Thank you for purchasing the Hioki 3638-20 LEAK
LOGGER. To obtain maximum performance from the
instrument, please read this manual first, and keep it handy
for future reference.
Inspection
When you receive the instrument, inspect it carefully to
ensure that no damage occurred during shipping. If damage
is evident, or if it fails to operate according to the
specifications, contact your dealer or Hioki representative.
Accessories
9632 CONNECTION CABLE
Instruction Manual
LR03 alkaline battery
(built into this unit, for monitor)
4
X
i
Before using the product the first time, verify that it
operates normally to ensure that the no damage occurred
during storage or shipping. If you find any damage, contact
your dealer or Hioki representative.
Testing monitor batteries installed in the unit may possibly be
weak. Replace batteries before extended measurement usage.
Before using the product, make sure that the insulation on the
sensor cables is undamaged and that no bare conductors are
improperly exposed. Using the product in such conditions could
cause an electric shock, so contact your dealer or Hioki
representative for repair.
This instrument is designed to conform to IEC 61010
Safety Standards, and has been thoroughly tested for
DANGE
Safety Symbols
safety prior to shipment. However, mishandling
during use could result in injury or death, as well as
damage to the instrument. Be certain that you
understand the instructions and precautions in the
manual before use. We disclaim any responsibility for
accidents or injuries not resulting directly from
product defects.
This manual contains information and warnings essential
for safe operation of the instrument and for maintaining it
in safe operating condition. Before using the instrument,
be sure to carefully read the following safety notes.
Thesymbol printed on the instrument indicates that
the user should refer to a corresponding topic in the
manual (marked with the
relevant function.
In the manual, thesymbol indicates particularly
important information that the user should read before
using the instrument.
The following symbols in this manual indicate the relative
importance of cautions and warnings.
Indicates that incorrect operation presents an extreme
DANGE
WARNING
CAUTION
NOTE
hazard that could result in serious injury or death to
the user.
Indicates that incorrect operation presents a
significant hazard that could result in serious injury or
death to the user.
Indicates that incorrect operation presents a
possibility of injury to the user or damage to the
instrument.
Advisory items related to performance or correct
operation of the instrument.
Accuracy
The specifications in this manual include figures for
"measurement accuracy" when referring to digital
measuring instruments, and for "measurement tolerance"
when referring to analog instruments.
f.s.
(maximum display or scale value, or length of scale)
Signifies the maximum display (scale) value or the length
of the scale (in cases where the scale consists of unequal
increments or where the maximum value cannot be
defined).
In general, this is the range value (the value written on
the range selector or equivalent) currently in use.
rdg.
(displayed or indicated value)
This signifies the value actually being measured, i.e., the
value that is currently indicated or displayed by the
measuring instrument.
dgt.
(resolution)
Signifies the smallest display unit on a digital measuring
instrument, i.e., the value displayed when the last digit on
the digital display is "1".
To ensure safe operation of measurement product, IEC
61010 establishes safety standards for various electrical
environments, categorized as CAT I to CAT IV, and called
measurement categories. These are defined as follows.
CAT I
CAT II
CAT III
CAT IV
Secondary electrical circuits connected to an AC
electrical outlet through a transformer or similar
device.
Primary electrical circuits in equipment connected
to an AC electrical outlet by a power cord (portable
tools, household appliances, etc.)
Primary electrical circuits of heavy equipment
(fixed installations) connected directly to the
distribution panel, and feeders from the distribution
panel to outlets.
The circuit from the service drop to the service
entrance, and to the power meter and primary
overcurrent protection device (distribution panel).
Higher-numbered categories correspond to electrical
environments with greater momentary energy. So a
measurement device designed for CAT III environments can
endure greater momentary energy than a device designed for
CAT II. Using a measurement product in an environment
designated with a higher-numbered category than that for
which the product is rated could result in a severe accident,
and must be carefully avoided.
Never use a CAT I measuring product in CAT II, III, or IV
environments. The measurement categories comply with the
Overvoltage Categories of the IEC60664 Standards.
Follow these precautions to ensure safe operation and to
obtain the full benefits of the various functions.
This product cannot be used on following voltage lines.
v
DANGE
If the voltage exceeds, there will be a short-circuit
accident or electrocution accident will result.
CAT III, AC300 Vrms for 9657
CAT III, AC150 Vrms for 9658
To avoid short circuits and potentially life-threatening
hazards, never attach the clamp-on leak sensor to a
circuit that operates at more than the following
maximum rated voltage to earth.
AC300 Vrms for 9657
AC150 Vrms for 9658
Clamp-on leak sensor should only be connected to the
secondary side of a breaker, so the breaker can prevent
an accident if a short circuit occurs. Connections
should never be made to the primary side of a breaker,
because unrestricted current flow could cause a
serious accident if a short circuit occurs.
For model 9657
When conductors being measured carry in excess of
the safe voltage level (SELV-E)
and not more than 300
V, to prevent short circuits and electric shock while the
clamp core jaw is open, make sure that conductors to
be measured are insulated with material conforming to
(1) Measurement Category III
, (2) Pollution Degree 2,
and (3) Basic Insulation Requirements for Working
Voltages of 300 V.
When conductors being measured carry in excess of
the safe voltage level (SELV-E)
and not more than 150
V, to prevent short circuits and electric shock while the
clamp core jaw is open, make sure that conductors to
be measured are insulated with material conforming to
(1) Measurement Category III
, (2) Pollution Degree 2,
and (3) Basic Insulation Requirements for Working
Voltages of 150 V.
Refer to the following standards regarding the
meanings of underlined terms.
IEC61010-1
IEC61010-2-031
IEC61010-2-032
To avoid electric shock, do not allow the product to
WARNING
get wet, and do not use it when your hands are wet.
When measuring the ground conductor on a
transformer Class 2 connection site, be careful not to
approach high voltage devices or conductors. Also, if
close to high voltage charging devices or if
measurement is otherwise difficult, first change the
route of the grounding wire.
To avoid damage to the product, protect it from
vibration or shock during transport and handling, and
be especially careful to avoid dropping.
This is a precision instrument: do not clamp any
foreign objects in the end of the clamp core, or insert
anything in the core gap.
Do not exert excessive pressure on the clamp sensor
or attempt to wedge the sensor into a tight spot for
measurement.
To avoid electric shock when measuring live lines, wear
appropriate protective gear, such as insulated rubber
gloves, boots and a safety helmet.
This product is designed for indoor use, and operates
reliably from 0
to 50.
Do not store or use the product where it could be exposed
to direct sunlight, high temperature or humidity, or
condensation. Under such conditions, the product may be
damaged and insulation may deteriorate so that it no longer
meets specifications.
Accurate measurement may be impossible in the presence of
strong magnetic fields, such as near transformers and highcurrent conductors, or in the presence of strong electromagnetic
fields such as near radio transmitters.
Set current time when using 3638-20 for the
first time or replacingbatteries. See 2.3.
See 2.4.
Setting 3638-20 with
COMMUNICATION BASE
・Interval
・Current time
・Start control
・Recording method
Quick start
Starts measurement
immediately.
One time
When recording data exceeds 32000(
overwrites previously recorded data.
Press REC/STOP button to stop measurement.
Replace batteries
COMMUNICATION BASE and PC
・Interval ・Current time
・Start control ・Range
・Recording method
・Number of channels
・Recording mode
・Alarm ・Comment
No
recordingmethod
Settingstart control
Start control options
Starts measurement at
prescheduled time.
Setting
Setting 3638-20 with
Endless
16000 in 2 channels
See 2.1.
Yes
Prescheduled start
), it
After measurement is complete, disconnect sensor cables.
Use COMMUNICATION BASE to send measurement data to personal computer.
(See COMMUNICATION BASE instruction manual.)
3638-20 LEAK LOGGER with clamp on leak sensor
records 1 or 2 channels of data at leakage current value.
It measures and records AC RMS current values of up to 1
Arms at fixed intervals.
With two optional recording modes; recording maximum
value and recording average value, extended duration of
current recording is possible with battery operation.
Data is saved in nonvolatile memory when batteries are
weak or removed for replacement.
Interval and maximum recording time (when power save
function is valid) are as follows.
When 3638-20 LEAK LOGGER is used in 2 channels,
maximum recordable data is 16000 per unit.
INTVLMaximum Recording Time
1s4h26min40s
2s8h53min20s
5s22 h 13 min 20 s
10 s1 day 20 h 26 min 40 s
15 s2 day 18 h 40 min
20 s3 day 16 h 53 min 20 s
30 s5 day 13 h 20 min
1min11 day 2 h 40 min
2min22 day 5 h 20 min
3
5min55 day 13 h 20 min
10 min111 day 2 h 40 min
15 min166 day 16 h
20 min222 day 5 h 20 min
30 min333 day 8 h
60 min666 day 6 h
*
*
*
*
*
*
*: Maximum recording time is limited by battery charge
To calculate effective value, measurement signal is sampled
every 334 µs and operation is carried out from data with
maximum 135 points to seek effective value.
Maximum 135 points partitioned
at com
lete cycle
334µs
Recording
To record measurement value, calculate effective value per
1 second or 0.2 second and record average or maximum
value of all data within interval in memory.
After recording starts, average or maximum value within
interval is recorded, so the initial data is not recorded
immediately after recording starts but is recorded from the
following interval.
Clamp the sensor on the conductor, so that the conductor
passes through the center of the clamp core. For
measurement of grounded leads, clamp the sensor on one
lead only (see a). For overall measurements, clamp the
sensor on the entire circuit path (see b).
NOTE
When the current to be measured is unknown, begin
measurement with the 1000 mA range selected.
Single-phase 3-lead circuits
Transformer
Ig: leak Current
Class 2 grounding line
Three-phase 3-lead
circuits
Ig
a
E
2
Class 2 grounding line
a
Ig
E
Clamp all three leads of the circuit
b
Clamp all three leads
b
2
of the circuit
b
Load
device
b
Load
device
a
E
Class 3 grounding line
3
NOTE
For measurement of single-phase 2-lead circuits, clamp both
leads of the circuit.
For measurement of three-phase 4-lead circuits, clamp all four
leads of the circuit. If this is not possible, the measurement can
also be carried out on the ground lead of the equipment.
If a strong current (on the order of 100 A) is flowing in an
adjacent circuit, accurate measurement may not be possible.
Perform the measurement at a sufficient distance from other
current-carrying conductors.
Do not input a current which exceeds the maximum continuous
input rating.
When the clamp is open and while the range is being changed,
an indicated value in the range of tens may appear; this is not a
malfunction.
The indication goes progressively to zero, but this make take a
little time. Beginning measurement before the display reaches
zero has no effect on measurement.
Checking for insulation faults
Normally, for a type 2 grounding installation of a
transformer, the measurement will first be made to check
for overall circuit leak current in the ground lead (a).
Current changes can be used to diagnose the leak current
condition.
When leak current has been detected, the measurement
should proceed from the power source towards the load,
using overall measurement.
1. If an insulation fault in the wiring has occurred at
position A in the illustration, leak current will be
detected at position b using overall measurement, but not
at position b'.
2. If an insulation fault in the load equipment has occurred
at position B in the illustration, leak current will be
detected at position c using overall measurement, but not
at position c'.
Be sure to clamp only one lead of the conductor. Clamp so
that the conductor is in the center of the sensor aperture.
OKNG
NOTE
The frequency of special waveforms such as at the secondary
side of an inverter can not be indicated correctly.
When the current to be measured is unknown, begin
measurement with the 1000 mA range selected.
Do not input a current which exceeds the maximum continuous
input rating.
To avoid electric shock when replacing the batteries,
WARNING
first disconnect the connection cables from the object
to be measured.
During battery replacement, use caution not to put any
foreign materials such as a metal object into the unit
to avoid damage to the unit.
Before using the product after replacing the batteries,
replace the cover and screw.
Do not mix old and new batteries, or different types of
batteries. Also, be careful to observe battery polarity
during installation. Otherwise, poor performance or
damage from battery leakage could result.
To avoid the possibility of explosion, do not short
circuit, disassemble or incinerate batteries.
To avoid corrosion from battery leakage, remove the
batteries from the product if it is to be stored for a
long time.
When exchanging the batteries, the circuit may sometimes
short circuit due to static electricity. As far as possible, do not
touch the base board with bare hands.
Installing new batteries ensures about one month of
recording (when interval is set at 1 sec) and about 10 days
of recording (when interval is set at 0.2 sec). (reference
value with power save setting set to valid and at 20
Remaining battery power indicator (
)indicates
)
remaining battery life reducing incrementally from right.
Empty battery power indicator (
) indicates time to
replace batteries.
1. Remove back cover screw to remove cover. Verify
polarity and install four new LR03 alkaline batteries.
Display window is automatically turned off in
approximately 15 seconds after last key entry. (Sleep)
However, while recording, REC/AV/// AL mark
shows each conditions.
Sleeping......
Press any button to turn display on to display measurement
value or to set settings.
Note when interval setting display is on, sleep does not
engage with no button press.
Initially, power save function is on. To turn off power save
function, follow the instructions below.
When power save function is off, maximum continuous
duration is approximately 10 days.
1. Connect logger, COMMUNICATION BASE and
personal computer.
2. Start up application software packaged with
COMMUNICATION BASE.
3. Go to Communications on the menu bar and select
Power Save Options. Choose Off to turn off power save
function.
When replacing 3638-20 LEAK LOGGER batteries or
using 3638-20 stand-alone (with manual operation) for the
first time, connect with COMMUNICATION BASE and set
current time.
See how to set current time in COMMUNICATION BASE
instruction manual.
2.4Connecting Sensor Cable
Maximum input current is 1 Arms. Do not exceed the
WARNING
continuous maximum input current.
This product cannot be used on following voltage
lines. If the voltage exceeds, there will be a shortcircuit accident or electrocution accident will result.
CAT III, AC300 Vrms for 9657
CAT III, AC150 Vrms for 9658
When connecting the clamp on leak sensor, if a
metallic part of the sensor exposed while the clamp is
open should touch the two wires of the line, or if the
sensor is used on a bare conductor, a shorting
accident or electric shock may result.
The 3638-20 measurement range is set at 100 mA, do
not exceed 100 mArms.
To avoid damaging the unit, do not use any other
sensors except 9657/9658 CLAMP ON LEAK SENSOR
as sensor connector.
3638-20 LEAK LOGGER enables 2 channels of data
measurement when connected with 9657/9658 CLAMP ON
LEAK SENSOR.
9657/9658 CLAMP ON LEAK SENSOR
13
9658
NOTE
9657
Carefully read 9657/9658 CLAMP ON LEAK SENSOR
instruction manuals before use.
Improper measurement method may result in accidental injury,
death or damage in the unit.
Logger stand-alone manual settings and settings in
combination with COMMUNICATION BASE with
measurement conditions stored in memory loaded from
personal computer.
15
3638-20+
3638-20
1. Start recordingValidValidValid
2. Stop recordingValid------
3. Interval settingValidValidValid
4. Current time setting---ValidValid
5. Start control---ValidValid
6. Recording method setting---ValidValid
7. Measurement interval settingValid---Valid
8. Range settingValid---Valid
9. Setting number of channelsValid---Valid
10. Recording mode settingValid---Valid
11. Alarm setting------Valid
12. Comments------Valid
13. Minimum/Maximum value
display
NOTE
・
When PC, COMMUNICATION BASE and 3638-20 are
Valid------
COMMUNICATION
BASE
3638-20+
COMMUNICATION
BASE+PC
connected, separate range can be set in each channel. Once
settings are set, range setting with manual operation sets
common range.
・
Alarm and comment setting are available when personal
computer is connected to both logger and COMMUNICATION
BASE.
button for 1 second or initiate by prescheduled start set
using COMMUNICATION BASE.
When time scheduled start is engaged, clock icon appears
in display. When batteries are weak, recording does not
start. During recording, weak battery interrupts recording.
2. Stop recording
Stop recording by pressing logger
REC/STOP
button for 1
second. Or recording stops automatically when data is full
when set to recording method: one time.
3. Interval setting
Set interval with logger alone or using COMMUNICATION
BASE.
(1/2/5/10/15/20/30 s, 1/2/5/10/15/20/30/60 min)
4. Current time setting
To set current time, see COMMUNICATION BASE
instruction manual.
5. Start control
Set specific recording date and time using
COMMUNICATION BASE to engage time scheduled start.
When time scheduled start is engaged, clock icon appears
in display.
6. Recording method setting
Set recording method using COMMUNICATION BASE.
Choose either one time or endless recording method.
Default setting is one time.
One time: Ends recording when data reaches 32000 (2 ch:
16000).
Endless : Overwrites previously recorded data when data
You can set the measurement interval on the 3638-20 or
through the COMMUNICATION BASE. Select between an
interval of 0.2 sec (when measuring just 1 channel) or 1
second.
The factory setting is 1 second.
8. Range setting
With 3638-20 stand-alone, or connected with
COMMUNICATION BASE and PC, range setting is
available.
Two measurement range options are 100 mA range and
1000 mA.
Select and set in accordance with measurement current.
When manually operated, common settings apply to channel
1 and channel 2.
When 3638-20 measurement range is set at 100 mA, do not
exceed 100 mArms.
NOTE
When PC, COMMUNICATION BASE and 3638-20 are
connected, separate range can be set in each channel. Once
settings are set, range setting with manual operation sets
common range.
9. Setting number of channels
With 3638-20 stand-alone, or connected with
COMMUNICATION BASE and PC, setting is available.
Select input for channel 1 or channel 2.
At factory shipment settings, both channel 1 and channel 2
are valid.
With 3638-20 stand-alone, or connected together with
COMMUNICATION BASE and PC, setting is available.
Two recording mode options are maximum value recording
and average value recording (
displayed).
AV
At factory shipment setting, average value recording is
selected.
NOTE
NOTE
Maximum value recording mode records maximum value within
interval, so the initial data is not recorded immediately after
recording starts but is recorded from the following interval.
For example, when measurement started at 12:00 at 1 minute
intervals, the first data is data recorded when 1 minute has
passed at 12:01.
11. Alarm setting
Alarm setting is available when 3638-20,
COMMUNICATION BASE and personal computer are
connected.
Designate upper limit and lower limit value of measurement
value with personal computer to output alarm signal when
measurement value goes outside of set range. Two channels
pass individual judgment.
During signal output, AL appears on the display.
Alarm signal output terminal is an open drain output. During
signal output, signal side and GND side are internally grounded
and any other time, it remains open. When external power
supply is supplied, alarm signal can control relay or sequencer.
Maximum rate: 30 V, 200 mA (allowable loss 200 mW)
Set comments entered by personal computer to logger using
COMMUNICATION BASE. When sorting collected
recording data, comments are helpful.
Comment setting is available when personal computer is
connected to both logger and COMMUNICATION BASE.
13. Maximum/Minimum value display
3638-20 stand-alone enables maximum/minimum value
display settings. Maximum value or minimum value of
recorded data is alternately displayed.
NOTE
・
Maximum value and minimum value are from data recorded in
memory and may vary from each displayed value per second.
・
When recording method is set as endless, maximum value and
minimum value from the beginning of recording are displayed.
When old data is overwritten after extended recording, data
currently recorded may differ from displayed maximum value
and minimum value.
button on monitor screen to select 100
mA or 1000 mA.
When measurement range is reselected, decimal point
moves its position.
When manual operation is conducted, common range
settings apply to both channel 1 and channel 2.
100 mA range1000 mA range
(5) Starting and ending recording
Press
REC/STOP
data and start recording. (
Press
REC/STOP
button for 1 second to clear last recorded
appears.)
REC
button for 1 second to stop recording.
When memory is full, recording automatically stops when
recording method: one time is selected.
When batteries are weak, recording does not start. During
recording, weak batteries interrupt recording.
(6) Maximum/Minimum value display
Press
MAX/MIN
button to display maximum or minimum
value from recorded data.
Each button press alternates display.
・Previously recorded logger data is erased when recording is
resumed. Be sure to load data to be saved to
COMMUNICATION BASE or to personal computer before
recording.
・At any other time even when interval setting display is not
shown, except during recording and waiting for recording,
communication with COMMUNICATION BASE is available.
However communication is disabled when logger is set to sleep.
3638-20 settings in application software
COMMUNICATION UTILITY packaged with
COMMUNICATION BASE are as follows.
Go to 'Communication' on the menu bar in
COMMUNICATION UTILITY and select 'Set measurement
condition'. When measurement condition setting window is
open, select '3638-20' setting items to set settings.
interval, range and alarm setting are only available in 3638-20
setting items. Personal computer, COMMUNICATION BASE
and 3638-20 must be connected during setting.
・Common settings are available to be set in '3911-20, 3912-20'
Number of input2 channels (1 or 2 channels settable)
Measurement range0 to 1000 mA
Range structure100.0 mA, 1000 mA
Measurement methodTrue effective value calculation
Measurement interval0.2 second (1 ch only), 1 second (1 or 2 ch)
Measurement accuracy
Unit + Sensor
9657:
9658:
Operating temperature
and humidity for
guaranteed accuracy
Period of guaranteed
accuracy
Temperature coefficientAccuracy X 0.05/
LCD displayMeasurement value, Interval, Battery status
Recording modeRecording maximum value (records maximum value
Interval1/2/5/10/15/20/30 s, 1/2/5/10/15/20/30/60 min
Recording capacity32,000 data X 1 channel
9657/9658 CLAMP ON LEAK SENSOR
AC35 mVrms, 50 mV peak value
Unit:
1% rdg. 5 dgt. (50/60 Hz)
2% rdg. 10 dgt. (100 mA range, 50/60 Hz)
2% rdg. 6 dgt. (1000 mA range, 50/60 Hz)
4.5% rdg. 10 dgt. (100 mA range, 50/60 Hz)
4.5% rdg. 6 dgt. (1000 mA range, 50/60 Hz)
235(739 ), 80%RH or less
1 year
(remaining battery power indicator: 4 phases)
Unit (A), recording (REC), prescheduled(),
average value recording (AV), maximum value (MAX),
minimum value (MIN), alarm (AL)
within interval)
Recording average value (records average value
within interval)
Alarm outputTurns ON (open drain output) when value goes
Data backupAvailable
InterfaceInfrared optical data transfer
Power supplyLR03 alkaline battery X 4 (1.5 VDC X 4)
Maximum rated power 0.1 VA
Battery lifeAbout 10 days
DimensionsApprox. 57W X 86H X 30D mm
MassApprox. 130 g (4.6 oz) (including batteries)
Location for useIndoors, altitude up to 2000 m
Operate temperature
and humidity range
Storage temperature
and humidity range
AccessoriesLR03 alkaline battery X 4
Options3911-20 COMMUNICATION BASE
Standards Applying
Safety
Displays maximum value and minimum value.
outside previously set upper limit or lower limit.
(Data not erased by weak batteries or battery
replacement)
(temperature at 20 , power save function: valid,
measurement interval: 0.2 second)
About 1 month
(temperature at 20 , power save function: valid,
measurement interval: 1 second)
(excluding projections)/ 2.24"W X 3.39"H X 1.18"D
0to50 , 80% RH or less (no condensation)
(32 to 122 )
-10 to 60 , 80% RH or less (no condensation)
(14 to 140 )
9632 CONNECTION CABLE
Instruction Manual
3912-20 COMMUNICATION BASE
9657 CLAMP ON LEAK SENSOR
9658 CLAMP ON LEAK SENSOR
To clean the product, wipe it gently with a soft cloth
moistened with water or mild detergent. Never use solvents
such as benzene, alcohol, acetone, ether, ketones, thinners
or gasoline, as they can deform and discolor the case.
Service
If the product seems to be malfunctioning, confirm that the
batteries are not discharged, and that the sensor cables are
not open circuited before contacting your dealer or Hioki
representative.
Error Messages
The following error may be displayed on the LCD of the
main instrument as shown below.
Error MessageMeaning
ROM error
RAM error
Adjustment data error
Adjustment data is no longer available. Please send it for
repair. Contact your dealer or Hioki representative.
All inquiries to International Sales and Marketing Department
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL: +81-268-28-0562 / FAX: +81-268-28-0568
E-mail: os-com@hioki.co.jp
URL http://www.hioki.com/
Printed in Japan 3638A981-05
All reasonable care has been taken in the production of this manual,
but if you find any points which are unclear or in error, please
contact your supplier or the International Sales and Marketing
Department at HIOKI headquarters.
In the interests of product development, the contents of this manual
are subject to revision without prior notice.
The content of this manual is protected by copyright. No
reproduction, duplication or modification of the content is permitted
without the authorization of Hioki E.E. Corporation.
Page 44
HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp / URL http://www.hioki.com/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
3638A981-05 08-06H
Printed on recycled paper
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.