Hioki 3470-01, 3470-02 Instruction Manual

Instruction Manual
3470
MAGNETIC FIELD HiTESTER
December 2008 Revised edition 4 3470A981-04 08-12H
Contents
i
Contents
Introduction ................................................................................1
Inspection ..................................................................................2
Safety Information ................... .................... ................... ...........4
Operating Precautions ................. ................... .................... .......6
1Overview 9
1.1 Product Overview ...............................................9
1.2 Features ........................................................... 11
1.3 Measurement Example ....................................13
1.4 Names and Functions of Parts .........................17
1.4.1 3470 Magnetic Field HiTester ............................17
1.4.2 3471 Magnetic Field Sensor ..............................21
1.4.3 3472 Magnetic Field Sensor ..............................22
1.5 Definition of Measurement Items ......................23
1.6 Screen Configuration and Key Operation
Work Flow ........................................................24
1.7 Measurement Work Flow ................................. 27
2 Measurement Preparations 29
1 2 3 4 5 6 7 8
2.1 Supplying Power ..............................................29
2.1.1 Connecting the AC Adapter ..............................29
2.1.2 Inserting and Replacing Batteries ......................30
2.2 Connecting a Magnetic Field Sensor ...............32
2.3 Turning the Power On and Off .........................33
3 Measurement Procedure 35
3.1 Measurement Preparations ..............................3 5
3.2 Pre-measurement Check .................................37
3.3 Measuring Magnetic Flux Density .................... 38
3.4 Measuring Exposure Level (General Public) .... 41
10 11
APP
Index
ii
Contents
3.4.1 Preparing for Exposure Measurements .............41
3.4.2 Exposure Level ..................................................43
3.4.3 Measurement Procedure ...................................45
3.5 Measuring Exposure Level (Occup.) ................ 48
3.6 Holding Maximum Value of the Resultant
RMS Values .....................................................51
3.7 Checking and Deleting Saved Data .................52
3.7.1 Checking Saved Data ........................................52
3.7.2 Deleting Saved Data ..........................................54
3.7.3 Deleting All Saved Data .....................................57
4 Advanced Functions 59
4.1 Checking Remaining Battery Power ................. 59
4.2 Function for Saving Settings (Saving Set
Modes and Measurement Ranges) .................. 60
4.3 Initializing Function (Returning Settings to
Their Factory Defaults) .....................................61
4.4 Other Functions ................................................62
4.4.1 Setting Auto Power Off (Sav ing Battery Power) 62
4.4.2 Setting Audible Key Feedback ..........................64
4.4.3 Changing Unit Indication ....................... ............66
4.4.4 Setting the Slow Function ..................................68
4.4.5 Turning on the MEM. MODE key in MAX.
HOLD or output mode .......................................70
5 Using Application Software 73
5.1 Overview ..........................................................73
5.2 Installing ........................................................... 74
5.2.1 Installing the Driver ............................................74
5.2.2 Installing software ..............................................82
5.3 Graphic Representation of Measurement
Values ..............................................................85
5.4 Transferring Recorded Data to a PC ................ 88
5.5 Configuring the 3470 ........................................91
Contents
5.6 Viewing Version Information .............................93
6 Advanced Measurements 95
6.1 Outputting Waveform and Resultant
RMS Values .....................................................95
7 Specifications 101
iii
1 2
7.1 3470 Magnetic Field HiTester ........................101
7.1.1 Basic Specifications .........................................101
7.1.2 General Specifications .....................................103
7.1.3 Application Software ................................. .......104
7.1.4 Equations and Functions .................................105
7.2 3471 Magnetic Field Sensor ...........................107
7.3 3472 Magnetic Field Sensor ...........................108
8 Maintenance and Service 109
8.1 Troubleshooting ..............................................109
8.2 Error Indication ...............................................110
8.3 Version Display ..............................................110
8.4 Cleaning ......................................................... 110
8.5 Service ...........................................................110
Appendix 111
Appendix 1 ICNIRP Guideline (Basic Restrictions) ..111 Appendix 2 Coupling Factor in the ICNIRP
Guidelines .............................................112
Appendix 3 Graph Showing the Range of
ICNIRP 1998 Guidelines ....................... 113
3 4 5 6 7 8
10 11
APP
Index
iv
Contents

Introduction

Thank you for purchasing the HIOKI “Model 3470 Magnetic Field HiT­ester.” To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference.
1
Introduction
This instruction manual contains instructions for use of the device with the3471 and 3472 Magnetic Field Sensor.
Note
While every effort has been made to ensure that the content of this man­ual is correct, we would appreciate it if you notified your dealer or local Hioki agent should you notice any unclear, incorrect or missing informa tion.
Trademarks in this manual
Microsoft, Windows, and Windows NT are registered trademarks of Microsoft Corporation in the United States and/or other countries.
Notation
• Unless otherwise specified, “Windows” represents Windows 98, Me, Windows NT, Windows 2000, Windows XP.
• Dialog box represents a Windows dialog box.
• Menus, commands, dialogs, buttons in a dialog, and other names on the screen and the keys are indicated in brackets.
Mouse Operation
Click: Press and quickly release the left button of the mouse. Right-click: Press and quickly release the right button of the mouse. Double click: Quickly click the left button of the mouse twice. Drag: W hile ho lding dow n the left button of the mouse, move the
Activate: Click on a window on the screen to activate that window.
mouse and then releas e the left button to deposit the cho­sen item in the desired position.
-
2 3 4 5 6 7 8 9 10 11 12 13
2

Inspection

Inspection
Verifying Package Contents
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki rep resentative.
3470-01 Magnetic Fiel d HiTester
-
3470 (1) USB cable (1)CD (PC application
software) (1)
LR6 alkaline battery (4)
3471 Magnetic Field Sensor (1)
9445-02 AC Adapter
or
9445-03 AC Adapter (EU)
(1)
Carrying Case (1)
Instruction manual
(1)
Inspection
3470-02 Magnetic Fiel d HiTester
3
3470 (1) USB cable (1)CD (PC application soft-
LR6 alkaline battery (4)
9758 Extention Cable
(1) 9759 Output Cable (1 ) Carrying Case (1)
ware) (1)
9445-02 AC Adapter
or
9445-03 AC Adapter (EU)
(1)
3471 Magnetic Field Sensor (1) 3472 Magnetic Field
Instruction man-
ual (1)
Sensor (1)
2 3 4 5 6 7 8
This instrument has been calibrated to operate with the provided sen­sor and should be used only with this sensor. Thus if multiple instru­ments are used, Before use make sure that the matching number (up to the hyphen) on the tester and the sensor are identical.
Shipping precautions
Use the original packing materials when transporting the instrument, if possible.
9 10 11 12 13
4

Safety Information

Safety Information
This instrument is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the instrument. Be certain that you understand the ins truct i ons and precautions in the manual before use. We disclaim any responsibility for accidents or injuries not resulting directly from instrument defects.
This manual contains information and warnings essential for safe opera­tion of the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the f ollowing sa fety precautions.
Safety Symbols
In the manual, the symbol indicat es p art icul ar ly im por ­tant information that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the man­ual (marked with the symbol) before using the relevant function.
Indicates DC (Direct Current).
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a signifi­cant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possi­bility of injury to the user or damage to the instru­ment.
Indicates advisory items related to performance or correct operation of the instrument.
Other Symbols
(⇒ p. )
*
5
Safety Information
Indicates the location of reference information. Indicates that descriptive information is provided below.
Indicates a prohibited action.
Measurement Categories (Overvoltage categories)
To ensure safe operation of measurement instruments, IEC 61010 establishes safety standards for various electrical environments, catego rized as CA T I to C AT IV, and call ed measurement cat egories. The se are defined as follows.
CAT I : Secondary electrical circuits connected to an AC electrical CAT II : Primary electrical circuits in equipment connected to an AC
CAT III : Primary electrical circuits of heavy equipment (fixed installa-
CAT IV : The circuit from the service drop to the service entrance, and
Higher-numbered catego­ries correspond to electri­cal environments with greater momentary energy. So a measure ment device designed for CAT III envir onment s can endure greater momen tary energy than a device designed for CAT II. Using a measurement instrument in an environment designated with a higher-numbered category than that for which the instrument is rated could result in a severe accident, and must be carefully avoided. Never use a CAT I measuring instrument in CAT II, III, or IV enviro nment s. The measurement categories comply with the Overvoltage Categories of the IEC60664 Standards.
outlet through a transformer or similar device. electrical outlet by a power cord (portable tools, household
appliances, etc.) tions) connected directly to the distribution panel, and feeders
from the distribution panel to outlets. to the power meter and primary overcurrent protection device
(distribution panel).
-
-
2
-
3 4 5 6 7 8 9 10
Do not touch high voltage lines with the sensor or the instrument. They are not insulated and the application of high volt age to this instrument may expose testing personnel to high voltage that may at worst lead to electric shock and death.
-
11 12 13
6

Operating Precautions

Operating Precautions
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Preliminary Checks
• Before using the instrument the first time, verify that it operates nor­mally to ensure that the no damage occurred during storage or ship­ping. If you find any damage, contact your dealer or Hioki representative.
• Before using the instrument, make sure that the insulation on the cables is undamaged and that no bare conductors are improperly exposed. Using the instrument in such conditions could cause an electric shock, so contact your dealer or Hioki repr esent ative for rep air.
Installation Environment
Operating temperature/humidity range: 0 to 40°C, 80%RH or less (no condensation) Temperature and humidity for guaranteed accuracy: -10 to 50°C, 80%RH or less (no condensation)
Avoid the following locations that could cause an accident or damage to the instru­ment.
Exposed to direct sunlight Exposed to high tempera­ture
Exposed to liquids Exposed to high humidity or condensation
Exposed to high levels of particulate dust
In the presence of corro­sive or explosive gases
Exposed to strong elec­tromagnetic fields Near electromagnetic radiators
Subject to vibration
Connections
Connect only designated devices to the instrument terminals. To pre­vent damage, do not apply a voltage that is outside the specified range to the instrument.
Handling
7
Operating Precautions
• The instrument and the 3471 and 3472 Magnetic Field Sensors are built mainly from nonmagnetic metals and contain hardly any mag netic materials. Thus if the inst rument is heated by an all metal IH and other induction heating system the sensor may melt or be damaged, because of the nonmagnetic metals.
• The 3471 and 3472 Magnetic Field Sensors have a plastic housing that may melt if brought into contact with a hot measurement object (cookpot or other hot object).
• Do not expose cables t o exces siv e str ess as they may br eak . Esp ec iall y, the 9758 Exten tion Ca ble and 975 9 Output C able shou ld be hand led with care.
Handling with the CD
• Always hold the disc by the edges, so as not to make fingerprints on the disc or scratch the printing.
• Never touch the recorded side of the disc. Do not place the disc directly on anything hard.
• Do not wet the disc with volatile alcohol or water, as there is a possi­bility of the label printing disappearing.
• To write on the disc label surface, use a spirit-based felt pen. Do not use a ball-point pen or hard-tipped pen, because there is a danger of scratching the surface and corrupting the data. Do not use adhesive labels.
• Do not expose the disc directly to the sun's rays, or keep it in condi­tions of high temperature or humidity, as there is a danger of warping, with consequent loss of data.
• To remove dirt, dust, or fingerprints from the disc, wipe with a dry cloth, or use a CD cleaner. Always wipe radially from the inside to the outside, and do no wipe with circular movements. Never use abra sives or solvent cleaners.
• Hioki shall not be held liable for any problems with a computer system that arises from the use of this CD, or for any problem related to the purchase of a Hioki product.
-
-
2 3 4 5 6 7 8 9 10 11 12 13
8
Operating Precautions

1.1 Product Overview

Overview 1
1.1 Product Overview
The 3470 Magnetic Field HiTester is designed to measure magnetic flux density and level of magnetic field exposure.
• It can be used to assess conformance to ICNIRP 1998 and EN50366 (IEC62233) and in research on magnetic field exposure.
• The instrument is designed to be used with the 3471 and 3472 Mag­netic Field Sensors. Make sure that the matching number (up to the hyphen) on the tester and the magnetic field sensor are identical.
9
1 2 3 4
• Magnetic field exposure refers to the exposure that may affect human health. The 3470 Magnetic Field HiTester can be used to measure conform­ance to Guidelines for Limiting Exposure to Time-varying Elec- tric, Magnetic, and Electromagnetic Fields (ICNIRP 1998) and a standard for measuring magnetic fields of electrical household appliances (EN50366).
For more information on ICNIRP: "Appendix" (P.111)
• This tester can measure magnetic flux density and magnetic field exposure generated by products manufactured at a customer site or produced by devices in a customer environment. It can measure magnetic flux density and magnetic field exposure generated by products manufactured at a customer site or produced by devices in a customer environment.
5 6 7 8 9 10 11 12 13
10
1.1 Product Overview
Functions Functions in detail Reference
Basic Magnetic flux
Application Output
density
measurements
Exposure
level
measurements
functions
Application
software use
*1: Available units are T, G, and A/m and the magnetic permeability of air being 4π×10-7 H/m, *2: This value is obtained by inverse filtering of magnetic flux density values stated in
Measuring the magnetic flux density of appli-
*1
ances and environments
Magnetic flux density (10 Hz to 400 kHz): For wide range measurement s of magne tic fl ux densit y
Magnetic flux density (10 Hz to 2 kHz): For magnetic flux density of power frequencies
Magnetic flux density (2 kHz to 400 kHz): For magnetic flux density of hig h frequencies other than power frequencies
*2
Measuring the exposure level of appliances and environmen t s
Exposure level (general public): This level is used f or measuring exposur e to mag­netic fields prevalent in environments where we can assume the presence of people of all age groups and health conditions (people who are not normally aware of the presence of magne tic field s) as well as exposure to magnetic fields gener ated by appli anc es and other consumer devices manufactured to conform to EN50366 (or IEC6233) and other st an dards. This level corresponds t o the general public exp o­sure level in ICNIRP 1998.
Exposure level (occupational): This level applies to measurements of exposure to magnetic fields in environments that involve adults only who are generally exposed under known conditions and are trained to be aware of potential risk and to take ap propriate prec autions. This level corres ponds to the occupational ex po sure level in ICNIRP 1998.
A function that makes it possible to output
3.3 (P . 38)
3.4 (P . 41)
-
-
-
6.1 (P . 95)
3.5 (P . 48)
waveforms and resultant RMS value of mag­netic flux density or exposure levels.
• This function allows you to do the following:
5 (P. 73)
• to set up the 3470 from your PC
• to upload data saved on the 3470 to your PC
• to monitor RMS values
4
the relationships between th e differ ent uni ts are as fol lows : 1T=10 the ICNIRP1998 Guideline.
Use the exposure level figures indicated for the general public in IEC62233 or Since the reference levels for magnetic field exposure indicated in ICNIRP1998 for occu pational exposure is approximately five times that of general public exposure, the in­dicated values for occupational exposure is roughly a fifth of general public exposure.
"Appendix" (P.111)
G, 1 A/m=4π×10-7 T.
-
How big is the magnetic field?
1.2 Features
11

1.2 Features

Easy to view LCD displ ay
The large LCD display makes for ease of viewing measurement values.
Automatic recognition of magnetic field sensors
Two magnetic field sensors of different sizes, the 3471 and 3472, are available for the tester
The 3471, inte nded fo r normal appli cation , has a cr oss- sectio nal a rea of
2
100 cm The 3472 sensor has a cross-sectional area of 3 cm2 and is designed
for measuring d et a ile d di stri but ion .
The 3470 automatically recognizes an attached sensor so there is no need to set a conversion rate whenever a sensor is connected.
Switchable measu re ment methods
The tester allows the user to switch between the following measurement modes: magnetic flux density, exposure (General Public), and exposure (Occup.).
Switchable axis
The user can switch between different axes (x-axis, y-axis and z-axis and resultant values) in each measurement mode
.
3471
3472
1 2 3 4 5 6 7 8 9
Waveform and resultant RMS value output
Both waveform ( ) and resultant RMS ( ) output can be selected. With a 9759 Output Cable you can connect the 3470 to an oscilloscope or recorder for output of waveforms and resultant RMS values.
Waveform out­put :
Resultant RMS value output :
enables measurement of magnetic field waveforms when the tester is connected to an oscilloscope or recorder.
permits recording of resultant values on a recorder or log­ger, which is convenient for long-term measurements.
10 11 12 13
12
1.2 Features
Memory function
Records up to 99 measurement data.
Function for saving settings
Measuring conditions can be stored to enable quick access to the same operating conditions each time you power up.
Dual power supply
The tester can be powered both by battery and AC power to enable measurements in the field as well as permitting installation for long-term measurements.

1.3 Measurement Example

1.3 Measurement Example
This section provides a measurement example using this tester.
Objective: to measure magnetic field exposure generated by products to determine
whether they pass or fail the requirements to obtain a CE marking.
Measurement location : Anechoic chamber*1 Device under test : exposure level (general public) of magnetic
Measurement location, distance*2: around, 30 cm
State of device under test *2 : Half-filled with water, and maxi­Measurement time : a few seconds at each measurement point
fields generated by an electric rice cooker
*3
mum heat setting
DUT (device under test)
13
1 2 3 4 5
Be sure to measure areas that generate maximum exposure to magnetic fields.
In this example, r1 = 30 cm.
Storage of measurement data: tester memory and PC memory Required equipment : 3470
*1: A location where magnetic fields other than that generated by the DUT is minimal
and where the re are no magnet ic fie lds or me tal ob jects in the vi cinity that c ould be affected by the measurements. An anechoic chamber is not required if the above conditions can be met. Measurements performed in an anechoic chamber should be performed at a cer­tain distance from the walls, which are made of a magnetic material.
*2 : Since this state depends on the DUT, refer to the IEC62233, EN50366 or other rel-
evant specifications to determine the required state of the DUT during testing.
*3 : This value comes from IEC62233 Ed.1.0. Note that it may change when the stan-
dard is revised.
3471 or 3472 sensor Batteries AC adapter Product whose magnetic field is to be tested PC
6 7 8 9 10 11 12 13
14
1
1.3 Measurement Example
Preparations
1. Insert batteries in the tester or connect it to the AC adapter and fold out the stand. Also connect the 3471 or 3472 M agnetic Field Sensors.
Insert th e sensor in the sensor terminal.
Plug the adapter into the AC adapter terminal and the mains plug into an AC wall outlet.
Insert batteries in the battery compartment on the rear and fold out the stand.
2. Connect the tester to a PC using a USB cable.
USB cable
3. Install the driver and the PC software on the PC.
15
2
3
1.3 Measurement Example
Setup, measurements and data storage
1. Select Exposure (general public) measurement mode.
2. Select Auto range. ([AUTO] on the display lights. )
3. Place the magnetic field sensor near the DUT. (A distance of 30 cm if the DUT is an electric cooker)
Read the indicated value and press to save the m easurement data.
4. Read the indicated value and store the measurement data in the tester.
1 2 3 4 5
Processing data after measurement
1. Start up the software and transfer the measurement data stored in the instrument to the PC.
2. Analyze the data transferred to the PC using Excel or other spread­sheet program.
6 7 8 9 10 11 12 13
16
1.3 Measurement Example
Analysis flow chart
START
Measurement data ± accuracy
< 100%
*
Yes
100%
(Measurement data ± accuracy) ×
coupling factor (0 to 1)
100%*
Yes
100%
Exceeds
Does not comply
* : Whether tester accur acy sho uld be added or subt racted depen ds on th e sta ndpoint
of testing personnel. Refer to the instructions in the IEC62233 standards. For details on the coupling factor, refer to Appendix 2 "Coupling Fa ctor in the IC­NIRP Guidelines" (P.112).
*2 : This cannot be measured by a magnetic field tester.
Calculate the induced current
density with the human body
model and a numerical analysis.
Evaluation goes below the basic
restrictions.
Yes
Complies

1.4 Names and Functions of Parts

1.4 Names and Functions of Parts

1.4.1 3470 Magnetic Field HiTester

17
1
Front
Display
(P. 20)
AC adapter terminal
Connect the AC adapter here.
2.1.1 (P. 29)
Back
Matching number
Use the tester with a sensor that has an identical matching number (up to the hyphen).
Sensor terminal
Connect the 3471, 3472 Magnetic Fi eld Sensor o r a 9758 Extent ion Ca ble.
2.2 (P. 32)
Keys
(P. 18)
Output terminal
Connect a 9759 Output Cable.
USB terminal
Connect a USB cable.
Stand
When setting up the meter on a desk or other flat surface fold out the stand to make sure that the magnetic field sensor does not come into contact with the surface underneath.
6.1 (P. 95)
Camera tripod screw
Attach a commercially avail­able camera tripod here.
2 3 4 5 6 7 8 9 10
When using the handle as a stand for the device, do not press down too hard on the device as this can dam­age the handle.
Battery compartment lid
Open to insert or replace bat­teries.
2.1.2 (P. 30)
11 12 13
18
1.4 Names and Functions of Parts
keys
1
7
8 2
3 5
4
Keys Description
1
2 Switches magnetic fiel d mea sur em ent modes as sho w belo w:
3
4
5
6 Shows saved measurement data.
• Turns the power on and off.
• Holding down while turning on the power initia lizes the memory and the saved settings (returning them to their factory defaults).
For details on the factory defaults: 4.3 (P. 61)
• Holding down th e whi le power in g up al lows you t o set the Au to po we r off, audible key feedback, unit indication, slow function, and MEM.MODE
key
enable/disable.
4.4 (P. 62)
Magnetic flux density (10 Hz to 400 kHz) magnetic flux density (10 Hz to 2 kHz) magnetic flux density (2 kHz to 400 kHz) sure level (Occup.) magnetic field density (10 Hz to 400 kHz) ...
3.3 (P. 38)
• Swit ches between Auto and Manual ranges. Manual range: (magnetic flux density mode) AUTO r0 (2 μT/20 mG/1.6 A/m) r1 (20 μT/200 mG/16 A/m) r2 (200 μT/2 G/160 A/m) r3 (2 mT/
G/1600 A/m) AUTO...
20 (exposure level mode) AUTO r0 (20%) r1 (200%) r2 (2000%) AUTO... Auto range: automatically selects appropriate range during measurements.
• Switches measurement ranges duri ng ma nual range.
3.3 (P. 38)
• Magnetic flux density switching is performed by converting from T (Tesla). For this reason, for example, 1.592 A/m is the maximum indicated value in the 1.6 A/m range.
• Use to output.
• Switches between the following output functions. OFF Waveform output r e s u ltant RMS va lue output → OFF...
• This mode disables the auto power off mode.
6.1 (P. 95)
• Holds the maximum value. (Each new maximum value updates the indicated value.)
• Press this key again to unhold.
3.6 (P. 51)
3.7 (P. 52)
exposure level (General Public) expo-
6
1.4 Names and Functions of Parts
Keys Description
7
8 Switches the displayed axis as shown below:
• Stores measurement data in memory.
• Clears displayed measurement data when viewing memory data.
3.7 (P . 52)
XYZ→X→Y→Z→XYZ...(XYZ indicates a resultant value
1.5 "Definition of Measurement Items" (P.23)
*
) *:
X2+Y2+Z
19
1
2
2 3 4 5 6 7 8 9 10 11 12 13
20
.
1.4 Names and Functions of Parts
Display
Note: Icons no t de scr i b ed be lo w c a nno t b e u s ed wi t h t his t e st e r
76
8
1, 2
3, 4
9 10
11 12
5
Memory View Mode
Displays saved data.
3.7 (P . 52)
Magnetic flux density mode
Displays t he s ele c t ed ma gn eti c f lux d en s i­ty mode.
3.3 (P. 38)
Icons Description
1 Lights when Auto Range is selected. 2 Lights when measurement data is being stored. 3 Shows displayed axis. (XYZ indicates a resultant value*.)
*
1.5 "Definition of Measurement Items" (P.23)
4 Lights when holding a maximum value. 5 Lights when bat ter y vo ltage becomes too l ow. (Replace the batteries w hen t hi s
6 Lights during output of waveform. 7 Lights during output of a resultant RMS value.
8 Lights when automatic power off is on. 9 Indicates the number of the measurement data stored in memory.
10 Indicates that a measurement value is an RMS value. 11 Indicates the unit used for indicating exposure level.
12
3.6 "Holding Maximum Value of the Resultant RMS Values" (P.51 )
icon lights.)
"Function for checking low battery voltage" (P.36)
3.7 "Checking and Deleting Saved Data" (P.52)
Flashes during the slow function is on.
3.4 "Measuring Exposure Level (General Public)" (P .41) ,3.5 "Measuring Exposure Level
(Occup.)" (P. 48)
• Indicates the unit used for indicating magnetic flux density.
3.3 "Measuring Magnetic Flux Density" (P.38)
• "mV" indication isn’t used in this instrument.
X2+Y2+Z
2
Exposure level mode
Displays the selected exposure level modes as follows: General Public Occup. (Occupational)
3.4 (P. 41) ,3.5 (P. 48)
1.4 Names and Functions of Parts

1.4.2 3471 Magnetic Field Sensor

z
yx
Sensor
Output conn ector
Connects to the 3470 sensor connector.
2.2 (P. 32)
Matching number
Use the tester with a sensor that has an identical matching number (up to the hyphen).
21
1 2 3 4
x-axis coil
The dotted line indicates the x-axis coil.
x-axis
z-axis coil
z-axis
The dotted line indicates the z-axis coil.
y-axis coil
The dotted line indicates the y-axis coil.
y-axis
5 6 7 8 9 10 11 12 13
22
1.4 Names and Functions of Parts

1.4.3 3472 Magnetic Field Sensor

Output connector
Connects to the 3470 sen­sor connector.
2.2 (P. 32)
The line indicates the x-axis coil.
z yx
Matching number
Sensor
y-axis coil z-axis coil
The line indicates the z-axis coil.
Use the tester with a sensor that has an identical matching number (up to the hyphen).
y-axis
x-axis
x-axis coil
z-axis
The line indicates the z-axis coil.

1.5 Definition of Measurement Items

)(tx)(ty)(tz)(ˆtx)(ˆty)(ˆt
)
ˆ
tR
)
ˆ
tR
)(ˆtx)(ˆty)(ˆtz)(tx)(ty)(t
1.5 Definition of Measurement Items
Instantaneous values obt ained from the x- axis, y-axis and z-axis sensors at time t are referred to as, respectively, RMS values are respectively , and . The RMS values used
by this tester represent the values shown below. (T when the slow func­tion is turned off is about 0.12 s and about 1 s when the function is on.)
t
ˆ
tx
)(
ˆ
ty
)(
ˆ
tz
)(
1
=
=
=
Tt
T
t
1
Tt
T
t
1
Tt
T
2
dttx
)}({
2
dtty
)}({
2
dttz
)}({
, and while the
z
23
1 2 3 4
A resultant RMS value represents the following.
ˆ
• The IEC62233 and EN50366 standards state that "Transient mag­netic fields with a duration of less than 200 ms, e.g. during switching events, are disregarded. If a switching action o ccurs d uring t he mea surements, the measurement has to be repeated." Transient responses may not be correctly measured when slow function is off. Make a new measurement if this happens.
• T o make a 1 s measurement that includes transient responses, mea­sure using the slow function while being mindful of the timing that transient responses occur.
4.4.4 "Setting the Slow Function" (P.68)
The above can be summarized as shown below.
Resultant RMS value :
RMS value (per axis):
Instantaneous value (per axis):
These are the same both in magnetic flux density mode and exposure level mode.
Resultant RMS and RMS values (for each axis) can be viewed on a PC using the application software (on CD) supplied with the tester .
5.1 "Overview" (P.73)
ˆ
{}{}{}
)(
)(
(
ˆ
)(
222
ˆ
)(
tztytxtR ++=
(
, , :
, , :
-
LCD reading, RMS
:
output ( ) LCD reading
z
Waveform output (
)
5 6 7 8 9 10 11 12 13
24
.
.

1.6 Screen Configuration and Key Operation Work Flow

1.6 Screen Configuration and Key Operation Work Flo w
3.3 (P . 38)
Magnetic flux density mode (10 Hz to 400 kHz)
3.3 (P . 38)
Press .
Magnetic flux density mode (10 Hz to 2 kHz)
Measurement mode
Tindicates selected mode.
Magnetic flux density
mode
3.5 (P. 48)
Exposur level (Occup.) mode
Exposure
level mode
Press
3.3 (P . 38)
Press .
Magnetic flux density mode (2 kHz to 400 kHz)
Press .
3.4 (P. 41)
Exposure level (General Public) mode
Press
1.6 Screen Configuration and Key Operation Work Flow
Screens for verifying and setup
25
Screen name Screen access method Screen displayed
Check Battery status
Check Saved Data
Set Auto Power Off When the power is off, hold
Press to view battery status.
3.1 (P. 35), 4.1 (P. 59)
Press in each measure­ment mode.
3.7.1 (P. 52)
down and and press or to turn
on
[APS].
4.4.1 (P. 62)
1 2 3 4 5 6 7 8
Set Audible Key
Feedback
When the power is off, hold down and . And
press or to turn
BEEP].
on [
4.4.2 (P. 64)
9 10 11 12 13
26
1.6 Screen Configuration and Key Operation Work Flow
Screen name Screen access method Screen displayed
Change Unit Indication When the pow e r is of f, ho l d
down and . And press or to turn
on [
Unit].
4.4.3 (P.66)
Set Slow Function When th e p o wer i s off , h o l d
down and . And press or to turn
SLou].
on [
4.4.4 (P. 68)
Enable/disable MEM.
MODE Key in MAX.
HOLD or Output Mode
QuickSet
The slow function is on in Exposure l evel (General Public) mode when power­ing on.
When the power is off, hold down and . And
press or to turn on [
MEM.M].
4.4.5 (P. 70)
When the power is off, hold down and .
(P. 42)

1.7 Measurement Work Flow

1.7 Measurement Work Flow
This section describes measurement work flow.
Supplying power
1.
2.1.1 "Connecting the
AC Adapter" (P.29)
2.1.2 "Inserting and
Replacing Batteries" (P.30)
Connect a mag-
2.
netic field sensor.
2.2 "Connecting a
Magnetic Field Sen­sor" (P.32)
27
1 2 3 4 5 6 7
Turn on the power.
3.
2.3 "Turning the Power
On and Off" (P.33)
8 9 10 11 12 13
28
5.
1.7 Measurement Work Flow
4. Pre-measurement check
3.2 "Pre-measurement Check" (P.37)
Measuring mag­netic fields
3.3 "Measuring Mag-
netic Flux Density" (P.38)
Transfer the mea-
6.
surement data to a PC and use the application soft ware to analyze the data.
5.3 "Graphic Repre-
sentation of Measure­ment Values" (P.85)
5.4 "Transferring
Recorded Data to a PC" (P.88)
-

2.1 Supplying Power

9
Measurement Preparations 2
2.1 Supplying Power

2.1.1 Connecting the AC Adapter

Use only the specified Model 9445-02 AC Adapter (SA110-09S-I, SINO-AMERICAN) AC adapter input voltage range is 100 to 240 VAC (with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage to the instrument, do not apply voltage outside of this range.
• Make sure the power is turned off before connecting or disconnect­ing the AC adapter.
• Use the AC adapter for long-term measurements.
Connect the AC adapter according to the steps below
1. Connect the 9445-02 AC Adapter to the AC adapter terminal.
29
2
2 3 4 5 6 7 8
445-02 AC Adapter
9 10
2. Connect th e AC adapter plug to an AC wall outle t .
This completes AC adapter connection.
11 12 13
30
2.1 Supplying Power

2.1.2 Inserting and Replacing Batteries

• Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result.
• To avoid the possibility of explosion, do not short circuit, disas­semble or incinerate batteries.
• Handle and dispose of batteries in accordance with local regula­tions.
• The indicator appears when battery voltage becomes low. Replace the batteries as soon as possible.
• Use battery power to run the tester for brief measurement sessions in locations without AC power.
• Disconnecting and connecting the AC adapter during measure­ments when the tester is both battery and AC powered may cause the measurement values to fluctuate greatly.
Use the following steps to insert (replace) the batteries. Required batteries: LR6 alkaline battery
1. Remove th e battery compartment li d.
Remove
Battery compartment lid
2. When replacing batteries, first remove the old batteries.
31
2.1 Supplying Power
3. Insert the batteries taking care to match the pola rit y correctly.
2
2 3
4. Attach the battery compartment lid.
Attach
Battery compartment lid
This completes the battery insertion (replacement) procedure.
4 5 6 7 8 9 10 11 12 13
32

2.2 Connecting a Magnetic Field Sensor

2.2 Connecting a Magnetic Field Sensor
Use the sensor within its specifie d range. In a high fre­quency and strong magnetic field, the sens or w ill output a high voltage that may cause electric shock or sensor damage when someone touches an unconnected output terminal. Observe the rating and the derating curve of the sensor during operation.
Connect the magnetic field sensor (3471 or 3472) accor ding to the st eps below.
The 3471, intended for normal application, has a cross-sectional area of 100 cm The 3472 sensor has a cross-sectional area of 3 cm2 and is
designed for measuring detailed magnetic field distribution.
2
.
Locate the Z-axis of the magn etic field sen sor facing forwa rd, insert the output connector of the sensor in the 3470 sensor connector and screw the nut of the se nsor in to firmly at tach the sensor. The sensor automatically recognize s the connected m agnetic field sensor.
Output connector
Sensor connector
To avoid damaging the output cable, grasp the connector, not the cable, when unplugging the cable.
This completes connection of 3471 and 3472 magnetic field sensors.

2.3 Turning the Power On and Off

2.3 Turning the Power On and Off
Turn the power on (off) according to the steps below.
T urning on the power
Press .
33
2
2 3 4 5
T urning the power off
Hold down for 2 seconds or m ore.
2 seconds or more
6 7 8 9 10 11 12 13
34
2.3 Turning the Power On and Off

3.1 Measurement Preparations

Measurement Procedure 3
3.1 Measurement Preparations
35
Before the start of measurements, make the following preparations in the stated order.
1. Press to turn on the power.
2. Make sure th at all s egm ents on the LCD display light.
After this it will display the model name (3470) and battery sta­tus.
2 3
3 4 5 6 7 8 9 10 11 12 13
36
3.1 Measurement Preparations
Function for checking low battery voltage
When battery voltage goes low and lights, the power is forcibly turned off. At this time, [BAtt] [Lo] also appear.
3. The magnetic flux den si ty m ode screen appears.
This ends measurement preparations.
• When a setup has been saved (for details refer to 4.2 "Function for Saving Settings (Saving Set Modes and Measurement Ranges)" (P.60)), the set measurement mode screen appears.
• To initialize the tester (to return it to its factory defaults), turn off the power and then hold down
4.3 "Initializing Function (Returning Settings to Their Factory Defaults)" (P.61)
and press .
3.2 Pre-measurement Ch eck
This instrument has been calibrated to operate with the optional sen­sors and should be used only with these sensors. Thus if multiple instruments are used, make sure that each instrument is combined with the correct sensor.
Check the following before using the tester.
Inspection flow chart
1.
Check whether the mag­netic field sensor, the tester case, the 9759 Out­put Cable and the USB cable are damaged.
Damaged

3.2 Pre-measurement Check

Do not use the tester if any of the cables are da maged and the cable metal core is visible.
37
2 3
3 4 5
No
damage
2.
Check if the battery is depleted after powering up. Check battery status.
4.1 "Checking Remaining Battery
Power" (P.59)
Sufficient
battery
power
Start
measurements
Magnetic field sensor and tester case must be repaired. Contact your dealer (agent) or local sales office.
Replace the 9759 Output Cable and USB cable with new cables.
Insufficient
battery
power
lights when battery power is low. The tester is not able to make accu­rate measurements when battery power is low. Replace the battery.
2.1.2 "Inserting and Replac-
ing Batteries" (P.30)
6 7 8 9 10 11 12 13
38

3.3 Measuring Magnetic Flux Density

3.3 Measuring Magnetic Flux Density
Measure magnetic flux density according to the steps below.
5
21
1. Press to open the magne ti c flux density mode screen.
: wide range measurements
: power frequency range measurements
: extended power frequency range measurements
3.3 Measuring Magnetic Flux Density
2. Press to select measure m ent range.
Range number
Ranges are switched in the order shown belo w.
AUTO r0 r1 r2 r3
39
2 3
3
Range type Range number Range
Auto-range
Manual range r0 2 μT/20 mG/1.6 A/m
Magnetic flux density switching is performed by converting from T (Tesla). For this reason, for example, 1.592 A/m is the maximum indicated value in the 1.6 A/m range.
• The ranges for the x, y and z axes are interlocked with maximum range and are automatically switched.
Range of guaranteed accuracy
At resultant value R, a count of less than 100 is outside guaranteed accuracy. Use an appropriate range.
Guaranteed accuracy range for resultant value R
Magnetic flux density (10 Hz - 400 kHz, 10 Hz - 2 kHz, 2 kHz - 400 kHz)  Fixed range
0.1 uT
The magnetic flux densi t y for 3472 (10 Hz - 400 kHz, 10 Hz - 2 kHz) is 0.4 uT.
1 uT 20 uT
2 uT
3.464 uT
10 uT
appears on the
display
r1 20 μT/200 mG/16 A/m r2 200 μT/2 G/160 A/m r3 2 mT/20 G/1600 A/m
0.1 mT 2.0 mT 200 uT
34.64 uT
Automatically selects
appropriate range dur
ing measurements.
346.4 uT
3.464 mT
-
r3 r2 r1 r0
4 5 6 7 8 9 10 11 12 13
40
3.3 Measuring Magnetic Flux Density
3. Place the mag netic field se nsor near the DUT.
• For accurate measurements, the tester should be located further away from the DUT than the sensor.
• Although few magnetic materials are used in the magnetic field tester, the battery is a magnetic material and may affect measure ments if located close to the DUT. In this situation, it may be a good idea to use an extension cable to increase the distance between the sensor and the tester or to remove the battery and use the AC adapter.
-
4. Read the indicated value.
5. Press to save the mea s urement data.
This ends magnetic flux density measurements.

3.4 Measuring Exposure Level (General Public)

3.4 Measuring Exposure Level (General Public)
Use this mode to make measurements that conform to EN50366 and IEC62233.

3.4.1 Preparing for Exposure Measurements

41
2
Use the flow chart below to determine whether to use the slow function setting.
Slow function setting assessment flow chart
Use this test to determine whether the test of a household appliance should conform to IEC62233 or EN50366.
START
Waveform of mag-
netic field to be mea-
sured is known
Yes
Yes No
Turn off the slow
function
*: Since turning the slow function on or off will not change the value of a magnetic field
that periodically fluctuates by 10 Hz or more, turn the slow function off to speed up measurements. This is beca use movi ng t he sen sor caus es th e det ec tion of AC curr ent gene rate d by a DC magnetic field such as the terrestrial magnetic field and other AC magnetic fields. Twice the integral time is then required including processing time for the mea surement value to stabilize.
A magnetic field that
periodically fluctu-
ates by 10 Hz or
more?*
No
Turn on the slow
function
-
3
3 4 5 6 7 8 9 10 11 12 13
42
3.4 Measuri ng Exposure Level (General Public)
The result of the flo w chart assessment is "Turn on the slow function."
Use the steps below to open the "QuickSet" screen. (An exposure level (General Public) measurement can be started after turning on the slow function.)
When the power is of f, hold dow n and press .
The "QuickSet" screen opens
After this follow the procedures 2. (P.46) described in 3.4.3 "Measurement Procedure" to perform an exposure level (Gen­eral Public) measurement.
The result of the flow chart assessment is "Turn off the slow function."
The result of the flow chart assessment is "Turn off the slow function." Turn off the slow function as described in 4.4.4 "Setting the Slow Func­tion" (P.68). (The default setting is "Off.")
43
g
3.4 Measuring Exposure Level (General Public)
The table below shows the measurement and internal processing times with the slow function on and when it is off.
Slow function on
1 s 1 s 1 s 1 s 1 s 1 s 1 s 1 s
Measure-
Internal
ment
processing
Measure­ment
Internal processing
Measure­ment
Internal processing
Measure­ment
Internal processin
2
Slow function off
0.12 s 0.13 s 0.12 s 0.13 s 0.12 s 0.13 s 0.12 s 0.13 s
Measure-
Internal processing
ment

3.4.2 Exposure Level

• Values obtained in an exposure measurement are not measures of the risk a specific magnetic field poses.
• In Exposure Level mode, first order filters are used to approximate magnetic density flux levels in the ICNIRP guideline. Thus corner frequencies cause a difference of at most 3 dB, but this corresponds to the "smoothed edges" referred to in IEC62233 or EN50366.
• A discussion of acute health effects of exposure must acco unt for cou­pling factors such as the human body and magnetic fields, threshold values under basic restrictions and safety factors (reduction factor) for general public and occupational exposure.
Reference: In IEC62233, the coupling factor at 10 cm as described in ICNIRP is "0.14" for small and "0.16" for large ones while the safety factor (reduction factor) (the reduction factor for public exposure is "50" (1/ 50 of the threshold value) for threshold values for occupational expo sure is "10" (1/10 of the threshold value).
For details, see "Appendix" (P . 111).
Measure­ment
Internal processing
Measure­ment
1 s
Internal processing
Measure­ment
Internal processing
3
3 4 5 6 7 8 9 10 11
-
12 13
44
3.4 Measuri ng Exposure Level (General Public)
• In measuring non-uniform magnetic fields (see figure below) where the sensor may come into contact with a source, it is recommended that measurements be conducted at a distance from the three coils
and at an angle (indicated by ropy of each axis.
The IEC62233 and EN50366 standards do not mention sensor angle.
Reference
Uniform magnetic field
in the figure below) to maintain isot-
Non-uniform magnetic
field
Magnetic field s ensor
Magnetic
field
field
Magnetic
Magnetic field
Magnetic fiel d
Source
Measure at an angle away from the three coils
The locatio ns indicat ed by (8 points on t he spheri cal sens or surfac e) should be maintained in a vertical attitude relative to the DUT.
3.4 Measuring Exposure Level (General Public)

3.4.3 Measurement Procedure

Measure exposure level (General Public) according to the steps below
45
5
21
1. Press to open the exposure level (General Public)
mode screen.
2 3
3 4 5 6 7 8 9 10 11 12 13
46
accuracy range
3.4 Measuri ng Exposure Level (General Public)
2. Press to select meas urement range.
Range number
Ranges are switched in the order shown below .
AUTO r0 r1 r2
Range type Range number Range
Auto-range
Manual range r0 20%
appears on the
display
r1 200% r2 2000%
Automatically selects
appropriate range dur-
ing measurements.
Range of guaranteed accuracy
At resultant value R, a count of less than 100 is outside guaranteed accuracy. As a result, when R is 99% or less in range r2, it is not within guaran­teed accuracy . Use an appropriate range.
Guaranteed accuracy range for resultant value R
Exposure level (General Public, Occup.) Fixed range
10% 200%
1% 20%
Guaranteed
100% 2000%
346.4%
34.64%
Conditional guaranteed accuracy range (when none of x, y, z are O.L.)
3464%
r2 r1 r0
3.4 Measuring Exposure Level (General Public)
DUT
m
3. Place the magnetic field sensor near the D U T.
47
2 3
3
• For accurate measurements, the tester should be located further away from the DUT than the sensor.
• Although few magnetic materials are used in the magnetic field tester, the battery is a magnetic material and may affect measure ments if located close to the DUT. In this situation, it may be a good idea to use an extension cable to increase the distance between the sensor and the tester or to remove the battery and use the AC adapter.
• For details on measuring distance from DUT, sensor location and
operating conditions, refer to the following documents. (IEC62233 or EN50366)
• The center of the 3471 sensor is located at right
angles to the x, y and z axes. Since the dis­tance from the center to the spherical surface is 61 mm and the distance to the DUT is a mm, locate the tester so that the distance from the surface of the DUT to the center of the sensor is (a + 61 mm). Calculate the measurement dis tance from the surface of protrusions on the DUT.
-
4. Read the indicated value.
5. Press to save the measurement data.
4
-
5 6
a mm
61 m
7 8 9 10 11
This ends exposure (General Public) measurements.
12 13
48

3.5 Measuri ng Exposure Level (Occup.)

3.5 Measuring Exposure Level (Occup.)
Measuring Exposure Level (Occup.)
5
21
1. Press to open the exposure level (Occup.) mode screen.
3.5 Measuring Exposure Level (Occup.)
accuracy range
2. Press to select measure m ent range.
Range number
Ranges are switched in the order shown below.
AUTO r0 r1 r2
Range type Range number Range type
Auto-range
Manual range r0 20%
appears on the display
r1 200% r2 2000%
Automatically selects
appropriate range dur-
ing measurements.
49
2 3
3 4 5 6
Range of guaranteed accuracy
At resultant value R, a count of less than 100 is outside guaranteed accuracy. As a result, when R is 99% or less in range r2, it is not within guaran­teed accuracy. Use an appropriate range.
Guaranteed accuracy range for resultant value R
Exposure level (General Public, Occup.) Fixed range
10% 200%
1% 20%
Guaranteed
100% 2000%
346.4%
34.64%
Conditional guaranteed accuracy range (when none of x, y, z are O.L.)
3464%
r2 r1 r0
7 8 9 10 11 12 13
50
3.5 Measuri ng Exposure Level (Occup.)
3. Place the mag netic field se nsor near the DUT.
• For accurate measurements, the tester should be located further away from the DUT than the sensor.
• Although few magnetic materials are used in the magnetic field tester, the battery is a magnetic material and may affect measure ments if located close to the DUT. In this situation, it may be a good idea to use an extension cable to increase the distance between the sensor and the tester or to remove the battery and use the AC adapter.
-
4. Read the indicated value.
5. Press to save the m easurement data.
This ends exposure level (Occup.) measurements.
51

3.6 Holding Maximum Value of the Resultant RMS Values

3.6 Holding Maximum Value of the Resultant RMS Values
This function holds the maximum value of the resultant RMS value R and the values of each axis.
1, 2
2 3
3
1. Press to hold the maximum value of the resultant RMS
value R and the icon lights.
The update of the maximum value will occur as a bigger resultant RMS value is observed. Therefore it is possible that the value of x, y, or z axis may decrease.
2. Press again to turn off an d clear held maximum
data value. The screen returns t o the st ate i t had befor e was pres sed.
4 5 6 7 8 9 10 11 12 13
52

3.7 Checking and Deleting Saved Data

3.7 Checking and Deleting Saved Data

3.7.1 Checking Saved Data

Use the following steps to check saved measurement data. You can save up to 99 data.
3
2
1, 2
1. Press to open the scree n w it h th e s av ed data.
53
3.7 Checking and Deleting Saved Data
2. Press (S) or (T) to select the memory number of
measurement data you wish to check.
Memory number
3. Press to return to the meas urement mode screen.
This completes saving measurement data
blinks when measurement data is saved with the slow function
ON.
2 3
3 4 5 6 7 8 9 10 11 12 13
54
3.7 Checking and Deleting Saved Data

3.7.2 Deleting Saved Data

Use the following steps to delete saved measurement data.
3
2
1, 2
1. Press to open the scree n w it h th e s av ed data.
3.7 Checking and Deleting Saved Data
2. Press and hold for 2 seconds or more.
This deletes the measur ement data with the highest memory number. Data can deleted only when moved to the highest memory number.
Highest memory number
[dAtA] [cLr] appea r when data is deleted.
55
2 3
3 4 5 6
The data for the next highest m em ory number appears.
7 8 9 10 11 12 13
56
3.7 Checking and Deleting Saved Data
3. Press to return to th e m eas urement mode screen.
This completes deleting measurement data.

3.7.3 Deleting All Saved Data

Use the following steps to delete all saved measurement data.
When the power is off, hold down and press . This deletes all saved data.
[dAtA] [cLr] appea r when data is deleted.
3.7 Checking and Deleting Saved Data
57
2 3
3 4 5 6
Then the model name (3470), battery status and the measure­ment mode screen appear.
This completes deleting all saved data.
7 8 9 10 11 12 13
58
3.7 Checking and Deleting Saved Data

4.1 Checking Remaining Battery Power

Advanced Functions 4
4.1 Checking Remaining Battery Power
59
Use the procedure below to check remaining battery power.
Press .
The measurement mode screen appears after showing bat­tery status.
2 3 4
4 5 6 7 8 9
The screen indicates 100% battery power regardless of battery status when the 9445-02 AC Adapter is connected. (A reading of 100% may not appear when the USB cable is connected.)
10 11 12 13
60

4.2 Function for Saving Settings (Saving Set Modes and Measurement Ranges)

4.2 Function for Saving Settings (Saving Set Modes and Measurement Ranges)
A frequently used mode can be saved for immediate access. The saved mode appears as soon as the tester is powered up.
Follow the steps given below.
Select the mode or range y ou wish to access upon po wering up and hold down
[SAvE] [End] appear and the tester returns to the screen prior to saving settings.
for 2 seconds or more.
Approx. 2 s
61

4.3 Initializing Function (Returning Settings to Their Factory Defaults)

4.3 Initializing Function (Returning Settings to Their Factory Defaults)
This function initializes all settings returning them to their factory default s and deletes all saved data. Follow the steps given below.
When the power is off , h old dow n to press .
[All clr] appears.
After this it displays the model name (3470) and battery status, and is restarted using the factory default settings.
The factory default settings are listed in the table below.
Item Settings
Mode Magnetic flux density
Range AUTO Output None Auto power off Enabled Buzzer Enabled Axis XYZ Slow function OFF Unit T
(10 Hz to 400 kHz)
2 3 4
4 5 6 7 8 9 10 11 12 13
62

4.4 Other Functions

4.4 Other Functions
The following sections describe other functions that a user should know .

4.4.1 Setting Auto Power Off (Saving Battery Power)

• This funct ion saves bat tery power w hen the user forgets to tur n off the power.
• The tester is automatically turned off 10 m inutes a fter the last key press.
lights when th e au to p o we r o f f fu n ctio n is on. B efo re tu rni ng o ff th e tester, the icon flashes and sounds a buzzer tone (for about 30 s).
Pressing any key during this interval p ostpone s shutdo wn by 10 minute s.
Follow the steps given below.
1, 4
3
2
1, 2
1. When the power is off, hold down and pr ess .
4.4 Other Functions
2. Press (S) or (T) to turn on .
63
2 3
3. Press and select [y] or [n].
[y] : set auto power off. [n] : cancel auto power off.
y
n
4. Press to save the se tting.
[SAvE] [End] appear and the pow er is tu rned off.
4
4 5 6 7 8 9 10 11 12 13
64
4.4 Other Functions

4.4.2 Setting Audible Key Feedback

This function causes the keys to generate a beep when pressed. Follow the steps given below.
1, 4
3
2
1, 2
1. When the power is off, hold down and pr ess .
2. Press (S) or (T) to turn on [bEEP].
3. Press and select [y] or [n].
[y] : audible key feedback is on. [n] : audible key feedback is off.
y
65
4.4 Other Functions
2
n
3
4. Press to save the se tting.
[SAvE] [End] appear and the pow er is tu rned off.
4
4 5 6 7 8 9 10 11 12 13
66
4.4 Other Functions

4.4.3 Changing Unit Indication

Use this function to change the unit indication. (The default unit is T.) Follow the steps given below.
1, 4
3
2
1, 2
1. When the power is off, hold down and pr ess .
2. Press (S) or (T) to turn on [Unit].
4.4 Other Functions
3. Press and select [T], [G], or [A/m].
67
4. Press to save the se tting.
[SAvE] [End] appear and the pow er is tu rned off.
T, G, or A/m
2 3 4
4 5 6 7 8 9 10 11 12 13
68
4.4 Other Functions

4.4.4 Setting the Slow Function

This function performs 1 s sampling and then displays the measurement value.
Follow the steps given below.
1, 4
3
2
1, 2
1. When the power is off, hold down and pr ess .
2. Press (S) or (T) to turn on [SLou].
3. Press and select [y] or [n].
[y] : sets the slow function. [n] : cancels the slow function.
y
69
4.4 Other Functions
2
n
3
4. Press to save the se tting.
[SAvE] [End] appear and the pow er is tu rned off.
blinks while measuring with the slow function on.
4
4 5 6 7 8 9 10 11 12 13
70
4.4 Other Functions

4.4.5 Turning on the MEM. MODE key in MAX. HOLD or output mode

Use this function to turn on or of f the during MAX. HOLD or during output.
Checking saved data during output results in output off
Follow the steps given below.
1, 4
3
2
1, 2
1. When the power is off, hold down and press .
2. Press (S) or (T) to turn on [MEM.M].
3. Press and select [y] or [n].
71
4.4 Other Functions
[y] :
[n] :
enables in MAX. HOLD mode and output modes.
disables in MAX. HOLD mode and output modes.
y
n
4. Press to save the se tting.
[SAvE] [End] appear and the pow er is tu rned off.
2 3 4
4 5 6 7 8 9 10 11 12 13
72
4.4 Other Functions

5.1 Overview

Using Application Software 5
5.1 Overview
73
The application software supplied with the 3470 provides the following functions.
Functions Functions in detail See:
Graphic representation of measurement values
Transfer recorded data to a PC
Configure the 3470 from a PC
View application soft­ware version
The supplied PC driver and the ap plic ation software must be installed o n a PC to enable downloading data from the 3470 tester.
5.2 "Installing" (P.74)
Operating environment
Supported operating system
Hard disk capacity
Interface
USB hubs with a power consumption exceeding 100 mA must be self­power hubs.
CD configuration
Folder
File
Select "small font" for the operating system since "large font" may cause irregularities in screen indications.
Japanese English
HiUsbSgl.inf File for installing driver HiUsbSgl.sys 3470 USB device driver accuracy.dat Accuracy file
Analyzing measurement data on a PC5.3 (P. 85)
5.4 (P. 88)
Configuring the 3470 from a PC 5.5 (P. 91)
Checking version of application soft­ware
• Windows 98, Me, 2000, XP
• The hardware environment comprising CPU, memory, display, etc. must conform to operating system recommendations.
• 10 MB or more free disk space
• USB Ver. 1.1 or greater
• One 3470 unit can be connected at any one time
Folder containing the application configu­ration file
5.6 (P. 93)
2 3 4 5
5 6 7 8 9 10 11 12 13
74
1
2

5.2 Installing

5.2 Installing
The major installation procedures are listed below.
Install the driver.
5.2.1 "Installing the Driver" (P.74)
Install the PC software.
5.2.2 "Installing software" (P.82)

5.2.1 Installing the Driver

The driver installation procedures are described below.
Do not connect and disconnect the USB cable during operations. Dis­connecting the USB cable will prevent normal completion of the instal­lation procedure.
• Only one 3470 tester can be connected to one PC.
• You may be informed that a new device has been detected also when a 3470 tester with a different serial number is connected. Install the device driver according to the on-screen instructions.
1. Turn on the tester.
2. Use the supplied USB cable to connect the 3470 tester wit h
the PC where the driver is to be ins talled .
(PC operations) The 3470 tester is recognized as "HIOKI Single USB Products." After the "Found New Hardware" screen, the "Fou nd New Hardware Wiz-
ard" appears.
The procedures for "Found New Hardware Wizard" depends on the PC operating system.
75
5.2 Installing
Insert the CD according to the instructions that appear to install the driver. Installation procedures for each operating system are provided below.
"WindowsXP" (P.75) "Windows2000" (P.76) "WindowsMe" (P.78) "Windows98" (P.79)
WindowsXP
1. Insert the supplied C D in the CD-ROM drive.
2. Select "Install the software automatically" and click [Next].
3. Click [Continue].
The "Found New Hardware Wizard" appears.
2 3 4 5
5 6 7 8 9
Clicking this button starts file copying. When the operating system detects the software, a message noti fies the use r that this sof tware i s not recognized by Microsoft. Select [
Continue Anyway].
10 11 12 13
76
5.2 Installing
4. Click [Finish].
5. Remove the CD from the CD-ROM drive when the "Found
Windows2000
1. Click [Next] when the "Found New Hardware Wizard" appears.
New Hardware Wizard" screen closes.
To install further software, go to Section 5.2.2 "Installing software" (P.82)
2. Insert the supplied CD in the CD-ROM drive.
77
5.2 Installing
3. Select "Search for a suitable driver for my device" and click
[
Next].
2 3 4
4. Select "CD-ROM drives" (deselect other alternatives) and click
Next].
[
5
5 6
5. Click [Next].
7 8 9 10 11 12 13
78
5.2 Installing
6. Click [Finish].
7. Remove the CD from the CD-ROM drive when the "Found
WindowsMe
New Hardware Wizard" screen closes.
To install further software, go to Section 5.2.2 "Installing software" (P.82)
1. Insert the supplied CD in the CD-ROM drive.
The "Add New Hardware Wizard" appears.
2. Select "Automatic search for a better driver" and click [Next].
3. Click [Finish].
79
5.2 Installing
2 3
4. Remove the CD from the CD-ROM drive when the "Add New
Windows98
1. Click [Next] when the "Add New Hardware Wizard " appears.
Hardware Wizard" screen closes.
To install further software, go to Section 5.2.2 "Installing software" (P.82)
4 5
5 6 7 8 9 10 11 12 13
80
5.2 Installing
2. Select "Search for the b est driver for your device" and click
[
Next].
3. Insert the supplied CD in the CD-ROM drive.
4. Select "CD-ROM drive" (deselect other alternatives) and click
[
Next].
5. Click [Next].
Clicking this button starts file copying.
5.2 Installing
6. Click [Finish] in the nex t sc reen that appears.
81
2 3
7. Remove the CD from the CD-ROM drive when the "Add New
Hardware Wizard" screen closes.
To install further software, go to Section 5.2.2 "Installing software" (P.82)
4 5
5 6 7 8 9 10 11 12 13
82
5.2 Installing

5.2.2 Installing software

The software installation procedures are described below.
1. Close all currently active applic at ions on the PC.
2. Insert the CD (PC application software) supplied with the
instrument into the CD-ROM drive.
3. Open the folder [Eng lish] included on the CD, then execute
Setup.exe.
This will activate the installer and open the "DATA VIEWER for 3470
" window.
(EN)
4. Click [Next].
4.
83
5.2 Installing
5. Click [Change], then specify the installing destination folder.
Click [Next].
Click this button when specify the installing destination.
6. Click [Install].
2 3 4 5
5 6
7. When installati on ends, click [Finish] in the checking window.
Installation is now complete.
7 8 9 10 11 12 13
84
5.2 Installing
Uninstallation
1. Select [Setup] - [Control Panel] from the [Start] menu, then
2. Select "
3. Uninstall the program by following the procedure appearing on
Select [Program] - [HIOKI] - [DATA VIEWER for 3470] - [DATA VIEWER for
3470 (EN)
click [
[
Change/Remove].
the screen.
] from the [Start] menu of Windows to activate the application.
Add/Remove Programs].
DATA VIEWER for 3470 (EN)
," then click [
Remove
] or

5.3 Graphic Re presentation of Measurement Values

5.3 Graphic Representation of Measure­ment Values
Measurement values obtained using the 3470 tester can be imported to a PC for graphic representation.
85
1. Turn on the 3470 and select [HIOKI]-[DATA VIEW ER for 3470]-
[
DA TA VIEWER for 3470 (EN)] from the Start button on the PC.
2 3
2. Connect the tester to a PC using the supplied USB cable.
(Push the cable plugs all the way in.)
4 5
5
USB cable
6
on the 3470 lights an d all key operations o ther than
the are disabled.
• Only one 3470 tester can be connected to a PC at one time.
• The effective value logger can obtain up to 32000 data.
7 8 9 10 11 12 13
3.
7.
2
86
5.3 Graphic Representation of Measurement Values
3. Click the "Effective Value Logger" tab in the "HIOKI DATA
VIEWER for 3470" window to open the "Effective Value Logger" screen.
6.
Menu item Description
Fixed Interval Samples at fixed intervals to obtain 20 points per
Average Calculates the average of sampling data obtained in
Maximum Indicates data when resultant RMS value R is at its
Drop-down list box for
sampling interval
R/ x/ y/ z Turn R, x, y and z data indication On and Off.
File Read Represents saved data in graphic form and saves it
CSV Save Saves graph data.
Start Starts measurements. Stop Stops measurements. Print Prints graphs
division.
drawing 1 point. (Equation)
t
ˆ
x
ˆ
y
ˆ
z
ˆ
maximum among the sampling data obtained in drawing 1 point.
Selects sampling interval. When the slow function is set to ON, 1 s/div and 5 s/div are not available.
2
dtx
Tt
T
t
2
dty
Tt
T
t
2
dtz
Tt
T
22
ˆˆ
ˆ
zyxR ++=
N
11
2
==
==
==
ˆ
x
n
N
n
=
1
N
11
2
ˆ
y
n
N
n
=
1
N
11
2
ˆ
z
n
N
n
=
1
as graph data.
87
5.3 Graphic Re presentation of Measurement Values
4. Set sampling method, sampling interval and start measure-
ments.
Sampling method
Sampling interval
2
5. Click the mo use to mo ve the ver tical line s in the grap h and to
indicate the data for the corresponding time interval at the bot tom of the screen.
3
-
4 5
5 6 7
6. Click [Stop] to stop measuring. Then click [CSV Save] to save
the graph data.
Click [File Read] to read out saved dat a. Or click [Print] to print the graph.
8 9
7. Disconnect the USB cable from the tester and the PC and turn
off the tester.
10 11 12 13
88

5.4 Transf erring Recorded Data to a PC

5.4 Transferring Recorded Data to a PC
Use this function to transfer data from the 3470 tester to a PC. The data can be saved in the CSV format on the PC.
1. Turn on the 3470 and select [HIOKI]-[DATA VIEWER for 3470]-
DA TA VIEWER for 3470 (EN)] from the Start button on the PC.
[
2. Connect the tester to a PC using the supplied USB cable.
(Push the cable plugs all t he way in.)
USB cable
on the 3470 ligh ts and all key operation s other than
the are disabled.
• Only one 3470 tester can be connected to a PC at one time.
• The prescribed accuracy range is not considered in R+Accuracy and R-Accuracy*. *: R+Accuracy and R-Accuracy are the results of + and - accuracy
calculations of resultant RMS values and indicates that R is within this range. These values may be used in the standard.
89
5.4 Transferring Recorded Data to a PC
3. Click the "Entire Read" tab in the "HIOKI DATA VIEWER for
3470" window to open the "Entire Read" screen.
3.4. 6. 5.
2 3 4
Screen item Description
Read Transfers data from the 3470 tester to a
Memory Clear Deletes data recorded on the 3470.
CSV Save Saves displayed data.
File Read Displays saved data.
Unit Change drop-down list box Selects unit for magnetic flux density data.
Unit Change Changes unit for indicated magnetic flux
Displayed Value on the 3470
Rxyz /
Print Prints displayed data.
PC.
density data to unit selected in the Unit Change drop-down list box.
Selects Rxyz or value displayed on the
3470.
4. Click [Read] to transfer rec orded data to a PC.
Do not disconnect the USB cable during t ransfer or the tra nsfer will not end normally.
5
5 6 7 8 9 10 11 12 13
90
5.4 Transf erring Recorded Data to a PC
5. Then click [CSV Save] to save the displayed data.
Click [File Read] to read out saved data. Or click [Print] to print the dis­played data.
6. Use [Memory Clear] to delete recorded data on the 3470 that
is no longer needed.
7. Disconnect the USB cable from the tester and the PC and turn
off the tester.
5.5 Configuring the 3470
Use the instructions below to configure and store the settings that the 3470 use when powered up.
1. Turn on the 3470 and select [HIOKI]-[DATA VIEW ER for 3470]-
DA TA VIEWER for 3470 (EN)] from the Start button on the PC.
[
91

5.5 Configuring the 3470

2
2. Connect the tester to a PC using the supplied USB cable.
(Push the cable plugs all the way in.)
USB cable
on the 3470 lights an d all key operations o ther than
the are disabled.
Only one 3470 tester can be connected to a PC at one time.
3 4 5
5 6 7 8 9 10 11 12 13
92
5.5 Configuring the 3470
3. Click the "Configuration" tab in the "HIOKI DATA VIEWER for
3470" window to open the "Configuration" screen.
3.4. 5.6.
Screen item Description
Configure the 3470 Configures the 3470.
Initialize the 3470 Returns the 3470 to its factory defaults.
Setting Save Saves made settings.
Setting Re ad Reads saved settings.
4. Click [Configure the 3470] to set up the 3470 tester.
5. Then click [Setting Save] to save the displayed data.
Click [Setting Read] to read out saved data.
Be sure not to change the file extension as this will prevent normal reading of file data.
6. Use [Initialize the 3470] to return the 3470 settings to their fac-
tory defaults.
7. Disconnect the USB cable from the tester and the PC and turn
off the tester.

5.6 Viewing Version Information

5.6 Viewi ng Version Information
Use this function to view the version of the "HIOKI DATA VIEWER for 3470".
1. Turn on the 3470 and select [HIOKI]-[DATA VIEW ER for 3470]-
[
DA TA VIEWER for 3470 (EN)] from the Start button on the PC.
2. Connect the tester to a PC using the supplied USB cable.
(Push the cable plugs all the way in.)
USB cable
93
2 3 4 5
5
on the 3470 lights an d all key operations o ther than
the are disabled.
Only one 3470 tester can be connected to a PC at one time.
3. Click the "Version" tab in the "HIOKI DATA VIEWE R for 3470"
window to open the "
Version" screen.
6 7 8 9 10 11 12 13
94
5.6 Viewing Version Information
4. Disconnect the USB cable from the tester and the PC and turn
off the tester.
Loading...