Hioki 8860-50, 8861-50, 8860-51, 8861-51 Input Module Guide

8860-50 8861-50
Input Module Guide
8860-51 8861-51
MEMORY HiCORDER
2
This Guide describes the optional input modules, related cable connection procedures, and their settings and specifications.
Dec. 2016 Revised edition 8 8860B985-08 16-12H
EN
* 6 0 0 1 9 3 0 4 8 *
Notes for Installing the Input Units
When installing 4 or more Model 8946 and Model 8958 at the same time
1
UNIT
45 8
1
UNIT
45 8 1
UNIT
45 8
OK
Do not install 4 units of Model 8946 to the UNIT 1-4 slots.
1
UNIT
45 8
4 units of Model 8946 can be installed to the UNIT 1-4 slots.
Other than Model 8958
Only a maximum of 3 units of Model 8946 can be installed in the UNIT 1-4 slots.
When Model 8958 is not installed
OK
8946 8946 8958
8946 89468958 8946 8958
When Model 8958 16ch Scanner Unit and 4 or more units of Model 8946 4ch Analog Unit are installed in Model 8861-50/8861-51 at the same time, only a maximum of 3 units of Model 8946 can be installed to the UNIT 1-4 slots of Model 8861-50/8861-51.
in Model 8861-50/8861-51
When Model 8961 High Voltage Unit are installed
Only a maximum of FOUR units of Model 8961 High Voltage Unit may be installed in Model 8861-50/8861-51 Memory HiCorder. To avoid the danger of electric shock, install no more than four 8961s.
When Model 8946 4ch Analog Unit are installed
8860B985-08
Please refer to the following flow chart to when installing input units:
As much as possible, install an equal number of units to the UNIT 1-4 and 5-8 slots.
Yes
Installation completed.
No
Up to 3 units of Model 8946 can be installed to the UNIT 1-4 slots. As much as possible, install an equal number of units to the UNIT 1-4 and 5-8 slots.
Will model 8958 be installed?
Will 4 or more units Model 8946 be installed?
As much as possible, install an equal number of units to the UNIT 1-4 and 5-8 slots.
1
UNIT
45 8 1
UNIT
45 8
8958
1
UNIT
45 8
8958
89468946
Installation example
Yes
No
Installation exampleInstallation example
Number of input units
1
UNIT
45 8 1
UNIT
45 8
5
6
7
8
1
2
3
4
Recommended positions for installing input units
Balanced installation of input units in the UNIT 1-4 and 5-8 slots is recommended in order to effectively utilize the internal memory of Model 8861-50/8861-51.
Number of input units
Contents
Introduction ........................................................................................1
Structure of this Document.................................................................3
Chapter 1
Overview___________________________________ 5
1.1 Product Overview ................................................................5
1.2 Input Module Usage List ...................................................10
1.3 List of Input Modules, Cables, Probes and Clamp
Combinations .................................................................... 13
i
Contents
Chapter 2
Connections _______________________________ 15
2.1 Installing Input Modules (Adding or Replacing) ................16
2.2 Connecting the Cables ......................................................18
2.2.1 Connection Preparations .......................................................18
2.2.2 Connecti ng to an Analog Unit
(Models 8936, 8938, 8946, 8956 and 8957) ..........................19
2.2.3 Connecti ng to the Model 8937 Voltage/Temp Unit ................20
2.2.4 Connecti ng to a Strain Unit (Models 8939 and 8960) ............22
2.2.5 Connecti ng to the Model 8940 F/V Unit ................. .. ..............23
2.2.6 Connecti ng to the Model 8947 Charge Unit .........................26
2.2.7 Connecti ng to the Model 8958 16-Ch Scanner Unit ..............28
2.2.8 Connecti ng to the Model 8959 DC/RMS Unit ........................30
2.2.9 Connecti ng to the Model 8961 High Voltage Unit ..................31
2.3 Connecting Clamps ...........................................................32
2.4 Connecting a Differential Probe ........................................34
2.5 Connecting Attenuating Probes ........................................38
2.6 Connecting Logic Probes ..................................................40
2.7 Supplying power from the Model 9687 Probe Power Unit 44
Chapter 3
Input Channel Settings ______________________ 45
3.1 Analog Unit Settings (Models 8936, 8946 and 8956) .......46
3.2 Model 8937 Voltage and Temperature Unit Settings ........47
3.2.1 Voltage Measurement ...........................................................47
3.2.2 Temperature Measurement ...................................................48
3.3 Model 8938 FFT Analog Unit Settings ..............................50
ii
Contents
3.4 Strain Unit Settings (Models 8939 and 8960) ...................51
3.5 Model 8940 F/V Unit Settings ...........................................52
3.5.1 Measuring Freq uency, 50/60 Hz and Rotation Rate .............52
3.5.2 Pulse Count Measurement ...................................................54
3.5.3 Pulse Duty M e a su r e m e n t ......................... .. .. ............. ... .. ....... 55
3.5.4 Voltage Measurement ................. ..........................................56
3.5.5 Current Measurement ........................................................... 57
3.6 Model 8947 Charge Unit Settings .....................................58
3.6.1 Voltage Measurement ................. ..........................................58
3.6.2 Acceleration Measurement (Charge, Preamp) .....................59
3.7 Model 8957 High Resolution Unit Settings .......................61
3.8 Model 8958 16-Ch Scanner Unit Settings ........................62
3.8.1 Voltage Measurement ................. ..........................................62
3.8.2 Temperatu r e M e a s urement ............ ............. .. .. .............. .. .. ... 63
3.9 Model 8959 DC/RMS Unit Settings ..................................65
3.10 Model 8961 High Voltage Unit Settings ............................66
3.11 About Setting Contents ............................. ............ ............67
3.11.1 Measurement Range Setting ..................... .. ........................67
3.11.2 Setting Input Coupling ...........................................................68
3.11.3 Low-Pass F ilter (LPF) Sett in g s ....... ............. .. ....................... 69
3.11.4 Anti-Aliasing Filter (AAF) Settings ......................................... 70
3.11.5 Digital Filter (Digital F) Settings ............................................ 70
3.11.6 Thermocouple (Sensor) Type Setting ...................................71
3.11.7 Reference Junction Compensation Setting ................... ........71
3.11.8 Drift Corre ct io n Se t ting ......... ........................................ .. .. ..... 72
3.11.9 Disconnect (Burn-Out) Detection Setting ..............................72
3.11.10 Hold Setting .......................................................................... 73
3.11.11 Pu ll-Up Setting .............. .. ... ............. .. .. ............. ... .. ............. .. . 74
3.11.12 Th reshold Sett in g ............... .. .. ............. .. ... ............. .. .. ............ 75
3.11.13 Response (Timing) Setting ..................... .. ........................ .... 75
3.11.14 Se n s o r S e n sitivity Setti n g ...... ............. .. ................................ 76
3.11.15 Probe Attenuation Selection ..................... .. ........................ ..78
3.11.16 Bridge Voltage Setting ................. .........................................78
3.11.17 Ex e c u ting Zero Adjustment ................................. .. ................ 79
3.11.18 Ex e cuting Offse t C a nc e llation ....... .......................... .. ... ......... 79
3.11.19 Ex e cuting Auto- B a la n c e ............. ... .. ............. .. .. .............. .. .. ... 81
Chapter 4
Logic Input Settings_________________________83
4.1 Operation Overview ..........................................................83
4.2 Setting Procedure .............................................................84
Chapter 5
Specifications______________________________ 87
5.1 Logic Input Section ...........................................................87
5.1.1 Model 9327 Logic Probe ...................... ........................ ..........87
5.1.2 Model MR9321-01 Logic Probe ................................... ..........88
5.1.3 Model 9320-01 Logic Probe ..................... .. ........................ ...89
5.2 Analog Input Section .........................................................90
5.2.1 Model 8936 Analog Unit ........................................................90
5.2.2 Model 8937 Volt age/Temp Unit .............................................92
5.2.3 Model 8938 FFT Analog Unit .................................................94
5.2.4 Model 8939 Str a in Unit ..........................................................95
5.2.5 Model 8940 F/V Unit ............................ .................. ................96
5.2.6 Model 8946 4-Ch Analog Unit ............................. .. ................98
5.2.7 Model 8947 Charge Uni t ...................... ........................ ..........99
5.2.8 Model 8956 Analog Unit ......................................................101
5.2.9 Model 8957 High Resoluti on Unit ................................. .. .....102
5.2.10 Model 8958 16-Ch Scanner Unit .........................................103
5.2.11 Model 8959 DC/RMS Unit ................................... ................105
5.2.12 Model 8960 Strain Unit ................................... .. ...................106
5.2.13 Model 8961 High Voltage Unit .................. .. ........................ .107
iii
Contents
Index ________________________________ Index1
iv
Contents

Introduction

1
Introduction
The following document s are pr ovided with the Model 8860-50, the 8861- 50 , the 8860-51 or the 8861-51 Memory HiCorders. Refer to them as appropriate for your application.
Document Description
Read this first.
Quick Start Manual
1
Input Module Guide
2
(This document)
It describes preparations for use, basic operati ng p ro­cedures and usage methods.
To connect input modules and measurement cables, and when making input channel set­tings;
this Guide describes th e optional input modul es, relat­ed cable connection procedures, and their settings and specifications.
To obtain setting details;
Instruction Manual
3
Analysis and Communication
4
Supplement
Which input module and cables to use with the in st ru men t dep end o n yo ur mea­surement application. Refer to this as appropriate for your application. (p. 3)
• In this document, th e “instrument” means the Model 8860- 50, the 8861-50, the 8860-51 and the 8861-51 Memory HiCorder.
• “Clamp” refers to one of our optional clamp-on sensor products.
• Unless otherwise no ted in this manual, informa tion provided for the 886 0-50 also applies to the 8860-51, and information provided for the 8861-50 also applies to the 8861-51.
this Manual describes details of the functions and op­erations of the instrument, and its specifications.
To analyze measurement data using the cal­culation functions, and to communicate with the instrument;
this supplement describes the procedures for analyz­ing data using numerical calculations, waveform cal­culations and FFT fun ctions, and h ow to communi cate with the instrument using a computer.
2
Introduction
Safety Symbols
In the manual, the symbol indicates particularly important informa­tion that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates DC (Direct Current).
Indicates AC (Alternating Curre nt).
Indicates both DC (Direct Current) and AC (Alternating Current).
Indicates a grounding terminal.
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrec t operation presents an ex treme ha zard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument.
Indicates advisory items related to performance or correct operation of the instrument.
Other Symbols
Indicates a prohibited action.
(p. )
*
[ ]
CURSOR
(Bold
characters)
Indicates the location of reference information. Indicates quick refere nces for ope ration and r emedies for troubleshoot -
ing. Indicates that descriptive information is provided below.
Screen labels such as menu items, page titles, setting items, dialog titles and buttons are indicated by square brackets [ ].
Bold characters within the text indicate operating key labels.
Accuracy We define measuremen t tolerances in terms of f.s. (full scal e) values, with the
following meanings: f.s. (maximum display value or scale length) The maximum disp la yable valu e or scale l ength. T his is usu ally the na me of th e currently selected rang e. In this instrumen t, the maxim um di spla yable val ue is th e ran ge (V /div ) times the number of divisions (20) on the vertical axis. Example: For the 1 V/div range, f.s. = 20 V

Structure of this Document

3
Structure of this Document
Preparation
Overview of Input
Modules and Logic
Groups
Procedure
Connections
Turn Power On
Instrument Settings
(input channel
settings)
Reference Location
About the overview of in put mod­ules and logic groups
Overview of usage, connections and settings
Input module connection s Specifications of input modules
and logic probes
Connecting the input modules, cables and logic probes to the instrument
Refer to the Quick Start Manual for other connections.
(When using an input module) Set analog channels
(When using a logic probe) Set logic channels
In addition to the above, perform necessary measurement settings such as measure­ment criteria and trigger settings. Refer to the Quick Start Manual for overviews of all operations. Refer to the Instruction Manual for details of settings other than those related to input channels.
"1.1 Product Overview" (p. 5)
"1.2 Input Module Usage List" (p. 10)
"1.3 List of Inpu t Modu les, Ca bles, Probes and Clamp Combinations" (p. 13)
"Chapter 5 Specifications" (p. 87)
Reference Location
"Chapter 2 Connections" (p. 15)
"Chapter 3 Input Channel Settings" (p. 45)
"Chapter 4 Logic Input Settings" (p. 83)
Measure
4
This Instrument
Input Module
GND
400 VDC max
370 VAC/DC
H
L
370 VAC/DC
Model 8936 Analog Unit Model 8938 FFT Analog Unit Model 8959 DC/RMS Unit
*2
*2
*1
This Instrument
Input Module
GND
400 VDC max
300 VAC/DC
H
L
300 VAC/DC
Model 8956 Analog Unit Model 8957 High Resolution
Unit
*1
*2
*2
This Instrument
Input Module
GND
30 Vrms or 60 VDC
30 Vrms or 60 VDC
H
L
30 Vrms or 60 VDC
Model 8937 Voltage/Temp Unit
Model 8946 4-Ch Analog Unit
*2
*2
*1
This Instrument
Input Module
GND
10 VDC max
30 Vrms or 60 VDC
H
L
30 Vrms or 60 VDC
Model 8939 Strain Unit
*2
*2
*1
This Instrument
Input Module
GND
DC10 V max
33 Vrms or 70 VDC
H
L
33 Vrms or 70 VDC
Model 8960 Strain Unit
*2
*2
*1
This Instrument
Input Module
GND
30 Vrms or 60 VDC
30 Vrms or 60 VDC
H
L
30 Vrms or 60 VDC
30 Vrms or 60 VDC
H
L
Connectors
BNC
Model 8940 F/V Unit
*2
*2
*1 *2
*1
This Instrument
Input Module
GND
Maximum applied elec­tric charge (p. 99)
30 Vrms or 60 VDC
H
L
30 Vrms or 60 VDC
30 Vrms or 60 VDC
H
L
BNC
Miniature
Connector
Common
Model 8947 Charge Unit
*2
*1
*2
This Instrument
Input Module
GND
40 VDC max
33 Vrms or 70 VDC
+
33 Vrms or 70 VDC
Model 8958 16-Ch Scanner Unit
*2
*1
*2
This Instrument
Input Module
GND
1000 VDC max
H
L
Model 8961 High Voltage Unit
*2
*2
*1
1000 VAC/DC (CAT II), 600 VAC/DC (CAT III)
Structure of this Document
Maximum Input Voltage*1 of input module and Maximum Rated Voltage to Ground*
2
5

1.1 Product Overview

Overview Chapter 1
1.1 Product Overview
This chapter pr ovides an ove rview of the opti onal input mo dules and in put cables that can be us ed with this instrument. Refer to "Appendix 6 Disposin g of the Instrument" in the Inst ruction Manual for a full list of option s for this instrument.
Voltage Measurement
Input Module
8956 Analog Unit 8957 High Resolution Unit 8936 Analog Unit 8946 4-Ch Analog Unit
2 channels, 20 MS/s, 12-bit, maximum input voltage: 400 VDC
High-speed type
8938 FFT Analog Unit 8959 DC/RMS Unit 8940 F/V Unit 8937 Voltage/Temp Unit
2 channels, 1 MS/s, 12-bit, maximum input voltage: 400 VDC
(Includes internal anti-alias­ing filter.)
8958 16-Ch Scanner Unit 8947 Charge Unit
16 channels, 20 S/s, 16-bit, maximum input voltage: 40 VDC
In addition to voltage, also supports temperature mea­surement.
2 channels, 2 MS/s, 16-bit, maximum input voltage: 400 VDC
Supports high -precision mea­surements. (Incl udes inte rnal anti-aliasing fi lte r.)
2 channels, 1 MS/s, 12-bit, maximum input voltage: 400 VDC
Measures both normal and RMS voltages.
2 channels, 1 MS/s, 12-bit, maximum input voltage: 30 Vrms/60 VDC
In addition to voltage, also supports measurements with piezoelectric acceleration sensors. (Includes internal anti-aliasing filter.)
2 channels, 1 MS/s, 12-bit, maximum input voltage: 400 VDC
2 channels, 1 MS/s, 12-bit, maximum input voltage: 30 Vrms/60 VDC
In addition to voltage, also measures current, frequen­cy, rotation rate, pulse total­ization and duty cycle.
8961 High Voltage Unit
2 channels, 2 MS/s, 16-bit, maximum input voltage: 1000 VDC
Use to measure high voltage power lines.
4 channels, 1 MS/s, 12-bit, maximum input voltage: 30 Vrms/60 VDC
2 channels, 1 MS/s, 12-bit, maximum input voltage: 30 Vrms/60 VDC
Measures temperature in addition to voltage.
Also refer to "Maximum Input Voltage*1 of input module and Maximum Rated Voltage to Ground*2" (p. 4) For the measurement resolution of each input module, see "Appendix 2 Overview of Settings and Functions" - "Measure- ment Range Setting" in the Quick Start Manual.
6
1.1 Product Overview
Input cables for voltage measurement with input modules (except the Model 8958 16-Ch Scanner Unit and 8961 High Voltage Unit)
L9197 Connection Cord
9197 Connection Cord
For high voltage, maximum input volt­age: 500 V (large alligator clips)
9322 Differential Probe 9665 10:1 Probe 9666 100:1 Probe
For high voltage, maximum input volt­age: (CAT II): 2000 VDC, 1000 VAC, (CAT III): 600 VAC/DC
Applicable Modules:
• Model 8940 F/V Unit
• Input modules for voltage measurement (except the Model 8958 16-Ch Scanner Unit)
To connect to an input modu le, the fol­lowing separate items are required:
• Model 9325 Power Cord (when using the Model 8940)
• Model 9418-15 AC Adapter
• Model 9248 Power Cord (when using the Model 9687)
L9198 Connection Cord L9217 Connection Cord
For low voltage, maxi mum input volt­age: 300 V (small alligator clips)
Maximum input voltage: 1 kVrms (500 kHz or l ess)
Applicable Modules :
• Model 8936 Analog Unit
• Model 8938 FFT Analog Unit
• Model 8956 Analog Unit
• Model 8957 High Resolution Unit
• Model 8959 DC/RMS Unit
Maximum input voltage: 300 V (for BNC output)
Maximum input voltage: 5 kVpeak (1 MHz or less)
Applicable Modules:
• Model 8936 Analog Unit
• Model 8938 FFT Analog Unit
• Model 8956 Analog Unit
• Model 8957 High Resolution Unit
• Model 8959 DC/RMS Unit
For Model 8961 High Voltage Unit
9242 Connection Cord
For Model 8961, Maximum input volt­age: 1000 V, Dolphin Clip
Model 9243 Grabber Clip (supplied)
Refer to "1.3 List of Input Modules, Ca bles, Probes and Clamp Combinati ons" (p. 13) for c ombinations of cables and pro bes to connect to an input module.
Input Module
9018-50
9132-50
3273-50 and 3276
3274 and 3275
9270 and 9271
9272
9277 9278
9279
8940 F/V Unit
7
1.1 Product Overview
Current Measurement
Input modules for voltage measurement
(except the Model 8958 16-Ch Scanner Unit
and 8961 High Voltage Unit)
Current can be measured using a clamp. 2 channels, 1 MS/s, 12-bit, clamp input In addition to current, also measures voltage,
Refer to "Voltage Measurement" (p. 5) for input modules.
Measures in [Voltage] mode. frequency, rotation rate, pulse totalization and duty cycle.
Clamps compatible with the above input modules.
3273-50 Clamp-On Probe* 9018-50 Clamp-On Probe 9132-50
Clamp-On Probe*
3
For AC Outputs 0.2 V AC waveform of com­mercial power line current. 9018-50: 10 to 500 A, 40 Hz to 3 kHz 9132-50: 20 to 1000 A, 40 Hz to 1 kHz
3274 Clamp-On Probe *
3275 Clamp-On Probe *
3276 Clamp-On Probe *
AC/DC Supports high-accuracy observation over broad ranges from DC to MHz. 3273-50: 30 A*
2
, DC to 50 MHz 3274: 150 A, DC to 10 MHz 3275: 500 A, DC to 2 MHz 3276: 30 A, DC to 100 MHz
1 1 1 1
9270 Clamp-On Sensor * 9271 Clamp-On Sensor * 9272 Clamp-On Sensor *
1, *3 1, *3 1, *3
For AC Supports inverter secondary current measurement. 9270: 20 A, 5 Hz to 50 kHz 9271: 200 A, 5 Hz to 50 kHz 9272: 20/200 A, 5 Hz to 10 kHz
For users of the following legacy products:
• Model 9018 Clamp-On Probe
• Model 9132 Clamp-On Probe Usable with this instrument by connecting the Model 9199 Conversion
9277 Universal Clamp-On CT * 9278 Universal Clamp-On CT *
1
1
9279 Universal Clamp-On CT *1,*
3
Adapter.
AC/DC Low zero drift supports stable long-
term measurements. 9277: 20 A, DC to 100 kHz
For measur ing leakage current:
The Model 9657-10 Clamp-On Leak Sensor can also be used.
9278: 200 A, DC to 100 kHz 9279: 500 A, DC to 20 kHz
*1. Except for Models 9018-50 and 913 2-50 , a s ep arat e c on ve rsi on cable or power su ppl y is re qui red when connecting to
an input module. Refer to "1.3 List of Input Modules, Cables, Probes and Clamp Combinations" (p. 13) for details. *2. Up to 15 A when combined with the Model 8940 F/V Unit. *3. Not applicable to CE Marking
8
1.1 Product Overview
Temperature Measurement (Thermocouple Inputs)
Input Module
8937 Voltage/Temp Unit 8958 16-Ch Scanner Unit
2 channels, 4 kS/s, 12-bit Thermocouple Types: K, E, J, T, N, R, S, B
In addition to temperature, also sup­ports voltage measurement.
16 channels, 20 S/s, 16-bit Thermocouple Types: K, E, J, T, N, R, S, B, W
In addition to temperature, also sup­ports voltage measurement.
Frequency, Rotation Rate, Commercial Line
Frequency (50/60 Hz), Pulse Count,
Pulse Duty Measurement
Input Module
8940 F/V Unit 8947 Charge Unit
2 channels, 1 MS/s, 12-bit
• Frequency: c an b e me as ured from the inp ut pulse corresponding t o the measurement waveform.
• Rotation Rate: can be measured from the input pulse correspond­ing to the measurement waveform.
• 50/60 Hz: Frequencies near 50/60 Hz can be measured from the input pulse corresponding to the measurement waveform.
• Totalization: cumulative count of input pulses.
• Pulse Duty: measured as the percentage of High level of a single pulse waveform.
Acceleration Measurement
2 channels, 1 MS/s, 12-bit
• Electric Charge: can be measured using a voltage-output type acceleration pick-up sensor.
• Preamp: ca n be measured usin g a n a cc el ­eration pick-up sensor with built-in preamp.
Also supports voltage mea sur em ent.
Electric Charge, Pressure, Torque,
Displacement Measurement
Input Module
8939 Strain Unit 8960 Strain Unit
2 channels, 1 MS/s, 12-bit Vibration and displacement can be measured using a strain gauge trans­ducer. (conversion cable supplied)
2 channels, 200 kS/s, 16-bit Vibration and displacement can be measured using a strain gauge trans­ducer. Supports high-precision mea­surements. (Includes internal anti­aliasing fil ter.) (conversi on cable sup­plied)
Logic Signal Measurement
Logic probes that can be connected to this instrument
9320-01 Logic Probe MR9321-01 Logic Probe 9327 Logic Probe
9
1.1 Product Overview
Measures digital signals and on/off switching of non-voltage contacts.
For users of the following legacy products:
Models 9306, 9307, 9320, 9321 Logic Probes Usable with this instrument by connecting the Model 9323 Conversion Cable.
Detects the presence of AC and DC voltages. Measures activation timing of relay sequencing circuits. Maximum input voltage: 250 Vrms (HIGH range)
Measures faster sig nal s than the Mod­el 9320-01 Logic Probe.
10

1.2 Input Module Usage List

1.2 Input Module Usage List
Voltage Measurement
To Perform Thi s Measurement
To measure 400 volts or
Up to 1000 V
more by direct connec­tion.
Up to 2 channels per module
For faster sampling Model 8956 Analog Unit To see high-precision
voltage values
Up to 400 V
To perform FFT analysis
To also see RMS voltage
For additional channels (four)
Up to 30 Vrms or 60 V DC
To measure with high sensitivity (500 µV/div)
Recommended Input
Module
Model 8961 High Voltage Unit
Model 8936 Analog Unit
Model 8957 High Resolu­tion Unit
Model 8957 High Resolu­tion Unit
Model 8938 FFT Analog Unit
Model 8959 DC/RMS Unit Model 8961 High Voltage Unit
Model 8946 4-Ch Analog Unit
Model 8937 Voltage/ Temp Unit
Model 8940 F/V Unit Model 8947 Charge Unit
Use to connect
Model 9242 Connec­tion Cord
Model L9197 Connec­tion Cord, Model 9197 Connection Cord (Up to 500 V)
Model L9198 Connec­tion Cord (Up to 300 V)
Model L9217 Connec­tion Cord (Up to 300 V)
Model 9242 Connec­tion Cord (for Model 8961 High Voltage Unit)
Connection
Procedure
(p. 31) (p. 66)
(p. 19) (p. 46)
(p. 19) (p. 46)
(p. 19) (p. 61)
(p. 19) (p. 61)
(p. 19) (p. 50)
(p. 30) (p. 31)
(p. 19) (p. 46)
(p. 20) (p. 47)
(p. 23) (p. 56) (p. 26) (p. 58)
Setting
Procedure
(p. 65) (p. 66)
Up to 40 V
When measured voltage exceeds the maximum in­put for the Input module
When slow sampling is acceptable, but more channels desired
Up to 2000 V DC or 1000 V AC (depending o n mea­surement site (p. 34) )
Up to 1000 Vrms (@500 kHz max.)
Up to 5000 Vpeak (@1 MHz max.)
Model 8958 16-Ch Scan­ner Unit
Above Input Modules
(except the Model 8958 16­Ch Scanner Unit and Model 8961 High Voltage Unit)
Model 8936 Analog Unit Model 8956 Analog Unit Model 8957 High Resolu­tion Unit Model 8938 FFT Analog Unit
Input Cable
Model 9322 Differe ntial
1
Probe *
Model 9665 10:1 Probe
2
*
Model
9666 100:1
2
Probe *
(p. 28) (p. 62)
(p. 34) (p. 78)
(p. 38) (p. 78)
(p. 38) (p. 78)
Model 8959 DC/RMS Unit
Overview of the above input modules (p. 5), description of cables (p. 6) Procedures for settings unrelated to input channels are the same. Refer to the Instruction Ma nua l for details.
*1. Voltage to ground complies with Model 9322 specifications. (p. 34) *2. Voltage to ground complies with the specifications of the input module used. (p. 90)
Temperature Measurement
11
1.2 Input Module Usage List
To Perform This
Measurement
To measure temperature
T o meas ure multi-channe l temperature
Recommended Input
Module
Model 8937 Voltage/ Temp Unit
Model 8958 16 -Ch Scan­ner Unit
Use to connect
Thermocouple
Thermocouple
Connection
Procedure
(p. 20) (p. 47)
(p. 28) (p. 62)
Setting
Procedure
Remarks
Up to 2 channels can be measured per module.
Up to 16 channels can be measured per module.
Overview of the above input modules (p. 8) Procedures for settings unrelated to input channels are the same. Refer to the Instruction Manual for details.
Current Measurement
To Perform This
Measurement
To read current values di­rectly without comp licated settings such as scaling
To measure using a
voltage measurement
module
Recommended
Input Module
Model 8940 F/V Unit
Voltage Measure­ment Input Module
Use to connect
Model 3273-50 Clamp-On Probe
Models 9270 to 9272 Clamp-On Sensor
Models 9277 to 9279 Universal Clamp-On CT
Models
9018-50/ 9132­50 Clamp-On Probe
Models 3273-50 to 3276 Clamp-On Probe
Models 9270 to 9272 Clamp-On Sensor
Models 9277 to 9279 Universal Clamp-On CT
Connection
Procedure
(p. 23) (p. 32)
(p. 32)
Setting
Procedure
(p. 52)
Refer to the voltage measurement description for each module
A conversion cable is required for connection (p. 13).
Scaling is required. Depending on the clamp, a power supply or sensor unit may be required for connection (p. 13).
Remarks
Overview of the above input modules (p. 7), about Clam ps (p. 7) Procedures for settings unrelated to input channels are the same. Refer to the Instruction Manual for details.
Other Measurements
To Perform This Measurement Recommended Input Module Use to connect
To measure electric charge, accelera­tion, pressure, torque or displacement using a strain sensor.
Model 8939 Strain Unit Model 8960 Strain Unit
Conversion Cable (sup­plied) Strain Gauge
Transducer To measure acceleration Model 8947 Charge Unit Piezoelectric Sensor To measure frequency
To measure pulse counts (totalization), pulse duty and similar values
Model 8940 F/V Unit
Models
L9198 Connection Cord
L9217 Connection Cord
Overview of the above input modules (p. 8) Procedures for settings unrelated to input channels are the same. Refer to the Instruction Manual for details.
Connection
Procedure
(p. 22) (p. 51)
(p. 26) (p. 59)
(p. 23) (p. 52)
Setting
Procedure
12
1.2 Input Module Usage List
Logic Measurements
To Perform This Measurement Recommended Probe
To measure digital signals and on/off switching of non-voltage contacts
T o mea sure larger si gnals than the above Model 9327 Logic Probe To measure the presence or absence of
AC or DC voltage
Model 9320-01 Logic Probe
Model MR9321-01 Logic Probe
Connection Procedure
(p. 40), (p. 42)
(p. 40), (p. 42)
(p. 40), (p. 43)
Setting
Procedure
(p. 83)

1.3 List of Input Modules, Cables, Probes and Clamp Combinations

*1. Although compatible, the L9198 is recommended instead. To connect to the input module, the following separate items
are required: *2. Model 9325 Power Cord , 9418-15 AC Adapter, or 9248
Power Cord (when using the Model 9687)
*3. Model 9418-15 AC Adapter or 9248 Power Cord (when
using the Model 9687) *4. Model 9318 Conversion Cable or 9555-10 Sensor Unit *5. Model 9319 Conversion Cable, 3272 Power Supply, 3269
Power Supply, or 9687 Probe Power Unit *6. Model 9555-10 Sensor Unit*
9
*7. Model 3272 , 3269 Power Supply, or 9687 Probe Power
Unit
*8. Set the instrument to [Voltage] measurement mode
when using combinations that do not include the Model 8940, and with the c ombinatio n of the 8940 and Model 9018-50 or 9132-50. (p. 56)
*9. Up to six units can be used with the 9325
Up to eight clamps can be used When the 9325 and clamps are used at the same
time, up to eight can be used *10. Not applicable to CE Marking *1 1: Th e Model 9 243 Gra bbe r Clip ca n also be conn ected .
About Scaling settings: Voltage acquired from the sensor is converted to the corresponding physical measurement units for display.
See "5.4 Converting Input Values (Scaling Function)" in
the Instruction Manual
1.3 List of Input Modules, Cables, Probes an d Clamp Combinatio n s
O = Compatible, = Incompatible, Δ = Compatible, but scaling required
Measurement
Parameter
Voltage
RMS Values Same as above FFT Analysis
(with AAF installed)
Current
Acceleration Piezoelectric Sensor Temperature Thermocouple
Frequency, Pulse Totalization, Pulse Duty
Use to connect
L9197 9197
Connection Cables
Differential Probe
Attenuating Probes
Input Cable
Same as above
Clamp-On Sensors
Universal Clamp-On
8
CTs *
Clamp-On Probes
*
Connection Cable (Sensor)
L9198 L9217
9242 9322 9665
9666
9270 * 9271 *
8
*
9272 * 9277
9278 9279 *
3273-50 3274
3275
8
3276 9018-50
9132-50
8936 8937 8938 8939 8940 8946 8947 8956 8957 8958 8959 8960 8961
OO
*1O
OOO OOO
*11
−−−−−−−−−−−−
O
*3O
O
*3O
3
*
O
O
*2,*
−−−−
−−−−−−−−−
−−−−−−−−−−
−−
10 10
Δ***
10
Δ***
10
Δ***
ΔΔΔ ΔΔΔ
O
−−−
6
6
7
*4,*
*4,*
*5,*
−−−−−−
O
−−−−−−−
−−−−
Input Module
*1O
*1O
*1OO
OOOOO OOOOO
O
O
*3O
9
O
Δ****
9
O
Δ****
9
O
9
Δ****
7
Δ*
*3O
O
OO
*3O
3
*
O
6
6
7
ΔΔΔΔΔ ΔΔΔΔΔ
O
−−−−−−
O
−−−−−−−−
O
O
O
3
O
*
O
O
−−−
O
−−
6
Δ*
6
Δ*
7
Δ*
Δ
Δ
O
−−−
13
−−
−−
−−
O
−−
−−
O
O
−−
−−
−−
−−
14
*1. Although compatible, the L9198 is recommended instead. To connect to the input module, the following separate items
are required: *2. Model 9325 Power Cord , 9418-15 AC Adapter, or 9248
Power Cord (when using the Model 9687)
*3. Model 9418-15 AC Adapter or 9248 Power Cord (when
using the Model 9687) *4. Model 9318 Conversion Cable or 9555-10 Sensor Unit *5. Model 9319 Conversion Cable, 3272 Power Supply, 3269
Power Supply, or 9687 Probe Power Unit *6. Model 9555-10 Sensor Unit*
9
*7. Model 3272 , 3269 Power Supply, or 9687 Probe Power
Unit
*8. Set the instrument to [Voltage] measurement mode
when using combinations that do not include the Model 8940, and with the c ombinatio n of the 8940 and Model 9018-50 or 9132-50. (p. 56)
*9. Up to six units can be used with the 9325
Up to eight clamps can be used When the 9325 and clamps are used at the same
time, up to eight can be used *10. Not applicable to CE Marking *1 1: Th e Model 9 243 Gra bbe r Clip ca n also be conn ected .
About Scaling settings: Voltage acquired from the sensor is converted to the corresponding physical measurement units for display.
See "5.4 Converting Input Values (Scaling Function)" in
the Instruction Manual
1.3 List of Input Modules, Cables, Probes and Clamp Combinations
Measurement
Parameter
Electric Charge, Acceleration, Pressure, Torque, Displacement
Use to connect
Strain Gauge Transducer
Input Module
8936 8937 8938 8939 8940 8946 8947 8956 8957 8958 8959 8960 8961
−−−
O
−−−−−−−
O
15
• Model 8936 Analog Unit (p. 19)
• Model 8937 Voltage/Temp Unit
(p. 20)
• Model 8938 FFT Analog Unit (p. 19)
• Model 8939 Strain Unit (p. 22)
• Model 8940 F/V Unit (p. 23)
• Model 8946 4-Ch Analog Unit (p. 19)
• Model 8947 Charge Unit (p. 26)
• Model 8956 Analog Unit (p. 19)
• Model 8957 High Resolution Unit
(p. 19)
• Model 8958 16-Ch Scanner Unit
(p. 28)
• Model 8959 DC/RMS Unit (p. 30)
• Model 8960 Strain Unit
(p. 22)
• Model 8961 H igh Voltage Unit (p. 31)
Connections Chapter 2
This chapter descr ibes the installation and connection of input modules to the instrument. Refer to the Quick Start Manual for other con necti ons .
Input modules for measurement (analog inputs)
1 Install the input module(s) in the instrument’s input module compartment.
(p. 16)
2 Connect the cables and sensors to the input module(s).
Connection items d iffer according to your appli cation. Refer to the descri ption of eac h input module.
Also refer to the following sections when measuring voltage and current:
• Using Connection Cables (p. 18) (Voltage measurement)
• Using Differential Probes (p. 34) (Voltage measurement)
• Using Attenuating Probes (p. 38) (Voltage measurement)
• Using Clamps (p. 32) (Current measurement)
• Supplying the power from the Model 9687 Probe Power Unit (p. 44)
3 After making connections, make instrument settings (p. 45).
This Guide describes on ly the procedures for setting the inp ut channels of ea ch input mod­ule. Refer to the Quick Start Manual for an overview of all settings, and to the Instruction Manual for the details of each setting.
Measurements with logic probes (logic inputs)
1 Connect the logic probes to the LOGIC receptacles on the instrument
(p. 40).
2 After connecting, perform settings on this instrument (p. 83).
16
1
UNIT
45 8
8958
1
UNIT
45 8 1
UNIT
45 8
Installation examples (installing four Model 8946s, one 8958 and either additional input modules or blank panels)
8946 8946 8946 89468958 8946 8958
OKOK

2.1 Installing Input Modules (Adding or Replacing)

2.1 Insta lling Input Modules (Adding or Replacing)
Input modules specifi ed at the time the i nstr ume nt is orde re d are s upp li ed pr ein­stalled. Use the following procedures to add or replace input modules, or to remove them from the instrument.
Preparations
• To avoid electric shock accident, before removing or replacing an input module, confirm that the instrument is turned off and that the connection cords are disconnected.
• The mounting screws must be firmly tightened or the input module may not perform to specifications, or may even fail.
When an input module is not used
• To avoid the da nger of elect ric shock, never opera te the instrume nt with an input module removed. To use the instrument after removing an input module, install a blank panel over the opening of the removed module.
When the Model 8961 High Voltage Unit is used
• Only a maximum of FOUR units of Model 8961 High Voltage Unit may be installed in Model 8861-50/8861-51 M emory HiCorder. To avoid the dan­ger of electric shock, install no more than four 8961s.
To avoid da magin g input m odule s, do no t touch the inpu t modul e connec tors on this instrument.
Using the Model 8958 16-Ch Scanner Unit
In the following case s, the Model 8958 must be adjusted before use with this instrument.
See "Scanner Unit Zero Position Adjustmen t" (p. 29)
• When installing the Model 8958 16-Ch Scanner Unit in this instrument
• When the zero position has shifted due to aging or environmental changes
When using the Model 8958 16- Ch Scanner Unit togeth er with four Model 894 6 4-Ch Analog Units at th e same time, the 8946s cann ot all be installed in UNIT locations 1 to 4 in the Model 8861-50/8861-51. In this case, no more than three 8946s can be installed in UNIT locations 1 to 4.
Adding an Input Module
A confirmation dialo g asks whether to reset the settings on th e Sheet Settings screen at starting (power-on) time. When the settings are reset, waveforms from all installed input modules (including added modules) are set to display. Until the settings are reset, the Sheet Settings screen is unaffected by added modules. To display waveforms from an added input module, select the corresponding dis­play channel on the Sheet Settings screen. Even when waveform display is enabled ([On]), a waveform is not displa yed unle ss it has be en enable d for dis­play on the Sheet Settings screen.
See "7.2.6 Assigning Display Channels to Graphs (Analog Channels)" in the Instruction
Manual
17
Required item: One Phillips-head screwdriver
1 Turn the instrument’s POWER switch Off. 2 With attention to the orientation of the input module,
insert it firmly all the way in.
Make certain that the labe ls on the input mo dule’s panel fa ce the same direction as the l abels on the rig ht side of the inst ru­ment.
3 Using the Phillips screwdriver, tighten the two input
module mounting screws.
When the instrument is positioned vertically as illustrated, module number o ne is at the left, and c hannel num ber one is at the bottom.
Information about the input modules installed in the instru­ment can be verified on the initial screen that appears when power is t ur n ed on , a nd i n t h e [Config] menu on the System screen.
Channel Configuration
Module No. (UNIT)
Channel
(CH)
Input Module
Compartment
UNIT1
CH1
CH2
CH3
CH4
UNIT3UNIT2
UNIT4
CH1
CH2
8947
8946
CH1
CH4
CH16
8958
Installing an input module
Right Side
Removing an input module
1 Turn the instrumen t’s POWER switch Off. 2 Remove any cables or thermocouples connected to
the input module.
3 Remove the power cord. 4 Using the Phillips screwdriver, loosen the two input
module mounting screws.
5 Grasp the handle and pull the module out.
If the input module has no handle, pull it out by grasp­ing the connectors.
Right Side
Required item: One Phillips-head screwdriver
Handle
(Example: 8956)
Blank panel
Install a blank panel. Using the Phillips screwdr iver, tighte n the two moun t­ing screws.
Measurements made without a blank panel installed may fail to meet specificatio ns because of temperature instability within the instrument.
If not installing another input module after removal
Right Side
2.1 Installing Input Modules (Adding or Replac ing)
18

2.2 Connecting the Cables

2.2 Connecting the Cables

2.2.1 Connection Preparations

Read the following carefully before making connections.
For voltage measurement
About maximum input ratings of input modules and cables
• The maximum input volt a ge is the lesser rating of either the input module or the cable to be used. Do not measure if this voltage would be exceeded, as damage the instrument or a serious accident may result. Example: When connecting the Model L9197, 9197 Connection Cord (500 Vmax) to the Model 8936 Analog Unit (400 Vmax), the maximum input voltage for measurement is 400 V DC.
For details of maximum input voltages, refer to "1.1 Product Overview"
(p. 5)
and "Chapter 5 Specifications" (p. 87).
• When connecting clip-type test leads to live termina ls, be very ca reful to avoid accidentally shorting conductors together and causing a serious accident.
• When measuring voltage in a power line, connection cables should only be connected to the sec ond a ry s ide o f a b re ake r, so the breaker can pre­vent an accident if a short circuit occurs. Connections should never be made to the primary side of a bre aker, because unrestricted curre nt flow could cause a serious accident if a short circuit occurs.
• Do not permanently connect the device in an en vi ron men t w here vo ltage surges exceeding the maximum input voltage may occur. Failure to observe this precaution could result in damage to the device and per­sonal injury.
• Do not use cables and clips other than those specified by Hioki. The specified cables include insulated plugs to avoid electric shock. Bare metal plugs may cause electric shock or damage to the connector.
• Do not connect a cable to the instrument while it is connected to the object to be measured. Otherwise, an electric shock accident may occur.
Use only the specif ied connection cords and clips. Usin g a non-specified c able may result in incorrect measurements due to poor connection or other reasons.
Using cables with BNC connectors
When disconnecting the BNC connec tor, be sure to release the lock be fore pull­ing off the connector. Forcibly pulling the connector without releasing the lock, or pulling on the cable, can damage the connector.
When using the Model 8958 16-Ch Scanner Unit together with other input modules
Do not attempt to measure the same signal with the 8958 and another input module at the same time. Channel switching noise from the 8958 may affect the other input module.
2.2 Connecting the Cables
Input modules
• Model 8936 Analog Unit
• Model 8956 Analog Unit
• Model 8946 4-Ch Analog Unit *
1
• Model 8938 FFT Analog Unit
• Model 8957 High Resolution Unit Maximum Input Voltage (p. 4)
Any of the following connects to a BNC jack on an input module.
• Model L9197 Connection Cord Model 9197 Connection Cord
Maximum input voltage: 500 V (large alligato r clip s)
• Model L9198 Connection Cord
Maximum input voltage: 300 V (small allig ator clips)
• Model L9217 Connection Cord
Maximum input voltage: 300 V (for BNC output)
If the voltage to be measured exceeds the maximum input rating of the input module being used:
• Model 9665 10:1 Probe*
2
(p. 38)
• Model 9666 100:1 Probe*
2
(p. 38)
• Model 9322 Differential Probe*
3
(p. 34)
*1. Use Model L9198 Connection Cords
for the Model 8946 4-Ch Analog Unit.
*2. Incompatible with the Model 8946 4-
Ch Analog Unit.
*3. The Model 9418-15 AC Ada pter op tion
is required for connection (p. 36).
1 Connect the BNC plug on the cabl e to a
BNC jack on the input module. Align the slots in the BNC plug with the
guide pins on the jack on the input module, then push and twist the plug clockwise until it locks.
2 Connect the cable clips to the measure-
ment object.
Disconnecting BNC connectors
Push the BNC p lug, twist it countercloc k­wise, and pull it out.
Example: Model 8936 Analog Unit
Required item: One of the above cables
BNC jack
Connection Procedure
Lock Input module connec­tor guide pins
BNC plug slots
Connect to the measurement object
Connecting the cable

2.2.2 Connecting to an Analog Unit (Models 8936, 8938, 8946, 8956 and 8957)

Be sure to read "2.2.1 Connection Preparations" (p. 18) before connecting.
19
20
Input module
Model
8937 Voltage/Temp Unit
Maximum Input Voltage (p. 4)
Voltage Measurement: Any of the following connects to a BNC jack on an input
module.
Temperature Measurement: Connect the thermocouple to the terminal block.
(Recommended cables (p. 21))
• Model L9198 Connection Cord
Maximum input voltage: 300 V (small alligator clips)
• Model L9217 Connection Cord
Maximum input voltage: 300 V (for BNC output)
If the voltage to be measured exceeds the maximum input rating of the input module being used:
• Model 9322 Differential Probe(p. 34) *
The Model 9418-15 AC Adapter option
is required for connection (p. 36).
2.2 Connecting the Cables

2.2.3 Connecting to the Model 8937 Voltage/Temp Unit

For voltage measurement, be sure to read "2.2.1 Connection Preparations" (p. 18) before connecting.
About Connections and Inputs
When measuring temperature
The ground (GND) sides of the voltage and temperature inputs for each channel are common. Do not connect and apply both voltage and temperature in puts at the same time. The measurement object could be damaged. Voltage and temperature cannot both be measured simultaneously on one chan­nel.
Observe the following to avoid electric shock.
• When measuring temperature with an uninsulated thermocouple at a point that has a non-zero electric potential, be careful to avoid touching the input module’s terminal block. The terminals may have voltage present.
• The voltage and temperature inputs are electrically isolated from one another.
• The push-button te rminals on the Model 8937 Voltage/Temp Unit ar e for the r-
• Use a tool (screwdriver) to connect and disconnect thermocouples.
• Confirm the po larity o f a thermoc ouple b efore conn ection. Correct v alues ar e
mocouples only. Do not connect anything except a specified thermocouple (type K, J, E, T, N, R, S or B).
not displayed if the thermocouple is connected with reverse polarity.
21
1 Connect the BNC plug on the cabl e to a
BNC jack on the input module.
Align the slot s in the BNC plug with the guide pins on the jack on the inp ut modu le, then p ush and twist the plug clockwise until it locks.
2 Connect the cable clips to the measure-
ment object.
Disconnecting BNC connectors
Push the BNC p lug, twist it countercloc k­wise, and pull it out.
Required item: Connection cord
BNC jack
Connection Procedure: Voltage Measurement
Lock Input module connector guide pins
BNC plug slots
Connect to the measurement object
Connecting the cable
1 Strip insulation from the thermocouple
wires as shown at the left.
Stripping length: approx. 10 mm
2 Push the blade of a flat screwdriver into the
button on the terminal block of the input module.
3 Insert each thermocouple wire into the
appropriate terminal hole while pressing the button. Confirm proper polarity.
4 Release the button.
The thermocouple is conne cte d.
5 Attach to the measurement object.
To remove the thermocouple
Hold the button while pulling the thermo­couple wire out.
Terminal Block
Connection Holes
Required item: Thermocouple, flat-blade screwdriver (2.6-mm blade)
Recommended wire: Compatible wire: Single-strand thermocouple wire,
0.4 to 1.2-mm diameter Stripping length: 10 mm
Connection Procedure: Temperature Measurement
Outer Insul a ti o n
10 mm
25 mm
Thermocouple single wires
Inner insulation
Connect to the measurement object
Connecting a Thermocouple
2
1
3
4
5
2.2 Connecting the Cables
22
Input module
Model 8939 Strain Unit
(conversion cable supplied)
Maximum Input Voltage (p. 4)
Connect the strain gauge converter* to the input module jack.
(Depending on the sensor, the supplied conversion cable may be needed.) A bridge box* is required for measurement using a strain gauge.*
The strain gauge converter, strain gauge and bridge box may be
common types available in the market.
Connect to the measurement object
Connecting the Conversion Cable
1 Connect the supplied c onversio n cable to
a receptacle on the input module.
Align the to oth on the inside of the recept acle with the slot in the conversion cable plug, and insert the plug. Inse rt it all the way to th e end of the slot in the plug ring (shaded part in the illus­tration at the left). When inserted, twist the ri ng clockwise to lock.
2 Connect the strain gauge transducer to
the conversion cable.
3 Attach to the measurement object.
To remove the conversion cable
Push and twist the plug ring counterclock­wise to unlock it, then pull the plug out.
Example: Connecting the strain gauge converter with the supplied conversion cable
Required item: Conversion Cable (sup plied ), Strain Gaug e Conv erter
Connection Procedure: Vibration and displacement measurement
Connect to the Strain Gauge Converter
Bridge voltage 8939: 2 V is supplied. 8960: Selectable (2 V/ 5 V/ 10 V)
Input voltage
+
­+
Connector Pinout
Tooth inside
receptacle
Lock
Connector
guide pin
Plug ring
Slot
3
2
1
The grounds of the instrument and the Con­nector Shell of the 8960 are not isolated .
-
2.2 Connecting the Cables

2.2.4 Connecting to a Strain Unit (Mode ls 89 39 and 8960)

When using a conversion cable
• Do not connect a sensor other than the Model 89 39 Strain Unit to the su ppl ie d conversion cable.
• When removing the conversion c able, be su re to relea se the lock b efore pull­ing the plug out.
23
Input module
Model 8940 F/V Unit Maximum Input Voltage (p. 4)
Frequency, Pulse Totalization and Duty, and Voltage Measurement:
Any of the following connects to a BNC jack on an input module.
Current Measurement: Connect to the BNC and sensor jacks on the input mod-
ule.
(The connection procedure depends on the clamp being used.)
• Model L9197 Connection Cord Model 9197 Connection Cord
Maximum input voltage: 500 V (large alligato r clips )
• Model L9198 Connection Cord
Maximum input voltage: 300 V (small alligato r clips)
• Model L9217 Connection Cord
Maximum input voltage: 300 V (for BNC output)
If the voltage to be measured exceeds the maximum input rating of the input module being used:
• Model 9322 Differential Probe*
1
(p. 34)
• Model 3273-50*5Clamp-On Probe*
2
• Model 3274, 3275, or 3276 Clamp-On Probe*
3
• Model 9270, 9271, or 9272 Clamp-On Sensor*
4
• Model 9277, 9278, or 9279 Universal Clamp-On CT*
4
*1. The optional Model 9325 Pow er Cord ,
Model 9418-15 AC Adapter, or Model 9248 Power Cord (when using the Model 9687) is required for connec­tion. (p. 35)
The following items are required to connect a clamp.
*2. Any of the following models:
9319 Conversion Cable, the 3272 or 3269 Power Supply, or Model 9248 Power Cord (when using the Model
9687)
*3. The Model 3272 or 3269 Power Supply
or Model 9248 Power Cord (when us­ing the Model 9687)
*4. The Model 9318 Conversion Cable or
the Model 9555-10 Sensor Unit
*5. When the Models 8940 and 3273-50
are used tog ether, the input rating of the 3273-50 is 15 A rms.
2.2 Connecting the Cables

2.2.5 Connecting to the Model 8940 F/V Unit

For voltage measurement, be sure to read "2.2.1 Connecti on Preparations" (p. 18) before connecting. For current measurement, read also "2 .3 Conn ectin g Clamp s" (p . 32) .
When measuring current
• Connect the clamp using the Model 9318 or 9319 Conversion Cable The grounds of this instrument and the clamp are not isolated. Pay care­ful attention to the connections to avoid damage to the equipment and serious injury.
• When us ing the Model 3273-50 Clamp-On Probe To avoid short circuits and serious injury when opening the sensor jaws to measure, if the voltage on the conductor to be measured may exceed safe voltage 1400 V AC withstand voltage), comply with pollution level 2 double (strength) insulation rated for 300 V do not clamp over bare conductors. The core and shield cover are not insulated.
• To avoid electric shock when using a clamp, be careful not to damage the insulation on conductors being measured.
up to 300 V, the conductor should be well-insulated (with
operation. To maint ain safety,
and have
For information about other clamps, see "2.3 Connecting Clamps" (p. 32).
24
1 Connect the BNC plug on the cabl e to a
BNC jack on the input module.
Align the slot s in the BNC plug with the guide pins on the jack on the inp ut modu le, then p ush and twist the plug clockwise until it locks.
2 Connect the cable clips to the measure-
ment object.
Disconnecting BNC connectors
Push the BNC p lug, twist it countercloc k­wise, and pull it out.
Required item: Connection Cables
BNC jack
Connection Procedure: Frequency, Pulse Totalization and Duty, and Voltage Measurement
Lock Input module connector guide pins
BNC plug slots
Connect to the measurement object
Connecting the cable
1
2
1 Align the slot in the co nversion ca ble plug
with the sensor receptacle on the input module, then insert a nd twist the p lug unti l it locks.
2 Align the slot in the sens or cab le plu g with
the conversion cable receptacle, then insert and twist the plug until it locks.
3 Clamp the sensor around the measure-
ment object.
To remove the conversion cable
Push and twist the plug to unlo ck it, then pull it out.
Example: When connecting the Model 9272 Clamp-On Sensor
Required item: Model 9318 Conversion Cable and 9272 Clamp-On Sensor
Sensor connector
Connection Procedure: Current measurement (with Model 9318 Conversion Cable)
(for Model 9270 to 9272 Clamp-On Sensors and Model 9277 to 9279 Universal Clamp-On CTs)
Connect to the measurement object
Connecting the clamp and the
Mode 9318 Conversion Cable
Conversion Cable Recep­tacle
2
1
Conversion Cable Plug
9272
Clamp-On Sensor
Clamp-On Sensor Plug
3
2.2 Connecting the Cables
T o maint ain safety, always use the optional Model 9318 or 9319 Conversion Cables when using one of theses clamp models: 3 273-50, 9270 – 9272 o r 9277 – 9279.
25
Connection Procedure: Current measurement (with Model 9319 Conversion Cable)
(for Model 3273-50 Clamp-On Probe)
1 Align the slots in the clamp- on probe plug
with the pins on the input module’s BNC jack, press and twist.
2 Align the groove in the conversion cable
plug with the sensor receptacle on the input module, then insert and twist the plug until it locks.
3 Align the groove in the probe’s power
cable plug with the conversion cable receptacle, and mate the connectors.
4 Clamp the sensor around the measure-
ment object.
To remove the conversion cable
Push and twist the plug to unlo ck it, then pull it out.
Example: When using the Model 3273-50 Clamp-On Probe
Required item: Models 9319 Conversion Cable and 3273-50 Clamp-On Probe
BNC jack
Sensor Receptacle
Connect to the measurement object
Connecting the clamp and the
Model 9319 Conversion Cable
Conversion Cable Receptacle
Measurement Signal Plug Alignment Notch
1
2
4
3273-50 Clamp-On Probe
1
2
3
Conversion Cable Plug
Power Cable Plug
2.2 Connecting the Cables
26
Input module
Model 8947 Charge Unit Maximum Input Voltage (p. 4)
Measurement using a piezoelectric acceleration sensor: Connect to the BNC jack or miniature receptacle
(according to the particular sensor).
Voltage measurement*
2
:
Connect to the BNC jack on an input module.
Model L9198 Connection Cord (recommended)
• Preamplified acceleration sensor *
1
(BNC connector, 2-mA, 15-V power supply)
• Charge-output type piezoelectric acceleration sensor *
1
(Miniature #10-32 connector)
*1. A commercially available acceleration
sensor can be used.
*2. For voltage measuremen t, the connec-
tion procedure is described in "2.2.2
Connecting to an Analog Unit (Models 8936, 8938, 8946, 8956 an d 8957 )" (p.
19).
2.2 Connecting the Cables

2.2.6 Connecting to the Model 8947 Charge Unit

Before measuring voltage and connecting to the BNC jack, be sure to read "2.2.1 Connec-
tion Preparations" (p. 18).
Connection Preparations
• Ne ver connect the cable t o the ins trument while c onnected to a mea sure­ment object. An electric shock could result.
• The BNC jack for each channel and the miniature receptacles all share common ground. To avoid short circuits, do not connect to two recepta­cles at the same time.
• To avoid electric shock or damage to the measurement object, turn the instrument off while connecting a sensor or probe to the BNC jack. When the [Pream p] measurement mode is selected, internal power (15 V @ 2 mA) is applied to the BNC jack when measurement starts.
About preamplified acceleration sensors
The sensor should be compatible with the spe cification (15 V @ 2 mA) o f the Model 8947 Charge Unit. Incompatible sensors are likely to be damaged.
27
1 Connect the BNC plug from th e se nsor to
a BNC jack on the input module.
Align the slots in the BNC plug with the guide pins on the jack on the input m odule, then push and twist the plug clockwise until it locks.
2 Connect the sensor to the measur ement
object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock­wise, and pull it out.
Required item: Preamplified acceleration sensor
BNC jack
Connection Procedure: Using a preamplified acceleration sensor
When using a sensor/preamp that has no BNC plug Use an adapter or conversion cable to connect to the BNC jack.
Lock Input module connector guide pins
BNC plug slots
Connect to the measurement object
Connecting a preamplified
acceleration sensor
1
2
1 Connect the miniature plug from the
charge-output acceleration sensor to the receptacle on the input module, and twist the plug clockwise until tight.
2 Connect the sensor to the measur ement
object.
To disconnect the miniature connector, twist the plug counte rclockwi se and pull it out.
Required item: Charge-output accelera tion sensor
Miniature Connector
Connection Procedure: Using a charge-output acceleration sensor
With a charge-output sensor that does not have a (#10-32) miniature connector Use an adapter or conversion cable to connect to the miniature connector.
Connect to the measurement object
Connecting the charge-output
acceleration sensor
1
2
2.2 Connecting the Cables
28
Input module
Model 8958 16-Ch Scanner Unit Maximum Input Voltage (p. 4)
Connect to the terminal block on the input module.
Temperature Measuremen t: Thermoc oupl e Voltage Measurement: Input Cable
Recommended cables: Solid 0.14 to 1.5 mm
2
Stranded 0.14 to 1 mm
2
16 to 26 AWG
Connect to the measurement object
Connecting to the terminal block
1 Strip back the insulation from the thermo-
couple wires as shown at the left.
Stripping length: approx. 5 mm
2 Remove the terminal block cover from the
input module.
3 Loosen the screws in the input module’s
terminal block using the screwdriver.
4 Insert the thermocouple wires in to the con-
nection holes. Confirm proper polarity.
5 Tighten the screws in the terminal block
using the screwdriver to affix the thermo­couple leads.
6 Replace the terminal block cover. 7 Attach the thermocouple to the measure-
ment object.
Terminal Block
Example: Connecting a thermo couple for temperatu re measureme nt (the c onnection pr ocedure i s the same for voltage measurement)
Connection Holes
Required item: Thermocouple, flat-blade d screw driv er (sup pl ied )
Connection Procedure (for both temperature and voltage measurements)
5 mm
Thermocouple single wires
Screw
Thermocouple
Terminal Block
Cover
The terminal block can be de­tached from the Model 8958 16­Ch Scanner Unit.
How to detach:
Loosen the screws at the top and bottom of the terminal block (left illustration). When attaching, be careful not to overtighten the screws .
With cover in place
Screw
Terminal Block
Turn the cover over and place the terminal block on it as shown above for stability while making connections.
Terminal Block
Cover
If difficulties are encountered connecting input cables or thermocouples
1
2
3
4
6
7
To remove the thermocouple
Loosen the screws on the terminal block and pull the wires out.
2.2 Connecting the Cables

2.2.7 Connecting to the Model 8958 16-Ch Scanner Unit

2.2 Connecting the Cables
The zero position of the Model 8958 16-Ch Scanner Unit needs to be ad justed in the following cases. If the adjustment is not performed, the accuracy specification may not be satisfied. Allow one hour
warm-up after turning power on before adjusting.
• When the Model 89 58 16-Ch Sc an ner Un it has jus t be en ins talle d i n th e i nstr ume nt (a me ssage requesting adjustment appears when the instrument is turned on)
• When the zero pos ition has shifted due to aging or environmental changes
Init Settings Screen
1 Short the + and – terminals of c hannels 1
and 9 on the Mode l 8958 16-Ch Scanner Unit to be adjusted.
2 Press F7 [System] on the Openin g screen,
or place the curs or on the Function menu of the Waveform or Settings screen, and press the F7 [System] key.
The System screen appears.
3 Press the SUB MENU keys to select the
[Init] menu item.
The Initialization (Init) Settings screen appears.
4 Using the CURSOR keys, place the cursor
on the [Adjust Scanner Unit] button, and press the F1 [Execute] key.
The Adjustment dialog box appears.
5 Select the module to be adjusted, and
press the F7 [Execute] key. To cancel, press the F8 [Cancel] key.
A processing messa ge ap pears when execute d, and the 8958 16-Ch Scanner Unit is adjusted.
Adjustment is complete when “Completed nor­mally” appears.
Waveform or Settings Screen
Function Menu
Supplied shorting bar or tinned wire.
CH1
CH9
Opening Screen
Scanner Unit Zero Position Adjustment
Tighten the screws securely.
29
30
Input module
Model 8959 DC/RMS Unit Maximum Input Voltage (p. 4)
Any of the following connects to a BNC jack on an input module.
• Model L9197 Connection Cord Model 9197 Connection Cord
Maximum input voltage: 500 V (large alligator cl ips)
• Model L9198 Connection Cord
Maximum input voltage: 300 V (small alligator clips)
• Model L9217 Connection Cord
Maximum input voltage: 300 V (for BNC output)
If the voltage to be measured exceeds the maximum input rating of the input module being used:
• Model 9665 10:1 Probe (p. 38)
• Model 9666 100:1 Probe (p. 38)
• Model 9322 Differential Probe*
1
(p. 34)
*1. The Model 9418-15 AC Ada pter op tion
is required for connection (p. 36).
1 Connect the BNC plug on the cabl e to a
BNC jack on the input module.
Align the slot s in the BNC plug with the guide pins on the jack on the inp ut modu le, then p ush and twist the plug clockwise until it locks.
2 Connect the cable clips to the measure-
ment object.
Disconnecting BNC connectors
Push the BNC p lug, twist it countercloc k­wise, and pull it out.
Required item: One of the above cables
BNC jack
Connection Procedure
Lock Input module connector guide pins
BNC plug slots
Connect to the measurement object
Connecting the cable
1
2
2.2 Connecting the Cables

2.2.8 Connecting to the Model 8959 DC/RMS Unit

Be sure to read "2.2.1 Connection Preparations" (p. 18) before connecting.
2.2 Connecting the Cables
Input module
Model 8961 High Voltage Unit Maximum Input Voltage (p. 4)
Connect to the terminal on the input module.
Model 9242 Connection Cord
Maximum input voltage: 1000 V (dolphin clips)
The supplied Model 9243 Grabber Clip can also be connected.
Model 9242 Connection Cord
1 Connect the cable plugs (at the end with fer-
rite beads) to the input module terminal.
2 Insert the other ends of the cables into the
supplied or optional clips.
3 Connect the cable clips to the measurement
object.
Required item: above cable
Connection Procedure
Connect to the measurement object
1
2
3
Ferrite beads
Supplied clips
Supplied Model 9243 Grabber Clip
Securely attach the clips to the screws on the terminal strip or bus bar.
Connection Example
Connect each plug to the terminal of the same color.

2.2.9 Connecting to the Model 8961 High Voltage Unit

Be sure to read "2.2.1 Connection Preparations" (p. 18) before connecting.
Connection Preparations
• For safety reasons, when taking measurements, only use th e specified HIOKI 9242 Connection Cord or optional 9243 Grabber Clip. Using a non­specified cable may cause an electric shock.
• Only a maximum of FOUR units of Model 8961 High Voltage Unit may be installed in Model 8861-50/8861- 51 Memory HiCorder. To avoid the d an­ger of electric shock, install no more than four 8961s.
31
32

2.3 Connecting Clamps

2.3 Connecting Clamps
Because it measures on ly voltag e input, curr ent is not me asured dire ctly by this instrument. Use a clamp that provides voltage output for current measurement. Refer to "1.3 List of Input Modules, Cables, Probes and Clamp Combinations" (p.
13) for compatible combinations of Hioki clamps and input modules.
Refer also to the clamp’s Instruction Manual for clamp-specific details.
Preparations for Using Clamps
Connect the clamp-on sensors to the instrument first, and then to the active lines to be measured. O bserve the follo wing to avoid ele ctric shock and short circuits.
• To avoid short circuits and potentially life-threatening hazards, never attach the clamp to a circuit that operates at more than the maximum rated voltage, or over bare conductors.
• Clamps s hould only be connected to the secondary side of a breaker, so the breaker can prevent an accident if a short circuit occurs. Connec­tions should never be made to the primary side of a breaker, because unrestricted current flow could cause a serious accident if a short circuit occurs.
• When the clamp sensor is opened, do not allow the metal part of the clamp to touch any exposed metal, or to short between two lines, and do not use over bare conductors.
To avoid electric shoc k when measuring live lines, wear appropriate pro­tective gear, such as insulated rubber gloves, boots and a safety helmet.
• To prevent damage to the instrume nt and clam p, neve r connect or di sconnect a sensor while the po wer is on, or while the sensor is cl amped around a c on­ductor.
• Be careful to avoid dropping the clamps or otherwise subjecting them to mechanical shock , which could damage the mati ng surfaces of the core and adversely affect measurement.
• Measurements are a ffected by the combi ned accuracy of thi s instrument an d
the clamp.
• The measured value is displayed as voltage [V], although this can be
converted for display as current [A] using the Scaling function.
See "5.4 Converting Input Values (Scaling Function)" in the Instruction Manual
• In the following cases, scaling setup is necessary:
(1) When using a clamp with a voltage measurement input module other than
the Model 8940 F/V Unit
(2) When connecting a clamp to the Model 8940 F/V Unit without the Mode l
9318 or 9319 Conversion Cable
For details about connecting clamps to the Model 8940 F/V UNIT:
See "2.2.5 Connecting to the Model 8940 F/V Unit" (p. 23).
33
1 Connect the probe to a BNC jack on the
input module.
2 Clamp the sensor around the measure-
ment object. Disconnecting BNC connectors
Push the BNC plug, twist it c ounterclock­wise, and pull it out.
Example: Connecting the Model 9018-50 Clamp-On Probe to the 8936 Analog Unit
Required item: 9018-50 Clamp-On Probe
BNC jack
Connection Procedure
L
O
A
D
S
O
U
R
C
E
Conductor
Current flow direc-
tion arrow
The arrows on the clamp indicating the direction of current fl ow should point towar d the load side.
9018-50 Clamp-On Probe
Attach the clamp around only one conductor. Single-phase (2-wire) or three-phase (3­wire) cables clamped together will not pro­duce any reading.
OK
Lock
Input module connector guide pins
BNC plug slots
Connect to the measurement object
Connecting Clamps
For connection details, refer to "2.2.5 Connecting to the Model 8940 F/V Unit" (p. 23).
1
2
2.3 Connecting Clamps
34
Grabber clip
H
L
(CAT II) 2000 V DC, 1000 V AC (CAT III) 600 V AC/DC
(CAT II) 1500 V AC/DC (CAT III) 600 V AC/DC
Between H-L H-to-ground
L-to-ground
H
L
(CAT II) 2000 V DC, 1000 V AC (CAT III) 600 V AC/DC
(CAT II) 1000 V AC/DC (CAT III) 600 V AC/DC
Between H-L H-to-ground
L-to-ground
Alligator clip

2.4 Connecting a Differential Probe

2.4 Connecting a Differential Probe
Input modules that are compatible with the Model 9322 Differential Probe:
Model 8940 F/V Unit*
• Voltage measurement input modules (BNC jack) other than the Model 8940 *
*1. The Model 9325 Power Cord or Model 9418-1 5 AC Ad ap ter i s requ ire d fo r c onn ec tion . *2. The Model 9418-15 AC Adapter is required for connection.
Refer to the instruction manual for the 9322 Differential Probe.
Note the following maximum input voltage and maximum rated voltage to earth. If their voltages are exceede d, this device will be damaged and per­sonal injury will result. Therefore, do not perform measurement in this case.
• Maximum input voltage (CAT II) 2000 V DC, 1000 V AC, (CAT III) 600 V AC/DC
• Maximum rated voltage to earth When using the Grabber clip: (CAT II) 1500 V AC/DC, (CAT III) 600 V AC/DC When using the Alligator clip: (CAT II) 1000 V AC/DC, (CAT III) 600 V AC/DC
1
2
When using an input module (BNC jack) other than the Model 8940 F/V Unit
• Turn the instrument off before connecting the AC adapter to the 9322 and to AC power.
• Use only the specified Model 9418-15 AC Adapter. AC adapter input volt­age range is 100 to 240 VAC (with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this range.
35
1 Connect the power cord to the power jack
on the probe.
2 Connect the other end of the power cord
to the sensor receptacle on the 8940 F/V Unit
3 Connect the BNC plug on the probe to a
BNC jack on the input module.
4 Connect an alligator clip or grabber clip to
the input lead of the probe.
5 Attach to the measurement object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock­wise, and pull it out.
Required item: Model 9322 Differential Probe and 9325 Power Cord
9325 Power Cord
BNC jack
Connection Procedure: When connecting to the Model 8940 F/V Unit (using the 9325 Power Cord)
Connect to the measurement object
Power jack
Input leads
Sensor connector
2
4
5
Lock
Input module connector guide pins
BNC plug slots
BNC connector
3
1
Set the [Probe] selection on the Channel Setting screen to [9322+9325].
When connecting to the Model 8940 F/V Unit (using the 9418-15 AC Adapter)
Set the [Probe] selection on the Channel Setting screen to [9322+9418].
Refer to "Connection Procedure: Measur-
ing voltage with the input module (using the Model 9418-15 AC Adapter)" (p. 36) for
connection details.
2.4 Connecting a Differential Probe
36
1 Connect the BNC plug on the probe to a
BNC jack on the input module.
2 Connect the AC adapter to the power jack
on the probe, and plug it into an AC outlet.
3 Connect an alligator clip or grabber clip to
the input lead of the probe.
4 Connect to the measurement object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock­wise, and pull it out.
Required item: 9322 Differential Probe and 9418-15 AC Adapter
BNC jack
Connection Procedure: Measuring voltage with the input module (using the Model 9418-15 AC Adapter)
Connect to the measurement object
Power jack
Input leads
1
2
3
4
Lock
Input module connector guide pins
BNC plug slots
BNC connector
Example: Connecting to the Model 8936 Analog Unit
Connect the Model 9418-15 AC Adapter
Connect to an AC outlet
Set the [Probe] selection on the Channel Setting screen to [1000:1].
2.4 Connecting a Differential Probe
37
Model 9687 Probe
Power Unit
1 Connect the BNC plug on the probe to a
BNC jack on the input module.
2 Connect the Model 9248 Power Cord to
the power jack on the probe, and plug it into the Model 9687 Probe Power Unit.
3 Connect an alligator clip or grabber cli p to
the input lead of the probe.
4 Connect to the measurement object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock­wise, and pull it out.
Required item: Model 9322 Differential Probe and Model 9248 Power Cord
BNC jack
Connection Procedure: Supplying power from the Model 9687 Probe
Power Unit (Using the Model 9248 Power Cord)
Connect to the measurement object
Power jack
Input leads
1
2
3
4
Lock
Input module connector guide pins
BNC plug slots
BNC connector
Example: Connecting to the Model 8936 Analog Unit
Connect the Model 9248 Power Cord
Connect the Model 9687 Probe Power Unit
Set the [Probe] selection on the Channel Setting screen to
[1000:1].
When using the Model 8940, select
[9322+9325] or [9322+9418]. Either setting is
acceptable.
2.4 Connecting a Differential Probe
38
1 Connect the BNC plug on the probe to a
BNC jack on the input module.
2 Connect the ground lead to the probe. 3 Connect the hook clip to the probe. 4 Connect to the measurement object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock­wise, and pull it out.
Required item: 9665 10:1 Probe
BNC jack
Connection Procedure
Connect to the measurement object
Ground lead
1
2
3
Lock
Input module connector guide pins
BNC plug slots
BNC connector
Example: Connecting the Model 9665 10:1 Probe to the 8936 Analog Unit
Hook Clip
4

2.5 Connecting Attenuating Probes

2.5 Connecting Attenuating Probes
The Model 9665 10:1 Probe and 966 6 100:1 Probe can be connected t o the following input modules:
• Model 8936 Analog Unit
• Model 8938 FFT Analog Unit
• Model 8956 Analog Unit
• Model 8957 High Resolution Unit
• Model 8959 DC/RMS Unit
Refer to the probe’s instruction manual for details.
Note the following maximum input voltage and maximum rated voltage to earth. If their voltages are exceede d, this device will be damaged and per­sonal injury will result. Therefore, do not perform measurement in this case. The measurement category ( overvoltage category) must correspond with that of the input module being used.
• Maximum input voltage Model 9665 10:1 Probe 1000 Vrms (500 kHz max) Model 9666 100:1 Probe 5000 Vpeak (1 MHz max)
• Maximum rated voltage to earth The maximum rated voltage to ground of the input module being used
Calibrate the attenuating probe before measurement. Procedure:
"Calibration" (p. 39)
1 Connect the metal pins or lead wires to the
TRIG.OUT
/CAL and GND I/O terminals.
While pressing the terminal button with the screw­driver, insert the metal pin or lead wire. Removing the screwdriver affixes the pin or wire.
2 Connect the hook clip of the probe to the
metal pin or lead wire in the TRIG.OUT
/CAL terminal, and the ground lead to the GND ter­minal.
3 Select the [Ext Term] menu from the System
screen, and select [Calibration] for the
[TRIG.OUT
/CAL] setting.
4 Select the following for each settings screen.
5 Press the instrument’s START key to begin
recording.
A 1-kHz, 5-V square wave is displayed and output at the terminals.
6 Adjust the trimmer on the prob e to obta in t he
proper waveform on the screen.
Settings Screen Setting Item Settings Function MEM
Status
Timebase 100
μS/div
Shot (recording length)
25 div
Channel
Range 1V/div Coupling DC LPF OFF Probe "3.11.15" (p. 78)
Trigger
Trigger Mode Repeat
[Analog] page
Type Level Level 2.5V
Slope
Calibration
Right Side
Waveforms indicating calibration required
Calibrated waveform
Required item: Flat-blade screwdriver (2.6-mm blade) Connect the probe between the TRIG.OUT/CAL and GND external I/O terminals using metal pins or two short leads Metal pin example: Mac8
• DH-4-20 pin for surface mounting
• Terminal for hybrid ICs CD-5-20, CF-1-15 and CA-7-3
(Attenuating probes do not connect directly to the external I/O terminals.)
Ext Term Settings Screen
5 V
500 μs
1
2
TRIG.OUT/CAL
Hook Clip
Ground Lead
2
Press the DISP key, and the F7 [System] key. The System screen appears. Press the SUB MENU keys to se lect the [Ext Term] (Ex- ternal terminal) menu.
3
Attenuating probe
1
To Insert
3
Push
Pull
External I/O Terminals
(Any GND terminal may be used)
Calibration
Trimmer
6
GND
39
2.5 Connecting Attenuating Probes
40
Measurement Object
Memory
HiCorder
Logic probe
Logic probe
Functional Earth Terminals
GND
Connect this instrument to the same outlet as the measurement ob ject us­ing the (supplied) grounding polar­ized power cord.
Connect the measurement object’s ground to the GND terminal of this in­strument. (Always obtain power from the same mains circuit.)
About Functional Earth:
Memory
HiCorder
Measurement Object
See "3.5 Grounding the Instrument’s Functional Earth" in the Quick Start Manual

2.6 Connecting Logic Probes

2.6 Connecting Logic Probes
Up to four logic probes can be connected to the LOGIC receptac les on the right side of the instrument. Each logic probe provides four logic input channels, so up to 16 logic waveforms can be recorded.
The following logic probes are supported:
• Model 9327 Logic Probe
• Model MR9321-01 Logic Probe
• Model 9320-01 Logic Probe Refer to the instruction manual supplied with the logic probe for specific details.
Do not connect logic probes other than those supplied by Hioki to the logic inputs.
For users of the following legacy models:
Models: 9306, 9307, 9320 and 9321 These can be used with this instrument by connecting the Model 9323 Conver-
sion Cable.
Before connecting a logic probe to the measurement object
To avoid electric shock and short circuit accidents or damage to the instru­ment, pay attention to the following:
• The ground pin in the LOGIC connector (plug) of the Model 9320-01 and 9327 Logic Probes (and legacy Models 9306 and 9320) is not isolated
from this instrument’s ground (common ground). Use grounding-type polarized power cords for the measurement object and this instrument, and obtain power from the same main s ci rcuit. Connecting to different mains circuits or using a non-grounding power cord may cause damage to the measurement object or this instrument because of current flow through the logic probes resulting from potent ial difference between the grounds of the different wiring systems. To avoid these problems , we recommend the following c onnection pro­cedure:
41
Connection Procedure
Connect to the measurement object
See "Logic Probe Usage Procedures" (p. 42)
Right Side
Example: Connecting the Model 9327 Logic Probe
1 Connect the logic probe by aligning the
groves on the plug and a LOG IC recepta­cle.
2 Connect to the measurement object.
Required item: Model 9327 Logic Probe
LOGIC receptacles
2.6 Connecting Logic Probes
• Maximum logic probe input voltages are as follo ws. Do not measure if the maximum voltage would be exceeded, as damage the instrument or personal injury may result.
Model 9327 Logic Probe: +50 VDC Model 9320-01 Logic Probe: +50 VDC Model MR9321-01 Logic Probe: 250 Vr ms ( HI GH ra ng e ) , 150 Vrms (LOW range)
• Do not allow the metal tip of a logic probe to cause a short between con­ductors on the measurement object. Never tou ch the metal tip of a probe.
When a logic probe is not connected, the corresponding logic waveform appears at HIGH level on the waveform screen.
42
Using Model 9327 and 9320-01 Logic Probes
Connect to the measurement object
When measuring digital signals (Digital input)
1 Connect the IC clip leads to the logic
probe.
2 Set the input selector to DIGITAL. 3 Connect the alligator clip to the circuit
ground.
4 Use the threshold value selector to select
the threshold value.
5 Connect the IC clips to the meas urement
object.
When measuring the contact signal (Contact input)
1 Connect the Alli gator cli p leads to the logic
probe.
2 Set the input selector to CONTACT. 3 Connect the alligator cl ips to the measure-
ment object.
IC clip leads
Alligator clip leads
1
3
3
4
Alligator clip lead
5
Connect to the measurement object
GND
Input selector
1
2
Input selector
2
Connect to the LOGIC receptacle
Connect to the LOGIC receptacle
Threshold value selector
2.6 Connecting Logic Probes
Logic Probe Usage Procedures
Range T able
Range Digital input (Threshold v alue) Contact input (Detecting resistance value)
1.4 V 1.4 V±0.3 V
2.5 V 2.5 V±0.4 V
4.0 V 4.0 V±0.5 V
more than 1.5 kΩ less than 500Ω
more than 3.5 kΩ less than 1.5 kΩ
more than 2.5 kΩ less than 8 kΩ
opened (Output L) shorted (Output H)
opened (Output L) shorted (Output H)
opened (Output L) shorted (Output H)
43
Using the Model MR9321-01 Logic Probe
Connect to the measurement object
Input selector
Connect to the LOGIC receptacle
Clips
1 Set the input selector in accordance with
the measured voltage. LOW range: On/off for 100 VAC and 24 VDC, etc. HIGH range: On/off for 200 VAC, etc.
2 Connect the clips to the measurement
object.
2.6 Connecting Logic Probes
The number of channels 4 (Isolated between channels and the instrument) Range LOW HIGH Input resistance 30 k Sensitivity Output: L 0 to 10 VAC
Output: H 60 to 150 VAC
Response time
Maximum input voltage 150 V rms 250 V rms Maximum rated voltage
to earth Dielectric strength Insulation resistance
↑ ↓
For other models, refer to the specifications (p. 88).
• Inputs are non-polarized, so polarity can be disregarded.
• Inputs are isolated, so each channel can be connect ed to a point of inde pen­dent potential.
• This instrument detects absolute values, so DC of either polarity can be input.
• AC voltage is evaluated as a 50/60 Hz standard sine wave.
Ω min 100 kΩ min
0 to 30 VAC
±(0 to 15) VDC
±(20 to 150) VDC Less than 1 ms Less than 1 ms Less than 3 ms Less than 3 ms with 100 VDC with 200 VDC
250 V rms
(between unit and channels, between channels) More than 100 MΩ / 500 VDC
(between unit and channels, between channels)
±(0 to 43) VDC 170 to 250 VAC
±(70 to 250) VDC
44
Connect the power cables from the probe s to the output connectors of the Model 9687 Probe Power Unit.
Power Output Connector
Connections

2.7 Supplying power from the Model 9687 Probe Power Unit

2.7 Supplying powe r from the Model 9687 Probe Power Unit
The optional Model 9 687 Prob e Power Unit c an be used to s upply p ower to our optional clamp-on and differential probes.
Probe models compatible with the Model 9687 Probe Power Unit
• Models 3273-50, 3274, 3275, 3276 Clamp-On Probes
• Model 9322 Differential Probe (using Model 9248 Power Cord) The power output connectors on the Model 9687 are especially designed for our
clamp-on and differential probes. To avoid damage, do not use the power supply for any purpose other than supplying power.
There are limits to how many probes can be used with the instrument with a 9687 installed. Avoid exceeding the quanti ties in the following table for all probes to be used. When mixing probe types, the lower quantity of usable probes applies.
Probe Usable Quantity 3273-50, 3274, 3276 Clamp-On Probe 6 3275 Clamp-On Probe 5 9322 Diffe rential Probe 8
9687 Probe Power Unit Specifications
See "9687 Probe Power Unit (specify option when ordering)" in "(8) Power Supply
Options" in "15.1 General Specifications" in the Instruction Manual. (p. 400)
45
• Model 8936 Analog Unit (p. 46)
• Model 8937 Voltage/Temp Unit (p. 47)
• Model 8938 FFT Analog Unit (p. 50)
• Model 8939 Strain Unit (p. 51)
• Model 8940 F/V Unit (p. 52)
• Model 8946 4-Ch Analog Unit (p. 46)
• Model 8947 Charge Unit (p. 58)
• Model 8956 Analog Unit (p. 46)
• Model 8957 High Resolution Unit (p. 61)
• Model 8958 16-Ch Scanner Unit (p. 62)
• Model 8959 DC/RMS Unit (p. 65)
• Model 8960 Strain Unit (p. 51)
• Model 8961 High Voltage Unit (p. 66)
Level Monitor
Shows the input level whi le makin g settings.
See "5.3 Monitoring Input Status" in the
Instruction Manual
Comments can be entered for each channel.
See "5.2 Adding Comments" in the Instruction
Manual
Select the module and channel number to be set.
The display color of acquired waveforms, zero posi-
tion, vertical magnific ation and display range ca n be
changed.
See "7.1 Making Input Waveform Display Settings
(Analog Waveforms)" in the Instruction Manual
Set the scaling if you want to convert the measured
units for display.
Measured val ues ( outp ut vo ltag e) fr om se nsor s can
be converted to physical values appropriate to the
measurement object.
See "5.4 Converting Input Values (Scaling Func-
tion)" in the Instruction Manual
Set up the input channel.
Display appearance depends on the particular input
module. Refer to the description of each module.
Make settings on the Channel Settings screen.
Measurement Range
Indicates the selected range.
1 Press the SET key. 2 Press the SUB MENU keys to select the [Channel] menu. 3 Press the SHEET/PAGE keys to select the [One Ch] page. 4 Use the CURSOR keys to move among the setting items,
and press the F1 to F8 keys to select a setting.
Displayed Range
Indicates the ra nge displa yed on the scre en. To c hange t he displ ay pos ition or range, change the waveform display settings.
Input Channel Settings Chapter 3
After turning power on, set the measurement criteria before measuring. This chapter des crib es the i npu t channel setting procedu re s f or each in put mod ­ule. Refer to "3.11 About Setting Contents" (p. 67) for details of each setting.
Refer to the Instruction Manual for other settings.
46
Example: Setting a channel on the Model 8936 Analog Unit
8
7
Execute to adjust input signal offset. (p. 79)
12 3
4 5
([One Ch] Page of Channel Setting Screen)
6

3.1 Analog Unit Settings (Models 8936, 8946 and 8956)

3.1 Analog Unit Settings (Mo dels 8936, 8946 and
8956)
Set the input channel for voltage me asurement when using the following input modules:
• Model 8936 Analog Unit
• Model 8956 Analog Unit
• Model 8946 4-Ch Analog Unit
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 ke ys .
Step
Select the module No.
1
Select the channel No .
2
Confirm the input module and mea-
3
surement mode
Select the measurement range
4
Select the input coupling method
5
Select the low-pass filter
6
Select the probe attenuation
7
Perform zero adjustment
8
Setting Item
[Unit ]
[Ch ]
[Module]
[Mode]
[Range]
[Coupling]
[LPF]
[Probe]
[
Zero-Adjust
Set up scaling
Description Select the number of the input module to set up.
("Channel Configuration" (p. 17)) Select the channel number on the input module.
Confirm the type of the se lected module.
• Model 8936 Analog Unit..............[Analog]
• Model 8956 Analog Unit..............[H-Speed]
• Model 8946 4-Ch Analog Unit.....[4CH]
Confirm that the display is set for [Voltage] Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 90).
Select the coupling method for input signals (p. 68).
[ DC, AC or GND ]
To suppress hi gh- freq uenc y com ponen ts, select a low- pas s filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
Select according to the cable or probe being used (p. 78).
[ 1:1, 10:1 , 100:1 or 1000:1 ]
]
Perform at least 30 minutes after turning power on
(p. 79).
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
.
(in the Model 8936 case)

3.2 Model 8937 Voltage and Temperature Unit Settings

10
([One Ch] Page of Channel Setting Screen)
8
Execute to adjust input signal offset. (p. 79)
12 3
5 6
4
9
Displayed when the measurement range is 500 µV to 2 mV.
7
3.2 Model 8937 Voltage and Temperature Unit Settings

3.2.1 Voltage Measurement

47
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Select the module No.
1
Select the channel No.
2
Confirm the module
3
Select the measurement mode
4
Select the measurement range
5
Select the input coupling method
6
Select the low-pass filter
7
Select the probe attenuation
8
Setting Item
[Unit ]
[Ch ]
[Module]
[Mode]
[Range]
[Coupling]
[LPF]
[Probe]
Description Select the number of the input module to set up.
("Channel Configuration" (p. 17)) Select the channel number on the input module.
Confirm that the display is set for [Volt/Temp].
Select [Voltage].
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 92).
Select the coupling method for input signals (p. 68).
[ DC, AC or GND ]
To suppress high-frequency components, select a low-pass filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
Select according to the cable or probe being used (p. 78).
[ 1:1, 10:1, 100:1 or 1000:1 ]
Set the digital filter
9
Perform zero adjustment
10
Set up scaling
[Digital F]
[
Zero-Adjust
(Effective measurement ranges: 500 µV to 2 mV) Set the digital filter function to be enabled or disabled. Set it to ON to suppress noise components (p. 70).
[ Off or On ]
]
Perform at least one hour after turning power on
(p. 79).
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
48
([One Ch] Page of Channel Setting Screen)
8
12 3
5
6
4
9
The [Voltage] measurement mode is selected upon system reset.
7
3.2 Model 8937 Voltage and Temperature Unit Settings

3.2.2 Temperature Measurement

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 ke ys .
Step
Setting Item
Description
Select the module No.
1
Select the channel No .
2
Confirm the module
3
Select the measurement mode
4
Select thermocouple
5
Select the measurement range
6
Select the low-pass filter
7
Set the reference junction compensa-
8
tion
[Unit ]
[Ch ]
[Module]
[Mode]
[Sensor]
[Range]
[LPF]
[RJC]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [Volt/Temp].
Select [Temp].
Select the type of thermocouple (p. 71).
[ K, J, E, T, N, R, S or B ]
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67).
[ 10, 20, 50 or 100°C/div ]
To suppress hi gh- freq uenc y com ponen ts, select a low- pas s filter within the input module (p. 69).
[ Off, 5Hz or 500Hz ]
Select whether reference junction compensation should be provided internally (p. 71).
[ Internal or External ]
Set drift correction
9
Set up scaling
[Drift]
Set whether to cancel variations in th e reference volt age. Set this On to improve the temperature characteristic (p. 72).
[ Off or On ]
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
49
3.2 Model 8937 Voltage and Temperature Unit Settings
• The upper and lowe r limits of measure ment input depend on the measure­ment range. Refer to "About Measurement Rang e, Upper and Lower Input Limits and the Waveform Display:" (p. 67).
• When the input terminals of the Model 8937 Voltage/Temp Unit are exposed to strong drafts: Loss of thermal e quilibrium of the input se ction may result in meas urement errors. T o measure in such environments, reposition the equipment so that the input terminals are protected from the drafts.
• When the ambient temperature changes rapidly: Loss of thermal e quilibrium may r esult in measurem ent errors. If this occurs, allow about an hour for the equip ment to acclim atize to the new te mperature, then take measurements after thermal equilibrium is achieved.
• In an environment wit h fluc tuating ambie nt temp erature, when r ecordi ng for a long time using a thermoco uple with weak thermo elect romotiv e force (R, S or B sensors): Temperature characteristic can be improved by setting drift correction On.
•The [Voltage] measurement mode is selected upon system reset.
50
9
([One Ch] Page of Channel Setting Screen)
7
Execute to adjust input signal offset. (p. 79)
12 3
4 5
8
6

3.3 Model 8938 FFT Analog Unit Settings

3.3 Model 8938 FFT Analog Unit Settings
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 ke ys .
Step
Setting Item
Description
Select the module No.
1
Select the channel No .
2
Confirm the input module and mea-
3
surement mode
Select the measurement range
4
Select the input coupling method
5
Select the low-pass filter
6
Select the probe attenuation
7
Set the anti-aliasing filter
8
[Unit ]
[Ch ]
[Module] [Mode]
[Range]
[Coupling]
[LPF]
[Probe]
[AAF]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [FFT]. Confirm that the display is set for [Voltage].
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 94).
Select the coupling method for input signals (p. 68).
[ DC, AC or GND ]
To suppress hi gh- freq uenc y com ponen ts, select a low- pas s filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
Select according to the cable or probe being used
(p. 78).
[ 1:1, 10:1, 100:1 or 1000:1 ]
Turn this on to suppress anti-aliasing distortion, such as for FFT analysis (p. 70).
[ Off or On ]
Perform zero adjustment
9
Set up scaling
[
Zero-Adjust
]
Perform at least 30 minutes after turning power on
(p. 79).
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
51
([One Ch] Page of Channel Setting Screen)
12 3
4
7
5
6
8939
8960
7
8 8
(Model 8960 only)
(Model 8960 only)

3.4 Strain Unit Settings (Models 8939 and 8960)

3.4 Strain Unit Settings (Models 89 39 and 8960)
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
Select the module No.
1
Select the channel No.
2
Confirm the input module and mea-
3
surement mode
Select the measurement range
4
Select the bridge voltage
5
Set the anti-aliasing filter
6
Select the low-pass filter
7
Execute auto-balance
8
Set up scaling
[Unit ]
[Ch ]
[Module] [Mode]
[Range]
[Bridge]
[AAF]
[LPF]
[Auto­Balance]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [Strain].
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. (Units: [µ] Microstrain = 1/2 µV/V) The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 95).
Select the bridge voltage (p. 78).
[ 2V, 5V, 10V ]
Turn this on to suppress anti-aliasing distortion, such as for FFT analysis (p. 70).
[ Off or On ]
To suppress high-frequency components, select a low-pass filter within the input module (p. 69).
8939: [ Off, 10Hz, 30Hz, 300Hz or 3kHz ] 8960: [ OFF, 5Hz, 10Hz, 100Hz, 1kHz ]
Sets the reference outpu t level of th e transducer to the spec­ified zero position Execute one hour after turning the power on, with the sens or in the unloaded state.
Set as occasion demands.
See
"Scaling Setting Examples" in "5.4 Converting Input Values (Scaling Function)" in the Instruction Manual
(p. 81).
52
([One Ch] Page of Channel Setting Screen)
8
12 3
5
4
9
Example: Measuring frequency
6
7
10
The [Frequency] measurement mode is selected upon system reset.

3.5 Model 8940 F/V Unit Settings

3.5 Model 8940 F/V Unit Settings
Settings are related to the follo win g mea su rements. One measurement type ca n be set per channel.
• Measure freq uency from the input p ulse corresponding t o the measurement waveform (p. 52)
• Measure frequencies near 50/60 Hz from the input pulse corresponding to the measurement waveform (p. 52)
• Measure rotation rate from the input pulse corresponding to the measurement waveform (p. 52)
• Count of input pulses (p. 54)
• Measure the perc en tage of Hig h level of a single pulse w av eform ( pu ls e dut y)
(p. 55)
• Measure voltage (p. 56)
• Measure current (p. 57)

3.5.1 Measuring Frequency, 50/60 Hz and Rotation Rate

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 ke ys .
Step
Select the module No.
1
Select the channel No .
2
Setting Item
[Unit ]
[Ch ]
Description Select the number of the input module to set up.
("Channel Configuration" (p. 17)) Select the channel number on the input module.
Confirm the module
3
Select the measurement mode
4
Select the measurement range
5
Select the low-pass filter
6
[Module]
[Mode]
[Range]
[LPF]
Confirm that the display is set for [F/V].
Select the measurement mode.
[ Frequency, Rotation, 50/60 Hz ]
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 96).
To suppress high-frequency components, select a low-pass filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
53
3.5 Model 8940 F/V Unit Settings
Select the probe attenuation
7
Set the threshold level
8
Set pull-up
9
Set the Hold function
10
Set up scaling
[Probe]
[Threshold]
[Pull-Up]
[Hold]
Select according to the cable or probe being used
(p. 78).
[ 1:1, 10:1, 100:1, 1000:1(9322+ 93 25) or 1000:1 (9322 +941 8) ]
Set the threshold value (p. 75).
Enable (set On) th e pull-up resist ance when connecting to an open-collector output signal. Disable it (Off) for normal mea­surements (p. 74).
[ Off or On ]
Select whether to h old the last measure d v al ue unti l t he nex t frequency or rotation rate is determined (p. 73).
[ On, 10-ms Off or 1-s Off ]
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
Difference between frequency measurement and 50 /60 Hz f reque ncy measurement
The frequency measurement function consists of selecting a range from 50 mHz to 5 kHz and observing changes in frequency. The 50/60 Hz frequen cy measurement function cons ists of setting either 50 or 60 Hz as the zero position, and observing deviations from that frequency.
54
([One Ch] Page of Channel Setting Screen)
8
12 3
5
4
6
7
9
The [Frequency] measurement mode is selected upon system reset.
3.5 Model 8940 F/V Unit Settings

3.5.2 Pulse Count Measurement

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 ke ys .
Step
Setting Item
Description
Select the module No.
1
Select the channel No .
2
Confirm the module
3
Select the measurement mode
4
Select the measurement range
5
Select the low-pass filter
6
Select the probe attenuation
7
Set the threshold level
8
[Unit ]
[Ch ]
[Module]
[Mode]
[Range]
[LPF]
[Probe]
[Threshold]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [F/V].
Select [Count].
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. (Units: c = count) The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 96).
To suppress hi gh- freq uenc y com ponen ts, select a low- pas s filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
Select according to the cable or probe being used
(p. 78).
[ 1:1, 10:1, 100: 1, 1000: 1(9322+ 9325) or 100 0:1(932 2+9418 ) ]
Set the threshold level (p. 75).
Set pull-up
9
Set up scaling
[Pull-Up]
Enable (set On) the pu ll-up resistanc e when connectin g to an open-collector output sign al. Disab le it (O ff) for normal mea ­surements (p. 74).
[ Off or On ]
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
3.5 Model 8940 F/V Unit Settings
([One Ch] Page of Channel Setting Screen)
7
12 3
4
5
6
8
The [Frequency] measurement mode is selected upon system reset.
Calculation (50 µs)
(dead time)
Waveform Acquisition
Waveform Acquisition
Ignored

3.5.3 Pulse Duty Measurement

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
55
Select the module No.
1
Select the channel No.
2
Confirm the module
3
Select the measurement mode
4
Select the low-pass filter
5
Select the probe attenuation
6
Set the threshold level
7
Set pull-up
8
Set up scaling
[Unit ]
[Ch ]
[Module]
[Mode]
[LPF]
[Probe]
[Threshold]
[Pull-Up]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [F/V].
Select [Duty].
To suppress high-frequency components, select a low-pass filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
Select according to the cable or probe being used
(p. 78)
[ 1:1, 10:1, 100:1, 1000: 1(9322+9 325) o r 1000 :1(9 322+9418 ) ]
Set the threshold level (p. 75).
Enable (set On) the pull-up resistanc e when connectin g to an open-collector outp ut si gna l. Dis abl e it (Off) for normal mea­surements
[ Off or On ]
Set as occasion demands.
See
(p. 74).
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
Pulse duty measurement
If a rising puls e (at 2 0 kHz o r above) is appl ied during calcu lation (“ dead time ”), the duty during th e dead time cann ot be measured. D uty is measured f rom the pulse following the dead time.
56
10
([One Ch] Page of Channel Setting Screen)
8
Execute to adjust input signal offset. (p. 79)
12 3
5 6
4
9
The [Frequency] measurement mo de is selected upon system rese t.
7
3.5 Model 8940 F/V Unit Settings

3.5.4 Voltage Measurement

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 ke ys .
Step
Setting Item
Description
Select the module No.
1
Select the channel No .
2
Confirm the module
3
Select the measurement mode
4
Select the measurement range
5
Select the input coupling method
6
Select the low-pass filter
7
Select the probe attenuation
8
[Unit ]
[Ch ]
[Module]
[Mode]
[Range]
[Coupling]
[LPF]
[Probe]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [F/V].
Select [Voltage].
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 96).
Select the coupling method for input signals (p. 68).
[ DC, AC or GND ]
To suppress hi gh- freq uenc y com ponen ts, select a low- pas s filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
Select according to the cable or probe being used
(p. 78).
[ 1:1, 10:1, 100: 1, 1000: 1(9322+ 9325) or 100 0:1(932 2+9418 ) ]
Set pull-up
9
Perform zero adjustment
10
Set up scaling
[Pull-Up]
[
Zero-Adjust
Enable (set On) the pu ll-up resistanc e when connectin g to an open-collector output sign al. Disab le it (O ff) for normal mea ­surements (p. 74).
[ Off or On ]
]
Perform at least 30 minutes after turning power on
(p. 79).
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
3.5 Model 8940 F/V Unit Settings
9
5
Execute to adjust input signal offset. (p. 79)
12 3
6 7
4
The [Frequency] measurement mode is selected upon system reset.
([One Ch] Page of Channel Setting Screen)
8

3.5.5 Current Measurement

When using the Mode l 9018-50 and 9132-50 Clamp-On Probe, s elect [Voltage] mode.
See "3.5.4 Voltage Measurement" (p. 56)
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
57
Select the module No.
1
Select the channel No.
2
Confirm the module
3
Select the measurement mode
4
Select a sensor
5
Select the measurement range
6
Select the input coupling method
7
[Unit ]
[Ch ]
[Module]
[Mode]
[Sensor]
[Range]
[Coupling]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [F/V].
Select [Current].
Select the clamp to be used.
[ 3273-50, 9270, 9271, 9272LO* 9279, 20A/2V*2, 200A/2V*2, 500A/2V*2 ]
*1 9272HI: 200A-range, 9272LO: 20A-range *2 When connecting a sensor (Hioki Clamp-On Sensor or
AC/DC Current Sensor) with a model name that cannot be selected in this screen, select a conversion ratio. Re­fer to the instruction manual to determine whether the sensor can be connected to the instrument or not.
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 96).
Select the coupling method for input signals (p. 68).
[ DC, AC or GND ]
1
, 9272HI*1, 9277, 9278 or
Select the low-pass filter
8
Perform zero adjustment
9
[LPF]
[
Zero-Adjust
• Up to eight channels can be used simultaneously when measuring with the Model 9318 and 9319 Conversion Cables.
• When using a clamp model that is not di splayed in the sensor sel ection list, measure in [Voltage] mode, and apply the scaling function.
See:"5.4 Converting Input Va lu es (Sca lin g Func tio n)" in the Instruction Manual
To suppress high-frequency components, select a low-pass filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
]
Perform at least 30 minutes after turning power on
(p. 79).
58
10
([One Ch] Page of Channel Setting Screen)
8
Execute to adjust input signal offset. (p. 79)
12 3
5 6
4
9
The [Charge] measurement mode is selected upon system reset.
7

3.6 Model 8947 Charge Unit Settings

3.6 Model 8947 Charge Unit Settings
Make settings regarding input channels for measuring voltage or acceleration (Charge and Preamp). One type can be measured per channel.

3.6.1 Voltage Measurement

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 ke ys .
Step
Select the module No.
1
Select the channel No .
2
Confirm the module
3
Select the measurement mode
4
Select the measurement range
5
Select the input coupling method
6
Select the low-pass filter
7
Select the probe attenuation
8
Setting Item
[Unit ]
[Ch ]
[Module]
[Mode]
[Range]
[Coupling]
[LPF]
[Probe]
Description Select the number of the input module to set up.
("Channel Configuration" (p. 17)) Select the channel number on the input module.
Confirm that the display is set for [Charge].
Select [Voltage].
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 99).
Select the coupling method for input signals (p. 68).
[ DC, AC or GND ]
To suppress hi gh- freq uenc y com ponen ts, select a low- pas s filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
Select according to the cable or probe being used (p. 78).
[ 1:1, 10:1, 100:1 or 1000:1 ]
Set the anti-aliasing filter
9
Perform zero adjustment
10
Set up scaling
[AAF]
[
Zero-Adjust
Turn this on to suppress anti-aliasing distortion, such as for FFT analysis (p. 70).
[ Off or On ]
]
Perform at least one hour after turning power on
(p. 79).
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
3.6 Model 8947 Charge Unit Settings
10
([One Ch] Page of Channel Setting Screen)
9
12 3
6 7
4
5
The [Charge] measurement mode is selected upon system reset.
8

3.6.2 Acceleration Measurement (Charge, Preamp)

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
59
Select the module No.
1
Select the channel No.
2
Confirm the module
3
Select the measurement mode
4
Set sensor sensitivity
5
Select the measurement range
6
Select the input coupling method
7
[Unit ]
[Ch ]
[Module]
[Mode]
[Sensitivity]
[Range]
[Coupling]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [Charge].
Select the measurement mode.
[Charge]: When using a charge-output acceleration pick­up sensor
[Preamp]: When using an acceleration pick-up sensor equipped with an internal preamp
Set the value per m/ s2 defined for the acceleration sensor to be used
"Sensor Sensitivity Setting Examples" (p. 76)
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. (Units: m/s The RANGE/POSN knob selects the range. Refer to the specifications for available ranges (p. 99).
Select the coupling method for input signals (p. 68).
[ AC or GND ]
(p. 76).
2
= acceleration)
Select the low-pass filter
8
Set the anti-aliasing filter
9
Perform zero adjustment
10
Set up scaling
[LPF]
[AAF]
[
Zero-Adjust
To suppress high-frequency components, select a low-pass filter within the input module (p. 69).
[ Off, 500Hz or 5kHz ]
Turn this on to suppress anti-aliasing distortion, such as for FFT analysis
[ Off or On ]
]
Perform at least one hour after turning power on
(p. 79).
Set as occasion demands.
See
"To convert [m/s2] units to [G] for disp lay " (p . 76 ) in th is ma n­ual, "5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
(p. 70).
60
5 to 6 seconds
3.6 Model 8947 Charge Unit Settings
Supplement ___________________________________________________
Before connecting sensors and probes
To avoid electric shock or damage to the measurement object, turn the instrument off before connecting a sensor or probe to the BNC jack. When the [Preamp] measurement mode is selected, internal power (15 V @ 2 mA) is applied to the BNC jack when measurement starts.
Range Setting
Auto-ranging cannot be used on channels for which the measurement mode is set to
[Charge] or [Preamp].
For more on auto-ranging, refer to "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the Instruction Manual.
Measuring Timing
In the following cases, five or six seconds should be allowed for input stabilization:
• Measurement Mode: When switching the mea suremen t rang e betwe en one of th e six h igh- sensiti v­ity ranges and one of the six low-sensitivity ranges.
• Measurement Mode: After selecting preamp mode and before starting a new measurement
[Charge]
[Preamp]

3.7 Model 8957 High Resolution Unit Settings

9
([One Ch] Page of Channel Setting Screen)
7
Execute to adjust input signal offset. (p. 79)
12 3
4 5
8
6
3.7 Model 8957 High Resolution Unit Settings
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
61
Select the module No.
1
Select the channel No.
2
Confirm the input module and mea-
3
surement mode
Select the measurement range
4
Select the input coupling method
5
Select the low-pass filter
6
Select the probe attenuation
7
Set the anti-aliasing filter
8
[Unit ]
[Ch ]
[Module] [Mode]
[Range]
[Coupling]
[LPF]
[Probe]
[AAF]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [High Res]. Confirm that the display is set for [Voltage].
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 102).
Select the coupling method for input signals (p. 68).
[ DC, AC or GND ]
To suppress high-frequency components, select a low-pass filter within the input module (p. 69).
[ Off, 5Hz, 50Hz, 500Hz, 5kHz or 50kHz ]
Select according to the cable or probe being used
(p. 78).
[ 1:1, 10:1, 100:1 or 1000:1 ]
Turn this on to suppress anti-aliasing distortion, such as for FFT analysis (p. 70).
[ Off or On ]
Perform zero adjustment
9
Set up scaling
[
Zero-Adjust
]
Perform at least 30 minutes after turning power on
(p. 79).
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
62
([One Ch] Page of Channel Setting Screen)
Execute to adjust input signal offset. (p. 79)
12 3
5
4
6

3.8 Model 8958 16-Ch Scanner Unit Settings

3.8 Model 8958 16-Ch Scanner Unit Settings
Before Setting
The zero position of the Model 8958 16-Ch Scanner Unit needs to be adjusted in the following cases.
See "Scanner Unit Zero Position Adjustmen t" (p. 29)
• When the Model 8958 16-Ch Scanner Unit has just been installed in the
instrument (a message requesting adjustment appears when the instrument is turned on)
• When the zero position has shifted due to aging or environmental changes

3.8.1 Voltage Measurement

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 ke ys .
Step
Select the module No.
1
Select the channel No .
2
Confirm the module
3
Select the measurement mode
4
Select the measurement range
5
Set the digital filter
6
Set up scaling
Setting Item
[Unit ]
[Ch ]
[Module]
[Mode]
[Range]
[Digital F]
Description Select the number of the input module to set up.
("Channel Configuration" (p. 17)) Select the channel number on the input module.
Confirm that the display is set for [Scanner].
Select [Voltage].
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 103).
Set the digital filter function to be enabled or disabled. Set it to ON to suppress noise components (p. 70).
[ Off, 60Hz, 50Hz or 10Hz ]
Set as occasion demands.
"5.4 Converting Input Values (Scaling Function)" in the
See
Instruction Manual
3.8 Model 8958 16-Ch Scanner Unit Settings
([One Ch] Page of Channel Setting Screen)
8
12 3
5
6
4
9
7
The [Voltage] measurement mode is selected upon system reset.

3.8.2 Temperature Measurement

Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
63
Select the module No.
1
Select the channel No.
2
Confirm the module
3
Select the measurement mode
4
Select thermocouple
5
Select the measurement range
6
Set the digital filter
7
Set the reference junction compensa-
8
tion
[Unit ]
[Ch ]
[Module]
[Mode]
[Sensor]
[Range]
[Digital F]
[RJC]
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [Scanner].
Select [Temp].
Select the type of thermocouple (p. 71).
[ K, J, E, T, N, R, S, B or W ]
Select the appropriate measurement range. The setting val­ue is the units per division on the vertical axis. The RANGE/POSN knob selects the range (p. 67).
[ 10, 50 or 100°C/div ]
Set the digital filter function to be enabled or disabled. Set it to ON to suppress noise components (p. 70).
[ Off, 60Hz, 50Hz or 10Hz ]
Select whether reference junction compensation should be provided internally
[ Internal or External ]
(p. 71).
Set disconnect detection
9
Set up scaling
[Burn-Out
]
Select thermocouple broken-wire detection (p. 72). When enabled (On), the waveform trace of a broken wire stays at the top of the screen and the cursor value is dis­played as “+OVER” when a broken wire is present.
[ Off or On ]
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
64
3.8 Model 8958 16-Ch Scanner Unit Settings
• The upper and l ower limits of measurement input depend on the measure-
ment range. Refer to "About Measurement Range, Upper and Lower Input
Limits and the Waveform Display:" (p. 67).
• When the input terminals of the Mod el 8958 16-Ch Scanner Unit are subject
to strong drafts: Loss of thermal eq uilibrium of the input section may result in measurement errors. T o measure in such environments, install the supplied cover or relocate the equipment out of the wind.
• When the ambient temperature changes rapidly:
Loss of thermal equilibrium may result in measurement erro rs. If this occurs, allow about an hour for the equip ment to acclima tize to the new tem perature, then take measurements after thermal equilibrium is achieved.

3.9 Model 8959 DC/RMS Unit Settings

10
([One Ch] Page of Channel Setting Screen)
8
Execute to adjust input signal offset. (p. 79)
12 3
5 6
4
9
Example: Measuring RMS values
The [DC] measurem ent mode is se­lected upon system reset.
7
3.9 Model 8959 DC/RMS Unit Settings
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
65
Select the module No.
1
Select the channel No.
2
Confirm the module
3
Select the measurement mode
4
Select the measurement range
5
Select the input coupling method
6
Select the low-pass filter
7
Select the probe attenuation
8
[Unit ]
[Ch ]
[Module]
[Mode]
[Range]
[Coupling]
[LPF]
[Probe]
Select the number of the input module to set up. ("Channel Configurati on" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [DC/RMS].
Select the me asurement mode.
[DC]: Voltage measurement
[RMS]: RMS measurement Select the voltage per division on the voltage (vertical) axis.
The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 105).
Select the coupling method for input signals (p. 68).
[ DC, AC or GND ]
To suppress high-frequency components, select a low-pass filter within the input module (p. 69).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
Select according to the cable or probe being used
(p. 78).
[ 1:1, 10:1, 100:1 or 1000:1 ]
Set the response speed
9
Perform zero adjustment
10
Set up scaling
[Response]
[
Zero-Adjust
Select the response speed (p. 75).
[ Fast, Medium or Slow ]
]
Perform at least 30 minutes after turning power on
(p. 79).
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
66
8
([One Ch] Page of Channel Setting Screen)
Execute to adjust input signal offset. (p. 79)
12 3
5 6
4
Example: Measuring RMS values
The [DC] measurem ent mode is se­lected upon system reset.
7

3.10 Model 8961 High Voltage Unit Settings

3.10 Model 8961 High Voltage Unit Settings
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
Select the module No.
1
Select the channel No.
2
Confirm the module
3
Select the measurement mode
4
Select the measurement range
5
Select the input coupling method
6
Select the low-pass filter
7
Perform zero adjustment
8
[Unit ]
[Ch ]
[Module]
[Mode]
[Range]
[Coupling]
[LPF]
[
Zero-Adjust
Select the number of the input module to set up. ("Channel Configuration" (p. 17))
Select the channel number on the input module.
Confirm that the display is set for [High Volt].
Select the measurement mode.
[DC]: Voltage measurement
[RMS]: RMS measurement Select the voltage per division on the voltage (vertical) axis.
The RANGE/POSN knob selects the range (p. 67). Refer to the specifications for available ranges (p. 107).
Select the coupling method for input signals (p. 68).
[ DC or GND ]
To suppress high-frequ ency components, select a low-pass filter within the input module (p. 69).
[ Off, 5Hz, 50Hz, 500Hz or 5kHz ]
]
Perform at least 30 minutes after turning power on
(p. 79).
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
3.11 About Setting Contents
Minimum Resolution

3.11.1 Measurement Range Setting

Select the approp riate measurement range. The setting value is the units per division on the vertical axis. Use the RANGE/POSN k nob to make the setti ng independ ently from the cu rsor position.
Setting Item: [Range]
•Refer to "Chapter 5 Specifications" (p. 87) for the extents of the measureme nt ranges.
• If the range is undetermined, auto-range selection can be enabled. Refer to "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the
Instruction Manual
• When data has been stored in the instrument’s memory, an asterisk “∗” appears beside the data value’s measurement range.
67

3.11 About Setting Contents

If the range is changed while measuring, acquired data is deleted and data recording is restarted.
Voltage Axis Range
When variable auto adjustment is enabled [On] ( de faul t s etting) an d the va riabl e function is enabl ed, the varia ble s etting bec omes linked for measure ment ran ge changes. If you do not want the display area of the screen to change, disable variable auto adjustment [Off].
See "13.2.8 Performing Variable Function Auto Adjustment" in the Instruction Manual
Temperature Measurement About Measurement Range, Upper and Lower Input Limits and the Wave-
form Display:
The upper and lower limits of meas urement input depend on the meas urement range. Measurement is no t available outsid e of the temperature range limits in the fol­lowing tables:
Model 8937 Voltage/Temp Unit
Measurement Range [°C/div ] 10 20 50 100
Upper Limit [°C] 400 800 2000 4000 Lower Limit [°C] -92 -184 -460 -920
Model 8958 16-Ch Scanner Unit
Measurement Range [°C/div] 10 50 100
Upper Limit [°C] 200 1000 2000 Lower Limit [°C] -100 -200 -200
68
Measurement range: 50°C/div (-460 to 2000°C) When T measuremen t input range exce eded
Measurement range: 20°C/div (-184 to 800°C) Lower limit of measurement range (-184°C) exceeded
Upper limit of mea­surement input range exceeded
Lower limit of mea­surement range ex­ceeded
Upper limit of T
measurement input
Lower limit of T
measurement input
Lower limit of mea­surement input range exceeded
Thermocouple: T Measurement Range
3.11 About Setting Contents
Example: Zero position = 50%, Magnification = ×1, Thermocouple = type T (using Model 8973) (Measurement input range of T = -200 to 400°C

3.11.2 Setting Input Coupling

Select the coupling method for input signals Setting Item: [Coupling]
Selections Description
DC Coupling (Default setting) Select this to acquire both DC and AC components of an input signal.
AC Coupling Select this to eliminate any DC component from an input signal. Use this to measure only the ripple component superimposed on pulsating currents.
The input signal is disconnected. Zero position can be confirmed.
Voltage DC AC GND Temp DC
Frequency, Rotation, 50/60Hz, Count, Duty
Voltage DC AC GND Current
Voltage DC AC GND Charge, Preamp
DC
AC
GND
Available selections depend on the input module and measurement mode.
Input modules Modes Selections 8936 Voltage DC AC GND
8937
8938 FFT DC AC GND
8940
8946 Voltage DC
8947
8959 DC, RMS DC AC GND 8956 Voltage DC AC GND 8957 Voltage DC AC GND 8961 DC, RMS DC
−−
DC
depends on the particular clamp
−−
AC GND
GND
GND

3.11.3 Low-Pass Filter (LPF) Settings

To suppress high-frequency components, select a low-pass filter within the input module
Setting Item: [LPF]
Selections Description
Off 5Hz 500Hz 5kHz 100kHz
Available selections depend on the input module and measurement mode.
Input modules Modes Selections [Hz]
The low-pass filter is disabled. (Default setting) Applies a filter with 5-Hz cutoff frequency. Applies a filter with 500-Hz cutoff frequency. Applies a filter with 5-kHz cutoff frequency. Applies a filter with 100-kHz cutoff frequency.
69
3.11 About Setting Contents
8936 Voltage Off, 5, 500, 5 k, 100 k 8937 * Voltage Off, 5, 500, 5 k, 100 k
Temp Off, 5, 500 8938 Voltage Off, 5, 500, 5 k, 100 k 8939 Strain Off, 10, 30, 300, 3 k 8940 Voltage Off, 5, 500, 5 k, 100 k 8946 Voltage Off, 5, 500, 5 k, 50 k 8947 Voltage Off, 5, 500, 5 k, 100 k
Charge, Preamp Off, 500, 5 k 8956 Voltage Off, 5, 500, 5 k, 1 M 8957 Voltage Off, 5, 50, 500, 5 k, 50 k 8959 DC, RMS Off, 5, 500, 5 k, 100 k 8960 Strain Off, 5, 10, 100, 1 k 8961 DC, RMS Off, 5, 50, 500, 5 k
* When the digital filter is enable d [On], only O ff, 5 Hz and 500 Hz selection s are available .
70
3.11 About Setting Contents

3.11.4 Anti-Aliasing Filter (AAF) Settings

Enable the anti-al iasing filter to remove ali asing distortio n. The cutoff fr equency is automatically set according to the time axis range setting.
Setting Item: [AAF]
Selections Description
Off The anti-aliasing filter is disabled. (Default setting)
The anti-aliasing filter is enabled.
On
To perform sampling for analysis without being affected by aliasing distortion
• We recommend using an input module that supports anti-aliasing filtering. (Supporting input module models: 8938, 8947 and 8957)
• Refer to "3.11 FFT Definitions" in the Analysis and Communication Supplement for details of the Aliasing and Anti-aliasing filters.
(When the Recorder func tion or External sam pling is used, th e anti­aliasing filter (AAF) is not available.)
Printing setting data (List Print)
Settings can be printed as follows: (Example) When enabled: “AAF ON”

3.11.5 Digital Filter (Digital F) Settings

Noise can be removed by applying additional averaging to the measurement data within the input module. With the 8937 Voltage /Temp Unit, digital filtering is avail able only in the mea­surement ranges from 500 µV to 2 mV/div. The digital filter is not available in other ranges.
Setting Item: [Digital F]
Selections Description
8937
8958
Off
On Off The digital filter function is disabled. (Default setting)
60Hz, 50Hz or 10Hz
The digital filter function is disabled. (Default setting) The digital filter function is enabled.
(Data refresh rate: Approx. 100 µs)
The digital filter function is enabled. The data refresh rate is affected by the filter setting.
See "5.2.10 Model 8958 16-Ch Scanner Unit" (p. 103)
Printing setting data (List Print)
Settings can be printed as follows: Example: Enabled = “Digital F ON"
3.11 About Setting Contents

3.11.6 Thermocouple (Sensor) Type Setting

Set to match the type of thermocouple being used. Setting Item: [Sensor]
8937 Voltage/Temp Unit
71
Selections
K -200 to 1350°C R 0 to 1700°C J -200 to 1100°C S 0 to 1700°C E -200 to 800°C B 300 to 1800°C T N -200 to 1300°C
Measurement Range
-200 to 400°C
Selections
Measurement Range
8958 16-Ch Scanner Unit
Selections
K -200 to 1350°C R 0 to 1700°C J -200 to 1200°C S 0 to 1700°C E -200 to 1000°C B 400 to 1800°C T -200 to 400°C W 0 to 2000°C N -200 to 1300°C
Measurement Range
Selections
Measurement Range

3.11.7 Reference Junction Compensation Setting

When connecting a thermocouple directly to the input module, select [Internal]. Reference junction comp ensati on is perform ed with in the inp ut modu le. When connecting t hrough a reference j unction device (e .g., a 0°C control tan k), select [External].
Setting Item: [RJC]
Selections Description
Reference junctio n co mpensation is pro vi ded w i thi n the i npu t mod-
Internal
External
ule. (Default setting) (Measurement Accuracy: Th e sum of the ac curacies of th e temper­ature measurement and the reference junction compensation.)
Reference junction compensation is not provided within the input module. (Measurement Accuracy: The accuracy of temperature measure­ment only)
72
3.11 About Setting Contents

3.11.8 Drift Correction Setting

This function period ically co rrects fo r drift of the reference p otential ( about onc e per second) to improve thermal characteristics in thermocouple mode.
Setting Item: [Drift]
Selections Description
Off On
• Data refresh rate is about once per second when drift correction is enabled.
• Drift correction is a vailable only wh ile measuring. It does not affect the Le vel Monitor on the Waveform and Channel Setting screens.
Drift correction is disabled. (Default setting) Drift correction is enabled.

3.11.9 Disconnect (Burn-Out) Detection Setting

A broken thermocoupl e wire c an be d etected during temper ature mea sure ment. Normally when a thermocouple wire breaks, measured values exhibit random instability.
Setting Item: [Burn-Out]
Selections Description
Off
On
Broken wires are not detected. Broken wires are detected.
When a broken wire is pre sent, t he w aveform trace st ays a t th e top of the screen and the cursor value is displayed as “+OVER”. Wire breakage is detected by sensing a miniscule current flow (about 200 nA) through the thermocouple. If the thermocouple wires are long or composed of a high-resistance material, set
[Burn-Out] to [Off] to avo id me asu rem en t errors.

3.11.10 Hold Setting

Rotation Stops
Hold: On
The previously measured value is retained continuously until the next cycle is determined. Because the next cycle never occurs, the stopping condition is not detected.
Hold: Off
If the cycle is not detected within the specified period (10 ms or 1 s), half of the value of the previous signal is dis­played. As a result, the displa yed valu e graduall y approac hes zero, allowing the stopping condition to be estimated.
Rotation Stops
During measurement, the measured frequency value can be retained by the Hold function until the frequency of the next cycle has been determined. Also, if the value i s not ret ained a nd the n ext value canno t be dete rmin ed within a specified period, one half of the last measured value can be displayed.
Setting Item: [Hold]
Selections Description
73
3.11 About Setting Contents
On
10-ms Off
1-s Off
The on/off settings of the Hold and Pull-Up function s are not indicated on the Waveform screen. The settings are indicated on the Channel Settings screen.
Retains the measured v alue until the next frequen cy value is deter­mined.
If the next frequency va lue is not de termin ed wit hin 10 m s, the last value is divided in half. (Default setting)
If the next frequency value is no t determin ed within 1 s , the last val­ue is divided in half.
What is the difference between Hold On and Off?
When measuring frequency and rotation rate, results are displayed only after the value of one cycle has been determined. When Hold is O n, the previ ously meas ured va lue i s reta ined, and w ith Ho ld O ff, one half of the previous value is displayed if a new value cannot be determined within the specified period.
Example: Measuring the phenome na of stop pin g a rotating body
74
Actual Phenomena
Frequency drops from a certain cycle to 0 Hz, and then increases from that point.
Hold: 10-ms Off
At frequencies below 100 Hz, the displayed value is suc­cessively halved to 0 Hz. However, until the frequency reaches 100 Hz again, the halved signal is now added.
Hold: On
The previous value is retained until the 0-Hz event.
Hold: 1-s Off
At frequencies below 1 Hz, the displayed value is succes­sively halved to 0 Hz. However, until the frequency reaches 1 Hz again, the halved signal is now added.
200
100
0
[Hz]
2
1
0
[Hz]
3.11 About Setting Contents
Example: Measuring frequency dropping to 0 Hz, and then increasing

3.11.11 Pull-Up Setting

The pull-up resistance function is used when connecting to an open-collector output signal. The input terminal is pulled up to +5 V. Set Pull-Up to Off for normal measurements.
Setting Item: [Pull-Up]
Selections Description
Off Pull-up resistance is disable (Off).
On
Pull-up resistanc e is ena bled. (for c onnecti on to open -collecto r out­put)

3.11.12 Threshold Setting

5
[V]
0
-5
Threshold
Threshold Setting Example
Sets threshold value. Measures when the waveform crosses a preset threshold. Setting Item: [Threshold]
Selections Description
75
3.11 About Setting Contents
↑↑ ↑ ↓ ↓↓
Values can also be entered by numeric keypad.
See "3.3.3 Entering Text and Numbers" in the Instruction Manual
Measurement results may differ according to the threshold setting. To obtain the correct measurement results, set the threshold level to suit the input waveform.
Increases value by large steps. Increases value by small steps. Decreases value by small steps. Decreases value by large steps.

3.11.13 Response (Timing) Setting

Response can be set to three speeds: Fast, Medium and Slow. Normally set to [Fast], this can b e c han ged to [Medium] or [Slow] to sta bil iz e t he display when measuring low frequencies, or when severe fluctuations are present.
Setting Item: [Response]
Selections Description
Fast Medium Slow
(The response time of the 8961 High Voltage Unit is fixed at about one second.)
Sets the response time to about 100 ms. Sets the response time to about 800 ms. Sets the response time to about 5 s.
76
3.11 About Setting Contents

3.11.14 Sensor Sensitivity Setting

Sets the value per m/s2 defined for the acceleration sensor to be used.
Setting Item: [Sensitivity]
Setting Range : 0.1 to 10 (pC/(m/s
When using a sensor with specified value per G:
See Example 2 of the "Sensor Sensitivity Setting Examples" (p. 76)
To use a sensor outside of the setting range:
See "To Use a Sensor Outside of the Setting Range" (p. 77)
The measurement r ange can be affected by th e senso r s ens it iv ity. So sensor sensitivit y sh o ul d be se t be fo r e se tt in g th e measurement ra ng e.
Sensor Sensitivity Setting Examples
Setting Example 1:
When using a sensor with sensitivity specified in units of m/s
If the specified sensor sensitivity is 1.08 pC/(m/s2) Sensor sensitivity setting value: 1.08
Setting Example 2:
When using a sensor with sensitivity specified in units of G
An acceleration sensor with sensitivity specified in units of G can be set by divid­ing the specified sensitivity value by 9.8 (m/s If the specified sensor sensitivity is 64.0 (pC/G): 64.0 / 9.8 = 6.53 (pC/(m/s Sensor sensitivity setting value: 6.53
2
)) [Charge], 0.1 to 10 (mV/(m/s2) [Preamp]
2
2
).
2
))
To convert [m/s2] units to [G] for display
This instr u men t m e as ur es ch ar ge va lu e i n un it s o f m/ s be used to convert to charge value in units of G.
See "5.4 Converting Input Values (Scaling Function)" in the Instruction Ma nua l
Set up scaling as follows.
• Conversion ratio setting [Ratio]
Conversion ratio: 1/9.8 = 0.1020E+00 Offset: 0.0000E+00 Units: G
• Two-point setting [2-Point]
Entry 1: 9.8000E+00 Physical val ue 1: 1.0000E+00 Entry 2: 0.0000E+00 Physical val ue 2: 0.0000E+00 Units: G
2
. The scaling function can
To Use a Sensor Outside of the Setting Range
The scaling function can be employed to use a sensor outside of the setting range.
1 On the Channel S ettings screen, select the chann el on whi ch to use th e sensor
outside of the setting range.
2 Move the cursor to [Sensitivity] to set the sensor sensitivity.
Enter a multiplier val ue that wil l produc e a settable value ( 0.1 to 10) plied by the actual sensitivity of the sensor.
3 Move the cursor to [Scaling], and set it to [On]. 4 Set the conversion ratio to the sensitivity multiplier value.
77
3.11 About Setting Contents
when multi-
Setting Example 1: If the specified sensor sensitivity is 23.4 pC/(m/s Set 1/2.34 as the sensor sensiti vity multiplier, and 10 (pc/m/s
sensitivity. Scaling is set as follows to display measured values multiplied by 1/2.34.
• Conversion ratio setting [Ratio]
Conversion ratio: 10/23.4 = 0.4274E+00 Offset: 0.0000E+00 Units: m/s
• Two-point setting [2-Point]
Input 1: 2.3400E+01 Scale 1: 1.0000E+01 Input 2: 0.0000E+00 Scale 2: 0.0000E+00 Units: m/s
Setting Example 2: If the specified sensor sensitivity is 0.05 pC/(m/s
Set the sensor sensitiv ity multiplier to 2, an d the sensor sens itivity to 0.1 (pc/m/
2
s
)).
Scaling is set as follows to display measured values multiplied by 2.
2
2
2
):
2
)
2
)) as the sensor
• Conversion ratio setting [Ratio]
Conversion ratio: 0.1/0.05 = 2.0000E+00 Offset: 0.0000E+00 Units: m/s
• Two-point setting [2-Point]
Input 1: 0.0500E+00 Scale 1: 0.1000E+00 Input 2: 0.0000E+00 Scale 2: 0.0000E+00 Units: m/s
2
2
78
3.11 About Setting Contents

3.11.15 Probe Attenuation Selection

Probe attenuation can be selec ted when measuring using a connection cable, differential probe or attenuating probe. (Only for voltage measurement)
By matching the i nput channel probe setting to the a ttenuation ratio of a probe connected to an analog input module, the voltage axis range is automatically converted for direct reading of numerical values. Each channel should be set to match the attenuation ratio of its input probe.
Setting Item: [Probe]
Selections Description
1:1
10:1
100:1
1000:1
Select this setting when measuring with the Model L9197. 9197, L9198 or L9217 cable connected to the input module.
Select this setting when measuring with the Mo del 9665 10:1 Probe connected to the input module.
Select this setting when measuring with the Model 9666 100:1 Probe connected to the input module.
Select this setting when using the 9322 Differential Probe.
When using the Model 8940 F/V Unit with the 9322 Differential Probe:
9322+9325
9322+9418
When supplying power from the Model 9687 Probe Power Unit, either of the above settings can be selected. The factory default and system reset default setting is 1:1.
Some probes cannot be used with certain input modules.
See "1.3 List of Input Modules, Cables, Probes and Clamp Combinations" (p. 13)
Select this setting when using the Model 9325 Power Cord to con­nect the 9322 Differential Probe.
Select this setting when using the Model 9418-15 AC Adapter to connect the 9322 Differential Probe.

3.11.16 Bridge Voltage Setting

Set to suit the sensor to be used. Setting Item: [Bridge]
Selections Description
2V Se ts the brid ge voltage to 2 V (Bridge resistance = 120 Ω to 1 kΩ) 5V Se ts the brid ge voltage to 5 V (Bridge resistance = 350 Ω to 1 kΩ) 10V Sets the bridge voltage to 10 V (Bridge resistance = 350 Ω to 1 kΩ)
This setting should not exceed the allowable applied voltage for the strain gauge type converter to be used. Other wise, converter specifica tions will not be satis­fied, and correct measurements will not be possible.

3.11.17 Executing Zero Adjustment

The input module can apply an internal offset to set the reference potential of the instrument to zero volts.
Before executing zero adjustment
• Turn power on and wait 30 minutes to allow the internal temperature of the input module to stabilize. (Some input modules may require one hour warm-up.)
• Zero adjustment cannot be performed while measuring.
• Key operations are not accepted while zero adjustment is executing.
To execute zero adjustment
Move the cursor to the [Zero-Adjust] button, and select F1 [Execute]. Zero adjustment can be performed from the CH SET dialog on the Wave form screen.
Zero adjustment is not applicable to the following modules:
• The [Temp] mode of the Model 8937 Voltage/Temp Unit
• Model 8939 Strain Unit
• Modes other than [Voltage] and [Current] of the Model 8940 F/V Unit
• Model 8958 16-Ch Scanner Unit
79
3.11 About Setting Contents
In the following cases, zero adjustment should be executed again.
• After an input module has been removed or inserted
• After power has been turned off and on
• After comp lete reinitialization
• When ambient temperature has changed significantly The zero position may have drifted.
• When measurement mode is changed to the following mode:
8959 DC/RMS Unit [DC] [RMS]8961 High Voltage Unit [DC] [RMS]8940 F/V Unit [Voltage] [Current]8947 Charge Unit [Voltage] [Charge]
When using the Model 8958 16-Ch Scanner Unit, zero position adjustment of the module from the System screen may be needed.
See "Scanner Unit Zero Position Adjustment" (p. 29)

3.11.18 Executing Offset Cancellation

The input value ca n be forc ed to dis pl ay as ze ro vo lts . Input signal bias, such a s sensor emf, ca n be adjusted to display as zero volt s. This function applies to voltage and current measurement.
See " What is the difference between Offset Cancellation and Zero Adjustment?" (p. 80)
Before executing offset cancellation
Connect the probe or clam p to the measur ement object and app ly the input s ig­nal that is supposed to measure zero volts.
See "About input voltage during offset cancellation" (p. 80)
When also using scaling, execute offset cancellation before setting the scaling.
To execute offset cancellation
Move the cursor to the [Offset Cancel] button, and select F1 [Execute]. An asterisk "" appears on the [Offset Cancel] button when offset canc ellat ion i s active. To revert to the offset value prior to executing offset cancellation, select F2
[Reset].
80
When input voltage is constant When input voltage is fluctuating
Offset voltage +V
1
is set to display as
zero regardless of execution time.
The offset voltage set to display as zero depends on the timing of offset cancellation execution.
Offset cancellation referenc e potential is +V.
Waveform after executing offset cancellation
Input Signal Waveform
3.11 About Setting Contents
After executing offset cancellation, the offset value will be calculated aut omati­cally if the rang e is changed. When a precise offset value is required, execu te the offset cancellation again.
Offset cancellation cannot be executed in the following cases:
• When measuring other than voltage or current
• When the input voltage is more than ±10 divisions from zero
About input voltage during offset cancellation
The maximum input voltage and maximum rated voltage to ground are unaffected by offset cancellation of an input voltage to zero volts. There­fore, careful attention is necessary during measurement.
A constant input voltage is required to execute offset cancellation. If the input voltage is fluctuating, the offset to zero volts depends on the timing of offset cancellation execution.
Offset cancellation
Offset cancellation is not r eset even when power is turned o ff. It is reset when the input module configuration is changed (after an input module has been inserted or removed).
If “Warning: Offset Cancel failed.” appears
This appears when executing offset cancellation if the input voltage is more than ±10 divisions from zero volts.
To avoid the warning, set the input signal within ±10 divisions of zero volts.
What is the difference between Offset Cancellation and Zero Adjust­ment?
Offset cancellation a djusts for input signal bia s if there is a potential d ifference between the referenc e potential of the measurement obj ect and the reference potential (0 V) of this instr ument , so that th e referen ce pote ntial o f the meas ure­ment object appears as zero volts.
Zero adjustment sets the input m odule’s intern al bias so the re ference potent ial of the instrument is zero volts.

3.11.19 Executing Auto-Balance

Auto-balance sets the reference output level of a transducer to the specified zero position. It is applicable only to a strain module.
Before executing auto-balance
• Turn power on and wait 1 hour to allow the internal temper ature of the input module to stabilize.
• With the sensor conn ected to the input module, execute auto- balance under stable input conditions.
• Auto-balance cannot execute during measurement.
• Key operations are not accepted while auto-balance is executing.
To execute auto-balance
Move the cursor to the [Auto-Balance] button, and selec t F1 [Execute]. Auto-balance can be performed from the CH SET dialog on the Waveform screen (when the unit and channel numbers of the strain module is selected).
81
3.11 About Setting Contents
In the following cases, auto-balance should be executed again.
• After changing the measurement range
• After an input module has been removed or inserted
• After the strain gauge transducer has been replaced
• After power has been turned off and on
• After performing a system reset
• When ambient temper ature has changed si gnificantly (the ze ro position may drift)
If “Warning: Auto balance failed.” appears
The channel on which auto-balance failed is displayed. Verify the following, and execute again:
• Is the sensor in a disc harged sta te? (Make sure tha t it is not bein g subje ct to vibration, etc.)
• Is the sensor connected correctly?
82
3.11 About Setting Contents

4.1 Operation Overview

Logic Input Settings Chapter 4
After turning power on, set the measurement criteria before measuring. This chapter describes the settings related to logic inputs. Refer to "7.3 Displaying L ogic Waveforms" in the In struction Manual for setti ng details. Measurement status and trigger settings are the same as for analog inputs. Refer to the related chapters in the Instruction Manual.
4.1 Operation Overview
83
(Select the Function)
Select the measurement status and
channels
Set up input channels
Set trigger criteria
Waveform display settings
Start Measurement
To record simultaneously with an analog signal, select a function that suits the analog input. When recording only logic signals, the function can be set freely.
(For Memory Function)
[Use Ch] page on Status Settings screen
Enable (set to On) the logic channels to be used.
See "4.2.1 Selecting Channels to Use" in the Instruction Manual
[Logic] page on Channel Settings screen
Select whether to enable (On) or disable (Off) the input waveform for each logic channel prob e.
(Default setting: All Off) Select waveform colors as desired.
See "7.3 Displaying Logic Waveforms" in the Instruction Manual
[Logic] page on Trigger Settings screen
To apply a trigger event for measurements, set the logic trigger for each channel.
See "6.8 Triggering by Logic Signals (Logic Trigger)" in "Chapter 6 Trigger
Settings" in the Instruction Manual
[Logic] page on Sheet Settings screen
Select the channels to appear on the wave form disp la y. (Default setting: All On) Set the waveform display position and height for each channel as needed.
See "7.2 Setting the Screen Layout of the Waveform Screen (Sheet Set-
tings Screen)" and "7.3 Dis pla ying Logic Wav efo rms" in the Instruction Manual
84
Display Settings Screen
Press the SET key.
The Settings screen appears. Function setting is not needed.
Selecting channels to use (when displayed with the memory function)
Status Settings Screen
Logic Channel A to D
1 Press the SUB MENU keys to select the [Status]
menu item.
2 Press the SHEET/PAGE keys to select the [Use Ch]
page.
To record simultaneously with analog input:
See "4.2.1 Selecting Channels to Use" in the Instruction
Manual
3 Use the CURSOR keys to move the cursor to the Set-
ting column of the logic channel, and press the F2
[On] key.
1
2
3
LOGIC Receptacles
CH D CH C CH B CH A
Logic Channels
(Default setting: depends on the settings of the channels to be used)

4.2 Setting Procedure

4.2 Setting Procedure
85
3
4
Select the time axis range and recording length
Status Setting Screen (with memory function)
1 Press the SUB MENU keys to select the [Status]
menu item.
2 (When using the memory function)
Press the
SHEET/PAGE keys to select the [Basic]
page.
3 (To record only logic inputs)
Use the CURSOR keys to move the cursor to [Time
base], and press an F key to select the time per divi-
sion for the horizontal axis. (When the Timebase 2 is enabled [On], and only logic
inputs are assigned)
Set the sampling rate.
4 Use the CURSOR keys to move the cursor to [Shot]
(Recording Length), an d press an F ke y to select the number of divisions for the recording time.
About measurement configuration (timebase, recording length, etc.)
See "Chapter 4 Measurement Configuration Settings" in
the Instruction Manual
1
2
3
Select the input channels
Channel Settings Screen
Probes
Logic Channels
1 Press the SUB MENU keys to select the [Channel]
menu item.
2 Press the SHEET/PAGE keys to sele ct the [Logic]
page.
The [Logic] page of the Channel Setting screen appears.
3 Use the CURSOR keys to move the cu rsor to the
Setting column for each probe (1 to 4) of the logic channel, and press the F2 [On] key.
(On = display waveform, Off = do not display waveform)
To change a wavefo rm’s color, move the curso r to the waveform display color setting column and select the color with the F1 or F2 key.
See "7.3.1 Setting the Waveform Display" in the
Instruction Manual
If you want to e nte r c omme nts f or ind iv idu al channels: Select the [Comment] page.
See "5.2 Adding Comments" in the
Instruction Manual
1
2
(Default setting: [Off])
Waveform Display On/Off
Waveform Display Color
4.2 Setting Procedure
86
2
Set the trigger conditions (if applying a trigger)
Trigger Settings Screen
1 Press the SUB MENU keys to select the [Trigger]
menu item.
2 Press the SHEET/PAGE keys to select the [Logic]
page.
The [Logic] p age o f the Trigger Settings scr een appears.
3 Use the CURSOR keys to move to each channel
to be triggered, and set each as needed.
See "6.8 Trigg erin g by Logic Sign als ( Logic Trigger)" in
the Instruction Manual
4 Set triggering as occasion demands.
See "Chapter 6 Trigger Settings" in the Instruction
Manual
1
Logic Channels
Set the trigger pattern for each probe
Select AND/OR trigger conditions.
Set the trigger filter width. Prevents noise from causing false triggers.
(Default setting: Logic trigger [Off])
3
Set the trigger detection method.
Set the waveform display method (as occasion demands)
Sheet Settings Screen
1 Press the SUB MENU keys to select the [Sheet]
menu.
2 Press the SHEET/PAGE keys to select the [Logic]
page.
The [Logic] page of the Sheet Settings screen appears.
3 Use the CURSOR keys to move to a logic channel
to be displayed , an d pr es s the F keys to select t he setting contents for each item.
See "7.3.2 Setting the Display Position" and "7.3.3 Set-
ting the Display Height" in the Instruction Ma nua l
Setting other display sheets
See "Chapter 7 Wa ve form Displ ay Setti ng s" in the
Instruction Manual
1
2
3
(Default setting: All logic channels displayed [On])
Set storage and printing as occasion demands, then start measuring.
Logic Channels
Set display of each channel on or off.
Waveform display positions can be modified.
Display height can be modified.
4.2 Setting Procedure

5.1 Logic Input Section

Digital input Threshold value
Contact input Detecting resistance value
1.4 V range 1.4 V±0.3 V more than 1.5 kΩ less than 500Ω
opened (Output L) shorted (Output H)
2.5 V range 2.5 V±0.4 V more than 3.5 kΩ less than 1.5 kΩ
opened (Output L) shorted (Output H)
4.0 V range 4.0 V±0.5 V more than 25 kΩ less than 8 kΩ
opened (Output L) shorted (Output H)
Specifications Chapter 5
Refer to "Safety Informati on" for the “Measurement Category” in the Quick Start Manual.
5.1 Logic Input Section

5.1.1 Model 9327 Logic Probe

87
Temperature and humidity range for guaranteed accu­racy Period of guaranteed accu­racy
The number of input chan­nels
Input type Digital input / Contact input
Input resistance 1 MΩ±5% (Digital input: 0 to +5 V)
Pull-up resistance 2 kΩ (Contact inputs: internally connected through pull-up resistance to +5 V) Detecting level
Response pulse width 100 ns or more Maximum input voltage 0 to +50 VDC Operating temperature and
humidity ranges Storage temperature and
humidity ranges Operating environment Indoors, Pollution degree 2, Altitude up to 2,000 m (6,562-ft.) Dimensions Approx. 62W x 94H x 20D mm (2.44"W x 3.70"H x 0.79"D) (sans protrusions) Connector cable length Approx. 1500 mm (59") Probe tip cable length Approx. 300 mm (11.8") Mass Approx. 150 g (5.3 oz.) (including connector cable, excluding input leads) Accessories IC clip leads, Alligator clip leads, Carrying case, Instruction Manual
23 ±5°C(73±9°F), 35 to 80% RH
1 year
4 channels (Common ground between chassis and between channels)
Input type can be selected for each channels. Open collector outputs can be directly measured using contact inputs.
Ω or more (Digital input: +5 to +50 V)
500 k
0 to 40°C (32 to 104°F), 80%RH or less (non-condensating)
-10 to 50°C (14 to 122°F), 90%RH or less (non-condensating)
88
5.1 Logic Input Section

5.1.2 Model MR9321-01 Logic Probe

Temperature and humidity range for guaranteed accu­racy
Period of guaranteed accu­racy
The number of input chan­nels
Input voltage range LOW HIGH Input resistance More than 30 kΩ More than 100 kΩ Detecting level
(Output: L) (Output: H)
Response time
(Rising) (Falling)
Maximum input voltage 150 V rms 250 V rms Maximum rated voltage to
earth Dielectric strength 2.224 kVAC for 15 s (between each input channel and chass is, and between inp ut chan-
Insulation resistance At least 100 MΩ at 500 V DC (between each input channel and chassis, and between
Operating temperature and humidity ranges
Storage temperature and humidity ranges
Operating environment Indoors, Pollution degree 2, Altitude up to 2,000 m (6,562-ft.) Dimensions Approx. 62W x 128H x 20D mm (2.44"W x 5.04"H x 0.79"D) (sans protrusions) Connector cable length Approx. 1500 mm (59") Probe tip cable length Approx. 1200 mm (47") Mass Approx. 320 g (11.3 oz.) Accessories Carrying case, Instruction Manual
23 ±5°C(73±9°F), 35 to 80% RH
1 year 4 channels (Isolated from chassis and between channels)
0 to 10 VAC ±(0 to 15) VDC 60 to 150 VAC ±(20 to 150) VDC
Less than 1 ms Less than 3 ms at 100 V DC
250 V rms
nels)
input channels) 0 to 40°C (32 to 104°F), 80%RH or less (non-condensating)
-10 to 50°C (14 to 122°F), 90%RH or less (non-condensating)
0 to 30 VAC ±(0 to 43) VDC 170 to 250 VAC ±(70 to 250) VDC
Less than 1 ms Less than 3 ms at 200 V DC
• Absolute values are detected, so either polarity DC may be input. AC voltage is evaluated as a 50/60-Hz standard sine wave.
• On/off judgment of the input signal is performed within the above detection range. On/off judgment of input signals cannot be performed correctly with input voltage between L and H detection levels.

5.1.3 Model 9320-01 Logic Probe

Digital input Threshold value
Contact input Detecting resistance val ue
1.4 V range 1.4 V±0.3 V more than 1.5 kΩ less than 500Ω
opened (Output L) shorted (Output H)
2.5 V range 2.5 V±0.4 V more than 3.5 kΩ less than 1.5 kΩ
opened (Output L) shorted (Output H)
4.0 V range 4.0 V±0.5 V more than 25 kΩ less than 8 kΩ
opened (Output L) shorted (Output H)
89
5.1 Logic Input Section
Temperature and humidity range for guaranteed accu­racy Period of guaranteed accu­racy
The number of input chan­nels
Input type Digital input / Contact input
Input resistance Digital input: 1 MΩ±5% (0 to +5 V), 500 kΩ or more (+5 to +50 V)
Detecting level
Response pulse width 500 ns or more Maximum input voltage 0 to +50 VDC Operating temperature and
humidity ranges Storage temperature and
humidity ranges Operating environment Indoors, Pollution degree 2, Altitude up to 2,000 m (6,562-ft.) Dimensions Approx. 62W x 94H x 20D mm (2.44"W x 3.70"H x 0.79"D) (sans protrusions) Connector cable length Approx. 1500 mm (59") Probe tip cable length Approx. 300 mm (11.8") Mass Approx. 150 g (5.3 oz.) (including connector cable, excluding input leads) Accessories IC clip leads, Alligator clip leads, Carrying case, Instruction Manual
23 ±5°C(73±9°F), 35 to 80% RH
1 year
4 channels (Common ground between chassis and between channels)
Input type can be selected for each channel. Open collector outputs can be directly measured using contact inputs.
Contact inputs: Approx. 2 k tance)
0 to 40°C (32 to 104°F), 80%RH or less (non-condensating)
-10 to 50°C (14 to 122°F), 90%RH or less (non-condensating)
Ω (input terminals are c onn ec ted to +5 V th roug h 2 -kΩ resis-
90
Measurement Error
Es 1
Rin
Rs Rin+
-----------------------
[V]
Es: signal voltage Rs: signal source impedance Rin: input impedance
=

5.2 Analog Input Section

5.2 Analog Input Section
Input module specifications are described here.
Measurement error caused by sig­nal source impedance
• Errors occur when th e signal source impedance is greater than the i nput impedance of the module.
• The input impedance of the Model 8936 Analog Unit is 1 MΩ. Errors of about 0.1% occur when the signal source impedance is 1 kΩ.

5.2.1 Model 8936 Analog Unit

Temperature and humidity range for guaranteed accuracy
Period of guaranteed accuracy
Measurement ranges 5 m, 10 m, 20 m, 50 m, 100 m, 200 m, 500 m, 1, 2, 5, 10, 20 V/div DC amplitude accuracy ±0.4%f.s. Zero position accuracy ±0.1%f.s. (after zero adjustment) Temperature charact eristic Gain: ±0.025%f.s./°C
Frequency characteristic DC coupling: DC to 400 kHz ±3 dB
Noise 450 μV p-p typ., 750 μV p-p max.
Common mode rejection ratio
Low-pass filter OFF, 5±50%, 500±50%, 5k±50%, 100k±50% (Hz) -3 dB Input type Unbalanced (input isolated from output) Input coupling DC, GND, AC Input resistance 1 MΩ±1% Input capacitance 30 pF±10 pF (at 100 kHz) A/D resolution 12 bits Voltage axis resolution 80 points/div (with 1X magnification) Maximum sampling rate 1 MS/s (sampling period: 1 μs) Input terminals Insulated BNC terminal Maximum input voltage 400 VDC Maximum rated voltage to
earth Operating temperature and
humidity ranges Operating environment Same as the host Memory HiCorder Storage temperature and
humidity ranges Effect of radiated radio-
frequency electromagnetic field
23 ±5°C(73±9°F), 35 to 80% RH (when zero adjustment is executed 30 minutes after power on)
1 year
Zero position: ±0.02%f.s./°C (after zero adjustment)
AC coupling: 7 Hz to 400 kHz ±3 dB (low cut-off frequency: 7 Hz±20%)
(sensitivity range, with input shorted) (S/N 2004-041018235 and later) 80 dB minimum (at 50/60 Hz and with signal source resistance 100 Ω maximum)
370 VAC/DC (between each input channel and chassis, and between input channels), Measurement category II (anticipated transient overvoltage: 4000 V)
Same as the host Memory HiCorder
-10 to 50°C (14 to 122°F), 80% RH or less (non-condensating)
±2%f.s. at 3 V/m max.
91
5.2 Analog Input Section
Effect of conducted radio­frequency electromagnetic field
Dielectric strength 3.7 kVAC for 1 minute (between input m odu le an d ch as si s, an d between input modules) Dimensions Approx. 170W x 20H x 148.5D mm (6.69”W x 0.79”H x 5.85”D) (sans protrusions) Mass Approx. 290 g (10.2 oz.) Applicable Standards
Safety EMC
±40%f.s. at 3 V max. (100 mV/div with 1 V DC input)
EN 61010 EN 61326 Class A
92
+1
3+1−
3
5.2 Analog Input Section

5.2.2 Model 8937 Voltage/Temp Unit

General specifications
Temperature and humidity range for guaranteed accu­racy
Period of guaranteed accu­racy
Common mode rejection ra­tio
Input type Unbalanced (input isolated from output) A/D resolution 12 bits Maximum sampling rate 1MS/s (However, update rate differs with temperature input.) Maximum input voltage 30 V rms or 60 VDC (voltage and temperature inputs) Maximum rated voltage to
earth Operating temperature and
humidity ranges Operating environment Same as the host Memory HiCorder Storage temperature and
humidity ranges Effect of radiated radio-
frequency electromagnetic field
Effect of conducted radio­frequency electromagnetic field
Dimensions Approx. 170W x 20H x 148.5D mm (6.69”W x 0.79”H x 5.85”D) (sans protrusions) Mass Approx. 300 g (10.6 oz.) Applicable Standards
Safety EMC
23 ±5°C(73±9°F), 35 to 80% RH (when zero adjustment is executed 1 hour after power on)
1 year 80 dB min. (at 50/60 Hz and with signal source resistance 100 Ω max.)
30 V rms or 60 VDC (voltage and temperature inputs) Anticipated transient overvoltage: 330 V
Same as the host Memory HiCorder
-10 to 50°C (14 to 122°F), 80% RH or less (non-condensating)
±2% f.s. at 3 V/m (in 5-mV/div range)
±5% f.s. at 3 V (100 mV/div with 1 V DC input)
EN 61010 EN 61326 Class A
Voltage input
Measurement ranges 500 μ, 1 m, 2 m, 5 m, 10 m, 20 m, 50 m, 100 m, 200 m, 500 m, 1, 2 V/div DC amplitude accuracy ±0.4%f.s. Zero position accuracy ±0.15%f.s. (after zero adjustment) Temperature charact eristic Gain: ±0.02%f.s./°C, Zero position: ±0.03%f.s./°C Frequency characteristic DC to 400 kHz dB (in 500 μV to 2 V/div range)
With digital filter ON: DC to 3 kHz dB (Data update rate: 100 μs±20% (in 500 μV to 2 mV/div range))
Noise 75 μVp-p typ.,120 μVp-p max. with digital filter OFF (in 500 μV/div range)
20 μVp-p typ., 30 μVp-p max. with digital filter ON (in 500 μV/div range) (S/N 2004-041135258 and later)
Input terminals BNC terminal Input resistance 1 MΩ±1% Input capacitance 50 pF±20 pF (at 100 kHz) Input coupling DC, GND, AC Low-pass filter OFF, 5±50%, 500±50%, 5k±50%, 100k±50% (Hz) -3 dB
Loading...