Thank you for purchasing the HIOKI "9605-01 HARMONIC
MEASUREMENTS UNIT." To obtain maximum performance from the
product, please read this manual first, and keep it handy for future reference.
・ The 3194 must be upgraded to support the 9605-01.
NOTE
・ This product is designed for installation in the 3194 MOTOR/HARMONIC
HiTESTER to enable harmonic analysis. For details on operation of the
3194, please refer to its instruction manual.
・ This product is a factory-fitted option. When the 3194 product is powered
on, you can check the installation on the screen. (Same as the 9605.)
・ For operating environment, maintenance, and disposal at end of life, the
same conditions apply as to the main 3194 product.
・ Because of differences in measurement principle, frequency response, and
accuracy, the values measured by the 9605-01 (effective voltage value,
effective current value, active power, and phase difference) may not agree
with values measured by 9600/9601/9602 products installed in the same
3194 product.
・ The 9605-01 does not save analysis data if there is a power failure. All data
values from before the power failure are treated as zero.
・ The D/A output is not possible for the analyzed data by the 9605-01.
i
Beforeuse
Before using the product, inspect it and check the operation to make sure
that the product was not damaged due to poor storage or transport
conditions. If damage is found, contact your dealer or HIOKI representative.
This product is designed to conform to I EC 61010 Safety Standards, and
DANGE
has been thoroughly tested for safety prior to shipment. However,
mishandling during use could result in injury or death, as well as damage
to the product. Be certain that you understand the instructions and
precautions in the manual before use. We disclaim any responsibility for
accidents or injuries not resulting directly from product defects.
This Instruction Manual provides information and warnings essential for
operating this equipment in a safe manner and for maintaining it in safe
operating condition. Before using this equipment, be sure to carefully read
the following safety notes.
Read the Instruction Manual supplied with the 3194 product very carefully,
and follow the indications given under "
CAUTION
"
," and "
NOTE
."
DANGER
WARNING
," "
,"
Safety Symbols
This symbol is affixed to locations on the equipment where
the operator should consult corresponding topics in this
manual (which are also marked with thesymbol) before
using relevant functions of the equipment.
In the manual, this mark indicates explanations which it is
particularly important that the user read before using the
equipment.
The following symbols are used in this Instruction Manual to indicate the
relative importance of cautions and warnings.
Indicates that incorrect operation presents extreme danger of
DANGE
accident resulting in death or serious injury to the user.
Indicates that incorrect operation presents significant danger
ARNING
CAUTION
of accident resulting in death or serious injury to the user.
Indicates that incorrect operation presents possibility of
injury to the user or damage to the equipment.
NOTE
Denotes items of advice related to performance of the
equipment or to its correct operation.
Installing the 9605-01
MOTOR/HARMONIC HiTESTER enables voltage, current, and power
harmonic analysis to be carried out on the voltage and current waveforms
captured by an input unit, for single phase to three-phase four-wire lines. It
also provides list, a range of display options, including graph and vector
displays.
Installing the 9605-01 HARMONIC MEASUREMENTS UNIT in a 3194
MOTOR/HARMONIC HiTESTER adds the following functions to the basic
functions of the 3194.
(1) Support for range of power lines
Harmonic analysis of voltage, current, and power can be carried out on lines
from single phase to three-phase four-wire. If the frequencies are the same,
up to three single phase lines can be analyzed simultaneously.
(On screen; however, the amount of data displayed on one page is limited to
50 records)
(2) A maximum of 3000th of analysis
A maximum of 3000th of high frequency analysis on basic frequency is
possible. Analysis of the secondary inverter including carrier frequency can
also be done.
(3) High frequency analysis using external synchronization signals
High frequency analysis using external synchronization signals as the base
standard is possible. For this reason, measurement of the phase change
amount against the base standard signals can also be done.
(4) Range of display options
The amplitude value, proportions, phase angle, and distortion of each
harmonic for voltage, current, and power can be found. The analysis results
can be displayed in a list, as a bar graph, or as a vector display.
(5) Floppy disk support
The analysis data can be saved to the floppy disk drive in the 3194, and a
periodic automatic save function is also available.
(6) Built-in printer (option)
This can be used to print required data, and for periodic automatic printing.
There is also a screen copy function.
For details on how to operate keys of the 3194, refer to Chapter 2 "Name
and Functions of Parts" in the instruction manual of the 3194.
2.2Screen Configuration
When the 9605-01
3194, the following screen is added to the 3194 functionality.
Refer to Section 2.3.1, "Screen Configuration" in the instruction manual of
the 3194.
This chapter describes the function settings when using the 9605-01.
For other functions, refer to the Instruction Manual supplied with the 3194
product.
NOTE
The following settings have no effect on harmonic analysis.
Selecting RMS/MEAN value, setting the phase polarity discrimination filter,
switching the waveform peak value, setting the response, setting the equation
for reactive power/apparent power, and setting the indications for out-of
range inputs
3.1Setting the Wiring Mode (1P2W to 3P4W)
Set the wiring mode of the 3194.
For details, refer to Section 4.1, "Setting the Wiring Mode" in the instruction
manual of the 3194.
You can select three input unit channels for harmonic analysis from the
input units (maximum three channels) in the 3194. If the frequencies are the
same, analysis on three single-phase channels is possible.
NOTE
1. Press the
STATUS
to move the cursor to "
key, then use the
HARM
," to display the
PAGE
keys
harmonic measurement setting screen. This
displays the selected wiring mode.
2. Use the
CURSOR
keys to select the "
WIRING
"
item.
3. Move the cursor to the channel to be analyzed
with the
CURSOR
keys, then press theF1"
SET
key to complete the setting.
・ The channel selection is always of adjacent input units, corresponding to the
wiring modes set on the 3194 product. If the 3194 product wiring mode is
set to 3P3W for each of channels 1 and 2 and channels 3 and 4, then the
selection becomes channels 1, 2, and 3. In this case, the third channel is
treated as single phase for analysis.
・ Up to three channels can be selected; a single-channel analysis specification
A PLL source is required for analysis, and may be provided in one of the
following three ways.
(1) Input voltage or current serves as the source:
Waveforms are sampled at a frequency synchronized with the measured
voltage or current waveform to facilitate correct measurement.
(2) Internal fixed clock [Clock] frequency serves as the source:
Use this options when the input signal is not suitable as a source of PLL
synchronization, and the fundamental frequency of the signal for analysis is
50 Hz. In this case, measurement accuracy cannot be assured at frequencies
other than integer multiples of 50 Hz.
(3) An external sync signal [Ext(CH6) or Ext(Con)] serves as the source:
PLL-synchronized measurements for harmonic analysis can be made using an
external signal other than the input voltage or current waveform. Refer to
Section 4.7, "Measurement using an External Sync Signal."
NOTE
The PLL indicator at the upper right of the screen appears when the PLL is
unlocked, and is not visible when the PLL is locked. The source setting for
frequency measurement shares the same indicator.
1. Press the
STATUS
to move the cursor to "
key, then use the
HARM
," to display the
PAGE
keys
harmonic measurement setting screen.
2. Moving the cursor to "
CURSOR
keys displays the settings available for
PLL SOURCE
" with the
the PLL source in a window.
3. Use the
F1 "↓"
and
F2 "↑"
key to specify the
required item.
・ If the signal selected for the source setting is very distorted, if the level is
low for the range setting, or the signal frequency is not stable, the PLL
circuit may not function. In such cases accurate analysis is not possible.
・ If the three channels selected in Section 3.2 are separate systems with
different signal frequencies, the analysis results are only valid for the channel
for which the PLL source is set.
For the channel of which the analysis results are shown on the screen only,
press the
SHIFT
key then use the
It is also possible to change the setting in the STATUS/Units screen.
Refer to Section 4.2, "Setting the Coupling Mode" in the instruction manual
of the 3194.
CURSOR
key to switch.
NOTE
・ When AC mode (coupling mode) is selected on the 3194 product, a DC
blocking filter is required on the input unit. Therefore, if the measurement
frequency is 10 Hz or below, or the waveform has a superimposed DC
component (e.g. a half-wave rectified signal), errors may become large.
・ Analysis results in the DC mode and analysis results in the AC+DC mode
are the same.
・ When the AC+DC or DC mode is selected, a value is shown with a
harmonic index of "0." This shows the DC component included in the
analysis data. However, the polarity is not displayed for the voltage and
current since the effective value is converted.
Zero adjustment must be carried out after warming up.
NOTE
1. Press the
STATUS
to move the cursor to "
key, then use the
HARM
," to display the
PAGE
keys
harmonic measurement setting screen.
2. Move the cursor to "
3. Press theF5"
ZERO ADj
WIRING.
"
" key to start zero
adjustment.
・ The operation is carried out for 3 channels simultaneously.
・ The operation takes about 3 seconds.
・ Carry out the operation with no inputs on either voltage or current.
・ Carry out the operation after degaussing (DMAG) when using the 9600 or
By setting an external trigger, analysis can be started when the trigger signal
is input. One window of analysis data is displayed, then the process stops.
By synchronizing the timing with an external device, a number of 3194
products can perform analysis simultaneously.
NOTE
1. Press the
STATUS
to move the cursor to "
key, then use the
HARM
," to display the
PAGE
keys
harmonic measurement setting screen.
2. Use the
CURSOR
keys to select the "
EXT TRIG
item.
3. Press theF2"ON" key, putting the product on
hold for a trigger.
When a trigger signal is input from the outside,
analysis starts.
・ The analysis always starts from a point at which the PLL source waveform
crosses the zero value, and there is therefore a delay of a maximum of one
cycle of the waveform with respect to the trigger signal.
・ This does not affect data on the 3194 product (other than the 9605-01).
・ The PLL should be locked.
Measurements are sorted by the specified parameter, and other parameters
are sorted for the respective degree.
Unsorted
NOTE
Sorted
The graph, list and vector displays are by harmonic order rather than by
degree. Therefore, in the vector display case, multiple degrees can be
displayed on the same screen.
.
First refer to Chapter 3 "Preparations for Measurement" in the instruction
manual of the 3194. Pay special attention to the precautions listed in this
chapter.
The 9605-01 HARMONIC MEASUREMENTS UNIT applies FFT (FastFourier Transform) analysis to waveforms from the A/D converter for each
window, and can analyze voltage, current and power harmonics. Up to 3,000
degrees can be analyzed, and fundamental frequencies from 10 Hz to 4.5
kHz are supported.
NOTE
・ Because of differences in measurement principle, frequency response, and
accuracy, the values measured by the 9605-01 and 9600/9601/9602 products
installed in the same 3194 product may not agree. Similarly, values may not
agree with measurements made by other devices.
・ If the PLL synchronization range is 10 Hz to 17.5 Hz or 17.5 Hz to 35 Hz,
then the anti-aliasing filter of Fc ≒ 15 kHz is inserted. If the PLL
synchronization range is 35 Hz or more, then the anti-aliasing filter of Fc
≒ 120 kHz is inserted.
・ Although harmonics can be analyzed at up to 3,000 degrees, only a
maximum of 50 items can be displayed on the screen.
・ Because items not displayed are calculated internally at the same time, by
switching to the hold state and switching the screen, it is possible to read
simultaneous values for other analysis data.
・ Be sure to carry this out with the PLL in the locked state.
・ The harmonics which can be analyzed depend on the frequency of the PLL
source. For details see the Specifications.
・ Color specifications are not possible.
・ The zeroth index is displayed when using a 9600/9602 with AC+DC or DC
mode selected. This zeroth index is the DC component of the analyzed
waveform. In the voltage/current mode, effective values are computed, and
there is therefore no polarity. The accuracy of the zeroth index is not
specified.
・ When using 3P3W/3V3A mode, analysis results for power signify the total
values (ΣP). The power analysis results are shown for each channel, but
these have no significance. The total value (ΣP) is the same result for 3P3W
and 3V3A.
・ The phase angle indicates the phase angle of all data, taking the phase angle
of the fundamental from the PLL source as 0。. For the total value (ΣP) , the
phase angle is not displayed.
・ The phase difference indicates the phase difference of the current with
respect to the voltage on the same channel for each harmonic.
・ At less than 0.1% of range, internal circuit noise may appear in the
display.
・ When the rms value of voltage or current exceeds 130% of range, or when
the peak value of voltage or current exceeds crest factor, the value of
analysis data and power analysis data is shown in red.
Effective voltage value,
effective current value,
active power
Fundamental
(component, level)
Harmonic (component, level)
Harmonic content
Harmonic phase angle
Harmonic voltage phase
angle, harmonic current
phase angle
Harmonic voltage-current
phase difference
Total harmonic
Found from waveform sampling at a sampling
frequency determined from the frequency of the
fundamental. Includes harmonics above the analyzed
number.
Fourier coefficient of fundamental component
Fourier coefficient second and subsequent harmonics
Taking the fundamental component level as 100%,
indicates the other harmonic components as
proportions.
Generic term for harmonic voltage phase angle,
harmonic current phase angle, and harmonic voltage
current phase difference
Taking the phase angle of the fundamental from the
PLL source as 0。, indicates the phase difference of
other voltage and current harmonics.
The phase difference between the voltage and current
for a particular harmonic
The total value of the harmonics being analyzed (from
2nd to limit of analysis)
Total harmonic distortion
PLL (Phase Locked Loop)
Anti-aliasing filter
THD-F: total harmonic content as proportion of the
fundamental.
THD-R: total harmonic content as proportion of the
effective voltage (current) value
This generates a clock signal synchronized to the
fundamental component of the measured waveform.
The waveform is sampled using this clock signal. If
the PLL does not function correctly, accurate
harmonic analysis is not possible.
When digital sampling is applied to signal waveforms,
if over half of the frequency composition of the
sampling frequency is included in the measurement
signals, repeated skew (area sync) occurs and so
correct frequency analysis cannot be done. This
problem can be prevented by inserting a low pass
filter that cuts the sampling frequency at the front
section of the A/D converter by half. This filter is
called an anti-aliasing filter.
key.
Move the cursor to the display item box, and
use the
F1 "↑"
and
F2 "↓"
keys to make a
selection.
After specifying, press the
F5 "RETURN"
key.
(4) Selecting the display analysis information
Select the information from the analysis from Amplitude, Proportion, and
Phase.
NOTE
Press the
F5 "SELECT"
key.
Move the cursor to the analysis item box, and
use the
F1 "↑"
and
F2 "↓"
keys to make a
selection.
After specifying, press the
F5 "RETURN"
・ This specification affects the list display in all screens.
・ Selecting "
PHASE
" has the following significance:
For voltage (U) or current (I), display the phase angle of the fundamental
waveform with respect to the PLL source.
For power (P), display the voltage-current phase difference
Select a linear (LINEAR) or logarithmic display (LOG) for the vertical axis.
NOTE
Press the
F5 "SELECT"
key.
Move the cursor to the LINEAR/LOG box, and
use the
F1 "↑"
selection. After specifying, press the
and
F2 "↓"
keys to make a
F5 "RETURN"
key.
It is not possible to make individual LINEAR/LOG selections. The vector
display also changes accordingly.
(6) Data read-out with the cursor
On a displayed graph, you can use the cursor to
select the harmonics to be read. Use the
keysandto move the cursor ▼ on the
screen, and display the selected data.
The harmonic list display shows the amplitude value, proportion, phase
angle, and distortion for each harmonic of voltage, current, and power. There
are two display formats: format 1 and format 2.
Format 1Format 2
Only the analysis data for the channel synchronized to the PLL source is
valid.
When sorting is enabled, display order is by descending amplitude.
(1) Display procedure
In harmonic mode, press theF2"
LIST
" key, to switch to the List screen.
(2) Selecting the display format (two methods)
Method 1
Press the
to cycles through the format.
Method 2
Press the
Move the cursor to the display item box, and
use the
selection. After specifying, press the
key.
Specify information from the analysis to be displayed.
NOTE
Press the
F5 "SELECT
"key.
Move the cursor to the display item box, and use
F1 "↑"
the
After specifying, press the
and
F2 "↓"
keys to make a selection.
F5 "RETURN"
key.
・ This specification affects the list display in all screens.
・ Select the analysis information from the amplitude, proportion, and phase
angle.
・ Selecting "
PHASE
" has the following significance:
For voltage (U) or current (I), display the phase angle of the fundamental
waveform with respect to the PLL source.
For power (P), display the voltage-current phase difference
・ For the total value (ΣP) , the phase angle and total harmonic distortion are
The harmonic vector display shows the voltage, current, and phase angle for
each harmonic, making clear the phase relationship between voltage and
current. The numeric values of the displayed harmonics also appear.
For details on wiring mode and vector display, see Appendix.
NOTE
・ It is only possible to display the voltage and current for a single system on
the screen. Again, only one harmonic can be selected for display. Therefore,
for a single phase system two lines (voltage and current) appear, and for a
three-phase four-wire system, six lines (voltage ×3, current ×3) are
displayed.
・ By setting the hold state, and switching the screen, it is possible to read
other analysis data for the same time instant.
・ The distortion is not displayed.
・ The voltage/current phase angle indicates the phase angle taking the phase
angle of the fundamental from the PLL source as 0。.
・ The power phase difference indicates the phase difference of the current with
respect to the voltage on the same channel for each harmonic.
・ This reflects the harmonic selected on the harmonic graph display.
・ When sorting is enabled, the same degree may not be displayed.
The waveform display shows one cycle of the voltage and current waveforms
being measured. There are five formats.
Format 1
Shown the waveforms
(voltage or current) for each
system being measured
together.
Format 2
Shown the waveforms
(voltage or current) for each
system being measured
together. The display is
smaller than the graph of
format 1, but the effective
and peak values are shown.
Format 3
The display is divided into
two, and the voltage
waveform (above) and
current waveform (below)
for the system being
measured are shown
together.
Voltage and current
waveforms can be zoomed.
The zoom area can be
panned by the left and right
cursor keys, and the amount
of magnification set by the
up and down keys. Pressing
theF4key switches between
UandI.
10 to 70 HzX 16 max.
70 tp 140 HzX 8 max.
140 to 280 HzX 4 max.
280 to 560 HzX 2 max.
NOTE
560to4500HzX 1 max. (can not be changed)
FFT
All voltage and current
waveform analysis data can
be graphed. The upper part
is voltage and the lower part
is current. Degree and
frequency can be read by the
CURSOR
andkeys.
・ The cycle displayed in one screen is different depending on the PLL
synchronization range. See Chapter 9, "Specifications" for more information.
・ The waveform zero-crossing point is not correlated with frequencies.
・ The waveform starts from a point at which the voltage waveform (or current
waveform) specified as the PLL source waveform crosses the zero value, and
displays the first cycle from the window being analyzed. The other waveforms
are displayed for the same time interval as this waveform.
・ It is not possible to select individual items for display.
・ Because items not displayed are calculated internally at the same time, by
switching to the hold state and switching the screen, it is possible to
display simultaneous parts of other waveforms.
・ The power waveform is not displayed.
・ Only waveforms for channels synchronized to (at the same frequency as)
the PLL source channel are valid.
・ Cursors are not available for data readout.
・ Effective power is not displayed.
・ The vertical axis is fixed as LOG.
・ FFT format can not be selected when sorting is enabled.
・ The vertical axis is fixed as 100% when the FFT format is selected.
External signals can be synchronized with PLL for high frequency analysis
of voltage, current, and electric power. Also, measurement of the phase
amount change of voltage/current based on external signals is possible.
4.7.1Theory of Operation
If the pulse count is a multiple of the fundamental waveform frequency,
calculation is based on the rising edge of the pulse nearest the rising edge
zero crossing of the U1 fundamental waveform component.
Pressing the
standard to zero.
Single pulse
MEAS
key forces the phase difference (θ0) between U1 and the
・ The input connector for external synchronization signals is not insulated
from the main product of the 3194.
・ See the "EXT.CLOCK" item in Chapter 7, "External Control" for
information on the pin assignment of the connector.
・ Use Ext (ch6) when there is a need to insulate external synchronization
signals from the 3194 main product. Because this requires use of a dedicated
product, consult your local dealer for more information on how to obtain this
product.
・ As a condition, external synchronization signals require sine wave in the
range of 1 to 10 Vrms. However, as other distorted waveforms are also
possible, the PLL circuitry many not operate correctly, and measurement
cannot be made as well.
・ The phase difference between external synchronization signals and
voltage/current includes the internal phase difference of θ. Because of this
reason, the accuracy of the phase is determined as the variation amount.
Also, accuracy is determined by the sine wave and frequency ratio (1/1).
・ When the frequency of external synchronization signals changes, or the
signals stop and then the PLL circuitry stops, the phase compensation value
using the MEAS key becomes invalid.
・ The frequency of external synchronization signals must be integer multiples
(from 1 to 255 times) of the frequency of the voltage/current to measure.
The external synchronizing signal must be applied to the EXT CONTROL
terminal before connecting input voltage and current for measurement.
1. Set the PLL source to the voltage measured
with the STATUS screen, and measure the
frequency of the voltage from the "Harmonic"
screen on the MEAS screen.
2. Set the PLL source to [Ext(Con)] from the
"STATUS" screen, to measure the frequency of
the external sync signal on the "MEAS" screen.
3. Divide the external sync signal frequency by the
frequency of the displayed voltage, and enter
the resulting value as the [Ext Divide]. For
example, if the frequency of the displayed
voltage is 50 Hz and the frequency of the
external sync signal is 300 Hz, enter 1/6 as the
Before Modification
[Ext Divide].
After Modification
4. From this state, return to the "MEAS" screen
and press the
MEAS
key again to phase lock
with the synchronizing signal by zeroing out
any UI phase difference, and to normalize the
display.
Press the panel
this case, by switching the screen selection, you can view harmonic data for
the same time interval which was not displayed.
Since internally the measurement continues, each time you press the
key the values at that time are displayed. To end the hold function, hold
down the
The following measurements are also possible.
When combined with the interval timer, the results of analysis at the end of
each interval can be displayed.
・ The hold function applies to the whole 3194 product.
Refer to the following sections in the instruction manual of the 3194.
Section 6.1, "Hold Function", and 7.2, "Setting the Control Time."
・ Averaging is not supported.
・ Peak Hold function is not supported.
・ If using the sort function, output is in maximum sort order but the hamonic
order are not output.
SHIFT
key to freeze the display of all items on the screen. In
7. Press the
delete the item which has been set, press the
"OFF"
8. Pressing the
F2
key.
"ON"
key to make a setting. To
F3
"LINE"
key sets the items on
F1
the specified line to on or off.
NOTE
9. When the settings are completed, press the
"RETURN"
key.
・ For example, when the output item is set to ODD, and "MIN ORDER" is set
F5
to 1, and "MAX ORDER" is set to 10, this selects the data items for odd
harmonics in the range 1 to 10.
Even if the waveform output is set to ON, the data is not output on a
printer.
・ When sorting is enabled, the data is output in the displayed order.
You can set control times to output automatically at those times.
1. Set the control time.
2. Select the output items.
NOTE
3. Press the
・ For the basic method of using the timer control, refer to the Instruction
START/STOP
Manual for the 3194 product. Section 7.2, "Setting the Control Time"
・ The minimum interval which can be specified is varied automatically,
depending on the number of output items.
6.2.2Relationship Between the Number of Output Items and
key to start operation.
Interval Time
The maximum number of items to be output and corresponding minimum
interval times for output to floppy diskette or printer are determined
automatically as follows.
Number of setting itemsMinimum interval time
FDD only1 to 70,
71 to 400
400 or more
Printer only1 to 60 lines
every 60 lines
FDD+Printer1to60
every 60
NOTE
・ For integration data, one data value is regarded as two values.
・ Waveform data cannot be output to a printer or floppy disk.
Specifications of the floppy disk data for the 9605-01(Harmonic Analysis)
Harmonic data is output appended to the end of a line of ordinary
measurement data. The output sequence is as shown in the table on the
following page. The data for each of the harmonics in the list is output
consecutively. If the time averaging or peak hold functions are active, the
corresponding values (the same as on the screen) are output.
10 s
20 s
(20 s + 10 s)/370 items
10 s
+10 s
Total value of FDD only
and printer only
NOTE
・ The suffixes (n) are output as two digit values (xx), with a leading space
in the case of a single-digit value.
Example "U1( 3),I2(23)"
The channel numbers, such as 1, 2, and 3 in "U1," "U2," and "U3"
change, for example to "U4," "U5," and "U6," depending on the analysis
channel settings.
・ The suffixes (n) output by the sort function indicate the sequence. The
Voltage (U1) rms value
Voltage (U2) rms value
Voltage (U3) rms value
Current (I1) rms value
Current (I2) rms value
Current (I3) rms value
Active power (P1)
Active power (P2)
Active power (P3)
Active power (ΣP)
Voltage (U1) positive peak value
Voltage (U2) positive peak value
Voltage (U3) positive peak value
Current (I1) positive peak value
Current (I2) positive peak value
Current (I3) positive peak value
Voltage (U1) negative peak value
Voltage (U2) negative peak value
Voltage (U3) negative peak value
Current (U1) negative peak value
Current (U2) negative peak value
Current (U3) negative peak value
Voltage (U1) total harmonic distortion ratio (THD-R)
Voltage (U2) total harmonic distortion ratio (THD-R)
Voltage (U3) total harmonic distortion ratio (THD-R)
Current (I1) total harmonic distortion ratio (THD-R)
Current (I2) total harmonic distortion ratio (THD-R)
Current (I3) total harmonic distortion ratio (THD-R)
Voltage (U1) total harmonic distortion ratio (THD-F)
Voltage (U2) total harmonic distortion ratio (THD-F)
Voltage (U3) total harmonic distortion ratio (THD-F)
Current (I1) total harmonic distortion ratio (THD-F)
Current (I2) total harmonic distortion ratio (THD-F)
Current (I3) total harmonic distortion ratio (THD-F)
The n-th harmonic voltage (U1) rms value
The n-th harmonic voltage (U2) rms value
The n-th harmonic voltage (U3) rms value
The n-th harmonic current (I1) rms value
The n-th harmonic current (I2) rms value
The n-th harmonic current (I3) rms value
The n-th harmonic power value (P1)
The n-th harmonic power value (P2)
The n-th harmonic power value (P3)
The n-th harmonic power value (ΣP)
The n-th harmonic voltage (U1) contents
The n-th harmonic voltage (U2) contents
The n-th harmonic voltage (U3) contents
The n-th harmonic current (I1) contents
The n-th harmonic current (I2) contents
The n-th harmonic current (I3) contents
The n-th harmonic power value (P1) contents
The n-th harmonic power value (P2) contents
The n-th harmonic power value (P3) contents
The n-th harmonic power value (ΣP) contents
The n-th harmonic voltage (U1) contents
The n-th harmonic voltage (U2) contents
The n-th harmonic voltage (U3) contents
The n-th harmonic current (I1) contents
The n-th harmonic current (I2) contents
The n-th harmonic current (I3) contents
The n-th harmonic power value (P1) contents
The n-th harmonic power value (P2) contents
The n-th harmonic power value (P3) contents
The n-th harmonic power value (ΣP) contents
The measurement input terminal and chassis of the 3194 are not
DANGE
isolated from each other. Do not exceed the maximum rated working
voltage. Doing so can damage the product or cause a serious accident.
With a connection to the external output connector on the rear panel of the
3194 product, various external control functions are available. The following
controls apply to the 9605-01.
NOTE
・ Be sure to observe the items under the
headings in Chapter 9, "External Output/External Control Terminals" in the
Instruction Manual supplied with the 3194 product.
・ The FDD/PRINTER.START and EXT.HOLD controls are common to overall
3194 operation.
7.1EXT.CLOCK
The PLL sync signal is set to a multiple of the external clock for analyzing
voltage, current and power.
Pin numberTerminal
24FDD/PRINTER. START
25EXT.HOLD
48EXT.CLOCK
49TRIGGER.IN
50TRIGGER.OUT
WARNING,CAUTION
,and
NOTE
NOTE
・ Set the PLL source to [Ext(Con)].
・ Set the [Ext Divide] to match the fundamental voltage frequency to the clock
frequency.
・ Synchronization is not possible with distorted waveforms.
・ Refer to Section 4.6, "Measurement using an External Sync Signal."
When an external control signal is input, a single measurement is made and
displayed. This operation is repeated each time the control signal is input.
This is controlled by triggering at the edge of the input waveform, or
shorting between 48-pin and 47-pin.
NOTE
To use this control, the external trigger setting must be set to "
7.3TRIGGER.OUT
This outputs a clock signal with the same frequency as the signal waveform
selected as the PLL source. For example, when a 50 Hz sine wave is input,
this outputs a 50 Hz clock signal.
NOTE
・ If the PLL circuit is not functioning correctly, normally no wave-shaped
clock signal is output.
・ The duty factor of the clock signal is not specified.
・ The point at which the signal waveform crosses the zero value is not
synchronized to the rising edge of the clock signal.
In order to avoid the possibility of an electric shock, unplug the power
meter's power cord and disconnect the other wiring before
connecting the GP-IB or RS-232 cable to the interface connector.
・ Turn the power off when connecting the personal computer to the
power meter. Connecting or disconnecting cables while the power is on
could damage the equipment.
・ After connecting the GP-IB or RS-232C cable, always be sure to
secure the connection with the screws on the connector.
The 3194 MOTOR/HARMONIC HiTESTER is fitted as standard with a GPIB/RS-232C interface. Using this interface, all of the functions of the
product can be controlled from a personal computer by remote control, for
the acquisition of harmonic measurement data.
This section lists the extra commands added for harmonic measurement.
This section explains only commands added for the 9605-01.
For details on operation of the GP-IB or RS-232C interface, first refer to
Chapter 12, "GP-IB/RS-232C Interface" in the instruction manual of the
This section explains each command in the harmonic analysis mode.
The 9605-01 is only capable of simultaneous analysis on three channels.
Therefore, when channel are specified in a command, regardless of the
channels on which the 9605-01 is operating, these are specified as channels
1, 2, and 3.
For example, if channels 3, 4, and 5 of the 3194 product are used for
analysis by the 9605-01, then the specifications to get measurement values
are "HU1", "HTFI3", and so forth.
:Command
Indicates functions of message reference
Syntax
Response
syntax
Example
: Indicates the command syntax.
<>
: (Data portion)
Indicates the data format for a
command that includes data.
<NR1>= integer data
: Indicated only for commands for
which a response message is returned.
: Shows a simple example illustrating
the usage of the command. All
transmissions are indicated in "short
form."
As the default output items to the
floppy disk drive or printer for the
normal measurement, the level,
percentage, and phase angle for U1
and P1 are specified.
Sets the display item on the harmonic
waveform screen.
The number of graphs displayed may
vary, depending on the number of input
units and the wiring mode.
Queries the display item on the
harmonic waveform screen.
:DISPlay?
Queries the screen displays.
Syntax
Response
syntax
Example
Transmission
Response
:DISPlay?
":DISPLAY <character>"
<character>
H_GRAPH
H_LIST
H_VECTOR
H_WAVE
":DISPlay?"
":DISPLAY H_GRAPH"
:HARMonic:CHANnel
Sets the harmonic analysis channel.
Syntax
Example":HARMonic:CHANnel 1"
:HARMonic:CHANnel <NR1>
<NR1>= 1 to 5
Harmonic analysis for channel 1, 2,
and 3 is carried out.
Function
Note
Function
Note
Queries the current screen displayed.
The harmonic screen is added to the
existing ":DISPLAY" command. There
is no change in the response messages
for the screen.
Specifies the first channel for which
harmonic analysis is to be carried out.
Some channel numbers cannot be
specified as the first channel, depending
on the input unit configuration and
wiring mode of the 3194. In this case,
an execution error occurs.
Obtains the frequency dividing ratio of
the external sync signal.
According to the harmonic waveform
analysis results, selects an object to be
averaged.
Queries the average setting.
Syntax
Response
syntax
Example
Transmission
Response
:HARMonic:LPF?
:HARMONIC:LPF <0/1/2>
":HARMonic:LPF?"
":HARMONIC:LPF 1"
Amplitude has been selected as the
object to be averaged.
:HARMonic:PHASezero
Phase angle zero adjust of external sync signal
Syntax
Example":HARMonic:PHASezero"
:HARMonic:PHASezero
Function
Function
Note
Obtains the object selected to be
averaged.
When measuring with an external sync
signal, the phase angle of U1 is
normalized to zero degrees. Other
measurement values are adjusted relative
to the amount of U1 normalization.
Selects the PLL source of the harmonic
analysis. This selects the PLL source for
harmonic analysis. Analysis is based on
the PLL source selected by this
command.
Some channel numbers cannot be
specified as the first channel, depending
on the input unit configuration and
wiring mode of the 3194. In this case,
an execution error occurs.
Queries the current selecting channel as
PLL source.
:HARMonic:RTC
Sets the harmonic RTC counter.
Syntax
:HARMonic:RTC <NR1>
<NR1> = 0 to 10000
Example":HARMonic:RTC 50"
Sets the harmonic RTC counter to 50.
:HARMonic:RTC?
Queries the harmonic RTC counter.
Syntax
Response
syntax
Example
Transmission
Response
:HARMonic:RTC?
":HARMONIC:RTC <0-10000>"
":HARMonic:RTC?"
":HARMONIC:RTC 50"
Function
Note
Function
Theharmonicdisplayupdatetimingis
counted the specified number of times,
and is indicated in UE bit (bit 0) of
ESR0. When set to 0, this is off.
If the setting value is other than 0 to
10000, an execution error occurs.
Queries the counter value set by the
harmonic RTC counter.
HU1 data is sorted and other
parameters are output for the
respective degree.
:HARMonic:SORT?
Query the sorting basis for harmonic analysis.
Syntax
Response
syntax
Example
Transmission
Response
:HARMonic:SORT?
:HARMonic:SORT <data>
":HARMonic:SORT?"
":HARMonic:SORT HU1"
:HARMonic:TRIGger
Function
Function
Select the basis for sorting harmonic
wave analysis results.
Obtains the sorting basis for the
specified harmonic analysis data.
Enables or disables the trigger mode.
Syntax
Example":HARMonic:TRIGger ON"
:HARMonic:TRIGger <ON/OFF>
:HARMonic:TRIGger?
Queries the trigger mode.
Syntax
Response
syntax
Example
Transmission
Response
:HARMonic:TRIGger?
:HARMONIC:TRIGGER <ON/OFF>
":HARMonic:TRIGger?"
":HARMONIC:TRIGGER ON"
Function
Function
Selects whether or not to switch to the
trigger pending state. When this is ON,
the trigger pending state is entered.
When OFF, the trigger pending state is
left, and the system switches to normal
harmonic analysis.
For details, refer to Chapter 7,
"TRIGGER IN."
Queries whether the current state is the
trigger pending state. When this is ON,
it indicates the trigger pending state.
When it is OFF, it indicates not the
trigger pending state.
Display blank +6666.6E+99
Calculation impossible +7777.7E+99
Input over +9999.9E+99
Character
HU1, HU2, HU3 Voltage rms value
HI1, HI2, HI3 Current rms value
HP1, HP2, HP3, HPSUM Active power
HPUP1, HPUP2, HPUP3 Voltage (+) peak value
HMUP1, HMUP2, HMUP3 Voltage (-) peak value
HPIP1, HPIP2, HPIP3 Current (+) peak value
Current (-) peak value
Voltage total harmonic distortion ratio (rms reference)
Voltage total harmonic distortion ratio (fundamental waveform reference)
Current total harmonic distortion ratio (rms reference)
Current total harmonic distortion ratio (fundamental waveform reference)
Frequency
Function ① Default mode
If no parameters are specified in the
data section, then this mode is used.
Default item data specified by the
":MEASure:ITEM:HARMonic"
command is created. In this case the
data order is fixed.
② Data (parameter) specification mode
If one or more parameters are specified
in the data section, then this mode is
used. Measurement item data specified
by <data> is created.
If data is specified which cannot be
selected, because of the number of input
units or the channels being used for
harmonic analysis, an execution error
results.
The order of arranging the data
(parameters) is freely selectable, and
data is created in the specified order.
Note
・ Up to 70 items can be responded,
however, in the data section
specification mode, the harmonic level,
harmonic proportion, and harmonic
phase angle cannot be obtained. Select
the output items with the
":MEASure:ITEM:HARMonic:..."
command, and get the measurement
values in the default mode.
・ To change the NR3 numerical data
format, see the ":TRANsmit:COLumn"
command.
Headers for harmonic level, harmonic percentage, harmonic phase angle are shown below. When headers
are ON, headers are affixed to all harmonic measurement value. The value of the last two digits of
characters are shown harmonic order.
HU1L00 to HU1L50, HU2L00 to HU2L50, HU3L00 to HU3L50
HI1L00 to HI1L50, HI2L00 to HI2L50, HI3L00 to HI3L50
HP1L00 to HP1L50, HP2L00 to HP2L50, HP3L00 to HP3L50, HPSUML00 to HPSUML50
HU1D00 to HU1D50, HU2D00 to HU2D50, HU3D00 to HU3D50
HI1D00 to HI1D50, HI2D00 to HI2D50, HI3D00 to HI3D50
HP1D00 to HP1D50, HP2D00 to HP2D50, HP3D00 to HP3D50, HPSUMD00 to HPSUMD50
HU1P00 to HU1P50, HU2P00 to HU2P50, HU3P00 to HU3P50
HI1P00 to HI1P50, HI2P00 to HI2P50, HI3P00 to HI3P50
HP1P00 to HP1P50, HP2P00 to HP2P50, HP3P00 to HP3P50
Queries the harmonicdata.
Creates data in the default order
Response
syntax
Example
Transmission
Response
Headers: ON
<header1> <data1 degree>;<data1>;
<header2> <data2 degree>;<data2>; ;
Headers: OFF
<NR3>;<NR3>;<NR3>,...
":MEASure:HARMonic:SORT?"
":HU1L01S 80;1.50E+02"
Note
specified by
"MEASure:ITEM:HARMonic:SORT"
and "MEASure:ITEM:HARMonic:LIST"
commands.
・ Up to 70 items can be responded.
・ Refer to "MEASure:HARMonic?" for
data beside the sort.
・ Select the output items with the
":MEASure:ITEM:HARMonic:SORT...."
Character
HU1L ch1 voltage level
01S First sort position
80 80 degrees
1.50E+02 150 V
or ":MEASure:ITEM:HARMonic:LIST"
command.
・ To change the NR3 numerical data
format, see the ":TRANsmit:COLumn"
command.
・ When sorting is disabled, the data is
invalid.
Headers for harmonic level, harmonic percentage, harmonic phase angle are shown below. When
headers are ON, headers are affixed to all harmonic measurement value. The two numeric digits
before the last character of the string (nnS) indicate the sort order. Up to fifty items can be sorted,
so the last three characters range from 01S to 50S.
Level
Percentage
Phase
angle/
difference
Voltage
Current
Power
Voltage
Current
Power
Voltage
Current
Power
HU1L01S to HU1L50S, HU2L01S to HU2L50S, HU3L01S to HU3L50S
HI1L01S to HI1L50S, HI2L01S to HI2L50S, HI3L01S to HI3L50S
HP1L01S to HP1L50S, HP2L01S to HP2L50S, HP3L01S to HP3L50S, HPSUML01S to
HPSUML50S
HU1D01S to HU1D50S, HU2D01S to HU2D50S, HU3D01S to HU3D50S
HI1D01S to HI1D50S, HI2D01S to HI2D50S, HI3D01S to HI3D50S
HP1D01S to HP1D50S, HP2D01S to HP2D50S, HP3D01S to HP3D50S, HPSUMD01S to
HPSUMD50S
HU1P01S to HU1P50S, HU2P01S to HU2P50S, HU3P01S to HU3P50S
HI1P01S to HI1P50S, HI2P01S to HI2P50S, HI3P01S to HI3P50S
HP1P01S to HP1P50S, HP2P01S to HP2P50S, HP3P01S to HP3P50S
first <NR1> = 0 to 50 (lower limit
order)
second <NR1> = 0 to 50 (upper limit
order)
Example":MEASure:ITEM:HARMonic:ORDer
1,15,ODD"
Sets the default output item to odd
data up to the 15th.
:MEASure:ITEM:HARMonic:ORDer?
Queries the output order of the harmonic data.
Function
Function
Note
Clears all output items set by the
":MEASure:ITEM" command.
Sets the default items (harmonic output
order for the level, percentage, and
phase angle) to be transferred in the
response message to the
":MEASure:HARMonic?" query in the
default mode.
Some harmonics cannot be specified,
depending on the measurement
frequency.
Queries the default items (harmonic
output order for the level, percentage,
and phase angle) to be transferred in the
response message to the ":MEASure?"
query in the default mode.
Sets the default items (harmonic list for
the level, percentage, and phase angle)
to be transferred in the response
message to the
":MEASure:HARMonic?" query in the
default mode.
The item is set as shown below by
setting bits, to specify a single
numerical value.
For the harmonics to be output, it is
necessary beforehand to issue a
":MEASure:ITEM:HARMonic:ORDer"
command.
Queries the output item for the harmonic measurement value.
Sets the default items (rms value, active
power, total harmonic distortion ratio) to
be transferred in the response message
to the ":MEASure:HARMonic?" query
in the default mode.
The item is set as shown below by
setting bits, to specify a single
numerical value.
If the setting value is out of range, an
execution error occurs.
Specifies the (highest) harmonic order of analysis during sorting.
Syntax
:MEASure:ITEM:HARMonic:SORT
Function
<NR1>
<NR1> = 0 to 50
Example":MEASure:ITEM:HARMonic:SORT 5"
Specify sorted results for the 1st
through 5th harmonics.
Note ・An execution error results if the
:MEASure:ITEM:HARMonic:SORT?
Query the (highest) harmonic order of analysis when sorted.
Syntax
Response
syntax
:MEASure:ITEM:HARMonic:SORT?
":MEASure:ITEM:HARMonic:SORT
<data1>" <0-50>"
Function
Specifies the highest harmonic order for
returning data in response to
"MEASure:HARMonic:SORT?". Items
to be output are according to
"MEASure:HARMonic:LIST?".
numeric value specified is out of range.
・Sortingisbylevel,andthecontentand
phase angle are output as degree data
for the respective level.
・All harmonic orders from the 1st to the
specified order will be output.
Intermediate orders cannot be skipped.
・The outputs items are always as
indicated by the
"MEASure:HARMonic:LIST?"
command.
Obtains the harmonic orders specified
by the
"MEASure:ITEM:HARMonic:SORT"
command.
Every 1 window (excluding when FD/printer output and communication)
12 bits
32 bits (floating-point calculations)
Selectable from the following sources:
U of the selected combination of channels for measurement
I of the selected combination of channels for measurement
Derived from an external sync signal
Possible to synchronize with external synchronization signals input from the
3194 main product's analog connector for U/I/P analysis.
Input level (for sine wave, 1 to 10 Vrms, 100 kHz or less)
Frequency splitting function included (1/1 to 1/255)
Accuracy requirement of analysis count is limited by the frequency
indicated in the parentheses.
The waveform number displayed on the screen is different depending on
the frequency.
If the fixed clock approach is used, the basic frequency is set to 50 Hz.
* The valid input range is from 5% to 110%.
* The voltage or current level selected for the PLL source is at least 10% of
range.
* Accuracy requirement is set to below 10 kHz if the PLL synchronization
range is from 10 Hz to 17.5 Hz or from 17.5 Hz to 35 Hz, or below 100
kHz if the synchronization range is 35 Hz or higher.
* Accuracy requirement is not set if the basic frequency synchronized with
PLL is not integer multiples of frequency composition.
* If the fixed clock is used, frequency composition other than integer
multiples of 50 Hz does not have accuracy requirement set up.
* In actual use, the rdg accuracy of the combined input unit is added to the
above analysis accuracy. Also, when it is used together with a clamp
(9270, 9271, 9272, 9277, 9278, or 9279), the accuracy of the clamp and
the frequency characteristics add to the analysis accuracy.
* If LPF of the input unit is used, the accuracy requirement is set to below
1/10 of LPF.
* While the phase angle is expressed using the PLL source's basic wave as
the base standard, the accuracy requirement for the phase angle of the high
frequency composition for the basic wave composition is not set.
* The phase for external synchronization signals is defined as variance
amount with sine wave (0.1 to 10 Vrms) input to the external
synchronization signals and Ext frequency ratio (1/1) setting conditions.
Accuracy requirement for other waveforms is not set.
* In combination with the 9601, the phase accuracy above 5 kHz is not
・(i), (i+1), and (i+2) are the measurement channel numbers.
・M is the number of samples; s is the sample point number.
・The active power of 3P3W and 3V3A channel are displayed but there is no meaning. However, only a
Note 1: (i), (i+1), and (i+2) are the measurement channel numbers.
Note 2: The subscript "k" on U, I, and P indicates the harmonic number, so for example "U
" is the
1
fundamental component of voltage. A "K" indicates the total harmonic analyzed.
Note 3: The subscripts "r" and "i" on Uk and Ik indicate the real and imaginary components of the results
of FFT analysis.
Note 4: The harmonic voltage phase angle and harmonic current phase angle are corrected and displayed
taking the PLL source forming the phase reference as 0
。
. A positive sign indicates the phase
leading, and a negative sign the phase lagging. (Figure 1 on the next page)
Note 5: The harmonic voltage-current phase difference indicates the difference between the harmonic
voltage phase angle and the harmonic current phase angle for the same harmonic. (Figure 2 on the
next page)
Voltage is measured between lines and current is measured with phase, so
U2 is phased 60°from U1. When the load is purely resistive, I1 lags U1 by
30°,andI2leadsU2by30°.
② With an inductive load such as a motor, when the effective power of ch1 is
negative, the phase of I1 lags U1 by more than 90°.
Based on 3P3W wiring, the three channels of voltage and current are
summed. The figures above show the state when the load is purely resistive.
② With an inductive load such as a motor
The effective power of ch1 appears negative because I1 lags U1 by more
than 90°. However, the summed data for ch(3) is not related to power
measurement, so wiring can be reversed.
To measure voltage and current phase, the phase of the voltage of each
channel is shifted 120°. When the load is purely resistive, as shown above,
the phase difference of the voltage and current of each channel is 0.
Compared to 3P3W (3V3A) wiring, ch(2) and ch(3) are switched.
② With an inductive load such as a motor, the phase of the current lags the
Voltage vector diagram after conversion
Psum = P1 + P2 +P3
NOTE
■ Conversion Theory
The momentary waveforms of the line voltages (U1, U2, and U3) are
converted into momentary waveforms for phase voltages (u1, u2, and u3)
using the following operation expressions ("s" refers to value).
u1s = (U1s-U3s)/3, u2s = (U3s+U2s)/3, u3s = (-U2s-U1s)/3
Δ-Y
Δ-Y
・ Since N is a virtual neutral point, phase voltages may differ from actual
values.
・ Following conversion, the three power meter method is applied.
・ Following conversion, the 9605-01 vector screen changes as follows. (Load:
Voltage vector diagram after conversio
Psum = P1 + P3
LOAD
NOTE
■ Conversion Theory
The momentary waveforms for phase voltages (U1, U2, and U3) are
converted into momentary waveforms for line voltages (u1, u2, and u3)
using the following operation expressions ("s" refers to sampling value).
u1s = (U1s-U2s), u2s = (U3s-U1s) , u3s = (U3s-U2s)
Y-Δ
Y-Δ
・ Following conversion, the two power meter method is applied.
・ Following conversion, the 9605-01 vector screen changes as follows. (Load:
Run the motor by operating the load (motor), without supplying voltage from
the inverter to the motor, and measure the induced voltage generated at the
motor terminal.
■ Connection
1. Connect voltage and current lines as shown below, in the same way as
when the motor is operated using an inverter.
2. The rotation-sensor signal must be connected to pins 48/47 of the "EXT
CONTROL" terminal on the rear panel of the 3194. The required input
level is a 1- to 10-Vrms sine wave or a TTL-level rectangular wave. Please
note that this input terminal is not insulated from the chassis. In addition,
set the PLL source to "EXT (con)."
3. To insulate the rotation-sensor signal, set the PLL source to "EXT (CH6)";
it will be taken in as a CH6 voltage input. In such a case, make sure the
rotation-sensor signal level is more than 50% of the voltage range of the
unit connected to the CH6.
The 3194 can be used for the following measurement when a rotation-sensor
signal (e.g., magnetic-pole position signal) is input.
(1) Zero-correction of phase difference
The 3194 is capable of zero-correction of the phase difference (electrical
angle) between the rotation-sensor signal and the induced voltage
fundamental wave. If this correction is made, line voltages generated when
the motor is driven by an inverter, and the line current vectors can be
evaluated by evaluating the rotation-sensor signals instead. Accurate
correction will be made even if the induced voltage is distorted, as the 3194
extracts the fundamental wave from the induced voltage using its harmonic
analysis function.
NOTE
NOTE
To ensure reproducibility, the motor must have the same rotational pulse
count as when the 3194 is used with a dividing setting of 1/1. For example,
the pulse count must be 2 pulses/rotation in the case of a 4-pole motor.
(2) Harmonic measurement of induced voltage
The 3194 is capable of harmonic measurement of the induced voltage. Using
the harmonic analysis function, the 3194 measures the rms value,
fundamental wave component, and harmonic of the induced voltage (line
voltage). In addition, using the Δ - Y conversion function, the induced
voltage may be converted to a phase voltage.
・For the phase difference (θ), the absolute accuracy is not specified. The
accuracy will be specified as a deviation, based on a comparison of one
cycle of induced voltage with one cycle of the rotation-sensor signal. In
addition, the 3194 measures the distance between the rising edge of the
rotation-sensor signal and the rising zero-crossing of the induced-voltage
fundamental wave.
・The zero-correction setting is valid for a single piece of data only. If the
frequency changes, zero-correction must be performed again.
・The zero-correction setting is cleared when power to the 3194 is turned off.
Rotation-sensor signal
(Magnetic-pole position signal)
When a motor is actually run by an inverter, changes in the control of the
inverter and the motor can be determined easily, without significant change
in the wiring for the induced-voltage measurement.
If zero-correction is performed on the phase difference between the rotationsensor signal and the induced voltage during the motor induced-voltage
measurement, the relationship between the fundamental wave phases
(electrical angle) of the line voltages and phase currents can be easily seen
on the vector screen, with the induced-voltage phase used as a reference. By
checking the screen, therefore, changes in the control of the inverter and the
motor as a result of fluctuations in the load can be easily determined. The
line voltages may be also expressed as phase voltages (virtual neutral) using
the Δ - Y conversion function.
3-phase
power
supply
Inverter
3194
Motor
l
a
n
g
i
s
r
o
s
n
e
s
n
o
i
t
a
t
o
R
Torque
sensor
)
l
a
n
g
i
s
n
o
i
t
i
s
o
p
e
l
o
p
-
c
i
t
e
n
g
a
M
(
l
a
n
g
i
s
r
e
t
e
m
o
h
c
a
T
When the rotationsensor signal is used
as a reference, if
zero-correction is
performed on the
rotation-sensor signal
phase and the
induced-voltage
phase, the inducedvoltage phase will
coincide with the
reference.
With this method, the rotation-synchronizing signal also undergoes FFT
analysis. In addition, using the fundamental wave as a reference, the absolute
phase of the signal of the line to be analyzed is measured. If the duty ratio
of the rotation-synchronizing signal is 50%, when the zero-crossing of the
rotation-synchronizing signal coincides with the zero-crossing of the line to
be analyzed, the phase difference will be 0 deg. If the duty ratio is other than
50%, even if the phase difference is indicated as 0 deg., the zero-crossing
points will not coincide.
External synchronizing signal
Signal waveform
Conditions for 0-degree phase difference with a duty ratio of 50%
External synchronizing signal
Signal waveform
Conditions for 0-degree phase difference with a duty ratio of 50%