Thank you for purchasing the HIOKI “Model 3664 OPTICAL
POWER METER.” To obtain maximum performance from the
instrument, please read this manual first, and keep it handy for
future reference.
Trademarks:
Microsoft and Windows are either registered trademarks or trademarks of Microsoft Corporation in the United States and other
countries.
1
1
2
Inspection
Verifying Package Contents
Options
• When you receive the instrument, inspect it carefully to ensure
that no damage occurred during shipping. In particular, check the
accessories, panel switches, and connectors. If damage is evi
dent, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
• Use the original packing materials when transporting the instrument, if possible.
Model 9246CARRYING CASE
Model 9742OPTICAL SENSOR
Model 9742-10 OPTICAL SENSOR
Model 9743OPTICAL SENSOR
Model 9743-10 OPTICAL SENSOR
Model 9445-02 AC ADAPTER
Model L9094OUTPUT CORD
3
4
-
5
6
7
8
9
10
11
2
Safety Information
This instrument is designed to comply with IEC 61010 Safety
Standards, and has been thoroughly tested for safety prior to
shipment. However, mishandling during use could result in
injury or death, as well as damage to the instrument. Be cer
tain that you understand the instructions and precautions in
the manual before use. We disclaim any responsibility for
accidents or injuries not resulting directly from instrument
defects.
Safety Symbols
This manual contains information and warnings essential for safe
operation of the instrument and for maintaining it in safe operating
condition. Before using it, be sure to carefully read the following
safety precautions.
In the manual, the symbol indicates particularly
important information that the user should read before
using the instrument.
The symbol printed on the instrument indicates that
the user should refer to a corresponding topic in the
manual (marked with the
relevant function.
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents an
extreme hazard that could result in serious injury or
death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or
death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument.
Indicates advisory items related to performance or
correct operation of the instrument.
-
symbol) before using the
Other Symbols
Indicates the prohibited action.
Indicates the location of reference information.
Operating Precautions
Direct
Avoid the following
Sunlight
High Temperature
High Humidity
3
Follow these precautions to ensure safe operation and to obtain
the full benefits of the various functions.
Observe the following to avoid damage to the instrument.
• Do not allow the instrument to get wet, and do not take measurements with wet hands. This may cause an electric shock.
• Do not use the instrument where it may be exposed to corrosive
or combustible gases. The instrument may be damaged or
cause an explosion.
• To avoid damage to the instrument, protect it from physical
shock when transporting and handling. Be especially careful to
avoid physical shock from dropping.
• Do not store or use the instrument where it could be exposed to
direct sunlight, high temperature or humidity, or condensation.
Under such conditions, the instrument may be damaged and
insulation may deteriorate so that it no longer meets specifica
tions.
• This instrument is designed for use indoors. It can be operated
at temperatures between 0 and 40°C without degrading safety.
• After use, always turn OFF the power.
• The “ ” indicator appears when battery voltage becomes low.
Replace the batteries as soon as possible.
• Use the specified battery (AA alkaline battery) only. If a nonspecified battery such as a manganese battery is used, the
service hours may be shortened.
• When connecting the instrument to a PC with a USB cable,
use a cable of 2 m or shorter to prevent the influence of noise.
1
2
3
4
5
-
6
7
8
9
10
11
4
Cords and Cables
Optical Sensor
• Avoid stepping on or pinching cables, which could damage the
cable insulation.
• To avoid breaking the cables, do not bend or pull them.
• To avoid damaging the output cable, grasp the connector, not
the cable, when unplugging the cable.
• Use only the specified optical sensor. Using a non-specified sensor may result in incorrect measurements due to poor connection or other reasons.
• To prevent damage to the instrument and sensor, never connect
or disconnect a sensor while the power is on, or while the sensor
is connected around a conductor.
• To avoid damaging the optical sensor and to ensure accurate
measurements, avoid dropping or applying any physical shock
to the optical sensor.
• Avoid touching the detector window with your bare hands. The
detector window must be clean for the sensor to meet the
specified performance parameters.
• Avoid scratching the detector window with sharp or pointed
objects (e.g., the tips of tweezers) or against hard surfaces.
Damage to the detector window may prevent the sensor from
meeting specified performance parameters.
• Make sure no stress is applied to the repeater cable of the 9742-
10, 9743-10 between the sensor main unit and its detector
element.
To prevent dust accumulation, stains, and damage, cover the
detector window with the sensor cover when the sensor is not in
use.
5
Care and Handling of CD-R
• Always hold the disc by the edges, so as not to make fingerprints on the disc or scratch the printing.
• Never touch the recorded side of the disc. Do not place the disc
directly on anything hard.
• Do not wet the disc with volatile alcohol or water, as there is a
possibility of the label printing disappearing.
• To write on the disc label surface, use a spirit-based felt pen. Do
not use a ball-point pen or hard-tipped pen, because there is a
danger of scratching the surface and corrupting the data. Do not
use adhesive labels.
• Do not expose the disc directly to the sun's rays, or keep it in
conditions of high temperature or humidity, as there is a danger
of warping, with consequent loss of data.
• To remove dirt, dust, or fingerprints from the disc, wipe with a dry
cloth, or use a CD cleaner. Always wipe radially from the inside
to the outside, and do no wipe with circular movements. Never
use abrasives or solvent cleaners.
• Hioki shall not be held liable for any problems with a computer
system that arises from the use of this CD-R, or for any problem
related to the purchase of a Hioki product.
Preliminary Checks
• Before using the instrument, make sure that the insulation on the
optical sensor cable and USB cable is undamaged and that no
bare conductors are improperly exposed. Using the instrument
in such conditions could cause an electric shock, so contact your
dealer or Hioki representative for repair.
• Before using the instrument the first time, verify that it operates
normally to ensure that the no damage occurred during storage
or shipping. If you find any damage, contact your dealer or Hioki
representative.
1
2
3
4
5
6
7
8
9
10
11
6
1.1 Product Overview
Overview1
1.1Product Overview
The 3664 OPTICAL POWER METER is an instrument for measuring the optical power of a spatial light. It is suitable for measurement of the optical power of a laser light source and an LED light
source. The 3664 is a hand-held meter equipped with various mea
surement functions and interfaces that meet diversified measurement needs. The optical sensors (optional) for the power meter are
interchangeable, as the sensitivity of each sensor is adjusted inde
pendently.
7
1
2
3
-
4
-
5
6
7
8
9
10
11
8
1.2 Features
1.2Features
Scaling function
The reading of the 3664 OPTICAL POWER METER can be converted to the reading of another optical power meter by a simple
operation.
Various measurement functions
In addition to the optical power measurement (absolute-value measurement), the 3664 power meter has relative-measurement, maximum-value-hold, minimum-value-hold, and averaging functions.
With all these functions, the display unit can be switched between
W and dBm (dB).
Automatic correction of sensitivity through input of
the measurement wavelength
When the measurement wavelength is entered, the wavelength
sensitivity of the sensor connected to the 3664 is automatically cor
rected for measurement display.
Interface with a computer
The USB interface is provided as standard equipment. The 3664
can be controlled on the PC, its settings can be edited on the PC,
or its measurement data can be uploaded to the PC.
-
Analog output terminal
The 3664 has an analog output terminal. It outputs in the voltage
form the output signal of the optical sensor. Use this terminal, for
example, when performing monitoring using a DMM or recording
data for extended periods.
Auto-power-save function
When the 3664 is running on the battery, if there is no key operation or USB communication for approximately 10 minutes, the
power to the 3664 will be automatically turned off to conserve bat
tery power.
-
1.3 Names and Functions of Parts
Relative-measurement
mode
Front Panel
Operating Keys
Display
Strap installation
Slot
Relative-measurement
indicator
Measurement
wavelength
Display
Battery mark
Auto power save mark
Maximum, Minimum, Average
Range
Main display
Sub display
1.3Names and Functions of Parts
9
1
2
3
4
5
6
7
8
9
10
11
10
1.3 Names and Functions of Parts
Operating Keys
POWER
λ
REL
RANGE
SET
SAVE
W/dBm
▼
MAX/MIN/AVE
▲
Turns ON/OFF power to the 3664.
Power ON: Hold down the key for approximately 2 seconds to
turn the power ON (auto-power-save function ON).
Hold down the key for approximately 5 seconds to
turn the power ON and the auto-power-save function
is turned OFF.
Power OFF:Hold down the key for approximately 1 second to turn
the power OFF.
Sets the measurement wavelength to the preset wavelength.
Switches between optical-power measurement mode, relativemeasurement mode (measured-value basis) and relative-mea
surement mode (input-value basis).
Switches between auto range and manual range.
Moves to the setting edit screen. "SEt" appears on the screen.
Moves the digit to edit the setting (the digit blinks) during setting
edit.
Confirms the setting during setting edit and returns the meter to
measurement mode.
Switches the unit of the reading between "W" and "dBm." In relative-measurement mode, the unit is switched between "W" and
"dB."
Decreases the number during setting edit.
Switches the value to be displayed between the measurement,
maximum, minimum, and average value.
Increases the number during setting edit.
-
1.3 Names and Functions of Parts
Upper Panel
Analog output terminal
Connect the L9094 OUTPUT CORD
included among the accessories to
this terminal. Output signals from the
optical sensor are converted to a voltage and output from this terminal.
AC-adapter terminal
Connect the 9445-02 or 9445-03 AC
ADAPTER included among the accessories to this terminal. When the AC adapter
is used, power is supplied to the 3664 from
a commercial power supply.
Optical-sensor
terminal
Connect an optional
optical sensor to this
terminal.
Bottom Panel
USB terminal
Connect a USB cable to this terminal.
11
1
2
3
4
5
6
7
8
9
10
11
12
1.3 Names and Functions of Parts
2.1 Attaching the Strap
1. Insert each end of the Strap through
an installation slot on the instrument.
2. Feed each end of the Strap through its
clasp.
3. Feed each end through its stopper.
Confirm that the strap is not
slackened or twisted.
Model 3664
Measurement Preparations2
2.1Attaching the Strap
Use the strap to carry the 3664 power meter or hang it on a hook.
13
1
2
Be sure to attach the strap to the 3664 power meter in place (2 locations) properly. If the strap is not properly attached, the meter may drop
and be broken.
3
4
5
6
7
8
9
10
14
1. Connect the output plug of the AC
adapter to the AC-adapter terminal.
2. Connect the AC adapter to an AC
outlet.
AC Adapter Connection Terminal
2.2 Supplying Power
2.2Supplying Power
2.2.1Using the AC Adapter
• Use only the specified Model 9445-02 AC ADAPTER. AC
adapter input voltage range is 100 to 240 VAC (with ±10%
stability) at 50/60 Hz. To avoid electrical hazards and dam
age to the instrument, do not apply voltage outside of this
range.
• Turn the instrument off before connecting the AC adapter
to the instrument and to AC power.
To avoid damaging the cord, grasp the plug, not the cord, when unplugging it.
-
• When the AC adapter is connected and the battery is installed,
the AC adapter is used. In case of power outage, the 3664 is
automatically switched to the battery drive.
• Make sure the power is turned off before connecting or disconnecting the AC adapter.
2.2.2Using the Batteries
1. Confirm that the power to the 3664 is
turned OFF.
2. Disconnect all cables.
3. Turn over the 3664 and loosen the
screw on the battery cover.
4. Remove the battery cover.
5. Insert the four AA alkaline batteries
(LR6) into the battery case, noting the
correct polarity. (When replacing the
batteries, replace all of them.)
6. Place the battery cover back on, and
tighten the screw.
Tools and materials:
Four AA alkaline batteries (LR6) and a Phillips screwdriver
Screw
15
2.2 Supplying Power
Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery
leakage could result.
• The “ ” indicator appears when battery voltage becomes low.
Replace the batteries as soon as possible.
• Do not use other than the specified type of batteries (LR6 alkaline batteries). Using manganese batteries will result in a
shorter operating time than when using alkaline batteries.
• The 3664 power meter runs on battery or AC power. When the
AC adapter is connected and the battery is installed, the AC
adapter is used.
1
2
3
4
5
6
7
8
9
10
16
Grasp the connector.
Optical Sensor
Optical Sensor Connection Terminal
2.3 Connecting the Optical Sensor
2.3Connecting the Optical Sensor
• Use only the specified optical sensor. Using a non-specified sensor may result in incorrect measurements due to poor connection or other reasons.
• To prevent damage to the instrument and sensor, never connect
or disconnect a sensor while the power is on, or while the sensor
is connected around a conductor.
• To avoid damaging the sensor cable, grasp the plug, not the
cable, when unplugging it.
• To avoid ensuring accurate measurements, avoid dropping or
applying any physical shock to the optical sensor.
Insert an optical sensor, with the surface marked with an arrow facing upward, into the optical-sensor terminal of the 3664.
Always grasp the connector with the arrow to connect or disconnect the optical sensor. (You must grasp the connector to disconnect the sensor.)
2.4 Connecting the Analog Output Cord
Analog Output Cord
2.4Connecting the Analog Output Cord
17
To avoid damage to the instrument, do not short-circuit the output terminal and do not input voltage to the output terminal.
By connecting the analog output cord to the analog output terminal,
the optical power detected by the optical sensor can be converted
to a voltage in order to monitor the signal.
Connect the analog output cord to the analog output terminal.
Signals output from the analog output terminal are detected by
an optical sensor, with its wavelength sensitivity uncorrected.
This signal, therefore, is not proportional to the measurement
displayed on the screen.
1
2
3
4
5
6
7
8
9
10
18
AC Adapter Connection Terminal
2.5 Turning the Power On and Off
2.5Turning the Power On and Off
Power On
Hold down the POWER key for approximately 2 seconds. The
screen lights up and the power is turned ON.
If the battery indicator appears (" " at the top right of the
screen) after the power has been turned ON, the battery is running down. Replace the battery (when the 3664 is running on the
battery).
See Section 7.7 "Replacing the Batteries" (54 page)
Power OFF
Hold down the POWER key for approximately 1 second. The
screen goes out and the power is turned OFF.
Auto-power-save function
• This function automatically turns OFF the power to the 3664 if
there is no key operation or USB communication for approxi
mately 10 minutes, when it is running on the battery. (The indication "APS" lights up at the top right of the screen.)
• To turn OFF the auto-power-save function, hold down the
POWER key for 5 seconds when turning ON the power. "APS"
goes out and the auto-power-save function is turned OFF.
• When the 3664 is running on AC power, the auto-power-save
function is automatically turned OFF and the function is not avail
able.
-
-
• While power is supplied from the battery, when the power is
turned ON, the auto-power-save function is turned ON.
• When the USB is used, use the AC adapter for the power supply.
• After use, always turn OFF the power.
Measurement3
Input of light to the sensor
Setting of the range
Setting of the measurement
(20 page)
(21 page)
(24 page)
(26 page)
Setting of the measurement wavelength
• Optical-power measurement and relative measurement of special light are performed using the 3664 power meter.
• The measurement wavelength may be set to an arbitrary wavelength.
• The measurement wavelength may also be set to a previously
set wavelength.
Connect an optional optical sensor to the 3664. To operate the
3664, an optical sensor must be connected to it.
19
1
2
4
Flow of measurement
5
6
7
8
9
10
20
The 3664 sets the optical power
of the incident light to the optical
sensor as the reference value,
and measures the difference between the set reference value
and the current incident optical
power.
The 3664 measures the difference between the reference value arbitrarily set by key entry
and the current incident optical
power.
Optical-power measurement
Relative measurement (measured-value basis)
Relative measurement (input-value basis)
The 3664 performs optical-power measurement.
3.1 Setting the Measurement Mode
3.1Setting the Measurement Mode
Each time the REL key is pressed, the measurement mode
changes as shown below.
3.2Setting the Range
Auto Range
Manual Range
21
3.2 Setting the Range
Each time the RANGE key is pressed, auto range and manual
range alternate.
3.2.1Auto Range
The appropriate range for the incident optical power is automatically selected. The range display area will appear as shown below.
1
2
4
5
6
7
8
When the RANGE key is pressed, "MANUAL" appears in the range
display area and the range is fixed to that displayed.
9
10
22
Blinking
3.2 Setting the Range
3.2.2Manual Range
The range is set manually. "MANUAL" appears in the range display
area.
1.Press the SET key to display the setting edit screen.
2.Press the RANGE key to display the setting edit screen for
the manual range. The range display area will blink.
3.Use the ▲ and ▼ keys to change the range setting.
▲ : Increases the sensitivity range by one increment
▼ : Decreases the sensitivity range by one increment
The range No. displayed on the screen is the range No. of the internal
circuit. The range is divided into seven levels, from 1 (lowest
sensitivity) to 7 (highest sensitivity).
23
Fixed range
"Lo" Display
"Hi" Display
3.2 Setting the Range
4.Press the SAVE key to save the setting. "SAVE" is displayed;
the 3664 will return to the measurement screen. The range
is now fixed.
1
2
4
5
When the range is fixed through manual range setting, the measurement display shows "Lo" if the incident optical power falls
below the measurement range, and it shows "
exceeds the range.
Hi" if the power
6
7
8
9
10
24
Blinking
3.3 Setting the Wavelength
3.3Setting the Wavelength
The following two methods can be used:
• Set an arbitrary measurement wavelength.
• Retrieve a wavelength from the measurement-wavelength memory, and set it as the measurement wavelength.
• The selectable wavelength range depends on the specification
of the optical sensor connected to the 3664.
• To ensure proper measurement, set the wavelength to the
wavelength of the light used.
3.3.1Setting an Arbitrary Measurement Wavelength
1.Press the SET key to display the setting edit screen.
2.Press the λ key to display the setting edit screen for the
measurement wavelength.
3.Use the ▲ and ▼ keys to change the measurement wave-
length.
▲ : Increases the number
▼ : Decreases the number
3.3 Setting the Wavelength
Blinking
The measurement wavelength set
4.Press the key to change the blinking digit.
Repeat steps 3 and 4 above to set the measurement wavelength.
5.Press the SAVE key to complete setting.
"SAVE" appears; the 3664 will then return to the measurement
screen. The measurement wavelength is now set.
25
1
2
4
5
6
• An arbitrarily set measurement wavelength is automatically
saved in the measurement-wavelength memory.
• The measurement-wavelength memory holds up to 10 wavelengths (including preset wavelengths for each optical sensor).
• Each optical sensor has preset wavelengths. These preset
wavelengths cannot be edited.
• Through the setting operation for the arbitrary measurement
wavelength, the setting data is automatically saved in the mea
surement-wavelength memory. When it is necessary to save a
new measurement wavelength to the measurement-wave
length memory when the memory already contains 10 wavelengths, the oldest data is deleted in order to make space for
the new data.
• To reset the data in the measurement-wavelength memory,
perform measurement-wavelength-memory reset. This will
return the memory to its default state for each sensor.
See Section 7.4.3 "Measurement-Wavelength Memory Reset" (52
page)
7
8
9
-
-
10
26
1.When using the 9742 OPTICAL
SENSOR, slide the sensor cover
of the optical sensor until it clicks
to lock the cover. When using the
9742-10, 9743 and 9743-10
OPTICAL SENSOR, remove the
optical-sensor cover.
When using the 9742-10, 9743
and 9743-10 OPTICAL SENSOR,
remove the cover of the optical
sensor.
2.Inject light.
3.The 3664 displays the measured
optical power.
• Removed sensor cover (Be
careful to avoid misplacing.)
• Mount the sensor cover as
shown in the figure to the left.
Model 9742-10, 9743 and 9743-10
OPTICAL SENSOR
Optical Sensor
Cover
Optical Sensor
Cover
The 9742-10 OPTICAL SENSOR is
shown in the diagram.
Model 9742 OPTICAL SENSOR
3.4 Starting Measurement
3.3.2Retrieving the Wavelength Setting from the Measurement-Wavelength Memory
Each time the λ key is pressed, the measurement wavelengths
stored in the memory are retrieved individually in order.
To reset the data in the measurement-wavelength memory, perform measurement-wavelength-memory reset. This will return
the memory to its default state for each sensor.
See Section 7.4.3 "Measurement-Wavelength Memory Reset" (52 page)
3.4Starting Measurement
27
3.4 Starting Measurement
• To ensure accurate measurements, make sure the detector
window is free of dust, stains, and any damage.
• Confirm that the light to be measured fits completely within the
detector window.
• If the optical power is too intense for the unit area, such as
when a condensed beam on the focus plane is measured, the
optical sensor may be saturated and the 3664 may not be able
to conduct accurate measurement. This problem can be
solved by changing the distance between the light source and
the detector window of the optical sensor.
• The operation of the light source may become unstable due to
the reflection of light from the detector window of the optical
sensor. This problem can be solved by angling the detector
window slightly to prevent the reflection from returning to the
light source.
• If background light, sunlight, illumination light, or any other light
than the light to be measured is overly intense, the 3664 may
not be able to conduct accurate measurement. Take appropri
ate light-shielding measures.
1
2
4
-
5
6
7
8
9
10
28
Optical-power
measurement
3.5 Setting Example
3.5Setting Example
3.5.1Optical-Power Measurement
The 3664 performs optical-power measurement.
(1) Set the measurement mode.
Press the REL key to select the optical-power measurement mode.
(2) Set the measurement wavelength.
Set the measurement wavelength to an arbitrary wavelength or
one retrieved from the memory. The sensitivity is corrected by the
set measurement wavelength, and the 3664 displays the mea
sured value adjusted in accordance with the corrected sensitivity.
-
Set the measurement wavelength to the wavelength of the light
used. If the wavelengths differ, the 3664 is not able to conduct
accurate measurement.
3.5 Setting Example
Optical Sensor
Cover
1.When using the 9742 OPTICAL SEN-
SOR, slide the sensor cover of the
optical sensor until it clicks to lock the
cover. When using the 9742-10, 9743
and 9743-10 OPTICAL SENSOR,
remove the optical-sensor cover.
When using the 9742-10, 9743 and
9743-10 OPTICAL SENSOR, remove
the cover of the optical sensor
2.Inject light.
3.In the relative-measurement (mea-
sured-value basis) mode, hold down the
SAVE key for approximately 3 seconds.
• Removed sensor cover (Be
careful to avoid misplacing.)
• Mount the sensor cover as
shown in the figure to the left.
Optical Sensor
Cover
Model 9742 OPTCAL SENSOR-
The 9742-10 OPTICAL SENSOR is
shown in the diagram.
Model 9742-10, 9743 and 9743-10
OPTICAL SENSOR
3.5.2Relative Measurement (Measured-value basis)
The 3664 sets the optical power of the incident light to the optical
sensor as the reference value, and measures the difference
between the set reference value and the current incident optical
power.
(1) Set the measurement mode.
Press the REL key to select the relative-measurement (measuredvalue basis) mode.
29
1
2
4
(2) Set the measurement wavelength.
(3) Loading the reference value/relative measurement (when the optional
Set the measurement wavelength to an arbitrary wavelength or
one retrieved from the memory.
Set the measurement wavelength to the wavelength of the light
used. If the wavelengths differ, the 3664 is not able to conduct
accurate measurement.
9742, 9742-10, 9743 and 9743-10 OPTICAL SENSOR is used)
5
6
7
8
9
10
30
Mode
Reference Value
3.5 Setting Example
4."SAVE" is displayed on the screen, and the relative-mea-
surement result is then displayed.
The result is saved as the optical-power measurement reference value.
5."0" appears on the left side of the sub-display area, indicat-
ing the relative-measurement (measured-value basis)
mode, and the saved reference value is displayed on the
right side of the sub-display area.
The main display area shows the relative-measurement
value.
In the relative-measurement (measured-value basis) mode,
the value shown below is displayed.
(Current measurement) - (measurement set as the reference value)
• Only one reference value is saved, regardless of the wavelength setting. When a different measurement wavelength is
used, set a new reference value.
• When the maximum-value-hold, minimum-value-hold, or averaging function is ON, "(calculation result) - (reference value)" is
displayed.
• When the scaling function is ON, "(value after scaling) - (reference value)" is displayed.
• When the reference value is input, if the main display area
shows "
The relative measurement is not calculated, and "
played on the screen.
Lo" or "Hi," Lo or Hi is saved as the reference value.
- - - -" is dis-
3.5 Setting Example
Blinking
Blinking
Blinking
3.5.3Relative Measurement (Input-value basis)
31
The 3664 measures the difference between the reference value
arbitrarily set by key entry and the current incident optical power.
Entering the Reference Value Using the Keys/Relative
Measurement
1.Press the SET key to display the setting edit screen.
2.Press the REL key to display the reference-value input
screen.
3.Use the ▲ and ▼ keys to change the position of the decimal
point and the unit.
▲ : Increases the number
▼ : Decreases the number
1
2
4
5
6
7
8
4.Press the key to change the blinking digit.
9
10
32
Blinking
Reference Value
Mode
3.5 Setting Example
5.Use the ▲ and ▼ keys to change the blinking number.
▲ : Increases the number
▼ : Decreases the number
Repeat steps 4 and 5 above to set the reference value.
6.Press the SAVE key to complete setting.
"SAVE" appears and then the 3664 returns to the measurement screen. The measurement wavelength is now set and
the relative-measurement result is displayed.
The number "1" appears on the left side of the sub-display area,
indicating the relative-measurement (input-value basis) mode, and
the saved reference value is displayed on the right side of the subdisplay area.
The main display area displays the relative-measurement value.
In the relative-measurement (input-value basis) mode, the value
shown below is displayed.
(Current measurement) - (value entered using the keys as the reference value)
• Only one reference value is saved regardless of the wavelength setting.
• When a different measurement wavelength is used, set a new
reference value.
• The values that can be entered using the keys as the reference
value range from 0.001 nW to 1.9999 W (-90.00 dBm to 33.0
dBm).
• When the maximum-value-hold, minimum-value-hold, or averaging functions is ON, "(calculation result) - (reference value)"
is displayed.
• When the scaling function is ON, "(value after scaling) - (reference value)" is displayed.
4.1 Scalling Function
Functions4
• Using the scaling function, a measured value can be converted
to an arbitrary value and the converted value displayed.
• The maximum-value-hold, minimum-value-hold, and averaging
functions are available.
33
1
2
4.1Scalling Function
• A measured value is converted by scaling to an arbitrary value
and displayed. When the scaling function is turned ON, "(mea
sured value) x (internal coefficient)" is displayed.
• The internal scaling coefficient for each wavelength is saved in
the memory.
• When the display value after scaling is set, the internal coefficient
is automatically calculated and saved.
4.1.1Turning ON the Scaling Function
Save the internal scaling coefficient, and turn ON the scaling function.
1.Input the light to be measured to the optical sensor.
2.Press the SET key to display the setting edit screen.
-
3
4
5
6
7
8
9
10
34
Scaling function
Measured value immediately before
the setting edit screen appears
Blinking
Blinking
Sub Display
Blinking
4.1 Scalling Function
3.Press the λ key to display the setting edit screen for the
measurement wavelength.
The value displayed first in the sub-display area is the display value with the scaling OFF (internal coefficient of
1.000).
4.Use the key to move the cursor to a digit in the sub-dis-
play area.
5.Use the ▲ and ▼ keys to change the blinking number.
▲ : Increases the number
▼ : Decreases the number
Repeat steps 4 and 5 above to set the display value after
scaling.
35
Scaled Value
Blinking
4.1 Scalling Function
6.Press the SAVE key to save the setting.
"SAVE" is displayed, and the 3664 then returns to the measurement screen. The scaling coefficient is automatically
calculated and saved as an internal setting.
7.The indicator "λ" starts blinking and the scaled value is dis-
played.
1
2
3
4
• While the scaling function is ON, the indication "λ" blinks.
• The internal scaling coefficient ranges from 0.100 to 10.000.
The scaling function is not available under the following conditions, and "Hi" or "Lo" appears instead of the display value after
scaling (in the sub-display area).
• The measured value is displayed as "Hi" or "Lo" in the manual
range.
• The measured value is very large or very small in the auto
range.
5
6
7
8
9
10
36
Display value before scaling
Blinking
Scalling Function
4.1 Scalling Function
4.1.2Turning OFF the Scaling Function
1.Press the SET key to display the setting edit screen.
2.Press the λ key to display the setting edit screen for the
measurement wavelength.
3.Press the SAVE key without changing the display value
before scaling.
"SAVE" is displayed, and the 3664 then returns to the measurement screen. A scaling coefficient of 1.000 is set as an
internal setting, and the scaling function is turned OFF.
4.The indication "λ" remains ON as a solid light, and the
measured value that is not scaled is displayed.
To return to the measurement screen with the scaling ON:
Do not press the SAVE key in step 4 above; instead, press the
SET key twice.
4.2 Maximum-Value Hold / Minimum-Value Hold / Averaging
The calculation function is
disabled.
The maximum value among
the current measurements
is displayed.
The minimum value among
the current measurements
is displayed.
The moving average for the
set number of averaging
steps is displayed.
No calculation
Maximum-value hold
Minimum-value hold
Average-value hold
4.2Maximum-Value Hold / Minimum-Value
Hold / Averaging
37
Each time the MAX/MIN/AVE key is pressed, the calculation setting
changes as shown below.
1
2
3
4
5
6
7
8
To reset the display value, press the MAX/MIN/AVE key again and
start calculation again.
9
10
38
Blinking
Blinking
4.2 Maximum-Value Hold / Minimum-Value Hold / Averaging
Setting the Number of Averaging Steps
1.Press the SET key to display the setting edit screen.
2.Press the MAX/MIN/AVE key to display the setting edit screen
for the number of averaging steps.
3.Use the ▲ and ▼ keys to change the blinking number.
▲ : Increases the number
▼ : Decreases the number
A digit may be chosen by moving the cursor using the
key.
4.Press the SAVE key to save the setting.
"SAVE" is displayed, and the 3664 then returns to the measurement screen. The number of averaging steps is now
set.
• After returning to the measurement screen, "AVE" continues to
blink until the number of averaging steps reaches the set num
ber.
• The number of averaging steps is selectable between 2 and
100.
• If the MAX/MIN/AVE key, λ key, RANGE key, or SET key is
pressed during measurement with the calculation function ON,
the calculation is reset.
-
Control by Computer5
By connecting the 3664 power meter to a PC, measurements can
be read and the 3664 can be set up on the PC.
• When connecting the 3664 to a PC with a USB cable, use a
cable of 2 m or shorter to prevent the influence of noise.
• The 3664 is in remote mode when connected to a PC. In the
remote mode, a "r" appears on the bottom left corner of the dis
play, and only the POWER key operation is available. To deactivate the remote mode, disconnect the USB cable.
39
1
2
3
-
4
6
7
8
9
10
40
5.1 Overview of Control
5.1Overview of Control
• By connecting the 3664 power meter to a PC with a USB interface and calling API functions stored in the DLL from the application, the 3664 can be remotely controlled, and data can be
exchanged between the 3664 and the PC.
System requirements
OS
HDD space10 MB or more of free disk space
InterfaceUSB Ver. 1.1 or later
CD-R contents
EnglishDriver
Windows XP, Windows Vista, Windows 7
(64-bit support limited to Windows 7)
Hardware requirements, such as those involving
the CPU, RAM, and display monitor, shall con
form to the system requirements of the OS.
DLL is a 32-bit library. It is only compatible with
32-bit applications.
(connectable to only one PC)
-
• For details, see the "API Instructions Manual (3664api.pdf)" on
the supplied CD-R.
41
5.1 Overview of Control
The table below shows a list of the API functions.
API list
Name of functionDescription
RESET functions
ResetDeviceUSB()Resets the communications function
ResetAllData()Resets the setting data
ResetAllWave()Resets the measurement wavelength
SET functions
SET_Rel0()Sets the reference value for relative measurement
(measured-value basis)
SET_Rel1()Sets the reference value for relative measurement
(input-value basis)
QUES functions
QUES_IDN()Queries the device data
QUES_Sensor()Queries the optical-sensor model No.
QUES_AllWave()Queries all measurement wavelengths stored in
the measurement-wavelength memory
QUES_MeasData()Queries the measurement data
QUES_Rel()Queries the reference value
SQ functions
SQ_Mode()Sets or queries the mode
SQ_Wave()Sets (arbitrary setting) or queries the measure-
ment wavelength
SQ_Ave()Sets or queries the number of averaging steps
SQ_Range()Sets or queries the range
SQ_Aps()Sets or queries the APS (auto power save) func-
tion
1
2
3
4
6
7
8
An application for controlling the 3664 shall be created by the
user.
9
10
42
5.2 Installing the Driver
5.2Installing the Driver
Perform this procedure before connecting the main unit to the PC
with a USB cable. If already connected, disconnect the USB cable
while you perform the procedure.
Install
1. Log into the PC using the "administrator" account or other
account with administrative privileges.
2. Exit all applications running on the PC before installing the
hardware drivers.
3. Execute the following software from the Hioki 3664 Driver
Software (CD):
X:\English\driverSetup.msi
(where X: indicates the CD-ROM drive)
After executing driverSetup.msi, follow the instructions on the
screen to continue the installation.
A warning message will be displayed since the software has
not been certified under the Windows
logo program. Con-
tinue the installation.
4. Once the installation is complete, the 3664 will be automati-
cally detected when the main unit is connected to the PC
with a USB cable.
If the "Found New Hardware" wizard dialog box is displayed, select [No, not this time] under "Can Windows con-
nect to Windows Update to search for software?" and
select [Install the software automatically].When a 3664 unit with
a different serial No. is connected, the instrument may
indicate that a new device has been detected.In such case,
install the device driver by following the instructions
appearing on the screen.
A warning message will be displayed since the software has
not been certified under the Windows
logo program. Con-
tinue the installation.
Uninstall
Select [Control Panel]-[Add and Remove Programs] and delete HIOKI
USB Single Driver
.
6.1 Measurement Specifications
Specifications6
43
1
6.1Measurement Specifications
Optical-power
measurement
Range selection Automatic and manual
Accuracy± 0.7% (± 5% in combination with an optional optical sensor)
Maximum-value
hold
Minimum-value
hold
Average-value
display
Wavelength
sensitivity
correction
Scaling[Display value] = [Measured value] x [internal setting]
Relative
measurement
Unit: W/dBm
Displays the maximum value among the current measurements
Displays the minimum value among the current measurements
Displays the moving average for the set number of averaging
steps
The number of averaging steps is selectable between 2 and 100.
The setting is editable every 1 nm. The sensitivity of the sensor is
corrected automatically. Up to 10 wavelengths can be preset
(each sensor has unchangeable presets).
The internal setting is set for each wavelength and is selectable
between 0.1 and 10. The resolution is the same as that for opticalpower measurement.
Displays a relative value to a measured or set value
[Display value] = [Measured value] - [Reference value]
The reference value is set based on the measured value or
through key entry by the user. The selectable value ranges from
0.001 nW to 1.9999 W (-90.00 dBm to 33.00 dBm).
2
3
4
5
6
7
8
9
10
Display update
rate
Approx. 330 ms
44
6.1 Measurement Specifications
Analog outputOutput level:Depends on the optional optical sensor.
9742, 9742-10: Approx. 1 V at the sensor calibration point
9743, 9743-10: Approx. 0.7 V at the sensor calibration point *
Output resistance: 50 ± 5%
Connector: 3.5 Mini-jack
* The calibration point of the sensor is determined based on the
calibration wavelength and the calibration power specified in the
specification of the optional optical sensor.
Display monitor Monitor: LCD
Communications function
Functions• Auto power save
Measurement display: 4-1/2 digits in the unit of nW, W, mW,
dBm, or dB
Wavelength display: 4 digits in the unit of nm
Communications data:
Interface: USB Ver. 1.1 or later
Automatically turns OFF the power if there is no key operation or
USB communication for approximately 10 minutes. To turn OFF
the function, hold down the POWER key for 5 sec. when turning
ON the power. The auto-power-save function is available only
when the 3664 is running on the battery.
• Setting backup
Saves the settings when the power is turned OFF, and starts up
the 3664 with the saved settings when the power is turned ON
again.
• Battery check
Displays the battery indicator when the supply voltage falls
below 4.0 V
When the battery indicator is displayed, accuracy cannot be
guaranteed.
Accuracy guarantee for temperature and
humidity
Guaranteed
accuracy period
Product warranty period
• 1.5 V DC × 4 (LR6 alkaline batteries × 4)
• AC Adapter 9 V DC, 0.2 A
Model 9445-02 AC ADAPTER:
Rated supply voltage: 100 V to 240 V AC (Voltage fluctuations of
±10% from the rated supply voltage are taken into account.),
50 Hz/60 Hz
1.6 VA
Approx. 9 hours
(Ambient temperature 23°C(74°F) reference value)
Model 9742 OPTICAL SENSOR
Model 9742-10 OPTICAL SENSOR
Model 9743 OPTICAL SENSOR
Model 9743-10 OPTICAL SENSOR
Model 9445-02 AC ADAPTER
Model L9094 OUTPUT CORD
7.1 Cleaning
Maintenance and Service7
47
1
Never modify the instrument. Only Hioki service engineers should disassemble or repair the instrument. Failure to observe these precautions
may result in fire, electric shock, or injury.
7.1Cleaning
7.1.1Cleaning of the 3664
To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
2
3
4
5
6
8
9
10
48
7.2 Service
7.1.2Cleaning of the Optical Sensors
• Avoid touching the detector window with your bare hands. The
detector window must be clean for the sensor to meet the speci
fied performance parameters.
• With the exception of ethyl alcohol, avoid using organic solvents
to clean the detector window. These solvents can damage the
detector window.
• Avoid scratching the detector window with sharp or pointed
objects (e.g., the tips of tweezers) or against hard surfaces.
Damage to the detector window may prevent the sensor from
meeting specified performance parameters.
Take the cleaning method of the 9742 OPTICAL SENSOR as an
example.
1.Open the sensor cover.
(When using the 9742-10, 9743 and 9743-10 OPTICAL
SENSOR, take care to avoid misplacing the sensor cover.
2.Wipe the detector window of the sensor using lens cleaning
paper or other lint-free material.
If lint remains on the detector window, blow off with an optical lens airbrush.
If the detector window is soiled, fluff the tip of a cotton swab.
Moisten the cotton tip in ethyl alcohol and wipe the surface.
7.2Service
-
• If the instrument seems to be malfunctioning, confirm that the
batteries are not discharged, and that the USB cable is not open
circuited before contacting your dealer or Hioki representative.
• When sending the instrument for repair, remove the batteries and
pack carefully to prevent damage in transit. Include cushioning
material so the instrument cannot move within the package. Be
sure to include details of the problem. Hioki cannot be responsi
ble for damage that occurs during shipment.
-
7.3Error Display
49
7.3 Error Display
In the event of an error, one of the following messages is displayed.
Err0Reference-value save error
An improper value has been saved as the
reference value. If a reference value is
saved while the measured value is "Lo" or
"Hi" in the optical-power measurement
mode, "REL" and "SAVE" remain on the
screen and "Err0" is displayed. The refer
ence value is then updated to Hi or Lo, and
the relative measurement display shows "
- -
".
Err1Wavelength error
There is a problem with the measurementwavelength memory. Reset the measure
ment-wavelength memory.
See Section 7.4.3 Measurement-Wavelength
Memory Reset (page 52)
If the measurement-wavelength-memory
reset does not solve the problem, conduct a
system reset.
See Section 7.4.1 System Reset (page 51)
If the system reset does not solve the problem, the 3664 may be malfunctioning.
Contact your dealer or Hioki representative.
Err1 + no SEnSSensor data not compatible
1
2
3
-
4
- -
5
-
6
8
9
The 3664's internal sensor data is not compatible with the currently connected sensor. Please download the 3664 Sensor
Data Setup Software from our website
(http://www.hioki.co.jp) and perform the
setup of your 3664.
10
50
7.3 Error Display
Err2Setting-data error
There is a problem with the setting data.
Conduct a system reset.
See Section 7.4.1 System Reset (page 51)
If the system reset does not solve the problem, the 3664 may be malfunctioning.
Contact your dealer or Hioki representative.
Err3Memory error
There is a problem with writing into the
memory. Conduct a system reset.
See Section 7.4.1 System Reset (page 51)
If the system reset does not solve the problem, the 3664 may be malfunctioning.
Contact your dealer or Hioki representative.
Err4System error
The 3664 may be malfunctioning. Contact
your dealer or Hioki representative.
"1"Display over range
The value displayed after wavelength sensitivity correction or scaling exceeds the
upper limit that can be displayed by the
3664 (1.9999 W).
"no SEnS"Sensor disconnected
A sensor is not connected. Connect a sensor. The 3664 cannot be operated without a
sensor connected.
The number on the upper line is the sensor
data compatibility number. Please refer to
7.6 "Confirming Sensor Data Compatibility"
(page 53.) regarding the sensor data compatibility number.
7.4Reset
Blinking
Blinking
51
7.4 Reset
There are three types of reset operations for the 3664.
• System reset
Returns all settings to their defaults
• Setting-data reset
Returns all settings except those for the wavelength data to their
defaults
• Measurement-wavelength memory reset
Returns the measurement-wavelength data stored in the measurement-wavelength memory to its default state.
7.4.1System Reset
1.While holding down the SET key and key, turn ON the
power.
2."CLr" and "ALL" start blinking, and all settings are returned
to their defaults.
1
2
3
4
5
6
8
7.4.2Setting Data Reset
1.While holding down the SET key and SAVE key, turn ON the
power.
2."CLr" and "SEt" start blinking, and all setting data except
those for the wavelength is returned to its defaults.
9
10
52
Blinking
7.4 Reset
7.4.3Measurement-Wavelength Memory Reset
1.With an optional optical sensor connected to the 3664,
while holding down the
power.
2."CLr" and "λ" start blinking, and the measurement-wave-
length data is returned to the default wavelength setting of
the connected sensor.
The measurement-wavelength memory reset cannot be performed if a
sensor is not connected to the 3664.
SET key and λ key, turn ON the
7.5 Confirming Software Version
software version
sensor data
compatibility number
7.5Confirming Software Version
53
The software version is displayed on screen immediately after turning on the power.
• The software version is shown beneath the "3664". This example
shows Version 1.02.
7.6Confirming Sensor Data Compatibility
If the 3664 is turned on with no sensor connected, a no sensor
message is displayed along with the sensor data compatibility
number.
• Beneath the software version number, the sensor data compatibility number is shown as the 4th digit of the sensor model number. "2" is displayed for models 9742 and 9742-10, while "3" is
displayed for models 9743 and 9743-10.
The diagram shows an example where the unit is compatible
with model 974
2 (or 9742-10) and model 9743 (or 9743-10).
1
2
3
4
5
6
8
The sensor data compatibility number is displayed with software
Version 1.02 or higher. Versions 1.01 or lower will only display
"no SEnS".
9
10
54
1. Make sure the power to the 3664 is
OFF.
2. Disconnect all cables.
3. Turn over the 3664 and loosen the
screw on the battery cover.
4. Remove the battery cover.
5. Insert the four AA alkaline batteries
(LR6) into the battery case, noting the
correct polarity. (When replacing the
batteries, replace all of them.)
6. Place the battery cover back on, and
tighten the screw.
Tools and materials:
Four AA alkaline batteries (LR6) and a Phillips screwdriver
7.7 Replacing the Batteries
7.7Replacing the Batteries
• To avoid electric shock, turn off the power switch and disconnect the USB cable before replacing the batteries.
• After replacing the batteries, replace the cover and screws
before using the instrument.
• Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during
installation. Otherwise, poor performance or damage from
battery leakage could result.
• To avoid the possibility of explosion, do not short circuit,
disassemble or incinerate batteries.
• Handle and dispose of batteries in accordance with local
regulations.
•The “ ” indicator appears when battery voltage becomes low.
Replace the batteries as soon as possible.
• Use the specified battery (AA alkaline battery) only. If nonspecified battery such as a manganese battery is used, the
service hours may be shorter than when alkaline batteries are
used.
Appendix
P [dBm] = 10log
10
1 [mW]
E [mW]
55
Explanations
Optical-power measurement
Assuming that the energy of the input light measures E [mW], the
measured value P is expressed as follows.
P: Measured value [dBm]
E: Energy of input light [mW]
1
2
4
5
6
7
8
9
10
11
APPX
APP
56
Standard Optical Sensor for Model 3664
Standard Optical Sensor for Model 3664
Blue-violet Optical Sensor for Model 3664
Blue-violet Optical Sensor for Model 3664
Options
Model 9742 OPTICAL SENSOR
Model 9742-10 OPTICAL SENSOR
Model 9743 OPTICAL SENSOR
Model 9743-10 OPTICAL SENSOR
Model 9246 CARRYING CASE
Carrying case holding the 3664, the complete accessory set, and
an optional optical sensor
Model 9445-02 AC ADAPTER
Model L9094 OUTPUT CORD
57
• For 9742-10 and 9743-10 OPTICAL SENSOR, because of the
sensor wire is thin and breakable, do not put into the carrying
case.
• Put carefully the light-receiving part of the 9742 and 9743
OPTICAL SENSOR so that no stress is applied.
1
2
4
5
6
7
8
9
10
11
APPX
APP
58
59
1
2
4
5
6
7
8
9
10
11
APPX
APP
60
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.