Thank you for purchasing the HIOKI “Model BT3561A, BT3562A, BT3562, BT3562-01, BT3563A,
BT3563, and BT3563-01 BATTERY HiTESTER. ” To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference.
Please review the separate “Operating Precautions” before using the instrument.
This manual uses the following conventions:
The Model BT3561A, BT3562A, BT3562, BT3562-01, BT3563A, BT3563, and BT3563-01 are referred to as
"the instrument."
Unless otherwise noted, the Model BT3562, BT3562-01, and BT3562A are referred to collectively as "Model
BT3562," while the Model BT3563, BT3563-01, and BT3563A are referred to collectively as "Model BT3563."
Trademarks
Microsoft, Windows, Visual Studio, Visual Basic, Edge, and Visual C# are either registered trademarks or
trademarks of Microsoft Corporation in the United States and other countries.
2
HIOKI BT3562A981-09
Verifying Package Contents
Verifying Package Contents
When you receive the instrument, inspect it carefully to ensure that no damage occurred during
shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident,
or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Use the original packing materials when transporting the instrument, if possible.
Confirm that these contents are provided. (One each)
Model BT3561ABATTERY HiTESTER
Model BT3562BATTERY HiTESTER
Model BT3562-01(GP-IB version) BATTERY HiTESTER
Model BT3562A(LAN version) BATTERY HiTESTER
Model BT3563BATTERY HiTESTER
Model BT3563-01(GP-IB version) BATTERY HiTESTER
Model BT3563A(LAN version) BATTERY HiTESTER
Power Cord
Instruction Manual
Operating Precautions (0990A903)
OptionsThe following options are available for the instrument. Contact your authorized
Hioki distributor or reseller when ordering.
The options are subject to change. Visit our website for updated information.
Model L2107 CLIP TYPE LEAD (60 VDC or less)
Model 9453 FOUR TERMINAL LEAD (60 VDC or less)
Model 9467 LARGE CLIP TYPE LEAD (50 VDC or less)
Model 9770 PIN TYPE LEAD (60 VDC or less)
Model 9771 PIN TYPE LEAD (60 VDC or less)
Model L2100 PIN TYPE LEAD (1000 VDC or less)
Model L2110 PIN TYPE LEAD (1000 VDC or less)
Model Z5038 0 ADJ BOARD (for model L2100 and L2110)
Model 9637 RS-232C CABLE (9-pin to 9-pin/cross cable)
Model 9638 RS-232C CABLE (9-pin to 25-pin/cross cable)
Model 9151-02 GP-IB CONNECTOR CABLE (2 m)
Model 9642 LAN CABLE
Safety Information
HIOKI BT3562A981-09
This instrument is designed to comply with IEC 61010 Safety Standards, and
has been thoroughly tested for safety prior to shipment. However, mishandling
during use could result in injury or death, as well as damage to the instrument.
Be certain that you understand the instructions and precautions in the manual
before use. We disclaim any responsibility for accidents or injuries not resulting
directly from instrument defects.
This manual contains information and warnings essential for safe operation of
the instrument and for maintaining it in safe operating condition. Before using it,
be sure to carefully read the following safety precautions.
3
Safety Information
In the manual, the symbol indicates particularly important information
that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should
refer to a corresponding topic in the manual (marked with the symbol)
before using the relevant function.
The following symbols in this manual indicate the relative importance of cautions
and warnings.
Symbols for Various Standards
Other Symbols
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Indicates that incorrect operation presents an extreme hazard that
could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that
could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to
the user or damage to the instrument.
Indicates advisory items related to performance or correct operation of the instrument.
This symbol indicates that the product conforms to regulations set out by
the EU Directive.
Indicates a prohibited action.
(p. )
*
Indicates the location of reference information.
Indicates quick references for operation and remedies for troubleshooting.
Indicates that descriptive information is provided below.
4
HIOKI BT3562A981-09
Safety Information
Screen display
The screen of this instrument displays characters in the following manner.
Measurement categories
To ensure safe operation of measurement instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as CAT II to
CAT IV, and called measurement categories.
CAT II
CAT III
CAT IV
Primary electrical circuits in equipment connected to an AC electrical outlet
by a power cord (portable tools, household appliances, etc.)
CAT II covers directly measuring electrical outlet receptacles.
Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel
to outlets.
The circuit from the service drop to the service entrance, and to the power
meter and primary overcurrent protection device (distribution panel).
Accuracy
Using a measurement instrument in an environment designated with a highernumbered category than that for which the instrument is rated could result in a
severe accident, and must be carefully avoided.
Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV
measurement applications could result in a severe accident, and must be carefully avoided.
Hioki expresses accuracy as error limit values specified in terms of
percentages relative to reading and full scale, and digits.
Full scale
Reading
Digits
(Maximum displayable value)
Refers to the maximum displayable value of each measurement range.
The instrument has measurement ranges whose values are equal to the maximum displayable values. The limit values of full-scale errors are expressed in
percent of full scale (% of full scale, % f.s.).
(Displayed value)
Refers to the displayed value of the measuring instrument.
The limit values of reading errors are expressed in percent of reading (% of
reading, % rdg).
(Resolution)
Refers to the smallest change in the indication on the digital measuring instrument, i.e., the numeral one in the rightmost place. The limit values of digit
errors are expressed in terms of digits (dgt).
Operating Precautions
HIOKI BT3562A981-09
Follow these precautions to ensure safe operation and to obtain the full benefits
of the various functions.
Use of the instrument should confirm not only to itsspecifications, but also to the
specifications of all accessories, options, and other equipment in use.
Instrument Installation and Operating Environment
Operating temperature and humidity:
0 to 40°C (32 ± 104°F), 80%RH or less (non-condensating)
Temperature and humidity range for guaranteed accuracy:
23 ± 5°C (73 ± 9°F), 80% RH or less (non-condensating)
Avoid the following locations that could cause an accident or damage to the
instrument.
Exposed to direct sunlight
Exposed to high temperature
5
Operating Precautions
In the presence of corrosive or explosive gases
To avoid electric shock, do not remove the instrument's case. The internal components of the instrument carry high voltages and may become very hot during
operation.
Avoid using near electrically noisy devices, as the noise may impinge upon the
test object and cause unreliable measurements.
Installation Precautions
• The instrument should be operated only with the bottom downwards.
• Do not place the instrument on an unstable or slanted surface.
Exposed to water, oil,
other chemicals, or
solvents
Exposed to high humidity or condensation
Exposed to high levels of particulate dust
Exposed to strong electromagnetic fields
Near electromagnetic
radiators
Subject to vibration
50 mm or more
The instrument can be used with the stand. (p.15)
It can also be rack-mounted. Appendix (p. A17)
50 mm or more
10 mm or more
Rear
6
HIOKI BT3562A981-09
Operating Precautions
Preliminary Checks
Before using the instrument the first time, verify that it operates normally to
ensure that the no damage occurred during storage or shipping. If you find any
damage, contact your dealer or Hioki representative.
Before using the instrument, make sure that the insulation on the power cord
and test leads is undamaged and that no bare conductors are improperly
exposed. Using the instrument in such conditions could cause an electric
shock, so contact your dealer or Hioki representative for replacements.
Measurement Precautions
• To avoid electrical shock, be careful to avoid shorting live lines with the test
leads.
• Do not use the instrument to measure circuits that exceed its ratings or specifications. Damage to the instrument or overheating can cause
shock.
• The maximum input voltage and maximum rated voltage are shown in the
following table.
anelectric
ModelMaximum input voltage
BT3561A
BT3562
BT3562-02
BT3562A
BT3563
See:"9.1 Basic Specifications" (p.181)
• Never connect a battery cell or module to a motor or other load while it is
being measured. Doing so may result in a surge voltage, which may damage
the instrument or cause injury.
• To prevent electrical shock, verify the ratings of the measurement leads
before measurement and exercise care not to measure voltages that exceed
those ratings.
• Do not touch the metallic tip of probes after measuring high-voltage batteries. Doing so may result in electrical shock since internal instrument components could retain a charge under those conditions. (Internal discharge time:
Approx. 20 sec.)
• To avoid short-circuit accidents, connect the probe's banana terminals to the
instrument before connecting the probes to the battery.
±
60 V DC
±
60 V DC
±
100 V DC
±
300 V DC
Maximum rated voltage
to earth
±
60 V DC
±
70 V DC
±
100 V DC
±
300 V DC
7
HIOKI BT3562A981-09
Operating Precautions
• Use only the specified test leads and cables. Using a non-specified cable may
result in incorrect measurements due to poor connection or other reasons.
• To ensure certified measurement accuracy, allow at least 30 minutes warmup. After warm-up, be sure to execute self-calibration.
See "4.10 Self-Calibration" (p.73).
• The input circuitry includes a protective fuse. Measurement is not possible
when the fuse is blown.
• This instrument internally stores (backs up) all settings (except memory function and measurement values), such as measurement range, comparator
settings and etc., but only when no operation is performed for a certain time.
Therefore, to preserve settings, do not turn the power off for a short time
(about five seconds) after changing a setting. However, measurement settings made through the RS-232C, LAN, or GP-IB interface and measurement
settings loaded by LOAD
rized.
• Select an appropriate measurement range when measuring batteries. Using
a low range such as 3 m to measure a button cell or other battery that has
high internal resistance may result in an open-terminal voltage (approx. 4
V), causing the battery to be charged.
signals of the EXT I/O connector are not memo-
Before Connecting and Powering On
• Before turning the instrument on, make sure the supply voltage matches that
indicated on the its power connector. Connection to an improper supply voltage may damage the instrument and present an electrical hazard.
• To avoid electrical accidents and to maintain the safety specifications of this
instrument, connect the power cord provided only to a 3-contact (two-conductor + ground) outlet.
To suppress noise, the instrument needs to be set to match the frequency of
the power source. Before operating, set the instrument to the frequency of your
commercial power. If the supply frequency is not set properly, measurements
will be unstable.
See "2.5 Selecting the Line Frequency" (p.24).
Make sure the power is turned off before connecting or disconnecting the
power cord.
Handling the Instrument
• To avoid damage to the instrument, protect it from physical shock when
transporting and handling. Be especially careful to avoid physical shock from
dropping.
• Do not apply heavy downward pressure with the stand extended. The stand
could be damaged.
This instrument may cause interference if used in residential areas. Such use
must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
8
HIOKI BT3562A981-09
Operating Precautions
Handling the Test Leads and Cables
• To avoid breaking the test leads and cables, do not bend or pull them.
• Avoid stepping on or pinching cables, which could damage the cable insula-
tion.
• To avoid equipment failure, do not disconnect the communications cable
while communications are in progress.
• Use a common ground for both the instrument and the computer. Using different ground circuits will result in a potential difference between the instrument's ground and the computer's ground. If the communications cable is
connected while such a potential difference exists, it may result in equipment
malfunction or failure.
• Before connecting or disconnecting any the communications cable, always
turn off the instrument and the computer. Failure to do so could result in
equipment malfunction or damage.
• After connecting the communications cable, tighten the screws on the connector securely. Failure to secure the connector could result in equipment
malfunction or damage.
9
HIOKI BT3562A981-09
1.1 Product Overview
1
OverviewChapter 1
1.1Product Overview
The Model BT3561A, BT3562 and BT3563 Battery Hitester measure battery
internal resistance using a four-terminal, 1-kHz AC method, while simultaneously
measuring DC voltage (electromotive force [emf]). The high-precision, fast measurement performance and extensive interface capabilities make these models
ideal for incorporating into battery testing production lines.
Chapter 1 Overview
10
HIOKI BT3562A981-09
1.2 Features
1.2Features
Simultaneously Measures Battery Internal Resistance and Voltage
The four-terminal AC method measures resistance and DC voltage simultaneously, so battery internal resistance and emf are measured and judged at once.
High-Precision Measurements
Resistance measurement resolution of 1 or 0.1 and voltage measurement resolution of 10 V. The voltage measurement accuracy of ±0.01% of
reading is achieved.
High-Speed Measurements
Simultaneous resistance and voltage measurements can be performed as fast
as once every 20 ms.
(Response time of approx. 10 ms + sampling time of approx. 8 ms)
High-voltage measurement
The Model BT3563 supports measurement of high-voltage batteries of up to 300 V.
(the Model BT3561A, BT3562, and BT3562-01 supports measurement of up to
60 V) (the Model BT3562A supports measurement of up to 100 V)
Comparator Functions
Resistance and voltage measurement values are judged in three categories (Hi,
IN and Lo), with results clearly displayed. A comparator judgment beeper also
provides distinct sounds to indicate pass/fail judgments and to facilitate correct
recognition of judgment results.
Statistical Calculation Functions
Maximum, minimum and average measurement values, standard deviation, process capability indices and other values can be automatically calculated for
applications such as production management. Calculation results can also be
applied as comparator setting values.
Measurement Value Memory Function
The instrument includes a Memory function and storage capacity for up to 400
pairs of measurement values. When making many sequential measurements at
high speed and sending the measured values to a PC after each measurement,
the time to switch test objects can become unsatisfactorily long. The Memory
function can avoid the slow-down by sending stored measurements in batches
during idle times.
EXT I/O Interface
EXT I/O and RS-232C interfaces are equipped as standard, supporting transfer
rates up to 38,400 bps.
• Model BT3562-01 and BT3563-01 also supports GP-IB and analog output.
• Model BT3561A, BT3562A, and BT3563A also supports LAN and analog out-
put.
Printing Measurement Values and Statistical Results
Connect the printer to print measurement values and statistical calculation
results.
11
HIOKI BT3562A981-09
1.3 Names and Functions of Parts
1.3Names and Functions of Parts
Front Panel
Input Terminals (INPUT)
Connect the optional test leads.
See "2.3 Connecting the Optional Test Leads" (p.21)
Main Display
(p.12)
SOURCE-H
SOURCE-L
SENSE-H
SENSE-L
1
Chapter 1 Overview
Operating Keys
(p.13)
Sub Display
(p.12)
POWER Switch
Turns the instrument on and off (Standby).
Turns the power on and off (standby).
Off (standby)
On (press and hold for 1 second) Off (standby)
(The main power switch is located on the back of the instrument.)
See "2.4 Turning the Power On and Off" (p.22)
On (cancel standby)
12
HIOKI BT3562A981-09
1.3 Names and Functions of Parts
Main Display
The current measurement mode is indicated while measuring, and the setting item is displayed while
making settings.
(Upper row)
AUTO
Lit when measuring with Auto-Ranging.
EX.FAST, FAST, MED, SLOW
The selected Sampling Rate is lit.
0 ADJ
MEM
EXT TRIG
Lit when measuring
voltage.
Indicates percentage
units during relative
value comparator operation.
Lit when measuring in a range for
which Zero-Adjustment has been performed.
Lit when the Memory function is enabled.
Lit when the External Trigger function
is enabled.
Units of displayed measurement
Shows measured
value or setting item.
V
m
(Lower row)
V
STAT
AVG
LOCK
REMOTE
Unit of voltage
Unit of resistance (lit when the 3
to 3000 range is selected)
Unit of resistance (lit when the 3
m300 m range is selected)
Lit when the V (Resistance and
Voltage measurement) mode is selected.
Lit when the Statistical Calculation
function is enabled.
Lit when measuring with the Averaging setting enabled.
Lit when the keys are locked.
Lit during communications.
Shows Comparator Decision Result.
Indicates that the measured value is
Hi
above the upper threshold.
Indicates that the measured value is be-
IN
tween the upper and lower thresholds.
Indicates that the measured value is be-
Lo
low the lower threshold.
Sub Display
Upper and lower thresholds and other settings are displayed (when set).
Indicates Voltage measurement mode
Indicates percentage
units during relative value
comparator operation
While measuring, indicates the Comparator
function is enabled.
HIGH, LOWIndicates that absolute value comparator operation is enabled
(while measuring), and also when setting.
REF, %Indicates that relative value comparator operation is enabled
(while measuring), and also when setting.
VIndicates voltage measurement units.
13
HIOKI BT3562A981-09
1.3 Names and Functions of Parts
Operating Keys
To use a function marked on a
key, just press the key.
To use a function printed under a key (blue letter), press
the SHIFT key first (and confirm the SHIFT lamp is lit),
and then the key.
SHIFT Lamp
[ ]: Enabled after pressing the SHIFT key (SHIFT lamp lit).
Operating
Key
Description
Operating
Key
1
Chapter 1 Overview
Use numeric keypads to enter numerical values.
(Numerical values
can be used with
the RANGE key.)
Description
V/ / V
[0 ADJ]
LOAD
[SAVE]
TRIG
[INT/EXT]
VIEW
STAT
[DELAY]
SMPL
[AVG]
COMP
[SET]
LOCAL
Selects Measurement mode.
(Resistance and voltage measurement, Resistance measurement or
Voltage measurement)
Executes Zero-Adjustment.
Loads a saved measurement configu-
ration (Panel settings).
Saves the current measurement con-
figuration (Panel settings).
Executes a Manual Trigger event.
Selects internal/external triggering.
Switches the view mode of the V
mode.
Displays and sets Statistical Calculation results.
Sets the Trigger Delay.
Selects the Sampling Rate.
Activates Averaging function settings.
Switches the Comparator function on
and off.
Activates Comparator function setting.
Cancels remote control (RMT) and reenables key operations.
PRINT
AUTO
[LOCK]
ENTER
[MENU]
RANGE
[V RANGE]
SHIFT
Sends measurement values and statistical calculation results to the printer.
Switches between Auto and Manual
range selection.
Switches the Key-Lock function on and
off.
Applies settings.
Selects various operating functions
and settings.
Up/Down:
Changes setting value or numerical
value, and sets the resistance measurement range.
Left/Right:
Moves the setting item or digit.
Up/Down:
Sets voltage measurement range.
• Enables the functions of the operat-
ing keys marked in blue.
The lamp is lit when the SHIFT
state is active.
• Cancels settings in various setting
displays. (Returns to the Measurement display without applying settings.) However, this does not apply
to Menu display. However, from a
menu item display, changed settings are not canceled, but
accepted as the display returns to
measurement display (except after
Zero-Adjustment clear or resetting).
14
HIOKI BT3562A981-09
1.3 Names and Functions of Parts
Rear Panel
* The illustration shows the Model BT3563-01 Battery Hitester (GP-IB version).
2
1
3
4
* The illustration shows the Model BT3563A Battery Hitester (LAN version).
2
1
5
6
3
7
8
15
HIOKI BT3562A981-09
1.3 Names and Functions of Parts
No.
Main power switch
1
Power InletConnect the supplied power cord here.
2
Analog output connector
(Model BT3561A, BT3562A, BT3562-
3
01, BT3563A, and BT3563-01 only)
EXT I/O ConnectorConnect here to use the EXT I/O interface.
4
GP-IB Connector
5
(Model BT3562-01, BT3563-01 only)
RS-232C ConnectorConnection for the printer or RS-232C interface.
6
LAN Connector
(Model BT3561A, BT3562A, and
7
BT3563A oniy)
MAC addressMAC address of the LAN.
8
MameDescription
: Main power off
:Main power on
Connect when using analog output (of resistance
measured values).
Connect here to use the GP-IB interface.
Connect here to use the LAN interface.
Refer-
rence
p. 22
p. 20
p. 99
p. 80
p. 102
p. 102
p. 102
1
Chapter 1 Overview
Side View
Stand
Can be opened to tilt the
front panel upwards.
Do not apply heavy downward pressure with the stand extended. The stand
could be damaged.
16
HIOKI BT3562A981-09
1.4 Menu Display Sequence (SHIFT ENTER)
1.4Menu Display Sequence (SHIFT ENTER)
Various auxiliary settings can be performed from the menu item displays.
(SHIFT Lamp lit)
The Menu display appears.
(Main Display)
Zero-Adjustment Clear
display (p.35)
Measurement Value Output
function setting display
(p.74)
Key Beeper setting
display (p.75)
The up/down RANGE key changes the setting
Interface Selection
display (p.105)
EOM-signal setup dis-
play (p.84)
Line Frequency set-
ting display
(p.24)
shown on the Sub Display.
Self-Calibration setting
display (p.73)
ERR Output Selection
display (p.84)
Measurement Current
Pulse Output
display(p.61)
Configuring the Absolute
Value Judgment display
(p.55)
Settings on the menu item displays are applied and saved internally when
changed.
Reset display
(p.76)
Pressing this key returns to the
previous item display.
17
HIOKI BT3562A981-09
1.5 Measurement Flowchart
1.5Measurement Flowchart
The basic measurement process flow is as follows:
Measurement Preparations
Connecting the power cord (p.20)
Connecting the test leads (p.21)
Turning the power on (p.22)
Selecting the line frequency (p.24)
Instrument’s Settings
Selecting measurement mode (p.30)
1
Chapter 1 Overview
Selecting measurement range (p.31)
Selecting sampling rate (p.34)
Zero-Adjustment
Short the test leads together (p.35)
Executing zero-adjustment
Measurement Start
Connect the test leads to a test object.
Read the measured value (p.38)
For details about the functions that can be applied to measurement values such
as comparator, trigger and averaging functions, refer to "Chapter 4 Applied
Measurement" (p.41).
18
HIOKI BT3562A981-09
1.5 Measurement Flowchart
Measurement
HIOKI BT3562A981-09
19
2.1 Preparation Flowchart
Preparations
Chapter 2
2.1Preparation Flowchart
This procedure describes instrument preparations such as making connections and turning power on.
(p.21)(p.20)
1
Rear Panel
3
Connecting the power cord.
1
2
Front Panel
(p.22)
5
(p.20)
2
Chapter 2 Measurement Preparations
4
Connect the test leads to the instrument.
2
Connect the EXT I/O connector and inter-
3
face connector.
Turn the power on.
4
Set measurement settings.
5
Start measurement.
6
Verify that the instrument’s line frequency is correctly set when using it for the
first time and after initialization following repair or recalibration.
See "2.5 Selecting the Line Frequency" (p.24).
(p.21)
(p.102)
(p.22)
(p.25)
20
HIOKI BT3562A981-09
2.2 Connecting the Power Cord
2.2Connecting the Power Cord
To avoid electrical accidents and to maintain the safety specifications of this
instrument, connect the power cord provided only to a 3-contact (two-conductor + ground) outlet.
To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it from the power outlet.
To suppress noise, the instrument needs to be set to match the line frequency.
Before operating, set the instrument to the frequency of your commercial
power. If the supply frequency is not set properly, measurements will be unstable.
See "2.5 Selecting the Line Frequency" (p.24).
Make sure the power is turned off before connecting or disconnecting the
power cord.
Rear Panel
1
2
1. Confirm that the instrument's Main power
switch (rear panel) is OFF(
).
2. Check that the power supply voltage ( 100
V to 240 V) is correct, and connect the
power cord to the power inlet socket on
the rear of the instrument.
3. Plug the power cord into the AC outlet.
2.3 Connecting the Optional Test Leads
HIOKI BT3562A981-09
2.3Connecting the Optional Test Leads
• To prevent an accident caused by short-circuiting the battery, be sure to verify
that nothing is connected to the tips of the measurement leads before connecting the leads to or disconnecting them from the instrument. (Contact between
the banana terminals while the tips of the measurement leads are connected
to the battery will short-circuit the battery, possibly resulting in serious injury.)
• To prevent electrical shock, verify the ratings of the measurement leads before
measurement and exercise care not to measure voltages that exceed those
ratings.
Test leads are not included as standard accessories with the instrument, so the
appropriate options need to be purchased separately or constructed according to
the user’s application requirements. To construct custom test leads, refer to "Precautions for Making Custom Test Leads"(p.A1). The resistance measurement terminals on this instrument consist of four separate banana jacks.
See "Appendix 1 Precautions for Making Custom Test Leads"(p.A1).
21
2
Chapter 2 Measurement Preparations
1. Confirm that the instrument's Power
switch is OFF.
2. Verify that nothing is connected to the tips
of the four-terminal measurement leads.
3. Connect four-terminal test leads such as
Red Lead
Black Lead
Example: Optional model L2107 CLIP TYPE LEAD
About Test Leads ______________________________________________
(Example: Model L2107 CLIP TYPE LEAD)
The side with “V” mark is SENSE.
the L2107 CLIP TYPE LEAD to INPUT.
Plug the mark on the red lead into the
red marked jack on the instrument,
and plug the mark on the black lead into
the black marked jack on the instrument.
SENSE
SOURCE
SENSE
SOURCE
Red
Black
When clipping a thin line
(Clip the line at the tip,
serrated part of the
jaws.)
Red
SENSE
SOURCE
SENSE
SOURCE
Black
When clipping a thick line
(Clip the line at the
deep, non-serrated part
of the jaws.)
22
HIOKI BT3562A981-09
2.4 Turning the Power On and Off
2.4Turning the Power On and Off
Before turning the instrument on, make sure the supply voltage matches that
indicated on the its power connector. Connection to an improper supply voltage may damage the instrument and present an electrical hazard.
• The measurement setting state is the same as when the power was previ-
ously turned off (backup).
To preserve changes to settings, wait a short time (about five seconds) after
changing a setting before turning power off.
• However, measurement settings made through the RS-232C, LAN, or GPIB interface and measurement settings loaded by LOAD
I/O connector are not memorized.
• Before starting to measure, allow 30 minutes for warm-up.
After warm-up, be sure to perform a self-calibration.
See "4.10 Self-Calibration" (p.73).
signals of the EXT
Turning On () the Main Power Switch (Rear of Instrument)
Turn on the main power switch on ( ) the rear of the instrument.
The instrument will start up in the standby state in which it was last
turned off. (The instrument ships in the standby state.)
Power ON
Turning the Power Off
Turn off the main power switch on the rear of the instrument.().
Power OFF
Cancelling the Standby State
HIOKI BT3562A981-09
Press the power switch on the front of the instrument while it is in the
standby state.
23
2.4 Turning the Power On and Off
(Main Display)
Model name
(Main Display)
Software version
(Sub Display)
Line frequency
Interface
The measurement display appears.
Placing the Instrument in the Standby State
Press and hold the power switch on the front of the instrument for
approximately 1 second while it is in the operating state.
2
Chapter 2 Measurement Preparations
24
HIOKI BT3562A981-09
2.5 Selecting the Line Frequency
2.5Selecting the Line Frequency
The instrument's power supply frequency must be set in order to eliminate noise.
Although the power supply frequency setting is configured automatically ("AUTO") by default, it can also be
set manually. Measured values will not stabilize if the power supply frequency is not set properly.
1
2
3
(The SHIFT indicator lights up.)
The Menu display appears.
(Main Display)
(Sub Display)
Select the Line Frequency setting display.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
(Main Display)
(Sub Display)
Select the frequency of the AC mains supply being used.
4
(Main Display)
(Sub Display) flashing
AUTO.... Automatic configuration of power supply frequency
50 .......... 50 Hz
60 .......... 60 Hz
Applies settings and returns to the Measurement display.
• When set to automatic configuration (AUTO), a power supply frequency of
either 50 Hz or 60 Hz will be automatically detected whenever the instrument is turned on or reset.
• Changes in the power supply frequency occurring at other times will not be
detected.
• The power supply frequency will be set to either 50 Hz or 60 Hz, whichever
is closer.
3.1 Pre-Operation Inspection
HIOKI BT3562A981-09
MeasurementChapter 3
Before starting measurement, please read Operating Precautions (Page 5) and
"Chapter 2 Measurement Preparations" (p.19).
• To avoid electrical shock, be careful to avoid shorting live lines with the test
leads.
• Do not use the instrument to measure circuits thatexceed its ratings or specifications. Damage to the instrument or overheating can cause anelectric
shock. The maximum input voltage and maximum rated voltage are shown in
the following table. see: "9.1 Basic Specifications" (p.181)
25
3
Chapter 3 Measurement
Model
BT3561A±60 V DC±60 V DC
BT3562
BT3562-02
BT3562A±100 V DC±100 V DC
BT3563±300 V DC±300 V DC
To prevent electrical shock, verify the ratings of the measurement leads before measurement and exercise care not to measure voltages that exceed those ratings.
Maximum input
voltage
±60 V DC±70 V DC
Maximum rated
voltage to earth
3.1Pre-Operation Inspection
Before using the instrument for the first time, verify that it operates normally to ensure that no damage
occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Before using the instrument, perform the following inspection to ensure that it is operating properly.
Check PointCheck Contents
Instrument Chassis
(both front and rear panels)
Test Leads and Power
Cord
Good Test Sample• Measures as good and displays the correct
Bad Test Sample• Measures as bad and displays the correct
If the inspection reveals a defect, stop using the instrument and contact your
dealer or Hioki representative.
• No damage or cracks
• No internal circuitry is exposed
• Metal parts that should be insulated are not
exposed
measurement value
measurement value
26
HIOKI BT3562A981-09
3.2 Basic Measurement Example
3.2Basic Measurement Example
The following example describes the measurement process.
Example: Measuring resistance and voltage of a 30 m lithium-ion battery
Required items:Lithium-ion battery (30 m)
Test leads: Model 9770 PIN TYPE LEAD are used here.
Measurement
conditions:
Preparations
Measurement mode.................. V (Resistance and Voltage measurement)
Range ....................................... 30 m
Sampling rage .......................... SLOW
Zero adjustment........................ Enabled
6 V
1
2
Connect the power cord.
See "2.2 Connecting the Power Cord" (p.20).
3
Power ON
Connect the test leads.
See "2.3 Connecting the Optional Test Leads" (p.21).
Red Lead
Black Lead
2
1
4
3
4
Example: Model 9770
Turn the main power switch on.
See "2.4 Turning the Power On and Off" (p.22).
See "2.5 Selecting the Line Frequency" (p.24).
Cancel the standby state.
See "2.4 Turning the Power On and Off" (p.22).
Instrument Settings
HIOKI BT3562A981-09
27
3.2 Basic Measurement Example
5
6
Confirm the SHIFT lamp is not lit.
If this is lit, press the SHIFT key to turn it off.
SHIFT not lit
Select the Resistance Measurement mode.
(Here, resistance and voltage measurement is selected.)
See "3.3 Selecting Measurement Mode" (p.30).
V lit
The measurement mode changes each time
you press this key. Select to measure only
resistance, or V to measure only voltage.
V
3
Chapter 3 Measurement
V
7
8
Set the measurement range. (Here, 30 m range is selected.)
See "3.4 Setting Measurement Range" (p.31).
m lit
Increase the resistance measurement range.
Decrease the resistance measurement range.
(SHIFT Lamp lit)
Set the voltage measurement range. (Here, the 6 V setting has been selected.)
See "Voltage measurement range" (p.32)
V lit
Increase the voltage measurement range.
Decrease the voltage measurement range.
28
HIOKI BT3562A981-09
3.2 Basic Measurement Example
9
Set the sampling rate. (Here, SLOW is selected.)
See "3.5 Setting Sampling Rate" (p.34).
Zero-Adjustment
10
Short the test leads together.
Proper Zero-Adjustment is not possible with incorrect wiring.
See "3.6 Zero-Adjust Function" (p.35).
Example: Model 9770 Pin Type Lead
SLOW lit
The sampling rate changes each time
you press this key.
EX.FAST
FAST
MED
SLOW
11
Internal
conductor
Bring the pins into contact at 3 points.
• Internal conductor and internal conductor
External
conductor
(The SHIFT indicator lights up.)
Execute Zero-Adjust.
After zero-adjustment, the display returns to the measurement mode.
• Internal conductor and external conductor
• External conductor and external conductor
Model 9770
0ADJ lit
“Err.02” appears if Zero-Adjustment fails. Verify that the test lead tips are properly shorted,
and try zero-adjustment again.
Measurement
HIOKI BT3562A981-09
29
3.2 Basic Measurement Example
12
13
Connect the test leads to a battery.
Open-terminal voltages for the instrument are as follows:
and 30 m ranges: 25 V peak
3 m
range: 7 V peak
300 m
to 3000 range: 4 V peak
3
These voltages derive from the load associated with charging the 1.2 F
capacitor inside the instrument.
• The open-terminal voltage for the 3 m
4 V approximately 500 ms after the terminal is placed in the open state.
• When building a measurement line using scanners, use a relay with a dielectric strength that is greater than or equal to the open-terminal voltage for the
range being used.
Read the measured resistance and voltage.
3
Chapter 3 Measurement
, 30 m, and 300 m ranges peaks at
Measured Resistance
Measured Voltage
See "3.7 Displaying Measurement Results" (p.38).
See "10.3 Error Display" (p.191).
Please refer to "Before returning for repair." of "Measured value is unstable."
(p.190) as a measurement and attention.
30
HIOKI BT3562A981-09
3.3 Selecting Measurement Mode
3.3Selecting Measurement Mode
Select the measurement mode from V (both resistance and voltage measurement), (resistance measurement only) or V (voltage measurement only).
1
2
“V” indicates the
V mode is selected
Confirm the SHIFT lamp is not lit.
If this is lit, press the SHIFT key to turn it off.
SHIFT not lit
Switches the displayed measurement mode.
Each key-press switches the measurement mode.
V mode
(Resistance and Voltage measurement) measurement)
“” or “m” lit
mode
(Resistance measurement)
V mode
(Voltage measurement)
The fastest measurements are provided by selecting the or V mode when
measuring resistance or voltage, respectively.
See " Sampling Time" (p.182).
3.4 Setting Measurement Range
HIOKI BT3562A981-09
3.4Setting Measurement Range
This section describes how to set the measurement range for resistance or voltage measurement.
Choose a resistance measurement range from among the following ranges:
• Six ranges, including the 30 mΩ through 3000 Ω ranges (For the BT3561A)
• Seven ranges, including the 3 mΩ through 3000 Ω ranges (For the BT3562 and BT3563)
Choose a voltage measurement range from among the following ranges:
• Two ranges, including the 6 V through 60 V ranges (For the BT3561A, BT3562, and BT3562-01)
• Three ranges, including the 6 V through 100 V ranges (For the BT3562A)
• Three ranges, including the 6 V through 300 V ranges (For the BT3563)
There is also an auto-range function, which determines the optimal range automatically.
Resistance measurement range
31
3
Chapter 3 Measurement
1
2
Select the resistance measurement range.
The position of the decimal point and unit of measurement on the display will
be switched according to the selected range.
Select the range to use.
When the 3 range is selected
lit
Increase the resistance measurement range.
When the 300 m
selected
range is
Decrease the resistance measurement range.
m lit
Pressing the up or down keys while in auto-range mode will cancel auto-ranging,
leaving the current measurement range as the manually set range.
32
HIOKI BT3562A981-09
3.4 Setting Measurement Range
Voltage measurement range
1
Select the range to use.
2
Select the voltage measurement range.
The position of the decimal point and unit of measurement on the display will
be switched according to the selected range.
When the 60 V range is selected
When the 6 V range is
selected
V lit
Increase the voltage measurement range.
Decrease the voltage measurement range.
V lit
Auto-Ranging
HIOKI BT3562A981-09
33
3.4 Setting Measurement Range
When manual range selection is enabled, pressing this enables auto-ranging.
The most suitable measurement range is then selected automatically.
AUTO lit
3
Chapter 3 Measurement
The auto-range setting (on/off) for the V function applies to both resistance and
voltage measurement.
Switching from
Auto
-ranging back to
Manual
range selection
Press the AUTO key again. The range can now be changed manually.
• Depending on the state of the test object, auto-ranging may be unstable. In
this case, select the range manually, or increase the Delay time.
• Auto-ranging is not available when Comparator or Memory functions are
enabled (ON).
• Refer to "Specifications" (p.181) for details about accuracy.
Resistance Measurement Mode
RangeDisplayed Values
Measured CurrentOpen-Terminal Voltage
3 m-0.1000 to 3.1000 m100 mA 25 V peak
30 m-1.000 to 31.000 m100 mA25 V peak
300 m-10.00 to 310.00 m10 mA7 V peak
3 -0.1000 to 3.1000 1 mA4 V peak
30 -1.000 to 31.000100 A4 V peak
300 -10.00 to 310.00 10 A4 V peak
3000 -100.0 to 3100.0 k10 A4 V peak
6 V-6.00000 V to ±6.00000 V----
60 V-60.0000 V to ±60.0000 V----
*1
100 V
300 V
-100.000 V to ±100.000 V----
*2
-300.000 V to ±300.000 V----
*1: BT3562A only
*2: BT3563A, BT3563 or BT3563-01 only
34
HIOKI BT3562A981-09
3.5 Setting Sampling Rate
3.5Setting Sampling Rate
The sampling rate can be selected from EX.FAST, FAST, MEDIUM and SLOW.
Slower sampling rates generally provide greater measurement precision.
Selects the sampling rate
EX.FASTFASTMEDIUMSLOW
• Measurements are especially susceptible to interference from the environ-
ment when EX.FAST is selected, so countermeasures such as shielding or
twisting of test leads, cables and wiring around the test object may be necessary.
See "Appendix 1 Precautions for Making Custom Test Leads" (p.1)
• When SLOW sampling is selected, self-calibration is executed during each
measurement. At other sampling rates, self-calibration is executed manually
or automatically every 30 minutes.
See "4.10 Self-Calibration" (p.73).
• Refer to the specifications for details of sampling rates.
See " Sampling Time" (p.182).
3.6Zero-Adjust Function
HIOKI BT3562A981-09
Execute zero adjustment before measuring to nullify any residual offset voltage
from the instrument or measurement environment. Measurement accuracy specifications are applicable after zero adjustment. Zero adjustment can also be executed by the 0ADJ terminal of the EXT I/O connector.
See "5.2 Signal Descriptions" (p.80).
35
3.6 Zero-Adjust Function
Wiring Method for Zero-Adjustment
Before executing zero adjustment, connect the test leads (probes) as follows:
1. Connect SENSE-H to SENSE-L.
2. Connect SOURCE-H to SOURCE-L.
3. Connect the joined SENSE and SOURCE leads together as shown below.
SENSE-HSENSE-L
Executing Zero-Adjustment
1
Position the measurement leads in the actual measurement state.
Since the amount of zero adjustment varies with the position and state of the mea-
surement leads (probes) (i.e., their length, shape, position, etc.), the measurement leads must be positioned in the actual measurement state before performing
zero adjustment.
SOURCE-HSOURCE-L
Connection
3
Chapter 3 Measurement
Connection
at zero adjustment at measurement
These variations are particularly pronounced in the 3 m and 30 m ranges, so
be sure to position the leads in same state as will be used to perform actual measurement when using those configurations.
36
HIOKI BT3562A981-09
3.6 Zero-Adjust Function
2
Short the test leads together.
Proper zero adjustment is not possible with incorrect wiring.
Example: Model L2107 CLIP TYPE LEAD
Correct
SENSE
SOURCE
Red
Bring the "V" marks together at the
same position.
SENSE
SOURCE
Black
Incorrect
SENSE
SOURCE
Model 9453 (Option)
Model 9771 (Option)
When the resistance measurement value is displayed
as "-----", change the facing
direction.
Red
Perform zero adjustment with the alligator clips and lead
rods placed as
above.
Black
SOURCE
SENSE
Model 9770 (Option)
Bring the pins into contact at 3 points.
Internal
conductor
Let the two points of the pin
tip touch the spring part perpendicularly (be careful not
to short the springs).
Model L2100 (Option)
Line
Model Z5038 0 ADJ Board
Each SENSE pin has a line affixed to its
base. When using the zero-adjust feature,
orient the test lead so that the surface the
same direction.
Choose a hole suited to the distance between the terminals on the battery subject to
measurement and hold the test lead against
the zero-adjust board so that it remains symmetrical to the central plus sign (+) on
the board, inserting each SENSE (the side
with a line) pin into the larger side of each
elongate hole.
External conductor
Model L2110 (Option)
Flat
surface
Model Z5038 0 ADJ Board
The SENSE side of the tip of the test lead
has a flat surface. When using the zeroadjust feature, orient the test lead so that
the surface the same direction.
Choose a hole suited to the distance between the terminals on the battery subject
to measurement and hold the test lead
against the zero-adjust board so that it remains symmetrical to the central plus sign
(+) on the board, inserting each SENSE
(the side with the surface) pin into the larger side of each elongate hole.
37
HIOKI BT3562A981-09
3.6 Zero-Adjust Function
3
(The SHIFT indicator lights up.)
Zero-adjust display appears.
After measurement, the measured value of the compensation applied by the zero-adjust
function is displayed.
The range of zero adjustment is up to 1000 digits.
Clearing Zero-Adjustment
1
(The SHIFT indicator lights up.)
The Menu display appears.
0ADJ lit
3
Chapter 3 Measurement
2
If Err02 is displayed
(Main Display)
(Sub Display) flashing
The zero-adjust value is cleared. (0ADJ not lit)
(Main Display)
Indicates that zero adjustment could not be executed, either because the range
to be adjusted exceeds
The zero adjust function is canceled, so repeat the operation after correcting
the cause of the error.
• Zero adjustment is limited to ±1000 digits (all ranges)
• Perform zero adjustment for each range that will be used in measurement.
• When using the auto-range function, perform zero adjustment for all ranges.
• When using the V function, the 0ADJ indicator lights up or turns off accord-
ing to the resistance measurement range zero-adjust state.
• Zero-adjustment values are retained even when power is turned off.
• The 0ADJ terminal of the EXT I/O connector also executes zero adjustment.
See "5.2 Signal Descriptions" (p.80).
±1000 digits or a measurement fault condition exists.
38
HIOKI BT3562A981-09
3.7 Displaying Measurement Results
3.7Displaying Measurement Results
In the V mode, resistance measurements appear on the upper display, and
voltage measurements appear on the lower display.
Measured Resistance
Measured Voltage
In the
mode, resistance measurements appear on the upper display.
Measured Resistance
In the V mode, voltage measurements appear on the upper display.
Measured Voltage
Measurement Fault Detection
HIOKI BT3562A981-09
If a measurement does not execute properly, a measurement fault
“- - - - -” is indicated on the display.
In addition, a measurement fault signal (ERR) is output at the EXT I/O connector.
See " ERR Output" (p.83).
A measurement fault is displayed in the following cases.
• When a test lead is not connected to the test object
• When the resistance of the measured object is over-range
Example: Attempting to measure 30 with the 300 m range selected.
• When there is a break in a probe wire
• When the contact resistance is high due to probe wear, dirt, or other factors,
or when the wiring resistance is high (see chart below)
• If the circuit protection fuse is blown
See "10.1 Troubleshooting" (p.189).
Levels at which a measurement fault is detected
A measurement fault will result when the resistance values (contact resistance +
wiring resistance + test object resistance) between the source H and L or the
sense H and L leads is greater than or equal to the values in the following table:
39
3.7 Displaying Measurement Results
3
Chapter 3 Measurement
BT3562, BT3562-01, BT3563, BT3563-01
RangeSOURCE H-LSENSE H-L
3 m2 2
2 2
30 m
300 m15 15
3 150 15
1.5 k150
30
6 k1.5 k
300
6 k15 k
3000
* Large contact resistance and/or wiring resistance values may increase the error
component in measured values. (Accuracy is not guaranteed when the sum of
contact resistance and wiring resistance is greater than or equal to 20 [for the
3 m and 30 m ranges, 2 ].)
* The instrument may be unable to detect measurement faults when the mea-
surement lead capacitance is greater than or equal to 1 nF.
40
HIOKI BT3562A981-09
3.7 Displaying Measurement Results
BT3561A, BT3562A, BT3563A
30 m5.5 6.5
300 m15 30
3000 6 k25 k
• A high contact resistance or a high wiring resistance may cause an increase in
measurement errors.
Hioki cannot guarantee the measurement accuracy if the sum of contact resistance and wiring resistance reaches 40 (4 for the 3 m and 30 m
ranges).
• The instrument may not detect measurement malfunctions if the measurement lead has a capacitance of 1 nF or more.
RangeSOURCE H-LSENSE H-L
3 m5.5 6.5
3 150 30
30 1.5 k150
300 6 k2.5 k
Overflow Display
Overflow is indicated by “OF” or “-OF” on the display, caused by one of the following:
DisplayCondition
• The measured value exceeds the limit of the current
measurement range
OF
-OF
• The test object impedance exceeds the input level.
• When the result of relative value calculation is larger
than +99.999%.
• The measured value is below the limit of the current
measurement range
• The test object impedance exceeds the input level (in
the negative direction).
• When the result of relative value calculation is smaller
than -99.999%.
Applied
HIOKI BT3562A981-09
41
Measurement
This chapter describes advanced operations employing the Comparator, Statistical Calculation and Memory functions.
Judge measurement values
against specified thresholds
Measure when trigger events occurTrigger Function
Output averaged measurement valuesAveraging Function
Display the results of calculation expressions applied to measurement values
Store measurement valuesMemory Function
Lock the keysKey-Lock Function
Save measurement configurationsPanel Save Function
Load saved measurement configurations Panel Load Function
Chapter 4
Comparator Function
Statistical Calculation Functions
(p.42)
(p.59)
(p.63)
(p.64)
(p.68)
(p.70)
(p.71)
(p.72)
4
Chapter 4 Applied Measurement
Increase measurement precisionSelf-Calibration
Output measurement values via the RS232C interface according to trigger input
timing
Enable/disable key-press beepsKey Beeper Setting
Re-initialize the instrumentReset Function
Measurement Value
Output Function
(p.73)
(p.74)
(p.75)
(p.76)
42
HIOKI BT3562A981-09
4.1 Comparator Function
4.1Comparator Function
The comparator function compares measured values to preset upper and lower
thresholds, judges the measurements according to their relative levels within the
preset range, and indicates the results of the comparisons.
Comparator thresholds can be set either by specifying upper and lower thresholds, or by specifying a reference value and tolerance.
Comparator results can be indicated by the Hi, IN and Lo LEDs, beeper sound
and signal output at the EXT I/O connector.
See "Chapter 5 External Control (EXT I/O)" (p.79).
The comparator setting process flow is as follows:
Display the comparator settings
1
Set the comparator judgment beeper
2
Select the comparator execution mode
3
(Auto or Manual/External)
Select resistance measurement
(If you do not need to configure resistance settings,
4
proceed to step 7.)
Select the resistance comparison method.
5
(absolute or relative value) for the comparator
Specify the resistance upper and lower thresholds
6
(or reference value and tolerance).
Select voltage measurement
7
]
Select the voltage comparison method
8
(absolute or relative value) for the comparator
Specify the voltage upper and lower thresholds
9
(or reference value and tolerance)
Apply your comparator settings
10
Enable the Comparator function
11
4.1 Comparator Function
HIOKI BT3562A981-09
Comparator Setting Example 1 (Upper and Lower Threshold Judgment)
This example describes the comparator setting method.
Example:
Set the upper and lower thresholds for resistance and voltage in the V mode (300
m range), and indicate whether the measurement value exceeds the upper or
lower thresholds by sounding the beeper.
Resistance: Upper threshold value 150.00 m, Lower threshold value 100.00 m
Voltage: Upper threshold value 15.2000 V, Lower threshold value 15.0000 V
1
Confirm that the Comparator function is OFF.
First make sure the Comparator function is disabled. Settings cannot be changed while
the Comparator function is enabled. Press the COMP key, if necessary, to disable the
Comparator function.
43
4
Chapter 4 Applied Measurement
2
3
COMP not lit
Select the V measurement mode.
V lit
Select the Resistance measurement range (for this example, the 300 m range).
m lit
Increase the resistance measurement range.
Decrease the resistance measurement range.
4
Select the voltage measurement range (for this example, the 60 V range).
V lit
Increase the resistance measurement range.
Decrease the resistance measurement range.
44
HIOKI BT3562A981-09
4.1 Comparator Function
5
6
The Comparator setting display appears.
oFF flashing
Set the comparator judgment beeper (for this example, select HL).
HL flashing
oFF........ no beeps sound
HL.......... beeps repeatedly (when measurements are Hi or Lo)
in ...........beeps continuously (when measurements are IN)
btH1.......beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2.......beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
7
8
Press so that the indicated position blinks, and select the comparator execution
mode (for this example, Auto).
A flashing
A............ Auto Comparator (default setting)
E ............ Manual Comparator
Press so that the indicated position blinks, and select resistance.
r flashing
r .............Resistance
u ............Voltage
45
HIOKI BT3562A981-09
4.1 Comparator Function
9
10
Or
numeric key-
pads
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, HIGH/LOW).
HIGH & LOW flashing
HIGH, LOW..... Compare by upper and lower thresholds (default setting)
REF, % ............ Compare by reference value and tolerance
Switch to the upper/lower threshold setting display, and specify the thresholds.
For this example,
Upper Threshold: 150 m
Upper Threshold: 100 m
Using the RANGE keys:Using the numeric keypads:
Select a digit to change by mov-
ing the blinking location, then select the new numerical value.
Press the numeric keys corresponding to the digits to
be entered.
4
Chapter 4 Applied Measurement
11
Select a digit
Select numerical value
To enter the current measurement as the setting value: AUTO key
(Press on a screen other than the upper/lower threshold setting display.)
To enter the result of statistical calculation as the setting value: STAT key
(Press on a screen other than the upper/lower threshold setting display.)
See " Upper and Lower Thresholds Setting (by Reference Value and Tolerance)" (p.54).
Press so that the indicated position blinks, and select voltage.
u flashing
r............. Resistance
u ............ Voltage
46
HIOKI BT3562A981-09
4.1 Comparator Function
12
13
14
Or
numeric key-
pads
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, HIGH/LOW).
HIGH & LOW flashing
HIGH, LOW ..... Compare by upper and lower thresholds (default setting)
REF, %............. Compare by reference value and tolerance
Switch to the upper/lower threshold setting display, and specify the thresholds.
For this example,
Upper Threshold: 15.2 V
Upper Threshold: 15 V
Applies setting and returns to the Measurement display.
The comparator function is enabled.
15
COMP lit
To cancel the settings: SHIFT key
Connect a test object and judge the measured value.
Measured Resistance
Judgment Result
Measured Voltage
In the V mode, you can verify comparator settings by pressing the VIEW key.
See " Switching Between Measurement Value and Comparator Setting Displays" (p.58).
Upper Threshold Value < Measured Value
Lower Threshold Value
Measured Value < Lower Threshold Value
Measured Value Upper Threshold Value
4.1 Comparator Function
HIOKI BT3562A981-09
• The upper and lower thresholds are saved as the displayed counts (independent of measurement mode and range). Therefore, changing the measurement mode or range results in the same display counts representing different
absolute values.
Example:
To specify the lower threshold as 150 m
Switching to the 3
old to 1.5 .
• The instrument can also base judgments on the absolute value of voltage
measured values (to prevent Lo judgments when the positive and negative
terminals are connected backwards).
See "Configuring the Absolute Value Judgment Function (Voltage)" (p.55)
range after making this setting changes the lower thresh-
in the 300 m range, enter “15000”.
Comparator Setting Example 2 (Reference Value and Tolerance Judgment)
This example describes the comparator setting method.
47
4
Chapter 4 Applied Measurement
Example:
Set a reference value and tolerance in the V mode (3 range), and set the beeper
to sound while measured values are within tolerance.
Resistance: Reference value 1.5 , Tolerance 5%
Voltage: Reference value 4.2 V, Tolerance 0.5%
1
Confirm that the Comparator function is OFF.
First make sure the Comparator function is disabled. Settings cannot be changed while
the Comparator function is enabled. Press the COMP key, if necessary, to disable the
Comparator function.
COMP not lit
Select the V measurement mode.
2
V lit
3
Select the measurement range (for this example, the 3 range).
lit
Increase the resistance measurement range.
Decrease the resistance measurement range.
48
HIOKI BT3562A981-09
4.1 Comparator Function
4
5
6
Select the voltage measurement range (for this example, the 6 V range).
V lit
The Comparator setting display appears.
oFF flashing
Set the comparator judgment beeper (for this example, select IN).
7
in flashing
oFF........ no beeps sound
HL.......... beeps repeatedly (when measurements are Hi or Lo)
in ...........beeps continuously (when measurements are IN)
btH1.......beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2.......beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
Press so that the indicated position blinks, and select the comparator execution
mode (for this example, Auto).
A flashing
A............ Auto Comparator (default setting)
E ............ Manual Comparator
49
HIOKI BT3562A981-09
4.1 Comparator Function
8
9
Press so that the indicated position blinks, and select resistance.
r flashing
r............. Resistance
u ............ Voltage
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, REF/%).
4
Chapter 4 Applied Measurement
REF & % flashing
HIGH, LOW..... Compare by upper and lower thresholds (default setting)
REF, % ............ Compare by reference value and tolerance
10
Or
numeric
keypads
Switch to the Ref/% threshold setting display, and specify the thresholds.
For this example,
Reference value: 1.5
Tolerance: 5%
Using the RANGE keys:Using the numeric keypads:
Select a digit to change by mov-
ing the blinking location, then select the new numerical value.
Select a digit
Select numerical value
Press the numeric keys corresponding to the digits to
be entered.
To enter the current measurement as the setting value: AUTO key
(Press on a screen other than the upper/lower threshold setting display.)
To enter the result of statistical calculation as the setting value: STAT key
(Press on a screen other than the upper/lower threshold setting display.)
See " Upper and Lower Thresholds Setting (by Reference Value and Tolerance)" (p.54).
50
HIOKI BT3562A981-09
4.1 Comparator Function
11
12
Press so that the indicated position blinks, and select voltage.
u flashing
r .............Resistance
u ............Voltage
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, REF/%).
REF & % flashing
HIGH, LOW ..... Compare by upper and lower thresholds (default setting)
REF, %............. Compare by reference value and tolerance
13
14
Or
numeric
keypads
Switch to the Ref/% threshold setting display, and specify the thresholds.
For this example,
Reference value: 4.2 V
Tolerance: 0.5%
Applies setting and returns to the Measurement display.
The comparator function is enabled.
COMP lit
To cancel the settings: SHIFT key
51
HIOKI BT3562A981-09
4.1 Comparator Function
15
Connect a test object and judge the measured value.
Resistance measurements are displayed as
their relative percentage offset from the reference value (%)
Judgment Result
Relative
percentage
Voltage measurements are displayed as their relative percentage offset
from the reference value (%)
In the V mode, you can verify comparator settings by pressing the VIEW key.
See " Switching Between Measurement Value and Comparator Setting Displays" (p.58).
Upper Threshold Value of setting range < Measured value
Lower Threshold Value of setting range Measured value Upper
Threshold Value of setting range
Measured value < Lower Threshold Value of setting range
Measured resistance - Reference value
=
Reference value
× 100
4
Chapter 4 Applied Measurement
The instrument can also base judgments on the absolute value of voltage measured values (to prevent Lo judgments when the positive and negative terminals
are connected backwards).
See "Configuring the Absolute Value Judgment Function (Voltage)" (p.55)
52
HIOKI BT3562A981-09
4.1 Comparator Function
Comparator Judgment Beeper Setting
Four beeper settings are available to audibly indicate comparator judgment results.
1
2
(The SHIFT indicator lights up.)
The Comparator setting display appears.
Set the comparator judgment beeper.
(Main Display)
oFF........ no beeps sound
HL.......... beeps repeatedly (when measurements are Hi or Lo)
in ...........beeps continuously (when measurements are IN)
btH1.......beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2.......beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
• The beeper does not sound when the comparator judgment beeper setting
is disabled (oFF).
• The beeper does not sound when there is no judgment result.
See " Comparator Judgment Results" (p.57).
Comparator Execution Mode Setting
Comparator judgment execution is selected by setting the auto or manual/external comparator mode. Comparator judgment can be enabled and disabled by EXT I/O signals. Refer to Input Signals (Page 81).
(The SHIFT indicator lights up.)
1
The Comparator setting display appears.
2
Press so that the indicated position blinks, and set the comparator execution
mode.
(Main Display)
A............ Auto comparator (comparator results are always output [default setting])
E ............Manual comparator (comparator results are output only when the MANU EXT I/
O input is enabled [ON])
The auto setting is appropriate for normal use. Use the manual/external setting
when you need to control comparator judgment timing.
Comparator Threshold Method Selection
HIOKI BT3562A981-09
Two methods are available for setting comparator thresholds.
53
4.1 Comparator Function
1
2
(The SHIFT indicator lights up.)
The Comparator setting display appears.
Press so that the indicated position blinks, and set the comparator threshold
method.
4
Chapter 4 Applied Measurement
HIGH, LOW..... Compare against specified upper and lower thresholds (default setting
method)
REF, % ............ Compare against upper and lower thresholds internally calculated from a
specified reference value and tolerance
About comparisons based on a reference value and tolerance
When the reference value and tolerance method is selected, thresholds are calculated as
follows:
Measured values are displayed as a percentage relative to the reference value, calculated as follows:
Relative value = (measured value - reference value) / reference value× 100 [%]
54
HIOKI BT3562A981-09
4.1 Comparator Function
Upper and Lower Thresholds Setting
(by Reference Value and Tolerance)
1
2
3
(The SHIFT indicator lights up.)
The Comparator setting display appears.
Press so that the indicated position blinks, and select resistance or voltage.
r .............Resistance
u ............Voltage
Select the threshold setting display, and enter upper and lower threshold values.
Or
numeric
keypads
For example,
Upper Threshold: 150 m
Lower Threshold: 100 m
Using the RANGE keys:Using the numerc keypads
Select a digit to change by moving the blinking location, then select the new numerical value.
Select a digit
Select numerical value
Press the numeric keys corresponding to the digits to
be entered.
4.1 Comparator Function
HIOKI BT3562A981-09
To enter the current measurement as the setting value: AUTO key
Press on a screen other than the upper/lower threshold (reference value/tolerance) setting display. This key is used as a numeric key on the upper/lower
threshold (reference value/tolerance) setting display.
The current measurement value is set as the upper or lower threshold (during
upper/lower threshold setting), or as the reference value (during reference
value and tolerance setting). If the measured value is faulty or
ignored (not entered).
To enter a statistical calculation result as the setting value: STAT key
Press on a screen other than the upper/lower threshold (reference value/tolerance) setting display. This key is used as a numeric key on the upper/lower
threshold (reference value/tolerance) setting display.
The result of statistical calculation is set as follows:
± OF, it is
55
During upper/lower
threshold setting
During reference value
and tolerance setting
Where “
No setting occurs if statistical calculation is disabled and no statistical calcula-
tion result exists.
See "4.5 Statistical Calculation Functions" (p.64).
Setting thresholds from the AUTO and STAT keys is possible only when the
selected (blinking) character is non-numeric.
Threshold and reference values can be set from 0 to 99999 (or 999999 for voltage), and tolerance can be set from 0.000 to 99.999%. Negative values are
not settable. Entries using statistical calculation results that exceed the valid
range are restricted to the range limit.
” represents population standard deviation (
Upper threshold = average value + 3
Lower threshold = average value - 3
Reference value = average value
Tolerance = 3
/ average value ×100%
Configuring the Absolute Value Judgment Function (Voltage)
This section describes how to configure functionality for acquiring the absolute
value of the voltage measured value when judging comparators, allowing a judgment to be made based on the absolute value of the voltage even if polarity is
reversed when the probes are connected to the battery.
4
Chapter 4 Applied Measurement
).
n
Ordinarily, connecting the probes with the polarity reversed results in a negative
voltage measured value, yielding a Lo comparator judgment result. To generate
an IN judgment whenever the reading falls within the specified range, even if the
probes have been connected backwards (resulting in a negative voltage measured value), set the absolute value judgment function to "On."
This function is configured on the menu screen.
(The SHIFT indicator lights up.)
1
The menu screen is displayed.
56
HIOKI BT3562A981-09
4.1 Comparator Function
2
Display the absolute value judgment function configuration screen.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16)
3
Set the absolute value judgment function to either "On" or "Off."
on
.........................
oFF................... Absolute value judgment function off.
Accept the setting and return to the measurement screen.
Absolute value judgment function on.
4
For example, the following judgment results would be obtained when connecting
the probes backwards to a 3.7 V battery (resulting in a displayed voltage measured value of -3.7 V) with an upper threshold of 3.9 V and a lower threshold of
3.6 V:
Absolute value judgment function off: Lo
Absolute value judgment function on: IN
Enabling and Disabling the Comparator Function
Enables the comparator
COMP lit
When the comparator is enabled, the following key operations are disabled
to avoid inadvertent operations.
V//V key (Measurement mode setting)
•
• SHIFT
• SHIFTCOMP key (Comparator setting)
• AUTO key (Auto-ranging setting)
• SMPL key (Sampling rate setting)
• SHIFT
• SHIFTTRIG key (Trigger source setting)
• SHIFT
• SHIFT
• Range keys
When the comparator is enabled, auto-ranging is automatically disabled.
Resistance and voltage measurements are judged independently. Both judgment results are indicated on the display.
Judgment OperationThe comparator compares measured values with the preset threshold values,
and judges whether the measurement is within the thresholds. Resistance and
voltage measurements are judged independently.
The absolute value of the measurement is compared to the upper and lower
thresholds.
When the absolute value judgment function is on, the absolute value of the measured value is compared to the upper and lower thresholds.
PASS/FAIL
Judgment Output
Upper Threshold Value < Measured Value
Lower Threshold Value
Measured Value < Lower Threshold Value
Resistance
V : Measurement
Measurement fault values are judged as follows:
DisplayJudgment
- - - - -
OF
-OF
Judgment results (Hi, IN or Lo for both resistance and voltage) are output to EXT
I/O connectors.
Additionally, the instrument can generate PASS/FAIL judgment output to facilitate easy judgments. In this configuration, it outputs a PASS judgment when the
resistance and voltage are both IN and otherwise a FAIL judgment.
See " Output Signals" (p.82).
No judgment
Hi (exceeds the upper threshold)
Lo (less than the lower threshold)
Measured Value Upper Threshold Value
4
Chapter 4 Applied Measurement
With the relative value comparison method (thresholds defined by a reference
value and tolerance), the upper and lower thresholds are calculated internally
for comparison with measurements. Therefore, even if a relative display value
is equal to a judgment threshold (tolerance limit), it may be judged Hi or Lo.
58
HIOKI BT3562A981-09
4.1 Comparator Function
Switching Between Measurement Value and Comparator Setting Displays
In V mode, both resistance and voltage measurement values are displayed.
Although comparator setting values are not normally displayed when the comparator is enabled, they can be displayed for confirmation by the display switching function.
Example:
Resistance: Upper threshold value 150.00 m
Voltage: Upper threshold value 15.2000 V, Lower threshold value 15.0000 V
Press this key to switch the display between measurement values
and comparator setting values.
Resistance and voltage measurement display
(Shows resistance and voltage measurement values simultaneously)
, Lower threshold value 100.00 m
Resistance measurement and comparator
display (Shows resistance measurement and
resistance comparator setting values)
Voltage measurement and comparator display (Shows voltage measurement and voltage comparator setting values)
Measurement display switching is available only with the comparator
enabled, and in the V mode.
Use it to confirm comparator setting values.
4.2Trigger Function
HIOKI BT3562A981-09
Trigger Source Settings
Two trigger sources are available: internal and external.
Internal TriggerTrigger signals are automatically generated internally.
External TriggerTrigger signals are provided externally or manually.
(The SHIFT indicator lights up.)
59
4.2 Trigger Function
(free-run)
Measurement with
External Triggering
Switches the selected trigger source.
EXT TRIG lit
EXT.TRIG lit................. External triggering is selected.
EXT.TRIG not lit........... Internal triggering is selected.
An external trigger can be applied in three ways.
• Applying a trigger manually by operating key
Pressing the TRIG key causes one measurement.
• Applying a trigger at the EXT I/O connector.
Shorting the TRIG
rear panel causes one measurement.
See " Input Signals" (p.81).
• Applying a trigger through RS-232C or GP-IB interface
Sending the
measurement.
• When Internal triggering is enabled, external input at the EXT I/O TRIG ter-
minal and the
• The normal state of operation with the front panel controls is continuous
measurement. Setting the trigger source to Internal enables the free-run
condition in which triggering occurs continuously. When the trigger source is
set to External, a measurement occurs each time an external trigger is
applied. Continuous measurement can be disabled via RS-232C, LAN, or
GP-IB interface signals, in which case triggering occurs only when signaled
by the external host (PC or PLC).
See " Triggering System Description" (p.153).
terminal to the ISO_COM of the EXT I/O connector on the
TRG command via the RS-232C or GP-IB interface causes one
TRG command are ignored.
4
Chapter 4 Applied Measurement
60
HIOKI BT3562A981-09
4.2 Trigger Function
Trigger Delay Settings
Specify the delay from the moment a trigger is applied to the start of measurement. By using this function,
even when a trigger is applied immediately after connecting a test object, the start of measurement can be
delayed to allow sufficient time for the measurement value to stabilize. Trigger delay can be set with 1 ms resolution from 0.000 to 9.999 seconds.
(The SHIFT indicator lights up.)
1
The Trigger Delay setting display appears.
(Main Display)
(Sub Display)
The current setting blinks.
2
Select ON.
3
The numerals indicating the trigger delay blink.
4
Set the trigger delay.
Or numeric keypads
5
Applies setting and returns to the Measurement display.
To cancel the settings:
Disabling the Trigger Delay Function
(The SHIFT indicator lights up.)
1
(Sub Display)
(Sub Display)
SHIFT key
The Trigger Delay setting display appears.
Select OFF.
2
(Sub Display)
The Trigger Delay is disabled.
3
61
HIOKI BT3562A981-09
4.3 Measurement Current Pulse Output Function
4.3Measurement Current Pulse Output Function
When using multiple instruments to take measurements at the same time, you
may experience interference as a result of electromagnetic induction caused by
the devices' measurement currents, causing measured values to gradually oscillate between increasingly large values. To prevent this phenomenon, the measurement current pulse output function can be used to apply the measurement
current only while measurement is being performed. (This functionality is only
available during external trigger measurement or while continuous measurement
is off.)
When this function is turned on, the measurement current is only applied at the
time of measurement. By staggering the timing of trigger measurement, measurement current interference can be eliminated.
4
1
2
3
4
(The SHIFT indicator lights up.)
The menu screen is displayed.
Display the measurement current pulse function configuration screen.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16)
Set the constant current pulse function to either "On" or "Off."
on.................... Measurement current pulse function on.
oFF.................. Measurement current pulse function off.
Accept the setting and return to the measurement screen.
Chapter 4 Applied Measurement
62
HIOKI BT3562A981-09
4.3 Measurement Current Pulse Output Function
Measurement current application timing when measurement current pulse mode is on
Measurement processing triggered by external TRIG signal or "READ?" command
Measurement
Measurement
current
Applied only during measurement (including trigger delay time).
Measurement and current application when using external trigger measurement
or when continuous measurement is off (:INIT:CONT OFF)
Measurement
Measurement
current
Measurement current is applied continuously.
Measurement current application when using normal internal trigger measurement
(The measurement current is applied continuously when measurement current pulse mode is off.)
• When using the instrument with the measurement current pulse output func-
tion enabled, always enable the trigger delay function and set the delay time
to 10 ms or more.
See "Trigger Delay Settings" (p.60)
• When the measurement current pulse output function is on, the below is
added to the sampling time.
When using the V mode: +1 ms
When using the mode: +4 ms
63
HIOKI BT3562A981-09
4.4 Averaging Function
4.4Averaging Function
The Averaging Function averages measurement values for output. This function can minimize instability of
displayed values. The number of samples to average can be set from 2 to 16.
1
2
3
4
numeric keypads
Or
(The SHIFT indicator lights up.)
The Averaging Function setting display appears.
(Main Display)
(Sub Display)
The current setting blinks.
Select ON.
(Sub Display)
The number of samples to average setting blinks.
Select the number of samples to average.
4
Chapter 4 Applied Measurement
5
The Average Measurement display appears. (AVG lit)
To cancel the settings:
Disabling the Averaging Function
1
(The SHIFT indicator lights up.)
The Averaging Function setting display appears.
Select OFF.
2
3
The Averaging Function is disabled. (AVG not lit)
When the internal trigger is used for continuous measurement (free-run), the
display shows the moving average. Otherwise, the display shows the integrating average.
See "4.2 Trigger Function" (p.59).
SHIFT key
(Sub Display)
64
n
x
x
n
xnx
22
122
n
xnx
s
16
n
LoHi
Cp
1
6
2
n
xLoHiLoHi
CpK
HIOKI BT3562A981-09
4.5 Statistical Calculation Functions
4.5Statistical Calculation Functions
The mean, maximum, minimum, standard deviation of population, standard deviation of sample and process capability indices are calculated and displayed for
up to 30000 measurement values.
The calculation formulas are as follows:
Mean
Standard deviation of population
(=
)
n
Standard deviation of sample(=
Process capability index (dispersion)
Process capability index (bias)
• In these formulas, n represents the number of valid data samples.
• Hi and Lo are the upper and lower thresholds of the comparator.
• The process capability indices represent the quality achievement capability
created by a process, which is the breadth of the dispersion and bias of the
process' quality. Generally, depending on the values of Cp and CpK, process
capability is evaluated as follows:
Cp, CpK>1.33........................ Process capability is ideal
1.33 Cp, CpK>1.00............. Process capability is adequate
1.00 Cp, CpK...................... Process capability is inadequate
• When only one valid data sample exists, standard deviation of sample and
process capability indices are not displayed.
• When
• The upper limit of Cp and CpK is 99.99. Values of Cp and CpK>99.99 are
displayed as 99.99.
• Negative values of CpK are handled as CpK=0.
• When comparator, range or auto-ranging settings are changed while statistical data is displayed, the display of Cp and CpK values changes to “- - . - -”.
• When normal measurement values and relative display values (%) are
mixed, correct calculation results cannot be obtained.
is 0, Cp and CpK are 99.99.
n-1
)
n-1
4.5 Statistical Calculation Functions
HIOKI BT3562A981-09
Enabling/Disabling the Statistical Calculation Function
65
1
2
(press three times)
3
The Statistical Calculation display appears.
(Main Display)
(Sub Display)
The function enable/disable display appears.
(Sub Display)
Enable or disable the Calculation Function on the Sub Display.
on.......... enables the calculation function on.
oFF ....... disables the calculation function off.
Applies setting and returns to the Measurement display.
To cancel the settings:
• Statistical Calculation function setting (ON, OFF) is not available when the
Comparator is enabled.
• If Statistical Calculation is turned off and then back on without first clearing
calculation results, it resumes calculating from the point when it was turned
off.
• The Statistical Calculation function slows measurements when it is ON.
SHIFT key
4
Chapter 4 Applied Measurement
Clearing Statistical Calculation Results
The Statistical Calculation display appears.
1
The Clearing screen will appear.
2
(press once)
Clears statistical calculation results.
3
(Main Display)
(Sub Display)
(Sub Display)
66
HIOKI BT3562A981-09
4.5 Statistical Calculation Functions
Automatic Clearing of Statistical Calculation Results after Printing
The instrument can be set to automatically clear statistical calculation results after results are output to the
printer.
1
2
3
4
(Press twice)
The Statistical Calculation display appears.
(Main Display)
(Sub Display)
Bring up Auto Clearing After Printing in the Setup screen.
(Sub Display)
Turn Automatic Clearing After Printing on or off.
on.......... Automatically clears statistical calculation results after they are
output to the printer.
oFF........ Does not clear the results themselves.
Applies setting and returns to the Measurement display.
To cancel the settings:
SHIFT key
Importing Data
Pressing the TRIG key while Statistical Calculation is ON executes one of the following operations:
• External Trigger: Takes one measurement and performs statistical calcu-
lation on the result
• Internal Trigger: Performs statistical calculation on the value displayed
immediately after pressing
• TRG command executes the same operation.
• Shorting the TRIG
cutes the same operation.
terminal to the ISO_COM of the EXT I/O connector exe-
Confirming Statistical Calculation Results
1
The Statistical Calculation display appears.
2
The indication on the display changes as follows with each
key-press.
Example: when the V mode is selected
HIOKI BT3562A981-09
(not displayed in V mode)
Total data count of resistance measurement
Mean of resistance measurementMaximum of resistance measurement
67
4.5 Statistical Calculation Functions
Valid data
Mean
Minimum of resistance measurementStandard deviation of population
of resistance measurement
Minimum
Data Sample
Process capability indices of resistance
measurement
Cp
CpK
(not displayed in mode)
Total data count of voltage measurement
Minimum of voltage
measurement
Process capability indices
of voltage measurement
Standard deviation of
population of voltage
Mean of voltage
measurement
Maximum
Data Sample
No.
Standard deviation of sample of resistance measurement
4
Chapter 4 Applied Measurement
Maximum of voltage
measurement
Standard deviation of
sample of voltage
measurement
ON/OFF settingAuto Clearing After Printing setupClear setup
• When a valid data count (measurement fault other than ± OF) is zero, no
calculation result is displayed.
• When only one valid data sample exists, standard deviation of sample and
process capability indices cannot be displayed.
• When comparator, range or auto-ranging settings are changed while statistical data is displayed, the display of Cp and CpK values changes to “- - . - -”.
Sending Statistical Calculation Results to the Printer
With the statistical calculation results displayed, press the PRINT key.
The statistical calculation results are output to the optional printer.
See "Chapter 6 Printing" (p.91).
68
HIOKI BT3562A981-09
4.6 Memory Function
4.6Memory Function
The Memory function is only available via communication commands.
When the Memory function is enabled, measurement values are stored in the
instrument’s internal memory according to trigger input sequence (up to 400 values). Stored data can be downloaded later upon command.
When measuring using a scanner to switch multiple test objects, switching time
can be quite long if measurement values are downloaded to the PC after each
measurement. Test cycle time can be minimized by using this function to store
measurement values internally until all channel measurements are finished, at
which time the stored values are downloaded together during the next idle
period.
1
2
3
4
Select the RS-232C, LAN, or GP-IB interface.
See " Selecting the Interface" (p.105).
Send the command to enable the Memory function.
:MEMory:STATe ON
The MEM indicator lights.
MEM lit
Measurement values are stored.
When a trigger is applied by the TRIG key, TRIG EXT I/O input signal or
TRG command, the MEM indicator blinks once and the measured value is
stored.
MEM lit
If an external trigger source is selected, one measurement is stored after
each trigger event. In the internal triggering case, the first measurement value after triggering is stored. Apply a trigger as many times as is necessary.
69
HIOKI BT3562A981-09
4.6 Memory Function
5
Send the command to download the data from memory.
:MEMory:DATA?
The stored measurement values are returned in response.
The “END” character is sent as the last line of the data.
To download stored data one measurement at a time, send this command:
:MEMory:DATA? STEP
The instrument sends one stored data object and enters the wait state.
When the instrument receives an “N” from the PC or other device, the next
stored data object is sent.
Repeat until the last data object is downloaded.
When all stored data has been downloaded, the instrument sends an “END”
character.
:MEM:DATA? STEP
1, 290.60E-3, 1.3924E+0
N
(sent from PC)
2, 290.54E-3, 1.3924E+0
N
(sent from PC)
3, 290.50E-3, 1.3923E+0
N
(sent from PC)
4, 290.43E-3, 1.3923E+0
N
(sent from PC)
5, 290.34E-3, 1.3924E+0
N
(sent from PC)
END
4
Chapter 4 Applied Measurement
6
To clear the instrument's memory, send it the following command.
:MEMory:CLEAr
Unless the memory is cleared, measurement data continues to be stored
upon each trigger event.
• The instrument's memory storage capacity is 400 measurements. Be aware
that attempting to store more data (by applying a trigger) results in nothing
further being stored.
• Refer to Chapter 8 RS-232C/GP-IB/LAN Interfaces (Page 99),
for details about the communication methods and sending and receiving
commands.
• When the Memory function is enabled, auto-ranging is not available.
• Memory contents are cleared when performing the following operations:
When enabling the Memory function (off to on)
When changing the measurement range
When changing comparator settings
When sending the
When Reset is executed from the menu display
When sending
When sending :SYSTem:RESet
When turning power on
:Memory:Clear command
RST
70
HIOKI BT3562A981-09
4.7 Key-Lock Function
Disabling the Memory Function
1
2
Send the command to enable the Memory function Off.
:MEMory:STATe OFF
The Memory function is disabled. (MEM not lit)
4.7Key-Lock Function
Executing Key-Lock disables the operating keys on the front of the instrument.
This function can be useful for protecting settings.
(The SHIFT indicator lights up.)
Enable the Key-Lock function.
LOCK lit
Disabling Key-Lock
• Even if the power supply is interrupted, the Key-Lock function is not can-
celed.
• The TRIG key remains operational.
(The SHIFT indicator lights up.)
Disable the Key-Lock function. (
When communicating by remote control, the remote control status is canceled.
LOCK not lit)
4.8Panel Save Function
HIOKI BT3562A981-09
The current measurement setting state is stored (saved) in non-volatile memory.
Up to 126 sets of measurement states can be saved.
The measurement settings (state) at the time this function is executed are
saved.
Saved measurement states can be reloaded using the Panel Load function,
described later.
71
4.8 Panel Save Function
1
2
numeric keypads
3
Or
(The SHIFT indicator lights up.)
The Panel Saving display appears.
(Main Display)
(Sub Display)
The panel number blinks.
Select the panel number to save.
(Sub Display)
(To save measurement settings as Panel No. 3)
When selecting a saved panel, “USEd” is displayed.
Saves the measurement setting state and returns to the Measurement display.
To cancel the settings:
SHIFT key
4
Chapter 4 Applied Measurement
Saved Items
• If you select a Panel number that was previously saved and press the
ENTER key, the contents are overwritten.
• The Key-Lock state can be saved only by the
command.
(The absolute value judgment function setting is not saved.)
:SYSTem:SAVE remote
72
HIOKI BT3562A981-09
4.9 Panel Load Function
4.9Panel Load Function
Loads the measurement settings saved by the Panel Save function from internal
non-volatile memory.
1
2
numeric keypads
3
Or
The Panel Loading display appears.
(Main Display)
(Sub Display)
The panel number blinks.
Select the panel number to load.
(Sub Display)
(To load measurement settings from Panel No.3)
Loads the measurement setting state and returns to the Measurement
display.
To cancel the settings:
• If an unsaved Panel No. is selected, a warning beep sounds when you
press ENTER key.
• When selecting a Panel No. with the up/down RANGE keys, only the numbers of previously saved panels appear.
• Loading can also be executed using the TRIG
LOAD6
See " Input Signals" (p.81).
pins of the EXT I/O interface.
SHIFT key
signal and the LOAD0 to
4.10 Self-Calibration
HIOKI BT3562A981-09
The self-calibration function adjusts offset voltage and gain drift of the instrument’s internal circuitry to improve measurement precision.
The instrument’s measurement accuracy specifications depend on self-calibration, so it must be executed frequently. In particular, always execute self-calibration after warm-up and when the ambient temperature changes by more than
2
°C. However, regardless of this setting, self-calibration is executed during
every measurement when SLOW sampling is used.
Self-calibration can be executed by the following two methods:
73
4.10 Self-Calibration
1
2
AutoExecutes self-calibration automatically once every 30 min-
utes.
ManualSelf-calibration can be executed manually by applying a CAL
input signal (shorting the CAL terminal to the ISO_COM of the
EXT I/O connector).
It can also be executed with the SYSTem:CALibration command.
(The SHIFT indicator lights up.)
The Menu display appears.
The Self-Calibration setting display appears.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
(p.147)
(Main Display)
(Sub Display)
The current setting blinks.
4
Chapter 4 Applied Measurement
3
4
Select Auto or Manual on the Sub Display.
AUto ..... Auto self-calibration
in........... Manual self-calibration
Applies setting and returns to the Measurement display.
Self-calibration requires about 55 ms, during which measurement processing
is temporarily suspended.
74
HIOKI BT3562A981-09
4.11 Measurement Value Output Function
4.11Measurement Value Output Function
This function causes output of measured values via the RS-232C or LAN interface in the same sequence as trigger input.
This function is useful when measuring using internal (free-run) triggering, and
for obtaining measured values on a PC when using a footswitch for triggering.
1
2
3
4
(The SHIFT indicator lights up.)
The Menu display appears.
The Measurement Value Output function setting display appears.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
(Main Display)
(Sub Display)
The current setting blinks.
Turn Measurement Value Output Function on or off.
on.......... enables the measurement value output function on.
oFF........ disables the measurement value output function off.
Applies setting and returns to the Measurement display.
5
The measured value is output from the RS-232C or LAN interface
when you press the
I/O TRIG
Set the PC to the receiving state beforehand. When a measurement
value is received, the PC should perform appropriate processing
such as recording or displaying, then re-enable the receiving state.
• When external triggering is enabled, a measurement is performed and the
value is sent after each trigger event. When internal triggering is enabled,
the first value measured after triggering is sent.
• The measurement output function is not applicable to the GP-IB interface or
printer.
terminal.
TRIG key or when a signal is applied to the EXT
4.12 Key Beeper Setting
HIOKI BT3562A981-09
Select whether a beep sounds when an operating key on the front of the instrument is pressed.
75
4.12 Key Beeper Setting
1
2
3
4
(The SHIFT indicator lights up.)
The Menu display appears.
The Key Beeper setting display appears.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
(Main Display)
(Sub Display)
The current setting blinks.
Select the key beeper state on the Sub Display.
on.......... Key beeper enabled.
oFF ....... Key beeper disabled.
Applies setting and returns to the Measurement display.
4
Chapter 4 Applied Measurement
76
HIOKI BT3562A981-09
4.13 Reset Function
4.13 Reset Function
The reset function can be used to re-initialize current measurement settings
(excluding saved panel data) to their factory defaults, or to re-initialize all measurement settings including saved panel data to factory defaults.
1
2
3
4
(The SHIFT indicator lights up.)
The Menu display appears.
The Reset display appears.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
(Main Display)
(Sub Display)
The current setting blinks.
Select the Reset method on the Sub Display.
SEt ........ Reset (initializes measurement settings other than those stored
with Panel Save)
SYS....... System Reset (initialize all measurement settings)
ENTER blinks.
(Sub Display)
5
When SYS (system reset) is selected
Executes the Reset.
To cancel the settings:
System Reset also initializes Panel Save data.
Executing either of them cannot initialize the LAN interface settings. To initialize the LAN interface settings, execute the reset from the SEt item of the LAN
settings. (p.105)
SHIFT key
Initial Factory Default Settings
HIOKI BT3562A981-09
DescriptionDefault
Measurement Mode
77
4.13 Reset Function
V
Resistance Measurement Range
Voltage Measurement Range6 V
Auto RangeON
Zero-AdjustOFF
DelayOFF
Delay Time0.000s
Sampling RateSLOW
Averaging FunctionOFF
Average Times2
Self-CalibrationAUTO
Continuous MeasurementON
Trigger SourceInternal trigger
Line FrequencyAUTO
Key Beeper SettingON
Key-Lock FunctionOFF
ComparatorOFF
Comparator Threshold Method
(resistance and voltage)
Comparator Upper Threshold
(resistance and voltage)
Comparator Lower Threshold
(resistance and voltage)
Comparator Judgment BeeperOFF
Comparator Execution ModeAUTO
Statistical Calculation FunctionsOFF
Automatic Clearing of Statistical
Calculation Results
InterfaceRS-232C
Baud Rate9600 bps
GP-IB Address1
GP-IB DelimiterLF
Print Interval0 (The interval print disabled)
Error OutputASync
Measurement Value Output FunctionOFF
EOM Output
3 m (BT3562, BT3563)
(BT3561A)
30 m
Hi, Lo
0
0
OFF
HOLD
4
Chapter 4 Applied Measurement
EOM Pulse Width1 ms
Measurement current pulse functionOFF
Comparator absolute value judgment
function
OFF
78
HIOKI BT3562A981-09
4.13 Reset Function
External Control
HIOKI BT3562A981-09
79
5.1 Overview
(EXT I/O)
5.1Overview
External Control
Input Functions
External Output
Terminal Func-
tions
Chapter 5
• External trigger input (TRIG)
• Select Panel No. to load (LOAD0 to LOAD6)
• Zero-adjust signal input (0ADJ
• Print Signal input (PRINT
• Self-calibration signal input (CAL)
• Manual comparator judgment input (MANU
• End-of-Conversion signal output (EOM)
• Reference signal output (INDEX
• Measurement Fault signal output (ERR)
• Comparator decision signal output (R-Hi
)
FAIL
)
)
)
, R-IN, R-Lo, V-Hi, V-IN, V-Lo, PASS,
)
5
Chapter 5 External Control (EXT I/O)
To avoid electric shock or damage to the equipment, always observe the
following precautions when connecting to the EXT I/O terminals.
• Always turn off the power to the instrument and to any devices to be
connected before making connections.
• During operation, a wire becoming dislocated and contacting another
conductive object can be serious hazard. Make sure that connections
are secure and use screws to secure the external connectors.
• Ensure that devices and systems to be connected to the EXT I/O ter-
minals are properly isolated.
To avoid damage to the instrument, observe the following cautions:
• Do not apply voltage or current to the EXT I/O terminals that exceeds their
ratings.
• When driving relays, be sure to install diodes to absorb counter-electromotive force.
Be careful not to short-circuit ISO_5V to ISO_COM.
•
See: "5.2 Signal Descriptions" (p.80)
80
HIOKI BT3562A981-09
5.2 Signal Descriptions
5.2Signal Descriptions
Pinout
(Reserved)
(Reserved)
PASS
(Reserved)
(Reserved)
(Reserved)
V_IN
R_LO
R_HI
ISO_COM
ERR
ISO_5V
MANU
LOAD3
LOAD5
(Reserved)
LOAD1
12345678910111213141516171819
(Reserved)
TRIG
Connector: (Instrument Side)
202122232425262728293031323334353637
37-pin D-sub female with #4-40 screws
FAIL
LOAD0
LOAD2
LOAD4
LOAD6
PRINT
ISO_COM
EOM
INDEX
R_IN
V_HI
V_LO
(Reserved)
(Reserved)
(Reserved)
(Reserved)
Mating Connectors:
DC-37P-ULR (solder type) /
DCSP-JB37PR (pressure weld type)
Japan Aviation Electronics Industry Ltd.
EXT I/O Connector (Instrument Side)
Pos: positive, Neg: negative,: not applicable
Pin
Signal nameI/OFunctionLogicPinSignal nameI/OFunctionLogic
1TRIG
(Reserved)
2
(Reserved)
3
LOAD1
4
LOAD3
5
LOAD5
6
7
MANU
8
ISO_5V
9
ISO_COM
10
ERR
11
R_HI
12
R_LO
13
V_IN
(Reserved)
14
(Reserved)
15
(Reserved)
16
(Reserved)
17
PASS
18
(Reserved)
19
External trigger
IN
Load no. bit 1Neg Level
IN
Load no. bit 3Neg Level
IN
Load no. bit 5Neg Level
IN
Comparator manu-
IN
al control
Isolated 5 V power
output
Isolated common
signal ground
Measurement fault Neg Level
OUT
HI resistance
OUT
judgment result
LO resistance judg-
OUT
ment result
IN voltage judg-
OUT
ment result
OUT
OUT
OUT
OUT
PASS judgment re-
OUT
sult
OUT
Pos/
Edge
Neg
Neg Level
Neg Level
Neg Level
Neg Level
Neg Level
20 0ADJ
CAL
21
LOAD0
22
LOAD2
23
LOAD4
24
LOAD6
25
26
PRINT
27
ISO_COM
28
EOM
29
INDEX
30
R_IN
31
V_HI
32
V_LO
(Reserved)
33
(Reserved)
34
(Reserved)
35
(Reserved)
36
FAIL
37
Zero adjustmentsNeg
IN
Self-calibration
IN
execution
Load no. bit 0Neg Level
IN
Load no. bit 2Neg Level
IN
Load no. bit 4Neg Level
IN
Load no. bit 6Neg Level
IN
Print measured val-
IN
ue
Isolated common
-
signal ground
End of measurement Neg Edge
OUT
Analog measure-
OUT
ment finished
IN resistance
OUT
judgment result
Hi voltage judgment
OUT
result
Lo voltage judgment
OUT
result
Judgment result
OUT
FAIL
Reserved pins are not connected inside the instrument.
Do not connect to reserved pins.
0ADJ
CAL
Neg
Neg Edge
Neg Edge
Neg Level
Neg Level
Neg Level
Neg
Edge
Edge
Level
The connector frame is connected to (continuous with) both the instrument's
case (the metal cabinet surrounding the instrument) and the power inlet's protective ground pin. Note that the frame is not isolated from the ground.
Input Signals
HIOKI BT3562A981-09
81
5.2 Signal Descriptions
LOAD0 to
LOAD6
Select a Panel No. to load and apply a TRIG signal to load the selected Panel
No. and measure. LOAD0 is the LSB, and LOAD6 is the MSB.
When a TRIG signal is applied, if LOAD0 through LOAD6 are unchanged from
the previous trigger event, panel settings are not loaded. In this case, using
external triggering, one measurement is taken as usual when the TRIG
applied.
Panel
No.
1
2
3
4
5
6
7
8
...
122
123
124
125
126
0: HIGH: Open or from 5 V to 24 V
1: LOW: 0 V to 0.9 V
When a TRIG signal is applied with LOAD0 to LOAD6 set to all 1's or all 0's,
no Panel Load occurs.
• At least 70 ms is required for the settings to change after executing a Panel
Load (the actual time depends on the particular
rate
).
• When set to external trigger mode, one measurement is taken upon load completion.
function, range and sampling
5
Chapter 5 External Control (EXT I/O)
TRIGWhen the external trigger, one measurement is taken each time the TRIG signal
transitions from High to Low.
This trigger signal is ignored when internal triggering is enabled.
Trigger functions are also available for statistical calculation, recording to memory and output of measured values (valid also with internal triggering).
CALWhen manual self-calibration is selected with EX.FAST, FAST or MEDIUM sam-
pling rate, self-calibration begins when the CAL signal transitions from High to
Low.
Self-calibration takes about 55 ms.
When SLOW sampling is selected, the CAL
See "4.10 Self-Calibration" (p.73).
signal is ignored.
82
HIOKI BT3562A981-09
5.2 Signal Descriptions
0ADJZero adjustment executes once when the 0ADJ signal transitions from High to
Low.
PRINTThe current measurement value prints when the PRINT signal transitions from
High to Low.
MANUWhen the MANU comparator mode is selected, comparator judgment is enabled
while the MANU signal is Low.
See " Comparator Execution Mode Setting" (p.52).
Output Signals
ERRIndicates a measurement fault.
The Synchronous ERR output setting causes ERR output to be synchronous
with EOM
output to follow actual (asynchronous) contact of the probes with the test object.
See " ERR Output" (p.83).
output, while with the Asynchronous ERR output setting causes ERR
INDEXThe INDEX signal is output during the Trigger Wait, Delay, Self-Calibration and
Calculation states.
This signal is not output while measuring the resistance of test objects. This signal transitions from Hi (Off) to Lo (On) to indicate that the test object can be
removed.
EOMThis signal indicates the end of a measurement (End-Of-Conversion).
This signal indicates when comparator judgment results and ERR output (when
SYNC is enabled) are available.
R-Hi, R-IN, R-Lo
These are the results of comparator decision.
V-Hi, V-IN, V-Lo
PASSThis signal indicates when both resistance and voltage judgment results are IN
V mode).
(
It transitions to Low (ON) when both the resistance and voltage judgment results
are IN.
and V modes, this signal is the same as R-IN and V-IN outputs, respec-
In the
tively.
FAILThis signal transitions to Low (ON) when PASS is High (OFF).
• I/O signals should not be used while measurement settings have been
changed.
• The EOM
• If it is not necessary to change the measurement conditions, set LOAD0
through LOAD6 to either Hi or Lo.
• To avoid erroneous comparator judgments, both the PASS
should be checked.
and INDEX signals are initialized HIGH (OFF) at power on.
and FAIL signals
ERR Output
HIOKI BT3562A981-09
83
5.2 Signal Descriptions
The ERR output signal indicates the occurrence of measurement fault conditions
(such as open test leads, or a bad contact).
There are two ERR output methods.
Synchronized
with EOM Output (SYNC)
Asynchronous
with EOM Output (ASYNC)
Measurement faults detected while measuring (not while awaiting trigger or
during delay or calculation intervals), are indicated by ERR output synchronous
with EOM output (the end-of-measurement signal).
ERR Output Low (On): A measurement fault has prevented correct measurement
ERR Output High (Off): Correct measurement obtained (OF or -OF: Out-ofrange cases are included)
Measurement faults (test lead connection conditions) are output in real time. The
output is asynchronous with the TRIG signal and EOM output.
ERR Output Low (On): Measurement fault condition (open test leads, or a bad
contact)
ERR Output High (Off): Test lead connections are normal
5
Chapter 5 External Control (EXT I/O)
84
HIOKI BT3562A981-09
5.2 Signal Descriptions
Instrument Settings
Measurement Fault Output Signal (ERR) Setting
1
2
3
4
(The SHIFT indicator lights up.)
The Menu display appears.
Select the ERR Output Selection display.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
Select the type of signal to be output on the Sub Display.
SynC .......... Synchronous output (synchronized with EOM output)
ASynC ....... Asynchronous output (not synchronized with EOM output)
Applies settings and returns to the Measurement display.
Setting the EOM Signal
(Main Display)
(Sub Display)
The current setting blinks.
1
2
3
4
Or numeric keypads
(The SHIFT indicator lights up.)
The Menu display appears.
Select the EOM-signal setup display.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
(Main Display)
(Sub Display)
The current setting blinks.
Choose the output method for the EOM signal.
HoLd .......... Holds the EOM signal after measurement.
Go to Step 5.
PULSE ....... Outputs the specified pulse after measurement.
Go to the next step.
(When PULSE is selected)
The number representing the pulse width of the EOM signal will start blinking.
Set the pulse width in ms.
5
Applies settings and returns to the Measurement display.
5.3Timing Chart
HIOKI BT3562A981-09
External Trigger Timing Chart
85
5.3 Timing Chart
Contact State
*3 TRIG
INDEX
EOM
Output
Input
Output
Output
Output
*1 ERR
(Err Output ASYNC Setting)
Measurement start signal
Reference Signal
End-of-Measurement Signal
Comparator Result
*2 ERR
(Err Output SYNC Setting)
Contact
t1t1
ONOFF
ON
t2
Measuring
t3
t4t5t6
OFF
Open
ON
End of Conversion
ON
OFF
OFF
5
Chapter 5 External Control (EXT I/O)
*1: For details, see “"ERR Output" (p.83).”
*2: When ERR output is set to the SynChronous mode, measurement fault detection results can be obtained
when measurement is finished, as with comparator results.
*3: After connecting to the test object, wait for longer than the response time (approximately 10 ms) before
inputting the TRIG signal (It is necessary to wait out the response time for the measurement values to stabilize after connection. Response times depend on the test object).
Internal Trigger Timing Chart
* INDEX Output
Reference Signal
Output
* EOM
End-of-Measurement
Signal
Comparator Result
* When the EOM signal is set to PULSE, the signal will remain on only for the specified period upon com-
pletion of conversion.
ON
t3
MeasuringMeasuring
OFF
OFF
t4t5
ON
t6
86
HIOKI BT3562A981-09
5.3 Timing Chart
Description
t1
ERR Output response
*1
time
t2Measurement trigger
pulse width
t3Delay Timeper setting
t4Measurement time
t5Calculation time
t6EOM Output pulse widthWhen the external trigger is selected
*2
*3
Time
1.5 ms
0.5 ms or more.
See " Trigger Delay Settings" (p.60).
V mode
EX.FAST7.8 ms
FAST23.8 ms
MEDIUM83.8 ms
69.8 ms
SLOW258.8 ms
252.2 ms
0.3 ms
HOLD setting: Holds until the next trigger is detected
PULSE setting : Remains only for the specified pulse width
See " Instrument Settings" (p.84).
When the internal trigger is selected
HOLD setting: EX.FAST 1 ms, FAST 5 ms, MEDIUM 20 ms (50 Hz line
frequency setting)/ 16 ms (60 Hz line frequency setting), SLOW 50 ms
PULSE setting : Remains only for the specified pulse width
mode or V mode
EX.FAST3.4 ms
FAST11.4 ms
MEDIUM41.4 ms (50 Hz line fre-
quency setting)
34.4 ms (60 Hz line frequency setting)
SLOW156.4 ms (50 Hz line fre-
quency setting)
149.8 ms (60 Hz line frequency setting)
*1: For details, see “"ERR Output" (p.83).”
: About t4 measurement time
*2
When averaging is enabled, the running average is obtained with internal triggering, so measurement time
t4 does not change. The measurement time for external triggering is as follows:
With SLOW sampling
V(t4 - 57.8) × n + 57.8 ms (50 Hz)
(t4 - 51.2) × n + 51.2 ms (60 Hz)
or V(t4 - 56.4) × n + 56.4 ms (50 Hz)
(t4 - 49.8) × n + 49.8 ms (60 Hz)
With other than SLOW sampling
V(t4 - 2.8) × n + 2.8 ms
or V(t4 - 1.4) × n + 1.4 ms
(n represents the number of values averaged)
*3
: About t5 calculation time
In the following cases, add the indicated times to calculation time t5:
When the Statistical Calculation function is enabled 0.3 ms
When the reference value/tolerance method of
Maximum load voltage30 V
Maximum output current50 mA/ch
Residual voltage1 V (10 mA), 1.5 V (50 mA)
Output Voltage4.5 to 5.0 V
Maximum output current100 mA
External power inputnone
Connection Examples
Input Circuit
Connection Examples
Instruments
(source input, active-low)
(current sink)
Instruments
Input
Switch Connections
Instruments
InputInputOutput
Common
PLC Output (Sink Output) Connections
Input
Relay Connections
Instruments
Common
Output
PLC Output (Source Output) Connections
Output Circuit
HIOKI BT3562A981-09
Connection Examples
Instruments
BT3562,BT3563
50mA max
Output
Instruments
BT3562,BT3563
Output
89
5.4 Internal Circuitry
ISO_COM
Relay Connections
Instruments
BT3562,BT3563
Output
ISO_COM
Active-Low Logic Output
Instruments
BT3562,BT3563
30 V max
Active-Low
Logic Output
PLC
ISO_COM
LED Connection
Instruments
BT3562,BT3563
Output
Output
ISO_COM
Wired OR
5
Chapter 5 External Control (EXT I/O)
Common
Output
Input
ISO_COM
PLC Input (Source Input) Connections
Instruments
BT3562,BT3563
Output
ISO_COM
PLC Input (Sink Input) Connections
PLC
Input
Common
90
HIOKI BT3562A981-09
5.5 External Control Q&A
5.5External Control Q&A
Common QuestionsAnswers
How do I connect external trigger input?
Which pins are common ground for input and
output signals?
Are the common (signal ground) pins shared
by both inputs and outputs?
How do I confirm output signals?
How do I troubleshoot input (control) signal
issues?
Are the comparator decision signals retained
during measurement (or can they be off)?
Why would the EOM signal not be detected?
Connect the (active low) TRIG input pin to an ISO_COM pin using a
switch or open-collector output.
The ISO_COM pins.
Both common ground pins can be shared by inputs and outputs.
Confirm voltage waveforms with an oscilloscope. To do this, the output pins such as EOM
pulled up (through several k
For example, if triggering does not operate properly, bypass the
PLC and short the TRIG pin directly to an ISO_COM pin. Be careful
to avoid power shorts.
The state is determined at the end of measurement, and is off once
at the start of measurement.
Try using the Pulse setting for EOM output.
When the measurement time is short and EOM output is set to
Hold, the time to de-assert may be too short to be detected by the
PLC. When the EOM output is set to Pulse, the signal is asserted
(ON) for the specified pulse width before turning off.
and comparator decision outputs need to be
).
What situations cause measurement faults to
occur?
Is a connector or flat cable for connection provided?
Is direct connection to a PLC possible?
Can external I/O be used at the same time as
RS-232C or other communications?
How should external power be connected?
Can free-running measured values be
acquired using a footswitch?
An error is displayed in the following cases:
• A probe is not connected
• A contact is unstable
• A probe or measurement object is dirty or corroded
• Measurement object resistance is much higher than the measure-
ment range
A solder-type connector is supplied. The cable must be prepared at
the user's side.
Direct connection is supported for relay or open-collector outputs
and positive-ground optocoupler inputs. (Before connecting, confirm
that voltage and current ratings will not be exceeded.)
After setting up communications, it is possible to control measurement with the TRIG
communications interface.
The instrument's external I/O input and output signals all operate
from an internal isolated power source, so power must not be supplied from the PLC side.
Please use the free software for acquiring measured values available for download from our website.
signal while acquiring measurement data via a
6.1 Connecting the Printer
HIOKI BT3562A981-09
PrintingChapter 6
6.1Connecting the Printer
Before connecting the printer
Because electric shock and instrument damage hazards are present, always
follow the steps below when connecting the printer.
• Always turn off the instrument and the printer before connecting.
• A serious hazard can occur if a wire becomes dislocated and contacts
another conductor during operation. Make certain connections are secure.
91
Recommended printer
• As much as possible, avoid printing in hot and humid environments. Other-
wise, printer life may be severely shortened.
• Use only compatible recording paper in the printer. Using non-specified
paper may not only result in faulty printing, but printing may become impossible.
• If the recording paper is skewed on the roller, paper jams may result.
The requirements for a printer to be connected to the instrument are as follows.
Confirm compatibility and make the appropriate settings on the printer before
connecting it to the instrument.
• Control codes ........................ Capable of directly printing plain text
The optional printer model 9670 is no longer available. Their model 9670 printers
can still use.
6
Chapter 6 Printing
92
HIOKI BT3562A981-09
6.1 Connecting the Printer
Connecting the PRINTER to the Instrument
Model BT3562 (BT3562-01)
4
Printer
3
2
1. Confirm that the instrument and
Printer are turned off.
2. Connect the AC Adapter to the
Printer, and insert the power
plug into an outlet.
3.
Connect the RS-232C Cable
the RS-232C connectors on the
instrument and printer.
to
RS-232C Cable
Connector Pinouts
1 2 3 4 5
6 7 8 9
Instrument (9-pin) Connector
Function
Receive DataRxD2
Transmit DataTxD3
Signal or Common GroundGND5
Signal
Name
AC Adapter
Pin
4. Turn the instrument and printer
on.
13 ........................................ 1
25 ................................ 14
Printer (25-pin) Connector (Example)
Pin
Signal
Name
2TxDTransmit Data
3RxDReceive Data
7GNDSignal or Common Ground
4RTSRequest to Send
5CTSClear to Send
Function
6.2Selecting the Interface
HIOKI BT3562A981-09
93
6.2 Selecting the Interface
1
2
3
Or
numeric keypads
(SHIFT Lamp lit)
The Menu display appears.
Select the Interface Selection display.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
(Main Display)
(Sub Display)
The current setting blinks.
Select Printer on the Sub Display.
rS ............ RS-232C
LAn......... LAN
GP-Ib...... GP-IB
Prn.......... Printer
Set the print interval time.
0000 .........................Interval printing is OFF. (Printing is carried out once when
PRINT key is pressed.)
0001 to 3600 ............Sets the print interval time in seconds.
6
Chapter 6 Printing
4
Applies setting and returns to the Measurement display.
94
HIOKI BT3562A981-09
6.3 Printing
6.3Printing
Printing Measured Values and Decision Results ____________________
From the Measurement display, press the PRINT key or short the PRINT pin to
the ISO_COM of the EXT I/O connector to print the measured value and decision result.
• When using the external trigger, if you want to print after a triggered measurement finishes, connect the EOM signal of the EXT I/O to the PRINT signal.
• To print all measurements continuously, connect the EOM
PRINT signal and enable the internal trigger.
• When the statistical calculation function is on and the internal trigger is
selected, the TRIG key or TRIG
printing of the current measurement value.
• Valid counts are 1 to 30000. Above 30000, the count returns to 1.
This function allows you to automatically print out measurement results at preset
intervals. The print interval time must be set from the Interface Selection display.
See "6.2 Selecting the Interface" (p.93).
The setting range is 1 to 3600 seconds.
When the print interval time is set to "0", interval printing is disabled, and only
normal printing is carried out.
Operation when interval printing is selected:
1. Start printing by pressing the PRINT key or sending the PRINT
EXT I/O.
2. Elapsed time (hours/minutes/seconds) and measurement values are printed
automatically at intervals corresponding to the preset interval time.
3. Stop printing by pressing the PRINT key or sending the PRINT
I/O again.
• When the printed elapsed time reaches 100 hours, it resets to 00:00:00 and
continues from zero.
(Example)
After 99 hours, 59 minutes and 50 seconds: 99:59:50
After 100 hours, 2 minutes and 30 seconds: 00:02:30
• Selecting a display other than the measurement display causes interval
printing to stop.
From the Statistical Calculation display, press the PRINT key to print statistical
calculation results. If no valid data exists, only the data count is printed. When
only one valid data sample exists, standard deviation of sample and process
capability indices cannot be printed.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.