Warranty
Warranty malfunctions occurring under conditions of normal use in
conformity with the Instruction Manual and Product Precautionary
Markings will be repaired free of charge. This warranty is valid for a
period of one (1) year from the date of purchase. Please contact
the distributor from which you purchased the product for further
information on warranty provisions.
User Memo
Write the follows when you purchase this product.
This memo will be surely helpful to send the product to service.
The Production Number
The date of purchase
Shop Name
TEL
1FC6P110006705November 2013 Revised edition 5
21
The Temperature Logger
(Button-typed)
3650
Instruction Manual
Thank you for the purchase of our product.
Before using this system, read this User's Manual
for the proper operation.
This manual should be retained along with the
Operation Manual.
Those documents will be surely helpful for the
future use.
2
Instruction Manual
Before Use
Safety Instructions
Part Names and Functions
Features
Operation Instruction
Precautions for Use
Maintenance
The management of
the solid wastes
Specifications
Declaration of Conformity
Contents
3650 Instruction Manual (this document)
1
2
3
4
5~8
9
10
10
1 1
12
dealer or the user support.
※The product is shipped after the careful examination. If any defect is found, contact with the
Part Names and Functions
Main body
YYWW
000000
ABCDEF
3650
RR
The ID number
The product-specific
number
(12
The product number
digits
000000
ABCDEF
3650
)
Holder
Holder(Small)
Threaded screw hole
000000
ABCDEF
Holder(Large)
3
Magnet
Operation Instruction 1
1. Before use
Set the 3650 in preparation for the start of measurements (3650 Initial Settings)
The 3920-01 Data Reader (not included in this system ) and a personal computer are used
for the 3650 to start measuring temperature.
Refer to the 3920-01 Instruction Manual and the Software Operation Manual in details.
Features
Measuring temperature: Measures the temperature of the environment where the
3650 is placed.
Measuring temperature range: The measuring temperature range is as follows.
3650 : -40
Recording Duration
The following table shows the number of data that can be recorded in the memory
when the temperature is continuously measured.
Measuring
Time Interval
1 minute
3 minutes
5 minutes
10 minutes
30 minutes
1 hour
Possible recording counts: Over 500 thousand times
(4 years when measured in 5 minutes interval)
The usage in a place where the ambient temperature is over 50 may reduce the
defined recording counts.
※ For example,the battery will be exhausted if the 3650 is left at more than 80℃
for approximately half a year.
1 1.00
10.00
9.00
8.00
7.00
6.00
5.00
4.00
3.00
MINIMUM PRODUCT LIFETIME(YEARS)
2.00
1.00
0.00
Characteristics of relationship between battery life and operating temperature(example)
EVERY MINUTE
NO SAMPLES
EVERY 3 MINUTES
EVERY 10 MINUTES
-40-30-20-
10 10 20 30 40 50 60 70 800
The data recorded in the 3650 can be retrieved with the 3920-01 Data Reader
(Refer to the 3920-01 3920-01 Instruction Manual and the Software Operation Manual in details),
※ Recorded date cannot be read out when the battery power is close to nil.
Operation Instruction 2
3.Initial settings 2 to start measuring temperature
(Example) The setting screen to start measuring temperature using the analysis software for
Windows (3920-01)
・Adjust the date and time to the computer.
・ Set the starting date and time to 19 minutes later than the current time.
・ Set the Measuring Time Interval to 1 minute.
・ et the High Temperature alarm at 40.0 and the Low Temperature at -10.0 .
℃ to +85℃
Data Counts
1.4 day
5.2 days
7 days
14 days
42 days
85 days
TEMPERATURE(℃)
4
other metals or batteries.
DO NOT keep the product with
Sanitation Law.
to food. It may violate the Food
DO NOT expose the magnet directly
Warning
a case happens.
Contact a doctor immediately if such
to avoid accidental swallowing.
mouth.Keep it away from children
DO NOT put the product in the
service personnel.
No Disassembling
fire disaster. Refer to the qualified
may cause abnormal overheat or
or service the products. Such an act
DO NOT attempt to disassemble
Instructions for Use
● After reading this manual, keep it in a place where anyone can have access at any time.
● The mark within a black circle denotes the required procedure.
consequences in some cases when the product is not handled properly.
indicates a potentially hazardous situation which could result in serious
serious injury when the product is not handled properly.
indicates a potentially hazardous situation which could result in death or
Caution
Warning
situation which, if not avoided, could result in serious consequences in some cases.
if not avoided, could result in death or serious injury. CAUTION also indicates such a hazardous
WARNING and CAUTION. WARNING indicates a potentially hazardous situation which,
prevent the risk of harming or damaging people. Safety instructions are divided into two sections;
● Safety Instructions written in this manual are intended for the proper use of our product to
● Before using, read this 'Safety Instructions' thoroughly for the proper operation.
2
1
1
1
1
Safety Instructions
4×12
Screws for fixing holders
Magnet
Holder(Small)
Holder (Large)
3650
ABCDEF
000000
YYWW
RR
3650 Temperature Logger
Before Use
2. Initial settings 1 to start measuring temperature
●
Basic Information
The ID number of the 3650 is unique throughout all the 3650s.
The number will be used when the analysis software reads the data.
Write down the number in advance before the setting.
●
Items to set
・Set the Date and Time
The 3650 internal clock is displayed. Mark off the checkbox next to "Adjust to PC Time"
to adjust to the clock of computer.
・Set the Date to Start Measuring
Set the date and time to start measuring temperature.
・Set the Measuring Interval
Set the time interval between the consecutive measurements of temperature.
・Set the Recording Mode
The 3650 can record two thousand and forty eight (2048) data at maximum. There are
two possible modes to choose from in the recording.
Enable memory overwriting
"Yes" :Record the newest data until now (max. 2048)
"No" : Record the oldest data from start of measuring (max. 2048)
・Set the Temperature Alarm Warning
Set the warning temperature when measuring temperature.
Data outside of this range will be recorded as the alarm temperature time.
Take out the user's manual and make sure that the following items are included in your package.
Check the Packaging
65

Operation Instruction 3
Operation Instruction 4
Specifications
●
Fitting in the Holder
① Fit the holder (Small) in the frame.
② Press the protruding part of the holder (Small).
②Press
①Put
● Removing the Holder
① Take hold of the holder with your finger
inserting into the depression of
the either side of the holder.
② Pull the holder (Small) toward you.
③ Remove the holder (Small).
②Pull
③Remove
① Inserting finger
into the
depression
The included magnet or screws are used depending on the place to fix
the holder.
● Attaching the magnet
Use the scotch tape at the back of the
magnet to attach to the frame at the
back side of the holder (Large) firmly.
● Fixing with screws
Screw the holder (Large) as illustrated
below.
Remove the 3650 out of the holder when used without the holder.
● Removing the 3650
①Remove
Connect to the 3920-01 Data Reader
● Used with the holder
Connect to the 3920-01 Data Reader with
the 3650 in the holder (Small)
● Used without the holder
Remove the 3650 from the holder and
connect to the 3920-01 Data Reader
● Connect to the 3920-01 Data Reader
Connect the 3920-01 Data Reader with the 3650 fit in the holder (Large)
Product Name
Model Name
Measuring Temperature Count
Measuring Temperature Range
Operating Temperature Range
Storage Temperature Range
Resolution
Maximum Measuring Counts
Accuracy of measurements
Accuracy of clock
Resistance to water
Conformity
Dimension
Weight
The Declaration of Conformity for instruments that comply to CE mark
requirements may be downloaded from the HIOKI website.
Temperature Logger (button-typed)
(The usage in a place where the ambient temperature
is over 50℃ may reduce the specified count.
Battery life is approximately half a year when
operating continuously or storing for a long time
at more than 80℃.)
- 10.0℃ to 30.0℃ ±1.5℃
0℃~45℃ ±1 minute per month
other - 8minutes per month max.
17.35 mm across and 5.89 mm thick
3650
1
- 40℃ to 85℃
- 40℃ to 85℃
- 40℃ to 85℃
0.5℃
Over 500,000
- 40.0℃ to ー10.5℃ ±2℃
30.5℃ to 85.0℃ ±2℃
Ordinary water resistance
EN61326
3.3g
7
Precautions for Use
Precautions for Use
Avoid using the 3650 in any place where
it will be subject to vibration.
Otherwise, it may cause failure.
Avoid using the 3650 in water.
Otherwise, it may cause failure.
※When installing the 3650 in water or in an environment where water always splashes,
please use the 3650 well sealed in a plastic bag.
● Before Use
Before you use the 3650 for a long time, be sure to check the total counts.
If the count is close to five hundred thousand, the battery may wear out before the
measurements are complete.
Before you use the 3650, it is also recommended that it be tested in a normal
temperature for several times to confirm its proper operation.
※The usage in an environment where the ambient temperature is over 50℃ may shorten
battery life.
※Recorded date cannot be read out when the battery power is close to nil.
Avoid exposing the 3650 directly to the sun
or using it near fire. Otherwise, it may cause
failure.
Avoid using or storing the 3650 in any place
exposed to rain. Otherwise, it may cause
failure.
8 11
Maintenance
●
Regular cleaning is recommended to keep the 3650 clean all the time.
●
Don't wipe with polish, powder soap, benzene, oil or hot water.
Those substances may damage coating or plastics.
Brush
Polish
Holder, the 3650
1
Stop the operation of the 3650 and wipe with a dry cloth.
Check after the maintenance
2
After finishing the maintenance, check the following items for a proper operation.
●
There is no fissure in the holder.
●
No equipment is abnormally heated.
If anything seems to be wrong, feel free to ask to the store where you purchased
this product or the service center.
Thinner
Benzine
Acids
Gasoline
Hot Water
The management of the solid wastes
Instructions for disposal
Attention
Dispose the wastes as noncombustible. However, under various state and local laws,
it may be illegal to dispose of the wastes that way. Check with your local solid waste
officials for details in your area for proper disposal.
9
10