Hioki CT9667-02 User Manual

CT9667-01
k
CT9667-02 CT9667-03
Instruction Manual
June 2015 Edition 1 Printed in Japan CT9667B981-00 15-06H
AC FLEXIBLE CURRENT SENSOR
EN
Introduction
Thank you for purchasing the Hioki CT9667-01, CT9667-02, CT9667-03 AC Flexible Current Sensor. To obtain maximum
performance from the device, please read this manual rst,
and keep it handy for future reference.
Be sure to also read the separate booklet “Current Sensor Operating Precautions” before use.
Use Environment of the Device
WARNING
Although part of this device (the exible loop part
only) is designed to resist the ingress of dust and dripping water, it is not entirely waterproof or dustproof, so to avoid electric shock or damage, do not use it in a wet or dusty environment.
Troubleshooting
If the device seems to be malfunctioning, conrm that the
batteries are not discharged before contacting your authorized Hioki distributor or reseller.
Overview
This device measures large currents of up to 5000 A AC.
The air core coil makes the sensor unit highly exible,
allowing it to be used for clamping in narrow spaces with crowded wiring.
Specications
General Specications
CT9667-01 CT9667-02 CT9667-03
Operating environment
Operating temperature and humidity
Temperature Humidity
(no condensation)
(When using batteries, AC adapter, or external power supply, depends on
the power supply’s specications.)
Storage temperature and humidity
Indoors, pollution degree 2, altitude up to 2000 m (6562 ft.)
25°C to 65°C (−13°F to 149°F)
Less than 40°C (104°F): 80% RH or less From 40°C to 65°C (104°F to 149°F):
Maximum relative humidity declining linearly from 80% RH at 40°C (104°F) to 25% RH at 65°C (149°F)
−30
°C to 70°C (−22°F to 158°F),
80% RH or less (no condensation)
10°C to 50°C
(14°F to 122°F)
From 40°C to 50°C (104°F to 122°F): Maximum relative humidity declining linearly from 80% RH at 40°C (104°F) to 50% RH at 50°C (122°F)
−20
°C to 60°C (−4°F to 140°F), 80% RH or less
Dustproof and waterproof
Standards Dielectric
strength
Power supply
Continuous operating time
Dimensions
(circuit box)
Mass
Cable length
Flexible loop
length
Flexible loop
cross-sectional diameter
Flexible loop
end cap diameter
Product warranty period
Accessories
Options
CT9667-01 CT9667-02 CT9667-03
IP54 (EN60529) (Flexible loop only)
Safety: EN61010 EMC: EN61326, EN61000-3-2, EN61000-3-3
8.54 kV AC rms for 1 minute (at 50 Hz/ 60 Hz)
(between exible loop and output connector)
• LR6 (AA size) alkaline battery × 2 Rated supply voltage: 1.5 V DC × 2 Maximum rated power: 35 mVA
• 9445-02, 9445-03 (for EU), Z1012 AC adapter (option) Rated supply voltage: 100 V to 240 V AC
(Voltage uctuations of ±10% from the rated supply
voltage are taken into account.) Rated supply frequency: 50 Hz/60 Hz Anticipated transient overvoltage: 2500 V
• External power supply Rated supply voltage: 5 V to 15 V DC Maximum rated power: 0.2 VA
When LR6 (AA size) alkaline battery × 2 are used (at 23
°C
)
Approx. 168 hours (7 days)
Approx. 35W × 120.5H × 34D mm (1.38″W × 4.74″H ×
1.34″D) (excluding protruding parts)
Approx. 280 g (9.9 oz.) (Flexible loop + circuit box, including batteries)
Approx. 2 m (78.74″) (between exible loop and circuit box) Approx. 1 m (39.37″) (output cable)
Approx. 390 mm
(15.35″)
Approx.
Approx.
1 year
• LR6 (AA size) alkaline battery × 2
• Instruction Manual
• Current Sensor Operating Precautions
• 9445-02 AC Adapter
• 9445-03 AC Adapter (for EU)
• Z1012 AC Adapter (Operating temperature:
• 9704 Conversion Adapter
7.4 mm (0.29″)
φ
9.9 mm (0.39″)
φ
70°C [−40°F to 158°F]) (BNC female to banana male)
Approx. 630 mm
(24.80″)
Approx. 470 (
16.6
Approx. 930 mm
(36.61″)
Approx. φ13 mm
(0.51″)
Approx. φ20 mm
(0.79″)
oz.)
−40
g
°C to
Input Specications, Output Specications, and Measurement Specications
(1) Basic specications
CT9667-01 CT9667-02 CT9667-03
Output connector
Rated measurement current
Output rate 1 mV/A (500 A range)
Maximum
measurement current
Frequency
band
Output impedance
Measurable
conductor diameter
Maximum rated
voltage to earth
Frequency derating (continuous, design values)
測定電流 [A]
Measurement current [A]
(2) Accuracy specications
f.s. (range): The currently selected range. rdg. (reading or displayed value): The value currently being
Conditions of guaranteed accuracy
Measurement accuracy
Amplitude accuracy
Phase accuracy
Temperature
coefcient
Effect of conductor position
Effect of
external magnetic eld
Offset voltage ±1 mV or less
BNC
500 A AC (500 A range) 5000 A AC (5000 A range)
0.1 mV/A (5000 A range) RMS value, continuous:
see "Frequency derating" below. Peak value:
under the RMS value conditions described above.
1500 A peak (500 A range) 15000 A peak (5000 A range)
10 Hz to 20 kHz (within ±3 dB)
50
(±5%)
Ω
100 mm (3.94″)
φ
or less 1000 V AC (Measurement category III)
600 V AC (Measurement category IV) (Anticipated Transient Overvoltage: 8000 V)
15,0 00
10,0 00
5,00 0
0
10 100 100 0 100 00 100 000
180 mm (7.09″)
φ
or less
1 k 10 k 100
Frequency [Hz]
周波数 [Hz]
measured and indicated on the measuring instrument.
CT9667-01 CT9667-02 CT9667-03
Guaranteed accuracy period: 1 year Guaranteed accuracy period
after adjustment made by Hioki: Opening and closing of the
exible loop: Accuracy guarantee for
temperature and humidity: (With no exible loop stretching, damage, or
cross-sectional deformation in shape)
±2% rdg. ±0.3% f.s. (at 45 Hz to 66 Hz, at exible loop center)
Within ±1.0° (at 45 Hz to 66 Hz)
In the operating temperature range, add 0.05 × specied
accuracy/°C (at temperatures other than 23°C ±5°C).
Within ±3% (deviation from center)
1.5% f.s. or less. (400 A/m, 50 Hz/ 60 Hz)
1 year
10000 times or less
23°C±5°C (73°F±9°F),
80% RH or less
254 mm (10.00″)
φ
or less
1 2 3 4
Parts Names
LOCK
UNLOCK
LOCK
UNLOCK
LOCK
LOCK
Example: CT9667-01
Cable
Connected
instrument
Flexible loop
BNC connector
Circuit box (Battery cover [on back])
Output cable
LED indicator
Lights up
(Green)
POWER
LED
OVER
LED
* When using battery power.
Power on. Battery life is low.*
Coupling
Lights up
(Red)
Range exceeded (peak value of at least 3 × range).
OVER LED POWER LED
Switch
AC adapter connection terminal
Turns off
There is no battery life remaining*, or power off.
Normal measure-
or power off.
ment
,
Measurement Methods
Inspection Before Use
Verify that the device operates normally to ensure that no
damage occurred during storage or shipping. If you nd any
damage, contact your authorized Hioki distributor or reseller.
Check Items Remedy
Is the exible loop or
cable insulation torn, or is any metal exposed?
Is there a broken connec­tion involving the connec­tor or sensor base?
When the switch is placed in any position other than
OFF, does the POWER
LED turn green?
Device damage may result in electric shock. Contact your authorized Hioki distributor or reseller.
Broken connections will make proper measurement impossible. Discontinue use and contact your authorized Hioki distributor or reseller.
Lights up (red): Battery life is low.
→ Replace the batteries soon.
Turns off: There is no battery life
remaining.
→ Replace the batteries immediately.
• When using an AC adapter, always use the optional AC adapter.
• When using the AC adapter and batteries at the same time, the AC adapter takes precedence. Switching power supplies during measurement may introduce noise into the device’s output.
• When using the AC adapter for continuous monitoring, we recommend that you also use batteries to prevent interruptions due to instantaneous power outages.
• After use, always turn off the power.
• Be aware of the following precautions to avoid damage to the device:
Do not pull with excessive force.
NO
Wire band
NO
Do not store while bent.
Do not force
the exible
loop into a bent position.
NO
Insert/ Replace batteries
1
Necessary tool: • Two LR6 (AA size) alkaline batteries
• Phillips screwdriver
Turn off the switch on the circuit box and then disconnect the AC adapter.
Rear of Circuit box
Loosen the
1
screw.
2
Remove the battery cover.
Connect the BNC connector to the
2
+
Install or replace the
3
batteries.
+
4
Mount the battery cover and tighten the retaining screw.
connected instrument
1
Align and insert the connector.
2
Lock
Disconnect the exible loop from the
3
coupling
Example: CT9667-01, CT9667-02
UNLOCK
Rotate the dial until the blue mark
1
is in the UNLOCK position.
Clamp the conductor
4
Clamp only the conductor you wish
Load
Source
Direction of
current ow
Connect the exible loop to the coupling
5
Example: CT9667-01, CT9667-02
Electric conductor
Current direction indicator
Rotate the dial until the blue mark
2
is in the LOCK position.
to measure with the current direc­tion indicator pointing toward the load side. (To ensure that the measured cur­rent and sensor output have the same phase.)
UNLOCK
1
Remove
2
Insert
Pulling on the exible loop with a large amount
of force while in the locked state may cause it to become disconnected from the coupling.
Select the range depending on the current
6
value to be measured
Select the desired current range with the switch.
Memo
• Attach the clamp around only one conductor. If you clamp single-phase (2-wire) or three-phase (3-wire) conductors together, the device will not be able to make a measurement.
OK
5 6 7 8
NO
1
Align and insert the connector.