
.
Open circuit voltage2.5 V DC or less
Maximum capacity load 100 mF
Maximum inductive load 10 H
86
Page 90
Accuracy Table
Conductance (nS)
8
RangeAccuracyMeasurement current
600.00 nS±1.5% rdg. ±10 dgt.96 nA ±10%
• The accuracy is specifi ed for humidity 60%RH or less.
• For more than 300 nS, ±20 dgt. is added.
• The accuracy is specifi ed in 20.00 nS or more.
Open circuit voltage2.5 V DC or less
Maximum capacity load 100 mF
Maximum inductive load 10 H
Electrostatic capacity
9
RangeAccuracy
1.000 nF±1% rdg. ±20 dgt.
10.00 nF±1% rdg. ±5 dgt.
100.0 nF±1% rdg. ±5 dgt.
1.000 F±1% rdg. ±5 dgt.32 A ±10%2.5 V DC or less
10.00 F±2% rdg. ±5 dgt.
100.0 F±2% rdg. ±5 dgt.
1.000 mF ±2% rdg. ±5 dgt.
10.00 mF ±2% rdg. ±5 dgt. *3680 A ±20%2.1 V DC or less
100.0 mF ±2% rdg. ±20 dgt.
*1: In the case of the 100 nF range or less, the accuracy is specifi ed after the
REL function is activated.
*2: The accuracy is specifi ed in 0.22 nF or more.
*3: Adds “Measurement accuracy × 0.3/°C” (except 23°C ± 5°C)
• When the manual range is set, the accuracy is specifi ed in 1% or more of
the range.
• The accuracy is specifi ed for series resistance of 5
• Maximum count for each range: 1100 (excluding 100.0 mF)
*2
Measurement
*1
*1
*1
*3
*3
*3
*3
current
32 A ±10%2.5 V DC or less
32 A ±10%2.5 V DC or less
32 A ±10%2.5 V DC or less
680 A ±20%3.1 V DC or less
680 A ±20%3.1 V DC or less
680 A ±20%2.1 V DC or less
680 A ±20%2.1 V DC or less
or less.
Open circuit
voltage
87
1
2
3
4
5
6
7
Appx.Ind.
Page 91
Accuracy Table
Temperature (type K thermocouple)
10
RangeAccuracy
-40.0 to 800.0°C±0.5% rdg. ±3°C
-40.0 to 1472.0°F
*1: In an environment where the temperature of the instrument is ±1°C and
stable, the accuracy is specifi ed.
*2: Using the power-on option setting, °F display is activated.
Fahrenheit °F = (9/5) × Celsius°C + 32
• The optional type K thermocouple are used.
• The accuracy does not include the error of the type K thermocouple.
• Standard contact temperature compensation stability time
When the instrument environmental temperature fl uctuates ±5°C or more:
120 minutes
After current measurement: 30 minutes
DC (DCμA, DCmA, DCA)
11
DCA
DCmA
DCA
(DT4282)
*1: Accuracy rule when using the slow display update rate. Add ±20 dgt. when
using the normal rate.
*2
±0.5% rdg. ±5.4°F
RangeAccuracyShunt
600.00 A±0.05% rdg. ±5 dgt.*1101
6000.0 A±0.05% rdg. ±5 dgt.101
60.000 mA ±0.05% rdg. ±5 dgt.
600.00 mA ±0.15% rdg. ±5 dgt.1
6.0000 A±0.2% rdg. ±5 dgt.
10.000 A±0.2% rdg. ±5 dgt.10m
*1
Fuse
resistance
*1
1
Approx.
1.2
*1
10m
0.1 or
less
88
Page 92
Accuracy Table
AC (ACμA, ACmA, ACA)
12
*1
[Hz]
Accuracy
Over 65
to 1 k
[Hz]
±0.6% rdg.
±20 dgt.
±0.6% rdg.
±5 dgt.
±0.6% rdg.
±20 dgt.
±0.6% rdg.
±5 dgt.
±0.8% rdg.
±20 dgt.
±0.8% rdg.
±5 dgt.
Approx. 1.2
Approx. 0.1
Over 1 k
to 10 k
[Hz]
±2% rdg.
±20 dgt.
±2% rdg.
±5 dgt.
±1% rdg.
±20 dgt.
±1.5% rdg.
±10 dgt.
Not
specifi ed
Not
specifi ed
*2
Over 10 k
to 20 k
[Hz]
±4% rdg.
±20 dgt.
±4% rdg.
±5 dgt.
±2% rdg.
±20 dgt.
Not
specifi ed
Not
specifi ed
Not
specifi ed
Range
[A]
600.00
ACA
6000.0
60.000 m
ACmA
600.00 m
6.0000
*3
ACA
10.000
ACA600.00 A101
ACmA60.000 mA1
*3
ACA
*1: The accuracy is specifi ed in 5% or more of the range. (Less than 5% of
each range is outside of the guaranteed accuracy.)
*2: For more than 300 mA, the accuracy is specifi ed in 5 kHz or less.
*3: Only the DT4282
*4: The accuracy is specifi ed in 2 A or more.
Crest factor3 or less (Note that it applies to 1/2 of the range.)
20 to
below 45
[Hz]
±1.0% rdg.
±20 dgt.
±1.0% rdg.
±5 dgt.
±1.0% rdg.
±20 dgt.
±1.0% rdg.
±5 dgt.
Not
specifi ed
Not
*4
specifi ed
RangeShuntFuse resistance
6.0000 mA101
45 to 65
±0.6% rdg.
±20 dgt.
±0.6% rdg.
±5 dgt.
±0.6% rdg.
±20 dgt.
±0.6% rdg.
±5 dgt.
±0.8% rdg.
±20 dgt.
±0.8% rdg.
±5 dgt.
600.00 mA1
6.0000 A10 m
10.000 A10 m
1
2
3
4
5
6
7
Appx.Ind.
89
Page 93
Accuracy Table
Accuracy guarantee
range for frequency
AC clamp (DT4281)
13
The accuracy is specifi ed in 15% or more of
Range
10.00 A±0.6% rdg. ±2 dgt.±0.9% rdg. ±2 dgt.0.05
20.00 A±0.6% rdg. ±4 dgt.±0.9% rdg. ±4 dgt.0.10
50.00 A±0.6% rdg. ±10 dgt.±0.9% rdg. ±10 dgt.0.25
100.0 A±0.6% rdg. ±2 dgt.±0.9% rdg. ±2 dgt.0.5
200.0 A±0.6% rdg. ±4 dgt.±0.9% rdg. ±4 dgt.1.0
500.0 A±0.6% rdg. ±10 dgt.±0.9% rdg. ±10 dgt.2.5
1000 A±0.6% rdg. ±2 dgt.±0.9% rdg. ±2 dgt.5
• The optional 9010-50, 9018-50, or 9132-50 clamp-on probe is used.
• For accuracy in combination with the clamp, add the accuracy of the clampon probe.
Input impedance1 M ±4%, 100 pF or less
Crest factor3 or less
Accuracy guarantee
range for frequency
the range. (Less than 15% of each range is
outside of the guaranteed accuracy.)
20 Hz to 20 kHz
(A measured value outside of the accuracy
guarantee range for frequency is also displayed.)
Accuracy (only the instrument)
Conversion
rate
(A/mV)
40 to 65 [Hz]Over 65 to 1 k [Hz]
40 Hz to 1 kHz
(A measured value outside of the accuracy
guarantee range for frequency is also displayed.)
90
Page 94
Frequency (Hz)
14
(In the case of ACV, DC+ACV, ACA, ACmA, or ACA)
RangeAccuracy
99.999 Hz±0.005% rdg. ±3 dgt.
999.99 Hz±0.005% rdg. ±3 dgt.
9.9999 kHz±0.005% rdg. ±3 dgt.
99.999 kHz±0.005% rdg. ±3 dgt.
500.00 kHz±0.005% rdg. ±3 dgt.
*1
*1
Accuracy Table
1
2
*1: Measurement range where the accuracy is specifi ed in the range of
99.999 kHz/500.00 kHz
Up to 200 kHz for ACV or ACA
Up to 50 kHz for DC+ACV
Up to 100 kHz for the range from 60.000 mV AC to 600.00 mV AC
Up to 30 kHz for ACmA or ACA
Measurement
range
Pulse width1 s or more (duty ratio: 50%)
• With the fi lter ON, the accuracy is specifi ed in 100 Hz or less.
• During DC+ACV measurement, it is in accordance with the attenuator range
where the input component is large.
0.5 Hz or more ([-----] appears for less than 0.5 Hz.)
91
3
4
5
6
7
Appx.Ind.
Page 95
Accuracy Table
Minimum sensitivity voltage (During DC+ACV measurement, it indicates
an RMS of the AC component.)
Range
[Hz]
99.999
*2,*3
999.99
9.9999 k
99.999 k
500.00 k
60.000 mV 600.00 mV 6.0000 V60.000 V600.00 V1000.0 V
10.000 mV 100.00 mV 1.0000 V 10.000 V 100.00 V 100.0 V
10.000 mV 100.00 mV 1.0000 V 10.000 V 100.00 V 100.0 V
10.000 mV 100.00 mV 1.0000 V 10.000 V 100.00 V 100.0 V
10.000 mV 100.00 mV 1.0000 V 10.000 V Not
30.000 mV 150.00 mV 1.5000 V Not
AC voltage range (sine wave)
specifi ed
specifi ed
Not
specifi ed
Not
specifi ed
Not
specifi ed
*2: The minimum sensitivity voltage which is less than 5 Hz is tripled.
*3: With the fi lter ON, the minimum sensitivity voltage is multiplied by 0.7.
Minimum sensitivity current
Range
[Hz]
99.999
*4
999.99
9.9999 k
99.999 k
500.00 k
600.00 A 6000.0 A60.000 mA 600.00 mA 6.0000 A 10.000 A
60.00 A600.0 A6.000 mA60.00 mA0.6000 A 4.000 A
60.00 A600.0 A6.000 mA60.00 mA0.6000 A 4.000 A
60.00 A600.0 A6.000 mA60.00 mA0.6000 A 4.000 A
60.00 A600.0 A6.000 mA*660.00 mA
100.00 A*51000.0 A
AC range (sine wave)
*5
Not
specifi ed
Not
specifi ed
Not
*6
specifi ed
Not
specifi ed
Not
specifi ed
Not
specifi ed
The maximum input is within each range.
*4: The minimum sensitivity current which is less than 5 Hz is tripled.
*5: Specifi ed in 200 kHz or less.
*6: Specifi ed in 30 kHz or less.
at 80%RH or less (non-condensating)
40°C to 45°C (104°F to 113°F):
at 60%RH or less (non-condensating)
45°C to 55°C (113°F to 131°F):
at 50%RH or less (non-condensating)
-30°C to 60°C (-22°F to 140°F),
at 80%RH or less (non-condensating)
Indoors, pollution degree 2, altitude up to 2000 m
(6562 ft.)
1 m on concrete
1
2
3
4
5
6
7
Appx.Ind.
95
Page 99
General Specifi cations
Power supply
BatteriesAA size alkaline (LR6) battery or AA size manganese
Rated power supply
voltage
Maximum rated
power supply
voltage
Maximum rated
power
Rated power0.2 VA (typ) supply voltage 6.0 V, DCV measurement,
Power during OFF/
APS
Continuous
operating time
(Representative value:
DCV function)
(Using new battery)
Dielectric strength
Maximum rated
voltage between
terminals
Maximum rated
current between
terminals
Maximum rated
voltage between
measurement
terminals and
ground
(R6) battery x 4
1.5 V DC × 4
6.8 V
0.5 VA (max) supply voltage 6.0 V, Continuity
measurement, Input short circuit, Backlight on
Input short circuit, Backlight off
0.1 mVA (max) supply voltage 6.0 V
AA size
battery
AlkalineApprox. 100 hours Approx. 30 hours
Manganese Approx. 30 hours Approx. 10 hours
Between all measurement terminals and case: 8.54 kV AC
(sine wave, 50 Hz/60 Hz, 60 seconds)
Between the V and COM terminals: 1000 V DC/AC
7
or 2 × 10
Between the A/mA and COM terminals: 600 mA
DC/600 mA AC
Between the A and COM terminals: 10 A DC/10 A AC
(continuous)
1000 V (Measurement category III)
600 V (Measurement category IV)
Anticipated transient overvoltage: 8000 V
V • Hz (whichever is the lower value.)
Backlight
offon
96
Page 100
General Specifi cations
DimensionsApprox. 93W × 197H × 53D mm
MassApprox. 650 g (22.9 oz.) (including batteries)
Applicable
standards
Accessories • L9207-10 Test Lead
Replacement parts • A/mA terminal fuse (DT4281, DT4282)
OptionsSee “Options (sold separately)” (p. 2).
(3.66W × 7.76H × 2.09D) (excluding the projection)
• Safety: EN61010
• EMC: EN61326
• Dustproof and waterproof: IP40 (EN60529)
IMPORTANT
To avoid any failure, do not allow the
instrument to get wet. If the instrument gets
wet, have your authorized Hioki distributor or
reseller inspect or repair it, if necessary.
• Instruction Manual
• LR6 alkaline battery × 4 (unloaded)
ManufacturerHOLLYLAND
Rating630 mA/1000 V
Breaking
characteristics
Breaking capacity 50 kA AC/ 30 kA DC
Size
ResistanceApprox. 1.2
Printed colorBlue
• A terminal fuse (DT4282 only)
ManufacturerHOLLYLAND
Rating11 A/1000 V
Breaking
characteristics
Breaking capacity 50 kA AC/ 30 kA DC
Size
Resistance0.1
Printed colorRed
Fast-blow type
10.3 mm × 38 mm
Fast-blow type
10.3 mm × 38 mm
or less
1
2
3
4
5
6
7
Appx.Ind.
97
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.