Hioki 3664 Instruction Manual

Page 1
INSTRUCTION MANUAL
3664
OPTICAL POWER METER
www. .com
information@itm.com1.800.561.8187
Page 2
X
Contents
Introduction .................................................................. 1
Inspection ..................................................................... 1
Safety Information ........................................................2
Operating Precautions .................................................4
1Overview 7
i
Contents
1 2
1.1 Product Overview ...............................................7
1.2 Features ............................................................. 8
1.3 Names and Functions of Parts ........................... 9
2 Measurement Preparations 13
2.1 Attaching the Strap ...........................................13
2.2 Supplying Power ..............................................14
2.2.1 Using the AC Adapter ........................................14
2.2.2 Using the Batteries ............................................15
2.3 Connecting the Optical Sensor .........................16
2.4 Connecting the Analog Output Cord ................17
2.5 Turning the Power On and Off ......................... 18
3 Measurement 19
3.1 Setting the Measurement Mode ....................... 20
3.2 Setting the Range .............................................21
3.2.1 Auto Range .......................................................21
3.2.2 Manual Range ...................................................22
3.3 Setting the Wavelength .................................... 24
3.3.1 Setting an Arbitrary Measurement Wa v elength .24
3.3.2 Retrieving the Wavelength Setting from t he
Measurement-Wavelength Memory ..................26
3.4 Starting Measurement ......................................26
3.5 Setting Example ............................................... 28
3.5.1 Optical-Power Measurement .............................28
3.5.2 Relative Measurement (Measured -v alue basis) 29
3.5.3 Relative Measurement (Input-value basis) ........31
3 4 5 6 7 8 9 10 11
APP
IDX
www. .com
information@itm.com1.800.561.8187
Page 3
ii
Contents
4 Functions 33
4.1 Scalling Function .............................................. 33
4.1.1 Turning ON the Scaling Function ......................33
4.1.2 Turning OFF the Scaling Function .....................36
4.2 Maximum-Value Hold / Minimum-Value Hold
/ Averaging ....................................................... 37
5 Contr ol by Comp uter 39
5.1 Overview of Control ..........................................40
5.2 Installing the Driver ...........................................41
6 Specifications 51
6.1 Measurement Specifications ............................ 51
6.2 General Specifications .....................................53
7 Maintenance and Service 55
7.1 Cleaning ........................................................... 55
7.1.1 Cleaning of the 3664 .........................................55
7.1.2 Cleaning of the Optical Sensors ........................56
7.2 Service .............................................................56
7.3 Error Display ..................................................... 57
7.4 Reset ................................................................ 59
7.4.1 System Reset ....................................................59
7.4.2 Setting Data Reset ............................................59
7.4.3 Measurement-Wavelength Memory Reset ........60
7.5 Replacing the Batteries .................................... 61
Appendix 63
Explanations .............................................................. 63
Options .......................................................................64
www. .com
information@itm.com1.800.561.8187
Page 4

Introduction

Thank you for purchasing the HIOKI “Model 3664 OPTICAL POWER METER.” To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference.
1
1
Registered Trademarks:
• Windows is a registered trademark of Microsoft Corporation in the United States and/or other countries.
• Adobe, the Adobe logo, Acrobat, the Acrobat logo are trade­marks of Adobe Systems Incorporated.

Inspection

• When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evi­dent, or if it fails to operate according t o the specifications, con­tact your dealer or Hioki representative.
• Use the original packing materials when transporting the instru­ment, if possible.
Verifying Package Contents
Model 9445-02 AC ADAPTER or Model 9445-03 AC ADAPTER (1)
Model 3664 OPTICAL POWER METER (1)
2 3 4 5 6 7 8
Model 9094 OUTPUT CORD (1)
9
Driver CD-R (1)
Options
Strap (1)
USB Cable (1)
LR6 alkaline batteries (4)
Model 9246 CARRYING CASE............................................... .. ...1
Model 9742 OPTICAL SENSOR..................................................1
Model 9742-10 OPTICAL SENSOR.............................................1
www. .com
Instruction Manual (1)
information@itm.com1.800.561.8187
Page 5
2

Safety Information

This instrument is designed to comply with IEC 61010 Safety Standards, and has been tho roughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. Be certain that you understand the instructions and precautions in the manual before use. We disclaim an y responsibility for accidents or injuries not resulting direct ly from instrument defects.
Safety Symbols
This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions.
In the manual, the symbol indicates particularly important information that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the
manual (marked with the symbol) before using the relevant function.
The following symbols in this manual indicate the relative impor­tance of cautions and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a signifi­cant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possi­bility of injury to the user or damage to the instru ­ment.
Indicates advisory items related to performance or correct operation of the instrument.
Other Symbols
Indicates the prohibited action.
Indicates the location of reference information.
www. .com
information@itm.com1.800.561.8187
Page 6
Overvoltage Categories (CAT )
To ensure safe operation of measurement instruments, IEC 60664 establishes safety standards for various electrical environments, categorized as CAT I to CAT IV, and called overvoltage categories. These are defined as follows.
3
1
CAT I Secondary electrical circuits connected to an AC
electrical outlet through a transformer or similar device.
CAT II Primary electrical circuits in equipment connected to
an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
CAT III Primary electrical circuits of heavy equipment (fixed
installations) connected directly to the distribution panel, and feeders from the distribution panel to out­lets.
CAT IV The circuit from the service drop to the service
entrance, and to the power meter and primary over­current protection device (distribution panel).
Higher-numbered categories correspond to electrical environments with greater momentary energy. So a measurement device designed for CAT III environments can endure greater momentary energy than a device designed for CAT II. Using a measurement instrument in an environment designated with a higher-numbered category than that for which the instrument is rated could result in a severe accident, and must be carefully avoided.
2 3 4 5 6 7 8 9 10 11
www. .com
information@itm.com1.800.561.8187
Page 7
4

Operating Precautions

Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Observe the following to avoid damage to the instrument.
• Do not allow the instrument to get wet, and do not take measure­ments with wet hands. This may cause an electric shock.
Avoid the following
Direct Sunlight
High Temp erature High Humidity
• Do not use the instrument where it may be exposed to corrosive or combustible gases. The instrument may be damaged or cause an explosion.
• To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially care ful to avoid physical shock from dropping.
• Do not store or use the instrument wh ere it could b e exposed to direct sunlight, high temperature or humidity, or condensation. Under such conditions, the instrument may be damaged and insulation may deteriorate so that it n o longer meets specifica­tions.
• This instrument is designed for use indoors. It can be operated at temperatures between 0 and 40°C without degrading safety.
• After use, always turn OFF the power.
• The “ ” indicator appears when batter y voltage becomes low.
Replace the batteries as soon as possible.
• Use the specified battery (AA alkaline battery) only. If a non-
specified battery such as a manganese battery is used, the service hours may be shortened.
• When connecting the instrument to a PC with a USB cable,
use a cable of 2 m or shorter to prevent the influence of noise.
www. .com
information@itm.com1.800.561.8187
Page 8
Cords and Cables
5
• Avoid stepping on or pinching cables, which could dam age the cable insulation.
• To avoid breaking the cables, do not bend or pull them.
• To avoid damaging the output cable, grasp the connector, not the cable, when unplugging the cable.
Optical Sensor
• Use only the specified optical sensor. Using a non-specified sen­sor may result in incorrect measurements due to poor connec­tion or other reasons.
• To prevent damage to the instrument and sensor, never connect or disconnect a sensor while the power is on, or while the sensor is connected around a conductor.
• To avoid damaging the optical sensor and to ensure accurate measurements, avoid dropping or applying any p hysical shock to the optical sensor.
• Avoid touching the detector window with your bare hands. The detector window must be clean for the sensor to meet the speci­fied performance parameters.
• Avoid scratching the detector window with sharp or pointed objects (e.g., the tips of tweezers) or against hard surfaces. Damage to the detector window may prevent the sensor from meeting specified performance parameters.
• Make sure no stress is applied to the repeater cable of the 9742­10 between the sensor main unit and its detector element.
1 2 3 4 5 6 7 8 9
To prevent dust accumulation, stains, and damage, cover the detector window with the sensor cover when the sensor is not in use.
www. .com
information@itm.com1.800.561.8187
Page 9
6
Care and Handling of CD-R
• Always hold the disc by the edges, so as not to make finger­prints on the disc or scratch the printing.
• Never touch the recorded side of the disc. Do not place the disc directly on anything hard.
• Do not wet the disc with volatile alcohol or water, as there is a possibility of the label printing disappearing.
• To write on the disc label surface, use a spirit-based felt pen. Do not use a ball-point pen or hard-tippe d pen, because there is a danger of scratching the surface and corrupting the data. Do not use adhesive labels.
• Do not expose the disc directly to the sun's rays, or keep it in conditions of high temperature or humidity, as there is a danger of warping, with consequent loss of data.
• T o remove dirt, dust, or fi ngerprints from the disc, wipe with a dry cloth, or use a CD cleaner. Always wipe radially from the inside to the outside, and do no wipe with circular movements. Never use abrasives or solvent cleaners.
• Hioki shall not be held liable for any problems with a com puter system that arises from the use of this CD-R, or for any problem related to the purchase of a Hioki product.
Preliminary Checks
• Before using the instrument, make sure that the insulation on the optical sensor cable and USB cable is undamaged and that no bare conductors are improperly exposed. Using the instrument in such conditions could cause an electric shock, so contact your dealer or Hioki representative for repair.
• Before using the instrument the first time, verify that it ope rates normally to ensure that the no damage occur red during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
www. .com
information@itm.com1.800.561.8187
Page 10

1.1 Product Overview

7

Overview

1.1 Product Overview
The 3664 OPTICAL POWER METER is an instrumen t for measur­ing the optical power of a spatial light. It is suitable for measure­ment of the optical power of a laser light source and an LED light source. The 3664 is a hand-held meter equipped with various mea­surement functions and interfaces that meet diversified measure­ment needs. The optical sensors (optional) for the power meter are interchangeable, as the sensitivity of each sensor is adjusted inde­pendently.
1
1 2 3 4 5 6 7 8
www. .com
9 10 11
information@itm.com1.800.561.8187
Page 11
8

1.2 Features

1.2 Features
Scaling function
The reading of the 3664 OPTICAL POWER METER can be con­verted to the reading of another optical power meter by a simple operation.
Various measurement functions
In addition to the optical power measurement (absolute-value mea­surement), the 3664 power meter has relative-measurement, max­imum-value-hold, minimum-value-hold, and averaging functions. With all these functions, the display unit can be switched between W and dBm (dB).
Automatic correction of sensitivity through input of the measurement wavelength
When the measurement wavelength is entered, the wavelength sensitivity of the sensor connected to the 3664 is automatically cor­rected for measurement display.
Interface with a computer
The USB interface is provided as standard equipment. The 3664 can be controlled on the PC, its settings can be edited on the PC , or its measurement data can be uploaded to the PC.
Analog output terminal
The 3664 has an analog output terminal. It outputs in the voltage form the output signal of the optical sensor. Use this terminal, for example, when performing monitoring using a DMM or reco rding data for extended periods.
Auto-power-save funct ion
When the 3664 is running on the battery, if there is no key opera­tion or USB communication for approximately 10 minutes, the power to the 3664 will be automatically turned off to conserve bat­tery power.
www. .com
information@itm.com1.800.561.8187
Page 12

1.3 Names and Functions of Parts

1.3 Names and Functions of Parts
9
Front Panel
Operating Keys
Strap installation Slot
Display
1 2 3 4 5 6 7 8
Display
Main display
Relative-measurement indicator
Range
Sub display
Relative-measurement mode
www. .com
Maximum, Minimum, Ave ra ge
Auto power save mark
Battery mark
Measurement wavelength
information@itm.com1.800.561.8187
9 10 11
Page 13
10
1.3 Names and Functions of Parts
Operating Keys
POWER
λ
REL
RANGE
SET
SAVE
W/dBm
MAX/MIN/AVE
Turns ON/OFF power to the 3664. Power ON: Hold down the key for approximately 2 seconds to
Power OFF: Hold down the key for approximate ly 1 second to turn
Sets the measurement wavelength to the preset wavelength. Switches between optical-power measurement mode, relative-
measurement mode (measured-value basis) and relative-mea­surement mode (input-value basis).
Switches between auto range and manual range. Moves to the setting edit screen. " Moves the digit to edit the setting (the digit blinks) during setting
edit. Confirms the setting during setting edit and returns the meter to
measurement mode. Switches the unit of the reading between "W" and "dBm." In rela-
tive-measurement mode, the unit is switched between "W" and "dB." Decreases the number during setting edit.
Switches the value to be displayed between the measurement, maximum, minimum, and average value. Increases the number during setting edit.
turn the power ON (auto-power-save function ON). Hold down the key for approximately 5 seconds to turn the power ON and the auto-power-save function is turned OFF.
the power OFF.
" appears on the screen.
SEt
www. .com
information@itm.com1.800.561.8187
Page 14
Upper Panel
11
1.3 Names and Functions of Parts
Analog output terminal
Connect the 9094 OUTPUT CORD in­cluded among the accessories to this terminal. Output signals from the opti­cal sensor are converted to a voltage and output from this terminal.
Optical-sensor terminal
Connect an optional optical sensor to this terminal.
Bottom Panel
AC-adapter terminal
Connect the 9445-02 or 9445-03 AC ADAPTER included among the accesso­ries to this terminal. When the AC adapter is used, power is supplied to the 3664 from a commercial power supply.
1 2 3 4 5 6 7 8
www. .com
USB terminal
Connect a USB cable to this terminal.
information@itm.com1.800.561.8187
9 10 11
Page 15

2.1 Attaching the Strap

13

Measurement Preparations

2.1 Attaching the Strap
Use the strap to carry the 3664 power meter or hang it on a hook.
Be sure to attach the strap to the 3664 power meter in place (2 loca­tions) properly . If the strap is not properly attached, the meter m ay drop and be broken.
Model 3664
Insert each end of the Strap through
1.
an installation slot on the instrument.
Feed each end of the Strap through its
2.
clasp.
Feed each end through its stopper.
3.
2
1 2 3 4 5 6 7 8
Confirm that the strap is not slackened or twisted.
www. .com
9 10
information@itm.com1.800.561.8187
Page 16
14

2.2 Supplying Power

2.2 Supplying Power

2.2.1 Using the AC Adapter

• Use only the specified Model 9445-02 ADAPTER (SA10­9010N,SINO-AMERICAN) or Model 9445-03 AC ADAPTER (CEE, SA10-9010G, SINO-AMERICAN) AC adapter input voltage range is 100 to 240 VAC (with ±10% stability) at 50/ 60 Hz. To avoid electrical hazards and damage to the instru­ment, do not apply voltage outside of this range.
• Turn the instrument off before connecting the AC adap ter to the instrument and to AC power.
To avoid damaging the cord, grasp the plug, not the cord, when unplug­ging it.
AC Adapter Connection Term i na l
• When the AC adapter is connected and the battery is installed,
the AC adapter is used. In case of power outage, the 3664 is automatically switched to the battery drive.
• Make sure the power is turned off before connecting or discon-
necting the AC adapter.
Connect the output plug of the AC
1.
adapter to the AC-adapter terminal.
Connect the AC adapter to an AC
2.
outlet.
www. .com
information@itm.com1.800.561.8187
Page 17

2.2.2 U sing the Batteri es

15
2.2 Supplying Power
Screw
Do not mix old and new batteries, or different types of batter­ies. Also, be careful to observe ba ttery polarity during instal­lation. Otherwise, poor performance or da mage from battery leakage could result.
• The “ ” indicator appears when battery voltage becomes low. Replace the batteries as soon as possible.
• Do not use other than the specified type of batteries (LR6 alka­line batteries). Using manganese batteries will result in a shorter operating time than when using alkaline batteries.
• The 3664 power meter runs on battery or AC power. When the AC adapter is connected and the battery is installed, the AC adapter is used.
Tools and materials: Four AA alkaline batteries (LR6) and a Phillips screwdriver
Confirm that the power to the 3664 is
1.
turned OFF.
Disconnect all cables.
2.
Turn over the 3664 and loosen the
3.
screw on the battery cover.
1 2 3 4 5 6 7 8
Remove th e ba t te r y cover.
4.
Insert the four AA alkaline batteries
5.
(LR6) into the battery case, noting the correct polarity. (When replacing the batteries, replace all of them.)
Place the battery cover back on, and
6.
tighten the screw.
www. .com
9 10
information@itm.com1.800.561.8187
Page 18
16

2.3 Connecting the Optical Sensor

2.3 Connecting the Optical Sensor
• Use only the specified optical sensor. Using a non-specified sen­sor may result in incorrect measurements due to poor connec­tion or other reasons.
• To prevent damage to the instrument and sensor, never connect or disconnect a sensor while the power is on, or while the sensor is connected around a conductor.
• To avoid damaging the sensor cable, grasp the plug, not the cable, when unplugging it.
• To avoid ensuring accurate measurements, avoid dropping or applying any physical shock to the optical sensor.
Insert an optical sensor, with the surface marked with an arrow fac­ing upward, into the optical-sensor terminal of the 3664.
Optical Sensor
Grasp the connector.
Optical Sensor Connection Te rmi na l
Always grasp the connector with the arrow to connect or discon­nect the optical sensor. (You must grasp the connector to discon­nect the sensor.)
www. .com
information@itm.com1.800.561.8187
Page 19

2.4 Connecting the Analog Output Cord

2.4 Connecting the An alog Output Cord
17
To avoid damage to the instrument, do not short-circuit the o utput termi­nal and do not input voltage to the output terminal.
By connecting the analog output cord to the analog output terminal, the optical power detected by the optical sensor can be converted to a voltage in order to monitor the signal.
Connect the analog output cord to the analog output terminal.
Analog Output Cord
Signals output from the analog output terminal are detected by an optical sensor, with its wavelength sensitivity uncorrected. This signal, therefore, is not proportional to the measurement displayed on the screen.
1 2 3 4 5 6 7 8
www. .com
9 10
information@itm.com1.800.561.8187
Page 20
18

2.5 Turning the Power On and Off

2.5 Turn in g the Power On and Off
Power On
Hold down the POWER key for approximately 2 seconds. The screen lights up and the power is turned ON.
If the battery indicator appears (" " at the top right of the screen) after the power has been turned ON, the battery is run­ning down. Replace the battery (when the 3664 is running on the battery).
See Section 7.5 "Replacing the Batteries" (61 page)
Power OFF
Hold down the POWER key for approximately 1 second. The screen goes out and the power is turned OFF.
Auto-power-save funct ion
• This function automatically turns OFF the power to the 3664 if there is no key operation or USB communication for approxi­mately 10 minutes, when it is running on the battery. (The indica­tion "
• To turn OFF the auto-power-save function, hold down the
POWER key for 5 seconds when turning ON the power. "
goes out and the auto-power-save function is turned OFF.
• When the 3664 is running on AC power, the auto-power-save function is automatically turned OFF and the function is not avail­able.
" lights up at the top right of the screen.)
APS
APS
"
• While power is supplied from the battery, when the power is turned ON, the auto-power-save function is turned ON.
• When the USB is used, use the AC adapter for the power sup­ply.
• After use, always turn OFF the power.
AC Adapter Connection Terminal
www. .com
information@itm.com1.800.561.8187
Page 21
19

Measurement

• Optical-power measurement and relative measurement of spe­cial light are performed using the 3664 power meter.
• The measurement wavelength may be set to an arbitrary wave­length.
• The measurement wavelength may also be set to a previously set wavelength.
Connect an optional optical sensor to the 3664. To operate the 3664, an optical sensor must be connected to it.
Flow of measurement
Setting of the measurement
Setting of the range
Setting of the measurement wavelength
(2 0 page)
(21 page)
(24 page)
3
1 2 3 4 5 6 7 8
Input of light to the sensor
(2 6 page)
9 10
www. .com
information@itm.com1.800.561.8187
Page 22
20

3.1 Setting the Measurement Mode

3.1 Setting the Measurement Mode
Each time the changes as shown below.
Optical-power measurement
Relative measurement (measured-value basis)
Relative measurement (input-value basis)
key is pressed, the measurement mode
REL
The 3664 performs optical- po w ­er measurement.
The 3664 sets th e optical p ower of the incident light to the optical sensor as the reference value, and measures th e difference be­tween the set reference va lue and the current incident o ptical power.
The 3664 measures the differ­ence between the refe re nce v al­ue arbitrarily set by key entry and the current incident optical power.
www. .com
information@itm.com1.800.561.8187
Page 23
3.2 Setting the Range
21

3.2 Setting the Range

Each time the range alternate.

3.2.1 Auto Range

The appropriate range for the incident optical power is automati­cally selected. The range display area will appear as shown below.
RANGE
Auto Range
key is pressed, auto range and manual
Manual Range
1 2 3 4 5 6 7 8
When the RANGE key is pressed, "MANUAL" appears in the range display area and the range is fixed to that displayed.
www. .com
information@itm.com1.800.561.8187
9 10
Page 24
22
3.2 Setting the Range

3.2.2 Manual Range

The range is set manually. " area.
1.
Press the
2.
Press the the manual range. The range display area will blink.
key to display the setting edit screen.
SET
RANGE
MANUAL
key to display the setting edit screen for
" appears in the range display
Blinking
3.
Use the ▲ and ▼ keys to change the range s et t ing.
: Increases the sensitivit y range by one increment
: Decreases the sensiti vity range by one increment
The range No. displayed on the screen is the range No. of the internal circuit. The range is divided into seven levels, from 1 (lowest sensitivity) to 7 (highest sensitivity).
www. .com
information@itm.com1.800.561.8187
Page 25
3.2 Setting the Range
4.
Press the the 3664 will return to the meas urement scree n. The range is now fixed.
key to save the setting. "
SAVE
" is displayed;
SAVE
23
1 2 3 4
Fixed range
When the range is fixed through manual range setting, the mea­surement display shows " below the measurement range, and it shows "Hi" if the power exceeds the range.
"Lo" Display
" if the incident optical power falls
Lo
"Hi" Display
5 6 7 8 9 10
www. .com
information@itm.com1.800.561.8187
Page 26
24

3.3 Setting the Wavelength

3.3 Setting the Wavelength
The following two methods can be used:
• Set an arbitrary measurement wavelength.
• Retrieve a wavelength from the measurement-wavelength mem­ory, and set it as the measurement wavelength.
• The selectable wavelength range depends on the specification of the optical sensor connected to the 3664.
• To ensure proper measurement, set the wavelength to the wavelength of the light used.

3.3.1 Setting an Arbitrary Measurement Wavelength

1.
Press the
key to display the setting edit screen.
SET
2.
Press the λ key to display the setting edit screen for the measurement wavelength.
Blinking
3.
Use the ▲ and ▼ keys to change the measurement wave­length.
: Increases the number
: Decreases the numbe r
www. .com
information@itm.com1.800.561.8187
Page 27
3.3 Setting the Wavelength
4.
Press the key to change the blinking digit.
Blinking
Repeat steps 3 and 4 above to set t he measurement w av e­length.
5.
Press the
"SAVE" appears; the 3664 will then return to the measurement screen. The measurement wavelength is now set.
key to complete setting.
SAVE
The measurement waveleng th set
25
1 2 3 4 5 6
• An arbitrarily set measurement wavelength is automatically saved in the measurement-wavelength memory.
• The measurement-wavelength memory holds up to 10 wave­lengths (including preset wavelengths for each optical sensor).
• Each optical sensor has preset wavelengths. These preset wavelengths cannot be edited.
• Through the setting operation for the arbitrary measurement wavelength, the setting data is automatically saved in the mea­surement-wavelength memory. When it is necessary to save a new measurement wavelength to the measurement-wave­length memory when the memory already contains 10 wave­lengths, the oldest data is deleted in order to make space for the new data.
• To reset the data in the measurement-wavelength memory, perform measurement-wavelength-memory reset. This will return the memory to its default state for each sensor.
See Section 7.4.3 "Measurement-Wavelength Memory Reset" (60
page)
7 8 9 10
www. .com
information@itm.com1.800.561.8187
Page 28
26

3.4 Starting Measurement

3.3.2 Retrieving the Wavelength Setting from the Mea­surement-Wavelength Memory
Each time the λ key is pressed, the measurement wavelengths stored in the memory are retrieved individually in order.
To reset the data in the measurement-wavelength memory, per­form measurement-wavelength-memory reset. This will return the memory to its default state for each sensor.
See Section 7.4.3 "Measurement-Wavelength Memory Reset" (60 page)
3.4 Starting Measurement
1.
When using the 9742 OPTICAL
Model 9742 OPTI­CAL SENSOR
Optical Sensor Cover
SENSOR, slide the sensor cover of the optical sensor until it clic ks to lock the cover. When using the 9742-10 OPTICAL SENSOR, remove the optical-sensor cover.
Model 9742-10 OP­TICAL SENSOR
Optical Sensor Cover
• Removed sensor cover (B e
• Mount the sensor cover as
careful to avoid misplacing.) shown in the figure to the
left.
When using the 9742-10 OPTI­CAL SENSOR, remove the cover of the optical sensor.
2.
Inject light.
3.
The 3664 displays the measured optical power.
www. .com
information@itm.com1.800.561.8187
Page 29
27
3.4 Starting Measurement
• To ensure accurate measurements, make sure the detector window is free of dust, stains, and any damage.
• Confirm that the light to be measured fits completely within the detector window.
• If the optical power is too intense for the unit area, such as when a condensed beam on the focus plane is measured, the optical sensor may be saturated and the 3664 may not be able to conduct accurate measurement. This problem can be solved by changing the distance between the light source and the detector window of the optical sensor.
• The operation of the light source may become unstable due to the reflection of light from the detector window of the optical sensor. This problem can be solved by angling the detector window slightly to prevent the reflection from ret urning to the light source.
• If background light, sunlight, illumination light, or any other light than the light to be measured is overly intense, the 3664 may not be able to conduct accurate measurement. Take appropri­ate light-shielding measures.
1 2 3 4 5 6 7 8 9 10
www. .com
information@itm.com1.800.561.8187
Page 30
28

3.5 Setting Example

3.5 Setting Example

3.5.1 Optical-Power Measurement

The 3664 performs optical-power measurement.
(1) Set the measurement mode.
Press the
(2) Set the measurement wavelength.
Set the measurement wavelength to an arbitrary wavelength or one retrieved from the memory. The sensitivity is corrected by the set measurement wavelength, and the 3664 displays the mea­sured value adjusted in accordance with the corrected sensitivity.
Set the measurement wavelength to the wavelength of the light used. If the wavelengths differ, the 3664 is not able to conduct accurate measurement.
key to select the optical-power measurement mode.
REL
Optical-power measurement
www. .com
information@itm.com1.800.561.8187
Page 31
3.5 Setting Example

3.5.2 R ela tive Mea surem ent (Meas ur ed-va lue basi s)

29
(1) Set the measurement mode.
(2) Set the measurement wavelength.
(3) Loading the reference value/relative measurement (when the optional
Model 9742 OPTI­CAL SENSOR
Model 9742-10 OP­TICAL SENSOR
The 3664 sets the optical power of the incident light to the optical sensor as the reference value, and measures the difference between the set reference value and the current incident optical power.
Press the value basis) mode.
Set the measurement wavelength to an arbitrary wavelength or one retrieved from the memory.
Set the measurement wavelength to the wavelength of the light used. If the wavelengths differ, the 3664 is not able to conduct accurate measurement.
9742 or 9742-10 OPTICAL SENSOR is used)
Optical Sensor Cover
• Removed sensor cove r (Be careful to avoid misplacing.)
• Mount the sensor cover as shown in the figure to the left.
key to select the relative-measurement (measured-
REL
1.
When using the 9742 OPTICAL SENSOR, slide the sensor cover of the optical sensor until it clicks to lock the cover. When usi ng the
Optical Sensor Cover
9742-10 OPTICAL SENSOR, remove the optical-sensor cover.
When using the 97 42-10 OPTICAL SENSOR, remove the cover of the optical sensor.
2.
Inject light.
3.
In the relative-measure ment (mea­sured-value basis) mode, hold down the mately 3 seconds.
key for approxi-
SAVE
1 2 3 4 5 6 7 8 9 10
www. .com
information@itm.com1.800.561.8187
Page 32
30
3.5 Setting Example
4.
"
" is displayed on the screen, and the relative-mea-
SAVE
surement result is then displayed. The result is save d as the optical-pow er measuremen t ref­erence valu e.
Mode
5.
"0" appears on the left side of t he sub-dis play are a, indicat -
Reference Value
ing the relative-measurement (measured-value basis) mode, and the saved reference value is displayed on the right side of the sub-display area. The main display area shows the relative-measurement value. In the relative-mea surem ent (me asur ed- value basis ) mo de, the value show n below is displayed. (Current measurement) - (measurement set as the refer­ence value)
• Only one reference value is saved, regardless of the wave­length setting. When a different measurement wavelength is used, set a new reference value.
• When the maximum-value-hold, minimum-value-hold, or a ver­aging function is ON, "(calculation result) - (reference value)" is displayed.
• When the scaling function is ON, "(value af ter scaling) - (refe r­ence value)" is displayed.
• When the reference value is input, if the main display area shows " The relative measurement is not calculated, and " played on the screen.
" or "Hi," Lo or Hi is saved as the reference value.
Lo
- - - -
" is dis-
www. .com
information@itm.com1.800.561.8187
Page 33
3.5 Setting Example

3.5.3 Relative Measurement (Input-value basis)

31
The 3664 measures the difference between the reference value arbitrarily set by key entry and the current incident optical power.
Entering the Reference Value Using the Keys/Relative Measurement
1.
Press the
2.
Press the screen.
3.
Use the ▲ and ▼ keys to change the position of the decimal point and the unit.
: Increases the number
: Decreases the number
key to display the setting ed it screen.
SET
key to display the reference-value input
REL
Blinking
Blinking
1 2 3 4 5 6 7 8 9
4.
Press the key to change the blinking digit.
www. .com
information@itm.com1.800.561.8187
Blinking
Page 34
32
3.5 Setting Example
5.
Use the ▲ and ▼ keys to change the blink ing number.
: Increases the number
: Decreases the numbe r
Repeat steps 4 and 5 ab ov e t o s et the r ef erence value.
6.
Press the "
" appears and then the 3664 returns to the measure -
SAVE
ment screen. The mea surem ent wa velen gth i s now se t an d the relative-measurem ent result is dis played.
key to comp l ete setting.
SAVE
Blinking
Mode
The number "1" appears on the left side of the sub-display area, indicating the relative-measurement (input-value basis) mode, and the saved reference value is displayed on the right side of the sub­display area. The main display area displays the relative-measurement value. In the relative-measurement (input-value basis) mode, the value shown below is displayed. (Current measurement) - (value entered using th e keys as the ref­erence value)
• Only one reference value is saved regardless of the wave­length setting.
• When a different measurement wavelength is used, set a new reference value.
• The values that can be entered using the keys as the reference value range from 0.001 nW to 1.9999 W (-90.00 dBm to 33.0 dBm).
• When the maximum-value-hold, minimum-value-hold, or a ver­aging functions is ON, "(calculation result) - (reference value)" is displayed.
• When the scaling function is ON, "(value af ter scaling) - (refe r­ence value)" is displayed.
Reference Value
www. .com
information@itm.com1.800.561.8187
Page 35

4.1 Scalling Function

33

Functions

• Using the scaling function, a measured value can be converted to an arbitrary value and the converted value displayed.
• The maximum-value-hold, minimum-value-hold, and averaging functions are available.
4.1 Scalling Function
• A measured value is converted by scaling to an arbitrary value and displayed. When the scaling function is turned ON, "(mea­sured value) x (internal coefficient)" is displayed.
• The internal scaling coefficient for each wavelength is saved in the memory.
• When the display value after scaling is set, the internal coe ffi­cient is automatically calculated and saved.

4.1.1 Turning ON the Scaling Function

Save the internal scaling coefficient, and turn ON the scaling func­tion.
1.
Input the light to be measu red to the optical sensor.
2.
Press the
key to display the setting ed it screen.
SET
4
1 2 3 4 5 6 7 8
www. .com
9 10
information@itm.com1.800.561.8187
Page 36
34
4.1 Scalling Function
3.
Press the λ key to display the setting edit screen for the measurement wavelength.
Blinking
Scaling function
The value displaye d first in the sub-displa y area is the dis­play value with the scaling OFF (internal coefficient of
1.000).
4.
Use the key to move the cursor to a digit in the s ub-dis­play area.
5.
Use the ▲ and ▼ keys to change the blink ing number.
: Increases the number
: Decreases the numbe r
Repeat steps 4 and 5 above to set the display value after scaling.
Measured value immediately before the setting edit screen appears
Blinking
Sub Display
Blinking
www. .com
information@itm.com1.800.561.8187
Page 37
4.1 Scalling Function
6.
Press the "
" is displayed, an d the 3664 then returns to the me a-
SAVE
surement screen. The scaling coefficient is automatically calculated and saved as an i nt ern al setting.
7.
The indicator "λ" starts blinking and t he scaled value is dis­played.
key to save the setting.
SAVE
35
1
Blinking
Scaled Value
• While the scaling function is ON, the indication "λ" blinks.
• The internal scaling coefficient ranges from 0.100 to 10.000. The scaling function is not available under the following condi-
tions, and " scaling (in the sub-display area).
• The measured value is displayed as " range.
• The measured value is very large or very small in the auto range.
" or "Lo" appears instead of the display value after
Hi
" or "Lo" in the manual
Hi
2 3 4 5 6 7 8 9
www. .com
information@itm.com1.800.561.8187
Page 38
36
4.1 Scalling Function

4.1.2 Turning OFF the Scaling Function

1.
Press the
2.
Press the λ key to display the setting edit screen for the measurement wavelength.
key to display the setting edit screen.
SET
Blinking
Scalling Function
3.
Press the before scaling. "
" is displayed, and the 3664 then re turns to the mea-
SAVE
surement screen. A scaling coefficient of 1. 000 is set as an internal s etting, and the s c aling functi on is turned OFF.
4.
The indication "λ" remains ON as a solid light, and the measured value that is n ot scaled is displayed.
To return to the measurement screen with the scaling ON: Do not press the
key twice.
SET
key without changing the display value
SAVE
key in step 4 above; instead, press the
SAVE
Display value before scaling
www. .com
information@itm.com1.800.561.8187
Page 39

4.2 Maximum-Value Hold / Minimum-Value Hold / Averaging

4.2 Maximum-Value Hold / Minimum-Value Hold / Averaging
37
Each time the changes as shown below.
No calculation
Maximum-value hold
Minimum-value hold
MAX/MIN/AVE
key is pressed, the calculation setting
The calculation functi on is disabled.
The maximum value among the current measurements is displayed.
The minimum value among the current measurements is displayed.
1 2 3 4 5 6 7 8
Average-value hold
The moving average for the set number of averaging steps is displayed.
To reset the display value, press the start calculation again.
www. .com
MAX/MIN/AVE
information@itm.com1.800.561.8187
9 10
key again and
Page 40
38
4.2 Maximum-Value Hold / Minimum-Value Hold / Averaging
Setting the Number of Averaging Steps
1.
Press the
2.
Press the for the number of averaging steps.
3.
Use the ▲ and ▼ keys to change the blink ing number.
: Increases the number
: Decreases the numbe r
A digit may be chos en by moving the cursor using the key.
key to display the setting edit screen.
SET
MAX/MIN/AVE
key to display the setting edit screen
Blinking
Blinking
4.
Press the
" is displayed, and the 3664 then re turns to the mea-
SAVE
"
key to save the setting.
SAVE
surement screen. The number of averaging steps is now set.
• After returning to the measurement screen, " blink until the number of averaging steps reaches the set num­ber.
• The number of averaging steps is selectable between 2 and
100.
• If the
MAX/MIN/AVE
pressed during measurement with the calculation function ON, the calculation is reset.
key, λ key,
RANGE
" continues to
AVE
key, or
SET
key is
www. .com
information@itm.com1.800.561.8187
Page 41
39

Control by Computer

By connecting the 3664 power meter to a PC, measurement s can be read and the 3664 can be set up on the PC.
When connecting the 3664 to a PC with a USB cable, use a cable of 2 m or shorter to prevent the influence of noise.
5
1 2 3 4 5
5
6 7 8
www. .com
9 10
information@itm.com1.800.561.8187
Page 42
40

5.1 Overview of Control

5.1 Overview of Control
• By connecting the 3664 power meter to a PC with a USB inter­face and calling API functions stored in the DLL from the applica­tion, the 3664 can be remotely controlled, and data can be exchanged between the 3664 and the PC.
• The table below shows a list of the API functions. For details, see the "API Instructions Manual (3664api.pdf)" on the supplied CD­R.
• To view the 3664api.pdf, Acrobat Reader must be installed on your PC. If it is not installed, first install the Acrobat Reader from the supplied CD-R.
API list
Name of function Description
RESET functions ResetDeviceUSB() Resets the communications function ResetAllData() Resets the setting dat a ResetAllWave() Resets the measurement w avelength SET functions SET_Rel0() Sets the reference valu e for rel ative measur ement
SET_Rel1() Sets the reference valu e for rel ative measur ement
QUES functions QUES_IDN() Queries the device data QUES_Sensor() Queries the optical-sensor model No. QUES_AllWave() Queries all measurement wavelengths stored in
QUES_MeasData() Queries the measurement data QUES_Rel() Queries the reference value SQ functions SQ_Mode() Sets or queries the mode SQ_Wave() Sets (arbitrary setting) or queries the measure-
SQ_Ave() Sets or que r ies the number of averaging steps SQ_Range() Sets or queries the range SQ_Aps() Sets or queries the APS (auto power save) func-
(measured-value basis)
(input-value basis)
the measurement-wavelength memory
ment wavelength
tion
An application for controlling the 3664 shall be created by the user.
www. .com
information@itm.com1.800.561.8187
Page 43
5.2 Installing th e Driver
41

5.2 Installing the Driver

To connect the 3664 to a PC using the USB interface, the driver must be installed.
System requirements
OS Windows 98SE, Windows Me, Windows 200 0,
Windows XP Hardware requirements, such as those involving the CPU, RAM, and display monitor, shall con-
form to the system requirements of the OS. HDD space 10 MB or more of free disk space Interface USB Ver. 1.1 or higher
(connectable to only one PC)
CD-R contents
HiUsbSgl.inf File for driver installation HiUsbSgl.sys Device driver for the USB of the 3664
1 2 3 4 5
5
6 7 8
www. .com
9 10
information@itm.com1.800.561.8187
Page 44
42
5.2 Installing the Driver
Do not plug in or unplug the USB cable while the instrument is operat­ing.
• The PC and 3664 should be connected on a one-on-one basis. Multiple 3664 units cannot be connected to the PC at a time.
• When a 3664 unit with a different serial No. is connected, the instrument may indicate that a new device has been detected. In such case, install the device driver by following the instruc­tions appearing on the screen.
Turn on the power of the instrument.
1.
Connect the PC on which you intend to install the driver and the
2.
3664 by using the USB cable supplied with the instrument. (PC side)
The 3664 will be recognized by the device name of "
OPTICAL POWER METER
After displaying " the "
Found New Hardware Wizard
The "
Found New Hardware Wizard
depending on the OS. Insert the CD-R into the PC according to instructions given in the dialog box, then install the driver. The following describes how to install the driver for each type of OS.
See Section "For Windows XP" (page 43)See Section "For Windows 2000" (page 45)See Section "For Windows Me" (page 47)See Section "For Windows 98SE" (page 48)
."
Found New Hardware
."
" requires different procedures
" on the screen, the PC starts
HIOKI 3664
www. .com
information@itm.com1.800.561.8187
Page 45
For Windows XP
Insert the CD-R supplied with the instrument into the CD-ROM
1.
drive. The "
5.2 Installing the Driver
Found New Hardware Wizard
" window appears.
43
1
2.
3.
Check "
Click the [
Install the software automatically
Continue Anyway
] button.
," then click the [
Next
] button.
2 3 4 5
5
6 7 8 9
Clicking the button initiates copying of the file. When this software is recognized, the screen will display a message stating that the software is not authorized by Microsoft. Ignore the message and continue operation.
www. .com
information@itm.com1.800.561.8187
Page 46
44
5.2 Installing the Driver
Click the [
4.
When the "
5.
CD-R from the CD-ROM drive.
] button.
Finish
Found New Hardware Wizard
" window closes, remove the
www. .com
information@itm.com1.800.561.8187
Page 47
For Windows 2000
45
5.2 Installing the Driver
When the "
1.
[
Next
Insert the CD-R supplied with the instrument into the CD-ROM
2.
drive.
Check "
3.
button.
Found New Hardware Wizard
] button.
Search for a suitable driver fo r my device
" window appears, click the
," then click the [
Next
1 2 3 4 5
5
6
]
7 8
www. .com
9 10
information@itm.com1.800.561.8187
Page 48
46
5.2 Installing the Driver
4.
5.
Check " click the [
Click the [
CD-ROM drives
] button.
Next
] button.
Next
" (thus deselecting the other items), then
Click the [
6.
When the " CD-R from the CD-ROM drive.
] button.
Finish
Add New Hardware Wizard
" window closes, remove the
www. .com
information@itm.com1.800.561.8187
Page 49
For Windows Me
Insert the CD-R supplied with the instrument into the CD-ROM
1.
drive. The "
Add New Hardware Wizard
5.2 Installing the Driver
" window appears.
47
1
2.
3.
Check " ton.
Click the [
Automatic search for a better driver
Finish
] button.
," then click the [
Next
] but-
2 3 4 5
5
6 7 8 9
When the "
4.
CD-R from the CD-ROM drive.
Add New Hardware Wizard
www. .com
" window closes, remove the
information@itm.com1.800.561.8187
Page 50
48
5.2 Installing the Driver
For Windows 98SE
1.
2.
When the " [
] button.
Next
Check " button.
Search for the best d riv er f or your device
Add New Hardware Wizard
" window appears, click the
," then click the [
Next
]
Insert the CD-R supplied with the instrument into the CD-ROM
3.
drive.
www. .com
information@itm.com1.800.561.8187
Page 51
5.2 Installing the Driver
49
4.
5.
Check " the [
Click the [
CD-ROM drive
Next
] button.
] button.
Next
" (thus deselecting the other items), then click
1 2 3 4 5
5
6 7
Clicking the button initiates copying of the file.
8 9 10
www. .com
information@itm.com1.800.561.8187
Page 52
50
5.2 Installing the Driver
When the following window subsequently appears, click the [
6.
button.
When the "
7.
CD-R from the CD-ROM drive.
Add New Hardware Wizard
" window closes, remove the
Finish
]
www. .com
information@itm.com1.800.561.8187
Page 53

6.1 Measurement Specifications

51

Specifications

6
6.1 Measurement Specifications
Optical-power measurement
Range selection Automatic and manual Accuracy ± 0.7% (± 5% in combination with an optional optical sensor) Maximum-value
hold Minimum-value
hold Average-value
display
Wavelength sensitivity correction
Scaling [Display value] = [Measured value] x [internal setting]
Unit: W/dBm
Displays the maximum value among the current measurements
Displays the minimum value among the current measurements
Displays the moving average for the set number of averaging steps The number of averaging steps is selectable between 2 and 100.
The setting is editable every 1 nm. The sensitivity of the sensor is corrected automatically. Up to 10 wavelengths can be preset (each sensor has unchangeable presets).
The internal setting is set for each wavelength and is selectable between 0.1 and 10. The resolution is the same as that for optical­power measurement.
1 2 3 4 5 6 7 8
Relative measurement
Display update rate
Analog output
Displays a relative value to a measured or set value [Display value] = [Measured value] - [Reference value] The reference value is set based on the measured value or through key entry by the user. The selectable value ranges from
0.001 nW to 1.9999 W (-90.00 dBm to 33.00 dBm). Approx. 330 ms
Output level: Approx. 1 V at the sensor calibration point Output resistance: 50 Ω ± 5% Connector: φ 3.5 Mini-jack * The calibration point of the sensor is determined based on the
calibration wavelength and the calibration power specified in the specification of the optional optical sensor.
www. .com
information@itm.com1.800.561.8187
*
9 10
Page 54
52
6.1 Measurement Specifications
Display monitor Monitor: LCD
Communica­tions function
Functions • Auto power save
Measurement display: 4-1/2 digits in the unit of nW, dBm, or dB Wavelength display: 4 digits in the unit of nm
Communications data:
Output of measured values at the request of the PC Change in setting at the command of the PC
Interface: USB Ver. 1.1
Automatically turns OFF the power if there is no key operation or USB communication for approximately 10 minutes. To turn OFF the function, hold down the POWER key for 5 sec. when turning ON the power. The auto-power-save function is available only when the 3664 is running on the battery.
• Setting backup Saves the settings when the power is turned OFF, and starts up the 3664 with the saved settings when the power is turned ON again.
• Battery check Displays the battery indicator when the supply voltage falls below 4.0 V
W, mW,
μ
www. .com
information@itm.com1.800.561.8187
Page 55
6.2 General Specifications
53

6.2 General Specifications

Rated supply voltage
Maximum rated power
Operating time • With maximum load
Dimensions Approx. 85W × 160H × 35D mm (3.35"W × 6.30"H × 1.38"D)
Mass Approx. 270 g (9.5 oz.) (without batteries) Operating
temperature and humidity
Storage temperature and humidity
Accuracy guar­antee for tem­perature and humidity Guaranteed accuracy period
Operating environment
• 1.5 V DC × 4 (LR6 alkaline batteries × 4)
• AC Adapter 9 V DC, 0.2 A Model 9445-02 AC ADAPTER or Model 9445-03 AC ADAPTER: Rated supply voltage: 100 to 240 V AC (Voltage fluctuations of ±10% from the rated supply voltage are taken into account.), 50/ 60 Hz, Maximum rated current 250 mA)
1.6 VA
Approx. 9 hours (Ambient temperature 23°C(74°F) reference value)
• Specified measurement state Approx. 60 hours (Ambient temperature 23°C(74°F) reference value)
(excluding projections)
0 to 40°C (32 to 104°F), 80%RH or less (non-condensating)
-10 to 50°C (14 to 122°F), 80%RH or le ss (non-condensating)
23 ± 5°C (73 ± 9°F), 80%RH or less (non-con densating)
One year
Indoors, altitude up to 2000 m (6562-ft.)
1 2 3 4 5 6 7 8 9 10
Applicable standards
Safety : EN61010-1:2001 Pollution degree II EMC : EN61326:1997+A1:1998+A2:2001+A3:2003
www. .com
EN61000-3-2:2000 EN61000-3-3:1995+A1:2001
information@itm.com1.800.561.8187
Page 56
54
6.2 General Specifications
Accessories Model 9445-02 AC ADAPTER
Options Model 9246 CARRYING CASE ........................................... 1
or Model 9445-03 AC ADAPTER
CAN)
......................................................................................... 1
Model 9094 OUTPUT CORD................................................... 1
Driver CD-R ............................................................................. 1
LR6 alkaline batteries .............................................................. 4
USB cable................................................................................ 1
Strap ........................................................................................ 1
Instruction manual.................................................................... 1
Model 9742 OPTICAL SENSOR.......................................... 1
Model 9742-10 OPTICAL SENSOR .................................... 1
(UL, SA10-9010N, SINO-AMERICAN)
(CEE, SA10-9010G, SINO-AMERI-
www. .com
information@itm.com1.800.561.8187
Page 57

7.1 Cleaning

55

Maintenance and Service

Never modify the instrument. Only Hioki service engineers should dis­assemble or repair the instrument. Failure to observe t hese precautions may result in fire, electric shock, or injury.
7.1 Cleaning

7.1.1 Cleaning of the 3664

To clean the instrument, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
7
1 2 3 4 5 6 7 8
www. .com
9 10
information@itm.com1.800.561.8187
Page 58
56

7.2 Service

7.1.2 Cleaning of the Optical Sensors

• Avoid touching the detector window with your bare hands. The detector window must be clean for the sensor to meet the speci­fied performance parameters.
• With the exception of ethyl alcohol, avoid using organic solvents to clean the detector window. These solvents can damage the detector window.
• Avoid scratching the detector window with sharp or pointed objects (e.g., the tips of tweezers) or against hard surfaces. Damage to the detector window may prevent the sensor from meeting specified performance parameters.
Ta ke the cleaning method of the 9742 OPTICAL SENSOR as an example.
1.
Open the sensor cove r. (When using the 9742-10 OPTICAL SENSOR , take care to avoid misplacing the sensor cover.
2.
Wipe the detector w indow of the sensor usin g lens cleaning paper or other lint-free ma t erial. If lint remains on the detector wind ow, blow off with an opti­cal lens airbrush. If the detector window is soil ed, fl uf f th e tip of a cot ton swab. Moisten the cotton tip in et hy l alcohol and wipe the surface .
7.2 Service
• If the instrument seems to be malfunctioning, confirm that the batteries are not discharged, and that the USB cable is not open circuited before contacting your dealer or Hioki representative.
• When sending the instrument for repair, remove the batteries and pack carefully to prevent damage in transit. Include cushioning material so the instrument cannot move within the package. B e sure to include details of the problem. Hioki cannot be responsi­ble for damage that occurs during shipment.
www. .com
information@itm.com1.800.561.8187
Page 59
7.3 Error Display
57

7.3 Error Display

In the event of an error, one of the following messages is dis­played.
Err0 Reference-value save error
An improper value has been saved as the reference value. If a reference value is saved while the measured value is "Lo" or "Hi" in the optical-power measurement mode, "REL" and "SAVE" remain on the screen and "Err0" is displayed. The refer­ence value is then updated to Hi or Lo, and the relative measurement display shows "
- -
".
Err1 Wavelength error
There is a problem with the measurement­wavelength memory. Reset the measure­ment-wavelength memory.
See Secti on 7.4.3 Measu rement-Wavelength
Memory Reset (page 60)
If the measurement-wavelength-memory reset does not solve the problem, conduct a system reset.
See Section 7.4.1 System Reset (page 59)
If the system reset does not solve the prob­lem, the 3664 may be malfunctioning. Contact your dealer or Hioki representative.
Err2 Setting-data error
- -
1 2 3 4 5 6 7 8 9
There is a problem with the setting data. Conduct a system reset.
See Section 7.4.1 System Reset (page 59)
If the system reset does not solve the prob­lem, the 3664 may be malfunctioning. Contact your dealer or Hioki representative.
www. .com
information@itm.com1.800.561.8187
Page 60
58
7.3 Error Display
Err3 Memory error
There is a problem with writing into the memory. Conduct a system reset.
S ee Section 7.4.1 System Reset (page 59)
If the system reset does not solve the prob­lem, the 3664 may be malfunctioning. Contact your dealer or Hioki representative.
Err4 System error
The 3664 may be malfunctioning. Contact your dealer or Hioki representative.
"1" Display over range
The value displayed after wavelength sensi­tivity correction or scaling exceeds the upper limit that can be displayed by the 3664 (1.9999 W).
"no SEnS" Sensor disconnected
A sensor is not connected. Connect a sen­sor. The 3664 cannot be operated without a sensor connected.
www. .com
information@itm.com1.800.561.8187
Page 61
7.4 Reset
59

7.4 Reset

There are three types of reset operations for the 3664.
•System reset Returns all settings to their defaults
• Setting-data reset Returns all settings except those for the wavelength data to their defaults
• Measurement-wavelength memory reset Returns the measurement-wavelength data stored in the mea­surement-wavelength memory to its default state.

7.4.1 System Reset

1.
While holding down the power.
2.
"CLr" and "ALL" start blinking, and a ll settin gs ar e r eturn ed to their defaults.
key and key, tu rn ON the
SET
Blinking
1 2 3 4 5 6 7 8

7.4.2 Setting Data Reset

1.
While holding down th e power.
2.
"CLr" and "SEt" start blinking, and all setting d ata except those for the wavelength is ret urned to its defaults.
www. .com
key and
SET
key, turn ON the
SAVE
Blinking
information@itm.com1.800.561.8187
9 10
Page 62
60
7.4 Reset

7.4.3 Measurement-Wavelength Memory Reset

1.
With an optional optical sensor connected to the 3664,
key and λ key, turn ON the
while holding down the power.
2.
"CLr" and "λ" start blinking, and the measurement-wave­length data is returned to th e default wavelength setting of the connected sensor.
The measurement-wavelength memory reset cannot be performed if a sensor is not connected to the 3664.
SET
Blinking
www. .com
information@itm.com1.800.561.8187
Page 63

7.5 Replacing the Batteries

7.5 Replacing the Batteries
61
• To avoid electric shock, turn off the power switch and dis­connect the USB cable before replacing the batteries.
• After replacing the batt eries, replace the cover and screws before using the instrument.
• Do not mix old and new batteries, or diffe rent types of bat­teries. Also, be careful to observe battery polarity during installation. Otherwise, poor perform ance or damage from battery leakage could result.
• To avoid the possibility of explosion, do not short circu it, disassemble or incinerate batteries.
• Handle and dispose of batteries in accordance with local regulations.
• The “ ” indicator appears when battery voltage becomes low. Replace the batteries as soon as possible.
• Use the specified battery (AA alkaline battery) only. If non­specified battery such as a manganese battery is used, the service hours may be shorter than when alkaline batteries are used.
Tools and materials: Four AA alkaline batteries (LR6) and a Phillips screw­driver
Make sure the power to the 3664 is
1.
OFF. Disconnect all cables.
2.
1 2 3 4 5 6 7 8 9
Turn over the 3664 and loosen the
3.
screw on the battery cover. Remove the battery cover.
4.
Insert the four AA alkaline batteries
5.
(LR6) into the battery case, noting the correct polarity. (When replacing the batteries, replace all of them.)
Place the battery cover back on, and
6.
tighten the screw.
www. .com
information@itm.com1.800.561.8187
Page 64
X

Appendix

63

Explanations

Optical-power measurement
Assuming that the energy of the input light measures E [mW], the measured value P is expressed as follows.
P [dBm] = 10log
P: Measured value [dBm] E: Energy of input light [mW]
10
E [mW]
1 [mW]
1 2
APP
4 5 6 7 8 9 10 11
APP
APP
www. .com
information@itm.com1.800.561.8187
Page 65
64

Options

Model 9742 OPTICAL SENSOR
Optical sensor for the 3664
Model 9742-10 OPTICAL SENSOR
Optical sensor for the 3664
Model 9246 CARRYING CASE
Carrying case holding the 3664, the complete accessory set, and an optional optical sensor
www. .com
information@itm.com1.800.561.8187
Page 66
4 5 6 7 8 9 10 11
www. .com
information@itm.com1.800.561.8187
Loading...