HIOKI 9500-10 User manual

Page 1
9500, 9500-10
4-TERMINAL PROBE
Instruction Manual
September 2013 Revised edition 5 Printed in Japan
9500A980-05 13-09H
Probe tip
Levers
9500
9500-10
BNC plug
Orienting the probe so that the side printed with the model number is on top, plug it into the instrument's measurement terminals and secure in place with the levers on the left and right.
BNC plug of H
cur
, H
pot
, L
cur
and
L
pot
connected properly to the measurement terminals of each of the instruments
9500
9500-10
Short the probe's H
cur
and H
pot
terminals
(red) with one short bar and its L
cur
and
L
pot
terminals (black) with the other, so that there is no connection between the high and low terminals. Perform open correction.
*1: Leave the high and low terminals as
far apart as they will be when con-
nected to the measurement sample.
Short bar
Red (HIGH)
Black (LOW)
*1
Short the probe's H
cur
, H
pot
, L
pot
, and L
cur
terminals in this order, with a short bar and perform short correction.
Short bar
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Warranty
Warranty malfunctions occurring under conditions of normal use in con­formity with the Instruction Manual and Product Precautionary Mark­ings will be repaired free of charge. This warranty is valid for a period of one (1) year from the date of purchase. Please contact the distribu­tor from which you purchased the product for further information on warranty provisions.
Introduction
Thank you for purchasing the HIOKI Model 9500, 9500-10 4- Terminal Probe. To obtain maximum performance from the de­vice, please read this manual first, and keep it handy for future reference.
Initial Inspection
When you receive the device, inspect it carefully to ensure that no damage occurred during shipping. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Overview
The 9500 4-Terminal Probe, alligator-clip-type 4-terminal probe connect
s directly to Model 3532-80 Chemical Impedance Meter and Model RM3543 Resistance HiTester, and The9500-10 4-Terminal Probe, allig
ator-clip-t measurement terminals on the HIOKI IM Series (LCR Meters, Impe
dance Analyzers)
Safety Information
Follow these precautions the full benefit
Mishandling this device during use could result in injury or death, as well as damage to the device. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from device defects.
the measurement terminals on the HIOKI
ype 4-terminal probe connects directly to the
s of the various functions.
to ensure safe operation and to obtain
Safety Symbols
In the manual, the symbol indicates particularly impor­tant information that the user should read before using the device. The symbol p user should refer to a corresponding topic in the manual (marked with the sy function.
Indicates DC (Direct Current).
Indicates that incorrect operation presents a significant haz­ard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of
njury to the
i
Indicates advisory items related to performance or correct operation of the device.
rinted on the device indicates that the
mbol) before using the relevant
user or damage to the device.
Symbols for Various Standards
This symbol indicates that the product conforms to safety regulations set out by the EC Directive.
WEEE marking: This symbol indicates that the electrical and electronic appli August 13, 2005, and producers of the Member States are required to display it on the appliance under Article 11.2 of Directive 2002/96/EC (WEEE).
ance is put on the EU market after
Usage Notes
Before using the device for the first time, verify that it operates normally to ensure that no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Before using the device, make sure that the insulation on the probes is undamaged and that no bare conductors are improperly exposed. Using the device in such condi tions could cause an electric shock, so contact your
Hioki representative for repair.
dealer or
• Do not store or use the device where it could be exposed to direct sunlight, high temperature or humidity, or condensa tion. Under such conditions, the device may be damaged and insulation ma specifications.
• Do not use the device where it may be exposed to corrosive or
combustible gases.
• This device is not designed to be entirely water- or dust-proof. Do not
use it in an especially dusty environment, nor where it
might be splashed with liquid. This may cause damage.
• To avoid damage to the device, protect it from physical shock when transporting to avoid physical shock from dropping.
• Stop use and contact your dealer or Hioki representative if t
he probe is cont
oil, or dust.
• Which 4-terminal probe can be used depends on the mea­suring device to which this product is connected. Us wrong probe may increase the measurement error.
For more information about how to use the instrument with which you are using the probe, refer to that instrument's instruction manual.
y deteriorate so that it no longer meets
The device may be damaged.
and handling. Be especially careful
aminated with an excess amount of water,
ing a
Specifications
Probe Rubber-sheathed alligator-clip-type Frequency mea-
surement range Maximum rated
voltage Maximum rated
current
Length
Measurable conductor diameter
Mass
Cable
Oper
ating temper-
ature and humidity Storage tempera-
ture and humidity Operating
environment Accessories Instruction manual Applicable
standards (Safety)
9500: DC to 1 MHz 9500-10: DC to 200 kHz
±40 V DC (42 Vpeak (Measurement signal + bias voltage))
1 Apeak (Measurement signal + bias current)
Approx. 1000 mm (39.37”) (not including connection terminal)
diameter 0.3 to 2 mm
9500: Approx. 230 g (8.1 oz.) 9500-10: Approx. 140 g (4.9 oz.)
9500: 75 Ω c 9500
0 to 40°C (32 to 104°F), 80%RH or less. (non-condensing)
-10 to 55°C (14 to 131°F), 80%RH or less. (non-condensing)
Indoors, Pollution Degree 2 altitude up to 2000 m (6562 feet)
EN61010
oaxial cable
-10: 50 Ω coaxial cable
Names of Parts
-
-
Connection Methods
Open and Short Correction
When using the Model 3532-80 Chemical Impedance Meter and IM Series , perform open and short correction to increase mea­surement precision.
You will need: Two short bars A short bar is a device that sho terminals. Use short bars with as low an impedance as possible. When using a metal wire or similar object as the short bar, use as thick and short a wire as possible.
measurement cable's
rts the
Open correction
Short correction
• To perform zero adjustment with the Model RM3543 Resis­tance HiTester, follow the directions in the RM3543 instruc­tion manual.
• For more information about instrument operation, refer to the instr
ument's instruction
manual.
Measurement Procedures
Attach the rubber-sheathed alligator clips at the end of the probe t
o the measurement sa
• Exercise caution as dirt on the connection surfaces of elec­trodes or the measurement sample may prevent proper con­tact, making accurate measurement impossible.
• Open correction and measurement of high-impedance ele­ments are susceptible to the effects of external induced noise and str
ay capacitance. It for example by making measurements on a metal plate that has been connected to the guard terminal. (For more infor­mation about guarding, refer to the instrument's instruction manual
.)
mple and make measurements.
is recommended to use guarding,
Maintenance and Service
• To clean
• If the device seems to be malfunctioning, contact your dealer
the device, wipe it gently with a soft cloth moistened
with
water or mild benzene, alcohol, acetone, ether, ketones, thinners or gaso­line, as they can deform and discolor the case.
or Hioki repr tain damage during shipping, and include a description of existing
damage. We
age incurred during shipping.
detergent. Never use solvents such as
esenta
tive. Pack the device so that it will not sus-
do not take any responsibility for dam-