Hioki 3911-20 Instruction Manual

Page 1
INSTRUCTION MANUAL
For...は専用機種。複数の場合は「/」で区切る。
不要の場合はとる。
形名を入力。 複数の場合は「/」で区切る。
3911-20
品名を入力。
COMMUNICATION BASE
Page 2
Page 3
Contents
Safety Notes ii
Notes on Use iii
Chapter 1 Overview 1
1.1 Product Overview 1
1.2 Name and Functions of Parts
Chapter 2 Application 3
2.1 Replacing the Battery 3
2.2 Installing Application Software
2.3 Connecting and Setting PC to the 3911-20
2.4 Connecting the 3911-20 with Logger
Chapter 3 Transferring Data 9
3.1 Measuring with Logger 9
3.2 Setting the 3911-20
3.3 Sending Setting Data
3.4 Receiving Measurement Data
3.5 Sending Measurement Data
3.6 Confirming and Clearing Loaded Data
10
11
12
13
14
2
5
6
7
Chapter 4 Specifications 15
Chapter 5 Maintenance and Service 17
Page 4
Page 5
_____________________________________________________________________
Introduction
Thank you for purchasing the HIOKI "3911-20 COMMUNICATION BASE." To obtain maximum performance from the product, please read this manual first, and keep it handy for future reference.
Inspection
When you receive the product, inspect it carefully to ensure that no damage occurred during If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Accessories
Instruction Manual LR03 alkaline battery X 4 COMMUNICATION UTILITY Operation Guide CD-R (COMMUNICATION UTILITY)
i
Note on shipping
Use the original packing materials when reshipping the product, if possible.
Registered trademarks
Windows is a registered trademark of Microsoft Corporation. Pentium is a registered trademarks of Intel Corporation.
The latest version of COMMUNICATION UTILITY is now available at our web site for downloading. Visit our site at: http://www.hioki.co.jp/
______________________________________________________________
Inspection
Page 6
ii
_____________________________________________________________________
SafetyNotes
This product is designed to conform to IEC 61010
WARNING
Safety Symbols
Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the product. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from product defects.
This manual contains information and warnings essential for safe operation of the product and for maintaining it in safe operating condition. Before using the product, be sure to carefully read the following safety notes.
The symbol printed on the product indicates that the user should refer to a corresponding topic in the manual (marked with the relevant function. In the manual, the symbol indicates particularly important information that the user should read before using the product.
symbol) before using the
Indicates DC (Direct Current).
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents a
WARNING
CAUTION
______________________________________________________________
Safety Notes
significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the product.
Page 7
_____________________________________________________________________
iii
Notes on Use
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
CAUTION
The unit should always be operated in a range from -20 to 70 and 80% RH or less. During data transfer when connected to personal computer (using RS-232C), operating temperature range is from 0 If used outside the specified environmental ranges for operation (or storage), the operation of the unit cannot be guaranteed. This product is not designed to be entirely water- or dust­proof. To avoid damage, do not use it in a wet or dusty environment. Take care to avoid condensation. In particular, if there is a sudden change of temperature (for example moving from a cold place to a warm one), condensation is likely to occur. Do not use the product where it may be exposed to corrosive or combustible gases. The product may be damaged or cause an explosion. To avoid corrosion from battery leakage, remove the batteries from the product if it is to be stored for a long time. Leaving the unit connected to an activated personal computer hastens battery discharge. When not using the unit, disconnect it from personal computer.
to 40 .
______________________________________________________________
Notes on Use
Page 8
iv
_____________________________________________________________________
Care and handling of CD-R
To keep disc clean from finger prints and prevent label fading, carefully hold disc on the outer edge when handling. To prevent damage to label, do not expose disc to alcohol based solvents or water. When writing on label, use an oil base felt pen. Do not write with a ball point pen or other hard tipped pens to prevent scratching disc. To prevent disc warping, do not expose disc to direct sunlight, high temperature and high humidity.
Note on Used for COMMUNICATION UTILITY
Before using the software, please read these notes carefully.
HIOKI E.E. Corporation is the author of COMMUNICATION UTILITY. Except for the purposes of processing data from the 3911-20 or controlling the 3911-20, copying, reproduction, or amendment in whole or in part of COMMUNICATION UTILITY is prohibited by law. For the purposes of product development, COMMUNICATION UTILITY is subject to upgrading without notice. To publish material referencing COMMUNICATION UTILITY, the prior consent of HIOKI is required. The trademark "HIOKI" may not be used. HIOKI cannot accept any responsibility whatever for the results of a customer's operation of COMMUNICATION UTILITY.
______________________________________________________________
Notes on Use
Page 9
_____________________________________________________________________
1
Chapter 1
Overview
1.1 Product Overview
The 3911-20 COMMUNICATION BASE loads data from logger series and sends loaded data to personal computer. Packaged application software allows personal computer to load data from the 3911-20 and analyze loaded data.
Features
(1) Capable of collecting up to 16 channels of data when
connected to logger.
(2) Capable of storing setting conditions (clock, interval,
recording start time, etc.) in memory and loading settings to logger when connected to logger.
(3) LED power indicator indicates battery replacement. (4) Packaged software allows personal computer to load up
to 16 channels of data from logger series. Loaded data can be applied to display graphs and calculate maximum value, minimum value, etc.
______________________________________________________________
Overview
Page 10
2
_____________________________________________________________________
1.2 Name and Functions of Parts
1. SEND button Sends the 3911-20 settings to logger series.
2. RECEIVE
button
3. Optical data
transfer port
Loads recorded data in logger to the 3911-20.
Enables optical data transfer to logger series.
LED condition
POWER Turns on when button is pressed or
connected to personal computer. Starts blinking to indicate weak batteries.
TRANSMITTING Blinks during data transfer.
ERROR Starts blinking to indicate data transfer
error. Press either SEND button or RECEIVE button to correct error.
FULL Blinks to indicate when all 16 channels
(8 channels maximum for the 3636-20/ 3637-20/3638-20/3639-20/3640-20/ 3645-20) of data have been loaded to the 3911-20.
______________________________________________________________
Overview
Page 11
_____________________________________________________________________
3
Chapter 2
Application
2.1 Replacing the Battery
To avoid electric shock when replacing the batteries,
WARNING
first disconnect the cable from the object to be measured. Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result. Handle and dispose of batteries in accordance with local regulations.
Battery Indicator
Under normal conditions LED power indicator is on. When batteries are weak, LED power indicator starts blinking. LED power indicator turns off after approximately 15 seconds with no operation. When LED is blinking, data transfer is enabled with personal computer but not available to logger.
______________________________________________________________
Application
Page 12
4
_____________________________________________________________________
Note on Battery Replacement
Replace batteries when LED Power indicator is OFF. Be sure to send data to be saved to personal computer before replacing batteries although the 3911-20 data and clock settings are saved for approximately one minute after batteries are removed. Do not remove batteries when LED power is on or blinking to avoid losing the 3911-20 data and clock settings. LED power indicator is turned off after approximately 15 seconds with no operation.
(1) Remove cover. Verify polarity and install four new
LR03 alkaline batteries
(2) Close cover.
______________________________________________________________
Application
Page 13
_____________________________________________________________________
5
2.2 Installing Application Software
Packaged application software must be installed in personal computer to use the 3911-20 and logger series.
Operating environment
Computer with Pentium 90 MHz CPU or higher Microsoft Windows98/Me/2000/XP/Vista (32 bit version) At least 32 MB of main memory A display with 800 X 600 dot resolution At least 256 colors At least 4 MB of empty space on the hard disk COM port is necessary
Recommended environment
Computer with Pentium 200 MHz CPU or higher Microsoft Windows2000/XP/Vista (32 bit version) 32 MB of main memory A display with 800 X 600 dot resolution At least 65536 colors At least 4 MB of empty space on the hard disk COM port is necessary
See COMMUNICATION UTILITY operation guide to install and use application software.
______________________________________________________________
Application
Page 14
6
_____________________________________________________________________
2.3 Connecting and Setting PC to the 3911-20
(1) Use RS-232C Cross Cable (the 3911-20 side: D-SUB9
pin) to connect personal computer to the 3911-20.
(2) Start up COMMUNICATION UTILITY from the
Windows start menu.
(3) Go to Communications on the menu bar and choose
Select Communication Port to specify the same port (COM1 to COM6) as serial port on personal computer connected to the 3911-20.
The same communication port settings remain valid until communication is reestablished. Serial port is designated for communications. Locate the mark port in the back of personal computer.
(4) Go to Communications and select Set Time to send
time of personal computer to the 3911-20. Make sure to set correct time in personal computer before sending. After installing and/or replacing batteries, make sure to connect personal computer to the 3911-20 and set current time in the 3911-20 with COMMUNICATION UTILITY before using the 3911-20.
Corresponded cable
The 9637 RS-232C CABLE (1.8 m)
Optional connection cables compatible with the 3911-20 (PC/AT) Connection: reversed
D-sub
9pin female
1DCD 2RxD 3TxD 4DTR 6DSR 7RTS 8CTS 5 GND SHELL
______________________________________________________________
Application
D-sub
9pin female
1DCD 2RxD 3TxD 4DTR 6DSR 7RTS 8CTS 5 GND SHELL
Page 15
_____________________________________________________________________
7
2.4 Connecting the 3911-20 with Logger
Align and insert logger prongs to the 3911-20 slots and fasten logger securely to the 3911-20.
Logger prong
Improper connection between logger and the 3911-20 results communications failure and the 3911-20 error indicator LED appears.
When error occurs in the 3911-20, Press SEND button or RECEIVE button on the 3911-20 to reset. Logger returns to normal status. Open interval setting display again (INTVL displayed) and restart communication.
______________________________________________________________
Application
Page 16
8
_____________________________________________________________________
______________________________________________________________
Application
Page 17
_____________________________________________________________________
Chapter 3
Transferring Data
3.1 Measuring with Logger
Steps for basic measurement with logger
9
Prepare to measure.
See individual logger instruction manual.
Set the 3911-20 measurement conditions.
Use personal computer with application software COMMUNICATION UTILITY to set measurement conditions. After completing, measurement conditions are sent to the 3911-20.
Send setting conditions from the 3911-20 to logger.
Connect the 3911-20 with logger and send setting conditions.
Measure data.
To start measurement with logger, options are manual start and prescheduled start. To end measurement, options are manual stop and one time stop.
Retrieve measurement data from logger with the 3911-20.
Connect the 3911-20 and logger to retrieve measurement data.
See individual logger instruction manuals.
Load batteries in logger and the 3911-20.
Connect sensor and connection cable, etc.
Install application software included with the 3911-20.
Retrieve measurement data from the 3911-20 with personal computer.
Connect personal computer and the 3911-20 to retrieve measurement data. Analyze and save measurement data with COMMUNICATION UTILITY.
______________________________________________________________
Transferring Data
Page 18
10
_____________________________________________________________________
3.2 Setting the 3911-20
Common settings
Time Interval Recording method Start control
(1) Connect personal computer to the 3911-20.
(See Section 2.3,"Connecting and Setting PC to the 3911-20")
(2) Start up COMMUNICATION UTILITY on personal
computer.
(3) Go to 'Communications' on the menu bar and choose
'Set Measurement Conditions'.
(4) When measurement condition setting window is open,
go to '3911 setting item' and configure common logger series settings such as interval and recording preferences.
(5) After setting, select 'Send' to send settings to the 3911-
20.
Each logger requires specific settings and measurement conditions. See Measurement condition settings in COMMUNICATION UTILITY operation guide.
______________________________________________________________
Transferring Data
Page 19
_____________________________________________________________________
3.3 Sending Setting Data
Common settings
Time Interval Recording method Start control
(1) Press logger INTERVAL button lightly to display LCD. (2) When REC mark or clock mark is displayed on logger
LCD, press REC/STOP button for more than one second to stop recording. During recording or waiting for recording start time, error occurs.
(3) Press logger INTERVAL button to open interval setting
display (INTVL displayed).
(4) Connect the 3911-20 to logger
series.
11
(5) Press the SEND button on the 3911-20 momentarily so
that the LED lights, then press and hold the same button for one second to transfer logger series common settings to the logger.
Previously recorded logger data is erased when recording is resumed. Be sure to load data to be saved to the 3911-20 or to personal computer before recording. Communication with the 3911-20 is available even when interval setting display is not displayed except during recording or waiting for recording start time. However, data transfer is not possible when logger is set sleep.
______________________________________________________________
Transferring Data
Page 20
12
_____________________________________________________________________
3.4 Receiving Measurement Data
Collected data (with serial number ) Comments (when set)
(1) Press logger INTERVAL button lightly to display LCD. (2) When REC mark or clock mark is displayed on logger
LCD, press REC/STOP button for more than one second to stop recording. During recording or waiting for recording start time, error occurs.
(3) Press logger INTERVAL button to open interval setting
display (INTVL displayed).
(4) Connect the 3911-20 to logger
series.
(5) Press the RECEIVE button on the
3911-20 momentarily so that the LED lights, then press and hold the same button for one second to transfer measurement data from the logger to the 3911-
20. Logger type determines number of channels applicable in the 3911-20.
Model 3631-20, 3641-20 8000 data X 16 ch (8 units)
Model 3632-20 to 3635-20 16000 data X 16 ch (16 units)
Model 3636-20, 3638-20 (1 ch) 32000 data X 8 ch (8 units)
Model 3636-20, 3638-20 (2 ch) 16000 data X 16 ch (8 units)
Model 3637-20, 3639-20 Model 3640-20, 3645-20
______________________________________________________________
Transferring Data
32000 data X 8 ch (8 units)
Page 21
_____________________________________________________________________
13
3.5 Sending Measurement Data
Collected data
(up to 16 channels)
Comments
(when set)
(1) Connect personal computer to the 3911-20.
(See Section 2.3,"Connecting and Setting PC to the 3911-20")
(2) Start up COMMUNICATION UTILITY. (3) Go to 'Communication' on the menu bar and choose
'Load Data from 3910,11'. Sends measurement data from the 3911-20 to personal computer.
Saving Measurement Data (1) After measurement data is sent from the 3911-20 to
personal computer, data is displayed in graph automatically.
(2) To save data, go to 'File' on the menu bar and choose
'Save'.
(3) When window appears, select save format and save
data. Binary: Saves data in format compatible to
COMMUNICATION UTILITY.
Text : Saves data in text format that can be exported
to spread sheet and other application software. Cannot be exported to COMMUNICATION UTILITY.
See Save measurement data in COMMUNICATION UTILITY operation guide to save data.
______________________________________________________________
Transferring Data
Page 22
14
g
_____________________________________________________________________
3.6 Confirming and Clearing Loaded Data
Press SEND button and Load button briefly simultaneously (within one second) . With a beep sound, LED turns on or starts blinking. Number of channels loaded when LED is on or blinking is displayed. See below for details. When LED is on or blinking, press SEND button for more than one second to clear all data with a beep sound.
:OFF
: Blinkin
:ON
No data is saved.
Data equal to 1 channel is saved.
Data equal to 2 channel is saved.
Data equal to 3 channel is saved.
Data equal to 4 channel is saved.
Data equal to 5 channel is saved.
Data equal to 6 channel is saved.
Data equal to 7 channel is saved.
Data equal to 8 channel is saved.
Data equal to 9 channel is saved.
Data equal to 10 channel is saved.
Data equal to 11 channel is saved.
Data equal to 12 channel is saved.
Data equal to 13 channel is saved.
Data equal to 14 channel is saved.
Data equal to 15 channel is saved.
Data equal to 16 channel is saved.
With Models 3636-20 and 3638-20, when measurement data from a single channel (CH1 or CH2) is received by the 3911-20, the LEDs on the 3911-20 indicate that data for two channels is present.
______________________________________________________________
Transferring Data
Page 23
_____________________________________________________________________
Chapter 4
Specifications
15
LED condition
Buzzer sound
Connection
Transfer rate (reference value)
Recording capacity
Power supply
Maximum rated power
Battery life
Dimensions
Mass
Location for use
Operate temperature and humidity range
POWER, TRANSMITTING, ERROR, FULL
Once (when executing key click) Three times (when error occurs)
Logger: Infrared optical data transfer (synchronous serial connection, three wire type) Personal computer: RS-232C (asynchronous serial connection, two wire
type) 
The 3911-20 to Logger 250 data/s The 3911-20 to PC 1000 data/s (19200 bps)
16 channels (16,000 data X 16 channels) 8 channels (32,000 data X 8 channels)
LR03 alkaline battery 1.5 V X4
0.2 VA
About 80 days (at 20 , when not in operation) Transferring data about 100 times (at 20
, receiving and sending maximum 16000 X 16 channels of data from logger to personal computer)
Approx.68.5W X 92H X 36D mm (excluding projections) (2.7"WX 3.62"H X 1.42"D)
Approx.150 g (5.3 oz.) (including batteries)
Indoors, altitude up to 2000 m (6562-ft.)
-20 to 70 (-4 to 158 ), 80% RH or less (no condensation) 0to40 transfer with personal computer using RS­232C
(32 to 104 ) during data
______________________________________________________________
Specifications
Page 24
16
_____________________________________________________________________
Storage temperature and humidity range
Water Resilience
Accessories
Option
Standards
Applying 
Safety EMC
-20 to 70 (-4 to 158 ), 80% RH or less (no condensation)
none
LR03 alkaline battery X 4 Instruction Manual CD-R
9637 RS-232C CABLE
EN61010 Pollution Degree 2 EN61326
______________________________________________________________
Specifications
Page 25
_____________________________________________________________________
17
Chapter 5
Maintenance and
Service
Cleaning
To clean the product, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
Service
If the product seems to be malfunctioning, confirm that the batteries are not discharged before contacting your dealer or Hioki representative. Pack the product carefully so that it will not be damaged during shipment, and include a detailed written description of the problem. Hioki cannot be responsible for damage that occurs during shipment.
______________________________________________________________
Maintenance and Service
Page 26
18
_____________________________________________________________________
Troubleshooting
Problem
Button press does not turn LED power indicator on.
Check and see if the batteries are properly installed and not weak.
Problem
Button press does not enable sending settings to logger nor receiving data from logger.
Is logger properly connected to the 3911-20? When logger is not properly connected, error occurs during optical data transfer turning LED error indicator on.
Is LED power indicator blinking? When batteries are weak, LED power indicator starts blinking and communication with logger cannot be established.
Is logger in recording status? When logger is recording, data cannot be transferred. Executing data transfer during recording results in error and LED error indicator starts blinking.
Is logger set to sleep? Press logger INTERVAL button and display interval setting display.
Are logger batteries weak? Communication cannot be established when logger batteries are weak. Replace with new batteries.
Is LED datafull indicator blinking? LED datafull indicator starts blinking to indicates maximum data has been loaded to the 3911-20. Transfer data to personal computer. Although LED datafull indicator is not blinking, when the 3911-20 already has 15 channels of data attempting to load data from the 3631-20 (2 ch), the 3641-20 (2 ch), the 3636-20/3637­20/3638-20/3639-20/3640-20/3645-20 results in error and LED error indicator starts blinking.
When interval setting saved in the 3911-20 is set at 1 second, interval setting cannot be sent to the 3631-20 to the 3635-24/-25/­26 and the 3641-20 which minimum interval is at 2 seconds.
______________________________________________________________
Maintenance and Service
Page 27
_____________________________________________________________________
-MEMO-
19
______________________________________________________________
Maintenance and Service
Page 28
20
_____________________________________________________________________
-MEMO-
______________________________________________________________
Maintenance and Service
Page 29
Page 30
Page 31
HIOKI 3911-20 COMMUNICATION BASE
Instruction Manual
Publication date: May 2008 Revised edition 9
Edited and published by HIOKI E.E. CORPORATION
Technical Support Section
All inquiries to International Sales and Marketing Department
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL: +81-268-28-0562 / FAX: +81-268-28-0568
E-mail: os-com@hioki.co.jp
URL http://www.hioki.com/
Printed in Japan 3911A981-09
All reasonable care has been taken in the production of this manual, but if you find any points which are unclear or in error, please contact your supplier or the International Sales and Marketing Department at HIOKI headquarters. In the interests of product development, the contents of this manual are subject to revision without prior notice. The content of this manual is protected by copyright. No reproduction, duplication or modification of the content is permitted without the authorization of Hioki E.E. Corporation.
Page 32
HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp / URL http://www.hioki.com/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA TEL +1-609-409-9109 / FAX +1-609-409-9108
3911A981-09 08-05H