To evaluate whether batteries have deteriorated, measure the
internal resistance of a fresh battery. Deteriorated batteries will have
about 1.5 to 2 times (values for reference purposes) as high internal
resistance as a fresh one. Use these values as guidelines when
determining deterioration evaluation values.
Example: Changes in the internal resistance and voltage
associated with battery deterioration
Resistance has
doubled . . .
Deteriorated
Internal resistance: 0.5 m
Voltage: 2.0 V
(Example values)
Deterioration
evaluation value
of resistance
Initial value × 2
Initial value × 1.5
BT3554F961-00
FAIL
WARNING
PASS
Initial value
Fresh
Need
replacing
1.0 m
Ω
(Doubled)
Ω
1.8 V (Decreased by 10%)
Internal
resistance
Available
charge
amount
PassPassPassPassFail
HIOKI BT3554F961-00HIOKI BT3554F961-00
Basic instruction on how to use the instrument
Connect the test lead to the
1
instrument.
Turn on the instrument.
2
Set the date and time when using
the instrument for the rst time.
(p. 43)
Choose a resistance range.
3
(p. 47)
Choose a voltage range.
4
(p. 47)
Enable the automatic hold and
5
automatic memory functions.
(The
segments will appear.)
and
When the measured values become stable, the
display will automatically freeze them.
See “3.6 Automatic Hold Function” (p. 59).
Immediately after the display freezes measured
values, the instrument will automatically save them.
See “5.3 Automatic Memory Function” (p. 87).
(Black)(Red)
Hold down for
at least 1 s.
Connect the test lead to a battery under measurement.
HIOKI BT3554F961-00HIOKI BT3554F961-00
6
Saved in
internal
memory.
+
(Black)(Red)
−
HIOKI BT3554F961-00HIOKI BT3554F961-00
Convenient functions
Comparator function
You can set threshold values to evaluate battery deterioration using
the comparator function. (p. 67)
Example of setting deterioration evaluation values
PASSWARNINGFAIL
FreshIn use
Acceptance value
Resistance: 0.5 m
Voltage: 2.0 V
Hold down for
at least 1 s.
Warning threshold
0.75 m
Ω
(Fresh battery’s value × 1.5)
1.8 V
(Fresh battery’s value × 0.9)
Ω
Downloading measured values to your computer
You can connect the instrument and your computer using the
accompanying USB cable to download measured values. (p. 103)
Need
replacing
Rejection value
1.0 m
Ω
(Fresh battery’s value × 2)
Prole information (p. 82)
HIOKI BT3554F961-00HIOKI BT3554F961-00
12
Where this battery
has been installed?
3
Prole information
comes in handy.
How can I help you?
4
Prole information?
Let me know the
details.
“Hiokichi-kun”
Hioki’s corporate mascot
The use of prole information lets you save measured data
Resistance value
Voltage value
Temperature
Comparator threshold
values
Comparison resultPASS, WARNING, or FAIL
(yyyy/mm/dd hh:mm)
m
.
.
.
, mΩ, V
m
Ω
Ω
V
°C
HIOKI BT3554F961-00HIOKI BT3554F961-00
Measurement recording prompt (p. 95)
12
Let’s get down
to work.
3
The memory contains
49 data sets only . . .
One short!
The 50th battery.
That’s it.
4
I have to start from
scratch again.
The measurement recording prompt uses voice prompts and the
display to inform you of the battery number you are to measure
next. You can get your jobs done neatly, which will no longer require
you to start all over again.
Calibrating the resistance measurement unit .........162
Calibrating the voltage measurement unit .............. 163
Index 165
WarrantyCerticate
v
Contents
HIOKI BT3554F961-00HIOKI BT3554F961-00
vi
Introduction
HIOKI BT3554F961-00HIOKI BT3554F961-00
Introduction
Thank you very much for choosing the Hioki BT3554-50, BT355451, BT3554-52 Battery Tester. To ensure your ability to get the most
out of this instrument over the long term, please read this manual
carefully and keep it available for future reference.
Model
number
(Order code)
BT3554-50
BT3554-519465-10 Pin Type Lead
BT3554-52L2020 Pin Type Lead
Hereafter, the model number is referred to as the one that appears
on the product, BT3554-50.
Model name printed on
the instrument
BT3554-50
Standard accessory lead
None
Trademarks
• Android, Google Play, and Google Chrome are trademarks of Google, Inc.
• IOS is a registered trademark of Cisco Systems, Inc. and/or its aliates in
the United States and certain other countries.
• Other products and company names are trade names, registered
trademarks, or trademarks of their respective owners.
1
Verifying Package Contents
HIOKI BT3554F961-00HIOKI BT3554F961-00
Verifying Package Contents
When you open the package, carefully inspect the instrument to
ensure that everything is in good condition, and that no damage
occurred during shipping. In particular, check the accessories, panel
switches, and connectors. If the instrument seems to have been
damaged or does not work as specied, contact your authorized
The latest version can be downloaded from our website.
*:
2
USB cable
Zero-adjustment board
Z5050 Fuse Set
Pin Type Lead
9465-10 (For BT3554-51)
L2020 (For BT3554-52)
Not included
(For BT3554-50)
BT3554-50 Battery Tester
(With the Z5041 Protector
attached)
C1014 Carrying Case
Neck strap
(p. 44)
Options
HIOKI BT3554F961-00HIOKI BT3554F961-00
Options
The options listed below are available for the instrument. To order
an option, contact your authorized Hioki distributor or reseller.
Options are subject to change. Please check Hioki’s website for
the latest information. Please check Hioki’s website for the latest
information.
Model 9465-10 Pin Type Lead
This pin type lead has a fourterminal structure.
Model L2020 Pin Type Lead
This pin type lead has a fourterminal structure and can be used
for hard-to-reach measurement
targets.
Model 9465-90 Tip Pin
The 9465-90 is a replacement tip
pin for the 9465-10 and L2020 Pin
Type Lead.
2.7 mmφ1.27 mm
φ
3
Options
HIOKI BT3554F961-00HIOKI BT3554F961-00
Model 9772 Pin Type Lead
This pin type lead has parallel
arranged pins. The pins possess
high-strength wear-resistant.
This lead, having the pins that
can be inserted into a hole 5 mm
in diameter, lets you perform
measurement without removing
terminal covers. You can also
perform measurement in virtually
any location because the pins can
be inserted diagonally in hard-toreach places.
Model 9772-90 Tip Pin
The 9772-90 is a replacement tip
pin for the 9772 Pin Type Lead.
Model 9460 Clip Type Lead with
Temperature Sensor
Using the 9460 lets you measure
resistance, voltage, and
temperature simultaneously.
4.3 mm
2.5 mm
9.15 mm
1.8 mm
φ
Clip (Black)
Temperature
sensor
4
Mini plug
(Connect to TEMP.SENSOR)
Options
HIOKI BT3554F961-00HIOKI BT3554F961-00
Model 9466 Remote Control
Switch
Attaching the 9466 to the test lead
lets you freeze displayed values
during measurement.
Supported models:
• Model 9465-10 Pin Type Lead
• Model 9772 Pin Type Lead
• Model L2020 Pin Type Lead
Model 9467 Large Clip Type Lead
The 9467 can clip on thick rodlike
terminals of measurement targets.
You can perform four-terminal
measurement just by clipping the
lead on to the target.
Approx. φ29 mm
Model 9451 Temperature Probe
(Cable length: 1.5 m)
Connect the 9451 to the TEMP.
SENSOR terminal on the top face
of the instrument.
Mini plug φ2.5 mm
(Connect to EXT.HOLD)
Switch
5
Options
HIOKI BT3554F961-00HIOKI BT3554F961-00
Model 9451S Temperature Probe
Order code: 9451-01
(Cable length: 0.1 m)
Connect the 9451S to the TEMP.
SENSOR terminal on the top face
of the instrument.
Model Z5038 0 Adj Board
(For the 9465-10, L2020, and
9772)
Hook-and-loop fasteners are
separately necessary to stick the
Z5038 to the carrying case. Please
use commercially available hookand-loop fasteners.
Model Z5050 Fuse Set
Always use the specied fuse.
Model Z3210 Wireless Adapter
Model C1014 Carrying Case
Model Z5041 Protector
6
Safety Information
HIOKI BT3554F961-00HIOKI BT3554F961-00
Safety Information
This instrument is designed to conform to IEC 61010 Safety
Standards, and has been thoroughly tested for safety prior to
shipment. However, using the instrument in a way not described in
this manual may negate the provided safety features.
Carefully read the following safety notes before using the
instrument.
DANGER
Mishandling the instrument could result in bodily
injury or even death, as well as damage to the
instrument. Familiarize yourself with the instructions
and precautions in this manual before use.
WARNING
Electricity can potentially cause serious events such
as an electric shock, heat generation, re, and an arc
ash due to a short-circuit. If you have not used any
electrical measuring instruments before, you should
be supervised by a technician who has experience in
electrical measurement.
Protective gear
Performing measurement using this instrument
involves live-line work. To prevent an electric shock,
use appropriate protective insulation and adhere to
applicable laws and regulations.
WARNING
7
Safety Information
HIOKI BT3554F961-00HIOKI BT3554F961-00
Symbols and abbreviations
In this document, the severity levels of risk and hazard are classied
as follows.
Indicates an imminently hazardous situation
DANGER
WARNING
CAUTION
IMPORTANT
that will result in death of or serious injury to the
operator.
Indicates a potentially hazardous situation that
may result in death of or serious injury to the
operator.
Indicates a potentially hazardous situation
that may result in minor or moderate injury to
the operator or damage to the instrument or
malfunction.
Indicates information or content that is particularly
important from the standpoint of operating or
maintaining the instrument.
Indicates useful advice concerning instrument
performance and operation.
Indicates a high voltage hazard. Failure to verify
safety or improper handling of the instrument
could lead to an electric shock, burn, or death.
Indicates an action that must not be performed.
Indicates an action that must be performed.
HOLD
[HOLD]
8
Indicates a control key.
Indicates elements on the display.
Symbols on the instrument
HIOKI BT3554F961-00HIOKI BT3554F961-00
Indicates the presence of a potential hazard. When the symbol
is printed on the instrument, refer to a corresponding topic in the
Instruction Manual.
Indicates a fuse.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates the power button that switches the instrument between
on and o states.
Symbols for various standards
Indicates that the product is subject to the Waste Electrical and
Electronic Equipment (WEEE) Directive in EU member nations.
Dispose of the product in accordance with local regulations.
Indicates that the product complies with standards imposed by
EU directives.
Safety Information
9
Safety Information
HIOKI BT3554F961-00HIOKI BT3554F961-00
Character expression
The instrument’s display expresses the alphanumeric characters as
follows.
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
1 2 3 4 5 6 7 8 9 0
Some dierent expressions are used as below:
Indicates that the saved data has been deleted.
Indicates that the comparator buzzer is set to
FAIL.
Indicates that the A/D converter communications
error occurs.
Accuracy labeling
The instrument accuracy is expressed by dening a percentage of
the reading and a limit value for errors in terms of digits.
readingDisplay value
Indicates the value displayed by the instrument. Limit values
for reading errors are expressed as a percentage of the
reading (“% of reading”).
digitsResolution
Indicates the minimum display unit (in other words, the
smallest digit that can have a value of 1) for a digital
measuring instrument. Limit values for digit errors are
expressed using digits.
10
Operating Precautions
HIOKI BT3554F961-00HIOKI BT3554F961-00
Operating Precautions
Observe the following precautionary information to ensure that the
instrument can be used safely and in a manner that allows it to
perform as described in its specications.
Use of the instrument should conrm not only to its specications,
but also to the specications of all accessories, options, LR6
Alkaline batteries, and other equipment in use.
Installing the instrument
CAUTION
Installing the instrument in inappropriate locations could
cause a malfunction of the instrument or an accident.
• Exposed to direct sunlight or high temperature
• Exposed to corrosive or combustible gases
• Exposed to a strong electromagnetic elds or
electrostatic charges
• Near induction heating systems (such as high-frequency
induction heating systems and IH cooking equipment)
• Susceptible to vibration
• Exposed to water, oil, chemicals, or solvents
• Exposed to high humidity or condensation
• Exposed to high concentrations of dust particles
Do not place the instrument on an unstable or uneven
surface. Doing so could cause the instrument to fall or turn
over, causing bodily injury or damage to the instrument.
11
Operating Precautions
HIOKI BT3554F961-00HIOKI BT3554F961-00
Preliminary checks
DANGER
If the test lead or the instrument is damaged, there
is a risk of an electric shock. Perform the following
inspection before use:
• Check that the insulation of the test lead is neither
ripped nor torn and that no metal parts are exposed.
Replace the test lead with a one specied by Hioki.
• Check the instrument for any damage that may
have occurred during storage or shipping, and
perform functional checks before use. If you nd
any damage to the instrument, please contact your
authorized Hioki distributor or reseller.
Precautions for Transportation
During shipment of the instrument, handle it carefully so that it is not
damaged due to a vibration or shock.
12
Handling the instrument
HIOKI BT3554F961-00HIOKI BT3554F961-00
To prevent an electric shock, do not remove the
instrument’s case. The internal components of the
instrument carry high voltages and may become very
hot during operation.
To avoid damage to the instrument, do not subject it to
vibration or mechanical shock during transportation and
handling. Exercise particular care to avoid subjecting the
instrument to mechanical shock, for example by dropping.
Operating Precautions
DANGER
CAUTION
13
Operating Precautions
HIOKI BT3554F961-00HIOKI BT3554F961-00
Precautions for measurement
DANGER
To prevent an electric shock, be careful to avoid
shorting live lines with the test lead tip.
WARNING
• Do not use the instrument to measure circuits that
exceed its ratings or specications. Damage to the
instrument can cause an electric shock.
• Do not measure any voltage that would exceed the
instrument’s maximum input voltage (terminal-toterminal) or maximum rated line-to-ground voltage
of 60 V.
Maximum input voltage
(terminal-to-terminal)
60 V DC
• Do not measure AC voltage.
Maximum rated
voltage to earth
60 V DC
14
Operating Precautions
HIOKI BT3554F961-00HIOKI BT3554F961-00
WARNING
• Connect the test lead correctly.
• Wear gloves of rubber or similar material during
measurement.
• Ventilate the rooms where batteries have been
installed before measuring batteries to prevent
explosions. Sparks may occur when a test lead is
connected to a battery to be measured, which can
ignite any accumulated inammable gases such as
hydrogen.
CAUTION
• After measuring a high-voltage battery, rst short-circuit
the test lead to discharges the DC elimination capacitor
connected across the lead before continuing to measure
a low-voltage battery. Otherwise the low-voltage battery
may be subject to an excess voltage, causing damage
to the battery.
• To avoid damage to the instrument, do not apply voltage
to the EXT.HOLD and TEMP.SENSOR terminal.
15
Operating Precautions
HIOKI BT3554F961-00HIOKI BT3554F961-00
IMPORTANT
• Do not place the test lead in contact with the measurement
terminals of a leaky battery. Doing so may cause a degradation
in the instrument functionality due to exposure to electrolyte
from the leaky battery.
• Subjecting the test lead to an excessive common-mode voltage
could cause the following issues:
(1) Unstable measured values
(2) Wire-break detection display ([−−−−])
Attaching ferrite cores around the test lead or placing the
instrument some distance away from the oor may have a
lightening the eect.
16
Operating Precautions
HIOKI BT3554F961-00HIOKI BT3554F961-00
IMPORTANT
• If there is a potential dierence between the SOURCE−
terminal and SENSE − terminal or between the SOURCE +
terminal and SENSE + terminal, the instrument cannot perform
correct measurement.
SOURCE +SOURCE −
SENSE +SENSE −
OK
SENSE +SENSE −
SOURCE +SOURCE −
NO
17
Operating Precautions
HIOKI BT3554F961-00HIOKI BT3554F961-00
Handling the test lead
Do not subject the pin type lead tip to force when it is in
contact with the battery under measurement at a tilted
angle.
CAUTION
Avoid subjecting the temperature probe tip to physical
shock, and avoid sharp bends in the lead. These may
damage the probe or break a wire.
IMPORTANT
Use only the Hioki-specied test lead. Using another test lead
may result in incorrect measurements due to loose connections
or other reasons. Besides, Hioki does not guarantee the accuracy
and proper operation.
Zero-adjustment board
To prevent short-circuit accidents, do not place the
zero-adjustment board on top of a battery under
measurement.
18
OK
NO
WARNING
Instrument’s batteries and fuse
HIOKI BT3554F961-00HIOKI BT3554F961-00
WARNING
• To avoid an electric shock, disconnect the test lead
from the object under measurement before open the
cover to replace the LR6 Alkaline batteries or fuse.
• To prevent instrument damage or an electric shock,
use only the screw for securing the fuse cover in
place that is originally installed. If you have lost the
screw or nd that the screw is damaged, please
contact your Hioki distributor for a replacement.
• Use a fuse only specied by Hioki.
Failure to observe this could damage the
instrument, resulting in bodily injury.
Specied fuse: Model Z5050 Fuse Set (216.630,
Littelfuse Inc., fast-acting, rating: 250 V / F 630 mA,
interrupting rating: 1500 A)
• Do not short-circuit, recharge ,or disassemble
LR6 Alkaline batteries, or dispose of them in re.
Batteries may explode if mistreated.
• Do not use the instrument with the fuse holder
short-circuited.
Failure to observe this could damage the
instrument, resulting in bodily injury.
Operating Precautions
19
Operating Precautions
HIOKI BT3554F961-00HIOKI BT3554F961-00
CAUTION
Poor performance or damage from LR6 Alkaline battery
leakage could result. Observe the cautions listed below:
• Do not mix old and new LR6 Alkaline batteries, or
dierent types of LR6 Alkaline batteries.
• Be careful to observe the battery polarity during
installation.
• Do not use LR6 Alkaline batteries after their
recommended expiry date.
• Do not allow weak LR6 Alkaline batteries to remain in
the instrument.
• Replace LR6 Alkaline batteries only with ones of the
specied type.
• Remove the LR6 Alkaline batteries from the instrument
if they are to be stored for a long time.
Handle and dispose of LR6 Alkaline batteries in accordance with
local regulations.
Precautions related to use of the CD
• Exercise care to keep the recorded side of the disc free of dirt
and scratches. When writing text on the disc’s label, use a pen or
marker with a soft tip.
• Keep the disc inside a protective case and do not expose to direct
sunlight, high temperature, or high humidity.
• Hioki is not liable for any issues your computer system
experiences in the course of using this disc.
20
Overview
HIOKI BT3554F961-00HIOKI BT3554F961-00
1
1.1 Evaluating Battery Deterioration
IMPORTANT
To evaluate whether batteries have deteriorated, rst measure the
internal resistance of a fresh or non-defective battery.
If a battery has deteriorated, the internal resistance will increase
by 50 to 100 percent (value for reference purposes) from its initial
value.
The graph below shows the relation between the amount of battery
charge available and the initial value of internal resistance in a leadacid battery. “CS,” “HS,” and “MSE” denote lead-acid battery types
according to Japanese Industrial Standard (JIS).
The internal resistance of an MSE (sealed stationary lead-acid
battery) can be graphically read at approximately 1 m
and approximately 0.13 mΩ (1000 Ah).
)
(m
Ω
CSClad-type stationary
lead-acid battery
HSHigh-rate-discharge
paste-type stationary
lead-acid battery
MSE Sealed stationary lead-
acid battery
CS
HS
Internal resistance value
MSE
(100 Ah)
Ω
Lead-acid battery capacity
(Ah)
21
Evaluating Battery Deterioration
HIOKI BT3554F961-00HIOKI BT3554F961-00
• For an MSE (sealed stationary lead-acid battery), the warning
threshold (WARNING) of the internal resistance is dened as
about 1.5 times its initial value. The rejection value (FAIL) varies
depending on the manufacturer.
• Initial values of internal resistance may vary among batteries
under measurement with the same capacity, depending on the
model or manufacturer. Use the graph on the previous page for
reference purposes.
• Warning thresholds (WARNING) and rejection values (FAIL) of
internal resistance vary depending on the manufacturer.
Source: Battery technician certication textbook, Battery Association
of Japan (BAJ)
The battery measurement values can be compared
with the present threshold values using the comparator
function to determine the scope in which the values fall
within: PASS, WARNING, or FAIL.
See “4 Comparator Function (Evaluation Based on
Threshold Values)” (p. 67).
Changes in the internal resistance of open (liquid) stationary
lead-acid batteries, including CS and HS, and alkaline lead-acid
batteries, may be smaller than sealed lead-acid batteries. Thus, it is
sometimes dicult to determine whether such batteries have been
deteriorated.
22
Evaluating Battery Deterioration
HIOKI BT3554F961-00HIOKI BT3554F961-00
Measuring lithium-ion batteries
The instrument measures internal resistance and voltage of
batteries using AC having a frequency of 1 kHz. The internal
resistance of lithium-ion batteries can also be measured; however,
that of a packed battery will include the protective resistance, which
is installed in packed batteries. Besides, the instrument may not be
able to diagnose and evaluate the deterioration of such batteries
because changes in the internal resistance of lithium-ion batteries
due to deterioration may be smaller than lead-acid batteries.
23
Overview
HIOKI BT3554F961-00HIOKI BT3554F961-00
1.2 Overview
The instrument measures internal resistance, voltage, and terminal
temperature of lead-acid, nickel-cadmium, nickel-hydrogen, and
other types of batteries, enabling you to determine if batteries have
deteriorated.
To measure terminal temperature, the 9460 Clip Type Lead with Temperature
Sensor (option) is required. To measure an ambient temperature, the
9451/9451S Temperature Probe (option) is required.
After measurement, connecting the instrument to your computer
with the accompanying USB cable enables measured data
transfer to the computer. Also, you can browse through and record
measured data on your smart phone or tablet using the wireless
communications function.
24
Using the wireless communications function
HIOKI BT3554F961-00HIOKI BT3554F961-00
requires the Z3210 Wireless Adapter.
Measured data
Measured data
Prole information
Threshold values setting
Overview
Threshold values setting
Prole information
25
Features
HIOKI BT3554F961-00HIOKI BT3554F961-00
1.3 Features
Facilitated data management
Measured data can be saved in connection with
prole information.
The instrument can save up to 6000 data sets consisting of
presently measured data (resistance, voltage, temperature,
comparison result). This equates to up to 12 units, each of which
consisting of 500 cell cubicles.
The instrument can be populated 100 sets of prole information
(comments, such as location information and device information,
and the buttery number information).
Saving measured data in connection with prole information
makes it possible for you to manage measurement locations,
UPSs, and batteries easily.
Measurement recording prompt
The instrument and your mobile device with GENNECT Cross
installed, using the display and voice prompts, respectively, can
inform you of comparison results and the battery number you
are to measure next. This enables you to record measured data
quickly.
Automatic hold function and automatic memory
function
When these functions are enabled, the instrument can
automatically, at the instance when the display freezes measured
values, save the measured values in its internal memory. This
can lead to increased operational eciency.
Measurement without the need for UPS system
shutdown
The instrument uses high-precision AC low-resistance
measurement technology and noise reduction technology. Time
26
Features
HIOKI BT3554F961-00HIOKI BT3554F961-00
required for measurement will be reduced since the instrument is
capable of measuring live wires without requiring a UPS system
shutdown.
Reliable measured values
The instrument is capable of obtaining reliable measured values
without being aected by wire resistance of the test lead or
contact resistance because it uses the AC four-terminal method
to measure internal resistance.
Simultaneous display of resistance, voltage, and
temperature
Without switching functions, the instrument can display
battery internal resistance, voltage, and terminal temperature
simultaneously. To measure terminal temperature, the 9460 Clip
Type Lead with Temperature Sensor (option) is required. To
measure an ambient temperature, the 9451/9451S Temperature
Probe (option) is required.
Comparator function
The comparator function enables you to set threshold values
for internal resistance and voltage. This can evaluate whether
batteries have deteriorated more easily.
Computer interface
Measured data can be transferred to your computer.
Wireless communications function
Connecting the Z3210 Wireless Adapter (option) enables you
to browse through and record measured values with your smart
phone and tablet.
You can utilize the measurement recording prompt in sync with
your mobile device with GENNECT Cross installed.
Holding down the key for 3 s to toggle between Celsius
2
and Fahrenheit.
Press the key to conrm your choice.
3
−
−
31
Names and Functions of Parts
HIOKI BT3554F961-00HIOKI BT3554F961-00
Display
Data number
Measured resistance
value and its unit
Measured voltage value
and its unit
Wireless
communications
function enabled.
Freezes measured
values.
Input overows.
Comparator buzzer
enabled.
Communicating
through USB.
Instrument’s battery
level
Measurement
recording prompt
enabled.
32
Measured temperature
value and its unit
Memory deleted.
Chosen memory
number occupied.
Chosen memory
number contains
prole information.
Prole number
Zero adjustment
enabled
Noise-frequency
reduction enabled.
Function being set.
Names and Functions of Parts
HIOKI BT3554F961-00HIOKI BT3554F961-00
Measurement
recording voice
prompt enabled.
Automatic hold
enabled.
Automatic memory
enabled.
Automatic power
saving enabled.
Next battery number
to be measured and
recorded
(When measurement
recording prompt
enabled)
Memory number to
be saved
Memory number to
be loaded
Comparator number
Comparator
threshold values
being set.
Comparator
enabled.
PASS judgment
given.
WARNING result
given.
FAIL judgment given.
When all segments on the display are shown, segments other than
those listed above also appear, which are not used.
33
Names and Functions of Parts
HIOKI BT3554F961-00HIOKI BT3554F961-00
Top
EXT.HOLD
terminal
USB terminal
TEMP.SENSOR terminal
EXT.HOLD terminalConnect the 9466 Remote Control Switch (option)
USB terminalConnect the USB cable here.
TEMP.SENSOR
terminal
34
here.
Connect the mini plug of the 9460 Clip Type Lead
with Temperature Sensor (option) here.
Connect the 9451/9451S Temperature Probe here.
Rear
HIOKI BT3554F961-00HIOKI BT3554F961-00
Names and Functions of Parts
Fuse cover (
Screw
Serial
number*
Battery cover (
* The serial number consists of nine digits. The rst two (from the
left) indicate the year of manufacture, and the next two indicate
the month of manufacture. Required for production control. Do not
peel o the label.
p. 19, p. 147)
p. 19, p. 147)
35
Outer Dimension Drawing
HIOKI BT3554F961-00HIOKI BT3554F961-00
1.5 Outer Dimension Drawing
60.6 ±3 mm
199 ±3 mm
132 ±3 mm
36
Preparing for Measurement
HIOKI BT3554F961-00HIOKI BT3554F961-00
2
2.1 Installing/Replacing LR6 Alkaline
Batteries
When using the instrument for the rst time, insert eight LR6
Alkaline batteries. Before attempting measurement, check the
instrument’s battery level. If the instrument’s batteries are running
low, replace the LR6 Alkaline batteries with fresh ones.
When the segment blinks, which indicates that the
instrument’s batteries are running low, replace them as soon as
possible.
Turn o the instrument and
1
remove the test lead.
1
Remove the battery cover on
2
the rear of the instrument.
Ensure correct polarity and
3
insert eight LR6 Alkaline
batteries.
Install the battery cover.
4
The batteries can continuously
provide power to the instrument
with the Z3210 not installed for
about 8.3 hours (with the Z3210
installed for communications,
about 8.2 hours).
4
3
2
37
Attaching the Z5041 Protector
HIOKI BT3554F961-00HIOKI BT3554F961-00
2.2 Attaching the Z5041 Protector
If the Z5041 Protector has been removed from the instrument,
attach it by following the instruction below.
Turn o the instrument and
1
remove the test lead.
Insert the instrument into the
2
Z5041 Protector.
Observe the proper orientation.
Push the instrument into the
3
Protector in the direction of the
arrow.
OK
NO
38
(Completed)
Connecting the Z3210 Wireless Adapter (Option)
HIOKI BT3554F961-00HIOKI BT3554F961-00
2.3 Connecting the Z3210 Wireless
Adapter (Option)
Installing the Z3210 Wireless Adapter (option) in the instrument lets
you use the wireless communications function.
See “7.2 Communicating With a Mobile Device” (p. 106).
WARNING
Turn o the instrument and remove the test lead.
Failure to do so could cause an electric shock.
CAUTION
After touching any metallic part, such as a doorknob,
to eliminate static electricity from your body, connect/
disconnect the Z3210.
Failure to do so could cause static electricity to damage
the Z3210.
Turn o the instrument and
1
remove the test lead.
Remove the Z5041 Protector
2
while pressing down it as
shown.
Remove the protective cap with
3
a athead screw driver.
Insert the Z3210, observing the
4
correct orientation, as far as it
will go.
Attach the Protector.
5
2
2
3
4
• Store the removed protective cap.
• When removing the Z3210, install the protective cap.
39
Connecting the Test Lead
HIOKI BT3554F961-00HIOKI BT3554F961-00
2.4 Connecting the Test Lead
WARNING
To avoid an electric shock, be sure to connect the test
lead properly.
This section describes how to connect the test lead to the
instrument. Connect the test lead’s connectors to all four terminals:
the SOURCE terminals (positive and negative) and the SENSE
terminals (positive and negative).
(Black)
(Red)
Align the marks
of the same color.
SOURCE
SENSE
(Red)(Black)
When using the 9460 Clip Type Lead with Temperature Sensor
(option), connect the mini plug to the TEMP.SENSOR terminal.
See “3.9 Measuring Temperature” (p. 66).
40
Connecting the Test Lead
HIOKI BT3554F961-00HIOKI BT3554F961-00
Bundling up the Pin Type Lead and the 9466
Remote Control Switch
You can bundle up the Pin Type Lead (the 9465-10, 9772, L2020)
and the 9466 Remote Control Switch (option).
Attach the Remote Control Switch to the probe of the Pin Type
Lead. Bundle up the two cables using the spiral tubes.
Model 9465-10
Model L2020
(1) Model 9466 Remote Control Switch
(2) Probe
(3) Spiral tube (small)
Bundle up the cables at the center of the lead between the
probe and probe-side junction with the spiral tube.
(4) Junctions
(5) Spiral tubes (large)
Bundle up the cables between the junctions if you wish.
(1)(2)(3)(4)
(4)(5)
41
Turning On/O the Instrument
HIOKI BT3554F961-00HIOKI BT3554F961-00
2.5 Turning On/O the Instrument
Hold down the key for at least 1 s to turn on/o the instrument.
Conrm the date and time settings when using the instrument for
the rst time.
Hold down for at least 1 s.
When the segment blinks, which indicates that the
instrument’s batteries are running low, replace them as soon as
possible.
See “9.5 Default Settings and Resettable Settings” (p. 137).
42
Setting the Date and Time
HIOKI BT3554F961-00HIOKI BT3554F961-00
2.6 Setting the Date and Time
The instrument can show the date and time. Conrm the date and
time settings when using the instrument for the rst time. The time
is displayed using a 24-hour clock. The instrument’s calendar can
automatically recognize leap years.
1
2
3
You can also set the date and time using GENNECT ONE
or GENNECT Cross.
(Hold down for at least
1 s.)
Let the date and time
appear.
Holding down the key at
least 1 s again will hide the
date and time.
Enter the date and time (in
the yyyy/mm/dd hh:mm
format).
Conrm your entry.
Date and time will not be
set if you switch over the
display from the clock
setup mode without
pressing the ENTER key.
43
Attaching the Neck Strap
HIOKI BT3554F961-00HIOKI BT3554F961-00
2.7 Attaching the Neck Strap
You can hang the instrument from your neck using the neck strap.
Attach the neck strap as described below.
Turn o the instrument and
1
remove the test lead.
Pass the webbing through the
2
openings of the instrument and
x the webbing with the triglide slides.
(There are two openings on
each of the left and right side)
Adjust the length of the neck
3
strap.
You can put the instrument in
the carrying case with the neck
strap attached.
Check that the neck strap is not
4
removed even if you pull it.
Face the
webbing
through
the tri-glide
outside.
44
Measurement
HIOKI BT3554F961-00HIOKI BT3554F961-00
3
To ensure safe operation, be sure to read “Operating Precautions”
(p. 11) before starting the measurements.
• There is a considerable dierence in the battery’s internal
resistances between a full-charge state and a discharge state.
To improve evaluation accuracy, perform measurements under
a constant condition (for example, in a full-charge state).
• Terminals of lead-acid batteries (objects under measurement)
have a high resistance. Thus, resistance values may dier
between contact locations, the case side and the tip of the
terminals. Place the test lead into contact with the terminals at
the same location every time.
See “11.4 Eects of Current Density” (p. 157).
• Use the 9460 Clip Type Lead with Temperature Sensor
(option) to measure the temperature of the battery terminals.
Otherwise, use a non-contact thermometer, such as a radiation
thermometer, for safety.
• If terminals are covered with an insulating layer, the
measurement current cannot ow suciently, resulting in
measurement failure. In such a case, clean the terminals to
remove the insulation layer before measurement.
45
Inspection Before Measurement
HIOKI BT3554F961-00HIOKI BT3554F961-00
3.1 Inspection Before Measurement
Check the instrument for any damage that may have occurred
during storage or shipping, and perform functional checks before
use. If you nd any damage to the instrument, please contact your
authorized Hioki distributor or reseller.
Inspection itemChecking method
Has the fuse
blown?
Has the test lead
broken?
Place the test lead into contact with the zeroadjustment board. If the resistance readout
remains the [−−−−] segments, the fuse may
have blown or the test lead may have broken.
Replace the fuse or test lead with a new one.
Does the
instrument’s
battery level
remain enough?
Inspecting
batteries to be
measured
46
The display, in the top right, includes the
battery level indicator
batteries. If the indicator exhibits , the
batteries should be replaced early. Please
have spare LR6 Alkaline batteries ready.
If terminals are covered with an insulating
layer, the measurement current cannot ow
suciently, resulting in measurement failure.
In such a case, clean the terminals to remove
the insulation layer before measurement.
for the instrument’s
Setting the Measurement Ranges
HIOKI BT3554F961-00HIOKI BT3554F961-00
3.2 Setting the Measurement Ranges
This section describes how to set resistance and voltage
measurement ranges.
Resistance
ranges
Voltage
ranges
Temperature
range
Press the
key repeatedly to cycle through the ranges.
3 mΩ, 30 mΩ, 300 mΩ, 3
6 V, 60 V
(Single range)
Since the instrument has a single temperature
measurement range, setting the temperature
range is not required.
key or V key to display the present settings. Press the
Ω
Ω
47
Setting the Measurement Ranges
HIOKI BT3554F961-00HIOKI BT3554F961-00
Resistance ranges
3.000 m
3.000
Ω
Ω
30.00 m
300.0 m
Ω
Ω
Voltage ranges
6.000 V60.00 V
After a certain period of inactivity, the instrument will conrm your
entry and the display will return to measurement mode.
48
Noise-Frequency Reduction Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
3.3 Noise-Frequency Reduction
Function
Enabling the noise-frequency reduction function can reduce the
eects of noise exhibited in measurement environments, resulting in
smaller variation in measured values. Resistance measured values
will become more stable.
(Noise reduction technology)
Noise-frequency reduction
function disabled
When the segment pops up:
The noise-frequency reduction function is enabled.
When the segment is blinking:
The noise frequencies are being avoided.
Disabling the noise-frequency reduction function
Cycling the instrument will disable the function.
Noise-frequency reduction
function enabled
• It may take longer to perform measurement with the noise-
frequency reduction function enabled.
This time, the instrument will blink the
• It may not be possible to avoid all noise depending on the noise
frequency.
segment.
49
Adjusting the Zero Point (Zero Adjustment)
HIOKI BT3554F961-00HIOKI BT3554F961-00
3.4 Adjusting the Zero Point
(Zero Adjustment)
When the zero adjustment function executed, the instrument will
regard the measured values (correction values) as zero to display
subsequent measurement results.
Only if an accessory or optional test lead is used, the instrument
can meet the accuracy specications even without executing the
zero adjustment.
However, execute the zero adjustment in the following cases:
• When you wish to increase measurement accuracy
For the 3 m
on whether zero adjustment has been performed.
See “9.3 Accuracy Specications” (p. 123).
• When using a test lead, including Hioki product, that is not an
accessory or option of the instrument or whose length has been
extended
IMPORTANT
Use only the Hioki-specied test lead. Hioki does not guarantee
the accuracy and proper operation if any test lead not specied by
the company is used.
• Performing zero adjustment adjusts the zero points of all
ranges.
• Even after the instrument has been turned o, the correction
values will be retained and with the zero adjustment function
remaining enabled.
• After replacing the test lead, always perform zero adjustment
prior to measurement.
• Always use the included or optional zero-adjustment board
when performing zero adjustment.
• Keep the test lead short-circuited during zero adjustment.
• Keep the tips of the test lead away from any metal components.
range, the accuracy specications dier depending
Ω
50
Adjusting the Zero Point (Zero Adjustment)
HIOKI BT3554F961-00HIOKI BT3554F961-00
How to short-circuit various test lead
For Pin Type Lead
Use the included or optional zero-adjustment board. The zero
adjustment can be achieved based on the AC four-terminal method.
Choose two holes on the zero-
1
adjustment board whose
interval is nearly equal to the
distance between the two
terminals on a battery to be
measured.
Press the test lead against the zero-adjustment board in a
2
vertical direction.
Model 9465-10 or
Model L2020 Pin Type Lead
Model 9772 Pin Type Lead
Insert the pins with the marked
(engraved) side of the probe facing you.
51
Adjusting the Zero Point (Zero Adjustment)
HIOKI BT3554F961-00HIOKI BT3554F961-00
• Keep the zero-adjustment board at least 10 cm away from the
instrument.
• Always use the accompanying or optional zero-adjustment
board when performing zero adjustment.
• Insert the pin tips into the hole on the zero-adjustment board
and place each of the SOURCE and SENSE terminals into
contact with the hole on the board. (See the gure below.)
SOURCE
SENSE
• Do not place the zero-adjustment board on top of the battery
or any metal. Electromagnetic induction eect could result in
unstable measurement values. In such a case, keep the zeroadjustment board away from any metal.
• If zero adjustment is performed with the Pin Type Lead short-
circuited at the tips or using a metal sheet other than the zeroadjustment board intended for this purpose, the instrument
cannot adjust the zero point accurately.
• When the distance between the terminals on the battery (target
under measurement) is more than that between the holes on
the zero-adjustment board, use the outermost holes to perform
zero adjustment.
• Consider the zero-adjustment board to be a consumable. It
is recommended to replace it with a new one after the use of
around 700 times.
52
Adjusting the Zero Point (Zero Adjustment)
HIOKI BT3554F961-00HIOKI BT3554F961-00
For Clip Type Lead
Engage the red and black clips together, and then perform zero
adjustment.
Model 9460 Clip Type Lead with Temperature Sensor
(Red)(Black)
Model 9467 Large Clip Type Lead
(Red)(Black)
SOURCE (+)SOURCE (−)
SENSE (+)SENSE (−)
SOURCE (+)SOURCE (−)
SENSE (+)SENSE (−)
53
Adjusting the Zero Point (Zero Adjustment)
HIOKI BT3554F961-00HIOKI BT3554F961-00
Performing zero adjustment
Check if the test lead has been
1
connected properly.
Disconnect the test lead from the
measurement target if connected.
Press the 0ADJ key.
2
The instrument will be put
on standby for acquiring the
correction value.
While the [0AdJ] segment is
3
blinking, short-circuit the test
lead using the zero-adjustment
board.
See “How to short-circuit various
test lead” (p. 51).
Failure to short-circuit the test
lead while the display is blinking
will result in an error.
(Black)
1
Blinks for
about 10s.
For Pin Type Lead
(Red)
2
54
Adjusting the Zero Point (Zero Adjustment)
HIOKI BT3554F961-00HIOKI BT3554F961-00
The instrument will automatically start obtaining correction
values.
When the zero adjustment has been completed, the instrument
will display the [0ADJ] segment and return the display to
measurement mode.
3 mΩ range30 mΩ range
• Keep the test lead short-circuited until the zero adjustment is
completed.
• The zero adjustment will start even if the key is pressed after
the test lead has been short-circuited.
Zero adjustment
completed
300 mΩ range
3
range, voltage range
Ω
55
Adjusting the Zero Point (Zero Adjustment)
HIOKI BT3554F961-00HIOKI BT3554F961-00
Troubleshooting on zero adjustment
Check itemSolution
Has the fuse blown?Check if the fuse has blown. (p. 147)
Do the obtained
correction values
exceed 300 counts in
either resistance or
voltage range?
Did you short-circuit
the test lead properly
while the instrument
was in standby for
obtaining correction
values?
Reconnect the test lead to the instrument.
The test lead may be broken. Replace
the fuse or test lead with a new one.
Remove dirt from the zero-adjustment
board.
While the instrument is in standby for
obtaining correction values (for about
10 s), short-circuit the test lead using the
zero-adjustment board to perform zero
adjustment.
Canceling the zero adjustment
Press the 0ADJ key for at least 1 s with the zero adjustment
function enabled to cancel the zero-adjustment.
Zero adjustment enabledZero adjustment disabled
56
Hold down for at least 1 s.
Using the Hold Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
3.5 Using the Hold Function
This section describes how to freeze measured values on the
display using the hold function. Press the HOLD key. The [HOLD]
segment will appear and the display will freeze measured values.
• When a warning display appears or the voltage readout exhibits
the [----] segments, the instrument cannot freeze the displayed
value.
• When you change any settings, the instrument will disable the
hold function.
• Turning o the instrument will disable the hold function.
Using the automatic hold function can automatically freeze
measured values after they have stabilized.
See “3.6 Automatic Hold Function” (p. 59).
Hold function disabledHold function enabled
Disabling the hold function
Press the HOLD key again to disable the hold function.
57
Using the Hold Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
Freezing measured values using the 9466 Remote
Control Switch
Model 9466
The 9466 Remote Control Switch
(option) is available for the operation
in the same way as when using the
HOLD key.
Disconnect the test lead
1
from the battery under
measurement.
Insert the mini-plug of the 9466
2
Remote Control Switch into the
EXT.HOLD terminal.
Remote Control Switch
NO
Top of the
instrument
Connect the connectors of the
3
test lead to the instrument.
Press the PRESS button on the
4
9466 Remote Control Switch.
The instrument will freeze
measured values.
(Black)(Red)
Disabling the hold function
Press the PRESS button on the 9466 Remote Control Switch or the
HOLD key on the instrument.
58
Automatic Hold Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
3.6 Automatic Hold Function
This section describes how to automatically freeze measured values
after they have stabilized.
Press the A HOLD/MEMO key several times to display the [AUTO
HOLD] segment.
Automatic holdAutomatic memory
Automatic hold and
Disabled
To disable the hold function, press the HOLD key or the PRESS
button on the 9466 Remote Control Switch.
automatic memory
59
Automatic Hold Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
The automatic hold will not work in the following cases:
• When the resistance readout exhibit the [−−−−] segments
• When the [OVER] segment and the maximum display value of
the resistance are blinking
Using the automatic memory function in combination with
the automatic hold function can automatically freeze and
save measurement values.
Canceling the automatic hold function
Press the A HOLD/MEMO key several times to hide the [AUTO
To evaluate whether batteries have deteriorated, rst, measure
the internal resistance of a fresh or non-defective battery, and then
decide the battery-deterioration evaluation values.
Deteriorated batteries will have about 1.5 to 2 times (values for
reference purposes) as high internal resistance, generating only
0.9 times as high voltage as a fresh one. Use these values as
guidelines when determining deterioration evaluation values.
Example of deterioration evaluation values
PASSWARNINGFAIL
Resistance:
Voltage:
The above values will vary depending on the manufacturer and
battery model.
See “1.1 Evaluating Battery Deterioration” (p. 21).
New fuseIn use
Initial value
0.5 m
Ω
Warning
threshold
0.75 m
2.0 V1.8 V
Need
replacing
Rejection value
Ω
1.0 m
Ω
61
Measuring Batteries (Inspection)
HIOKI BT3554F961-00HIOKI BT3554F961-00
3.8 Measuring Batteries (Inspection)
Prepare for measurement. (p. 37)
1
Set the resistance and voltage ranges. (p. 47)
2
Perform zero adjustment. (p. 50)
3
Connect the test lead to a battery under measurement.
4
Place all pins into contact
with the battery.
Read the measured values.
5
62
(Red)
+
(Black)
−
Internal resistance of the battery
Voltage value
Measuring Batteries (Inspection)
HIOKI BT3554F961-00HIOKI BT3554F961-00
Use the measured values to evaluate whether the battery
6
has deteriorated.
Example:
Voltage
Initial
value
(2.0 V)
Present
value
(1.8 V)
PASS
WARNING
Initial value
(0.5 mΩ)
Present value
(1.0 m
As shown above, this battery needs replacing.
Ω
FAIL
)
Resistance
63
Measuring Batteries (Inspection)
HIOKI BT3554F961-00HIOKI BT3554F961-00
To freeze measured
values
To save measured
values.
To load the saved
data to your
computer
To set threshold
values to evaluate
whether a battery has
deteriorated
See “3.5 Using the Hold Function”
(p. 57).
Measurement values can be saved
by pressing the MEMO key while
the display is freezing the values.
See “5.2 Saving Data in the
Memory” (p. 85).
See “7 Communications
Function” (p. 103).
Based on the deterioration
evaluation values, you can set
threshold values to evaluate
whether batteries have
deteriorated.
See “4 Comparator Function
(Evaluation Based on Threshold
Values)” (p. 67).
64
Measuring Batteries (Inspection)
HIOKI BT3554F961-00HIOKI BT3554F961-00
Measurement error
Even if the [−−−−] segments appear and the [OVER] segment
blinks on the display (at the same time, the maximum display value
blinks), this does not indicate an error.
[−−−−] • If the resistance readout exhibits the [----] segments,
Blinking the
[OVER] segment
and maximum
display value
IMPORTANT
When measuring contact resistances of relays or connectors, be
aware that the instrument will generate an open-terminal voltage
of 5 V at maximum. The open-terminal voltage may damage the
oxidized coating on the contacts of targets under measurement,
leading to incorrect measurements.
the test lead may be open-circuited.
Otherwise, the instrument cannot perform
measurement because of a failure, such as no current
ows due to a broken test lead.
• The test lead may not be connected correctly to the
target under measurement.
• The resistance of the target under measurement
signicantly exceeds the measurement range.
• Resistance, voltage, or temperature may exceed each
measurable range.
Warning display
If an overvoltage is input, the instrument will show the [OVER]
segment and the blinking maximum display value on the display,
backlight the display in red, and produce beeps.
65
Measuring Temperature
HIOKI BT3554F961-00HIOKI BT3554F961-00
3.9 Measuring Temperature
Use the 9460 Clip Type Lead with Temperature Sensor (option) to
measure battery temperature.
Use the 9451/9451S Temperature Probe (option) to measure an
ambient temperature.
See “Options” (p. 3).
Connect the connectors
1
of the 9460 Clip Type Lead
with Temperature Sensor to
the instrument.
Connect the mini-plug of the
2
9460 Clip Type Lead with
Temperature Sensor to the
TEMP.SENSOR terminal.
Connect the mini-plug of the
9451/9451S Temperature
Probe to the TEMP.SENSOR
terminal.
The instrument will detect
the temperature sensor and
automatically display the
temperature.
You can toggle between the two temperature measurement units:
Celsius and Fahrenheit.
See “How to toggle the temperature measurement units” (p. 31).
(Black)(Red)
NO
Top of the
instrument
66
Comparator Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
(Evaluation Based on
4
Threshold Values)
4.1 Overview
The comparator function compares measured values of batteries
with the preset threshold values to evaluate batteries on a threelevel basis consisting of PASS, WARNING, and FAIL.
Comparator threshold values
Set the resistance warning threshold, resistance rejection value,
and voltage warning threshold. Up to 200 comparison conditions
can be set.
For information about how to set the thresholds, see “1.1 Evaluating
Battery Deterioration” (p. 21).
Comparator buzzer
The instrument with the default settings will produce beeps when
the comparator yields the WARNING or FAIL result.
See “4.4 Setting the Comparator Buzzer” (p. 78).
Voltage
PASS
Initial value
Warning
threshold
WARNING
FAIL
FAIL
Initial
value
Warning
threshold
Rejection
value
Resistance
67
Enabling the Comparator Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
4.2 Enabling the Comparator Function
1
2
3
When the comparator function is enabled, the measurement
range will switch over to that specied in the comparator setting
with the chosen number.
Press the key.
A blinking comparator number
will appear.
Press the COMP key again to
return to normal measurement
mode.
Choose a comparator
number.
You can choose a number from
between 1 and 200.
Conrm your entry.
The comparator function will be
enabled.
68
Setting Threshold Values for the Comparator
HIOKI BT3554F961-00HIOKI BT3554F961-00
4.3 Setting Threshold Values for the
Comparator
This section describes how tot set threshold values for the
comparator (resistance warning threshold, resistance rejection
value, voltage warning threshold).
You can set the threshold values through GENNECT ONE
or GENNECT Cross.
See GENNECT ONE (contained in the accompanying CD)
or User Manual on GENNECT Cross.
Example: Threshold values for a battery that initially has
an internal resistance of 0.4
a voltage of 2 V.
Resistance warning threshold: 0.6 Ω (1.5 times the initial value)
Resistance rejection value: 0.8 Ω (2 times the initial value)
Voltage warning threshold: 1.8 V
*: The initial values mean the resistance value a fresh battery or
non-defective battery has and the initial voltage value such a
battery can generate.
and generates
Ω
69
Setting Threshold Values for the Comparator
HIOKI BT3554F961-00HIOKI BT3554F961-00
Choosing a comparator number
1
2
3
Hold down the key for at least
1 s.
A blinking comparator number
will appear.
Press the COMP key again to
return to normal measurement
mode.
Choose a comparator
number.
You can choose a number from
between 1 and 200.
Conrm your entry.
The display will return to range
setup mode.
70
Setting the range
HIOKI BT3554F961-00HIOKI BT3554F961-00
1
Choose a resistance range.
(The decimal point position will
move.)
Setting Threshold Values for the Comparator
2
3
Choose a voltage range.
(The decimal point position will
move.)
Conrm your entry.
The resistance warning
threshold and the [WARNING]
segment will blink.
71
Setting Threshold Values for the Comparator
HIOKI BT3554F961-00HIOKI BT3554F961-00
Setting the threshold values
1
2
3
4
5
6
Set the resistance warning
threshold.
Conrm your entry.
The resistance rejection value
and the [FAIL] segment will
blink.
Set the resistance rejection
value.
Conrm your entry.
The voltage warning threshold
and the [WARNING] will blink.
Set the voltage warning
threshold.
Conrm your entry.
72
Setting Threshold Values for the Comparator
HIOKI BT3554F961-00HIOKI BT3554F961-00
7
Set the voltage evaluation
method.
When you have chosen the
[PoL], reversely connecting the
red and black wires of the test
lead will cause the [WARNING]
result.
[WARNING]
(Black)(Red)(Red)(Black)
Voltage
evaluation
method
[AbS]
(Default
setting)
[PoL]
Comparison method
Evaluates absolute voltage
values, regardless of whether
they are positive or negative.
The [WARNING] result will be
given for the negative voltage
value.
If the test lead is placed into
contact in reverse polarity with
a battery (the red and black
probes with the negative and
positive terminals, respectively),
the [WARNING] result will be
given.
Data to be
saved
Signed
(minus sign
only)
Signed
(minus sign
only)
You can set the evaluation method using version 1.8 or
later of GENNECT Cross.
73
Setting Threshold Values for the Comparator
HIOKI BT3554F961-00HIOKI BT3554F961-00
8
If you set the resistance rejection value that is less than the
resistance warning threshold, the warning threshold will change
to the same value as the resistance rejection value.
Conrm your entry.
The display will return to
measurement mode with the
comparator function enabled.
The settings are now saved.
74
Setting Threshold Values for the Comparator
HIOKI BT3554F961-00HIOKI BT3554F961-00
When a PASS judgment is given
Voltage
Initial value
Voltage
warning
threshold
PASS
Measured
value
WARNING
FAIL
FAIL
When a WARNING result is given
Voltage
Initial value
Voltage
warning
threshold
When a FAIL judgment is given
Voltage
Initial value
Voltage
warning
threshold
Initial
value
PASS
WARNING
Initial
value
PASS
WARNING
Initial
value
Resistance
warning
threshold
Measured
value
Resistance
warning
threshold
Resistance
warning
threshold
Resistance
rejection
value
FAIL
FAIL
Resistance
rejection
value
FAIL
Measured
FAIL
value
Resistance
rejection
value
Resistance
Resistance
Resistance
75
Setting Threshold Values for the Comparator
HIOKI BT3554F961-00HIOKI BT3554F961-00
Comparison table for the comparator
The instrument will show a result and produce beeps as shown in
the following table:
Resistance
warning threshold
Resistance
rejection value
Resistance
Voltage (high)PASSWARNINGFAIL
Voltage warning threshold
Voltage (low)WARNINGWARNINGFAIL
The boundary conditions are as follows:
Resistance
PASS
Voltage
WARNING
Resistance
≤
warning
threshold
Voltage
<
warning
threshold
76
(low)
<
≤
Resistance
WARNING
Voltage
PASS
Resistance
(medium)
Resistance
≤
rejection value
Resistance
(high)
Resistance
<
FAIL
Setting Threshold Values for the Comparator
HIOKI BT3554F961-00HIOKI BT3554F961-00
Examples of how to read the comparator output table
Example 1:
If the measured resistance is less than or equal to the resistance
warning threshold, and the measured voltage is more than or equal
to the voltage warning threshold, the [PASS] segment will appear.
Example 2:
If the measured resistance exceeds the resistance warning
threshold but is less than or equal to the resistance rejection value,
and the measured voltage exceeds the voltage warning threshold,
the instrument will show the [WARNING] segment and produce
beeps.
When the resistance warning threshold and resistance are set to the same
value, the boundary conditions are as shown below:
Resistance
PASS
Resistance
≤
warning threshold
Resistance
=
rejection value
Resistance
<
FAIL
77
Setting the Comparator Buzzer
HIOKI BT3554F961-00HIOKI BT3554F961-00
4.4 Setting the Comparator Buzzer
This section describes how to set the instrument to produce beeps
according to the comparison result. The comparator buzzer can be
set in the following states: By default, the instrument has been set
to produce beeps when the comparator gives a WARNING result or
a FAIL judgment.
Disabled
Sounds for
PASS
Sounds
for FAIL
judgments
The instrument will not produce beeps regardless of
comparison results.
The instrument will produce beeps when the comparator
gives a PASS judgment.
The instrument will backlight the display in red and
produce beeps when the comparator gives a WARNING
result or a FAIL judgment.
When you press the key, the present comparator buzzer setting
will appear. Press the key repeatedly to cycle through the settings.
Disabled
Sounds for PASS
After a certain period of inactivity, the instrument will conrm your
entry and the display will return to measurement mode.
The key tone cannot be set.
78
Sounds for FAIL
Canceling the Comparator Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
4.5 Canceling the Comparator Function
When the comparator is enabled, pressing the COMP key can
disable the comparator function.
Comparator function enabledComparator function disabled
• The range keys cannot be used while the comparator function
is enabled.
• If there are no measurement values, the [−−−−] segments will
appear and the comparator will not work.
• Even when the instrument is turned o, the comparator settings
are saved; the comparator will be enabled when the instrument
is turned on again.
79
Canceling the Comparator Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
80
Memory Function
HIOKI BT3554F961-00HIOKI BT3554F961-00
5
5.1 Overview
The instrument can store up to 6000 data sets consisting of
presently measured values*.
After measurement, you can browse and send saved data to your
computer or mobile device.
The internal-memory structure is as follows:
*: Date and time, resistance, voltage, temperature, comparator
threshold values, and comparison results
Memory structure
Unit label (12 units)Memory number (500 cells)
A001. . .499 500
B001. . .499 500
C001. . .499 500
D001. . .499 500
(Display example)
E001. . .499 500
F001. . .499 500
G001. . .499 500
H001. . .499 500
J001. . .499 500
L001. . .499 500
N001. . .499 500
P001. . .499 500
81
Overview
HIOKI BT3554F961-00HIOKI BT3554F961-00
Prole information
The instrument can be populated 100 sets of prole information
(comments, such as location information and device information,
and the battery numbers of batteries under measurement). You
can save measured data in connection with prole information
previously registered.
Concept for registering prole information in the memory
Prole information
Prole number*
Location information*
Device information*
Battery number*
*1: Can be chosen from between 1 and 100.
*2: Any comment can be registered. Strings at most 72 bytes
*3: The number assigned to the battery under measurement. Can be chosen
from between 1 and 500.
Prole number
1
2
3
4
5
1
2
2
3
Memory number
A.001
B.001
C.001
D.001
E.001
1
HIOKI 1F UPS ROOM
Measured data to
be saved
UPS 1-1Memory number
Start1
End50
HIOKI 1F UPS ROOM
A.050
B.050
C.050
D.050
E.050
A.001
A.050
82
Overview
HIOKI BT3554F961-00HIOKI BT3554F961-00
You can save measured data associating with prole information
previously registered. This makes the control of information,
including the UPSs equipped with batteries under measurement and
their locations, easier.
Concept of a data set to be saved
Prole
information
Measured dataMemory numberA.001
Prole number1
Location informationHIOKI 1F UPS ROOM
Device informationUPS 1-1
Battery number1
Date and time2020/4/20 13:00:00
Resistance value
Voltage value
Temperature
Comparator threshold
values
Comparison resultPASS, WARNING, or FAIL
(yyyy/mm/dd hh:mm)
m
.
.
.
, mΩ, V
m
Ω
Ω
V
°C
When you choose a memory number that has already been
assigned to a prole information set, the [PROFILE] segment will
appear.
83
Overview
HIOKI BT3554F961-00HIOKI BT3554F961-00
How to register prole information
ComputerUse GENNECT ONE to register.
The prole information can be transferred via
the USB interface.
Smart phone /
Tablet
• Operating the instrument only cannot register or delete prole
information.
• Details of prole information transferred to the instrument
cannot be seen on the instrument’s display.
To transfer prole information registered in the
instrument to your computer or start phone
You can load the information on GENNECT ONE or
GENNECT Cross.
See GENNECT ONE (contained in the accompanying CD)
or User Manual on GENNECT Cross.
Use version 1.8 or later of GENNECT Cross
to register.
The prole information can be wirelessly
transferred.
The instrument requires the Z3210 Wireless
Adapter to be connected.
84
Saving Data in the Memory
HIOKI BT3554F961-00HIOKI BT3554F961-00
5.2 Saving Data in the Memory
Pressing the MEMORY key stores the present measurement values.
For information about handy features, see “5.3 Automatic Memory
Function” (p. 87)
1
2
3
4
Enable the memory function.
Choose a memory number.
After a certain period of
inactivity, the instrument will
conrm your entry and return
the display to measurement
mode. When the memory
function is enabled, you can
always choose a memory
number.
Conrm your entry.
[USED]
[PROFILE]
Freeze the measured values.
When showing the [−−−−]
segments on the display, the
instrument cannot freeze the
values.
Measured values has already been saved.
(The data will be overwritten)
Prole information has already been
registered.
85
Saving Data in the Memory
HIOKI BT3554F961-00HIOKI BT3554F961-00
5
• Holding down the CLEAR key for at least 1 s can delete the
data saved most recently. However, you can delete immediately
after the data has been saved only.
Every time you hold down the key for at least 1 s, the
instrument will delete the data that is the latest at the time and
return the display to the rst memory number in the unit to
display the contents.
• While the measurement recording prompt is operating, the
instrument will return the memory number corresponding to the
rst battery number to display the contents.
Save measured values.
The instrument will assign
the chosen number to the
measured values and save
them.
After the data has been saved,
the subsequent available
memory number will appear.
The hold function will be
disabled.
86
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.