Thank you for purchasing the HIOKI "3540 mΩ HiTESTER." To
obtain maximum performance from the product, please read this
manual first, and keep it handy for future reference.
Inspection
When you receive the product, inspect it carefully to ensure that
no damage occurred during shipp ing. In particular, check the
accessories, panel switches, and connectors. If damage is evident,
or if it fails to operate according to the specifications, contact your
dealer or Hioki representative.
Accessories
9287-10 CLIP TYPE LEAD 1
9451 TEMPERATURE PROBE
Instruction Manual
R6P manganese batteries (monitor batteries)
Spare fuse to protect the circuit
This product is designed to conform to IEC 61010
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling
during use could result in injury or death, as well as
damage to the product. Be certain that you
understand the instructions and precautions in the
manual before use. We disclaim any responsibility for
accidents or injuries not resulting directly from
product defects.
Safety symbols
This manual contains information and warnings essential for safe
operation of the product and for maintaining it in safe operating
condition. Before using the product, be sure to carefully read the
following safety notes.
・ The
user should refer to a corresponding topic in the
manual (marked with the
relevant function.
・ In the manual, the
important information that the user should read before
The following symbols in this manual indicate the relative
importance of cautions and warnings.
iii
WARNING
CAUTION
NOTE
Indicates that incorrect operation presents a
significant hazard that could result in serious injury or
death to the user.
Indicates that incorrect operation presents a
possibility of injury to the user or damage to the
product.
Advisory items related to performance or correct
operation of the product.
Measurement categories
To ensure safe operation of measurement instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement
categories. These are defined as follows.
CAT II: Primary electrical circuits in equipment connected to an
AC electrical outlet by a power cord (portable tools,
household appliances, etc.)
CAT II covers directly measuring electrical outlet
receptacles.
CAT III: Primary electrical circuits of heavy equipment (fixed
installations) connected directly to the distribution panel,
and feeders from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent
protection device (distribution panel).
Using a measurement instrument in an environment designated
with a higher-numbered category than tha t for which the
instrument is rated could result in a severe accident, and must be
carefully avoided.
Use of a measurement instrument that is not CAT-rated in CAT II
to CAT IV measurement applications could result in a severe
accident, and must be carefully avoided.
Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions.
v
WARNING
CAUTION
Use either the specified Hioki model 9445-02/03 AC
ADAPTER.
・ Never apply an external voltage to the SENSE and
SOURCE terminals of the instrument.
・ To properly suppress noise, this product must be set to
match the power supply frequency. Before using the
product, make sure the power supply frequency selector
is set correctly, to avoid erroneous readings.
(Refer to 4.2.1 Setting the Power Supply Frequency.)
・ Do not store or use the product where it could be
exposed to direct sunlight, high temperature or humidity,
or condensation. Under such conditions, the product
may be damaged and insulation may deteriorate so that
it no longer meets specifications.
・ To avoid damage to the product, protect it from
vibration or shock during transport and handling, and be
especially careful to avoid dropping.
・ The sensor used in the temperature probe is a thin,
precision platinum film.
Please note that excessive voltage pulses or static
discharges can destroy the film.
To avoid damage or malfunction, avoid hitting the tip of
the temperature probe and overly bending the leads.
When measuring high temperatures, do not let the
handle of the temperature probe or the compensation
lead wire exceed the temperature range.
The battery indicator lights up when the remaining battery
capacity is low. In this case, the instrument's reliability is not
guaranteed. Replace the battery immediately.
(Refer to "8.1 Battery Replacement Procedure.")
Be sure to turn the power switch OFF () when not using the
instrument.
Warm up the instrument for at least 30 minutes prior to use, to
attain proper measurement accuracy.
This product should be installed and operated indoors only,
between 0 and 40
and 80 %RH or less.
A fuse is provided in the current source (SOURCE) to protect the
circuit. If the fuse burns out, measurements cannot be made.
Refer to "8.2 Fuse Replacement procedure" on how to check for a
burned-out fuse.
This instrument should not be used with relays or other devices
that handle small signals, since it may damage their contact
coating.
Do not measure points which have a voltage across them. The
main unit of the 3540 will be damaged by an induced voltage if a
measurement is made immediately after a temperature rise test or
dielectric test of a motor or transformer.
Accurate measurement may be difficult to obtain if the instrument
is used near equipment that generates noise.
Also, the indicator may sometimes fluctuate if the device-undertest picks u p noise. Therefore, do not use the instrument in
environments with excessive electrical noise.
The measurement indication may sometimes fluctuate due to
noise pick-up if the temperature probe is touched or held with
bare fingers.
Temperature correction is not possible when the temperature
probe is in contact with the surface of the device to be measured.
Note that the temperature probe is only designed to measure
ambient air temperature.
Significant measurement error will result if the device to be
temperature corrected and the temperature probe are not at the
same ambient air temperature.
Large measurement error will result if the temperature probe is
not inserted fully into the tc sensor jack.
Make sure the power is turned off before connecting or
disconnecting the AC adapter.
The AC adapter may pick up noise which will affect the
measurement. In such a case, operate the instrument from battery
power.
Measurement range , comparator settings and all settings of the
3540 (except for the measured value) are backed up internally,
but this backup occurs only after a certain amount of time has
elapsed without any operation. Therefore, after changing the
settings, wait a few moments (about 5 seconds) before turning off
the power.
Because the 3540 uses direct current to make measurements,
thermoelectromotive effects can result in measurement errors.
Refer to "1.3 Effects of T hermoelectromotive Force" for details.
When the one that includes the L component such as the
transformer for the power supply a lot is measured a measured
value may not stabilize.
The 3540 mΩ HiTESTER is an four-terminal method tester
designed to accurately measure the coil resistance in motors and
transformers, the contact resistance of relays, switches a nd
connectors, and the trace resistance on printed circuit boards.
In addition, a temperature correction function, comparator function
and data output function are provided, making the 3540 mΩ
HiTESTER ideal for use in production and inspection lines and
systems.
1.1Four-terminal Method
1
To obtain accurate results when measuring resistance values that
are very small, the Four-terminal method must be used. As shown
in figure 1, in the Two-terminal method, the resistance of the test
leads adds to the resistance of the device being measured, resulting
in an erroneous measurement. However, in the Four-terminal
method shown in the Figure 2, the input consists of two current
terminals to which a constant current is supplied, and two voltage
terminals measured the voltage drop. The voltmeter has a high
input impedance so that essentially no current flows through the
leads connected between the device-under-test and the voltage
terminals. As a result, there is almost no voltage drop across the
resistances r
resistances and contact resistances is very small, and these can be
canceled out.
The temperature sensor used in the 3540's temperature probe is a
thin platinum film whose resistance changes according to
temperature. The resistance of the film is detected and converted to
a temperature value by the CPU.
This section explains use of the 3540's temperature coefficient
correction function.
Since the resistance of copper wire is relatively susceptible to
changes in temperature, that fact must be kept in mind when
measuring its resistance. Using the temperature probe, the
resistance value of copper wire can easily be converted to its 20
equivalent for display.
In general, the relationship between the resistance of copper wire
and temperature is as indicated by the following expression.
=Rt0x {1 + α
R
t
Here, α
is referred to as the temperature coefficient, which is
t0
expressed as follows.
= 1 / [{1 / (0.00393 x α }+(t0- 20)] (2)
α
t0
Here, σ is the conductivity of copper wire. From expressions (1)
and (2), the temperature coefficients of various types of wire with
various conductivities can be calculated, and their resistance values
at a particular temperature obtained. Table 1 shows the
conductivity of various types of copper wire.
For 3540 temperature correction, conductivity σ is calculated as
1. Accordingly, when the resistance value R
the time of temperature correction is taken as R
that is displayed at
20
, the measured
t
resistance at the current ambient temperature is expressed by the
following expression. (Temperature coefficient: α
-6
ppm = x 10
R
)
=Rt/{1+α
20
x (t - 20)} (3)
20
=3930ppm,
20
Error occurs during temperature correction because the temperature
coefficient calculated according to expression (2) differs when the
conductivity σ is other than 1.
For example, when the tin-plated soft copper wire (diameter 0.10
to 0.26) of Table 1 is measured, σ = 0.93 gives α
of 3650 ppm,
t0
so that the 3540 temperature value contains error.
Thermoelectromotive force is the potential difference at the
junction of two dissimilar metals. If this emf is large, measurement
errors can result. As the 3540 uses constant direct current flowing
through the object being measured, readings can be affected by
even slight thermoelectromotive force.
Furthermore, the quantity of thermoelectromotive force is
dependent upon the temperature of the measurement environment,
with the force generally being greater at higher temperature.
Thermoelectromotive force occurs at the junction of dissimilar
metals and between the probes of the 3540 and the contacts on the
object being measured. The following figure illustrates
thermoelectromotive force. The battery symbols represent a
junction of dissimilar metals, and the probe symbols represent the
thermoelectromotive force.
Metal AThermoelectromotive force
Metal:B
5
3540
As an example of the error effects of thermoelectromotive force, if
the force is 10 μV and the measured resistance is 3 Ω, the current
is 1 mA in the 3 Ω range, so the measured value displayed on the
3540 is actually
(3 Ω x 1mA+10μV) / 1 mA = 3.010 Ω. In t hi s s ituation,
changing the probe direction to HI-LO leaves the polarity of the
thermoelectromotive force unaffected, so the measured value is
now (3 Ω
On power-up, the LCD and LED's light and the instrument
performs an internal check and init ialization of internal analog
circuits. An error code is displayed if an internal error is detected
during the check. (Refer to "8.4 Error Code Table".)
Upon completion of the internal check, the LCD displays the
instruments power supply frequency setting and the 3540 version
number.
2. Measurement display area
Displays the measurement count, decimal point, unit of
measurement, and the measurement range.
3.
The
keys Page 26
(up) and(down) keys are used to select the
resistance measurement range.
4. AUTO
Pressing the AUTO
key, AUTO mark Page 26
key turns on automatic range selection,
lighting the AUTO mark on the LCD. The measurement range for
resistance is then automatically selected according to the resistance
of the resistor being measured.
5. HOLD
Pressing the HOLD
key, HOLD mark Page 28
key turns on the hold mode, lighting the
HOLD mark on the LCD and fixing the currently displayed
measurement value.
6. 0ADJ
key Page 27
To zero-adjust the instrument, short the test leads and press
SHIFT
+ 0ADJ . (Zero adjustment is only possible with a
reading of 100 counts or less.)
7. TC
key, TC mark Page 35
When the temperature probe is connected, pressing the TC
key
converts the measured resistance value of copper wire to its 20
equivalent resistance value and displays it.
Pressing SHIFT
selects the temperature measurement mode. Pressing SHIFT
TEMP
a second time returns measurement to the resistance
+ TEMP with the temperature probe connected
+
mode.
9. LOCK
Pressing the LOCK
/ UNLOCK key, LOCK mark
key lights the LOCK mark on the LCD and
locks out key input. With key input locked, all of the keys on the
instrument are disabled except for the POWER
SHIFT
10. SAMPL
key. To unlock the keys, press SHIFT + UNLOCK .
key, FAST mark Page 28
Sampling speed is switched using the SAMPL
switch and the
key. With fast
sampling, FAST is displayed on the LCD and sampling takes place
at the rate 16 samples per second. When slow sampling is selected,
sampling takes place at the rate of 4 samples per second.
11. 50/60Hz key Page 25
Power supply frequency is selected by pressing SHIFT
50/60Hz
. Select the frequency that matches that of your power
supply.
12. SHIFT
Pressing the SHIFT
key, SHIFT mark
key lights the SHIFT mark on the LCD and
puts the keys in the shift mode. In the shift mode, pressing any
key activates the function whose name is printed in blue below
that key. To cancel the shift mode, press the SHIFT
time.
13. COMP
Pressing the COMP
key, comparator display Page 31
key turns on the comparator, lighting the
comparator display on the LCD and enabling comparison
measurement. To end comparison measurement, press the COMP
key a second time.
14. TABLE
Pressing the TABLE
key, Table No. display Page 31
key switches the internal comparator table
number. With the 3540, up to seven different comparator
configurations can be saved.
Upper (HIGH) and lower (LOW) comparator limits and the
comparator reference value (REF) and range (%) can be set by
pressing keys corresponding to the various digits of the comparator
configuration display on the LCD
17. Buzzer key and buzzer mode display
Page 32
Pressing the buzzer key switches the buzzer mode that is set
dependent upon comparator results. When the
buzzer mark is
lit on the LCD, the buzzer is in the HL mode, and sounds when
the comparator result is HIGH or LOW. When the
buzzer
mark is lit on the LCD, the buzzer is in the IN mode, and sounds
when the comparator result is IN. When no buzzer mark is lit on
the LCD, the buzzer is in the OFF mode, and does not sound.
18. Battery mark
Page 77
The battery indicator appears when battery voltage becomes low.
Replace the batteries as soon as possible.
(Refer to "8.1 Battery Replacement Procedure".)
Selects the AUTO mode, for continuous comparator output
operation("normal" mode), or the MANU mode for comparator
output only when the MANU and GND terminals of the external
connector on the rear panel are shorted.
2. AC adapter jack
The 3540 can be operated from an AC power source by connecting
an AC adapter.
When using an AC adapter, use only the specified HIOKI model
9445-02, 9445-03 AC ADAPTER.
An AC adapter rated at 9 VDC and 1.4 A to the AC adapter
socket. The polarity of socket is center-plus.
Resistance measurement 3500 counts
Temperature measurement 999 counts
Provided (disabled when comparator is on)
"OF" display
"----" displayed (CCERR: external output*2)
Resistance measurement
"SLOW" 4 samples/sec
"FAST" 16 samples/sec
Temperature sampling 1 sample/sec
Resistance measurement
"SLOW" 300 msec
"FAST" 80 msec
Table storage of up to 7 configurations (table
externally selectable
Comparator results displayed by LED and
externally output (open collector*2)
Comparator modes (Hi-Lo/REF-%), Buzzer mode
(HL/IN/OFF)
Interface
Printer
Power supply frequency
Overvoltage protection
Operating
temperature/humidity
Storage temperature/humidity
Operating Environment
Power source
AC adapter (option)
Maximum rated power
Continuous operating time
Dimensions
Mass
RS-232C interface
Centronics interface
*5
*4
50/60 Hz, switchable
30 VDC or ACpeak (circuit protection by fuse)
0to40oC (32 to 104), 80 %RH or less
(No condensing)
-10 to 50oC (14 to 122), 80 %RH or less
(No condensing)
Indoor < 2000 m ASL (6566 feet)
Six LR6 alkaline batteries or six R6P manganese
Batteries
Rated supply voltage 1.5 VDC x 6
9445-02 AC ADAPTER
9445-03 AC ADAPTER
Rated supply voltage of the AC adapter is 100 to
240 VAC. (Voltage fluctuations of 10% from the
rated supply voltage are taken into account.)
Rated supply frequency is 50/60 Hz.
5VA
LR6 Alkaline Batteries
Approx.7 hours
(30 m, 300 mΩ range, LED and buzzer: ON)
Approx.18 hours
(other ranges, LED and buzzer: ON)
R6P manganese Batteries
Approx.1.5 hours
(30 m, 300 mΩ range, LED and buzzer: ON)
Approx.6 hours
(other ranges, LED and buzzer: ON)
Approx. 215W x 61H x 213D mm
(8.46"W x 2.40"H x 8.39"D)
Temperature measurement and temperature correction
6month
Temperature correction accuracy
Temperature range
-10.0 to 39.9oC
(14.0 to 103.9
40.0 to 99.9oC
(104.0 to 211.9
Temperature
measurement accuracy
0.3 %rdg.0.5
)
0.3 %rdg.1.0
)
(Add the following values to the
accuracy specifications of the
resistance measurement)
o
C
o
C
0.3 %
0.6 %
1year
Temperature correction accuracy
Temperature range
-10.0 to 39.9oC
(14.0 to 103.9
40.0 to 99.9oC
(104.0 to 211.9
Temperature
measurement accuracy
0.45 %rdg.0.8
)
0.45 %rdg.1.5
)
(Add the following values to the
accuracy specifications of the
resistance measurement)
o
C
o
C
0.4 %
0.8 %
* For 3540 only, accuracy is0.2oC for 6months ( 0.3oC for 1year) when connected
to a manufacturer-recommended temperature sensor (Pt). The temperaturemeasurement accuracy and temperature-correction accuracy above are those of 3540
combined with the results of the 9541 TEMPERATURE PROBE.
We define measurement tolerances in terms of rdg. (reading) and
dgt. (digit) values, with the following meanings:
rdg. (reading or displayed value)
The value currently being measured and indicated on the
measuring product/ instrument.
dgt. (resolution)
The smallest displayable unit on a digital measuring product/
instrument, i.e., the input value that causes the digital display to
show a "1" as the least-significant digit.
The 3540 works on battery power.
Refer to the section on battery replacement when installing
batteries into the battery compartment while.
(Refer to "8.1 Battery Replacement Procedure".)
When connecting the AC adapter, first make sure the POWER
switch is OFF, then insert the jack of AC adapter socket, connect
the AC adapter body to the AC power source, and finally turn ON
the
Since power consumption is high in the case of 3540-01, 3540-02
and 3540-03, batteries will be exhausted quickly.
For exhausted manganese batteries, the battery voltage fluctuates
greatly depending on the current retrieved.
Therefore, when measuring using manganese batteries, even when
the battery voltage is adequately secured at a measurement current
of 1 mA, at 100 mA the battery voltage will not be sufficient,
causing the power supply to cut off or the LCD to keep repeating
a cycle of blinking and disappearing.
Connect the leads as shown in the following figure :
The side with "v" mark is SENSE.
ENSE
Red
OURCE
SENSE
SOURCE
Black
When clipping a thin line
(Clip the line at the tip,
serrated part of the jaws.)
The cable part of the lead is shielded.
For user-made leads, take the following precautions when the leads :
(1) A shield must be applied. (Refer to the follow figure.)
(2) The cable length must be 5 m or less (The resistance of the
When using the 9451
TEMPERATURE PROBE, loop
the probe cable once around the
provided ferrite clamp and fasten it
as shown in the figure.
4.1.3 Instrument Handle
CAUTION
When using the handle as a stand for the device, do not
press down too hard on the device as this can damage
the handle.
The handle can be used as a stand. Pull both ends of the handle
outward to release it and rotate it to the desired position. Then,
push the ha ndle inward to lock it in place. The handle can be
locked at interval of 22.5 degrees.
Make the connection by mating the red
Make the connection by mating the black
and the leads. (See the Figure below.)
(2) Select the range.
(3) Zero adjust (0 ADJ) the instrument referring to subsection 4.2.3.
(4) Connect the lead clips to the device to be measured, and read the
To properly suppress noise, this product must be set to
match the power supply frequency. Before using the
product, make sure the power supply frequency selector
is set correctly, to avoid erroneous readings.
First press the SHIFT key, then press 50/60Hz .Aftera
moment, the selected power supply frequency appears on the LCD
as shown in the example below.
The power supply setting also appears for a moment after the 3540
is powered up, allowing you to verify the current setting.
The measurement range is changed by pressing the
key.
Pressing the
key cycles the range selection through the
sequence from 30 mΩ, to 300 mΩ, and so forth up to 30 kΩ.
Pressing the
key cycles the range selection through the
sequence from 30 kΩ,to3kΩ, and so forth down to 30 mΩ.
Auto range
Pressing the AUTO
key toggles auto range selection on or off.
When auto range selection is on, the AUTO mark lights on the
LCD as shown below and the measurement range is selected
automatically as appropriate for the value of the resistance being
measured.
Auto range selection can be turned off by pressing the
keys.
or
or
Auto range selectionManual range selection
NOTE
With the 3540-01, the measurement range can be selected by
range control signal through the external connector on the rear
panel. However, when the measurement range is selected by range
control signal, the setting cannot be changed from the front panel.
Auto range selection is not possible during comparator operation.
Zero adjustment is performed by shorting the test leads, then
pressing first the SHIFT
is only possible with a reading of 100 counts or less.)
Connect the test leads as shown below. The connection must be
made exactly as shown; otherwise, altered lead resistance will
make i t impossible to obtain correct measurement.
With the 3540-01, 3540-02 and 3540-03, zero adjustment can also
be performed by shorting the 0 ADJ and GND terminals on the
rear panel terminal strip.
key, then the 0ADJ .(Zeroadjustment
SENSE
SOURCE
Red
NOTE
NOTE
SENSE
SOURCE
Black
Bring the "v" marks together at the same position.
SENSE
SOURCE
Red
Wrong connectionRight connection
SOURCE
SENSE
Black
If leads other than the specified ones are used, connect as shown
in the following figure.
SOURCE : HSENSE : L
SOURCE : LSENSE : H
connectionconnection
The following is displayed on the LCD during 0 ADJ.
The zero adjust value is maintained internally even when the
power is turned off. However, note that zero adjustment must be
performed for each measurement range to be used.
4.2.4 Switching the Sampling Speed
Pressing the SAMPL
two settings, FAST and SLOW.
FAST setting
16 samples/sec FAST mark lights on LCD
SLOW setting
4 samples/sec FAST mark does not light on LCD
key toggles the sampling speed between
SLOW settingFAST setting
4.2.5 Hold Function
Pressing the HOLD
holds the current measurement value of the display.
With the 3540-01, 3540-02 and 3540-03, shorting the TRIG and
GND terminals on the rear panel terminal strip with display in the
hold state results in taking of one measurement, after which
display returns to the hold state.
Normal (free-running) measurement resumes when the hold state is
canceled.
If the input overloads, the following mark is displayed on the
LCD.
4.2.7 Current Abnormality (CCERR) Detection
Function
If any abnormality is detected in the regularity of current in the
power supply, the curr ent abnormality detection circuit operates
and the symbols indicating CCERR, as shown below, is displayed
on the LCD to advise of current abnormality.
With the 3540-01, 3540-02 and 3540-03, the CCERR signal is
output from the CCERR terminal on the rear panel. (Refer to
"Chapter 5 External Control Features".)
Conditions when current abnormality display
(1) When the resistance being measured is large with respect to the
range
Example: When a 30 Ω resistance is measured in the 30 mΩ
range.
(2) When the test leads are shorted (and only SOURCE is
detected)
(3) When any of the four terminals on the front panel is not
properly connected
(4) When there is an open lead condition
(5) When the fuse is burned out
The 3540 allows storing of up to 7 comparator configuration
tables. Each table can hold the comparator configuration for
measurement range, comparator mode, buzzer and mode.
The results of co mparison are indicated by buzzer, as well as by
lighting of the Hi, IN, and Lo LEDs. With the 3540-01 and 354002, results ca n also be output through the terminal strip on the rear
panel by open collector.
Refer to "5.4.2 O u tputting Measurement Results" regarding
comparator output to the external terminal strip.
NOTE
Auto range selection is canceled if used together with the
comparator.
Since the measurement range is also saved with the comparator
table, switching the comparator table also switches the
measurement range.
If the measurement range is switched during comparator
operation, the measurement range information in the
corresponding comparator table is also changed.
With the 3540-01, the comparator table can be controlled with the
comparator control signal through the external connector on the
rear panel.
However, when the comparator table is selected by the comparator
control signal, the selection cannot be changed with the keys on
the front panel.
Pressing the COMP key lights the comparator display on the
LCD and starts the comparator function, allowing you to make
comparative measurements
Pressing the COMP
key a second time turns off the comparator.
4.3.2 Selecting the Comparator Table
The 3540 allows saving up to 7 tables of comparator
configurations. The comparator configuration is switched by
pressing the TABLE
numbers in the sequence from 1 through 7, then to 1 again.
key to cycle through the comparator table
4.3.3 Selecting the Comparator Mode
Pressing the HI/REF
currently selected comparator table.
Comparator modes that can be selected are the Hi-Lo mode, in
which comparison is done usi ng upper (HIGH) and lower (LOW)
limits, and the REF-% mode, in which comparison is based on a
reference value (REF) or range (%). With operation in the REF-%
mode, the measured value is displayed as a deviation (as a
percentage of reference value).
Pressing the buzzer key switches the buzzer mode of the currently
displayed comparator table.
Available buzzer modes are the HL mode, in which the buzzer
sounds when comparison results are "Hi" or "Lo"; the IN mode, in
which the buzzer sounds when the comparison result is "IN"; and
the OFF mode, in which the buzzer never sounds.
The upper (HIGH) and lower (LOW) limit values are set using the
comparator configuration keys. The configuration range in counts
is from 0 to 9999. (A count is the number resulting after any
decimal point and unit are eliminated from a numeric value
With the REF-% comparator
The reference value (REF) and range (%) are set using the
comparator configuration keys. The configuration range in counts
is 0 to 9999 counts for the reference value (REF), and from 0.0%
to 999.9% for the range (%).
For the REF-% comparator, the following upper and lower limits
are used for comparison.
Upper limit = REF + (REF x % / 100)
Lower limit = REF - (REF x % / 100)
The standard for comparator evaluation is as shown below.
Range of measured valuesComparator result
Upper limit < Measured valueHi
Lower limit ≦ Measured value ≦ Upper limitIN
Measured value < Lower limitLo
LED display for the various comparator results is as follows.
LoINHi
NOTE
If the value set for the lower limit is greater than that set for the
upper limit, the 3540 reverses the values for comparison, taking
the lower limit as the upper limit and vice versa.
If the input limit is exceeded, the comparator result is Hi.
No comparator result is produced if there is a current abnormality.
Comparator results can be output using either of two modes: the
auto mode or the manual mode.
With the 3540-01, 3540-02 and 3540-03, the external control
mode can be selected using the AUTO/MANU selector switch on
the rear panel.
With the 3540, output always uses the auto mode.
Auto mode (AUTO)
During comparator operation, comparator results are output
continuously through the LED, buzzer, and (with the 3540-01,
3540-02 and 3540-03) the comparator result signal terminals (Hi,
IN, Lo) on the rear panel.
Manual mode (MANU)
In the manual mode, comparator output to an external source is
canceled. Comparison results for any desired period can be
obtained by shorting the MANU and GND terminals on the rear
panel terminal strip. Output is canceled when the short is broken.
This function uses the principle of temperature correction (refer to
"1.2 Temperature Correction Function") to convert the resistance
of copper wire to its 20
Connect the 9451 TEMPERATURE PROBE to the TC sensor jack
on the rear panel. For connection p rocedures, refer to "4.5
Temperature Measurement".
When the 9451 TEMPERATURE PROBE is connected, pressing
the TC
temperature correction (display as s hown below).
If the temperature probe is not connected or is connected
incorrectly, when the TC
displayed instead. If temperature correction is not performed as
expected, check connection of the temperature probe.
For error messages, refer to "8.4 Error Code Table".
The temperature probe is not designed to sense surface
temperature. It should only be used to sense ambient air
temperature.
Note also that unless both the sample and the probe have
completely adjusted to the ambient air temperature, the reading
error will be large.
Prior to use, the temperature probe should be connected to the
3540, and both should be allowed to warm up at least 30 minutes.
The temperature probe should be connected or disconnected with
the power of the unit OFF.
Note also that the temperature probe is not watertight. Therefore,
do not allow the instrument to get wet or be immersed in water or
any other fluid.
Large measurement error will result if the temperature probe is not
Turn the power off. And connect the TC adapter as a following
this Figure.
Temperature probe
TC sensor jack
Plug into the TC sensor jack on the panel
NOTE
After turning on the power a nd pressing SHIFT
TEMP
, ambient temperature is sensed by the temperature probe
followed by
and displayed as follows.
If the temperature probe is not connected or is connected
incorrectly, temperature measurement is not performed when the
TEMP
key is pressed, and an "
o
" is displayed instead.
C
If temperature correction is not performed as expected, check
connection of the temperature probe.
For error messages, refer to "8.4 Error Code Table".
After completing the temperature measurement, to proceed to the
resistance measurement, press the SHIFT
TEMP
key.
key followed by
Accurate temperature measurement are not possible if the sheath
of the temperature probe is held with bare fingers.
The temperature probe should be connected or disconnected with
the power of the unit OFF.
Note also that the temperature probe is not watertight. Therefore,
do not allow the instrument to get wet or be immersed in water or
any other fluid.
This chapter explains external control features of the 3540-01,
3540-02 and3540-03.
The rear panel of the 3540-01 is equipped with external connectors
(for BCD output, range and comparator control, etc.), and a n
external terminal strip (for trigger input, comparator output, and so
forth). The 3540-02 and 3540-03 also has an external terminal
strip for functions such as trigger input and comparator output.
The 3540-01, 3540-02 and 3540-03 are equipped wit h external
connectors and/or terminals. Signals assigned to these terminals
can be used to control operation of the 3540 or determine its
status.
Functions of the various terminals and procedures for using the
corresponding signals are described below.
WARNING
Always observe the following precautions when
connecting to an external terminal or external
connectors. Failure to do so may result in electric
shock or damage to the equipment.
(1) Use suitable wires bared at their ends for a length of about 10 mm.
(2) As shown in Figure, depress the knob on the terminal with a
screwdriver, and push the end of the wire into the connection hole.
(3) Release the screwdriver, and the wires will be locked into place.
(4) Use the same procedure to remove the wires
Recommended wire
Usable limits
Strand diameter
Standard insulation stripping length
(1) Make connections to the compatible pins as appropriate for the pin
assignments of the external connector.
(2) Plug the connector wired in (1) firmly into the external connector.
(3) Fasten the plug to the external connector with screws (M2.6).
About the accessory plug
The connector plug (FDCD-37P) provided with the 3540-01 is
equipped with a flat cable pressure connector. Use the following
flat ribbon cable with this connector.
Cable pitch
Core wire
Insulation thickness
: 1.27 mm or 1.38 mm
: AWG#26 to #28 (stranded or single core)
: 0.8 mm to 1 mm
NOTE
A special tool is required in order to make flat cable pressure
connections.
Consult the connector maker for the tool and how to make
connections.
A wide variety of other connectors are available from various
manufacturers which can be used in place of the connector
provided. These include solder connection type connectors,
pressure fitted connectors, and connectors with various types of
covers.
See the various manufacturers' catalogs for connectors that can be
used in place of the one provided.
The maxi mum capacity of this power supply is about 200 mA.
In situati ons requiring more power, use an external power supply.
If transitory current is required, insert an electrolytic capacitor
between 5 V and GND.
Since GND is insulated from the measurement circuit, do not
connect the measurement system to GND.
The voltage (5 V) varies a maximum of approximately 20%
according to the size of the load current on the power supply.
(Approximately 4 V to approximately 6 V) For example, the
power supply voltage will fall below 5 V if the load current
increases.
Power supply (GND + Approx. 5 V)
Approx. 200 mA max.
The ratings given here are absolute maximums. This
means that exceeding these values, even momentarily,
may result in damage to the circuits. Always ensure
that applied voltage and current are below the rated
values.
However, with TTL output, never apply polarized
voltage or current.
Input/output type
TTL input
TTL output5V0V
Absolute maximum
rating
5 V, 20 mA max.
Open collector output35 V, 50 mA max.ONOFF
Signal logic
ValidInvalid
0V5V
Signal logic refers to the signal state in which the function
indicated by the signal is becomes valid.
With open collector signals, the output transistor acts as a switch
between the output signal and GND inside the 3540. When output
is "valid", the switch goes ON and current flows inside the 3540
from the output signal to GND.
Accordingly, this type of output can be used to control connected
LEDs or contact relays, provided that s uch devices operate on less
than 35 V, 50 mA. However, if a relay is connected, be sure to
insert a diode to absorb counter electromotive force.
Multiple open collector outputs can be connected together for use.
This produces a wired OR configuration in which a "valid" signal
any one on the connected outputs will produce a "valid" result.
For example, if a "valid" result is desired from the Hi and Lo
comparator results, connect the Hi and Lo open collector outputs.
This section explains how to use the 3540's various signals and
shows the signals' timing charts.
NOTE
Since the timing charts indicate the logic of the signals, the high
line positions are "valid" and the low line positions are "invalid."
Note that the positions of the lines are not related to the signal
(voltage) levels.
In the I/O code explanation s (for measurement range and
comparator), "0" indicates 0 V, and "1" indicates 5 V.
For an explanation of the signal logic, refer to "5.1 Connectors"
and "5.3 Electrical Specifications".
Refer to "Chapter 4 Operating Procedure" regarding instrument
settings such as the comparator configuration.
The timings shown in the timing charts assume that no key input
is taking place.
In the hold state, measurement starts when TRIG becomes valid,
then EOC becomes valid when measurement ends. EOC then
remains valid until the next time TRIG becomes valid.
TRIG
EOC
Measurement Measurement
5msmin.
FAST: 65 ms max.
SLOW: 260 ms max.
Until TRIG becomes valid
NOTE
FAST: 25 ms min.
SLOW: 115ms min.
During measurement, EOC is also invalidated when the next
TRIG signal is input.
With free-running measurement
With free-running measurement, the 3540 measures repeatedly at
the internal sampling rate, with EOC becoming valid each time
measurement ends.
After switching range codes, up to 550 ms are required for the
internal circuitry to stabilize. Therefore, no measurement should
be taken for at least 550 ms after switching range codes.
Zero adjustment
Zero adjustment is possible using 0 ADJ.
0 ADJ
Zero adjustment Measurement
5msmin.
10 ms max.
NOTE
Measurement is not possible during zero adjustment.
Wait at least 10 ms after completing zero adjustment before
taking measurements.
After switching comparator table, up to 550 ms are required for
the internal circuitry to stabilize. Therefore, no measurement
should be taken for at least 550 ms after switching comparator
table.
Printing (with the 3540-02 only)
Measurements displayed can be output to a printer connected to
the printer connector using the PRINT signal.
Refer to "Chapter 7 Printers".
PRINT
Measurement result Measurement result a Measurement result b Result c
Printout of a Printout of b
5msmax.
NOTE
Output of any previous printout must be completed before
printing the next measurement.
Print time varies according to the speed of the connected printer.
Printout of the measurement displayed starts when the PRINT
signal becomes valid. Before starting printing, check the EOC
signal to verify that measurement has been completed.
Measurement output in the auto mode
(outputting comparator results)
In the auto mode, comparator results (Hi, IN, Lo) and result data
of measurements (BCD, DP) are output when EOC becomes valid
after measurement is completed.
Comparator results and measurement data should be taken after
EOC becomes valid (at the signal's rising edge).
EOC
Measurement result
1msmax.
Previous measurement
result
Measurement a
Result of measurement a
Measurement b
Result of measurement b
Measurement c
Measurement output in the manual mode
(outputting comparator results)
In the manual mode, comparator results (Hi, IN, Lo) are output
only when the MANU signal becomes valid. Measurement data
(BCD, DP) is output continuously.
When a current abnormality occurs (when the resistance being
measured is unchucked), the CCERR signal becomes valid. When
this happens, the all comparator result output becomes invalid.
CCERR
Comparator result
Measurement result
Chucked Unchecked Chucked
1msmax.
1msmax.
NOTE
Comparator results are output even if CCERR becomes valid.
However, comparator results are not output if CCERR is valid
when measurement starts.
About the measurement data
Measurement data is output as 4 digits (each of which is
represented by 4 bits) of BCD data indicating the display count of
the measured value, and as a 3-bit DP code indicating the range.
Meanings of the BCD and DP output codes are as follows.
The RS-232C settings of the 3540-03 are as follows. Since the
3540 settings are fixed and cannot be changed, these settings must
be matched on the computer side.
Transmission mode:
Transfer rate:
Data length:
Parity:
Stop bit:
Hand shake:
Delimiter:
Connect the 3540-03 to the computer using a cross cable.
RS-232C
cross cable
3540
Perform the RS-232C settings on the computer side.
For details on how to make the settings, refer to the instruction
manual for the computer.
6.2.2 Command Transfer Method
The command is issued from the computer.
When the 3540 receives the incoming command from the
computer, it executes the processing specified by the command.
When 3540 has completed processing of the command, it always
returns a response to the computer.
When the computer has confirmed the response, it sends the next
command.
NOTE
Every time the computer has sent a command, a response is always
returned.
Make sure that the computer only sends the next command after it
has received the response to the previous command issued from the
3540.
If multiple commands are sent consecutively, the 3540 may not
execute the commands or command errors may occur.
The command and the parameter are separated by ""(one
character space).
If there is no parameter, send the delimiter after the command.
The command may consist of both upper and lower case letters.
Make sure to use one character space as the separator between the
command and the parameter.
When the command contains a parameter
In the case of
"rng 0" (+ delimiter)
the command format consists of the command "rng" followed by
the separator "". Then follows the parameter "0" . Following the
parameter comes the delimiter.
When the command contains no parameter
"adj" (+ delimiter)
the command format consists of the command "adj" immediately
followed by the delimiter.
NOTE
The meaning of the delimiter is to separate commands and data.
When the 3540 receives the delimiter, it starts analysis of the
command.
Describes the syntax of the command.
Explains the parameter data.
Explains the received data.
Explains the actions specified by the command.
Describes errors that may occur when the command is executed.
Command execution examples.
PC> denotes command from the computer.
3540> denotes command from 3540.
NOTE
The setting conditions and comparator data are written to the nonvolatile memory in the 3540 main unit 5 seconds after the settings
have been made (no key operations should be performed during this
interval).
Measurement mode Resistance measurement mode
Measurement range 30 mΩ range
Sampling speed SLOW
Hold function Free-running (hold state canceled)
Comparator Comparator not used
Temperature measurement value (See "6.3.3 Received Data".)
Obtains the latest temperature measurement value
Obtaining temperature measurement values.
PC >TMES Obtain measurement value.
3540 > 25.6 Temperature measurement value 25.6
PC >TMES Obtain measurement value
3540 >- 5.1 Temperature measurement value -5.1
o
C
o
C
TRG
Trigger measurement
Syntax
Response
<data>
Function
Example
TRG
<rdata>
<rdata>
Resistance measurement value (See "6.3.3 Received Data".)
During measurement hold, measurement is performed once and
this result is captured.
In the free-running state, the newest resistance measurement value
is captured.
Obtaining the resistance measurement value.
PC >TRG Obtain measurement value
3540 > 15.72E-03 Resistance measurement value
ON Resistance measurement completed
OFF Resistance measurement unfinished
Checks whether the resistance measurement is completed.
The condition of the measurement completed is cleared when the
data is readout with this command.
Checks whether the resistance measurement is completed.
PC >EOC Checks whether the resistance measurement is
completed.
3540 >ON Resistance measurement is completed.
PC >EOC Checks whether the resistance measurement is
completed.
3540 >OFF Resistance measurement is not completed.
The EOC command is ineffective in versions 1.02 or before.
In this case, the command error "CMD ERR" occurs when the
EOC command is transmitted.
The version number of 3540 is displayed on the lower right of the
LCD, when powering on.
When attempts are made to perform zero adjustment with a
reading exceeding 100 counts, the zero adjustment error is
generated and zero adjustment is prevented. In these cases, the
execution error is returned as the response.
Executing zero adjustment.
PC >ADJ Execute zero adjustment.
3540 >OK Zero adjustment execution completed.
PC >ADJ Execute zero adjustment.
3540 >EXEC ERR Zero adjustment execution error
FUNC
Function settings
Syntax
<data>
Response
Function
Example
FUNC <func data><func data >
0: resistance measurement mode
1: temperature measurement mode
OK Function switch completed.
Sets the measurement mode to resistance measurement or
temperature measurement.
Setting to temperature measurement mode.
PC >FUNC 1 Set to temperature measurement mode.
3540 >OK Setting completed.
Checks the condition of the current for resistance measurement.
Unless the constant-current flows correctly, the resistance value
cannot be measured.
Checking the current constant-current status.
PC >CCC Find the constant-current status.
3540 >CC ERR Current abnormality (CCERR)
PC >CCC Find the constant-current status.
3540 >CC OK Current normal.
LOCK
Key lock setting
Syntax
<data>
Response
Function
Example
LOCK <on/off><on/off>
0: OFF
1: ON
OK Key lock setting completed.
Toggles the key lock function ON or OFF.
Locking the keys.
PC >LOCK 1 Turn key lock ON.
3540 >OK Setting completed.
This chapter discusses printer output from the 3540-02.
With the 3540-02, measured values can be printed out using the
optional 9203 DIGITAL PRINTER or other general purpose
Centronics printer.
This instruction manual explains how to print out measured values
to a general purpose Centronics printer. For printing using the
9203, refer to the 9203 Instruction Manual.
In order to output measured values to the printer, short the PRINT
signal on the rear panel terminal strip to GND.
For printing using the PRINT signal, refer to "5.4.1 Measurement
Control".
7.1Making Connections
Connect the printer to the printer connector on the rear panel using
the optional 9425 CONNECTION CABLE as shown below.
The connection cable is equipped with two locking claws. Be sure
to press the connector in firmly until these claws click into place.
Always observe the following safety precautions
when connecting a printer. Failure to observe these
safety precautions may result in electrocution or
damage to the equipment.
Always turn off the product and the printer before
making any connections.
Because of the inherent dangers of such situations,
be careful to prevent the connections from coming
loose or leads from coming into contact with other
conductors. Make sure the connections are secure.
This section explains printing to a general purpose Centronics
printer. For printing us ing the 9203, refer to the 9203 Instruction
Manual.
A printout example is shown below, along with the meanings of
the printed data.
NOTE
30.00mohm IN
300.0mohm Hi
3.000 ohm
30.00 ohm
300.0 ohm Lo
3.000kohm
30.00kohm
OF
‑‑
100.0 %Lo
25.0 C
Function
Resistance
measurement
Temperature
measurement
Measurement
Current Abnormality (CCERR)
Resistance measurement
(Standard deviation)
range
30 m
300 mmΩ300.0 mm
3
30
300
30 k
300 k
‑25.0
measured
value
30.00 m
Ω
3.000
Ω
30.00
Ω
300.0
Ω
30.00 k
Ω
300.0 k
Ω
Over
Ω
Ω
Ω
Ω
Ω
Ω
Ω
o
C
Comparator
IN
Hi
Unused
Lo
Unused
Lo
‑
The data outputted from the instrument to printer is ASCII text
only and does not include any control command (except for
CR+LF).
Only the printer, which can print ASCII text directly, is
connectable. (ex. ESC/P printer).
Please be careful that the type of printer, which needs the
exclusive commands to print ASCII text, is not connectable.
To avoid electric shock when replacing the batteries
first disconnect the AC adapter and leads from the
object to be measured. Also, after replacing the
batteries always replace the cover before using the
unit.
Do not mix old and new batteries, or different types
of batteries. Also, be careful to observe battery
polarity during installation. Otherwise, poor
performance or damage from battery leakage could
result.
To avoid the possibility of explosion, do not short
circuit, disassemble or incinerate batteries.
Handle and dispose of batteries in accordance with
local regulations.
The battery mark on the LCD lights when the battery is exhausted.
When this occurs, change the battery as follows.
(1) Remove the battery cover
(2) Replace the batteries with new ones, observing the correct polarity.
(3) Replace the battery cover.
Battery cover
R6P manganese batteries or
LR6 alkaline batteries
NOTE
For exhausted manganese batteries, the battery voltage fluctuates
greatly depending on the current retrieved.
Therefore, when measuring using manganese batteries, even when
the battery voltage is adequately secured at a measurement current
of 1 mA, at 100 mA the battery voltage will not be sufficient,
causing the power supply to cut off or the LCD to keep repeating
To avoid danger of electrical shock when changing
the fuse, disconnect the AC adapter and all test
leads and other wiring from the 3540 before making
replacement.
After replacing the fuse, be sure to close the case
before using the instrument.
Replace the fuse only with one of the specified
characteristics and voltage and current ratings.
Using a non-specified fuse or shorting the fuse
holder may cause a life-threatening hazard.
Fuse type: F1.0 AH/250 V 20x5 mm dia.
To check whether the fuse is burned out set the instrument to the
resistance measurement mode. If NG is displayed on the LCD
when the leads are shorted, the fuse is burned out. The fuse is
mounted on the printed circuit board of the ma in unit. To access
the fuse, remove the screws on the bottom of the case and remove
the upper part of the case. Then, replace the fuse with the spare
fuse in the battery compartment.
If damage is suspected, check the "Troubleshooting" section before
contacting your dealer or Hioki representative.
When sending the product for repair, remove the batteries an d
pack carefully to prevent damage in transit. Include cushioning
material so the instrument cannot move within the package. Be
sure to include details of the problem. Hioki cannot be responsible
for damage that occurs during shipment.
8.6Cleaning
To clean the product, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
Wipe the LCD gently with a soft, dry cloth.