Thank you for purchasing the Hioki CM4375, CM4376 AC/DC Clamp Meter. To obtain maximum
performance from the instrument over the long term, be sure to read this manual carefully and
keep it handy for future reference.
Read the separate document “Operating Precautions” carefully before using the instrument.
1
Target audience
This manual has been written for use by individuals who use the product in question or who teach
others to do so. It is assumed that the reader possesses basic electrical knowledge (equivalent
to that of someone who graduated from the electrical program at a technical high school).
Trademark
• Bluetooth® is a registered trademark of Bluetooth SIG, Inc.(USA). The trademark is used by
HIOKI E.E. CORPORATION under license.
• Android, Google Play, and Google Chrome are trademarks of Google, Inc.
• IOS is a registered trademark of Cisco Systems, Inc. and/or its afliates in the United States
and certain other countries.
• iPhone, iPad, iPad miniTM, iPad Pro, and iPod touch are trademarks of Apple Inc.
• The App Store is a service mark of Apple Inc.
• Any other products and company names are generally either trade names, registered
trademarks or trademarks of respective companies.
1
2
3
4
5
6
索
引
Page 6
Notation
Notation
Symbols afxed to the instrument
Indicates cautions and hazards. Refer to the "Usage Notes" (p. 6) section of the instruction
manual and the included “Operating Precautions” for more information.
Indicates that the instrument may be connected to or disconnected from a live conductor.
Screen display
The instrument screen displays the alphanumeric characters as follows.
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
1 2 3 4 5 6 7 8 9 0
A different display is used in the case below.
: Wiring break detected
2
Page 7
Notation
Accuracy
We dene measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values with the
following meanings:
f.s. (maximum display value/range)
The maximum displayable value. This is usually the name of the currently selected range.
1
rdg. (displayed value)
The value currently being measured and displayed on the measuring instrument.
dgt. (resolution)
The smallest displayable unit on a digital measuring instrument, i.e., the input value that causes
the digital display to show a “1” as the least-signicant digit.
2
3
4
5
6
索
引
3
Page 8
Verifying Package Contents
L4937マグネットアダプタ
L4934小ワニグチ
マグネ付ストラップ
Verifying Package Contents
Model CM4375 or CM4376
AC/DC Clamp Meter
Model L9207-10 Test Lead
Model C0203 Carrying Case
LR03 Alkaline battery ×2
Instruction Manual*
Operating Precautions (0990A907)
Precautions Concerning Use of
Equipment that Emits Radio Waves (only
for model CM4376)
* Instruction manuals may also be available in other languages.
Please visit our website at http://www.hioki.com
4
Page 9
Options (sold separately)
9243 graber
L4937マグネットアダプタ
L4934小ワニグチ
マグネ付ストラップ
DM4910熱電対
DT4911TestLead
DT4912TestLead
L4930接続ケーブル
9243 graber
L4937マグネットアダプタ
L4934小ワニグチ
マグネ付ストラップ
L9207-10
DM4910熱電対
DT4911TestLead
DT4912TestLead
L4930接続ケーブル
L4931renketu
9243 graber
L4937マグネットアダプタ
L4934小ワニグチ
マグネ付ストラップ
L9207-10
DM4910熱電対
DT4911TestLead
DT4912TestLead
L4930接続ケーブル
L4931renketu
L4931延長ケーブル
L4932(+9207-10cap)
L4933コンタクトピン
L4935ワニ口
9243 graber
L4937マグネットアダプ
9243 graber
L4937マグネットアダプタ
L4934小ワニグチ
マグネ付ストラップ
L9207-10
DM4910熱電対
DT4911TestLead
DT4912TestLead
L4930接続ケーブル
L4931renketu
L4931延長ケーブル
L4932(+9207-10cap)
L4933コンタクトピン
L4935ワニ口
L4936バスバー
L4937マグネットアダプタ
9243 graber
L4937マグネットアダプタ
L4934小ワニグチ
マグネ付ストラップ
L9207-10
DM4910熱電対
DT4911TestLead
DT4912TestLead
L4930接続ケーブル
L4931renketu
L4931延長ケーブル
L4932(+9207-10cap)
9243 graber
L4937マグネットアダプタ
L4934小ワニグチ
マグネ付ストラップ
L9207-10
DM4910熱電対
DT4911TestLead
DT4912TestLead
L4930接続ケーブル
L4931renketu
L4931延長ケーブル
L4932(+9207-10cap)
L4933コンタクトピン
L4937マグネットアダプタ
L4934小ワニグチ
L4937マグネットアダプタ
L4934小ワニグチ
マグネ付ストラップ
9243 graber
L4937マグネットアダプタ
DM4910熱電対
Options (sold separately)
L9207-10 Test Lead
L4930 Connection Cable Set
(Length: 1.2 m)
L4931 Extension Cable Set *
(Length: 1.5 m, with the coupling
connector)
DT4910 Thermocouples (K)
C0203 Carrying Case
*1: CAT IV 600 V/ CAT III 1000 V/ CAT II 1000 V*5: CAT III 300 V/ CAT II 600 V
*2: CAT IV 600 V/ CAT III 1000 V*6: 33 V AC/ 70 V DC
*3: CAT III 1000 V*7: CAT III 600 V/ CAT II 600 V
*4: CAT III 600 V*8: CAT IV 1000 V
1
*
2
*
2
L4933 Contact Pin Set
L4934 Small Alligator Clip Set
L4935 Alligator Clip Set
9243 Grabber Clip
L4936 Bus Bar Clip Set
L4937 Magnetic Adapter Set
9804 Magnetic Adapter *
L4932 Test Pin Set
L4938 Test Pin Set
L4939 Breaker Pin Set
6
*
2
*
3
*
4
*
8
1
*
7
*
1
5
*
2
3
3
*
4
5
4
*
6
索
引
5
Page 10
Usage Notes
Usage Notes
Follow these precautions to ensure safe operation and to obtain the full benets of the various
functions. Read the separate document “Operating Precautions” carefully before using the
instrument. Ensure that your use of the product falls within the specications not only of the
instrument itself, but also of any accessories, options, batteries, and other equipment being
used.
DANGER
• To prevent an electric shock, do not touch any areas beyond the barrier
while the instrument is in use.
See: "1.2 Part Names" (p. 12)
• The maximum measurement current varies with the frequency, and the
current that can be measured continuously is limited. Operating the
instrument at less than this limitation is referred to as derating. Do not
measure currents in excess of the derating curve. Doing so may result in
instrument damage or malfunction, a re, or burn due to sensor heating.
• To prevent an electric shock, conrm that the white portion (insulation layer)
inside the cable is not exposed. If a color inside the cable is exposed, do not
use the cable.
6
Page 11
Usage Notes
WARNING
• Do not allow the instrument to get wet, and do not take measurements with
wet hands. This may cause an electric shock. (This precaution does not
apply to insulated conductors.)
• To prevent an electric shock, do not exceed the lower of the ratings shown
on the instrument and test leads.
1
2
CAUTION
Do not place any foreign object between the jaws or any insert foreign
object into the gap of the sensor head. Doing so may worsen the
performance of the sensor or the opening-closing operation of the
sensor head.
Avoid dropping or jarring the instrument, which could damage the jaw,
adversely affecting measurement.
Attach the clamp around only one conductor. If you clamp single-
phase (2-wire) or three-phase (3-wire) conductors together, the
instrument will not be able to make a measurement.
3
4
5
6
索
引
7
Page 12
Usage Notes
Test Lead
To prevent an electric shock, when measuring the voltage of a power line use a
test lead that satises the following criteria:
The optional test leads provided for the instrument conform to the safety
standard EN61010. Use a test lead in accordance with its dened measurement
category and rated voltage.
The cable is hardened in freezing temperatures. Do not bend or pull it to avoid tearing
its shield or cutting cable.
8
WARNING
• Conforms to safety standards IEC61010 or EN61010
• Measurement category III or IV
• Its rated voltage is higher than the voltage to be measured
• To prevent a short-circuit accident, be sure to use the test leads with the
sleeves attached when performing measurements in the CAT III and CAT IV
measurement categories.
• If the sleeves are removed during measurement, stop the measurement.
CAUTION
Page 13
Model L4937, 9804 Magnetic Adapter Set (optional)
Usage Notes
DANGER
P
ersons wearing electronic medical devices such as a pacemaker should not
use the Magnetic Adapter Set. Such persons should avoid even proximity to the
Magnetic Adapter Set, as it may be dangerous. Medical device operation could be
compromised, presenting a hazard to human life.
CAUTION
• Do not subject the Magnetic Adapter Set to mechanical shock, for example, due to
dropping it. Shock can cause it to be chipped or cracked.
• Do not use the Magnetic Adapter Set in locations where it may be exposed to
rainwater, dust, or condensation. In those conditions, the Magnetic Adapter Set
may be decomposed or deteriorated. The magnet adhesion may be diminished. In
such case, the instrument may not be hung in place and may fall.
• Do not bring the Magnetic Adapter Set near magnetic storage device such as
oppy disks, magnetic cards, pre-paid cards, or magnetized tickets. Doing so may
corrupt and may render them unusable. Furthermore, if the Magnetic Adapter
Set is brought near precision electronic equipment such as PCs, TV screens, or
electronic wrist watches, they may fail.
1
2
3
4
5
6
索
引
9
Page 14
Usage Notes
10
Page 15
1
Overview
1.1 Product Overview and Features
This instrument is a clamp meter that can
perform true RMS measurement of current
simply by clamping it around a circuit. In
addition to current, it provides voltage
measurement, frequency measurement, rush
current measurement, resistance measurement,
diode measurement, capacitance measurement,
temperature measurement, and DC power
measurement.
Model CM4376 also provide Bluetooth®
communications functionality, allowing
measurement data to be monitored and logged
from a mobile device.
Measurement function list
DC current/DC voltage, DC power
Capacitance, temperature
Continuity check, resistance, diode
AUTO AC/DC, AC voltage, DC
voltage, AC+DC voltage, frequency
AUTO AC/DC, AC current, DC
current, AC+DC current, frequency
11
1
2
3
4
5
6
索
引
Page 16
Overview
1.2 Part Names
FrontRear
1
2
3
(CM4376)
12
1 Operation grip
2 Jaw open/closed mark
(The jaws are open if the mark is not
4
5
6
7
8
9
10
11
12
showing.)
3 SHIFT key
(Selects function indicated in blue lettering.)
4 Jaw
5 Barrier
6 Serial number (The serial number consists of
9 digits. The rst two (from the left) indicate
the year of manufacture, and the next two
indicate the month of manufacture.)
7 Rotary switch
8 HOLD key
9 LCD
10 Battery cover
11 Operation keys
12 Measurement terminals
Page 17
2
NO
NONO
Making Measurements
2.1 Inspection Before Measurement
Check if there is any damage to the instrument occurred during storage or shipping and
verify that instrument operates normally before using it. If you nd any damage, contact your
authorized Hioki distributor or reseller.
Check item
The battery cover is closed and its screw has
been securely tightened.
There is no foreign matter on the
measurement terminals. (p. 12)
The test leads are not broken.
1 Ω or less
OK
There is no damage to the test lead
insulation, and neither the white sheathing nor
metal conductor inside the wire are exposed.
The instrument is neither damaged nor cracked.
No indicators are missing.
(All lit up)
13
1
2
3
4
5
6
索
引
Page 18
Current Measurement
2.2 Current Measurement
1
2
Press for 1 sec.
Zero adjustment
Frequency detection range of AC current
5.0 A or more
DC current polarity detection function
(p. 31)
If the measured value is negative, the
buzzer will sound, and the display will turn
red (threshold: −10 A).
14
3
Clamp the instrument around a conductor.Turn the rotary switch.
4
AUTO
(AUTO AC/DC)
e.g.: 3P4W (3-phase, 4-wire) circuit
breaker (AC current measurement)
Be sure that the currents
direction points the arrow.
(AC A)
(AC+DC A)(DC A)
Hz
(Frequency)
Page 19
Manual Hold / Auto Hold
Current Measurement
Manual holdAuto hold
Press for 1 sec.
Measured value
retains.
Pressing the HOLD key again cancels
the measured value hold function.
(Measured value stabilizes.)
Measured value automatically retains.
Pressing the HOLD key for 1 sec. cancels the auto
hold function.
1
blinks.
2
Clamp the instrument around a
conductor.
3
4
Disconnect
5
6
索
引
15
Page 20
Current Measurement
Auto hold conditions
Display value updates are stopped when the following two conditions are satised:
• When the measured value exceeds the threshold value described in the table in the next page. (voltage,
current). When the measured value is less than the threshold value described in the table in the next page.
(resistance, continuity, diode)
• When the range over which the measured value is uctuating stabilizes within the uctuation range
described in the table in the next page.
Held measured value
Measured
value
Auto hold
e.g.: 100.0 A
Fluctuation range
Fluctuation range
Auto hold
e.g.: 99.0 A
Fluctuation range
Threshold
e.g.: 12.0 A
Start
End
If the measured value falls below the threshold value (voltage, current) or exceeds the threshold value
(resistance, continuity, diode) after display value updates are stopped, the display value update is restarted.
Display value updates will stop if the two conditions are satised once again.
End
Start
Measurement
End
Start
Time
16
Page 21
Current Measurement
Measurement functionFluctuation rangeThreshold value
AUTO A
AC current
DC current
AC+DC current
AC voltage
DC voltage
(excluding the 600.0 mV range)
AC+DC voltage
Resistance
Continuity
Diode1.800 V range: within 40 counts1.800 V range: 1460 counts
The auto hold function only operates for the above measurement functions.
Within 120 counts120 counts
6.000 V/60.00 V/600.0 V range:
within 120 counts
1000 V range: within 20 counts
1500 V range: within 30 counts
/6.000 kΩ/60.00 kΩ/
600.0
Ω
600.0 k
Ω range: within 100 counts
6.000 V/60.00 V/600.0 Vrange:
120 counts
1000 V range: 20 counts
1500 V range: 30 counts
600.0 Ω/6.000 kΩ/60.00 kΩ/
600.0 k
Ω range: 4900 counts
17
1
2
3
4
5
6
索
引
Page 22
Current Measurement
Filter Function
FILTER OFFFILTER ON
Measured value including noise
Press for 1 sec.
Measured value with reduced noise
Frequency characteristics when using the
lter function (100 A input)
FILTER OFF
FILTER ON Passband 100 [Hz]
Measured value [A]
Frequency [Hz]
18
Turn off the lter function when performing
measurement of power supply frequencies in excess
of 100 Hz, for example on an aircraft or ship.
Filter OFF
Page 23
MAX value/MIN value/AVG value/PEAK value
1
Clamp the instrument around a conductor.
Current Measurement
1
2
3
4
The instrument
performs true RMS
measurement.
“AVG” indicates
the average of all
measured values.
(AUTO AC/DC)
Measured value retains.
(AC A)
PEAK MIN
(AC+DC A)(DC A)
PEAK MAX
MAX
HzAUTO
(Frequency)
MIN
Can not be used at
AUTO AC/DC.
Press for 1 sec.
Cancel
Waveform
Measured value
(RMS value)
Display refresh interval
19
2
3
4
5
6
索
引
Page 24
Current Measurement
Rush current (INRUSH)
1
Turn off the motor.
2
Turn the rotary switch.
3
Press for 1 sec.
Zero adjustment
4
Clamp the instrument around a
conductor.
Trigger level: ±10 A
20
5
Press for 1 sec.
INRUSH ON
6
Turn on the motor.
RMS value (INRUSH)
Example
screen: 100 A
AC current
Example screen: 250 A
Maximum wave height (PEAK)
Interval during which rush
current occurs
(Dozens to hundreds of
milliseconds in duration)
(Rush current
occurrence)
RMS value (INRUSH)
DC current
Maximum wave height (PEAK)
Interval during which rush
current occurs
(Dozens to hundreds of
milliseconds in duration)
Page 25
2.3 Other Measurement Functions
Other Measurement Functions
Voltage
e.g.: commercial power supply (AC voltage measurement)
23
1
RedBlack
Do not touch.No overvoltage
NO
(Red ash)
NO
DC voltage polarity detection function (p. 31)
If the measured value is negative, the buzzer will sound,
and the display will turn red (threshold: −10 V).
AUTO
Hz
Switch the range
(AUTO AC/DC)
(AC V)
(DC V)
(AC+DC V)
(Frequency)
1
2
3
4
5
6
索
引
21
Page 26
Other Measurement Functions
Continuity Check
1
Zero adjustment
Press for 1 sec.
2
Red
BlackBlackBlack
(Red display)
Resistance
1
Zero adjustment
Press for 1 sec.
2
RedRed
22
Diode
Page 27
Other Measurement Functions
Capacitance
RedBlack
Temperature
Model DT4910
Thermocouples (K)
−
+
Red
Black
: DT4910 is broken.
To change the temperature display unit: p. 32
23
1
2
3
4
5
6
索
引
Page 28
Other Measurement Functions
Simultaneous display of
DC current and DC voltage
Red
Black
24
e.g.: Solar power system maintenancee.g.: Checking a car battery
DC power
Black
Red
Page 29
Bluetooth® Communications (only for model CM4376)
2.4 Bluetooth® Communications (only for model CM4376)
The CM4376 are clamp-style meters with Bluetooth (Bluetooth low energy) support. When the Bluetooth
function is enabled, you can review measurement data and create measurement reports on mobile devices
(iPhone, iPad, iPad mini
functionality, see the Help function in the application software GENNECT Cross.
Install the GENNECT Cross on your mobile device. (p. 26)
1
Enable the Bluetooth function on the CM4376. (p. 27)
2
Launch the GENNECT Cross and pair it with the CM4376. (p. 28)
3
Select the General Measurement, Logging (Recording), or
4
Waveform Graph function. (p. 29)
TM
, iPad Pro, iPod touch, and AndroidTM devices). For more information about this
Press for 1 sec.
25
1
2
3
4
5
6
索
引
Page 30
Bluetooth® Communications (only for model CM4376)
Installing the application software GENNECT Cross
Search for “GENNECT Cross” on the App Store from your iPhone, iPad or other Apple device, or on Google
TM
from your Android device. Then download and install the GENNECT Cross. You will need an Apple ID
Play
to download the app on the App Store, or a Google account to download the app on Google Play. For more
information about how to register an account, contact the store at which you purchased your device.
• Because the CM4376 emit radio waves, use in a country or region where they have not been approved may
be subject to nes or other penalties as a violation of applicable laws or regulations. For more information,
see the attached “Precautions Concerning Use of Equipment that Emits Radio Waves” or go to our website.
• The CM4376 availability is limited to certain countries. For more information, contact your authorized Hioki
distributor or reseller.
• The distance over which data can be sent and received using Bluetooth varies greatly depending on
whether there are any obstructions between the paired instruments (for example, walls, metal barriers, etc.)
and on the distance between the instrument and the oor (or ground). To ensure stable measurement, verify
adequate signal strength.
• Although this application software is provided free of charge, downloading or using the application software
may incur Internet connection charges. Such charges are the sole responsibility of the user.
• This application software is not guaranteed to operate on all mobile devices.
26
Page 31
Bluetooth® Communications (only for model CM4376)
Turning on the Bluetooth function
Bluetooth function OFFBluetooth function ON
1
2
Press for 1 sec.
lights up: Bluetooth function ON
ashes: Sending/receiving data
27
3
4
5
6
索
引
Page 32
Bluetooth® Communications (only for model CM4376)
Pairing the app with the CM4376
2
3
1
• When the app is launched for the rst time (before being paired with any instrument), the Instrument
Settings screen will be displayed.
• While the mobile device is displaying the Instrument Settings screen, simply move it close to a CM4376 to
automatically pair it with the instrument (the app can be paired with up to 8 instruments).
• Allow about 5 to 30 seconds for the CM4376 to pair with the app after being turned on. If the instrument fails
to pair within 1 minute, relaunch GENNECT Cross and cycle the instrument’s power.
28
Page 33
Bluetooth® Communications (only for model CM4376)
Making measurements with the Bluetooth function
Select the General Measurement, Logging (Recording), or Waveform Graph function on the Home screen.
For more information about each function, see the Help function in the GENNECT Cross.
General Measurement
Saves measured values from
multiple channels
Logging (Recording)
Simple logging (up to 24 hours)
Waveform Graph
Simple oscilloscope
(voltage/current)
29
1
2
3
4
5
6
索
引
Page 34
Backlight / Auto Power Save (APS)
2.5 Backlight / Auto Power Save (APS)
BacklightAuto Power Save (APS)
Backlight OFF
Backlight ON
Automatically switched off when
the instrument is not in use for
40 sec.
The instrument is automatically turned off.
To restart the instrument, briey set the rotary switch to “OFF.”
30
(Always on)
Cancelation method: (p. 31)
No operation
for 15 min.
You can turn the display back on
by pressing a key or by turning the
rotary switch.
No operation for 45 min.
Page 35
2.6 Power-on Option Table
Power-on Option Table
+
Canceling the auto power save (APS)
function (OFF)
DC current and DC voltage polarity
detection function (ON/OFF)
Displaying all indicators
(Version of software/Model number/
Serial number)
Buzzer sound (ON/OFF)
Automatic backlight deactivation
(ON/OFF)
Move the rotary switch from the “OFF” position to any of the test mode positions
while pressing an operation key.
SettingOperating instruction
Factory
setting
ON
OFFYes
––
ONYes
ONYes
retained?
(Set each time)
Setting
No
31
1
2
3
4
5
6
索
引
Page 36
Power-on Option Table
SettingOperating instruction
Switching the temperature unit
32
Press for 1 sec.
To change the
temperature unit:
To save the setting:
Press for 1 sec.
Factory
setting
°CYes
Setting
retained?
Page 37
3
Specications
3.1 General Specications
Operating environment Indoors, pollution degree 2, altitude up to 2000 m (6562 ft.)
Operating temperature
and humidity
Storage temperature
and humidity
Dustproof and
waterproof
StandardsSafety EN 61010
Power supplyLR03 alkaline battery ×2
Continuous operating
time
Interface (only for
CM4376)
−25°C to 65°C (−13°F to 149°F), 90% RH or less (no condensation)
−30°C to 70°C (−22°F to 158°F), 90% RH or less
(no condensation, when batteries are removed)
IP20 (Measuring voltage or current in a hazardous live conductor while
completely dry)
IP50 (Measuring resistance while completely dry)
IP54 (During storage and while measuring current in an insulated conductor)
EMC EN 61326
Rated supply voltage: 1.5 V DC ×2
Approx. 40 hours (Bluetooth communication OFF)
Approx. 20 hours (Bluetooth communication ON)
(100 A AC measurement, LCD backlight OFF, at 23°C [ 73.4°F ])
1 year (accuracy shown in accuracy table)
3 years (accuracy shown in accuracy table × 1.5) (reference
values)
1 year
23°C±5°C (73°F±9°F), 90% RH or less (no condensation)
Input specications/Measurement specications
41
1
2
3
4
5
6
索
引
Page 46
Accuracy Table
3.3 Accuracy Table
(1) AUTO A (AC/DC current automatic detection)
During AC detection: Conforms to accuracy specications described in "(4) AC+DC current" (p. 44).
During DC detection: Conforms to accuracy specications described in "(3) DC current" (p. 43).
(2) AC current
Measurement value/MAX/MIN/AVE
Accuracy guarantee range
(Resolution)
1.0 A to 30.0 A (0.1 A)10 Hz ≤ f < 45 Hz±1.8% rdg.±1.0 A±2.3% rdg.±1.0 A
30.1 A to 900.0 A (0.1 A)10 Hz ≤ f < 45 Hz±1.8% rdg.±0.5 A±2.3% rdg.±0.5 A
900.1 A to 999.9 A (0.1 A)10 Hz ≤ f < 45 Hz±2.3% rdg.±0.5 A±2.8% rdg.±0.5 A
Accuracy guarantee
frequency range
45 Hz ≤ f ≤ 66 Hz±1.3% rdg.±0.8 A±1.8% rdg.±0.8 A
66 Hz < f ≤ 1 kHz±2.0% rdg.±1.0 A–
45 Hz ≤ f ≤ 66 Hz±1.3% rdg.±0.3 A±1.8% rdg.±0.3 A
66 Hz < f ≤ 1 kHz±2.0% rdg.±0.5 A–
45 Hz ≤ f ≤ 66 Hz±1.8% rdg.±0.3 A±2.3% rdg.±0.3 A
66 Hz < f ≤ 1 kHz±2.5% rdg.±0.5 A–
Measurement accuracy
Filter OFFFilter ON
42
Page 47
PEAK MAX/PEAK MIN
Accuracy guarantee range (Resolution)
±10 A to ±1000 A (1 A)10 Hz ≤ f < 45 Hz±1.8% rdg.±7 A
±1001 A to ±1500 A (1 A)10 Hz ≤ f < 45 Hz±2.3% rdg.±7 A
Accuracy guarantee
frequency range
45 Hz ≤ f ≤ 66 Hz±1.3% rdg.±7 A
66 Hz < f ≤ 1 kHz±2.0% rdg.±7 A
45 Hz ≤ f ≤ 66 Hz±1.8% rdg.±7 A
66 Hz < f ≤ 1 kHz±2.5% rdg.±7 A
Measurement accuracy
Accuracy Table
1
2
(3) DC current
Measurement value/MAX/MIN/AVE
Accuracy guarantee range (Resolution)Measurement accuracy
±1.0 A to ±30.0 A (0.1 A)±1.3% rdg.±0.8 A
±30.1 A to ±999.9 A (0.1 A)±1.3% rdg.±0.3 A
PEAK MAX/PEAK MIN
Accuracy guarantee range (Resolution)Measurement accuracy
±10 A to ±1000 A (1 A)±1.3% rdg.±7 A
±1001 A to ±1500 A (1 A)±1.8% rdg.±7 A
43
3
4
5
6
索
引
Page 48
Accuracy Table
(4) AC+DC current
Measurement value/MAX/MIN/AVE
Accuracy guarantee
range (Resolution)
1.0 A to 30.0 A (0.1 A)10 Hz ≤ f < 45 Hz±1.8% rdg.±1.2 A±2.3% rdg.±1.2 A
30.1 A to 900.0 A (0.1 A)10 Hz ≤ f < 45 Hz±1.8% rdg.±0.7 A±2.3% rdg.±0.7 A
900.1 A to 999.9 A (0.1 A)10 Hz ≤ f < 45 Hz±2.3% rdg.±0.7 A±2.8% rdg.±0.7 A
Accuracy guarantee
frequency range
DC, 45 Hz ≤ f ≤ 66 Hz±1.3% rdg.±1.8 A±1.8% rdg.±1.8 A
66 Hz < f ≤ 1 kHz±2.0% rdg.±1.2 A–
DC, 45 Hz ≤ f ≤ 66 Hz±1.3% rdg.±1.3 A±1.8% rdg.±1.3 A
66 Hz < f ≤ 1 kHz±2.0% rdg.±0.7 A–
DC, 45 Hz ≤ f ≤ 66 Hz±1.8% rdg.±1.3 A±2.3% rdg.±1.3 A
66 Hz < f ≤ 1 kHz±2.5% rdg.±0.7 A–
Measurement accuracy
Filter OFFFilter ON
PEAK MAX/PEAK MIN
Accuracy guarantee range (Resolution)
±10 A to ±1000 A (1 A)10 Hz ≤ f < 45 Hz±1.8% rdg.±7 A
±1001 A to ±1500 A (1 A)10 Hz ≤ f < 45 Hz±2.3% rdg.±7 A
Accuracy guarantee
frequency range
DC, 45 Hz ≤ f ≤ 66 Hz±1.3% rdg.±7 A
66 Hz < f ≤ 1 kHz±2.0% rdg.±7 A
DC, 45 Hz ≤ f ≤ 66 Hz±1.8% rdg.±7 A
66 Hz < f ≤ 1 kHz±2.5% rdg.±7 A
Measurement accuracy
44
Page 49
Accuracy Table
(5) Current frequency/Voltage frequency
Range (Auto-range threshold)
9.999 Hz (more than 9999 counts) 1.000 Hz to 9.999 Hz (0.001 Hz)±0.1% rdg.±0.003 Hz
99.99 Hz (more than 9999 counts/
less than 900 counts)
999.9 Hz (less than 900 counts)1.0 Hz to 999.9 Hz (0.1 Hz)±0.1% rdg.±0.1 Hz
Accuracy guarantee range
(Resolution)
1.00 Hz to 99.99 Hz (0.01 Hz)±0.1% rdg.±0.01 Hz
Measurement accuracy
(6) INRUSH (Rush current)
INRUSH measurement value
Accuracy guarantee range (Resolution)
10.0 A to 999.9 A (0.1 A)DC, 20 Hz ≤ f ≤ 500 Hz±5.0% rdg.±1.3 A
Accuracy guarantee
frequency range
Measurement accuracy
1
2
3
4
INRUSH PEAK value
Accuracy guarantee range (Resolution)
±10 A to ±1000 A (1 A)DC, 20 Hz ≤ f ≤ 500 Hz±6.0% rdg.±10 A
±1001 A to ±1500 A (1 A)DC, 20 Hz ≤ f ≤ 500 Hz±8.0% rdg.±10 A
Accuracy guarantee
frequency range
Measurement accuracy
45
5
6
索
引
Page 50
Accuracy Table
(7) AUTO V (AC/DC voltage automatic detection)
During AC detection: Conforms to accuracy specications described in "(10) AC+DC voltage" (p. 51).
During DC detection: Conforms to accuracy specications described in "(9) DC voltage" (p. 49).
(8) AC voltage
Measurement value/MAX/MIN/AVE
Range
(Auto-range
threshold)
6.000 V (more
than 6000
counts)
Accuracy
guarantee range
(Resolution)
0.000 V to 0.299
V (0.001 V)
0.300 V to 6.000
V (0.001 V)
Accuracy guarantee
frequency range *
15 Hz ≤ f < 45 Hz±1.5% rdg.
45 Hz ≤ f ≤ 66 Hz±0.9% rdg.
66 Hz < f ≤ 1 kHz±1.5% rdg.
15 Hz ≤ f < 45 Hz±1.5% rdg.
45 Hz ≤ f ≤ 66 Hz±0.9% rdg.
66 Hz < f ≤ 1 kHz±1.5% rdg.
Measurement accuracy
1 *2
Filter OFFFilter ON
±0.015 V
±0.013 V
±0.015 V
±0.005 V
±0.003 V
±0.005 V
±2.0% rdg.
±0.015 V
±1.4% rdg.
±0.013 V
–
±2.0% rdg.
±0.005 V
±1.4% rdg.
±0.003 V
–
impedance *
3.2 M
3.2 M
46
Input
±5%
Ω
±5%
Ω
3
Page 51
Accuracy Table
Range
(Auto-range
threshold)
60.00 V
(more than
6000 counts/
less than 540
counts)
600.0 V
(more than
6000 counts/
less than 540
counts)
1000 V
(less than 540
counts)
*1: Frequency range of 15 Hz ≤ f < 20 Hz is designed value.
*2: Within the frequency range of f < 45 Hz, the accuracy guarantee assumes a superposed DC voltage of
less than 500 V.
*3: At 50 Hz AC.
Accuracy
guarantee range
(Resolution)
3.00 V to 60.00 V
(0.01 V)
30.0 V to 600.0 V
(0.1 V)
50 V to 1000 V
(1 V)
Accuracy guarantee
frequency range *
15 Hz ≤ f < 45 Hz±1.5% rdg.
45 Hz ≤ f ≤ 66 Hz±0.9% rdg.
66 Hz < f ≤ 1 kHz±1.5% rdg.
15 Hz ≤ f < 45 Hz±1.5% rdg.
45 Hz ≤ f ≤ 66 Hz±0.9% rdg.
66 Hz < f ≤ 1 kHz±1.5% rdg.
15 Hz ≤ f < 45 Hz±1.5% rdg.
45 Hz ≤ f ≤ 66 Hz±0.9% rdg.
66 Hz < f ≤ 1 kHz±1.5% rdg.
Measurement accuracy
1 *2
Filter OFFFilter ON
±0.05 V
±0.03 V
±0.05 V
±0.5 V
±0.3 V
±0.5 V
±5 V
±3 V
±5 V
±2.0% rdg.
±0.05 V
±1.4% rdg.
±0.03 V
–
±2.0% rdg.
±0.5 V
±1.4% rdg.
±0.3 V
–
±2.0% rdg.
±5 V
±1.4% rdg.
±3 V
–
Input
impedance *
±5%
3.1 M
Ω
±5%
3.0 M
Ω
±5%
3.0 M
Ω
3
47
1
2
3
4
5
6
索
引
Page 52
Accuracy Table
PEAK MAX/PEAK MIN
Range
Accuracy guarantee range
(Resolution)
6.000 V0 V to ±12.00 V (0.01 V)15 Hz ≤ f < 45 Hz±1.8% rdg.±0.07 V
60.00 V±3.0 V to ±120.0 V (0.1 V)15 Hz ≤ f < 45 Hz±1.8% rdg.±0.7 V
3
600.0 V±30 V to ±1000 V *
1000 V±50 V to ±1000 V *
(1 V)15 Hz ≤ f < 45 Hz±1.8% rdg.±7 V
4
(1 V)15 Hz ≤ f < 45 Hz±1.8% rdg.±7 V
*1: Frequency range of 15 Hz ≤ f < 20 Hz is designed value.
*2: Within the frequency range of f < 45 Hz, the accuracy guarantee assumes a superposed DC voltage of
less than 500 V.
*3: Values of up to ±1200 V are displayed, but accuracy is not dened for display values in excess of 1000 V
(which are provided as reference values).
*4: Values of up to ±1700 V are displayed, but accuracy is not dened for display values in excess of 1000 V
(which are provided as reference values).
Accuracy guarantee
frequency range *
Measurement accuracy
1 *2
45 Hz ≤ f ≤ 66 Hz±1.5% rdg.±0.07 V
66 Hz < f ≤ 1 kHz±1.8% rdg.±0.07 V
45 Hz ≤ f ≤ 66 Hz±1.5% rdg.±0.7 V
66 Hz < f ≤ 1 kHz±1.8% rdg.±0.7 V
45 Hz ≤ f ≤ 66 Hz±1.5% rdg.±7 V
66 Hz < f ≤ 1 kHz±1.8% rdg.±7 V
45 Hz ≤ f ≤ 66 Hz±1.5% rdg.±7 V
66 Hz < f ≤ 1 kHz±1.8% rdg.±7 V
48
Page 53
Accuracy Table
(9) DC voltage
Measurement value/MAX/MIN/AVE
Range (Auto-range threshold)
600.0 mV
(more than 6000 counts)
6.000 V (more than 6000 counts/
less than 540 counts)
60.00 V (more than 6000 counts/
less than 540 counts)
600.0 V (more than 6000 counts/
less than 540 counts)
1500 V (less than 540 counts)0 V to ±1000 V *
*1: In the 1500 V range, the instrument can withstand input of up to 1000 V continuously or input in excess of
1000 V for no greater than 1 minute.
Accuracy guarantee range
(Resolution)
0.0 mV to ±600.0 mV
(0.1 mV)
0.000 V to ±6.000 V
(0.001 V)
0.00 V to ±60.00 V
(0.01 V)
0.0 V to ±600.0 V
(0.1 V)
(1 V)
±1001 V to ±1700 V *
(1 V)
1
1
Measurement
accuracy
±0.5% rdg.
±0.5 mV
±0.5% rdg.
±0.003 V
±0.5% rdg.
±0.03 V
±0.5% rdg.
±0.3 V
±0.5% rdg.
±3 V
±2.0% rdg.
±5 V
Input impedance (DC
input)
±5%
6.7 M
Ω
±5%
6.7 M
Ω
±5%
6.1 M
Ω
±5%
6.0 M
Ω
6.0 M
±5%
Ω
49
1
2
3
4
5
6
索
引
Page 54
Accuracy Table
PEAK MAX/PEAK MIN
RangeAccuracy guarantee range (Resolution)Measurement accuracy
600.0 mV0 mV to ±1200 mV (1 mV)±1.0% rdg.±7 mV
6.000 V0.00 V to ±12.00 V (0.01 V)±1.0% rdg.±0.07 V
60.00 V0.0 V to ±120.0 V (0.1 V)±1.0% rdg.±0.7 V
600.0 V0 V to ±1000 V (1 V)±1.0% rdg.±7 V
1500 V0 V to ±1000 V (1 V)±1.0% rdg.±7 V
±1001 V to ±1200 V (1 V)±5.0% rdg.±7 V
±1001 V to ±1700 V (1 V)±5.0% rdg.±7 V
50
Page 55
(10) AC+DC voltage
Measurement value/MAX/MIN/AVE
Range
(Auto-range
threshold)
6.000 V (more
than 6000
counts)
60.00 V
(more than
6000 counts/
less than 540
counts)
Accuracy
guarantee range
(Resolution)
0.000 V to 0.299 V
(0.001 V)
0.300 V to 6.000 V
(0.001 V)
3.00 V to 60.00 V
0.01 V)
Accuracy guarantee
frequency range *
10 Hz ≤ f < 45 Hz±1.5% rdg.
DC,
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz±1.5% rdg.
10 Hz ≤ f < 45 Hz±1.5% rdg.
DC,
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz±1.5% rdg.
10 Hz ≤ f < 45 Hz±1.5% rdg.
DC,
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz±1.5% rdg.
Measurement accuracy
1
Filter OFFFilter ON
±0.023 V
±1.0% rdg.
±0.023 V
±0.023 V
±0.013 V
±1.0% rdg.
±0.013 V
±0.013 V
±0.13 V
±1.0% rdg.
±0.13 V
±0.13 V
±2.0% rdg.
±0.023 V
±1.5% rdg.
±0.023 V
–
±2.0% rdg.
±0.013 V
±1.5% rdg.
±0.013 V
–
±2.0% rdg.
±0.13 V
±1.5 %rdg.
±0.13 V
–
Accuracy Table
Input
Ω
Ω
Ω
Ω
Ω
Ω
2
±5%
±5%
±5%
±5%
±5%
±5%
impedance *
DC: 6.7 M
AC: 3.2 M
DC: 6.7 M
AC: 3.2 M
DC: 6.1 M
AC: 3.1 M
51
1
2
3
4
5
6
索
引
Page 56
Accuracy Table
Range
(Auto-range
threshold)
600.0 V
(more than
6000 counts/
less than 540
counts)
1000 V
(less than 540
counts)
*1: Frequency range of 10 Hz ≤ f < 20 Hz is designed value.
*2: At DC input, 50 Hz AC input.
Accuracy
guarantee range
(Resolution)
30.0 V to 600.0 V
(0.1 V)
50 V to 1000 V
(1 V)
Accuracy guarantee
frequency range *
10 Hz ≤ f < 45 Hz±1.5% rdg.
DC,
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz±1.5% rdg.
10 Hz ≤ f < 45 Hz±1.5 %rdg.
DC,45 Hz ≤ f ≤ 66 Hz±1.0 %rdg.
66 Hz < f ≤ 1 kHz±1.5 %rdg.
52
Measurement accuracy
1
Filter OFFFilter ON
±0.7 V
±1.0% rdg.
±0.7 V
±0.7 V
±7 V
±7 V
±7 V
±2.0% rdg.
±0.7 V
±1.5% rdg.
±0.7 V
–
±2.0 %rdg.
±7 V
±1.5 %rdg.
±7 V
–
Input
impedance *
DC: 6.0 M
AC: 3.0 M
DC: 6.0 M
AC: 3.0 M
Ω
Ω
Ω
Ω
2
±5%
±5%
±5%
±5%
Page 57
Accuracy Table
PEAK MAX/PEAK MIN
Range
6.000 V0.00 V to ±12.00 V
60.00 V±3.0 V to ±120.0 V
600.0 V±30 V to ±1000 V *
1000 V±50 V to ±1000 V *
*1: Frequency range of 10 Hz ≤ f < 20 Hz is designed value.
*2: Values of up to ±1200 V are displayed, but accuracy is not dened for display values in excess of 1000 V
(which are provided as reference values).
*3: Values of up to ±1700 V are displayed, but accuracy is not dened for display values in excess of 1000 V
(which are provided as reference values).
Accuracy guarantee range
(Resolution)
(0.01 V)
(0.1 V)
2
(1 V)
(1 V)
3
Accuracy guarantee
frequency range *
10 Hz ≤ f < 45 Hz±1.5% rdg.±0.07 V
DC, 45 Hz ≤ f ≤ 66 Hz±1.0% rdg.±0.07 V
66 Hz < f ≤ 1 kHz±1.5% rdg.±0.07 V
10 Hz ≤ f < 45 Hz±1.5% rdg.±0.7 V
DC, 45 Hz ≤ f ≤ 66 Hz±1.0% rdg.±0.7 V
66 Hz < f ≤ 1 kHz±1.5% rdg.±0.7 V
10 Hz ≤ f < 45 Hz±1.5% rdg.±7 V
DC, 45 Hz ≤ f ≤ 66 Hz±1.0% rdg.±7 V
66 Hz < f ≤ 1 kHz±1.5% rdg.±7 V
10 Hz ≤ f < 45 Hz±1.5% rdg.±7 V
DC, 45 Hz ≤ f ≤ 66 Hz±1.0% rdg.±7 V
66 Hz < f ≤ 1 kHz±1.5% rdg.±7 V
Measurement accuracy
1
53
1
2
3
4
5
6
索
引
Page 58
Accuracy Table
(11) Continuity check
RangeAccuracy guarantee range (Resolution) Measurement current Measurement accuracy
600.0
0.0 Ω to 600.0 Ω (0.1 Ω)200 µA±20%±0.7% rdg.±0.5
Ω
Ω
(12) Resistance
Range (Auto-range threshold)
600.0
Ω
(more than 6000 counts)
6.000 k
Ω
(more than 6000 counts/less than 540 counts)
60.00 k
Ω
(more than 6000 counts/less than 540 counts)
600.0 k
Ω
(less than 540 counts)
Accuracy guarantee
range (Resolution)
0.0
to 600.0 Ω
Ω
)
(0.1
Ω
0.000 k
to 6.000 kΩ
Ω
(0.001 k
0.00 k
(0.01 k
0.0 k
(0.1 k
)
Ω
to 60.00 kΩ
Ω
)
Ω
to 600.0 kΩ
Ω
)
Ω
Measurement
current
200 µA±20%±0.7% rdg.
100 µA±20%±0.7% rdg.
10 µA±20%±0.7% rdg.
1 µA±20%±0.7% rdg.
Measurement
accuracy
±0.5
Ω
±0.005 k
Ω
±0.05 k
Ω
±0.5 k
Ω
(13) Diode
Range Accuracy guarantee range (Resolution)Short-circuit currentMeasurement accuracy
1.800 V 0.000 V to 1.800 V *
*1: Beeping buzzer tone at forward connection (0.15 V to 1.8 V). Continuous buzzer tone and red backlight
lights up if less than 0.15 V.
1
(0.001 V)200 µA±20%±0.7% rdg.±0.005 V
54
Page 59
(14) Capacitance
Range (Auto-range threshold)
1.000 µF
(more than 1100 counts)
10.00 µF
(more than 1100 counts/less than
100 counts)
100.0 µF
(more than 1100 counts/less than
100 counts)
1000 µF
(less than 100 counts)
RangeAccuracy guarantee range (Resolution)Measurement accuracy *
°C
2
°F *
*1: Conditions (In an environment where the temperature of the instrument is ±1°C and stable)
*2: Instrument can be made to display readings in Fahrenheit (°F) by means of special operation.
40.0°C to 400.0°C (0.1°C)±0.5% rdg.±3.0°C
40.0°F to 752.0°F (0.1°F)±0.5% rdg.±5.4°F
1
55
5
6
索
引
Page 60
Accuracy Table
(16) DC power
Voltage range *1
(Input voltage range)
600.0 mV (0.0 mV to ±600.0 mV)0.000 kVA to ±0.600 kVA (0.001 kVA)±2.0% rdg.±0.020 kVA
6.000 V (±0.540 V to ±6.000 V)0.00 kVA to ±6.00 kVA (0.01 kVA)±2.0% rdg.±0.20 kVA
60.00 V (±5.40 V to ±60.00 V)0.0 kVA to ±60.0 kVA (0.1 kVA)±2.0% rdg.±2.0 kVA
600.0 V (±54.0 V to ±600.0 V)0 kVA to ±600 kVA (1 kVA)±2.0% rdg.±20 kVA
1500 V (±540 V to ±1000 V)0 kVA to ±1000 kVA (1 kVA)±2.0% rdg.±20 kVA
1500 V (±1001 V to ±1700 V)0 kVA to ±1700 kVA (1 kVA)±4.0% rdg.±20 kVA
*1: The DC power range is selected automatically based on the voltage range.
Accuracy guarantee range (Resolution) Measurement accuracy
56
Page 61
4
Repairs, Inspections, and Cleaning
4.1 Troubleshooting
SymptomVerication and/or Solution
• The instrument is indicating an
abnormal measured value.
• Is the measured current value too small for the instrument’s
measurement range?
• Wrap the wire around the jaw one or more times. Each additional
wrap of the wire will increase the measured value, so that wrapping
it once yields a measured value that is twice the actual value and
wrapping it twice yields a measured value that is three times the
actual value.
• Is the tip of the jaw open?
• Is the jaw damaged?
• If the jaw is damaged or cracked, it will not be able to measure current
accurately. Send the instrument for repair.
• Displayed values can frequently uctuate due to induction potential
even when no voltage is applied. This, however, is not a malfunction.
57
1
2
3
4
5
6
索
引
Page 62
Troubleshooting
SymptomVerication and/or Solution
• When readings from the
instrument are compared with
those of another clamp-on
current meter, the measured
values differ.
• The current value is larger than
expected.
• A current value is displayed
even though there is no input.
• A sound is being emitted by the
instrument’s jaw.
• The measured value does not
appear.
• No measured value is displayed,
even when the test leads are
shorted.
• Zero adjustment is impossible.
58
• The instrument cannot accurately measure waveforms that contain a
component that falls outside the frequency characteristics range.
• Since the instrument performs true RMS measurement, it can
accurately measure distorted waveforms. When measuring a distorted
waveform, the measured value will differ from a clamp-on current
meter that uses the averaging method.
• The instrument cannot perform measurement accurately in the
presence of a strong magnetic eld from a source such as a nearby
transformer or high-current circuit or in the presence of a strong
electric eld from a source such as a wireless device.
• The jaw may emit sound when measuring AC currents in excess of
approx. 500 A, however, there is no effect on the measurement.
• Check the continuity of the test leads. (p. 24)
If a wiring break is found, replace the test leads.
• Insert the test leads all the way.
Use the proper measurement method.
If no problem can be found, the instrument may be damaged. Send
the instrument for repair.
• When performing current measurement, perform zero-adjustment
while no measurement target is being clamped.
Page 63
4.2 Error display
Error display
Error displayDescriptionSolution
Err 001
Err 002
Err 005
Err 008
ROM error
Program
ROM error
Adjustment data
ADC error
Hardware malfunction
Bluetooth error
Hardware malfunction
(only for model CM4376)
When the error appears in the display,
it is necessary to repair the instrument.
Please contact your authorized Hioki
distributor or reseller.
59
1
2
3
4
5
6
索
引
Page 64
Insert/Replace Batteries
4.3 Insert/Replace Batteries
WARNING
• To avoid electric shock, turn off the instrument and disconnect the test leads
before installing or replacing the batteries.
• Handle and dispose of batteries in accordance with local regulations.
• To prevent instrument damage or an electric shock, use only the screw that
are originally installed for securing the battery cover in place. If you have
lost a screw or nd that a screw is damaged, please contact your authorized
Hioki distributor or reseller.
The indicator lights up when the battery charge diminishes. Replace the batteries as
soon as possible. The batteries may die if the backlight turns on or the buzzer sounds. After
use, be sure to turn off the instrument.
60
Page 65
Cleaning
Screw used to adjust
measured values ×3
(Do not adjust)
4
1
5
Do not adjust any screws other than the screw holding the battery cover in place. Do not adjust
the three screws underneath the cover, which are used to adjust measured values, as doing so
may prevent accurate measurement.
2
3
1
2
3
4
4.4 Cleaning
To clean the instrument, wipe it gently with a soft cloth moistened with water or mild detergent.
61
5
6
索
引
Page 66
Insert/Replace Batteries
62
Page 67
Index
A
AC current ............................................ 17, 42
AC+DC current ..................................... 17, 44
AC+DC voltage ..................................... 17, 51
AC voltage ............................................ 17, 46
Auto hold .............................................. 15, 16
Auto Power Save (APS) ....................... 30, 31
AVG (average) value .................................. 19