The source-related button functions are now enabled.
4.Press the [MODE] button as necessary to select
the source mode or the parameter on the secondary display.
5.Making the settings with the [W], [X], [S], or [T]
button.
Press the [OUTPUT] button and output the setting
value. Verify thatis
displayed on the LCD.
(Ex. Voltage Source)
7.Press the [OUTPUT] button again to disable out-
put. Verify that is no longer displayed.
6
1.2 Constant-Voltage Source (DCCV)
1.2Constant-Voltage Source (DCCV)
This mode outputs a constant current at the set
value regardless of load.
Procedure
1.
2.
3.
Set the slide switch.
Set the function selector.
Press the [SHIFT] button.
[M/S]
[CV]
Confirm that
is displayed.
Source Mode
• Constant-voltage
source 1.5 V range
• Constant-voltage
Press the [
4.
essary to select the source mode
on the secondary display.
MODE
] button as nec-
source 15 V range
• Constant-voltage
SCAN source 1.5 V
range
• Constant-voltage
SCAN source 15 V
range
• Constant-voltage
RAMP source 1.5
V range
• Constant-voltage
RAMP source 15 V
range
1.2 Constant-Voltage Source (DCCV)
7
5.
6.
Select a digit to change.
Set a value.
Settanble Ranges
[+0.0000 V]
(1.5 V range)
[+00.000 V]
(15 V range)
With the [W] or [X]
button
With the [S] or [T]
button
-1.5000 V to 1.5000 V
-15.000 V to 15.000 V
7.
8.
Press the [OUTPUT] button to output the set voltage value.
Press the [OUTPUT] button again
to disable output.
Verify that is
displayed.
Verify that
is no longer displayed.
8
1.3 Constant-Current Source (DCCC)
1.3Constant-Current Source (DCCC)
This mode outputs a constant current at the set
value regardless of load.
Procedure
1.
2.
3.
4.
Set the slide switch[M/S]
Set the function selector.[CC]
Press the [SHIFT] button.
Press the [
essary to select the source mode
on the secondary display.
MODE
] button as nec-
Confirm that
is displayed
Source Mode
• Constant-current
source
• Constant-current
5.
Select a digit to change.With the [W] or [X]
Set current value.With the [S] or [T]
6.
SCAN source
• Constant-current
RAMP source
button
button
1.3 Constant-Current Source (DCCC)
Settable range
[
+00.000 mA
(25 mA range)
9
]
-25.000 mA to 25.000 mA
7.
8.
Press the [OUTPUT] button to output the set current value.
Press the [OUTPUT] button again
to disable output.
Verify that is
displayed.
Verify that is
no longer displayed.
10
1.4 Pulse Source (PULSE)
1.4Pulse Source (PULSE)
• In addition to flow meter calibration, the pulse
source can be used for checking and calibrating
counters, rotation counters, oscilloscopes, frequency converters and other frequency-input
systems.
• Frequency, duty cycle, pulse width and amplitude
can be set.
Procedure
1.
2.
3.
4.
Set the slide switch.[M/S]
Set the function selector.[PULSE]
Press the [SHIFT] button.
Press the [MODE] button to
select a parameter for setting.