Thank you for purchasing the HIOKI "8807-01, 8808-01,
8807-51, 8808-51 MEMORY HiCORDER." To obtain
maximum performance from the product, please read this
manual first, and keep it handy for future reference.
Inspection
When you receive the product, inspect it carefully to ensure
that no damage occurred during sh ipping. In particular,
check the panel switches, and connectors. If damage is
evident, or if it fails to operate according to the
specifications, contact your dealer or Hioki representative.
The connector cable is not included with the product.
Use optional 9197, 9198, or 9217 CONNECTION CORD.
Maximum input voltage:
9197: 500 V AC/DC
9198: 300 V AC/DC
9217: 300 V AC/DC
9626 PC CARD 32M
9627 PC CARD 64M
9726 PC CARD 128M
9727 PC CARD 256M
9728 PC CARD 512M
9729 PC CARD 1G
*: no CE marking
NOTE
The 9270 to 9272 CLAMP ON SENSORs and 9277 to 9279
UNIVERSAL CLAMP ON CTs are designed for use only with
the 9555 SENSOR UNIT, and cannot be used alone.
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling during
use could result in injury or death, as well as damage
to the product. Be certain that you understand the
instructions and precautions in the manual before use.
We disclaim any responsibility for accidents or injuries
not resulting directly from product defects.
This manual contains information and warnings essential
for safe operation of the product and for maintaining it in
safe operating condition. Before using the product, be sure
to carefully read the following safety notes.
Thesymbol printed on the product indicates that the
user should refer to a corresponding topic in the
manual (marked with the
relevant function.
In the manual, thesymbol indicates particularly
important information that the user should read before
using the product.
symbol) before using the
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Indicates both DC (Direct Current) and AC (Alternating
This instrument complies with CAT II safety requirements.
To ensure safe operation of measurement instruments, IEC
61010 establishes safety standards for various electrical
environments, categorized as CAT I to CAT IV, and called
measurement categories. These are defined as follows.
CAT I
CAT II
CAT III
CAT IV
Secondary electrical circuits connected to an
AC electrical outlet through a transformer or
similar device.
Primary electrical circuits in equipment
connected to an AC electrical outlet by a power
cord (portable tools, household appliances, etc.)
Primary electrical circuits of heavy equipment
(fixed installations) connected directly to the
distribution panel, and feeders from the
distribution panel to outlets.
The circuit from the service drop to the service
entrance, and to the power meter and primary
overcurrent protection device (distribution
panel).
Higher-numbered categories correspond to electrical
environments with greater momentary energy. So a
measurement device designed for CAT III environments can
endure greater momentary energy tha n a device designed for
CAT II.
Using a measurement instrument in an environment
designated with a higher-numbered category than that for
which the instrument is rated could result in a severe
accident, and must be carefully avoided.
Never use a CAT I measuring instrument in CAT II, III, or
IV environments.
The measurement categories comply with the Overvoltage
Categories of the IEC60664 Standards.
We define measurement tolerances in terms of f.s. (full
scale), rdg. (reading) and dgt. (digit) values, with the
following meanings:
f.s. (maximum display value or scale length)
The maximum displayable value or the full length of the
scale.
This is usually the maximum value of the currently selected
range.
rdg. (reading or displayed value)
The value currently being measured and indicated on the
measuring product.
dgt. (resolution)
The smallest displayable unit on a digital measuring
product, i.e., the input value that causes the digital display
to show a "1".
Follow these precautions to ensure safe operation and to
obtain the full benefits of the various functions.
To avoid the risk of electric shock and damage to the
DANGE
product
Do not apply more than 450 V AC or DC between the
analog inputs and product, or between the analog
inputs.
Use only the specified Model 9418-15 AC ADAPTER
(SA145A-1240V-6, SINO AMERICAN). AC adapter input
voltage range is 100 to 240 V AC (with
10% stability)
at 50/60 Hz. To avoid electrical hazards and damage to
the product, do not apply voltage outside of this range.
Terminal inputs should never exceed the limits
indicated below.
Input/output terminalsMaximum input voltage
Analog input450 V AC/DC
EXT TRIG-5 V to +10 V
TRIG OUT
-20Vto+30V
500 mA max, 200 mW max
The maximum voltage applied to the probe and cables
connecting to the analog input of the product must
never exceed the limits in the following table.
Notwithstanding these limits, the voltage at the analog
input must not exceed 450V AC or DC.
Maximum input voltage
9197 CONNECTION CORD
9217 CONNECTION CORD
9322DIFFERENTIAL
When using the batteries, use LR6 alkaline batteries
ix
ARNING
or 9447 BATTERY PACK. In this case, do not use
new batteries with old ones. Also, do not mix
batteries of different types. We take no responsibility
for an accident using the batteries and battery packs
which are not specified.
The logic inputs are not floating. Although two sets
of logic probes can be connected, they all have a
common ground with the main product.
CAUTION
Use only the specified input cord. Using a non-specified
cable may result in incorrect measurements due to poor
connection or other reasons.
To avoid damage to the product, do not input a
voltage/current exceeding the rated maximum to the
external input terminals.
This product should be installed and operated indoors only,
between 5 and 40
Do not store or use the product where it could be exposed
to direct sunlight, high temperature or humidity, or
condensation. Under such conditions, the product may be
damaged and insulation may deteriorate so that it no longer
meets specifications.
Do not use the product where it may be exposed to
corrosive or combustible gases. The product may be
damaged.
This product is not designed to be entirely water- or dustproof. To avoid damage, do not use it in a wet or dusty
environment.
Preliminary Checks
Before using the product the first time, verify that it
operates normally to ensure that the no damage occurred
during storage or shipping. If you find any damage, contact
your dealer or Hioki representative.
Check that the 9418-15 AC ADAPTER (SA145A-1240V-6,
SINO AMERICAN), 9447 BATTERY PACK, or LR6
alkaline batteries are correctly fitted.
When using an AC adapter and if properly grounded three-
pin outlet is available, then using the three-core power cord
provides automatic grounding.
(2) Using the Printer
Avoid using the printer in hot, h umid environments, as this
can greatly reduce printer life.
Printing is not possible if the recording paper is loaded
wrong-side up.
(3) Recording Paper
This product uses a thermal printer. The recording paper
supplied has characteristics finely tuned for use with the
printer.
Using recording paper of a different specification may not
only result in impaired printing quality, but even prevent
the printer from operating. Always use the 9234
RECORDING PAPER.
(4) Storage
Always place the product in the head-up condition when it
is stored for an extended period of time. If the product is
left idle for a long time with the head pressing on the roller,
the roller may be deformed, resulting in uneven prin ting.
(5) Shipment
Use the original packing materials when reshipping the
product, if possible.
Always hold the disc by the edges, so as not to make
fingerprints on the disc or scratch the printing.
Never touch the recorded side of the disc. Do not place the
disc directly on anything hard.
Do not wet the disc with volatile alcohol or water, as there
is a possibility of the label printing disappearing.
To write on the disc label surface, use a spirit-based felt
pen. Do not use a ball-point pen or hard-tipped pen,
because there is a danger of scratching the surface and
corrupting the data. Do not use adhesive labels.
Do not expose the disc directly to the sun's rays, or keep it
in conditions of high temperature or humidity, as there is a
danger of warping, with consequent loss of data.
To remove dirt, dust, or fingerprints from the disc, wipe with
a dry cloth, or use a CD cleaner. Always wipe radially from
the inside to the outside, and do no wipe with circular
movements. Never use abrasives or solvent cleaners.
Hioki shall not be held liable for any problems with a
computer system that arises from the use of this CD, or for
any problem related to the purchase of a Hioki product.
This manual consists of the following chapters.
"Introduction", "Safety Notes", "Notes on Use" describe
precautions on use, overview, and features of this product.
Be sure to read them all. Next, check Chapter 1 to 3 and
the product to confirm your understanding of the function.
For details on the harmonic wave analysis function of the
8807-51, 8808 -51, refer to the supplement guide "Harmonic
Wave Analysis Function."
Chapter 1 Product Overview
Contains an overview of the product and its features.
Chapter 2 Installation and Preparation
Explains how to set the product up for measurement.
Chapter 3 Basic Button Operation
Explains how to operate the buttons for carrying out basic
measurement functions.
Chapter 4 Memory Recorder Function
Explains how to use the Memory Recorder functions of the
product.
Chapter 5 Recorder Function
Explains how to use the Recorder functions of the product.
Chapter 6 RMS Recorder Function
Explains how to use the RMS Recorder functions of the
product.
Chapter 7 Input Channel Settings
Explains how to make settings using the channel setting
screen.
Chapter 8 Trigger Functions
Explains how to use the Trigger functions of the product.
The 5.7-inch color LCD display has a resolution of 320 x
240 dots.
(2) Three functions support a variety of measurements
Memory recorder with up to 2.5 μs (all channels
simultaneously) sampling period (400 kS/s maximum
sampling rate).
Real-time recording capability to paper in Recorder function
RMS recorder function for recording rms values of AC
power supply lines and DC sources.
(3) Powerful trigger functions
Level trigger, window-in trigger, window-out trigger,
voltage drop trigger, waveform detection trigger, pattern
trigger (logic)
(4) Thermal printing with 8992 optional printer
4
4
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
(5) Scaling function
By setting the physical amount and the product to be used
for 1 V input, the measurement result can be converted into
any desired scale.
Two of the HIOKI CLAMP ON PROBEs, the 9018 and the
9132, feature a clamp mode that allows electric current
values to be read directly.
ADAPTER (SA145A-1240V-6, SINO AMERICAN). AC
adapter input voltage range is 100 to 240 V AC (with
10% stability) at 50/60 Hz. To avoid electrical hazards
and damage to the product, do not apply voltage
outside of this range.
Use the specified 9447 BATTERY PACK or LR6 alkaline
batteries. Do not use old and new batteries or batteries
of different types together.
The manufacturer disclaims all responsibility for any
accident or injury except that resulting due to defect in
its specified batteries or Battery packs.
Do not use old and new batteries or batteries of
different types together. If intend to, performance of
the product may suffer.
If not recharge, the 9447 can not be used.
For the details on charging batteries, refer to Section 16.2,
"9447 BATTERY PACK."
If the battery voltage drops while the product is running off of
battery power, the product will automatically switch off the
power supply. If the product is left in this state for an extended
period of time, the battery pack may be discharged too much.
Be sure to turn OFF the power switch on the product.
When the battery capacity is reduced, the power is automatically
turned off. Before turning on the power again with AC adapter
or new batteries, turn the power off and leave for about 30
seconds. If this time is shorter than 30 seconds, the power may
be immediately turned off even if powering on.
When not in use for a long time, to prevent possible corrosion
caused by battery leakage, remove the batteries before storage.
Actual performance will change from values indicated in this table
according to the types of batteries, temperature , and printing rate.
The printer cannot be used with alkaline batteries: use either the
9418-15 AC ADAPTER (SA145A-1240V-6, SINO AMERICAN)
or 9447 BATTERY PACK when using the printer.
The backlight is automatically dimmed when the printer is in use.
Check the following points before the power switch is
turned on.
Power supply matches Rated supply voltage (100 to
240 V AC: Voltage fluctuations of
supply voltage are taken into account.) and rated
supply frequency (50/60 Hz).
The Product is correctly installed (See Section 2.1).
Depending on the type and status of power supply of the
product, one of the following messages appears on the
Waveform Screen.
Operating with the AC ADAPTER.
Operating with batteries.
When the product is turned off, it memorizes the currently used
settings and reestablishes the same settings the next time the
product is turned on again. (Backup function)
other end is connected to the measurement object,
as you could receive an electric shock.
Use only the specified connection cable.
To prevent electric shock, an insulated BNC
connector is used. There is the possibility of shock
if a metal BNC is used because the input L terminal
and the metal part of the BNC connector have the
same electric potential.
CAUTION
When disconnecting the BNC connector, be sure to release
the lock before pulling off the connector. Forcibly pulling the
connector without releasing the lock, or pulling on the
cable, can damage the connector.
Using a connection cable other than that specified may
result in damage to the BNC connector, poor contact, or
other symptoms preventing accurate measurement.
Use the optional 9197, 9198, or 9217
CONNECTION CORD
to connect to the analog input terminals.
CONNECTION CORDs
Maximum
input voltage:
91979198
500 V300 V300 V
9217
Connecting to the main product
1. Align the grooves in the BNC
Groove o
the BNC
connector with the guide pins on
the product input connector, then
turn clockwise while pressing in
to lock the connector.
2. To remove from the product, turn
Connectorguide
the BNC connector
counterclockwise to release the
lock, then pull it.
For the 9322, the maximum rated voltage to earth is as
DANGE
follows.
When using grabber clips: 1500 V AC or DC (CAT ll) /
600 V AC or DC (CAT lll)
When using alligator clips: 1000 V AC or DC (CAT lI) /
600 V AC or DC (CAT lll)
To avoid electrical shock and possible damage to the
product, never apply voltage exceeding these limits
between the input terminals and the chassis, or
between the inputs of two 9322s.
The maximum input voltage is 1000 V AC or 2000 V DC
(CAT ll) / 600 V AC or DC (CAT lll).
Do not attempt to measure voltage exceeding these
limits, as doing so may damage the equipment or
cause an accident that might result in injury or death.
High voltage can be present at the inputs to the 9322.
To avoid electrical hazard, be sure to turn off the
equipment to be measured before connecting the test
leads. Turn the power on only after the leads have been
securely connected.
Before using the product, make sure that the sheathing
on the probes is not damaged and that no bare wire is
exposed. If there is damage, using the product could
cause electric shock. Contact your dealer or HIOKI
representative.
The Model 9322 is a differential probe that connects to
input modules for the 8800 MEMORY HiCORDER series.
High voltage applied between the H and L terminals is
divided by 1/1000 and output as a voltage signal relative to
GND.
Power jack
(Connect the 9418-10
AC ADAPTER
BNC connecto
(Connect to the analog
input terminal
Grabber cli
Alligator cli
Connection Method
1. Connect the alligator or grabber clips to the 9322.
For the 9322, the maximum rated voltage to earth is as
follows.
When using grabber clips: 1500 V AC or DC (CAT ll) /
600 V AC or DC (CAT lll)
When using alligator clips: 1000 V AC or DC (CAT lI) /
600 V AC or DC (CAT lll)
2. Connect the AC ADAPTER to power jack of the 9322.
The indicator lights when power is applied.
One 9418-15 AC ADAPTER (SA145A-1240V-6, SINO
AMERICAN) is required for each 9322 to be used.
Refer to the 9322 DIFFERENTI AL PROBE Instruction Manual for
details.
The ground sides of the logic inputs are connected in
WARNING
common to the frame ground of the product. Therefore,
if the product is powered by the AC adapter and the
object to be measured is powered from a different
power source, the possibility exists for electric shock
or damage to the equipment. Furthermore, even when
power is supplied from the same source, a potential
difference can exist due to different ground-side wiring ,
resulting in current flow through the logic probes that
could result in electric shock or damage.
To avoid these hazards, before connecting a logic
probe to the product and the object to be measured,
confirm that power is sup plied from the same
grounded outlet using 3-wire grounded power cords.
Do not connect logic probes other than supplied by
HIOKI to the logic inputs.
Object to
be measured
Outlet
MEMORY
HiCORDER
Power suppl
Ground
The logic input is located on the
top side of the product. Since one
logic probe can record 4 channels,
the combined maximum recording
capability for logic waveforms is 8
channels.
If no logic probe is connected, the corresponding logic
waveform is displayed on the screen at high level.
Carefully read the instruction manual supplied with the 9320-01
or 9321-01 LOGIC PROBE.
Always connect the clamp-on probe to the secondary
DANGE
side of a breaker. On the secondary side of a breaker,
even if the lines are shorted the breaker can trip and
prevent an accident. On the primary side, however, the
current capacity may be large, and in the event of a
short-circuit there may be a serious accident.
Before using the 9018-10, 9132-10 CLAMP ON PROBE,
be sure to carefully read its instruction manual and
familiarize yourself with the operation principles of this
product.
This product measures input voltage, so although current
cannot be measured directly, current can be measured by
the voltage output from the HIOKI 9018-10, 9132-10
CLAMP ON PROBEs.
Connect the BNC connector of the CLAMP ON PROBE to
the analog input terminal. For details, refer to Section 2.4.
To measure current, set the input coupling of the product to
"CLAMP." For details, refer to Section 7.3.4.
When the clamp-on probe is used for measurement, the
measurement precision will be affected both by the product
precision and clamp-on probe precision ratings.
When using the 9199 CONVERSION ADAPTOR, the 9018,
9132 CLAMP ON PROBEs can be used.
When measuring electric current, make sure that the direction of
the CLAMP ON PROBE matches the direction of electric
current.The direction of connection for the 9132-10 is opposite
from the 9018-10. If the CLAMP ON PROBE is faced in the
opposite direction, the phase will shift 180 degrees from the
actual value.
2.7 Connecting the 9018-10, 9132-10 CLAMP ON PROBE
Before connecting or disconnecting the printer unit, make
sure that the power switch is off. When the printer is not
connected, be sure to install the cover.
1. Using a plastic screwdriver,
remove the two screws that secure
the product's rear cover and lift off
the cover.
2. Insert the 8992 PRINTER UNIT.
3. Secure with the two screws
provided.
If the connector does not mate easily,
rock the printer slightly as shown
while pressing the products together.
The printer cannot be used with alkaline batteries: use either the
9418-15 AC ADAPTER (SA145A-1240V-6, SINO
AMERICAN) or 9447 BATTERY PACK when using the
printer.
When the stock cover is fully open, attempting to force it open
further may damage the cover.
The paper may jam if i t is not aligned with the roller.
Always place the product in the head-up condition when it is to
be transported or stored for an extended period of time. If the
product is left idle for a long time with the head pressing on the
roller, the roller may be deformed, resulting in uneven printing.
Be careful to correctly orient the recording paper. Nothing will
be printed if the recording paper is loaded with the wrong side
facing the print head.
We recommend using only the specified 9234 RECORDING
PAPER.
While unopened, thermal paper will not be affected by the
environment, provided that ambient temperature and
humidity do not exceed normal levels. For long-term
storage, temperature should be lower than 40
Low temperatures cause no problem.
After opening, protect the paper from strong light, to prevent
discoloration.
Storing data
The recording paper uses a thermochemical reaction.
Note the following points:
To avoid discoloration, do not leave recording paper in
direct sunlight.
Store at not more than 40(104 ) and 90% RH.
To keep definitive data, make photocopies of the recordings.
Thermal paper will blacken when brought into contact with
alcohol, ester, ketone, or other volatile organic substances.
If the thermal paper absorbs an organic solvent such as
alcohols or ketones it may no longer develop properly, and
recorded information may fade. Soft PVC film and
transparent contact adhesive tape contain such solvents, so
avoid using them with recordings.
Avoid interleaving the thermal r ecordings with damp diazo
copies.
To avoid the risk of electric shock and damage to the
23
DANGE
product
Do not apply more than 450 V AC or DC between the
analog inputs and product, or between the analog
inputs.
Terminal inputs should never exceed the limits
indicated below.
Input/output terminalsMaximum input voltage
Analog input450 V AC rms, DC
EXT TRIG-5 V to +10 V
TRIG OUT-20Vto+30V
500 mA max, 200 mW max
The maximum voltage applied to the probe and cables
connecting to the analog input of the product must
never exceed the limits in the following table.
Notwithstanding these limits, the voltage at the analog
input must not exceed 450V AC or DC.
The zero position may be offset when the input coupling is
DC or GND and the input is open circuit in the highsensitivity (10 mV) range.
The accuracy specifications apply after 30-minute warm-up
following power on. During warm-up, the zero position
drifts as indicated below.
The measurement start/stop operation is separately
determined by each measurement function.
For details, refer to "Start and Stop Operation," which
explains each measurement function.
1
2
3
START
STOP
START
button
STOP buttonPressing this button stops the measurement.
Pressing this button starts measurement.
The LED on the upper part of the switch
lights green during measurement.
3.7Other Buttons Operation
CARD
button
Manual trigger
CARD buttonUsing PC card:
When Manual Trigger is enabled (ON) on
the Trigger Screen, the trigger is activated
by pressing this button during the trigger
wait state.
Pressing the
cursor aligned with a setting item displays
the setting items.
This function disables all front-panel button controls of the
product. The function serves to prevent unintended changes
to settings during a measurement.
Press both cursor buttons (
) simultaneously for 3
seconds to switch the key lock on/off.
The key lock function will not be canceled by turning the
pow er off and on.
When the key lock function is active, the indication "
" is shown on the display.
LOCK
KEY
If the backlight saver function is used and the display
backlight turns off, it can be turned on again by touching
any button. The function assigned to the button will not be
activated.
Press the SET button to
display the Channel Screen.
Set the waveform display color,
voltage axis range, zero position,
etc. Refer to Chapter 7.
The scaling function can be used to
convert the output voltage
from a sensor, for example, to a
physical property of the measured
object.
Refer to Section 7.5.
35
SET
1
2
3
4
5
6
Settings on the Trigger Screen
SettingsontheTriggerScreen
Start measurement
Press the START button to start measurement.
To avoid inadvertent measurements, the START button
activation condition can be selected. Refer to Section 10.1.1,
for details.
Zero adjustment is performed automatically for every
measurement.
Press the SET button to
display the Trigger Screen.
Set the trigger source, type and
condition, and pre-trigger condition.
Refer to Chapter 8.
The 8807-01, 8808-01 has three function modes.
Additionally, the 8807-51, 8808-51 also offers a harmonic
wave analysis function. Select the Memory recorder
function.
Procedure
Status/ Channel/ Trigger Screen
1. Press the
SET
button to display the Status, Channel, or
Trigger Screen.
2. Move the flashing cursor to the position shown in the
figure below and select
Memory Recorder Function
Analog input signals are converted by the A/D converter to digital values
for display and storage in the internal memory.
X-Y combined plotting of multiple channels is possible.
2. Move the flashing cursor to
display format using the
Single
Dual
Quad
X-Y (Dot)
X-Y (Line)
SET
button to display the Status Screen.
Display Format
and select the
buttons.
Flashing cursor
Display and record as one graph.
Display and record as two graphs.
Display and record as four graphs.
Display and record the waveform as an X-Y (Dot)
Screen.
Display and record the waveform as an X-Y (Line)
Screen.
When the Dual or Quad display is selected, the channel to
be displayed in each screen partition can be set.
For details, refer to Section 7.3.2, and for the settings on
the X-Y Screen, refer to Section 4.4.
When the function is enabled, measurement data are
automatically stored on PC card after they are captured.
This function stores a file in the directory currently selected
on the File Screen.
Procedure
(1) Select the data format for Auto Save.
Status Screen
SET
1. Press the
button to display the Status Screen.
2. Move the flashing cursor to
saving using the
OFF
Binary
Text
Auto save is disabled.
Data are stored as binary data.
Data are stored as text data.
Data stored in the text format is not readable by
the product.
(2) If using the A/B cursors on the screen, specify the
area to save.
Flashing cursor
Move the flashing cursor to the position shown in the
figure below and use the
keys to make a setting.
ALL
A-B
Store data for all captured waveforms.
If A cursor is in use, data is saved from the
cursor to the end of input.
If both A and B cursors are in use, data between
the cursors is saved.
(3) Enter the file name to be used for Auto Save.
Once the file name has been entered, measurement data is
automatically saved as the entered name.
When multiple files are saved in succession, numbers are
appended to the file name. If the file name entry is left
blank, the file name is automatically set as "AUTO,
AUTO0001", etc.
1. Move the flashing cursor to
File Name
and use the
buttons to display the Comment Entry Screen.
2. Move the cursor to the desired character to enter using the
buttons. For details, refer to Section 11.3.3.
When the text data format is selected, enable t hinning.
Move the flashing cursor to
Thin Out
and use the
buttons to make a setting.
Flashing cursor
NOTE
Data for the channel for which the waveform is displayed is
saved.
During measurement, if remaining PC card capacity runs low
enough for a warning message to appear, press the STOP button,
then remove the PC card once measurement ha s been suspended.
Data may be lost if the PC card is inserted or removed during
measurement.
Files will be saved in the currently selected directory.
When "Wave" is selected, Smooth Print can be set only for the
Memory Recorder function. Instantaneous waveforms recorded
by the RMS Recorder can be smoothed.
When "Logging" is selected, if the print interval is set longer
than the record length, only the first point will print.
This function is available when the 8992
connected. Printout is carried out automatically after a
waveform has been captured for the specified recording
length.
Procedure
Status Screen
1. Press the
2. Move the flashing cursor to
using the
NOTE
OFF
ON
When cursor A and B are enabled, partial printing is executed.
For using the A/B cursors, refer to Section 9.1.
When the roll mode is enabled and the time-axis range is 10
ms/DIV to 5 min/DIV, data is displayed and printed
simultaneously.
The printer cannot be used with alkaline batteries: use either the
9418-15 AC ADAPTER (SA145A-1240V-6, SINO
AMERICAN) or 9447 BATTERY PACK when using the
printer.
X-Y combined plotting can be performed when the Display
Layout is set to X-Y display on the Status Screen. For the
8808-01, 8808-51 when the X axis is assigned to one
analog channel, up to three X-Y combined plots are
possible.
Voltage axis magnification/compression is active also when
using X-Y combined plotting.
Using the A/B cursors, it is possible to specify the data
between the cursors for partial plotting.
Procedure
Status/ Channel Screen
1. Press the
2. Move the flashing cursor to
X-Y DotorX-Y Line
(
3. Press the
SET
button to display the Status Screen.
Display Format
and select format
) using thebuttons.
Flashing cursor
SET
button with the Status Screen displayed to
display the Channel Screen.
4
5
4. Select the display color using thebuttons.
5. Move the flashing cursor to the channel to use as the X
axis, and select it using the
buttons.
Channels other than the one selected for the X axis are the
Y-axis channels.
Using the A/B cursors, it is possible to specify a range for
partial X-Y plotting. Refer to Section 9.1. (Normal X-Y
plotting covers all data of the recording length.)
Procedure
Status/ Channel/ Waveform Screen
1. Display the input waveforms, except for the X-Y display.
2. Using the A/B cursors, specify the desired portion for
plotting.
SET
3. Press the
4. Select format (
button to display the Status Screen.
X-Y dotorX-Y line
) on the
Display Format
item.
5. On the Channel Screen, select the channel to use as the x
axis, using the
Partial X-Y plotting is disabled when using the horizontal
cursor.
When using only cursor A, X-Y combined plotting is applied
only to the range from cursor A to the end of the
measurement.
When the number of channels to be used is set to one, X-Y
combined plotting is not available.Refer to Section 4.8.1.
This function automatically selects the time axis range,
voltage axis and zero position (excluding when the setting
of input coupling is "CLAMP").
The time axis range is applied to the lowest numbered
channel with waveform display enabled, and is
automatically set to record 1 to 2.5 cycles on the Waveform
Screen.
Procedure
Waveform Screen
1. Press the
button to display the Waveform Screen.
DISP
2. Press both sides of the
seconds to activate auto-ranging.
3. Press the
START
to cancel. To cancel forcibly during auto-range operation,
press the
button. If measurement has started using the
STOP
auto-range function:
Conditions related to the input
channels
Voltage axis range, zero position:
Value set automatically
Magnification/compression ratio along
the voltage axis:
Single display x1
Other displays x1/2
Low-pass filter, Input coupling:
OFF, DC
Status conditions
Time axis range (time/div):
Value set automatically
(Magnification/compression ratio: x1
Sequential save: OFF
(all channels)
TIME/DIV
button for at least 2
button to execute, or press the
Trigger conditions (one channel only)
Trigger mode: AUTO
AND/OR for internal and external
triggers: OR
Pre-trigger: 20%
Internal trigger:
Enables only the waveform display
channel with the lowest number. (If
the difference between maximum and
minimum is 2 divisions or less, the
next channel is enabled.)
Trigger type : Level
Slope:rising
Trigger level: value set automatically
Filter: OFF
STOP
button
NOTE
When the auto-range function is engaged a trigger output signal
is generated. Take suitable precautions when using the autorange function concurrently with the trigger output connector.
button activation
condition can be selected. Refer to Section 10.1.1.
When pre-trigger is activated, the trigger is not
enabled until the specified pre-trigger data has been
read.("Pre-trig wait" is displayed.)
"Wait for trig" is displayed until the trigger condition
occurs.
Data recording starts when the trigger conditions are
met. When trigger mode is AUTO, the product waits
for about 1 second for trigger conditions to be met.
After this interval, data recording starts, regardless of
trigger state.
During recording, "Storing" is displayed.
When the Roll mode is enabled, measurement data is
displayed (scrolling) as it is recorded. When automatic
printing is also enabled, data prints at the same time it
is displayed.
Measurement can be aborted by pressing the
STOP
button twice during measurement.
Then the waveform data until the
STOP
button is
pressed is displayed and saved. When the auto saving
is enabled, the data until a forced termination is
automatically saved. (automatic printing will not
AUTO
execute)
Trigger mode: REP/AUTO
Activates the trigger waiting condition. Input resumes
when the trigger conditions are met.
STOP
Trigger mode: SING
Measurement can be aborted by pressing the
STOP
button twice (automatic printing will not execute).
Trigger mode: REP/AUTO
During measurement, pressing the
STOP
button once
records up to the set recording length (waveform
display, printing and auto-save are executed).
Measurement can be aborted by pressing the
STOP
button twice (automatic printing will not execute).
If measurement is aborted after a trigger event occurs, waveform
data up to the abort are retained. However, no data is retained if
measurement is aborted before a trigger event. If the trigger
mode is Continuous or Auto, and waveforms remain from a
previous trigger event, the waveforms are displayed (excluding
maximum recording length).
Set segmenting of measurement data memory.
The record length can be made longer when using fewer
channels.
Procedure
Extended Status Screen
1. Press the
2. Move the flashing cursor to
Extended Status Screen using the
3. Move the flashing cursor to
using the
1CH
2CH
4CH
button to display the Status Screen.
SET
Extended Status
and display the
buttons.
Use Channel
and select channel
buttons.
Flashing cursor
Internal memory is allocated to channel 1, so
measurement data on channels 2 and above
cannot be stored to internal memory. Only
Channel A can be used for logic.
Internal memory is allocated to channel 1 and 2, so
measurement data on channels 3 and above cannot
be stored to internal memory. Channels A and B
can be used for logic.
Internal memory is allocated for the four channels
(Model 8808-01, 8808-51).
Channels A and B can be used for logic.
This mode can be used at a time axis range setting of 10
ms/DIV to 5 min/DIV. When roll mode is set to ON, the
waveform is displayed immediately at the start of recording
(the screen scrolls).
When the 8992 PRINTER UNIT is connected and the Roll
Mode and Auto Print are enabled, printing occurs
concurrently with waveform display.
Procedure
NOTE
Extended Status Screen
1. Press the
2. Move the flashing cursor to
Extended Status Screen using the
3. Move the flashing cursor to
using the
OFF
ON
Disables the roll mode.
Enables the roll mode.
Even if set to ON, the Roll Mode is inoperative when the time
axis range is set faster than 10 ms/div.
When Roll mode is set to ON, the settings for sequential saving
is automatically turned OFF.
This function divides the memory into separate blocks, each
of which can be used for waveform recording.
Input signal capture is carried out continuously using the
trigger, storing waveform data successively in each block.
During recording, no display or printout is carried out.
This reduces dead time (non-sensitivity periods due to
display and printing delays).
When continuous print (auto print) is being performed in
REPEAT trigger mode
RecordinglengthRecordinglengthRecordinglength
Briefphenomena maynot register
55
Using sequential save
Recordinglength RecordinglengthRecordinglength
Data are divided into blocks and recorded in memory.
Dead time: (interval in which no sampling occurs due to display
andprintprocessin
Select whether waveforms should be
displayed on screen after recording to each
block.
Select which blocks to display on screen
after measurement.
Select the block at which to start
measurement.
Select the block at which to end
measurement.
Press the SET button to return to the Status
Screen.
NOTE
Dead time during which no further data is sampled, in order to
permit display and recording after the acquired data is recorded
to one block, is about 4 ms.
Dead time, however, may slightly fluctuate by key operation.
While the roll mode is being used in memory recorder function
(on the Extended Status Screen), the sequential save function is
disabled.
When setting the recording length for Sequential Save, the
recording length setting (on the Status Screen) is
automatically set to the same length.
The variable recording length can be set. Refer to Section
The recording length and maximum number of divisions are
automatically determined according to the number of
available channels, as shown in the tables below. Refer to
Section 4.8.1 for details.
Maximum number of divisions
59
Recordingchannels
length (DIV)
Number of
124
2032323216
503232168
100321684
20016842
400842--80042------
16002---------
*This shows the number of divisions for the instantaneous
waveform recording (Memory Recording) of the RMS Recorder
function.
Instantaneous
waveform
recording *
(5) Follow-up waveform display
Enabling Subsequent Waveform Display causes the
waveforms that were input to each block following a trigger
event during Sequential Save to be displayed one by one
(resulting in longer dead time).
Move the flashing cursor to
to be displayed using the
Display Block
buttons to make a setting.
and set the block
Flashing cursor
After measurement starts, the displayed block is updated by
the recorded block. When measurement is completed, the
most recently recorded block is displayed.
(7) Set the first block
Move the flashing cursor to
block to be displayed using the
Relation between trigger mode and sequential save function
63
Measurement
start
First block
Last block
Waveform display
Auto saving
Auto print
SINGLE
Trigger
mode
REPEAT
AUTO
STOP
Press the
START
button to start measurement, and
LED lights in green. Data recording starts when trigger
conditions are met.
First block is stored by trigger setting.
The following waveform display ON:
waveform is displayed.
The following waveform display OFF:
waveform is not displayed.
Last block is stored by trigger setting.
The following waveform display ON:
waveform is displayed.
The following waveform display OFF:
waveform is not displayed.
Displays last block waveform.
When auto saving is enabled, waveforms in all blocks
are saved in a batch.
When auto printing is enabled, all blocks are printed
from the first block while individual blocks are
displayed.
Trigger mode: REP/AUTO
Activates the trigger waiting condition. Input resumes
when the trigger conditions are met.
Trigger mode: SING
STOP
button once
Measurement
end
Measurement can be aborted by pressing the
button twice (automatic printing will not execute).
Trigger mode: REP/AU TO
During measurement, pressing the
STOP
records up to the end of the block (waveform display,
printing and auto-save are executed).
Measurement can be aborted by pressing the
STOP
button twice (automatic printing will not execute).
Numerical calculations apply to recording by the Memory
Recorder and Instantaneous waveform recording in RMS
Recorder. Results of calculations are displayed numerically
(by printing or saving). Waveforms stored by the
Instantaneous waveform recording in RMS Recorder are
calculated using the Memory Recorder function. When the
A/B cursors are used, calculations are performed on the
data between the cursors.
The following eleven types of calculations can be
performed.
(1) Average value (2) Peak-to-peak value (3) Maximum value
(4) Time to maximum value(5) Minimum value
(6) Time to mini mum value (7) Rms value
(8) Period (9) Frequency (10) Area value
(11) XY area value
Settings Screen
Display
ON/OFF Setting
Calculation Selection
Channel Selection
for Calculation
Print Calculation
Results
Save Calculation
Results
Calculation
Execution
Displaythe Numerical calculation Screen.
OFF
ON
Average
Average
Average
Average
Value
Value
Value
Value
CH1
CH1
CH1
OFF
ON
OFF
ON
New Data
Existing
Data
Numerical calculation is not performed.
Numerical c alculation is performed.
Up to four simultaneous calculations
(numbered 1 to 4) can be set.
Channels which are disabled for
displaying and recording are also
disabled for calculations.
Calculation results are not printed.
Calculation results are printed.
Calculation results are not saved to PC
card.
Calculation results are saved to PC card.
Display the Waveform Screen, and press
the START button to start calculations.
Waveform input data is subject to
calculation.
Calculation is applied to the waveform
data in internal memory. Move the
flashing cursor to Execute, and execute
Calculates the data stored in memory and read from a PC
card.
Move the flashing cursor to
Execute
and use the
buttons to make a setting.
Flashing cursor
Operational Example of Numerical Calculation
Partitioning Display
NOTE
Full Screen Display
Refer to Section 9.5 for information about waveform display
area setting.
Numerical calculation results cannot be displayed when the
Settings window is open on the Waveform Screen: press the
Channel select button (CH1 to CH4) to close the Settings
window to display calculation results.
button to
begin measurement. Data input begins when trigger
conditions are met. After the specified data record amount
is input, numerical calculation starts, and when calculation
is finished the results are displayed. If printing or saving of
calculation results has been enabled, those processes
execute.
When starting to input subsequent data, the previous data is
cleared.
The calculation area can be specified by the A/B cursors. If only
cursor A is used, the calculation area extends from the cursor to
the end of the measurement data. The (horizontal) line cursor is
not used to specify calculation area.
When both numerical calculation and sequential save are
enabled at the same time, calculation starts after sequential save
finishes. All blocks are processed by the calculation, so the
calculation result is refreshed and the final calculation result is
that of the last block. To view the results of all calculations,
enable printing or saving of the calculation results.
(1) Real time display and printing of the input signal
(2) Real time continuous recording of the input signal
(3) All input channel data are recorded on the same time axis.
Since data for all channels can be superimposed, the relative
relationship between input signals can be observed visually.
(4) Time axis setting
100 ms/DIV to 1 h/DIV (14 steps)
(5) Time axis resolution 80 points/DIV (printer)
(6) Sampling period 2.5 μsfixed
(7) Waveform magnification/compression display and printout
Time axis direction: x1tox1/500
Voltage axis direction: x10 to x1/2 (single)
(8) Display layout
Time axis waveform: single, dual, and quad displays
X-Y waveform: single display (dot/line)
(9) Storage capacity
400 divisions of the data are stored in memory.
(10) Logging function
Numeric printout of waveform data
(11) Reprinting capability
The last 400-division quantity of data is stored in memory,
Press the START button to start measurement.
To avoid inadvertent measurements, the START button
activation condition can be selected. Refer to Section 10.1.1,
for details.
Zero adjustment is performed automatically for every
measurement.
Press the SET button to
display the Channel Screen.
Set the waveform display color,
voltage axis range, zero position,
etc. Refer to Chapter 7 for details.
The scaling function can be used to
convert the output voltage
from a sensor, for example, to a
physical property of the measured
object.
Refer to Section 7.5.
Press the SET button to
display the Trigger Screen.
Set the trigger source, type and
condition, and pre-trigger condition.
Refer to Chapter 8.
The 8807-01, 8808-01 has three function modes.
Additionally, the 8807-51, 8808-51 also offers a harmonic
wave analysis function. Select the Recorder function.
Procedure
Status, Channel, Trigger Screen
1. Press the
SET
button to display the Status, Channel, or
Trigger Screen.
2. Move the flashing cursor to the position shown in the
figure below and select
Recorder Function Description
Input signals converted to digital data are displayed and printed in real
time (when the 8992 PRINTER UNIT is connected).
When measurement is finished, the last 400-division quantity is stored in
internal memory, so reverse scroll viewing and reprinting are possible.
Set the speed for inputting and storing the waveform of the
input signal. Time axis range setting expresses the time for
1 division. The sampling period is fixed to 2.5 μs (400
kS/s).
Status Screen
1. Press the
2. Move the flashing cursor to
axis range using the
Pressing the
automatically to the Time Axis position for setting.
Waveform Screen
1. Press the
2. Move the flashing cursor to the position shown in the
figure below and use the
SET
button to display the Status Screen.
Time/DIV
buttons.
TIME/DIV
DISP
button to display the Waveform Screen.
button moves the flashing cursor
buttons to make a setting.
and select the time
Flashing cursor
The data interval for the
time axis range setting is
1/80 (1 division=80 data
points).