Hioki 3638-20 Instruction Manual

Page 1
INSTRUCTION MANUAL
For...は専用機種。複数の場合は「/」で区切る。
不要の場合はとる。
形名を入力。 複数の場合は「/」で区切る。
3638-20
品名を入力。
LEAK LOGGER
Page 2
Page 3
Contents
Chapter 1 Product Outline 1
1.1 Name and Functions of Parts 2
1.2 Interval and Maximum Recording Time
1.3 Measurement Value Recording Modes
1.4 Leak Current Measurement
1.5 Load current measurement
Chapter 2 Set Up 9
2.1 Replacing the Battery 9
2.2 Power Save Function
2.3 Setting Current Time
2.4 Connecting Sensor Cable
11
12
12
3
5
6
8
Chapter 3 Settings 15
3.1 Setting Items 15
3.2 Manual Setting
3.3 Setting by COMMUNICATION BASE
20
22
Chapter 4 Specifications 25
Chapter 5 Maintenance and Service 27
Page 4
Page 5
_____________________________________________________________________
Introduction
Thank you for purchasing the Hioki 3638-20 LEAK LOGGER. To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference.
Inspection
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Accessories
9632 CONNECTION CABLE Instruction Manual LR03 alkaline battery (built into this unit, for monitor)
4
X
i
Before using the product the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
NOTE
______________________________________________________________
Testing monitor batteries installed in the unit may possibly be weak. Replace batteries before extended measurement usage. Before using the product, make sure that the insulation on the sensor cables is undamaged and that no bare conductors are improperly exposed. Using the product in such conditions could cause an electric shock, so contact your dealer or Hioki representative for repair.
Inspection
Page 6
ii
R
_____________________________________________________________________
SafetyNotes
This instrument is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for
DANGE
Safety Symbols
safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from product defects.
This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before using the instrument, be sure to carefully read the following safety notes.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with the relevant function. In the manual, the symbol indicates particularly important information that the user should read before using the instrument.
symbol) before using the
Indicates DC (Direct Current).
______________________________________________________________
Safety Notes
Page 7
_____________________________________________________________________
R
iii
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents an extreme
DANGE
WARNING
CAUTION
NOTE
hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument.
Advisory items related to performance or correct operation of the instrument.
Accuracy
The specifications in this manual include figures for "measurement accuracy" when referring to digital measuring instruments, and for "measurement tolerance" when referring to analog instruments.
f.s.
(maximum display or scale value, or length of scale) Signifies the maximum display (scale) value or the length of the scale (in cases where the scale consists of unequal increments or where the maximum value cannot be defined). In general, this is the range value (the value written on the range selector or equivalent) currently in use.
rdg.
(displayed or indicated value) This signifies the value actually being measured, i.e., the value that is currently indicated or displayed by the measuring instrument.
dgt.
(resolution) Signifies the smallest display unit on a digital measuring instrument, i.e., the value displayed when the last digit on the digital display is "1".
______________________________________________________________
Safety Notes
Page 8
iv
_____________________________________________________________________
Measurement categories (Overvoltage categories)
To ensure safe operation of measurement product, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT I to CAT IV, and called measurement categories. These are defined as follows.
CAT I
CAT II
CAT III
CAT IV
Secondary electrical circuits connected to an AC electrical outlet through a transformer or similar device.
Primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets.
The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel).
Higher-numbered categories correspond to electrical environments with greater momentary energy. So a measurement device designed for CAT III environments can endure greater momentary energy than a device designed for CAT II. Using a measurement product in an environment designated with a higher-numbered category than that for which the product is rated could result in a severe accident, and must be carefully avoided. Never use a CAT I measuring product in CAT II, III, or IV environments. The measurement categories comply with the Overvoltage Categories of the IEC60664 Standards.
______________________________________________________________
Safety Notes
Page 9
_____________________________________________________________________
R
Notes on Use
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
This product cannot be used on following voltage lines.
v
DANGE
If the voltage exceeds, there will be a short-circuit accident or electrocution accident will result. CAT III, AC300 Vrms for 9657 CAT III, AC150 Vrms for 9658 To avoid short circuits and potentially life-threatening hazards, never attach the clamp-on leak sensor to a circuit that operates at more than the following maximum rated voltage to earth. AC300 Vrms for 9657 AC150 Vrms for 9658 Clamp-on leak sensor should only be connected to the secondary side of a breaker, so the breaker can prevent an accident if a short circuit occurs. Connections should never be made to the primary side of a breaker, because unrestricted current flow could cause a serious accident if a short circuit occurs. For model 9657 When conductors being measured carry in excess of the safe voltage level (SELV-E)
and not more than 300 V, to prevent short circuits and electric shock while the clamp core jaw is open, make sure that conductors to be measured are insulated with material conforming to (1) Measurement Category III
, (2) Pollution Degree 2, and (3) Basic Insulation Requirements for Working Voltages of 300 V.
______________________________________________________________
Notes on Use
Page 10
vi
R
_____________________________________________________________________
For model 9658
DANGE
When conductors being measured carry in excess of the safe voltage level (SELV-E)
and not more than 150
V, to prevent short circuits and electric shock while the clamp core jaw is open, make sure that conductors to be measured are insulated with material conforming to (1) Measurement Category III
, (2) Pollution Degree 2, and (3) Basic Insulation Requirements for Working Voltages of 150 V. Refer to the following standards regarding the meanings of underlined terms. IEC61010-1 IEC61010-2-031 IEC61010-2-032
To avoid electric shock, do not allow the product to
WARNING
get wet, and do not use it when your hands are wet.
When measuring the ground conductor on a
transformer Class 2 connection site, be careful not to approach high voltage devices or conductors. Also, if close to high voltage charging devices or if measurement is otherwise difficult, first change the route of the grounding wire.
To avoid damage to the product, protect it from
vibration or shock during transport and handling, and be especially careful to avoid dropping.
This is a precision instrument: do not clamp any
foreign objects in the end of the clamp core, or insert anything in the core gap.
Do not exert excessive pressure on the clamp sensor
or attempt to wedge the sensor into a tight spot for measurement.
______________________________________________________________
Notes on Use
Page 11
_____________________________________________________________________
vii
CAUTION
NOTE
To avoid electric shock when measuring live lines, wear appropriate protective gear, such as insulated rubber gloves, boots and a safety helmet. This product is designed for indoor use, and operates reliably from 0
to 50 . Do not store or use the product where it could be exposed to direct sunlight, high temperature or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate so that it no longer meets specifications.
Accurate measurement may be impossible in the presence of strong magnetic fields, such as near transformers and high­current conductors, or in the presence of strong electromagnetic fields such as near radio transmitters.
______________________________________________________________
Notes on Use
Page 12
viii
_____________________________________________________________________
Measurement Flow Chart and Reference Guide
Is there sufficient
battery charge remaining
to operate 3638
Setting current time
Connecting sensor
Setting 3638-20 stand-alone
IntervalNumber of
channels
Recording modeRange
See 3.2.
Press REC/STOP button to start measurement.
See 3.2.
When recording data reaches 32000 (
16000 in 2 channels
measurement stops.
),
-20?
Set current time when using 3638-20 for the first time or replacingbatteries. See 2.3.
See 2.4.
Setting 3638-20 with COMMUNICATION BASE
IntervalCurrent timeStart controlRecording method
Quick start
Starts measurement immediately.
One time
When recording data exceeds 32000( overwrites previously recorded data.
Press REC/STOP button to stop measurement.
Replace batteries
COMMUNICATION BASE and PC
Interval Current timeStart control RangeRecording methodNumber of channelsRecording modeAlarm ・Comment
No
recordingmethod
Settingstart control
Start control options
Starts measurement at prescheduled time.
Setting
Setting 3638-20 with
Endless
16000 in 2 channels
See 2.1.
Yes
Prescheduled start
), it
After measurement is complete, disconnect sensor cables. Use COMMUNICATION BASE to send measurement data to personal computer. (See COMMUNICATION BASE instruction manual.)
______________________________________________________________
Measurement Flow Chart and Reference Guide
Page 13
_____________________________________________________________________
1
Chapter 1
Product Outline
3638-20 LEAK LOGGER with clamp on leak sensor records 1 or 2 channels of data at leakage current value. It measures and records AC RMS current values of up to 1 Arms at fixed intervals. With two optional recording modes; recording maximum value and recording average value, extended duration of current recording is possible with battery operation. Data is saved in nonvolatile memory when batteries are weak or removed for replacement.
NOTE
______________________________________________________________
3638-20 LEAK LOGGER cannot be set with 3910-20 COMMUNICATION BASE.
Product Outline
Page 14
2
_____________________________________________________________________
1.1 Name and Functions of Parts
1
2
3
6
7
4
8
5
1. LCD Displays measurement value and settings.
2. Optical data transfer ports
3. INTERVAL button Calls up interval setting display to set interval,
4. MAX./MIN. button Displays maximum value or minimum value of
5. Clamp connection terminal
6. REC/STOP button Pressing more than 1 second initiates or stops
(SELECT) button Interval is selected in interval setting display.
7. Alarm output terminal
8. 100/1000 mA button
Enables optical data transfer to COMMUNICATION BASE.
recording mode and measurement channels.
recorded data.
Connects 9657/9658 CLAMP ON LEAK SENSOR.
recording.
Connects 9632 CONNECTION CABLE and outputs alarm signal (open drain output).
Changes measurement range from 100/1000 mA.
______________________________________________________________
Product Outline
Page 15
_____________________________________________________________________
1.2 Interval and Maximum Recording Time
Interval and maximum recording time (when power save function is valid) are as follows. When 3638-20 LEAK LOGGER is used in 2 channels, maximum recordable data is 16000 per unit.
INTVL Maximum Recording Time
1s 4h26min40s
2s 8h53min20s
5s 22 h 13 min 20 s
10 s 1 day 20 h 26 min 40 s
15 s 2 day 18 h 40 min
20 s 3 day 16 h 53 min 20 s
30 s 5 day 13 h 20 min
1min 11 day 2 h 40 min
2min 22 day 5 h 20 min
3
5min 55 day 13 h 20 min
10 min 111 day 2 h 40 min
15 min 166 day 16 h
20 min 222 day 5 h 20 min
30 min 333 day 8 h
60 min 666 day 6 h
*
*
*
*
*
*
*: Maximum recording time is limited by battery charge
condition.
______________________________________________________________
Product Outline
Page 16
4
_____________________________________________________________________
When 3638-20 LEAK LOGGER is used in 1 channel, maximum recordable data is 32000 per unit.
INTVL Maximum Recording Time
1s 8h53min20s
2s 17 h 46 min 40 s
5s 1 day 20 h 26 min 40 s
10 s 3 day 16 h 26 min 40 s
15 s 5 day 13 h 20 min
20 s 7day9h46min40s
30 s 11 day 2 h 40 min
1min 22 day 5 h 20 min
2min 44 day 10 h 40 min
5min 111 day 2 h 40 min
10 min 222 day 5 h 20 min
15 min 333 day 8 h
20 min 444 day 10 h 40 min
30 min 666 day 6 h
60 min 1333 day 8 h
*
*
*
*
*
*
*
*: Maximum recording time is limited by battery charge
condition.
______________________________________________________________
Product Outline
Page 17
_____________________________________________________________________
p
1.3 Measurement Value Recording Modes
Effective value calculation
To calculate effective value, measurement signal is sampled every 334 µs and operation is carried out from data with maximum 135 points to seek effective value.
Maximum 135 points partitioned at com
lete cycle
334µs
Recording
To record measurement value, calculate effective value per 1 second or 0.2 second and record average or maximum value of all data within interval in memory. After recording starts, average or maximum value within interval is recorded, so the initial data is not recorded immediately after recording starts but is recorded from the following interval.
Interval
1 or 0.2 second interval
5
Effective
value
Recordingstarts
Average
value
Interval
Recording average value
Recording
Records average/maximum value
Effective
value
Maximum
value
Interval
Recording maximum value
______________________________________________________________
Product Outline
Recording
Page 18
6
_____________________________________________________________________
1.4 Leak Current Measurement
Clamp the sensor on the conductor, so that the conductor passes through the center of the clamp core. For measurement of grounded leads, clamp the sensor on one lead only (see a). For overall measurements, clamp the sensor on the entire circuit path (see b).
NOTE
When the current to be measured is unknown, begin measurement with the 1000 mA range selected.
Single-phase 3-lead circuits
Transformer
Ig: leak Current
Class 2 grounding line
Three-phase 3-lead circuits
Ig
a
E
2
Class 2 grounding line
a
Ig
E
Clamp all three leads of the circuit
b
Clamp all three leads
b
2
of the circuit
b
Load device
b
Load
device
a
E
Class 3 grounding line
3
NOTE
For measurement of single-phase 2-lead circuits, clamp both leads of the circuit. For measurement of three-phase 4-lead circuits, clamp all four leads of the circuit. If this is not possible, the measurement can also be carried out on the ground lead of the equipment.
______________________________________________________________
Product Outline
Page 19
_____________________________________________________________________
7
NOTE
If a strong current (on the order of 100 A) is flowing in an adjacent circuit, accurate measurement may not be possible. Perform the measurement at a sufficient distance from other current-carrying conductors. Do not input a current which exceeds the maximum continuous input rating. When the clamp is open and while the range is being changed, an indicated value in the range of tens may appear; this is not a malfunction. The indication goes progressively to zero, but this make take a little time. Beginning measurement before the display reaches zero has no effect on measurement.
Checking for insulation faults
Normally, for a type 2 grounding installation of a transformer, the measurement will first be made to check for overall circuit leak current in the ground lead (a). Current changes can be used to diagnose the leak current condition. When leak current has been detected, the measurement should proceed from the power source towards the load, using overall measurement.
1. If an insulation fault in the wiring has occurred at position A in the illustration, leak current will be detected at position b using overall measurement, but not at position b'.
2. If an insulation fault in the load equipment has occurred at position B in the illustration, leak current will be detected at position c using overall measurement, but not at position c'.
b
b
Single-phase 3-lead circuits
c
c
a
Ig
E
2
A
Load
device
Load
device
B
______________________________________________________________
Product Outline
Page 20
8
_____________________________________________________________________
1.5 Load current measurement
Be sure to clamp only one lead of the conductor. Clamp so that the conductor is in the center of the sensor aperture.
OK NG
NOTE
The frequency of special waveforms such as at the secondary side of an inverter can not be indicated correctly. When the current to be measured is unknown, begin measurement with the 1000 mA range selected. Do not input a current which exceeds the maximum continuous input rating.
______________________________________________________________
Product Outline
Page 21
_____________________________________________________________________
9
Chapter 2
Set Up
2.1 Replacing the Battery
To avoid electric shock when replacing the batteries,
WARNING
first disconnect the connection cables from the object to be measured. During battery replacement, use caution not to put any foreign materials such as a metal object into the unit to avoid damage to the unit. Before using the product after replacing the batteries, replace the cover and screw. Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result. To avoid the possibility of explosion, do not short circuit, disassemble or incinerate batteries. To avoid corrosion from battery leakage, remove the batteries from the product if it is to be stored for a long time.
CAUTION
______________________________________________________________
When exchanging the batteries, the circuit may sometimes short circuit due to static electricity. As far as possible, do not touch the base board with bare hands.
Set Up
Page 22
10
_____________________________________________________________________
Installing new batteries ensures about one month of recording (when interval is set at 1 sec) and about 10 days of recording (when interval is set at 0.2 sec). (reference value with power save setting set to valid and at 20 Remaining battery power indicator (
)indicates
)
remaining battery life reducing incrementally from right. Empty battery power indicator (
) indicates time to
replace batteries.
1. Remove back cover screw to remove cover. Verify polarity and install four new LR03 alkaline batteries.
2. Fit cover properly and tighten screw.
______________________________________________________________
Set Up
Page 23
_____________________________________________________________________
11
2.2 Power Save Function
Display window is automatically turned off in approximately 15 seconds after last key entry. (Sleep) However, while recording, REC/AV/ / / AL mark shows each conditions.
Sleeping......
Press any button to turn display on to display measurement value or to set settings. Note when interval setting display is on, sleep does not engage with no button press.
Initially, power save function is on. To turn off power save function, follow the instructions below. When power save function is off, maximum continuous duration is approximately 10 days.
1. Connect logger, COMMUNICATION BASE and
personal computer.
2. Start up application software packaged with
COMMUNICATION BASE.
3. Go to Communications on the menu bar and select
Power Save Options. Choose Off to turn off power save function.
NOTE
______________________________________________________________
See COMMUNICATION BASE instruction manual to connect logger and to install application software. To use application software, see operation guide.
Set Up
Page 24
12
_____________________________________________________________________
2.3 Setting Current Time
When replacing 3638-20 LEAK LOGGER batteries or using 3638-20 stand-alone (with manual operation) for the first time, connect with COMMUNICATION BASE and set current time.
See how to set current time in COMMUNICATION BASE instruction manual.
2.4 Connecting Sensor Cable
Maximum input current is 1 Arms. Do not exceed the
WARNING
continuous maximum input current. This product cannot be used on following voltage lines. If the voltage exceeds, there will be a short­circuit accident or electrocution accident will result. CAT III, AC300 Vrms for 9657 CAT III, AC150 Vrms for 9658 When connecting the clamp on leak sensor, if a metallic part of the sensor exposed while the clamp is open should touch the two wires of the line, or if the sensor is used on a bare conductor, a shorting accident or electric shock may result. The 3638-20 measurement range is set at 100 mA, do not exceed 100 mArms. To avoid damaging the unit, do not use any other sensors except 9657/9658 CLAMP ON LEAK SENSOR as sensor connector.
______________________________________________________________
Set Up
Page 25
_____________________________________________________________________
Connecting clamp on leak sensor
3638-20 LEAK LOGGER enables 2 channels of data measurement when connected with 9657/9658 CLAMP ON LEAK SENSOR.
9657/9658 CLAMP ON LEAK SENSOR
13
9658
NOTE
9657
Carefully read 9657/9658 CLAMP ON LEAK SENSOR instruction manuals before use. Improper measurement method may result in accidental injury, death or damage in the unit.
______________________________________________________________
Set Up
Page 26
14
_____________________________________________________________________
Connecting 9632 CONNECTION CABLE
Connecting 9632 CONNECTION CABLE to alarm output terminal.When connecting connection cable, securely insert
connection cable to unit as designated by triangle mark
on connection terminal. Improper connection results in failure to output accurate signal.
9632 CONNECTION CABLE
9632 CONNECTION CABLE
GND
Non connection
Signal
Cord length: 1000 mm
White: Signal Black: GND
______________________________________________________________
Set Up
Page 27
_____________________________________________________________________
Chapter 3
Settings
3.1 Setting Items
Logger stand-alone manual settings and settings in combination with COMMUNICATION BASE with measurement conditions stored in memory loaded from personal computer.
15
3638-20+
3638-20
1. Start recording Valid Valid Valid
2. Stop recording Valid --- ---
3. Interval setting Valid Valid Valid
4. Current time setting --- Valid Valid
5. Start control --- Valid Valid
6. Recording method setting --- Valid Valid
7. Measurement interval setting Valid --- Valid
8. Range setting Valid --- Valid
9. Setting number of channels Valid --- Valid
10. Recording mode setting Valid --- Valid
11. Alarm setting --- --- Valid
12. Comments --- --- Valid
13. Minimum/Maximum value display
NOTE
When PC, COMMUNICATION BASE and 3638-20 are
Valid --- ---
COMMUNICATION
BASE
3638-20+
COMMUNICATION
BASE+PC
connected, separate range can be set in each channel. Once settings are set, range setting with manual operation sets common range.
Alarm and comment setting are available when personal computer is connected to both logger and COMMUNICATION BASE.
______________________________________________________________
Settings
Page 28
16
_____________________________________________________________________
1. Start recording
Start manual recording by pressing logger
REC/STOP
button for 1 second or initiate by prescheduled start set using COMMUNICATION BASE. When time scheduled start is engaged, clock icon appears in display. When batteries are weak, recording does not start. During recording, weak battery interrupts recording.
2. Stop recording
Stop recording by pressing logger
REC/STOP
button for 1 second. Or recording stops automatically when data is full when set to recording method: one time.
3. Interval setting
Set interval with logger alone or using COMMUNICATION BASE. (1/2/5/10/15/20/30 s, 1/2/5/10/15/20/30/60 min)
4. Current time setting
To set current time, see COMMUNICATION BASE instruction manual.
5. Start control
Set specific recording date and time using COMMUNICATION BASE to engage time scheduled start. When time scheduled start is engaged, clock icon appears in display.
6. Recording method setting
Set recording method using COMMUNICATION BASE. Choose either one time or endless recording method. Default setting is one time. One time: Ends recording when data reaches 32000 (2 ch:
16000).
Endless : Overwrites previously recorded data when data
exceeds 32000 (2 ch: 16000).
______________________________________________________________
Settings
Page 29
_____________________________________________________________________
17
7. Setting the measurement interval
You can set the measurement interval on the 3638-20 or through the COMMUNICATION BASE. Select between an interval of 0.2 sec (when measuring just 1 channel) or 1 second. The factory setting is 1 second.
8. Range setting
With 3638-20 stand-alone, or connected with COMMUNICATION BASE and PC, range setting is available. Two measurement range options are 100 mA range and 1000 mA. Select and set in accordance with measurement current. When manually operated, common settings apply to channel 1 and channel 2. When 3638-20 measurement range is set at 100 mA, do not exceed 100 mArms.
NOTE
When PC, COMMUNICATION BASE and 3638-20 are connected, separate range can be set in each channel. Once settings are set, range setting with manual operation sets common range.
9. Setting number of channels
With 3638-20 stand-alone, or connected with COMMUNICATION BASE and PC, setting is available. Select input for channel 1 or channel 2. At factory shipment settings, both channel 1 and channel 2 are valid.
______________________________________________________________
Settings
Page 30
18
_____________________________________________________________________
10. Recording mode setting
With 3638-20 stand-alone, or connected together with COMMUNICATION BASE and PC, setting is available. Two recording mode options are maximum value recording and average value recording (
displayed).
AV
At factory shipment setting, average value recording is selected.
NOTE
NOTE
Maximum value recording mode records maximum value within interval, so the initial data is not recorded immediately after recording starts but is recorded from the following interval. For example, when measurement started at 12:00 at 1 minute intervals, the first data is data recorded when 1 minute has passed at 12:01.
11. Alarm setting
Alarm setting is available when 3638-20, COMMUNICATION BASE and personal computer are connected. Designate upper limit and lower limit value of measurement value with personal computer to output alarm signal when measurement value goes outside of set range. Two channels pass individual judgment. During signal output, AL appears on the display.
Alarm signal output terminal is an open drain output. During signal output, signal side and GND side are internally grounded and any other time, it remains open. When external power supply is supplied, alarm signal can control relay or sequencer. Maximum rate: 30 V, 200 mA (allowable loss 200 mW)
______________________________________________________________
Settings
Page 31
_____________________________________________________________________
19
12. Comments
Set comments entered by personal computer to logger using COMMUNICATION BASE. When sorting collected recording data, comments are helpful. Comment setting is available when personal computer is connected to both logger and COMMUNICATION BASE.
13. Maximum/Minimum value display
3638-20 stand-alone enables maximum/minimum value display settings. Maximum value or minimum value of recorded data is alternately displayed.
NOTE
Maximum value and minimum value are from data recorded in
memory and may vary from each displayed value per second.
When recording method is set as endless, maximum value and minimum value from the beginning of recording are displayed. When old data is overwritten after extended recording, data currently recorded may differ from displayed maximum value and minimum value.
______________________________________________________________
Settings
Page 32
20
_____________________________________________________________________
3.2 Manual Setting
3638-20 LEAK LOGGER stand-alone manual operation settings are shown below.
(1) Interval setting
Press
INTERVAL
display to interval setting display. ( Press
SELECT
Press
INTERVAL
(2) Setting number of channels
Press
INTERVAL
Each
100/1000 mA
When channel 1 is selected, only display " andwhenchannel2isselected,display" When both channel 1 and channel 2 are selected, display "
" is turned on. In these cases, the measurement interval is 1 second. When neither "
"nor" " is lit, the measurement interval is 0.2
second (measurement on CH1 only).
button to switch measurement value
INTVL
appears.)
button to designate interval.
button to complete setting.
button to display interval setting display.
button press selects channel to be used.
"isturnedon
" is turned on.
(3) Setting recording mode
Press
INTERVAL
Each
MAX/MIN
maximum value recording and average value recording (
button to display interval setting display.
button press alternates and selects
AV
displayed). To set maximum value recording and average value recording, see 1.3 Measurement Value Recording Modes.
______________________________________________________________
Settings
Page 33
_____________________________________________________________________
21
(4) Setting range
Press
100/1000 mA
button on monitor screen to select 100 mA or 1000 mA. When measurement range is reselected, decimal point moves its position. When manual operation is conducted, common range settings apply to both channel 1 and channel 2.
100 mA range 1000 mA range
(5) Starting and ending recording
Press
REC/STOP
data and start recording. ( Press
REC/STOP
button for 1 second to clear last recorded
appears.)
REC
button for 1 second to stop recording. When memory is full, recording automatically stops when recording method: one time is selected. When batteries are weak, recording does not start. During recording, weak batteries interrupt recording.
(6) Maximum/Minimum value display
Press
MAX/MIN
button to display maximum or minimum value from recorded data. Each button press alternates display.
______________________________________________________________
Settings
Page 34
22
_____________________________________________________________________
3.3 Setting by COMMUNICATION BASE
1. Press logger
INTERVAL
2. When logger LCD shows
REC/STOP
button for more than 1 second to stop
button lightly to display LCD.
mark or clock icon, press
REC
recording. During recording or waiting time before recording start time, data transfer cannot be established with COMMUNICATION BASE.
3. Press logger
INTERVAL
button to display interval setting
display. (INTVL appears.)
4. Connect COMMUNICATION BASE with logger.
5. Press COMMUNICATION BASE
SEND
button for more
than 1 second to send data settings to logger.
NOTE
______________________________________________________________
Settings
Previously recorded logger data is erased when recording is
resumed. Be sure to load data to be saved to COMMUNICATION BASE or to personal computer before recording.
At any other time even when interval setting display is not
shown, except during recording and waiting for recording, communication with COMMUNICATION BASE is available. However communication is disabled when logger is set to sleep.
Page 35
_____________________________________________________________________
23
3638-20 settings in application software COMMUNICATION UTILITY packaged with COMMUNICATION BASE are as follows.
Go to 'Communication' on the menu bar in COMMUNICATION UTILITY and select 'Set measurement condition'. When measurement condition setting window is open, select '3638-20' setting items to set settings.
NOTE
______________________________________________________________
Comment, recording mode, measurement channel, measurement
interval, range and alarm setting are only available in 3638-20 setting items. Personal computer, COMMUNICATION BASE and 3638-20 must be connected during setting.
Common settings are available to be set in '3911-20, 3912-20'
setting items.
Settings
Page 36
24
_____________________________________________________________________
______________________________________________________________
Settings
Page 37
_____________________________________________________________________
Chapter 4
Specifications
25
Sensor types
Input AC voltage (AC25 mV/A)
Maximum input current
Number of input 2 channels (1 or 2 channels settable)
Measurement range 0 to 1000 mA
Range structure 100.0 mA, 1000 mA
Measurement method True effective value calculation
Measurement interval 0.2 second (1 ch only), 1 second (1 or 2 ch)
Measurement accuracy
Unit + Sensor
9657:
9658:
Operating temperature and humidity for guaranteed accuracy Period of guaranteed accuracy
Temperature coefficient Accuracy X 0.05/
LCD display Measurement value, Interval, Battery status
Recording mode Recording maximum value (records maximum value
Interval 1/2/5/10/15/20/30 s, 1/2/5/10/15/20/30/60 min
Recording capacity 32,000 data X 1 channel
9657/9658 CLAMP ON LEAK SENSOR
AC35 mVrms, 50 mV peak value
Unit:
1% rdg. 5 dgt. (50/60 Hz)
2% rdg. 10 dgt. (100 mA range, 50/60 Hz) 2% rdg. 6 dgt. (1000 mA range, 50/60 Hz)
4.5% rdg. 10 dgt. (100 mA range, 50/60 Hz)
4.5% rdg. 6 dgt. (1000 mA range, 50/60 Hz)
23 5 (73 9 ), 80%RH or less
1 year
(remaining battery power indicator: 4 phases) Unit (A), recording (REC), prescheduled( ), average value recording (AV), maximum value (MAX), minimum value (MIN), alarm (AL)
within interval) Recording average value (records average value within interval)
16,000 data X 2 channels
______________________________________________________________
Specifications
Page 38
26
_____________________________________________________________________
Recording start Manual start, Prescheduled start
Recording stop Manual stop, Memory full
Recording method One time, Endless
Displaying Max/Min value
Alarm output Turns ON (open drain output) when value goes
Data backup Available
Interface Infrared optical data transfer
Power supply LR03 alkaline battery X 4 (1.5 VDC X 4)
Maximum rated power 0.1 VA
Battery life About 10 days
Dimensions Approx. 57W X 86H X 30D mm
Mass Approx. 130 g (4.6 oz) (including batteries)
Location for use Indoors, altitude up to 2000 m
Operate temperature and humidity range
Storage temperature and humidity range
Accessories LR03 alkaline battery X 4
Options 3911-20 COMMUNICATION BASE
Standards Applying
Safety
Displays maximum value and minimum value.
outside previously set upper limit or lower limit.
(Data not erased by weak batteries or battery replacement)
(temperature at 20 , power save function: valid, measurement interval: 0.2 second) About 1 month (temperature at 20 , power save function: valid, measurement interval: 1 second)
(excluding projections)/ 2.24"W X 3.39"H X 1.18"D
0to50 , 80% RH or less (no condensation) (32 to 122 )
-10 to 60 , 80% RH or less (no condensation) (14 to 140 )
9632 CONNECTION CABLE Instruction Manual
3912-20 COMMUNICATION BASE 9657 CLAMP ON LEAK SENSOR 9658 CLAMP ON LEAK SENSOR
EMC
EN61326 EN61010 Pollution Degree 2
______________________________________________________________
Specifications
Page 39
_____________________________________________________________________
27
Chapter 5
Maintenance and Service
Cleaning
To clean the product, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
Service
If the product seems to be malfunctioning, confirm that the batteries are not discharged, and that the sensor cables are not open circuited before contacting your dealer or Hioki representative.
Error Messages
The following error may be displayed on the LCD of the main instrument as shown below.
Error Message Meaning
ROM error
RAM error
Adjustment data error
Adjustment data is no longer available. Please send it for repair. Contact your dealer or Hioki representative.
______________________________________________________________
Maintenance and Service
Page 40
28
_____________________________________________________________________
______________________________________________________________
Maintenance and Service
Page 41
Page 42
Page 43
HIOKI 3638-20 LEAK LOGGER
Instruction Manual
Publication date: June 2008 Revised edition 5
Edited and published by HIOKI E.E. CORPORATION
Technical Support Section
All inquiries to International Sales and Marketing Department
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL: +81-268-28-0562 / FAX: +81-268-28-0568
E-mail: os-com@hioki.co.jp
URL http://www.hioki.com/
Printed in Japan 3638A981-05
All reasonable care has been taken in the production of this manual, but if you find any points which are unclear or in error, please contact your supplier or the International Sales and Marketing Department at HIOKI headquarters. In the interests of product development, the contents of this manual are subject to revision without prior notice. The content of this manual is protected by copyright. No reproduction, duplication or modification of the content is permitted without the authorization of Hioki E.E. Corporation.
Page 44
HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp / URL http://www.hioki.com/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA TEL +1-609-409-9109 / FAX +1-609-409-9108
3638A981-05 08-06H
Printed on recycled paper
Loading...