Hioki 8841, 8842 Instruction Manual

Page 1
INSTRUCTION MANUAL
8841 8842
MEMORY HiCORDER
Page 2
Page 3
Contents
Introduction i Inspection i Safety Notes iii Notes on Use v Chapter Summary ix
1.1 Major Features 1
1.2 Identification of Controls and Indicators
2.1 Installation of the Unit 7
2.2 Power Supply and Ground Connections
2.2.1 Connecting the AC Power Supply 9
2.2.2 Connecting the 9433 DC POWER ADAPTER 10
2.2.3 Functional Grounding of the 8841/8842
2.3 Power On/Off 13
2.4 Connection of the Input Unit
2.4.1 8936/ 8938/ 8946 ANALOG UNITs 14
2.4.2 8937 VOLTAGE/TEMP UNIT 15
2.4.3 8939 STRAIN UNIT
2.4.4 8940 F/V UNIT
2.4.5 8947 CHARGE UNIT 19
2.5 Logic Probe Connection 20
2.6 9018-10/9132-10 CLAMP ON PROBE
12
14
16 17
21
3
9
2.7 9322 DIFFERENTIAL PROBE Connection
2.8 Loading Recoding Paper
2.9 Care of Recording Paper
2.10 Notes on Measurement
2.10.1 Maximum Input Voltage 28
2.10.2 Using a Voltage Transformer
22 23 25 26
29
Chapter 3 Basic Key Operation 31
3.1 Basic Key Operation 31
3.1.1 Basic Display Operation (MENU) 31
3.1.2 Printer Key Operation 31
3.1.3 Setting the Items 32
3.1.4 Jog/Shuttle Controls and Select Key 32
3.1.5 Basic Input Operation 33
3.1.6 Measurement Start and Stop Operation 33
Page 4
3.2 Other Keys Operation 34
3.3 Screen Configuration
35
Chapter 4 Memory Recorder Function 37
4.1 Overview of the Memory Recorder Function 37
4.2 Operation Sequence (MEM)
4.3 STATUS Settings (MEM) 41
4.3.1 Setting the Function Mode 42
4.3.2 Setting the Time Axis Range 42
4.3.3 Setting the Recording Length 43
4.3.4 Setting the Display Layout 44
4.3.5 Setting the Interpolation (dot-line) 46
4.3.6 Setting the Print Mode 46
4.3.7 Setting the Roll Mode 47
4.3.8 Setting the Auto Print Function 47
4.3.9 Setting the Auto Save Function 48
4.3.10 Setting the Overlay Function 50
4.3.11 Setting the Averaging Function 51
4.3.12 Setting the Waveform Evaluation 53
39
4.4 Using the X-Y Waveform Plots (MEM) 54
4.5 Settings on the Waveform Display Screen (MEM)
4.6 Auto Range Function
4.7 Other Screen Settings
4.8 Start and Stop Operation (MEM)
57 58 60 61
Chapter 5 Recorder Function 63
5.1 Overview of the Recorder Function 63
5.2 Operation Sequence (REC)
5.3 STATUS Settings (REC)
5.3.1 Setting the Function Mode 68
5.3.2 Setting the Time Axis Range 68
5.3.3 Setting the Sampling Period 69
5.3.4 Setting the Recording Length 69
5.3.5 Setting the Display Layout
5.3.6 Setting the Print Mode
5.3.7 Setting the Additional Recording Function 74
5.3.8 Setting the Printer Function (Real Time Printing) 75
5.3.9 Setting the Auto Save Function 76
5.3.10 Setting the Interpolation (dot-line, X-Y only) 77
5.3.11 Setting the Display Clear Function (X-Y only) 77
65 67
70 73
5.4 Using the X-Y CONT Recorder 78
5.4.1 Setting the Status Screen 79
Page 5
5.4.2 Setting the Channel Screen 79
5.4.3 Setting the Display Screen
80
5.5 Settings on the Waveform Display Screen (REC) 81
5.6 Start and Stop Operation (REC)
82
Chapter 6 RMS Recorder Function 83
6.1 Overview of the RMS Recorder Function 83
6.2 Operation Sequence (RMS) 84
6.3 STATUS Settings (RMS)
6.3.1 Setting the Function Mode 87
6.3.2 Setting the Time Axis Range 87
6.3.3 Setting the Frequency
6.3.4 Setting the Recording Length
6.3.5 Setting the Display Layout 89
6.3.6 Setting the Print Mode 91
6.3.7 Setting the Additional Recording Function 92
6.3.8 Setting the Printer Function (Real Time Printing)
6.3.9 Setting the Auto Save Function
6.4 Settings on the Waveform Display Screen (RMS) 95
86
88 88
93 93
6.5 Start and Stop Operation (RMS)
96
Chapter 7 Recorder & Memory Function 97
7.1 Overview of the Recorder & Memory Function 97
7.2 Operation Sequence (REC&MEM)
7.3 STATUS Settings (REC&MEM)
7.3.1 Setting the Function Mode 101
7.3.2 Setting the Time Axis Range 102
7.3.3 Setting the Recording Length
7.3.4 Display Function
7.3.5 Setting the Display Layout 105
7.3.6 Setting the Print Mode 106
7.3.7 Setting the Additional Recording Function 107
7.3.8 Setting the Printer Function (Real-time Printing)
7.3.9 Setting the Auto Save Function
7.3.10 Setting the Real-Time Save 111
7.4 Settings on the Waveform Display Screen (REC& MEM)
98
100
103 104
108 109
117
7.5 Start and Stop Operation (REC&MEM)
118
Page 6
Chapter 8 FFT Function 119
8.1 Overview of the FFT Function 119
8.2 Operation Sequence (FFT)
8.3 STATUS1 Settings (FFT)
8.3.1 Setting the Function Mode 123
8.3.2 Setting the FFT Channel Mode 123
8.3.3 Setting the Frequency Range 124
8.3.4 FFT Number of Points Setup 125
8.3.5 Setting the Window Function 125
8.3.6 Setting the Display Format 126
8.3.7 Selecting Reference Data 128
8.3.8 Setting the FFT Analysis Mode 129
8.3.9 Setting the X-axis and Y-axis Displays 130
8.3.10 Setting the Analysis Channel 132
8.3.11 Setting the Display Scale 133
8.3.12 Octave Filter Setting 134
120 122
8.4 STATUS2 Settings (FFT) 135
8.4.1 Setting the Averaging Function 136
8.4.2 Setting the Interpolation (dot-line) 139
8.4.3 Setting the Print Mode 140
8.4.4 Setting the Auto Print Function 140
8.4.5 Setting the Auto Save Function
8.4.6 Setting the Waveform Evaluation
141 142
8.5 Settings on the Waveform Display Screen (FFT) 143
8.6 Start and Stop Operation (FFT)
8.7 FFT Analysis Function
8.7.1 Storage Waveform [STR] 145
8.7.2 Linear Spectrum [LIN] 146
8.7.3 RMS Spectrum [RMS] 148
8.7.4 Power Spectrum [PSP] 150
8.7.5 Auto Correlation [ACR] 152
8.7.6 Histogram [HIS]
8.7.7 Transfer Function [TRF]
8.7.8 Cross Power Spectrum [CSP] 156
8.7.9 Cross Correlation [CCR] 158
8.7.10 Unit Impulse Response [IMP] 159
8.7.11 Coherence [COH] 160
8.7.12 Octave Analysis [OCT] 161
144 145
153 154
Page 7
Chapter 9 Input Channel Settings 165
9.1 Overview 165
9.2 Operation Procedure (Input Channel Setting)
9.3 Setting the CHANNEL Screen
9.3.1 Setting the Waveform Display Color 169
9.3.2 Setting the Waveform Display Graph Position 169
9.3.3 Setting the Voltage Axis Range 170
9.3.4 Setting the Input Coupling 170
9.3.5 Setting the Magnification/Compression Ratio Along the Voltage Axis 171
9.3.6 Setting the Zero Position
9.3.7 Setting the Low-pass Filter
9.3.8 Setting the Logic Inputs 174
166 168
172 173
9.4 Copying Channel Settings 175
9.5 Zero Settings
9.5.1 Zero Adjustment 176
9.5.2 Zero Offset Setup
176
177
9.6 Channel Guide Function 179
9.7 Setting the Variable Function
180
9.8 Scaling Function (SYSTEM2)
9.8.1 Setting the Scaling Function 183
9.8.2 Scaling Setting Example 186
182
9.9 Comment Function (SYSTEM 3) 187
9.9.1 Title Comment Entry 187
9.9.2 Analog/Logic Channel Comment Entry 188
9.9.3 Character Entry Procedure
189
9.10 Setting the Waveform Display Screen 191
9.10.1 Entering by CH.SET Key 191
9.10.2 Setting the Vernier Function 192
9.10.3 Direct Channel Setting 193
9.11 Setting the 8937 VOLTAGE/TEMP UNIT 194
9.11.1 Making the Settings of Voltage Measurement 194
9.11.2 Making the Settings of Temperature
Measurement 196
9.12 8938 FFT ANALOG UNIT 199
9.13 Setting the 8939 STRAIN UNIT
200
9.14 8940 F/V UNIT
9.14.1 Frequency Measurement 202
9.14.2 Setting the Count Mode 205
9.14.3 Setting the Pulse Duty Ratio Mode 207
9.14.4 Setting the Voltage Mode
201
209
Page 8
9.14.5 Setting the Current Measurement 211
9.15 Setting the 8947 CHARGE UNIT 212
9.15.1 Acceleration Measurement (Charge/Preamp) Selection 213
9.15.2 Setting the Sensor Sensitivity
9.15.3 Setting the Voltage Measurement
215 216
Chapter 10 Trigger Functions 217
10.1 Overview of the Trigger Functions 217
10.2 Operation Sequence (Trigger mode setting)
10.3 TRIGGER Screen Organization
10.4 Trigger Mode
10.5 Pre-trigger
10.6 Trigger Timing (REC only)
10.7 Trigger Source AND/OR Linking
10.8 Analog Trigger
10.8.1 Level Trigger 226
10.8.2 Window Trigger 229
10.8.3 Voltage Drop Trigger 231
10.8.4 Period Trigger 233
10.8.5 RMS Level Trigger 236
218 220 221 222 224 225 226
10.9 Logic Trigger 238
10.10 External Trigger Function
10.11 Timer Trigger Function
10.12 Manual Trigger
10.13 Trigger Output Terminal
240 241 244 244
Chapter 11Waveform Display Screen Operation 245
11.1 Scrolling the Waveform 246
11.2 Using the AB Cursors
11.2.1 Using the Line Cursors (Vertical) 248
11.2.2 Using the Line Cursors (Horizontal) 249
11.2.3 Using the Trace Cursors 250
11.2.4 Using the A・B Cursors (X-Y Screen)
11.3 Magnification/compression Ratio Along the Time Axis 253
11.4 Zoom Function (MEM only)
11.5 Input Level Monitor Function
11.6 View Function (VIEW key)
11.6.1 Position Display 256
11.6.2 Block Display (MEM only)
247
252
254 255 256
257
Page 9
Chapter 12 SYSTEM Screen Settings 259
12.1 Overview 259
12.2 SET UP Screen (SYSTEM 1)
12.2.1 Using Channel (MEM only) 260
12.2.2 Start Backup
12.2.3 Backlight Saver
12.2.4 Display Color
12.2.5 Beep Sound
12.2.6 Display Language 263
12.2.7 Grid Type 263
12.2.8 Channel Marker 263
12.2.9 Time Value
12.2.10 List & Gauge
12.2.11 Printer Density 264
12.2.12 Upper-lower Values Print 264
12.2.13 0 Position Comment 265
12.2.14 Counter Print
12.2.15 Numerical Input
12.2.16 Variable Auto Calibration 266
260
261 261 262 262
264 264
265 265
12.3 SCALING Screen (SYSTEM 2) 266
12.4 COMMENT Screen (SYSTEM 3)
12.5 INTERFACE Screen (SYSTEM 4)
12.5.1 Output Destination by the COPY Key 269
12.5.2 Output Destination by the PRINT Key 269
12.5.3 Interface Settings 269
12.5.4 Setting the SCSI ID 270
12.5.5 Setting the Internal MO (when the 9607 is installed) 270
12.5.6 Setting the LAN 271
267 268
12.6 INITIALIZE Screen 278
12.6.1 Setting the Clock 278
12.6.2 Clearing the Waveform Data 278
12.6.3 System Reset 279
12.6.4 List System Configuration 279
12.7 Self-check 280
12.7.1 ROM/RAM Check 280
12.7.2 Printer Check
12.7.3 Display Check 281
12.7.4 Key Check 281
12.7.5 PC Card Check 281
280
Page 10
Chapter 13 Printout of Measurement Data 283
13.1 Printout of Measurement Data Operating Procedure
284
13.2 Setting the STATUS Screen (printout)
13.2.1 Setting the Display Format 286
13.2.2 Setting the Print Mode
286
287
13.3 Setting the CHANNEL Screen (printout) 288
13.3.1 Setting the Print Density 288
13.3.2 Setting the Waveform Display Graph Position 289
13.4 Setting the SYSTEM Screen (printout) 290
13.4.1 SET UP Screen (SYSTEM 1) 290
13.4.2 SCALING Screen (SYSTEM 2) 292
13.4.3 COMMENT Screen (SYSTEM 3)
13.4.4 Setting the Output Destination by the COPY Key
13.4.5 Setting the Output Destination by the PRINT Key 294
292
293
13.5 Example of Printer Output 296
13.6 Printing Procedure
13.6.1 Manual Print 303
13.6.2 Auto Print 304
13.6.3 Real Time Print 305
13.6.4 Partial Print 306
13.6.5 Screen Hard Copy 307
13.6.6 List Print 307
13.6.7 Report Print 307
303
13.7 External Printer (color print) 308
Chapter 14 Storing Measurement Data 311
14.1 Floppy Disk 312
14.2 9607 MO UNIT (Option)
14.2.1 Setting the Internal MO 314
14.3 PC Card 315
14.3.1 Using the PC Card Slot (PC Card with a Cable Only) 316
14.4 SCSI Interface 317
14.4.1 SCSI Interface Specifications 317
14.4.2 Setting the MO Disk Drive
14.4.3 Setting the SCSI ID
14.5 Selecting the Media Type 320
14.6 Initializing (FORMAT)
14.7 Saving the Data (SAVE)
313
318 319
321 322
Page 11
14.7.1 Setting the Data to Store 324
14.8 Loading the Data (LOAD) 331
14.9 File Information (INFO) 335
14.10 Deleting the Data (DELETE)
14.11 Sorting Files (SORT)
14.12 Operating the Directory
14.12.1 Making a Directory (MK DIR) 338
14.12.2 Changing a Directory (CH DIR) 339
14.12.3 Deleting a Directory (RM DIR) 340
336 337 338
14.13 Printing the File List 341
Chapter 15 Memory Segmentation Function 343
15.1 Using the Sequential Save Function 344
15.2 Using the Multi-Block Function (MEM)
349
Chapter 16 Waveform Operation Function 353
16.1 Waveform Parameter Calculation (MEM) 353
16.1.1 Making Settings for Waveform Parameter Calculation
16.1.2 Making Settings for Waveform Parameter Evaluation
16.1.3 Setting the Waveform Parameter GO/NG Stop Mode 358
16.1.4 Executing Waveform Parameter Calculation 359
355
357
16.2 Calculating Waveform Data 361
16.2.1 Preparing for Waveform Processing 362
16.2.2 Defining the Processing Equation 363
16.2.3 Copying an Equation 366
16.2.4 Setting the Channel for Recording Processing Results 367
16.2.5 Setting the Display Scale and Number of Moved Points
16.2.6 Perform Waveform Processing
368 370
16.3 Waveform GO/NG Evaluation (MEM, FFT) 372
16.3.1 Setting the Waveform Area 374
16.3.2 Setting the Waveform Evaluation Mode 375
16.3.3 Setting the GO/NG Stop Mode 375
16.3.4 Creating the Evaluation Area 376
16.3.5 Editor Command Details 376
Page 12
Chapter 17 External Input/ Output Terminals/ Key Lock
Function 379
17.1 External Input/Output Terminals 379
17.1.1 Connecting the Terminals 379
17.1.2 External Trigger Input Terminal [EXT TRIG] 380
17.1.3 Trigger Output Terminal [TRIG OUT] 380
17.1.4 External Sampling Terminal [EXT SMPL] 381
17.1.5 External Print Terminal / External Start/Stop Terminal 381
17.1.6 GO/NG Evaluation Output Terminal 382
17.2 Using the Key Lock Function 383
Chapter 18 Specifications 385
18.1 General Specifications 385
18.2 External Data Storage and Interface
18.3 Trigger Unit
18.4 Memory Recorder Function
18.5 Recorder Function
18.6 RMS Recorder Function
18.7 Recorder & Memory Function
18.8 FFT Function
18.9 Auxiliary Functions
18.10 Others
18.11 System Operation
387 389 390 391 392 393 394 395 396 397
18.12 Maximum Recording Length for Time Axis Settings
18.13 Memory Capacity and Recording Length
398 400
Chapter 19 Logic and Analog Inputs 401
19.1 Logic Inputs 401
19.2 Analog Inputs
19.2.1 8936 ANALOG UNIT 403
19.2.2 8937 VOLTAGE / TEMP UNIT
19.2.3 8938 FFT ANALOG UNIT
19.2.4 8939 STRAIN UNIT 407
19.2.5 8946 4ch ANALOG UNIT 408
19.2.6 8940 F/V UNIT 409
19.2.7 8947 CHARGE UNIT 411
403
404 406
Page 13
Chapter 20 Maintenance and Service 415
20.1 Maintenance and Inspection 415
20.2 Replacing the Input Units
20.3 Replacing the Fuse of the 9433
20.4 Removing the Battery Before Discarding the Unit
417 418 419
20.5 Troubleshooting 420
Appendix APPENDIX 1
Appendix 1 Error Messages APPENDIX 1 Appendix 2 Glossary Appendix 3 Reference
Appendix 3.1 Sampling APPENDIX 8 Appendix 3.2 Aliasing APPENDIX 8 Appendix 3.3 Measurement Limit Frequency APPENDIX 9 Appendix 3.4 Recorder Function APPENDIX 10 Appendix 3.5 RMS Recorder Function APPENDIX 11 Appendix 3.6 Averaging Equations APPENDIX 12 Appendix 3.7 "2-point method" Scaling Equation
APPENDIX 12
Appendix 3.8 Waveform Parameter Calculation Details
APPENDIX 13 Appendix 3.9 Details on Operators Appendix 3.10 FFT Function
APPENDIX 6 APPENDIX 8
APPENDIX 16 APPENDIX 19
Appendix 4 Size of a Waveform File (Binary data)
APPENDIX 26
Appendix 5 Size of a Waveform File (Reference value
of text data)
Appendix 6 FFT File Size
APPENDIX 28 APPENDIX 30
Appendix 7 Real-Time Save Files Size (Memory data)
Appendix 8 Waveform Viewer (Wv)
Appendix 8.1 Starting the Waveform Viewer
APPENDIX 33 Appendix 8.2 Waveform Viewer Menus Appendix 8.3 Using the Waveform Viewer Appendix 8.4 Conversion to CSV Format APPENDIX 39 Appendix 8.5 Batch Conversion APPENDIX 40
APPENDIX 31 APPENDIX 32
APPENDIX 34 APPENDIX 36
Index INDEX 1
Page 14
Page 15
────────────────────────────────────────────────────
Introduction
Thank you for purchasing the HIOKI "8841/42 MEMORY HiCORDER." To obtain maximum performance from the product, please read this manual first, and keep it handy for future reference.
Inspection
When the unit is delivered, check and make sure that it has not been damaged in transit. In particular, check the accessories, panel switches, and connectors. If the unit is damaged, or fails to operate according to the specifications, contact your dealer or HIOKI representative.
Accessories
i
Power cord 1 9231 RECORDING PAPER 1 Protective cover 1 Roll paper attachment 2 PC card protector 1 Connector cable label 1 Instruction Manual 1 Application Disk (CD-R) 1
Options
Input units
Cards
8936 ANALOG UNIT 8937 VOLTAGE/TEMP UNIT
8938 FFT ANALOG UNIT 8939 STRAIN UNIT 8940 F/V UNIT
*1
8946 4 ch ANALOG UNIT 8947 CHARGE UNIT
9557 RS-232C CARD 9558 GP-IB CARD 9559 PRINTER CARD 9578 10BASE-T LAN CARD 9607 MO UNIT (with eject pin) 9608 MEMORY BOARD (24M-WORD)total 32 M words 9626 PC CARD 32M 9627 PC CARD 64M 9726 PC CARD 128M 9727 PC CARD 256M 9728 PC CARD 512M 9729 PC CARD 1G
Cables and adapter
9433 DC POWER ADAPTER 9197 CONNECTION CORD (for high voltage, maximum input voltage 500 V) 9198 CONNECTION CORD (for low voltage, maximum input voltage 300 V) 9199 CONVERSION ADAPTOR 9217 CONNECTION CORD (isolated between BNC and BNC) 9320 LOGIC PROBE (maximum input voltage 50 V) 9321 LOGIC PROBE (maximum input voltage 250 V) 9322 DIFFERENTIAL PROBE
────────────────────────────────────────────────────
Introduction
Page 16
ii
────────────────────────────────────────────────────
9324 POWER CORD (for logic connector) 9325 POWER CORD (for 8940 F/V UNIT sensor connector) 9665 10:1PROBE 9666 100:1PROBE
*2
9303 PT 9318 CONVERSION CABLE (for 9270-72, 9277-79) 9319 CONVERSION CABLE (for 3273)
Clamps
3273 CLAMP ON PROBE (DC to 50 MHz) 9018-10 CLAMP ON PROBE (10 to 500 A, 40 Hz to 3 kHz)
*2
9132-10 CLAMP ON PROBE (20 to 1000 A, 40 Hz to 1 kHz)
*2
9270 CLAMP ON SENSOR (20 A, 5 Hz to 50 kHz)
*2
9271 CLAMP ON SENSOR (200 A, 5 Hz to 50 kHz)
*2
9272 CLAMP ON SENSOR (20/200 A 9277 UNIVERSAL CLAMP ON CT (20 A, DC to 100 kHz) 9278 UNIVERSAL CLAMP ON CT (200 A, DC to 100 kHz)
*2
9279 UNIVERSAL CLAMP ON CT (500 A, DC to 20 kHz)
*2
9555 SENSOR UNIT (used with the 9270 to 9272, and the 9277 to 9279)
Others
8910 CAN ADAPTER 9333 LAN COMMUNICATOR 9335 WAVE PROCESSOR 9397-01 CARRYING CASE (for 8841) 9349 CARRYING CASE (for 8842) 9231 RECORDING PAPER (6 rolls)
*2
220H PAPER WINDER
*1: Only used with 8841/ *2: no CE marking
NOTE
The 9270 to 9272 CLAMP ON SENSORs and 9277 to 9279 UNIVERSAL CLAMP ON CTs are designed for use only with the 9555 SENSOR UNIT, and cannot be used alone.
,5Hzto10kHz)
────────────────────────────────────────────────────
Inspection
Page 17
iii
────────────────────────────────────────────────────
Safety Notes
This manual contains information and warnings essential for safe operation of the product and for maintaining it in safe operating condition. Before using the product, be sure to carefully read the following safety notes.
DANGER
Safety symbols
This product is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the product. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from product defects.
In the manual, the symbol indicates particularly important
information that the user should read before using the product.
The symbol printed on the product indicates that the user should refer to a corresponding topic in the manual (marked with the symbol) before using the relevant function.
Indicates a grounding terminal.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates both DC (Direct Current) and AC (Alternating Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
────────────────────────────────────────────────────
Safety Notes
Page 18
iv
────────────────────────────────────────────────────
The following symbols are used in this Instruction Manual to indicate the relative importance of cautions and warnings.
DANGER
Indicates that incorrect operation presents extreme danger of accident resulting in death or serious injury to the user.
WARNING
Indicates that incorrect operation presents significant danger of accident resulting in death or serious injury to the user.
CAUTION
NOTE
Indicates that incorrect operation presents possibility of injury to the user or damage to the equipment.
Denotes items of advice related to performance of the equipment or to its correct operation.
Measurement categories (Overvoltage categories)
This instrument conforms to the safety requirements for CAT II (370 V) measurement products (using with the 8936 ANALOG UNIT). To ensure safe operation of measurement instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT I to CAT IV, and called measurement categories. These are defined as follows.
CAT
Secondary electrical circuits that are connected to a wall outlet through a transformer or similar device.
CAT
Primary electrical circuits in equipment connected to a wall outlet via a power cord (portable tools, household appliances, etc.)
CAT
Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders between the distribution panel and outlets.
CAT
The circuit from the service drop to the service entrance, then to the power meter and to the primary overcurrent protection device.
Higher-numbered categories correspond to electrical environments with greater momentary energy. So a measurement device designed for CAT III environments can endure greater momentary energy than a device designed for CAT II. Using a measurement instrument in an environment designated with a higher-numbered category than that for which the instrument is rated could result in a severe accident, and must be carefully avoided. Never use a CAT I measuring instrument in CAT II, III, or IV environments.
The measurement categories comply with the Overvoltage Categories of the IEC60664 Standards.
────────────────────────────────────────────────────
Safety Notes
Page 19
v
────────────────────────────────────────────────────
Notes on Use
In order to ensure safe operation and to obtain maximum performance from the unit, observe the cautions listed below.
Inspection before Use
Before using the product the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
DANGER
Power Supply Connections
Before connecting the unit, make sure that the power supply voltage matches the rated power supply voltage of the unit.
Probe Connection, Measurement Voltage Input
Maximum input voltage ratings for the input unit and the input terminals
of the unit are shown below. To avoid the risk of electric shock and damage to the unit, take care not to exceed these ratings.
The maximum rated voltage to earth of the input unit (voltage
between input terminals and main unit frame ground, and between inputs of other analog units) is shown below. To avoid the risk of electric shock and damage to the unit, take care that voltage between channels and between a channel and ground does not exceed these ratings.
The maximum rated voltage to earth rating applies also if an input
attenuator or similar is used. Ensure that voltage does not exceed these ratings.
When measuring power line voltages with the 8936 or 8938, always
connect the probe to the secondary side of the circuit breaker. Connection to the primary side involves the risk of electric shock and damage to the unit.
Always use the supplied connector cables. Any exposed metal
sections in a connector cable consist a risk of electric shock. The logic units all have and the unit have a common ground. The external I/O terminal and the unit have a common GND.
Input/output terminal Maximum input voltage Maximum rated voltage to earth
8936 (input) 400 VDC max. 370 VAC/DC
8937 (input) 30 Vrms or 60 VDC 30 Vrms or 60 VDC
8938 (input) 400 VDC max. 370 VAC/DC
8939 (input) 10 VDC max. 30 Vrms or 60 VDC
8940 (input) 30 Vrms or 60 VDC
(BNC/sensor connector terminal)
8946 (input) 30 Vrms or 60 VDC 30 Vrms or 60 VDC
8947 (input)
*1: The maximum allowable charge that can be applied to the miniature connection terminals is 500 pC at the most sensitive of the six ranges, and 50,000 pC at the least sensitive range.
────────────────────────────────────────────────────
*1
30 Vrms or 60 VDC (BNC terminal) 30 Vrms or 60 VDC (BNC terminal)
30 Vrms or 60 VDC (BNC terminal) Not insulated (Sensor connector terminal)
Notes on Use
Page 20
vi
────────────────────────────────────────────────────
Input/output terminal Maximum input voltage Maximum rated voltage to earth
9322
EXT TRIG/ START・STOP/ PRINT/ EXT SMPL
TRIG OUT/ GO/ NG
2000 VDC, 1000 VAC (CAT II)
600 VAC/DC (CAT III)
-5 to +10 VDC
-20 V to +30 VDC
500 mA max./ 200 mW max.
When using grabber clips 1500 VAC/DC (CAT II), 600 VAC/DC (CAT III) When using alligator clips 1000 VAC/DC (CAT II), 600 VAC/DC (CAT III)
Not insulated
DANGER
Logic Probe Connection
The logic input and 8841/42 Unit share a common ground. Therefore, if
power is supplied to the measurement object of the logic probe and to the 8841/42 from different sources, an electric shock or damage to the equipment may result. Even if power is supplied from the same system, if the wiring is such that a potential difference is present between the grounds, current will flow through the logic probe so that the measurement object and 8841/42 could be damaged. We therefore recommend the following connection method to avoid this kind of result. Refer to Section 2.5, "Logic Probe Connection" for details.
(1) Before connecting the logic probe to the measurement object, be sure
that power is supplied from the same outlet box to the measurement object and the 8841/42 using the supplied power cord.
(2) Before connecting the logic probe to the measurement object,
connect the ground of the measurement object to the 8841/42 ground terminal. Also in this case, power should be supplied from the same source. Refer to Section 2.2, "Power Supply and Ground Connections" for grounding terminal details.
Replacing the Input Units
In order to avoid accidents from electric shock, before removing or
replacing an input unit, check that the connector cables and thermocouple are disconnected, turn off the power, and remove the power cable.
Normally keep all input units installed permanently. If a unit is not
fitted, it must be replaced by a blanking panel. If the unit is operated with an input unit not in place it poses a shock hazard.
Differential Probe Connection
When using grabber clips, the 9322's maximum rated voltage to
earth is 1500 VAC/DC; when using alligator clips, it is 1000 VAC/DC. To avoid electrical shock and possible damage to the unit, never apply voltages greater than these limits between the input channel terminals and chassis, or across the inputs of two 9322s.
Maximum input voltage is 1000 VAC/2000 VDC. Do not measure
voltage in excess of these limitations, as doing so may damage the unit or cause an accident that might result in injury or death.
────────────────────────────────────────────────────
Notes on Use
Page 21
vii
────────────────────────────────────────────────────
WARNING
Installation Environment
To avoid electric shock, do not allow the product to get wet, and do
not use it when your hands are wet.
Grounding the Unit
The unit is constructed so as to be connected to a ground line via a
three-core power cord that is supplied with the unit. In order to
avoid electric shock, connect the unit to a properly grounded (3-pin)
outlet using the power cord provided.
Before Powering on
Check that the power supply is correct for the rating of the unit.
(The AC fuse is integrated in the unit.)
Usage Precautions for the Internal MO Drive (option)
Please do not attempt to disassemble the MO drive.
Laser radiation can be emitted when the MO drive is open. Avoid
looking directly into the laser when the MO drive is open. Maximum
laser output is 50 mW (at 685 nm, pulsed).
A laser warning label is attached to the bottom of the 8841/42 Unit.
Refer to Section 1.2, "Identification of Controls and Indicators" for
the label location.
CAUTION
Installation Environment
This product should be installed and operated indoors only, between 5
o
C and 35% to 80%RH.
to 40 This product is not designed to be entirely water- or dust-proof. To avoid
damage, do not use it in a wet or dusty environment. Do not store or use the product where it could be exposed to direct
sunlight, high temperature or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate so that it no longer meets specifications.
Using a printer
Using the printer in a high-temperature or high-humidity environment should be avoided at all costs. This can seriously reduce the printer life.
Using the connecting cables
Use only the specified connection cable. Using a non-specified cable may result in incorrect measurements due to poor connection or other reasons.
Recording paper
This unit uses a thermal printer. The recording paper supplied has characteristics finely tuned for use with the printer. Using recording paper of a different specification may not only result in impaired printing quality, but even prevent the printer from operating. Always use the HIOKI specified product.
Printing is not possible if the recording paper is loaded wrong-side up. (Refer to Section 2.8, "Loading Recoding Paper.")
o
C
────────────────────────────────────────────────────
Notes on Use
Page 22
viii
────────────────────────────────────────────────────
CAUTION
Storing
When the unit is not to be used for an extended period, set the head up/down lever to the "head up" position. This will protect the printer head and prevent deformation of the rubber roller.
Precautions on carrying this equipment
The terminal guard of the equipment protects the inputs. Do not hold this guard when carrying the equipment. To carry this equipment, use the handle. See Section 1.2.
Shipping
Remove the printer paper from the unit. If the paper is left in the unit,
paper support parts may be damaged due to vibrations.
Remove the PC card, floppy disk, MO disk and SCSI cable from the unit.
If reshipping the unit, preferable use the original packing.
Do not transport using the 9397-01 or 9349 CARRYING CASEs.
Others
In the event of problems with operation, first refer to Section 20.5,
"Troubleshooting".
Carefully read and observe all precautions in this manual.
Handling the CD-R
Always hold the disc by the edges, so as not to make fingerprints on the
disc or scratch the printing.
Never touch the recorded side of the disc. Do not place the disc directly
on anything hard.
Do not wet the disc with volatile alcohol or water, as there is a possibility
of the label printing disappearing.
To write on the disc label surface, use a spirit-based felt pen. Do not use
a ball-point pen or hard-tipped pen, because there is a danger of scratching the surface and corrupting the data. Do not use adhesive labels.
Do not expose the disc directly to the sun's rays, or keep it in conditions
of high temperature or humidity, as there is a danger of warping, with consequent loss of data.
To remove dirt, dust, or fingerprints from the disc, wipe with a dry cloth,
or use a CD cleaner. Always wipe radially from the inside to the outside, and do no wipe with circular movements. Never use abrasives or solvent cleaners.
Hioki shall not be held liable for any problems with a computer system
that arises from the use of this CD-R, or for any problem related to the purchase of a Hioki product.
────────────────────────────────────────────────────
Notes on Use
Page 23
ix
────────────────────────────────────────────────────
Chapter Summary
Chapter 1 Product Overview
Contains an overview of the unit and its features.
Chapter 2 Installation and Preparation
Explains how to set the unit up for measurement.
Chapter 3 Basic Key Operation
Explains how to operate the keys and the Jog/Shuttle control for carrying out basic measurement functions.
Chapter 4 Memory Recorder Function
Explains how to use the memory recorder functions of the unit.
Chapter 5 Recorder Function
Explains how to use the recorder functions of the unit.
Chapter 6 RMS Recorder Function
Explains how to use the RMS recorder functions of the unit.
Chapter 7 Recorder and Memory Recorder Function
Explains how to use the recorder and memory recorder functions of the unit.
Chapter 8 FFT Function
Explains how to use the FFT functions of the unit.
Chapter 9 Input Channel Settings
Explains how to make settings using the channel setting screen.
Chapter 10 Trigger Functions
Explains how to use the trigger functions of the unit.
Chapter 11 Waveform Display Screen Operation
Explains how to perform waveform scrolling, how to use the A・B cursors and so.
Chapter 12 SYSTEM Screen Settings
Explains how to make settings using the system setting screen.
Chapter 13 Printout of Measurement Data
Explains how to print out measurement data and how to read printed charts.
Chapter 14 Storing Measurement Data
Explains how to store, recall, and delete measurement data and measurement settings.
Chapter 15 Memory Segmentation Function
Explains how to use the Memory Segmentation Function.
────────────────────────────────────────────────────
Chapter Summary
Page 24
x
────────────────────────────────────────────────────
Chapter 16 Waveform Operation Function
Explains how to use the Calculating, Waveform Parameters/Evaluating Parameter value and Waveform GO/NG Evaluation.
Chapter 17 External Input/ Output Terminals/ Key Lock Function
Gives specifications and usage details of the external input/output terminals, and explains how to use the key lock function.
Chapter 18 Specifications
Contains general specifications and detailed function specifications.
Chapter 19 Logic and Analog Inputs
Contains specifications and precautions for logic input section and input amplifier units.
Chapter 20 Maintenance and Service
Describes maintenance procedures.
Appendix
Contains information that is necessary for using this unit, including a description of error messages, a glossary, and an explanation how to increase memory.
────────────────────────────────────────────────────
Chapter Summary
Page 25
1
────────────────────────────────────────────────────
Chapter 1
1
2
1.1 Major Features
(1) Waveform recording performance
For the 8841, using the 8946 4ch ANALOG UNIT, waveform recording can be performed in up to 16 channels with 12-bit resolution (8841). With the 8842, waveform recording can be performed for up to 16 channels at 12-bit resolution, for any combination of input units.
(2) Vertical display
The 10.4-inch vertically-mounted TFT color LCD display has a resolution of 640×480 dots.
Product Overview
3
4
5
6
7
8
(3) Five functions
Memory recorder with up to 1μs (all channels simultaneously) sampling
period.
Real-time recording capability to paper in recorder function
RMS recorder function for recording rms values of AC power supply lines
and DC sources.
Recorder & Memory Recorder function provides combined recorder and
memory recorder functions
FFT function offers 12 types of analysis functions
(4) Storage capacity
The 8841/8842 has a standard storage capacity of 8 M words, expandable to 32 M words with memory upgrades.
(5) Trigger function
Digital trigger circuit
Trigger types: level trigger, window-in trigger, window-out trigger, voltage
drop trigger, RMS level trigger, period trigger, logic trigger (pattern trigger)
9
10
11
12
13
14
A
────────────────────────────────────────────────────
1.1 Major Features
Page 26
2
────────────────────────────────────────────────────
(6) Simple function key interface (GUI)
Thanks to its GUI-inspired design using large function key graphics, the unit is easy to set up and operate.
(7) On-line help
On-line help guides the user through operation steps and various functions.
(8) Scaling function
By setting the physical amount and the unit to be used for 1 V input, the measurement result can be converted into any desired scale.
(9) Additional recording function
When enabled, the memory is regarded as printer paper.
(10) Input units
The analog inputs are floating, and so each input can be connected to its own independent potentials.
Select input units suitable for measurements.
(11) Built-in thermal printer
Thermosensitive recording method using a thermal line head
The built-in printer delivers waveform printouts on the spot.
The printer can also be used to print screen shots and parameter information.
Report print (A4 size) can be printed.
(12) External storage means
The waveform data and/or setup conditions can be stored on a floppy disk, MO disk, or PC card (SRAM, flash ATA, or HDD card).
(13) SCSI interfaces
If a MO drive is connected to the SCSI interface, the waveform data and/or setup conditions can be stored on a MO disk.
(14) GP-IB, RS-232C, and LAN (10BASE-T) interface
Remote control including input unit is possible.
(15) External print
If a color printer is connected to the printer card, the data can be printed in color.
(16) Dual-language capability
Display language is switchable between Japanese and English.
────────────────────────────────────────────────────
1.1 Major Features
Page 27
3
────────────────────────────────────────────────────
1.2 Identification of Controls and Indicators
1
(example: 8841)
4 5
1 2
3 6
2
3
19
20
21
22
23
24
25
7 8 9
10
11
12
13
14
4
5
6
7
15
8
1
STATUS key
2
CHAN key
3
DISP key
4
SYSTEM key
5
TRIG key
26 27 28
Front Panel
Causes the display to show the STATUS screen which serves for setting most measurement parameters.
Causes the display to show the CHANNEL or Variable screen which serves for making input channel settings.
Causes the display to show measurement and analysis results. Causes the display to show the SYSTEM screen which serves for making
system-wide settings such as for the scaling function.
Causes the display to show the TRIGGER screen. Setting the trigger functions.
16
17
18
9
10
11
12
13
14
A
────────────────────────────────────────────────────
1.2 Identification of Controls and Indicators
Page 28
4
────────────────────────────────────────────────────
6
FILE key
Causes the display to show the FILE screen which serves for reading, storing, etc. the waveform data etc.
7
PRINT key
Serves to print out stored waveforms. Output destination can be set on the SYSTEM4 screen.
8
COPY key
Serves to print out a hard copy of the current screen display. Copy destination can be set on the SYSTEM4 screen
9
FEED key
10
VALUE key
11
WAVE key
12
ABCSRkey
13
Jog
Causes the printer paper to advance for as long as the key is pressed.
Uses to select the numerical values setting.
Uses to select the the waveform scrolling.
Uses to select the the A・B cursor moving.
Rotary control knob that serves to change values, move the A・B cursors, and scroll the waveform.
14
Shuttle
Concentric ring that serves to change values, move the A・B cursors, and to scroll the waveform. The speed of movement is proportional to the rotation angle.
15
CURSOR key
16
STOP key
These keys serve to move the flashing cursor in the four directions.
Stops measurement and analysis. Pressing this key twice stops measurement.
17
START key
Initiates the measurement and analysis. During measurement, the LED above the key is lit.
18
TIME/DIV key
19
MANU TRIG key
20
CH.SET key
Serves to set the speed for inputting and storing the input signal.
Serves to cause manual triggering.
Sets the input channel on the display screen Toggles the screen among those for analog channels (1 to 16) and logic channels (A to D).
21
LEVEL MONl key
22
VIEW key
Serves to check the input signal level.
Serves to indicate the position of the currently displayed screen information in relation to the entire recording length.
Call up information about the status of memory block.
23
AUTO key
Pressing this key activates automatic setting of time axis range and voltage range values of input waveform.
24
HELP key
25
F1 to F5 key
26
Channel select/ input value keys
Provides on-line help.
Serve to select setting items.
Selects channel or sets numerical value
27
RANGE knob
28
POSITION knob
────────────────────────────────────────────────────
1.2 Identification of Controls and Indicators
Sets the measurement range for the channel.
Sets the zero position for the channel.
Page 29
5
────────────────────────────────────────────────────
1
2
3
4
1
AC power switch
2
AC connector
3
Function ground terminal (GND)
5
20
6
7
8
9
1
2
3
4
18
19
5
10
Right Side Panel
Switches on or off the AC power supply.
The supplied power cord must be plugged in here.
Connects to the earth.
11
12 13
14
15
17
16
6
7
4
DC connector
5
Logic probe connectors
6
Input unit slots
7
Fastening screw
8
Analog input connector
9
FD slot
10
SCSI connector
11
PC card slot
12
Eject button
13
External start/stop terminals
14
External print terminal
15
GO evaluation output terminal
16
NG evaluation output terminal
17
Ground terminal (GND)
18
External sampling terminal
19
Trigger terminals
20
Blowing slot
Connects to the optional 9433 DC POWER ADAPTER.
Input connector for the logic input section, designed for the dedicate logic probes (CH A to D).
These slots accept input units.
Secures the plug-in unit.
Unbalanced analog input. (on ANALOG UNIT)
Floppy disk is inserted.
An MO drive can be connected.
Inserts the PC card.
Removes the PC card.
Start and stop operation can be controlled.
Print operation can be controlled.
When the waveform evaluation has resulted in GO, a signal is output from this terminal.
When the waveform evaluation has resulted in NG, a signal is output from this terminal.
Uses with⑬to⑲(except⑰)terminals.
Allows input of an external sampling signal. (in Memory recorder and FFT functions)
Can be used to synchronize multiple units, using the EXT TRIG input and TRIG OUT output.
8
9
10
11
12
13
14
A
────────────────────────────────────────────────────
1.2 Identification of Controls and Indicators
Page 30
6
────────────────────────────────────────────────────
2 23 1 2
Left Side Panel
4
2
5
Upper Panel
Bottom Panel
Rear Panel
1
Handle
2
Ventilation slots
3
Printer
4
Stand
5
Laser warning label (when the 9607 MO UNIT is installed)
────────────────────────────────────────────────────
1.2 Identification of Controls and Indicators
Serves for transporting the 8841/42.
The unit can also be propped up at an angle using the stand.
Page 31
7
────────────────────────────────────────────────────
Chapter 2
1
2
Installation and Preparation
2.1 Installation of the Unit
Installation orientation
Install the unit on a flat, level surface. The unit can also be propped up at an angle, using the stand.
3
4
5
6
7
8
NOTE
Stand
When the MO drive is used, do not use the stand to prop up the unit.
9
10
11
12
13
14
A
────────────────────────────────────────────────────
2.1 Installation of the Unit
Page 32
8
V
V
V
────────────────────────────────────────────────────
Ambient conditions
Temperature 5 to 40 ,23±5 recommended for high-precision
measurements.
Humidity 35 to 80%RH (no condensation); 50±10%RH (no
condensation) recommended for high-precision measurements.
Ventilation Take care not to block the ventilation openings and assure
proper ventilation. When using the unit in an upright position, take care not to block the openings on both side.
entilation
Blowing slot
entilation
Right side
Upper side
Avoid the following locations:
Subject to direct sunlight.
Subject to high levels of dust, steam, or corrosive gases (Avoid using the
entilation
Left side
Bottom side
equipment in an environment containing corrosive gases (e.g., H2S, SO2, NI2,andCI2) or substances that generate harmful gasses (e.g., organic silicones, cyanides, and formalins).
Subject to vibrations.
In the vicinity of equipment generating strong electromagnetic fields.
────────────────────────────────────────────────────
2.1 Installation of the Unit
Page 33
9
────────────────────────────────────────────────────
2.2 Power Supply and Ground Connections
1
2.2.1 Connecting the AC Power Supply
WARNING
Check the following points before connecting the unit to a power
supply. Take care never to exceed the power supply ratings given below, to avoid the risk of electric shock and damage to the unit.
Power supply matches Rated supply voltage (100 to 240 VAC:
Voltage fluctuations of taken into account.) and rated supply frequency (50/60 Hz).
The AC power switch of the 8841/42 is set to OFF.
Use only the supplied AC power cord.
When supplying power with an inverter or an uninterruptible power
supply (UPS), use a device that complies with the following conditions. To avoid the risk of electric shock and damage to the unit, do not use devices that have a voltage frequency outside the specified range, or that output square waves.
Voltage: AC100 V to 240 V
Power frequency: 50/60 Hz
Sine wave output (Do not use devices that have an unstable
output, even if the output is sinusoidal.)
To avoid electric shock and ensure safe operation, connect the
power cable to a grounded (3-contact) outlet.
10% from the rated supply voltage are
2
3
4
5
6
7
8
NOTE
1. Verify that the AC power switch of the 8841/42 is set to OFF.
2. Plug the grounded three-core power cord supplied into the AC power connector on the right side of the 8841/42.
3. Plug the power cord into an AC outlet corresponding to the rating of the 8841/42.
The fuse is incorporated in power supply. It is not user-replaceable. If a problem is found, contact your nearest dealer.
9
10
11
12
13
14
A
────────────────────────────────────────────────────
2.2 Power Supply and Ground Connections
Page 34
10
────────────────────────────────────────────────────
2.2.2 Connecting the 9433 DC POWER ADAPTER
DANGER
WARNING
Before making connections, make sure that the 9433 DC POWER ADAPTER switch is turned off. If the 8841/8842 is connected to a device such as a battery while the switch is in the on position, a spark may be given off, damaging the 8841/8842.
The rated supply voltage of the 9433 is 10 to 28 VDC (Voltage
fluctuations of account.).
The 9433 DC POWER ADAPTER generates heat. To provide
adequate heat dissipation, avoid placing objects on top of the 8841/8842 or mounting it in confined spaces.
When connecting the 9433, take care not to mix up the red (+) and
black (-) lead. If polarity is reversed, the 9433 may be damaged.
When wishing to extend DC cable, use a cable of identical or better
rating as the cable of the input side.
The 9433 is made for use specifically with the 8841/8842 MEMORY
HiCORDER. Do not use with any other equipment.
10% from the rated supply voltage are taken into
NOTE
1. Verify that the 9433 switch is set to OFF.
2. Align the ridge of the connector on the unit with the groove in the plug and insert the plug fully, and rotate the plug of the connector to lock.
3. Connect the red input cable to the positive side (+) and the black cable to the negative side (-).
4. To remove the cable, rotate it.
This unit is not equipped to charge an external battery.
When using a battery, take care not to deplete it completely.
When the 8841/8842 detects an overcurrent or overvoltage on the output, it will shut down. To turn power back on, turn off 9433 DC POWER ADAPTER switch, wait approximately one minute, then turn on the switch.
If both the AC power supply and 9433 are connected to the 8841/8842, the AC power supply has priority. However, even when the 8841/8842 runs off the AC power supply, the 9433 remains in standby mode and consumes some power. Keep this in mind when connecting the 9433 to the battery while the 8841/8842 is connected to an AC power source.
Input cord rating: 37 A (allowable current)
Failure of the 9433 to power on may indicate a blown fuse.
For replacing the fuse, see Section 20.3.
────────────────────────────────────────────────────
2.2 Power Supply and Ground Connections
Page 35
11
────────────────────────────────────────────────────
Estimated battery operation hours (at room temperature)
Battery type: 12 V, 38 Ah, fully charged
Units
Operation condition
Printer not used
(trigger waiting)
Printer used
(recorder function
500 ms/DIV,
all store)
Actual running time may differ, depending on battery age, charge condition, ambient temperature, and other factors.
If the 9433 is connected to the 8841/8842, and the adapter switch is turned on, even when the 8841/8842 runs off the AC power supply, the 9433 remains in standby mode and consumes some power. In this case, the battery operation time is 95 hours.
Specification of the 9433 DC POWER ADAPTER
Accuracy at 23
8936 installed 8946 installed 8936 installed
Approx. 9 h Approx. 7 h 30 min Approx. 7 h
Approx. 2 h 50 min Approx. 2 h 40 min Approx. 2 h 40 min
±
, 35% to 80%RH, after 30-minutes warming-up time.
5
8841 8842
1
2
3
4
5
6
7
Input voltage range
Rated output voltage
Output voltage accuracy
Rated output current
Efficiency
Maximum rated power
Temperature and humidity range for use
Temperature and humidity range for storing
Location for use
Dielectric strength
Insulation voltage
Dimensions
Mass
10 to 28 VDC
24 VDC
24 VDC±2%
7A
70% or more(12 VDC input, rated output
250 VA
0 to 40 (32 to 104 ), 35% to 85%RH (with no condensation)
-20 to 60 (-4 to 140 ) 10% to 90%RH (with no condensation)
According to specifications for the 8841/8842 MEMORY HiCORDER
500 VDC for a minute (between input and output, between input and unit)
100 MΩor more / 500 VDC (between input and output, between input and unit)
Approx. 110W×65H×150D mm (4.33"W
2.56"H×5.91"D) Input cable: 2000 mm (78.7") Output cable: 500 mm (19.69")
Approx. 1250 g (44.1 oz) approx.
×
8
9
10
11
12
13
14
Accessory
────────────────────────────────────────────────────
Spare fuse(class A melting fuse (NM) 30 A/125 V, 6.4 dia.×31.8 mm
2.2 Power Supply and Ground Connections
A
Page 36
12
────────────────────────────────────────────────────
2.2.3 Functional Grounding of the 8841/8842
When the 8841/8842 is used in noise-prone environments, connect the functional grounding terminal to improve noise characteristics.
Functional grounding terminal
────────────────────────────────────────────────────
2.2 Power Supply and Ground Connections
Page 37
13
────────────────────────────────────────────────────
2.3 Power On/Off
WARNING
Check the following points before the power switch is turned on.
Power supply matches Rated supply voltage (100 to 240 VAC:
Voltage fluctuations of
taken into account.) and rated supply frequency (50/60 Hz). The Unit is correctly installed (Section 2.1). Power cord is correctly connected (Section 2.2). The unit is properly grounded.
There is no need for the user to manually select AC or DC power.
When both AC and DC power are connected, AC power has priority.
10% from the rated supply voltage are
1
2
3
4
5
6
NOTE
If both the AC power supply and 9433 DC POWER ADAPTER are connected to the 8841/8842, the AC power supply has priority. However, even when the 8841/8842 runs off the AC power supply, the 9433 remains in standby mode and consumes some power. Keep this in mind when connecting the 9433 to the battery while the 8841/8842 is connected to an AC power source.
After the power switch is turned on, wait approximately 30 min for the 8936, 8938, 8946 and approximately 1 hour for the 8937, 8939, to stabilize the inside temperature of the connected input unit in order to obtain accurate waveforms. Then, make a zero adjustment of the 8936, 8937, 8938, 8946 or perform an auto balancing of the 8939 prior to measurement. Zero adjustment →see Section 9.5.1. Auto balancing →see Section 9.13.
When the unit is turned off, it memorizes the currently used settings and reestablishes the same settings the next time the unit is turned on again.
7
8
9
10
11
12
13
14
────────────────────────────────────────────────────
2.3 Power On/Off
A
Page 38
14
4
────────────────────────────────────────────────────
2.4 Connection of the Input Unit
2.4.1 8936/ 8938/ 8946 ANALOG UNITs
WARNING
CAUTION
NOTE
Connector Cable
Never connect the probe to the 8841/42 while the probe is already
connected to the measurement object. Otherwise there is a risk of electric shock.
Use only the specified connection cables.
An insulated BNC connector is used for the specified connection cables to prevent electric shock. If a metal BNC connector is used, electric shock may result, as the input L-terminal and the metal part of the BNC connector will have the same potential.
When disconnecting the BNC connector, be sure to release the lock before pulling off the connector. Forcibly pulling the connector without releasing the lock, or pulling on the cable, can damage the connector.
Use only the specified connection cables. Using a non-specified cable may result in incorrect measurements due to poor connection or other reasons. In addition, the BNC connector may be damaged. For safety reasons, only use the optional 9197, 9198, or 9217
CONNECTION CORD for connection to the analog input units. (For the 8946, use the 9198 CONNECTION CORD.)
9197
(500 V)
(Maximum input voltage :)
Groove of the BNC
Connector guide
9198 (300 V)
9217 (300 V)
Connecting to the main unit
1. Align the BNC connector with the guide groove of the unit input connector, and turn clockwise while pressing in to lock the connector.
2. To remove from the unit, turn the BNC connector counterclockwise to release the lock, then pull it.
Setting channels of the 8946 4 ch ANALOG UNIT (8841 only)
When the 8946 4-channel analog unit is used, input channels 1 to 4 are not
set in this customary sequence on the 8841. The slot position at which the 8946 is mounted determines the set input channel (see the illustration below). To confirm the position of the setting channels, see Section 9.6.
Illustrates active channels when the 8946 is mounted
9
2
1
3
6
5
8
7
11
13
15
10
12
14
16
in each slot.
────────────────────────────────────────────────────
2.4 Connection of the Input Unit
Page 39
15
────────────────────────────────────────────────────
2.4.2 8937 VOLTAGE/TEMP UNIT
WARNING
CAUTION
Outer insulation (mantle)
Thermocouple
Inner insulation
25 mm
10 mm
leads
Never connect the cable to the unit while the cable is already
connected to the measurement object. Otherwise there is a risk of electric shock.
When an uninsulated thermocouple is used to measure temperature
at a point carrying electric potential, take care not to touch the terminals and connector screws. Otherwise there is a risk of electric shock.
Between voltage and temperature inputs and unit is insulated.
When disconnecting the BNC connector, be sure to release the lock before pulling off the connector. Forcibly pulling the connector without releasing the lock, or pulling on the cable, can damage the connector. A common GND is used for voltage and temperature input on all channels. Never input voltage and temperature simultaneously, since doing so could result in damage to the sample being tested.
Connecting to the connector cable (voltage measurement)
1. Align the BNC connector with the guide groove of the 8841/42 input connector, and turn clockwise while pressing in to lock the connector. (Use the 9198 CONNECTION CORD for voltage measurement with the
8937.)
2. To remove from the unit, turn the BNC connector counterclockwise to release the lock, then pull it.
Thermocouple connection (temperature measurement)
1. Strip off the insulation as shown in the illustration.
2. Push the tab with a flatblade screwdriver or similar.
3. While keeping the tab depressed, insert a stripped wire into the connector opening.
4. Release the tab to lock the wire.
5. To remove the thermocouple, pull it out with the button depressed.
NOTE
To mount and remove the thermocouple, use a tool such as a screwdriver.The push-button terminal block of the 8937 VOLTAGE/TEMP UNIT is
specialized for the thermocouple. Do not use thermocouples other than the specifiedtypes(K,J,E,T,N,R,S,B).
If the thermocouple is connected in reverse, the temperature reading will not
be correct.
────────────────────────────────────────────────────
2.4 Connection of the Input Unit
Page 40
16
++−
────────────────────────────────────────────────────
2.4.3 8939 STRAIN UNIT
CAUTION
Connect only the sensor to the conversion cable supplied with the 8939 STRAIN UNIT. To disconnect the conversion cable, always unlock the plug and pull out the cable.
Connecting to the main unit
1. Align the projection on the unit connector with the cutout on the conversion cable, and insert the plug into the connector.
2. Turn the fixing guide (the colored area in the figure below) so that it engages with the connector guides on the unit, fully insert the fixing guide, and turn it clockwise to lock the plug.
3. To remove the conversion cable from the unit, turn the fixing guide (the colored area in the figure below) counterclockwise to unlock the plug, and pull out the plug.
Connector guide
Projection on the unit connector
Connector
Bridge voltage:2V (Apply voltage)
Cutout on the conversion cable
Fixing guide
Input voltage
────────────────────────────────────────────────────
2.4 Connection of the Input Unit
Page 41
17
────────────────────────────────────────────────────
2.4.4 8940 F/V UNIT
DANGER
To avoid electrical accidents, make sure that the MEMORY HiCORDER and the equipment being measured are powered off before making connections. Do not make connections with the power turned on.
When using the 9318/9319 CONVERSION CABLE, there is no isolation between GND the MEMORY HiCORDER and GND of the clamp on sensor/probe. Exercise extreme care in connection to avoid possible damage to the equipment or personal injury.
When connecting 8940 F/V UNIT to 3273, and conductors being measured carry in excess of the safe voltage level (SELV-E) and not more than 300 V, to prevent short circuits and electric shock while the core section is open, make sure that conductors to be measured are insulated with material conforming to (1) Measurement Category l, (2) Double Insulation (Reinforced insulation) Requirements for Working Voltage of 300 V, and (3) Pollution Degree 2. For safeties sake, never use this sensor on bare conductors. The core and shield case are not insulated.
When connecting 8940 F/V UNIT to 3273, do not damage insulation sheathing on testing device.
Refer to the following standards regarding the meanings of underlined terms. IEC 61010-1, IEC 61010-2-031, IEC 61010-2-032
WARNING
CAUTION
When using the clamp-on sensor or clamp-on probe, be sure to use the optional 9318 or 9319 CONVERSION CABLE.
When disconnecting the BNC connector, be sure to release the lock before pulling off the connector. Forcibly pulling the connector without releasing the lock, or pulling on the cable, can damage the connector.
Connection cable connection (Frequency, count, pulse duty ratio and voltage measurement)
Groove of the BNC
Connector guide
Use the optional 9198 CONNECTION CORD for connection to the F/V UNIT.
1. Align the BNC connector with the guide groove of the 8841 input connector, and turn clockwise while pressing in to lock the connector.
2. To remove from the unit, turn the BNC connector counterclockwise to release the lock, then pull it.
────────────────────────────────────────────────────
2.4 Connection of the Input Unit
Page 42
18
────────────────────────────────────────────────────
Clamp connection (Current measurement)
The following clamp-on sensors and clamp-on probes can be connected using the 9318 and 9319 CONVERSION CORDs.
9318
It can be connected to a maximum of 4 channels.
Unit's sensor connector
Conversion cable plug
Conversion cable connector
Adapted clamp's
Projection on the unit connector
Groove on the 3273 CLAMP ON PROBE's termination connector
Sensor connector
on the unit
Conversion cable plug
9270, 9271, 9272, 9277, 9278, 9279
Connecting the 9318 CONVERSION CABLE
1. Align the groove on the conversion cable plug with the sensor connector on the F/V unit and push inward until the connector locks into place.
2. Align the groove on the conversion cable connector
plug
with the adapted clamp on sensor plug and push inward until the connector locks into place.
3. To unplug the cables, slide the lock ring on each plug outward to unlock it, then pull out the plug.
Connecting the 9319 CONVERSION CABLE
1. Align the groove on the 3273 CLAMP ON PROBE's termination connector with the pin on the BNC connector on the F/V unit, then slide the termination connector over the BNC connector and turn to lock it in place.
2. Align the groove on the conversion cable plug with the sensor connector on the F/V unit and push inward until the connector locks into place.
3. Unlock the conversion cable connector and the power plug on the 3273 before unplugging the cable.
4. To unplug the cables, slide the lock ring on each plug outward to unlock it, then pull out the plug.
9319
3273
────────────────────────────────────────────────────
2.4 Connection of the Input Unit
Page 43
19
────────────────────────────────────────────────────
2.4.5 8947 CHARGE UNIT
WARNING
Never connect the connection cable to the unit while it is connected
to the measurement object, to avoid electric shock.
The BNC connectors for each channel and the miniature connection
terminals share a common ground. Do not connect both types of terminals at the same time.
When the PREAMP measurement mode is selected, voltage (15 V at
2 mA) is applied internally to the BNC connector when measurement starts. To avoid electric shock and damage to measurement objects, select a measurement mode other than PREAMP or turn the unit off when connecting a sensor or probe to the BNC terminals.
Before using an Internal Preamp Type Acceleration Sensor, be sure
that it conforms to the 8947 specifications (15 V at 2 mA). Using a non-conforming sensor could result in damage to the sensor.
Acceleration sensors compatible with the 8947 are as follows: Acceleration sensors with internal preamp (BNC connector with 15 V at 2 mA operating power) Charge-output type accelerator sensors (miniature 10-32 connection terminal)
Connecting an Internal Preamp Type Acceleration Sensor (PREAMP and VOLTAGE modes)
Use a cable with BNC connector to connect the pickup sensor to the unit in PREAMP mode. We recommend the Model 9198 CONNECTION CORD for the VOLTAGE mode.
1. Align the BNC connector with the guide groove of the 8841/42 input connector, and turn clockwise while pressing in to lock the connector.
2. To remove from the unit, turn the BNC connector counterclockwise to release the lock, then pull it.
Connecting a Charge-Output Type Acceleration Sensor (CHARGE mode)
Use a cable with a plug to mate with the miniature connector to connect the pickup sensor to the unit in CHARGE mode (10-32 miniature connection terminal).
1. Insert the miniature connector plug into the miniature connector on the unit, and turn it clockwise until tight.
2. To remove the miniature connector, turn the plug counterclockwise.
────────────────────────────────────────────────────
2.4 Connection of the Input Unit
Page 44
20
y
────────────────────────────────────────────────────
2.5 Logic Probe Connection
DANGER
The logic input and 8841/42 Unit share a common ground. Therefore, if power is supplied to the measurement object of the logic probe and to the 8841/42 from different sources, an electric shock or damage to the equipment may result. Even if power is supplied from the same system, if the wiring is such that a potential difference is present between the grounds, current will flow through the logic probe so that the measurement object and 8841/42 could be damaged. We therefore recommend the following connection method to avoid this kind of result.
(1) Before connecting the logic probe to the measurement object, be
sure that power is supplied from the same outlet box to the measurement object and the 8841/42 using the supplied power cord.
(2) Before connecting the logic probe to the measurement object,
connect the ground of the measurement object to the 8841/42 ground terminal. Also in this case, power should be supplied from the same source. Refer to Section 2.2, "Power Supply and Ground Connections" for grounding terminal details.
Object to be measured
8841/42
Object to be measured
8841/42
NOTE
GND
Functionground terminal
In this case too, obtain power from the same supply.
Power suppl
Figure 2
Outlet
Figure 1
Ground
The logic input is located on the rear of the unit. Up
to four probes can be connected.
Since one logic probe can record 4 channels, the
combined maximum recording capability for logic waveforms is 16 channels.
Connect the probe by aligning the groove on the plug
with the ridge on the connector.
If no logic probe is connected, the corresponding logic waveform is
displayed on the screen at high level.
Carefully read the instruction manual supplied with the probe.Do not connect logic probes other than supplied by HIOKI to the logic
inputs.
────────────────────────────────────────────────────
2.5 Logic Probe Connection
Page 45
21
────────────────────────────────────────────────────
2.6 9018-10/9132-10 CLAMP ON PROBE
DANGER
Clamp-on probe should only be connected to the secondary side of a breaker, so the breaker can prevent an accident if a short circuit occurs. Connections should never be made to the primary side of a breaker, because unrestricted current flow could cause a serious accident if a short circuit occurs.
Before using the 9018-10/9132-10 CLAMP ON PROBE, be sure to carefully read its instruction manual and familiarize yourself with the operation principles of this product.
HIOKI
The
9018-10/9132-10 CLAMP ON PROBE can be used to perform current measurement. This probe has a voltage output. The 8841/42 is designed for voltage input. It cannot be used to perform current measurements by itself.
Connections
Connect the BNC connector of the 9018-10/9132-10 CLAMP ON PROBE to the analog input of the 8841/42. Refer to "2.4 Connection of the Input Unit."
9132-10
NOTE
9018-10
The 8841/42 will indicate measurement results as voltage.Using the scaling function, units can be converted to "A" for display.When the clamp-on probe is used for measurement, the measurement
precision will be affected both by the 8841/42 precision and clamp-on probe precision ratings. The same is true for cases where other clamps are used.
When using the 9018/9132 CLAMP ON PROBE, always use the 9199
CONVERSION ADAPTOR.
────────────────────────────────────────────────────
2.6 9018-10/9132-10 CLAMP ON PROBE
Page 46
22
────────────────────────────────────────────────────
2.7 9322 DIFFERENTIAL PROBE Connection
DANGER
When using grabber clips, the 9322's maximum rated voltage to earth is 1500 VAC/DC; when using alligator clips, it is 1000 VAC/DC. To avoid electrical shock and possible damage to the unit, never apply voltages greater than these limits between the input channel terminals and chassis, or across the inputs of two 9322s.
Maximum input voltage is 1000 VAC/2000 VDC. Do not measure voltage in excess of these limitations, as doing so may damage the unit or cause an accident that might result in injury or death.
9322 is a differential probe that connects to input of 8841/42 MEMORY HiCORDER input unit. For more details, refer to its instruction manual.
────────────────────────────────────────────────────
2.7 9322 DIFFERENTIAL PROBE Connection
Page 47
23
A
────────────────────────────────────────────────────
2.8 Loading Recoding Paper
1. Press the stock cover and open it.
Stock cover
Head up/down
lever
2. Put down the head up/down lever.
Holder
Printer roller
3. Insert the attachments into the ends of the roll of recording paper and set the paper into its holder.
ttachment
4. Insert the leading edge of the recording paper from above into the gap behind the printer roller, and pull it out to the other side.
Back side of
paper
────────────────────────────────────────────────────
2.8 Loading Recoding Paper
Page 48
24
────────────────────────────────────────────────────
5. Align the edges of the recording paper you pulled out of the printer with the edges of the recording paper set into the holder. If the edges of the recording paper are not aligned exactly, the paper will come out crooked when printing.
6. Raise the head up/down lever.
7. Pull the recording paper to the outside through the printer exit slot in the stock cover.
NOTE
8. Close the stock cover, and finish by tearing off the recording paper against the edge of the printer exit slot.
Make sure that the recording papers positioned quite straight, if the
recording paper is slanted with respect to the roller there is a danger that later a paper jam will occur.
Always put the unit in the head up condition when it is to be transported or
if it is to be stored for a long period of time. If the unit is left to lie in the state where the roller is being subjected to pressure by the head, then the roller may become deformed or the characters may become uneven.
Particularly care should be taken not to put the recording paper in back to
front by mistake, because if this happens the waveform cannot be drawn.
────────────────────────────────────────────────────
2.8 Loading Recoding Paper
Page 49
25
────────────────────────────────────────────────────
2.9 Care of Recording Paper
Care of recording paper
While unopened, thermal paper will not be affected by the environment,
provided that ambient temperature and humidity do not exceed normal levels. For long-term storage, temperature should be lower than 40 temperatures cause no problem.
After opening, protect the paper from strong light, to prevent discoloration.
Storing data
The recording paper uses a thermochemical reaction. Note the following points:
To avoid discoloration, do not leave recording paper in direct sunlight.Store at not more than 40 and 90%RH.To keep definitive data, make photocopies of the recordings.Thermal paper will blacken when brought into contact with alcohol, ester,
ketone, or other volatile organic substances.
If the thermal paper absorbs an organic solvent such as alcohols or ketones it
may no longer develop properly, and recorded information may fade. Soft PVC film and transparent contact adhesive tape contain such solvents, so avoid using them with recordings.
Avoid interleaving the thermal recordings with damp diazo copies.
.Low
Direct sunlight Organic solvent
────────────────────────────────────────────────────
2.9 Care of Recording Paper
Page 50
26
────────────────────────────────────────────────────
2.10 Notes on Measurement
DANGER
Maximum input voltage ratings for the input unit and the input terminals of the unit are shown below. To avoid the risk of electric shock and damage to the unit, take care not to exceed these ratings.
The maximum rated voltage to earth of the input unit (voltage
between input terminals and main unit frame ground, and between inputs of other analog units) is shown below. To avoid the risk of electric shock and damage to the unit, take care that voltage between channels and between a channel and ground does not exceed these ratings.
The maximum rated voltage to earth rating applies also if an input
attenuator or similar is used. Ensure that voltage does not exceed these ratings.
When measuring power line voltages with the 8936 or 8938, always
connect the probe to the secondary side of the circuit breaker. Connection to the primary side involves the risk of electric shock and damage to the unit.
Input/output terminal Maximum input voltage Maximum rated voltage to earth
8936 (input) 400 VDC max. 370 VAC/DC
8937 (input) 30 Vrms or 60 VDC 30 Vrms or 60 VDC
8938 (input) 400 VDC max. 370 VAC/DC
8939 (input) 10 VDC max. 30 Vrms or 60 VDC
8940 (input) 30 Vrms or 60 VDC
(BNC/sensor connector terminal)
8946 (input) 30 Vrms or 60 VDC 30 Vrms or 60 VDC
8947 (input)
9322
EXT TRIG/ START STOP/ PRINT/ EXT SMPL
TRIG OUT/ GO/ NG
The maximum allowable charge that can be applied to the miniature connection terminals is 500 pC at the most sensitive of the six ranges, and 50,000 pC at the least
*1
30 Vrms or 60 VDC (BNC terminal)
2000 VDC, 1000 VAC (CAT II) 600 VAC/DC (CAT III)
-5 to +10 VDC
-20 V to +30 VDC 500 mA max./ 200 mW max.
30 Vrms or 60 VDC (BNC terminal) Not insulated connector terminal)
30 Vrms or 60 VDC (BNC terminal)
When using grabber clips 1500 VAC/DC (CAT II) 600 VAC/DC (CAT III) When using alligator clips 1000 VAC/DC (CAT II) 600 VAC/DC (CAT III)
(
Not insulated
Sensor
In order to avoid accidents from electric shock, before removing or
replacing an input unit, check that the connector cables and thermocouple are disconnected, turn off the power, and remove the power cable.
Normally keep all two input units installed permanently. If a unit is
not fitted, it must be replaced by a blanking panel. If the unit is operated with an input unit not in place it poses a shock hazard.
────────────────────────────────────────────────────
2.10 Notes on Measurement
Page 51
27
────────────────────────────────────────────────────
CAUTION
NOTE
Before using the unit, make sure that the sheathing on the input cables is not damaged and that no bare wire is exposed. If there is damage, using the unit could cause electric shock. Replace with the specified 9197 or 9198 CONNECTION CORD. When making measurements on an AC power line for example, using a voltage transformer, be sure to connect the voltage transformer ground terminal to ground.
Strong wind striking the input terminal can disrupt the thermal balance of
the input circuit, resulting in incorrect readings. When taking measurements in windy environments, arrange the equipment to prevent wind from directly striking the input terminal of the 8937.
Abrupt changes on ambient temperature can also disrupt the thermal
balance of the input circuit. To prevent measurement error, allow the unit to adjust to the new temperature for about one hour before starting measurement.
Difference between "370 VAC/DC" and "400 VDC max." indication
370 VAC/DC: RMS value is displayed. 400 VDC max.: Instantaneous value is displayed.
The maximum input voltage (400 VDC max.) is defined as the superposition of DC component and AC peak, as shown in the figure below.
AC DC DC+AC
+400 VDC
0V
-400 VDC
────────────────────────────────────────────────────
2.10 Notes on Measurement
Page 52
28
────────────────────────────────────────────────────
2.10.1 Maximum Input Voltage
8936 ANALOG UNIT, 8938 FFT ANALOG UNIT
8841/42
GND
8937 VOLTAGE/TEMP UNIT
8841/42
GND
8939 STRAIN UNIT
8841/42
GND
8940 F/V UNIT
8936 8938
8937 8946
8939
H
L
H
L
H
L
400 VDC max.
370 VAC/DC
30 Vrms or 60 VDC
30 Vrms or 60 VDC
10 VDC max.
30 Vrms or 60 VDC
370 VAC/DC
30 Vrms or 60 VDC
30 Vrms or 60 VDC
8841/42
GND
8947 CHARGE UNIT
8841/42
GND
8940
8947
Connector
BNC
BNC
Miniature connector
H
L
H
L
H
L
H
L
30 Vrms or 60 VDC
30 Vrms or 60 VDC
30 Vrms or 60 V
30 Vrms or 60 VDC
Maximum input load
30 Vrms or 60 VDC
Maximum input load
±
500 pC(High sensitivity: 6 range
±
50,000 pC(Low sensitivity: 6 ranges
Common
30 Vrms or 60 VDC
30 Vrms or 60 VDC
────────────────────────────────────────────────────
2.10 Notes on Measurement
Page 53
29
────────────────────────────────────────────────────
2.10.2 Using a Voltage Transformer
When making measurements on an AC power line for example, using a voltage transformer, be sure to connect the voltage transformer ground terminal to ground.
When the voltage transformer has a ground terminal
8841/42
GND terminal
8936
Ground
H L
PT
Voltage transformer (PT) ground terminal
When the voltage transformer has no ground terminal
8841/42
GND terminal
8936
Ground
H L
PT
Input
Input
────────────────────────────────────────────────────
2.10 Notes on Measurement
Page 54
30
────────────────────────────────────────────────────
────────────────────────────────────────────────────
2.10 Notes on Measurement
Page 55
31
────────────────────────────────────────────────────
Chapter 3
1
2
Basic Key Operation
3.1 Basic Key Operation
3.1.1 Basic Display Operation (MENU)
STATUS
CHAN
STATUS
CHAN
key
key
Displays the STATUS screen. Serves to switch pages of the STATUS screen. Serves to make main settings for various functions on the STATUS screen.
Displays the CHANNEL screen. Serves to set measurement range, position, etc. for input channels. Serves to switch pages of the CHANNEL screen.
3
4
5
6
7
8
9
DISP
DISP
SYSTEM
TRIG
FILE
key
SYSTEM
TRIG
key
FILE
key
key
3.1.2 Printer Key Operation
PRINT
COPY
key
key
PRINT
COPY
Displays the display screen. Serves to display and observe waveforms.
Displays the SYSTEM screen. Serves to switch pages of the SYSTEM screen. Serves to make common settings for all functions (clock setting, comment input, etc.) on the SYSTEM screen.
Displays the TRIGGER screen and serves to set trigger. Serves to switch pages of the trigger screen.
Displays the FILE screen. Serves to read and store waveform data on the FILE screen. See Chapter 14.
Serves to print out the waveform. Output destination by the See Section 12.5.2.
Produces a hard copy of the display content. Copy destination by the See Section 12.5.1.
COPY
key can be selected.
PRINT
key can be selected.
10
11
12
13
14
A
FEED
────────────────────────────────────────────────────
FEED
key
Forwards the paper while the key is held down.
3.1 Basic Key Operation
Page 56
32
────────────────────────────────────────────────────
3.1.3 Setting the Items
F1 F5
to
F1 to F5
(Function keys)
CURSOR
keys
The respective items are shown in the function key display. Select and set the items.
Serve to move the flashing cursor or use the key lock function.
3.1.4 Jog/Shuttle Controls and Select Key
Use the select key when the screen is in measurement display mode. Selected key lights LED.
VALUE
WAVE
A.B CSR
When the LED "VALUE" is lit and the following function keys are displayed, values and items may be set with the Jog/Shuttle control.
Entering the numerical values.
Waveform scrolling (See Section 11.1)
Movement of A・B cursors (See Section 11.2)
Increases in number.
Move the item selection cursor up in the selection window.
Decreases in number.
Increases in number, large step
Increases in number, small step
Decreases in number, small step
Decreases in number, large step
Move the item selection cursor down in the selection window.
Increases in number, 10-units
Increases in number, 1-units
Decreases in number, 1-units
Decreases in number, 10-units
────────────────────────────────────────────────────
3.1 Basic Key Operation
Page 57
33
────────────────────────────────────────────────────
3.1.5 Basic Input Operation
Set the basic settings for analog channels. See Chapter 9.
1
2
3
CH1 - CH16
RANGE
POSITION
TIME/DIV
knob
knob
key
keys
Used to select channels and input values for connected input units In the DISPLAY or CHANNEL screen, these keys are used to select channels or set channels directly by means of the
POSITION
In the value input, these keys are used to set values.
Serves to set the measurement range for the channel selected by the channel select key.
Serves to set the zero position for the channel selected by the channel select key.
Serves to set the input signal capture speed.
and
RANGE
knobs.
3.1.6 Measurement Start and Stop Operation
LED: light during measurement
START
STOP
key
key
Press the to trigger standby.
Press the to stop the measurement.
START
STOP
key to initiate measurement or set the unit
key during measurement and trigger waiting
4
5
6
7
8
9
10
NOTE
────────────────────────────────────────────────────
The measurement start/stop operation is separately determined by each measurement function. For details, see "Start and Stop Operation," which explains each measurement function.
3.1 Basic Key Operation
11
12
13
14
A
Page 58
34
y
y
────────────────────────────────────────────────────
3.2 Other Keys Operation
MANU
TRIG
CH.SET
LEVEL
MONI
VIEW
AUTO
HELP
MANU TRIG
Manual trigger
CH. SET
Channel settings
LEVEL MONI
VIEW
AUTO
HELP
key
key
key
key
key
key
When the unit is in trigger standby mode, pressing this key causes manual triggering. See Section 10.12.
Enables the measurement conditions for each channel on the display screen to be set or changed. Settings remain in effect even during startup. (Settings are not in effect during waveform parameter calculation setup.) The setup screen for "ch 1 to 16," or "logic A to D" can be selected on the channel and trigger screens. See Section 9.3.
Displays the level monitor. See Section 11.5.
Shows at the bottom of the screen the position with respect to the recording length of the displayed part of the waveform. Call up information about the status of memory block. See Section 11.6.
Automatically sets time axis range, measurement range, and zero position of the input waveforms with the memory recorder function. See Section 4.6.
An explanation of the display screen or the item currently selected by the cursor appears. Press the
key to bring up a Help window
HELP
containing explanatory information. Pressing any key exits help screen.
On-line Help
A brief explanation of the item currently selected by the flashing cursor is displayed by pressing the
HELP
key.
Press any key to cancel the help screen.
Press anyke
Press the
HELP
ke
────────────────────────────────────────────────────
3.2 Other Keys Operation
Page 59
35
g
────────────────────────────────────────────────────
3.3 Screen Configuration
DISPLAY screen Channel setting screen
s34-1 s34-2
The input channel can be set. Press the
CH.SET
key to change
the input channel
s34-3 s34-4
DISP
settin
screen.
CH.SE
s34-5 s34-6 s34-7
SYSTE
STATU TRIG CHAN
Level monitor screen
The input level can be viewed and set.
LEVEL.
MONI
View screen
The waveform display position and block occupation status can be viewed.
VIEW
s34-8
SYSTEM screen
Each settings
STATUS screen
SYSTE STATU
Status
Scaling
Comment
Interface
Initialize
Self check
Memory segment
Parameter calculation
Waveform calculation
In Memory recorder function, these selection can be made by using the function keys on page 1/4.
TRIGGER screen CHANNEL screen
TRIG
Channels 1 - 16
Channels A - D
or
CH.SET
CHAN
Channels 1 - 8
Channels 9 - 16
Channels A - D
CH.SET
Channels 1 - 16
Variable function
────────────────────────────────────────────────────
3.3 Screen Configuration
Page 60
36
────────────────────────────────────────────────────
────────────────────────────────────────────────────
3.3 Screen Configuration
Page 61
37
────────────────────────────────────────────────────
Chapter 4
1
1
2
2
Memory Recorder Function
4.1 Overview of the Memory Recorder Function
The memory recorder function has the following features.
(1) After being stored in the internal memory, input signal data can be displayed
and printed.
(2) All input channel data are recorded on the same time axis.
Since data for all channels can be superimposed, the relative relationship between input signals can be observed visually.
(3) Time axis setting
100 μs/DIV to 5 min/DIV
(4) Time axis resolution 100 points/DIV
(5) Storage capacity
3
3
4
4
5
5
6
6
7
7
8
8
9
9
Number of channels 16 8 4 2
8 M words (DIV)
32 M words (DIV)
(6) Waveform magnification/compression display and print
Time axis direction ×10 to ×1/10000 (16 steps)Voltage axis direction ×10 to×1/2 (single),×5to×1/4 (2 to 8, X-Y single,
dual screens) 5 steps
With the variable function, vernier function, zoom function
(7) Display format
Time axis waveform: single, dual, quad, oct screen display (LCD), hex
screen display (when printing)
X-Y waveform: single, dual screen display (dot, line)
(8) Printing
Auto print, Manual print, Partial print, Report print, Screen hard copy.Multiple printing possible.
(9) High-quality print
Smooth print function approximates analog waveform.
5000
20000
10000 40000
20000 80000
40000
160000
10
10
11
11
12
12
13
13
14
14
A
A
────────────────────────────────────────────────────
4.1 Overview of the Memory Recorder Function
Page 62
38
────────────────────────────────────────────────────
(10) Logging function
Numeric printout of waveform data
(11) Memory segmentation function
Helps to reduce dead time of continuous recording. (Sequential save
function)
Memory is divided into blocks which can freely selected by the user for
storing measurement data. (Multi-block function)
(12) Processing function
Waveform processing (arithmetic processing, differential processing etc.)Waveform parameter processing (frequency measurement, rms measurement
etc.)
(13) Averaging
This makes it possible to eliminate noise and irregular signal components.
(14) Waveform evaluation function detects abnormal waveforms.
────────────────────────────────────────────────────
4.1 Overview of the Memory Recorder Function
Page 63
39
────────────────────────────────────────────────────
4.2 Operation Sequence (MEM)
1
Preparation
See Chapter 2
Basic Settings
Settings on the STATUS1 screen
See
Section 4.3
Press the the STATUS screen.
STATUS
key to display
Installation of the Unit
Turn on the power
Connect the inputs
Function
Time axis range
Recording length
Display format
Interpolation
Print mode (Waveform/ Numeric print)
Roll mode
Select the memory recorder function.
Determine the speed with which the input signal waveform is captured.
Determine the length of the data sample for a single measurement.
Select the format type for the waveform display screen.
Interpolation can be set when set the X-Y graph format.
Select the format, waveform, or numerical value should be used to output measured data and calculation results.
Select whether or not to display the waveform at the same time as it is captured.
2
3
4
5
6
7
8
s-42-1
Advanced Settings
(Use this function, if necessary.)
Settings on the STATUS2 screen
Auto print function
Auto save function
Overlay function
Averaging count
Waveform evaluation
Memory segmentation
See Chapter 15
Set the waveform calculation
See Chapter 16
Select whether or not to print out the waveform automatically after it is captured.
Select whether or not to save the waveform automatically after it is captured.
Select whether or not to overlay.
Set the averaging type and count.
Set the waveform evaluation mode
Set the GO/NG stop mode
Sequential save function
Multi-block function
Parameter calculation
Waveform processing calculation
9
10
11
12
13
14
A
────────────────────────────────────────────────────
4.2 Operation Sequence (MEM)
Page 64
40
────────────────────────────────────────────────────
Setting the input channel
Settings on the channel screen
Settings on the variable screen
Settings the advanced function
Setting the trigger function
(See Chapter 10)
(See Chapter 9)
Waveform display color Waveform display graph Voltage axis range Input coupling Logic input Magnification/compression along voltage Zero position Zero adjustment Offset cancel Low-pass filter
Variable function (See Chapter 9.7)
Scalingfunction(SeeChapter9.8) Comment function (See Chapter 9.9) Vernier function (See Chapter 9.10.2)
Trigger mode Trigger source Pre-trigger Trigger selection
Start measurement
Press theWhen the trigger conditions are met, measurement start.Press the
Processing measurement data
Measurements on display screen
Stop measurement
Setting the system
Press the SYSTEM screen. (See Chapter 12)
SYSTEM
key to display the
START
STOP
key and the LED lights.
key and the LED goes out after measurement has finished.
Printing the measurement data (See Chapter 13) Saving the measurement data (See Chapter 14)
Waveform scrolling (See Section 11.1)
Using the AB cursor (See Section 11.2)
Magnification/compression ratio along time axis
(See Section 11.3)
Zoom function (See Section 11.4)
Set up (See Section 12.2) Scaling (See Section 12.3) Comment (See Section 12.4) Interface (See Section 12.5) Initialize (See Section 12.6) Self-check (See Section 12.7)
────────────────────────────────────────────────────
4.2 Operation Sequence (MEM)
Page 65
41
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
1
s43-1
Press the how to set the STATUS screen of the memory recorder function. See the corresponding sections for items that can be set in the Waveform display or CHANNEL screens. Waveform display screen: See Section 4.5 CHANNEL screen: See Section 9.3
STATUS
key to access the STATUS1 screen. This section explains
See Sections
4.3.1 
4.3.2
4.3.3
4.3.4
4.3.5
4.3.6
4.3.7
4.3.8
4.3.9
2
3
4
5
6
7
8
NOTE
4.3.10
4.3.11
4.3.12
The settings in Section 4.3.5 are displayed only when the display format has been set to "X-Y."
9
10
11
12
13
14
A
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 66
42
────────────────────────────────────────────────────
4.3.1 Setting the Function Mode
The 8841/42 has five function modes. Select the Memory recorder function.
Procedure
Screen: STATUS1, CHANNEL, TRIGGER, Waveform display
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the top position.
3. Select the
Function
display
MEMORY
Meaning
:
Memory recorder function
:
Recorder function
:
RMS recorder function
:
Recorder & Memory recorder function
:
FFT function
function key display.
4.3.2 Setting the Time Axis Range
Set the speed for inputting and storing the waveform of the input signal. Time axis range setting expresses the time for 1 division. The sampling period is 1/100th of the set value for the time axis range. (100 samples/DIV)
Procedure1
Screen: STATUS1, Waveform display
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
3. Use the Jog/Shuttle control, the function keys or the the selection. When "EXT." is selected, the external sampling can be used. For details, see Section 17.1.4.
Data points per division are set when external sampling is selected.
1. Move the flashing cursor to the the left.
2. Use the
JOG/SHUTTLE
Setting range is 10 to 1000.
Procedure 2
Using the TIME/DIV key
1. Use the Menu keys to display the desired screen.
2. Use the The
TIME/DIV
TIME/DIV
located.
NOTE
The symbol "*" in the selection window indicates the time axis of the data stored to current memory. (If no data is present in memory, this symbol is not displayed.)
time/DIV
samples/DIV
item.
TIME/DIV
key to make
item, as shown in the figure on
control or the function keys to make the selection.
key to make the selection.
key can be used regardless of where the flashing cursor is
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 67
43
────────────────────────────────────────────────────
4.3.3 Setting the Recording Length
The length of recording for one measurement operation (number of DIV) can be set.
SELECT Select the recording length. ARBITRARY Variable recording length can be selected by the user.
1
2
Screen: STATUS1, Waveform display
Procedure 1 Constant recording length mode
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
3. Use the Jog/Shuttle control or the function keys to make a setting.
Function
display
Meaning
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
item and select
shot
Procedure 2 Variable recording length mode
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
3. Use the Jog/Shuttle control or the function keys to make a setting. To change the column, use the cursor keys ( ).
Function
display
Meaning
item and select
shot
SELECT
ARBITRARY
.
3
4
5
6
7
.
8
9
NOTE
:
Increases in number.
:
Decreases in number.
Depending on the number of channels, the recording length is limited.
(For setting the number of channels, see Section 12.2.1.)
If the recording length is changed during measurement, measurement is
restarted using the newly set recording length.
In fixed-length recording mode, the symbol "*" in the selection window
indicates the recording length of the data stored to current memory. (If no data is present in memory, this symbol is not displayed.) In any recording­length mode, the symbol "#" is displayed before the recording length.
The relation between number of channels and maximum recording length
Memory capacity
(words)
8M
32 M
16 channels 8 channels 4 channels 2 channels
5000
20000
Maximum recording length (DIV)
10000 40000
20000 80000
40000
160000
10
11
12
13
14
A
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 68
44
────────────────────────────────────────────────────
4.3.4 Setting the Display Layout
The layout can be set for showing input signals on the screen display and
recording them on the printer.
The following layout is available: single, dual, quad, oct (Waveform display
screen), hex (print only, oct graph on Waveform display screen) X-Y single, X-Y dual (dot/line).
The voltage per division is automatically changed according to the display
format.
Procedure
Screen: STATUS1
1. Press the
STATUS
2. Move the flashing cursor to the
key to display the STATUS1 screen.
format
item.
3. Use the function keys to select the display format.
4. Set the graph type when the display format is set to Dual, Quad, Oct or Hex screen display. For the setting of the X-Y screen, refer to Section 4.4.
5. Press the
CHAN
key to display the CHANNEL screen.
6. Move the flashing cursor to the point in the figure below. The figure shows the setting for the channel 1 (CH1). Setting for the channel 2 to 16 should be made in the same way.
Flashing cursor
7. Use the function keys to select.
Function
display
Meaning
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
In the cases of HEX, as far as the printer recording output is concerned, the waveforms for each channel are automatically distributed on each graph according to the table below:
Graph
1 2 3 4 5 6 7 8
Analog
channel
CH 1
CH 2
CH 3
CH14
CH15
CH16
CH17
CH18
Graph
9 10 11 12 13 14 15 16
Analog
channel
CH19
CH 10
CH 11
CH 12
CH 13
CH 14
CH 15
CH 16
For details on logic channels, refer to Section 9.3.8.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 69
45
────────────────────────────────────────────────────
Single graph
Display and record as one graph. (At the most, 16 analog + 16 logic signals)
s434-2, 3, 4, 5, 6, 7
Dual graph
Display and record as two
graphs. (At the most, 16 analog +16logicsignals) Specify which input channel to
use for waveform graph display and recording.
Quad graph
Display and record as four
graphs. (At the most, 16 analog + 8 logic signals)
Specify which input channel to
use for waveform graph display and recording.
Oct graph
Record as eight graphs. (At the
most, 16 analog + 4 logic signals)
Specify which input channel to
use for waveform graph display and recording.
X-Y single graph (dot)
X-Y dual graph (line)
HEX graph
Display is same as Oct screen display.
Record as sixteen graphs. (At the most, 1 analog + 2 logic signals)
Channel positions are automatically distributed.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 70
46
────────────────────────────────────────────────────
4.3.5 Setting the Interpolation (dot-line)
Interpolation can be set when set the X-Y graph format.This setting determines whether the input waveform (sampling data) is to be
displayed and printed as a series of dots or a line using linear interpolation.
Procedure
Screen: STATUS1
1. Press the
STATUS
key to display the STATUS1 screen.
2. Move the flashing cursor to the dot-line item.
3. Use the function keys to make a setting.
Function
display
Meaning
:
Linear interpolation is not performed.
:
Linear interpolation is performed.
4.3.6 Setting the Print Mode
Select the format, waveform, or numerical value should be used to output measured data and calculation results.
Waveform
Numerical value The data spacing interval also must be set.
The smooth print function can be used, but print speed will
decrease.
Procedure
(1) Setting the printer format
(2) Setting the smooth printing and print interval
Screen: STATUS1
1. Press the
STATUS
2. Move the flashing cursor to the
key to display the STATUS1 screen.
print mode
item.
3. Use the function keys to make a setting.
Function
display
Meaning
Measurement data and the result of calculation
:
are printed as a waveform.
Measurement data and the result of calculation
:
are printed as numeric data.
1. When the waveform format is selected, determine whether to use the smooth printing or not. When the numerical value is selected, set the print interval. (unit: divisions)
2. Move the flashing cursor to
smooth printorinterval
.
3. Use the function keys to make a setting.
Since 1 division represents 100 samples, the print interval "0.01" refers to a printout of every sample (no print interval). If the set print interval exceeds recording length, only the first point is printed.
NOTE
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
On X-Y screen, smooth printing cannot be specified.
Page 71
47
────────────────────────────────────────────────────
4.3.7 Setting the Roll Mode
This mode can be used at a time axis range setting of 10 ms/DIV or slower.In normal recording, the waveform is displayed only after all data of the
recording length have been captured. At low sampling speed settings, this will cause a considerable delay between the start of measurement and the appearance of the waveform on the display.
When roll mode is set to ON, the waveform is displayed immediately at the
start of recording (the screen scrolls).
When the time axis range of 10 ms/DIV or faster is set, normal recording is
carried out even if roll mode is set to ON.
Procedure
Screen: STATUS1
1. Press the
STATUS
key to display the STATUS1 screen.
2. Move the flashing cursor to the
roll mode
3. Use the function keys to make a setting.
Function
NOTE
display
Roll mode cannot be set together with the external sampling.
Meaning
:
Normal recording is carried out.
The waveform is displayed immediately at the
:
start of recording.
When Roll Mode is set to ON, the settings for Overlay, Averaging,
Sequential Saving, Calculating Waveform data, and Waveform evaluation are automatically turned OFF. (When the display format is set to "X-Y," you can turn ON Overlay with Roll Mode ON.)
4.3.8 Setting the Auto Print Function
When the function is enabled, printout is carried out automatically after a measurement data is captured.
item.
Procedure
Screen: STATUS1
1. Press the
STATUS
key to display the STATUS1 screen.
2. Move the flashing cursor to the
auto print
item.
3. Use the function keys to make a setting.
Function
NOTE
display
When cursor A and B are enabled, partial printing is executed.
Meaning
:
Auto print is disabled.
:
Automatically printed to the internal printer.
A LAN is used to transfer printing images to the
:
9333 LAN COMMUNICATOR.
When the roll mode is enabled and the time-axis range is lower than 10
ms/DIV, data is displayed and printed simultaneously.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 72
48
────────────────────────────────────────────────────
4.3.9 Setting the Auto Save Function
When the function is enabled, measurement data are automatically stored on a floppy disk, PC card, MO disk or connected SCSI device after they are captured. The Auto Save function stores a file in the directory currently selected on the file screen. See Section 14.12.2.
Procedure
Screen: STATUS1
1. Press the
STATUS
2. Move the flashing cursor to the
key to display the STATUS1 screen.
auto save
item.
3. Use the function keys, select the media for auto saving.
Function
display
Meaning
:
Auto save is disabled.
:
Automatically stored on floppy disk
:
Automatically stored on PC card.
:
Automatically stored on connected SCSI device.
:
Automatically stored on MO disk.
(When using a LAN card)
:
A LAN is used to transfer data to the 9333 LAN COMMUNICATOR.
4. When Media is selected, the Storage method item appears.
Function
display
Meaning
:
When the media becomes full, automatic storage stops.
When the media becomes full, old files are deleted to make room for automatic storage. With the binary format selected, the file with the extension 'MEM' is
:
deleted, and if sequential save is selected, files with the extension 'SEQ' are also deleted. With the text format selected, files with the extension 'TXT' are deleted.
5. Move the flashing cursor to
Function
display
Meaning
:
Data are stored as binary data.
:
Dataarestoredastextdata.
, and select the format.
type
Data stored in the text format is not readable by the 8841/42.
6. When the data format to be saved is set to text data, the intermittent setting item is displayed. Use the function keys or Jog/Shuttle control to make a setting.
Function
display
Meaning
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 73
49
────────────────────────────────────────────────────
7. Set the file name. For the input method, refer to Section 9.9.3. When using auto-save, a number is appended to the name you specify as the file name. This then becomes the file name. If you start procedures before specifying a file name, AUTO is automatically enabled.
File names cannot contain the characters +, -, ], or [. Please note that file names containing these characters will not be saved.
NOTE
File names consist of 8 characters (or 4 double-byte characters). With auto-
save, since numbers are attached to the end of file names, long file names are truncated when this number is appended to the end of the file name.
When auto-save is enabled, the storage channel cannot be selected. Data for
the channel for which the waveform is displayed is saved.
When both auto-print and auto-save are enabled, auto-save usually takes
precedence. However, if roll mode is enabled, auto-print will execute first.
For details on connected SCSI device, refer to Section 14.4.When cursor A and B are enabled, partial saving is executed.During automatic storage, if the
STOP
key is pressed twice to interrupt measurement, waveforms taken prior to the interruption are stored automatically.
The directory and the number of files that can be stored in the directory are
limited. For details, see Section 14.7.
When "COMMUNI" (communications) is selected as Media, Storage Method
and Type are not displayed.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 74
50
────────────────────────────────────────────────────
4.3.10 Setting the Overlay Function
Overlay is performed without clearing the currently displayed waveform (if
trigger mode is REPEAT or AUTO). This allows comparison to the immediately preceding waveform.
If trigger mode is SINGLE, measurement terminates after one set of data has
been collected. Therefore the overlay setting is invalid.
Procedure
s4310-1, 2
Screen: STATUS1
1. Press the
STATUS
2. Move the flashing cursor to the
key to display the STATUS1 screen.
overlay
3. Use the function keys to make a setting.
Function
display
Meaning
:
Overlay is not performed.
:
Overlay is performed.
item.
NOTE
Normal Display
While overlay is being executed, operations on the display screen (waveform
Overlaid Display
scrolling, change in time-axis magnification/compression ratio, change in zero position, and jumping to another display screen using VIEW function (see Section 11.6)) are disabled.
When manual printing or trace cursor reading carried out, only the last
waveform will be done.
Use Report Print to print overlaid waveforms. (See Section 13.6.7, "Report
Print.")
If one of the following settings is changed, the overlay waveform display
terminates and only the last waveform is shown:
 ① STATUS1: Display format  ② CHANNEL: Input settings
The overlay function cannot be set together with roll mode.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 75
51
────────────────────────────────────────────────────
4.3.11 Setting the Averaging Function
The averaging function allows capturing several instances of a waveform and
determining the average.
This makes it possible to eliminate noise and irregular signal components.The higher the number of averaging instances, the more effectively will noise
be suppressed.
Procedure
s4311-1, 2
Screen: STATUS1
1. Press the
STATUS
2. Move the flashing cursor to the
key to display the STATUS1 screen.
averaging
item.
3. Use the function keys or Jog/Shuttle control to make a setting.
Function
display
Meaning
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
After starting the measurement, the averaging count and the current waveform data count are shown on the screen.
After Averaging
NOTE
Before Averaging
When the averaging function is used, logic waveform is not displayed.When the memory segmentation function is used, averaging is not available.Averaging and waveform processing cannot be carried out simultaneously.The averaged waveform becomes available for waveform processing when
the averaging setting is turned OFF following measurement.
When the averaging function is used, the maximum recording length is
reduced to 25% of the normal value.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 76
52
────────────────────────────────────────────────────
Averaging and trigger mode
(1) Trigger mode: SINGLE
1. After the
START
key was pressed, data are captured whenever the trigger
conditions are fulfilled, and summing averaging is carried out.
2. When the specified number of data has been captured, measurement stops automatically.
3. If the measurement was stopped prematurely with the
STOP
key, the
averaging result up to that point is displayed.
Waveform averaging count = specified number
Input
Input
Trigger conditions
fulfilled
Waveform capture
Averaging
(summing)
Display End
(for recording length)
Waveform averaging count = less than specified number
(2) Trigger mode: REPEAT
1. After the
START
key was pressed, data are captured whenever the trigger conditions are fulfilled, and summing averaging is carried out until the specified averaging count. The averaging result is shown on the display.
2. After the specified averaging count was reached, exponential averaging is carried out whenever data are captured, and the averaging result is shown on the display.
3. If the measurement was stopped prematurely with the
STOP
key, the
averaging result up to that point is displayed.
The STOP key is pressed
Trigger conditions
fulfilled
Waveform capture
(for recording length)
After specified count
Averaging
(summing)
Averaging
(exponential)
Display End
(3) Trigger mode: AUTO
When the
START
key is pressed, data are captured even if trigger conditions are not fulfilled after a certain interval. If averaging is applied to unsynchronized input signals, the result will be meaningless.
NOTE
For details on summing averaging and exponential averaging, refer to Appendix 3.6.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 77
53
────────────────────────────────────────────────────
4.3.12 Setting the Waveform Evaluation
Display format can be set on single screen or X-Y single screen.GO (pass) or NG (fail) evaluation of the input signal waveform can be
performed using an evaluation area specified by the user. This can serve to detect irregular waveforms. After the evaluation result is generated, signals are output from the GO/NG terminal. All the channels being displayed are evaluated.
NOTE
When a waveform evaluation area is created by the memory recorder
function, the waveform evaluation area created by the FFT function is cleared.
For details on the waveform evaluation, refer to Section 16.3.
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
Page 78
54
────────────────────────────────────────────────────
4.4 Using the X-Y Waveform Plots (MEM)
Setting the display format to X-Y in status screen allows X-Y waveforms to
be combined. Any of channels 1 to 16 can be selected for each of the X and Y axis. Up to four X-Y plots can be made simultaneously.
Voltage axis magnification/compression is active also when using X-Y
combined plotting.
Using the AB cursors, it is possible to specify the data between the cursors
for partial plotting.
#s434- 6, 7copy# s44-1, 2
X-Y Single
A single plot is displayed and recorded. X-Y waveforms of graphs 1 to 4 are displayed in one screen.
X-Y Dual
Two single plot is displayed and recorded. X-Y waveforms of graphs 1 and 3 are displayed in the upper section, and those of graphs 2 and 4 are displayed in the lower section.
X-Y(dot)
The sampled data is displayed and
recorded just as it comes.
Linear interpolation is not performed.
X-Y(line)
Linear interpolation is performed.
The display becomes easier to read, but
display speed is slower compared to dot display.
────────────────────────────────────────────────────
4.4 Using the X-Y Waveform Plots (MEM)
Page 79
55
────────────────────────────────────────────────────
Procedure
s44-3
Screen: STATUS1
1. Press the
STATUS
key to display the STATUS1 screen.
2. Move the flashing cursor to the format item.
3. Use the function keys to select the
X-Y single or X-Y dual format.
4. Set the interpolation. (See Section 4.3.5)
5. Press the
6. Press the
CHAN
CH.SET
key to display the CHANNEL screen.
key to display page 3/3.
7. Move the flashing cursor to desired channel, and use the function keys to set the waveform color (including displays waveform ON/OFF).
Function
display
Meaning
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
:
Waveform is performed.
Procedure
:
Waveform is not performed.
8. Specify the X-axis channel. Move the cursor to the channel to be used as X axis. And use the function keys or the Jog/Shuttle control to select X axis.
9. Specify the Y-axis channel. This is done in the same way as the X-axis setting.
10. For graph 2 to graph 4, the settings are made in an identical.
Partial X-Y plot
Using the AB cursors, it is possible to specify a range for partial X-Y
plotting. (Normal X-Y plotting covers all data of the recording length.)
Screen: Waveform display (excluding X-Ysingle and X-Ydual)
1. Display the captured waveform data, using a format other than X-Ysingle and X-Ydual.
2. Use the AB cursors to specify the desired portion for plotting (see Section
11.2).
3. Press the STATUS
key to display the STATUS1 screen.
4. Carry out combined plotting as described above.
In partial X-Y plot, the screen displays the period of time passed after the completion of the trigger specified using the cursors A and B.
────────────────────────────────────────────────────
4.4 Using the X-Y Waveform Plots (MEM)
Page 80
56
────────────────────────────────────────────────────
Switching the X-Y1 Screen Display (Wide Display)
Setting the display format to X-Y1 in the STATUS1 screen allows you to switch between wide screen display and normal screen display each time
DISP
you press the
key.
Normal Display Screen Wide Display Screen
────────────────────────────────────────────────────
4.4 Using the X-Y Waveform Plots (MEM)
Page 81
57
────────────────────────────────────────────────────
4.5 Settings on the Waveform Display Screen (MEM)
Explains the setting items on the Waveform display screen. For details on setting, refer to Section 4.3. When want to use the Jog/Shuttle control, press the selection window is not displayed.)
Setting items Selection Explanation
1.
1. Function
MEM, REC, RMS, REC&MEM, FFT
VALUE select key.
Select function.
The
2.
2. Trigger mode
3.
3. Analog trigger
4.
4. Pre-trigger
5.
Time Axis Range
5.
6.
7.
8.
6. Magnification/
compression along
9.
the time axis
7. Zoom function
8. Recording Length
SINGLE, REPEAT, AUTO
OFF, LEVEL, OUT, IN, V-DROP, CYCLE
0 to 100%, -95%
100μs/DIV to 5
min/DIV, EXT
×10 to ×1/10000
Changes in the magnification in the
range of ×1/5000 to ×
10
SELECT: 25 DIV to 5000 DIV ARBITRARY: 1 DIV to 5000 DIV
Select trigger mode.
Set the analog trigger.
Set the Pre-trigger.
Set the speed for inputting and storing the waveform of the input signal. Time axis range setting expresses the time for 1 DIV.
By magnifying the waveform, detailed observations can be made. By compressing the waveform, an entire change can be promptly apprehended.
Splits the Waveform display screen into two halves, upper and lower screens, expanding the waveform in the upper screen along the time axis and displaying the expanded waveform in the lower screen.
Using channels: 16 ch Capacity: 8 M words The length of recording for one measurement operation (the number of DIV) can be set.
9. Cursor Measurement
Input channel settings
OFF, , ,
Analog input Logic input X, Y axis (X-Y format)
The A・B cursors can be used.
Enables the measurement conditions for each channel on the display screen to be set or changed. See Section 9.10.
Input level monitor function
VIEW function
Press the LEVEL MONI key. See Section 11.5.
Press the VIEW key. See Section 11.6.
Channels that may be changed with the RANGE and
POSITION
knob (zero position).This channel display is selected with
knob (measurement range)
channel-select keys CH1 to CH16. Changing the set channel in the CHANNEL or CH.SET screens modifies the setting accordingly.
────────────────────────────────────────────────────
4.5 Settings on the Waveform Display Screen (MEM)
Page 82
58
────────────────────────────────────────────────────
4.6 Auto Range Function
This function automatically selects the time axis range, measurement range and zero position. Taking the lowest numbered channel among the channels for which waveform display is on, 1 to 2.5 cycles are automatically set to be recorded as 15 divisions.
Procedure
Screen: Waveform display
1. Press the
2. Press the
key to display the Waveform display screen.
DISP
AUTO
key.
3. Use the function keys to make a setting.
Function
display
Meaning
:
Execute the Auto Range function.
:
Cancel the Auto Range function.
If measurement has started using the auto-range function:
・ Conditions related to the input units (all channel)
Voltage axis range, zero position
Magnification/compression ratio along the voltage axis
Low-pass filter, Input coupling
Value set automatically
Single screen:×1 Other screens:×1/2
OFF, DC
・ Trigger conditions (one channel only)
Trigger mode
AND/OR for internal trigger and external trigger
Pre-trigger
Internal trigger
Trigger type : Level
・ Status conditions
Time axis range (time/div)
Memory segmentation
AUTO
OR
20%
Enables only the waveform display channel with the lowest number. (If the difference between maximum and minimum is 2 divisions or less, the next channel is enabled.)
Slope:
rising Trigger level: value set automatically Filter: OFF
Value set automatically Magnification/compression ratio:×1
OFF
────────────────────────────────────────────────────
4.6 Auto Range Function
Page 83
59
────────────────────────────────────────────────────
NOTE
Because the auto-range function performs automatic setting for the input
signal present at the time the function is executed, input a signal before executing the function.
If for the input signal for this channel there is only a small difference
between the maximum value and the minimum value in the range of highest sensitivity (5 mV/DIV), the setting is made by taking the next higher channel.
If the range cannot be determined, for all channels for which the waveform
display is on. A warning message appears, and measurement is abandoned.
When the auto range function is activated by pressing the
AUTO
key, a trigger output signal is generated. This should be taken into consideration when using both the trigger output and the auto range function.
The auto-range function does not operate on channels for which the
CHARGE or PREAMP measurement mode is selected on the 8947 CHARGE UNIT.
────────────────────────────────────────────────────
4.6 Auto Range Function
Page 84
60
────────────────────────────────────────────────────
4.7 Other Screen Settings
The status screen for the memory recorder function contains three more setup screens. For detailed setup procedures, see the associated sections.
STATUS2 screen: Memory Segmentation Screen (See Chapter 15) STATUS3 screen: Parameter Calculation Screen (See Section 16.1) STATUS4 screen: Calculating Waveform Data Screen (See Section 16.2)
s47-1, 2, 3, 4
STATUS 1 Screen
STATUS 3 Screen
STATUS 2 Screen
STATUS 4 Screen
NOTE
The STATUS screen can also be changed from the function display field "1/4" on the upper right of the STATUS screen.
────────────────────────────────────────────────────
4.7 Other Screen Settings
Page 85
61
────────────────────────────────────────────────────
4.8 Start and Stop Operation (MEM)
Setting the 8841/42
STATUS Settings Channel Settings Trigger Settings
Measurement start
Trigger condition
are met
Recording start
See Section 4.3, "STATUS Settings", Chapter 9, "Input Channel Settings", Chapter 10, "Trigger Functions."
Press the
START
key and LED light. When the pre-trigger is set, the trigger will not be registered for a certain period after the start of measurement. (During this interval, "Pre-trigger standby" is shown on the display.) When the trigger can be registered, the indication "Waiting for trigger" is shown on the display.
Data recording starts when trigger condition are met. When trigger mode selected AUTO, unit waits for about 1 second for trigger conditions to be met. After this interval, data recording starts, regardless of trigger state.
Displays "Storing". When roll mode is set to ON, the waveform is displayed (scrolling) immediately at the start of recording Using the
VIEW
key to display the time (from measurement started time). When the
STOP
key is pressed twice during measurement, the
8841/42 is forcibly stopped. Then the waveform data until the
STOP
key is pressed is displayed and saved. When the auto saving is set to ON, the data until a forced termination is automatically saved. (Auto printout is not executed.)
Recording end
Waveform displays after data corresponding to recording length have been stored in memory. Auto printout and auto save are executed.
Measurement end
Trigger mode SINGLE
REPEAT AUTO
End of measurement
Each time when trigger conditions are met, data are recorded and memory contents are overwritten.
End of measurement in REPEAT and AUTO trigger modes When the 8841/42 acquires measurement data in an amount corresponding to the set recording length, and the measurement is stopped. (Waveform display, auto printout, and auto save are executed.)
When the time axis range is set to the slow range (not more than 10 ms/DIV), the waveforms measured until the pressed are displayed.
When the time axis range is set to the fast range (greater than 10 ms/DIV), waveforms are not displayed. (Waveforms are displayed only if measured data having recording length equal to the previous setting exist.)
STOP
key is pressed once during measurement, the
STOP
key is
────────────────────────────────────────────────────
4.8 Start and Stop Operation (MEM)
Page 86
62
────────────────────────────────────────────────────
────────────────────────────────────────────────────
4.8 Start and Stop Operation (MEM)
Page 87
63
────────────────────────────────────────────────────
Chapter 5
1
2
Recorder Function
5.1 Overview of the Recorder Function
The recorder function has the following features.
(1) Real time display and printing of the input signal
(2) Real time continuous recording of the input signal
(3) All input channel data are recorded on the same time axis.
Since data for all channels can be superimposed, the relative relationship between input signals can be observed visually.
(4) Time axis setting
20 ms/DIV to 1 h/DIV
(5) Time axis resolution 100 points/DIV (printer)
(6) Sampling period
1, 10, 100μs, 1, 10, 100 msCan be selected, from 1/100 of the time axis setting
(7) Memory capacity
8 M words: 2000 DIV32 M words: 10000 DIVArbitrarily (Set from 1 division to the maximum number of divisions at 1-
division intervals)
3
4
5
6
7
8
9
10
11
(8) Waveform magnification/compression display and printout
Time axis direction: ×1to×1/500Voltage axis direction: ×10 to ×1/2 (single)
×5to×1/4 (dual to hex, X-Y single/dual) 5 steps
With the variable function, vernier function
(9) Display format
Time axis waveform: single, dual, quad, oct screen display (LCD), hex
screen displayprinter only
X-Y waveform: single, dual
────────────────────────────────────────────────────
5.1 Overview of the Recorder Function
12
13
14
A
Page 88
64
────────────────────────────────────────────────────
(10) Scrollable display
The most recent 2000 (32 M words: 10000) divisions of the data are stored
in memory.
It is possible to scroll back for easy review.
(11) Additional recording function
The first set of measurement data is preserved, and recording of the second
set of measurement data starts after the first set.
(12) Logging function
Numeric printout of waveform data.
(13) Reprint function
The most recent 2000 (32 M words: 10000) divisions of the data stored in memory can be printed as many times as required.
(14) Print
Real-time print, manual print, partial print, report print, screen hard copy can be printed.
(15) X-Y CONT Recorder
This function allows X-Y plot between channels in real time.
────────────────────────────────────────────────────
5.1 Overview of the Recorder Function
Page 89
65
────────────────────────────────────────────────────
5.2 Operation Sequence (REC)
1
Preparation
See Chapter 2
Basic Settings
Settings on the STATUS screen
See
Section 5.3
Press the the STATUS screen.
STATUS
key to display
Installation of the Unit
Turn on the power
Connect the inputs
Function
Time axis range
Sampling period
Recording length
Display format
Print mode
Additional recording
Print function
Select the recorder function.
Determine the speed with which the input signal waveform is captured.
Determine the length of the data sample for a single measurement.
Select the format type for the waveform display screen.
Select the waveform or numerical value to be output the measured data Set the print interval (numerical format)
Set whether additional recording is required.
Select whether or not to print out simultaneously with screen display.
2
3
4
5
6
7
8
Auto save function
Interpolation
Waveform clear
Select whether or not to save the waveform automatically
Select dot or line in X-Y display format.
Select whether or not to clear waveform in X-Y display format.
9
10
11
12
13
14
────────────────────────────────────────────────────
5.2 Operation Sequence (REC)
A
Page 90
66
────────────────────────────────────────────────────
Setting the input channel
Settings on the channel screen
Settings on the variable screen
Settings the advanced function
Setting the trigger function
(See Chapter 10)
(See Chapter 9)
Waveform display color Waveform display graph Voltage axis range Input coupling Logic input Magnification/compression along voltage Zero position Zero adjustment Offset cancel Low-pass filter
Variable function (See Chapter 9.7)
Scaling function (See Chapter 9.8) Comment function (See Chapter 9.9) Vernier function (See Chapter 9.10.2)
Trigger mode Trigger source Trigger timing Trigger selection
Start measurement
Press the
When the trigger conditions are met, measurement start.
Press the STOP
Processing measurement data
Measurements on display screen
Stop measurement
Setting the system
Press the SYSTEM screen. (See Chapter 12)
SYSTEM
key to display the
START
key and the LED lights.
key and the LED goes out after measurement has finished.
Printing the measurement data (See Chapter 13) Saving the measurement data (See Chapter 14)
Waveform scrolling (See Section 11.1) Using the A/B cursor (See Section 11.2) Magnification/compression ratio along time axis
(See Section 11.3)
Set up (See Section 12.2) Scaling (See Section 12.3) Comment (See Section 12.4) Interface (See Section 12.5) Initialize (See Section 12.6) Self-check (See Section 12.7)
────────────────────────────────────────────────────
5.2 Operation Sequence (REC)
Page 91
67
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
1
Press the how to set the STATUS screen of the recorder function. See the corresponding sections for items that can be set in the Waveform display or CHANNEL screens. Waveform display screen: see Section 5.5 CHANNEL screen: see Section 9.3
STATUS
key to access the STATUS screen. This section explains
See Sections
5.3.1
5.3.2
5.3.3
5.3.4
5.3.5
5.3.10 (when format:X-Y)
5.3.6
5.3.7
5.3.8
5.3.9
2
3
4
5
6
7
8
NOTE
5.3.11 (when format:X-Y)
The settings in Section 5.3.10 and 5.3.11 are displayed only when the display format has been set to "X-Y."
9
10
11
12
13
14
A
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 92
68
────────────────────────────────────────────────────
5.3.1 Setting the Function Mode
The 8841/42 has five function modes. Select the Recorder function.
Procedure
Screen: STATUS, CHANNEL, Waveform display, TRIGGER
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the top position.
3. Select the RECORDER function key display.
Function
display Meaning
:
Memory recorder function
:
Recorder function
:
RMS recorder function
:
Recorder & Memory recorder function
:
FFT function
5.3.2 Setting the Time Axis Range
Set the speed for inputting and storing the waveform of the input signal. Time axis range setting expresses the time for 1 division.
Screen: STATUS, Waveform display
Procedure 1 1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the time/DIV item.
3. Use the Jog/Shuttle control, the function keys or the TIME/DIV the selection.
Procedure 2
Using the TIME/DIV key
1. Use the Menu keys to display the desired screen.
2. Use the TIME/DIV key to make the selection. The TIME/DIV
key can be used regardless of where the flashing cursor is
located.
NOTE
While the printer always outputs the data at the measurement magnification in recording mode, the waveform on the screen is reduced in size at the ratio shown in the table below, depending on the time-axis range.
50 ms/DIV
→×
1/2, 20 ms/DIV
→×
1/10 The symbol "*" in the selection window indicates the time axis of the data stored to current memory. (If no data is present in memory, this symbol is not displayed.)
key to make
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 93
69
────────────────────────────────────────────────────
5.3.3 Setting the Sampling Period
The available range depends on the selected time axis range (input signal waveform capture rate).
1
2
Procedure
NOTE
Screen: STATUS
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the sampling item.
3. Use the Jog/Shuttle control or the function keys to make the selection.
Function
display
When a short sampling period is set and the input waveform changes slightly, a sudden disturbance such as noise will increase the difference the between the maximum and minimum values. To eliminate this phenomenon, set a long sampling period. For details, see Appendix 3.4.
Meaning
:
Increases in number.
:
Decreases in number.
5.3.4 Setting the Recording Length
The length of recording for one measurement operation (number of DIV) can be set.
SELECT ARBITRARY
Select the recording length.
Variable recording length can be selected by the user.
3
4
5
6
7
8
9
Screen: STATUS, Waveform display
Procedure 1 Constant recording length mode
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
3. Use the Jog/Shuttle control or the function keys to make a setting.
Function
display
Meaning
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
shot item and select SELECT.
10
11
12
13
14
A
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 94
70
────────────────────────────────────────────────────
Procedure 2 Variable recording length mode
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
shot
item and select
ARBITRARY
.
3. Use the Jog/Shuttle control or the function keys to make a setting. To change the column, use the cursor keys ( ).
NOTE
Function
display
When the recording length is set to CONT.
Meaning
:
Increases in number.
:
Decreases in number.
ms/DIV, the printer setting (real-time print) is automatically set to OFF.
When the recording length is set to to OFF.
The maximum recording length is 2000 DIV when the memory capacity is 8 M words, and 10000 DIV when the memory capacity is 32 M words (when memory extended)
In fixed-length recording mode, the symbol "*" in the selection window indicates the recording length of the data stored to current memory. (If no data is present in memory, this symbol is not displayed.) In any recording­length mode, the symbol "#" is displayed before the recording length.
5.3.5 Setting the Display Layout
in a range of 20 ms to 200
CONT., auto-saving is automatically set
Procedure
1. Press the STATUS key to display the STATUS screen.
2. Move the flashing cursor to the format item.
3. Use the function keys to select the display format.
4. Set the graph type when the display format is set to DUAL, QUAD, OCT or
5. Press the CHAN key to display the CHANNEL screen.
6. Move the flashing cursor to the point in the figure below.
The layout can be set for showing input signals on the screen display and recording them on the printer.
The following layout is available: single, dual, quad, oct (Waveform display screen), hex (Print only, Display oct style)X-Y single, X-Y dual (dot/line)
The voltage per division is automatically changed according to the display format.
Screen: STATUS
HEX screen display.
The figure shows the setting for the channel 1 (CH1). Setting for the channel 2 to 16 should be made in the same way.
Flashing cursor
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 95
71
────────────────────────────────────────────────────
7. Use the function keys to select.
Function
display
Meaning
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
##F2
In the cases of HEX, as far as the printer recording output is concerned, the waveforms for each channel are automatically distributed on each graph according to the table below: For details on logic channels, refer to Section 9.3.8.
Graph
1
2
3
4
5
6
7
8
Analog
channel
CH 1
CH 2
CH 3
CH14
CH15
CH16
CH17
CH18
Graph
10
11
12
13
14
15
16
Analog
channel
9
CH19
CH 10
CH 11
CH 12
CH 13
CH 14
CH 15
CH 16
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 96
72
────────────────────────────────────────────────────
Single graph
Display and record as one graph. (At the most, 16 analog + 16 logic signals)
s535-2, 3, 4, 5, 6, 7
Dual graph
Display and record as two graphs. (At the most, 16 analog +16logicsignals)
Specify which input channel to use for waveform graph display and recording.
Quad graph
Display and record as four
graphs. (At the most, 16 analog + 8 logic signals)
Specify which input channel to
use for waveform graph display and recording.
Oct graph
Record as eight graphs. (At the most, 16 analog + 4 logic signals)
Specify which input channel to use for waveform graph display and recording.
X-Y single graph (dot)
X-Y dual graph (line)
HEX graph
Display is same as Oct screen display.
Record as sixteen graphs. (At the most, 1 analog + 2 logic signals)
Channel positions are automatically distributed.
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 97
73
────────────────────────────────────────────────────
5.3.6 Setting the Print Mode
Select the format, waveform, or numerical value should be used to output measured data.
Waveform Numerical value
The waveform is printed.
The data spacing interval also must be set.
Procedure
(1) Setting the printer format
1. Press the STATUS key to display the STATUS screen.
2. Move the flashing cursor to the print mode item.
3. Use the function keys to make a setting.
(2) Setting the print interval
1. When the numerical value is selected, set the print interval. (unit: DIV)
2. Move the flashing cursor to the interval item.
3. Use the function keys to make a setting.
Screen: STATUS
Function
display
Meaning
Measurement data and the result of calculation
:
are printed as a waveform.
Measurement data and the result of calculation
:
are printed as numeric data.
Since 1 division represents 100 samples, the print interval "0.01" refers to a printout of every sample (no print interval). When the print interval longer than the recording length is set, only the first dot is printed.
NOTE
Function
display Meaning
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
The maximum and minimum values are printed (see Appendix 3.4).
The print intervals of 0.01 to 0.5 DIV can be selected, only when there are waveform data present.
When numerical values are output for real-time printing, the minimum print interval is 1 division.
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 98
74
e
────────────────────────────────────────────────────
5.3.7 Setting the Additional Recording Function
This records, regarding the memory as though it were recording paper. The memory capacity of 8 M words can accommodate up to 2000 divisions of waveform data. With 32 M words (after optional memory expansion), waveform data of up to 10000 divisions can be saved. The waveform can be scrolled and printed out. Switching the additional recording on and off affects the use of memory as shown below.
Additional recording: OFF
Additional recording: ON
Procedure
1. Press the STATUS key to display the STATUS screen.
2. Move the flashing cursor to the record add item.
3. Use the function keys to make a setting.
1. Recording 25 divisions of waveform
2000 DIV(No memoryexpansion
25 DIV
2. Recording another 25 divisions of waveform The first set of measurement data is discarded, and recording of the second set of measurement data starts again from the beginning of memory.
25 DIV
Screen: STATUS
1. Recording 25 divisions of waveform
2000 DIV(No memoryexpansion
25 DIV
2. Recording another 25 divisions of waveform The first set of measurement data is preserved, and recording of the second set of measurement data starts after th first set.
25 DIV
25 DIV
The first and second sets of waveforms can be observed by scrolling or printing the waveform.
Function
display
Meaning
:
Additional recording is disabled.
:
Additional recording is enabled.
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 99
75
────────────────────────────────────────────────────
NOTE
Time values output to the printer and displayed on the screen with the additional recording function enabled are equal to those of the most recently acquired waveforms. Therefore, when measuring waveforms in a different time axis range, always take that difference into consideration.
When the 2000 (32 M words: 10000) DIV has been reached, old data will be overwritten.
In the recorder, the trigger mark ( ) is written as the start position mark. In an additional recording, the trigger mark is displayed in front of the most recently entered data.
If auto-save is enabled, only newly acquired waveforms are saved. Even when A and B cursors appear on the screen, no partial save is performed, because the setting is disabled when the
START
key is pressed.
5.3.8 Setting the Printer Function (Real Time Printing)
The input waveform is continuously printed in real time.
Procedure
1. Press the STATUS key to display the STATUS screen.
2. Move the flashing cursor to the printer item.
3. Use the function keys to make a setting.
NOTE
Screen: STATUS
Function
display
At a time axis range setting of 200 ms/DIV or faster, the waveform data
Meaning
:
Printing is disabled.
:
Printing is enabled.
will be printed out later. See Section 18.5, "Recorder Function."
While the printer always outputs the data at the measurement magnification in recording mode, the waveform on the screen is reduced in size at the ratio shown in the table below, depending on the time-axis range. 50 ms/DIV:×1/2, 20 ms/DIV:×1/10
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Page 100
76
────────────────────────────────────────────────────
5.3.9 Setting the Auto Save Function
When the function is enabled, measurement data are automatically stored on a floppy disk, PC card, MO disk or connected SCSI device after they are captured.
The Auto Save function stores a file in the directory currently selected on the file screen. See Section 14.12.2.
Procedure
Screen: STATUS
1. Press the STATUS key to display the STATUS scr een.
2. Move the flashing cursor to the auto save item.
3. Use the function keys, select the media for auto saving.
Function
display Meaning
:
Auto save is disabled.
:
Automatically stored on floppy disk
:
Automatically stored on PC card.
:
Automatically stored on connected SCSI device.
:
Automatically stored on MO disk.
(When using a LAN card)
:
A LAN is used to transfer data to the 9333 LAN COMMUNICATOR.
4. When Media is selected, the Storage Method item appears.
Function
display Meaning
:
When the media becomes full, automatic storage stops.
When the media becomes full, old files are deleted to make room for automatic storage.
:
With the binary format selected, the file with the extension 'REC' is deleted. With the text format selected, files with the extension 'TXT' are deleted.
5. Move the flashing cursor to type
Function
display Meaning
:
Data are stored as binary data.
:
Dataarestoredastextdata.
, and select the format.
Data stored in the text format is not readable by the 8841/42.
6. When the data format to be saved is set to text data, the intermittent setting item is displayed. Use the function keys or Jog/Shuttle control to make a setting.
7. Set the file name. For the input method, refer to Section 9.9.3. When using auto-save, a number is appended to the name you specify as the file name. This then becomes the file name. If you start procedures before specifying a file name, AUTO is automatically enabled.
File names cannot contain the characters +, -, ], or [. Please note that file names containing these characters will not be saved.
────────────────────────────────────────────────────
5.3 STATUS Settings (REC)
Loading...