Hioki CM4141, CM4142 Instruction Manual

CM4141 CM4142
AC CLAMP METER
Feb. 2019 Edition 1 CM4141A961-00 19-02H
Instruction Manual
EN

Contents

Introduction ..................................................................................................... 1
Notations ........................................................................................................2
Verifying Package Contents ...........................................................................4
Options (sold separately)................................................................................5
Usage Notes ................................................................................................... 6
1 Overview 11
1.1 Product Overview and Features ......................................................... 11
1.2 Part Names ........................................................................................12
1
2
3
2 Making Measurements 13
2.1 Inspection Before Measurement ........................................................13
2.2 Current Measurement ........................................................................14
Manual Hold / Auto Hold ......................................................................15
Switching the range ............................................................................ 18
MAX value/MIN value/AVG value/PEAK value ...................................19
Filter Function.....................................................................................20
Rush current (AC INRUSH)................................................................21
2.3 Other Measurement Functions ........................................................... 22
CM4141A961-00
4
5
6
i
Contents
2.4 Backlight / Auto Power Save (APS) .....................................................26
2.5 Power-on Option Table ....................................................................... 27
2.6 Bluetooth® Communications (only for model CM4142) ...................... 29
3 Specications 35
3.1 GeneralSpecications .......................................................................35
3.2 Inputspecications/Measurementspecications ............................... 37
3.3 Accuracy Table ................................................................................... 44
4 Repairs,Inspections,andCleaning 59
4.1 Troubleshooting .................................................................................. 59
4.2 Error display ....................................................................................... 61
4.3 Insert/Replace Batteries ..................................................................... 62
4.4 Cleaning ............................................................................................. 64
Index 65
WarrantyCerticate
ii

Introduction

Thank you for purchasing the Hioki CM4141, CM4142 AC Clamp Meter. To obtain maximum performance from the instrument over the long term, be sure to read this manual carefully and keep it handy for future reference. Read the separate document “Operating Precautions” carefully before using the instrument.
Target audience
This manual has been written for use by individuals who use the product in question or who teach others to do so. It is assumed that the reader possesses basic electrical knowledge (equivalent to that of someone who graduated from the electrical program at a technical high school).
Trademark
• Bluetooth® is a registered trademark of Bluetooth SIG, Inc.(USA). The trademark is used by HIOKI E.E. CORPORATION under license.
• Android, Google Play, and Google Chrome are trademarks of Google, Inc.
• IOS is a registered trademark of Cisco Systems, Inc. and/or its afliates in the United States and certain other countries.
• iPhone, iPad, iPad miniTM, iPad Pro, and iPod touch are trademarks of Apple Inc.
• The App Store is a service mark of Apple Inc.
• Any other products and company names are generally either trade names, registered trademarks or trademarks of respective companies.
1

Notations

Notations
Concerning Safety
In this document, the risk seriousness and the hazard levels are classied as follows.
DANGER
WARNING
CAUTION
Indicates an imminently hazardous situation that will result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in death or serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the instrument or malfunction.
IMPORTANT
Indicates information related to the operation of the instrument or maintenance tasks with which the operators must be fully familiar.
Indicates prohibited actions.
Indicates an action that must be performed.
Symbols Afxed to the Instrument
Indicates cautions and hazards. Refer to the “Usage Notes” (p. 6) section of the instruction manual and the included “Operating Precautions” for more information.
Indicates that the instrument may be connected to or disconnected from a live conductor.
2
Notations
Screen display
The instrument screen displays the alphanumeric characters as follows.
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
A different display is used in the case below.
1 2 3 4 5 6 7 8 9 0
: Wiring break detected
Other
Indicates a buzzer sound (either intermittent or continuous).
Accuracy
We dene measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values with the
following meanings:
f.s. (maximum display value/range)
The maximum displayable value. This is usually the name of the currently selected range.
rdg. (displayed value)
The value currently being measured and displayed on the measuring instrument.
dgt. (resolution)
The smallest displayable unit on a digital measuring instrument, i.e., the input value that causes
the digital display to show a “1” as the least-signicant digit.
3

Verifying Package Contents

L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
Verifying Package Contents
Model CM4141 or CM4142
AC Clamp Meter
Model L9207-10 Test Lead
Model C0203 Carrying Case
LR03 Alkaline battery ×2
Instruction Manual*
Operating Precautions (0990A907)
Precautions Concerning Use of Equipment that Emits Radio Waves (only for model CM4142)
* Instruction manuals may also be available in other languages.
Please visit our website at http://www.hioki.com
4

Options (sold separately)

9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
L4931 延長ケーブル
L4932(+9207-10cap)
L4933 コンタクトピン
L4935 ワニ口
9243 graber
L4937 マグネットアダプ
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
L4931 延長ケーブル
L4932(+9207-10cap)
L4933 コンタクトピン
L4935 ワニ口
L4936 バスバー
L4937 マグネットアダプ
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
L4931 延長ケーブル
L4932(+9207-10cap)
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
L4931 延長ケーブル
L4932(+9207-10cap)
L4933 コンタクトピン
L4937 マグネットアダプ
L4934 小ワニグチ
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
9243 graber
L4937 マグネットアダプ
DM4910 熱電対
Options (sold separately)
L9207-10 Test Lead
L4930 Connection Cable Set (Length: 1.2 m)
L4931 Extension Cable Set * (Length: 1.5 m, with the coupling connector)
DT4910 Thermocouples (K)
C0203 Carrying Case
*1: CAT IV 600 V/ CAT III 1000 V/ CAT II 1000 V *5: CAT III 300 V/ CAT II 600 V *2: CAT IV 600 V/ CAT III 1000 V *6: 33 V AC/ 70 V DC *3: CAT III 1000 V *7: CAT III 600 V/ CAT II 600 V *4: CAT III 600 V *8: CAT IV 1000 V
1
*
2
*
2
L4933 Contact Pin Set
L4934 Small Alligator Clip Set
L4935 Alligator Clip Set
9243 Grabber Clip
L4936 Bus Bar Clip Set
L4937 Magnetic Adapter Set
9804 Magnetic Adapter *
L4932 Test Pin Set
L4938 Test Pin Set
L4939 Breaker Pin Set
6
*
2
*
3
*
4
*
8
1
*
7
*
5
*
3
*
4
*
5

Usage Notes

Usage Notes
Follow these precautions to ensure safe operation and to obtain the full benets of the various
functions. Read the separate document “Operating Precautions” carefully before using the
instrument. Ensure that your use of the product falls within the specications not only of the
instrument itself, but also of any accessories, options, batteries, and other equipment being used.
DANGER
To prevent an electric shock, do not touch any areas beyond the barrier
while the instrument is in use.
See: “Part Names” (p. 12)
The maximum measurement current varies with the frequency, and the
current that can be measured continuously is limited. Operating the instrument at less than this limitation is referred to as derating. Do not measure currents in excess of the derating curve. Doing so may result in instrument damage or malfunction, a re, or burn due to sensor heating.
• Never apply a voltage to the instrument when the resistance, continuity
check, diode check, capacitance, or temperature functions are selected. Damage to the instrument can cause bodily injury. To avoid electrical accidents, turn off the circuit before measuring it.
6
Usage Notes
DANGER
To prevent an electric shock, conrm that the white portion (insulation layer)
inside the cable is not exposed. If a color inside the cable is exposed, do not use the cable.
WARNING
Do not allow the instrument to get wet, and do not take measurements with wet hands. This may cause an electric shock. (This precaution does not apply to
insulated conductors.)
To prevent an electric shock, do not exceed the lower of the ratings shown on the instrument and test leads.
CAUTION
Do not place any foreign object between the jaws or any insert foreign object into the gap of the sensor head. Doing so may worsen the performance of the sensor or the opening-closing operation of the sensor head.
7
Usage Notes
8
CAUTION
Avoid dropping or jarring the instrument, which could damage the jaw, adversely affecting measurement.
IMPORTANT
Clamp the instrument around only one conductor. The instrument will not be able to make a measurement if you clamp it around two or more wires together, regardless of whether they are part of a single­phase or three-phase circuit.
Test Lead
Usage Notes
WARNING
To prevent an electric shock, when measuring the voltage of a power line use a test lead that satises the following criteria:
• Conforms to safety standards IEC61010 or EN61010
• Measurement category III or IV
Its rated voltage is higher than the voltage to be measured
The optional test leads provided for the instrument conform to the safety standard EN61010. Use a test lead in accordance with its dened measurement
category and rated voltage.
To prevent a short-circuit accident, be sure to use the test leads with the
sleeves attached when performing measurements in the CAT III and CAT IV
measurement categories.
If the sleeves are removed during measurement, stop the measurement.
CAUTION
The cable is hardened in freezing temperatures. Do not bend or pull it to avoid tearing its shield or cutting cable.
9
Usage Notes
Model L4937, 9804 Magnetic Adapter Set (optional)
DANGER
P
ersons wearing electronic medical devices such as a pacemaker should not use the Magnetic Adapter Set. Such persons should avoid even proximity to the Magnetic Adapter Set, as it may be dangerous. Medical device operation could be compromised, presenting a hazard to human life.
CAUTION
• Do not subject the Magnetic Adapter Set to mechanical shock, for example, due to dropping it. Shock can cause it to be chipped or cracked.
• Do not use the Magnetic Adapter Set in locations where it may be exposed to rainwater, dust, or condensation. In those conditions, the Magnetic Adapter Set may be decomposed or deteriorated. The magnet adhesion may be diminished. In such case, the instrument may not be hung in place and may fall.
• Do not bring the Magnetic Adapter Set near magnetic storage device such as
oppy disks, magnetic cards, pre-paid cards, or magnetized tickets. Doing so may
corrupt and may render them unusable. Furthermore, if the Magnetic Adapter Set is brought near precision electronic equipment such as PCs, TV screens, or electronic wrist watches, they may fail.
10
1

Overview

1.1 Product Overview and Features

This instrument is a clamp meter that can perform true RMS measurement of current simply by clamping it around a circuit. In addition to current, it provides voltage measurement, frequency measurement, rush current measurement, resistance measurement, diode measurement, capacitance measurement, and temperature measurement. Model CM4142 also provide Bluetooth® communications functionality, allowing measurement data to be monitored and logged from a mobile device.
Measurement function list
1
Temperature
Capacitance, diode
Continuity check, resistance
AUTO V, AC voltage, DC voltage, AC+DC voltage
AC current
11
Overview

1.2 Part Names

Front Rear
1
2
3
(CM4141)
12
1 Operation grip
2 Jaw open/closed mark
(The jaws are open if the mark is not
4
5
6
7
8
9
10
11
12
showing.)
3 SHIFT key
(Selects function indicated in blue lettering.)
4 Jaw (p. 14)
5 Barrier
6 Serial number (The serial number consists of
9 digits. The rst two (from the left) indicate
the year of manufacture, and the next two indicate the month of manufacture.)
7 Rotary switch
8 HOLD key
9 LCD
10 Battery cover
11 Operation keys
12 Measurement terminals
2
NO
NONO

Making Measurements

2.1 Inspection Before Measurement

Check if there is any damage to the instrument occurred during storage or shipping and
verify that instrument operates normally before using it. If you nd any damage, contact your
authorized Hioki distributor or reseller.
Check item
The battery cover is closed and its screw has
been securely tightened.
There is no damage to the test lead
insulation, and neither the white sheathing nor
metal conductor inside the wire are exposed.
2
There is no foreign matter on the
measurement terminals. (p. 12)
The test leads are not broken.
1 Ω or less
OK
The instrument is neither damaged nor cracked.
No indicators are missing.
(All lit up)
13
Current Measurement

2.2 Current Measurement

1
Turn the rotary switch.
2
Clamp the instrument around a conductor.
e.g.: 3P4W (3-phase, 4-wire) circuit
breaker (AC current measurement)
14
3
Hz
+
(AC A)
(Frequency)Hz(Frequency)
Frequency detection range of AC current
3 A or more (60.00 A range) 30 A or more (600.0 A range) 200 A or more (2000 A range)
Area mark
Jaw center (maximum precision)
Maximize measurement precision by positioning the wire on the center side of the area mark.
Range (p. 18)
Default setting is auto range. Press the RANGE key to switch to manual range.

Manual Hold / Auto Hold

Manual hold Auto hold
Press for 1 sec.
Blinks
Measured value
retains.
Pressing the HOLD key again cancels the measured value hold function.
measured value stabilizes.)
Blinks
Measured value automatically retains.
Pressing the HOLD key for 1 sec. cancels the auto hold function.
(
lights up when
Current Measurement
blinks.
2
Clamp the instrument around a conductor.
Disconnect
15
Current Measurement
Auto hold conditions
Display value updates are stopped when the following two conditions are satised:
• When the measured value exceeds the threshold value described in the table in the next page. (voltage, current). When the measured value is less than the threshold value described in the table in the next page.
(resistance, continuity, diode)
• When the range over which the measured value is uctuating stabilizes within the uctuation range described in the table in the next page.
Held measured value
Measured value
Auto hold e.g.: 100.0 A
Fluctuation range
Fluctuation range
Auto hold e.g.: 99.0 A
Fluctuation range
Threshold e.g.: 12.0 A
Start
End
If the measured value falls below the threshold value (voltage, current) or exceeds the threshold value (resistance, continuity, diode) after display value updates are stopped, the display value update is restarted.
Display value updates will stop if the two conditions are satised once again.
End
Start
Measurement
End
Start
Time
16
Current Measurement
Measurement function Fluctuation range Threshold value
AC current 60.00 A range: within 400 counts
AUTO V AC voltage DC voltage AC+DC voltage
Resistance Continuity
Diode 1.800 V range: within 40 counts 1.800 V range: 1460 counts
The auto hold function only operates for the above measurement functions.
600.0 A range: within 400 counts 2000 A A range: within 40 counts
6.000 V/60.00 V/600.0 V range: within 120 within 1000 V range: within 20 counts 1500 V range: within 30 counts
/6.000 kΩ/60.00 kΩ/
600.0
Ω
range: within 100 counts
600.0 k
Ω
60.00 A range: 100 counts
600.0 A range: 120 counts 2000 A range: 40 counts
6.000 V/60.00 V/600.0 Vrange: 120 counts 1000 V range: 20 counts 1500 V range: 30 counts
600.0
/6.000 kΩ/60.00 kΩ/
Ω
range: 4900 counts
600.0 k
Ω
2
17
Current Measurement

Switching the range

Auto range
(default setting)
18
Manual
60.00 A range
Manual
600.0 A range
Manual
2000 A range

MAX value/MIN value/AVG value/PEAK value

1
Clamp the instrument around a conductor.
Current Measurement
2
3
4
The instrument performs true RMS measurement.
“AVG” indicates the average of all measured values.
See: “Switching the range” (p. 18)
In auto-range mode, the instrument will be automatically set to the 2000 A range.
Press for 1 sec.
Measured value retains.
PEAK MIN
PEAK MAX
MAX
MIN
Waveform
Measured value (RMS value)
Display refresh interval
Cancel
2
19