Test lead (Yellow, with 15-V zener diode)............ 1
AC Adapter (SA-141A0F-1 1)
(with power cord)............1
Thank you for purchasing the HIOKI Model 7016 SIGNAL
SOURCE. To obtain maximum performance from the instrument,
please read this manual first, and keep it handy for future reference.
1
Introduction
1
Verifying Package Conten ts
When you receive the instrument, inspect it carefully to ensure that
no damage occurred during shipping. In particular, check the
accessories, panel switches, and connectors. If damage is evident,
or if it fails to operate according to the specifications, contact your
dealer or Hioki representative.
Instrument
Accessories
2
3
4
5
6
7
8
9
11
Page 8
2
Carrying Case ............1
Battery pack (Ni-MH battery, installed) .... 8
Instruction manual
Operation guide
Connection band ............................ 1
Verifying Package Contents
The supplied battery pack (Ni-MH battery, installed) is not charged
when shipped from the factory. Charge it using the supplied AC
adapter.
❖ 2.2 "Charging the Internal Batteries" (20 p.)
Page 9
10
Safety Information
This manual contains information and warnings essential for safe
operation of the instrument and for maintaining it in safe operating
condition. Before using it, be sure to carefully read the following
safety precautions.
3
Safety Information
1
This instrument is designed to comply with IEC 61010 Safety
Standards, and has been thoroughly tested for safety prior to
shipment. However, mishandling during use could result in
injury or death, as well as damage to the instrument. U sing
the instrument in a way not described in this manual may
negate the provided safety features. Be certain that you
understand the instructions and precautions in the manual
before use. We disclaim any responsibility for accidents or
injuries not resulting directly from instrument defects.
Safety Symbols
The following symbols in this manual indicate the relative importance of cautions and warnings.
In the manual, the symbol indicates particu larly impo rtant information that the user should read before using the instrument.
The symbol printed on the instrument indicate s that the user
should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates a grounding termin al.
Indicates a double-insulated device.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates that incorrect operation presents an extreme hazard
that could result in serious injury or death to the user.
2
3
4
5
6
7
8
9
Indicates that incorrect op eration presents a significan t hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of
injury to the user or damage to the instrument.
Indicates advisory items related to performance or correct
operation of the instrument.
11
Page 10
4
Safety Information
Other Symbols
Indicates the reference.
❖
Indicates that descriptive information is provided bellow.
*
Indicates quick references for operation and remedies for troubleshooting.
Measurement categories
This instrument complies with CAT II safety requirements.
To ensure safe operation of measurement instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement categories. These are defined as follows.
CAT IIPrimary electrical circuits in equipment connected to an AC electrical outlet
CAT III Primary electrical circuits of heavy equipment (fixed installations) connected
CAT IV The circuit from the service drop to th e service entrance , and to the power
Using a measurement instrument in an environment designated
with a higher-numbered category than that for which the instrument
is rated could result in a severe accident, and must be carefully
avoided.
Use of a measurement instrument that is not CAT-rated in CAT II to
CAT IV measurement applications could result in a severe accident, and must be carefully avoided.
by a power cord (portable tools, household appliances, etc.)
CAT II covers directly measuring electrical outlet receptacles.
directly to the distr ibution panel, and f eeders from th e distribution pan el to
outlets.
meter and primary overcurrent protection device (distribution panel).
Page 11
10
Accuracy
Safety Information
We define measurement tolerances in terms of f.s. (full scale), rdg.
(reading) and dgt. (digit) values, with the following meanings:
f.s.
(maximum display value or
scale length)
rdg.
(reading or displayed value)
dgt. (resolution)
The maximum displ ayab le val ue or s cale length. Thi s
is usually the name of the currently selected range.
The value curre nt ly b ei ng m ea sur ed a nd i nd icat ed on
the measuring instrument.
The smallest displayable unit on a digital measuring
instrument, i.e., the input value that causes the digital
display to show a "1" as the least-signific a nt digit.
5
1
2
3
4
5
6
7
8
9
11
Page 12
6
Operating Precautions
Operating Precautions
Follow these precautions to ensure safe operation and to obtain
the full benefits of the various functions.
• Befo re connecting the test leads, check the position of the
function selector. Applying voltage when a non-voltage
measuring range is selected ma y cause bodily injury and
damage to the meter. Also, remove the test leads fr om the
device under test before switching the function selector.
• Never apply voltage to the test leads when the Resistance
Measurement (OHM), Current Measurement (DCA, ACA,
ACDCA), Continuity or Diode Test are selected. Doing so
may damage the instrument and result in personal injury.
To avoid electrical accidents, remove power from the circuit
before measuring.
• Use only the supplied AC Adapter (SA-141A0F-11). AC
Adapter input voltage range is 100 to 240 VA C (with ±10%
stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this
range.
• To avoid the possibility of electric shock, always con nect
the green wire from the AC Adapter to ground.
• Do not use an AC Adapter other than the one supplied.
Doing so could caus e da mage to th e m eter or ot her elec trical mishap.
• Do not use the supp lied AC Adapter with another device. It
could damage the device or cause another electrical mishap.
• When replacing the batteries, do not mix old and new
batteries, or different types of batteries. Also, be careful to
observe battery polarity during installation. Otherwise,
poor performance or damage from battery leakage could
result.
• Do not allow the instrument to get wet, and do not take
measurements with wet hands. This may cause an electric
shock.
• Do not apply voltage or current exceeding the measurement range of each terminal. H ioki can not take responsibility for accidents involving electric shock or insulation
failures that could occur as a result of exceeding the specified range limitations.
Page 13
Operating Precautions
10
• This instrument is designed for use in doors. It can be op erate d at
temperatures between 0 and 40°C without degrading safety.
• Calibration and repair of this instrument should be performed only
under the supervision of qualified technicians knowledgeable
about the dangers involved.
• If the protective functions of the instrumen t are damaged, either
remove it from service o r mark it clea rly so that ot her s do not use
it inadvertently.
• Do not store or use the instrum ent where it could be ex posed to
direct sunlight, high temperature or humidity, or condensation.
Under such conditions, the instrument may be damaged and insulation may deteriorate so that it no longer meets specifications.
• This instrument is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it
might be splashed with liquid. This may cause damage.
• For safety reasons, when taking measurements, only use the
L9207-10 provided with the instrument.
• To avoid d ama ge to th e inst ru men t, prot ect it fro m ph ysical shock
when transporting and handling. Be especially careful to avoid
physical shock from dropping.
• To avoid corrosion from battery leakage, remove the batteries
from the instrument if i t is to be s tore d for a long time.
• Because the input and output terminals are isolated from each
other, signal source and DMM functions can operate simultaneously as two roles of the single instrument.
• Data can be transferred to a PC using the Model 3856-02
COMMUNICATION PACKAGE (U SB).
• Correct source and measurement may be impossible in the
presence of strong magnetic fields, such a s near transformers
and high-current conductors, or in the presence of strong electromagnetic fields such as near radio transmitters.
7
1
2
3
4
5
6
7
8
9
Before Use
Before using the instrument the first time, verify that it operates
normally to ensure that the no damage occurred during storage or
shipping. If you find any damage, contact your dealer or Hioki representative.
11
Before using the instrument, mak e s ure that the insulation
on the test leads is un damag ed and that no bar e c ondu ctors
are improperly exposed. Using the product in such conditions could cause an electric shock, so contact your dealer
or Hioki representative for replacements. (Model L9207-10)
Page 14
8
Operating Precautions
Pre-Operation Inspection
Before using the instrument, verify that the instrument operates
normally to ensure that the no damage.
Page 15
1.1 Product Overview
10
Overview1
1.1Product Overview
Handy multi-f unction calibrator
A programmable voltage/current source that includes a highperformance DMM.
9
1
2
3
Many sourcing functions
In addition to constant voltage and current sourcing, voltage pulses
can also be sourced, suitable for applications such as flow meter
calibration.
Extensive measu ring functions
DMM functions including AC/DC voltage and current, resistance,
frequency, temperature measurement, diode and continuity test
are fully supported.
Supports various maintenance, adjustment and
testing applications
The multi-function high-precision source supports maintenance of
instrumentation systems and plants, calibration of industrial
measuring instruments, and electrical circuit testing.
PC control and data acquisition with an optional
communications package
Using the separately available Model 3856-02 COMMUNICATION
PACKAGE(USB), signal source settings can be made on a PC,
and measured values transferred to a PC.
4
5
6
7
8
9
11
Page 16
10
1.2 Features
1.2Features
Sources bipolar DC voltage from -15.000 to 15.000 V (@100 μV
finest resolution) and DC current from -25 .000 to 25.000 mA (@1
μA finest resolution), including sink (absorbed current) and source
output.
Voltage pulses at one of 28 frequencies from 0.5 to 48000 Hz can
be generated with settable duty cycle, pulse width and amplitude.
For each function and range, source settings can be stored in
memory for up to 16 steps, which can then be output in sequence
continuously, singly or manually using the Scan sourcing function.
Measures true effective value (rms) AC/DC voltage and current
(with crest factor less than three).
As well as measuring AC/DC voltage and current, the meter also
tests continuity and measures resistance, diodes, frequency and
temperature (using an IEC584-1:1995-compliant type K
thermocouple).
Page 17
10
1.3Names of Parts
Display (LCD)
❖ 1.4 "Display (LCD)" (14.p)
Front
Input terminal
❖
2.3 "Connecting the Test Leads"
(21.p)
Output terminal
❖
2.3 "Connecti ng th e Test Leads"
(21.p)
Buttons
❖ (12 p.)
Function
selector
❖ 2.5 "Function
Selection" (24.p)
11
1.3 Names of Parts
1
2
3
4
5
6
7
8
9
11
Page 18
12
Buttons
SHIFT butt on
MODE, AC/DC button
, REL, TEMP button
, DUAL button
❖ Detail→ "Source Mode Button Functions" (27.p),
"Measurement Mode Button Functions" (29.p)
OUTPUT, Hz button, HOLD, MAX MIN button
, RANGE, AUTO button
1.3 Names of Parts
Important!
• Source-function related button operations are indicated by
yellow labels on the meter.
• The [
SHIFT] button switches between source and
measurement functions.
Page 19
10
Side
Jack for AC adapter
❖ 2.1 "Connecting the AC
Adapter" (19.p)
Slide switch
❖ 2.2 "Charging the
Internal Batteries" (20.p)
Back
Battery cover
❖ 8.1 "Replacing the
Batteries" (96.p)
Charging lamp
❖ 2.2 "Charging the
Internal Batteries" (20.p)
Connector for RS-232C
❖ 5.3 "RS-232C
Communication Functions"
(75.p)
1.3 Names of Parts
13
1
2
3
4
5
6
7
8
9
11
Page 20
14
Secondary Display
Contents and Units
Indicators for
Secondary Display
Primary Displa y
Contents and Units Indicators
for Primary Display
Source Function Indicators
Source Function Status
and Mode Indicators
Meter status display
Measurement Func tion
Application Indicators
1.4 Display (LCD)
1.4Display (LCD)
Page 21
10
Meter status display
0°C
0-20
4-20
Indicates low battery v oltage (below the guaranteed accu racy
voltage - approximately 8 V). It blinks.
Indicates the Auto Power Off function is enabled.
15
1.4 Display (LCD)
1
Primary Display
Shows measured value.
Content s and Units Indicators for Primary Display
DC
AC
ACDC
°C
%
m sec
mA
mV , V
Hz, kHz
M
Ω, kΩ,
Ω
Indicates measurement of DC voltage (DCV) or current (DCA).
Indicates measuremen t of AC voltage (AC V) or cur ren t (A CA ).
Indicates measurement of DC + AC voltage (ACDCV) or current
(ACDCA).
Unit of temperature measurement
Indicates percentage measurement of duty cycle or 0-20 or 4-20
mA.
Unit of pulse width measurement
Indicates negative tr igger slope when mea suring duty cycle or
pulse width.
Indicates positive tri gger slope when measuring duty cycle or
pulse width.
Unit of current measurement
Unit of voltage measurement
Unit of frequency measurement
Unit of resistance measurement
Indicates standard connection is normalized to 0°C.
Indicates displayed v alue is percentage of 0-20 mA measure-
ment current.
Indicates displayed v alue is percentage of 4-20 mA measure-
ment current.
2
3
4
5
6
7
8
9
11
Page 22
16
0-20
4-20
1.4 Display (LCD)
Secondary Display
Contents and Units Indicators for Se co ndary Display
DC
AC
ACDC
°C
%
m sec
mA
mV, V
Hz, kHz
Shows either measured or sourced value.
Indicates measurement of DC voltage (DCV) or current (DCA).
Indicates measurement of AC voltage (ACV) or current (ACA).
Indicates measurement of DC + AC voltage (ACDCV) or current
(ACDCA).
Unit of temperature measurement
Indicates percentage measurement of duty c ycle or setting of
pulse duty cycle.
Unit of pulse width measurement and setting
Unit of current measurement and constant current (CC) sourcing
Unit of voltage measur ement and constant voltage (CV) sourc-
ing
Unit of frequency measurement and pulse frequency setting
Indicates displayed val ue is percentage of 0-20 mA measure-
ment current.
Indicates displayed val ue is percentage of 4-20 mA measure-
ment current.
Source Function Indicators
Pulse source function is active.
Constant current (CC) source function is active.
Constant voltage (CV) sour ce function is active.
Pulse source frequency setting
Pulse source duty cycle setting
Pulse sour ce pulse width settin g
Pulse source amplitude setting
Page 23
1.4 Display (LCD)
10
Source Function Status and Mode Indicators
Indicates no signal is being output.
17
Indicates a signal is being output (sourced or generated).
Indicates the [SHIFT] button is pressed (S ource Function set-
tings enabled).
Indicates ramp generatio n.
Indicates scan generation.
Measurement Function Application Indicators
+
DH
MAX A VG MIN
MAX
AVG
MIN
DH MAX
DH MIN
Diode test is active.
Continuity test is active.
Data Hold and Refresh Hold are enabled.
Recording function is active.
Maximum value is displayed.
Average value is displayed.
Minimum value is displayed .
Maximum value of 1-ms Peak Hold is displayed.
Minimum value of 1-ms Peak Hold is displayed.
• Use only the supplied AC adapter (SA-141A0F-11). AC
adapter input voltage range is 100 to 240 VAC (with ±10%
stability) at 50/60 Hz. To avoid electrical hazards and
damage to the instrument, do not apply voltage outside of
this range.
• To avoid the possibility of electric shock, always connect
the green wire of the AC adapter to ground.
• Do not use an AC adapter other than the one supplied.
Doing so could cause damage to the instrument or other
electrical mishap.
• Do not use the supplied AC adapter with another device. It
could damage the device or cause another electrical
mishap.
19
1
2
3
4
5
6
Connection Procedures
1.
2.
Set the slide switch on the meter as follows:
•[CHARGE]: to charge the batteries
•[
M]: to measure only
•[M/S] : to source only or to measure and
Connect the output plug from
the AC adapter to the jack on
the side of meter.
source simultaneously
Slide Switch
3. Verify that the mains voltage matches the inp ut voltage of the
AC adapter, and plug the adapter into an outlet .
7
8
9
11
Page 26
20
Charge lamp
2.2 Charging the Internal Batteries
2.2Charging the Internal Batteries
Do not attempt to charge any batteries other than the
supplied battery pack (Ni-MH battery). The
instrument could overheat and the batteries could
explode.
While using the supplied battery pack (8 Ni-MH batteries), use the
following procedure to charge the batteries when the battery warn-
ing () indicator appears on the LCD.
Charging Procedure
1.
Turn the function selector to [OFF].
2.
Move the slide switch to [CHARGE].
3. Connect the AC adapter t o th e jac k on t he side of the meter.
4. The charge lamp lights red as ch arging begins.
5. The charge lamp turns gre en when charging is finished.
• The supplied AC adapter operates from 100 to 250 VAC @47
to 63 Hz). The jack on the side of the meter accepts 14 VDC
±10%, center contact positive (+).
• When you use the instrument for the first time or when you
have not used it for a long time, the batteries might not be fully
charged. Please charge and discharge the batteries a few
times repeatedly.
Page 27
2.3 Connecting the Test Leads
10
L9207-10 TEST LEAD (one each red and black, for Input terminal)
Test leads
(one each red and black, for Output terminal)
Test lead (Yellow, with 15-V zener diode)
For Input Terminals
For Output Terminals
2.3Connecting th e Test Leads
21
The meter has two pair of terminals, labeled Input and Output. The supplied Model
L9207-10 TEST LEAD with probes (one each red and black) are for connection to
the Input terminals, and the other supplied test leads with clips (one each, red, black
and yellow) are for the Output terminals.
To prevent an electric shock accident, confirm that the white or red portion (insulation layer) inside the cable is not exposed. If a color inside
the cable is exposed, do not use the cable.
1
2
3
4
5
6
7
8
Removable sleeves are attached to the metal pins at the ends of the test leads.
The test leads can also be used with the sleeves removed.
Removing and attaching the sleeves
The tips of the metal pins are sharp, so take care not to injure yourself.
Removing the sleeves
Gently hold the bottom of the sleeves and pull the sleeves off.
Safely store the removed sleeves so as not to lose them.
Attaching the sleeves
Insert the metal pins of the test leads into the holes of the sleeves,
all the way in.
9
11
and firmly push them
Page 28
22
Input terminals
L9207-10
(black)
L9207-10
(red)
Output terminals
Test lead
(black)
Test lead
(red or yellow)
2.3 Connecting the Test Leads
2.3.1Connections
1.
Connect the L9207-1 0 TEST LEAD to the Input
terminals.
Connect the red Model L9207-10 TEST LEAD to
the "+" INPUT terminal, and the black Model L920710 TEST LEAD to the "-" INPUT terminal.
2.
Connect the test leads to th e Output terminals
Connect red test lead to the "+" OUTPUT terminal,
and the black test lead to the "-" OUTPUT terminal.
Only make connections when the source
function is OFF.
❖ Using the yellow test lead→ 6.2 "Loop Test" (81 p.)
Page 29
10
2.4Turning Power On / Off
The meter is turned on and off by both the slide switch and the
function selector.
Turning Power On
23
2.4 Turning Power On / Off
1
Measure only
Source only,
or
Source and
measure
simultaneously
Turning Power Off
1.Set the slide switch to [M].
2.Set the function selector to the
❖ 2.5 "Function Selection" (24 p.)
1.Set the slide switch to[M/S].
2.Set the function selector to the
❖ 2.5 "Function Selection" (24 p.)
Set the function selector to [OFF].
desired measurement function.
desired source or measurement
function.
2
3
4
5
6
7
8
9
• To prevent battery discharge when the meter is not in use, set
the function selector to [OFF] or set the slide switch to [M] or
[
M/S].
• Although measurement-only operation is possible with the
slide switch set to [
be shortened.
M/S], operating time per battery charge will
11
Page 30
24
: Source functions
2.5 Function Selection
2.5Function Selection
Select the desired function.
Important!
• The outer yellow labels around the function selector are for
source-related functions.
• The inner reverse-color labels are for measurement-related
functions.
• The [
SHIFT] button switches between source- and measure-
ment-related button functions.
2.5.1Source Function Settings
There are three types of source functions:
• Constant-voltage (
• Constant-current (CC) source
• Pulse source (
CV) source
PULSE)
The source functions are indicated by yellow labels
tion selector to the function you want to set up.
. Set the func-
Page 31
10
Setting Procedure
When constant-voltage sourcing (CV) is selected
(The indicator is displayed)
Constant-voltage source
1.5 V range
Constant-voltage RAMP source
15 V range
Constant-voltage source
15 V range
Constant-voltage SCAN source
1.5 V range
Constant-voltage SCAN source
15 V range
Constant-voltage RAMP source
1.5 V range
SCAN source: The indicator is displayed.
RAMP source: The indicator is displayed.
1. Press the [SHIFT] button (so theindicator is displayed).
25
2.5 Function Selection
This enables the source-relate d butto n functi on s.
2. Press the [MODE] button as necessary to select the desired
source mode.
The source mode selection order is as follows.
1
2
3
4
5
6
7
8
9
11
Page 32
26
When constant current sourcing (CC) is selected
(The indicator is displayed)
Constant current source
Constant current SCAN source
Constant current RAMP source
SCAN source: The indicator is displayed.
RAMP source: The indicator is displayed.
When pulse source (PULSE) is selected
(The indicator is displayed)
Pulse source frequency
Pulse source duty cycle
Pulse source amplitude
Pulse source pulse width
Frequency source : The indicator is displayed.
Duty cycle source : The indicator is displayed.
Pulse width source: The indicator is displayed.
Amplitude source : The indicator is displayed.
• Switches the source mode when sourcing constant voltage, current or pulses.
• Holding this button for one second activates the memory source setting mode.
❖ Detail → 3.4 "Using Memory Function (SCAN and RAMP)" (39 p.)
• Select the digit or polarity(±) to be changed.
• Press these buttons with memory source SCAN or RAMP output to select the
source mode.
❖
Detail → 3.4.1 "Using the SCAN Function" (40 p.), 3.4.2 "Using the RAMP Function" (45
p.)
• Increments (decrements) the blinking digit by 1.
• Switches output polarity(±) .
• When pressed ( or ) continuously, the value is decreased (increased)
automatically.
• Press this button to activate the SHIFT mode (appears).
• While is displayed, all buttons perform their source-related functions.
• Holding this button for one second toggles the LCD backlight on and off (even
when the [SHIFT] button is not pressed, the backlight turns off automatically
after a preset time.
❖ To set the backlight power-on option→ 5.4 "Power-On Options" (76 p.)
• Toggles source output on and off.
source output is inactive.
• Press this button to store settings for later output as SCAN or RAMP memory
source output values (proceeding to another step without pressing this button
invalidates the settings up to this point).
❖ Detailed Descriptions and Operation Procedures→ 3.4 "Using Memory Function
(SCAN and RAMP)" (39 p.)
27
2.5 Function Selection
1
2
3
4
5
6
7
8
9
is displayed when source output is active, andis displayed when
11
Page 34
28
Vo l tage measu r ement (mV),
Temperature measurement
Current
measurement
Resistance
measurement
Current
measurement
Diode test
2.5 Function Selection
2.5.2Measurement Function Settings
There are five types of measurement functions.
Voltage measurement (V, mV)
Current measurement
Resistance measurement
Diode test
Temperature measurement
Measurement functions are indicated by dark labels (on wh ite
background). Set the function selector to the function you want to
• When measuring resistance, enables and disables continuity (CONT)
testing.
• When measuring duty cycle (D UTY) or pulse width (WIDTH), holding this
button for one second toggles the trigger slope between rising and
falling edges.
• Holding this button for one second toggles the 1- ms Peak Hold fun ction on
and off.
• Enables and disables the Data Hold function.
• Enables the Refresh Hold function.
❖ 4.8.2 "Data Hold / Refresh Hold Function" (66 p.)
• Toggles through MAX, AVG and MIN.
❖ 4.8.1 "Recording Function" (63 p.)
• Enables and disables relative (relative value display) function.
• When mV measurement is selected, holding this button for one second
enables temperature (TEMP) measurement.
• Toggles between auto- and manual ranging.
• Changes ranges when manual ranging is enabled.
• Selects various Dual Display* combinations.
• Toggles the SHIFT mode (appears when active).
• Whileis displayed, all buttons perform their source- related functio ns,
so be sure thatis not displayed before performing button operations
for measurement functions.
• Holding this button for one second toggles the LCD backlight on and off
(even when the [SHIFT] button is not pressed, the backlight turns off
automatically after a preset time).
❖ 5.4 "Power-On Options" (76 p.)
• Changes the measurement mode (frequency, duty cycle or pulse width)
when using the Dual Display function.
❖ 4.8.5 "Dual Display Function" (70 p.)
29
2.5 Function Selection
1
2
3
4
5
6
7
8
9
11
Page 36
30
2.5 Function Selection
Page 37
10
Using Source Functions3
Please observe the following precautions t o avoid damage to the inst rument:
• Do not connect a device to be measured while the instrument is
turned off.
• Do not turn the instrument off and on or move the function switch
while a test device is connected for measurement.
• Always make sure to turn the source output off before changing
the connections at the OUTPUT terminals or device under test.
• Never connect the instrument to another voltage source when
using it to source voltage, or to another current source when
using it to source current. The instrument or device under test
could be damaged.
• Use the supplied yellow test lead when performing loop testing
of internally powered instruments.
• To use this instrument’s source function, set the slide switch to
the [
M/S] position. Source output is disabled in the other switch
positions.
• Holding the [/] buttons changes the setting continuously
within the settable range.
• In source (output) mode, the occurrence of electrostatic discharge at the output terminals can cause output to be inter-
rupted and the standby mode enabled (appears).
• In source (output) mode, if battery voltage falls below the level
necessary to maintain stable output, output is interrupted and
the standby mode is enabled (appears).
31
1
2
3
4
5
6
7
8
9
Important!
• The yellow labels around the outer circumference of the
function selector indicate signal source/generation
functions.
source-related button functions. ( is displayed on the
LCD.)
11
Page 38
32
3.1 Constant-Voltage Source (DCCV)
3.1Constant-Voltage Source (DCCV)
This mode outputs a constant voltage at the set value regardless of
load.
Procedure
1.
Set the slide switch to the [M/S] po s iti on.
2.
Set the function selector to the
CV] position.
[
Select one position.
When source only, any [
selected.
CV] position can be
3.
Press the [SHIFT] button and confirm that
is displayed on the LCD.
The source-related button functions are now enabled.
Page 39
33
10
Selecting the Source Mode
(The indicator is displayed)
Constant-voltage source
1.5 V range
Constant-voltage RAMP source
15 V range
Constant-voltage source
15 V range
Constant-voltage SCAN source
1.5 V range
Constant-voltage SCAN source
15 V range
Constant-voltage RAMP source
1.5 V range
SCAN source: The indicator is displayed.
RAMP source: The indicator is displayed.
3.1 Constant-Voltage Source (DCCV)
4. Press the [MODE] button as necessary to select the source
mode on the secondary display.
Pressing the [MODE] button repeat edly cha nge s the sou rce mo de
as shown below.
1
2
3
4
5
6
7
8
5. Press the [] or [] button to select a digit to change.
6. Press the [] or [] button to set a value.
Settable Ranges
[+0.0000 V]
[+00.000 V]
-1.5000 V to 1. 5000 V
-15.000 V to 15 .000 V
9
11
Page 40
34
3.1 Constant-Voltage Source (DCCV)
Press the [OUTPUT] button to output the
7.
set voltage value. Verify thatis displayed on the LCD.
8. Press the [OUTPUT] button again to dis able output. Verify that
is no longer displayed.
• To change the setting value, repeat Steps 5 and 6.
• If an overload such as a short circuit occurs while sourcing
voltage, output is disabled and the standby state is enabled
( lit), or the meter may be turned off by the internal protection circuitry. If the power is turned off, wait a short time bef ore
turning it back on.
Page 41
3.2 Constant-Current Source (DCCC)
10
3.2Constant-Current Source (DCCC)
This mode outputs a constant current at the set value regardless of
load.
Procedure
35
1
1. Set the slide switch to the [M/S] position.
2.
Set the function selector to the
[
CC] position.
Select one position.
When source only, any [
selected.
CC] position can be
3. Press the [SHIFT] button and confirm that is displayed
on the LCD.
4. Press the [MODE] button as necessary to select the source
mode on the secondary display.
Pressing the [MODE] button repeat edly cha nge s the sou rce mo de
as shown (page 36).
2
3
4
5
6
7
8
9
11
Page 42
36
Selecting the Source Mode
(The indicator is displayed)
Constant current source
Constant current SCAN source
Constant current RAMP source
SCAN source: The indicator is displayed.
RAMP source: The indicator is displayed.
3.2 Constant-Current Source (DCCC)
5. Press the [] or [] button to select a digit to change.
6. Press the [] or [] button to set a value.
Settable range
[+00.000 mA]
7. Press the [OUTPUT] button to output the set current value.
Ve rify thatis dis played on the LCD.
8. Press the [OUTPUT] button again to dis able output. Verify that
is no longer displayed.
• To change the setting value, repeat Steps 5 and 6.
• In current source (output) mode, if overloading occurs (such as
from an open-circuit load), output voltage is maintained at the
instrument’s maximum limit.
-25.000 mA to 25.000 mA
Page 43
10
3.3Pulse Source (PULSE)
• In addition to flow meter calibration, the pulse source can be
used for checking and calibrating counters, rotation counters,
oscilloscopes, frequency converters and other frequency-input
systems.
• Frequency, duty cycle, pulse width and amplitude can be set.
Procedure
37
3.3 Pulse Source (PULSE)
1
2
1. Set the slide switch to the [M/S] position.
2.
Set the function selector to the
[
PULSE] position.
Select one position.
When source only, any [
can be selected.
PULSE] position
3. Press the [SHIFT] button and confirm that is displayed
on the LCD.
4. Press the [MODE] button to select a parameter for setting .
Pressing the [MODE] butto n re peated ly c hange s the param eter as
shown (page 38).
3
4
5
6
7
8
9
11
Page 44
38
Selecting the Parameter
(The indicator is displayed)
Pulse source frequency
Pulse source duty cycle
Pulse source amplitude
Pulse source pulse width
Frequency source: The indicator is displayed., Duty cycle source: The indicator is displayed.
Pulse width source: The indicator is displayed., Amplitude source: The indicator is displayed.
3.3 Pulse Source (PULSE)
The values displayed above are the default settings.
As there are 256 settable steps, each step is 0.390625%,
which is displayed as 0.39% per step
increments.
Example: If an ini tial value of 5 0.00% is incremen ted by
As there are 256 settable steps, the pulse width per
step is 1/(256 ∗ frequency).
+5V, ±5 V, +12 V, ±12 V
one step, the display shows 50.39%.
6. Press the [OUTPUT] button to output the set pu lse value. Ver-
ify thatis disp lay ed on the LCD.
7. Press the [OUTPUT] button again to dis able output. Verify that
is no longer displayed.
If an overload such as a short circuit occurs while sourcing voltage,
output is disabled and the standby state is enabled ( lit), or the
meter may be turned off by the internal protection circuitry. If the
power is turned off, wait a short time before turning it back on.
Page 45
3.4 Using Memory Function (S CAN and RAMP)
10
division M division m
Start levelEnd level
Number of Number of
Example:
Number of division M: 20 divisions
Number of division m: 10 divisions
End level
3.4Using Memory Function (SCAN and
RAMP)
39
These two functions are convenient for constant-voltage and
constant-current sourcing.
SCAN
Function
RAMP
Function
* Dual slope: two slopes.
The number of divisions between the start and end levels, and
from the end level to the next start level can be set for each slope.
These functions are described next.
Up to 16 steps of diffe rent source va lues
and time intervals can be set.
A dual slope* can be generated to
simulate a linear waveform. The num ber
of divisions in each slope can be set.
❖ 3.4.1 (40 p.)
❖ 3.4.2 (45 p.)
1
2
3
4
5
6
7
8
9
11
Page 46
40
Voltage source SCAN±1.5 V
Current source SCAN±25 mA
Selecting the Sourc e Mo de
SCAN±15 V
3.4 Using Memory Function (SCAN and RAMP)
3.4.1Using the SCAN Function
Procedure
1. Set the slide switch to the [M/S] pos it ion.
2. Set the function selector to the [CV] position to source vo lt age,
or to the [
CC] position to source current.
3. Press the [SHIFT] button and confirm that is displayed
on the LCD.
4. Press the [MODE] button as necessary to select the source
mode on the sec ondary display
Pressing the [MODE] button repea tedly ch ang es the sour ce mode
as shown below.
During SCAN function setup,is display e d.
Page 47
3.4 Using Memory Function (S CAN and RAMP)
10
SCAN function factory default settings
41
RangeSCAN±1.5000 VSCAN±15.000 VSCAN±25.000 mA
StepOutput ValueOutput ValueOutput Value
1+1.5000 V+15.000 V+00.000 mA2 S
2+1.2000 V+12.000 V+04.000 mA2 S
3+0.9000 V+09.000 V+08.000 mA2 S
4+0.6000 V+06.000 V+12.000 mA2 S
5+0.3000 V+03.000 V+16.000 mA2 S
6+0.0000 V+00.000 V+20.000 mA2 S
7-0.3000 V-03.000 V+16.000 mA2 S
8-0.6000 V-06.000 V+12.000 mA2 S
9-0.9000 V-09.000 V+08.000 mA2 S
10-1.2000 V-12.000 V+04.000 mA2 S
11-1.5000 V-15.000 V+00.000 mA2 S
12+0.0000 V+00.000 V+04.000 mA0 S
13+0.0000 V+00.000 V+08.000 mA0 S
14+0.0000 V+00.000 V+12.000 mA0 S
15-1.5000 V-15.000 V+16.000 mA0 S
16+0.0000 V+00.000 V+20.000 mA0 S
Do you want to change the values
from the factory default settings?
❖ "Changing Data Settings" (44 p.)
Time
Interval
1
2
3
4
5
6
7
8
9
11
Page 48
42
Continuous (Cont)
Cycle (CyCLE)
Step (StEP )
[] []
[] []
[] []
Selecting the SCAN Source Mode
3.4 Using Memory Function (SCAN and RAMP)
5. When finished maki ng SCAN function settings, pres s the []
or [
] button to select the SCAN source mode.
Pressing the [] or [] button changes the secondary display as
follows.
Cont
• Output proceeds sequentially fr om t he first step to the last, then
returns to the first step and repeats continuously.
• However, if the time interval of the last s tep is set to 0 s econ ds ,
it is ignored, and output continues from the first step .
Example: When the factory default settings for Step 12 is 0
CyCLE
StEP
• Output proceeds sequentially fr om t he first step to the last, then
continues output at the settings of the last step.
• However, if the time i nterva l of a step i s set to 0 secon ds, out put
continues according to the settings of the previous step.
Example: When the factory default settings for Step 12 is 0
Pressing the [] or [] button moves to the previ ous or next output
step. The value set for each step is output until the next step is selected.
seconds, after ou tput from S tep 11, outp ut continue s
from Step 1.
seconds, output from Step 11 is maintained continuously.
Page 49
3.4 Using Memory Function (S CAN and RAMP)
10
Continuous
(Cont)
Cycle
(CyCLE)
Step
(StEP)
Time interval
(second)
43
1
2
3
4
6. Press the [OUTPUT] button to output the set voltage value.
Verify thatis displayed on the LCD.
7. Press the [OUTPUT] button again to disable out put. Verify that
is no longer displayed.
• Continuous (CONT) and Cycle (CyCLE) source modes always
start at Step 1.
• If the time interval of Step 1 is set to 0s, the source out put is
not activated. (Except for STEP source mode)
• When Continuous (
are stopped by the [OUTPUT] button, subsequent source output
resumes at Step 1.
CONT) and Cycle (CyCLE) source modes
5
6
7
8
9
11
Page 50
44
Step number
Output value
Time interval
Step numberTime intervalOutput value
Setting Display
Each press of button
3.4 Using Memory Function (SCAN and RAMP)
Changing Data Settings
1. With the SCAN function active, hold the [MODE] button for one
second to enter the data setting mode for each source step.
2. Set the time interval and so urce output value for each step.
The secondary display shows the output value, the leftmost
two digits of the primary disp lay show the step number, and
the rightmost two digits show the time interval.
3. Press the [MODE] button to select the step number, time
interval or output value. The param eter digit to be set blinks.
• Press the [] or [] button to increment or decrement the value,
and press the [
put value.
• Time intervals can be set from 0 to 99 seconds.
4. Press the [OUTPUT] button to save your settin gs.
5. To clear all settings, hold the [] button for one second. Time
interval and output values are cleared to zero.
6. Hold the [MODE] button for one second to exit t he step data
setting m o de.
] or [] button to move to another digit of the out-
Page 51
3.4 Using Memory Function (S CAN and RAMP)
10
Voltage source RAMP±1.5 VCurrent source RAMP±25 mA
Selecting the Source Mode
RAMP±15 V
3.4.2Using the RAMP Function
Procedure
45
1
1. Set the slide switch to the [M/S] position.
2. Set the function selector to the [CV] position to source voltage,
or to the [
CC] position to source current.
3. Press the [SHIFT] button and confirm that is displayed
on the LCD.
4. Press the [MODE] button as necessary to select the source
mode on the secondary display.
Pressing the [MODE] button repeat edly cha nge s the sou rce mo de
as shown below.
During RAMP function setup,is displayed.
2
3
4
5
6
7
8
9
11
Page 52
46
Continuous (Cont)
Cycle (CyCLE)
[
] []
[
] []
Selecting the R AMP Source Mode
3.4 Using Memory Function (SCAN and RAMP)
RAMP function factory default settings
±1.5000 V±15.000 V±25.000 mA
Output
value
Start-1.5000 V015-15.000 V015-25.000 mA025
End+1.5000 V015+15.000 V015+25.000 mA025
Number
of
division
Output
value
Do you want to change the values
from the factory default settings?
Number
of
division
Output
value
❖ "Changing Data Settings" (47 p.)
Number
of
division
5. When finished making source mode settings, press the [] or
] button to select the RAMP source mode.
[
Pressing the [] or [] button changes the secondary display as
follows.
Cont
CyCLE
• After the first slope is output (start to end values), the second
slope is output to return to the start value, forming one repeatable cycle.
• The interval of the first slope = (end output value - start output
value), divisions.
Example: The factory de fault settings w ith±1.5000 V range
is (1.5 + 1.5) / 15 steps = 0.2 V.
• The interval of the second slope = (starting output value - end
output value) / divisions.
After the first slop e is output (s tart to end va lues), the s tarting value
of the second slope is output, continuing output according to the
settings.
Page 53
47
10
Start positionStart position
End position
The output value at each interval =
(end output value - start output value) / divisions
S - XXX
No. of divisions
X 0.33s
E - XXX
No. of divisions
X 0.33s
First slope
Second slope
3.4 Using Memory Function (S CAN and RAMP)
6. Press the [OUTPUT] button to output the set voltage value.
Verify thatis displayed on the LCD.
The output value changes according to settings every 0.33
seconds.
7. Press the [OUTPUT] button again to disable out put. Verify that
is no longer displayed.
Changing Data Settings
1. With the RAMP function active, hold the [MODE] button for
one second to enter the data setting mode for each source
step.
2. The RAMP function involves dual slope generation*.
Set the number of division s from the start to the end position,
and from the end to the start pos itio n.
Also set the out put values of the start and end positions.
* Dual slope generation: generating two slopes.
1
2
3
4
5
6
7
8
9
11
Page 54
48
Start position
Output value at start position
Number of division
(1 to 999)
Start position [S] Number of division Output value at st a rt posi ti on
Output value at end position
End position Number of division
(1 to 999)
Each press of [], [] button
Setting Display
When the start position [S] or end position [
E] is blinking, press the [] or []
button.
End position [E] Number of divisionOutput value at end p o sition
Each press of button
Each press of button
3.4 Using Memory Function (SCAN and RAMP)
3. Set the division number and output value for both the start and
end positions.
Secondary display
Leftmost digit of primary display
Rightmost three digits of primary display
Output value
Start [S] or End [E] positi on
Number of divisions
Page 55
49
10
3.4 Using Memory Function (S CAN and RAMP)
4. Press the [] or [] button to select the start positio n [S] or
end position [
between the number of divisions and output value. The
parameter digit to be set blinks.
• Press the [] or [] button to increment or decrement the value,
and press the [
• The number of divisions can be set from 1 to 999.
E]. Then press the [MODE] button to switch
] or [] button to move to another digit to set.
1
2
5. Press the [OUTPUT] button to save your settings.
6. Hold the [MODE] button for one secon d to exit the step data
setting mode.
3
4
5
6
7
8
9
11
Page 56
50
3.4 Using Memory Function (SCAN and RAMP)
Page 57
10
Using Measurement
51
Functions
Observe the following precautions to avoid electric shock.
• Before connecting the test leads, check the position of the
function selector.
• Also, remove the test leads from the device under test
before switching the function selector.
• Do not attempt to m easure a television’s horizontal output
signal.
The terminal section does not provide safe insulating
distance. So to avoid electric shock, disconnect mains
power before connecting test leads to the terminals.
• Set the slide switch to [M] for measuring only, or to [M/S] if you
want to measure and use source functions at the same time
• If the source functions are not used but the slide switch is set
to [
M/S], continuous operating time per battery charge will be
shortened.
• Malfunctions may occur when measuring a voltage with har monic
elements.
• Before measurement, verify that is no longer displayed.
When is displayed, please press the [SHIFT] button to cancel
the display.
4
1
2
3
4
5
6
7
8
9
11
Page 58
52
4.1 Measuring Voltage
4.1Measuring Voltage
Procedure
1. Set the slide switch to the [M] position.
2. Set the
3. Connect the red L9207-10 TEST LEAD to the (+) INPUT
4. Press the [AC/DC] button as needed to s elect the appropri ate
The maximum input voltage is 250 VDC, 250 Vrms (with
maximum 10
to measure voltage in excess of the maximum input could
destroy the instrument and result in personal injury or death.
•Setting [mV] is for 50 mV and 500 mV range measurement.
•Setting [
terminal, and the black L9207-10 TEST LEAD to the (-) INPUT
terminal.
voltage measurement function.
6
V•Hz voltage-frequency product). Attempting
function selector
/V] is for the range over 500 mV measurement .
to the
[
/V
] or [mV] position.
5. To enable man ual range selection, press the [RANGE/AUTO]
button (so theindicator is not displayed on the LCD).
To re-enable auto- ranging, hold the [RANGE/AUTO] button for one
second.
6. Connect the test leads to the circuit to be measured, and read
the displayed value.
Page 59
10
Measuring DC voltageMeasuring AC voltage
Display Example
After pressing the and buttons
After pressing the button
4.1 Measuring Voltage
• Press the [DUAL] button to activate the Dual Display mode.
• Press the [
duty cycle or pulse width) in the Dual Display mode.
❖ Detail→ 4.8.5 "Dual Display Function" (70 p.)
Hz] button to activate frequency measurement (or
53
1
2
3
4
5
6
7
8
9
11
Page 60
54
4.2 Measuring Current
4.2Measuring Current
Never apply voltage to the test leads when a current
measurement function is selected. Doing so may damage the
instrument and result in personal injury.
To avoid electrical accid ents, remove power from the circuit
before connecting the test leads.
To avoid electrical shock, do not use the instrument to
measure current in circuits of 250 V or greater. The current
function overload protection tri ps at 250 VDC, or 250 Vrms.
Procedure
1. Set the slide switch to the [M] position.
2. Set the function selector t o th e [/mA] position.
3. Connect the red L9207-10 TEST LEAD to the (+) INPUT
terminal, and the black L9207-10 TEST LEAD to the (-) INPUT
terminal.
4. Press the [AC/DC] button as needed to s elect the appropri ate
current measurement function.
5. To enable man ual range selection, press the [RANGE/AUTO]
button (so theindicator is not displayed on the LCD).
To re-enable auto- ranging, hold the [RANGE/AUTO] button for one
second.
Page 61
55
10
DC Current
Display
Display Switching
"4-20 mA%" or "0-20 mA%"
Display
AC+DC Current Display
AC Current
Display
4.2 Measuring Current
6. Connect the te s t le ads t o th e c irc uit to be measured, and read
the displayed value.
• Press the [DUAL] button to activate the Dual Display mode.
• Press the [
duty cycle or pulse width) in the Dual Display mode.
❖ Detail→ 4.8.5 "Dual Display Function" (70 p.)
• "4-20 mA%" displays 4 mA input as 0% and 20 mA input as
100%.
• "0-20 mA%" displays 0 mA input as 0% and 20 mA input as
100%.
• You can select between "4-20 mA%" and "0-20 mA%" display
modes in the 5.4 "Power-On Options" (76 p.).
Hz] button to activate frequency measurement (or
1
2
3
4
5
6
7
8
9
11
Page 62
56
Selecting the Measurement Mode
Hz
Measurement
Duty Cycle
Measurement
Pulse
Measurement
Press the button for
one second or more.
Selecting the Trigger Slope
4.3 Measuring Frequency
4.3Measuring Frequency
Procedure
1. Display voltage or current by t he procedures in 4.1 (52 p.) or
4.2 (54 p.).
2. Press the [Hz] button and read the displayed voltage.
3. Pressing the [Hz] button repeatedly toggles through Hz, duty
cycle and pulse measurement.
4. Hold the [Hz] but ton for one secon d to return to t he status in
Step 1.
Page 63
10
4.4Measuring Resistance
57
4.4 Measuring Resistance
Procedure
1. Set the slide switch to the [M] position.
2. Set the function se lector to the [Ω] positi o n
3. Connect the red L9207-10 TEST LEAD to the (+) INPUT
4. Connect the te s t le ads t o th e c irc uit to be measured, and read
5. To enable manual range selection, press the [ RANGE/AUTO]
Never apply voltage to the input terminals when the
Resistance function is selected. Doing so may damage th e
instrument and result in personal injury.
To avoid electrical accidents, remove power from the circuit
before measuring.
terminal, and the black L9207-10 TEST LEAD to the (-) INPUT
terminal.
the displayed value.
button (so theindic at or is not dis played on the LCD)
1
2
3
4
5
6
7
8
9
To re-enable auto-ranging, hold the [RANGE/AUTO] button for
one second.
6. Pressing the [AC/DC] button toggles the continuity test
function on and off (mark is displayed/not displayed).
At this time, auto-ranging is dis abled.
11
Page 64
58
Resistance Measurement Display Example
Continuity Test Display Example
4.4 Measuring Resistance
With the resistance measurement function active, short the
probe tips together and press the [
internal resistance of the probes.
4.4.1Continuity Test
Procedure
1. When the resistance measurement function is active, pressing
the [
AC/DC] button enables the cont inuity test function using
resistance measurement.
At this point, the 500 Ω range is selected, and if auto-ranging was
active, it is disabled and manual ranging is enabled.
REL] button to zero out the
• Pressing the [RANGE] button selects each range.
• The beeper sounds whenever the resistance is 1000 counts
or less in the selected range.
2. Press the [AC/DC] button agai n to exit the continuity testing
function.
• While the continuity test function is active, holding the [RANGE/
AUTO
] button for one second activates auto-ranging. In this
case, the beeper sounds whenever the measured resistance
value is about 10 Ω or less.
• When you cancel the continuity test function, the resistance
measurement range does not revert to the range that was
active at the time the function was enabled.
Page 65
10
4.5Diode Test
Diode Test Display Example
59
4.5 Diode Test
Procedure
1. Set the slide switch to the [M] position.
2. Set the function se lector to the [] position.
3.
4. Connect the te s t le ads t o th e c irc uit to be measured, and read
5. The display ap pears as follows:
Never apply voltage to the input terminals w hen the Diode
Test function is selected. Doing so may damage the
instrument and result in personal injury.
To avoid electrical accidents, remove power from the circuit
before measuring.
Connect the red L9207-10 TEST LEAD to the (+) INPUT
terminal, and the black L9207-10 TEST LEAD to the (-) INPUT
terminal.
the displayed value.
Forward direction
Reverse dire ction
Short-circuit
Open-circuit
0.3 to 0.8 V forward voltage drop displayed
[OL] displayed
Approx. 0 V displayed, and beeper sounds
[OL] displayed in forward direction
1
2
3
4
5
6
7
8
9
The diode function also works for continuity testing. The beeper
sounds at voltage levels of about 50 mV or less.
11
Page 66
60
Temperature Measurement Disp lay Example
4.6 Measuring Temperature
4.6Measuring Temperature
Temperature can be measured using an optional temperature
probe.
• To avoid damage to the instrument, do not apply voltage or current to temperature probe.
• Do not bend the thermocouple leads, sharp ly. Repeatedly bending the leads can break them.
Measurement Procedure
1. Set the slide switch to the [M] position .
2. Set the function selector t o th e [mV ] position.
3. Connect the temperature probe to the (+) and (-) INPUT
terminals.
4. Hold the [TEMP] button for one second to start temperature
measurement.
Pressing the [AC/DC] button displays [0°C] on the LCD .
At this point, the reference junction temperature is 0°C.
Page 67
4.7 Source and Measure Simultaneously
10
Measuring voltage[/V], or [mV]
Measuring current[
/mA]
Measuring resistance[Ω]
Diode Test[]
Measuring temperature [mV]
Measuring frequency[
/V], [mV], or
[
/mA]
ex. Voltage source and Measuring
resistance
4.7Source and Measure Simultaneously
61
Procedure
1. Set the slide switch to the [M/S] position.
Set the function selector.
2.
• Voltage source:
Set to [
function position (dark labels on
white background) that will be
measured.
• Current source:
Set to [
function position (dark labels on
white background) that will be
measured.
❖ Follow the appropriate separate procedure from here.
This instrument provides the following special measurement
functions:
FunctionUseReference
Recording
Data Hold and Refresh Hold
Relative (relative dis pl ay)
1-ms Peak Hold
Dual Display
For acquiring long-term data and checking
for transients.
Use to simplify reading fluctuating
measurements.
• When you want to zero out the internal
resistance of probes
• When you want to read only the di ffere nce
(offset) of a measured value from another
value
• When you want high-accuracy
measurement in the 50 mV DC or DCA
ranges
Use to determine the crest factor of a
waveform, where
crest factor = wavefo rm peak value / true
rms value
Use to monitors two different measurement
parameters (voltage, current, frequency,
duty cycle or pulse width) of one signal
simultaneously.
❖ 4.8.1 (63 p.)
❖ 4.8.2 (66 p.)
❖ 4.8.3 (67 p.)
❖ 4.8.4 (68 p.)
❖ 4.8.5 (70 p.)
Page 69
10
4.8.1Recording Function
During measurement, maximum (MAX), minimum (MIN) and
average (AVG) values can be automatically recorded in memory.
63
4.8 Special Measurement Functions
1
In what situations
is it used?
Procedure
While measuring, hold the [HOLD/MAX
1.
MIN] button for one second.
appears on the display as the recording
function starts.
For acquiring long-term data and checking for
transients.
2. Press the [HOLD/MAX MIN] button to change the display as
follows:
Display
Indication
Real-time measurement
value
Current measuremen t value
Maximum value
Minimum value
Average value
2
3
4
5
6
7
8
9
3. Hold the [HOLD/MAX MIN] button again for one second to stop
the recording function. The indicators
disappear from the display.
11
Page 70
64
4.8 Special Measurement Functions
• The beeper sounds and the relevant value is refreshed when
a new maximum or minimum value is recorded.
• When an overload value is measured, calculation of the
average value stops, and [
value.
• When the recording function is started with auto-ranging
enabled, the maximum (MAX), minimum (MIN) and average
(AVG) values can be in different ranges.
When the recording function is active, the power saver function
is disabled, and does no t appear on the display.
The average value is the average of all recorded measurements
since recording was started.
OL] is displayed for the average
Page 71
10
Recording Display Example
65
4.8 Special Measurement Functions
1
2
3
4
5
6
7
8
9
11
Page 72
66
Data Hold Function
Refresh Hold Function
4.8 Special Measurement Functions
4.8.2Data Hold / Refresh Hold Function
In what situations
is it used?
This function holds measured values on the display.
Use to simplify reading fluctuating measurements.
Procedure
Press the [HOLD/MAX MIN] button to
1.
activate the Data Hold function.
appears on the display, and the displayed
value is held (retained).
2. Press the [HOLD/MAX MI N ] but ton again to measure on e time
and hold the measurement data.
3. Hold the [HOLD/MAX MIN] but t on for one second to cancel the
Data Hold function . disappears from the display.
• This function automatically refreshes the (held) displayed value
when the measured value changes while Data Hold is active.
• The Data Hold function can be changed to the Refresh Hold
function in the Power-On Options.
❖ 5.4 "Power-On Options" (76 p.)
• If a change occurs that is greater than the count value (100 to
1000) set in the Power-On Options, the value is refreshed (while
is blinking).
Procedure
1. Press the [HOLD/MAX MIN] button to activate the Refresh
Hold function. appears on the display.
2. Press the [HOLD/MAX MIN] button for one second to cancel
the Refresh Hold function. disappears from the display.
Page 73
4.8 Special Measurement Functions
10
Zeroing Out Probe Resist ance
4.8.3Relative (relative display) Function
The function applies to all measurement functions, subtracting a
reference value from the measured value and displaying the
calculated result.
67
1
In what situations
is it used?
Procedure
1. Press the [REL] button to activate the Relative function.
2. Press the [REL] button again to cancel the Relat ive function.
• When you want to zero out the internal resistance
of probes
• When you want to read only the difference (offset)
of a measured value from another value
• When you want high-accuracy measurement in the
50 mV DC or DCA ranges
appears on the display.
At this point,
• The displayed value becomes the reference for subsequent
measurements, and is redisplayed as zero.
• If measured data is at overload le vel, [
disappears from the disp lay.
• The Relative function cannot be activated if an overload is
present ([
• With the resistance measurement function active, short the
probe tips together and press the [
internal resistance of the probes.
OL] displayed).
OL] appears on t he d isp lay.
REL] button to zero out the
2
3
4
5
6
7
8
9
11
Page 74
68
Display
indication
Measurement value
Maximum value
Minimum value
4.8 Special Measurement Functions
4.8.41-ms Peak Hold Function
• Works with voltage and current measurement functions
• Shows the peak value of an input signal waveform
Use to determine the crest factor of a waveform,
In what situations
is it used?
Procedure
Hold the [AC/DC] button for one second to
1.
activate the 1-ms Peak Hold function.
appears on the display.
where
crest factor = waveform peak value / true rms
value
At this point,
• The measurement range displayed when the function was
activated is held (retained).
• If auto-ranging was enabled, manual rangi ng is activate d with the
currently operative range retained.
• Pressing the [HOLD/MAX MIN] button updates the displayed
(held) MAX and MIN values.
• The beeper sounds when a new maximum or minimum
value is measured.
• Pressing the [
and restarts 1-ms Peak Hold operation.
The maximum (MAX) and minimum (MIN) display mode are
not changed.
DUAL] button resets the MAX and MIN values
Page 75
4.8 Special Measurement Functions
10
1-ms Peak Hold Function
Display State Switching
(function disabled state)
Hold for one second
(function enabled state)
Maximum value display
(function enabled state)
Minimum value display
Press the [AC/DC] button for one second
2.
to cancel the 1-ms Peak Hold function.
disappears from the displa y.
69
1
2
3
4
5
6
7
8
9
11
Page 76
70
4.8 Special Measurement Functions
4.8.5Dual Display Function
This function simultaneously monitors two different measurement
parameters (voltage, current, frequency, duty cycle or pulse width)
of one signal.
Procedure
1. Press the [DUAL] or [Hz] button to activate dual display.
(1) Pressing the [DUAL] button selects the Dual Display mode
Pressing it repeatedly changes the display as follows.
Function
Selector Position
/V
mV
mA
Primary DisplaySecondary Display
Frequency
Voltage measurement (V)
Voltage measurem ent (m V)
Current measurement (mA)
measurement
Source function display
Frequency
measurement
Source function display
Frequency
measurement
Source function display
Page 77
71
10
4.8 Special Measurement Functions
(2) Press the [Hz] button to activate frequency measurement (or
duty cycle or pulse width) in the Dual Display mode.
Pressing it repeatedly changes the display as follows.
Function
Selector Position
/V
mV
mA
• When measuring frequency or pulse width, pressing the
[
RANGE/AUTO] button toggles between auto- and manual
ranging.
• When measuring duty cycle or pulse width, holding the [
• When pressed [Hz] button at procedure 1. : Hold the [Hz] button
for one second.
(V)
(mV)
(mA)
1
2
3
4
5
6
AC/
7
8
9
11
Page 78
72
Display Example
After pressing the buttons
After pressing the button
4.8 Special Measurement Functions
Page 79
5.1 Auto Power Off Function
10
Functions5
5.1Auto Power Off Function
The Auto Power Off function is provided to minimize battery
consumption.
This function is automatically enabled when the meter is turned on
( appears on the display).
73
1
2
3
• During operation, if one of the following actions is not
performed within a specific period, the meter turns off
automatically.
1.Any key is pressed
2.The recording function is active
3.The 1-m s Peak H old function is active
4. is displayed with Source mode
5. If the Auto Power Off function has been disabled in the
Power-On Options
• To turn power back on after it has been turned off, move the
function selector to the [OFF] position, wait for one second,
then turn it back on.
When power is turned off by Auto Power Off, all settings are
reset.
When you will be using the meter for a long period, disable Auto
Power Off (so that does not appear on the display)
Disabling Auto Power Off
Timing and disabling of the Auto Power Off function is set in the
Power-On Options.
❖ 5.4 "Power-On Options" (76 p.)
To disable the function, set the time to [OFF m].
4
5
6
7
8
9
11
Page 80
74
5.2 Backlight Function
5.2Backlight Function
The LCD backlight can be enabled or disabled.
Procedure
1. Hold the [SHIFT] button for one second to turn on the LCD
backlight.
Once the backlight is turned on, it turns off automatically after a
preset time (set by the Backlight Auto-Off function).
2. Hold the [SHIFT] button for one second if you want to manually
turn off the LCD backlight.
Disabling the Backlight Auto-Off Function
Timing and disabling of the Backlight Auto-Off function are set in
the Power-On Options.
❖ 5.4 "Power-On Options" (76 p.)
To disable the function, set the time to [OFF sec].
Page 81
5.3 RS-232C Communication Functions
10
Connect to USB
connector on the PC
Connection band
5.3RS-232C Communication Functions
The meter includes a data transfer function using an RS-232C
interface, by which the meter can be connected to a PC to transfer
measurement data for collection and storage.
To use this function, the optional 3856-02 COMMUNICATION
PACKAGE (USB)
set, is required.
, a special communication cable and software
75
1
2
Transferring Data to a PC
1. The optical-connector end of the communication cable
connects to the socket on the back of the meter.
At this time, the "RS-232C INTERFACE" label should face
upward.
Connect the USB connector
2.
on the cable to a USB
interfa c e on the PC.
3. Install the supplied driv er.
3
4
5
6
7
8
9
4. Run the software to activate data acquisition.
• Refer to the instruction manual provided with the 3856-02 for
software usage procedures.
• When operating with the case cover folded back, install the
supplied connection band to lessen the chance of tipping over.
11
Page 82
76
5.4 Power-On Options
5.4Power-On Options
These settings are retained even when power is turned off and the
batteries removed.
The various settings are described in the table on the next page.
Setting Procedure
1. Turn the meter off (if on).
2. Hold the [AC/DC] button while turning the meter on (set the
function selector to any posit ion).
3. Press the [] or [] button to select the item y ou w ant to set.
4. Press the [] or [] button to change the setting.
5. Press the [Hz] button to accept your setting.
(The selected item ceases blinking and stays lit)
6. Press the [SHIFT] button. All LCD segments appear and
progress through the Power-On Options setting displays.
To set multiple items, repeat Steps 3 to 5 above.
If you change an item without pressing the [Hz] button, or if you
hold the [Hz] button for more than one second, the selected
setting remains unchanged.
Page 83
10
Power-On Options
77
5.4 Power-On Options
ItemDisplay
Echo
Data length
Parity
Baud Rate
Backlight
Auto power off
Refresh hold
Beep sound
Minimum Measur e-
ment Frequency
Default
(factory setting)
OFF
8 bit
None
9600 bps
30 sec
15 m
OFF
4800 Hz
0.5 Hz
Selection Contents
OFF/ON
With ECHO ON, the meter echoes
(returns) all the characters whatever it receives.
8bit/7bit
None/odd/EN (even)
2400 bps/4800 bps/9600 bps/19200
bps
The display shows [
bps].
[
OFF sec/01 to 99 sec
Setting OFF disables the Backlig ht
Auto Off function
OFF m/01 to 99 m
• Setting OFF disables the Auto
Power-Off function
• "m" means "minute".
OFF/100/200/300/400/500/600/700/
800/900/1000
Setting OFF enables Data Hold
OFF/600 Hz/1200 Hz/2400 Hz/4800
Hz
0.5 Hz/1 Hz/2 Hz
Hz] in place of
1
2
3
4
5
6
7
8
DCmA% Display
Printer
4-20 mA
OFF
4-20 mA/0-20 mA
OFF/ON
When turned ON, data f or every sample is output to the serial interface.
When turned ON, appears
blinking, and o ther serial commun ications are disabled.
9
11
Page 84
78
5.4 Power-On Options
Page 85
6.1 Transducer Calibration
10
Applications6
6.1Transducer Calibration
Transducers that accept pulse input and provide current output can
be calibrated.
79
1
2
Procedure
1. Set the slide switch to the [M/S] position.
Set the function selector to the
2.
[
PULSE] and [/mA] position.
3. Press the [SHIFT] button and confirm that is displayed
on the LCD.
4. Press the [MODE] button, and set the duty cycle, pulse width
and frequency.
3
4
5
6
7
8
9
5. While paying at tention to polarity, connect the meter ’s output
terminals to the transdu cer’s input te rminals, and the me ter’s
input terminals to the transducer’s output.
6. Press the [OUTPUT] button to begin pulse generation.
(During the sourcing, is displayed on the LCD.)
11
Page 86
80
6.1 Transducer Calibration
7. Read the displayed valu e
Verify that the frequency and current values meet the
transducer’s specifications.
8. Change the fr equency and observe the cur rent value.
Page 87
10
6.2Loop Test
Note which terminal has the higher voltage.
L9207-10
TEST LEAD (black)
L9207-10
TEST LEAD (red)
Device under test
(with internal power)
The meter’s constant current source function can be used to check
4- to 20-mA measurement devices.
Wiring Preparation
Set the slide switch to [M/S] and set the
1.
function selector as shown in the illustration.
2. Using the voltage measurement function of the meter (see 4.1
(52 p.)), measure the (loop) voltage between the input
terminals of the device under test.
81
6.2 Loop Test
1
2
3
4
5
6
7
8
9
11
Page 88
82
Test lead
(yellow)
Wiring
Test lead (black)
Lower-voltage terminal
Direction of
current flow
Direction of
current flow
Higher-voltage terminal
Device under test
(with internal power)
Wiring
6.2 Loop Test
1. Connect the black test lead from the negative OUTPUT
terminal of the meter to the higher-voltage terminal of the
device under test.
2. Connect the yellow test lead from the positive OUTPUT
terminal of the meter to the lower-voltage terminal of the
device under test.
3. Press the [SHIFT] button and confirm that is displayed
on the LCD.
Use the supplied red test lead when testing measurement
instruments that are not internally powered. The set current will
not flow if the yellow test lead is used.
Page 89
10
Use the cursor button s [///] to set
4.
the current to be sourced. The setting
value should have positive polarity.
Press the [
current.
(During the sourcing, is displayed
on the LCD.)
OUTPUT] button to output the
83
6.2 Loop Test
1
2
When the input impedance of the device under test is low
(approx. below 140 Ω and a voltage of 24 V), the internal
protection circuitry may prevent from being displayed even
when the [OUTPUT] button is pressed.
When this happens, insert a series resistance of approx. 100 200 Ω between the instrument and the device under test.
3
4
5
6
7
8
9
11
Page 90
84
6.2 Loop Test
Page 91
7.1 Product Specifications
10
Specifications7
7.1Product Specifications
Output methodBipolar sink/source output
Measurement methodDouble integral system
AC Measurement
method
Functions
Additional functions
SwitchesSlide switch, Function selector
Range selectionFull auto or manual
Display deviceLCD with backlight
Display contents
Maximum
measurement count
Polarity indicator"-" symbol displayed automatically
Auto power offSettable (0 to 99 minutes in 1-minute intervals)
Low battery indication Warning on LCD when battery voltage falls below 9 V
Sampling rate
Temperature
characteristic
Noise susceptibi lity
True RMS
DC constant voltag e (DCCV) sour ce, DC constant current (DC CC) source,
Pulse (PULSE) source, DC voltage (DCV) measurement,
AC voltage (ACV) measurement, AC + DC voltage (ACDCV)
measurement,
DC current (DCA) measurement, AC current (ACA) measurement,
AC + DC current (ACDCA) measurement, Resistance (OHM)
measurement,
Diode testing (DIODE), Continuity testing (CONT),
Temperature (TEMP) measurement, Frequency (FREQ) measurement,
Duty cycle (DUTY) measurement, Pulse width (WIDTH) measurement
• Settable duty cycle, pulse width and amplitude pulse generation
• Mem ory generation (16 memor y data settings per range); s can generation
(single/continuous)
• Ramp generation
• AC+DC RMS voltage/current measurement
• 1-ms Peak Hold (for voltage/current measurement)
• Recording
• Data Hold/Refresh Hold
• Relative display
• 4 - 20 mA current-loop percentage display, 0 - 20 mA percentage display
• RS-232C data communications
• Power-on options
Two 5-digit numeric digital displays (for source and measurement
functions), one large primary display and one small secondary display)
51,000 counts
Approx. 3/s (except AC+DC and frequency measurement)
Approx. 1/s for AC+DC and frequency measurement
Approx. 0.25 - 4/s for duty cycle and pulse width measurement
Add the following except within 23±5°C
Source functions: ±(50ppm of setting + 0.5dgt)/°C
Measurement functi ons: ±(measurement accuracy × 0.15)/°C
NMRR >60 dB (50/60 Hz)
CMRR >90 dB (50/60 Hz)
85
1
2
3
4
5
6
7
8
9
11
Page 92
86
7.1 Product Specifications
Over load protection
DCCV, DCCC, PULSE: ±30 VDC, fuse protected (250VT63 mAL fuse)
DCV, ACV, ACDCV : 250 VDC/250 Vrms at 10
voltage-frequency product
DCA, ACA, ACDCA : fuse protected (250VF630 mAH fuse)
OHM, DIODE, CONT, TEMP, FREQ: 250 Vrms at 10
6
V ・ Hz maximum
6
V ・ Hz maximum
voltage-frequency product
Maximum rated voltage
and current
• Maximum output voltage and current: ±15 V, ±25 mA
• Maximum input voltage and current: 250 V, 500 mA
• Maximum voltage to earth of the input terminal: 250 V
• Between case and combined powe r supply terminals: 510-Vrms 50/60-
Hz sine wave for one minute
• Between case and combined output ter minal s: 510- Vrms 50/ 60-H z sin e
wave for one minute
• Between combined input terminals and co mbined ou tpu t termi nals: 51 0-
Withstand voltage
Vrms 50/60-Hz sine wave for one minu te
• Between case and combined input terminals: 2.3-kVrms 50/60-Hz sine
wave for one minute
• Between combined power supply terminals and combined input
terminals: 2.3-kVrms 50/60-Hz sine wave for one minute
• Between combined output terminals and combined input terminals: 2.3-
kVrms 50/60-Hz sine wave for one minute
Operating temperatur e
& humidity
Storage temperature &
humidity
0 to 40°C, 0 to 80%RH (non-condensation)
However, above 30°C , humidity is derated to 50% RH at 40°C
-20 to 60°C, 0 to 80%RH (non-condensation, w/o batteries)
Operating locationIndoors, below 2000m ASL
Period of guaranteed
3856-02 COMMUNICATION PACKAGE (USB)
9180 SHEATH TYPE TEMPERATURE PROBE
9181 SURFACE TEMPERATURE PROBE
9182 SHEATH TYPE TEMPERATURE PROBE (Class 1)
9183 SHEATH TYPE TEMPERATURE PROBE (Class 1)
9472 SHEATH TYPE TEMPERATURE PROBE (Class 1)
9473 SHEATH TYPE TEMPERATURE PROBE (Class 1)
9474 SHEATH TYPE TEMPERATURE PROBE (Class 1)
9475 SHEATH TYPE TEMPERATURE PROBE (Class 1)
9476 SURFACE TEMPERATURE PROBE
Safety : EN61010
Pollution degree 2
Measurement category II (anticipated transient overvoltage
2,500V)
EMC: EN61326
EN61000-3-2
EN61000-3-3
87
1
2
3
4
5
6
7
8
9
11
Page 94
88
7.2 Accuracy
7.2Accuracy
7.2.1Source Range and Accuracy
Accuracy is specified at 23 ± 5°C, and 80%RH or less, after at
least five minutes warm-up.
DC const ant voltage (DCCV) source Fun ction
Range Output Ra nge ResolutionAccuracyRemarks
1.5000 V 0 to ±1.5000 V0.1 mV
• Load regulation = 0.012 mV/mA
• Maximum applied voltage = ±30 VDC
DC constant current (DCCC) source Function
Range Output Ra nge ResolutionAccuracyRemarks
25.000 mA
0 to ±25.000 mA
1 μA
±(0.03% of
setting+3 dgt.)
±(0.03% of
setting+5 dgt.)
Sink/source output
At least ±25mA max.
output15.000 V 0 to ±15.000 V1 mV
Sink/source output
At least ±12 V output
• Load regulation = 1 μA/V
• Maximum applied voltage = ±30 VDC
Pulse (PULSE) source Function
Setting Output RangeResolutionAccuracy
0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40,
Frequency
Duty0.39% ~ 99.60%0.390625 %
Pulse width
Output
voltage
50, 60, 75, 80, 100, 120, 150, 200,
240, 300, 400, 480, 600, 800, 1200,
1600, 2400, 4800 Hz
(0.39% ~ 99.60%)
× 1/frequency
5 V, 12 V-
±5 V, ±12 V-
-
1/(256 ×
Frequency)
• Duty cycle and pulse width must be set with at least 50 μs positive pulse width.
• Duty cycle and pulse width output voltage accuracy specification presumes at
least 50 μs positive pulse width.
• Maximum applied voltage is ±30 VDC.
±(0.005% of
setting+0.01 Hz)
±(0.01% of
setting+0.2%)
However, at 1 kHz
and above, add
±0.1% /kHz of
setting.
±(0.01% of
setting+0.3 ms)
±(2% of
setting+0.2 V)
±(2% of
setting+0.4 V)
Page 95
10
7.2.2Measurement Range and Accuracy
Accuracy is specified at 23 ± 5°C, and 80%RH or less, after at
least five minutes warm-up.
Accuracy of duty cycle a n d pu lse wi d th me a su rement specifications presume 5 V
pulses in the 5 V DC range.
DUTY CYCLE
Measurement range : 0.1% to 99.9% (at DC coupling)
Accuracy: < ±(0.3% / kHz+0.3% of f.s.)
PULSE WIDTH
Measurement range : 0.1 ms to 1999 ms
Accuracy: < ±(0.2% of rdg.+3 dgt.)
The specified me as ur ement range is determined by the signal f re quency when
the pulse width is greater than 10 μs.
(rms sine wave)
1 Hz to 100 kHz>100 kH z< 20 kHz 20 kHz to 200 kHz
5% to 95% (at AC coupling)
Trigger level (DC coupling)
3
4
5
6
7
8
9
(Current Measurement Sensitivity)
Minimum Sensitivity
Input range
50 mA2.5 mA
500 mA25 mA
(rms sine wave)
30 Hz to 20 kHz
11
Page 100
94
7.2 Accuracy
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.