Hioki 3550 User Manual

INSTRUCTION MANUAL
For...は専用機種。複数の場合は「/」で区切る。
不要の場合はとる。
形名を入力。 複数の場合は「/」で区切る。
3550
品名を入力。
BATTERY HiTESTER
Contents
Safety ii
Precautions v
Organization of this Manual viii
Chapter 1 Overview 1
Chapter 2 Names and Functions of Parts 3
Chapter 3 Specifications 9
3.1 General Specification 9
3.2 Measurement Range
3.2.1 Maximum Input Voltage 14
3.2.2 Floating Voltage
3.2.3 Dielectric Strength
13
15 15
Chapter 4 Standard Measurement 17
4.1 Preparing for Measurement 17
4.2 Measurement
19
Chapter 5 Advanced Measurement Functions 23
5.1 Comparator Function 23
5.1.1 What is the Comparator Function? 23
5.1.2 Comparator Settings
5.1.3 Comparator Decision Result Table
5.1.4 Switching the Comparator On and Off
5.1.5 Changing the Comparator Number
24 29 30 31
5.2 Measurement Value Memory Function 32
5.2.1 Memory 32
5.2.2 Clear
5.2.3 Readout
5.2.4 Printing
34 35 36
5.3 Beeper On/Off Function 38
5.4 Hold Function
5.5 Moving Average Function
5.6 Zero Adjust Function
5.7 Battery Low Warning
5.8 Auto Power Off
5.9 AC Four-Terminal Method
5.10 Internal Resistance of Lead-Acid Batteries
38
39
41
42
43
45
47
Chapter 6 Maintenance 49
6.1 Troubleshooting 49
6.2 Message Reference
6.3 Cleaning
51
51
___________________________________________________________________
Introduction
Thank you for purchasing the HIOKI "3550 BATTERY HiTESTER." To obtain maximum performance from the product, please read this manual first, and keep it handy for future reference.
Inspection
When you receive the product, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Accessories
9460 CLIP TYPE LEAD WITH TEMPERATURE SENSOR 9382 CARRYING CASE Six LR6 alkaline batteries Instruction Manual Dust cover (for the printer interface)
i
Shipping
Use the original packing materials when reshipping the product, if possible.
___________________________________________________________________
Inspection
ii
y
___________________________________________________________________
Safet
DANGER
This product is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the product. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from product defects.
Safety symbols
This manual contains information and warnings essential for safe operation of the product and for maintaining it in safe operating condition. Before using the product, be sure to carefully read the following safety notes.
The symbol printed on the product indicates that the user should refer to a corresponding topic in the manual (marked with the before using the relevant function. In the manual, the symbol indicates particularly important information that the user should read before using the product.
symbol)
Indicates DC (Direct Current).
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
___________________________________________________________________
Safety
___________________________________________________________________
The following symbols in this manual indicate the relative importance of cautions and warnings.
iii
DANGER
WARNING
CAUTION
NOTE
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the product.
Advisory items related to performance or correct operation of the product.
___________________________________________________________________
Safety
iv
___________________________________________________________________
Measurement categories (Overvoltage categories)
This product conforms to the safety requirements for CAT I
measurement products.
To ensure safe operation of measurement products, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT I to CAT IV, and called measurement
categories. These are defined as follows.
CAT I : Secondary electrical circuits connected to an AC
electrical outlet through a transformer or similar device.
CAT II : Primary electrical circuits in equipment connected to an
AC electrical outlet by a power cord (portable tools,
household appliances, etc.)
CAT III : Primary electrical circuits of heavy equipment (fixed
installations) connected directly to the distribution panel,
and feeders from the distribution panel to outlets.
CAT IV : The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent
protection device (distribution panel).
Higher-numbered categories correspond to electrical environments
with greater momentary energy, so a measurement product
designed for CAT III environments can endure greater momentary
energy than one designed for CAT II. Using a measurement
product in an environment designated with a higher-numbered
category than that for which the product is rated could result in a
severe accident, and must be carefully avoided.
Never use a CAT I measuring product in CAT II, III, or IV
environments.
The measurement categories comply with the Overvoltage
Categories of the IEC60664 Standards.
___________________________________________________________________
Safety
___________________________________________________________________
Precautions
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
v
DANGER
WARNING
The floating voltage between input terminals and ground is 450V DC/AC. Attempting to measure voltages exceeding 450V with respect to ground could damage the product and result in personal injury. When voltages higher than 30 Vrms, 42.4 Vpeak or battery circuits with more than 60 VDC are to be measured, be sure to establish the floating state first. Carrying out such measurements in the grounded state involves the risk of electric shock. When measuring batteries, always ensure sufficient ventilation. Sometimes sparks may occur when the test leads are connected to batteries, which can ignite any accumulated inflammable gases such as hydrogen. To avoid electric shock accidents, when measuring batteries being charged wear proper protective gear such as rubber gloves.
Do not allow the product to get wet, and do not take measurements with wet hands. This may cause an electric shock. Do not use the product where it may be exposed to corrosive or combustible gases. The product may be damaged or cause an explosion.
___________________________________________________________________
Precautions
vi
___________________________________________________________________
WARNING
Before using the product, make sure that the insulation on the leads is undamaged and that no bare conductors are improperly exposed. Using the product in such conditions could cause an electric shock, so contact your dealer or Hioki representative for replacements. (Model 9460) Be sure to connect the SOURCE and SENSE terminals (banana plugs), and TEMP SENSOR terminal (mini-plug) correctly. See Section 4.2, "Measurement" for details of the connections. To avoid injury or damage to the product, do not attempt to measure AC voltage, or DC voltage exceeding 50 V.
50 VDC max.
Do not apply a voltage between the SOURCE(+) and SENSE(+) or SOURCE(-) and SENSE(-) terminals. This could result in damage to the unit.
CAUTION
This product is not designed to be entirely water- or dust-proof. To avoid damage, do not use it in a wet or dusty environment.
___________________________________________________________________
Precautions
___________________________________________________________________
vii
CAUTION
NOTE
Do not store or use the product where it could be exposed to direct sunlight, high temperature or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate so that it no longer meets specifications. Do not use the product near a device that generates a strong electromagnetic field or electrostatic charge, as these may cause erroneous measurements.
Use the specified 9460 CLIP TYPE LEAD WITH TEMPERATURESENSORwhenthistesterisused.
Correct measurement may be impossible in the presence of strong magnetic fields, such as near transformers and high­current conductors, or in the presence of strong electromagnetic fields such as near radio transmitters.
Service
Before using the product the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
When sending the product for repair, remove the batteries and pack carefully to prevent damage in transit. Include cushioning material so the product cannot move within the package. Be sure to include details of the problem. Hioki cannot be responsible for damage that occurs during shipment.
___________________________________________________________________
Precautions
viii
___________________________________________________________________
Organization of this Manual
This manual consists of the following chapters.
"Introduction", "Inspection", "Safety", "Precautions" include some important notes which you should read before using the unit.
Chapter 1 Overview
describes an outline of the unit, and lists its features.
Chapter 2 Names and Functions of Parts
lists the names of the parts of the unit, and the functions of all of the indications, terminals, and switches.
Chapter 3 Specifications
lists the specifications of the unit.
Chapter 4 Standard Measurement
describes the basic operation of the unit.
Chapter 5 Advanced Measurement Functions
describes miscellaneous functions.
Chapter 6 Maintenance
gives troubleshooting information.
___________________________________________________________________
Organization of this Manual
___________________________________________________________________
Chapter 1
Overview
The 3550 is designed for measuring the internal resistance, open-circuit voltage, and terminal temperature of secondary batteries, including lead-acid batteries, nickel-cadmium batteries, nickel-metal hydride batteries, and lithium-ion batteries.
Features
Since it uses the AC four-terminal method to measure the internal resistance, it provides accurate results with the lead resistances and contact resistances eliminated. It is possible to display the readings for the battery internal resistance, voltage, and terminal temperature, without changing functions. Moreover, a composite comparator function, which can be set on resistance and voltage values, enables reliable detection of battery deterioration. The unit's memory function also allows a number of sets of readings to be stored at a touch, and later output to a printer. These functions make this an ideal tool for checking batteries which are under constant trickle charging and cannot be disconnected.
1
NOTE
___________________________________________________________________
Measurements are taken by using noise reduction technology to
attenuate noise at frequencies other than the measurement
frequency (1 kHz). Incorrect measurements may result if a large
1-kHz noise component from the charger is present at the battery
terminals.
Chapter 1 Overview
2
___________________________________________________________________
___________________________________________________________________
Chapter 1 Overview
___________________________________________________________________
Chapter 2
Names and Functions of
Parts
This chapter explains the keys, input and output terminals, display, LED indicators, and leads.
3
___________________________________________________________________
Chapter 2 Names and Functions of Parts
4
___________________________________________________________________
Front Panel
16
15
17
12
13
14
1 2 4 5 3 6 7 10 11 9
8
Side Panel
18 19 20
___________________________________________________________________
Chapter 2 Names and Functions of Parts
___________________________________________________________________
Keys and input/output terminals
POWER
1.
2.
300m
3.
0ADJ
4.
5.
HOLD
6.
7.
COMP
8.
30m
30
key Turns the power on or off.
key
3
3
key
keykey
V key V key
key
key
key
key
Selects the resistance range.
Selects the voltage range.
Zero adjustment key.
Turns the beeper on and off.
Locks out changes to the display.
Left cursor (flashing) key.
Switches the comparator on and off, and changes display to the comparator setting screen.
9.
10.
11.
12.
13.
14.
15.
16.
key
key
UP
DOWN
MEMO
CLEAR
READ
PRINT
SOURCE
key
key
key
key
key
Right cursor (flashing) key.
Increases the value of a numeric setting.
Decreases the value of a numeric setting.
Stores the display data in memory.
Clears data captured with the
Reads data captured with the
Prints data captured with the
MEMO
MEMO
MEMO
key.
key.
key.
Connects to the 9460 banana plug on the SOURCE side.
17.
SENSE
Connects to the 9460 banana plug on the SENSE side.
TEMP SENSOR
18.
EXT.MEMO
19.
Connects to the 9460 mini-plug.
Connects to an external foot switch or similar, and stores the displayed data in memory.
20.
PRINT
Connects to the printer.
5
___________________________________________________________________
Chapter 2 Names and Functions of Parts
6
___________________________________________________________________
LCD (view with all elements displayed)
1
2 3
Display
1. Measured resistance
2. Temperature measurement value [comparator resistance
lower limit setting]
3. Temperature measurement value [comparator resistance lower limit setting]
mΩ
V
HOLD
COMP
Indicates the unit of resistance.
Indicates the unit of voltage.
Indicates the unit of temperature.
Appears when the display is locked.
Appears when the comparator function is on.
Appears when the beeper is turned on.
DATA
No.
Indicates that data captured data with the
MEMO
key is present.
Indicates the number of data points captured
with the
MEMO
key.
Usually, this is the number of the last data
point captured.
[The comparator table number.]
[]: Appears in on the comparator setting screen.
___________________________________________________________________
Chapter 2 Names and Functions of Parts
___________________________________________________________________
7
MEMO
Flashes once when the
MEMO
pressed.
Appears when the battery voltage of the
tester is low, to prompt the user to replace
the battery.
COMP.SET
Appears during display of the comparator
setting screen.
OHM
Appears during display of the screen for
setting the comparator resistance upper and
lower limit values.
VOLT
Appears during display of the screen for
setting the comparator voltage threshold
value.
LEDs
PASS
Indicates that the tested battery is
satisfactory for operation.
key is
WARNING
Indicates that the tested battery is
beginning to deteriorate.
FAIL
Indicates that the tested battery has
deteriorated.
These indications appear when the upper and lower comparator limits for internal resistance and the comparator threshold value for voltage are all set.
___________________________________________________________________
Chapter 2 Names and Functions of Parts
8
___________________________________________________________________
9460 CLIP TYPE LEAD WITH TEMPERATURE SENSOR
Clip (red)
Temperature sensor
Mini-plug (TEMP.SENSOR)
Clip (black)
Banana plug (black)
Banana plug (red)
SOURCE SENSE
___________________________________________________________________
Chapter 2 Names and Functions of Parts
___________________________________________________________________
Chapter 3
Specifications
3.1 General Specification
9
Measurement method
A/D conversion
Display
Panel abbreviations
and symbols
Sampling rate
Open-circuit terminal
voltage
Input overflow
Battery low detection
Constant current fault
detection
Moving average
function
Resistance: AC four-terminal method. Temperature: platinum temperature sensor.
Double inte
gration method.
LCD and LEDs (comparator output)
HOLD,COMP
MEMO
COMP.SET,OHM,VOLT,V
,
No.
, ,
,m,Ω, ,
DATA
,
o
C
,
0.83 sets (resistance, voltage, and temperature measurements)/second.
5Vmax.
"OF" indication
" " indication (indicates at 6.4 V with LR6 X 6)
"----" indication.
measurement (can be set to on or off)
Zero adjustment
function
Hold function
___________________________________________________________________
Circuit offset voltage is displayed as 0V.
Display is held.
Chapter 3 Specifications
10
___________________________________________________________________
Beeper function
Auto power off
Comparator settings
Number of
comparator settings
Comparator output
Resistance
Voltage
LO
HI
Audible output for warning and fail results. Canbeturnedonandoff.
Power off automatically after about 30 minutes.
Resistance upper and lower limits and voltage lower limit.
Ten sets.
LEDs for (amber), and Audible tone for
PASS
FAIL
(green),
(red) results.
WARNING
WARNING
results.
LO IN HI
WARNING
(Amber)*1
PASS
(Green)
WARNING
(Amber)
WARNING
(Amber)
and
FAIL
(Red)
FAIL
(Red)
FAIL
*1 Voltage low and internal resistance low: amber flashing
Data save
Save measurement value in memory when
EXT.MEMO
MEMO
key pressed or
terminals shorted. Memory holds 260 sets of resistance, voltage, and temperature values, and comparator result. The LED for warning flashing is stored in memory as a WARNING result.
Data readout
Data in memory is read out to display.
Data clear
___________________________________________________________________
Chapter 3 Specifications
Clear the data in memory.
___________________________________________________________________
11
Printer interface
Operating temperature and humidity range
Storage temperature and humidity range
Power source
Rated supply voltage
Location for use
Maximum power consumption
Continuous operating time
Dimensions
Centronics (can be connected to the 9203 DIGITAL PRINTER)
o
Cto40
0
o
C(32。F to 104。F)
80 %RH (no condensation)
-10
o
Cto50
o
C(14。F to 122。F)
80 %RH (no condensation)
Six LR6 alkaline batteries or six R6P manganese batteries
1.5 VDC X 6 (Voltage fluctuations of 10% from the rated supply voltage are taken into account.)
Altitude up to 2000 m (6562-ft.)
1.8 VA
Approx. 7 hours (at 30 mrange, comparator ON, beeper ON, with LR6 batteries)
Approx. 196W X 130H X 50D mm (7.72"W X 5.12"H X 1.97"D)
Mass
Approx. 710 g (25.0 oz.) (Including batteries)
Effect of radiated radio-frequency electromagnetic field
at 3V/m Resistance measurement
3.0% f.s.
Voltage measurement
3.0%f.s.
___________________________________________________________________
Chapter 3 Specifications
12
___________________________________________________________________
Standards applying
EMC
EN61326:1997+A1:1998+A2:2001 +A3:2003
Safety
EN61010-1:2001 Pollution Degree 1, Measurement Category I (anticipated transient overvoltage 2500 V)
Accessories
9460 CLIP TYPE LEAD WITH TEMPERATURE SENSOR 9382 CARRYING CASE Instruction Manual Dust cover (for the printer interface) Six LR6 alkaline batteries
Options
9203 DIGITAL PRINTER 9425 CONNECTION CABLE (2-meter long for connecting to the
9203) 9233 RECORDING PAPER (ten 10-meter rolls for the 9203)
Accuracy
We define measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values, with the following meanings:
f.s. (maximum display value or scale length)
The maximum displayable value or the full length of the scale. This is usually the maximum value of the currently selected range.
rdg. (reading or displayed value)
The value currently being measured and indicated on the measuring product.
dgt. (resolution)
The smallest displayable unit on a digital measuring product, i.e., the input value that causes the digital display to show a "1".
___________________________________________________________________
Chapter 3 Specifications
___________________________________________________________________
13
3.2 Measurement Range
Conditions to guarantee accuracy:
o
Temperature :23 Humidity :80 %RH or less (no condensation) Zero adjustment :After zero adjustment for each range Warminguptime :Atleast10minutes Period of guaranteed accuracy :1 year
(1) Resistance Measurement
Temperature coefficient: ( 0.01% rdg. 0.5 dgt.)/ Measurement current frequency: 1 kHz
Range
30 m 30.00 m 10 µΩ 50 mA
300 m 300.0 m 100 µΩ 5mA
3 3.000 1m 500 µA
Maximum indication
Resolution
Measurement
current
o
C 5
C
Six months 1 year
Six months 1 year
Six months 1 year
o
C
30 Hz
Accuracy
0.8% rdg. 6 dgt.
1.2% rdg. 6 dgt.
0.8% rdg. 6 dgt.
1.2% rdg. 6 dgt.
0.8% rdg. 6 dgt.
1.2% rdg. 6 dgt.
(2) Voltage Measurement
o
Temperature coefficient: ( 0.005% rdg. 0.5dgt.)/
Range
3V 3.000 V 1mV
30 V 30.00 V 10 mV
Maximum indication
Resolution Accuracy
Six months 1 year
Six months 1 year
C
0.1% rdg. 6 dgt.
0.15% rdg. 6 dgt.
0.1% rdg. 6 dgt.
0.15% rdg. 6 dgt.
(3) Temperature Measurement
Measurement range Resolution Accuracy
-10oCto60oC 0.1oC
___________________________________________________________________
Six months 1 year
0.5% rdg. 10 dgt.
0.75% rdg. 15 dgt.
Chapter 3 Specifications
14
___________________________________________________________________
3.2.1 Maximum Input Voltage
DANGER
The maximum input voltage is 50 VDC. Attempting to measure voltage in excess of the maximum input could destroy the product and result in personal injury or death. To avoid electrical hazards and damage to the product, do not apply voltage exceeding the maximum input to the measurement terminals.
50 VDC max.
50 VDC maximum No AC voltage input
___________________________________________________________________
Chapter 3 Specifications
___________________________________________________________________
15
3.2.2 FloatingVoltage
Between input terminals and ground, 450 VDC/AC maximum.
DANGER
Do not input voltage in excess of this floating voltage to the measurement terminal. This could result in injury or damage to the unit. When voltages higher than 30 Vrms, 42.4 Vpeak or battery circuits with more than 60 VDC are to be measured, be sure to establish the floating state first. Carrying out such measurements in the grounded state involves the risk of electric shock.
3.2.3 Dielectric Strength
Between input terminals and output terminals (including the
EXT.MEMO
DANGER
terminal): 1350 Vrms maximum for 1 minute.
1350 Vrms max. for 1 minute
PRINT
___________________________________________________________________
Chapter 3 Specifications
16
___________________________________________________________________
___________________________________________________________________
Chapter 3 Specifications
___________________________________________________________________
17
Chapter 4
Standard Measurement
4.1 Preparingfor Measurement
To avoid electric shock when replacing the batteries,
WARNING
first disconnect the leads from the object to be measured. Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result. After replacing the batteries, replace the cover before using the product.
1. Remove the battery cover.
2. Insert the batteries into the battery compartment as shown in the figure below.
___________________________________________________________________
Chapter 4 Standard Measurement
18
___________________________________________________________________
To avoid the possibility of explosion, do not short
WARNING
circuit, disassemble or incinerate batteries. Handle and dispose of batteries in accordance with local regulations.
NOTE
The " " indicator appears when battery voltage becomes low.
Replace the batteries as soon as possible.
Removing the batteries clears all measurement data from
memory. (See Section 5.2, "Measurement Value Memory
Function.")
After replacing the batteries, the auto power off function is set
to 30 minutes, with the beeper on.
Be sure to press the
POWER
key to power the unit off before
removing the batteries. If the batteries are removed while the
unit is powered on, the previous comparator number may not be
recovered.
To avoid problems with battery operation, remove the batteries
from the product if it is to be stored for a week or more.
Even when the power to the main unit is switched off, a very
small current (approximately 0.7 mA) is drawn from the
batteries to back up internal data. Therefore if the unit is left
switched off with new batteries installed, they will become fully
drained within 2 to 2.5 months.
___________________________________________________________________
Chapter 4 Standard Measurement
___________________________________________________________________
19
4.2 Measurement
WARNING
CAUTION
To prevent electric shock, before using this tester, fit the supplied dust cover for the printer interface over the PRINT terminal. To avoid injury or damage to the product, do not attempt to measure AC voltage, or DC voltage exceeding 50 V.
Do not attempt to measure the voltage of a generator. This would result in an AC voltage being applied to the voltage-generating output terminals, which is dangerous. After measuring a high-voltage battery, before continuing to measure a low-voltage battery first short the measurement leads together. This will discharge the DC-elimination capacitor which is connected across the leads. Otherwise an excess voltage may be applied to the low-voltage battery, which is dangerous.
___________________________________________________________________
Chapter 4 Standard Measurement
20
___________________________________________________________________
1. Connect the 9460 leads as shown in the figure below. Connect leads to all five terminals; SOURCE +/-, SENSE +/-, and TEMP SENSOR.
Mini-plug
TEMP SENSOR terminal
2. Press the
SOURCE and SENSE terminals
POWER
keytoturnonthepowerandstart
Banana plugs
Black Red
measurement. After pressing the
POWER
key to power on, it is necessary to wait for ten minutes of warming-up time, to allow the unit to stabilize.
3. Connect the red clip to the positive (+) side of the battery to be tested and connect the black clip to the negative (-) side.
Make the connections so that the temperature sensor is firmly in contact with the battery terminal.
Battery
Black
Red
4. Using the range keys, select the voltage and resistance
___________________________________________________________________
Chapter 4 Standard Measurement
___________________________________________________________________
21
measurement ranges.
5. When the measurement is completed, disconnect the leads from the tested battery and press the
POWER
key to turn off
the power.
NOTE
When measuring the contact resistance of a relay or connector,
be careful of the open-circuit voltage across the test leads. It
may not always be possible to destroy an oxide layer on the
terminals of the object being measured in order to obtain an
accurate reading. When the indication for a measurement value is "OF", this
indicates that the measured voltage, resistance, or temperature
value is outside the measurement range.
A resistance indication "----" means that the measurement could
not be made because there is a break in the test lead circuit.
The "----" indication may also appear if the leads are not
making good contact with the object to be measured, or if its
resistance is extremely large compared with the measurement
range.
The "----" indication may also appear immediately after
changing the resistance or voltage measurement range.
___________________________________________________________________
Chapter 4 Standard Measurement
22
___________________________________________________________________
A temperature indication "---" means that the temperature sensor
is not connected. The "---" indication may also appear if there
is a break in the test lead circuit.
Except for setting the auto power off mode and the moving
average function, do not press the
POWER
key in combination
with other keys. (For details of the moving average settings, see
Section 5.5, "Moving Average Function", and for details of the
auto power off settings, see Section 5.8, "Auto Power Off".)
If you do press a combination of the
keys, and an "
press the
POWER
"or"
INSP
key to power off and on again. Otherwise,
" indication appears, immediately
Adju
POWER
key and other
continuing with key operations may destroy the calibration data
for the unit, and correct measurement will no longer be possible.
If the leads are open-circuit, a spurious voltage indication may
sometimes be given. This is not a malfunction.
___________________________________________________________________
Chapter 4 Standard Measurement
___________________________________________________________________
23
Chapter 5
Advanced Measurement
Functions
5.1 Comparator Function
5.1.1 What is the Comparator Function?
The comparator function compares the measurement values with preset lower and upper limit values for internal resistance and voltage level, and determines which range the measurement falls into, based on the preset conditions. It then lights the corresponding LED, and sounds a beeper for the WARNING and FAIL cases. (See Section 5.3 Beeper On/Off Function)
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
24
___________________________________________________________________
5.1.2 Comparator Settings
(1) Before changing the settings
To set the comparator (upper and lower resistance limits and voltage comparison value), do the following:
1. Press and hold the upper and lower resistance limit settings appear.
COMP
key for at least 3 seconds. The
2. "
Resistance range
COMP.SET
","
" appears at the lower left of the screen,
OHM
showing that you are ready to change the comparator settings.
Comparator number
Lower resistance limit Upper resistance limit
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
(2) Setting the comparator number
Change the comparator number with the
UP
and
DOWN
25
keys. You can select any comparator number up to 10.
(3) Setting the resistance range
Press the resistance range key (
) corresponding to the resistance to be measured.
3
RANGE:30m,300m
,
The currently selected resistance range is displayed at the center on the screen.
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
26
___________________________________________________________________
(4) Setting the resistance limits
1. Using the key, move the flashing number to the most significant digit of the lower resistance limit setting at the lower left of the screen.
Lower resistance limit Upper resistance limit
2. Set the lower and upper resistance limits with the
DOWN
keys.
UP
Both the upper and lower resistance limits can be set to any value in the range 0 to 3000.
and
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
(5) Setting the voltage range
27
1. Using the key, move the flashing number to the least significant digit of the upper resistance limit setting at the lower right of the screen.
2. Press the
key again to display the voltage comparison value setting screen will appear. "
COMP.SET
"and"
" appear while setting the voltage
VOLT
comparison value.
Voltage range
3. Press the voltage range key (
VRANGE:3
corresponding to the voltage to be measured.
The position of the decimal point corresponds to the currently selected voltage range.
,
30
)
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
28
___________________________________________________________________
(6) Setting the voltage comparison value
1. Using the key and the
UP
and
DOWN
keys, set the voltage comparison value at the lower left of the screen. The allowable range of settings for the voltage comparison value is -3000 to 3000.
Voltage range
2. Move the flashing number to the least significant digit with the
key.
3. Press the
key again to display the upper and lower
resistance limit settings.
4. Go on to set the next comparator number. You can make voltage comparison values for up to 10 comparator numbers.
(7) Ending setup
When done with comparator selection, press the
COMP
key.
Display returns to the previous measurement screen.
NOTE
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
When the settings are made by setting a comparator number,
they are saved in memory at the point at which the comparator
setting screen is exited.
Comparator setting is not possible while the display is locked or
while reading out measurement values from memory.
___________________________________________________________________
29
5.1.3 Comparator Decision Result Table
The decision result is indicated by the LEDs and by the beeper, as shown in the following table.
Resistance
Voltage
Voltage comparison value
*1 Voltage low and resistance low: amber flashing
Beeper sounds when the comparator result is WARNING or FAIL. (Refer to Section 5.3, "Beeper On/Off Function")
Lower resistance limit Upper resistance limit
LO IN HI
LO
HI
WARNING
Amber *1
PASS
Green
WARNING
Amber
WARNING
Amber
A"PASS" result is shown by the green LED, a "WARNING" by the amber LED, and a "FAIL" by the red LED. The boundary conditions are as follows.
Resistance LO Lower resistance limit Resistance IN
Resistance HI
Voltage HI
Example 1
Resistance IN Upper resistance limit Voltage LO Voltage comparison value
Interpreting the comparator output table
When the measured resistance is at or below the lower
resistance limit, and the measured voltage is greater than the voltage comparison value (that is, resistance: LO and voltage: HI), the LED for
(green) lights. The beeper
PASS
does not sound.
Example 2
When the measured resistance is greater than the lower
resistance limit and lower than the upper resistance limit value, and the measured voltage is greater than the voltage comparison value (that is, resistance: IN, voltage: HI) the LED for
WARNING
(amber) lights and the beeper sounds.
FAIL
Red
FAIL
Red
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
30
___________________________________________________________________
5.1.4 Switchingthe Comparator On and Off
Pressing the
and off. When the comparator is on, the "
COMP
key toggles the comparator function on
COMP
" indication appears in the display, and the comparator operates as measurements are taken. When the comparator is off, the "
COMP
" indication disappears from the display,
and the comparator does not operate.
When one of the range keys (
,
3
3
V,
30
V) is pressed and the range is
30m
Ω,
300m
,
changed, comparator operation stops even if the comparator function is being used. To use comparator function again, press the
COMP
key. The range reverts to the setting extant
prior to the change.
When you turn on the power, the comparator is always set
to ON.
NOTE
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
If there is no measurement value, "----" is displayed and
comparator operation is not enabled.
___________________________________________________________________
31
5.1.5 Changingthe Comparator Number
To change the comparator number, do the following procedure.
1. Press and hold the
COMP
key for at least 3 seconds. The
resistance comparator setting screen appears.
"
COMP.SET
" appears at lower left of the screen.
2. Select the comparator number with the You can select any number up to 10.
UP
and
DOWN
keys.
3. When the setting is completed, press the
NOTE
___________________________________________________________________
The selected comparator number remains in memory even when
the power is turned off.
Chapter 5 Advanced Measurement Functions
COMP
key.
32
y
___________________________________________________________________
5.2 Measurement Value MemoryFunction
WARNING
To avoid damage to the product, do not apply voltage
to the EXT.MEMO terminals.
The currently measured values (resistance value, voltage value, temperature, and comparator result), taken as a set can be stored: up to 260 such sets of values can be held, and later displayed and printed as required.
5.2.1 Memor
Store the current settings as follows.
Press the
"
MEMO
measurement value and comparator result are stored.
MEMO
key.
" appears at the upper right of the screen and the
MEMO Screen
"
" is displayed at the upper right of the screen and the
DATA
number of data points stored in memory is displayed.
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
33
Using the
MEMO
EXT. MEMO
key can be obtained.
terminals, the same effect as the
1. Using a blade screwdriver or a similar tool, insert the individual leads into the terminal apertures. (Leads are not supplied.) When inserting or removing the leads, push in the terminal button with a screwdriver as shown in the figure below.
Recommended wire size
(AWG #18)
Usable limits
Single strand: 0.4 to 1.0 mm dia.
Single strand: 1.0 mm dia.
multi-strand: 0.75 mm
2
(AWG #26 to #18) multi-strand: 0.3 to 0.75 mm (AWG #22 to #20) minimum strand diameter: 0.18 mm
Standard insulation stripping length
Button pressing tool
10 mm
Blade screw driver
(tip width 2.6 mm)
2
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
34
___________________________________________________________________
2. Contact the leads together for at least 100 ms (use a switch) to obtain the effect of pressing the The "
MEMO
" indication appears in the display, and one set
MEMO
key.
of measurement values and comparator results are saved in memory.
Contact for at least 100 ms
At least 2 s
Wait for at least 2 seconds until the next storing.
"
" is displayed at the upper right of the screen along
DATA
with the number of data points in memory.
NOTE
If while the
key is pressed, then the values may be saved twice.
Even when the unit is powered off the values are preserved in
memory.
If the batteries are removed, the values in memory are cleared.
5.2.2 Clear
Clear the latest value in memory as follows.
1. Press the data point from memory.
2. To clear all data, press and hold the seconds.
NOTE
Removing the batteries clears all data.
EXT.MEMO
CLEAR
terminals are shorted together another
key. This clears the most recently stored
CLEAR
key for five
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
35
5.2.3 Readout
Redisplay a measurement value from memory as follows.
1. Press the
READ
key.
NOTE
The "
MEMO
number of "
MEMO
" indication appears in the upper right and the
No.   
" indication flashes.
means that the currently displayed readings are values recalled from memory. The flashing number is the number of the set of values currently displayed.
2. Pressing the
UP
key or
DOWN
key increments or decrements this number, to display whichever set of data values is required.
The comparator result for previous data values read out from memory is also shown, but a "WARNING" result does not flash.
In the case of a "WARNING" the display is permanently on.
Data cannot be redisplayed when "
upper right of the screen (when there is no data in memory).
" does not appear at the
DATA
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
36
g
___________________________________________________________________
5.2.4 Printin
WARNING
For safety, never connect the 3550 to the printer during measurement. To prevent electric shock, before using this tester, fit the supplied dust cover for the printer interface over the PRINT terminal. To avoid damage to the product, do not apply voltage to the PRINT terminal.
Print the measurement value stored in memory as follows.
1. Disconnect the lead from the tested battery.
2. Connect this tester to the 9203 DIGITAL PRINTER with the 9425 CONNECTION CABLE (2 m). (Refer to the printer instruction manual for how to make printer settings.)
3. Press the
PRINT
key. "Prnt" appears on the screen and the
data is printed.
4. To stop the printing, press the
PRINT
key.
NOTE
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
Use the HIOKI 9203 DIGITAL PRINTER.
It is also possible to connect a general-purpose printer with a
Centronics interface. The connector used for this unit is a 20-
pin half-pitch D-subminiature connector, DHA-RC20-R132N
(Daiichi Electronics) or equivalent.
It is not possible to print data while measurement is taking
place. When printing comparator results, both a flashing "WARNING" and a constant "WARNING"areprintedas"Warn."
When "
(there is no data in memory), printing is not possible.
" does not appear at the upper right of the screen
DATA
___________________________________________________________________
37
Output Example of the 9203
Output Example of the General-purpose Printer
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
38
___________________________________________________________________
5.3 Beeper On/Off Function
Pressing the key toggles the beeper on and off. When the beeper is on, the "
" indication appears in the display, and the beeper sounds when there is a "Warning" or "Fail" result. When the beeper is off, the "
" indication disappears from the display, and the beeper
does not operate.
5.4 Hold Function
This suspends measurement, with the display values held the same. Press the
HOLD
and the display is locked to prevent it from changing.
key. "
" is displayed on the screen
HOLD
NOTE
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
While the display is locked, the resistance and voltage range keys
(
0ADJ
effective.
30m
,
,
300m
COMP
,
,
, ,
,
3
UP
3
,and
V,
DOWN
30
V), and the
keys are not
___________________________________________________________________
39
5.5 MovingAverage Function
If the resistance measurement value is unstable, this can be corrected with the moving average function. This takes the average of four samples of the resistance value and displays the result. The moving average takes the sum of the current value and the previous three sampled values, and divides by four. To take the moving average, use the following procedure.
1. Power the unit off.
2. Hold down the key while pressing the
3. Hold down the
key until a screen for setting the moving
average appears.
POWER
key.
4. Press the
___________________________________________________________________
key once more.
Chapter 5 Advanced Measurement Functions
40
___________________________________________________________________
NOTE
5. When the above screen appears, press the
POWER
key to
power the unit off.
6. Press the
POWER
key again to power the unit on.
7. To cancel the moving average function, in the moving average setting screen, press the
key to cancel the moving
average function.
If you hold down a key other than the key while powering on, an "INSP"or"Adju" indication may appear. In this case,
press the
again. Continuing with other key operations will result in the
calibration data being lost, as a result of which correct
measurement will no longer be possible.
It is not possible to change the number of samples (i.e. four)
used for calculating the moving average.
When using the moving average function, the time taken for the
measurement value to stabilize is increased.
POWER
key once more, and to power on the unit
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
41
5.6 Zero Adjust Function
The zero adjustment function adjusts the zero position of the resistance and voltage ranges of this unit. The value read during zero adjustment is taken as zero, and used to calibrate subsequent measurements.
1. Short the SOURCE and SENSE of the 9460 CLIP TYPE LEAD WITH TEMPERATURE SENSOR together as shown in the figure below.
2. Press the
0ADJ
key. During zero adjustment, "0Adj"is
displayed in the resistance measurement display position.
3. When "0Adj" disappears and measurement starts, connect the leads to the battery to be tested.
NOTE
___________________________________________________________________
Keep the leads shorted together throughout the zero adjustment
process.
The zero adjustment is valid for the currently selected range
only, as long as the power remains on. Powering on the unit
resets all zero adjustment values.
When the resistance or voltage value is displayed as "----", or whenthereadingismorethan"200", "FAIL" is displayed. The
zero adjustment is not carried out.
Chapter 5 Advanced Measurement Functions
42
g
___________________________________________________________________
Shorting only the
Always make sure that both the
are shorted together.
If the leads are shorted but their ends are brought close to metal
parts, the measured value may fluctuate as a result of
electromagnetic induction. In this case, move the ends of the
leads away from the metal parts.
SENSE
terminals will not display 0 V.
5.7 BatteryLow Warnin
When the remaining battery capacity is low, the " " indicator appears at the right of the display. After printing out any held data values, replace the battery, referring to Section 4.1, "Preparing for Measurement."
SENSE
and
SOURCE
terminals
NOTE
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
If the batteries are exhausted, you may be able to turn on the
power, but soon the "
off. (Momentary operation is possible because the battery partially
regains its former voltage after resting, but soon declines to the
exhausted state.) When the batteries wear out, replace them in
accordance with the specified procedure.
" mark will appear and the power will go
___________________________________________________________________
43
5.8 Auto Power Off
In the following states, if there is no switch operation for 30 minutes the unit automatically powers off.
When the resistance value is "----" indication. During hold On the comparator setting screen During the printing output During reading the memory data
For continuous measurement, in some cases it may be necessary to disable the auto power off function. Do this as follows.
1. Turn the power off.
2. Press the
HOLD
POWER
key simultaneously. Press the
while. The auto-power off setting screen will appear.
key on the unit while holding down the
HOLD
key for a
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
44
___________________________________________________________________
NOTE
3. Press the
HOLD
4. When the above screen appears, press the
key again.
POWER
off the power.
5. Press the
POWER
6. To set to the auto-power off, press the
key again to turn the power on.
HOLD
key in the auto power off setting screen and set the auto power off to 30 minutes.
If the
EXT.MEMO
the auto power off function does not operate.
If the
POWER
HOLD
key, is held down, and an "INSP"or"Adju"indication
appears, immediately press the
again. Otherwise, continuing with key operations may destroy
the calibration data for the unit, and correct measurement will
no longer be possible.
After the batteries are replaced, the auto power off is set to 30
minutes.
The set time of auto power off cannot be changed.
terminals are continuously shorted together,
key is pressed while another key, other than the
POWER
key to power off and on
key to turn
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
45
5.9 AC Four-Terminal Method
The 3550 uses the AC four-terminal method, so that resistance measurement can be carried out with the resistance of the leads, and the contact resistance between the object to be measured and the leads canceled out. The following figure shows the principle of the AC four-terminal measurement method.
Constant current source
Is
Voltmeter
DC-elimination capacitor
R
1
R
2
Resistance R
Resistance measurement circuit
V
IS
R
3
Values R1 to R4 are the resistances of the test leads plus contact resistances.
An AC current (Is) is supplied from the
SOURCE
terminals of
the 3550 across the tested battery.
The voltage drop across the internal impedance of the battery
) is measured by the
(V
IS
since the
SENSE
SENSE
terminals are connected to an internal
terminals. At this point,
voltmeter with a high impedance, almost no current flows through the resistances R2 and R3 which represent the lead resistances and contact resistances.
As a result, there is almost no voltage drop across the resistances R2 and R3. Thus the voltage drop due to the lead resistances and contact resistances is very small, and these can be canceled out.
R
4
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
46
___________________________________________________________________
NOTE
In the 3550, a synchronized wave detection system is used,
whereby the internal impedance is separated into effective
resistance and reactance, and the resistive component only
displayed.
Reactance
Impedance
Effective resistance
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
47
5.10 Internal Resistance of Lead-Acid Batteries
This shows the relationship between the capacity for lead­acid batteries and the initial value for internal resistance. (Quote: Text from battery fitters qualification lecture notes, Battery Industry Association) The terms CS, HS, and MSE appearing in the figure are battery types defined by JIS (Japanese Industrial Standards).
Temperature of battery25
)
Ω
m (
e c n a
t s
i s e
r l a n
r e
t n
I
CS
HS
MSE
Capacityof lead-acid batteryAh
CS : Clad type stationary lead-acid battery HS : High rate discharge paste type stationary lead-acid
battery
MSE : Seal type stationary lead-acid battery
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
48
___________________________________________________________________
From the figure, MSE internal resistance can be read as approximately 1 m(at 100Ah), or approximately 0.13 m (at 1000 Ah). When the battery deteriorates, internal resistance rises to 1.5 to 2 times the initial value (reference value). With an MSE (seal type stationary lead-acid battery), battery condition is regarded as marginal when internal resistance reaches 1.5 times the initial value, and complete deterioration is assumed when internal resistance reaches 2 times the initial value.
NOTE
Even if different batteries have the same capacity, initial internal
resistance may vary according to battery type and maker. The
figure is only for reference.The internal resistance warning
(WARN) and fail (FAIL) values vary according to maker.
___________________________________________________________________
Chapter 5 Advanced Measurement Functions
___________________________________________________________________
g
49
Chapter 6
Maintenance
If damage is suspected, check the "Troubleshooting" section before contacting your dealer or Hioki representative.
6.1 Troubleshootin
Symptom Cause Follow-up
Nothing appears on the screen when the
POWER
pressed.
Measurement values are wrong.
"---" is displayed. "OF" is displayed.
Comparator result is not correct.
key is
Batteries are exhausted, or not inserted.
Leads are not correctly connected.
There is an electrical discontinuity in the leads.
Zero adjustment is not correct.
The measurement range is not appropriate.
The comparator setting is not correct.
Replace the batteries or reinsert them correctly. See Section 4.1, "Preparing for Measurement."
Connect leads correctly. See Section 4.1, "Preparing for Measurement."
Replace with a new lead.
Perform correctly zero adjustment. See Section 5.6, "Zero Adjust Function."
Select the appropriate range with the range key. See Section 4.2, "Measurement."
Set the comparator correctly. See Section 5.1, "Comparator Function."
___________________________________________________________________
Chapter 6 Maintenance
50
___________________________________________________________________
Symptom Cause Follow-up
Data in memory is not displayed when the
READ
pressed.
Printing is not possible.
key is
Thereisnodatain memory.
Thereisnodatain memory.
Storedatainmemory. See Section 5.2, "Measurement Value Memory Function."
Storedatainmemory. See Section 5.2, "Measurement Value Memory Function."
NOTE
The comparator setting screen appears.
The connecting cable is not correctly connected.
The printer is not ready. Turn the printer power on and
Exit the comparator setting screen, to return to the measurement screen. See Section 5.1, "Comparator Function."
Correctly connect the connecting cable.
set correctly.
To avoid problems with battery operation, remove the batteries from the product if it is to be stored for a week or more.
When sending the product for repair, remove the batteries and pack carefully to prevent damage in transit. Include cushioning material so the instrument cannot move within the package. Be sure to include details of the problem. Hioki cannot be responsible for damage that occurs during shipment.
Never modify the product. Only a Hioki service engineer can disassemble or repair the product. Failure to observe these precautions may result in fire, electric shock, or injury.
___________________________________________________________________
Chapter 6 Maintenance
___________________________________________________________________
g
51
6.2 Message Reference
Message Meaning Follow-up
FAIL
Prnt
0Adj
INI
INSP
Adju
Er10 Er11 Er12 Er20 Er21 Er22 Er23 Er24
Zero adjustment cannot be carried out.
Data is being sent to the printer.
Zero adjustment is being carried out.
Initialization after battery replacement.
Inspection and adjustment modes for factory use.
Internal variable error. Servicing is required.
Connect correctly the lead and execute the zero adjustment. See Section 5.6, "Zero Adjust Function."
Disappears when data output to the printer is completed.
Disappears when zero adjustment is completed.
This is not a malfunction.
Press the power on.
POWER
key to turn the
6.3 Cleanin
To clean the product, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
When cleaning the LCD, gently wipe with a soft dry cloth.
___________________________________________________________________
Chapter 6 Maintenance
52
___________________________________________________________________
___________________________________________________________________
Chapter 6 Maintenance
HIOKI 3550 BATTERY HiTESTER
Instruction Manual
Publication date: September 2006 Revised edition 10
Edited and published by HIOKI E.E. CORPORATION
Technical Support Section
All inquiries to International Sales and Marketing Department
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL: +81-268-28-0562 / FAX: +81-268-28-0568
E-mail: os-com@hioki.co.jp
URL http://www.hioki.co.jp/
Printed in Japan 3550A981-10
All reasonable care has been taken in the production of this manual, but if you find any points which are unclear or in error, please contact your supplier or the International Sales and Marketing Department at HIOKI headquarters. In the interests of product development, the contents of this manual are subject to revision without prior notice. Unauthorized reproduction or copying of this manual is prohibited.
2003-01 改訂  枠消す
HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp / URL http://www.hioki.co.jp/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA TEL +1-609-409-9109 / FAX +1-609-409-9108
3550A981-10 06 -09H
Printed on recycled paper
Loading...