Thank you for purchasing the HIOKI “Model 3561, 3561-01 BATTERY HiTESTER.” To obtain
maximum performance from the instrument, please read this manual first, and keep it handy for
future reference.
Trademarks
Microsoft , Windows, and Visual Basic are either registered trademarks or trademarks of Microsoft
Corporation in the United States and other countries.
2
Verifying Package Contents
When you receive the instrument, inspect it carefully to ensure that no damage occurred during
shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident,
or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Use the original packing materials when transporting the instrument, if possible.
Power cord (1)
Model 3561/ Model 3561-01 (GP-IB version)
BATTERY HiTESTER (1)
Instruction manual (this manual/ 1)
Options
The options listed below are available for the instrument. To order an option, please contact
yourauthorized Hioki distributor or reseller.Options are subject to change. Please check Hioki's
website for the latestinformation.
3
4
5
6
7
8
9
Model L2107CLIP TYPE LEAD
Model 9452CLIP TYPE LEAD
Model 9453FOUR TERMINAL LEAD
Model 9455PIN TYPE LEAD (for ultra precision)
Model 9467LARGE CLIP TYPE LEAD
Model 9770PIN TYPE LEAD
Model 9771PIN TYPE LEAD
Model 9637RS-232C CABLE (9-pin to 9-pin/cross cable)
Model 9638RS-232C CABLE (9-pin to 25-pin/cross cable)
Model 9151-02GP-IB CONNECTOR CABLE (2 m)
10
11
Page 8
2
HIOKI 3561A981-10
Safety Information
Safety Information
This instrument is designed to comply with IEC 61010 Safety
Standards, and has been thoroughly tested for safety prior to
shipment. However, mishandling during use could result in
injury or death, as well as damage to the instrument. Using the
instrument in a way not described in this manual may negate the
provided safety features. Be certain that you understand the
instructions and precautions in the manual before use. We
disclaim any responsibility for accidents or injuries not resulting
directly from instrument defects.
This manual contains information and warnings essential for safe operation of the instrument and for
maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety
precautions.
In the manual, the symbol indicates particularly important
information that the user should read before using the
instrument.
The symbol printed on the instrument indicates that the user
should refer to a corresponding topic in the manual (marked with
the symbol) before using the relevant function.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents an extreme
hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant
hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of
injury to the user or damage to the instrument.
Indicates advisory items related to performance or correct
operation of the instrument.
Other Symbols
Indicates a prohibited action.
p.
Indicates the location of reference information.
Indicates quick references for operation and remedies for
troubleshooting.
Indicates that descriptive information is provided below.
*
Page 9
Safety Information
HIOKI 3561A981-10
Measurement categories
To ensure safe operation of measurement instruments, IEC 61010 establishes safety standards for
various electrical environments, categorized as CAT II to CAT IV, and called measurement
categories.
3
CAT II
CAT III
CAT IV
Using a measurement instrument in an environment designated with a higher-numbered category
than that for which the instrument is rated could result in a severe accident, and must be carefully
avoided.
Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV measurement
applications could result in a severe accident, and must be carefully avoided.
Primary electrical circuits in equipment connected to an AC
electrical outlet by a power cord (portable tools, household
appliances, etc.)
CAT II covers directly measuring electrical outlet receptacles.
Primary electrical circuits of heavy equipment (fixed
installations) connected directly to the distribution panel, and
feeders from the distribution panel to outlets.
The circuit from the service drop to the service entrance, and
to the power meter and primary overcurrent protection device
(distribution panel).
2
3
4
5
6
Accuracy
We define measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values,
with the following meanings:
f.s.
(maximum display value or scale length)
The maximum displayable value or scale length. This is usually the
name of the currently selected range.
rdg.
(reading or displayed value)
The value currently being measured and indicated on the measuring
instrument.
dgt.
(resolution)
The smallest displayable unit on a digital measuring instrument, i.e.,
the input value that causes the digital display to show a "1" as the
least-significant digit.
7
8
9
10
11
Page 10
4
HIOKI 3561A981-10
Operating Precautions
Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benefits of the various
functions.
Use of the instrument should confirm not only to itsspecifications, but also to the specifications of all
accessories, options, and other equipment in use.
Instrument Installation and Operating Environment
Operating temperature and humidity:
0 to 40°C (32 ± 104°F), 80%RH or less (non-condensating)
Temperature and humidity range for guaranteed accuracy:
23 ± 5°C (73 ± 9°F), 80% RH or less (non-condensating)
Avoid the following locations that could cause an accident or damage
to the instrument.
Exposed to direct
sunlight
Exposed to high
temperature
Exposed to liquids
Exposed to high
humidity or
condensation
Exposed to high
levels of particulate
dust
In the presence of
corrosive or explosive
gases
Exposed to strong
electromagnetic fields
Near electromagnetic
radiators
Subject to vibration
Avoid using near electrically noisy devices, as the noise may impinge
upon the test object and cause unreliable measurements.
Installation
Do not install the instrument with any side except the bottom facing down.
Page 11
5
HIOKI 3561A981-10
Operating Precautions
Preliminary Checks
Before using the instrument the first time, verify that it operates normally to ensure that the no
damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki
representative.
Before using the instrument, make sure that the insulation on the
power cord and test leads is undamaged and that no bare
conductors are improperly exposed. Using the instrument in
such conditions could cause an electric shock, so contact your
dealer or Hioki representative for replacements.
2
This instrument may cause interference if used in residential areas. Such
use must be avoided unless the user takes special measures to reduce
electromagnetic emissions to prevent interference to the reception of radio
and television broadcasts.
Measurement Precautions
• To avoid electrical shock, be careful to avoid shorting live lines
with the test leads.
• The maximum rated voltage between input terminals and
ground is
ing 60 V with respect to ground could damage the instrument
and result in personal injury.
• To avoid injury or damage to the instrument, do not attempt to
measure AC voltage and AC current, or DC voltage exceeding
22 V.
• To prevent an electric shock, do not exceed the lower of the
ratings shown on the instrument and test leads.
± 60 V DC. Attempting to measure voltages exceed-
3
4
5
6
7
±
8
• Use only the specified test leads and cables. Using a non-specified
cable may result in incorrect measurements due to poor connection or other reasons.
• To ensure certified measurement accuracy, allow at least 30 min-
utes warm-up. After warm-up, be sure to execute self-calibration.
See Section 4.9 Self-Calibration (P. 65).
• The input circuitry includes a protective fuse. Measurement is not
possible when the fuse is blown.
• This instrument internally stores (backs up) all settings (except
memory function and measurement values), such as measurement range, comparator settings and etc., but only when no operation is performed for a certain time. Therefore, to preserve settings,
do not turn the power off for a short time (about five seconds) after
changing a setting. However, measurement settings made through
the RS-232C or GP-IB interface and measurement settings loaded
LOAD signals of the EXT I/O connector are not memorized.
by
9
10
11
Page 12
6
HIOKI 3561A981-10
Operating Precautions
Before Connecting and Powering On
• Before turning the instrument on, make sure the supply voltage
matches that indicated on the its power connector. Connection
to an improper supply voltage may damage the instrument and
present an electrical hazard.
• To avoid electrical accidents and to maintain the safety
specifications of this instrument, connect the power cord
provided only to a 3-contact (two-conductor + ground) outlet.
To suppress noise, the instrument needs to be set to match the
frequency of the power source. Before operating, set the instrument
to the frequency of your commercial power. If the supply frequency is
not set properly, measurements will be unstable.
See Section 2.5 Selecting the Line Frequency (P. 22).
Make sure the power is turned off before connecting or disconnecting
the power cord.
Handling the Instrument
• Never modify the instrument. Only Hioki service engineers
should disassemble or repair the instrument. Failure to observe
these precautions may result in fire, electric shock, or injury.
• The GP-IB connector location is covered by a blank panel on
the Model 3561. To avoid electric shock, do not remove the
blank panel.
Blank Panel
* This model is the 3561 BATTERY
HiTESTER.
• To avoid damage to the instrument, protect it from physical shock
when transporting and handling. Be especially careful to avoid
physical shock from dropping.
• Do not apply heavy downward pressure with the stand extended.
The stand could be damaged.
This instrument may cause interference if used in residential areas. Such
use must be avoided unless the user takes special measures to reduce
electromagnetic emissions to prevent interference to the reception of radio
and television broadcasts.
Page 13
Operating Precautions
HIOKI 3561A981-10
Handling the Test Leads and Cables
• To avoid breaking the test leads and cables, do not bend or pull them.
• Avoid stepping on or pinching cables, which could damage the cable
insulation.
• To avoid equipment failure, do not disconnect the communications
cable while communications are in progress.
• Use a common ground for both the instrument and the computer.
Using different ground circuits will result in a potential difference
between the instrument's ground and the computer's ground. If the
communications cable is connected while such a potential difference
exists, it may result in equipment malfunction or failure.
• Before connecting or disconnecting any the communications cable,
always turn off the instrument and the computer. Failure to do so
could result in equipment malfunction or damage.
• After connecting the communications cable, tighten the screws on
the connector securely. Failure to secure the connector could result
in equipment malfunction or damage.
7
2
3
4
5
6
7
8
9
10
11
Page 14
8
HIOKI 3561A981-10
Operating Precautions
Page 15
1.1 Product Overview
HIOKI 3561A981-10
OverviewChapter 1
1.1Product Overview
9
1
Overview
2
The Model 3561 and 3561-01 BATTERY HiTESTERs measure battery
internal resistance using a four-terminal, 1-kHz AC method, while
simultaneously measuring DC voltage (electromotive force [emf]). The
high-precision, fast measurement performance and extensive
interface capabilities make these models ideal for incorporating into
battery testing production lines.
3
4
5
6
7
8
9
10
11
Page 16
10
HIOKI 3561A981-10
1.2 Features
1.2Features
Simultaneously Measures Battery Internal Resistance
and Voltage
The four-terminal AC method measures resistance and DC voltage
simultaneously, so battery internal resistance and emf are measured
and judged at once.
High-Precision Measurements
The instrument provides high-resolution resistance (0.01 m) and
voltage measurements (0.1 mV). High precision (
ensures accurate voltage measurements.
± 0.01% rdg.)
High-Speed Measurements
Simultaneous resistance and voltage measurements can be
performed as fast as once every 10 ms.
Comparator Functions
Resistance and voltage measurement values are judged in three
categories (Hi, IN and Lo), with results clearly displayed. A comparator
judgment beeper also provides distinct sounds to indicate pass/fail
judgments and to facilitate correct recognition of judgment results.
Statistical Calculation Functions
Maximum, minimum and average measurement values, standard
deviation, process capability indices and other values can be
automatically calculated for applications such as production
management. Calculation results can also be applied as comparator
setting values.
Measurement Value Memory Function
The instrument includes a Memory function and storage capacity for
up to 400 pairs of measurement values. When making many
sequential measurements at high speed and sending the measured
values to a PC after each measurement, the time to switch test objects
can become unsatisfactorily long. The Memory function can avoid the
slow-down by sending stored measurements in batches during idle
times.
EXT I/O Interface
EXT I/O and RS-232C interfaces are equipped as standard,
supporting transfer rates up to 38,400 bps. Model 3561-01 also
supports GP-IB.
Printing Measurement Values and Statistical Results
Connect the printer to print measurement values and statistical
calculation results.
Page 17
1.3 Names and Functions of Parts
HIOKI 3561A981-10
1.3Names and Functions of Parts
11
Front Panel
Input Terminals (INPUT)
Connect the optional test leads.
Connections:
See Section 2.3 Connecting the Optional
Test Leads (p.19).
SOURCE-H
SOURCE-L
SENSE-H
SENSE-L
Main Display
(p.12)
Operating Keys
(p.13)
1
Overview
2
3
4
5
6
Sub Display
(p.12)
POWER Switch
Turns the instrument on and off.
: Power OFF
: Power ON
See Section 2.4 Turning the Power On and Off (p.21).
7
8
9
10
11
Page 18
12
HIOKI 3561A981-10
1.3 Names and Functions of Parts
Main Display
The current measurement mode is indicated while measuring, and the setting item is displayed
while making settings.
(Upper row)
AUTO
EX.FAST, FAST, MED, SLOW
0 ADJ
MEM
EXT TRIG
Lit when measuring
voltage
Indicates percentage
units during relative
value comparator
operation
Shows measured
value or setting item.
Lit when measuring with Auto-Ranging.
The selected Sampling Rate is lit.
Lit when measuring in a range for
which Zero-Adjustment has been
performed.
Lit when the Memory function is
enabled.
Lit when the External Trigger function
is enabled.
Units of displayed measurement
V
m
(Lower row)
V
STAT
AVG
LOCK
REMOTE
Unit of voltage
Unit of resistance (lit when the
3 range is selected)
Unit of resistance (lit when the
300 m range is selected)
Lit when the V (Resistance and
Voltage measurement) mode is
selected.
Lit when the Statistical Calculation
function is enabled.
Lit when measuring with the
Averaging setting enabled.
Lit when the keys are locked.
Lit during communications.
Shows Comparator Decision Result.
Indicates that the measured value is
Hi
above the upper threshold.
Indicates that the measured value is
IN
between the upper and lower thresholds.
Indicates that the measured value is
Lo
below the lower threshold.
Sub Display
Upper and lower thresholds and other settings are displayed (when set).
While measuring,
indicates the Comparator
function is enabled.
Indicates Voltage
measurement mode
Indicates percentage
units during relative value
comparator operation
HIGH, LOW
REF, %
V
Indicates that absolute value comparator operation is enabled
(while measuring), and also when setting.
Indicates that relative value comparator operation is enabled
(while measuring), and also when setting.
Indicates voltage measurement units.
Page 19
Operating Keys
HIOKI 3561A981-10
To use a function marked on a
key, just press the key.
To use a function printed
under a key (blue letter),
press the SHIFT key first (and
confirm the SHIFT lamp is lit),
and then the key.
1.3 Names and Functions of Parts
Use as ten-keys to
enter numerical
values.
(Numerical values
can be used with
the RANGE key.)
13
1
Overview
2
3
SHIFT Lamp
[ ]: Enabled after pressing the SHIFT key (SHIFT lamp lit).
Operating
Key
V/ / V
[0 ADJ]
LOAD
[SAVE]
TRIG
[INT/EXT ]
VIEW
STAT
[DELAY]
SMPL
[AVG]
COMP
[SET]
LOCAL
Description
Selects Measurement mode.
(Resistance and voltage measurement,
Resistance measurement or Voltage
measurement)
Executes Zero-Adjustment.
Loads a saved measurement
configuration (Panel settings).
Saves the current measurement
configuration (Panel settings).
Executes a Manual Trigger event.
Selects internal/external triggering.
Switches the view mode of the V mode.
Displays and sets Statistical Calculation
results.
Sets the Trigger Delay.
Selects the Sampling Rate.
Activates Averaging function settings.
Switches the Comparator function on
and off.
Activates Comparator function setting.
Cancels remote control (RMT) and reenables key operations.
Operating
Key
PRINT
AUTO
[LOCK]
ENTER
[MENU]
RANGE
SHIFT
Description
Sends measurement values and
statistical calculation results to the printer.
Switches between Auto and Manual
range selection.
Switches the Key-Lock function on and off.
Applies settings.
Selects various operating functions and
settings.
Up/Down:
Changes setting value or numerical
value, and range selection.
Left/Right:
Moves the setting item or digit.
• Enables the functions of the operating
keys marked in blue.
The lamp is lit when the SHIFT state is
active.
• Cancels settings in various setting
displays. (Returns to the Measurement
display without applying settings.)
However, this does not apply to Menu
display.
However, from a menu item display,
changed settings are not canceled, but
accepted as the display returns to
measurement display (except after
Zero-Adjustment clear or resetting).
4
5
6
7
8
9
10
11
Page 20
14
HIOKI 3561A981-10
1.3 Names and Functions of Parts
Rear Panel
Power Inlet
Connect the supplied power cord here.
See Section 2.2 Connecting the Power Cord
(p.18).
EXT I/O Connector
Connect here to use the EXT
I/O interface.
GP-IB Connector(Model 3561-01 only)
Connect here to use the GP-IB interface.
7.3.1 Attaching the Connector (p.89).
RS-232C Connector
Connection for the printer or RS232C interface.
See Section 7.3.1 Attaching the
Connector (p.89).
Side View
* The illustration shows the Model 3561-01 BATTERY HiTESTER (GP-IB version).
The GP-IB connector location is covered by a blank panel on the
Model 3561. To avoid electric shock, do not remove the blank
panel.
Stand
Can be opened to tilt
the front panel upwards.
Do not apply heavy downward pressure with the stand extended. The
stand could be damaged.
Page 21
1.4 Menu Display Sequence (SHIFT ENTER)
HIOKI 3561A981-10
1.4Menu Display Sequence (SHIFT ENTER)
15
Various auxiliary settings can be performed from the menu item displays.
(SHIFT Lamp lit)
The Menu display appears.
(Main Display)
Zero-Adjustment Clear
display (p.30)
Measurement Value Output
function setting display
(p.66)
Interface Selection
EOC-signal setup
The up/down RANGE key changes the setting
shown on the Sub Display.
Self-Calibration setting
display (p.91)
ERR Output Selection
display (p.76)
1
Overview
2
3
4
display (p.65)
5
6
display (p.76)
7
Key Beeper setting
display (p.67)
Settings on the menu item displays are applied and saved internally
when changed.
Line Frequency
setting display
(p.22)
8
Reset display
(p.68)
9
Pressing this key returns to the
previous item display.
10
11
Page 22
16
HIOKI 3561A981-10
1.5 Measurement Flowchart
1.5Measurement Flowchart
The basic measurement process flow is as follows:
Measurement Preparations
Connecting the power cord (p.18)
Connecting the test leads (p.19)
Turning the power on (p.21)
Selecting the line frequency (p.22)
Instrument’s Settings
Selecting measurement mode (p.27)
Selecting measurement range (p.28)
Selecting sampling rate (p.30)
Zero-Adjustment
Short the test leads together (p.31)
Executing zero-adjustment
Measurement Start
Connect the test leads to a test object.
Read the measured value (p.33)
For details about the functions that can be applied to measurement
values such as comparator, trigger and averaging functions, refer to
Chapter 4 Applied Measurement (p.35).
Page 23
Measurement
HIOKI 3561A981-10
17
2.1 Preparation Flowchart
Preparations
Chapter 2
2.1Preparation Flowchart
This procedure describes instrument preparations such as making connections and turning power on.
(p.19)(p.18)
1
Rear Panel
2
Front Panel
3
5
(p.21)
4
1
2
Measurement Preparations
3
4
5
6
Connecting the power cord.
1
Connect the test leads to the instrument.
2
Connect the EXT I/O connector and
3
interface connector.
Turn the power on.
4
Set measurement settings.
5
Start measurement.
6
(p.18)
7
(p.19)
8
(p.89)
9
(p.21)
10
(p.23)
Verify that the instrument’s line frequency is correctly set when using
it for the first time and after initialization following repair or
recalibration.
See Section 2.5 Selecting the Line Frequency (p.22).
Page 24
18
HIOKI 3561A981-10
2.2 Connecting the Power Cord
2.2Connecting the Power Cord
To avoid electrical accidents and to maintain the safety
specifications of this instrument, connect the power cord
provided only to a 3-contact (two-conductor + ground) outlet.
To avoid damaging the power cord, grasp the plug, not the cord, when
unplugging it from the power outlet.
To suppress noise, the instrument needs to be set to match the line
frequency.
Before operating, set the instrument to the frequency of your
commercial power. If the supply frequency is not set properly,
measurements will be unstable.
See Section 2.5 Selecting the Line Frequency (p.22).
Make sure the power is turned off before connecting or disconnecting
the power cord.
Rear Panel
1. Confirm that the instrument's Power
switch is OFF.
2. Check that the power supply voltage
is correct, and connect the power
cord to the power inlet socket on the
rear of the instrument.
3. Plug the power cord into the AC
outlet.
Page 25
2.3 Connecting the Optional Test Leads
HIOKI 3561A981-10
2.3Connecting the Optional Test Leads
19
Red Lead
Test leads are not included as standard accessories with the
instrument, so the appropriate options need to be purchased
separately or constructed according to the user’s application
requirements. To construct custom test leads, refer to Appendix 1
Precautions for Making Custom Test Leads (p.171). The resistance
measurement terminals on this instrument consist of four separate
banana jacks.
See Appendix 1 Precautions for Making Custom Test Leads (p.171).
1. Confirm that the instrument's Power
switch is OFF.
2. Connect four-terminal test leads such
as the L2107 CLIP TYPE LEAD to
INPUT A.
1
2
Measurement Preparations
3
4
5
6
Plug the mark on the red lead into
the red marked jack on the
instrument,
and plug the mark on the black lead
into the black marked jack on the
Black Lead
Example: Optional model L2107 CLIP TYPE LEAD
instrument.
7
8
9
10
Page 26
20
HIOKI 3561A981-10
2.3 Connecting the Optional Test Leads
About Test Leads
(Example: Model L2107 CLIP TYPE LEAD)
The side with “V” mark is SENSE.
Red
SENSE
SOURCE
SENSE
SOURCE
Black
When clipping a thin line
(Clip the line at the tip,
serrated part of the jaws.)
Red
SENSE
SOURCE
SENSE
SOURCE
Black
When clipping a thick line
(Clip the line at the deep,
non-serrated part of the jaws.)
Page 27
2.4 Turning the Power On and Off
HIOKI 3561A981-10
2.4Turning the Power On and Off
21
Before turning the instrument on, make sure the supply voltage
matches that indicated on the its power connector. Connection
to an improper supply voltage may damage the instrument and
present an electrical hazard.
• The measurement setting state is the same as when the power
• However, measurement settings made through the RS-232C or
• Before starting to measure, allow 30 minutes for warm-up.
Turning the Power On
Turn the POWER switch ON.
was previously turned off (backup).
To preserve changes to settings, wait a short time (about five
seconds) after changing a setting before turning power off.
GP-IB interface and measurement settings loaded by
signals of the EXT I/O connector are not memorized.
After warm-up, be sure to perform a self-calibration.
See Section 4.9 Self-Calibration (p.65).
(Main Display)
Model name
LOAD
1
2
Measurement Preparations
3
4
5
6
Power ON
Turning the Power Off
(Main Display)
Software version
(Sub Display)
Line frequency
Interface
The measurement display appears.
Turn the POWER switch OFF.
7
8
9
10
Power OFF
Page 28
22
HIOKI 3561A981-10
2.5 Selecting the Line Frequency
2.5Selecting the Line Frequency
Verify that the instrument’s line frequency is correctly set when using it for the first time and after
initialization following repair or recalibration.
1
2
3
(SHIFT Lamp lit)
The Menu display appears.
(Main Display)
(Sub Display)
Select the Line Frequency setting display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display) flashing
Select the frequency of the AC mains supply being used.
4
(Main Display)
(Sub Display) flashing
50.......... 50 Hz
60.......... 60 Hz
Applies settings and returns to the Measurement display.
To suppress noise, the instrument needs to be set to match the
frequency of the power source.
Before operating, set the instrument to the frequency of your
commercial power. If the supply frequency is not set properly,
measurements will be unstable.
Page 29
3.1 Pre-Operation Inspection
HIOKI 3561A981-10
MeasurementChapter 3
23
1
Before starting measurement, please read Operating Precautions (p.4) and Chapter
2 Measurement Preparations (p.17).
• To avoid electrical shock, be careful to avoid shorting live lines with
the test leads.
• The maximum rated voltage between input terminals and ground is
60 V DC. Attempting to measure voltages exceeding 60 V with respect
to ground could damage the instrument and result in personal injury.
To avoid injury or damage to the instrument, do not attempt to measure
AC voltage and AC current, or DC voltage exceeding
± 22 V.
±
3.1Pre-Operation Inspection
2
3
Measurement
4
5
6
7
Before using the instrument, perform the following inspection to
ensure that it is operating properly.
Check PointCheck Contents
Instrument Chassis
(both front and rear
panels)
Test Leads and Power
Cord
Good Test Sample• Measures as good and displays the correct
Bad Test Sample• Measures as bad and displays the correct
If the inspection reveals a defect, stop using the instrument and
contact your dealer or Hioki representative.
• No damage or cracks
• No internal circuitry is exposed
• Metal parts that should be insulated are not
exposed
measurement value
measurement value
8
9
10
Page 30
24
HIOKI 3561A981-10
3.2 Basic Measurement Example
3.2Basic Measurement Example
The following example describes the measurement process.
Example: Measuring resistance and voltage of a 30 m lithium-ion battery
Required items:Lithium-ion battery (30 m)
Test leads: Model 9770 PIN TYPE LEAD are used here.
Measurement
conditions:
Preparations
Measurement mode ............. V (Resistance and Voltage measurement)
Range .................................. 300 m
Sampling rage...................... SLOW
Zero adjustment ................... Enabled
1
2
Connect the power cord.
See Section 2.2 Connecting the Power Cord (p.18).
1
Connect the test leads.
See Section 2.3 Connecting the Optional Test Leads (p.19).
Red Lead
Black Lead
2
3
3
Example: Model 9770
Turn the power on.
See Section 2.4 Turning the Power On and Off (p.21).
See Section 2.5 Selecting the Line Frequency (p.22).
Page 31
Instrument Settings
HIOKI 3561A981-10
25
3.2 Basic Measurement Example
4
5
Confirm the SHIFT lamp is not lit.
If this is lit, press the SHIFT key to turn it off.
SHIFT not lit
Select the Resistance Measurement mode.
(Here, resistance and voltage measurement is selected.)
See Section 3.3 Selecting Measurement Mode (p.27).
V lit
The measurement mode changes each time
you press this key. Select to measure only
resistance, or V to measure only voltage.
V
1
2
3
Measurement
4
5
V
6
6
7
Set the measurement range. (Here, 300 m range is selected.)
See Section 3.4 Setting Measurement Range (p.28).
m lit
Selects the 3 range ( lit)
Selects the 300 m range (m lit)
Set the sampling rate. (Here, SLOW is selected.)
See Section 3.5 Setting Sampling Rate (p.30).
SLOW lit
The sampling rate changes each time you
press this key.
EX.FAST
FAST
MED
7
8
9
10
SLOW
Page 32
26
HIOKI 3561A981-10
3.2 Basic Measurement Example
Zero-Adjustment
8
9
Short the test leads together.
Proper Zero-Adjustment is not possible with incorrect wiring.
See Section 3.6 Zero-Adjust Function (p.30).
Example: Model 9770 PIN TYPE LEAD
Internal
conductor
Bring the pins into contact at 3 points.
• Internal conductor and internal conductor
External
conductor
(SHIFT Lamp lit)
Execute Zero-Adjust.
After zero-adjustment, the display returns to the measurement mode.
• Internal conductor and external conductor
• External conductor and external conductor
Model 9770
“Err.02” appears if Zero-Adjustment fails. Verify that the test lead tips are properly shorted,
and try zero-adjustment again.
Measurement
10
11
Connect the test leads to a battery.
Read the measured resistance and voltage.
0ADJ lit
Measured Resistance
Measured Voltage
See Section 3.7 Displaying Measurement Results (p.33).
See Section 9.3 Error Display (p.170).
Page 33
3.3 Selecting Measurement Mode
HIOKI 3561A981-10
3.3Selecting Measurement Mode
27
Select the measurement mode from V (both resistance and voltage measurement), (resistance
measurement only) or V (voltage measurement only).
1
2
“V” indicates the
V mode is selected
Confirm the SHIFT lamp is not lit.
If this is lit, press the SHIFT key to turn it off.
Switches the displayed measurement mode.
Each key-press switches the measurement mode.
V mode
(Resistance and Voltage measurement)
“” indicates the 3 range is
selected
” indicates the 300 m range
“m
is selected
SHIFT not lit
1
2
3
Measurement
4
5
6
mode
(Resistance measurement)
V mode
(Voltage measurement)
The fastest measurements are provided by selecting the or V
mode when measuring resistance or voltage, respectively.
See Section Sampling Time (p.164).
7
8
9
10
Page 34
28
HIOKI 3561A981-10
3.4 Setting Measurement Range
3.4Setting Measurement Range
Select the resistance measurement range from 3 (“” indicator lit) or 300 m (“m” indicator lit).
The auto-ranging function can be enabled to automatically determine the most suitable range. The
voltage measurement range is fixed at 20 V.
3.4.1Manual Range Setting
1
2
When auto-ranging is enabled, pressing this disables it (AUTO not lit)
and enables manual range selection.
Select the range to use.
When the 3 range is selected
lit
3 range
When the 300 m
selected
range is
300 m
range
m lit
Page 35
3.4.2Auto-Ranging
HIOKI 3561A981-10
29
3.4 Setting Measurement Range
Switching from
Auto
-ranging back to
Manual
range selection
When manual range selection is enabled, pressing this enables auto-
ranging. The most suitable measurement range is then selected
automatically.
AUTO lit
Press the
• Voltage measurement has only one range. The range cannot be
changed.
• Depending on the state of the test object, auto-ranging may be
unstable. In this case, select the range manually, or increase the
Delay time.
See Section 3.4.1 Manual Range Setting (p.28).
• Auto-ranging is not available when Comparator or Memory
functions are enabled (ON).
• Refer to Chapter 8 Specifications (p.163) for details about accuracy.
AUTO key again. The range can now be changed manually.
1
2
3
Measurement
4
5
6
7
8
RangeDisplayed Values
Resistance Measurement Mode
Measured CurrentOpen-Terminal Voltage
300 m-10.00 to 310.00 m10 mA ± 10%7 Vpeak.
3 -0.1000 to 3.1000 1 mA ± 10%7 Vpeak.
20 V-19.9999 to 19.9999 V--7 Vpeak.
9
10
Page 36
30
HIOKI 3561A981-10
3.5 Setting Sampling Rate
3.5Setting Sampling Rate
The sampling rate can be selected from EX.FAST, FAST, MEDIUM
and SLOW. Slower sampling rates generally provide greater
measurement precision.
Selects the sampling rate
EX.FASTFASTMEDIUMSLOW
• Measurements are especially susceptible to interference from the
environment when EX.FAST is selected, so countermeasures
such as shielding or twisting of test leads, cables and wiring
around the test object may be necessary.
• When SLOW sampling is selected, self-calibration is executed
during each measurement. At other sampling rates, self-calibration
is executed manually or automatically every 30 minutes.
See Section 4.9 Self-Calibration (p.65).
• Refer to the specifications for details of sampling rates.
See Section Sampling Time (p.164).
3.6Zero-Adjust Function
Execute zero adjustment before measuring to nullify any residual
offset voltage from the instrument or measurement environment.
Measurement accuracy specifications are applicable after zero
adjustment. Zero adjustment can also be executed by the 0ADJ
terminal of the EXT I/O connector.
See Section 5.2 Signal Descriptions (p.72).
3.6.1Wiring Method for Zero-Adjustment
Before executing zero adjustment, connect the test leads (probes) as
follows:
1. Connect SENSE-H to SENSE-L.
2. Connect SOURCE-H to SOURCE-L.
3. Connect the joined SENSE and SOURCE leads together as shown
below.
SENSE-HSENSE-L
SOURCE-HSOURCE-L
Connection
Connection
Page 37
3.6.2Executing Zero-Adjustment
HIOKI 3561A981-10
31
3.6 Zero-Adjust Function
1
Short the test leads together.
Proper zero adjustment is not possible with incorrect wiring.
Example: Model L2107 CLIP TYPE LEAD
Correct
SENSE
SOURCE
Red
Bring the "V" marks together at the
same position.
SENSE
SOURCE
Black
Incorrect
SENSE
SOURCE
Model 9452 (Option)
Red
Black
Model 9453 (Option)
SOURCE
SENSE
Perform zero
adjustment with the
alligator clips and
lead rods placed as
above.
1
2
3
Measurement
4
5
6
Model 9770 (Option)
Bring the pins into
contact at 3 points.
Model 9771 (Option)
When the resistance measurement value
is displayed as "-----", change the facing
direction.
Let the two points of the pin tip touch the
spring part perpendicularly (be careful not
to short the springs).
Internal
conductor
External conductor
7
8
9
10
Page 38
32
HIOKI 3561A981-10
3.6 Zero-Adjust Function
2
(SHIFT Lamp lit)
Zero-adjust display appears.
After measurement, the measured value of the compensation applied by the zeroadjust function is displayed.
The range of zero adjustment is up to 1,000 dgt.
Clearing Zero-Adjustment
1
(SHIFT Lamp lit)
The Menu display appears.
0ADJ lit
2
If Err02 is displayed
(Main Display)
(Sub Display) flashing
The zero-adjust value is cleared. (0ADJ not lit)
(Main Display)
Indicates that zero adjustment could not be executed, either because
the range to be adjusted exceeds
fault condition exists.
The zero adjust function is canceled, so repeat the operation after
correcting the cause of the error.
• Zero adjustment is limited to ± 1,000 dgt.
• Both resistance (all ranges) and voltage are adjusted together for
all modes.
• Zero-adjustment values are retained even when power is turned off.
• The 0ADJ terminal of the EXT I/O connector also executes zero
adjustment.
See Section 5.2 Signal Descriptions (p.72).
• Zero adjustment is very difficult with the delicate probe tips of the
Model 9455 PIN TYPE LEAD. Refer to 3.6.1 Wiring Method for
Zero-Adjustment (p.30) to use other leads when executing zero
adjustment.
± 1,000 dgt, or a measurement
Page 39
3.7 Displaying Measurement Results
HIOKI 3561A981-10
3.7Displaying Measurement Results
33
In the V mode, resistance measurements appear on the upper
display, and voltage measurements appear on the lower display.
Measured Resistance
Measured Voltage
In the mode, resistance measurements appear on the upper display.
Measured Resistance
1
2
3
Measurement
4
5
6
In the V mode, voltage measurements appear on the upper display.
Measured Voltage
7
8
9
10
Page 40
34
HIOKI 3561A981-10
3.7 Displaying Measurement Results
3.7.1Measurement Fault Detection
If a measurement does not execute properly, a measurement fault
“- - - - -” is indicated on the display.
In addition, a measurement fault signal (ERR) is output at the EXT I/O
connector.
See Section 5.2.4 ERR Output (p.75).
A measurement fault is displayed in the following cases.
• When a test lead is not connected to the test object
• When the resistance of the measured object is over-range
Example: Attempting to measure 30 with the 300 m range
selected.
• If any of the following is open, or has a bad connection: SOURCE-
H, SOURCE-L, SENSE-H, SENSE-L
• When the resistance between SOURCE-H and SOURCE-L is 50
or more in the 300 m range (or 500 or more in the 3 range)
• When the resistance between SENSE-H and SENSE-L is greater
than about 20 (However, if the capacitance of the test leads is 1
nF or higher, the measurement fault may not be detected.)
• When a bad contact results from damage, excessive wear or
impurities on the test leads.
• If the circuit protection fuse is blown
See Section 9.1 Troubleshooting (p.169).
3.7.2Overflow Display
Overflow is indicated by “OF” or “-OF” on the display, caused by one of
the following:
DisplayCondition
OF
-OF
• The measured value exceeds the limit of the current
measurement range
• When the result of relative value calculation is larger
than +99.999%.
• The measured value is below the limit of the current
measurement range
• When the result of relative value calculation is smaller
than -99.999%.
Page 41
Applied
HIOKI 3561A981-10
35
Measurement
This chapter describes advanced operations employing the
Comparator, Statistical Calculation and Memory functions.
Judge measurement values
against specified thresholds
Measure when trigger events occur Trigger Function
Output averaged measurement
values
Display the results of calculation
expressions applied to
measurement values
Store measurement valuesMemory Function
Lock the keysKey-Lock Function
Chapter 4
Comparator Function
Averaging Function
Statistical Calculation
Functions
(36 page)
(53 page)
(55 page)
(56 page)
(60 page)
(62 page)
1
2
3
4
Applied Measurement
5
6
Save measurement configurations Panel Save Function
Load saved measurement
configurations
Increase measurement precisionSelf-Calibration
Output measurement values via the
RS-232C interface according to
trigger input timing
Enable/disable key-press beepsKey Beeper Setting
Re-initialize the instrumentReset Function
Panel Load Function
Measurement Value
Output Function
(63 page)
(64 page)
(65 page)
(66 page)
(67 page)
(68 page)
7
8
9
10
Page 42
36
HIOKI 3561A981-10
4.1 Comparator Function
4.1Comparator Function
The comparator function compares measured values to preset upper
and lower thresholds, judges the measurements according to their
relative levels within the preset range, and indicates the results of the
comparisons.
Comparator thresholds can be set either by specifying upper and
lower thresholds, or by specifying a reference value and tolerance.
Comparator results can be indicated by the Hi, IN and Lo LEDs,
beeper sound and signal output at the EXT I/O connector.
See Section Chapter 5 External Control (p.71).
The comparator setting process flow is as follows:
Display the comparator settings
1
Set the comparator judgment beeper
2
Select the comparator execution mode
3
(Auto or Manual/External)
Select resistance or voltage measurement
4
Select the comparison method
5
(absolute or relative value) for the comparator
Specify the upper and lower thresholds
6
(or reference value and tolerance)
Apply your comparator settings
7
Enable the Comparator function
8
Page 43
37
HIOKI 3561A981-10
4.1 Comparator Function
4.1.1Comparator Setting Example 1 (Upper and Lower
Threshold Judgment)
This example describes the comparator setting method.
Example:
Set the upper and lower thresholds for resistance and voltage in the V
mode (300 m range), and indicate whether the measurement value
exceeds the upper or lower thresholds by sounding the beeper.
Resistance: Upper threshold value 150.00 m, Lower threshold value 100.00 m
Voltage: Upper threshold value 15.2000 V, Lower threshold value 15.0000 V
1
Confirm that the Comparator function is OFF.
First make sure the Comparator function is disabled. Settings cannot be changed while
the Comparator function is enabled. Press the COMP key, if necessary, to disable the
Comparator function.
COMP not lit
1
2
3
4
Applied Measurement
5
6
2
3
4
Select the V measurement mode.
V lit
Select the measurement range (for this example, the 300 m range).
m lit
Selects the 3 range ( lit)
Selects the 300 m range (m lit)
The Comparator setting display appears.
oFF flashing
7
8
9
10
Page 44
38
HIOKI 3561A981-10
4.1 Comparator Function
5
6
Set the comparator judgment beeper (for this example, select HL).
HL flashing
oFF........ no beeps sound
In ...........beeps continuously (when measurements are IN)
HL.......... beeps repeatedly (when measurements are Hi or Lo)
btH1.......beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2.......beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
Press so that the indicated position blinks, and select the comparator execution
mode (for this example, Auto).
A flashing
7
8
A............ Auto Comparator (default setting)
E ............ Manual Comparator
Press so that the indicated position blinks, and select resistance.
r flashing
r .............Resistance
u ............Voltage
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, HIGH/LOW).
HIGH & LOW flashing
HIGH, LOW ..... Compare by upper and lower thresholds (default setting)
REF, %............. Compare by reference value and tolerance
Page 45
4.1 Comparator Function
HIOKI 3561A981-10
39
9
Or
ten-keys
Switch to the upper/lower threshold setting display, and specify the thresholds.
For this example,
Upper Threshold: 150 m
Upper Threshold: 100 m
Using the RANGE keys:Using the ten-keys:
Select a digit to change by
moving the blinking location, then
select the new numerical value.
Select a digit
Select numerical value
To enter the current measurement as the setting value: AUTO key
To enter the result of statistical calculation as the setting value: STAT key
See Section 4.1.6 Upper and Lower Thresholds Setting (by Reference Value and
Tolerance) (p.48).
Press the numeric keys
corresponding to the digits
to be entered.
1
2
3
4
Applied Measurement
5
6
10
11
Press so that the indicated position blinks, and select voltage.
7
u flashing
8
r............. Resistance
u ............ Voltage
9
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, HIGH/LOW).
10
HIGH & LOW flashing
HIGH, LOW..... Compare by upper and lower thresholds (default setting)
REF, % ............ Compare by reference value and tolerance
Page 46
40
HIOKI 3561A981-10
4.1 Comparator Function
12
13
Or
ten-keys
Switch to the upper/lower threshold setting display, and specify the thresholds.
For this example,
Upper Threshold: 15.2 V
Upper Threshold: 15 V
Applies setting and returns to the Measurement display.
The comparator function is enabled.
COMP lit
To cancel the settings: SHIFT key
14
Connect a test object and judge the measured value.
Measured Resistance
Judgment Result
Measured Voltage
In the
V mode, you can verify comparator settings by pressing the VIEW key.
See Section 4.1.9 Switching Between Measurement Value and Comparator
Setting Displays (p.52).
Upper Threshold Value < Measured Value
Lower Threshold Value
Measured Value < Lower Threshold Value
Measured Value Upper Threshold Value
The upper and lower thresholds are saved as the displayed counts
(independent of measurement mode and range). Therefore, changing the
measurement mode or range results in the same display counts representing
different absolute values.
Example:
To specify the lower threshold as 150 m in the 300 m range, enter “15000”.
Switching to the 3 range after making this setting changes the lower threshold
to 1.5 .
Page 47
4.1 Comparator Function
HIOKI 3561A981-10
4.1.2Comparator Setting Example 2 (Reference Value
and Tolerance Judgment)
41
This example describes the comparator setting method.
Example:
Set a reference value and tolerance in the V mode (3 range), and set
the beeper to sound while measured values are within tolerance.
Resistance: Reference value 1.5 , Tolerance 5%
Voltage: Reference value 4.2 V, Tolerance 0.5%
1
Confirm that the Comparator function is OFF.
First make sure the Comparator function is disabled. Settings cannot be changed while
the Comparator function is enabled. Press the COMP key, if necessary, to disable the
Comparator function.
COMP not lit
1
2
3
4
Applied Measurement
5
6
2
3
4
Select the V measurement mode.
V lit
Select the measurement range (for this example, the 3 range).
lit
Selects the 3 range ( lit)
Selects the 300 m range (m lit)
The Comparator setting display appears.
oFF flashing
7
8
9
10
Page 48
42
HIOKI 3561A981-10
4.1 Comparator Function
5
6
Set the comparator judgment beeper (for this example, select In).
In flashing
oFF........ no beeps sound
In ...........beeps continuously (when measurements are IN)
HL.......... beeps repeatedly (when measurements are Hi or Lo)
btH1.......beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2.......beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
Press so that the indicated position blinks, and select the comparator execution
mode (for this example, Auto).
A flashing
7
8
A............ Auto Comparator (default setting)
E ............ Manual Comparator
Press so that the indicated position blinks, and select resistance.
r flashing
r .............Resistance
u ............Voltage
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, REF/%).
REF & % flashing
HIGH, LOW ..... Compare by upper and lower thresholds (default setting)
REF, %............. Compare by reference value and tolerance
Page 49
4.1 Comparator Function
HIOKI 3561A981-10
43
9
Or
ten-keys
Switch to the Ref/% threshold setting display, and specify the thresholds.
Relative (%) calculation expression
Measured resistance - Reference value
Reference value
For this example,
Reference value: 1.5
Tolerance: 5%
Using the RANGE keys:Using the ten-keys:
Select a digit to change by
moving the blinking location, then
select the new numerical value.
Select a digit
Select numerical value
Press the numeric keys
corresponding to the digits
to be entered.
x 100
1
2
3
4
Applied Measurement
5
10
11
To enter the current measurement as the setting value: AUTO key
To enter the result of statistical calculation as the setting value: STAT key
See Section 4.1.6 Upper and Lower Thresholds Setting (by Reference Value and
Tolerance) (p.48).
Press so that the indicated position blinks, and select voltage.
u flashing
r............. Resistance
u ............ Voltage
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, REF/%).
6
7
8
9
10
REF & % flashing
HIGH, LOW..... Compare by upper and lower thresholds (default setting)
REF, % ............ Compare by reference value and tolerance
Page 50
44
HIOKI 3561A981-10
4.1 Comparator Function
12
13
Or
ten-keys
Switch to the Ref/% threshold setting display, and specify the thresholds.
Relative (%) calculation
Measured voltage - Reference value
Reference value
For this example,
Reference value: 4.2 V
Tolerance: 0.5%
Applies setting and returns to the Measurement display.
The comparator function is enabled.
COMP lit
To cancel the settings: SHIFT key
x 100
14
Connect a test object and judge the measured value.
Resistance measurements are displayed as
their relative percentage offset from the
reference value (%)
Judgment Result
Relative
percentag
Voltage measurements are displayed as their relative percentage offset
from the reference value (%)
V mode, you can verify comparator settings by pressing the VIEW key.
In the
See Section 4.1.9 Switching Between Measurement Value and Comparator Setting
Displays (p.52).
Upper Threshold Value of setting range < Measured value
Lower Threshold Value of setting range Measured value Upper
Threshold Value of setting range
Measured value < Lower Threshold Value of setting range
Measured resistance - Reference value
=
Reference value
x 100
Page 51
4.1 Comparator Function
HIOKI 3561A981-10
4.1.3Comparator Judgment Beeper Setting
Four beeper settings are available to audibly indicate comparator judgment results.
45
1
1
2
(SHIFT Lamp lit)
The Comparator setting display appears.
Set the comparator judgment beeper.
(Main Display)
oFF........ no beeps sound
In ........... beeps continuously (when measurements are IN)
HL.......... beeps repeatedly (when measurements are Hi or Lo)
btH1 ...... beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2 ...... beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
• The beeper does not sound when the comparator judgment
beeper setting is disabled (oFF).
• The beeper does not sound when there is no judgment result.
See Section 4.1.8 Comparator Judgment Results (p.51).
2
3
4
Applied Measurement
5
6
7
8
9
10
Page 52
46
HIOKI 3561A981-10
4.1 Comparator Function
4.1.4Comparator Execution Mode Setting
Comparator judgment execution is selected by setting the auto or manual/external comparator
mode. Comparator judgment can be enabled and disabled by EXT I/O signals.
Refer to 5.2.2 Input Signals (p.73).
1
2
(SHIFT Lamp lit)
The Comparator setting display appears.
Press so that the indicated position blinks, and set the comparator execution
mode.
(Main Display)
A............ Auto comparator (comparator results are always output [default setting])
E ............Manual comparator (comparator results are output only when the MANU EXT I/
O input is enabled [ON])
The auto setting is appropriate for normal use. Use the manual/
external setting when you need to control comparator judgment
timing.
Page 53
4.1 Comparator Function
HIOKI 3561A981-10
4.1.5Comparator Threshold Method Selection
Two methods are available for setting comparator thresholds.
47
1
1
2
(SHIFT Lamp lit)
The Comparator setting display appears.
Press so that the indicated position blinks, and set the comparator threshold
method.
HIGH, LOW..... Compare against specified upper and lower thresholds (default setting
method)
REF, % ............ Compare against upper and lower thresholds internally calculated from a
specified reference value and tolerance
About comparisons based on a reference value and tolerance
When the reference value and tolerance method is selected, thresholds are
calculated as follows:
Upper threshold = reference value
Lower threshold = reference value
Measured values are displayed as a percentage relative to the reference value,
calculated as follows:
Relative value = (measured value - reference value) / reference value
4.1.6Upper and Lower Thresholds Setting
(by Reference Value and Tolerance)
1
2
3
(SHIFT Lamp lit)
The Comparator setting display appears.
Press so that the indicated position blinks, and select resistance or voltage.
r .............Resistance
u ............Voltage
Select the threshold setting display, and enter upper and lower threshold values.
Or
ten-keys
For example,
Upper Threshold: 150 m
Lower Threshold: 100 m
Using the RANGE keys:Using the ten-keys:
Select a digit to change by
moving the blinking location, then
select the new numerical value.
Select a digit
Select numerical value
Press the numeric keys
corresponding to the digits
to be entered.
Page 55
4.1 Comparator Function
HIOKI 3561A981-10
To enter the current measurement as the setting value: AUTO key
The current measurement value is set as the upper or lower
threshold (during upper/lower threshold setting), or as the reference
value (during reference value and tolerance setting). If the
measured value is faulty or
To enter a statistical calculation result as the setting value:
The result of statistical calculation is set as follows:
± OF, it is ignored (not entered).
STAT key
49
1
During upper/lower
threshold setting
During reference value
and tolerance setting
Where “
No setting occurs if statistical calculation is disabled and no
statistical calculation result exists.
See Section 4.4 Statistical Calculation Functions (p.56).
Setting thresholds from the
when the selected (blinking) character is non-numeric.
Threshold and reference values can be set from 0 to 99999 (or
999999 for voltage), and tolerance can be set from 0.000 to
99.999%. Negative values are not settable. Entries using statistical
calculation results that exceed the valid range are restricted to the
range limit.
” represents population standard deviation (
Upper threshold = average value + 3
Lower threshold = average value - 3
Reference value = average value
Tolerance = 3
AUTO and STAT keys is possible only
/ average value X 100%
n
).
2
3
4
Applied Measurement
5
6
7
8
9
10
Page 56
50
HIOKI 3561A981-10
4.1 Comparator Function
4.1.7Enabling and Disabling the Comparator Function
Enables the comparator
COMP lit
When the comparator is enabled, the following key operations are
disabled to avoid inadvertent operations.
When the comparator is enabled, auto-ranging is automatically
disabled.
Page 57
4.1 Comparator Function
HIOKI 3561A981-10
4.1.8Comparator Judgment Results
Resistance and voltage measurements are judged independently.
Both judgment results are indicated on the display.
Judgment OperationThe comparator compares measured values with the preset threshold
values, and judges whether the measurement is within the thresholds.
Resistance and voltage measurements are judged independently.
The absolute value of the measurement is compared to the upper and
lower thresholds.
Upper Threshold Value < Measured Value
Lower Threshold Value
Measured Value < Lower Threshold Value
Measured Value Upper Threshold Value
51
1
2
3
4
Applied Measurement
AND Judgment
Output
Measurement fault values are judged as follows:
DisplayJudgment
- - - - -
OF
-OF
Judgment results (Hi, IN or Lo for both resistance and voltage) are
output to EXT I/O connectors. Also, to facilitate application of
judgment results, an AND output terminal indicates when both
resistance and voltage are IN (within the threshold range).
See Section 5.2.3 Output Signals (p.74).
With the relative value comparison method (thresholds defined by a
reference value and tolerance), the upper and lower thresholds are
calculated internally for comparison with measurements. Therefore,
even if a relative display value is equal to a judgment threshold
(tolerance limit), it may be judged Hi or Lo.
No judgment
Hi (exceeds the upper threshold)
Lo (less than the lower threshold)
5
6
7
8
9
10
Page 58
52
HIOKI 3561A981-10
4.1 Comparator Function
4.1.9Switching Between Measurement Value and
Comparator Setting Displays
In V mode, both resistance and voltage measurement values are
displayed.
Although comparator setting values are not normally displayed when
the comparator is enabled, they can be displayed for confirmation by
the display switching function.
Press this key to switch the display between measurement values and
comparator setting values.
Resistance and voltage measurement display
(Shows resistance and voltage measurement values simultaneously)
Resistance measurement and comparator display
(Shows resistance measurement and resistance
comparator setting values)
Voltage measurement and comparator display
(Shows voltage measurement and voltage
comparator setting values)
Measurement display switching is available only with the comparator
enabled, and in the V mode.
Use it to confirm comparator setting values.
Page 59
4.2Trigger Function
HIOKI 3561A981-10
4.2.1Trigger Source Settings
53
4.2 Trigger Function
1
Measurement with
External Triggering
Two trigger sources are available: internal and external.
Internal TriggerTrigger signals are automatically generated internally.
(free-run)
External TriggerTrigger signals are provided externally or manually.
(SHIFT Lamp lit)
Switches the selected trigger source.
EXT TRIG lit
EXT.TRIG
EXT.TRIG not lit........Internal triggering is selected.
An external trigger can be applied in three ways.
• Applying a trigger manually by operating key
Pressing the
• Applying a trigger at the EXT I/O connector
Grounding the
panel causes one measurement.
See Section 5.2.2 Input Signals (p.73).
• Applying a trigger through RS-232C or GP-IB interface
Sending the
causes one measurement.
lit..............External triggering is selected.
TRIG key causes one measurement.
TRIG terminal of the EXT I/O connector on the rear
TRG command via the RS-232C or GP-IB interface
2
3
4
Applied Measurement
5
6
7
8
9
• When Internal triggering is enabled, external input at the EXT I/O
TRIG terminal and the TRG command are ignored.
• The normal state of operation with the front panel controls is
continuous measurement. Setting the trigger source to Internal
enables the free-run condition in which triggering occurs
continuously. When the trigger source is set to External, a
measurement occurs each time an external trigger is applied.
Continuous measurement can be disabled via RS-232C or GP-IB
interface signals, in which case triggering occurs only when
signaled by the external host (PC or PLC).
See Section Triggering System Description (p.136).
10
Page 60
54
HIOKI 3561A981-10
4.2 Trigger Function
4.2.2Trigger Delay Settings
Specify the delay from the moment a trigger is applied to the start of measurement. By using this
function, even when a trigger is applied immediately after connecting a test object, the start of
measurement can be delayed to allow sufficient time for the measurement value to stabilize. Trigger
delay can be set with 1 ms resolution from 0.000 to 9.999 seconds.
1
2
3
(SHIFT Lamp lit)
The Trigger Delay setting display appears.
(Main Display)
(Sub Display)
The current setting blinks.
Select ON.
(Sub Display)
The numerals indicating the trigger delay blink.
(Sub Display)
4
Or ten-keys
5
Set the trigger delay.
Applies setting and returns to the Measurement display.
To cancel the settings: SHIFT key
Disabling the Trigger Delay Function
1
2
3
(SHIFT Lamp lit)
The Trigger Delay setting display appears.
Select OFF.
The Trigger Delay is disabled.
(Sub Display)
Page 61
55
HIOKI 3561A981-10
4.3 Averaging Function
4.3Averaging Function
The Averaging Function averages measurement values for output. This function can minimize
instability of displayed values. The number of samples to average can be set from 2 to 16.
1
1
2
3
4
Or
ten-keys
(SHIFT Lamp lit)
The Averaging Function setting display appears.
(Main Display)
(Sub Display)
The current setting blinks.
Select ON.
(Sub Display)
The number of samples to average setting blinks.
Select the number of samples to average.
2
3
4
Applied Measurement
5
6
7
5
Disabling the Averaging Function
1
2
3
The Average Measurement display appears. (AVG lit)
To cancel the settings: SHIFT key
(SHIFT Lamp lit)
The Averaging Function setting display appears.
Select OFF.
The Averaging Function is disabled. (AVG not lit)
When the internal trigger is used for continuous measurement (freerun), the display shows the moving average. Otherwise, the display
shows the integrating average.
See Section 4.2 Trigger Function (p.53).
8
9
10
(Sub Display)
Page 62
56
n
x
x
n
xnx
22
122
n
xnx
s
16
n
LoHi
Cp
1
6
2
n
xLoHiLoHi
CpK
HIOKI 3561A981-10
4.4 Statistical Calculation Functions
4.4Statistical Calculation Functions
The mean, maximum, minimum, standard deviation of population,
standard deviation of sample and process capability indices are
calculated and displayed for up to 30,000 measurement values.
The calculation formulas are as follows:
Mean
Standard deviation of
population
(=
)
n
Standard deviation of sample(=
Process capability index
(dispersion)
Process capability index
(bias)
• In these formulas, n represents the number of valid data samples.
• Hi and Lo are the upper and lower thresholds of the comparator.
• The process capability indices represent the quality achievement
capability created by a process, which is the breadth of the
dispersion and bias of the process' quality. Generally, depending on
the values of Cp and CpK, process capability is evaluated as
follows:
Cp, CpK>1.33................... Process capability is ideal
1.33 Cp, CpK>1.00........ Process capability is adequate
1.00 Cp, CpK................. Process capability is inadequate
• When only one valid data sample exists, standard deviation of
sample and process capability indices are not displayed.
• When
• The upper limit of Cp and CpK is 99.99. Values of Cp and
CpK>99.99 are displayed as 99.99.
• Negative values of CpK are handled as CpK=0.
• When comparator, range or auto-ranging settings are changed
while statistical data is displayed, the display of Cp and CpK
values changes to “- - . - -”.
• When normal measurement values and relative display values (%)
are mixed, correct calculation results cannot be obtained.
is 0, Cp and CpK are 99.99.
n-1
)
n-1
Page 63
4.4 Statistical Calculation Functions
HIOKI 3561A981-10
Enabling/Disabling the Statistical Calculation Function
57
1
2
(press three times)
3
The Statistical Calculation display appears.
(Main Display)
(Sub Display)
The function enable/disable display appears.
(Sub Display)
Enable or disable the Calculation Function on the Sub Display.
on.......... enables the calculation function (ON).
off ......... disables the calculation function (OFF).
Applies setting and returns to the Measurement display.
To cancel the settings: SHIFT key
• Statistical Calculation function setting (ON, OFF) is not available
when the Comparator is enabled.
• If Statistical Calculation is turned off and then back on without first
clearing calculation results, it resumes calculating from the point
when it was turned off.
• The Statistical Calculation function slows measurements when it is
ON.
1
2
3
4
Applied Measurement
5
6
7
8
Clearing Statistical Calculation Results
1
2
(press once)
3
The Statistical Calculation display appears.
The Clearing screen will appear.
Clears statistical calculation results.
(Main Display)
(Sub Display)
(Sub Display)
9
10
Page 64
58
HIOKI 3561A981-10
4.4 Statistical Calculation Functions
Automatic Clearing of Statistical Calculation Results after Printing
The instrument can be set to automatically clear statistical calculation results after results are
output to the printer.
1
2
3
4
(Press twice)
The Statistical Calculation display appears.
(Main Display)
(Sub Display)
Bring up Auto Clearing After Printing in the Setup screen.
(Sub Display)
Turn Automatic Clearing After Printing on or off.
on.......... Automatically clears statistical calculation results after they are
output to the printer.
off.......... Does not clear the results themselves.
Applies setting and returns to the Measurement display.
To cancel the settings: SHIFT key
Importing Data
Pressing the TRIG key while Statistical Calculation is ON executes
one of the following operations:
• External Trigger: Takes one measurement and performs statistical
calculation on the result
• Internal Trigger: Performs statistical calculation on the value
displayed immediately after pressing
• TRG command executes the same operation.
• Grounding the
the same operation.
TRIG terminal of the EXT I/O connector executes
Confirming Statistical Calculation Results
1
2
The Statistical Calculation display appears.
The indication on the display changes as follows with each keypress.
Page 65
Example: when the V mode is selected
HIOKI 3561A981-10
(not displayed in V mode)
Total data count of resistance
measurement
Mean of resistance measurementMaximum of resistance measurement
59
4.4 Statistical Calculation Functions
Valid data
Mean
Minimum of resistance measurementStandard deviation of population
Minumum
Data Sample
No.
Process capability indices of resistance
measurement
Cp
CpK
of resistance measurement
(not displayed in mode)
Total data count of
voltage measurement
Minimum of voltage
measurement
Process capability indices
of voltage measurement
Standard deviation of
population of voltage
measurement
Mean of voltage
measurement
Maximum
Data Sample
No.
Standard deviation of sample of
resistance measurement
Maximum of voltage
measurement
Standard deviation of
sample of voltage
measurement
1
2
3
4
Applied Measurement
5
6
ON/OFF settingAuto Clearing After Printing setupClear setup
• When a valid data count (measurement fault other than ± OF) is
zero, no calculation result is displayed.
• When only one valid data sample exists, standard deviation of
sample and process capability indices cannot be displayed.
• When comparator, range or auto-ranging settings are changed
while statistical data is displayed, the display of Cp and CpK
values changes to “- - . - -”.
Sending Statistical Calculation Results to the Printer
With the statistical calculation results displayed, press the PRINT key.
The statistical calculation results are output to the optional printer.
See Section Chapter 6 Printing (p.81).
7
8
9
10
Page 66
60
HIOKI 3561A981-10
4.5 Memory Function
4.5Memory Function
The Memory function is only available via communication commands.
When the Memory function is enabled, measurement values are
stored in the instrument’s internal memory according to trigger input
sequence (up to 400 values). Stored data can be downloaded later
upon command.
When measuring using a scanner to switch multiple test objects,
switching time can be quite long if measurement values are
downloaded to the PC after each measurement. Test cycle time can
be minimized by using this function to store measurement values
internally until all channel measurements are finished, at which time
the stored values are downloaded together during the next idle period.
1
2
3
4
Select the RS-232C or GP-IB interface.
See Section 7.3.2 Selecting the Interface (p.91).
Send the command to enable the Memory function.
:MEMory:STATe ON
The MEM indicator lights.
MEM lit
Measurement values are stored.
When a trigger is applied by the TRIG key, TRIG
TRG command, the MEM indicator blinks once and the measured value is
stored.
EXT I/O input signal or
MEM lit
If an external trigger source is selected, one measurement is stored after
each trigger event. In the internal triggering case, the first measurement
value after triggering is stored. Apply a trigger as many times as is
necessary.
Page 67
4.5 Memory Function
HIOKI 3561A981-10
61
5
Send the command to download the data from memory.
:MEMory:DATA?
The stored measurement values are returned in response.
The “END” character is sent as the last line of the data.
To download stored data one measurement at a time, send this command:
:MEMory:DATA? STEP
The instrument sends one stored data object and enters the wait state.
When the instrument receives an “N” from the PC or other device, the next
stored data object is sent.
Repeat until the last data object is downloaded.
When all stored data has been downloaded, the instrument sends an “END”
character.
:MEM:DATA? STEP
1, 290.60E-3, 1.3924E+0
N
2, 290.54E-3, 1.3924E+0
N
3, 290.50E-3, 1.3923E+0
N
4, 290.43E-3, 1.3923E+0
N
5, 290.34E-3, 1.3924E+0
N
END
(sent from PC)
(sent from PC)
(sent from PC)
(sent from PC)
(sent from PC)
1
2
3
4
Applied Measurement
5
6
6
To clear the instrument's memory, send it the following command.
:MEMory:CLEAr
Unless the memory is cleared, measurement data continues to be stored
upon each trigger event.
• The instrument's memory storage capacity is 400 measurements.
Be aware that attempting to store more data (by applying a trigger)
results in nothing further being stored.
• Refer to Chapter 7 RS-232C/GP-IB Interfaces (p.87),
for details about the communication methods and sending and
receiving commands.
• When the Memory function is enabled, auto-ranging is not
available.
• Memory contents are cleared when performing the following
operations:
When enabling the Memory function (off to on)
When changing the measurement range
When changing comparator settings
When sending the
When Reset is executed from the menu display
When sending
When sending :SYSTem:RESet
When turning power on
:Memory:Clear command
RST
7
8
9
10
Page 68
62
HIOKI 3561A981-10
4.6 Key-Lock Function
4.6Key-Lock Function
Executing Key-Lock disables the operating keys on the front of the
instrument. This function can be useful for protecting settings.
(SHIFT Lamp lit)
Enable the Key-Lock function.
LOCK lit
Disabling Key-Lock
• Even if the power supply is interrupted, the Key-Lock function is
not canceled.
• The
(SHIFT Lamp lit)
Disable the Key-Lock function. (LOCK is not lit)
TRIG key remains operational.
When communicating by remote control, the remote control status is
canceled.
Page 69
4.7Panel Save Function
HIOKI 3561A981-10
63
4.7 Panel Save Function
1
2
The current measurement setting state is stored (saved) in nonvolatile memory.
Up to 126 sets of measurement states can be saved.
The measurement settings (state) at the time this function is executed
are saved.
Saved measurement states can be reloaded using the Panel Load
function, described later.
(SHIFT Lamp lit)
The Panel Saving display appears.
(Main Display)
(Sub Display)
The panel number blinks.
Select the panel number to save.
1
2
3
4
Applied Measurement
5
6
3
Saved Items
Or
ten-keys
(Sub Display)
(To save measurement settings as Panel No. 3)
When selecting a saved panel, “USEd” is displayed.
Saves the measurement setting state and returns to the Measurement
display.
To cancel the settings: SHIFT key
• If you select a Panel number that was previously saved and press
ENTER key, the contents are overwritten.
the
• The Key-Lock state can be saved only by the
remote command.
Loads the measurement settings saved by the Panel Save function
from internal non-volatile memory.
1
2
3
Or
ten-keys
The Panel Loading display appears.
(Main Display)
(Sub Display)
The panel number blinks.
Select the panel number to load.
(Sub Display)
(To load measurement settings from Panel No.3)
Loads the measurement setting state and returns to the Measurement
display.
To cancel the settings: SHIFT key
• If an unsaved Panel No. is selected, a warning beep sounds when
you press
• When selecting a Panel No. with the up/down
the numbers of previously saved panels appear.
• Loading can also be executed using the
LOAD0 to LOAD6 pins of the EXT I/O interface.
See Section 5.2.2 Input Signals (p.73).
ENTER key.
RANGE keys, only
TRIG signal and the
Page 71
4.9Self-Calibration
HIOKI 3561A981-10
65
4.9 Self-Calibration
1
The self-calibration function adjusts offset voltage and gain drift of the
instrument’s internal circuitry to improve measurement precision.
The instrument’s measurement accuracy specifications depend on
self-calibration, so it must be executed frequently. In particular, always
execute self-calibration after warm-up and when the ambient
temperature changes by more than 2
setting, self-calibration is executed during every measurement when
SLOW sampling is used.
Self-calibration can be executed by the following two methods:
AutoExecutes self-calibration automatically once every 30
minutes.
ManualSelf-calibration can be executed manually by applying a
CAL input signal (grounding the CAL terminal of the EXT
I/O connector).
(SHIFT Lamp lit)
The Menu display appears.
°C. However, regardless of this
1
2
3
4
Applied Measurement
5
6
2
3
4
The Self-Calibration setting display appears.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display)
The current setting blinks.
Select Auto or Manual on the Sub Display.
AUto ..... Auto self-calibration
In........... Manual self-calibration
Applies setting and returns to the Measurement display.
Self-calibration requires about 55 ms, during which measurement
processing is temporarily suspended.
7
8
9
10
Page 72
66
HIOKI 3561A981-10
4.10 Measurement Value Output Function
4.10 Measurement Value Output Function
This function causes output of measured values via the RS-232C
interface in the same sequence as trigger input.
This function is useful when measuring using internal (free-run)
triggering, and for obtaining measured values on a PC when using a
footswitch for triggering.
1
2
3
4
(SHIFT Lamp lit)
The Menu display appears.
The Measurement Value Output function setting display appears.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display)
The current setting blinks.
Turn Measurement Value Output Function on or off.
on.......... enables the measurement value output function (ON).
off.......... disables the measurement value output function (ON).
Applies setting and returns to the Measurement display.
5
The measured value is output from the RS-232C interface when you
press the TRIG key or when a signal is applied to the EXT I/O TRIG
terminal.
Set the PC to the receiving state beforehand. When a measurement
value is received, the PC should perform appropriate processing such
as recording or displaying, then re-enable the receiving state.
• When external triggering is enabled, a measurement is performed
and the value is sent after each trigger event. When internal
triggering is enabled, the first value measured after triggering is
sent.
• The measurement output function is not applicable to the GP-IB
interface or printer.
Page 73
4.11 Key Beeper Setting
HIOKI 3561A981-10
67
4.11 Key Beeper Setting
1
2
3
Select whether a beep sounds when an operating key on the front of
the instrument is pressed.
(SHIFT Lamp lit)
The Menu display appears.
The Key Beeper setting display appears.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display)
The current setting blinks.
Select the key beeper state on the Sub Display.
on.......... Key beeper enabled
off ......... Key beeper disabled
1
2
3
4
Applied Measurement
5
6
4
Applies setting and returns to the Measurement display.
7
8
9
10
Page 74
68
HIOKI 3561A981-10
4.12 Reset Function
4.12 Reset Function
The reset function can be used to re-initialize current measurement
settings (excluding saved panel data) to their factory defaults, or to reinitialize all measurement settings including saved panel data to
factory defaults.
1
2
3
4
(SHIFT Lamp lit)
The Menu display appears.
The Reset display appears.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display)
The current setting blinks.
Select the Reset method on the Sub Display.
SEt ........ Reset (initializes measurement settings other than those stored
with Panel Save)
SYS....... System Reset (initialize all measurement settings)
ENTER blinks.
5
(Sub Display)
When SYS (system reset) is selected
Executes the Reset.
To cancel the settings: SHIFT key
System Reset also initializes Panel Save data.
Page 75
4.12 Reset Function
HIOKI 3561A981-10
Initial Factory Default Settings
DescriptionDefault
Measurement ModeV
69
Resistance Measurement RangeAUTO
Zero-AdjustOFF
Zero-Adjust Value0
DelayOFF
Delay Time0.000s
Sampling RateSLOW
Averaging FunctionOFF
Average Times2
Self-CalibrationAUTO
Continuous MeasurementON
Trigger SourceInternal trigger
Line Frequency50 Hz
Key Beeper SettingON
Key-Lock FunctionOFF
ComparatorOFF
Comparator Threshold Method
(resistance and voltage)
Hi, Lo
1
2
3
4
Applied Measurement
5
6
Comparator Upper Threshold
(resistance and voltage)
Comparator Lower Threshold
(resistance and voltage)
Comparator Judgment BeeperOFF
Comparator Execution ModeAUTO
Statistical Calculation FunctionsOFF
Automatic Clearing of Statistical
Calculation Results
InterfaceRS-232C
Baud Rate9600 bps
GP-IB Address1
GP-IB DelimiterLF
Print Interval0 (The interval print disabled)
Error OutputASync
Measurement Value Output Function OFF
EOC OutputHOLD
0
0
OFF
7
8
9
10
EOC Pulse Width1 ms
Page 76
70
HIOKI 3561A981-10
4.12 Reset Function
Page 77
5.1 Overview
HIOKI 3561A981-10
External ControlChapter 5
71
1
5.1Overview
External Control
Input Functions
External Output
Terminal
Functions
Connector Type57RE-40360-730B (D29) (manufactured by DDK)
RC30-36P (manufactured by HIROSE electric co.,ltd.)
or equivalent
PRINT)
LOAD0 to LOAD6)
0ADJ)
CAL)
MANU)
2
3
4
5
5
External Control
6
7
To avoid electric shock or damage to the equipment, always observe
the following precautions when connecting to EXT I/O connector.
• Always turn off the power to the instrument and to any devices
to be connected before making connections.
• Be careful to avoid exceeding the ratings of EXT I/O connector.
• During operation, a wire becoming dislocated and contacting
another conductive object can be serious hazard. Make sure
that connections are secure the EXT I/O connectors.
• The INT.GND terminals (Page 73) are grounded, so if an
external controller has a potential relative to ground,
connection could cause a short-circuit accident.
80
Be aware of the following to avoid damaging the instrument:
• When using relays, always include diodes to absorb back emf.
• Always provide protective grounding for devices the connect to the
EXT I/O connectors.
8
9
10
Page 78
72
HIOKI 3561A981-10
5.2 Signal Descriptions
5.2Signal Descriptions
5.2.1Pinout
118
3619
EXT I/O Connector
PinI/OSignal name
1INLOAD0
2IN
3IN
4IN
5IN
6IN
7INT.GND
8INT.GND
9OUTR-Hi
10OUTR-Lo
11OUTV-Hi
12OUTV-Lo
13OUTEOC
14OUTOUT0
15OUTOUT2
16OUTOUT4
17OUTOUT6
18OUTOUT8
LOAD2
LOAD4
LOAD6
CAL (IN1)
(IN3)
PRINT
PinI/OSignal name
19IN
20IN
21IN
22IN
23IN
24IN
25INT.GND
26INT.GND
27OUTR-IN
28OUTAND
29OUTV-IN
30OUTERR
31OUTINDEX
32OUTOUT1
33OUTOUT3
34OUTOUT5
35OUTOUT7
36OUTOUT9
LOAD1
LOAD3
LOAD5
TRIG
(IN0)
0ADJ (IN2)
(IN4)
MANU
Page 79
5.2.2Input Signals
HIOKI 3561A981-10
73
5.2 Signal Descriptions
LOAD0 to LOAD6Select a Panel No. to load and apply a TRIG signal to load the selected
Panel No. and measure.
When a TRIG signal is applied, if
unchanged from the previous trigger event, panel settings are not
loaded. In this case, using external triggering, one measurement is
taken as usual when the
Panel
No.
122
123
124
125
126
0: LOAD terminal is shorted to GND 1: LOAD terminal is unconnected, or
connected to 5 V
When a TRIG signal is applied with LOAD0 to LOAD6 set to all 1's or all
0's, no Panel Load occurs.
• At least 70 ms is required for the settings to change after executing
a Panel Load (the actual time depends on the particular
range and sampling rate
• When set to external trigger mode, one measurement is taken upon
load completion.
• The Panel Load function cannot be executed from
when controlling the instrument via RS-232C or GP-IB (Remote
State).
TRIGWhen the external trigger, one measurement is taken each time the
TRIG signal transitions from High to Low.
This trigger signal is ignored when internal triggering is enabled.
Trigger functions are also available for statistical calculation, recording
to memory and output of measured values (valid also with internal
triggering).
CALWhen manual self-calibration is selected with EX.FAST, FAST or
MEDIUM sampling rate, self-calibration begins when the
transitions from High to Low.
Self-calibration takes about 55 ms.
When SLOW sampling is selected, the
See Section 4.9 Self-Calibration p.65).
CAL signal is ignored.
CAL signal
10
Page 80
74
HIOKI 3561A981-10
5.2 Signal Descriptions
0ADJZero adjustment executes once when the 0ADJ signal transitions from
High to Low.
PRINTThe current measurement value prints when the PRINT signal
transitions from High to Low.
MANUWhen the MANU comparator mode is selected, comparator judgment
is enabled while the
See Section 4.1.4 Comparator Execution Mode Setting p.46).
MANU signal is Low.
IN0 to IN4The TRIG, CAL, 0ADJ, PRINT and MANU signals can also serve as
general-purpose input terminals, read with the
See Section EXT I/O Input p.135).
:IO:IN? command.
5.2.3Output Signals
ERRIndicates a measurement fault.
The Synchronous ERR output setting causes ERR output to be
synchronous with EOC output, while with the Asynchronous ERR
output setting causes ERR output to follow actual (asynchronous)
contact of the probes with the test object.
See Section 5.2.4 ERR Output p.75).
INDEXThe INDEX signal is output during the Trigger Wait, Delay, Self-
Calibration and Calculation states.
This signal is not output while measuring the resistance of test objects.
This signal transitions from Off to On to indicate that the test object
can be removed.
EOCThis signal indicates the end of a measurement (End-Of-Conversion).
This signal indicates when comparator judgment results and ERR
output (when SYNC is enabled) are available.
R-Hi, R-IN, R-Lo
These are the results of comparator decision.
V-Hi, V-IN, V-Lo
ANDThis signal indicates when both resistance and voltage judgment
results are IN (V mode).
In the and V modes, this signal is the same as R-IN and V-IN
outputs, respectively.
OUT0 to OUT9The output signals are controlled by the :IO:OUT command.
INT.GNDThis is the instrument's internal ground.
See Section EXT I/O Output p.135).
• I/O signals should not be used while measurement settings have
been changed.
• The EOC and INDEX signals are initialized (ON) at power on.
• If it is not necessary to change the measurement conditions, set
LOAD0 through LOAD6 to either Hi or Lo.
Page 81
5.2.4ERR Output
HIOKI 3561A981-10
The ERR output signal indicates the occurrence of measurement fault
conditions (such as open test leads, or a bad contact).
There are two ERR output methods.
75
5.2 Signal Descriptions
1
Synchronized
with EOC Output
(SYNC)
Asynchronous
with EOC Output
(ASYNC)
Measurement faults detected while measuring (not while awaiting
trigger or during delay or calculation intervals), are indicated by ERR
output synchronous with EOC output (the end-of-measurement
signal).
ERR Output On: A measurement fault has prevented correct
measurement
ERR Output Off: Correct measurement obtained (OF or -OF: Out-of-
range cases are included)
Measurement faults (test lead connection conditions) are output in real
time. The output is asynchronous with the
output.
ERR Output On: Measurement fault condition (open test leads, or a
bad contact)
ERR Output Off: Test lead connections are normal
TRIG signal and EOC
2
3
4
5
5
External Control
6
7
8
9
10
Page 82
76
HIOKI 3561A981-10
5.2 Signal Descriptions
5.2.5Instrument Settings
Measurement Fault Output Signal (ERR) Setting
1
2
3
4
(SHIFT Lamp lit)
The Menu display appears.
Select the ERR Output Selection display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) p.15).
Select the type of signal to be output on the Sub Display.
SynC .......... Synchronous output (synchronized with EOC output)
ASynC ....... Asynchronous output (not synchronized with EOC output)
Applies settings and returns to the Measurement display.
Setting the EOC Signal
(Main Display)
(Sub Display)
The current setting blinks.
1
2
3
4
(SHIFT Lamp lit)
The Menu display appears.
Select the EOC-signal setup display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) p.15).
(Main Display)
(Sub Display)
The current setting blinks.
Choose the output method for the EOC signal.
HoLd .......... Holds the EOC signal after measurement.
Go to Step 5.
PULSE ....... Outputs the specified pulse after measurement.
Go to the next step.
(When PULSE is selected)
The number representing the pulse width of the EOC signal will start blinking.
Set the pulse width in ms.
Or ten-keys
5
Applies settings and returns to the Measurement display.
Page 83
5.3Timing Chart
HIOKI 3561A981-10
77
5.3 Timing Chart
External Trigger Timing Chart
Contact State
*1 ERR Output
(Err Output ASYNC Setting)
*3 TRIG
Measurement start signal
INDEX Output
Reference Signal
End-of-Measurement Signal
Comparator Result
*2 ERR Output
(Err Output SYNC Setting)
Input
EOC Output
t1t1
1
Contact
Open
2
3
t2
4
Measuring
End of Conversion
5
5
External Control
6
t3
t4t5t6
*1: For details, see “5.2.4 ERR Output p.75).”
*2: When ERR output is set to the SynChronous mode, measurement fault detection results can be obtained
when measurement is finished, as with comparator results.
*3: After connecting to the test object, wait for longer than the response time (approximately 3 ms) before
inputting the TRIG signal (It is necessary to wait out the response time for the measurement values to
stabilize after connection. Response times depend on the test object).
Internal Trigger Timing Chart
* INDEX Output
Reference Signal
* EOC Output
End-of-Measurement
Signal
Comparator Result
t3
MeasuringMeasuring
t4t5
t6
7
8
9
10
* When the EOC signal is set to PULSE, the signal will remain on only for the specified period upon
completion of conversion.
Page 84
78
HIOKI 3561A981-10
5.3 Timing Chart
Description
t1
ERR Output response
*1
time
Measurement trigger
t2
pulse width
Delay Timeper setting
t3
Measurement time
t4
Calculation time
t5
EOC Output pulse widthWhen the external trigger is selected
t6
*2
*3
Time
1.5 ms
0.5 ms min.
See Section 4.2.2 Trigger Delay Settings p.54).
V mode
EX.FAST6.8 ms
FAST22.8 ms
MEDIUM82.8 ms
68.8 ms
SLOW257.8 ms
251.2 ms
0.3 ms
HOLD setting: Holds until the next trigger is detected
PULSE setting : Remains only for the specified pulse width
See Section 5.2.5 Instrument Settings p.76).
When the internal trigger is selected
HOLD setting: EX.FAST 1 ms, FAST 5 ms, MEDIUM 20 ms, SLOW 50 ms
PULSE setting : Remains only for the specified pulse width
mode or V mode
EX.FAST3.4 ms
FAST11.4 ms
MEDIUM41.4 ms (50 Hz line
frequency setting)
34.4 ms (60 Hz line
frequency setting)
SLOW156.4 ms (50 Hz line
frequency setting)
149.8 ms (60 Hz line
frequency setting)
*1: For details, see “5.2.4 ERR Output p.75).”
: About t4 measurement time
*2
When averaging is enabled, the running average is obtained with internal triggering, so
measurement time t4 does not change. The measurement time for external triggering is as
follows:
With SLOW sampling
V(t4 - 57.8) X n + 57.8 ms (50 Hz)
(t4 - 51.2) X n + 51.2 ms (60 Hz)
or V(t4 - 56.4) X n + 56.4 ms (50 Hz)
(t4 - 49.8) X n + 49.8 ms (60 Hz)
With other than SLOW sampling
V(t4 - 2.8) X n + 2.8 ms
or V(t4 - 1.4) X n + 1.4 ms
(n represents the number of values averaged)
: About t5 calculation time
*3
In the following cases, add the indicated times to calculation time t5:
When the Statistical Calculation function is enabled 0.3 ms
When the reference value/tolerance method of
comparator decision is selected
0.15 ms
Page 85
79
HIOKI 3561A981-10
5.4 Internal Circuitry
5.4Internal Circuitry
External Control and External Output Terminal Ratings
I/O typeLogicElectrical specification
OutputOpen drain30 V DC, 50 mA max.
InputC-MOSInverse logic H: 3.8 to 5.0 V, L: 0 to 1.2 V
External Control Terminals_______________________________________
Because electric shock and instrument damage hazards are
present, always follow the steps below when connecting the
printer.
• Always turn off the instrument and the printer before connecting.
• A serious hazard can occur if a wire becomes dislocated and
• As much as possible, avoid printing in hot and humid
• Use only compatible recording paper in the printer. Using non-
• If the recording paper is skewed on the roller, paper jams may
Recommended printer
The requirements for a printer to be connected to the instrument are
as follows.
Confirm compatibility and make the appropriate settings on the printer
before connecting it to the instrument.
contacts another conductor during operation. Make certain
connections are secure.
environments. Otherwise, printer life may be severely shortened.
specified paper may not only result in faulty printing, but printing
may become impossible.
result.
3
4
5
6
Printing
7
8
9
• Interface ............................ RS-232C
• Characters per line............ At least 45
• Communication speed ...... 9600 bps
• Data bits ............................ 8
• Parity ................................. none
• Stop bits ............................ 1
• Flow control....................... none
• Control codes ................... Capable of directly printing plain text
The optional printer model 9670 is no longer available. Their model
9670 printers can still use.
10
Page 88
82
HIOKI 3561A981-10
6.1 Connecting the Printer
6.1.1Connecting the PRINTER to the Instrument
Connection Methods
Model 3561 (3561-01)
4
Printer
1. Confirm that the instrument and
printer are turned off.
2. Connect the AC Adapter to the
printer, and insert the power
plug into an outlet.
3.
Connect the RS-232C cable
RS-232C connectors on the
instrument and printer.
3
2
to the
RS-232C cable
Connector Pinouts
1 2 3 4 5
6 7 8 9
Model 3561(-01) (9-pin) Connector
Function
Receive DataRxD2
Transmit DataTxD3
Signal or Common GroundGND5
Signal
Name
AC Adapter
Pin
4. Turn the instrument and printer
on.
13 ........................................ 1
25 ................................ 14
Printer (25-pin) Connector (Example)
Pin
Signal
Name
2TxDTransmit Data
3RxDReceive Data
7GNDSignal or Common Ground
4RTSRequest to Send
5CTSClear to Send
Function
Page 89
6.2Selecting the Interface
HIOKI 3561A981-10
83
6.2 Selecting the Interface
1
2
3
Or
ten-keys
(SHIFT Lamp lit)
The Menu display appears.
Select the Interface Selection display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display)
The current setting blinks.
Select Printer on the Sub Display.
rS ......... RS-232C
GP-Ib ... GP-IB
Prn....... Printer
Set the print interval time.
0000...................Interval printing is OFF. (Printing is carried out once
when PRINT key is pressed.)
0001 to 3600...... Sets the print interval time in seconds.
1
2
3
4
5
6
Printing
4
Applies setting and returns to the Measurement display.
7
8
9
10
Page 90
84
HIOKI 3561A981-10
6.3 Printing
6.3Printing
Printing Measured Values and Decision Results ____________________
From the Measurement display, press the PRINT key or ground the
PRINT pin in the EXT I/O connector to print the measured value and
decision result.
• When using the external trigger, if you want to print after a
triggered measurement finishes, connect the EOC signal of the
EXT I/O to the
• To print all measurements continuously, connect the EOC signal to
PRINT signal and enable the internal trigger.
the
• When the statistical calculation function is on and the internal
trigger is selected, the TRIG key or
statistical calculation and printing of the current measurement
value.
• Valid counts are 1 to 30000. Above 30000, the count returns to 1.
This function allows you to automatically print out measurement
results at preset intervals. The print interval time must be set from the
Interface Selection display.
See Section 6.2 Selecting the Interface (p.83).
The setting range is 1 to 3600 seconds.
When the print interval time is set to "0", interval printing is disabled,
and only normal printing is carried out.
Operation when interval printing is selected:
1. Start printing by pressing the PRINT key or sending the
signal via EXT I/O.
2. Elapsed time (hours/minutes/seconds) and measurement values
are printed automatically at intervals corresponding to the preset
interval time.
3. Stop printing by pressing the PRINT key or sending the
signal via EXT I/O again.
• When the printed elapsed time reaches 100 hours, it resets to
00:00:00 and continues from zero.
(Example)
After 99 hours, 59 minutes and 50 seconds: 99:59:50
After 100 hours, 2 minutes and 30 seconds: 00:02:30
• Selecting a display other than the measurement display causes
interval printing to stop.
From the Statistical Calculation display, press the PRINT key to print
statistical calculation results. If no valid data exists, only the data count
is printed. When only one valid data sample exists, standard deviation
of sample and process capability indices cannot be printed.
Page 91
85
HIOKI 3561A981-10
6.3 Printing
Example Printouts______________________________________________
Measurement values
(V mode)
1 298.60mOhm, 1.3924 V
2 0.2984 Ohm, 1.3924 V
3 - 3.35mOhm, 0.0000 V
4 - 0.0054 Ohm, 0.0000 V
5 299.10mOhm,- 1.3923 V
6 0.2984 Ohm,- 1.3923 V
7 3.57mOhm, 13.9071 V
8 - 16.89mOhm,-13.9088 V
With the Comparator ON
95 105.80mOhm Lo, 0.0000 V IN
96 213.15mOhm Hi
97 213.12mOhm IN
98 213.11mOhm Lo
99 10.0072 V Hi
100 10.0071 V IN
101 10.0070 V Lo
102 O.F. Hi, O.F. Hi
103 - 3.11mOhm Lo,- O.F. Lo
104 -------- --,-------- --
Statistical Calculations (Comparator ON)
*** RESISTANCE ***
Number 85
Valid 85
Average 13.06mOhm
Max 13.78mOhm( 74)
Min 12.10mOhm( 3)
Sn 0.38mOhm
Sn-1 0.38mOhm
Cp 1.32
CpK 0.09
Comp Hi 40
Comp IN 45
Comp Lo 0
Max/Min count
Measurement values
( mode)
15 209.98mOhm
16 0.2103 Ohm
Measurement values
(V mode)
3132 4.2019 V
3133 15.2084 V
With the relative value comparison method
(reference value and tolerance)
84 0.023 % Hi, 0.001 % IN
85 0.014 % IN, 0.000 % IN
86 - 0.019 % Lo, 0.002 % IN
00:00:00 13.74mOhm, 10.0138 V
00:00:01 13.87mOhm, 10.0138 V
00:00:02 13.67mOhm, 10.0139 V
00:00:03 13.47mOhm, 10.0138 V
00:00:04 13.58mOhm, 10.0139 V
00:00:05 13.58mOhm, 10.0139 V
00:00:06 13.68mOhm, 10.0139 V
1
2
3
4
5
6
Printing
7
8
*** VOLTAGE ***
Number 85
Valid 85
Average 10.0074 V
Max 10.0197 V ( 57)
Min 9.9938 V ( 31)
Sn 0.0068 V
Sn-1 0.0068 V
Cp 0.35
CpK 0.32
Comp Hi 10
Comp IN 59
Comp Lo 16
Measurement values indicated as "Invalid" cannot be displayed by
the instrument.
The number of statistical calculation results indicated as “Valid”
equals the count of valid data excluding measurement faults and
overflows.
9
10
Page 92
86
HIOKI 3561A981-10
6.3 Printing
Page 93
RS-232C/GP-IB
HIOKI 3561A981-10
87
7.1 Overview and Features
Interfaces
This chapter describes the GP-IB and RS-232C interfaces, using the
following symbols to indicate which information pertains to each
interface. Sections with neither of these symbols pertain to both
interfaces.
: GP-IB only
: RS-232C only
Before Use• GP-IB is available only on Model 3561-01.
• Always make use of the connector screws to affix the GP-IB or RS232C connectors.
• When issuing commands that contain data, make certain that the
data is provided in the specified format.
Chapter 7
7.1Overview and Features
1
2
3
4
5
6
All instrument functions other than power on/off switching can be
controlled via GP-IB/RS-232C interfaces.
• Resetting is supported.
• IEEE 488.2-1987 Common (essential) Commands are supported.
• Complies with the following standard:
Applicable standard IEEE 488.1-1987
• This instrument is designed with reference to the following standard:
Reference standard IEEE 488.2-1987
• If the output queue becomes full, a query error is generated and the
output queue is cleared. Therefore, clearing the output queue and
query error output from the deadlocked condition
IEEE 488.2 is not supported.
*1.ANSI/IEEE Standard 488.1-1987, IEEE Standard Digital Interface for
Programmable Instrumentation.
*2.ANSI/IEEE Standard 488.2-1987, IEEE Standard Codes, Formats,
Protocols, and Common Commands.
*1
*2
*3
as defined in
7
RS-232C/GP-IB Interfaces
8
9
10
*3.The situation in which the input buffer and the output queue become full,
so that processing cannot continue.
Page 94
88
HIOKI 3561A981-10
7.2 Specifications
7.2Specifications
7.2.1RS-232C Specifications
Transfer methodCommunications : Full duplex
Synchronization : Start-stop synchronization
Baud rate9600 bps/ 19200 bps/ 38400 bps
Data length8 bit
Paritynone
Stop bit1 bit
Message terminator
(delimiter)
Receiving: CR+LF, CR
Transmitting: CR+LF
Flow controlnone
Electrical specificationInput voltage levels5 to 15 V: ON,-15 to -5 V: OFF
Output voltage levels5 to 9 V: ON, -9 to -5 V: OFF
ConnectorRS-232C Interface Connector Pinout
(Male 9-pin D-sub, with #4-40 attachment screws)
The I/O connector is a DTE (Data Terminal Equipment) configuration
Recommended cables:
• Model 9637 RS-232C CABLE (for PC/AT-compatibles)
• Model 9638 RS-232C CABLE (for PC98-series)
See Section 7.3.1 Attaching the Connector (p.89).
7.2.2GP-IB Specifications (Model 3561-01 only)
GP-IB is available only on Model 3561-01.
Interface Functions
SH1All Source Handshake functions are supported.
AH1All Acceptor Handshake functions are supported.
T6Basic talker functions are supported. Serial poll function are
supported. No talk-only mode. The talker cancel function
with MLA (My Listen Address) is supported.
L4Basic listener functions are supported. No listen-only mode.
The listener cancel function with MTA (My Talk Address) is
supported.
SR1All Service Request functions are supported.
RL1All Remote/Local functions are supported.
PP0No Parallel Poll function.
DC1All Device Clear functions are supported.
DT1All Device Trigger functions are supported.
C0No Controller functions are supported.
Operating Code: ASCII codes
Page 95
89
HIOKI 3561A981-10
7.3 Selecting the Connections and Protocol
7.3 Selecting the Connections and Protocol
7.3.1Attaching the Connector
• Always turn both devices OFF when connecting and
disconnecting an interface connector. Otherwise, an electric
shock accident may occur.
• After connecting, always tighten the connector screws. The
mounting screws must be firmly tightened or the RS-232C
connector may not perform to specifications, or may even fail.
• To avoid damage to the instrument, do not short-circuit the
connector and do not input voltage to the connector.
RS-232C Connector
1 2 3 4 5
6 7 8 9
Male 9-pin D-sub
#4-40 attaching screws
Connect the RS-232C cable.
To connect the instrument to a controller
(DTE), use a crossover cable
the connectors on both the instrument and the
controller.
compatible with
1
2
3
4
5
6
The I/O connector is a DTE (Data Terminal Equipment) configuration.
This instrument uses only pins 2, 3 and 5. The other pins are
unconnected.
Pin
No.
CommonEIAJIS
1DCDCFCDUnusedNo connection
2RxDBBRDReceive Data
3TxDBASDTransmit Data
4
5GNDABSGSignal Ground
6DSRCCDRUnusedNo connection
7
8CTSCBCSUnusedNo connection
9RICECIUnusedNo connection
Signal Name
DTRCDER
RTSCARS
SignalNotes
Data Terminal
Ready
Request to SendInternally connected
Internally connected
to +5 V
to +5 V
7
RS-232C/GP-IB Interfaces
8
9
10
Page 96
90
HIOKI 3561A981-10
7.3 Selecting the Connections and Protocol
Connecting to a PC/ATCompatible (DOS/V)
Machine
Connecting to an NEC
PC9801 or PC9821
Series Desktop PC
(excluding NX)
Use a crossover cable with female 9-pin D-sub connectors.
Crossover Wiring
Female 9-pin
D-sub
Model 3561-end
Pin No.Pin No.
DCD11DCD
RxD22RxD
TxD33TxD
DTR44DTR
GND55GND
DSR66DSR
RTS77RTS
CTS88CTS
99
Female 9-pin
D-sub
PC/AT-end
Recommended
cable:
HIOKI
Model 9637 RS-232C
CABLE (1.8 m)
Use a crossover cable with a female 9-pin D-sub and a male 25-pinD-sub connector.
As the figure shows, RTS and CTS pins are shorted together and
crossed to DCD in the other connector.
Crossover Wiring
Female 9-pin
D-sub
Model 3561-end
Pin No.Pin No.
DCD1
RxD22TxD
TxD33RxD
DTR44RTS
GND55CTS
DSR66DSR
RTS77GND
CTS88DCD
920DTR
Male 25-pin
D-sub
PC-end
Recommended
cable:
HIOKI
Model 9638 RS-232C
CABLE (1.8 m)
Note that the combination of a dual male 25-pin D-sub cable and a 9to 25-pin adapter cannot be used.
GP-IB Connector
Connecting a GP-IB cable.
Recommended cable:
Model 9151-02 GP-IB CONNECTOR CABLE
(2 m)
Page 97
7.3 Selecting the Connections and Protocol
HIOKI 3561A981-10
7.3.2Selecting the Interface
91
1
2
3
(SHIFT Lamp lit)
The Menu display appears.
Select the Interface Selection display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display)
The current setting blinks.
Select RS-232C or GP-IB on the Sub Display.
......... RS-232C
rS
GP-Ib .... GP-IB (Model 3561-01 only)
Prn ....... Printer
When you select RS-232C, set the communications speed.
(Sub Display)
1
2
3
4
5
6
4
When selecting GP-IB, also set the
(Sub Display)
Message Terminator setting (LF/CRLF)
Address setting (0 to 30)
Selects the item to set
Applies settings and returns to the Measurement display.
Address and Message Terminator
Setting
7
.
RS-232C/GP-IB Interfaces
8
9
10
Page 98
92
HIOKI 3561A981-10
7.4 Communication Methods
7.4Communication Methods
Various messages are supported for controlling the instrument through
the interfaces.
Messages can be either program messages, sent from the PC to the
instrument, or response messages, sent from the instrument to the
PC.
Program Messages
Response Messages
Model 3561
(Model 3561-01)
Message types are further categorized as follows:
Messages
7.4.1Message Format
Program
Messages
Program messages can be either Command Messages or Query
Messages.
• Command Messages
Instructions to control the instrument, such as to change settings or
reset
Example: (instruction to set the measurement range)
:RESISTANCE:RANGE 100E-3
Command Message
Program Messages
Query Message
Response Messages
Header portion
• Query Messages
Requests for responses relating to results of operation or
measurement, or the state of instrument settings.
Example: (request for the current measurement range)
Space
Data portion
:RESISTANCE:RANGE?
Header portion
For details: See Section Headers (p.93), Separators (p.94) and Data
Formats (p.95).
Question Mark
Page 99
7.4 Communication Methods
HIOKI 3561A981-10
93
Response
Messages
When a query message is received, its syntax is checked and a
response message is generated.
The
:SYSTem:HEADer command determines whether headers are
(the current resistance measurement range is 300 m)
At power-on, Header OFF is selected.
If an error occurs when a query message is received, no response
message is generated for that query.
No header is applied to commands used only for queries, such as
:FETCH? and :CALCulate:LIMit:RESistance:RESult?.
Command SyntaxCommand names are chosen to mnemonically represent their
function, and can be abbreviated. The full command name is called
the "long form", and the abbreviated name is called the "short form".
The command references in this manual indicate the short form in
upper-case letters, extended to the long form in lower case letters,
although the commands are not case-sensitive in actual usage.
Response messages generated by the instrument are in long form and
FUNCTION
FUNC
FUNCT
FUN
in upper case letters.
OK (long form)
OK (short form)
Error
Error
1
2
3
4
5
6
HeadersHeaders must always be prefixed to program messages.
(1) Command Program Headers
There are three types of commands: Simple, Compound and
Standard.
• Headers for Simple Commands
This header type is a sequence of letters and digits
ESE 0
• Headers for Compound Commands
These headers consist of multiple simple command type headers
separated by colons ":"
:SAMPle:RATE
• Headers for Standard Commands
This header type begins with an asterisk "*",
indicating that it is a standard command defined by IEEE 488.2.
RST
(2) Query Program Header
These commands are used to interrogate the instrument about the
results of operations, measured values and the current states of
instrument settings.
As shown by the following examples, a query is formed by appending
a question mark "?" after a program header.
:FETCh?
:MEASure:RESistance?
7
RS-232C/GP-IB Interfaces
8
9
10
Page 100
94
HIOKI 3561A981-10
7.4 Communication Methods
Message
Terminators
Separators
This instrument recognizes the following message terminators:
• LF
• CR+LF
• EOI
• LF with EOI
From the instrument's interface settings, the following can be selected
as the terminator for response messages.
• LF with EOI (initial setting)
• LF with CR and EOI
See Section 7.3.2 Selecting the Interface (p.91).
• CR
• CR+LF
• CR + LF (initial setting)
(1) Message Unit Separator
Multiple message can be written in one line by separating them with
semicolons ";".
:SYSTEM:LFREQUENCY 60;IDN?
• When messages are combined in this way and if one command
• A query error occurs if a query command is combined with an
(2) Header Separator
In a message consisting of both a header and data, the header is
separated from the data by a space " ".
:SYSTEM:ELOCK ON
(3) Data Separator
In a message containing multiple data items, commas are required to
separate the data items from one another.
contains an error, all subsequent messages up to the next
terminator will be ignored.
immediately following semicolon and subsequent command.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.