Hioki 3645-20 Instruction Manual

Page 1
INSTRUCTION MANUAL
For...は専用機種。複数の場合は「/」で区切る。
不要の場合はとる。
形名を入力。 複数の場合は「/」で区切る。
3645-20
品名を入力。
VOLTAGE LOGGER
Page 2
Page 3
Contents
Chapter 1 Product Outline 1
1.1 Name and Functions of Parts 2
1.2 Interval and Maximum Recording Time
1.3 Timings of Measurement
Chapter 2 Set Up 7
2.1 Replacing the Battery 7
2.2 Power Save Function
2.3 Setting Current Time
2.4 Connecting 9639 CONNECTION CABLE
2.5 Preheat Signal
10
11
12
3
4
9
Chapter 3 Settings 15
3.1 Setting Items 15
3.2 Manual Setting
3.3 Setting by COMMUNICATION BASE
18
21
Chapter 4 Specifications 23
Chapter 5 Maintenance and Service 25
Page 4
Page 5
_____________________________________________________________________
Introduction
Thank you for purchasing this HIOKI "3645-20 VOLTAGE LOGGER." To get the maximum performance from the unit, please read this manual first, and keep this at hand.
Inspection
When the unit is delivered, check and make sure that it has not been damaged in transit. If the unit is damaged, or fails to operate according to the specifications, contact your dealer or HIOKI representative.
Accessories
9639 CONNECTION CABLE 9632 CONNECTION CABLE Instruction Manual LR03 alkaline battery X 4 (built into this unit, for monitor)
i
Before using the product the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
NOTE
______________________________________________________________
Testing monitor batteries installed in the unit may possibly be weak. Replace batteries before extended measurement usage. Before using the product, make sure that the insulation on the cables is undamaged and that no bare conductors are improperly exposed. Using the product in such conditions could cause an electric shock, so contact your dealer or Hioki representative for repair.
Inspection
Page 6
ii
R
_____________________________________________________________________
SafetyNotes
This equipment is designed according to IEC 61010 Safety Standards, and has been tested for safety prior
DANGE
Safety Symbols
to shipment. Incorrect measurement procedures could result in injury or death, as well as damage to the equipment. Please read this manual carefully and be sure that you understand its contents before using the equipment. The manufacturer disclaims all responsibility for any accident or injury except that resulting due to defect in its product.
This Instruction Manual provides information and warnings essential for operating this equipment in a safe manner and for maintaining it in safe operating condition. Before using this equipment, be sure to carefully read the following safety notes.
This symbol is affixed to locations on the equipment where the operator should consult corresponding topics in this manual (which are also marked with the symbol) before using relevant functions of the equipment. In the manual, this mark indicates explanations which it is particularly important that the user read before using the equipment.
Indicates DC (Direct Current).
______________________________________________________________
Safety Notes
Page 7
_____________________________________________________________________
R
iii
The following symbols are used in this Instruction Manual to indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents extreme
DANGE
WARNING
CAUTION
NOTE
danger of accident resulting in death or serious injury to the user.
Indicates that incorrect operation presents significant danger of accident resulting in death or serious injury to the user.
Indicates that incorrect operation presents possibility of injury to the user or damage to the equipment.
Denotes items of advice related to performance of the equipment or to its correct operation.
Accuracy
The specifications in this manual include figures for "measurement accuracy" when referring to digital measuring instruments, and for "measurement tolerance" when referring to analog instruments.
f.s.
(maximum display or scale value, or length of scale) Signifies the maximum display (scale) value or the length of the scale (in cases where the scale consists of unequal increments or where the maximum value cannot be defined). In general, this is the range value (the value written on the range selector or equivalent) currently in use.
rdg.
(displayed or indicated value) This signifies the value actually being measured, i.e., the value that is currently indicated or displayed by the measuring instrument.
dgt.
(resolution) Signifies the smallest display unit on a digital measuring instrument, i.e., the value displayed when the last digit on the digital display is "1".
______________________________________________________________
Safety Notes
Page 8
iv
_____________________________________________________________________
Measurement categories (Overvoltage categories)
This product complies with CAT I safety requirements. To ensure safe operation of measurement product, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT I to CAT IV, and called measurement categories. These are defined as follows.
CAT I
Secondary electrical circuits connected to an AC electrical outlet through a transformer or similar device.
CAT II
CAT III
CAT IV
Primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets.
The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel).
Higher-numbered categories correspond to electrical environments with greater momentary energy. So a measurement device designed for CAT III environments can endure greater momentary energy than a device designed for CAT II. Using a measurement product in an environment designated with a higher-numbered category than that for which the product is rated could result in a severe accident, and must be carefully avoided. Never use a CAT I measuring product in CAT II, III, or IV environments. The measurement categories comply with the Overvoltage Categories of the IEC60664 Standards.
______________________________________________________________
Safety Notes
Page 9
_____________________________________________________________________
R
Notes on Use
In order to ensure safe operation and to obtain maximum performance from the unit, observe the cautions listed below.
The maximum input voltage is 60 V DC. Attempting
v
DANGE
WARNING
to measure voltage in excess of the maximum input could destroy the product and result in personal injure or death.
To avoid electric shock, do not allow the product to get wet, and do not use it when your hands are wet. To avoid electric shock, shut off the line power supply before connecting the connection cable to the terminals.
CAUTION
To avoid electric shock, before replacing the connection cable, remove the all connection cable from the device to be measured. To avoid damaging connection cable (especially where the leads connect to the probes), do not kink or pull on the leads. This product is designed for indoor use, and operates reliably from 0
to 40 . Do not store or use the product where it could be exposed to direct sunlight, high temperature or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate so that it no longer meets specifications. This product is not designed to be entirely water- or dust­proof. To avoid damage, do not use it in a wet or dusty environment.
______________________________________________________________
Notes on Use
Page 10
vi
_____________________________________________________________________
Measurement Flow Chart and Reference Guide
Is there sufficient
battery charge remaining
to operate 3645-20
Setting current time
Connecting sensor
Setting 3645 stand­alone
IntervalPreheat timeRange
See 3.2.
Press REC/STOP button to start measurement.
See 3.2.
When recording data reaches 32000, measurement stops.
?
Set current time when using 3645-20 for the first time or replacingbatteries. See 2.3.
See 2.4.
Setting 3645 with COMMUNICATION BASE
IntervalCurrent timeStart controlRecording method
Manual start
Starts measurement immediately.
One time
When recording data exceeds 32000, it overwrites previously recorded data.
Replace batteries
COMMUNICATION BASE and
Interval Preheat timeCurrent time Start control Recording methodComment
No
recordingmethod
Settingstart control
Start control options
Starts measurement at prescheduled time.
Setting
Setting 3645 with
Endless
PC
Yes
Time prescheduled start
See 2.1.
Press REC/STOP button to stop measurement.
After measurement is complete, disconnect connection cord. Use COMMUNICATION BASE to send measurement data to personal computer. (See COMMUNICATION BASE instruction manual.)
______________________________________________________________
Measurement Flow Chart and Reference Guide
Page 11
_____________________________________________________________________
1
Chapter 1
Product Outline
3645-20 VOLTAGE LOGGER enables measurement with intervals and recording up to This logger is designed with four measurement ranges covering the outputs of various types of sensors, and the voltages of diverse devices and power supplies. The 3645-20 comes equipped with a Preheat function. When a synchronized Preheat signal is used in the measurement, the logger allows control of the power supply to various types of sensors. Data is saved in nonvolatile memory when batteries are weak or removed for replacement.
50 mV to 50 V, to enable measurement of
50 VDC.
NOTE
______________________________________________________________
3645-20 cannot be set with 3910-20 COMMUNICATION BASE. Use 3911-20, 3912-20 COMMUNICATION BASE to set 3645-20.
Product Outline
Page 12
2
(
_____________________________________________________________________
1.1 Name and Functions of Parts
LCD
Optical data transferports
INTERVAL button
Connection terminal
LCD Displays measurement value and settings.
Optical data transfer ports
INTERVAL button Calls up interval setting display to set interval
Connection terminal Connects 9639 CONNECTION CABLE.
REC/STOP button Pressing more than 1 second initiates or stops
(SELECT) button Interval is selected in interval setting display.
Preheat output terminal
RANGE button Changes measurement range from 50 mV/500
Enables optical data transfer to COMMUNICATION BASE.
and preheat time.
recording.
When the 9632 CONNECTION CABLE is plugged into this terminal, it outputs the Preheat signal (open-collector output).
mV/5 V/50 V.
REC/STOP button
SELECT)button
Preheat output terminal
Range button
______________________________________________________________
Product Outline
Page 13
_____________________________________________________________________
1.2 Interval and Maximum Recording Time
Interval and maximum recording time (when power save function is valid) are as follows. Maximum recording time is limited by battery charge condition. Maximum recordable data is 32000 per unit.
INTVL Maximum Recording Time
1s 8h53min20s
2s 17 h 46 min 40 s
5s 1 day 20 h 26 min 40 s
10 s 3 day 16 h 26 min 40 s
15 s 5 day 13 h 20 min
20 s 7day9h46min40s
30 s 11 day 2 h 40 min
1min 22 day 5 h 20 min
2min 44 day 10 h 40 min
5min 111 day 2 h 20 min
10 min 222 day 5 h 20 min
15 min 333 day 8 h
20 min 444 day 10 h 40 min
30 min 666 day 6 h
60 min 1333 day 8 h
3
______________________________________________________________
Product Outline
Page 14
4
_____________________________________________________________________
1.3 Timings of Measurement
When Preheat is OFF
(1) Recording ("
In the recording mode, the unit conducts measurement once the preset recording interval has elapsed. It then records the measurement value and updates the LCD indication. When the power-save function is disabled, or when the power-save function is active and any button is pressed to restart the LCD indicator, the unit conducts measurement every second and updates the LCD indication only, until the preset recording interval elapses.
Measurement
Indication updated
Recording
LCD
(2) Monitoring ("
In the monitoring mode, the unit conducts measurement every second and updates the LCD indication.
Measurement
Indication updated
LCD
" on)
REC
①→    ②→    ③→④→⑤→⑥→⑦→⑧→⑨→⑩→
①     ②     ③ ④ ⑤ ⑥ ⑦ ⑧ ①     ②     ③     ⑥     ⑨
Display OFF
Measurement (1 s)
Interval
" off)
REC
①→②→③→④→⑤→⑥→⑦→
① ② ③ ④ ⑤ ⑥ 
Display OFF
Display ON
Display ON
Measurement (1 s)
______________________________________________________________
Product Outline
Page 15
_____________________________________________________________________
When Preheat is ON
5
(1) Recording ("
In the recording mode, the unit first outputs the Preheat signal once the preset recording interval has elapsed, and then conducts measurement. It records the measurement value and updates the LCD indication.
Measurement
                  ②   
Indication updated
Recording
Preheat time
LCD
(2) Monitoring ("
In the monitoring mode, the unit conducts measurement every second and updates the LCD indication. In this condition, the unit delivers the Preheat signal permanently.
Measurement
Indication updated
Recording
Preheat time
LCD
" on)
REC
①→      ②→      ③→      ④→
①       ②       ③       ④
Display OFF
Preheat time
REC
Display OFF
" off)
①→②→③→④→⑤→⑥→⑦→⑧→⑨→⑩→⑪
Measurement (1 s)
Display ON
Measurement (1 s)
Interval
Display ON
NOTE
If the Preheat function is active, the measurement value recorded in the immediately preceding measurement is displayed in Recording ("REC" on) until the recording interval elapses, allowing the following measurement to be conducted. When the power-save function is active and any button is pressed to conduct monitoring, quite some time may be required for stable measurement values to be obtained, depending on the response time of the connected sensor (time required for the output to stabilize after the sensor is powered up).
______________________________________________________________
Product Outline
Page 16
6
_____________________________________________________________________
______________________________________________________________
Product Outline
Page 17
_____________________________________________________________________
7
Chapter 2
Set Up
2.1 Replacing the Battery
During battery replacement, use caution not to put any
WARNING
foreign materials such as a metal object into the unit to avoid damage to the unit. Before using the product after replacing the batteries, replace the cover and screw. Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result. Handle and dispose of batteries in accordance with local regulations.
CAUTION
When exchanging the batteries, the circuit may sometimes short circuit due to static electricity. As far as possible, do not touch the base board with bare hands.
Installing new batteries ensures about 1 year of recording. (reference value with interval setting set to 1 minute or more, preheat is OFF and power save setting set to valid and at 20
) Remaining battery power indicator ( ) indicates remaining battery life reducing incrementally from right. Empty battery power indicator (
) indicates time
to replace batteries.
______________________________________________________________
Set Up
Page 18
8
_____________________________________________________________________
1. Remove back cover screw to remove cover. Verify polarity and install four new LR03 alkaline batteries.
2. Fit cover properly and tighten screw.
______________________________________________________________
Set Up
Page 19
_____________________________________________________________________
9
2.2 Power Save Function
Display window is automatically turned off in approximately 15 seconds after last key entry. (Sleep) However, while recording, each conditions.
Press any button to turn display on to display measurement value or to set settings. Note when interval setting display is on, sleep does not engage with no button press.
Initially, power save function is on. To turn off power save function, follow the instructions below. When power save function is off, maximum continuous duration is approximately 20 days.
REC
/ / mark shows
Sleeping......
1. Connect logger, COMMUNICATION BASE and personal
computer.
2. Start up application software packaged with
COMMUNICATION BASE.
3. Go to Communications on the menu bar and select Power
Save Options. Choose Off to turn off power save function.
NOTE
______________________________________________________________
See COMMUNICATION BASE instruction manual to connect logger and to install application software. To use application software, see operation guide.
Set Up
Page 20
10
_____________________________________________________________________
2.3 Setting Current Time
When replacing 3645-20 VOLTAGE LOGGER batteries or using 3645-20 stand-alone (with manual operation) for the first time, connect with COMMUNICATION BASE and set current time.
See how to set current time in COMMUNICATION BASE instruction manual.
______________________________________________________________
Set Up
Page 21
_____________________________________________________________________
11
2.4 Connecting 9639 CONNECTION CABLE
If the end of the 9639 CONNECTION CABLE short-
WARNING
circuits lines with a voltage between them, this is very dangerous and can lead to a serious accident. Exercise great care when measuring voltages. The maximum input voltage is 60 VDC. Do not measure voltage in excess of these limitations, as doing so may damage the unit or cause an accident that might result in injury or death.
Connect 9639 CONNECTION CABLE to 3645-20.
9639 CONNECTION CABLE
3m
______________________________________________________________
Set Up
Page 22
12
_____________________________________________________________________
2.5 Preheat Signal
Plug in the 9632 CONNECTION CABLE. To do so, insert the cable with the mark on the connection terminal that matches the cable.
Inappropriate insertion of the cable is likely to result in malfunction of the unit. Particularly following communication with the COMMUNICATION BASE connected to the unit, make sure the connection terminal is not mistakenly lifted.
(1) 9632 CONNECTION CABLE
mark on the connection
GND
Preheat signal
White: Preheat signal Black: GND
Cable length: 1000 mm
(2) Internal Circuit (Preheat Signal)
FET
Preheat signal
Flow of current
GND
Open-drain output Rating: 30 V/20 mA max. (Allowable loss: 200 mW) GND Ground signal (0 V)
*Open-drain output The switch (FET), which is inserted between the Preheat signal and the GND signal, turns on when this signal becomes active (in Preheat mode). Make a connection such that the current will flow from the Preheat signal toward the GND.
______________________________________________________________
Set Up
Page 23
_____________________________________________________________________
13
(3) Example of a Preheat-Signal Connection Circuit
The Preheat signal controls the sensor power on/off. Connect the 3645-20 to a sensor, power supply, and relay in the manner illustrated below.
Counter-electromotive force protective diode
Confirm that the power supply to be used does not exceed the ratings (30 V, 20 mA max.) of the Preheat signal. When selecting a relay, note its ratings (voltages and currents to be imposed on the coil and the relay contacts) and the polarity of the coil. When connecting a relay, be aware that the circuit may be destroyed by a counter-electromotive force generated when the relay is turned on/off. To prevent such destruction of the circuit, always insert a protective diode in the manner illustrated above.
______________________________________________________________
Set Up
Page 24
14
_____________________________________________________________________
______________________________________________________________
Set Up
Page 25
_____________________________________________________________________
Chapter 3
Settings
3.1 Setting Items
Logger stand-alone manual settings and settings in combination with COMMUNICATION BASE with measurement conditions stored in memory loaded from personal computer.
15
NOTE
3645-20
1. Start recording Vaild Vaild Vaild
2. Stop recording Vaild --- ---
3. Interval setting Vaild Vaild Vaild
4. Preheat time setting Vaild --- Vaild
5. Current time setting --- Vaild Vaild
6. Start control --- Vaild Vaild
7. Recording method setting
8. Range setting Vaild --- Vaild
9. Comments --- --- Vaild
--- Vaild Vaild
3645-20
COMMUNICATION
BASE
3645-20
COMMUNICATION
BASEPC
The Setup of Preheat Time and Comment is valid only where the PC, the COMMUNICATION BASE, and the 3645 are interconnected.
______________________________________________________________
Settings
Page 26
16
_____________________________________________________________________
1. Start recording
Start manual recording by pressing logger REC/STOP button for 1 second or initiate by time prescheduled start set using COMMUNICATION BASE. When time scheduled start is engaged, "
" icon appears in display. When batteries are weak, recording does not start. During recording, weak battery interrupts recording.
REC
Manual start
During recording
REC
Time prescheduled start
NOTE
Prescheduled
The "
" icon will dim after recoring begins.
During recording
2. Stop recording
Stop recording by pressing logger REC/STOP button for 1 second.Or recording stops automatically when data is full when set to recording method: one time.
3. Interval setting
Set interval with logger alone or using COMMUNICATION BASE. (1/2/5/10/15/20/30 s, 1/2/5/10/15/20/30/60 min)
4. Setting of Preheat Time
The Preheat time can be set for the 3645 alone or when the 3645, the COMMUNICATION BASE, and a PC are interconnected. The Preheat time can be chosen from among eight settings (OFF/0.5 sec/1 sec/2 sec/5 sec/10 sec/30 sec/60 sec).
A Preheat time longer than the recording interval cannot be set.
______________________________________________________________
Settings
Page 27
_____________________________________________________________________
17
5. Current time setting
To set current time, see COMMUNICATION BASE instruction manual.
6. Start control
Set specific recording date and time using COMMUNICATION BASE to engage time scheduled start. When time scheduled start is engaged, " display. The "
" icon will dim after recoring begins.
" icon appears in
7. Recording method setting
Set recording method using COMMUNICATION BASE. Choose either one time or endless recording method. Default setting is one time. One time: Ends recording when data reaches 32000. Endless : Overwrites previously recorded data when data
exceeds 32000.
8. Range setting
The range can be set for the 3645 alone or when the 3645, the COMMUNICATION BASE, and a PC are interconnected. Four different measurement ranges, 50 mV/500 mV/5 V/50 V, are available. Select the one best suited to the voltage to be measured.
9. Comments
Set comments entered by personal computer to logger using COMMUNICATION BASE. When sorting collected recording data, comments are helpful. Comment setting is available when personal computer is connected to both logger and COMMUNICATION BASE.
______________________________________________________________
Settings
Page 28
18
_____________________________________________________________________
3.2 Manual Setting
3645-20 VOLTAGE LOGGER stand-alone manual operation settings are shown below.
(1) Interval setting
Press INTERVAL button to switch measurement value display to interval setting display. (" Press SELECT button to designate interval. Press INTERVAL button to complete setting.
(2) Setting of Preheat Time
Pressing the Range button on the Recording Interval Setup Screen ("
" displayed) displays and changes the
INTVL
Preheat time (0.5 sec displayed in the illustration). Pressing the [SELECT] button again returns you to the Recording Interval Setup Screen.
INTVL
" appears.)
______________________________________________________________
Settings
Page 29
_____________________________________________________________________
19
The preheat time cannot be defined to be longer than the recording interval. Conversely, in the setup of the recording interval, you cannot select a shorter recording interval than the preheat time that is currently set. If you choose a longer preheat time than the preset recording interval and press the [Recording Interval] button to complete the setup, the preset recording interval will automatically be rewritten with a longer time than the preheat time you have just chosen. When you press the [Recording Interval] button to complete the setup after setting a preheat time, the system will briefly display the length of the recording interval that is currently set. At this time you should verify the settings you have made.
(3) Range setting
Pressing the Range button on the Monitor screen switches the measurement range among 50 mV, 500 mV, 5 V, and 50 V. As the measurement range is changed, the decimal point will also change position.
50 mV range
5 V range
______________________________________________________________
500 mV range
50 V range
Settings
Page 30
20
_____________________________________________________________________
(4) Starting and ending recording
Press REC/STOP button for 1 second to clear last recorded data and start recording. ("
REC
" appears.) Press REC/STOP button for 1 second to stop recording. When memory is full, recording automatically stops when recording method: one time is selected. When batteries are weak, recording does not start. During recording, weak batteries interrupt recording.
______________________________________________________________
Settings
Page 31
_____________________________________________________________________
21
3.3 Setting by COMMUNICATION BASE
1. Press logger INTERVAL button lightly to display LCD.
2. When logger LCD shows " REC/STOP button for more than 1 second to stop recording. During recording or waiting time before recording start time, data transfer cannot be established with COMMUNICATION BASE.
3. Press logger INTERVAL button to display interval setting display. ("
INTVL
" appears.)
4. Connect COMMUNICATION BASE with logger.
5. Press COMMUNICATION BASE SEND button for more than 1 second to send data settings to logger.
"markor " " icon, press
REC
NOTE
______________________________________________________________
Previously recorded logger data is erased when recording is resumed. Be sure to load data to be saved to COMMUNICATION BASE or to personal computer before recording. At any other time even when interval setting display is not shown, except during recording and waiting for recording, communication with COMMUNICATION BASE is available. However communication is disabled when logger is set to sleep.
Settings
Page 32
22
_____________________________________________________________________
3645-20 settings in application software COMMUNICATION UTILITY packaged with COMMUNICATION BASE are as follows.
Go to 'Communication' on the menu bar in COMMUNICATION UTILITY and select 'Set measurement condition'. When measurement condition setting window is open, select '3645 setting items' to set settings.
NOTE
______________________________________________________________
Settings
Comment, range and preheat Time are only available in 3645 setting items. Personal computer, COMMUNICATION BASE and 3645-20 must be connected during setting. Common settings are available to be set in '3911, 3912 setting items'. This enables 3645-20 and COMMUNICATION BASE settings.
Page 33
_____________________________________________________________________
Chapter 4
Specifications
Input DC voltage
23
Maximum input voltage
Number of input 1ch
Measurement range DC 50 mV/ 500 mV/ 5V/ 50 V
Accracy 0.5% rdg. 5 dgt.
Guaranteed accuracy period
Effect of radiated radio-frequency electromagnetic field
LCD display Measurement value, Interval, Battery status
Interval 1/2/5/10/15/20/30 s, 1/2/5/10/15/20/30/60 min
Preheat time OFF/0.5/1/2/5/10/30/60 s
Recording capacity 32000 data
Recording start Manual start, Time prescheduled start
Recording stop Manual stop, Memory full
Recording method One time, Endless
Preheat output Open collector output (30 V, 20 mA MAX.)
Data backup Available (Data not erased by weak batteries or
Interface Infrared optical data transfer
Power supply LR03 alkaline battery X 4 (1.5 VDC X 4)
Maximum rated power 0.1 VA
60 VDC
Accuracy warranty range: 23 5 (73 9 ) Temperature coefficient: (0.02%rdg. 1.5 dgt.)/
1 year
20 dgt. at 3 V/m
(remaining battery power indicator: 4 phases) Unit (mV,V), recording (REC), prescheduled ( )
*: The preheat time cannot be defined to be longer
than the recording interval.
battery replacement)
______________________________________________________________
Specifications
Page 34
24
_____________________________________________________________________
Battery life About 1 year
(temperature at 20 , power save function: valid, interval: 1 minute, preheat: OFF) About 140 days (temperature at 20 , power save function: valid, interval: 1 minute, preheat: 5 s) About 20 days (temperature at 20 , power save function: OFF)
Dimensions Approx. 57W X 86H X 30D mm (excluding projections)
2.24"W X 3.39"H X 1.18"D
Mass Approx. 130 g (4.6 oz) (including batteries)
Location for use Indoors, altitude up to 2000 m
Operate temperature and humidity range
Storage temperature and humidity range
Accessories LR03 alkaline battery X 4
Options COMMUNICATION BASE
Standards Applying EMC EN61326
0to40 , 80% RH or less (no condensation) (32 to 122 )
-10 to 50 , 80% RH or less (no condensation) (14 to 140 )
9639 CONNECTION CABLE 9632 CONNECTION CABLE Instruction Manual
Safety EN61010 Measurement category l (anticipated transient overvoltage 330 V) Pollution Degree 2
______________________________________________________________
Specifications
Page 35
_____________________________________________________________________
25
Chapter 5
Maintenance and Service
Cleaning
To clean the product, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case. Wipe the LCD gently with a soft, dry cloth.
Service
If the product seems to be malfunctioning, confirm that the batteries are not discharged, and that the sensor is not open circuited before contacting your dealer or Hioki representative. Pack the product carefully so that it will not be damaged during shipment, and include a detailed written description of the problem. Hioki cannot be responsible for damage that occurs during shipment.
Error Message
The following error may be displayed on the LCD of the main instrument as shown below.
Error message Meaning
ROM error
RAM error
Adjustment data error
When this occurs, repair or check the device. Contact your dealer or Hioki representative.
______________________________________________________________
Maintenance and Service
Page 36
26
_____________________________________________________________________
______________________________________________________________
Maintenance and Service
Page 37
Page 38
Page 39
HIOKI 3645-20 VOLTAGE LOGGER
Instruction Manual
Publication date: May 2008 Revised edition 6
Edited and published by HIOKI E.E. CORPORATION
Technical Support Section
All inquiries to International Sales and Marketing Department
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL: +81-268-28-0562 / FAX: +81-268-28-0568
E-mail: os-com@hioki.co.jp
URL http://www.hioki.com/
Printed in Japan 3645A981-06
All reasonable care has been taken in the production of this manual, but if you find any points which are unclear or in error, please contact your supplier or the International Sales and Marketing Department at HIOKI headquarters. In the interests of product development, the contents of this manual are subject to revision without prior notice. The content of this manual is protected by copyright. No reproduction, duplication or modification of the content is permitted without the authorization of Hioki E.E. Corporation.
Page 40
HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp / URL http://www.hioki.com/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA TEL +1-609-409-9109 / FAX +1-609-409-9108
3645A981-06 08-05H
Printed on recycled paper
Loading...