Hioki 3931 Instruction Manual

INSTRUCTION MANUAL
3931
HIGH VOLTAGE
CONTACT CHECKER
Contents
Introduction.................................................................................1
Inspection ...................................................................................1
Safety Notes...............................................................................2
Usage Notes...............................................................................4
1.1 Product Overview..........................................................7
1.2 Principle ........................................................................8
1.3 Parts Names ...............................................................10
1.3.1 Front Panel ....................................................................10
1.3.2 Rear Panel ....................................................................12
1.3.3 H.V. TEST LEADs (9615-01, 9615-03) .........................13
i
Contents
1.4 External Dimensions ...................................................14
1.5 Measurement Flowchart..............................................15
Chapter 2 Testing Arrangements _____________ 17
2.1 Connecting the Protective Conductor Terminal ..........17
2.2 Connecting the Power Cord ........................................18
2.3 Connecting the External I/O Terminals .......................19
2.4 Connecting the H.V. TEST LEADs to the 3931...........20
2.5 Connecting the H.V. TEST LEADs to the Device to be Tested22
2.6 Powering On and Off the 3931....................................24
2.7 Startup Inspection .......................................................25
2.7.1 Cutoff Voltage ...............................................................25
2.7.2 Analog Voltmeter ...........................................................27
Chapter 3 Testing Method ___________________ 29
3.1 Test Startup Flow ........................................................30
3.1.1 Setting the Threshold Voltage .......................................31
3.2 Starting a Test.............................................................35
3.3 PASS or FAIL Determination ......................................37
3.3.1 PASS State ...................................................................37
3.3.2 FAIL State .....................................................................40
ii
Contents
Chapter 4 External Interface__________________ 43
4.1 External I/O Terminals................................................ 43
4.1.1 Signal Line ....................................................................44
4.1.2 Timing Chart of External I/O Signals ............................45
4.1.3 Example of Input Signal Connection ............................. 46
4.1.4 Example of Output Signal Connection .......................... 47
4.1.5 Example of Connection to the WITHSTANDING
VOLTAGE HiTESTERs ................................................48
Chapter 5 Specifications ____________________55
5.1 Basic Specifications.................................................... 55
5.1.1 Interface ........................................................................57
5.2 General Specifications................................................ 58
5.3 H.V. TEST LEADs (9615-01, 9615-03)
Specifications ............................................................. 59
Chapter 6 Maintenance and Service ___________61
6.1 Cleaning and Storage................................................. 61
6.2 Repair and Servicing .................................................. 61

Introduction

Warranty

Inspection

1
Introduction
Thank you for purchasing the HIOKI “3931 HIGH VOLTAGE CON­TACT CHECKER”. To obtain maximum performance from the product, please read this manual first, and keep it handy for future reference.
HOKI cannot be responsible for losses caused either directly or indi­rectly by the use of the 3931 with other equipment, or if ownership is transferred to a third party.
• When you receive the product, inspect it carefully to ensure that no dam­age occurred during shipping. In particular, check the accessories and con­nectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
• Use the original packing materials when reshipping the product, if possible.
Checking the contents of the package
Power cord (1)
3931 HIGH VOLTAGE CONTACT CHECKER (1)
Stopper plate (1)
(red)
9615-01 H.V. TEST LEAD (high voltage side, 1)
(black)
9615-03 H.V. TEST LEAD (low voltage, 1)
Instruction manual (1)
2

Safety Notes

Safety Notes
Safety Symbols
This product is designed to conform to IEC 61010 Safety Stan­dards, and has been thoroughly tested for safety prior to ship­ment. However, mishandling during use could result in injury or death, as well as damage to the product. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or inju­ries not resulting directly from product defects.
This manual contains information and warnings essential for safe operation of the product and for maintaining it in safe operating condi­tion. Before using the product, be sure to carefully read the following safety notes.
In the manual, the symbol indicates particularly important information that the user should read before using the product.
The symbol printed on the product indicates that the user should refer to a corresponding topic in the manual (marked
with the symbol) before using the relevant function.
Indicates that dangerous voltage may be present at this termi­nal.
Indicates a grounding terminal.
Indicates a protective conductor terminal.
Indicates AC (Alternating Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the product.
Advisory items related to performance or correct operation of the product.
Other Symbols
3
Safety Notes
Indicates the prohibited action.
Accuracy
We define measurement tolerances in terms of f.s. (full scale) and dgt. (digit) values, with the following meanings:
f.s. (maximum display value or scale length)
The maximum displayable value or the full length of the scale. This is usually the maximum value of the currently selected range.
dgt. (resolution)
The smallest displayable unit on a digital measuring product, i.e., the input value that causes the digital display to show a "1".
Overvoltage categories (CAT)
The input section conforms to CAT I (5000 V) safety requirements, and the power supply section conforms to CAT II (300 V) requirements. To ensure safe operation of measurement products, IEC 60664 estab­lishes safety standards for various electrical environments, catego­rized as CAT I to CAT IV, and called overvoltage categories. These are defined as follows.
Indicates the reference.
CAT I
CAT II
CAT III
CAT IV
Secondary electrical circuits connected to an AC electrical outlet through a transformer or similar device.
Primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the dis­tribution panel to outlets.
The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel).
Higher-numbered categories correspond to electrical environments with greater momentary energy, so a measurement product designed for CAT III environments can endure greater momentary energy than one designed for CAT II. Using a measurement product in an environ­ment designated with a higher-numbered category than that for which the product is rated could result in a severe accident, and must be carefully avoided.
4

Usage Notes

Usage Notes
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
To avoid electric shock
• Do not remove the product's case. The internal components of the product carry high voltages and may become very hot during operation.
• Do not touch the alligator clip while high voltage is being gen­erated. The vinyl sheath on the alligator clip of the H.V. TEST LEADs (9615-01, 9615-03) does not have an insulation with­stand voltage.
Check the voltage.
100 to 240 V AC
• To ensure the safety of the operator first, connect the protec­tive conductor terminal to a grounded conductor before mak­ing any other connections.
• To avoid electric shock and ensure safe operation, connect the power cord to a grounded (3-contact) outlet.
• Before turning the product on, make sure the source voltage matches that indicated on the product's power connector. Connection to an improper supply voltage may damage the product and present an electrical hazard.
• Turn off the power to all devices before plugging or unplugging any of the external I/O terminals.
• Make sure the power is turned off before connecting or disconnect­ing the power cord.
• If the protective functions of the product are damaged, either remove it from service or mark it clearly so that others do not use it inadvertently.
Setting up the 3931
• To avoid damage to the product, do not allow the product to
Do not get wet.
get wet, and do not use it when your hands are wet.
5
Usage Notes
Corrosive or explosive gases
Avoid the following:
Direct sunlight
High temperature High humidity
Electomagnetic radiation
Dust
• Do not use the product where it may be exposed to corrosive or combustible gases. The product may be damaged or cause an explosion.
• Do not store or use the product where it could be exposed to direct sunlight, high temperature or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate so that it no longer meets specifications.
• Do not use the product near a device that generates a strong elec­tromagnetic field or electrostatic charge, as these may cause erro­neous measurements.
• To avoid damage to the product, protect it from vibration or shock during transport and handling, and be especially careful to avoid dropping.
• This product is not designed to be entirely water- or dust-proof. To avoid damage, do not use it in a wet or dusty environment.
• This product should be installed and operated indoors only, between 0 and 40
°
C and 35 to 80% RH.
• Do not install the instrument upside-down, or stand it on its side.
6
Usage Notes
Handling the cords
• To avoid damaging the power cord, grasp the plug, not the cord, when unplugging the cord from the power outlet.
• To avoid damaging the H.V. TEST LEADs, do not bend or pull the test leads.
• Avoid stepping on or pinching the cable, which could damage the cable insulation.
• For safety reasons, when taking measurements, only use the H.V. TEST LEADs (9615-01, 9615-03) provided with the product.
Preliminary checks
Before using the product the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Before using the product, make sure that the insulation on the H.V TEST LEADs is undamaged and that no bare conductors are improperly exposed. Using the product under such conditions could result in electrocution. Replace the H.V TEST LEADs with the specified Hioki Model 9615-01, 9615-03.

1.1 Product Overview

7

Overview

1.1 Product Overview
The 3931 HIGH VOLTAGE CONTACT CHECKER is a contact checker used for the Hioki WITHSTANDING VOLTAGE HiTESTER with its LOW side grounded. Do not use the 3931 for any other pur­pose. The 3931 is not designed to serve as an isolation voltmeter.
(1) Monitoring the measurement terminal voltage
Because it monitors the voltage of the measurement terminal, the 3931 can detect a broken wire or contact failure during a withstand voltage test. As a re­sult, this checker does not require time only for a contact check and will im­prove the reliability of the test without increasing the tact time (minimum withstand voltage test time: 1 sec). The 3931 also detects a drop in test volt­age arising from fluctuations in the supply voltage.
Chapter 1
(2) Direct connection with the Hioki WITHSTANDING VOLTAGE
HiTESTER series
The 3931 is directly connectable to the Hioki WITHSTANDING VOLTAGE HiTESTER (3158, 3173, 3159, and 3153). These HiTESTERs require no set­ting, except that during use of the 3153, "Do not include the TEST blinking time in the
TEST
-signal ON output" must be selected using the optional function. If a broken wire or contact failure is detected, the error indication LED of the 3931 lights up, the buzzer sounds to indicate the occurrence of a failure, and the operation of the connected HiTESTER is stopped. The 3931 is not designed for insulation resistance testing and cannot be used with the continuous test function or the program test function of the 3159 and
3153. In addition, the 3931 is designed exclusively for AC withstand voltage testing and cannot be used for DC withstand voltage testing.
(3) Analog voltmeter
The measurement terminal voltage is checked on the analog voltmeter (the generated voltage is checked on the voltmeter of the WITHSTANDING VOLT­AGE HiTESTER). Switch over the meter to set the criterion voltage (threshold voltage).
(4) Warning on the application of a dangerous voltage
During the contact check (voltage monitoring), the DANGER lamp remains ON to indicate the application of a dangerous voltage. If a dangerous voltage is input while the contact check is not being performed, the DANGER lamp will start blinking (50 V to 100 V).
8

1.2 Principle

1.2 Principle
WITHSTANDING VOLTAGE HiTESTER
Voltage output terminals of the WITHSTANDING VOLTAGE HiTESTER
a
(contact checker)
d
Voltage-monitoring
c
b
Test current
Object to be tested
Voltmete r
current
h
Terminals of the voltmeter
e
f
g
Voltage application terminals of the object to be tested
9
1.2 Principle
The voltmeter (contact checker) shown in the diagram above is gener­ally not used for a withstand voltage test, and the output voltage between a and h of the WITHSTANDING VOLTAGE HiTESTER is applied to voltage application terminals b and g of the object to be tested. The test current flows from a b g h. The WITHSTAND­ING VOLTAGE HiTESTER monitors this current to determine whether the object to be tested has sufficient voltage withstand capability. However, if no current flows, the tester considers the object to have sufficient voltage withstand capability. Thus, if no current flows because a wire is broken somewhere in the current path of a → b → g h, the test is judged to have been passed.
To solve this problem, a voltmeter (contact checker) is connected to the voltage application terminals of the device to be tested, as in the WITHSTANDING VOLTAGE HiTESTER shown in the diagram. The voltmeter (contact checker) checks to determine whether the voltage output from the WITHSTANDING VOLTAGE HiTESTER is applied to the voltage application terminals. Assume that a wire is broken somewhere between a and b or g and h, and that the correct voltage is not applied to the voltage application terminal. In the event of a contact failure, the correct voltage is not applied to the voltage application terminal. The output voltage from the WITHSTANDING VOLTAGE HiTESTER (hereinafter referred to as the "threshold voltage") is compared with the voltage applied to the voltage application terminals. If the voltage applied to a voltage application terminal is less than the threshold volt­age, a broken wire or contact failure has occurred. If a wire is broken in the wiring while a voltmeter (contact checker) c and d or e and f, a failure is also detected. A broken wire between c and d or e and f does not directly affect the test. This is detected as a failure of the testing equipment.
An electric current also flows to the voltmeter (contact checker). Such a current may affect the test. The threshold current during the with­stand voltage test is normally approximately 0.1 mA to several tens of mA. If the current flowing to the voltmeter (contact checker) is no more than several tens of µA, it is considered to be low enough not to affect the withstand voltage test. Even if the current does affect the test, the current flowing to the voltmeter (contact checker) can always be esti­mated and added to the threshold current of the WITHSTANDING VOLTAGE HiTESTER.
When the WITHSTANDING VOLTAGE HiTESTER is of the supply­voltage boosting type, the output of which is not stabilized, the output voltage varies with fluctuations in the commercial supply voltage. As a voltmeter (contact checker) monitors the voltage of the voltage appli­cation terminals of the device to be tested, it detects an abnormal volt­age due to these voltage fluctuations, thereby further increasing the reliability of the test. In addition, as the voltmeter also conducts a contact check during the withstand voltage test, it does not require time only for a contact check and therefore will improve the reliability of the test without increasing the tact time on the production line.
10

1.3 Parts Names

1.3 Parts Names

1.3.1 Front Panel

RESET (STOP) switch
COMP LEVEL switch
To prevent electric shock, when the touch the input-voltage terminals (HIGH and LOW), the H.V. TEST LEADs, or the tested object.
DANGER
Analog voltmeter
lamp is lit, never
Power switch
Threshold adjustment knob
DANGER lamp
LOW terminal
HIGH terminal
1.3 Parts Names
RESET (STOP) switch Releases the stop state of the connected WITHSTANDING VOLT-
AGE HiTESTER or the FAIL state after a failure is detected.
11
COMP LEVEL switch Changes the meter indication.
OFF (normal) : Indicates the measurement terminal voltage. ON (depressed) : Indicates the set threshold voltage. When this switch is ON, a vice. As the FAIL judgment is available, if a failure is detected, the
FAIL
signal is output to the external device after the switch is turned OFF. This switch lamp lights up in green if the test is being conducted properly.
LOW terminal Low-voltage input terminal for measurement terminal voltage moni-
toring. Isolated from the housing.
HIGH terminal High-voltage input terminal for measurement terminal voltage moni-
toring.
DANGER lamp Lights up during a test.
Starts blinking to provide a warning if a dangerous voltage is input while the test is not being performed.
Threshold adjustment knob
Power switch Powers the 3931 on or off.
Sets the voltage used as a criterion for failure. The threshold voltage is normally set to the voltage output from the WITHSTANDING VOLTAGE HiTESTER.
The LED lights up when power is turned ON.
FAIL
signal is not output to an external de-
Analog voltmeter Indicates output-voltage when testing withstand voltage.
12
1.3 Parts Names

1.3.2 Rear Panel

Power inlet
External I/O terminals
To prevent electric shock, when the
DANGER
lamp is lit or blink­ing, never touch the input-voltage terminals (HIGH and LOW), the H.V. TEST LEADs, or the tested object.
Protective conductor terminal
Power inlet Supplies power. Connect the supplied power cord
External I/O terminals Used to connect the 3931 to a WITHSTANDING
Protective conductor terminal
In this manual, an input signal is expressed as "signal name_IN" and an output signal as "signal name_OUT." (Example)
output signal
STOP
input signal
FAIL
to this inlet.
VOLTAGE HiTESTER and control the 3931.
Used to earth a protective ground wire. Be sure to make grounding connections before starting a test.
STOP_OUT
FAIL_ IN
1.3 Parts Names

1.3.3 H.V. TEST LEADs (9615-01, 9615-03)

To prevent electric shock, do not touch the alligator clip while high voltage is being generated. The vinyl sheath on the alliga­tor clip of the H.V. TEST LEADs does not have an insulation withstand voltage.
13
Alligator clip
High-voltage input plug
Alligator clip
Low-voltage input plug
9615-01 (red)
9615-03 (black)
* The plugs for high voltage and low voltage are of the same size.
Alligator clip Attach to the test point on the device being mea-
sured.
High-voltage input plug (red)
Low-voltage input plug (black)
Connect to the HIGH terminal on the unit.
Connect to the LOW terminal on the unit.
14

1.4 External Dimensions

1.4 External Dimensions
0.08”)
±
2 mm (9.06”
±
0.08”) 230
±
2 mm (5.91”
±
150
130 ± 2 mm (5.12” ± 0.08”)

1.5 Measurement Flowchart

1.5 Measurement Flowchart
15
The flowchart below shows the flow of the test. For detailed information, see each chapter.
Preparations for the withstand voltage test
(WITHSTANDING VOLTAGE HiTESTER)
Connecting the protective conductor terminal
Connecting the power cord
Connecting the external I/O terminals
Connecting the H.V. TEST LEADs
(WITHSTANDING VOLTAGE HiTESTER and 3931)
Powering On
(WITHSTANDING VOLTAGE HiTESTER and 3931)
NG
Startup inspection
2.1
2.2
2.3
2.42.5
2.6
2.7
NG
Correct the problem.
Check the wiring.
OK
Correct the problem.
Withstand voltage failure
NG
Check the wiring again.
OK
Setting the threshold voltage (Test voltage)
Start the withstand voltage test.
NG
NG
Withstand voltage
Contact check NG
Withstand voltage test completed
OK
test NG
OK
OK
Continue the test.
Next test
3.1.1
3.2
3.3
OK
Malfunction
NO
Powering OFF
(WITHSTANDING VOLTAGE HiTESTER and 3931)
End the test.
16
1.5 Measurement Flowchart
Testing
17

2.1 Connecting the Protective Conductor Terminal

Arrangements
Chapter 2
2.1 Connecting the Protective Conductor Terminal
Observe the following precautions to avoid electric shock.
• Connect the protective conductor terminal to a grounded con­ductor before making any other connections.
• Be sure to connect the protective conductor terminal to a grounded conductor.
1.
Using a Phillips-head screwdriver, remove the screw of the protective conductor terminal from the rear of the unit.
2.
3.
Connect an electric wire with a sufficient current capacity to the pro­tective conductor terminal, and secure the wire using a Phillips-head
screwdriver.
Specified cable
Press-fit terminal
Ground the other end of the wire.
UL1015 AWG16
BT1.25_F4
Sumitomo Electric Industries, Ltd.'s or equivalent
Fuji Terminal Industry Co., Ltd.'s or equivalent
GND
18

2.2 Connecting the Power Cord

2.2 Connecting the Power Cord
• Before turning the product on, make sure the source voltage matches that indicated on the product's power connector. Connection to an improper supply voltage may damage the product and present an electrical hazard.
• To avoid electric shock and ensure safe operation, connect the power cord to a grounded (3-contact) outlet.
1.
2.
3.
Be sure that the power switch is turned to OFF( ).
Connect the power cord provided to the power inlet on the back of the unit.
Insert the plug into the grounded outlet.
3
2

2.3 Connecting the External I/O Terminals

2.3 Connecting the External I/O Terminals
Connect the external I/O terminals before turning on the power. If the external I/O terminals are installed or removed following startup, mal­function may result.
19
1.
2.
Press the top of the external I/O terminal with a flathead screwdriver.
While holding down the top of the terminal, insert a stripped (approxi­mately 1 cm) signal wire into the terminal.
3.
Draw out the screwdriver.
For the external I/O terminals specification, see
Terminals" (page 43)
• The 3931 starts failure detection upon receipt of a signal from the con­nected WITHSTANDING VOLTAGE HiTESTER. Confirm that the exter­nal I/O terminals are properly connected to the WITHSTANDING VOLTAGE HiTESTER.
.
4.1, "External I/O
20

2.4 Connecting the H.V. TEST LEADs to the 3931

2.4 Connecting the H.V. TEST LEADs to the 3931
• Before connecting the H.V. TEST LEADs, make sure that the insulation on the H.V. TEST LEADs is undamaged and that no bare conductors are improperly exposed. Using the product under such conditions could result in electrocution. Replace the H.V TEST LEADs with the specified Hioki Model 9615-01, 9615-03.
• To avoid electric shock, check the following to confirm that a high voltage is not applied to the measurement terminal before connecting the H.V. TEST LEADs to the 3931. (1) The analog voltmeter reads 0 kV. (2) The
DANGER
lamp is OFF.
0 kV
Lamp is OFF
• To ensure safety, be sure to use the stopper plate. If the H.V. TEST LEAD comes off and the stopper plate is not used, the voltage input terminal is exposed, which may result in electric shock.
• To avoid electrical accidents, confirm that the H.V. TEST LEAD is securely connected before starting the test. It is dangerous to output a voltage with the TEST LEAD not properly con­nected.
• Be careful to ensure that the polarity of the H.V. TEST LEAD is correct. The LOW side of the 3931 is isolated from the housing (protective earth potential). Although it withstands up to 5 kVrms, if the polarity is incorrect, electric shock may occur, which will also result in a measurement error.
2.4 Connecting the H.V. TEST LEADs to the 3931
21
1.
2.
3.
Remove the screws fastening the stopper plate.
Remove the stopper plate.
Connect the H.V. TEST LEADs. Be careful to ensure that the polarity is correct.
4.
5.
LOW terminal
Low-voltage side
(black)
Attach the stopper plate.
Fasten the stopper plate with the screws.
High-voltage side
(red)
HIGH terminal
22

2.5 Connecting the H.V. TEST LEADs to the Device to be Tested

2.5 Connecting the H.V. TEST LEADs to the Device to be Tested
Observe the following precautions to avoid electric shock.
• Check the following to confirm that a high voltage is not applied to the voltage input terminals before connecting the H.V. TEST LEADs to the device. (1) The analog voltmeter of the WITHSTANDING VOLTAGE HiT-
ESTER reads 0 kV.
(2)
DANGER
is off.
• Make sure that no high voltage is being applied between the input-voltage terminals (HIGH and LOW), before touching the input-voltage terminals, the H.V. TEST LEADs, or tested object. Even following a test, there may be a residual voltage at the input-voltage terminals.
• Be careful to ensure that the polarity of the H.V. TEST LEAD is correct. Connect the HIGH side (high-tension side, red) of the H.V. TEST LEAD to the HIGH side (high-tension side, red) of the device to be tested to which the WITHSTANDING VOLTAGE HiTESTER is connected, and the LOW side (low-tension side, black) of the H.V. TEST LEAD to the LOW side (GND, black) of the device. The LOW side of the 3931 is isolated from the housing (protec­tive earth potential). Although it withstands up to 5 kVrms, if the polarity is incorrect, electric shock may occur, which will also result in a measurement error.
lamp of the WITHSTANDING VOLTAGE HiTESTER
2.5 Connecting the H.V. TEST LEADs to the Device to be Tested
23
1.
2.
3.
Withstanding votage HiTESTER
Confirm that the analog voltmeter of the WITHSTANDING VOLTAGE HiTESTER reads 0 kV and that the
Connect the LOW side (black) of the H.V. TEST LEAD to the device to be tested.
Connect the HIGH side (red) of the H.V. TEST LEAD to the device to be tested.
Hioki WITHSTANDING VOLTAGE HiTESTER Instruction Manual
Confirm that the H.V. TEST LEADs connected to the HiTESTER and the 3931 do not touch each other. If the leads touch each other, the 3931 can­not detect contact failure, although it can detect a broken wire. Confirm that the probes are securely connected and that they will not come off dur­ing the test.
3931
Withstanding votage HiTESTER
DANGER
lamp is off.
3931
LOW
HIGH
Device to be tested
LOW
HIGH
LOW
HIGH
Device to be tested
LOW
HIGH
24

2.6 Powering On and Off the 3931

2.6 Powering On and Off the 3931
Before turning the product on, make sure the source voltage matches that indicated on the product's power connector. Con­nection to an improper supply voltage may damage the product and present an electrical hazard.
Allow 10 minutes warming up after powering on.
Powering on the unit
Turn the power switch to ON( ).
Powering off the unit
1.
2.
Upon completion of the test, check the following.
• The analog voltmeter of the WITHSTANDING VOLTAGE HiTESTER reads 0 kV.
DANGER lamp of the WITHSTANDING VOLTAGE HiTESTER is off.
0 kV
Lamp is OFF
3158
AC WITHSTANDING VOLTAGE HiTESTER
Turn the power switch to OFF( ).
If the analog voltmeter of the WITHSTANDING VOLTAGE HiTESTER does not read 0 kV or the test has been completed, the 3931 may malfunction. Turn off the power and contact your vender (our agent) or your nearest Hioki office.
DANGER
lamp is on despite the fact that the
2.7 Startup Inspection
To ensure safe testing, check the following before starting operation: For inspections of the WITHSTANDING VOLTAGE HiTESTER, see its operator's manual.

2.7.1 Cutoff Voltage

Check to confirm that the 3931 operates properly. After connecting the 3931 to the WITHSTANDING VOLTAGE HiTESTER, confirm that the 3931 enters the PASS state when the voltage exceeds the set thresh­old voltage and enters the FAIL state when the voltage is below the threshold. To create a FAIL state, set up the 3931 as shown in the example below.
25

2.7 Startup Inspection

1.
(Example)
Threshold voltage: 2.5 kV
• Check for the PASS state HiTESTER output voltage: 2.7 kV
• Check for the FAIL state HiTESTER output voltage: 2.3 kV
Connect the 3931 to the WITHSTANDING VOLTAGE HiTESTER in order to conduct a test. Do not forget to connect the external I/O termi­nals.
2.3, "Connecting the External I/O Terminals" (page 19)
Withstanding votage HiTESTER
HIGH
LOW
LOW
3931
HIGH
2.
Device to be tested
Set the threshold voltage to
3.1.1, "Setting the Threshold Voltage" (page 31)
2.5 kV
.
26
2.7 Startup Inspection
3.
4.
5.
Set the output voltage of the WITHSTANDING VOLTAGE HiTESTER
2.7 kV
to The output voltage may be any voltage, provided that it exceeds the threshold voltage. (In the example above, it shall be greater than 2.5 kV.)
Hioki WITHSTANDING VOLTAGE HiTESTER Instruction Manual
Start up the WITHSTANDING VOLTAGE HiTESTER to input the volt­age to the 3931. Turn off the test timer of the HiTESTER to make it easier to check that the 3931 operates normally before the test.
Confirm that the
DANGER lamp lights up (steady). (This indicates that the withstand
voltage test is being performed properly.)
.
COMP LEVEL
Lights up in green
switch lights up in green and that the
6.
7.
Lights up
Set the output voltage of the WITHSTANDING VOLTAGE HiTESTER
2.3 kV
to The output voltage may be any voltage, provided that it is lower than the threshold voltage. Normally, deduct 5% from the threshold voltage (deduct 50 V when the threshold is 1 kV or less).
When the 3158, 3159, or 3173 is used, gradually turn the output volt­age adjustment knob counterclockwise. Confirm that the 3931 enters the FAIL state and the WITHSTANDING VOLTAGE HiTESTER stops when the output voltage of the HiTESTER falls below 2.5 kV.
.
8.
When the 3153 is used, stop the 3153, set the output voltage to 2.3 kV, and then restart the 3153.
3.3.2, "FAIL State" (page 40)
4.1.5, "Example of Connection to the WITHSTANDING VOLTAGE HiTESTER" (page 48) "Connection to the 3153 (Using an automatic machine)"(page 52)

2.7.2 Analog Voltmeter

(Example)
Check to confirm that the margin of error of the analog volt­meter is within ± 5% f.s.
• HiTESTER output voltage: 2.5 kV
• Meter of the 3931: Shall read 2.25 kV to 2.75 kV
27
2.7 Startup Inspection
1.
2.
3.
Before turning on the power, confirm that the analog voltmeter reads
kV
.
0 kV
If the indicator does not point to zero, adjust the zero position using a flathead screwdriver.
Set the output voltage of the WITHSTANDING VOLTAGE HiTESTER
2.5 kV
to
Hioki WITHSTANDING VOLTAGE HiTESTER Instruction Manual
and start up the HiTESTER.
Withstanding votage HiTESTER
3931
0
LOW
HIGH
Device to be tested
LOW
HIGH
28
2.7 Startup Inspection
4.
Confirm that the meter of the 3931 reads the same voltage as the out­put voltage of the WITHSTANDING VOLTAGE HiTESTER. In this example, the meter shall indicate a value between
2.75 V
.
Do not twist together (twisted pair) the H.V. test leads connected to the 3931 and to the HiTESTER (HIGH with HIGH and LOW with LOW). If they are twisted together, due to influence of the capacity coupling, the meter may indicate a voltage even if no voltage is input to the 3931 (approxi­mately 10% of the input) applied during the withstand voltage test. Never twist HIGH and LOW leads together into a pair.
2.25 V
and
29

Testing Method

This chapter explains in detail the procedure for using the 3931 HIGH VOLTAGE CONTACT CHECKER (test condition setting procedure and testing procedure). Read Chapter 2, "Testing Arrangements" (page 17), and make the necessary preparations for testing.
Chapter 3
30

3.1 Test Startup Flow

3.1 Test Startup Flow
The flowchart below explains how a test is carried out.
(Example)
Check for broken wires and contact failures during the withstand voltage test using the 3931.
• HiTESTER output voltage : 2.5 kV
• Threshold voltage of the 3931: 2.5 kV
Setting the test parameters
Set the test conditions. Set the threshold voltage to 2.5 kV.
(2.4 kV if the threshold voltage is set)
3.1.1, "Setting the Threshold Voltage"
Start the test.
3.2, "Starting a Test"
1. Connect the WITHSTANDING VOLTAGE HiTESTER, the device to be tested, and the 3931.
2.
Set the output voltage of the WITHSTANDING VOLTAGE HiTESTER to
2.5 kV
.
3. Press the external START switch of the WITHSTANDING VOLTAGE HiTESTER; the test will start.
4. When the 3931 starts voltage monitoring, the
DANGER
lamp of the 3931 will
light up (and remain on).
* If a voltage of 50 V to 100 V is input while the 3931 is not yet in the START
state, the DANGER
PASS/FAIL judgment
• PASS: The test is being conducted properly.
Lights up in green
lamp will go out.
COMP LEVEL
The
Lights up
indicates that the voltages monitored in the set test time exceed the set threshold voltage. If the test is
3.3, "PASS or FAIL Determination"
switch will light up in green. This
completed in this state, the withstand voltage test is judged as PASS or FAIL by the WITHSTANDING VOLTAGE HiTESTER.
• FAIL: A failure is detected.
Lights up in red
Buzzer
RESET (STOP)
The
switch will light up in red and the buzzer will sound. The 3931 stops the WITH­STANDING VOLTAGE HiTESTER. Please note that the PASS judgment made by the WITHSTANDING VOLTAGE HiTESTER is meaningless if the 3931 has detected a failure.
3.1 Test Startup Flow

3.1.1 Setting the Threshold Voltage

• When the WITHSTANDING VOLTAGE HiTESTER is of the supply voltage boosting type, the output of which is not stabilized, the 3931 detects an abnormal output voltage due to fluctuations in the supply voltage. If the 3931 is judged as FAIL even within the allowable volt­age fluctuation range, use a stabilized power supply for the WITH­STANDING VOLTAGE HiTESTER or set the threshold voltage of the 3931 to a low value so as to allow for voltage fluctuations.
• To avoid errors, do not change the threshold voltage setting during testing.
Setting the Threshold by Applying Voltage from the HiTESTER
31
1.
2.
To avoid electric shock, never touch the input-voltage terminals (HIGH and LOW), the H.V. TEST LEADs, or tested object. This method prevents meter errors and internal errors and therefore enables more accurate setting than that made only by watching the meter. It is, however, more dangerous, as a high voltage is applied.
Set the output voltage of the WITHSTANDING VOLTAGE HiTESTER
2.5 kV
to
Hioki WITHSTANDING VOLTAGE HiTESTER Instruction Manual
Connect a non-defective device to be tested.
2.4, "Connecting the H.V. TEST LEADs to the 3931" (page 20)
2.5, "Connecting the H.V. TEST LEADs to the Device to be Tested"
.
(page 22)
Withstanding votage HiTESTER
3931
LOW
HIGH
Device to be tested
LOW
HIGH
32
3.1 Test Startup Flow
3.
4.
5.
Fully rotate the threshold adjustment knob of the 3931 counterclock­wise.
Start up the WITHSTANDING VOLTAGE HiTESTER and apply a test voltage.
Confirm that the
Lamp is OFF
RESET (STOP)
switch lamp is off.
Judged as FAIL
This is merely an internal error and not a failure of the 3931. Start up the WITHSTANDING VOLTAGE HiTESTER while holding down
COMP LEVEL
all steps through step 6. If a signal is properly input from the WITHSTANDING VOLTAGE HiT­ESTER, the 3931 will light up.
switch and perform
TEST
DANGER
lamp of the
6.
While holding down the adjustment knob clockwise.
Stop just before the switch lights up in red.
When the measurement terminal voltage falls below the threshold, the
RESET (STOP)
that in which the switch lamp is completely off (the closest position to that in which the lamp remains on, rather than blinking). When the threshold voltage is set by this method, the output voltage of the WITHSTANDING VOLTAGE HiTESTER can be set to approxi­mately -30 V. We recommend that the threshold voltage be set within this range.
switch lights up in red. The optimum setting position is
COMP LEVEL
switch, rotate the threshold
33
3.1 Test Startup Flow
• While the COMP LEVEL switch is on, failure detection is continued internally but a lights up during setting of the threshold, the 3931 enters the FAIL state after setting is completed and the COMP LEVEL switch is turned off (not depressed). However, this does not indicate a failure of the 3931. Press the RESET (STOP) switch to cancel the FAIL state.
• If the output of the WITHSTANDING VOLTAGE HiTESTER is affected by fluctuations in the supply voltage, this effect can be minimized by set­ting the threshold to a low value. However, this reduces the sensitivity of the 3931. We recommend that a stabilized power supply be used for the WITHSTANDING VOLTAGE HiTESTER, rather than setting the thresh­old to too low of a value.
• If a wire is broken when a large current (10 mA or below) flows during a withstand voltage test, the discharge of sparks is increased. When the difference between the threshold voltage and the test voltage is large and spark discharge occurs, a FAIL judgment may not be made due to the influence of noise. In such a case, set the threshold closer to the output voltage of the WITHSTANDING VOLTAGE HiTESTER.
FAIL
signal is not output. If the RESET (STOP) switch
34
3.1 Test Startup Flow
Setting the Threshold Without Applying Voltage from the HiTESTER
While holding down the adjustment knob to set the threshold voltage to
COMP LEVEL
switch, rotate the threshold
2.5 kV
. Set the thresh-
old voltage while watching the meter.
Set while watching
the meter
• With this method, the threshold can be set without applying a high volt­age to the 3931. However, the set threshold voltage may not be accu­rate due to an internal error. Adjust the setting as necessary.
• Ideally, the threshold voltage be the same as the voltage output by the WITHSTANDING VOLTAGE HiTESTER. In reality, we recommend that the output voltage be slightly less than the set threshold voltage.
• If the output of the WITHSTANDING VOLTAGE HiTESTER is affected by fluctuations in the supply voltage, this effect can be minimized by set­ting the threshold to a low level. However, this reduces the sensitivity of the 3931. We recommend that a stabilized power supply be used for the WITHSTANDING VOLTAGE HiTESTER, rather than setting the thresh­old to too low a level.
How is the Set Threshold voltage Viewed?
Press the viewed.
COMP LEVEL
switch; this will allow the set voltage to be
3.2 Starting a Test
Observe the following precautions to avoid electric shock.
• Confirm the following items, before a high voltage is applied: (1) The analog voltmeter of the WITHSTANDING VOLTAGE HiT-
ESTER reads 0 kV.
(2) DANGER lamp of the WITHSTANDING VOLTAGE HiTESTER
is off.
• Never touch the HIGH terminal, the H.V. TEST LEADs, or the device to be tested when the threshold voltage of the 3931 is set while a high voltage is output from the WITHSTANDING VOLTAGE HiTESTER.
35

3.2 Starting a Test

To avoid errors, do not change the threshold voltage setting during testing.
The 3931 starts failure detection upon receipt of a signal from the con­nected WITHSTANDING VOLTAGE HiTESTER. Once the setting has been made, it is not necessary to operate the 3931 to start the test. Start and stop the test on the WITHSTANDING VOLTAGE HiTESTER.
36
3.2 Starting a Test
1.
Connect the 3931 to the WITHSTANDING VOLTAGE HiTESTER. Do not forget to connect the external I/O terminals.
Withstanding votage HiTESTER
HIGH
LOW
Device to be tested
2.
Set the threshold voltage to
Chapter 2, "Testing Arrangements" (page 17)
3931
LOW
HIGH
3.1.1, "Setting the Threshold Voltage" (page 31)
Withstanding votage HiTESTER
HIGH
LOW
Device to be tested
2.5 kV
.
3931
LOW
HIGH
3.
4.
5.
Set the output voltage of the WITHSTANDING VOLTAGE HiTESTER
2.5 kV
to
Hioki WITHSTANDING VOLTAGE HiTESTER Instruction Manual
Start the withstand voltage test in accordance with the operating pro­cedure of the WITHSTANDING VOLTAGE HiTESTER. Upon receipt of a HiTESTER, the 3931 automatically starts failure detection.
Press the external HiTESTER; the withstand voltage test will start. When measurement has been started properly, the
As the external I/O of the WITHSTANDING VOLTAGE HiTESTER is used, the START key of the HiTESTER may be disabled. Start the test in accor­dance with the correct procedure.
.
TEST
signal from the WITHSTANDING VOLTAGE
START
Hioki WITHSTANDING VOLTAGE HiTESTER Instruction Manual
switch of the WITHSTANDING VOLTAGE
DANGER
lamp will light up.
How is the Test Interrupted?
Stop the WITHSTANDING VOLTAGE HiTESTER by pressing the
RESET (STOP)
HiTESTER. When the the 3931 transmits a HiTESTER and stops its voltage output. If no failure has been detected thus far, the 3931 judges that "no failure is detected."
switch of the 3931 or using the control method for the
RESET (STOP)
STOP
signal to the WITHSTANDING VOLTAGE
switch of the 3931 is pressed,

3.3 PASS or FAIL Determination

3.3 PASS or FAIL Determination

3.3.1 PASS State

• Even when a test has been terminated, there may still be volt­age in the input-voltage terminals. Confirm the following items, before touching the input-voltage terminals (HIGH and LOW), the H.V. TEST LEADs, or tested object. (1) The analog voltmeter of the WITHSTANDING VOLTAGE HiT-
ESTER reads 0 kV.
(2) DANGER lamp of the WITHSTANDING VOLTAGE HiTESTER
is off.
• When a high voltage is applied to a cable and a wire in the cable is broken, spark discharge occurs and the withstand voltage test may be affected by noise. Take measures for noise protection.
37
When the PASS state is indicated, i.e., the test is being performed properly. Upon receipt of a HiTESTER, the 3931 starts the test. After the test has been started, when the voltage between the measurement terminals exceeds the set threshold voltage, the
COMP LEVEL
TEST
Lights up in green
switch of the 3931 lights up in green, the
signal from the WITHSTANDING VOLTAGE
COMP LEVEL
switch lights up in green.
Lights up
38
3.3 PASS or FAIL Determination
• If the 3931 detects a broken wire/contact failure or an abnormal test voltage, the ESTER is masked. Even if the WITHSTANDING VOLTAGE HiTESTER has made a PASS judgment, the 3931 does not output a nal. When the 3931 outputs a properly connected to the WITHSTANDING VOLTAGE HiTESTER, the test has been completed properly.
• When a HiTESTER, if the 3931 detects a voltage of 50 V to 100 V, the DANGER lamp starts blinking. While the DANGER lamp is blinking, the judgment of the 3931 is not sent to the WITHSTANDING VOLTAGE HiTESTER, and the test is not determined to have been completed.
• If a ESTER, the DANGER lamp remains steadily lit. When the DANGER lamp is on and the COMP LEVEL switch is lit up in green, the test is being performed properly. Under these conditions, if the WITHSTAND­ING VOLTAGE HiTESTER makes a PASS judgment, the test has been completed without any failure such as a broken wire, contact failure, or voltage fluctuation. The 3931 will not control the WITHSTANDING VOLTAGE HiTESTER. Take the necessary steps in accordance with a PASS judgment.
•A put after the time set on the HiTESTER's timer has elapsed. When the timer is not used, a
TEST
TEST
signal is input from the WITHSTANDING VOLTAGE HiT-
PASS
signal from the WITHSTANDING VOLTAGE HiTESTER is out-
Hioki WITHSTANDING VOLTAGE HiTESTER Instruction Manual
PASS
signal from the WITHSTANDING VOLTAGE HiT-
PASS_OUT
PASS_OUT
signal (L level), if the 3931 is
sig-
signal is not input from the WITHSTANDING VOLTAGE
PASS
signal is not output.
Flow of Pass Determination
• The 3931 makes a PASS judgment and acts as shown below when the voltage between the measurement terminals exceeds the set threshold voltage.
39
3.3 PASS or FAIL Determination
Voltage
Threshold voltage
of the 3931
1. Lights up the
2. Receives a
3. Outputs a
WITHSTANDING VOLTAGE
• The diagram below shows the operation in detail, from start to com­pletion of the test. For details on PASS judgment, see the operator's manual of the WITHSTANDING VOLTAGE HiTESTER used.
DANGER lamp
Ramp-up time
of the 3153
HiTESTER
HiTESTER
blinking
COMP LEVEL
PASS
signal from the WITHSTANDING VOLTAGE
PASS_OUT
signal
PASS
Output voltage of the HiTESTER
DANGER lamp lights up
switch in green
signal
3931
PASS-OUT
signal
(e.g., lamp and buzzer)
DANGER lamp
blinking
Load
Ramp-down time
of the 3153
Approx. 50 V
HiTESTER judgment exexample
3931 judgment exexample
HiTESTER
START
signal
TEST
Undetermined
Approx. 0.3 s
The 3931 starts failure detection.
Undetermined
* The 3931 outputs a
HiTESTER
STOP
Failure detection time of the 3931
Te st
Normal judgment
PASS
signal from the HiTESTER (
Time
PASS
Undetermined*
Undetermined
PASS IN
).
40
3.3 PASS or FAIL Determination

3.3.2 FAIL State

• Even when a test has been terminated, there may still be volt-
• When a high voltage is applied to a cable and a wire in the
age in the input-voltage terminals. Confirm the following items, before touching the input-voltage terminals (HIGH and LOW), the H.V. TEST LEADs, or tested object. (1) The analog voltmeter of the WITHSTANDING VOLTAGE HiT-
ESTER reads 0 kV.
(2) DANGER lamp of the WITHSTANDING VOLTAGE HiTESTER
is off.
cable is broken, spark discharge occurs and the withstand voltage test may be affected by noise. Take measures for noise protection.
When the state is indicated. If the monitored voltage falls below the set threshold voltage during the test, the 3931 immediately sends a the WITHSTANDING VOLTAGE HiTESTER in order to stop the output of the voltage.
If the 3931 enters the FAIL state, press the cancel the FAIL state (the state continues until the switch is pressed).
RESET (STOP)
Lights up in red
switch of the 3931 lights up in red, the FAIL
STOP
signal to
Buzzer sounds
RESET (STOP)
switch to
Flow of Fail Determination
• If the voltage between the measurement terminals falls below the threshold voltage due to a broken wire, contact failure, or voltage fluctuation, the 3931 makes a FAIL judgment and acts as shown below.
41
3.3 PASS or FAIL Determination
Voltage
Threshold voltage
of the 3931
1. Lights up the
2. Masks (invalidates) the
RESET (STOP)
PASS
switch in red.
signal from the WITHSTANDING
VOLTAGE HiTESTER.
3. Sounds the buzzer.
4. Transmits a
STOP
signal to the connected WITHSTANDING
VOLTAGE HiTESTER
5. Transmits a
WITHSTANDING VOLTAGE
HiTESTER
FAIL_OUT
PASS
signal.
signal
3931
PAS S
signal
FAIL_OUT
• The diagram below shows the operation in detail
Down by approx. 5%
Output voltage of the HiTESTER
Automatic machine
Load
(e.g., lamp and buzzer)
signalSTOP signal
Approx. 50 V
TEST
HiTESTER judgment exexample
3931 judgment exexample
HiTESTER
START
HiTESTER
signal
DANGER lamp
blinking
Approx. 0.3 s
The 3931 starts failure detection.
Undetermined
Approx. 0.1 s
DANGER lamp lights up
HiTESTER
FAIL
lamp blinking
Failure detection time of the 3931
Te st
PASS
DANGER
Time
HiTESTER
STOP
UndeterminedUndetermined
FAIL
(Continues until STOP-switch input is made)
42
3.3 PASS or FAIL Determination
External
43

4.1 External I/O Terminals

Interface
4.1
External I/O Terminals
Chapter 4
The 3931 HIGH VOLTAGE CONTACT CHECKER automatically moni­tors the state of the test upon receipt of a signal from the WITH­STANDING VOLTAGE HiTESTER. If it is not properly connected to the WITHSTANDING VOLTAGE HiTESTER, the abnormal state detected by the 3931 will not be reflected in the test. Be sure to properly con­nect the external I/O signal lines to the WITHSTANDING VOLTAGE HiTESTER. All signals are based on negative logic and become effective at the LOW level.
In this manual, an input signal is expressed as "signal name_IN" and an output signal as "signal name_OUT." (Example)
STOP
output signal
FAIL
input signal
STOP_OUT
FAIL_IN
44
4.1 External I/O Terminals

4.1.1 Signal Line

Input Signal
Input Signal from an Automatic Machine
Connects chine, such as a computer and sequencer, to the WITH-
STOP_IN
Input Signal from the WITHSTANDING VOLTAGE HiTESTER
TEST_IN
PASS_IN
FAIL_IN
STANDING VOLTAGE HiTESTER. This signal and the
STOP
ic and output as a
Connects the VOLTAGE HiTESTER. When this signal is input, the 3931 considers the withstand voltage test to have start­ed and starts failure detection.
Connects the VOLTAGE HiTESTER. If the 3931 detects a failure, the 3931 invalidates the
PASS
ESTER and outputs a
Connects the VOLTAGE HiTESTER. The STANDING VOLTAGE HiTESTER and the due to a failure detected by the 3931 are combined by wired-OR logic and output as a
STOP
-signal output from an automatic ma-
signal of the 3931 are combined by wired-OR log-
STOP_OUT
TEST
signal of the WITHSTANDING
PASS
signal of the WITHSTANDING
signal from the WITHSTANDING VOLTAGE HiT-
PASS_OUT
FAIL
signal from the WITHSTANDING
signal.
signal.
FAIL
signal from the WITH-
FAIL_OUT
FAIL
signal.
signal
Output Signal
Output Signal to an Automatic Machine
If the 3931 detects a failure, the 3931 invalidates the
PASS
signal from the WITHSTANDING VOLTAGE HiT-
PASS_OUT
FAIL_OUT
Output Signal to the WITHSTANDING VOLTAGE HiTESTER
STOP_OUT
ESTER and outputs this signal to the automatic ma­chine. If the 3931 does not detect a failure, the 3931 outputs a VOLTAGE HiTESTER.
If the 3931 detects a failure or the WITHSTANDING VOLTAGE HiTESTER enters the FAIL state, this signal is output.
If a the RESET (STOP) switch of the 3931 is pressed, or if the 3931 detects a failure and enters FAIL state, the 3931 outputs a VOLTAGE HiTESTER in order to stop the test.
PASS
signal from the WITHSTANDING
STOP
signal is input from an automatic machine, if
STOP
signal to the WITHSTANDING
4.1 External I/O Terminals

4.1.2 Timing Chart of External I/O Signals

The diagram below shows the timing of external I/O signals and the start and stop of failure detection. For the timing relating to the with­stand voltage test, see the operator's manual of the WITHSTANDING VOLTAGE HiTESTER.
TEST IN
450 ms max.
45
State of the 3931
FAIL OUT
STOP IN
STOP switch
of the 3931
STOP OUT
Failure detection time
Failure detection
40 ms min.
50
µ
s max.
Tes t t ime
100 ms max.
50
µ
s max.
PASS IN
FAIL IN
PASS OUT
FAIL OUT
50
µ
s max.
46
4.1 External I/O Terminals

4.1.3 Example of Input Signal Connection

This section provides details of signals input to the 3931.
Input signals
Applied voltage
Voltage between V
and V
COM
HIGH level input voltage V
LOW level input voltage V
LOW level input current 1 mA min.
Signal names
EXT_V
(Insulated with a photo coupler)
EXT_V
EXT_COM
STOP_IN
TEST_IN
PASS_IN
FAIL_IN
Each signal is connected to V
Active low input
5 V DC min.
25 V DC max.
-1.0 min. V
-0.5 V min. V
: Stoppage of test
: Failure detection time
: Normal end of test
: Abnormal end of test
EXT_V
EXT_V terminal
+1.0 max.
EXT_V
EXT_V
-4.0 max.
at 3.66 kΩ.
7.32 K
• Design the connection to the input signals so as to prevent errors by using a relay, switch, or other circuit to prevent chattering.
• Design the connection so that a current of at least 1 mA flows when an input signal is active.
7.32 K
1 K
1 mA min.
Input signal
4.1 External I/O Terminals

4.1.4 Example of Output Signal Connection

This section provides details of signals output to the 3931.
47
Output signals
Applied voltage
Voltage between V
and V
HIGH level input voltage V
LOW level input voltage V
LOW level input current 10 mA min. (Sink current)
Signal names
EXT_DCV
COM
(Insulated with a photo coupler)
EXT_COM
STOP_OUT
TEST_OUT
PASS_OUT
FAIL_OUT
Each signal is connected to V
Active low input
5 V DC min.
25 V DC max.
-0.5 min. V
EXT_V
-0.5 V min. V
: Stoppage of test
: Failure detection time
: Normal end of test
: Abnormal end of test
EXT_V
EXT_COM
EXT_V
EXT_V terminal
max.
+0.5 V max.
at 10 kΩ.
10 K
10
10 mA max.
• The maximum sinkable current is 10 mA per signal.
• When a relay or other inductive load is connected, connect the diode in parallel with the coil.
10 K
Output terminal
EXT_COM terminal
48
4.1 External I/O Terminals
4.1.5 Example of Connection to the WITHSTAND­ING VOLTAGE HiTESTER
This section provides an example of connection to the Hioki WITH­STANDING VOLTAGE HiTESTER. The lines inside the broken line shall be connected by the user.
Connection to the 3173
Automatic machine (sequencer)
START
STOP
FAIL
PASS
TEST
(Using an automatic machine)
3173
START
STOP
FAIL
FAIL
PASS
PASS
TEST
COM
COM
+5 V
EXT_V
FAIL_OUT
PASS_OUT
STOP_IN
TEST_IN
STOP_OUT
FAIL_IN
PASS IN
GND
EXT_COM
3931
* The arrow indicates the direction of signal transmission.
The diagram above assumes that the operation is controlled from all automatic machines.
4.1 External I/O Terminals
49
Connection to the 3173
3931
STOP_OUT
PASS_OUT
+5 V
GND
EXT_V
EXT_COM
TEST_IN
STOP_IN
PASS_IN
FAIL_IN
FAI L_O UT
(Not using an automatic machine)
TEST STOP START COM
Current
amplification
circuit
Load
3173
PASS
FAIL
Current
amplification
circuit
Load: Relay, lamp, etc.
* The arrow indicates the direction of signal transmission.
PASS_IN
and
PASS_OUT
or lamp is not used.
Load
signals are not necessary if an external relay
50
4.1 External I/O Terminals
Connection to the 3158 and 3159
Automatic machine (sequencer)
START
STOP
EXT_E
INT_LOCK
READY
L_FAIL
U_FAIL
PASS
TEST
HV_ON
(Using an automatic machine)
ISO_DCV
START
STOP
EXT_E
INT_LOCK
READY
L_FAIL
U_FAIL
PASS
TEST
HV_ON
3158 3159
COM
ISO_COM
+5 V
EXT_V
STOP_IN
FAI L_O UT
PASS_OUT
STOP_OUT
FAIL_IN
TEST_IN
PASS_IN
GND
EXT_COM
3931
* The arrow indicates the direction of signal transmission.
The diagram above assumes that the operation is controlled from all automatic machines.
4.1 External I/O Terminals
51
Connection to the 3158 and 3159
3931
PASS_OUT
STOP_OUT
+5 V
EXT_V
FAI L_O UT
STOP_IN
TEST_IN
FAI L_I N
PASS_I N
EXT_COM
STOP switch
(Not using an automatic machine)
ISO_DCV
START switch
START
STOP
EXT_E
INT_LOCK
READY
L_FAIL
U_FAIL
PASS
TEST
HV_ON
3158 3159
GND
ISO_COM
Load: Relay, lamp, etc.
Current
amplification
circuit
Current
amplification
circuit
• The 3931 does not support insulation-resistance measurement of the 3159 and the 3153 and therefore cannot be used for the continuous test or program test.
• In the above connection, the START and STOP switches on the above diagram, and STOP switch of the 3158 or 3159 are valid. The START switch of the 3158 or 3159 is invalid.
• When a remote controller is used (9613, 9614), if the OPERATE switch is on, the START switch on the above diagram and the START switch of the 3158 or 3159 are invalid. The STOP switch on the above diagram and those of the remote controller and the 3931 are valid. If the OPER­ATE switch is off, any control from the remote controller is invalid.
PASS_IN
lamp is not used.
and
Load
Load
* The arrow indicates the direction of signal transmission.
PASS_OUT
signals are not necessary if an external relay or
52
4.1 External I/O Terminals
Connection to the 3153
Automatic machine (sequencer)
START
STOP
EXT_E
INT_LOCK
READY
L_FAIL
U_FAIL
PASS
TEST
HV_ON
COM
(Using an automatic machine)
ISO_DCV
EXT_DCV
START
STOP
EXT_E
INT_LOCK
READY
L_FAIL
U_FAIL
PASS
TEST
HV_ON
EXT_COM
ISO_GND
3153
+5 V
EXT_V
FAIL_OUT
PASS_OUT
STOP_IN
TEST_IN
STOP_OUT
FAI L_I N
PASS_IN
GND
EXT_COM
3931
* The arrow indicates the direction of signal transmission.
The diagram above assumes that the operation is controlled from all automatic machines.
4.1 External I/O Terminals
53
Connection to the 3153
3931
+5 V
EXT_V
PASS_OUT
FAI L_O UT
STOP_IN
STOP_OUT
TEST_IN
FAIL_IN
PASS_I N
EXT_COM
GND
(Not using an automatic machine)
START switch
STOP switch
ISO_DCV
EXT_DCV
START
STOP
EXT_E
INT_LOCK
READY
L_FAIL
U_FAIL
PASS
TEST
HV_ON
EXT_COM
ISO_GND
3153
Load: Relay, lamp, etc.
Current
amplification
circuit
Current
amplification
circuit
• The 3931 does not support insulation-resistance measurement of the 3159 and the 3153 and therefore cannot be used for the continuous test or program test.
• When the 3153 is used, the lamp-up/lamp-down function included in its output may affect the failure detection of the 3931. Using the optional function, set the included in the
3153 WITHSTANDING VOLTAGE HiTESTER Instruction Man-
Load
Load
* The arrow indicates the direction of signal transmission
TEST
signal output so that the TEST blinking time is not
TEST
signal ON time.
ual
• In the above connection, the START and STOP switches on the above diagram, and STOP switch of the 3153 are valid. The START switch of the 3153 is invalid.
• When a remote controller is used (9613, 9614), if the OPERATE switch is on, the START switch on the above diagram and the START switch of the 3153 are invalid. The STOP switch on the above diagram and those of the remote controller and the 3931 are valid. If the OPERATE switch is off, any control from the remote controller is invalid.
PASS_IN
lamp is not used.
and
PASS_OUT
signals are not necessary if an external relay or
54
4.1 External I/O Terminals

5.1 Basic Specifications

55

Specifications

Chapter 5
5.1 Basic Specifications
Input Unit for the Non-Insulation WITHSTANDING VOLTAGE HiTESTER
Maximum input voltage 5 kV AC, 50/60 Hz
Maximum rated voltage to the earth
Measurable range 200 V to 5 kV AC, 50/60 Hz
Waveform Commercial-power waveform, sine wave
Threshold voltage adjustment method
5 kV AC, 50/60 Hz
Manual adjustment
Voltmeter Analog voltmeter - average-value rectification rms type, 0 to 5 kV AC
±
5% f.s. (after more than 10 minutes of warm-up time)
Principle of operation Monitors the voltages of the measurement terminals during a withstand
voltage test, and detects abnormal voltages due to contact failure or a broken wire (also makes a FAIL judgment when the output voltage of the HiTESTER drops due to fluctuations in the supply voltage)
Input impedance
High-tension input unit, 150 M (HIGH side: 100 MΩ; LOW side: 50 MΩ)*
* This indicates that a 100 M resistor is inserted on the HIGH side,
and a 50 M resistor is inserted on the LOW side. It differs from the actual impedance, as a diode for rectification is inserted.
±
10%
56
5.1 Basic Specifications
Detection Unit for the Non-Insulation WITHSTANDING VOLTAGE HiTESTER
Threshold voltage
200 V to 5 kV (rms), meter reading, LED indication or VR
setting range
PASS guarantee range When the withstand voltage test has been completed with the monitored
voltage exceeding the set threshold voltage
FAIL guarantee range When the distortion rate is 5% or less, the monitored voltage is as follows:
5% or more of the set voltage when the set voltage is 1 kV to 5 kV 50 V or more when the set voltage is 200 V to 1 kV
When the monitored voltage remains smaller than the setting* for more than 100 ms (If the output voltage of the WITHSTANDING VOLTAGE HiTESTER remains larger than the setting and the current becomes larger than the threshold current (withstand voltage failure), the 3931 outputs a
FAI L
signal from the WITHSTANDING VOLTAGE HiTESTER
and does not make its own FAIL judgment.)
Judgment method Using an analog comparator
Judgment Monitors abnormal voltage during a test. Outputs a
PASS
signal or
FAIL
signal, depending on whether a failure is detected.
PASS
PASS Does not make its own PASS judgment. Outputs a
sig­nal from the WITHSTANDING VOLTAGE HiTESTER when a test is completed with the monitored voltage within the PASS guarantee range.
FAIL When the monitored voltage falls within the FAIL guarantee
range, masks AGE HiTESTER and transmits control master. Also outputs a STOP
PAS S
signals from the WITHSTANDING VOLT-
FAIL
signals to the HiTESTER
signal to the HiTESTER.
TEST
Judgment delay Start monitoring Within 450 ms after the
signal becomes LOW
FAIL Output FAILWithin 100 ms after FAIL is detected
Action Sounds a buzzer, lights up the LED, and outputs signals to EXT I/O ter-
minal in accordance with the judgment result.
* This value is not a value set by watching the meter, but by inputting a
reference voltage so that a FAIL judgment will not be made.

5.1.1 Interface

Rear Panel
57
5.1 Basic Specifications
External output power
Non-isolated 5 V DC 40 mA, without stabilization, with current control
supply
Between V
OUT
and V
COM
4.5 V DC or more when loaded with 40 mA
EXT_I/O input signal Isolated using a photo coupler. Each signal is pulled up internally to
V
EXT_DCV
TEST_IN STOP_IN
PAS S_IN FAIL_IN
with 3.66 kΩ.
Input level Determined by the external power supply (between
V Voltage between V HIGH V LOW V
EXT_V
and V
EXT_V
EXT_V
EXT_COM
) to determine the input level
EXT_V
+1.0 V max., V
- 4.0 V max., V
and V
EXT_COM
EXT_V
EXT_ COM
5 V min., 25 V max.
- 1.0 V min.
- 0.5 V min.
Minimum input current
1 mA (Source current at the LOW level)
EXT_I/O output signal Isolated using a photo coupler. Each signal is pulled up internally to
V
STOP_OUT
PAS S_OU T FAIL_OUT
EXT_V
with 10 kΩ.
Output level Determined by the external power supply (between
EXT_V
and V
EXT_COM
V Voltage between V HIGH V LOW V
EXT_V
EXT_ COM
) to determine the output level
and V
EXT_V
max., V
EXT_V
+0.5 V max., V
EXT_COM
- 0.5 V min.
5 V min., 25 V max.
EXT_ COM
- 0.5 V min.
Maximum output current
10 mA (Sink current at the LOW level)
58

5.2 General Specifications

5.2 General Specifications
Operating temperature and humidity
Storage temperature and humidity
Temperature and humidity for guaranteed accuracy Guaranteed accuracy period
Operating environment Indoors, altitude up to 2000 m (6562-ft.)
Power supply Rated voltage: 100 to 240 V AC (a change in voltage of ±10% taken
Dielectric strength
Maximum rated power 9 VA
Dimensions Approx. 130W X 150H X 230D mm (5.12”W X 5.91”H X 9.06”D)
0°C to 40°C (32°F to 104°F), 80% RH or less (non-condensating)
-10°C to 50°C (14°F to 122°F), 90% RH or less (no condensation)
23°C ± 5°C (73°F ± 9°F), 80% RH or less (non-condensating)
For 1 year
into consideration), 50/60 Hz
• 8.2 kV AC, 1 mA for one minute [voltage input terminals] - [instrument case]
• 1.35 kV AC, 10 mA for one minute [power supply terminals] - [instrument case]
(not including protrusions)
Mass Approx. 1.9 kg (67.02 oz.)
Accessories Power cord ....................................................................................... 1
9615-01 H.V. TEST LEAD (high voltage side).................................. 1
9615-03 H.V. TEST LEAD (low voltage side) ................................... 1
Stopper plate .................................................................................... 1
Instruction Manual ............................................................................ 1
Applicable Standards EMC EN61326:1997+A1:1998+A2:2001 CLASS A
EN61000-3-2:2000 EN61000-3-3:1995+A1:2001
Safety EN61010-1:1993+A2:1995
Power supply : Pollution Degree 2, Overvoltage Category II
(anticipated transient overvoltage 2500 V) EN61010-2-031:1994
Voltage input : Pollution Degree 2, Overvoltage Category I

5.3 H.V. TEST LEADs (9615-01, 9615-03) Specifications

59
5.3 H.V. TEST LEADs (
9615-01, 9615-03)
Specifications
Applicable Models 9615-01: Hioki WITHSTANDING VOLTAGE HiTESTER
9615-03: 3931 HIGH VOLTAGE CONTACT CHECKER
Rated voltage 5 kV AC or 5 kV DC
Rated current 150 mA AC or 150 mA DC
Dielectric strength 6.25 kV AC Sensitivity current 5 mA for one minute
Tested area (between cable conductor and cable sheath)
Operating temperature and humidity
Storage temperature and humidity
Operating environment Indoors, altitude up to 2000 m (6562-ft.)
Dimensions Approx. 1.5 m
0°C to 40°C (32°F to 104°F), 80% RH or less (non-condensating)
-10°C to 50°C (14°F to 122°F), 90% RH or less (no condensation)
Mass Approx. 50 g
Applicable Standards EN61010-2-031:1994 Pollution Degree 2, Overvoltage Category I
60
5.3 H.V. TEST LEADs (9615-01, 9615-03) Specifications
Maintenance
61

6.1 Cleaning and Storage

and Service
Chapter 6
6.1 Cleaning and Storage
Cleaning
To clean the product, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.

6.2 Repair and Servicing

• Adjustments and repairs should be made only by technically quali­fied personnel.
• If damage is suspected, check the "Troubleshooting" section before contacting your dealer or Hioki representative.
• Pack the product carefully so that it will not be damaged during shipment, and include a detailed written description of the problem. Hioki cannot be responsible for damage that occurs during ship­ment.
• Spare and replacement parts for this product are guaranteed to be avail­able only until 7 years after manufacture of this model is terminated.
• Use the original packing materials when reshipping the product, if possible.
• If the unit has been subject to moisture, or if oil and dust have accumulated in the unit interior, the danger of electrical shock or fires resulting from the deterioration of insulation increases greatly. If the unit is ever subject to excessive moisture, oil, or dust, cease use immediately, and return the unit to us for maintenance.
62
6.2 Repair and Servicing
Troubleshooting
Symptom Check Item Troubleshooting
If the product operates abnormally, check as described below. If the product continues to operate abnormally after taking the speci­fied corrective action, the product may be malfunctioning. Turn off the product's power switch immediately, then contact your dealer (Hioki agent) or the nearest Hioki sales office.
If any of the following should occur, stop using the unit, disconnect the power cord and the H.V. TEST LEADs and contact your dealer or Hioki representative.
• If you are certain that the unit is damaged.
• If the measurement is inoperative.
• If the unit was stored for a long period of time in high temperatures and humidity, or other undesirable conditions.
• If the unit was damaged in transit.
•The DANGER lamp does not light up or blink even if the measured voltage is input.
• The meter is not opera­tional.
• A FAIL judgment is not made even if the mea­sured voltage falls below the threshold voltage.
•The DANGER lamp contin- ues to blink after test has been started.
• The PASS or FAIL result is not output properly during inspections prior to daily operation.
Is the power cord disconnected? Connect the power cord.
Is the nected?
Is the external I/O disconnected?
Are the external I/O voltage level and current outside the specifica­tion?
Has the input?
Are the external I/O signal lines including the power supply discon­nected?
Are the external I/O voltage level and current outside the specifica­tion?
Is the threshold voltage correctly set?
Is the supply voltage of the WITH­STANDING VOLTAGE HiTESTER fluctuating?
TEST
signal line discon-
TEST
STOP_IN
signal been
signal line
Check the external I/O terminal for the
TEST
signal.
Connect
Check the voltage level and load cur­rent.
Make sure the input during the test.
Connect the signal lines correctly.
Check the voltage level and load cur­rent.
Set the threshold voltage correctly.
Use a stabilized power supply for the WITHSTANDING VOLTAGE HiT­ESTER, or set the threshold voltage lower to allow for fluctuations.
TEST
signal line correctly.
STOP_IN
signal is not
• The withstand voltage test result is NG (the WITH­STANDING VOLTAGE HiTESTER outputs an NG result) when the 3931 is connected.
Is there a problem with the H.V. TEST LEAD connected to the WITHSTANDING VOLTAGE HiT­ESTER or the 3931?
As the 3931 may be operating properly, inspect the H.V. TEST LEAD.
As the current flowing to the 3931 is very small (0.033 mA or less when 5 kV is input), it may still affect the with­stand voltage test. Increase the thresh­old current of the WITHSTANDING VOLTAGE HiTESTER by 0.1 mA and repeat the test.
HIOKI 3931 HIGH VOLTAGE CONTACT CHECKER
Instruction Manual
Publication date: September 2003 Edition 1
Edited and published by HIOKI E.E. CORPORATION Technical Support Section
All inquiries to International Sales and Marketing Department 81 Koizumi, Ueda, Nagano, 386-1192, Japan TEL: +81-268-28-0562 / FAX: +81-268-28-0568 E-mail: os-com@hioki.co.jp URL http://www.hioki.co.jp/
Printed in Japan 3931A981-00
• All reasonable care has been taken in the production of this manual, but if you find any points which are unclear or in error, please contact your supplier or the International Sales and Mar­keting Department at HIOKI headquarters.
• In the interests of product development, the contents of this manual are subject to revision without prior notice.
• Unauthorized reproduction or copying of this manual is prohibited.
HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp / URL http://www.hioki.co.jp/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA TEL +1-609-409-9109 / FAX +1-609-409-9108
3931A981-00 03-09H
Printed on recycled paper
Loading...