Hioki DT4221, DT4222 Instruction Manual

Page 1
Instruction Manual
DT4221 DT4222
DIGITAL MULTIMETER
June 2014 Revised edition 1 DT4221A981-01 14-06H
Page 2
Page 3

Contents

Introduction ................................................................... 1
Options (sold separately) ............................................ 2
Safety Notes .................................................................. 3
Usage Notes .................................................................. 8
1 Overview 11
1.1 Overview and Features ........................... 11
1.2 Parts Names and Functions ................... 12
1.3 Display ..................................................... 16
1.4 Alarm Display and Battery Indicator ..... 17
2 Preparation for Measurements 19
2.1 Measurement Workfl ow .......................... 19
2.2 Inserting/Replacing the Battery ............. 20
2.3 Using Test Leads ..................................... 22
2.4 Installation in Measurement Location ... 25
Hanging the instrument with the strap ................25
3 Performing Measurements 27
3.1 Inspection Before Use ............................ 27
3.2 Measuring Voltage................................... 31
Measuring AC voltage .........................................31
Measuring DC voltage ........................................32
Measurement using the AC and DC
automatic judgment (DT4221) ............................32
3.3 Measuring Frequencies .......................... 33
3.4 Checking Continuity ............................... 34
3.5 Measuring Diode (DT4222) ..................... 35
i
Page 4
Contents
3.6 Measuring Resistance (DT4222) ............ 36
3.7 Measuring Electrostatic Capacities
(DT4222) ................................................... 37
3.8 Checking the Electric Charge (DT4221) 38
4 Using Instrument Conveniently 39
4.1 Selecting the Measurement Range ........ 39
Measuring with the auto range ...........................39
Measuring with the manual range .......................39
4.2 Retaining the Measured Value ............... 40
4.3 Reducing the Noise (FILTER) ................. 41
4.4 Checking the Relative Value/
Performing Zero Adjustment .................. 43
Checking the relative value (REL) ......................43
Performing zero adjustment ................................44
4.5 Turning On the Backlight ........................ 45
4.6 Using the Auto Power Save (APS) ......... 45
4.7 Power-on Option Table ........................... 46
5 Specifi cations 49
5.1 General Specifi cations ........................... 49
5.2 Electrical Characteristics ....................... 51
5.3 Accuracy Table ........................................ 53
6 Maintenance and Service 59
6.1 Repair, Inspection, and Cleaning ........... 59
6.2 Troubleshooting ...................................... 60
6.3 Error Display ............................................ 62
Appendix Appx.1
Appx. 1 RMS and Average ....................... Appx.1
ii
Page 5

Introduction

Introduction
Thank you for purchasing the HIOKI DT4221, DT4222 Digital Multimeter. To obtain maximum performance from the product, please read this manual fi rst, and keep it handy for future reference.

Verifying Package Contents

When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifi cations, contact your authorized Hioki distributor or reseller.
Check the package contents as follows.
DT4221 or DT4222
DT4911 Test Lead (p. 22)
LR03 Alkaline battery × 1
Instruction Manual
DT4221 DT4222
The holster has been attached.
(English)
Instruction manuals may also be available in other languages. Please visit our website at
http://www.hioki.com.
1
Page 6

Options (sold separately)

Options (sold separately)
The following options are available for the instrument. Contact your authorized Hioki distributor or reseller when ordering.
Connecting cables (p. 22)
L4933
Contact Pin Set
L4934
Small Alligator Clip Set
DT4911
Test Lead
C0200 Carrying Case
The instrument, test leads, instruction manual, and others can be stored in the case.
Z5004 Magnetic Strap (p. 25)
Attach this strap to the instrument and secure it on the wall surface such as a metal plate for use.
33 V AC/70 V DC
CAT III 300 V/CAT II 600 V
2
Page 7

Safety Notes

Safety Notes
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, using the instrument in a way not described in this manual may negate the provided safety features.
Before using the instrument, be certain to carefully read the following safety notes.
DANGER
Mishandling during use could result in injury or death, as well as damage to the instrument. Be certain that you understand the instructions and precautions in the manual before use.
WARNING
With regard to the electricity supply, there are risks of electric shock, heat generation, fi re, and arc discharge due to short circuits. If persons unfamiliar with electricity measuring instruments are to use the instrument, another person familiar with such instruments must supervise operations.
Protective gear
WARNING
To avoid electric shock when measuring live lines, wear appropriate protective gear, such as insulated rubber gloves, boots and a safety helmet.
3
Page 8
Safety Notes
Notation
In this manual, the risk seriousness and the hazard levels are classifi ed as follows.
Indicates an imminently hazardous situation
DANGER
WARNING
CAUTION
IMPORTANT
that will result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the instrument or malfunction.
Indicates information related to the operation of the instrument or maintenance tasks with which the operators must be fully familiar.
Indicates a high voltage hazard. If a particular safety check is not performed or the instrument is mishandled, this may give rise to a hazardous situation; the operator may receive an electric shock, may get burnt or may even be fatally injured.
Indicates a strong magnetic-fi eld hazard. The effects of the magnetic force can cause abnormal operation of heart pacemakers and/or medical electronics.
Indicates prohibited actions.
Indicates the action which must be performed.
*
Additional information is presented below.
4
Page 9
Safety Notes
Symbols affi xed to the instrument
Indicates cautions and hazards. When the symbol is printed on the instrument, refer to a corresponding topic in the Instruction Manual.
Indicates that dangerous voltage may be present at this terminal.
Indicates a double-insulated device.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates DC (Direct Current) or AC (Alternating Current).
Symbols for various standards
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Indicates that the instrument conforms to safety regulations set out by the EC Directive.
Screen display
This instrument uses the following screen displays.
AB CD E F GH I J K L MNOPQR S T UVWXY Z
1234567890
5
Page 10
Safety Notes
Accuracy
We defi ne measurement tolerances in terms of rdg. (reading) and dgt. (digit) values, with the following meanings:
(Reading or displayed value)
rdg.
The value currently being measured and indicated on the measuring instrument.
(Resolution) The smallest displayable unit on a digital measuring
dgt.
instrument, i.e., the input value that causes the digital display to show a “1” as the least signifi cant digit.
6
Page 11
Safety Notes
Measurement categories
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT II to CAT IV, and called measurement categories.
DANGER
• Using a measuring instrument in an environment
designated with a higher-numbered category than that for which the instrument is rated could result in a severe accident, and must be carefully avoided.
• Using a measuring instrument without
categories in an environment designated with the CAT II to CAT IV category could result in a severe accident, and must be carefully avoided.
This instrument conforms to the safety requirements for CAT III 600 V, CAT IV 300 V measuring instruments.
CAT II: When directly measuring the electrical outlet receptacles of
the primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
CAT III: When measuring the primary electrical circuits of heavy equipment
(fi xed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets
CAT IV: When measuring the circuit from the service drop to the service
entrance, and to the power meter and primary overcurrent protection device (distribution panel)
Distribution panel
Service entrance
Service drop
CAT IV
Power meter
See: “2.3 Using Test Leads” (p. 22)
Internal wiring
CAT III
Fixed installation
CAT II
T
Outlet
7
Page 12

Usage Notes

Usage Notes
Follow these precautions to ensure safe operation and to obtain the full benefi ts of the various functions.
DANGER
If the test lead or the instrument is damaged, there is a risk of electric shock. Before using the instrument, perform the following inspection.
• Before using the instrument, check that the
coating of the test leads are neither ripped nor torn and that no metal parts are exposed. Using the instrument under such conditions could result in electrocution. Replace the test leads with those specifi ed by our company.
• Before using the instrument the fi rst time, verify
that it operates normally to ensure that no damage occurred during storage or shipping. If you fi nd any damage, contact your authorized Hioki distributor or reseller.
Installation
Installing the instrument in inappropriate locations may cause a malfunction of instrument or may give rise to an accident. Avoid the following locations. For details on the operating temperature and humidity, see the specifi cations. (p. 49)
CAUTION
• Exposed to direct sunlight or high temperature
• Exposed to corrosive or combustible gases
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to a strong electromagnetic fi eld or electrostatic charge
• Exposed to high quantities of dust particles
• Near induction heating systems (such as high-frequency induction heating systems and IH cooking equipment)
• Susceptible to vibration
8
Page 13
Handling the cables
WARNING
To prevent electric shock, when measuring the voltage of a power line use a test lead that satisfi es the following criteria:
• Conforms to safety standards IEC61010 or
EN61010
• Of measurement category III or IV
• Its rated voltage is higher than the voltage to be
measured
All of the optional test leads for this instrument conform to the safety standard EN61010. Use a test lead in accordance with its defi ned measurement category and rated voltage.
CAUTION
• Avoid stepping on or pinching the cable, which could damage the cable insulation.
• To avoid damaging the cables, do not bend or pull the leads and the probe bases.
The ends of the test leads are sharp. Be careful to avoid injury.
Precautions during measurement
Usage Notes
WARNING
If the instrument is used in locations where the rating indicated on the instrument or probes is exceeded, the instrument may be damaged resulting in personal injury. Do not use the instrument in such locations. See “Measurement categories” (p. 7).
9
Page 14
Usage Notes
CAUTION
• Do not input voltage exceeding the specifi ed measurement range. Doing so may damage the instrument.
• During the continuity check, diode test, or measurement of resistance or electrostatic capacity, measurement signals are generated in the terminals of the instrument. Depending on the target for measurement, the measurement signal may cause damage. Seeing “Measurement current” and “Open circuit voltage” in the accuracy table (p. 53), check, in advance, that there are no adverse effects of the measurement current and the open circuit voltage.
Precautions during shipment
Observe the following during shipment. Hioki cannot be responsible for damage that occurs during shipment.
CAUTION
• During shipment of the instrument, handle it carefully so that it is not damaged due to a vibration or shock.
• To avoid damage to the instrument, remove the accessories and optional equipment from the instrument before shipment.
If the instrument is not to be used for an extended period of time
IMPORTANT
To avoid corrosion and/or damage to the instrument due to battery leakage, remove the battery from the instrument if it is to be kept in storage for an extended period.
10
Page 15

Overview

1

1.1 Overview and Features

This measuring instrument is a multi-function digital multimeter that ensures both safety and durability.
Main features and functions
• Speedy display of the RMS measured value
• Environmental performance (can be used anywhere) (Operation temperature: -10 to 50°C)
• Filter function that controls the infl uence of noise
• Hold function that retains the screen display
• Solid body which can be used for an extended period of time (drop­proof)
• Speedy measurement via a fast response (0 V 100 V response approx. 0.6 seconds*) * Until the value falls within the accuracy specifi cation range.
Problem fi nding a suitable installation location?
The strap with magnet allows the instrument to be hung conveniently.
For various
Large, easily­viewable display
Backlighting to allow users to read the measurement values in dark environments
purposes
The measurement test leads and end pins can be selected.
11
Page 16
Parts Names and Functions

1.2 Parts Names and Functions

Front
Some indications are different between the DT4221 and DT4222.
DT4221 DT4222
Display
(p. 16)
Operation
keys (p. 13)
Rotary
switch
(p. 14)
12
Page 17
Operation keys
1
2
3
1
2
3
Parts Names and Functions
3
1
2
Normal
for at least 1 second
Pressed down
1
2
3
Sets the manual range/switches the range.
RANGE:AUTO / RANGE:MANUAL
Manually sets/ cancels the hold function for the displayed value.
lights up/
goes off.
Turns on/off the backlight.
Cancels the manual range.
Sets/cancels the display of the relative value (REL).
lights up/goes
off.
Switches/cancels the low pass fi lter and passband settings.
Power-on option (p. 46)
All LCDs light up and the software version is displayed.
Cancels the auto power save function (APS).
APS goes off.
Turns off the automatic backlight deactivation.
13
Page 18
Parts Names and Functions
7
8
9
1
1
3
4
5
6
2
5
2
3
1
2
3
4
5
6
7
8
9
Rotary switches and measurement descriptions
4
3
2
1
DT4221 DT4222
1
2
3
4
5
6
7
8
9
5
Function DT4221 DT4222
AC voltage measurement
DC voltage measurement
Frequency measurement
DC/AC voltage measurement (Automatic judgment) Input impedance 900k
Continuity check
Electrical charge measurement
Resistance measurement
Electrostatic capacity measurement
Diode test
2
6
1
±20%
5
3



-

-
-
-
-
7
8
9
14
Page 19
Rear
1
2
Parts Names and Functions
DT4221
Test lead holder
The test lead can be held.
1
Measurement terminals
The red test lead is connected. Hereafter referred to as “V
1
terminal”.
The black test lead is connected. Hereafter referred to as “COM
2
terminal”.
2
Strap hole
The optional Z5004 Magnetic Strap can be attached. (p. 25)
Battery cover
When replacing the battery (p. 20), remove the cover.
(See p. 20)
Serial number label
It is necessary for production control such as product warranty. Do not peel off the label.
(See p. 24)
15
Page 20
Display
1
2
3
4
5
6
7
8
9
10
11

1.3 Display

For error displays, see “6.3 Error Display” (p.62).
56789
1 2
3 4
AC, DC
1
2
3
4
5
6
7
8
9
automatic
*1
judgment
AC, DC
Continuity check (p.34)
Diode (p. 35)
Retention of the measured value (p. 40) Relative value display (p. 43) The fi lter function is activated. (p. 41) The auto power
save function is activated. (p. 45)
Battery indicator (p. 17)
Indication (example): In the
10
case of 30 V input in the 60 V range, the bar is displayed to the center of the scale.
Auto range, manual range
11
(p. 39)
*1: DT4221
10 11
16
Page 21
Alarm Display and Battery Indicator

1.4 Alarm Display and Battery Indicator

When the measured value exceeds the maximum input range in each range
Voltage measurement
The measured value and OVER blink.
Corrective action:
When the input exceeds the maximum rating, immediately move the test leads away from the measurement object.
Measurement other than voltage
The measured value and OVER blink.
Battery warning indicator
Fully charged.
As the battery charge diminishes, black charge bars disappear, one by one, from the left of the battery indicator. The battery voltage is low. Replace the battery as soon as possible.
(Blinks) The battery is exhausted. Replace the battery.
The charge is only a reference for the continuous operation time.
Power shutdown
When the charge is 0% (less than 1.15 V ± 0.2 V), “bAtt” appears in the display for 3 seconds and the power is shut down.
17
Page 22
Alarm Display and Battery Indicator
18
Page 23
Preparation for
2
1
2
Measurements
2.1 Measurement Workfl ow
Before using the instrument, be sure to read “Usage Notes” (p. 8).
Installation and connection
Insert the battery. (p. 20)
Perform the startup check. (p. 27)
As necessary, have other optional items available and ready.
Measurement
Turn on the power and select the measurement function.
Attach the test leads to the measurement terminals. (p. 22)
(As necessary, perform zero adjustment. (p. 44))
Connect the test leads to the measurement object.
(As necessary)
Hold the display of the measured value. (p. 40)
1
End of the measurement
Move the test leads away from the measurement object and then turn off the power.
2
19
Page 24
Inserting/Replacing the Battery

2.2 Inserting/Replacing the Battery

Before using the instrument, insert one LR03 alkaline battery. Before measurements, check that the battery level is suffi cient. When the battery charge is low, replace the battery.
Nickel-metal hydride batteries
Nickel-metal hydride batteries can be used. However, the discharge characteristic of these batteries is different from that of alkaline batteries. Be aware that the remaining battery power display does not function properly.
WARNING
To avoid electric shock, move the tips of the test leads away from the measurement object before replacing the battery.
To avoid the possibility of explosion, do not short circuit, charge, disassemble, or incinerate batteries.
After battery replacement but before using the instrument, reattach and screw down the battery cover.
CAUTION
Poor performance or damage from battery leakage could result. Observe the cautions listed below.
•Do not use old batteries.
• Be careful to observe the battery polarity during installation.
• Do not use batteries after their recommended expiry date.
• Do not allow the used battery to remain in the instrument.
• To avoid corrosion from battery leakage and/or damage to the instrument, remove the battery from the instrument if it is to be kept in storage for an extended period.
20
Page 25
Inserting/Replacing the Battery
• The indicator appears when the battery charge diminishes. Replace the battery as soon as possible. The power may be turned off when the backlight lights up or a buzzer sounds.
• After use, be sure to turn off the instrument.
• Handle and dispose of batteries in accordance with local regulations.
Have the following items
1
available and ready.
• Phillips screwdriver
• Alkaline (LR03) battery × 1
Move the tips of the test
2
Rear
Screw
leads away from the measurement object.
Set the rotary switch to
3
OFF.
Using a Phillips
4
screwdriver, remove the screw (1 location) from the battery cover on the rear of the instrument.
Remove the battery cover.
5
Remove the old battery.
6
Insert one new battery
7
(LR03), being careful of the battery polarity.
Reattach the battery cover.
8
Secure the cover with the
9
screw.
21
Page 26
Using Test Leads

2.3 Using Test Leads

The DT4911 Test Leads supplied with the instrument are used for measurements. Depending on measurement locations, use our optional parts. For details on the optional items, see “Options (sold separately)” (p. 2).
WARNING
• To prevent a short circuit accident, be sure to
use the test leads with the sleeves attached when performing measurements in the CAT III and CAT IV measurement categories. (See “Measurement categories” (p. 7))
• If the sleeves are inadvertently removed during
measurement, stop the measurement.
CAUTION
• To ensure safe operation, use only test leads specifi ed by our company.
• When carrying out measurements with the sleeves in place, be careful to avoid damaging the sleeves.
• The tips of the metal pins are sharp and may cause injury. Do not touch the tips.
22
Page 27
DT4911 Test Lead
Barriers
Using Test Leads
Sleeves
Plugs
Metal pin Connect to the object to be measured.
Approx. 3 mm (sleeve attached) Approx. 15 mm (sleeve removed)
Sleeve Attach to the metal pins to prevent short circuit accidents.
Barrier Represents the safe handling distance from the metal
pins.
Plug Connect to the measurement terminals on this instrument.
Cable Double sheathed cables (Length: approx. 540 mm)
Cables
During measurement, do not touch the area between the barrier and the tip of the sleeve.
When the white portion inside the cable is exposed, replace with a new DT4911 Test Lead.
Metal pins
Removing and attaching the sleeves
Removing the sleeves Attaching the sleeves
Gently hold the bottom of the sleeves and pull the sleeves off. Safely store the removed sleeves so as not to lose them.
Insert the metal pins of the test leads into the holes of the sleeves, and fi rmly push them all the way in.
23
Page 28
Using Test Leads
Connecting to the instrument
CAUTION
• When removing the test leads from the measurement terminals, hold the end of the plugs.
• Do not pull the cables with excessive force, as they may be broken.
24
BlackRed
When connecting the test leads to the instrument, push them as far as possible into the measurement terminals while holding the end of the plugs.
COM terminal Connect the black test lead. V terminal Connect the red test lead.
Page 29
Installation in Measurement Location

2.4 Installation in Measurement Location

Hanging the instrument with the strap

Attach the optional Z5004 Magnetic Strap to the instrument and attach the magnet to the wall surface (with metal plate affi xed).
DANGER
Those with medical electronics such as pacemakers should not use the Z5004 Magnetic Strap. Nor should such persons approach the Z5004. It is extremely dangerous. The electronics may not operate properly and the life of the operator may be put at great risk.
CAUTION
• Do not use the Z5004 in locations where it may be exposed to rainwater, dust, or condensation. In those conditions, the Z5004 may be decomposed or deteriorated. The magnet adhesion may be diminished. In such case, the instrument may not be hung in place and may fall.
• Do not bring the Z5004 near magnetic media such as fl oppy disks, magnetic cards, pre-paid cards, or magnetized tickets. Doing so may corrupt and may render them unusable. Furthermore, if the Z5004 is brought near precision electronic equipment such as PCs, TV screens, or electronic wrist watches, they may fail.
25
Page 30
Installation in Measurement Location
Strap holes
Front Rear
26
Magnet
Attach it to the wall surface
(with metal plate affi xed).
Page 31

Performing Measurements

3

3.1 Inspection Before Use

Before using the instrument the fi rst time, verify that it operates normally to ensure that no damage occurred during storage or shipping. If you fi nd any damage, contact your authorized Hioki distributor or reseller.
Appearance check of the instrument and test leads
Check item Action
The instrument is neither damaged nor cracked. The internal circuits are not exposed.
The terminals are not contaminated with debris.
The coating of the test leads is neither broken nor frayed, or the white portion or metal part within the lead is exposed.
Check when turning on the power
(Set the rotary switch to any position other than OFF.)
Check item Action
The battery voltage is suffi cient. When the
Visually check the instrument. If it is damaged, there is a risk of electric shock. Do not use the instrument but send it for repair.
Remove contamination with a cotton swab.
If the test lead is damaged, there is a risk of electric shock. Do not use the instrument but send it for repair.
indicator appears in the top right corner of the display, the battery voltage is low. Replace the battery as soon as possible. The power may be turned off when the backlight lights up or a buzzer sounds.
27
Page 32
Inspection Before Use
Check item Action
No indicators are missing. Display all indicators and ensure that
no indicators are missing. (p. 46) If any of the indicators are missing, send the instrument for repair.
Operation check
This section introduces some of the operation checks. Periodical calibration is necessary in order to ensure that this instrument operates according to its specifi cations.
Check that the test leads are not broken.
1
Check method Action
Regarding the continuity check, deliberately short circuit the test leads and then check the display.
For the DT4221 (The position of the rotary switch varies depending on the model.)
Normal:
A buzzer sounds and the value stabilizes at around 0
Abnormal:
A buzzer does not sound and a numeric value other than the above appears.
Corrective action:
The test leads may be broken. Replace with those specifi ed by our company. If the same phenomena persist even after the test leads are replaced, a malfunction may have occurred. Halt inspection and then send the instrument for repair.
.
28
Page 33
Inspection Before Use
Measure samples (such as battery, commercial power
2
supply, and resistor) of which values have already been known, and check that the appropriate values appear.
Check method Action
Example: Perform the AC voltage measurement to measure the commercial power supply, and then check the display.
Normal:
An already-known value appears. (In this example, the commercial voltage level should appear.)
Abnormal:
The measured value does not appear. A malfunction may have occurred. Stop the inspection and do not use the instrument.
29
Page 34
Inspection Before Use
Check that the electrical charge detection function
3
operates normally. (Only the DT4221)
Check method Action
Position the detector on a known power supply, such as a power outlet.
Normal:
A buzzer sounds.
Abnormal:
A buzzer does not sound and the display does not change.
Solution:
A malfunction may have occurred. Stop the inspection and do not use the instrument.
To check the electric charge properly, do not use the instrument with test leads wrapped around the instrument. The detection sensitivity of electric charge measurement deteriorates.
Before measurements
WARNING
Observe the following to avoid short circuit accidents.
• Always verify the appropriate setting of the
rotary switch before connecting the test leads.
• Disconnect the test leads from the measurement
object before switching the rotary switch.
• Operate or connect the instrument by following
the procedure of each measurement example (or procedure steps).
30
Page 35
Measuring Voltage
1
2

3.2 Measuring Voltage

AC/DC voltage measurement and measurement using the AC and DC automatic judgment (only the DT4221) can be performed.
Before measurements
WARNING
If the instrument is used in locations where the rating indicated on the instrument or probes is exceeded, the instrument may be damaged resulting in personal injury. Do not use the instrument in such locations. See “Measurement categories” (p. 7).
The autoranging function of this instrument automatically selects the optimum measurement range. To change the range arbitrarily, use the manual range. (p. 39)

Measuring AC voltage

Measure the AC voltage. The measured value is a true RMS. (p. Appx.1)
1
2
31
Page 36
Measuring Voltage
2
1
2
2
1

Measuring DC voltage

Measure the DC voltage.
1
2

Measurement using the AC and DC automatic judgment (DT4221)

The AC and DC are automatically judged and the voltage is measured. (The instrument does not measure both AC and DC at the same time)
32
voltage
AC
1
2
DC
voltage
1
2
Page 37
Measuring Frequencies
1
2

3.3 Measuring Frequencies

The frequency can be checked. The frequency display is autoranging. The AC voltage range can be changed by pressing the RANGE key.
• If signals out of the range of frequency measurement are measured, “-----” appears. Be aware of it. In a measurement environment with a large amount of noise, the
• frequency may be displayed even with no input. This does not indicate a malfunction of the instrument.
• The sensitivity of the frequency measurement is regulated by range. (Minimum sensitivity voltage (p. 54)) When the value is less than the minimum sensitivity voltage, the indicated value may fl uctuate. When the voltage range is lowered, the value stabilizes. This does not apply to cases where the value uctuates due to noise.
• During the measurement of low frequency voltage, if the auto range does not stabilize and the frequency cannot be measured, fi x the voltage range and measure again.
The frequency is measured.
1
2
33
Page 38
Checking Continuity
1
2

3.4 Checking Continuity

The input short circuit is detected and informed via a buzzer.
WARNING
Before measuring, be sure to turn off the power to the measurement circuit. Otherwise, electric shock may occur or the instrument may be damaged.
1
Detection Threshold Buzzer
Short circuit detection 25
Open detection 245
2
For the DT4221 (The position of the rotary switch varies depending on the model.)
± 10
± 10
Sounds (continuous buzzer sound)
Does not sound
34
Page 39
Measuring Diode (DT4222)
2
1

3.5 Measuring Diode (DT4222)

The forward voltage of the diode is measured. If the forward voltage is within the range from 0.15 V to 1.5 V, it is indicated via a buzzer (intermittent buzzer sound).
WARNING
Before measuring, be sure to turn off the power to the measurement circuit. Otherwise, electric shock may occur or the instrument may be damaged.
In the case of the opposite
AnodeCathode
2
1
The open terminal voltage is approx. 2.5 V or less. To avoid damage to the measurement object, check the specifi cations of the measurement object before use.
connection
35
Page 40
Measuring Resistance (DT4222)
2
1

3.6 Measuring Resistance (DT4222)

Resistance is measured. To measure the low resistance accurately, it is necessary to cancel the resistance of the test leads. Perform zero adjustment for the displayed value using the relative value display (relative function) in advance.
WARNING
Before measuring, be sure to turn off the power to the measurement circuit. Otherwise, electric shock may occur or the instrument may be damaged.
2
1
The open terminal voltage is approx. 1.8 V or less. The measurement current (DC) varies depending on the range. To avoid damage to the measurement object, check the specifi cations before use.
36
Page 41

Measuring Electrostatic Capacities (DT4222)

2
1
3.7 Measuring Electrostatic
Capacities (DT4222)
The capacity of the capacitor is measured.
WARNING
Before measuring, be sure to turn off the power to the measurement circuit. Otherwise, electric shock may occur or the instrument may be damaged.
Do not measure the capacitor which has been charged.
2
1
When measuring the polar capacitor
• Connect the V terminal (red test lead) to the + terminal of
the capacitor and the COM terminal (black test lead) to the
- terminal.
• For components on a circuit board, measurement may not
be possible due to the effect of the peripheral circuit.
37
Page 42

Checking the Electric Charge (DT4221)

2
1
3.8 Checking the Electric Charge
(DT4221)
Whether a power line is energized can be checked easily. If the power line is energized, it is indicated via a buzzer and display. Use this function for coated power lines. The detection may not be made depending on the measurement conditions.
WARNING
To avoid electric shock, do not use the instrument with test leads fi xed to the lead holders.
• To check the electric charge properly, do not use the instrument
with test leads wrapped around the instrument. The detection sensitivity of electric charge measurement deteriorates.
• Check that the detection function operates normally before use.
(p. 30)
Select the measurement
1
2
1
function.
Move the instrument
2
close to the power line.
When the detection level is exceeded, a buzzer sounds.
38
Reference detection level Reference detection level for the power line 80 V AC to 600 V AC
Page 43

Using Instrument Conveniently

4

4.1 Selecting the Measurement Range

Auto or Manual range can be selected. In the case of measurement where the desired range can be selected,
[RANGE:] lights up at the bottom of the display.
Auto range Sets the optimum range automatically in
• Manual range Sets the specifi c range manually.

Measuring with the auto range

When the measurement function is switched using the rotary switch, the auto range is enabled.

Measuring with the manual range

accordance with the actual measurement.
(When the relative value (REL) function is enabled, the range cannot be changed.)
[RANGE: AUTO] lights up.
Press .
[RANGE: MANUAL] lights up.
Each time When the key is pressed at the highest range, the lowest range is speci Example: When the range is 6.000 V to 600.0 V
is pressed, the range is switched.
ed once again.
6 V 60 V 600 V 6 V
To switch from the manual range to the auto range, press
for at least 1 second.
39
Page 44
Retaining the Measured Value

4.2 Retaining the Measured Value

The measured value is retained. (The bar graph is updated.)
To retain the measured value,
press
([HOLD] lights up and the measurement value is retained.)
To cancel the hold state, press it again.
([HOLD] goes off.)
.
40
Page 45
Reducing the Noise (FILTER)

4.3 Reducing the Noise (FILTER)

WARNING
To avoid electric shock or other personal injury, select the appropriate passband setting when measuring the AC voltage. If an inappropriate frequency is selected, the measured value displayed will not be correct.
The infl uence of high-frequency noise can be reduced with the low pass fi lter (digital fi lter). This function can be used when measuring the AC voltage, and the AC and DC automatic judgment. The passband setting for the low pass fi lter can be selected.
Example 1 (FILTER: OFF)
Press for at least 1
second.
(Current FILTER setting is displayed.)
Example 2 (FILTER: 100 Hz)
• When the desired passband setting is displayed for 2
seconds, the setting is applied and then the measurement display reappears.
• If the FILTER setting is changed, the relative value function
(REL) will be canceled.
Each time current FILTER setting is displayed, the passband setting is changed.
[OFF] [100 Hz] [500 Hz] [OFF]
is pressed while the
41
Page 46
Reducing the Noise (FILTER)
Example of frequency property when a fi lter is used
Measured
value [V]
140
140
120
120
100
100
80
80
60
60
40
40
20
20
(AC voltage 600.0 V range, 100 V input)
0
0
Filter OFF Passband 500 [Hz] Passband 100 [Hz]
10 100 1000 10000
Frequency at 100 V input (Hz)
Example: Power frequency on an aircraft or marine vessel is
400 Hz When voltage is 100 V
FILTER setting Displayed value
Normal
Abnormal 100 Hz Around 0 V
OFF
500 Hz
Approx. 100 V
42
Page 47

Checking the Relative Value/Performing Zero Adjustment

4.4 Checking the Relative Value/
Performing Zero Adjustment
The relative value comparing to the standard value can be checked (relative function). It can also be used as the zero adjustment function. Zero adjustment eliminates the infl uences of the test lead wiring resistance (continuity, resistance measurement) and the wiring capacity (capacitor measurement).
When the following measurement function is selected, this function is disabled. AUTO V, Frequency, Diode, Electric charge measurement

Checking the relative value (REL)

Example 1: DC voltage measurement
When the standard value is measured, press for at
least 1 second.
[ ] lights up.
The relative value is displayed.
To cancel the state, press it for at least 1 second again.
] goes off.)
([
43
Page 48
Checking the Relative Value/Performing Zero Adjustment
3
4
1
2
3
4
2

Performing zero adjustment

When performing zero adjustment, the condition of the test leads varies depending on the measurement function. Perform zero adjustment, referring to the table below.
Measurement function
Condition of the test leads Short circuit Open
2
1
4
2
1
4
V, ,
Example 1: Resistance measurement
Select the measurement
1
function. Connect the test leads
2
to the measurement
3
3
terminals. Allow the test leads to
3
short circuit.
Press for at
4
least 1 second.
(After zero adjustment: 0.0)
Measure the resistance.
5
Example 2: Capacitor measurement
Select the measurement
1
function. Connect the test leads
2
to the measurement terminals.
Allow the test leads to
3
open.
Press for at
4
least 1 second.
(After zero adjustment: 0.000 F)
Measure the capacitor.
5
(DT4222)
(DT4222)
44
Page 49
Turning On the Backlight

4.5 Turning On the Backlight

The backlight can be turned on/off by pressing .
The backlight automatically turns off if the instrument is not operated for 40 seconds. The automatic backlight deactivation function can be disabled. (p. 46)

4.6 Using the Auto Power Save (APS)

The auto power save function saves on battery consumption. If the instrument has not been operated for 15 minutes, it enters the sleep mode. When the sleep mode continues for approx. 45 minutes, the power turns off automatically. In the default setting, the auto power save function is set to
enabled. ( It is also possible to disable the auto power save function.
At 30 seconds before the instrument enters the sleep mode, the APS blinks to indicate its status. To continuously use the instrument, press any key or turn the rotary switch.
Auto power save function
• When the instrument is in the sleep mode, press any key or
turn the rotary switch to recover from the sleep mode.
• If the instrument will be used for an extended period of
time, disable the auto power save function. (p.46)
• After use, set the rotary switch to OFF. The sleep mode
consumes a small amount of current.
lights up.)
Recovering from a power shutdown
Set the rotary switch to OFF and turn on the power again.
45
Page 50
Power-on Option Table

4.7 Power-on Option Table

The settings in the instrument can be changed or checked. When the power is turned off, the settings are disabled. When the operation key is released after changing the setting, the regular display then reappears.
+
Canceling the auto power save function
(APS)
Turn on the power while pressing the operation key.
(Turn the rotary switch from OFF.)
Setting change
The APS and OFF displays are shown alternately.
Method
(APS goes off.) (See p. 45)
(See p. 45)
Disabling the automatic backlight deactivation
46
The backlight-auto and OFF displays are shown alternately.
Page 51
Power-on Option Table
Setting change
Checking the software version
Displaying all indicators
Method
(First position from OFF)
Example: Ver 1.00
(Second position from OFF)
Check that there are no missing indicators. If any indicator is missing, stop using the instrument and send it for repair.
47
Page 52
Power-on Option Table
48
Page 53
Specifi cations
5
5.1 General Specifi cations
Power supply LR03 Alkaline battery × 1
Battery indicator warning voltage
Dimensions Approx. 72 W × 149 H × 38 D mm (2.83 W ×
Mass Approx. 190 g (6.7 oz.) (including the battery and
Operating environment
Operating temperature and humidity
Storage temperature and humidity
Dustproof and waterproof
lights up.
lights up.
lights up.
blinks.
• Power shutdown Less than 1.15 V
*1: Error: ±0.2 V
5.87 H × 1.50 D) (including the holster and rotary switch)
holster)
Indoors, pollution degree 2, altitude up to 2,000 m (6,562-ft.)
• Temperature
-10°C to 50°C (14°F to 122°F)
• Humidity
-10°C to 40°C (14°F to 104°F): 80% RH or less (non-condensating) 40°C to 45°C (104°F to 113°F): 60% RH or less (non-condensating) 45°C to 50°C (113°F to 122°F): 50% RH or less (non-condensating)
-30°C to 60°C (-22°F to 140°F), 80% RH or less (non-condensating)
IP42 (EN60529)
1.4 V or more Less than 1.3 V to 1.4 V Less than 1.2 V to 1.3 V Less than 1.15 V to 1.2 V
*1
*1
*1
*1
*1
49
Page 54
General Specifi cations
Drop-proof
distance
Product warranty period
Accessories • DT4911 Test Lead
Options See: “Options (sold separately)” (p. 2)
Applicable standards
1 m on concrete (with the holster attached)
3 years (excluding the measurement accuracy)
• Holster (attached to the instrument, with a test lead holder)
• Instruction Manual
• LR03 Alkaline battery × 1 (not installed in the instrument)
• Safety: EN61010
• EMC: EN61326
50
Page 55
Electrical Characteristics

5.2 Electrical Characteristics

Noise suppression
NMRR
Noise suppression
CMRR
Response time
(Auto range)
Display update rate
Dielectric strength
Maximum rated voltage between terminals
Maximum rated voltage between input terminals and ground
• DCV: -60 dB or more (50 Hz/60 Hz)
• DCV: -100 dB or more (DC/50 Hz/60 Hz, 1k
unbalance)
• ACV: -60 dB or more (DC/50 Hz/60 Hz, 1k
unbalance)
• Power ON time: Within 2 seconds (When the range does not move until the measured value is displayed on the LCD screen)
• DCV: 0.7 to 0.8 seconds (0 V 100 V auto range operation)
• ACV: 0.6 to 0.7 seconds (0 V 100 V auto range operation)
•: 1.0 to 1.1 seconds (Infi nity  0 auto range operation)
*1
*1
*1
• Measured value: 5 times/s (excluding electrostatic capacity and frequency after the range is fi xed)
0.05 to 5 times/s (varies depending on the electrostatic capacity) 1 to 2 times/s (frequency)
*2
*2
*2
• Bar graph: Updated 40 times/s
Between the measurement terminal and case
7.06 kV AC sine wave (50 Hz/60 Hz, 60 seconds)
6
600 V DC/AC or 3 × 10
V • Hz
600 V AC (Measurement category III) 300 V AC (Measurement category IV) Anticipated transient overvoltage: 6000 V
51
Page 56
Electrical Characteristics
Rated power voltage
Maximum rated power
Rated power • 36 mVA +20% or less
Continuous operating time
*1: Until the values stabilize within the accuracy specifi cation range. *2: Measured within the measurement range (excluding range
movement).
1.5 V DC × 1 LR03 Alkaline battery × 1
220 mVA Power voltage 1.5 V, continuity measurement input short-circuited, backlight lit
Power voltage 1.5 V, DCV measurement, backlight off
• 6 mVA +20% or less Power voltage 1.5 V, auto power save function activated
LR03 alkaline battery, backlight off: Approx. 40 hours
52
Page 57
Accuracy Table

5.3 Accuracy Table

Accuracy warranty period 1 year
Regulated power supply range Until the power shutdown (1.15 V ± 0.2 V)
Accuracy guarantee for temperature and humidity
Temperature characteristic
• rdg. (reading or displayed value): The value currently being measured and displayed on the measuring instrument.
• dgt. (resolution): The smallest displayable unit, i.e., the input value that causes the digital display to show a “1”.
AC voltage
1
Range
6.000 V ±1.0% rdg. ±3 dgt. ±2.5% rdg. ±3 dgt. 11.2M
60.00 V ±1.0% rdg. ±3 dgt. ±2.0% rdg. ±3 dgt. 10.3M
600.0 V
40 to 500 Hz
±1.0% rdg. ±3 dgt. ±2.0% rdg. ±3 dgt. 10.2M
23°C ± 5°C (73°F ± 9°F), 80%RH or less (non-condensating)
•Adds “Measurement accuracy × 0.1/°C” (except 23°C ± 5°C (73°F ± 9°F))
•For resistance 60.00M range, adds “Measurement accuracy ×
0.4/°C”(except 23°C ± 5°C (73°F ± 9°F))
*1
Accuracy
Over 500 Hz to
1 kHz
Input
impedance
±2.0%
100 pF or less
±2.0%
100 pF or less
±1.5%
100 pF or less
• Overload protection: 660 V DC/660 V AC or 3 × 106 V • Hz (energized for 1 minute) Transient overvoltage: 6000 V
• Crest factor: The crest factor is 3 up to 4000 counts and reduces linearly to 2 at 6000 counts.
• Connection method: AC coupling
• Auto range movement threshold: 6,000 counts or more for upper range 540 counts or less for lower range
*1: The accuracy is specifi ed in 1% or more of the range, however, ±5
dgt. should be added to 5% or less of the range.
53
Page 58
Accuracy Table
• Accuracy guarantee range for frequency: 40 Hz to 1 kHz (Measured values outside the accuracy guarantee range for frequency are also displayed.) The accuracy is not specifi ed for strain waveforms outside the range of 40 Hz to 1 kHz.
• For 100 Hz with the fi lter ON, ±1.5% rdg. is added to the accuracy specifi cation between 40 Hz and 100 Hz and the accuracy is not specifi ed in 100 Hz or more.
• For 500 Hz with the fi lter ON, ±0.5% rdg. is added to the accuracy specifi cation between 40 Hz and 500 Hz and the accuracy is not specifi ed in 500 Hz or more.
Frequency
2
Range Accuracy
99.99 Hz ±0.1% rdg. ±2 dgt.
999.9 Hz ±0.1% rdg. ±2 dgt.
9.999 kHz ±0.1% rdg. ±2 dgt.
• Auto range movement threshold: 9,999 counts or more for upper range 900 counts or less for lower range
*1
Frequency minimum sensitivity voltage (sine wave)
Range
99.99 Hz
999.9 Hz
9.999 kHz
Measurement
range
5.00 Hz to
*1
99.99 Hz
100.0 Hz to
999.9 Hz
1.000 kHz to
9.999 kHz
6.000 V 60.00 V 600.0 V
0.600 V or more
0.600 V or more
0.600 V or more
AC voltage range
6.00 V or more 60.0 V or more
6.00 V or more 60.0 V or more
6.00 V or more 60.0 V or more
• The input is up to 3 × 106 V • Hz.
• “---- ” appears when no measurement can be made.
*1: The measurement range from 5.00 Hz is only for the 6.000 V range.
The measurement range for other voltage ranges is 40.00 Hz to 99.99 Hz.
54
Page 59
Accuracy Table
DC voltage
3
Range Accuracy Input impedance
600.0 mV ±0.5% rdg. ±5 dgt. 11.2M
6.000 V ±0.5% rdg. ±5 dgt. 11.2M
60.00 V ±0.5% rdg. ±5 dgt. 10.3M
600.0 V ±0.5% rdg. ±5 dgt. 10.2M
• Overload protection: 660 V DC/660 V AC or 3 × 10 for 1 minute)
• Auto range movement threshold: 6,000 counts or more for upper range 540 counts or less for lower range
AUTO V
4
*1
Range
600.0 V ±2.0% rdg. ±3 dgt. ±4.0% rdg. ±3 dgt. 900k ±20%
• Overload protection: 660 V DC/660 V AC or 3 × 10 for 1 minute) Transient overvoltage: 6000 V
• Crest factor: The crest factor is 3 up to 4000 counts and reduces linearly to 2 at 6000 counts.
• Connection method: DC coupling
*1: For AC voltage, the accuracy is specifi ed in 1% or more of the range,
however, ±5 dgt. should be added to 5% or less of the range.
• Accuracy guarantee range for frequency: 40 Hz to 1 kHz (Measured values outside the accuracy guarantee range for frequency are also displayed.) The accuracy is not specifi ed for strain waveforms outside the range of 40 Hz to 1 kHz.
• For 100 Hz with the fi lter ON, ±1.5% rdg. is added to the accuracy specifi cation between 40 Hz and 100 Hz and the accuracy is not specifi ed in 100 Hz or more.
• For 500 Hz with the fi lter ON, ±0.5% rdg. is added to the accuracy specifi cation between 40 Hz and 500 Hz and the accuracy is not specifi ed in 500 Hz or more.
DC, 40 to 500 Hz Over 500 Hz to 1 kHz
Accuracy
±2.0%
±2.0%
±2.0%
±1.5%
6
V • Hz (energized
Input impedance
6
V • Hz (energized
55
Page 60
Accuracy Table
Continuity
5
Range Accuracy Measurement current
600.0
• Open circuit voltage: 1.8 V DC or less
• Overload protection: 600 V DC/600 V AC or 3 × 10 for 1 minute) Current under overload: Steady state 15 mA or less, transient state 0.8 A or less
• Continuity ON threshold: 25
• Continuity OFF threshold: 245
• Response time: Open circuit or short circuit is detected for at least
0.5 ms.
• Accuracy guarantee condition: After zero adjustment has been performed
Resistance
6
Range Accuracy Measurement current
600.0
6.000k
60.00k
600.0k±0.9% rdg. ±5 dgt. 1 A ±20%
6.000M
60.00M
• Open circuit voltage: 1.8 V DC or less
• Overload protection: 600 V DC/600 V AC or 3 × 10 for 1 minute) Current under short circuit: 300 A or less Current under overload: Steady state 15 mA or less, transient state 0.8 A or less
• Maximum capacity load: 10 mF
• Maximum inductive load: 10 H
• Accuracy guarantee condition: After zero adjustment has been performed
• Auto range movement threshold: 6,000 counts or more for upper range 540 counts or less for lower range
±1.0% rdg. ±5 dgt. 200 A ±20%
±0.9% rdg. ±5 dgt. 200 A ±20%
±0.9% rdg. ±5 dgt. 100 A ±20%
±0.9% rdg. ±5 dgt. 10 A ±20%
±0.9% rdg. ±5 dgt. 100 nA ±20%
±1.5% rdg. ±5 dgt. 10 nA ±20%
6
V • Hz (energized
± 10 (continuous buzzer sound)
± 10
6
V • Hz (energized
56
Page 61
Accuracy Table
Electrostatic capacity
7
Range Accuracy Charging current
1.000 F ±1.9% rdg. ±5 dgt. 10 n/100 n/1 A ±20%
10.00 F ±1.9% rdg. ±5 dgt. 100 n/1 /10 A ±20%
100.0 F ±1.9% rdg. ±5 dgt. 1 /10 /100 A ±20%
1.000 mF ±1.9% rdg. ±5 dgt. 10 /100 /200 A ±20%
10.00 mF ±5.0% rdg. ±20 dgt. 100 /200 A ±20%
• Open circuit voltage: 1.8 V DC or less
• Overload protection: 600 V DC/600 V AC or 3 × 10 for 1 minute) Current under short circuit: 300 A or less Current under overload: Steady state 15 mA or less, transient state 0.8 A or less
• Maximum count for each range: 1100 (1000 for 10.00 mF)
• Auto range movement threshold: 1100 counts or more for upper range 100 counts or less for lower range
Diode
8
Range Accuracy
1.500 V ±0.9% rdg. ±5 dgt. 0.5 mA ±20%
• Overload protection: 600 V DC/600 V AC or 3 × 10 for 1 minute) Current under short circuit: 0.7 mA or less Current under overload: Steady state 15 mA or less, transient state 0.8 A or less
• During the forward connection, an intermittent buzzer sounds (threshold:
0.15 V to 1.5 V).
• A continuous buzzer sounds at 0.15 V or less.
Measurement
current
6
V • Hz (energized
Open circuit voltage
2.5 V DC or less Voltage drop due to battery consumption
6
V • Hz (energized
57
Page 62
Accuracy Table
Electric charge
9
Detection voltage range*1Detection target frequency
80 V AC to 600 V AC 50/60 Hz
• During voltage detection, a continuous buzzer sounds.
*1: In contact with the insulated wire that is equivalent to IV2 mm
2
.
58
Page 63

Maintenance and Service

6

6.1 Repair, Inspection, and Cleaning

DANGER
Customers are not allowed to modify, disassemble, or repair the instrument. Doing so may cause fi re, electric shock, or injury.
Calibrations
IMPORTANT
Periodic calibration is necessary in order to ensure that the instrument provides correct measurement results of the specifi ed accuracy.
The calibration frequency varies depending on the status of the instrument or installation environment. We recommend that the calibration frequency is determined in accordance with the status of the instrument or installation environment and that you request that calibration be performed periodically.
Cleaning
• To clean the instrument, wipe it gently with a soft cloth
moistened with water or mild detergent.
• Wipe the display gently with a soft, dry cloth.
IMPORTANT
Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
Disposal
Handle and dispose of the instrument in accordance with local regulations.
59
Page 64
Troubleshooting

6.2 Troubleshooting

• When a malfunction of the instrument is suspected, check the
information in “Before sending the instrument for repair” and then, if necessary, contact your authorized Hioki distributor or reseller.
• When sending the instrument for repair, remove the battery
and pack it carefully to prevent damage during transportation. Include cushioning material so the instrument cannot move within the package. Be sure to include details of the problem. Hioki cannot be responsible for damage that occurs during transportation.
Before sending the instrument for repair
Symptom Check and/or corrective action
Nothing appears in the display.
Or the display disappears after a short time.
The measurement value does not appear. Even after the measurement, 0 (zero) still appears.
Even after short circuit of the probe, the measured value does not appear.
Zero adjustment is not possible.
Check that the battey is not exhausted. Replace with a new battery. (p. 20)
Check that the auto power save function has not been activated. Check the setting of the auto power save function. (p. 45)
Check that the test lead is not broken. Perform the continuity check to confi rm the continuity of the test leads. (p. 28) If the test lead is broken, replace the lead.
Check that the test leads have been inserted at the ends. Check that the measurement method is correct. If no problems have been found, the instrument may be malfunctioning. Send the instrument for repair.
60
Page 65
Symptom Check and/or corrective action
The display does not stabilize and the value fl uctuates; it is diffi cult to read the value.
Turning on the power brings up the error display.
When nothing is connected, the error display appears.
Question Solution
Would like to perform zero adjustment.
Can rechargeable batteries be used?
Check that the input signal is within the input range for the instrument. If there is any infl uence from noise, use the lter function of the instrument. (p. 41)
Reset the instrument. If the same symptom still occurs even after resetting the instrument, send the instrument for repair.
Using the relative value display function, zero adjustment can be performed. (p. 44)
Rechargeable batteries can be used. However, the discharge characteristic of these batteries is different from that of alkaline batteries. Be aware that the remaining battery power display does not function properly.
Troubleshooting
61
Page 66
Error Display

6.3 Error Display

Error
display
Err 1
Err 2
Err 4
Err 5
Description Solution
ROM error Program
ROM error Adjustment data
EEPROM error Memory data
ADC error Hardware malfunction
When the error appears in the display repair the instrument. Contact your authorized Hioki distributor or reseller.
, it is necessary to
62
Page 67

Appendix

Appx. 1 RMS and Average

Difference between the RMS and Average
When converting AC to RMS, 2 methods are available, “True RMS method (True RMS indication)” and “Average method (Average rectifying RMS indication)”. In the case of the sine wave where no skew is included, the same values are indicated in both methods. However, if the waveform is skewed, a difference occurs between the 2 methods.
The true RMS method is applied to this instrument. In the true RMS method, the high frequency component is also included and displayed. In the average method, the input waveform is handled as a sine wave where no skew is included (only single frequency). The average of the AC signal is obtained, converted to the RMS, and then displayed. If the waveform is skewed, a greater measurement error occurs.
Measurement example True RMS Average rectifying
100 V sine wave 100 V 100 V
100 V square wave 100 V 111 V
V(t)
Vm
Sine wave
Vm: Maximum value, Vavg: Average value, Vrms: RMS, T: Time period
t
V(t)
Vm=Vavg=VrmsVavg Vrms
TT
Square wave
t
Appx.1
Page 68
RMS and Average
Appx.2
Page 69
13-09
Page 70
Page 71
Page 72
Loading...