Thank you for purchasing the HIOKI 3639-20 PULSE
LOGGER. To obtain maximum performance from the
product, please read this manual first, and keep it handy for
future reference.
Inspection
When you receive the product, inspect it carefully to ensure
that no damage occurred during shipping. If damage is
evident, or if it fails to operate according to the
specifications, contact your dealer or Hioki representative.
Accessories
9629 CONNECTION CABLE (for measurement)
9632 CONNECTION CABLE (for alarm output)
Instruction Manual
LR03 alkaline battery X 4
(built into this product, for monitor)
i
Testing monitor batteries installed in the
unit may possibly be weak. Replace
batteries before extended measurement
usage.
This product is designed to conform to IEC 61010
Safety Standards, and has been thoroughly tested for
DANGE
Safety Symbols
safety prior to shipment. However, mishandling
during use could result in injury or death, as well as
damage to the product. Be certain that you
understand the instructions and precautions in the
manual before use. We disclaim any responsibility for
accidents or injuries not resulting directly from
product defects.
This manual contains information and warnings essential
for safe operation of the product and for maintaining it in
safe operating condition. Before using the product, be sure
to carefully read the following safety notes.
Thesymbol printed on the product indicates that the
user should refer to a corresponding topic in the
manual (marked with the
relevant function.
In the manual, thesymbol indicates particularly
important information that the user should read before
using the product.
symbol) before using the
Indicates DC (Direct Current).
The following symbols in this manual indicate the relative
importance of cautions and warnings.
This product complies with CAT I safety requirements.
To ensure safe operation of measurement products, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT I to CAT IV, and called measurement
categories. These are defined as follows.
CAT
Secondary electrical circuits that are connected
Ⅰ
to a wall outlet through a transformer or similar
device.
CAT
Primary electrical circuits in equipment
Ⅱ
connected to a wall outlet via a power cord
(portable tools, household appliances, etc.)
CAT
Primary electrical circuits of heavy equipment
Ⅲ
(fixed installations) connected directly to the
distribution panel, and feeders between the
distribution panel and outlets.
CAT
The circuit from the service drop to the service
Ⅳ
entrance, then to the power meter and to the
primary overcurrent protection device.
Higher-numbered categories correspond to electrical environments
with greater momentary energy. So a measurement device
designed for CAT III environments can endure greater momentary
energythanadevicedesignedforCATII.
Using a measurement product in an environment designated with
a higher-numbered category than that for which the product is
rated could result in a severe accident, and must be carefully
avoided. Never use a CAT I measuring product in CAT II, III, or
IV environments. The measurement categories comply with the
Overvoltage Categories of the IEC60664 Standards.
Follow these precautions to ensure safe operation and to
obtain the full benefits of the various functions.
The maximum input voltage is 45 V DC. Attempting
v
WARNING
to measure voltage in excess of the maximum input
could destroy the product and result in personal
injure or death.
To avoid electric shock, do not allow the product to
get wet, and do not use it when your hands are wet.
CAUTION
Before using the product the first time, verify that it
operates normally to ensure that the no damage occurred
during storage or shipping. If you find any damage, contact
your dealer or Hioki representative.
Before using the product, make sure that the insulation on
the cables is undamaged and that no bare conductors are
improperly exposed. Using the product in such conditions
could cause an electric shock. Replace the test leads and
probes with the specified Hioki Model 9629 or 9632.
To avoid damaging the product, do not drop it or subject it
to intense shock.
This product is not designed to be entirely water- or dustproof. To avoid damage, do not use it in a wet or dusty
environment.
Do not store or use the product where it could be exposed
to direct sunlight, high temperature or humidity, or
condensation. Under such conditions, the product may be
damaged and insulation may deteriorate so that it no longer
meets specifications.
If the protective functions of the product are damaged,
either remove it from service or mark it clearly so that
others do not use it inadvertently.
Adjustments and repairs should be made only by
technically qualified personnel.
Use only the specified connection cable. Using a
non-specified cable may result in incorrect
measurements due to poor connection or other
reasons.
The service life of the batteries used in the 3639-20 varies
depending on the recording interval, power save function
setting, measurement-value display during recording,
temperature and humidity, and the like. The table below
gives the approximate battery service lives with an average
pulse count of 5 or less per second, and at 23
To avoid electric shock when replacing the batteries,
WARNING
first disconnect the connection cable from the object to
be measured.
During battery replacement, use caution not to put any
foreign materials such as a metal object into the
product to avoid damage to the product.
Before using the product after replacing the batteries,
replace the cover and screw.
Do not mix old and new batteries, or different types of
batteries. Also, be careful to observe battery polarity
during installation. Otherwise, poor performance or
damage from battery leakage could result.
Handle and dispose of batteries in accordance with
local regulations.
CAUTION
When exchanging the batteries, the circuit may sometimes
short circuit due to static electricity. As far as possible, do not
touch the base board with bare hands.
The installation of new batteries ensures more than six
months of recording (with the Power Save function turned
ON, the recording interval set to 10 seconds or longer, novoltage contact input with an average pulse count of 5 or
less per second, and at 23
The indicator of the remaining battery
power disappears successively from the
right as the remaining power expires.
Indicates the time remaining until battery
replacement
Recording cannot be started.
If the indicator looks like this during
measurement, the measurement will be
interrupted.
Replacing the Battery
(1) Remove back cover screw to remove cover. Verify
polarity and install four new LR03 alkaline batteries.
If no recording session is under way, the screen display is
automatically turned off approximately 15 seconds after the
last key entry (Sleep). However, the recording-interval setup
screen will not enter the Sleep state, even in the absence of
key operation.
During recording
Power Save function set to ON
The Power Save function mode is set to ON
Recording-Interval
SetupScreen
in the initial
settings. If the Power Save function is disabled, the battery
service life will be approximately 15 days. For the setting
procedure for the Power Save function, refer to the "391120 COMMUNICATION BASE Instruction Manual" and the
Measurement Guide.
When the Power Save function is set to ON (enabled)
before measurement starts, the measurement-value display
turns off 15 seconds after measurement starts, and the unit
automatically enters the Power Save function. To turn on
the measurement-value display again, press any key. The
measurement-value display turns on, then turns off after 15
seconds, and the unit automatically enters the Power Save
function again.
When the Power Save function is set to OFF (disabled)
before measurement starts, pressing the INTERVAL button
turns the Power Save function ON/OFF. When the
measurement-value display is on, the Power Save function
is set to OFF. When the measurement-value display is off,
the Power Save function is set to ON. For an extended
measurement operation, set the Power Save function to ON
(measurement-value display turned off).
When replacing 3639-20 PULSE LOGGER batteries or
using 3639-20 stand-alone (with manual operation) for the
first time, connect with 3911-20 and set current time.
See how to set current time in 3911-20
COMMUNICATION BASE instruction manual.
to measure voltage in excess of the maximum input
could destroy the product and result in personal
injure or death.
Connecting the CONNECTION CABLE
The 3639-20 PULSE LOGGER is capable of measuring one
channel of pulse signals with the 9629 CONNECTION
CABLE connected. To ensure correct measurement, insert
the CONNECTION CABLE as far as it will go.
Black
9629 CONNECTION CABLE
Red
The 9629 CONNECTION CABLE can be cut to any
desired length for use. If the cable was cut, use it after
stripping the covering with a wire stripper or another
appropriate tool.
To input a voltage to the 3639-20, connect the red
conductor to the positive (+) side of the pulse-generating
device, and connect the black conductor to the negative (-)
side. If the conductors are inversely connected,
measurement cannot be conducted.
For connection between the 9629
CONNECTION CABLE and the pulse-generating
device, refer to the operation manual of the
pulse-generating device.
Connecting 9632 CONNECTION CABLE to alarm output
terminal. When connecting connection cable, securely insert
connection cable to unit as designated by triangle mark
on connection terminal.
Improper connection results in failure to output accurate
signal.
Logger stand-alone manual settings and settings in
combination with 3911-20 COMMUNICATION BASE
with measurement conditions stored in memory loaded from
personal computer.
3639-20
1. Start recordingValidValidValid
2. Stop recordingValid------
3. Interval settingValidValidValid
4. Current time setting---ValidValid
5. Start control---ValidValid
6. Recording method setting---ValidValid
7. Comments------Valid
8. Filter Setting------Valid
9. Alarm Setting------Valid
3639-20
+
3911-20
3639-20
+
3911-20
+
PC
・
The settings Comments, Filter Setting, and
Alarm Setting are available only when the
three products of the PC, the 3911-20, and
the 3639-20 are interconnected.
Start manual recording by pressing logger REC/STOP
button for 1 second or initiate by prescheduled start set
using 3911-20 COMMUNICATION BASE.
When time scheduled start is engaged,
appears in
display.
If the INTERVAL button is pressed while
is displayed,
the measurement-value display can be turned on/off. When
measurement is to be conducted for an extended period, set
the measurement-value display to OFF in advance.
・
Recording cannot be started if the battery
power is low (indication). Moreover, if
the battery power becomes too low during
recording (indication), recording will be
interrupted.
2. Stop recording
To stop recording, turn on the measurement-value display,
then hold down the REC/STOP button on the 3639-20 for 1
second.
Or recording stops automatically when data is full when set
to recording method: one time.
3. Interval setting
Set interval with logger alone or using 3911-20.
(1/2/5/10/15/20/30 s, 1/2/5/10/15/20/30/60 min, 1 day)
Default setting is 1 sec.
4. Current time setting
To set current time, see 3911-20 COMMUNICATION
BASE instruction manual.
Set specific recording date and time using 3911-20 to
engage time scheduled start. When time scheduled start is
engaged,
appears in display.
When the Power Saving function is set to OFF before
measurement starts, pressing the INTERVAL button turns
the Power Saving function ON/OFF. When the
measurement-value display is on, the Power Saving
function is set to OFF. When the measurement-value
display is off, the Power Saving function is set to ON. For
an extended measurement operation, set the Power Saving
function to ON (measurement-value display turned off).
6. Recording method setting
Set recording method using 3911-20.
Choose either one time or endless recording method.
Default setting is one time
.
One time: Ends recording when data reaches 32000.
Endless : Overwrites previously recorded data when data
exceeds 32000.
7. Comments
The settings Comments is available only when the three
products of the PC, the 3911-20, and the 3639-20 are
interconnected.
Set comments entered by personal computer to logger using
3911-20. When sorting collected recording data, comments
are helpful.
The filter setting can be made when the 3639-20, the 391120, and the PC are interconnected. Either "Mechanicalcontact input" or "Electronic-contact input" can be selected
for the filter. Default setting is Mechanical-contact input.
In the case of the mechanical contacts of a relay or switch,
noise referred to as "chattering" may be heard when signals
are switched over. When the filter is set to "Mechanicalcontact input," noise components such as those of the
mechanical-contact chattering can be eliminated (see the
figure). If "Electronic-contact input" is selected, the
chattering cannot be eliminated. In the case illustrated
below, the measurement value will be 1 if "Mechanicalcontact input" is selected, or 5 if "Electronic-contact input"
is selected.
Chattering
Filter set to ON
9. Alarm setting
The Alarm Setting can be made when the 3639-20, the
3911-20, and the PC are interconnected. Default setting is
the alarm output set to OFF
.
The product issues an alarm signal if a measurement value
exceeds the preset upper-limit value. Decision of an alarm
and its output are made at each individual recording
interval. An alarm once issued will continue to be output
until the measurement value has fallen below the preset
upper-limit value (see the figure below). "AL" remains
displayed on the screen while the alarm signal is being
output.
open-drain output, the signaling side and the
GND side are internally short-circuited when
signals are output. In other cases, they
remain open. When an external power supply
is provided, a relay or sequencer can be
controlled through the use of alarm signals.
Maximum rating: 30 VDC, 200 mADC
(permissible loss: 200 mW).
If an alarm is issued immediately before a measurement
session is terminated, the alarm will continue to be output
even after termination of the measurement. The display of
"AL" and the alarm output will be turned off at the start of
3639-20 PULSE LOGGER stand-alone manual operation
settings are shown below.
(1) Interval setting
Press INTERVAL button to switch measurement value
display to interval setting display. (INTVL appears.)
Press SELECT button to designate interval.
Press INTERVAL button to complete setting.
(2) Starting and ending recording
Press REC/STOP button for 1 second to clear last
recorded data and start recording. (REC appears.)
Press REC/STOP button for 1 second to stop recording.
When memory is full, recording automatically stops
when recording method: one time is selected.
Recording cannot be started if the battery power is low
(
indication). Moreover, if the battery power
becomes too low during recording (
recording will be interrupted.
When the measurement value display is on, the
displayed values are updated at every recording interval.
At the outset of a measurement session, 0 will be
displayed.
When the Power Saving function is set to OFF before
measurement starts, pressing the INTERVAL button
turns the Power Saving function ON/OFF. When the
measurement-value display is on, the Power Saving
function is set to OFF. When the measurement-value
display is off, the Power Saving function is set to ON.
For an extended measurement operation, set the Power
Saving function to ON (measurement-value display
turned off).
3.3Setting by 3911-20
Preparation
(1) If the LCD of this unit remains off (Sleep state), briefly
press the INTERVAL button or SELECT button of the
3639-20 to display the LCD.
When logger LCD shows REC mark or
REC/STOP button for more than 1 second to stop
recording.
During sleeping, recording, or waiting time before
recording start time, data transfer cannot be established
with 3911-20.
(2) Press logger INTERVAL button to display interval
(1) Connect 3911-20 with logger.
(2) Press 3911-20 SEND button for more than 1 second to
send data settings to logger.
If an error is issued and the data cannot be transferred,
refer to Chapter 5. Maintenance and Service:
"Troubleshooting" in the 3911-20 Instruction Manual.
・Previously recorded logger data is erased
when recording is resumed. Be sure to load
data to be saved to 3911-20 or to personal
computer before recording.
・At any other time even when interval setting
display is not shown, except during recording
and waiting for recording, communication with
3911-20 is available. However communication
is disabled when logger is set to sleep.
3639-20 settings in application software
COMMUNICATION UTILITY for 3911-20 packaged with
3911-20 are as follows.
Go to 'Communication' on the menu bar in
COMMUNICATION UTILITY for 3911-20 and select
'Setting measurement condition'. When measurement
condition setting window is open, select '3639-20 setting
items' to set settings.
The following shows the procedure for connecting the
3639-20 PULSE LOGGER to a rain gauge for the
measurement of precipitation.
(1) If the product is being used for the first time or the
batteries have just been replaced, set the current time.
(2) Connect the 3639-20, 3911-20, and PC, and set the
filter to "For mechanical contacts" (default setting)
using the application software.
To output an alarm signal, set Alarm Output to ON and
specify an upper-limit value (default setting: Alarm
Output OFF).
(3) Make other settings (recording interval, etc.) as
necessary.
(4) Connect the rain gauge to the 3639-20. The 9629
CONNECTION CABLE can be cut to any desired
length. If the cable is cut, remove the sheath using a
wire stripper prior to use. Regarding connection of the
rain gauge and the 3639-20, refer to the operation
manual for the rain gauge.
(5) Start measurement.
The 3639-20 is not waterproof. Therefore, place the unit
inside the rain gauge or at a location where it is
protected from the rain.
(6) Upon completion of measurement, use the 3911-20 to
transfer the recorded data to the PC (see the "3911-20
Instruction Manual").
(7) Rain gauges are generally classified into two types:
those with 0.5 mm per pulse and those with 1 mm per
pulse. If a 1-mm-type rain gauge is used, the acquired
value represents the precipitation (mm). If a 0.5-mmtype rain gauge is employed, the precipitation (mm) is
obtained by dividing the acquired value by 2 (scaling).
Scaling is performed by the software supplied with the
3911-20.
3.4.2 Connecting the Product to an OpenCollector (Open-Drain) Output Device
Some flowmeters and wattmeters feature open-drain or
open-collector pulse output. In such a case, connect the
product as shown in the diagram. Insert a resistor between
the external power supply and open-collector output. Make
sure the resistor can turn on (allow a current to flow
through) the transistor used in the open collector. For an
open-drain pulse output, follow the same procedure.
Measurement
accuracy
Clock accuracyWithin 100 ppm (Relative to an error of100 µs
Display of
measurement
range
FilterFor mechanical contacts: Pulse width of 20 ms or
No-voltage contact input (switches, relays, etc.)
Counts when the terminals change from shorted
(short-circuited) to open
Voltage input H: +1.5 V to +45 V; L: +0.0 V to
+0.2 V (logic undefined for the range from +0.2 V
to +1.5 V)
Counts when the voltage level changes to H
Within1 dgt., Guaranteed accuracy period: 1 year
per second and 23
9999 counts/interval
* "OFF" displayed for 10,000 counts or more
(internal data also OFF)
more, pulse spacing of 40 ms or more (frequency of
25 Hz or less)
For electronic contacts: Frequency of 10 kHz or less
updated at the preset recording intervals), recording
status (recording/waiting for preset time), recording
interval (if preset), state of batteries (remainingbattery-power indicator: 4 steps), reservation start
mark, alarm setting
32,000 pieces of data
1/2/5/10/15/20/30 sec., 1/2/5/10/15/20/30/60 min., 1
day
One time: Stopped when the memory becomes full
Continuous: The oldest data is overwritten when the
memory becomes full.
Current time, recording interval, start of recording,
comments, switching of input filter, alarm output
on/off, alarm upper-limit value
200 mW of permissible loss)
If a measurement value exceeds the preset upper-limit
value, an alarm is decided at every interval and issued.
4.3Other Specifications
BackupData backup provided (data not lost due to dead
InterfaceOptical communication using an infrared light.
Power supplyLR03 alkaline battery X 4 (1.5 VDC X 4)
Battery lifeAbout six months or longer (with the power save
Maximum rated
power
DimensionsApprox. 57.5W X 86.5H X 30.0D mm (excluding
MassApprox. 130 g (4.6 oz) (including batteries)
Location for useIndoors, altitude up to 2000 m (6562 feet)
Operate
temperature and
humidity range
Storage
temperature and
humidity range
Guaranteed
accuracy period
Accessories9629 CONNECTION CABLE
Options3911-20 COMMUNICATION BASE
Standards
Applying
batteries or removal of batteries for replacement)
Optional 3911-20 COMMUNICATION BASE required.
function set to ON and the recording interval set to
10 s, at 23
of 5 or less per second)
0.1 VA
projections)/ 2.26"W X 3.41"H X 1.18"D
0to50 , 80% RH or less (no condensation)
(32 to 122
-10 to 60 , 80% RH or less (no condensation)
(14 to 140
1 year
9632 CONNECTION CABLE
LR03 alkaline battery X 4
Instruction Manual
EMC EN 61326
Safety EN 61010 Measurement category l
(anticipated transient overvoltage 330 V),
Pollution Degree 2
To clean the product, wipe it gently with a soft cloth
moistened with water or mild detergent. Never use solvents
such as benzene, alcohol, acetone, ether, ketones, thinners
or gasoline, as they can deform and discolor the case.
Wipe LCD lightly with the soft cloth which dried.
Service
If the product seems to be malfunctioning, confirm that the
batteries are not discharged, and that the connection cables
is not open circuited before contacting your dealer or Hioki
representative. Pack the product carefully so that it will not
be damaged during shipment, and include a detailed written
description of the problem. Hioki cannot be responsible for
damage that occurs during shipment.
Error Messages
The following error may be displayed on the LCD of the
main instrument as shown below.
Error messageMeaning
ROM error
RAM error
Adjustment data error
When this occurs, repair or check the device. Contact your
dealer or Hioki representative.
All inquiries to International Sales and Marketing Department
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL: +81-268-28-0562 / FAX: +81-268-28-0568
E-mail: os-com@hioki.co.jp
URL http://www.hioki.com/
Printed in Japan 3639A981-04
All reasonable care has been taken in the production of this manual,
but if you find any points which are unclear or in error, please
contact your supplier or the International Sales and Marketing
Department at HIOKI headquarters.
In the interests of product development, the contents of this manual
are subject to revision without prior notice.
The content of this manual is protected by copyright. No
reproduction, duplication or modification of the content is permitted
without the authorization of Hioki E.E. Corporation.
Page 40
HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp / URL http://www.hioki.com/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
3639A981-04 08-05H
Printed on recycled paper
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.