Hioki 3639-20 Instruction Manual

Page 1
INSTRUCTION MANUAL
For...は専用機種。複数の場合は「/」で区切る。
不要の場合はとる。
形名を入力。 複数の場合は「/」で区切る。
3639-20
品名を入力。
PULSE LOGGER
Page 2
Page 3
Contents
Chapter 1 Product Outline 1
1.1 Name and Functions of Parts 2
1.2 Interval and Maximum Recording Time
1.3 Battery Life
Chapter 2 Set Up 5
2.1 Replacing the Battery 5
2.2 Power Save Function
2.3 Setting Current Time
2.4 Connecting Sensor Cable
Chapter 3 Settings 11
3
4
7
8
9
3.1 Setting Items 11
3.2 Manual Setting
3.3 Setting by 3911-20
3.4 Application Example
3.4.1 Connecting the Product to a Rain Gauge 20
3.4.2 Connecting the Product to an Open-Collector (Open-Drain) Output Device
16
17
20
22
Chapter 4 Specifications 23
4.1 Measurement Specifications 23
4.2 Functional Specifications
4.3 Other Specifications
23
24
Page 4
Chapter 5 Maintenance and Service 25
Page 5
_____________________________________________________________________
Introduction
Thank you for purchasing the HIOKI 3639-20 PULSE LOGGER. To obtain maximum performance from the product, please read this manual first, and keep it handy for future reference.
Inspection
When you receive the product, inspect it carefully to ensure that no damage occurred during shipping. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Accessories
9629 CONNECTION CABLE (for measurement) 9632 CONNECTION CABLE (for alarm output) Instruction Manual LR03 alkaline battery X 4 (built into this product, for monitor)
i
Testing monitor batteries installed in the unit may possibly be weak. Replace batteries before extended measurement usage.
______________________________________________________________
Inspection
Page 6
ii
R
_____________________________________________________________________
SafetyNotes
This product is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for
DANGE
Safety Symbols
safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the product. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from product defects.
This manual contains information and warnings essential for safe operation of the product and for maintaining it in safe operating condition. Before using the product, be sure to carefully read the following safety notes.
The symbol printed on the product indicates that the user should refer to a corresponding topic in the manual (marked with the relevant function. In the manual, the symbol indicates particularly important information that the user should read before using the product.
symbol) before using the
Indicates DC (Direct Current).
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents a
WARNING
CAUTION
______________________________________________________________
Safety Notes
significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the product.
Page 7
_____________________________________________________________________
iii
Accuracy
We define measurement tolerances in terms of dgt. (digit) values, with the following meanings:
dgt.
(resolution) The smallest displayable unit on a digital measuring product, i.e., the input value that causes the digital display to show a "1".
______________________________________________________________
Safety Notes
Page 8
iv
_____________________________________________________________________
Measurement categories (Overvoltage categories)
This product complies with CAT I safety requirements. To ensure safe operation of measurement products, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT I to CAT IV, and called measurement categories. These are defined as follows.
CAT
Secondary electrical circuits that are connected
to a wall outlet through a transformer or similar device.
CAT
Primary electrical circuits in equipment
connected to a wall outlet via a power cord (portable tools, household appliances, etc.)
CAT
Primary electrical circuits of heavy equipment
(fixed installations) connected directly to the distribution panel, and feeders between the distribution panel and outlets.
CAT
The circuit from the service drop to the service
entrance, then to the power meter and to the primary overcurrent protection device.
Higher-numbered categories correspond to electrical environments with greater momentary energy. So a measurement device designed for CAT III environments can endure greater momentary energythanadevicedesignedforCATII. Using a measurement product in an environment designated with a higher-numbered category than that for which the product is rated could result in a severe accident, and must be carefully avoided. Never use a CAT I measuring product in CAT II, III, or IV environments. The measurement categories comply with the Overvoltage Categories of the IEC60664 Standards.
______________________________________________________________
Safety Notes
Page 9
_____________________________________________________________________
Notes on Use
Follow these precautions to ensure safe operation and to
obtain the full benefits of the various functions.
The maximum input voltage is 45 V DC. Attempting
v
WARNING
to measure voltage in excess of the maximum input could destroy the product and result in personal injure or death. To avoid electric shock, do not allow the product to get wet, and do not use it when your hands are wet.
CAUTION
Before using the product the first time, verify that it operates normally to ensure that the no damage occurred
during storage or shipping. If you find any damage, contact your dealer or Hioki representative. Before using the product, make sure that the insulation on the cables is undamaged and that no bare conductors are improperly exposed. Using the product in such conditions could cause an electric shock. Replace the test leads and probes with the specified Hioki Model 9629 or 9632. To avoid damaging the product, do not drop it or subject it to intense shock. This product is not designed to be entirely water- or dust­proof. To avoid damage, do not use it in a wet or dusty environment. Do not store or use the product where it could be exposed to direct sunlight, high temperature or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate so that it no longer meets specifications. If the protective functions of the product are damaged, either remove it from service or mark it clearly so that others do not use it inadvertently. Adjustments and repairs should be made only by technically qualified personnel.
Use only the specified connection cable. Using a non-specified cable may result in incorrect measurements due to poor connection or other reasons.
______________________________________________________________
Notes on Use
Page 10
vi
_____________________________________________________________________
Measurement Flow Chart and Reference Guide
Is there sufficient
battery charge remaining
to operate 3639-20?
Setting current time
Connecting sensor
Setting 3639-20 stand-alone
Interval
See 3.2.
Press REC/STOP button to start measurement.
See 3.2.
Make the ON/OFF setting of the Power Save function. (When the Power Save function is OFF before measurement starts)
When recording data reaches 32000, measurement stops.
Replace batteries
Set current time when using 3639-20 for the first time or replacingbatteries. See 2.3.
See 2.4.
Setting 3639-20 with 3911-20
IntervalCurrent timeStart controlRecording method
Quick start
Starts measurement immediately.
One time
Make the ON/OFF setting of the Power Save function. (When the Power Save function is OFF before measurement starts)
When recording data exceeds 32000, it overwrites previously recorded data.
Press REC/STOP button to stop measurement.
Setting 3639-20 with 3911-20 and PC
Interval Current timeStart control CommentRecording methodAlarm upper-limit valueAlarm output on/offFilter setting
No
recordingmethod
Settingstart control
Start control options
Starts measurement at prescheduled time.
Setting
Endless
Yes
See 2.1.
Prescheduled start
After measurement is complete, disconnect sensor cables. Use 3911-20 to send measurement data to personal computer. (See 3911-20 instruction manual.)
______________________________________________________________
Measurement Flow Chart and Reference Guide
Page 11
_____________________________________________________________________
1
Chapter 1
Product Outline
The 3639-20 PULSE LOGGER counts and records the
pulses of pulse signals (mechanical contact inputs of
switches, relays, etc., or voltage inputs from 1.5 V to 45 V)
at a specified interval. It can be used for a wide variety of
devices that deliver pulse signals, such as rain gauges, flow
meters, and the like. As data is stored in a nonvolatile
memory, it is not lost even if the batteries run out or are
removed for replacement.
3639-20 PULSE LOGGER cannot be set with 3910­20 COMMUNICATION BASE. Use 3911-20 COMMUNICATION BASE to set 3639-20.
______________________________________________________________
Chapter 1Product Outline
Page 12
2
_____________________________________________________________________
1.1 Name and Functions of Parts
1
2
3
5
6
4
1. LCD Displays measurement value and settings.
2. Optical data transfer ports
3. INTERVAL button Calls up interval setting display to set interval.
4. Connection terminal Connects 9629 CONNECTION CABLE.
5. REC/STOP button Pressing more than 1 second initiates or stops
(SELECT) button Interval is selected in interval setting display.
6. Alarm output terminal
Enables optical data transfer to 3911-20.
Make the ON/OFF setting of the Power Save function. (When the Power Save function is OFF before measurement starts)
recording.
Connects 9632 CONNECTION CABLE and outputs alarm signal (open drain output). (When it was set up in alarm output on.
______________________________________________________________
Chapter 1Product Outline
Page 13
_____________________________________________________________________
1.2 Interval and Maximum Recording Time
Interval and maximum recording time are as follows.
Maximum recording time is limited by battery charge condition.
The number of data that can be recorded on one single
product of 3639-20 PULSE LOGGER is 32,000.
INTVL Maximum Recording Time
1s 8h53min20s
2s 1 day 20 h 26 min 40 s
5s 1 day 20 h 26 min 40 s
10 s 3 day 16 h 26 min 40 s
15 s 5 day 13 h 20 min
20 s 7day9h46min40s
30 s 11 day 2 h 40 min
1min 22 day 5 h 20 min
2min 44 day 10 h 40 min
5min 111 day 2 h 40 min
10 min 222 day 5 h 20 min
15 min 333 day 8 h
20 min 444 day 10 h 40 min
30 min 666 day 6 h
60 min 1333 day 8 h
1 day 32000 day
3
______________________________________________________________
Chapter 1Product Outline
Page 14
4
_____________________________________________________________________
1.3 Battery Life
The service life of the batteries used in the 3639-20 varies depending on the recording interval, power save function setting, measurement-value display during recording, temperature and humidity, and the like. The table below gives the approximate battery service lives with an average pulse count of 5 or less per second, and at 23
Power Save Function
Recording Interval
Battery Life
During Recording (when measured at
the no-voltage contacts)
ON OFF
-- 1 sec
15 days or
longer
4months
or longer
10 min.
or more
6months
or longer
5 .
Standby for Recording Start (Power Save Function)
15 days or
longer
______________________________________________________________
Chapter 1Product Outline
Page 15
_____________________________________________________________________
5
Chapter 2
Set Up
2.1 Replacing the Battery
To avoid electric shock when replacing the batteries,
WARNING
first disconnect the connection cable from the object to be measured. During battery replacement, use caution not to put any foreign materials such as a metal object into the product to avoid damage to the product. Before using the product after replacing the batteries, replace the cover and screw. Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result. Handle and dispose of batteries in accordance with local regulations.
CAUTION
When exchanging the batteries, the circuit may sometimes short circuit due to static electricity. As far as possible, do not touch the base board with bare hands.
The installation of new batteries ensures more than six
months of recording (with the Power Save function turned ON, the recording interval set to 10 seconds or longer, no­voltage contact input with an average pulse count of 5 or less per second, and at 23
______________________________________________________________
5 .
Chapter 2Set Up
Page 16
6
_____________________________________________________________________
Display of remaining battery power
The indicator of the remaining battery power disappears successively from the right as the remaining power expires.
Indicates the time remaining until battery replacement Recording cannot be started. If the indicator looks like this during measurement, the measurement will be interrupted.
Replacing the Battery (1) Remove back cover screw to remove cover. Verify
polarity and install four new LR03 alkaline batteries.
(2) Fit cover properly and tighten screw.
______________________________________________________________
Chapter 2Set Up
Page 17
_____________________________________________________________________
(
)
7
2.2 Power Save Function
If no recording session is under way, the screen display is
automatically turned off approximately 15 seconds after the last key entry (Sleep). However, the recording-interval setup screen will not enter the Sleep state, even in the absence of
key operation.
During recording
Power Save function set to ON
The Power Save function mode is set to ON
Recording-Interval
SetupScreen
in the initial settings. If the Power Save function is disabled, the battery service life will be approximately 15 days. For the setting procedure for the Power Save function, refer to the "3911­20 COMMUNICATION BASE Instruction Manual" and the Measurement Guide. When the Power Save function is set to ON (enabled) before measurement starts, the measurement-value display turns off 15 seconds after measurement starts, and the unit automatically enters the Power Save function. To turn on the measurement-value display again, press any key. The measurement-value display turns on, then turns off after 15 seconds, and the unit automatically enters the Power Save function again. When the Power Save function is set to OFF (disabled) before measurement starts, pressing the INTERVAL button turns the Power Save function ON/OFF. When the measurement-value display is on, the Power Save function is set to OFF. When the measurement-value display is off, the Power Save function is set to ON. For an extended measurement operation, set the Power Save function to ON (measurement-value display turned off).
______________________________________________________________
Chapter 2Set Up
Page 18
8
_____________________________________________________________________
2.3 Setting Current Time
When replacing 3639-20 PULSE LOGGER batteries or using 3639-20 stand-alone (with manual operation) for the first time, connect with 3911-20 and set current time.
See how to set current time in 3911-20 COMMUNICATION BASE instruction manual.
______________________________________________________________
Chapter 2Set Up
Page 19
_____________________________________________________________________
9
2.4 Connecting Sensor Cable
The maximum input voltage is 45 V DC. Attempting
WARNING
to measure voltage in excess of the maximum input could destroy the product and result in personal injure or death.
Connecting the CONNECTION CABLE
The 3639-20 PULSE LOGGER is capable of measuring one
channel of pulse signals with the 9629 CONNECTION
CABLE connected. To ensure correct measurement, insert
the CONNECTION CABLE as far as it will go.
Black
9629 CONNECTION CABLE
Red
The 9629 CONNECTION CABLE can be cut to any
desired length for use. If the cable was cut, use it after stripping the covering with a wire stripper or another appropriate tool.
To input a voltage to the 3639-20, connect the red
conductor to the positive (+) side of the pulse-generating device, and connect the black conductor to the negative (-) side. If the conductors are inversely connected, measurement cannot be conducted.
For connection between the 9629 CONNECTION CABLE and the pulse-generating device, refer to the operation manual of the pulse-generating device.
______________________________________________________________
Chapter 2Set Up
Page 20
10
E
_____________________________________________________________________
Connecting 9632 CONNECTION CABLE
Connecting 9632 CONNECTION CABLE to alarm output terminal. When connecting connection cable, securely insert connection cable to unit as designated by triangle mark on connection terminal. Improper connection results in failure to output accurate signal.
9632 CONNECTION CABL
9632 CONNECTION CABLE
GND
non connection
Signal
Cord length: 1000 mm
White: Signal Black: GND
______________________________________________________________
Chapter 2Set Up
Page 21
_____________________________________________________________________
11
Chapter 3
Settings
3.1 Setting Items
Logger stand-alone manual settings and settings in
combination with 3911-20 COMMUNICATION BASE with measurement conditions stored in memory loaded from personal computer.
3639-20
1. Start recording Valid Valid Valid
2. Stop recording Valid --- ---
3. Interval setting Valid Valid Valid
4. Current time setting --- Valid Valid
5. Start control --- Valid Valid
6. Recording method setting --- Valid Valid
7. Comments --- --- Valid
8. Filter Setting --- --- Valid
9. Alarm Setting --- --- Valid
3639-20
3911-20
3639-20
3911-20
PC
The settings Comments, Filter Setting, and Alarm Setting are available only when the three products of the PC, the 3911-20, and the 3639-20 are interconnected.
______________________________________________________________
Chapter 3Settings
Page 22
12
_____________________________________________________________________
1. Start recording
Start manual recording by pressing logger REC/STOP button for 1 second or initiate by prescheduled start set using 3911-20 COMMUNICATION BASE. When time scheduled start is engaged,
appears in
display.
If the INTERVAL button is pressed while
is displayed, the measurement-value display can be turned on/off. When measurement is to be conducted for an extended period, set the measurement-value display to OFF in advance.
Recording cannot be started if the battery power is low ( indication). Moreover, if the battery power becomes too low during recording ( indication), recording will be interrupted.
2. Stop recording
To stop recording, turn on the measurement-value display, then hold down the REC/STOP button on the 3639-20 for 1 second. Or recording stops automatically when data is full when set to recording method: one time.
3. Interval setting
Set interval with logger alone or using 3911-20. (1/2/5/10/15/20/30 s, 1/2/5/10/15/20/30/60 min, 1 day) Default setting is 1 sec.
4. Current time setting
To set current time, see 3911-20 COMMUNICATION BASE instruction manual.
______________________________________________________________
Chapter 3Settings
Page 23
_____________________________________________________________________
13
5. Start control
Set specific recording date and time using 3911-20 to engage time scheduled start. When time scheduled start is engaged,
appears in display. When the Power Saving function is set to OFF before measurement starts, pressing the INTERVAL button turns the Power Saving function ON/OFF. When the measurement-value display is on, the Power Saving function is set to OFF. When the measurement-value display is off, the Power Saving function is set to ON. For an extended measurement operation, set the Power Saving function to ON (measurement-value display turned off).
6. Recording method setting
Set recording method using 3911-20. Choose either one time or endless recording method. Default setting is one time
. One time: Ends recording when data reaches 32000. Endless : Overwrites previously recorded data when data
exceeds 32000.
7. Comments
The settings Comments is available only when the three
products of the PC, the 3911-20, and the 3639-20 are interconnected. Set comments entered by personal computer to logger using 3911-20. When sorting collected recording data, comments are helpful.
______________________________________________________________
Chapter 3Settings
Page 24
14
_____________________________________________________________________
8. Filter setting
The filter setting can be made when the 3639-20, the 3911­20, and the PC are interconnected. Either "Mechanical­contact input" or "Electronic-contact input" can be selected for the filter. Default setting is Mechanical-contact input. In the case of the mechanical contacts of a relay or switch, noise referred to as "chattering" may be heard when signals are switched over. When the filter is set to "Mechanical­contact input," noise components such as those of the mechanical-contact chattering can be eliminated (see the figure). If "Electronic-contact input" is selected, the chattering cannot be eliminated. In the case illustrated below, the measurement value will be 1 if "Mechanical­contact input" is selected, or 5 if "Electronic-contact input" is selected.
Chattering
Filter set to ON
9. Alarm setting
The Alarm Setting can be made when the 3639-20, the 3911-20, and the PC are interconnected. Default setting is the alarm output set to OFF
. The product issues an alarm signal if a measurement value exceeds the preset upper-limit value. Decision of an alarm and its output are made at each individual recording interval. An alarm once issued will continue to be output until the measurement value has fallen below the preset upper-limit value (see the figure below). "AL" remains displayed on the screen while the alarm signal is being output.
______________________________________________________________
Chapter 3Settings
Page 25
_____________________________________________________________________
Alarm output
Measurement value
Upper-limit value
Time
Recording intervals
As the alarm-signal output terminals form an
open-drain output, the signaling side and the GND side are internally short-circuited when signals are output. In other cases, they remain open. When an external power supply is provided, a relay or sequencer can be controlled through the use of alarm signals. Maximum rating: 30 VDC, 200 mADC (permissible loss: 200 mW).
If an alarm is issued immediately before a measurement
session is terminated, the alarm will continue to be output even after termination of the measurement. The display of
"AL" and the alarm output will be turned off at the start of
the next recording session.
15
______________________________________________________________
Chapter 3Settings
Page 26
16
_____________________________________________________________________
3.2 Manual Setting
3639-20 PULSE LOGGER stand-alone manual operation settings are shown below.
(1) Interval setting
Press INTERVAL button to switch measurement value display to interval setting display. (INTVL appears.) Press SELECT button to designate interval. Press INTERVAL button to complete setting.
(2) Starting and ending recording
Press REC/STOP button for 1 second to clear last recorded data and start recording. (REC appears.) Press REC/STOP button for 1 second to stop recording. When memory is full, recording automatically stops when recording method: one time is selected. Recording cannot be started if the battery power is low (
indication). Moreover, if the battery power becomes too low during recording ( recording will be interrupted. When the measurement value display is on, the displayed values are updated at every recording interval. At the outset of a measurement session, 0 will be displayed.
indication),
______________________________________________________________
Chapter 3Settings
Page 27
_____________________________________________________________________
17
(3) Power save function ON/OFF
When the Power Saving function is set to OFF before measurement starts, pressing the INTERVAL button turns the Power Saving function ON/OFF. When the measurement-value display is on, the Power Saving function is set to OFF. When the measurement-value display is off, the Power Saving function is set to ON. For an extended measurement operation, set the Power Saving function to ON (measurement-value display turned off).
3.3 Setting by 3911-20
Preparation
(1) If the LCD of this unit remains off (Sleep state), briefly
press the INTERVAL button or SELECT button of the 3639-20 to display the LCD. When logger LCD shows REC mark or REC/STOP button for more than 1 second to stop recording. During sleeping, recording, or waiting time before recording start time, data transfer cannot be established with 3911-20.
(2) Press logger INTERVAL button to display interval
setting display. (INTVL appears.)
, press
______________________________________________________________
Chapter 3Settings
Page 28
18
_____________________________________________________________________
Connection and setting
(1) Connect 3911-20 with logger. (2) Press 3911-20 SEND button for more than 1 second to
send data settings to logger. If an error is issued and the data cannot be transferred, refer to Chapter 5. Maintenance and Service: "Troubleshooting" in the 3911-20 Instruction Manual.
Previously recorded logger data is erased
when recording is resumed. Be sure to load data to be saved to 3911-20 or to personal computer before recording.
At any other time even when interval setting
display is not shown, except during recording and waiting for recording, communication with 3911-20 is available. However communication is disabled when logger is set to sleep.
3639-20 settings in application software COMMUNICATION UTILITY for 3911-20 packaged with 3911-20 are as follows.
Go to 'Communication' on the menu bar in COMMUNICATION UTILITY for 3911-20 and select 'Setting measurement condition'. When measurement condition setting window is open, select '3639-20 setting items' to set settings.
______________________________________________________________
Chapter 3Settings
Page 29
_____________________________________________________________________
19
The settings of Comments, Filter Setting, and
Alarm Setting can be made only on the 3639-20 and are available only when the 3639-20, the 3911-20, and the PC are interconnected.
The settings common to all models are made in
"Setting Items of the 3911-20." In this case, the settings can be made with the 3639-20 and the 3911-20 interconnected.
______________________________________________________________
Chapter 3Settings
Page 30
20
_____________________________________________________________________
3.4 Application Example
3.4.1 Connecting the Product to a Rain Gauge
The following shows the procedure for connecting the 3639-20 PULSE LOGGER to a rain gauge for the measurement of precipitation.
(1) If the product is being used for the first time or the
batteries have just been replaced, set the current time.
(2) Connect the 3639-20, 3911-20, and PC, and set the
filter to "For mechanical contacts" (default setting) using the application software. To output an alarm signal, set Alarm Output to ON and specify an upper-limit value (default setting: Alarm Output OFF).
(3) Make other settings (recording interval, etc.) as
necessary.
(4) Connect the rain gauge to the 3639-20. The 9629
CONNECTION CABLE can be cut to any desired length. If the cable is cut, remove the sheath using a wire stripper prior to use. Regarding connection of the rain gauge and the 3639-20, refer to the operation manual for the rain gauge.
(5) Start measurement.
The 3639-20 is not waterproof. Therefore, place the unit inside the rain gauge or at a location where it is protected from the rain.
(6) Upon completion of measurement, use the 3911-20 to
transfer the recorded data to the PC (see the "3911-20 Instruction Manual").
______________________________________________________________
Chapter 3Settings
Page 31
_____________________________________________________________________
21
(7) Rain gauges are generally classified into two types:
those with 0.5 mm per pulse and those with 1 mm per pulse. If a 1-mm-type rain gauge is used, the acquired value represents the precipitation (mm). If a 0.5-mm­type rain gauge is employed, the precipitation (mm) is obtained by dividing the acquired value by 2 (scaling). Scaling is performed by the software supplied with the 3911-20.
Connect the tips
3639-20
Tipping-bucket rain gauge
______________________________________________________________
Chapter 3Settings
Page 32
22
_____________________________________________________________________
3.4.2 Connecting the Product to an Open­Collector (Open-Drain) Output Device
Some flowmeters and wattmeters feature open-drain or open-collector pulse output. In such a case, connect the product as shown in the diagram. Insert a resistor between the external power supply and open-collector output. Make sure the resistor can turn on (allow a current to flow through) the transistor used in the open collector. For an open-drain pulse output, follow the same procedure.
______________________________________________________________
Chapter 3Settings
Page 33
_____________________________________________________________________
Chapter 4
Specifications
4.1 Measurement Specifications
23
Input Pulse input: 1 channel
Measurement accuracy Clock accuracy Within 100 ppm (Relative to an error of 100 µs
Display of measurement range Filter For mechanical contacts: Pulse width of 20 ms or
No-voltage contact input (switches, relays, etc.) Counts when the terminals change from shorted (short-circuited) to open Voltage input H: +1.5 V to +45 V; L: +0.0 V to +0.2 V (logic undefined for the range from +0.2 V to +1.5 V) Counts when the voltage level changes to H
Within 1 dgt., Guaranteed accuracy period: 1 year
per second and 23 9999 counts/interval * "OFF" displayed for 10,000 counts or more
(internal data also OFF)
more, pulse spacing of 40 ms or more (frequency of 25 Hz or less) For electronic contacts: Frequency of 10 kHz or less
5 )
4.2 Functional Specifications
Display Measurement-value display (monitored values
Recording capacity Start of recording Manual start, reserved-time start End of recording Manual stop, memory full Recording interval Recording method
______________________________________________________________
updated at the preset recording intervals), recording status (recording/waiting for preset time), recording interval (if preset), state of batteries (remaining­battery-power indicator: 4 steps), reservation start mark, alarm setting 32,000 pieces of data
1/2/5/10/15/20/30 sec., 1/2/5/10/15/20/30/60 min., 1 day One time: Stopped when the memory becomes full Continuous: The oldest data is overwritten when the memory becomes full.
Chapter 4Specifications
Page 34
24
_____________________________________________________________________
Settings By key entry on the product:
Alarm output Provided (open-drain output, 30 VDC, 200 mADC,
Waterproofing None
Recording interval, manual start, manual stop, measurement-value display on/off
Using PC software:
Current time, recording interval, start of recording, comments, switching of input filter, alarm output on/off, alarm upper-limit value
200 mW of permissible loss) If a measurement value exceeds the preset upper-limit value, an alarm is decided at every interval and issued.
4.3 Other Specifications
Backup Data backup provided (data not lost due to dead
Interface Optical communication using an infrared light.
Power supply LR03 alkaline battery X 4 (1.5 VDC X 4) Battery life About six months or longer (with the power save
Maximum rated power Dimensions Approx. 57.5W X 86.5H X 30.0D mm (excluding
Mass Approx. 130 g (4.6 oz) (including batteries) Location for use Indoors, altitude up to 2000 m (6562 feet) Operate temperature and humidity range Storage temperature and humidity range Guaranteed accuracy period Accessories 9629 CONNECTION CABLE
Options 3911-20 COMMUNICATION BASE Standards Applying
batteries or removal of batteries for replacement)
Optional 3911-20 COMMUNICATION BASE required.
function set to ON and the recording interval set to 10 s, at 23 of 5 or less per second)
0.1 VA
projections)/ 2.26"W X 3.41"H X 1.18"D
0to50 , 80% RH or less (no condensation) (32 to 122
-10 to 60 , 80% RH or less (no condensation) (14 to 140
1 year
9632 CONNECTION CABLE LR03 alkaline battery X 4 Instruction Manual
EMC EN 61326 Safety EN 61010 Measurement category l (anticipated transient overvoltage 330 V), Pollution Degree 2
5 , and with an average pulse count
)
)
______________________________________________________________
Chapter 4Specifications
Page 35
_____________________________________________________________________
25
Chapter 5
Maintenance and Service
Cleaning
To clean the product, wipe it gently with a soft cloth
moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners
or gasoline, as they can deform and discolor the case.
Wipe LCD lightly with the soft cloth which dried.
Service
If the product seems to be malfunctioning, confirm that the
batteries are not discharged, and that the connection cables is not open circuited before contacting your dealer or Hioki representative. Pack the product carefully so that it will not be damaged during shipment, and include a detailed written description of the problem. Hioki cannot be responsible for damage that occurs during shipment.
Error Messages
The following error may be displayed on the LCD of the
main instrument as shown below.
Error message Meaning
ROM error
RAM error
Adjustment data error
When this occurs, repair or check the device. Contact your dealer or Hioki representative.
______________________________________________________________
Chapter 5Maintenance and Service
Page 36
26
_____________________________________________________________________
______________________________________________________________
Chapter 5Maintenance and Service
Page 37
Page 38
Page 39
HIOKI 3639-20 PULSE LOGGER
Instruction Manual
Publication date: May 2008 Revised edition 4
Edited and published by HIOKI E.E. CORPORATION
Technical Support Section
All inquiries to International Sales and Marketing Department
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL: +81-268-28-0562 / FAX: +81-268-28-0568
E-mail: os-com@hioki.co.jp
URL http://www.hioki.com/
Printed in Japan 3639A981-04
All reasonable care has been taken in the production of this manual, but if you find any points which are unclear or in error, please contact your supplier or the International Sales and Marketing Department at HIOKI headquarters. In the interests of product development, the contents of this manual are subject to revision without prior notice. The content of this manual is protected by copyright. No reproduction, duplication or modification of the content is permitted without the authorization of Hioki E.E. Corporation.
Page 40
HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp / URL http://www.hioki.com/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA TEL +1-609-409-9109 / FAX +1-609-409-9108
3639A981-04 08-05H
Printed on recycled paper
Loading...