Hioki 3561, 3561-01 Instruction Manual

Page 1
3561
HIOKI 3561A981-10
3561-01
Instruction Manual
BATTERY HiTESTER
Sept. 2020 Revised edition 10 3561A981-10 20-09H
Be sure to read this manual before using the instrument.
When using the instrument for the rst
time
Names and Functions of Parts
Measurement
p.11
p.23
Troubleshooting
Maintenance and Service
Error Display
p.2
p.169
p.170
EN
Page 2
HIOKI 3561A981-10
Page 3

Contents

HIOKI 3561A981-10
Contents
Introduction.................................................................................1
Verifying Package Contents .......................................................1
Safety Information ......................................................................2
Operating Precautions................................................................4
Chapter 1 Overview ___________________________________ 9
1.1 Product Overview..........................................................9
1.2 Features ......................................................................10
1.3 Names and Functions of Parts....................................11
1.4 Menu Display Sequence (SHIFT ENTER)..............15
1.5 Measurement Flowchart..............................................16
i
Chapter 2 Measurement Preparations___________________ 17
2.1 Preparation Flowchart .................................................17
2.2 Connecting the Power Cord ........................................18
2.3 Connecting the Optional Test Leads...........................19
2.4 Turning the Power On and Off ....................................21
2.5 Selecting the Line Frequency......................................22
Chapter 3 Measurement ______________________________ 23
3.1 Pre-Operation Inspection ............................................23
3.2 Basic Measurement Example .....................................24
3.3 Selecting Measurement Mode ....................................27
3.4 Setting Measurement Range ......................................28
3.4.1 Manual Range Setting ...................................................28
3.4.2 Auto-Ranging ................................................................29
3.5 Setting Sampling Rate ................................................30
3.6 Zero-Adjust Function...................................................30
3.6.1 Wiring Method for Zero-Adjustment ..............................30
3.6.2 Executing Zero-Adjustment ...........................................31
3.7 Displaying Measurement Results................................33
3.7.1 Measurement Fault Detection .......................................34
3.7.2 Overflow Display ...........................................................34
3561A981-10
Page 4
ii
HIOKI 3561A981-10
Contents
Chapter 4 Applied Measurement _______________________35
4.1 Comparator Function.................................................. 36
4.1.1 Comparator Setting Example 1
(Upper and Lower Threshold Judgment) ....................... 37
4.1.2 Comparator Setting Example 2
(Reference Value and Tolerance Judgment)................. 41
4.1.3 Comparator Judgment Beeper Setting ......................... 45
4.1.4 Comparator Execution Mode Setting ............................ 46
4.1.5 Comparator Threshold Method Selection ..................... 47
4.1.6 Upper and Lower Thresholds Setting
(by Reference Value and Tolerance)............................. 48
4.1.7 Enabling and Disabling the Comparator Function ........ 50
4.1.8 Comparator Judgment Results ..................................... 51
4.1.9 Switching Between Measurement Value and
Comparator Setting Displays ......................................... 52
4.2 Trigger Function ......................................................... 53
4.2.1 Trigger Source Settings ................................................ 53
4.2.2 Trigger Delay Settings .................................................. 54
4.3 Averaging Function..................................................... 55
4.4 Statistical Calculation Functions................................. 56
4.5 Memory Function........................................................ 60
4.6 Key-Lock Function...................................................... 62
4.7 Panel Save Function .................................................. 63
4.8 Panel Load Function................................................... 64
4.9 Self-Calibration........................................................... 65
4.10 Measurement Value Output Function......................... 66
4.11 Key Beeper Setting..................................................... 67
4.12 Reset Function ........................................................... 68
Chapter 5 External Control ____________________________71
5.1 Overview..................................................................... 71
5.2 Signal Descriptions..................................................... 72
5.2.1 Pinout ............................................................................ 72
5.2.2 Input Signals ................................................................. 73
5.2.3 Output Signals .............................................................. 74
5.2.4 ERR Output ..................................................................75
5.2.5 Instrument Settings ....................................................... 76
5.3 Timing Chart ............................................................... 77
5.4 Internal Circuitry ......................................................... 79
Page 5
Contents
HIOKI 3561A981-10
Chapter 6 Printing ___________________________________ 81
6.1 Connecting the Printer ................................................81
6.1.1 Connecting the PRINTER to the Instrument .................82
6.2 Selecting the Interface ................................................83
6.3 Printing ........................................................................84
Chapter 7 RS-232C/GP-IB Interfaces ____________________ 87
7.1 Overview and Features ...............................................87
7.2 Specifications ..............................................................88
7.2.1 RS-232C Specifications ................................................88
7.2.2 GP-IB Specifications (Model 3561-01 only) ..................88
7.3 Selecting the Connections and Protocol ....................89
7.3.1 Attaching the Connector ................................................89
7.3.2 Selecting the Interface ..................................................91
iii
7.4 Communication Methods ............................................92
7.4.1 Message Format ...........................................................92
7.4.2 Output Queue and Input Buffer .....................................97
7.4.3 Status Byte Register .....................................................98
7.4.4 Event Registers ...........................................................100
7.4.5 Initialization Items ........................................................103
7.4.6 Local Function .............................................................103
7.5 Message List .............................................................104
7.5.1 Standard Commands ..................................................104
7.5.2 Device-Specific Commands ........................................105
7.6 Message Reference ..................................................110
7.6.1 Standard Commands ..................................................111
7.6.2 Device-Specific Commands ........................................115
7.6.3 Measurement Value Formats ......................................142
7.6.4 Command Compatibility with the Model 3560 AC m
HiTESTER ...................................................................143
7.7 Basic Data Importing Methods ..................................147
7.8 Sample Programs .....................................................148
7.8.1 To be prepared in Visual Basic 5.0/6.0 .....................148
7.8.2 To be prepared in Visual Basic
7.8.3 Creation Procedure(Visual Basic
7.8.4 Sample Programs(Visual Basic
2005 ........................158
2005) ...................158
2005) ......................160
Page 6
iv
HIOKI 3561A981-10
Contents
Chapter 8 Specifications _____________________________ 163
8.1 Basic Specifications.................................................. 163
8.2 Accuracy................................................................... 167
8.3 General Specifications.............................................. 168
Chapter 9 Maintenance and Service ___________________169
9.1 Troubleshooting........................................................ 169
9.2 Cleaning ................................................................... 170
9.3 Error Display............................................................. 170
Appendix _________________________________ 171
Appendix 1 Precautions for Making Custom Test Leads .. 171
Appendix 2 AC Four-terminal Method............................... 173
Appendix 3 Synchronous Detection System ..................... 174
Appendix 4 Configuration and Extension of the Test Leads...... 175
Appendix 5 Effect of Eddy Currents .................................. 176
Appendix 6 Zero Adjustment............................................. 177
Appendix 7 Calibration Procedure..................................... 181
Appendix 8 Test Lead Options.......................................... 182
Appendix 9 Rack Mounting ............................................... 183
Appendix 10Dimensional Diagram..................................... 185
Index ______________________________________ i
Page 7
1
HIOKI 3561A981-10

Introduction

Introduction
Thank you for purchasing the HIOKI “Model 3561, 3561-01 BATTERY HiTESTER.” To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference.
Trademarks
Microsoft , Windows, and Visual Basic are either registered trademarks or trademarks of Microsoft Corporation in the United States and other countries.
2

Verifying Package Contents

When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative. Use the original packing materials when transporting the instrument, if possible.
Power cord (1)
Model 3561/ Model 3561-01 (GP-IB version)
BATTERY HiTESTER (1)
Instruction manual (this manual/ 1)
Options
The options listed below are available for the instrument. To order an option, please contact yourauthorized Hioki distributor or reseller.Options are subject to change. Please check Hioki's website for the latestinformation.
3
4
5
6
7
8
9
Model L2107 CLIP TYPE LEAD Model 9452 CLIP TYPE LEAD Model 9453 FOUR TERMINAL LEAD Model 9455 PIN TYPE LEAD (for ultra precision) Model 9467 LARGE CLIP TYPE LEAD Model 9770 PIN TYPE LEAD Model 9771 PIN TYPE LEAD Model 9637 RS-232C CABLE (9-pin to 9-pin/cross cable) Model 9638 RS-232C CABLE (9-pin to 25-pin/cross cable) Model 9151-02 GP-IB CONNECTOR CABLE (2 m)
10
11
Page 8
2
HIOKI 3561A981-10

Safety Information

Safety Information
This instrument is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. Using the instrument in a way not described in this manual may negate the provided safety features. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from instrument defects.
This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions.
In the manual, the symbol indicates particularly important information that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with
the symbol) before using the relevant function.
Indicates a grounding terminal.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument.
Indicates advisory items related to performance or correct operation of the instrument.
Other Symbols
Indicates a prohibited action.
p.
Indicates the location of reference information.
Indicates quick references for operation and remedies for troubleshooting.
Indicates that descriptive information is provided below.
*
Page 9
Safety Information
HIOKI 3561A981-10
Measurement categories
To ensure safe operation of measurement instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT II to CAT IV, and called measurement categories.
3
CAT II
CAT III
CAT IV
Using a measurement instrument in an environment designated with a higher-numbered category than that for which the instrument is rated could result in a severe accident, and must be carefully avoided. Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV measurement applications could result in a severe accident, and must be carefully avoided.
Primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.) CAT II covers directly measuring electrical outlet receptacles.
Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets.
The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel).
2
3
4
5
6
Accuracy
We define measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values, with the following meanings:
f.s.
(maximum display value or scale length) The maximum displayable value or scale length. This is usually the name of the currently selected range.
rdg.
(reading or displayed value) The value currently being measured and indicated on the measuring instrument.
dgt.
(resolution) The smallest displayable unit on a digital measuring instrument, i.e., the input value that causes the digital display to show a "1" as the least-significant digit.
7
8
9
10
11
Page 10
4
HIOKI 3561A981-10

Operating Precautions

Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions. Use of the instrument should confirm not only to itsspecifications, but also to the specifications of all accessories, options, and other equipment in use.
Instrument Installation and Operating Environment
Operating temperature and humidity: 0 to 40°C (32 ± 104°F), 80%RH or less (non-condensating) Temperature and humidity range for guaranteed accuracy: 23 ± 5°C (73 ± 9°F), 80% RH or less (non-condensating)
Avoid the following locations that could cause an accident or damage to the instrument.
Exposed to direct sunlight Exposed to high temperature
Exposed to liquids Exposed to high humidity or condensation
Exposed to high levels of particulate dust
In the presence of corrosive or explosive gases
Exposed to strong electromagnetic fields Near electromagnetic radiators
Subject to vibration
Avoid using near electrically noisy devices, as the noise may impinge upon the test object and cause unreliable measurements.
Installation
Do not install the instrument with any side except the bottom facing down.
Page 11
5
HIOKI 3561A981-10
Operating Precautions
Preliminary Checks
Before using the instrument the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Before using the instrument, make sure that the insulation on the power cord and test leads is undamaged and that no bare conductors are improperly exposed. Using the instrument in such conditions could cause an electric shock, so contact your dealer or Hioki representative for replacements.
2
This instrument may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
Measurement Precautions
• To avoid electrical shock, be careful to avoid shorting live lines with the test leads.
• The maximum rated voltage between input terminals and ground is ing 60 V with respect to ground could damage the instrument and result in personal injury.
• To avoid injury or damage to the instrument, do not attempt to measure AC voltage and AC current, or DC voltage exceeding 22 V.
• To prevent an electric shock, do not exceed the lower of the ratings shown on the instrument and test leads.
± 60 V DC. Attempting to measure voltages exceed-
3
4
5
6
7
±
8
• Use only the specified test leads and cables. Using a non-specified
cable may result in incorrect measurements due to poor connec­tion or other reasons.
• To ensure certified measurement accuracy, allow at least 30 min-
utes warm-up. After warm-up, be sure to execute self-calibration. See Section 4.9 Self-Calibration (P. 65).
• The input circuitry includes a protective fuse. Measurement is not possible when the fuse is blown.
• This instrument internally stores (backs up) all settings (except memory function and measurement values), such as measure­ment range, comparator settings and etc., but only when no opera­tion is performed for a certain time. Therefore, to preserve settings, do not turn the power off for a short time (about five seconds) after changing a setting. However, measurement settings made through the RS-232C or GP-IB interface and measurement settings loaded
LOAD signals of the EXT I/O connector are not memorized.
by
9
10
11
Page 12
6
HIOKI 3561A981-10
Operating Precautions
Before Connecting and Powering On
• Before turning the instrument on, make sure the supply voltage matches that indicated on the its power connector. Connection to an improper supply voltage may damage the instrument and present an electrical hazard.
• To avoid electrical accidents and to maintain the safety specifications of this instrument, connect the power cord provided only to a 3-contact (two-conductor + ground) outlet.
To suppress noise, the instrument needs to be set to match the frequency of the power source. Before operating, set the instrument to the frequency of your commercial power. If the supply frequency is not set properly, measurements will be unstable.
See Section 2.5 Selecting the Line Frequency (P. 22).
Make sure the power is turned off before connecting or disconnecting the power cord.
Handling the Instrument
• Never modify the instrument. Only Hioki service engineers should disassemble or repair the instrument. Failure to observe these precautions may result in fire, electric shock, or injury.
• The GP-IB connector location is covered by a blank panel on the Model 3561. To avoid electric shock, do not remove the blank panel.
Blank Panel
* This model is the 3561 BATTERY
HiTESTER.
• To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
• Do not apply heavy downward pressure with the stand extended. The stand could be damaged.
This instrument may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
Page 13
Operating Precautions
HIOKI 3561A981-10
Handling the Test Leads and Cables
• To avoid breaking the test leads and cables, do not bend or pull them.
• Avoid stepping on or pinching cables, which could damage the cable insulation.
• To avoid equipment failure, do not disconnect the communications cable while communications are in progress.
• Use a common ground for both the instrument and the computer. Using different ground circuits will result in a potential difference between the instrument's ground and the computer's ground. If the communications cable is connected while such a potential difference exists, it may result in equipment malfunction or failure.
• Before connecting or disconnecting any the communications cable, always turn off the instrument and the computer. Failure to do so could result in equipment malfunction or damage.
• After connecting the communications cable, tighten the screws on the connector securely. Failure to secure the connector could result in equipment malfunction or damage.
7
2
3
4
5
6
7
8
9
10
11
Page 14
8
HIOKI 3561A981-10
Operating Precautions
Page 15

1.1 Product Overview

HIOKI 3561A981-10
Overview Chapter 1
1.1 Product Overview
9
1

Overview

2
The Model 3561 and 3561-01 BATTERY HiTESTERs measure battery internal resistance using a four-terminal, 1-kHz AC method, while simultaneously measuring DC voltage (electromotive force [emf]). The high-precision, fast measurement performance and extensive interface capabilities make these models ideal for incorporating into battery testing production lines.
3
4
5
6
7
8
9
10
11
Page 16
10
HIOKI 3561A981-10

1.2 Features

1.2 Features
Simultaneously Measures Battery Internal Resistance and Voltage
The four-terminal AC method measures resistance and DC voltage simultaneously, so battery internal resistance and emf are measured and judged at once.
High-Precision Measurements
The instrument provides high-resolution resistance (0.01 m) and voltage measurements (0.1 mV). High precision ( ensures accurate voltage measurements.
± 0.01% rdg.)
High-Speed Measurements
Simultaneous resistance and voltage measurements can be performed as fast as once every 10 ms.
Comparator Functions
Resistance and voltage measurement values are judged in three categories (Hi, IN and Lo), with results clearly displayed. A comparator judgment beeper also provides distinct sounds to indicate pass/fail judgments and to facilitate correct recognition of judgment results.
Statistical Calculation Functions
Maximum, minimum and average measurement values, standard deviation, process capability indices and other values can be automatically calculated for applications such as production management. Calculation results can also be applied as comparator setting values.
Measurement Value Memory Function
The instrument includes a Memory function and storage capacity for up to 400 pairs of measurement values. When making many sequential measurements at high speed and sending the measured values to a PC after each measurement, the time to switch test objects can become unsatisfactorily long. The Memory function can avoid the slow-down by sending stored measurements in batches during idle times.
EXT I/O Interface
EXT I/O and RS-232C interfaces are equipped as standard, supporting transfer rates up to 38,400 bps. Model 3561-01 also supports GP-IB.
Printing Measurement Values and Statistical Results
Connect the printer to print measurement values and statistical calculation results.
Page 17

1.3 Names and Functions of Parts

HIOKI 3561A981-10
1.3 Names and Functions of Parts
11
Front Panel
Input Terminals (INPUT)
Connect the optional test leads.
Connections:
See Section 2.3 Connecting the Optional Test Leads (p.19).
SOURCE-H
SOURCE-L
SENSE-H
SENSE-L
Main Display
(p.12)
Operating Keys
(p.13)
1
Overview
2
3
4
5
6
Sub Display
(p.12)
POWER Switch
Turns the instrument on and off.
: Power OFF : Power ON
See Section 2.4 Turning the Power On and Off (p.21).
7
8
9
10
11
Page 18
12
HIOKI 3561A981-10
1.3 Names and Functions of Parts
Main Display
The current measurement mode is indicated while measuring, and the setting item is displayed while making settings.
(Upper row)
AUTO EX.FAST, FAST, MED, SLOW
0 ADJ
MEM
EXT TRIG
Lit when measuring voltage
Indicates percentage units during relative value comparator operation
Shows measured value or setting item.
Lit when measuring with Auto-Ranging.
The selected Sampling Rate is lit.
Lit when measuring in a range for which Zero-Adjustment has been performed.
Lit when the Memory function is enabled.
Lit when the External Trigger function is enabled.
Units of displayed measurement
V
m
(Lower row)
V
STAT
AVG
LOCK
REMOTE
Unit of voltage
Unit of resistance (lit when the 3 range is selected)
Unit of resistance (lit when the 300 m range is selected)
Lit when the V (Resistance and Voltage measurement) mode is selected.
Lit when the Statistical Calculation function is enabled.
Lit when measuring with the Averaging setting enabled.
Lit when the keys are locked.
Lit during communications.
Shows Comparator Decision Result.
Indicates that the measured value is
Hi
above the upper threshold.
Indicates that the measured value is
IN
between the upper and lower thresholds.
Indicates that the measured value is
Lo
below the lower threshold.
Sub Display
Upper and lower thresholds and other settings are displayed (when set).
While measuring, indicates the Comparator function is enabled.
Indicates Voltage measurement mode
Indicates percentage units during relative value comparator operation
HIGH, LOW
REF, %
V
Indicates that absolute value comparator operation is enabled (while measuring), and also when setting.
Indicates that relative value comparator operation is enabled (while measuring), and also when setting.
Indicates voltage measurement units.
Page 19
Operating Keys
HIOKI 3561A981-10
To use a function marked on a key, just press the key.
To use a function printed under a key (blue letter), press the SHIFT key first (and confirm the SHIFT lamp is lit), and then the key.
1.3 Names and Functions of Parts
Use as ten-keys to enter numerical values. (Numerical values can be used with the RANGE key.)
13
1
Overview
2
3
SHIFT Lamp
[ ]: Enabled after pressing the SHIFT key (SHIFT lamp lit).
Operating Key
V/ / V
[0 ADJ]
LOAD
[SAVE]
TRIG
[INT/EXT ]
VIEW
STAT
[DELAY]
SMPL
[AVG]
COMP
[SET]
LOCAL
Description
Selects Measurement mode. (Resistance and voltage measurement, Resistance measurement or Voltage measurement)
Executes Zero-Adjustment.
Loads a saved measurement configuration (Panel settings).
Saves the current measurement configuration (Panel settings).
Executes a Manual Trigger event.
Selects internal/external triggering.
Switches the view mode of the V mode.
Displays and sets Statistical Calculation results.
Sets the Trigger Delay.
Selects the Sampling Rate.
Activates Averaging function settings.
Switches the Comparator function on and off.
Activates Comparator function setting.
Cancels remote control (RMT) and re­enables key operations.
Operating Key
PRINT
AUTO
[LOCK]
ENTER
[MENU]
RANGE
SHIFT
Description
Sends measurement values and statistical calculation results to the printer.
Switches between Auto and Manual range selection.
Switches the Key-Lock function on and off.
Applies settings.
Selects various operating functions and settings.
Up/Down: Changes setting value or numerical value, and range selection. Left/Right: Moves the setting item or digit.
• Enables the functions of the operating keys marked in blue. The lamp is lit when the SHIFT state is active.
• Cancels settings in various setting displays. (Returns to the Measurement display without applying settings.) However, this does not apply to Menu display. However, from a menu item display, changed settings are not canceled, but accepted as the display returns to measurement display (except after Zero-Adjustment clear or resetting).
4
5
6
7
8
9
10
11
Page 20
14
HIOKI 3561A981-10
1.3 Names and Functions of Parts
Rear Panel
Power Inlet
Connect the supplied power cord here. See Section 2.2 Connecting the Power Cord
(p.18).
EXT I/O Connector
Connect here to use the EXT I/O interface.
GP-IB Connector (Model 3561-01 only)
Connect here to use the GP-IB interface.
7.3.1 Attaching the Connector (p.89).
RS-232C Connector
Connection for the printer or RS­232C interface.
See Section 7.3.1 Attaching the Connector (p.89).
Side View
* The illustration shows the Model 3561-01 BATTERY HiTESTER (GP-IB version).
The GP-IB connector location is covered by a blank panel on the Model 3561. To avoid electric shock, do not remove the blank panel.
Stand
Can be opened to tilt the front panel upwards.
Do not apply heavy downward pressure with the stand extended. The stand could be damaged.
Page 21
1.4 Menu Display Sequence (SHIFT ENTER)
HIOKI 3561A981-10
1.4 Menu Display Sequence (SHIFT ENTER)
15
Various auxiliary settings can be performed from the menu item displays.
(SHIFT Lamp lit)
The Menu display appears.
(Main Display)
Zero-Adjustment Clear
display (p.30)
Measurement Value Output
function setting display
(p.66)
Interface Selection
EOC-signal setup
The up/down RANGE key changes the setting
shown on the Sub Display.
Self-Calibration setting
display (p.91)
ERR Output Selection
display (p.76)
1
Overview
2
3
4
display (p.65)
5
6
display (p.76)
7
Key Beeper setting
display (p.67)
Settings on the menu item displays are applied and saved internally when changed.
Line Frequency
setting display
(p.22)
8
Reset display
(p.68)
9
Pressing this key returns to the previous item display.
10
11
Page 22
16
HIOKI 3561A981-10

1.5 Measurement Flowchart

1.5 Measurement Flowchart
The basic measurement process flow is as follows:
Measurement Preparations
Connecting the power cord (p.18)
Connecting the test leads (p.19)
Turning the power on (p.21)
Selecting the line frequency (p.22)
Instrument’s Settings
Selecting measurement mode (p.27)
Selecting measurement range (p.28)
Selecting sampling rate (p.30)
Zero-Adjustment
Short the test leads together (p.31)
Executing zero-adjustment
Measurement Start
Connect the test leads to a test object.
Read the measured value (p.33)
For details about the functions that can be applied to measurement values such as comparator, trigger and averaging functions, refer to Chapter 4 Applied Measurement (p.35).
Page 23
Measurement
HIOKI 3561A981-10
17

2.1 Preparation Flowchart

Preparations
Chapter 2
2.1 Preparation Flowchart
This procedure describes instrument preparations such as making connections and turning power on.
(p.19)(p.18)
1
Rear Panel
2
Front Panel
3
5
(p.21)
4
1
2

Measurement Preparations

3
4
5
6
Connecting the power cord.
1
Connect the test leads to the instrument.
2
Connect the EXT I/O connector and
3
interface connector.
Turn the power on.
4
Set measurement settings.
5
Start measurement.
6
(p.18)
7
(p.19)
8
(p.89)
9
(p.21)
10
(p.23)
Verify that the instrument’s line frequency is correctly set when using it for the first time and after initialization following repair or recalibration. See Section 2.5 Selecting the Line Frequency (p.22).
Page 24
18
HIOKI 3561A981-10

2.2 Connecting the Power Cord

2.2 Connecting the Power Cord
To avoid electrical accidents and to maintain the safety specifications of this instrument, connect the power cord provided only to a 3-contact (two-conductor + ground) outlet.
To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it from the power outlet.
To suppress noise, the instrument needs to be set to match the line frequency. Before operating, set the instrument to the frequency of your commercial power. If the supply frequency is not set properly, measurements will be unstable. See Section 2.5 Selecting the Line Frequency (p.22). Make sure the power is turned off before connecting or disconnecting the power cord.
Rear Panel
1. Confirm that the instrument's Power
switch is OFF.
2. Check that the power supply voltage
is correct, and connect the power cord to the power inlet socket on the rear of the instrument.
3. Plug the power cord into the AC
outlet.
Page 25

2.3 Connecting the Optional Test Leads

HIOKI 3561A981-10
2.3 Connecting the Optional Test Leads
19
Red Lead
Test leads are not included as standard accessories with the instrument, so the appropriate options need to be purchased separately or constructed according to the user’s application requirements. To construct custom test leads, refer to Appendix 1 Precautions for Making Custom Test Leads (p.171). The resistance measurement terminals on this instrument consist of four separate banana jacks. See Appendix 1 Precautions for Making Custom Test Leads (p.171).
1. Confirm that the instrument's Power
switch is OFF.
2. Connect four-terminal test leads such
as the L2107 CLIP TYPE LEAD to INPUT A.
1
2
Measurement Preparations
3
4
5
6
Plug the mark on the red lead into the red marked jack on the
instrument, and plug the mark on the black lead into the black marked jack on the
Black Lead
Example: Optional model L2107 CLIP TYPE LEAD
instrument.
7
8
9
10
Page 26
20
HIOKI 3561A981-10
2.3 Connecting the Optional Test Leads
About Test Leads
(Example: Model L2107 CLIP TYPE LEAD)
The side with “V” mark is SENSE.
Red
SENSE
SOURCE
SENSE
SOURCE
Black
When clipping a thin line (Clip the line at the tip, serrated part of the jaws.)
Red
SENSE SOURCE SENSE SOURCE
Black
When clipping a thick line (Clip the line at the deep, non-serrated part of the jaws.)
Page 27

2.4 Turning the Power On and Off

HIOKI 3561A981-10
2.4 Turning the Power On and Off
21
Before turning the instrument on, make sure the supply voltage matches that indicated on the its power connector. Connection to an improper supply voltage may damage the instrument and present an electrical hazard.
• The measurement setting state is the same as when the power
• However, measurement settings made through the RS-232C or
• Before starting to measure, allow 30 minutes for warm-up.
Turning the Power On
Turn the POWER switch ON.
was previously turned off (backup). To preserve changes to settings, wait a short time (about five seconds) after changing a setting before turning power off.
GP-IB interface and measurement settings loaded by signals of the EXT I/O connector are not memorized.
After warm-up, be sure to perform a self-calibration. See Section 4.9 Self-Calibration (p.65).
(Main Display) Model name
LOAD
1
2
Measurement Preparations
3
4
5
6
Power ON
Turning the Power Off
(Main Display) Software version
(Sub Display) Line frequency
Interface
The measurement display appears.
Turn the POWER switch OFF.
7
8
9
10
Power OFF
Page 28
22
HIOKI 3561A981-10

2.5 Selecting the Line Frequency

2.5 Selecting the Line Frequency
Verify that the instrument’s line frequency is correctly set when using it for the first time and after initialization following repair or recalibration.
1
2
3
(SHIFT Lamp lit)
The Menu display appears.
(Main Display)
(Sub Display)
Select the Line Frequency setting display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display) flashing
Select the frequency of the AC mains supply being used.
4
(Main Display)
(Sub Display) flashing
50.......... 50 Hz
60.......... 60 Hz
Applies settings and returns to the Measurement display.
To suppress noise, the instrument needs to be set to match the frequency of the power source. Before operating, set the instrument to the frequency of your commercial power. If the supply frequency is not set properly, measurements will be unstable.
Page 29

3.1 Pre-Operation Inspection

HIOKI 3561A981-10
Measurement Chapter 3
23
1
Before starting measurement, please read Operating Precautions (p.4) and Chapter 2 Measurement Preparations (p.17).
• To avoid electrical shock, be careful to avoid shorting live lines with the test leads.
• The maximum rated voltage between input terminals and ground is 60 V DC. Attempting to measure voltages exceeding 60 V with respect to ground could damage the instrument and result in personal injury.
To avoid injury or damage to the instrument, do not attempt to measure AC voltage and AC current, or DC voltage exceeding
± 22 V.
±
3.1 Pre-Operation Inspection
2
3

Measurement

4
5
6
7
Before using the instrument, perform the following inspection to ensure that it is operating properly.
Check Point Check Contents
Instrument Chassis (both front and rear panels)
Test Leads and Power Cord
Good Test Sample • Measures as good and displays the correct
Bad Test Sample • Measures as bad and displays the correct
If the inspection reveals a defect, stop using the instrument and contact your dealer or Hioki representative.
• No damage or cracks
• No internal circuitry is exposed
• Metal parts that should be insulated are not exposed
measurement value
measurement value
8
9
10
Page 30
24
HIOKI 3561A981-10

3.2 Basic Measurement Example

3.2 Basic Measurement Example
The following example describes the measurement process.
Example: Measuring resistance and voltage of a 30 m lithium-ion battery
Required items: Lithium-ion battery (30 m)
Test leads: Model 9770 PIN TYPE LEAD are used here.
Measurement conditions:
Preparations
Measurement mode ............. V (Resistance and Voltage measurement)
Range .................................. 300 m
Sampling rage...................... SLOW
Zero adjustment ................... Enabled
1
2
Connect the power cord.
See Section 2.2 Connecting the Power Cord (p.18).
1
Connect the test leads.
See Section 2.3 Connecting the Optional Test Leads (p.19).
Red Lead
Black Lead
2
3
3
Example: Model 9770
Turn the power on.
See Section 2.4 Turning the Power On and Off (p.21). See Section 2.5 Selecting the Line Frequency (p.22).
Page 31
Instrument Settings
HIOKI 3561A981-10
25
3.2 Basic Measurement Example
4
5
Confirm the SHIFT lamp is not lit.
If this is lit, press the SHIFT key to turn it off.
SHIFT not lit
Select the Resistance Measurement mode.
(Here, resistance and voltage measurement is selected.)
See Section 3.3 Selecting Measurement Mode (p.27).
V lit
The measurement mode changes each time you press this key. Select to measure only resistance, or V to measure only voltage.
V
1
2
3
Measurement
4
5
V
6
6
7
Set the measurement range. (Here, 300 m range is selected.)
See Section 3.4 Setting Measurement Range (p.28).
m lit
Selects the 3 range ( lit)
Selects the 300 m range (m lit)
Set the sampling rate. (Here, SLOW is selected.)
See Section 3.5 Setting Sampling Rate (p.30).
SLOW lit
The sampling rate changes each time you press this key.
EX.FAST
FAST
MED
7
8
9
10
SLOW
Page 32
26
HIOKI 3561A981-10
3.2 Basic Measurement Example
Zero-Adjustment
8
9
Short the test leads together.
Proper Zero-Adjustment is not possible with incorrect wiring.
See Section 3.6 Zero-Adjust Function (p.30).
Example: Model 9770 PIN TYPE LEAD
Internal conductor
Bring the pins into contact at 3 points.
• Internal conductor and internal conductor
External conductor
(SHIFT Lamp lit)
Execute Zero-Adjust. After zero-adjustment, the display returns to the measurement mode.
• Internal conductor and external conductor
External conductor and external conductor
Model 9770
Err.02” appears if Zero-Adjustment fails. Verify that the test lead tips are properly shorted, and try zero-adjustment again.
Measurement
10
11
Connect the test leads to a battery.
Read the measured resistance and voltage.
0ADJ lit
Measured Resistance
Measured Voltage
See Section 3.7 Displaying Measurement Results (p.33). See Section 9.3 Error Display (p.170).
Page 33

3.3 Selecting Measurement Mode

HIOKI 3561A981-10
3.3 Selecting Measurement Mode
27
Select the measurement mode from V (both resistance and voltage measurement), (resistance measurement only) or V (voltage measurement only).
1
2
V” indicates the
V mode is selected
Confirm the SHIFT lamp is not lit.
If this is lit, press the SHIFT key to turn it off.
Switches the displayed measurement mode.
Each key-press switches the measurement mode.
V mode
(Resistance and Voltage measurement)
” indicates the 3 range is selected
” indicates the 300 m range
“m is selected
SHIFT not lit
1
2
3
Measurement
4
5
6
mode
(Resistance measurement)
V mode (Voltage measurement)
The fastest measurements are provided by selecting the or V mode when measuring resistance or voltage, respectively. See Section Sampling Time (p.164).
7
8
9
10
Page 34
28
HIOKI 3561A981-10

3.4 Setting Measurement Range

3.4 Setting Measurement Range
Select the resistance measurement range from 3 (“” indicator lit) or 300 m (“m” indicator lit). The auto-ranging function can be enabled to automatically determine the most suitable range. The voltage measurement range is fixed at 20 V.

3.4.1 Manual Range Setting

1
2
When auto-ranging is enabled, pressing this disables it (AUTO not lit) and enables manual range selection.
Select the range to use.
When the 3  range is selected
lit
3 range
When the 300 m selected
range is
300 m
range
m lit
Page 35

3.4.2 Auto-Ranging

HIOKI 3561A981-10
29
3.4 Setting Measurement Range
Switching from Auto
-ranging back to
Manual
range selection
When manual range selection is enabled, pressing this enables auto-
ranging. The most suitable measurement range is then selected
automatically.
AUTO lit
Press the
• Voltage measurement has only one range. The range cannot be changed.
• Depending on the state of the test object, auto-ranging may be unstable. In this case, select the range manually, or increase the Delay time. See Section 3.4.1 Manual Range Setting (p.28).
• Auto-ranging is not available when Comparator or Memory functions are enabled (ON).
• Refer to Chapter 8 Specifications (p.163) for details about accuracy.
AUTO key again. The range can now be changed manually.
1
2
3
Measurement
4
5
6
7
8
Range Displayed Values
Resistance Measurement Mode
Measured Current Open-Terminal Voltage
300 m -10.00 to 310.00 m 10 mA ± 10% 7 Vpeak.
3 -0.1000 to 3.1000 1 mA ± 10% 7 Vpeak.
20 V -19.9999 to 19.9999 V -- 7 Vpeak.
9
10
Page 36
30
HIOKI 3561A981-10

3.5 Setting Sampling Rate

3.5 Setting Sampling Rate
The sampling rate can be selected from EX.FAST, FAST, MEDIUM and SLOW. Slower sampling rates generally provide greater measurement precision.
Selects the sampling rate
EX.FAST FAST MEDIUM SLOW
• Measurements are especially susceptible to interference from the environment when EX.FAST is selected, so countermeasures such as shielding or twisting of test leads, cables and wiring around the test object may be necessary.
• When SLOW sampling is selected, self-calibration is executed during each measurement. At other sampling rates, self-calibration is executed manually or automatically every 30 minutes. See Section 4.9 Self-Calibration (p.65).
• Refer to the specifications for details of sampling rates. See Section Sampling Time (p.164).

3.6 Zero-Adjust Function

Execute zero adjustment before measuring to nullify any residual offset voltage from the instrument or measurement environment. Measurement accuracy specifications are applicable after zero adjustment. Zero adjustment can also be executed by the 0ADJ terminal of the EXT I/O connector. See Section 5.2 Signal Descriptions (p.72).

3.6.1 Wiring Method for Zero-Adjustment

Before executing zero adjustment, connect the test leads (probes) as follows:
1. Connect SENSE-H to SENSE-L.
2. Connect SOURCE-H to SOURCE-L.
3. Connect the joined SENSE and SOURCE leads together as shown
below.
SENSE-H SENSE-L
SOURCE-H SOURCE-L
Connection
Connection
Page 37

3.6.2 Executing Zero-Adjustment

HIOKI 3561A981-10
31
3.6 Zero-Adjust Function
1
Short the test leads together.
Proper zero adjustment is not possible with incorrect wiring.
Example: Model L2107 CLIP TYPE LEAD
Correct
SENSE
SOURCE
Red
Bring the "V" marks together at the same position.
SENSE
SOURCE
Black
Incorrect
SENSE
SOURCE
Model 9452 (Option)
Red
Black
Model 9453 (Option)
SOURCE
SENSE
Perform zero adjustment with the alligator clips and lead rods placed as above.
1
2
3
Measurement
4
5
6
Model 9770 (Option)
Bring the pins into contact at 3 points.
Model 9771 (Option)
When the resistance measurement value is displayed as "-----", change the facing direction.
Let the two points of the pin tip touch the spring part perpendicularly (be careful not to short the springs).
Internal conductor
External conductor
7
8
9
10
Page 38
32
HIOKI 3561A981-10
3.6 Zero-Adjust Function
2
(SHIFT Lamp lit)
Zero-adjust display appears.
After measurement, the measured value of the compensation applied by the zero­adjust function is displayed. The range of zero adjustment is up to 1,000 dgt.
Clearing Zero-Adjustment
1
(SHIFT Lamp lit)
The Menu display appears.
0ADJ lit
2
If Err02 is displayed
(Main Display)
(Sub Display) flashing
The zero-adjust value is cleared. (0ADJ not lit)
(Main Display)
Indicates that zero adjustment could not be executed, either because the range to be adjusted exceeds fault condition exists. The zero adjust function is canceled, so repeat the operation after correcting the cause of the error.
• Zero adjustment is limited to ± 1,000 dgt.
• Both resistance (all ranges) and voltage are adjusted together for all modes.
• Zero-adjustment values are retained even when power is turned off.
• The 0ADJ terminal of the EXT I/O connector also executes zero adjustment. See Section 5.2 Signal Descriptions (p.72).
• Zero adjustment is very difficult with the delicate probe tips of the Model 9455 PIN TYPE LEAD. Refer to 3.6.1 Wiring Method for Zero-Adjustment (p.30) to use other leads when executing zero adjustment.
± 1,000 dgt, or a measurement
Page 39

3.7 Displaying Measurement Results

HIOKI 3561A981-10
3.7 Displaying Measurement Results
33
In the V mode, resistance measurements appear on the upper display, and voltage measurements appear on the lower display.
Measured Resistance
Measured Voltage
In the mode, resistance measurements appear on the upper display.
Measured Resistance
1
2
3
Measurement
4
5
6
In the V mode, voltage measurements appear on the upper display.
Measured Voltage
7
8
9
10
Page 40
34
HIOKI 3561A981-10
3.7 Displaying Measurement Results

3.7.1 Measurement Fault Detection

If a measurement does not execute properly, a measurement fault “- - - - -” is indicated on the display. In addition, a measurement fault signal (ERR) is output at the EXT I/O connector. See Section 5.2.4 ERR Output (p.75).
A measurement fault is displayed in the following cases.
• When a test lead is not connected to the test object
• When the resistance of the measured object is over-range
Example: Attempting to measure 30 with the 300 m range selected.
• If any of the following is open, or has a bad connection: SOURCE-
H, SOURCE-L, SENSE-H, SENSE-L
• When the resistance between SOURCE-H and SOURCE-L is 50
or more in the 300 m range (or 500 or more in the 3 range)
• When the resistance between SENSE-H and SENSE-L is greater
than about 20  (However, if the capacitance of the test leads is 1 nF or higher, the measurement fault may not be detected.)
• When a bad contact results from damage, excessive wear or
impurities on the test leads.
• If the circuit protection fuse is blown
See Section 9.1 Troubleshooting (p.169).

3.7.2 Overflow Display

Overflow is indicated by “OF” or “-OF” on the display, caused by one of the following:
Display Condition
OF
-OF
• The measured value exceeds the limit of the current measurement range
• When the result of relative value calculation is larger than +99.999%.
• The measured value is below the limit of the current measurement range
• When the result of relative value calculation is smaller than -99.999%.
Page 41
Applied
HIOKI 3561A981-10
35
Measurement
This chapter describes advanced operations employing the Comparator, Statistical Calculation and Memory functions.
Judge measurement values against specified thresholds
Measure when trigger events occur Trigger Function
Output averaged measurement values
Display the results of calculation expressions applied to measurement values
Store measurement values Memory Function
Lock the keys Key-Lock Function
Chapter 4
Comparator Function
Averaging Function
Statistical Calculation Functions
(36 page)
(53 page)
(55 page)
(56 page)
(60 page)
(62 page)
1
2
3
4

Applied Measurement

5
6
Save measurement configurations Panel Save Function
Load saved measurement configurations
Increase measurement precision Self-Calibration
Output measurement values via the RS-232C interface according to trigger input timing
Enable/disable key-press beeps Key Beeper Setting
Re-initialize the instrument Reset Function
Panel Load Function
Measurement Value Output Function
(63 page)
(64 page)
(65 page)
(66 page)
(67 page)
(68 page)
7
8
9
10
Page 42
36
HIOKI 3561A981-10

4.1 Comparator Function

4.1 Comparator Function
The comparator function compares measured values to preset upper and lower thresholds, judges the measurements according to their relative levels within the preset range, and indicates the results of the comparisons. Comparator thresholds can be set either by specifying upper and lower thresholds, or by specifying a reference value and tolerance. Comparator results can be indicated by the Hi, IN and Lo LEDs, beeper sound and signal output at the EXT I/O connector. See Section Chapter 5 External Control (p.71).
The comparator setting process flow is as follows:
Display the comparator settings
1
Set the comparator judgment beeper
2
Select the comparator execution mode
3
(Auto or Manual/External)
Select resistance or voltage measurement
4
Select the comparison method
5
(absolute or relative value) for the comparator
Specify the upper and lower thresholds
6
(or reference value and tolerance)
Apply your comparator settings
7
Enable the Comparator function
8
Page 43
37
HIOKI 3561A981-10
4.1 Comparator Function

4.1.1 Comparator Setting Example 1 (Upper and Lower Threshold Judgment)

This example describes the comparator setting method.
Example: Set the upper and lower thresholds for resistance and voltage in the V mode (300 m range), and indicate whether the measurement value exceeds the upper or lower thresholds by sounding the beeper.
Resistance : Upper threshold value 150.00 m, Lower threshold value 100.00 m Voltage : Upper threshold value 15.2000 V, Lower threshold value 15.0000 V
1
Confirm that the Comparator function is OFF.
First make sure the Comparator function is disabled. Settings cannot be changed while the Comparator function is enabled. Press the COMP key, if necessary, to disable the Comparator function.
COMP not lit
1
2
3
4
Applied Measurement
5
6
2
3
4
Select the V measurement mode.
V lit
Select the measurement range (for this example, the 300 m range).
m lit
Selects the 3 range ( lit)
Selects the 300 m range (m lit)
The Comparator setting display appears.
oFF flashing
7
8
9
10
Page 44
38
HIOKI 3561A981-10
4.1 Comparator Function
5
6
Set the comparator judgment beeper (for this example, select HL).
HL flashing
oFF........ no beeps sound
In ...........beeps continuously (when measurements are IN)
HL.......... beeps repeatedly (when measurements are Hi or Lo)
btH1.......beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2.......beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
Press so that the indicated position blinks, and select the comparator execution mode (for this example, Auto).
A flashing
7
8
A............ Auto Comparator (default setting)
E ............ Manual Comparator
Press so that the indicated position blinks, and select resistance.
r flashing
r .............Resistance
u ............Voltage
Press so that the indicated position blinks, and select the comparison method for the comparator (here, HIGH/LOW).
HIGH & LOW flashing
HIGH, LOW ..... Compare by upper and lower thresholds (default setting)
REF, %............. Compare by reference value and tolerance
Page 45
4.1 Comparator Function
HIOKI 3561A981-10
39
9
Or
ten-keys
Switch to the upper/lower threshold setting display, and specify the thresholds.
For this example, Upper Threshold: 150 m
Upper Threshold: 100 m
Using the RANGE keys: Using the ten-keys:
Select a digit to change by moving the blinking location, then select the new numerical value.
Select a digit
Select numerical value
To enter the current measurement as the setting value: AUTO key To enter the result of statistical calculation as the setting value: STAT key
See Section 4.1.6 Upper and Lower Thresholds Setting (by Reference Value and
Tolerance) (p.48).
Press the numeric keys corresponding to the digits to be entered.
1
2
3
4
Applied Measurement
5
6
10
11
Press so that the indicated position blinks, and select voltage.
7
u flashing
8
r............. Resistance
u ............ Voltage
9
Press so that the indicated position blinks, and select the comparison method for the comparator (here, HIGH/LOW).
10
HIGH & LOW flashing
HIGH, LOW..... Compare by upper and lower thresholds (default setting)
REF, % ............ Compare by reference value and tolerance
Page 46
40
HIOKI 3561A981-10
4.1 Comparator Function
12
13
Or
ten-keys
Switch to the upper/lower threshold setting display, and specify the thresholds.
For this example, Upper Threshold: 15.2 V
Upper Threshold: 15 V
Applies setting and returns to the Measurement display. The comparator function is enabled.
COMP lit
To cancel the settings: SHIFT key
14
Connect a test object and judge the measured value.
Measured Resistance
Judgment Result
Measured Voltage
In the
V mode, you can verify comparator settings by pressing the VIEW key.
See Section 4.1.9 Switching Between Measurement Value and Comparator
Setting Displays (p.52).
Upper Threshold Value < Measured Value
Lower Threshold Value
Measured Value < Lower Threshold Value
Measured Value Upper Threshold Value
The upper and lower thresholds are saved as the displayed counts (independent of measurement mode and range). Therefore, changing the measurement mode or range results in the same display counts representing different absolute values. Example: To specify the lower threshold as 150 m in the 300 m range, enter “15000”. Switching to the 3  range after making this setting changes the lower threshold to 1.5 .
Page 47
4.1 Comparator Function
HIOKI 3561A981-10

4.1.2 Comparator Setting Example 2 (Reference Value and Tolerance Judgment)

41
This example describes the comparator setting method.
Example: Set a reference value and tolerance in the V mode (3 range), and set the beeper to sound while measured values are within tolerance.
Resistance : Reference value 1.5 , Tolerance 5% Voltage : Reference value 4.2 V, Tolerance 0.5%
1
Confirm that the Comparator function is OFF.
First make sure the Comparator function is disabled. Settings cannot be changed while the Comparator function is enabled. Press the COMP key, if necessary, to disable the Comparator function.
COMP not lit
1
2
3
4
Applied Measurement
5
6
2
3
4
Select the V measurement mode.
V lit
Select the measurement range (for this example, the 3  range).
lit
Selects the 3 range ( lit)
Selects the 300 m range (m lit)
The Comparator setting display appears.
oFF flashing
7
8
9
10
Page 48
42
HIOKI 3561A981-10
4.1 Comparator Function
5
6
Set the comparator judgment beeper (for this example, select In).
In flashing
oFF........ no beeps sound
In ...........beeps continuously (when measurements are IN)
HL.......... beeps repeatedly (when measurements are Hi or Lo)
btH1.......beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2.......beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
Press so that the indicated position blinks, and select the comparator execution mode (for this example, Auto).
A flashing
7
8
A............ Auto Comparator (default setting)
E ............ Manual Comparator
Press so that the indicated position blinks, and select resistance.
r flashing
r .............Resistance
u ............Voltage
Press so that the indicated position blinks, and select the comparison method for the comparator (here, REF/%).
REF & % flashing
HIGH, LOW ..... Compare by upper and lower thresholds (default setting)
REF, %............. Compare by reference value and tolerance
Page 49
4.1 Comparator Function
HIOKI 3561A981-10
43
9
Or
ten-keys
Switch to the Ref/% threshold setting display, and specify the thresholds.
Relative (%) calculation expression
Measured resistance - Reference value
Reference value
For this example, Reference value: 1.5
Tolerance: 5%
Using the RANGE keys: Using the ten-keys:
Select a digit to change by moving the blinking location, then select the new numerical value.
Select a digit
Select numerical value
Press the numeric keys corresponding to the digits to be entered.
x 100
1
2
3
4
Applied Measurement
5
10
11
To enter the current measurement as the setting value: AUTO key To enter the result of statistical calculation as the setting value: STAT key See Section 4.1.6 Upper and Lower Thresholds Setting (by Reference Value and
Tolerance) (p.48).
Press so that the indicated position blinks, and select voltage.
u flashing
r............. Resistance
u ............ Voltage
Press so that the indicated position blinks, and select the comparison method for the comparator (here, REF/%).
6
7
8
9
10
REF & % flashing
HIGH, LOW..... Compare by upper and lower thresholds (default setting)
REF, % ............ Compare by reference value and tolerance
Page 50
44
HIOKI 3561A981-10
4.1 Comparator Function
12
13
Or
ten-keys
Switch to the Ref/% threshold setting display, and specify the thresholds.
Relative (%) calculation
Measured voltage - Reference value
Reference value
For this example, Reference value: 4.2 V
Tolerance: 0.5%
Applies setting and returns to the Measurement display. The comparator function is enabled.
COMP lit
To cancel the settings: SHIFT key
x 100
14
Connect a test object and judge the measured value.
Resistance measurements are displayed as their relative percentage offset from the reference value (%)
Judgment Result
Relative percentag
Voltage measurements are displayed as their relative percentage offset from the reference value (%)
V mode, you can verify comparator settings by pressing the VIEW key.
In the See Section 4.1.9 Switching Between Measurement Value and Comparator Setting
Displays (p.52).
Upper Threshold Value of setting range < Measured value
Lower Threshold Value of setting range Measured value Upper Threshold Value of setting range
Measured value < Lower Threshold Value of setting range
Measured resistance - Reference value
=
Reference value
x 100
Page 51
4.1 Comparator Function
HIOKI 3561A981-10

4.1.3 Comparator Judgment Beeper Setting

Four beeper settings are available to audibly indicate comparator judgment results.
45
1
1
2
(SHIFT Lamp lit)
The Comparator setting display appears.
Set the comparator judgment beeper.
(Main Display)
oFF........ no beeps sound
In ........... beeps continuously (when measurements are IN)
HL.......... beeps repeatedly (when measurements are Hi or Lo)
btH1 ...... beeps continuously while measurements are within the thresholds (IN), and
beeps repeatedly when measurements are Hi or Lo.
btH2 ...... beeps once when measurements move into the threshold range (IN), and beeps
repeatedly when measurements go Hi or Lo.
• The beeper does not sound when the comparator judgment beeper setting is disabled (oFF).
• The beeper does not sound when there is no judgment result. See Section 4.1.8 Comparator Judgment Results (p.51).
2
3
4
Applied Measurement
5
6
7
8
9
10
Page 52
46
HIOKI 3561A981-10
4.1 Comparator Function

4.1.4 Comparator Execution Mode Setting

Comparator judgment execution is selected by setting the auto or manual/external comparator mode. Comparator judgment can be enabled and disabled by EXT I/O signals. Refer to 5.2.2 Input Signals (p.73).
1
2
(SHIFT Lamp lit)
The Comparator setting display appears.
Press so that the indicated position blinks, and set the comparator execution mode.
(Main Display)
A............ Auto comparator (comparator results are always output [default setting])
E ............Manual comparator (comparator results are output only when the MANU EXT I/
O input is enabled [ON])
The auto setting is appropriate for normal use. Use the manual/ external setting when you need to control comparator judgment timing.
Page 53
4.1 Comparator Function
HIOKI 3561A981-10

4.1.5 Comparator Threshold Method Selection

Two methods are available for setting comparator thresholds.
47
1
1
2
(SHIFT Lamp lit)
The Comparator setting display appears.
Press so that the indicated position blinks, and set the comparator threshold method.
HIGH, LOW..... Compare against specified upper and lower thresholds (default setting
method)
REF, % ............ Compare against upper and lower thresholds internally calculated from a
specified reference value and tolerance
About comparisons based on a reference value and tolerance
When the reference value and tolerance method is selected, thresholds are calculated as follows: Upper threshold = reference value
Lower threshold = reference value Measured values are displayed as a percentage relative to the reference value, calculated as follows: Relative value = (measured value - reference value) / reference value
× (100 + tolerance [%]) / 100 ×(100 - tolerance [%]) / 100
× 100 [%]
2
3
4
Applied Measurement
5
6
7
8
9
10
Page 54
48
HIOKI 3561A981-10
4.1 Comparator Function

4.1.6 Upper and Lower Thresholds Setting (by Reference Value and Tolerance)

1
2
3
(SHIFT Lamp lit)
The Comparator setting display appears.
Press so that the indicated position blinks, and select resistance or voltage.
r .............Resistance
u ............Voltage
Select the threshold setting display, and enter upper and lower threshold values.
Or
ten-keys
For example, Upper Threshold: 150 m
Lower Threshold: 100 m
Using the RANGE keys: Using the ten-keys:
Select a digit to change by moving the blinking location, then select the new numerical value.
Select a digit
Select numerical value
Press the numeric keys corresponding to the digits to be entered.
Page 55
4.1 Comparator Function
HIOKI 3561A981-10
To enter the current measurement as the setting value: AUTO key
The current measurement value is set as the upper or lower threshold (during upper/lower threshold setting), or as the reference value (during reference value and tolerance setting). If the measured value is faulty or
To enter a statistical calculation result as the setting value:
The result of statistical calculation is set as follows:
± OF, it is ignored (not entered).
STAT key
49
1
During upper/lower threshold setting
During reference value and tolerance setting
Where “ No setting occurs if statistical calculation is disabled and no
statistical calculation result exists. See Section 4.4 Statistical Calculation Functions (p.56).
Setting thresholds from the when the selected (blinking) character is non-numeric.
Threshold and reference values can be set from 0 to 99999 (or 999999 for voltage), and tolerance can be set from 0.000 to
99.999%. Negative values are not settable. Entries using statistical calculation results that exceed the valid range are restricted to the range limit.
” represents population standard deviation (
Upper threshold = average value + 3 Lower threshold = average value - 3
Reference value = average value Tolerance = 3
AUTO and STAT keys is possible only
/ average value X 100%
n
).
2
3
4
Applied Measurement
5
6
7
8
9
10
Page 56
50
HIOKI 3561A981-10
4.1 Comparator Function

4.1.7 Enabling and Disabling the Comparator Function

Enables the comparator
COMP lit
When the comparator is enabled, the following key operations are disabled to avoid inadvertent operations.
V//V key (Measurement mode setting)
SHIFT V//V key (Zero-Adjustment) SHIFT COMP key (Comparator setting)
AUTO key (Auto-ranging setting)
SMPL key (Sampling rate setting)
SHIFT SMPL key (Averaging setting) SHIFT TRIG key (Trigger source setting)
SHIFT ENTER key (Menu display)
SHIFT STAT key (Delay setting)
• Range keys
When the comparator is enabled, auto-ranging is automatically disabled.
Page 57
4.1 Comparator Function
HIOKI 3561A981-10

4.1.8 Comparator Judgment Results

Resistance and voltage measurements are judged independently. Both judgment results are indicated on the display.
Judgment Operation The comparator compares measured values with the preset threshold
values, and judges whether the measurement is within the thresholds. Resistance and voltage measurements are judged independently. The absolute value of the measurement is compared to the upper and lower thresholds.
Upper Threshold Value < Measured Value
Lower Threshold Value
Measured Value < Lower Threshold Value
Measured Value Upper Threshold Value
51
1
2
3
4
Applied Measurement
AND Judgment
Output
Measurement fault values are judged as follows:
Display Judgment
- - - - -
OF
-OF
Judgment results (Hi, IN or Lo for both resistance and voltage) are output to EXT I/O connectors. Also, to facilitate application of judgment results, an AND output terminal indicates when both resistance and voltage are IN (within the threshold range). See Section 5.2.3 Output Signals (p.74).
With the relative value comparison method (thresholds defined by a reference value and tolerance), the upper and lower thresholds are calculated internally for comparison with measurements. Therefore, even if a relative display value is equal to a judgment threshold (tolerance limit), it may be judged Hi or Lo.
No judgment
Hi (exceeds the upper threshold)
Lo (less than the lower threshold)
5
6
7
8
9
10
Page 58
52
HIOKI 3561A981-10
4.1 Comparator Function

4.1.9 Switching Between Measurement Value and Comparator Setting Displays

In V mode, both resistance and voltage measurement values are displayed. Although comparator setting values are not normally displayed when the comparator is enabled, they can be displayed for confirmation by the display switching function.
Press this key to switch the display between measurement values and comparator setting values.
Resistance and voltage measurement display (Shows resistance and voltage measurement values simultaneously)
Resistance measurement and comparator display (Shows resistance measurement and resistance comparator setting values)
Voltage measurement and comparator display (Shows voltage measurement and voltage comparator setting values)
Measurement display switching is available only with the comparator enabled, and in the V mode.
Use it to confirm comparator setting values.
Page 59
4.2 Trigger Function
HIOKI 3561A981-10

4.2.1 Trigger Source Settings

53

4.2 Trigger Function

1
Measurement with
External Triggering
Two trigger sources are available: internal and external.
Internal Trigger Trigger signals are automatically generated internally.
(free-run)
External Trigger Trigger signals are provided externally or manually.
(SHIFT Lamp lit)
Switches the selected trigger source.
EXT TRIG lit
EXT.TRIG
EXT.TRIG not lit........Internal triggering is selected.
An external trigger can be applied in three ways.
• Applying a trigger manually by operating key Pressing the
• Applying a trigger at the EXT I/O connector Grounding the panel causes one measurement. See Section 5.2.2 Input Signals (p.73).
• Applying a trigger through RS-232C or GP-IB interface Sending the causes one measurement.
lit..............External triggering is selected.
TRIG key causes one measurement.
TRIG terminal of the EXT I/O connector on the rear
TRG command via the RS-232C or GP-IB interface
2
3
4
Applied Measurement
5
6
7
8
9
• When Internal triggering is enabled, external input at the EXT I/O
TRIG terminal and the TRG command are ignored.
• The normal state of operation with the front panel controls is continuous measurement. Setting the trigger source to Internal enables the free-run condition in which triggering occurs continuously. When the trigger source is set to External, a measurement occurs each time an external trigger is applied. Continuous measurement can be disabled via RS-232C or GP-IB interface signals, in which case triggering occurs only when signaled by the external host (PC or PLC). See Section Triggering System Description (p.136).
10
Page 60
54
HIOKI 3561A981-10
4.2 Trigger Function

4.2.2 Trigger Delay Settings

Specify the delay from the moment a trigger is applied to the start of measurement. By using this function, even when a trigger is applied immediately after connecting a test object, the start of measurement can be delayed to allow sufficient time for the measurement value to stabilize. Trigger delay can be set with 1 ms resolution from 0.000 to 9.999 seconds.
1
2
3
(SHIFT Lamp lit)
The Trigger Delay setting display appears.
(Main Display)
(Sub Display) The current setting blinks.
Select ON.
(Sub Display)
The numerals indicating the trigger delay blink.
(Sub Display)
4
Or ten-keys
5
Set the trigger delay.
Applies setting and returns to the Measurement display. To cancel the settings: SHIFT key
Disabling the Trigger Delay Function
1
2
3
(SHIFT Lamp lit)
The Trigger Delay setting display appears.
Select OFF.
The Trigger Delay is disabled.
(Sub Display)
Page 61
55
HIOKI 3561A981-10

4.3 Averaging Function

4.3 Averaging Function
The Averaging Function averages measurement values for output. This function can minimize instability of displayed values. The number of samples to average can be set from 2 to 16.
1
1
2
3
4
Or
ten-keys
(SHIFT Lamp lit)
The Averaging Function setting display appears.
(Main Display)
(Sub Display) The current setting blinks.
Select ON.
(Sub Display)
The number of samples to average setting blinks.
Select the number of samples to average.
2
3
4
Applied Measurement
5
6
7
5
Disabling the Averaging Function
1
2
3
The Average Measurement display appears. (AVG lit) To cancel the settings: SHIFT key
(SHIFT Lamp lit)
The Averaging Function setting display appears.
Select OFF.
The Averaging Function is disabled. (AVG not lit)
When the internal trigger is used for continuous measurement (free­run), the display shows the moving average. Otherwise, the display shows the integrating average. See Section 4.2 Trigger Function (p.53).
8
9
10
(Sub Display)
Page 62
56
n
x
x
n
xnx
22
122
n
xnx
s
16
n
LoHi
Cp
1
6
2
n
xLoHiLoHi
CpK
HIOKI 3561A981-10

4.4 Statistical Calculation Functions

4.4 Statistical Calculation Functions
The mean, maximum, minimum, standard deviation of population, standard deviation of sample and process capability indices are calculated and displayed for up to 30,000 measurement values.
The calculation formulas are as follows:
Mean
Standard deviation of population
(=
)
n
Standard deviation of sample (=
Process capability index (dispersion)
Process capability index (bias)
• In these formulas, n represents the number of valid data samples.
• Hi and Lo are the upper and lower thresholds of the comparator.
• The process capability indices represent the quality achievement capability created by a process, which is the breadth of the dispersion and bias of the process' quality. Generally, depending on the values of Cp and CpK, process capability is evaluated as follows:
Cp, CpK>1.33................... Process capability is ideal
1.33 Cp, CpK>1.00........ Process capability is adequate
1.00 Cp, CpK................. Process capability is inadequate
• When only one valid data sample exists, standard deviation of
sample and process capability indices are not displayed.
• When
• The upper limit of Cp and CpK is 99.99. Values of Cp and
CpK>99.99 are displayed as 99.99.
• Negative values of CpK are handled as CpK=0.
• When comparator, range or auto-ranging settings are changed
while statistical data is displayed, the display of Cp and CpK values changes to “- - . - -”.
• When normal measurement values and relative display values (%)
are mixed, correct calculation results cannot be obtained.
is 0, Cp and CpK are 99.99.
n-1
)
n-1
Page 63
4.4 Statistical Calculation Functions
HIOKI 3561A981-10
Enabling/Disabling the Statistical Calculation Function
57
1
2
(press three times)
3
The Statistical Calculation display appears.
(Main Display)
(Sub Display)
The function enable/disable display appears.
(Sub Display)
Enable or disable the Calculation Function on the Sub Display.
on.......... enables the calculation function (ON).
off ......... disables the calculation function (OFF).
Applies setting and returns to the Measurement display. To cancel the settings: SHIFT key
• Statistical Calculation function setting (ON, OFF) is not available when the Comparator is enabled.
• If Statistical Calculation is turned off and then back on without first clearing calculation results, it resumes calculating from the point when it was turned off.
• The Statistical Calculation function slows measurements when it is ON.
1
2
3
4
Applied Measurement
5
6
7
8
Clearing Statistical Calculation Results
1
2
(press once)
3
The Statistical Calculation display appears.
The Clearing screen will appear.
Clears statistical calculation results.
(Main Display)
(Sub Display)
(Sub Display)
9
10
Page 64
58
HIOKI 3561A981-10
4.4 Statistical Calculation Functions
Automatic Clearing of Statistical Calculation Results after Printing
The instrument can be set to automatically clear statistical calculation results after results are output to the printer.
1
2
3
4
(Press twice)
The Statistical Calculation display appears.
(Main Display)
(Sub Display)
Bring up Auto Clearing After Printing in the Setup screen.
(Sub Display)
Turn Automatic Clearing After Printing on or off.
on.......... Automatically clears statistical calculation results after they are
output to the printer.
off.......... Does not clear the results themselves.
Applies setting and returns to the Measurement display. To cancel the settings: SHIFT key
Importing Data
Pressing the TRIG key while Statistical Calculation is ON executes one of the following operations:
• External Trigger: Takes one measurement and performs statistical calculation on the result
• Internal Trigger: Performs statistical calculation on the value displayed immediately after pressing
TRG command executes the same operation.
• Grounding the the same operation.
TRIG terminal of the EXT I/O connector executes
Confirming Statistical Calculation Results
1
2
The Statistical Calculation display appears.
The indication on the display changes as follows with each key­press.
Page 65
Example: when the V mode is selected
HIOKI 3561A981-10
(not displayed in V mode)
Total data count of resistance measurement
Mean of resistance measurement Maximum of resistance measurement
59
4.4 Statistical Calculation Functions
Valid data Mean
Minimum of resistance measurement Standard deviation of population
Minumum Data Sample No.
Process capability indices of resistance measurement
Cp CpK
of resistance measurement
(not displayed in mode)
Total data count of voltage measurement
Minimum of voltage measurement
Process capability indices of voltage measurement
Standard deviation of population of voltage measurement
Mean of voltage measurement
Maximum Data Sample No.
Standard deviation of sample of resistance measurement
Maximum of voltage measurement
Standard deviation of sample of voltage measurement
1
2
3
4
Applied Measurement
5
6
ON/OFF setting Auto Clearing After Printing setup Clear setup
• When a valid data count (measurement fault other than ± OF) is
zero, no calculation result is displayed.
• When only one valid data sample exists, standard deviation of sample and process capability indices cannot be displayed.
• When comparator, range or auto-ranging settings are changed while statistical data is displayed, the display of Cp and CpK values changes to “- - . - -”.
Sending Statistical Calculation Results to the Printer
With the statistical calculation results displayed, press the PRINT key. The statistical calculation results are output to the optional printer.
See Section Chapter 6 Printing (p.81).
7
8
9
10
Page 66
60
HIOKI 3561A981-10

4.5 Memory Function

4.5 Memory Function
The Memory function is only available via communication commands. When the Memory function is enabled, measurement values are stored in the instrument’s internal memory according to trigger input sequence (up to 400 values). Stored data can be downloaded later upon command. When measuring using a scanner to switch multiple test objects, switching time can be quite long if measurement values are downloaded to the PC after each measurement. Test cycle time can be minimized by using this function to store measurement values internally until all channel measurements are finished, at which time the stored values are downloaded together during the next idle period.
1
2
3
4
Select the RS-232C or GP-IB interface.
See Section 7.3.2 Selecting the Interface (p.91).
Send the command to enable the Memory function.
:MEMory:STATe ON
The MEM indicator lights.
MEM lit
Measurement values are stored.
When a trigger is applied by the TRIG key, TRIG
TRG command, the MEM indicator blinks once and the measured value is
stored.
EXT I/O input signal or
MEM lit
If an external trigger source is selected, one measurement is stored after each trigger event. In the internal triggering case, the first measurement value after triggering is stored. Apply a trigger as many times as is necessary.
Page 67
4.5 Memory Function
HIOKI 3561A981-10
61
5
Send the command to download the data from memory.
:MEMory:DATA?
The stored measurement values are returned in response.
:MEM:DATA? 1, 290.60E-3, 1.3924E+0 2, 290.54E-3, 1.3924E+0 3, 290.50E-3, 1.3923E+0 4, 290.43E-3, 1.3923E+0 5, 290.34E-3, 1.3924E+0 END
The “END” character is sent as the last line of the data.
To download stored data one measurement at a time, send this command:
:MEMory:DATA? STEP
The instrument sends one stored data object and enters the wait state. When the instrument receives an “N” from the PC or other device, the next stored data object is sent. Repeat until the last data object is downloaded. When all stored data has been downloaded, the instrument sends an “END” character.
:MEM:DATA? STEP 1, 290.60E-3, 1.3924E+0
N 2, 290.54E-3, 1.3924E+0
N 3, 290.50E-3, 1.3923E+0
N 4, 290.43E-3, 1.3923E+0
N 5, 290.34E-3, 1.3924E+0
N END
(sent from PC)
(sent from PC)
(sent from PC)
(sent from PC)
(sent from PC)
1
2
3
4
Applied Measurement
5
6
6
To clear the instrument's memory, send it the following command.
:MEMory:CLEAr
Unless the memory is cleared, measurement data continues to be stored upon each trigger event.
• The instrument's memory storage capacity is 400 measurements. Be aware that attempting to store more data (by applying a trigger) results in nothing further being stored.
• Refer to Chapter 7 RS-232C/GP-IB Interfaces (p.87), for details about the communication methods and sending and receiving commands.
• When the Memory function is enabled, auto-ranging is not available.
• Memory contents are cleared when performing the following operations: When enabling the Memory function (off to on) When changing the measurement range When changing comparator settings When sending the When Reset is executed from the menu display When sending
When sending :SYSTem:RESet When turning power on
:Memory:Clear command
RST
7
8
9
10
Page 68
62
HIOKI 3561A981-10

4.6 Key-Lock Function

4.6 Key-Lock Function
Executing Key-Lock disables the operating keys on the front of the instrument. This function can be useful for protecting settings.
(SHIFT Lamp lit)
Enable the Key-Lock function.
LOCK lit
Disabling Key-Lock
• Even if the power supply is interrupted, the Key-Lock function is not canceled.
• The
(SHIFT Lamp lit)
Disable the Key-Lock function. (LOCK is not lit)
TRIG key remains operational.
When communicating by remote control, the remote control status is canceled.
Page 69
4.7 Panel Save Function
HIOKI 3561A981-10
63

4.7 Panel Save Function

1
2
The current measurement setting state is stored (saved) in non­volatile memory. Up to 126 sets of measurement states can be saved. The measurement settings (state) at the time this function is executed are saved. Saved measurement states can be reloaded using the Panel Load function, described later.
(SHIFT Lamp lit)
The Panel Saving display appears.
(Main Display)
(Sub Display) The panel number blinks.
Select the panel number to save.
1
2
3
4
Applied Measurement
5
6
3
Saved Items
Or
ten-keys
(Sub Display) (To save measurement settings as Panel No. 3)
When selecting a saved panel, “USEd” is displayed.
Saves the measurement setting state and returns to the Measurement display. To cancel the settings: SHIFT key
• If you select a Panel number that was previously saved and press
ENTER key, the contents are overwritten.
the
• The Key-Lock state can be saved only by the remote command.
• Measurement mode setting • Switching displays setting
• Range setting • Delay setting
• Auto-ranging setting • Zero-Adjust setting
• Sampling rate setting • Averaging setting
• Comparator settings • Key-Lock
• Internal/External trigger setting • Statistical Calculation setting
:SYSTem:SAVE
7
8
9
10
Page 70
64
HIOKI 3561A981-10

4.8 Panel Load Function

4.8 Panel Load Function
Loads the measurement settings saved by the Panel Save function from internal non-volatile memory.
1
2
3
Or
ten-keys
The Panel Loading display appears.
(Main Display)
(Sub Display) The panel number blinks.
Select the panel number to load.
(Sub Display) (To load measurement settings from Panel No.3)
Loads the measurement setting state and returns to the Measurement display. To cancel the settings: SHIFT key
• If an unsaved Panel No. is selected, a warning beep sounds when you press
• When selecting a Panel No. with the up/down the numbers of previously saved panels appear.
• Loading can also be executed using the
LOAD0 to LOAD6 pins of the EXT I/O interface.
See Section 5.2.2 Input Signals (p.73).
ENTER key.
RANGE keys, only
TRIG signal and the
Page 71
4.9 Self-Calibration
HIOKI 3561A981-10
65

4.9 Self-Calibration

1
The self-calibration function adjusts offset voltage and gain drift of the instrument’s internal circuitry to improve measurement precision. The instrument’s measurement accuracy specifications depend on self-calibration, so it must be executed frequently. In particular, always execute self-calibration after warm-up and when the ambient temperature changes by more than 2 setting, self-calibration is executed during every measurement when SLOW sampling is used.
Self-calibration can be executed by the following two methods:
Auto Executes self-calibration automatically once every 30
minutes.
Manual Self-calibration can be executed manually by applying a
CAL input signal (grounding the CAL terminal of the EXT
I/O connector).
(SHIFT Lamp lit)
The Menu display appears.
°C. However, regardless of this
1
2
3
4
Applied Measurement
5
6
2
3
4
The Self-Calibration setting display appears.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display) The current setting blinks.
Select Auto or Manual on the Sub Display.
AUto ..... Auto self-calibration
In........... Manual self-calibration
Applies setting and returns to the Measurement display.
Self-calibration requires about 55 ms, during which measurement processing is temporarily suspended.
7
8
9
10
Page 72
66
HIOKI 3561A981-10

4.10 Measurement Value Output Function

4.10 Measurement Value Output Function
This function causes output of measured values via the RS-232C interface in the same sequence as trigger input. This function is useful when measuring using internal (free-run) triggering, and for obtaining measured values on a PC when using a footswitch for triggering.
1
2
3
4
(SHIFT Lamp lit)
The Menu display appears.
The Measurement Value Output function setting display appears.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display) The current setting blinks.
Turn Measurement Value Output Function on or off.
on.......... enables the measurement value output function (ON).
off.......... disables the measurement value output function (ON).
Applies setting and returns to the Measurement display.
5
The measured value is output from the RS-232C interface when you press the TRIG key or when a signal is applied to the EXT I/O TRIG terminal. Set the PC to the receiving state beforehand. When a measurement value is received, the PC should perform appropriate processing such as recording or displaying, then re-enable the receiving state.
• When external triggering is enabled, a measurement is performed and the value is sent after each trigger event. When internal triggering is enabled, the first value measured after triggering is sent.
• The measurement output function is not applicable to the GP-IB interface or printer.
Page 73

4.11 Key Beeper Setting

HIOKI 3561A981-10
67
4.11 Key Beeper Setting
1
2
3
Select whether a beep sounds when an operating key on the front of the instrument is pressed.
(SHIFT Lamp lit)
The Menu display appears.
The Key Beeper setting display appears.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display) The current setting blinks.
Select the key beeper state on the Sub Display.
on.......... Key beeper enabled
off ......... Key beeper disabled
1
2
3
4
Applied Measurement
5
6
4
Applies setting and returns to the Measurement display.
7
8
9
10
Page 74
68
HIOKI 3561A981-10

4.12 Reset Function

4.12 Reset Function
The reset function can be used to re-initialize current measurement settings (excluding saved panel data) to their factory defaults, or to re­initialize all measurement settings including saved panel data to factory defaults.
1
2
3
4
(SHIFT Lamp lit)
The Menu display appears.
The Reset display appears.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display) The current setting blinks.
Select the Reset method on the Sub Display.
SEt ........ Reset (initializes measurement settings other than those stored
with Panel Save)
SYS....... System Reset (initialize all measurement settings)
ENTER blinks.
5
(Sub Display)
When SYS (system reset) is selected
Executes the Reset.
To cancel the settings: SHIFT key
System Reset also initializes Panel Save data.
Page 75
4.12 Reset Function
HIOKI 3561A981-10
Initial Factory Default Settings
Description Default
Measurement Mode V
69
Resistance Measurement Range AUTO
Zero-Adjust OFF
Zero-Adjust Value 0
Delay OFF
Delay Time 0.000s
Sampling Rate SLOW
Averaging Function OFF
Average Times 2
Self-Calibration AUTO
Continuous Measurement ON
Trigger Source Internal trigger
Line Frequency 50 Hz
Key Beeper Setting ON
Key-Lock Function OFF
Comparator OFF
Comparator Threshold Method (resistance and voltage)
Hi, Lo
1
2
3
4
Applied Measurement
5
6
Comparator Upper Threshold (resistance and voltage)
Comparator Lower Threshold (resistance and voltage)
Comparator Judgment Beeper OFF
Comparator Execution Mode AUTO
Statistical Calculation Functions OFF
Automatic Clearing of Statistical Calculation Results
Interface RS-232C
Baud Rate 9600 bps
GP-IB Address 1
GP-IB Delimiter LF
Print Interval 0 (The interval print disabled)
Error Output ASync
Measurement Value Output Function OFF
EOC Output HOLD
0
0
OFF
7
8
9
10
EOC Pulse Width 1 ms
Page 76
70
HIOKI 3561A981-10
4.12 Reset Function
Page 77

5.1 Overview

HIOKI 3561A981-10
External ControlChapter 5
71
1
5.1 Overview
External Control Input Functions
External Output Terminal Functions
Connector Type 57RE-40360-730B (D29) (manufactured by DDK)
Mating Connector 57-30360 (manufactured by DDK)
• External trigger input (TRIG)
• Select Panel No. to load (
• Zero-adjust signal input (
• Print Signal input (
• Self-calibration signal input (
• Manual comparator judgment input (
• End-of-Conversion signal output (EOC)
• Reference signal output (INDEX)
• Measurement Fault signal output (ERR)
• Comparator decision signal output (R-Hi, R-IN, R-Lo, V-Hi, V-IN, V­Lo, AND)
• General-purpose outputs (OUT0 to OUT9)
RC30-36P (manufactured by HIROSE electric co.,ltd.) or equivalent
PRINT)
LOAD0 to LOAD6)
0ADJ)
CAL)
MANU)
2
3
4
5
5

External Control

6
7
To avoid electric shock or damage to the equipment, always observe the following precautions when connecting to EXT I/O connector.
• Always turn off the power to the instrument and to any devices to be connected before making connections.
• Be careful to avoid exceeding the ratings of EXT I/O connector.
• During operation, a wire becoming dislocated and contacting another conductive object can be serious hazard. Make sure that connections are secure the EXT I/O connectors.
• The INT.GND terminals (Page 73) are grounded, so if an external controller has a potential relative to ground, connection could cause a short-circuit accident.
80
Be aware of the following to avoid damaging the instrument:
• When using relays, always include diodes to absorb back emf.
• Always provide protective grounding for devices the connect to the EXT I/O connectors.
8
9
10
Page 78
72
HIOKI 3561A981-10

5.2 Signal Descriptions

5.2 Signal Descriptions

5.2.1 Pinout

118
36 19
EXT I/O Connector
Pin I/O Signal name
1 IN LOAD0
2 IN
3 IN
4 IN
5 IN
6 IN
7 INT.GND
8 INT.GND
9 OUT R-Hi
10 OUT R-Lo
11 OUT V-Hi
12 OUT V-Lo
13 OUT EOC
14 OUT OUT0
15 OUT OUT2
16 OUT OUT4
17 OUT OUT6
18 OUT OUT8
LOAD2
LOAD4
LOAD6
CAL (IN1)
(IN3)
PRINT
Pin I/O Signal name
19 IN
20 IN
21 IN
22 IN
23 IN
24 IN
25 INT.GND
26 INT.GND
27 OUT R-IN
28 OUT AND
29 OUT V-IN
30 OUT ERR
31 OUT INDEX
32 OUT OUT1
33 OUT OUT3
34 OUT OUT5
35 OUT OUT7
36 OUT OUT9
LOAD1
LOAD3
LOAD5
TRIG
(IN0)
0ADJ (IN2)
(IN4)
MANU
Page 79

5.2.2 Input Signals

HIOKI 3561A981-10
73
5.2 Signal Descriptions
LOAD0 to LOAD6 Select a Panel No. to load and apply a TRIG signal to load the selected
Panel No. and measure. When a TRIG signal is applied, if unchanged from the previous trigger event, panel settings are not loaded. In this case, using external triggering, one measurement is taken as usual when the
Panel No.
122 123 124 125 126
0: LOAD terminal is shorted to GND 1: LOAD terminal is unconnected, or
connected to 5 V
 When a TRIG signal is applied with LOAD0 to LOAD6 set to all 1's or all
0's, no Panel Load occurs.
• At least 70 ms is required for the settings to change after executing a Panel Load (the actual time depends on the particular range and sampling rate
• When set to external trigger mode, one measurement is taken upon load completion.
• The Panel Load function cannot be executed from when controlling the instrument via RS-232C or GP-IB (Remote State).
LOAD6 LOAD5 LOAD4 LOAD3 LOAD2 LOAD1 LOAD0
1 2 3 4 5 6 7 8
...
1 1 1 1 1 1 1 1 1 1 1 1 1 0
1 1 1 1 1 0 1 1 1 1 1 1 0 0 1 1 1 1 0 1 1 1 1 1 1 0 1 0 1 1 1 1 0 0 1 1 1 1 1 0 0 0 1 1 1 0 1 1 1
0 0 0 0 1 0 1 0 0 0 0 1 0 0 0 0 0 0 0 1 1 0 0 0 0 0 1 0 0 0 0 0 0 0 1 0 0 0 0 0 0 0
LOAD0 is the LSB, and LOAD6 is the MSB.
LOAD0 through LOAD6 are
TRIG signal is applied.
function,
).
LOAD0 to LOAD4
1
2
3
4
5
5
External Control
6
7
8
9
TRIG When the external trigger, one measurement is taken each time the
TRIG signal transitions from High to Low.
This trigger signal is ignored when internal triggering is enabled. Trigger functions are also available for statistical calculation, recording to memory and output of measured values (valid also with internal triggering).
CAL When manual self-calibration is selected with EX.FAST, FAST or
MEDIUM sampling rate, self-calibration begins when the transitions from High to Low. Self-calibration takes about 55 ms. When SLOW sampling is selected, the
See Section 4.9 Self-Calibration p.65).
CAL signal is ignored.
CAL signal
10
Page 80
74
HIOKI 3561A981-10
5.2 Signal Descriptions
0ADJ Zero adjustment executes once when the 0ADJ signal transitions from
High to Low.
PRINT The current measurement value prints when the PRINT signal
transitions from High to Low.
MANU When the MANU comparator mode is selected, comparator judgment
is enabled while the
See Section 4.1.4 Comparator Execution Mode Setting p.46).
MANU signal is Low.
IN0 to IN4 The TRIG, CAL, 0ADJ, PRINT and MANU signals can also serve as
general-purpose input terminals, read with the
See Section EXT I/O Input p.135).
:IO:IN? command.

5.2.3 Output Signals

ERR Indicates a measurement fault.
The Synchronous ERR output setting causes ERR output to be synchronous with EOC output, while with the Asynchronous ERR output setting causes ERR output to follow actual (asynchronous) contact of the probes with the test object. See Section 5.2.4 ERR Output p.75).
INDEX The INDEX signal is output during the Trigger Wait, Delay, Self-
Calibration and Calculation states. This signal is not output while measuring the resistance of test objects. This signal transitions from Off to On to indicate that the test object can be removed.
EOC This signal indicates the end of a measurement (End-Of-Conversion).
This signal indicates when comparator judgment results and ERR output (when SYNC is enabled) are available.
R-Hi, R-IN, R-Lo
These are the results of comparator decision.
V-Hi, V-IN, V-Lo
AND This signal indicates when both resistance and voltage judgment
results are IN (V mode). In the and V modes, this signal is the same as R-IN and V-IN outputs, respectively.
OUT0 to OUT9 The output signals are controlled by the :IO:OUT command.
INT.GND This is the instrument's internal ground.
See Section EXT I/O Output p.135).
• I/O signals should not be used while measurement settings have been changed.
• The EOC and INDEX signals are initialized (ON) at power on.
• If it is not necessary to change the measurement conditions, set
LOAD0 through LOAD6 to either Hi or Lo.
Page 81

5.2.4 ERR Output

HIOKI 3561A981-10
The ERR output signal indicates the occurrence of measurement fault conditions (such as open test leads, or a bad contact). There are two ERR output methods.
75
5.2 Signal Descriptions
1
Synchronized with EOC Output (SYNC)
Asynchronous with EOC Output (ASYNC)
Measurement faults detected while measuring (not while awaiting trigger or during delay or calculation intervals), are indicated by ERR output synchronous with EOC output (the end-of-measurement signal). ERR Output On: A measurement fault has prevented correct
measurement
ERR Output Off: Correct measurement obtained (OF or -OF: Out-of-
range cases are included)
Measurement faults (test lead connection conditions) are output in real
time. The output is asynchronous with the output. ERR Output On: Measurement fault condition (open test leads, or a
bad contact)
ERR Output Off: Test lead connections are normal
TRIG signal and EOC
2
3
4
5
5
External Control
6
7
8
9
10
Page 82
76
HIOKI 3561A981-10
5.2 Signal Descriptions

5.2.5 Instrument Settings

Measurement Fault Output Signal (ERR) Setting
1
2
3
4
(SHIFT Lamp lit)
The Menu display appears.
Select the ERR Output Selection display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) p.15).
Select the type of signal to be output on the Sub Display.
SynC .......... Synchronous output (synchronized with EOC output)
ASynC ....... Asynchronous output (not synchronized with EOC output)
Applies settings and returns to the Measurement display.
Setting the EOC Signal
(Main Display)
(Sub Display) The current setting blinks.
1
2
3
4
(SHIFT Lamp lit)
The Menu display appears.
Select the EOC-signal setup display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) p.15).
(Main Display)
(Sub Display) The current setting blinks.
Choose the output method for the EOC signal.
HoLd .......... Holds the EOC signal after measurement.
Go to Step 5.
PULSE ....... Outputs the specified pulse after measurement.
 Go to the next step.
(When PULSE is selected)
The number representing the pulse width of the EOC signal will start blinking. Set the pulse width in ms.
Or ten-keys
5
Applies settings and returns to the Measurement display.
Page 83
5.3 Timing Chart
HIOKI 3561A981-10
77

5.3 Timing Chart

External Trigger Timing Chart
Contact State
*1 ERR Output
(Err Output ASYNC Setting)
*3 TRIG
Measurement start signal
INDEX Output
Reference Signal
End-of-Measurement Signal
Comparator Result
*2 ERR Output
(Err Output SYNC Setting)
Input
EOC Output
t1 t1
1
Contact
Open
2
3
t2
4
Measuring
End of Conversion
5
5
External Control
6
t3
t4 t5 t6
*1: For details, see “5.2.4 ERR Output p.75).” *2: When ERR output is set to the SynChronous mode, measurement fault detection results can be obtained
when measurement is finished, as with comparator results.
*3: After connecting to the test object, wait for longer than the response time (approximately 3 ms) before
inputting the TRIG signal (It is necessary to wait out the response time for the measurement values to stabilize after connection. Response times depend on the test object).
Internal Trigger Timing Chart
* INDEX Output
Reference Signal
* EOC Output
End-of-Measurement
Signal
Comparator Result
t3
Measuring Measuring
t4 t5
t6
7
8
9
10
* When the EOC signal is set to PULSE, the signal will remain on only for the specified period upon
completion of conversion.
Page 84
78
HIOKI 3561A981-10
5.3 Timing Chart
Description
t1
ERR Output response
*1
time
Measurement trigger
t2
pulse width
Delay Time per setting
t3
Measurement time
t4
Calculation time
t5
EOC Output pulse width When the external trigger is selected
t6
*2
*3
Time
1.5 ms
0.5 ms min.
See Section 4.2.2 Trigger Delay Settings p.54).
V mode
EX.FAST 6.8 ms FAST 22.8 ms MEDIUM 82.8 ms
68.8 ms
SLOW 257.8 ms
251.2 ms
0.3 ms
HOLD setting : Holds until the next trigger is detected PULSE setting : Remains only for the specified pulse width
See Section 5.2.5 Instrument Settings p.76).
When the internal trigger is selected HOLD setting : EX.FAST 1 ms, FAST 5 ms, MEDIUM 20 ms, SLOW 50 ms PULSE setting : Remains only for the specified pulse width
mode or V mode
EX.FAST 3.4 ms FAST 11.4 ms MEDIUM 41.4 ms (50 Hz line
frequency setting)
34.4 ms (60 Hz line frequency setting)
SLOW 156.4 ms (50 Hz line
frequency setting)
149.8 ms (60 Hz line frequency setting)
*1: For details, see “5.2.4 ERR Output p.75).”
: About t4 measurement time
*2
When averaging is enabled, the running average is obtained with internal triggering, so measurement time t4 does not change. The measurement time for external triggering is as follows: With SLOW sampling V (t4 - 57.8) X n + 57.8 ms (50 Hz)
(t4 - 51.2) X n + 51.2 ms (60 Hz)
or V (t4 - 56.4) X n + 56.4 ms (50 Hz)
(t4 - 49.8) X n + 49.8 ms (60 Hz)
With other than SLOW sampling
V (t4 - 2.8) X n + 2.8 ms or V (t4 - 1.4) X n + 1.4 ms
(n represents the number of values averaged)
: About t5 calculation time
*3
In the following cases, add the indicated times to calculation time t5:
When the Statistical Calculation function is enabled 0.3 ms When the reference value/tolerance method of
comparator decision is selected
0.15 ms
Page 85
79
HIOKI 3561A981-10

5.4 Internal Circuitry

5.4 Internal Circuitry
External Control and External Output Terminal Ratings
I/O type Logic Electrical specification
Output Open drain 30 V DC, 50 mA max.
Input C-MOS Inverse logic H: 3.8 to 5.0 V, L: 0 to 1.2 V
External Control Terminals_______________________________________
5 V
Circuit Diagram
5 k
H-CMOS
Input
1
2
3
Application Examples
5 k
Footswitch etc.
Input
GND
Switch Connection
0.1 F
100
GND
5 k
Input
GND
Photocoupler Connection
Photocoupler or equivalent within PLC
4
5
5
External Control
6
7
8
9
5 k
Input
GND
Relay Connection
Relay or equivalent within PLC
10
Page 86
80
HIOKI 3561A981-10
5.4 Internal Circuitry
External Output Terminals _______________________________________
Circuit Diagram
Output
GND
Open-Drain Output
Application Examples
3561
50 mA max
Output
ISO_COM
Relay Connection
3561
Output
ISO_COM
Active-Low Logic Output
3561
Active-Low Logic Output
PLC
3561
Output
ISO_COM
LED Connection
3561
Output
Output
ISO_COM
Wired-OR
Common
Output
Input
ISO_COM
PLC Input (Source Input) Connections
3561
Output
ISO_COM
PLC
Intput
Common
PLC Input (Sink Input) Connections
Page 87

6.1 Connecting the Printer

HIOKI 3561A981-10
Printing Chapter 6
6.1 Connecting the Printer
81
1
2
Before connecting the printer
Because electric shock and instrument damage hazards are present, always follow the steps below when connecting the printer.
• Always turn off the instrument and the printer before connecting.
• A serious hazard can occur if a wire becomes dislocated and
• As much as possible, avoid printing in hot and humid
• Use only compatible recording paper in the printer. Using non-
• If the recording paper is skewed on the roller, paper jams may
Recommended printer
The requirements for a printer to be connected to the instrument are
as follows.
Confirm compatibility and make the appropriate settings on the printer
before connecting it to the instrument.
contacts another conductor during operation. Make certain connections are secure.
environments. Otherwise, printer life may be severely shortened.
specified paper may not only result in faulty printing, but printing may become impossible.
result.
3
4
5
6

Printing

7
8
9
• Interface ............................ RS-232C
• Characters per line............ At least 45
• Communication speed ...... 9600 bps
• Data bits ............................ 8
• Parity ................................. none
• Stop bits ............................ 1
• Flow control....................... none
• Control codes ................... Capable of directly printing plain text
The optional printer model 9670 is no longer available. Their model
9670 printers can still use.
10
Page 88
82
HIOKI 3561A981-10
6.1 Connecting the Printer

6.1.1 Connecting the PRINTER to the Instrument

Connection Methods
Model 3561 (3561-01)
4
Printer
1. Confirm that the instrument and
printer are turned off.
2. Connect the AC Adapter to the
printer, and insert the power plug into an outlet.
3.
Connect the RS-232C cable RS-232C connectors on the instrument and printer.
3
2
to the
RS-232C cable
Connector Pinouts
1 2 3 4 5
6 7 8 9
Model 3561(-01) (9-pin) Connector
Function
Receive Data RxD 2
Transmit Data TxD 3
Signal or Common Ground GND 5
Signal Name
AC Adapter
Pin
4. Turn the instrument and printer
on.
13 ........................................ 1
25 ................................ 14
Printer (25-pin) Connector (Example)
Pin
Signal
Name
2 TxD Transmit Data
3 RxD Receive Data
7 GND Signal or Common Ground
4 RTS Request to Send
5 CTS Clear to Send
Function
Page 89
6.2 Selecting the Interface
HIOKI 3561A981-10
83

6.2 Selecting the Interface

1
2
3
Or
ten-keys
(SHIFT Lamp lit)
The Menu display appears.
Select the Interface Selection display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display) The current setting blinks.
Select Printer on the Sub Display.
rS ......... RS-232C
GP-Ib ... GP-IB
Prn....... Printer
Set the print interval time.
0000...................Interval printing is OFF. (Printing is carried out once
when PRINT key is pressed.)
0001 to 3600...... Sets the print interval time in seconds.
1
2
3
4
5
6
Printing
4
Applies setting and returns to the Measurement display.
7
8
9
10
Page 90
84
HIOKI 3561A981-10

6.3 Printing

6.3 Printing
Printing Measured Values and Decision Results ____________________
From the Measurement display, press the PRINT key or ground the
PRINT pin in the EXT I/O connector to print the measured value and
decision result.
• When using the external trigger, if you want to print after a triggered measurement finishes, connect the EOC signal of the EXT I/O to the
• To print all measurements continuously, connect the EOC signal to
PRINT signal and enable the internal trigger.
the
• When the statistical calculation function is on and the internal trigger is selected, the TRIG key or statistical calculation and printing of the current measurement value.
Valid counts are 1 to 30000. Above 30000, the count returns to 1.
PRINT signal.
TRIG signal will trigger
Interval Printing _______________________________________________
This function allows you to automatically print out measurement results at preset intervals. The print interval time must be set from the Interface Selection display. See Section 6.2 Selecting the Interface (p.83). The setting range is 1 to 3600 seconds. When the print interval time is set to "0", interval printing is disabled, and only normal printing is carried out.
Operation when interval printing is selected:
1. Start printing by pressing the PRINT key or sending the
signal via EXT I/O.
2. Elapsed time (hours/minutes/seconds) and measurement values
are printed automatically at intervals corresponding to the preset interval time.
3. Stop printing by pressing the PRINT key or sending the
signal via EXT I/O again.
• When the printed elapsed time reaches 100 hours, it resets to 00:00:00 and continues from zero.
(Example) After 99 hours, 59 minutes and 50 seconds: 99:59:50 After 100 hours, 2 minutes and 30 seconds: 00:02:30
• Selecting a display other than the measurement display causes interval printing to stop.
PRINT
PRINT
Printing Statistical Calculation Results ____________________________
From the Statistical Calculation display, press the PRINT key to print statistical calculation results. If no valid data exists, only the data count is printed. When only one valid data sample exists, standard deviation of sample and process capability indices cannot be printed.
Page 91
85
HIOKI 3561A981-10
6.3 Printing
Example Printouts______________________________________________
Measurement values (V mode)
1 298.60mOhm, 1.3924 V 2 0.2984 Ohm, 1.3924 V 3 - 3.35mOhm, 0.0000 V 4 - 0.0054 Ohm, 0.0000 V 5 299.10mOhm,- 1.3923 V 6 0.2984 Ohm,- 1.3923 V 7 3.57mOhm, 13.9071 V 8 - 16.89mOhm,-13.9088 V
With the Comparator ON
95 105.80mOhm Lo, 0.0000 V IN 96 213.15mOhm Hi 97 213.12mOhm IN 98 213.11mOhm Lo 99 10.0072 V Hi 100 10.0071 V IN 101 10.0070 V Lo 102 O.F. Hi, O.F. Hi 103 - 3.11mOhm Lo,- O.F. Lo 104 -------- --,-------- --
Statistical Calculations (Comparator ON)
*** RESISTANCE *** Number 85 Valid 85 Average 13.06mOhm Max 13.78mOhm( 74) Min 12.10mOhm( 3) Sn 0.38mOhm Sn-1 0.38mOhm Cp 1.32 CpK 0.09 Comp Hi 40 Comp IN 45 Comp Lo 0
Max/Min count
Measurement values ( mode)
15 209.98mOhm 16 0.2103 Ohm
Measurement values (V mode)
3132 4.2019 V 3133 15.2084 V
With the relative value comparison method (reference value and tolerance)
84 0.023 % Hi, 0.001 % IN 85 0.014 % IN, 0.000 % IN 86 - 0.019 % Lo, 0.002 % IN
With erroneous measurement values
10 O.F. , O.F. 11 - O.F. ,- O.F. 12 -------- ,-------­ 13 Invalid , Invalid 14 O.F. , 12.0097 V 15 - 19.82mOhm,- O.F.
Interval print
00:00:00 13.74mOhm, 10.0138 V 00:00:01 13.87mOhm, 10.0138 V 00:00:02 13.67mOhm, 10.0139 V 00:00:03 13.47mOhm, 10.0138 V 00:00:04 13.58mOhm, 10.0139 V 00:00:05 13.58mOhm, 10.0139 V 00:00:06 13.68mOhm, 10.0139 V
1
2
3
4
5
6
Printing
7
8
*** VOLTAGE *** Number 85 Valid 85 Average 10.0074 V Max 10.0197 V ( 57) Min 9.9938 V ( 31) Sn 0.0068 V Sn-1 0.0068 V Cp 0.35 CpK 0.32 Comp Hi 10 Comp IN 59 Comp Lo 16
Measurement values indicated as "Invalid" cannot be displayed by the instrument. The number of statistical calculation results indicated as “Valid” equals the count of valid data excluding measurement faults and overflows.
9
10
Page 92
86
HIOKI 3561A981-10
6.3 Printing
Page 93
RS-232C/GP-IB
HIOKI 3561A981-10
87

7.1 Overview and Features

Interfaces
This chapter describes the GP-IB and RS-232C interfaces, using the following symbols to indicate which information pertains to each interface. Sections with neither of these symbols pertain to both interfaces.
: GP-IB only
: RS-232C only
Before Use • GP-IB is available only on Model 3561-01.
• Always make use of the connector screws to affix the GP-IB or RS­232C connectors.
• When issuing commands that contain data, make certain that the data is provided in the specified format.
Chapter 7
7.1 Overview and Features
1
2
3
4
5
6
All instrument functions other than power on/off switching can be controlled via GP-IB/RS-232C interfaces.
• Resetting is supported.
• IEEE 488.2-1987 Common (essential) Commands are supported.
• Complies with the following standard: Applicable standard IEEE 488.1-1987
• This instrument is designed with reference to the following standard: Reference standard IEEE 488.2-1987
• If the output queue becomes full, a query error is generated and the output queue is cleared. Therefore, clearing the output queue and
query error output from the deadlocked condition IEEE 488.2 is not supported.
*1. ANSI/IEEE Standard 488.1-1987, IEEE Standard Digital Interface for
Programmable Instrumentation.
*2. ANSI/IEEE Standard 488.2-1987, IEEE Standard Codes, Formats,
Protocols, and Common Commands.
*1
*2
*3
as defined in
7

RS-232C/GP-IB Interfaces

8
9
10
*3. The situation in which the input buffer and the output queue become full,
so that processing cannot continue.
Page 94
88
HIOKI 3561A981-10

7.2 Specifications

7.2 Specifications

7.2.1 RS-232C Specifications

Transfer method Communications : Full duplex
Synchronization : Start-stop synchronization
Baud rate 9600 bps/ 19200 bps/ 38400 bps
Data length 8 bit
Parity none
Stop bit 1 bit
Message terminator (delimiter)
Receiving : CR+LF, CR Transmitting : CR+LF
Flow control none
Electrical specification Input voltage levels 5 to 15 V: ON,-15 to -5 V: OFF
Output voltage levels 5 to 9 V: ON, -9 to -5 V: OFF
Connector RS-232C Interface Connector Pinout
(Male 9-pin D-sub, with #4-40 attachment screws) The I/O connector is a DTE (Data Terminal Equipment) configuration Recommended cables:
• Model 9637 RS-232C CABLE (for PC/AT-compatibles)
• Model 9638 RS-232C CABLE (for PC98-series) See Section 7.3.1 Attaching the Connector (p.89).

7.2.2 GP-IB Specifications (Model 3561-01 only)

GP-IB is available only on Model 3561-01.
Interface Functions
SH1 All Source Handshake functions are supported.
AH1 All Acceptor Handshake functions are supported.
T6 Basic talker functions are supported. Serial poll function are
supported. No talk-only mode. The talker cancel function with MLA (My Listen Address) is supported.
L4 Basic listener functions are supported. No listen-only mode.
The listener cancel function with MTA (My Talk Address) is supported.
SR1 All Service Request functions are supported.
RL1 All Remote/Local functions are supported.
PP0 No Parallel Poll function.
DC1 All Device Clear functions are supported.
DT1 All Device Trigger functions are supported.
C0 No Controller functions are supported.
Operating Code: ASCII codes
Page 95
89
HIOKI 3561A981-10

7.3 Selecting the Connections and Protocol

7.3 Selecting the Connections and Protocol

7.3.1 Attaching the Connector

• Always turn both devices OFF when connecting and disconnecting an interface connector. Otherwise, an electric shock accident may occur.
• After connecting, always tighten the connector screws. The mounting screws must be firmly tightened or the RS-232C connector may not perform to specifications, or may even fail.
• To avoid damage to the instrument, do not short-circuit the connector and do not input voltage to the connector.
RS-232C Connector
1 2 3 4 5
6 7 8 9
Male 9-pin D-sub
#4-40 attaching screws
Connect the RS-232C cable.
To connect the instrument to a controller (DTE), use a crossover cable the connectors on both the instrument and the controller.
compatible with
1
2
3
4
5
6
The I/O connector is a DTE (Data Terminal Equipment) configuration. This instrument uses only pins 2, 3 and 5. The other pins are unconnected.
Pin
No.
Common EIA JIS
1 DCD CF CD Unused No connection
2 RxD BB RD Receive Data
3 TxD BA SD Transmit Data
4
5 GND AB SG Signal Ground
6 DSR CC DR Unused No connection
7
8 CTS CB CS Unused No connection
9 RI CE CI Unused No connection
Signal Name
DTR CD ER
RTS CA RS
Signal Notes
Data Terminal Ready
Request to Send Internally connected
Internally connected to +5 V
to +5 V
7
RS-232C/GP-IB Interfaces
8
9
10
Page 96
90
HIOKI 3561A981-10
7.3 Selecting the Connections and Protocol
Connecting to a PC/AT­Compatible (DOS/V) Machine
Connecting to an NEC PC9801 or PC9821 Series Desktop PC (excluding NX)
Use a crossover cable with female 9-pin D-sub connectors.
Crossover Wiring
Female 9-pin
D-sub
Model 3561-end
Pin No. Pin No.
DCD 1 1 DCD
RxD 2 2 RxD
TxD 3 3 TxD DTR 4 4 DTR GND 5 5 GND DSR 6 6 DSR
RTS 7 7 RTS CTS 8 8 CTS
9 9
Female 9-pin
D-sub
PC/AT-end
Recommended cable:
HIOKI
Model 9637 RS-232C CABLE (1.8 m)
Use a crossover cable with a female 9-pin D-sub and a male 25-pin D-sub connector. As the figure shows, RTS and CTS pins are shorted together and
crossed to DCD in the other connector.
Crossover Wiring
Female 9-pin
D-sub
Model 3561-end
Pin No. Pin No.
DCD 1
RxD 2 2 TxD
TxD 3 3 RxD DTR 4 4 RTS GND 5 5 CTS DSR 6 6 DSR
RTS 7 7 GND CTS 8 8 DCD
9 20 DTR
Male 25-pin
D-sub
PC-end
Recommended cable:
HIOKI
Model 9638 RS-232C CABLE (1.8 m)
Note that the combination of a dual male 25-pin D-sub cable and a 9­to 25-pin adapter cannot be used.
GP-IB Connector
Connecting a GP-IB cable.
Recommended cable: Model 9151-02 GP-IB CONNECTOR CABLE (2 m)
Page 97
7.3 Selecting the Connections and Protocol
HIOKI 3561A981-10

7.3.2 Selecting the Interface

91
1
2
3
(SHIFT Lamp lit)
The Menu display appears.
Select the Interface Selection display.
See Section 1.4 Menu Display Sequence (SHIFT ENTER) (p.15).
(Main Display)
(Sub Display) The current setting blinks.
Select RS-232C or GP-IB on the Sub Display.
......... RS-232C
rS
GP-Ib .... GP-IB (Model 3561-01 only)
Prn ....... Printer
When you select RS-232C, set the communications speed.
(Sub Display)
1
2
3
4
5
6
4
When selecting GP-IB, also set the
(Sub Display)
Message Terminator setting (LF/CRLF)
Address setting (0 to 30)
Selects the item to set
Applies settings and returns to the Measurement display.
Address and Message Terminator
Setting
7
.
RS-232C/GP-IB Interfaces
8
9
10
Page 98
92
HIOKI 3561A981-10

7.4 Communication Methods

7.4 Communication Methods
Various messages are supported for controlling the instrument through the interfaces. Messages can be either program messages, sent from the PC to the instrument, or response messages, sent from the instrument to the PC.
Program Messages
Response Messages
Model 3561
(Model 3561-01)
Message types are further categorized as follows:
Messages

7.4.1 Message Format

Program Messages
Program messages can be either Command Messages or Query Messages.
Command Messages Instructions to control the instrument, such as to change settings or reset
Example: (instruction to set the measurement range)
:RESISTANCE:RANGE 100E-3
Command Message
Program Messages
Query Message
Response Messages
Header portion
Query Messages Requests for responses relating to results of operation or measurement, or the state of instrument settings.
Example: (request for the current measurement range)
Space
Data portion
:RESISTANCE:RANGE?
Header portion
For details: See Section Headers (p.93), Separators (p.94) and Data
Formats (p.95).
Question Mark
Page 99
7.4 Communication Methods
HIOKI 3561A981-10
93
Response Messages
When a query message is received, its syntax is checked and a response message is generated. The
:SYSTem:HEADer command determines whether headers are
prefixed to response messages.
Header ON :RESISTANCE:RANGE 300.00E-3 Header OFF 300.00E-3
(the current resistance measurement range is 300 m)
At power-on, Header OFF is selected. If an error occurs when a query message is received, no response message is generated for that query. No header is applied to commands used only for queries, such as
:FETCH? and :CALCulate:LIMit:RESistance:RESult?.
Command Syntax Command names are chosen to mnemonically represent their
function, and can be abbreviated. The full command name is called the "long form", and the abbreviated name is called the "short form". The command references in this manual indicate the short form in upper-case letters, extended to the long form in lower case letters, although the commands are not case-sensitive in actual usage. Response messages generated by the instrument are in long form and
FUNCTION
FUNC
FUNCT
FUN
in upper case letters.
OK (long form)
OK (short form)
Error
Error
1
2
3
4
5
6
Headers Headers must always be prefixed to program messages.
(1) Command Program Headers
There are three types of commands: Simple, Compound and Standard.
Headers for Simple Commands This header type is a sequence of letters and digits
ESE 0
Headers for Compound Commands These headers consist of multiple simple command type headers separated by colons ":"
:SAMPle:RATE
Headers for Standard Commands This header type begins with an asterisk "*", indicating that it is a standard command defined by IEEE 488.2.
RST
(2) Query Program Header
These commands are used to interrogate the instrument about the results of operations, measured values and the current states of instrument settings. As shown by the following examples, a query is formed by appending a question mark "?" after a program header.
:FETCh? :MEASure:RESistance?
7
RS-232C/GP-IB Interfaces
8
9
10
Page 100
94
HIOKI 3561A981-10
7.4 Communication Methods
Message Terminators
Separators
This instrument recognizes the following message terminators:
• LF
• CR+LF
• EOI
• LF with EOI
From the instrument's interface settings, the following can be selected
as the terminator for response messages.
• LF with EOI (initial setting)
• LF with CR and EOI
See Section 7.3.2 Selecting the Interface (p.91).
• CR
• CR+LF
• CR + LF (initial setting)
(1) Message Unit Separator
Multiple message can be written in one line by separating them with semicolons ";".
:SYSTEM:LFREQUENCY 60;IDN?
• When messages are combined in this way and if one command
• A query error occurs if a query command is combined with an
(2) Header Separator
In a message consisting of both a header and data, the header is separated from the data by a space " ".
:SYSTEM:ELOCK ON
(3) Data Separator
In a message containing multiple data items, commas are required to separate the data items from one another.
contains an error, all subsequent messages up to the next terminator will be ignored.
immediately following semicolon and subsequent command.
Loading...