Thank you for choosing the Hioki CT6710, CT6710 Current
Probe. To ensure your ability to get the most out of this
device over the long term, please read this manual carefully
and keep it available for future reference.
Each model oers a dierent frequency band listed below:
Model CT6710: DC to 50 MHz
Model CT6711: DC to 120 MHz
The following manuals are provided along with these
models. Refer to manuals relevant to your purpose.
• Current Sensor Operating Precautions
Information on the device for safe operation
• Instruction Manual (this document)
Basic instructions and specications of the device
Please review the separate “Current Sensor Operating
Precautions” before using the device.
Target audience
This manual has been written for use by individuals who use
the product in question or who teach others to do so.
It is assumed that the reader possesses basic electrical
knowledge (equivalent to that of someone who graduated
from the electrical program at a technical high school).
1
Notations
HIOKI CT6710A961-01
Notations
Safety notations
This manual classies seriousness of risks and hazard
levels as described below.
Indicates an imminently hazardous situation
DANGER
WARNING
CAUTION
that, if not avoided, will result in death or
serious injury.
Indicates a potentially hazardous situation
that, if not avoided, could result in death or
serious injury.
Indicates a potentially hazardous situation
that, if not avoided, could result in minor or
moderate injury.
NOTICE
IMPORTANT
NOTE
Indicates a high-voltage hazard.
Failure to verify safety or improper handling of the device
could lead to electric shock, burn injury, or death.
2
Indicates potential risks of damage to the
supported product (or to other property).
Indicates information or content that is
particularly important from the standpoint of
operating or maintaining the device.
Indicates useful advice concerning device
performance and operation.
Indicates an action that must not be performed.
HIOKI CT6710A961-01
Indicates an action that must be performed.
Symbols shown on the device
Indicates the presence of a potential hazard. For more
information about locations where this symbol appears
on device components, see the “Usage Notes” section
(p. 8), warning messages listed at the beginning of
operating instructions, and the accompanying document
entitled “Current Sensor Operating Precautions.”
Indicates that the device can only be used at a location
on an insulated wire with sucient insulation for the
circuit voltage.
Symbols for various standards
Indicates the Waste Electrical and Electronic Equipment
Directive (WEEE Directive) in EU member states.
Indicates that the product complies with standards
imposed by EU directives.
Notations
3
Notations
HIOKI CT6710A961-01
Others
*
Indicates additional information is described below.
Bold
Indicates the names of the control keys.
Accuracy
Hioki denes tolerances for measured values in terms of
percentage of reading, as indicated below.
Reading (Displayed value)
Indicates the value the measuring instrument
rdg.
displays. Tolerances for reading errors are
expressed in percent of reading (% of reading,
% rdg).
4
Checking Package Contents
HIOKI CT6710A961-01
Checking Package Contents
When you receive the device, inspect it to ensure that
no damage occurred during shipment. Pay particular
attention to keys and connectors. If you nd any damage
or discover that the device does not perform as indicated
in its specications, please contact your authorized Hioki
distributor or reseller.
Check that the package contents are correct.
The carrying case contains the current probe, Instruction
Manual, and Current Sensor Operating Precautions.
Carrying case
(Accessory)
Model CT6710/CT6711 Current Probe
Instruction Manual
(Accessory)
Current Sensor Operating Precautions
(0990A901)
(Accessory)
5
Checking Package Contents
HIOKI CT6710A961-01
Option
The option below is available for the device. To order an
option, please contact your authorized Hioki distributor or
reseller.
Options are subject to change. Check Hioki’s website for the
latest information.
Model 3269 Power Supply
The 3269 Power Supply can provide the power to up to
two probes of the CT6710/CT6711.
(The 3272 Power Supply cannot be used as a power
supplying unit for the CT6710/CT6711.)
Precautions for transportation
• Store packaging materials for future use. You will need the
packaging materials when shipping the device.
• Transport the device in its carrying case.
6
Safety Notes
HIOKI CT6710A961-01
Safety Notes
The device has been designed in accordance with the IEC
61010 safety standard, and its safety has been veried by
means of testing prior to shipment. However, failure to follow
the information in this manual could render safety-related
functionality provided by the device ineective.
Please review the safety information below before using the
device.
DANGER
Read this manual carefully and ensure you
understand its contents before using the
device.
Improper use of the device could result in serious bodily
injury or damage to the device.
WARNING
If using an electrical measuring instrument
for the rst time, seek instruction from an
individual with electrical measurement
experience rst.
Failure to do so may lead to electric shock, overheating,
re, arcing due to a short-circuit, or other hazards.
7
Usage Notes
HIOKI CT6710A961-01
Usage Notes
Be sure to follow the precautions listed below in order to use
the device safely and in a manner that allows it to function
eectively.
Use of the device should conform not only to its
specications, but also to the specications of all
accessories, options, and other equipment in use.
DANGER
Do not use the device for measuring bare
conductors.
Take measurements at a location on an
insulated wire with sucient insulation for the
circuit voltage.
Doing so may result in electric shock or a short-circuit.
8
Usage Notes
HIOKI CT6710A961-01
DANGER
Do not remove any covers of the sensor,
junction box, and termination unit.
The internal components of the device carry high
voltages and may become very hot during operation.
Touching them could cause electric shock or burns.
Do not connect the device to the primary side
of a distribution panel.
If a short-circuit occurs on the primary side, an
unrestricted current ow can damage the device
and facilities, resulting in serious bodily injury. Even
if a short-circuit occurs on the secondary side of the
distribution panel, the panel will interrupt the short-
circuit current.
Do not use the device in powerful magnetic
elds.
Doing so could cause the sensor to become abnormally
hot, resulting in bodily injury, damage to the device, or
re.
Follow all operating precautions for a
waveform measuring instrument or any other
measuring instruments to which this device is
connected.
Failure to observe this could cause serious bodily injury
or damage to these instruments.
9
Usage Notes
HIOKI CT6710A961-01
Do not use the device to measure circuits that
Do not install the instrument in locations such
WARNING
exceed the ratings or specications of the
device.
Doing so could cause damage to the device or
overheating, resulting in bodily injury.
as the following:
• In locations where it would be subject to direct
sunlight
• In locations where it would be subject to high
temperatures
• In locations where it would be exposed to corrosive
or explosive gases
• In locations where it would be exposed to water, oil,
chemicals, or solvents
• In locations where it would be exposed to high
humidity or condensation
• In locations where it would be exposed to powerful
electromagnetic radiation
• Close to objects carrying an electric charge
• In locations with an excessive amount of dust
• Close to inductive heating devices (high-frequency
inductive heating devices, IH cooktops, etc.)
• In locations characterized by a large amount of
mechanical vibration
• Close to HF power supply units
Doing so could cause damage to the device or cause it
to malfucntion, resulting in bodily injury.
10
Usage Notes
HIOKI CT6710A961-01
WARNING
Ensure that the insulation on the cords is
undamaged and that no bare conductors are
improperly exposed before using the device.
Any damage to the cords leads to electric shock.
Contact your authorized Hioki distributor or reseller for
repair.
CAUTION
Do not place the device on an unstable stand
or angled surface.
Doing so could cause the device to fall or overturn,
resulting in bodily injury or damage to the device.
Properly connect the device to a circuit to
be measured and your waveform measuring
instrument.
Improperly connecting them could cause electric shock
or damage to the circuit, device, or your instrument.
11
Usage Notes
HIOKI CT6710A961-01
Do not store or use the device in locations
Do not apply force in the directions shown in
Do not
apply force.
Keep the upper jaw locked in place when the
NOTICE
subject to abrupt temperature changes.
Doing so could damage the sensor heads.
the gure below while the upper jaw is locked
in place.
Opening lever position
when the upper jaw is
locked in place
Doing so may damage the retracting/extending
mechanism.
device is not in use.
Leaving the upper jaw unlocked will cause dust or dirt to
settle on the facing core surfaces, resulting in damage
to the device.
12
1
HIOKI CT6710A961-01
Overview
1.1 Product Overview
Model CT6710/CT6711 is a clamp-on current probe that
features high current-detection sensitivity and broad
frequency band. The probe uses three current ranges to
detect current waveforms from 1 mA to 50 A.
You can directly connect the termination unit to a BNC input
terminal of your waveform measuring instrument such as
an oscilloscope and recorder, and then clamp the sensor
around a conductor to be measured to observe current
waveforms easily.
Sensor (p. 20)
Junction box (p. 18)
Termination unit (p. 16)
1
13
Product Features
HIOKI CT6710A961-01
1.2 Product Features
Clamp-on sensor heads (p. 21)
The clamp-on sensor heads allow current measurement
without the need to make physical contact with a
conductor to be measured or to disconnect it. You can
observe current waveforms while maintaining the ow of
electric current.
Sliding jaw retracting/extending mechanism
(p. 21)
This feature lets you easily retract, extend, and lock the
upper jaw in place to clamp the sensor around a conductor
to be measured.
Easy-to-connect output terminal (p. 17, p. 38)
Insert the output terminal into a BNC input terminal on
your waveform measuring instrument to connect the
termination unit.
Warning LED lights (p. 18)
The warning LEDs alert you to an overload (indicating
that an input current value exceeds the rated current of
the device) or jaw-unlocked condition (indicating that the
upper jaw has not been locked in place).
14
Product Features
HIOKI CT6710A961-01
Three current measurement ranges (p. 53)
You can choose from the three ranges according to the
magnitude of the current to be measured. This feature lets
you observe a wide range of currents, from 1 mA to 50 A.
Broad frequency band (p. 71)
Model CT6710: DC to 50 MHz
Model CT6711: DC to 120 MHz
Demagnetizing and automatic zero-adjustment
functions (p. 33)
You can use a single key to demagnetize the magnetic
cores and correct variations in oset voltage, both of
which are required before measuring.
Protection mode (p. 62)
The device automatically enters protection mode to
prevent damaging itself when overheating.
1
15
Name and Function of Each Part
HIOKI CT6710A961-01
1.3 Name and Function of Each Part
Termination unit
See p. 34 and p. 35.
Top view
12
5
16
3
4
6
Name and Function of Each Part
HIOKI CT6710A961-01
Output terminal
1
The device converts a captured current waveform into a
voltage waveform by multiplying a particular rate according
to a specied current range and outputs the voltage
waveform from this terminal. Connect this terminal to a BNC
input terminal of your waveform measuring instrument.
Unlock lever
2
To disconnect the output terminal, pull on the termination unit
while simultaneously pulling this lever.
Vents
3
The unit has the vents on the sides and bottom. Do not clog
them.
Serial number
4
The serial number consists of nine digits. The rst two digits
indicate the year of manufacture, while the second two
digits indicate the month of manufacture. Do not remove this
sticker as the number is important.
Power plug
5
Power is provided to the CT6710/CT6711 through this plug.
Connect this plug with the 3269 Power Supply.
Shell
6
Pull on the power plug while simultaneously pulling this shell
to disconnect the plug.
1
You can easily connect and disconnect
the termination unit with two ngers
even when multiple termination units
are connected to a waveform measuring
instrument.
17
Name and Function of Each Part
HIOKI CT6710A961-01
Junction box (keys, LEDs)
4
5
6
7
1
8
DEMAG / AUTO ZERO key
1
Hold down
(about 1 s)
Press momentarily
(less than 0.5 s)
(Higher range) key (p. 53)
2
Switch over to a next higher current range.
(Does not switch from the 30 A range to the 0.5 A range)
(Lower range) key (p. 53)
3
Switch over to a next lower current range.
(Does not switch from the 0.5 A range to the 30 A range)
18
2
3
Top view
Performs demagnetization followed by
automatic zero-adjustment. (p. 33)
Performs automatic zero-adjustment
alone. (p. 40)
Name and Function of Each Part
HIOKI CT6710A961-01
POWER LED (green light)
4
• Lights up when the power is on (p. 31). (Normal operation)
• Rapidly blinks when a checksum error has occurred (p. 95).
OVERLOAD LED (red light)
5
• Flashes three times when demagnetizing or automatic zeroadjustment cannot be performed (p. 91).
• Rapidly blinks when an overload state is detected (p. 90),
the device has entered protection mode (p. 62), or a
checksum error has occurred (p. 95).
JAW UNLOCKED LED (red light)
6
• Lights up when the upper jaw is unlocked.
• Flashes three times when demagnetizing or automatic zero-
adjustment cannot be performed (p. 91).
• Rapidly blinks when the device has entered protection mode
(p. 62) or a checksum error has occurred (p. 95).
DEMAG / AUTO ZERO LED (orange light)
7
• Slowly blinks when demagnetization or automatic zeroadjustment has not yet been performed (p. 33).
• Lights up when demagnetization and automatic zeroadjustment are
• Goes out when demagnetization and automatic zeroadjustment have been completed (p. 39).
• Flashes three times when demagnetization and automatic
zero-adjustment cannot be performed (p. 91).
• Rapidly blinks when the device has entered protection mode
(p. 62) or a checksum error has occurred (p. 95).
RANGE LEDs (green lights)
8
• Lights up when its current range is chosen (p. 53). (Normal
operation)
• Blinks rapidly when the device has entered protection mode
(p. 62) or a checksum error has occurred (p. 95).
in execution
(p. 39).
1
19
Name and Function of Each Part
HIOKI CT6710A961-01
Sensor
NOTICE
Do not store or use the device in locations
subject to abrupt temperature changes.
Do not apply force or mechanical shock to the
device.
Do not subject the device to static electricity.
Doing so could damage the device.
4
2
See p. 41.
1
3
4
20
5
6
See p. 41.
When the upper jaw is unlocked
See p. 41.
Top view
Name and Function of Each Part
HIOKI CT6710A961-01
Jaws
1
Clamp this part around a conductor to be measured by
operating the opening lever, which allows the upper jaw to
slide (retract/extend).
Sensor aperture
2
A conductor to be measured must pass through this
aperture.
Sensor heads
3
The sensor heads, located inside the jaws, detects currents.
Current direction indicator
4
Clamp the sensor around the conductor such that the
direction this arrow indicates matches the current direction of
a conductor to be measured.
JAW UNLOCKED indicator
5
If this indicator appears, the upper jaw is not locked in place.
Opening lever
6
To retract and extend the upper jaw, always operate this
lever. This lever allows you to retract and lock the upper jaw