Hioki CT7044, CT7045, CT7046 Instruction Manual

CT7044 CT7045 CT7046
Instruction Manual
2017
Nov.
Printed in Japan CT7044A961-01 17-11H
Revised edition 1
EN
16-01 EN
Introduction
Thank you for purchasing the Hioki CT7044, CT7045, CT7046 AC Flexible Current Sensor. To obtain maximum performance
from the device, please read this manual rst, and keep it
handy for future reference.
Be sure to also read the separate booklet “Current Sensor Operating Precautions” before use.
Use Environment of the Device
WARNING
Although part of this device is designed to resist the ingress of dust and dripping water, it is not entirely waterproof or dustproof, so to avoid electric shock or damage, do not use it in a wet or dusty environment.
Troubleshooting
If the device seems to be malfunctioning, contact your authorized Hioki distributor or reseller.
Overview
This device measures large currents of up to 6000 A AC.
The air core coil makes the sensor unit highly exible,
allowing it to be used for clamping in narrow spaces with crowded wiring. This current sensor has a Hioki PL14 output connector, enabling it to be automatically recognized when connected to a compatible instrument for simple setup.
Specications
General Specications
CT7044 CT7045 CT7046
Operating environment
Operating temperature and humidity
Temperature
Humidity Less than 40°C (104°F): 80% RH or less
Storage temperature and humidity
Dustproof and waterproof
Standards
Dielectric strength
Power consumption category
Dimensions (circuit box)
Mass
Cable length
Flexible loop length
Flexible loop cross-sectional diameter
Flexible loop end cap diameter
Product warranty period
Accessories
Indoors, pollution degree 2, altitude up to 2000 m (6562 ft.)
25°C to 65°C (−13°F to 149°F)
From 40°C to 65°C (104°F to 149°F): Maximum relative humidity declining linearly from 80% RH at 40°C (104°F) to 25% RH at 65°C (149°F)
(no condensation)
−30
°C to 70°C (−22°F to 158°F), 80% RH or less
(no condensation)
IP54 (EN60529) (when sensor is connected to a compatible instrument)
Safety: EN61010 EMC: EN61326
8.54 kV AC rms for 1 minute (at 50 Hz/ 60 Hz)
(between exible loop and output connector)
Sensor power consumption category: 1 (See the continuous operating time for the instrument to which the device is to be connected.)
Approx. 25W × 72H × 20D mm (0.98″W × 2.83″H ×
0.79″D) (excluding protruding parts)
Approx. 160 g (5.6 oz.)
Approx. 2300 mm (90.55″) (between exible loop and
circuit box)
Approx. 200 mm (7.87″) (output cable)
Approx. 390 mm
(15.35″)
Approx. φ7.4 mm (φ0.29″)
Approx. φ9.9 mm (φ0.39″)
1 year
• Instruction Manual
• Current Sensor Operating Precautions
Approx. 174 g (6.1 oz.)
Approx. 630 mm
(24.80″)
Approx. 186 g (6.6 oz.)
Approx. 870 mm
(34.25″)
Input Specications, Output Specications, and Measurement Specications
(1) Basic specications
CT7044 CT7045 CT7046
Output connector
Rated measurement current
Internal ranges 600 A AC / 6000 A AC
Maximum measurement current
Frequency band
Measurable conductor diameter
Maximum rated voltage to earth
Frequency derating (continuous, design values)
12000
10000
8000
6000
4000
2000
Measurement current [A]
(2) Accuracy specications
f.s. (range): The currently selected range. rdg. (reading or displayed value): The value currently being
Conditions of guaranteed accuracy
Measurement accuracy
Amplitude accuracy
Phase accuracy
Temperature
coefcient
Effect of conductor position
Effect of external
magnetic eld
Offset voltage ±1 mV or less
Hioki PL14
6000 A AC
*Range can be controlled from a connected instrument.
RMS value, continuous: see "Frequency derating" below.
Peak value: 1500 A peak (600 A range)
: 15000 A peak (6000 A range) under the RMS value conditions described below. 1000 A or less and 5×10 10000 A or less and 2×10
10 Hz to 50 kHz (within ±3 dB)
100 mm (φ3.94″)
φ
or less
1000 V AC (Measurement category III) 600 V AC (Measurement category IV) (Anticipated transient overvoltage: 8000 V)
6000 A range
600 A range
0
10 100 1 k 10 k 100 k
Frequency [Hz]
6
A・Hz or less (600 A range)
7
A・Hz or less (6000 A range)
180 mm (φ7.09″)
φ
or less
254 mm (φ10.00″)
φ
or less
measured and indicated on the measuring instrument.
CT7044 CT7045 CT7046
Guaranteed accuracy period: 1 year
Guaranteed accuracy period after adjustment made by Hioki: 1 year
Opening and closing of the exible loop:
10000 times or less
Accuracy guarantee for temperature and humidity: 23°C±5°C (73°F±9°F), 80% RH or less
(With no exible loop stretching, damage, or cross-sectional deformation in shape)
±1.5% rdg. ±0.25% f.s. (Full-scale value is determined by the selected internal range.)
(at 45 Hz to 66 Hz, at exible loop center)
Within ±1.0° (at 45 Hz to 66 Hz)
In the operating temperature range, add 0.05 × specied
accuracy/°C (at temperatures other than 23°C±5°C).
Within ±3.0% (deviation from center)
1.25% f.s. or less. 1.5% f.s. or less.
(400 A/m, 50 Hz/ 60 Hz)
Parts Names
LOCK
UNLOCK
LOCK
UNLOCK
LOCK
LOCK
Example: CT7044
Output connector
Circuit box
Cable
Coupling
Connect the output connector to the
1
connected instrument
Instrument
Align and insert the connector.
Align the arrow on the deviceʼs output connector with the
on the connected instrumentʼs sensor
input connector and insert the connector.
Disconnect the exible loop from the
2
coupling
Once measurement is complete, remove
5
the device from the conductor and disconnect it from the instrument.
When disconnecting the device from the instru­ment, grip the tip of the output connector (the part with the arrow) and pull the connector straight out.
Pulling forcibly on the base of the connector may damage the device.
Memo
Output cable
Flexible loop
Measurement Methods
Inspection Before Use
Verify that the device operates normally to ensure that no
damage occurred during storage or shipping. If you nd any
damage, contact your authorized Hioki distributor or reseller.
Check Items Remedy
Is the exible loop or
cable insulation torn, or is any metal exposed?
Is there a broken connec­tion involving the connec­tor or sensor base?
• Attach the clamp around only one conductor. If you clamp single-phase (2-wire) or three-phase (3-wire) conductors together, the device will not be able to make a measurement.
Device damage may result in electric shock. Contact your authorized Hioki distributor or reseller.
Broken connections will make proper measurement impossible. Discontinue use and contact your authorized Hioki distributor or reseller.
UNLOCK
Rotate the dial until the blue mark
1
is in the UNLOCK position.
Clamp the conductor
3
Clamp only the conductor you wish
Load
Source
Direction of
current ow
Connect the exible loop to the coupling
4
Electric conductor
Current direction indicator
to measure with the current direc­tion indicator pointing toward the load side. (To ensure that the measured cur­rent and sensor output have the same phase.)
Remove
2
OK
NO
• Be aware of the following precautions to avoid damage to the device:
Do not pull with excessive force.
NO
Wire band
NO
Do not store while bent.
Do not force
the exible
loop into a bent position.
NO
UNLOCK
Insert
1
Rotate the dial until the blue mark
2
is in the LOCK position.
Pulling on the exible loop with a large amount
of force while in the locked state may cause it to become disconnected from the coupling.