CM3281
[FILTER] : Not used
[HOLD]
[AUTO]
CM3291
AC CLAMP METER
Instruction Manual
Dec. 2019 Edition 1
CM3281A961-00 19-12H
Warranty
Warranty malfunctions occurring under conditions of normal use in
conformity with the Instruction Manual and Product Precautionary
Markings will be repaired free of charge. This warranty is valid for a
period of three (3) years from the date of purchase. Please contact
the distributor from which you purchased the product for further
information on warranty provisions.
Introduction
Thank you for purchasing the Hioki CM3281, CM3291 AC Clamp
Meter. To obtain maximum performance from the instrument, please
read this manual rst, and keep it handy for future reference. Be
sure to also read the separate document “Operating Precautions”
before use.
CM3281 Average value measurement RMS conversion model
CM3291 True RMS measurement model
Target audience
This manual has been written for use by individuals who use the
product in question or who teach others to do so. It is assumed that
the reader possesses basic electrical knowledge (equivalent to that
of someone who graduated from the electrical program at a technical high school).
EN
Safety Notes
Symbols affixed to the device
Precaution or hazard (See corresponding topic.)
The instrument can be connected to or disconnected from a
live conductor
The exible sensor can be connected to or disconnected from
live conductors when appropriate protective insulation is used.
The clamp meter and test leads can only be connected to or
disconnected from insulated conductors suited to the voltage
of the conductor under measurement.
DANGER
To avoid electric shock, do not touch the portion
beyond the protective barrier during use.
Do not subject the instrument to any voltages when
the resistance measurement or continuity check
function is selected. Doing so may damage the instrument and result in bodily injury. To avoid electrical accidents, turn off the circuit before measuring it.
WARNING
• To avoid electric shock, short circuits and damage
to the instrument, disconnect the test leads from
the measurement object before switching the rotary switch.
• To prevent electric shock, when measuring the
voltage of a power line use a test lead that satises
the following criteria:
• Conforms to safety standards IEC61010 or EN61010
• Of measurement category III or IV
• Its rated voltage is higher than the voltage to be
measured
• The optional test leads for this instrument conform
to the safety standard EN61010. Use a test lead in
accordance with its dened measurement category
and rated voltage.
• To prevent a short circuit accident, be sure to use
the L9208 test leads with the sleeves attached
when performing measurements in the CAT III measurement category. (For the measurement categories, see the “Measurement categories” section in
the separate document “Operating Precautions”.)
• If the sleeves are removed during measurement,
stop the measurement.
• Handle and dispose of batteries in accordance with
local regulations.
CAUTION
Do not place foreign objects between the jaw tips (or
exible loop couplings) or insert foreign objects into the
gaps of the jaws (or exible loop couplings). Doing so
may worsen the performances of the sensor or interfere
with clamping action.
• The indicator is displayed when the remaining battery
capacity is low. In this case, the accuracy of the instrument is not guaranteed. Replace the battery immediately.
• To avoid battery depletion, set the rotary switch in the
[OFF] position after use. (Even when the automatic
power-saving function is enabled, the instrument consumes a small amount of the battery power.)
Inspection Before Measurement
• Before using the instrument, check it and verify that it operates
properly to make sure that it suffered no damage during storage or
transportation.
• If damage is suspected, check the section below before contacting
your authorized Hioki distributor or reseller.
(1) Check the test lead for breaks.
If any, replace it with the new L9208 Test Lead.
(2) Check that the resistance measurement and continu-
ity check operates normally.
If any one of them does not operate normally, send the instrument for repair to the your authorized Hioki distributor or
reseller. The instrument may have been subject to a voltage of
greater than 600 V during resistance measurement or continu-
ity check.
(3) Check that the battery weakens.
If it weakens, replace the battery.
Functions
Automatic power-saving function
The instrument automatically turns off the LCD after it is not
operated for 30 minutes.
• To enable the function
• To restore the instrument from a non-displaying state
Set the rotary switch in the [OFF] position and then set the rotary
1
switch in a position other than [OFF].
• To cancel automatic power-saving function
Set the rotary switch in a position other than [OFF] while holding
1
down the HOLD key.
The text [APS] and [OFF] are displayed in turn on the LCD, and
the automatic power-saving function is disabled.
Auto-range function
The instrument automatically selects the most appropriate
measurement range.
The text [AUTO] is displayed on the LCD.
Manual-range function
Sets the measurement range manually.
Set the rotary switch in the [OFF] position and then set the rotary
1
switch in a position other than [OFF] while holding down the
key.
Press the key to switch the measurement range.
2
(Any ranges can be set except for the continuity check.)
Overflow indication
If an input exceeds the measurement range, the text [OF] or [−OF]
is displayed on the LCD.
Parts Names
LCD (with all segments turned on)
: Holds measured value
: Auto-range function
: Low battery warning
: Connection of exible loop
: Continuity check
The instrument screen displays the alphanumeric characters as follows.
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
1 2 3 4 5 6 7 8 9 0
AC Clamp Meter
Front
Jaw
Barrier
Operating
grip
LCD
key
HOLD key
(Press to hold the
displayed value.)
Rotary switch
Measurement
terminal
Plug of test lead or
exible sensor
Rear
Battery
cover
Serial number
The 9-digit serial number indicates
the year (rst two digits) and the
month of manufacture (next two
digits). Do not remove this sticker
as the number is important.
Carrying case storage
L9208 Test Lead’s
sleeves (accessory)
L9208 Test Lead
(red, black)
• Instruction Manual
• Download Guide
• Operating Precautions
CT6280 AC Flexible Current Sensor (optional)
Flexible loop
Cap
Cap
(Approx. φ7 mm)
Instruction
plate
Using the attachment
allows the exible
loop to weave through
cables easily.
Attachment
Cap
Measuring Methods
AC Current Measurement [ ]
Measuring current with the instrument
1 2
OK
NO
Measuring current with Model CT6280 AC Flexible
Current Sensor (optional)
Use protective gear
1 2 3
4
5
Cannot measure
current
OK
1 wire
NO
NO
Do not pull the exible loop
with excessive force.
Voltage Measurement
AC Voltage Measurement [
To avoid damage, . . .
Do not bend the
exible loop.
Wire binding band
Do not pack the exible
loop with it bent.
]
1 2 3 4
DC Voltage Measurement [
1 2 3 4
]
OK
NO
Black
(−)
Red
(−)
Red
(+)
Black
(+)
3
Resistance Measurement [ ]
21 3 4
5
Continuity Check [ ]
21 3
4
5
Cleaning
• Measurements are degraded by dirt on the mating surfaces of the
jaw (or exible loop coupling), so keep the surfaces clean by gently
wiping with a soft, dry cloth.
• To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent.
• Wipe the LCD gently with a soft, dry cloth.
Replacing Battery
Necessary items: Phillips screwdriver (No.1) and Coin cell lithium
battery (CR2032)
NO
2
CR2032 (new)
Do not loosen any screws but
OK
Do not turn any one of the 3 screws inside the battery case. Doing
so will cause the instrument to report abnormal measured values.
CALIFORNIA, USA ONLY
This product contains a CR Coin Lithium Battery which contains
Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
Specifications
Function Specifications
Display Maximum count: 4199 counts
Battery indicator
warning voltage
General Specifications
Operating
environment
Operating temperature and humidity
Temperature −25°C to 65°C (−13°F to 149°F)
The mark
2.3 V±0.15 V or less.
Indoors, pollution degree 2, altitude up to 2000 m (6562 ft.)
Ω
(For the 40 M
range: up to 40°C [104°F])
the one securing the cover.
is displayed at a battery voltage of
431
Humidity
(no condensation)
Storage temperature and humidity
Drop-proof Operate after a drop from 1 m on concrete
Standards Safety: EN61010
Power supply Coin cell lithium battery CR2032 ×1
Continuous
operating time
Dimensions
Dimensions (Jaw)
Mass
Product warranty
period
Accessories • Carrying Case
Options The following options are available for the instrument. To
Less than 40°C (104°F): 80% RH or less
At least 40°C (104°F) but less than 45°C (113°F):
60% RH or less
At least 45°C (113°F) but less than 50°C (122°F):
50% RH or less
At least 50°C (122°F) but less than 55°C (131°F):
40% RH or less
At least 55°C (131°F) but less than 60°C (140°F):
30% RH or less
At least 60°C (140°F) but less than 65°C (149°F):
25% RH or less
−25°C to 65°C (−13°F to 149°F),
80% RH or less (no condensation)
EMC: EN61326
Rated power voltage: 3 V DC
Maximum rated power: 15 mVA
CM3281: Approx. 120 hours
CM3291: Approx. 70 hours
(AC current measurement mode, continuous, unloaded,
under conditions of temperature and humidity for guaranteed accuracy)
• CM3281, CM3291: Approx. 57W×198H×16D mm
(2.24″W × 7.80″H × 0.63″D)
• CT6280: Approx. 42W×65H×18D mm
(1.65″W × 2.56″H × 0.71″D)
(excluding the exible loop and output cable)
Approx. 65W×13D mm (2.56
• CM3281, CM3291: Approx. 103 g (3.6 oz.) (including battery)
• CT6280: Approx. 71 g (2.5 oz.)
CM3281, CM3291, CT6280: 3 years
• L9208 Test lead
• Coin cell lithium battery CR2032 (Installed in the instru-
ment, for LCD)
• Instruction Manual (This document)
• Download Guide
• Operating Precautions (0990A909)
purchase an option, please contact your authorized Hioki
distributor or reseller. Options are subject to change.
Please check Hioki’s website for the latest information.
• CT6280 AC Flexible Current Sensor
(Attachment and C0205 Carrying Case are included)
• L4933 Contact Pin Set
(Can be connected to the tip of the L9208, which comes
with the instrument.)
• L4934 Small Alligator Clip Set
(Can be connected to the tip of the L9208, which comes
with the instrument.)
• L9208 Test Lead
″W × 0.51″D
)
Basic Specifications
Maximum input
current
Maximum input
voltage
Overload
protection
Maximum rated voltage to earth
CM3281,
CM3291 (jaw)/
CT6280
Voltage
measurement
terminal
AC measurement method
Display update
rate
Noise rejection
characteristics
Crest factor
Zero-display
range
• CM3281, CM3291 (Jaw): 2000 A AC, continuous (45 Hz
to 66 Hz)
• CT6280 (Flexible loop): 4200 A AC, continuous (50 Hz
to 60 Hz)
6
600 V AC/DC and 3×10
600 V AC/DC (ACV, DCV,
600 V AC (Measurement category III),
300 V AC (Measurement category IV)
(Anticipated transient overvoltage: 6000 V)
600 V AC (Measurement category II),
300 V AC (Measurement category III)
(Anticipated transient overvoltage: 4000 V)
CM3281: Average value measurement RMS conversion model
CM3291: True RMS measurement model
400 ms±25 ms
NMRR DCV −40 dB or more (50 Hz/60 Hz)
CMRR DCV −100 dB or more
(50 Hz/60 Hz, 1 k
ACV −60 dB or more
(50 Hz/60 Hz, 1 kΩ unbalance)
CM3291: For 2500 counts or less, 2.5
Reduces linearly to 1.5 or less at 4200 counts
But, 1.5 or less for 2000 A ACA range
5 counts (AC current measured with jaw or exible loop)
V·Hz or less (ACV, DCV)
Ω
, continuity)
Ω
unbalance)
But, −45 dB or more for 600 V range.
Effects of conductor position
Maximum measurable conductor diameter
Model CT6280
• CM3281, CM3291
11‑mm‑diameter [22 mm2] cable)
• CT6280: within ±5.0% (At any positions, based on the
center of sensor)
• CM3281, CM3291: φ46 mm or less
• CT6280: φ130 mm or less
Cross-section diameter of sensor cable:
Sensor-tip cap diameter:
Output cable length:
: within ±5.0% (Specied with a
Approx.
Approx.
Approx.
φ7.0 mm
800 mm
φ5.0 mm
Accuracy Specifications
rdg (reading or displayed value):The value currently being measured and
indicated on the measuring instrument.
dgt (resolution):The smallest displayable unit on a digital measuring instru-
ment, i.e., the input value that causes the digital display to show a “1” as
the least‑signicant digit.
Conditions of guaranteed accuracy
• Guaranteed accuracy period: 1 year (Number of jaw and exible loop open/
close cycles: 10,000 or less)
• Guaranteed accuracy period after adjustment made by Hioki: 1 year
• Temperature and humidity for guaranteed accuracy: 23°C±5°C (73°F±9°F),
80% RH or less
• Supply voltage for guaranteed accuracy: The mark is not displayed.
• Temperature characteristic: Measurement accuracy × 0.1/°C is added
(excluding 23°C±5°C)
• AC waveform: sine wave
AC Current Measured With Jaw (CM3281)
Range Accuracy range
42.00 A 4.00 A to 41.99 A
2000 A 100 A to 1999 A
AC Current Measured With Jaw (CM3291)
Range Accuracy range
42.00 A 4.00 A to 41.99 A
420.0 A 40.0 A to 419.9 A
2000 A 100 A to 1999 A
Accuracy is not dened for currents of 3×10
AC Current Measured With Flexible loop (CM3281)
Range Accuracy range
420.0 A 40.0 A to 419.9 A
4200 A 400 A to 4199 A
AC Current Measured With Flexible loop (CM3291)
Range Accuracy range
420.0 A 40.0 A to 419.9 A ±3.5% rdg
4200 A 400 A to 4199 A
AC Voltage
Range Accuracy range
0.400 V
to
4.200 V
42.00 V 4.00 V to 41.99 V 10 M
420.0 V 40.0 V to 419.9 V 10 M
600 V 400 V to 600 V 10 M
DC Voltage
Range Accuracy range Accuracy Input impedance
420.0 mV 40.0 mV to 419.9 mV
4.200 V 0.400 V to 4.199 V
42.00 V 4.00 V to 41.99 V 10 M
420.0 V 40.0 V to 419.9 V 10 M
600 V 400 V to 600 V 10 M
Resistance
Range Accuracy range Accuracy
Ω
420.0
Ω
4.200 k
Ω
42.00 k
Ω
420.0 k
4.200 MΩ0.400 MΩ to
Ω
42.00 M
Continuity Check
Range Accuracy
Ω
420.0
*1:
Includes accuracy of CT6280 AC Flexible Current Sensor, ±1.0% rdg
*2: Accuracy is not dened for a current of 1000 A or more or that of
5×105 A · Hz or more.
4.199 V
40.0 Ω to 419.9
0.400 kΩ to 4.199 k
4.00 kΩ to 41.99 k
40.0 kΩ to 419.9 k
4.199 MΩ±5.0% rdg ±4 dgt
4.00 MΩ to 41.99 M
±2.0% rdg ±4 dgt
Accuracy
50 Hz ≤ f ≤ 60 Hz
±1.5% rdg ±5 dgt420.0 A 40.0 A to 419.9 A
Accuracy
40 Hz ≤ f <
45 Hz
±2.0% rdg
±5 dgt
5
Accuracy
50 Hz ≤ f ≤ 60 Hz
±3.0% rdg ±5 dgt *
Accuracy
40 Hz ≤ f <
45 Hz
1, *2
±5 dgt *
Accuracy
45 Hz ≤ f ≤
66 Hz
±1.8% rdg
±7 dgt
±2.5% rdg
±5 dgt
±1.0% rdg
±3 dgt
Ω
Ω
±2.0% rdg ±4 dgt
Ω
Ω
Ω
±10.0% rdg ±4 dgt
Threshold for
buzzer sound
Ω
50
45 Hz ≤ f ≤
66 Hz
±1.5% rdg
±5 dgt
A · Hz or more.
1
45 Hz ≤ f ≤
66 Hz
±3.0% rdg
±5 dgt *
66 Hz < f ≤
500 Hz
±2.3% rdg
±8 dgt
±40 Ω or less 3.4 V or less
1
100 M
11 M
66 Hz < f ≤
1 kHz
±2.0% rdg
±5 dgt
66 Hz < f ≤
1 kHz
±3.5% rdg
1, *2
±5 dgt *
Input
impedance
Ω
±5%
11 M
Ω
±5%
Ω
±5%
Ω
±5%
Ω
or more
Ω
±5%
Ω
±5%
Ω
±5%
Ω
±5%
Open-circuit
voltage
3.4 V or less
Open-circuit
voltage