Hioki CM3286, CM3286-01 Instruction Manual

Contents
Introduction ............................................................................................................... 1
Options (sold separately) ........................................................................................ 4
Safety Notes .............................................................................................................. 5
Usage Notes .............................................................................................................. 9
Part Names .............................................................................................................. 15
Operation keys........................................................................................................ 16
■Rotary switch ................................................................................................... 17
■Power-on Option Table (buzzer sound, resetting the instrument to the
factory settings, etc.) ....................................................................................... 18
Insert / Replace Batteries....................................................................................... 20
Inspection Before Measurement ........................................................................... 21
Screen / Basic Operation ....................................................................................... 22
■Screen display ................................................................................................. 23
■Switching the information shown on the measurement display ....................... 24
CM3286A961-02
www. .com
i
information@itm.com1.800.561.8187
Contents
Connecting the Clamp and Clips .......................................................................... 27
Current/Voltage Measurement (Frequency)
Power Measurement (Power/Power Factor)......................................................... 29
■AC single-phase measurement (1P2W)
■AC single-phase measurement (1P3W)
■AC 3-phase measurement (3P3W, balanced)
■AC 3-phase measurement (3P3W, unbalanced)
■AC 3-phase measurement (3P4W, balanced)
■AC 3-phase measurement (3P4W, unbalanced)
Phase Detection [Phase Detect]........................................................................... 38
Single-phase Active Energy Measurement (Integrated Measurement)
[Setting Wh] ............................................................................................................ 40
Single-phase Energy Meter Comparison Function [Setting Wh] ...................... 42
■Setting the desired meter constant .................................................................. 46
Manual Hold / Auto Hold ........................................................................................ 48
Switching Ranges ................................................................................................... 51
MAX/ MIN/ AVG ........................................................................................................ 52
[ ] .................................28
[var VA W] ............................................... 29
[var VA W] ............................................... 30
[3PW] ............................................ 31
[3PW] ....................................... 32
[3PW] ............................................ 34
[3PW] ....................................... 35
ii
www. .com
information@itm.com1.800.561.8187
Contents
Backlight / Auto Power Save (APS) ...................................................................... 55
Measurement Using the Clamp Adapter .............................................................. 56
Bluetooth
Repairs, Inspections, and Cleaning ...................................................................... 63
Troubleshooting ..................................................................................................... 65
Specications ......................................................................................................... 71
Accuracy Table ....................................................................................................... 79
Equations ................................................................................................................ 91
®
Communications (only for CM3286-01) ............................................. 57
■Installing the application software GENNECT Cross....................................... 58
■Turning on the Bluetooth
■Pairing the app with the CM3286-01 ............................................................... 60
■Making measurements with the Bluetooth
■General Specications .................................................................................... 71
■Input/Output/Measurement Specications ...................................................... 73
■Harmonic Measurement Specications (only for CM3286-01) ........................ 76
■External Interface (Bluetooth®) Specications ................................................. 78
■Application Software Specications ................................................................ 78
®
function .................................................................. 59
®
function ....................................... 61
iii
www. .com
information@itm.com1.800.561.8187

Introduction

Introduction
Thank you for purchasing the Hioki CM3286/CM3286-01 AC Clamp Power Meter. To obtain maximum performance from the instrument over the long term, be sure to read this manual carefully and keep it handy for future reference.
This clamp power meter provides functionality for measuring AC current, voltage, power, and frequency as well as for detecting phase. The CM3286-01 also provides Bluetooth® communications functionality, allowing measurement data to be monitored and logged from a smartphone or tablet.
The instrument screen displays the alphanumeric characters as follows.
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
1 2 3 4 5 6 7 8 9 0
1
www. .com
information@itm.com1.800.561.8187
Introduction
Package contents
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is
evident, or if it fails to operate according to the specications, contact your authorized Hioki
distributor or reseller.
Model CM3286 or CM3286-01
AC Clamp Power Meter
Accessories
Model L9257 Connection Cord
(Model L4930 Connection Cable Set (1.2 m) + Model L4935 Alligator Clip Set) CAT IV 600 V/ CAT III 1000 V/ CAT II 1000 V
Model C0203 Carrying Case
LR03 Alkaline battery ×2
Instruction Manual
Precautions Concerning Use of Equipment
That Emits Radio Waves (only for the CM3286-01)
(p. 8)
Precautions during shipment
Handle it carefully so that it is not damaged due to a vibration or shock.
2
www. .com
information@itm.com1.800.561.8187
Introduction
Trademarks
• Bluetooth® is a registered trademark of Bluetooth SIG, Inc.(USA). The trademark is used by HIOKI E.E. CORPORATION under license.
• Android and Google Play are trademarks of Google, Inc.
• IOS is a registered trademark of Cisco in the U.S. and other countries.
• iPhone, iPad, iPad mini, iPad Pro, and iPod Touch are trademarks of Apple Inc.
Accuracy
We dene measurement tolerances in terms of f.s. (fullscale), rdg. (reading), and dgt. (digit) values, with the
following meanings:
f.s. (maximum display value or range)
rdg. (displayed value) The value currently being measured and indicated on the measuring instrument.
dgt. (resolution)
The maximum displayable value. This is usually the name of the currently selected range.
The smallest displayable unit on a digital measuring instrument, i.e., the input value
that causes the digital display to show a "1" as the least-signicant digit.
3
www. .com
information@itm.com1.800.561.8187

Options (sold separately)

9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
L4931 延長ケーブル
L4932(+9207-10cap)
L4933 コンタクトピン
L4935 ワニ口
9243 graber
L4937 マグネットアダプ
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
L4931 延長ケーブル
L4932(+9207-10cap)
L4933 コンタクトピン
L4935 ワニ口
L4936 バスバー
L4937 マグネットアダプ
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
L4931 延長ケーブル
L4932(+9207-10cap)
9243 graber
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
L9207-10
DM4910 熱電対
DT4911TestLead
DT4912TestLead
L4930 接続ケーブル
L4931renketu
L4931 延長ケーブル
L4932(+9207-10cap)
L4933 コンタクトピン
L4937 マグネットアダプ
L4934 小ワニグチ
L4937 マグネットアダプ
L4934 小ワニグチ
マグネ付ストラップ
Options (sold separately)
L4933 Contact Pin Set
L9207-10 Test Lead
1
*
L4934 Small Alligator Clip Set
L4935 Alligator Clip Set
L4930 Connection Cable Set
(1.2 m)
2
*
9243 Grabber Clip
L4936 Bus Bar Clip Set
L4937 Magnetic Adapter Set
L4931 Extension Cable Set
2
*
(1.5 m, with the coupling connector)
9804 Magnetic Adapter
L4932 Test Pin Set
L4938 Test Pin Set
( : See p. 8)
*1: CAT IV 600 V/ CAT III 1000 V/ CAT II 1000 V *6: 30 V AC/ 60 V DC *2: CAT IV 600 V/ CAT III 1000 V *7: CAT III 600 V/ CAT II 600 V *3: CAT III 1000 V *8: Can also be connected to the tip of *4: CAT III 600 V *5: CAT III 300 V/ CAT II 600 V
the L4932.
4
www. .com
8
*6,*
2
*
3
*
4
*
*
3
*
1
*
7
*
L4939 Breaker Pin Set
9290-10 Clamp On Adapter
information@itm.com1.800.561.8187
*5,*
3
8
4
*
4
*

Safety Notes

Safety Notes
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, using the instrument in a way not described in this manual may negate the provided safety features. Before using the instrument, be certain to carefully read the following safety notes.
DANGER
Mishandling during use could result in injury or death, as well as damage to the instrument. Be certain that you understand the instructions and precautions in the manual before use.
WARNING
With regard to the electricity supply, there are risks of an electric shock, a heat
generation, a re, and an arc ash due to short‑circuit. Individuals using an electrical measuring instrument for the rst time should be supervised by a
technician who has experience in electrical measurement.
5
www. .com
information@itm.com1.800.561.8187
Safety Notes
WARNING
Protective gear This instrument is measured on a live line. To prevent an electric shock, use
appropriate protective insulation and adhere to applicable laws and regulations.
Notation
In this document, the risk seriousness and the hazard levels are classied as follows.
Indicates information related to the operation of the instrument or maintenance tasks with which the operators must be fully familiar.
Indicates prohibited actions.
Indicates the action which must be performed.
DANGER
WARNING
CAUTION
Indicates an imminently
hazardous situation that will result
in death or serious injury to the operator.
Indicates a potentially hazardous
situation that may result in death or serious injury to the operator.
Indicates a potentially hazardous
situation that may result in minor or moderate injury to the operator or damage to the instrument or malfunction.
IMPORTANT
6
www. .com
information@itm.com1.800.561.8187
Symbols afxed to the instrument
Indicates cautions and hazards.
When the symbol is printed on the instrument, refer to a corresponding topic in the Instruction Manual.
Indicates that dangerous voltage may be present at this terminal.
Indicates that the instrument may be connected to or disconnected from a live conductor.
Indicates a instrument that has been protected throughout by double insulation or reinforced insulation.
Indicates that the product incorporates
®
wireless technology.
FCC ID
Bluetooth
Indicates the ID number of the
wireless module certied by the U.S.
Federal Communications Commission (FCC).
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates a grounding terminal.
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.
Indicates that the product conforms to regulations set out by the EU Directive.
Indicates the identication number of a
IC
wireless module approved by Industry Canada (IC).
Safety Notes
7
www. .com
information@itm.com1.800.561.8187
Safety Notes
Measurement categories
To prevent an electric shock, do not exceed the lower of the ratings shown on
the instrument and connecting cords.
Distribution panel
Service entrance
Service drop
CAT IV
(≤600 V)
Power meter
8
DANGER
Internal wiring
CAT III
(≤1000 V)
Fixed installation
CAT II
(≤1000 V)
T
Outlet
www. .com
information@itm.com1.800.561.8187

Usage Notes

Usage Notes
Follow these precautions to ensure safe operation and to obtain the full benets of the various
functions.
DANGER
• Do not use the instrument with circuits that exceed its
ratings or specications. Doing so may damage the
instrument or cause it to become hot, resulting in bodily injury.
• The instrument must not be used to measure current in
high‑voltage lines (1000 V or more). Attempting to do so could cause a short‑circuit or accident resulting in injury
or death. Also, do not perform measurement around a bare conductor.
To prevent an electric shock, do not touch any areas beyond the barrier while the instrument is in use.
Barrier
9
www. .com
information@itm.com1.800.561.8187
Usage Notes
10
DANGER
Do not short‑circuit two wires to be measured by bringing the clip or jaw
tip of the connecting cords into contact with them. Arcs or such grave
accidents are likely to occur.
To prevent an electric shock, be careful to avoid shorting live lines with the connecting cords tip.
To prevent a short‑circuit or an electric shock, do not touch the metal part of the connecting cords tip.
The maximum measurement current varies with the frequency, and the current that can be measured continuously is limited. Operating the instrument at less than this limitation is referred to as derating. Do not
measure currents in excess of the derating curve. Doing so may result in instrument damage or malfunction, a re, or a burn due to sensor heating.
• It is recommended to make measurements on the secondary side of the distribution panel. Making measurements on the primary side of the panel,
where currents are higher, poses a higher risk of instrument or equipment damage in the event of a short‑circuit.
www. .com
information@itm.com1.800.561.8187
Usage Notes
WARNING
• Installing the instrument in inappropriate locations may cause a malfunction
of instrument or may give rise to an accident. Avoid the following locations.
• Exposed to direct sunlight or high temperature
Exposed to corrosive or combustible gases
Exposed to a strong electromagnetic eld or electrostatic charge
Near induction heating systems (such as high‑frequency induction heating
systems and IH cooking equipment)
Susceptible to vibration
Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
Exposed to high quantities of dust particles
• Although this instrument is designed to resist the ingress of dust and
dripping water, it is not entirely waterproof or dustproof. Therefore, to prevent an electric shock, do not use it in a wet or dusty environment.
Battery may explode if mistreated. Do not short‑circuit, recharge, disassemble or dispose of in re.
11
www. .com
information@itm.com1.800.561.8187
Usage Notes
Use only the specied connection cords. Use of any connection cord not
To prevent an electric shock, set the rotary switch to the OFF position,
To prevent instrument damage or an electric shock, use only the screw for
Options may include connection cords which uses sleeves. To prevent
If the sleeves are inadvertently removed during measurement, stop the
12
WARNING
specied by our company does not allow safe measurements.
disconnect all connection cords, and remove the instrument from the
measurement object before replacing the batteries.
securing the battery cover in place that is originally installed. If you have lost a screw or nd that a screw is damaged, please contact your authorized Hioki
distributor or reseller for replacement.
a short‑circuit accident, be sure to use the connection cords with the sleeves attached when performing measurements in the CAT III or CAT IV
measurement categories. (See "Measurement categories" (p. 8))
measurement.
www. .com
information@itm.com1.800.561.8187
Usage Notes
CAUTION
• To avoid damage to the instrument, protect it from physical shock when transporting and handling it. Be especially careful to avoid physical shock due to dropping it.
• Do not place foreign objects between jaws or insert foreign objects into the gaps of the sensor head. Doing so may worsen the performances of the sensor or interfere with clamping action.
• Poor performance or damage from battery leakage could result. Observe the cautions listed below.
• Do no mix old and new batteries, or different types of batteries.
• Be careful to observe the battery polarity during installation.
• Do not use batteries after their recommended expiry date.
• Do not leave depleted batteries inside the instrument.
Replace batteries only with the specied type.
• Keep the jaw closed when not in use, to avoid accumulating dust or dirt on the facing core surfaces, which could interfere with clamp performance.
The cord is hardened in freezing temperature. Do not bend or pull it to avoid tearing its shield or cutting cord.
IMPORTANT
Inverter secondary-side waveforms and waveforms that include a large noise component may not be measured accurately.
13
www. .com
information@itm.com1.800.561.8187
Usage Notes
Current measurement precautions
14
Do not touch.
NO
Do not pinch wire between jaws.
NO
Do not clamp around two wires.
NO
Do not input
excessively high currents or voltages.
NO
(Flashes red)
www. .com
information@itm.com1.800.561.8187

Part Names

Front
Lever
(CM3286-01)
Jaw
Barrier*
Rotary switch
Part Names
Rear
Battery cover
(p. 12)
HOLD key
LCD
* Do not touch any areas beyond the barrier while the instrument is in use.
SHIFT key
(Selects function indicated in blue lettering.)
Operation keys
Voltage input terminal
(p. 9)
L9257 Connection Cord
Alligator clip
(p. 8)
Barrier*
Connector
Cord
Sleeve
15
www. .com
information@itm.com1.800.561.8187

Operation keys

Operation keys
Key Short press Long press (1 sec.)
Activates/cancels manual hold operation p. 48
HOLD
SHIFT
RANGE
MAX/MIN
16
Start/stop integration, clears the integrated energy value (during energy measurement) Switches the setting (when setting meter constants) Switches from the connection display to the measurement display (during 3-phase power measurement)
Switches the information shown on the measurement display
Switches ranges p. 51
Count up (when setting meter constants) p. 46
Displays and switches MAX/MIN/AVG value
Count down (when setting meter constants)
Toggles the display backlight on and off p. 55
p. 41
Activates/cancels automatic hold
p. 46
operation
p. 22
Switches between 3-phase/3-wire and 3-phase/4-wire measurement
p. 24
during 3-phase power measurement (setting is not stored)
High speed count up (when setting meter constants)
p. 52
Cancels the display of MAX/MIN/AVG value
High speed count down (when setting
p. 46
meter constants)
Enables/disables external communications (Bluetooth®) (only for CM3286-01, setting is stored)
p. 48
p. 34 p. 35
p. 52
p. 59
www. .com
information@itm.com1.800.561.8187

Rotary switch

"Phase Detection [Phase Detect]" (p. 38)
• "AC 3-phase measurement (3P3W, balanced) [3PW]" (p. 31)
• "AC 3-phase measurement (3P3W, unbalanced) [3PW]" (p. 32)
• "AC 3-phase measurement (3P4W, balanced) [3PW]" (p. 34)
• "AC 3-phase measurement (3P4W, unbalanced) [3PW]" (p. 35)
• "Single-phase Active Energy Measurement (Integrated Measurement) [Setting Wh]" (p. 40)
• "Single-phase Energy Meter Comparison Function [Setting Wh]" (p. 42)
• "AC single-phase measurement (1P2W) [var VA W]" (p. 29)
• "AC single-phase measurement (1P3W) [var VA W]" (p. 30)
"Current/Voltage Measurement (Frequency) [ ]" (p. 28)
Turns off the instrument.
Operation keys
When functions other than OFF is selected, the instrument turns on. Select the desired function.
17
www. .com
information@itm.com1.800.561.8187
Operation keys
Power‑on Option Table (buzzer sound, resetting the instrument to the
factory settings, etc.)
+
Switching the auto power save (APS) function (p. 55)
Displaying product information or displaying all indicators
(Display varies depending on the
position of the rotary switch.)
Switching between balanced and unbalanced operation (during AC
3‑phase power measurement)
Buzzer sound (ON/OFF)
Switching the auto backlight off
function (p. 55)
Turn on the power while pressing the operation key. (Turn the rotary switch from OFF.)
Setting See Operating instruction
18
Factory
setting
3PW: Serial number
Wh: Model number W: Version of software Besides the above: Displays all indicators
p. 32
p. 35
ON Yes
3PW
retained?
ON
ON Yes
Setting
www. .com
information@itm.com1.800.561.8187
Setting See Operating instruction
Operation keys
Factory
setting
Setting
retained?
Selecting the CT ratio p. 56
Reset to the factory setting
1/1 Ye s
19
www. .com
information@itm.com1.800.561.8187

Insert / Replace Batteries

Insert / Replace Batteries
Fully charged.
As the battery charge diminishes, black charge bars disappear, one by one, from the left of the battery indicator.
The battery voltage is low. Replace the batteries as soon as possible.
The instrument may lose power when the backlight turns on, when a buzzer sounds, etc.
The battery is exhausted. Replace with new batteries.
(Flashes)
Required items: No. 2 Phillips screwdriver and LR03 Alkaline battery ×2 Recommended screw tightening torque: 0.7 N・m
Exercise care to orient batteries properly.
3
20
4
1
2
5
www. .com
information@itm.com1.800.561.8187

Inspection Before Measurement

Inspection Before Measurement
Verify that the instrument operates normally to ensure that no damage occured during storage
or shipping. If you nd any damage, contact your authorized Hioki distributor or reseller.
Check item
The battery cover is closed and its screw
has been securely tightened.
There is no foreign matter on the voltage
input terminals. (p. 15)
The battery voltage (p. 20) is sufcient.
There is no damage to the connection cords insulation,
and neither the white sheathing nor metal conductor inside the wire are exposed.
The instrument is neither damaged nor cracked.
No indicators are missing.
(All indicators displayed)
21
www. .com
information@itm.com1.800.561.8187

Screen / Basic Operation

Screen / Basic Operation
Setting the rotary switch to a position other than OFF causes the instrument to turn on and the screen to activate. e.g.: During balanced 3-phase 3-wire active power measurement
Connection display Measurement display
Displays the connection locations for 3-phase power measurement and phase detection. Connects the wire.
Wire around which to clamp instrument
Black Red
Wires to which to clip connection cords
IMPORTANT
If measured with a wrong wire connection, a correct value does not appear.
22
Displays the measured value.
www. .com
information@itm.com1.800.561.8187

Screen display

(All indicators displayed)
"Error display" (p. 67) "Warning display" (p. 67)
* The key lock feature may be activated
according to the usage state of the application software.
Screen / Basic Operation
Appearing: Bluetooth® communications function enabled Flashing: Bluetooth (only for CM3286-01)
,
Connection type (not shown during single­phase measurement)
Unbalanced mode operation (not shown during balanced mode operation)
Manual range operation (not shown during auto-range operation)
CT ratio (not shown during 1/1)
Measured value held
Auto power-off enabled
Current RMS value or voltage RMS value exceeded range
Key lock enabled*
®
communications active
23
www. .com
information@itm.com1.800.561.8187
Screen / Basic Operation

Switching the information shown on the measurement display

Able to switch using the SHIFT key (Excluding the and functions).
How to use this chart:
(Sub1 display)
FREQ
U
(Main display)
I
RMS
(Sub2 display)
(Sub3 display)
I
|PEAK|
Sub1 display
Main display
Rotary switch
Current (frequency)
Voltage (frequency)
FREQ
: Current frequency
I
FREQ
: Voltage frequency
U
24
Current Voltage
FREQ
I
I
RMS
I
|PEAK|
I
RMS
U
RMS
FREQ
U
U
RMS
U
|PEAK|
: Current RMS value
: Voltage RMS value
Sub2 display
I
: Current peak value
|PEAK|
U
: Voltage peak value
|PEAK|
(All indicators displayed)
Sub3 display
www. .com
information@itm.com1.800.561.8187
Active power
Rotary switch
(Main display)
Voltage/Power
factor
PF PF PF P P
P S Q PF
U
RMSIRMS
PF
Balance
mode*
1
P
U
RMSIRMS
P
Unbalance
mode*
P
: Single-phase active power
S
: Single-phase apparent power
Q
: Single-phase reactive power
PF
: Power factor
: Zero-cross phase angle 1
φ
1
P
P
3
+
1
P
1
Screen / Basic Operation
Apparent power Reactive power Power factor Zero-cross phase angle
U
U
3
P
S
2
P1
: Active power 1
S
: Apparent power 1
1
Q
: Reactive power 1
1
PF
: Power factor 1
1
: Zero-cross phase
φ
2
angle 2
RMSIRMS
2
PF
*
S
3P
RMSIRMS
S
3
S
S
+
+
1
2
1
2
*
3P
3
P
+
2
U
RMSIRMS
2
PF
*
Q
3P
U
RMSIRMS
Q
3
S
3
S
Q
Q
+
+
1
2
Q
2
1
P2
: Active power 2
S2
: Apparent power 2
Q2
: Reactive power 2
PF2
: Power factor 2
: Zero-cross phase
φ
3
angle 3
U
RMSIRMS
P
3P
*2
PF
3P
U
RMSIRMS
PF
3
P
P
P
+
+
1
2
Q
3
S
Q
PF1PF
2
3
S
S
+
+
1
2
3
2
P3
: Active power 3
S3
: Apparent power 3
Q3
: Reactive power 3
PF3
: Power factor 3
φ
U
RMSIRMS
P
3P
φ
3P
U
RMSIRMS
φ
φ
1
*2
3
φ
2
25
www. .com
information@itm.com1.800.561.8187
Screen / Basic Operation
P3P:
Balanced 3-phase active power
S
:
Balanced 3-phase apparent power
3P
Q
:
Balanced 3-phase reactive power
3P
PF
:
Balanced 3-phase power factor
3P
: Zero-cross phase angle
φ
Notes(*) for table
*1 Value of the measured phase will be calculated and displayed. *2 Different calculation methods are used for 3-phase/3-wire and 3-phase/4-wire circuits. For more
information, see the list of equations.
*3 Only 3-phase active power (
3-phase/3-wire circuits.
P
P
+
), active power 1 (
1
2
P
:
P
P
Unbalanced 3-phase active power
+
+
1
2
3
S
:
S
S
Unbalanced 3-phase apparent power
+
+
1
2
3
Q
:
Q
Q
Unbalanced 3-phase reactive power
+
+
1
2
3
P
P
P
+
+
1
2
3
S
φ
3P
P
), and active power 2 (
1
: Unbalanced 3-phase power factor
S
S
+
+
1
2
3
: 3-phase zero-cross phase angle
P
) are measured for
2
26
P
P
+
1
2
P
P
2
1
www. .com
information@itm.com1.800.561.8187

Connecting the Clamp and Clips

Connecting the Clamp and Clips
Clamp
Current direction
Load
side
Power supply
Align the current direction mark with the direction of the current.
side
Clip (Alligator)
Connect to metal part.
Clip (Magnetic adapter)
Connect to metal part.
(If unable to connect the magnetic adapter so that it sits perpendicular to the terminal due to the weight of the voltage cord, connect it at an angle so as to balance it against the weight of the cord.)
27
www. .com
information@itm.com1.800.561.8187
Current/Voltage Measurement (Frequency) [ ]
N

Current/Voltage Measurement (Frequency)

Current measurement
1
L/R/A
Clamp
*
"Switching the information shown on the measurement display" (p. 24)
If the screen turns red: "Warning display" (p. 67)
The frequency display ashes
when frequency exceeds
999.9 Hz.
28
2
Power supply side
Voltage measurement
Red
L/R/A
Clip
2
Black
Power supply side
N
Load side
Load side
[ ]
Measurement display*
Measurement display*
www. .com
information@itm.com1.800.561.8187

Power Measurement (Power/Power Factor)

Power Measurement (Power/Power Factor)
AC single‑phase measurement (1P2W)
Clamp
2
1
and clip
L/R/A
Power supply side
*
"Switching the information shown on the measurement display" (p. 24)
If the screen turns red: "Warning display" (p. 67)
N
[var VA W]
Red
Black
Load side
Measurement display*
29
www. .com
information@itm.com1.800.561.8187
Power Measurement (Power/Power Factor)
AC single‑phase measurement (1P3W)
Clamp and clip
1
2
L1
Black
N
Power supply side
L2
*
"Switching the information shown on the measurement display" (p. 24)
If the screen turns red: "Warning display" (p. 67)
30
[var VA W]
Red
Measurement display*
Black
Load side
Red
3
Clamp and clip
www. .com
information@itm.com1.800.561.8187
Power Measurement (Power/Power Factor)
AC 3‑phase measurement (3P3W, balanced)
Clamp and clip
1
2
L1/R/A
Connection display (p. 22)
Black
[3PW]
Red
L2/S/B
Power supply side
L3/T/C
3
*
"Switching the information shown on the measurement display" (p. 24)
If the screen turns red: "Warning display" (p. 67)
When the balanced 3-phase 3-wire zero-cross phase angle is less than −90° or exceeds 90°, the measured
value appears “ – – – – ”.
Measurement display (calculation result)*
Load side
31
www. .com
information@itm.com1.800.561.8187
Power Measurement (Power/Power Factor)
AC 3‑phase measurement (3P3W, unbalanced)
1
Connection display 1 (p. 22)
Clamp and clip
2
L1/R/A
Red
Black
L2/S/B
Power supply side
L3/T/C
• Proceed to the next step after verifying that the measured values shown on measurement displays 1 and 2 are normal.
• If the screen turns red: "Warning display" (p. 67)
3
Measurement display (active power 1)
4
32
[3PW]
Clamp
5
and clip
Black
Red
Connection display 2
Load side
To next
page
www. .com
information@itm.com1.800.561.8187
Measurement display (active power 2)
6 7
3-phase active power
Active power 1
Active power 2
The measured value is cleared and returns to the initial connection display.
Power Measurement (Power/Power Factor)
Measurement display (calculation result) (3-phase active power)
When the active power is negative, “ − ” appears beside it.
Displays alternately
8
Long press
33
www. .com
information@itm.com1.800.561.8187
Power Measurement (Power/Power Factor)
AC 3‑phase measurement (3P4W, balanced)
1
More than 1 sec.
Connection display (p. 22)
2
L1/R/A
L2/S/B
L3/T/C
Power supply side
N
Clamp and clip
Red
Black
[3PW]
Measurement display (calculation result)*
3
If the screen turns red: "Warning display" (p. 67)
* "Switching the information shown on the measurement display" (p. 24)
34
Load side
www. .com
information@itm.com1.800.561.8187
Power Measurement (Power/Power Factor)
AC 3‑phase measurement (3P4W, unbalanced)
Clamp and clip
1
More than 1 sec.
Connection display 1 (p. 22)
2
Red
L1/R/A
L2/S/B
L3/T/C
Power supply side
N
Black
5
Red
Black
3
Measurement display 1
Proceed to the next step after verifying that the measured values shown on measurement display 1 is normal.
(active power 1)
4
[3PW]
Clamp and clip
Connection display 2
Load side
6
To next page
35
www. .com
information@itm.com1.800.561.8187
Power Measurement (Power/Power Factor)
Measurement display 2
(active power 2)
L1/R/A
L2/S/B
L3/T/C
Power supply side
N
7
8
Red
Black
Clamp and clip
Load side
Connection display 3
To
8
Proceed to the next step after verifying that the measured values shown on measurement displays 2 and 3 are normal.
36
9
Measurement display 3
(active power 3)
To next page
www. .com
information@itm.com1.800.561.8187
Measurement display 4 (calculation result) (3-phase active power)
Power Measurement (Power/Power Factor)
10
Displays
alternately
When the active power is negative, “ − ” appears beside it.
The measured value is cleared and returns to the initial connection display.
You can switch the information shown on the nal measurement display with the SHIFT key.
See "Switching the information shown on the measurement display" (p. 24)
• If the screen turns red: "Warning display" (p. 67)
11
Long press
Active power 3
3-phase active power
Active power 2
Active power 1
37
www. .com
information@itm.com1.800.561.8187

Phase Detection [Phase Detect]

Phase Detection
1
Connection display (p. 22)
• The instrument will display “– – – –” if open phase is detected or if it is unable to make a measurement.
• When the input is unstable, the second connection display will not show up.
* If not clipped within 10 seconds, it
is unable to make a measurement.
38
[Phase Detect]
Clip
2
L1/R/A
L2/S/B
Black
Power supply side
L3/T/C
Measurement display
Line voltage (1st measurement)
Red
Clip*
3
Black
Red
Connection display
Countdown display
Load side
To next page
www. .com
information@itm.com1.800.561.8187
Measurement display
Display of results
Normal phase
Phase Detection [Phase Detect]
Reverse phase
Display appears in the order of the arrow.
Line voltage (2nd measurement)
Goes back to the rst display when the HOLD key is pressed.
(Lights red)
39
www. .com
information@itm.com1.800.561.8187

Single-phase Active Energy Measurement (Integrated Measurement) [Setting Wh]

Single‑phase Active Energy Measurement
(Integrated Measurement)
1
Set the constant number to OFF.
2
or
40
3
L/R/A
Power supply side
N
[Setting Wh]
Clamp and clip
Red
Black
Load side
To next page
www. .com
information@itm.com1.800.561.8187
Single-phase Active Energy Measurement (Integrated Measurement) [Setting Wh]
Set the instrument to power integration mode.
4
Start integration.
5
Stop integration.
6
• When the HOLD key is pressed during integration stop, the integrated energy clears and returns to the display shown in Step 4.
• The measured values are automatically stored just before the instrument turns off due to low battery voltage. Next time the instrument is turned on, the saved values will be displayed. (The measured values can be cleared by pressing the HOLD key.)
Range is xed when integration starts. Change the range before starting to integrate a large power quantity or for an extended period of time.
• Only the positive (consumption) active power is added. Negative (generation) active power is not added.
Single-phase active power
Single-phase active energy
Elapsed time
41
www. .com
information@itm.com1.800.561.8187
Single-phase Energy Meter Comparison Function [Setting Wh]
Single‑phase Energy Meter Comparison Function
[Setting Wh]
This function allows you to compare the actual energy value (measured value) from an energy meter with the theoretical value.
There are two ways to start and stop integration:
• Start/stop at 1 cycle based on the energy meter’s instrument constant: 1-cycle mode
Start/stop based on a xed amount of energy as measured by the energy meter: Fixed energy mode
IMPORTANT
Energy may not be calculated properly in the following circumstances:
• If the power line of the measurement object, instrument connection, or meter constant (for the watt-hour meter) is set incorrectly.
• If integration is not started and stopped as described above.
• If the instrument is being used outside its operating temperature and humidity range.
• If the instrument is being used in close proximity to a device that emits powerful electromagnetic radiation or a device that carries an electrical charge.
If the instrument is being used in close proximity to a device that emits a strong magnetic eld, for example a transformer, high-current circuit, or wireless device.
42
www. .com
information@itm.com1.800.561.8187
1‑cycle mode
Mechanical meter
Once the disc has completed one revolution
Single-phase Energy Meter Comparison Function [Setting Wh]
Electronic meter
When the LED ashes once
Disc
Fixed energy mode
e.g.: With the xed energy set to 0.1 kWh
0.1 kWh digit
Start integration
Stop integration
LED
Start integration
Stop integration
Start integration
Stop integration
43
www. .com
information@itm.com1.800.561.8187
Single-phase Energy Meter Comparison Function [Setting Wh]
Clamp and clip
1
Select the constant number or
2
xed energy*.
or
3
L/R/A
Power supply side
N
* Select xed energy mode if the energy meter’s disc rotates, or if
its LED ashes, quickly.
The constant can be changed. "Watt-hour meter constants default setting value" (p. 47) "Setting the desired meter constant" (p. 46)
Black
44
Red
Load side
To next page
www. .com
information@itm.com1.800.561.8187
Single-phase Energy Meter Comparison Function [Setting Wh]
Set the instrument to power
4
integration mode.
Start integration.
5
• Only the positive (consumption) active power is added. Negative (generation) active power is not added.
Range is xed when integration starts. Change the range before starting to integrate a large power quantity or for an extended period of time.
Stop integration.
6
When the HOLD key is pressed during integration stop, the integrated energy clears and returns to the display shown in Step 4.
Single-phase active power (measured value)
Elapsed time
Energy (theoretical value)
Energy ratio (Measured value/theoretical value)
Energy difference (Measured value − theoretical
value)
Energy (Measured value)
45
www. .com
information@itm.com1.800.561.8187
Single-phase Energy Meter Comparison Function [Setting Wh]

Setting the desired meter constant

Set after conducting the procedures 1 and 2 of "Single-phase Energy Meter Comparison Function [Setting Wh]" (p. 42)
Select the portion you wish to change.
1
Change the value.
2
or
46
The selected portion will ash.
Set the value shown on the meter.
The set value will be stored.
• Press the SHIFT key to go back to the measurement
display.
The changed nal value will be the setting value.
• "Watt-hour meter constants default setting value" (p. 47)
www. .com
information@itm.com1.800.561.8187
Single-phase Energy Meter Comparison Function [Setting Wh]
Watt‑hour meter constants default setting value
No.01 to No.10: 1-cycle mode
0.10 kWh to 0.01 kWh: Fixed energy mode
Changing the
SET No. Setting value
oFF None (single-phase
energy measurement)
01 3200 cyc./1 kWh
02 1600 cyc./1 kWh
03 1200 cyc./1 kWh
04 1000 cyc./1 kWh
05 600 cyc./1 kWh
06 500 cyc./1 kWh
Updated settings are stored by the instrument.
setting value
Enable:
Disable: –
SET No. Setting value
07 300 cyc./1 kWh
08 250 cyc./1 kWh
09 150 cyc./1 kWh
10 125 cyc./1 kWh
0.10 kWh 0.10 kWh
0.05 kWh 0.05 kWh
0.01 kWh 0.01 kWh
Changing the
setting value
Enable:
Disable: –
47
www. .com
information@itm.com1.800.561.8187

Manual Hold / Auto Hold

Manual Hold / Auto Hold
MANUAL HOLD AUTO HOLD
To measurement object
>
Measured value is retained.
(
Pressing the HOLD key again cancels
the measured value hold function.
(
and disappears)
appears)
48
ashes.
Press for 1 sec.
To measurement object
(Measured value
stabilizes.)
Measured value is
automatically retains.
(
Pressing the HOLD key for 1 second cancels the auto hold
function. (
See the next page for auto hold conditions.
appears)
and disappears)
Disconnect
www. .com
information@itm.com1.800.561.8187
Manual Hold / Auto Hold
Auto hold conditions
Measured value is automatically retained when the following two conditions are satised:
When the range over which the measured value is uctuating stabilizes within the uctuation range described in the table in the next page.
• When the measured value exceeds the threshold value described in the table in the next page.
Ο: Held value
Auto hold e.g.: 99.0 A
Fluctuation range
Threshold value e.g.: 6.0 A
Measured value
Auto hold e.g.: 100.0 A
Fluctuation range
Fluctuation range
Start
If the measured value* (voltage, current, or active power) falls below the threshold value once and the two conditions are satised again after automatic retaining, the measured value at that point will retain automatically. * Either the current RMS value or voltage RMS value for power.
End
End
Start
Measurement
Start
End
Time
49
www. .com
information@itm.com1.800.561.8187
Manual Hold / Auto Hold
Measurement function*
AC current Current RMS value
AC voltage Voltage RMS value
Single-phase power, balanced 3-phase power
* No auto-hold function is available for single-phase active energy measurement.
Fluctuation range Threshold value
6.000 A range: within 60 counts
60.00 A range: within 60 counts
600.0 A range: within 60 counts
within 120 counts
Current and voltage RMS values satisfy above conditions, and active power is within 5 counts.
Current RMS value
6.000 A range: 59 counts
60.00 A range: 59 counts
600.0 A range: 59 counts
Voltage RMS value 799 counts
Current and voltage RMS values are within the above counts.
50
www. .com
information@itm.com1.800.561.8187

Switching Ranges

e.g.: During current measurement
AUTO range MANUAL 60.00 A range
MANUAL 600.0 A range
MANUAL 6.000 A range
Switching Ranges
51
www. .com
information@itm.com1.800.561.8187

MAX/ MIN/ AVG

MAX/ MIN/ AVG
1
2
• Switches to manual range when it is auto range. (
• The MAX/MIN/AVG measurement will be continued during hold function.
• The maximum, minimum, and average function cannot be used during Wh function and phase detect function operation.
• The maximum, minimum, and average values are automatically stored just before the instrument turns off due to low battery voltage. Next time the instrument is turned on, the saved values will be displayed. (The measured values can be cleared by pressing the HOLD key.)
To measurement object
Present value
: To switch the main display.
(long press) or (switching functions): MAX/MIN/AVG measurement function is cleared.
appears)
52
www. .com
information@itm.com1.800.561.8187
e.g.:
During current measurement
MAX/ MIN/ AVG
Maximum measured value of the current RMS*
Maximum measured value's time*
*1 The maximum, minimum, and average values for the main display’s measured value is shown.
(However, only the maximum and average values are shown during peak value measurement.
Also, only the maximum and minimum values are shown during zero-cross phase angle measurement.)
*2 Measured value's update time is displayed when maximum or minimum value is shown. Elapsed time
from the start of maximum, minimum, and average function is displayed when present or average value is shown.
2
1
53
www. .com
information@itm.com1.800.561.8187
MAX/ MIN/ AVG
|PEAK|
MAX
Measured value
(RMS value)
MAX/MIN/AVG of the RMS value
|PEAK|
AVG: Average value after pressing the MAX/MIN key MAX: Maximum value after pressing the MAX/MIN key MIN: Minimum value after pressing the MAX/MIN key |PEAK|: Maximum value of the absolute value of the waveform during the display update interval
MIN
Display refresh interval
54
Waveform
www. .com
information@itm.com1.800.561.8187

Backlight / Auto Power Save (APS)

30
sec
10
sec
10
sec
Backlight / Auto Power Save (APS)
Backlight Auto power save (APS)
Backlight OFF
min
Backlight ON
No operation for approx. 40 sec.: Backlight off
Switching the auto backlight off function: p. 18
15
No input and operation for approx. 15 min.
• Set the rotary switch to OFF when restarting the instrument after the APS function.
• The APS function is disabled while displaying the MAX/MIN/AVG value and during energy integration.
®
• Bluetooth
communications are treated as an operation for the purpose of the
APS function.
Flashes
Power OFF
(Ordinarily on)
Switching the function: p. 18
sec
30
Power OFF 30 sec. before
Power OFF 10 sec. before
sec
10
55
www. .com
information@itm.com1.800.561.8187

Measurement Using the Clamp Adapter

Measurement Using the Clamp Adapter
A clamp adapter (sold separately) can be used to measure currents that are larger than the rated input current.
Select the CT ratio.
1
Clamp
3
Clamp adapter (e.g.: Model 9290-10)
The instrument
56
Conductor
Clamp
2
Rotary switch CT ratio
A 1/1 (not shown)
W 1/10
Wh 1/100
3PW 1/1000
Set the CT ratio as appropriate for the clamp adapter. (e.g.: For the 9290-10 Clamp on Adapter, 1/10)
Perform measurement.
4
www. .com
information@itm.com1.800.561.8187

Bluetooth® Communications (only for CM3286-01)

Bluetooth® Communications
The CM3286-01 is a clamp-style meter with Bluetooth® low energy support. When the Bluetooth® function is enabled, you can review measurement data and create measurement reports on mobile devices (iPhone, iPad, iPad Mini, iPad Pro, iPod Touch, and Android™ devices). For more information about this functionality, see the Help function in the application software GENNECT Cross.
Install the GENNECT Cross on your mobile device. (p. 58)
1
Enable the Bluetooth® function on the CM3286‑01. (p. 59)
2
Launch the GENNECT Cross and pair it with the CM3286‑01.
3
(p. 60)
Select the General Measurement, Logging (Recording),
4
Waveform Graph, Electricity Theft Detection, or Harmonic
Analysis function. (p. 61)
(only for CM3286‑01)
Press for 1
sec.
57
www. .com
information@itm.com1.800.561.8187
Bluetooth® Communications (only for CM3286-01)

Installing the application software GENNECT Cross

Search for “GENNECT Cross” on the App Store from your iPhone, iPad or other Apple device, or on Google Play from your Android device. Then download and install the GENNECT Cross. You will need an Apple ID to download the app from the App Store, or a Google account to download the app from Google Play. For more information about how to register an account, contact the store at which you purchased your device.
• Because the CM3286-01 emit radio waves, use in a country or region where they have not been
approved may be subject to nes or other penalties as a violation of applicable laws or regulations. For
more information, see the attached “Precautions Concerning Use of Equipment That Emits Radio Waves” or go to our website.
The CM3286-01 availability is limited to certain countries. For more information, contact your authorized Hioki distributor or reseller.
®
• Bluetooth
etc.) as well as distance from the oor or ground. To ensure stable measurement, verify adequate signal
strength.
• Although this application software is provided free of charge, downloading or use of the application software may incur Internet connection charges. Such charges are the sole responsibility of the user.
• This application software is not guaranteed to operate on all mobile devices.
communications range varies greatly with distance from obstructions (walls, metal obstruction,
58
www. .com
information@itm.com1.800.561.8187
Bluetooth® Communications (only for CM3286-01)

Turning on the Bluetooth® function

Bluetooth® function OFF Bluetooth® function ON
Press for 1 sec.
icon will ash when the instrument is
connected to a mobile device.
59
www. .com
information@itm.com1.800.561.8187
Bluetooth® Communications (only for CM3286-01)
Pairing the app with the CM3286‑01
2
1
When the app is launched for the rst time (before being paired with any instrument), the Instrument
Settings screen will be displayed.
• While the mobile device is displaying the Instrument Settings screen, simply move it close to a CM3286-
01 to automatically pair it with the instrument (the app can be paired with up to 8 instruments).
• Allow about 5 to 30 seconds for the instrument to pair with the app after being turned on. If the instrument fails to pair within 1 minute, relaunch GENNECT Cross and cycle the instrument’s power.
3
60
www. .com
information@itm.com1.800.561.8187
Bluetooth® Communications (only for CM3286-01)

Making measurements with the Bluetooth® function

Select the General Measurement, Logging (Recording), Waveform Graph, Electricity Theft Detection, or Harmonic Analysis function on the Home screen. For more information about each function, see the Help function in the GENNECT Cross.
General Measurement
Saves measured values from
multiple channels.
Logging (Recording)
Simple logging
(up to 24 hours)
Waveform Graph
Simple oscilloscope
(voltage/current)
61
www. .com
information@itm.com1.800.561.8187
Bluetooth® Communications (only for CM3286-01)
Electricity Theft Detection
Creates a result report by measuring current and
energy.
62
Harmonic Analysis
Analyzes levels, content percentages, and total
distortion (voltage, current).
www. .com
information@itm.com1.800.561.8187

Repairs, Inspections, and Cleaning

Repairs, Inspections, and Cleaning
WARNING
Customers are not allowed to modify, disassemble, or repair the instrument.
Doing so may cause a re, an electric shock, or a injury..
Cleaning
• To clean the instrument, wipe it gently with a soft cloth moistened with water or mild detergent.
• Measurements are degraded by dirt on the mating surfaces of the jaw, so keep the surfaces
clean by gently wiping with a soft, dry cloth.
• Wipe the LCD gently with a soft, dry cloth.
Disposal
Handle and dispose of the instrument and batteries in accordance with local regulations.
Precautions during shipment
Be sure to observe the following precautions:
• To avoid damage to the instrument, remove the batteries, accessories, and options from
the instrument. Moreover, be sure to pack in a double carton. Damage that occurs during transportation is not covered by the warranty.
• When sending the instrument for repair, be sure to include details of the problem.
63
www. .com
information@itm.com1.800.561.8187
Repairs, Inspections, and Cleaning
Calibrations
The calibration period varies depending on the status of the instrument or installation environment. We recommend that the calibration period be determined in accordance with the status of the instrument or installation environment. Please contact your Hioki distributor to have your instrument periodically calibrated.
64
www. .com
information@itm.com1.800.561.8187

Troubleshooting

Troubleshooting
If damage is suspected, check the following before contacting your authorized Hioki distributor
or reseller.
Symptom Verication and/or Solution
• The instrument is indicating an
abnormal measured value for current.
• When readings from the
instrument are compared with those of another clamp-on current meter, the measured values differ.
• Is the measured current value too small for the instrument’s measurement range? Wrap the wire around the jaw one or more times. Each additional wrap of the wire will increase the measured value, so that wrapping it once yields a measured value that is twice the actual value and wrapping it twice yields a measured value that is three times the actual value.
• Are the tips of the jaw open?
• Is the jaw damaged? If the sensor is damaged or cracked, it will not be able to measure current accurately. Send the instrument for repair.
• The instrument cannot accurately measure waveforms that contain a component that falls outside the frequency characteristics range.
• Since the instrument performs true RMS measurement, it can accurately measure distorted waveforms. When measuring a distorted waveform, the measured value will differ from a clamp-on current meter that uses the averaging method.
65
www. .com
information@itm.com1.800.561.8187
Troubleshooting
Symptom Verication and/or Solution
• The current value is larger than expected.
• A current value is displayed even though there is no input.
• A sound is being emitted by the instrument’s jaw.
• The measured value does not appear.
• No measured value is displayed, even when the connection cords are shorted.
66
• The instrument cannot perform measurement accurately in the
presence of a strong magnetic eld from a source such as a nearby
transformer or high-current circuit or in the presence of a strong
electric eld from a source such as a wireless device.
• The jaw may emit sound when measuring AC currents in excess of approx. 500 A, however, there is no effect on the measurement.
• Insert the connection cords all the way.
• Use the proper measurement method.
If no measured value is displayed after attempting the above two solutions, the instrument may be broken. Have the instrument repaired.
www. .com
information@itm.com1.800.561.8187
Error display
Error
display
Err 001 ROM error Program
Err 002 ROM error Adjustment data
Err 005 ADC error Hardware malfunction
Err 008
Bluetooth error
Description Solution
®
Hardware malfunction (only for CM3286-01)
Warning display
Display Buzzer Cause Solution
Troubleshooting
Repair is necessary. Please contact your
authorized Hioki distributor or reseller.
Flashes
red
Measurement resulted in a negative active power value.
The instrument may not be connected properly. Reconnect the instrument to the circuit being measured.
67
www. .com
information@itm.com1.800.561.8187
Troubleshooting
Display Buzzer Cause Solution
e.g.: for current
measurement
Flashes
red
Intermittent
sound
A current or voltage exceeding the maximum input was input to the instrument.
Stop measurement immediately as the current or voltage cannot be measured by the instrument. For current measurement, the optional 9290-10 can be used to measure currents of up to 1000 A AC. When manual range is 6 A and 60 A range, this warning display will not appear.
e.g.: for current
measurement
68
Lights
red
A current or voltage
exceeding the range was input while using a manual range.
Change the measurement range or select the AUTO range.
www. .com
information@itm.com1.800.561.8187
Troubleshooting
Display Buzzer Cause Solution
Lights
red
Intermittent
sound
Phase detection indicated reverse phase.
69
www. .com
information@itm.com1.800.561.8187
Specications
General Specications
Operating environment Indoors, pollution degree 2, altitude up to 2000 m (6562 ft.)
Operating temperature and humidity
Storage temperature and humidity
Dustproof and waterproof
Temperature −25°C (−13°F) to 65°C (149°F)
Humidity −25°C (−13°F) or higher but less than 40°C (104°F):
80% RH or less 40°C (104°F) or higher but less than 45°C (113°F): 60% RH or less 45°C (113°F) to 65°C (149°F): 50% RH or less (no condensation)
Temperature −25°C (−13°F) to 65°C (149°F)
Humidity −25°C (−13°F) or higher but less than 40°C (104°F):
80% RH or less 40°C (104°F) or higher but less than 45°C (113°F): 60% RH or less 45°C (113°F) to 65°C (149°F): 50% RH or less (no condensation) Remove batteries before storing the instrument.
Grip except lever: IP54 (EN 60529) Jaw, barrier, lever: IP50 (EN 60529)
Specications
71
www. .com
information@itm.com1.800.561.8187
Specications
Standards (other than wireless communications functionality)
Power supply LR03 alkaline battery ×2
Continuous operating time
Interface (only for CM3286-01)
Dimensions Approx. 82W × 241H × 37D mm (3.23″W × 9.49″H × 1.46″D)
Jaw dimensions Approx. 79W × 20D mm (3.11″W × 0.79″D)
Maximum measurable conductor diameter
Mass Approx. 450 g (15.9 oz.) (including batteries)
Product warranty period 3 years
Accessories See p. 2
Options See p. 4
Safety: EN 61010 EMC: EN 61326
Rated supply voltage: 1.5 V DC ×2 Maximum rated power: 550 mVA
Approx. 25 hours (Backlight display off, Bluetooth as a referential) Approx. 18 hours (Backlight display off, Bluetooth as a referential)
®
Bluetooth
46 mm (1.81″)
4.0LE ( )
®
communication off, at 23°C,
®
communication on, at 23°C,
72
www. .com
information@itm.com1.800.561.8187
Input/Output/Measurement Specications
Basic Specications
Specications
Measurement items
Maximum input current See the frequency derating characteristics (p. 74).
Maximum measuring voltage
Maximum rated voltage to earth
Maximum rated voltage to terminal
Measurement method True RMS measurement with digital sampling
Measurement terminal COM terminal and V terminal
Input impedance 1 M
AC current RMS value/AC current peak value (no polarity)/AC current frequency AC voltage RMS value/AC voltage peak value (no polarity)/AC voltage frequency Single-phase active power/Single-phase apparent power/Single-phase reactive power/Single-phase power factor/Single-phase zero-cross phase angle Balanced 3-phase active power/Balanced 3-phase reactive power/Balanced 3-phase apparent power/Balanced 3-phase power factor/Balanced 3-phase zero-cross phase angle Single-phase active energy (only positive values added)/Phase detection
(up to 200 Hz, 600 A or less; above 200 Hz, 120000 A・Hz or less)
600 V AC
600 V AC (Measurement category IV) 1000 V AC (Measurement category III) Anticipated transient overvoltage 8000 V
or greater
Ω
73
www. .com
information@itm.com1.800.561.8187
Specications
Display update interval 2 times/sec.
Response time 1 sec.
Crest factor 3 or less for current 6 A and 60 A range
Zero-display range Voltage and current RMS values: 29 counts or less
Frequency derating characteristics
1.6 or less for current 600 A range and voltage 600 V range
If they fall within the zero-display range, current (voltage) peak values and active/apparent/reactive power values are shown as zero, while current (voltage) frequency, power factor, and zero-cross phase values are shown as “– – – –.”
A value of 0 is used in single-phase active energy calculations.
700
600
500
400
300
200
100
Maximum input current [A]
0
10 100 1000 10000
Frequency [Hz]
74
www. .com
information@itm.com1.800.561.8187
Accuracy Specications
Specications
Conditions of guaranteed accuracy
Input conditions for guaranteed accuracy
Effects of external magnetic elds
Effects of conductor position
Temperature coefcient Add “measurement accuracy × 0.1/°C” (excluding 23°C±5°C (73°F±9°F)).
Effects of sensor phase ±1° (50 Hz to 60 Hz)
See “Accuracy Table” (p. 79)
Guaranteed accuracy period: 1 year Guaranteed accuracy period after adjustment made by Hioki: 1 year Guaranteed accuracy for temperature and humidity: 23°C±5°C (73°F±9°F), 80% RH or less (no condensation) Number of jaw open/close cycles: 10000 times or less
Sine wave input
DC/AC 60 Hz, with a 400 A/m external magnetic eld:
0.10 A or less
At all positions around the jaw’s center-point reference: within ±0.5% (100 A input, f ≤100 Hz)
75
www. .com
information@itm.com1.800.561.8187
Specications
Harmonic Measurement Specications (only for CM3286-01)
All operations are performed by the GENNECT Cross application software. The following specications apply only to use of the GENNECT Cross’s harmonic analysis functionality. Perform sampling of data by instrument and harmonic analysis calculations by GENNECT Cross.
Measurement conditions
Measurement functionality
Analysis window width 1 cycle (50 Hz/60 Hz)
Window type Rectangular
Number of data points analyzed
Orders analyzed 1st to 30th
Items analyzed Harmonic level
Fundamental frequency 50 Hz/60 Hz
AC current/AC voltage (controlled by application software)
256
(RMS values for current harmonics [A], RMS values for voltage harmonics [V])
Harmonic content percentage (content percentages for current harmonics [%], content percentages for voltage harmonics [%])
Total harmonic distortion (THD-F and THD-R for current [%], THD-F and THD-R for voltage [%])
76
www. .com
information@itm.com1.800.561.8187
Range (maximum resolution)
Accuracy input range Input of 1% of range or greater for each order
Crest factor 3 or less for the 6 A and 60 A current ranges
Data refresh 5 s (reference value)
Measurement accuracy Harmonic level
AC current 600.0 A
AC voltage 600.0 V (0.1 V)
1.6 or less for the 600 A current range and the 600 V voltage range
(RMS value)
Harmonic content percentage
Total harmonic distortion
(0.1 A)
Order Accuracy
1 to 10 ±5.0% rdg. ±10 dgt.
11 to 20 ±10% rdg. ±10 dgt.
21 to 30 ±20% rdg. ±10 dgt.
±1 dgt. for calculations performed using measured values
±1 dgt. for calculations performed using measured values
60.00 A (0.01 A)
Specications
6.000 A (0.001 A)
77
www. .com
information@itm.com1.800.561.8187
Specications
External Interface (Bluetooth®) Specications
Display function Display of measured values on a iOS device or a Android device, using
Interface Bluetooth
Antenna power Maximum +0 dBm (1 mW)
Communications range 10 m (line of sight)
Communications prole
Supported devices Android 4.3 or later, iOS 10 or later (only for Bluetooth
®
Bluetooth
GATT (Generic Attribute Prole)
communications.
®
4.0 LE
®
low energy models)
Application Software Specications
Electricity theft detection function
Harmonic analysis function
Others
Creates an result report by measuring current and energy in conjunction with the instrument.
The app analyzes current or voltage harmonics up to the 30th order and displays the results in graph form.
See the GENNECT Cross specications (data saving, logging (recording), waveform graph function)
78
www. .com
information@itm.com1.800.561.8187

Accuracy Table

(1) AC Current Measurement
The current RMS (
Auto range threshold: Range up: Current RMS value greater than 6000 count
AC current RMS
I
)
(
RMS
I
) and current peak value (
RMS
Range down: Current RMS value less than 540 count
Range
(Accuracy guarantee
range)
6.000 A (0.060 A to 6.000 A)
60.00 A (0.60 A to 60.00 A)
600.0 A (6.0 A to 600.0 A)
I
) ranges will change at the same time.
|PEAK|
Resolution Accuracy
Display range
0.001 A
0.000 A to 6.000 A
0.01 A
0.00 A to 60.00 A
0.1 A
0.0 A to 600.0 A
45 Hz ≤ f ≤
66 Hz
±1.3% rdg. ±3 dgt.
±1.0% rdg. ±3 dgt.
66 Hz < f ≤
±2.0% rdg. ±5 dgt.
±1.5% rdg. ±5 dgt.
500 Hz
Accuracy Table
500 Hz < f ≤
1 kHz
±5.0% rdg. ±5 dgt.
±3.0% rdg. ±5 dgt.
79
www. .com
information@itm.com1.800.561.8187
Accuracy Table
AC current peak
I
I
|PEAK|
)
)
value (
Zero to Peak
No polarity
(absolute value of the maximum wave height during the display update interval)
AC current frequency
FREQ
(
80
Range
(Accuracy guarantee
range is specied in
terms of current RMS
values.)
6.000 A (0.060 A to 6.000 A)
60.00 A (0.60 A to 60.00 A)
600.0 A (6.0 A to 600.0 A)
Range
(Accuracy guarantee
range)
999.9 Hz (45.0 Hz to 999.9 Hz)
Resolution Accuracy
Display range
0.01 A
0.00 A to 18.00 A
0.1 A
0.0 A to 180.0 A
1 A
0 A to 1000 A
Resolution Accuracy
Display range
0.1 Hz
45.0 Hz to 999.9 Hz
45 Hz ≤ f ≤
66 Hz
±3.0% rdg. ±5 dgt.
±2.5% rdg. ±5 dgt.
Current frequency values are shown as “– – – –” when the current RMS value is less than 150 count. Current frequency values of less than
45.0 Hz are shown as “– – – –.”
±0.3% rdg. ±3 dgt.
66 Hz < f ≤
500 Hz
500 Hz < f ≤
1 kHz
±5.0% rdg. ±5 dgt.
±4.0% rdg. ±5 dgt.
www. .com
information@itm.com1.800.561.8187
(2) AC Voltage Measurement
AC voltage RMS
U
RMS
)
value (
Range
(Accuracy guarantee
range)
Resolution Accuracy
Display range
45 Hz ≤ f ≤
66 Hz
66 Hz < f ≤
500 Hz
Accuracy Table
500 Hz < f ≤
1 kHz
AC voltage peak
U
|PEAK|
)
value (
Zero to Peak
No polarity
(absolute value of the maximum wave height during the display update interval)
600 V (80.0 V to 600.0 V)
Range
(Accuracy guarantee
range is specied in
terms of current RMS
values.)
600 V (80.0 V to 600.0 V)
0.1 V
0.0 V to 600.0 V
Resolution Accuracy
Display range
1 V
0 V to 1000 V
±0.7% rdg. ±3 dgt.
45 Hz ≤ f ≤
66 Hz
±2.5% rdg. ±5 dgt.
±1.0% rdg. ±5 dgt.
66 Hz < f ≤
500 Hz
±3.0% rdg. ±5 dgt.
500 Hz < f ≤
1 kHz
±4.0% rdg. ±5 dgt.
81
www. .com
information@itm.com1.800.561.8187
Accuracy Table
AC voltage frequency
FREQ
)
(
U
Range
(Accuracy guarantee
range)
999.9 Hz (45.0 Hz to 999.9 Hz)
Resolution
Maximum
display
0.1 Hz
999.9 Hz
Voltage frequency values are shown as “– – – –” when the voltage RMS value is less than 150 count. Voltage frequency values of less than
45.0 Hz are shown as “– – – –.”
±0.3% rdg. ±3 dgt.
Accuracy
(3) Single-phase power measurement, balanced 3-phase/4-wire power measurement
Auto range threshold: Range up: Current RMS value greater than 6000 count
Range down: Current RMS value less than 540 count
Effective measuring range
82
Current RMS value
I
(
)
RMS
Voltage RMS value
U
(
)
RMS
Frequency 50 Hz/60 Hz
0.060 A to 600.0 A
Value must fall within the current measurement range’s guaranteed accuracy range.
80.0 V to 600.0 V
www. .com
information@itm.com1.800.561.8187
Single-phase active power/
Balanced 3-phase/4-wire active power
P
)
(P /
(3P4W)
Single-phase apparent power/ Balanced 3-phase/4-wire apparent power
S
)
(S /
(3P4W)
Single-phase reactive power/ Balanced 3-phase/4-wire reactive power
Q
)
(Q /
(3P4W)
Accuracy Table
Range conguration (minimum resolution) Current range
6.000 A 60.00 A 600.0 A
Voltage range 600.0 V Single-
phase
3-Phase 4-Wire
Accuracy (Power factor =1) Single-
phase
3-Phase 4-Wire
Accuracy ±1 dgt. relative to calculation from measured values
Range conguration For the above active power range conguration, the unit [W]
is replaced by [VA] for apparent power values. For reactive power values, the unit [W] is replaced by [var].
3.600 kW (0.001 kW)
10.80 kW (0.01 kW)
±2.0% rdg. ±7 dgt.
±2.0% rdg. ±3 dgt.
36.00 kW (0.01 kW)
108.0 kW (0.1 kW)
±1.7% rdg. ±5 dgt.
±1.7% rdg. ±2 dgt.
360.0 kW (0.1 kW)
1080 kW (1 kW)
83
www. .com
information@itm.com1.800.561.8187
Accuracy Table
Single-phase power factor/ Balanced 3-phase/4-wire power factor
PF / PF
(
(3P4W)
Zero-cross phase angle (φ)*
* Value is calculated based on the measurement of the zero-cross phase difference for the voltage and
current waveforms (positive [no sign] when the current lags the voltage and negative when the current leads the voltage).
Accuracy ±1 dgt. relative to calculation from measured values
Range conguration Regeneration −1.000 to −0.001
Consumption 0.000 to 1.000
)
Accuracy ±3°
Range conguration Lead −180.0° to −0.1°
Lag 0.0° to 179.9°
(4) Balanced 3-phase/3-wire power measurement
Auto range threshold: Range up: Current RMS value greater than 6000 count
Range down: Current RMS value less than 540 count
Effective measuring range
84
Current RMS
I
value (
Voltage RMS
U
value (
Frequency 50 Hz/60 Hz
0.060 A to 600.0 A
Value must fall within the current measurement range’s guaranteed accuracy range.
)
RMS
80.0 V to 600.0 V
)
RMS
www. .com
information@itm.com1.800.561.8187
Balanced 3-phase/3­wire active
P
power (
(3P3W)
Balanced 3-phase/3­wire apparent
S
power (
(3P3W)
Balanced 3-phase/3­wire reactive
Q
power (
(3P3W)
Balanced 3-phase/3­wire power factor
PF
)
(
(3P3W)
Balanced 3-phase/3­wire zero­cross phase angle (
φ
(3P3W)
Accuracy Table
Accuracy ±3.0% rdg. ±10 dgt. (Power factor =1)
Range conguration
)
(minimum resolution)
)
Accuracy ±1 dgt. relative to calculation from measured values
Range conguration
)
Accuracy ±3° ±2 dgt. (Calculated from the balanced 3-phase/3-wire zero-cross
Range conguration
Accuracy ±3°
Range conguration
)*
Voltage range 600.0 V
The unit [W] is replaced by [VA] for apparent power values.
For the above active power range conguration, the unit [W] is replaced by [var].
phase angle)
Regeneration −0.001
Consumption 0.000 to 1.000
Lead −90.0° to −0.1°
Lag 0.0° to 90.0°
6.000 A 60.00 A 600.0 A
7.200 kW (0.001 kW)
Current range
72.00 kW (0.01 kW)
720.0 kW (0.1 kW)
85
www. .com
information@itm.com1.800.561.8187
Accuracy Table
* Value is calculated based on the measurement of the zero-cross phase difference for the voltage and
current waveforms (positive [no sign] when the current lags the voltage and negative when the current leads the voltage).
(5) Single-phase active energy measurement (AC)
Effective measuring range
86
I
Current RMS value (
Voltage RMS value (
Frequency 50 Hz/60 Hz
) 0.060 A to 600.0 A
RMS
U
) 80.0 V to 600.0 V
RMS
Value must fall within the current measurement range’s guaranteed accuracy range.
www. .com
information@itm.com1.800.561.8187
Single-phase active energy (Wh)
Accuracy Table
Measurement method Positive active power values are added every 0.5 s.*
Range conguration
99.99 Wh 0.00 Wh to
999.9 Wh 100.0 Wh to
9.999 kWh 1.000 kWh to
99.99 kWh 10.00 kWh to
999.9 kWh 100.0 kWh to
9999 kWh 1000 kWh to
Display range • After the single-phase active power range is
99.99 Wh
999.9 Wh
9.999 kWh
99.99 kWh
999.9 kWh
9999 kWh
A value of zero is added when the active power is negative.
* When stopping integration, the energy measured
during the last 0.5 sec. is divided into 5 and added every 0.1 sec.
selected, integration starts with a value of 0.00 Wh. Only auto-range operation is supported for active energy measurement. When values exceed 9999 count, the range is switched to the next higher range.
The range is xed to the active power range in use when integrated began.
87
www. .com
information@itm.com1.800.561.8187
Accuracy Table
Integration time display
(6) Phase detection
Detected voltage range 80 V AC to 600 V AC
Detection target frequency 50 Hz/60 Hz (sine wave)
Phase order detection* Normal phase (Display: 123)
* After the second measurement display appears, and the second measurement value does not become
stable over 10 seconds, it is unable to make a measurement.
59:59 [min:sec] The time is incremented by 1 s from 00:00 [min:sec].
48:00 [hour:min]
Reverse phase (Display: 321) Open phase or unable to measure (Display: – – – –)
When 59:59 [min:sec] is exceeded, the range is switched to the 48:00 [hour:min] range. During integration using the 48:00 [hour:min] range,
“ : ”
the
display ashes every 0.5 s.
88
www. .com
information@itm.com1.800.561.8187
Range conguration when setting a CT ratio
CT ratio
Current RMS value
Current peak value
Single-phase active power
1/1
(default value)
600.0 A 6000 A
60.00 A 600.0 A 6000 A
6.000 A 60.00 A 600.0 A 6000 A
1000 A 10.00 kA
180.0 A 1800 A 18.00 kA
18.00 A 180.0 A 1800 A 18.00 kA
360.0 kW 3600 kW
36.00 kW 360.0 kW 3600 kW
3.600 kW 36.00 kW 360.0 kW 3600 kW
1/10 1/100 1/1000 Remarks
Accuracy Table
CT ratio 1/1 Same accuracy specications as 600.0 A. CT ratio 1/1 Same accuracy specications as 60.00 A. CT ratio 1/1 Same accuracy specications as 6.000 A. CT ratio 1/1 Same accuracy specications as 600.0 A. CT ratio 1/1 Same accuracy specications as 60.00 A. CT ratio 1/1 Same accuracy specications as 6.000 A. CT ratio 1/1 Same accuracy specications as 360.0 kW. CT ratio 1/1 Same accuracy specications as 36.00 kW. CT ratio 1/1 Same accuracy specications as 3.600 kW.
89
www. .com
information@itm.com1.800.561.8187
Accuracy Table
CT ratio
Balanced 3-phase/3-wire active power
Balanced 3-phase/4-wire active power
Add the accuracy of the appropriate CT.
The unit is replaced as below for apparent power and reactive power, relative to the active power range.
Apparent power: kVA Reactive power: kVA
*1 Multiply the dgt. error indicated in the accuracy specications noted in the “Remarks” column by 10.
1/1
(default value)
720.0 kW 7200 kW
72.00 kW 720.0 kW 7200 kW
7.200 kW 72.00 kW 720.0 kW 7200 kW
1080 kW 9999 kW*
108.0 kW 1080 kW 9999 kW*
10.80 kW 108.0 kW 1080 kW 9999 kW*
R
1/10 1/100 1/1000 Remarks
1
1
CT ratio 1/1 Same accuracy specications as 720.0 kW. CT ratio 1/1 Same accuracy specications as 72.00 kW. CT ratio 1/1 Same accuracy specications as 7.200 kW. CT ratio 1/1 Same accuracy specications as 1080 kW. CT ratio 1/1 Same accuracy
specications as 108.0 kW. CT ratio 1/1 Same accuracy
1
specications as 10.80 kW.
90
www. .com
information@itm.com1.800.561.8187

Equations

22
PS
S
P
{ }
(1) Single-phase power measurement
Equations
Apparent power
Reactive power
Power factor
S U
Q
PF
RMS
I
RMS
(2) Balanced 3-phase/3-wire power measurement
Balanced 3-phase/3-wire zero-cross phase angle
Balanced 3-phase/3-wire power factor
Balanced 3-phase/3-wire active power
Balanced 3-phase/3-wire apparent power
Balanced 3-phase/3-wire reactive power
φ
PF
P
S
(3P3W)
Q
(3P3W)
(3P3W)
(3P3W)
(3P3W)
[W]
[VA]
[var]
cos
3
PF
φ
−30°
φ
The active power P has no sign during
consumption and a negative sign during
generation.
Due to the effects of measurement error,
S=|P
| and Q=0 are used when S<|P|.
)33(
WP
SWP⋅
)33(
S3
2
PS
)33(2)33( WPWP
The symbol φ represents the zero-
cross phase angle of the voltage
U
and the current
12
The symbol S represents the
apparent power of the line voltage
U
and the wire current
12
I
.
1
I
.
1
91
www. .com
information@itm.com1.800.561.8187
Equations
P
(3) Balanced 3-phase/4-wire power measurement
Balanced 3-phase/4-wire active power
Balanced 3-phase/4-wire apparent power
Balanced 3-phase/4-wire reactive power
Balanced 3-phase/4-wire power factor
P
S
Q
PF
(3P4W)
(3P4W)
(3P4W)
(3P4W)
[W]
[VA]
[var]
P
3
S
3
Q
3
)43(WPWP
S
)43(
(4) Unbalanced 3-phase/3-wire power measurement
Unbalanced 3-phase/3-wire active power
P
(UB3P3W)
[W]
P
P
+
1
2
92
The symbol P represents the active power of
the phase voltage
The symbol S represents the apparent power of
the phase voltage
The symbol Q represents the reactive power of
the phase voltage
The active power P has no sign during
consumption and a negative sign during
generation.
Due to the effects of measurement error, S=|P|
and Q=0 are used when S<|P|.
The symbol
the line voltage
The symbol
the line voltage
The active power P has no sign during
consumption and a negative sign during
generation.
U
and the wire current
1
U
and the wire current
1
U
and the wire current
1
P
represents the active power of
1
U
and the wire current
21
P
represents the active power of
2
U
and the wire current
23
I
.
1
I
.
1
I
.
1
1
.
1
1
.
3
www. .com
information@itm.com1.800.561.8187
(5) Unbalanced 3-phase/4-wire power measurement
Equations
Unbalanced 3-phase/4-wire active power
Unbalanced 3-phase/4-wire apparent power
P
(UB3P4W)
S
(UB3P4W)
[W]
[VA]
The symbol
phase voltage
The symbol
+
P
3
phase voltage
The symbol
P
P
+
1
2
phase voltage
The active power P has no sign during consumption
P
represents the active power of the
1
U
and the wire current
1
P
represents the active power of the
2
U
and the wire current
2
P
represents the active power of the
3
U
and the wire current
3
I
.
1
I
.
2
I
.
3
and a negative sign during generation.
The symbol
phase voltage
The symbol
+
S
3
phase voltage
The symbol
S
S
+
1
2
phase voltage
Due to the effects of measurement error, S=|P| is
S
represents the apparent power of the
1
U
and the wire current
1
S
represents the apparent power of the
2
U
and the wire current
2
S
represents the apparent power of the
3
U
and the wire current
3
I
.
1
I
.
2
I
.
3
used when S<|P|.
93
www. .com
information@itm.com1.800.561.8187
Equations
)43(
WPUB
P
Unbalanced 3-phase/4-wire reactive power
Unbalanced 3-phase/4-wire power factor
94
Q
(UB3P4W)
PF
(UB3P4W)
[var]
The symbol
phase voltage
The symbol
Q
Q
+
1
2
phase voltage
Q
+
3
The symbol
phase voltage
Due to the effects of measurement error, Q=0 is
used when S<|P|.
)43(
WPUB
S
Q
represents the reactive power of the
1
U
and the wire current
1
Q
represents the reactive power of the
2
U
and the wire current
2
Q
represents the reactive power of the
3
U
and the wire current
3
I
.
1
I
.
2
I
.
3
www. .com
information@itm.com1.800.561.8187
(6) (Reference) Harmonic calculations
Calculated by GENNECT Cross
Equations
Harmonic current
RMS value
Harmonic content percentage for k th order
Total harmonic distortion
[A]
[%]
[%]
THD-F
THD-R
+
+
1
30
( )
=irk
2
1
30
( )
k
2
256
1
==n
II
kikr
II
+
+
+
( )
I
'
22
II +
kikr
22
[ ]
%100
×
212
ir
22
II
kikr
212
II
22
II
kikr
2
n
[ ]
100
×
%
[ ]
100
×
%
95
www. .com
information@itm.com1.800.561.8187
Equations
]
Harmonic voltage
Index
k
: Analyzed order
r
: Post-FFT resistance component
i
: Post-FFT reactance component
I’
: Current sampling value
U’
: Voltage sampling value
96
RMS value
Harmonic content percentage for k th order
Total harmonic distortion
[V]
[%]
[%]
THD-F
THD-R
+
+
30
( )
2
=irk
2
1
30
( )
2
k
256
( )
U
1
==n
22
UU +
kikr
22
UU
kikr
[ ]
%100
×
2
UU
121
ir
22
+
UU
kikr
2
+
UU
1
22
UU
+
kikr
2
'
n
[
100
×
%
[ ]
100
×
%
www. .com
information@itm.com1.800.561.8187
www. .com
information@itm.com1.800.561.8187
Loading...