Hioki 3454-11 Instruction Manual

Page 1
Introduction
3454-11
DIGITAL MΩ HiTESTER
Instruction Manua
January 2012 Revised edition 13
Printed in Japan
3454A981-13 12-01H
3.
2.
11.
10.
4. 5.
6.
1.
10.
9.
8.
7.
12.
12.
: Lights up when the measured value is held during
the resistance/insulation-resistance measurement.
: Lights up when 0Ω adjustment is made during
resistance measureme nt.
: Indicates that battery power is low. (during which
time accuracy cannot be guaranteed). : Lights up when auto power save is enabled. : Lights up when the comparat or fun ction is en abl ed.
Thank you for purchasing the HIOKI “3454-11 DIGITAL MΩ HiTESTER”. To obtain maximum performance from the instrument, please read this manual first, and keep it handy for future reference.
Initial Inspection
When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
Preliminary Checks
• Before using the instrument the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
• Before using the instrument, make sure that the insulation on the test leads and connection cords is undamaged and confirm that the white or red portion (insulation layer) inside the cable is not exposed. If a color inside the cable is exposed, do not use the cable. Using the product in such conditions could cause an electric shock, so contact your dealer or Hioki representative for replacements (Model L9787).
Maintenance and Service
• T o clean the instrument, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.
• If the instrument seems to be malfunctioning, confirm that the batteries are not discharged, and that the test leads and fuse are not open circuited before contacting your dealer or Hioki representative.
• When an indication Err.9 appears, send the instrument for repair.
Specifications
General Specifications
Guaranteed accuracy period
Display Max. 4000 LCD Sampling rate twice/second
Response time Operating Temperature
&Humidity Storage Temperature
&Humidity Operating Environment Effect of temperature
°C
0 to 18 (32 to 64°F, 82 to 104
Influence quantity
Degree of protection IP40 (condition which the test lead is connected) Power source Maximum rated power 3 VA
, 28 to 40
°F
)
1 year at
Insulation resistance ( 0, ∞ center value), resistance: Within 5 s, ACV: Within 2 s
0 to 40 (non-condensating)
-20 to 50 (non-condensating)
Indoors, altitude up to 2000 m (6562-ft.) ASL, Pollution Degree 2
±2% of reading ±5 dgt, plus basic allowance
°C
(Resistance Measurement 4 MΩ range: ±5% of reading plus basic allowance)
E E E (Influencing factor non-applicable for E4 to E10)
Rated power voltage: 1.5 V DC × 4, R6P manganese battery × 4 or LR6 alkaline battery × 4
23°C±5°C (73°F±9°F) and 90% RH
°C (32 - 104°F)
°C (-4 - 122°F)
(Position) : 0%
1
(Supply voltage) : Twice of the intrinsic uncertainty
2
(Temperature) : ±2%±5d
3
, 90%RH or lower
, 90%RH or lower
gt.
Continuous operating time
Possible number of measurements
Additional function
Dielectric strength Maximum input voltage
Maximum rated voltage to earth
Input error protection for 10 second (overvoltage protection)
Dimensions (excluding protrusions)
Mass 530 g, 18.7 oz. approx. (including batteries) Accessories
Options
Standards applying rdg. :reading value (The value currently being measured and indicated on the
measuring instrument)
dgt. :resolution (The smallest displayable unit, i.e., the input value that causes
the digital display to show a "1".)
Insulation Resistance Measurement
Rated output voltage (DC) Max. effective reading Center scale reading
250 V / 500 V 500 MΩ 10 MΩ
1000 V 4000 MΩ 100 MΩ
Rated output voltage Display range Max. reading Resolution
250 V / 500 V / 1000 V
250 V / 500 V 500 MΩ range 500 MΩ 1 MΩ
1000 V 4000 MΩ range 4000 MΩ 1 MΩ*
* Resistances of 1000 MΩ or higher are indicated in increments of 10 MΩ i.e.,
1010 MΩ, 1020 MΩ, etc .
Rated output
voltage
250 V / 500 V
1000 V
Operation uncertainty B: ±18% (1st effective measurement range) Measurement terminal voltage characteristic.
Rated output
voltage 250 V 1 to 1.2
500 V 0.5 MΩ
1000 V 1 MΩ
Effect of radiated radio-frequency electromagnetic field: 3 V/m, within +10%rdg.
Resistance Measurement
Display range Max. reading Resolution
Short circuit current: 200 mA or more, open circuit voltage: 5 VDC±1 V Operation uncertainty B: ±30% (0.2 to 400 Ω)
AC Voltage Measurement
Display range Max. reading Resolution
600 V range 750 V 1 V ±3%rdg. ±6dgt.
Input resistance: 100 kΩ or more, Frequency range: 50 to 60 Hz
Open circuit
voltage
times of nominal
test voltage
40 Ω range 40.00 Ω 0.01 Ω
400 Ω range 400.0 Ω 0.1 Ω
4 kΩ range 4.000 kΩ 0.001 kΩ
40 kΩ range 40.00 kΩ 0.01 kΩ
400 kΩ range 400.0 kΩ 0.1 kΩ
4 MΩ range 4.000 MΩ 0.001 MΩ
Insulation measurement at 250 V and 500 V for about 10 hours, at 1000 V for about 10 hours (with manganese battery)
250 V at 0.25 MΩ 300 times (R6P battery), 1600 times (LR6 battery) 500 V at 0.5 MΩ 250 times (R6P battery), 1400 times (LR6 battery) 1000 V at 1.0 MΩ 200 times (R6P battery), 1000 times (LR6 battery) Ω at 1.0 Ω 1000 times (R6P battery), 1000 times (LR6 battery)
Automatic power-saving mode, Comparator, High-volt­age warning, Warning indication of false voltage input, Data hold, Display lighting, 1000 V output error protec­tion, Zero adjustment
5550 V AC 50/60 Hz for one minute Between electric circuit and case
600 V AC, (Anticipated Transient Overvoltage: 6000 V)
600 V AC (ACV function: 800 V AC) (MΩ-1000 V function: 1200 V AC)
175W × 148H × 56D mm approx.
6.89"W × 5.83"H × 2.20"D approx.
L9787 Test Lead, Instruction Manual, R6P manganese battery × 4, Strap
L9787-91 Breaker Pin, 9804-02 Magnetic Adapter, L9787 Test Lead
Safety EN61010 EMC EN61326 EN 61557-1/-2/-4
400 MΩ range 400.0 MΩ 0.1 MΩ
Name of measurement
range
1st effective measurement range
2nd effective measurement range
other measurement range 1st effective
measurement range 2nd effective
measurement range other measurement range
resistance value to be maintained
Measurement Category III
4 MΩ range 4.000 MΩ 0.001 MΩ
40 MΩ range 40.00 MΩ 0.01 MΩ
Measurement range
0.200 to 50.0 MΩ ±3%rdg. ±4dgt.
50.1 to 500 MΩ 0 to 0.199 MΩ
0.200 to 200.0 MΩ ±3%rdg. ±4dgt.
200.1 to 4000 MΩ 0 to 0.199 MΩ
Lower limit measurement
rated output voltage
0.25 MΩ
Guaranteed at 600 V or less)
(
Rated
current
1 to 1.2 mA
Accuracy & Intrinsic
uncertainty A
(After 0 Ω Adjustment)
±3%rdg. ±6dgt.
±5%rdg. ±6dgt.
Accuracy
Accuracy & Intrins ic
uncertainty A
±5%rdg. ±5dgt.
±5%rdg. ±5dgt.
Short circuit
current
1.2 mA max.
Safety
This instrume nt is designed to co m ply with IEC 61010 Safety Stand a rd s , and has been thoroughly tested for safety prior to shipment. However, mishandling during use could result in injury or deat h, as well as damage to the instrument. Using the in s t ru m e nt in a way not desc ribe d in this manual may negate the provided safety features. Be certain that you understand the instructions and precautions in the manual before use. We disclaim any responsibility for accidents or inju­ries not resulting directly from instrume nt defect s.
Measurement Categories
This instrument complies with CAT III safety requirements. To ensure safe operation of measurement instruments, IEC 61010 establishes safety stan­dards for various electrical environments, categorized as CAT called measurement categories.
CAT II: Primary electrical circuits in equipment connected to an AC electri­cal outlet by a power cord ( portable tools, household app liances, etc.) CAT II co vers di rect ly measur ing el ect ri­cal outlet receptacles.
CAT III: Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets. CAT IV:The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel). Using a measurement instrument in an environment designated with a higher­numbered category than that for which the instrument is rated could result in a severe accident, and must be carefully avoided. Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV measurement applications could result in a severe accident, and must be care­fully avoided.
II
to CAT IV, and
Safety Symbol
This manual contains information and warnings essential for safe operation of the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions.
In the manual, the symbol indicates particularly important information that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with
the symbol) before using the relevant function. Indicates that dangerous voltage may be present at this terminal.
Indicates a double-insulated device. Indicates AC (Alternating Current).
Indicates DC (Direct Current).
The following symbols in this manual indicate the relative importance of cau­tions and warnings.
Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user. Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument Indicates advisory items related to performance or correct operation of the instrument.
Usage Notes
Follow these precautions to ensure safe operation and to obtain the full bene­fits of the various functions.
Observe the following precautions to avoid electric shock.
• Be sure to disconnect the test lead from the object to be measured and turn the function switch OFF before connecti ng or dis connecti ng the test lead from the MΩ HiTESTER.
• Always verify the appropriate setting of the function selector before connecting the test leads. Disconnect the test leads from the measurement object before switching the function selector.
• Do not use the instrument where it may be exposed to corrosive or com­bustible gases. The instrument may be damaged or cause an explosion.
• Do not allow the instrument to get wet, and do not take measurements with wet hands. This may cause an electric shock.
• Do not use any other electrical source other than the batteries. The use of any other sources may result in damage of the instrument or the object to be measured and also may cause electric shock.
• Before using the instrument, make sure that the insulation on the test leads and confirm that the white or red portion (insulation layer) inside the cable is not exposed. If a color inside the cable is exposed, do not use the cable. Using the product in such conditions could cause an electric shock, so contact your dealer or Hioki representative for replacements (Model L9787).
• Do not use the instrument where it may be exposed to oil, chemicals, or solvents. Contact wi th these substances may cause cracking in the instrument, re s ul ti ng in damage or electr ic s hock.
• If the protective functions of the instrument are damaged, either remove it from service or mark it clearly so that others do not use it inadvertently.
• This instrument is designed for use indoors. It can be operated at temper­atures between 0 and 40°C without degrading safety.
• Do not store or use the instrument where it could be exposed to direct sun­light, high temperature or humidity, or condensation. Under such condi­tions, the instrument may be damaged and insulation may deteriorate so that it no longer meets specifications.
• For safety reasons, when taking measurements, only use the L9787 Test Lead (or optional) provided with the instrument.
• To avoid damage to the instrument, protect it from physical shock when transporting and handling. Be especially careful to avoid physical shock from dropping.
• Calibration and repair of this the supervision of qualified technicians knowledgeable about the dangers involved.
• Removable sleeves are attached to the metal pins at the ends of the test leads. To prevent a short circui t accident, be sure to use the test lea ds with the sleeves attached when performing measur ements in the CAT III mea­surement category. Remove the sleeves from the test leads when perform­ing measurements in the CAT I and CAT II measurement categories. For details on measurement categories, see "Measurement categories" in the instruction manual.
• To avoid battery depletion, turn the function selector OFF after use (the Auto Power Save feature consumes a small amount of current).
• The safety sleeve is attached to the test lead plug. Remove the sleeve before connecting to the instrument.
instrument should be performed only under
Names and Functions of Parts
1. Function Selector: Selects among power ON/OFF, the output voltage for
insulation resistance measurement, ACV, or resistance (Ω).
2. MEASURE Key: Used to measure resistance and insulation resistance.
This key remains ON while it is held down.
3. LOCK Key: Used to measure resistance and insulation resistance. This key
switches ON if held down for more than 2 seconds. Press the key again to turn it OFF.
4. COMP Key: Used for the comparator function
5. LIGHT Key: Turns the display light ON/OFF. The light automatically
switches OFF after 30 seconds.
6. 0ΩADJ Key: Used for the zero-adjust function in resistance measurement
Used when "1000 V" is selected in insulation-resistance measurement Used to select the buzzer sound in the comparator function
7. High-voltage warning lamp: Begins flashing if the input voltage exceeds AC
70 V (±10 V) and during insulation resistance measurement.
8. EARTH Measurement Terminal: Connect the black test lead to this terminal.
9. LINE Measurement Terminal: Connect the red test lead to
this termina l.
10.Strap Hole: Pass the strap through this hole.
11. Test lead Housing: Houses the test leads. The test leads may be
housed without disconnecting them from the terminals after use.
12.Sleeve stand: Attach the sleeve removed from the tip of the
test lead.
Display Block
: Lights up when the measured value is less than the criterion for the comparator
function during insulation-resistance measurement, or when the measured value is greater than the criterion during resistance measurement.
Page 2
: Lights up when the measured value is equal to or greater than the criterion for
Source (primary side)
Always turn off the breaker or the measurement line.
Load (secondary side)
ON
OFF
Do not press the MEASURE key.
Always connect the test lead to the secondary side of the breaker
.
ON OFF
Source (primary side)
Load (secondary side)
Earth electrode
1. For safety, set the function selector to OFF and remove the test lead from the instrument.
2. Loosen the screw located at the center of the back of the instrument and remove the battery cover.
3. Replace all four batteries or fuse.
4. Reinstall the battery cover and
the comparator function during insulation-resistance measurement, or when the measured value is equal to or less than the criterion during resistance measure­ment.
: Lights up when the criterion for the comparator function is indicated. : Lights up when "1000 V" is selected, if the 0ΩADJ key has not been pressed.
: Lights up when the LOCK key is pressed to perform continuous measure-
: Indicated during insulatio n-r esistan ce mea sure me nt, or beg i ns flashi n g when the
input voltage exceeds AC 70 V (±10 V). : Overflow indicator. Indicated when the measurement exceeds the maximum indica-
ment of insulation resistance or resistance.
tion value.
Measurement Procedures
Pre-measurement inspection
Ensure that the test leads are not disconnected
1. Use the function selector to select Ω function.
2. Short the test lead tips.
3. Confirm that the reading is below 1 Ω when pressing the MEASURE key.
Insulation Resistance Measurement
To select 1000 V, set the function selector to 1000 V while press­ing the 0ΩADJ.
: Overflow indication
An abbreviation for overflow, the display OF is analogous to a indication in an analog insulation tester. When measurements are larger than the effective
maximum indicated values of each function, the display will indicate . [Measurement example] When the display indicates in the 1000 V function,
measurements are detected as larger than 4000 M to the test lead, is also displayed.
Display Function
250 V
500 V
1000 V 4000 M
To avoid electric shock, always verify the appropriate setting of the function selector before connecting the test leads.
• When measuring insulation resistance, dangerous voltage is applied to the measurement terminals. To avoid electric shock, do not touch the test lead.
• Never touch the object being measured immediately after measuring. There is a danger of electric shock from the change accumulating dur­ing high voltage testing. (See Discharge Function)
• Discharge the subject conductor after measurement.
• Do not attempt to measure insulation resistance on a live conductor. Doing so could damage the instrument or cause an accident that might result in injury or death. Always turn off power to the conductor being measured before starting
Preparing for Measurement
1. Set the function selector to 250 V, 500 V , or 1000 V. To select 1000 V, set the function selector to 1000 V while pressing the 0ΩADJ key. When the function switch is turned from OFF to 1000 V, hold down the 0ΩADJ key until "MΩ" appears on the display. If the 0ΩADJ key has not been pressed and "1000 V" is selected, appears on the display. In this case 1000 V is not applied when the MEASURE key or LOCK key is turned on.
2. Make sure that indicator does not appear . If the indicator appears, please replace the batteries.
3. Connect the black test lead to the measurement terminal on the earth side of the instrument. Connect the red test lead to the measurement terminal on the line side of instrument.
4. Connect the black test lead to the ground side of the object being measured. Except when measuring insulated resistance between ground and the object being measured, connect the black test lead to an optional point.
5. Connect the red test lead to the object being measured.
Measuring Only While the Key is Pressed
•Start
1. Press the MEASURE key. The high-voltage warning lamp begins flashing, and the indicator appears on the display.
Ω
. When nothing is connected
Effective maximum
indicated value
500 M
Ω
Ω
2. Read the measurement after it has stabilized.
•End
1. Release the MEASURE key to end measurement. The current measurement is automatically held.
2. When the object to be measured must be discharged, read and follow t he instructions given under "Discharge Function" below.
Measuring without Holding the Key (Continuous Measurement)
•Start
1. Hold down the LOCK key for more than 2 seconds. The high-voltage warning lamp begins flashing, and the indicator and appear on the display.
The MΩ HiTESTER continues with measurement even if the LOCK key or MEASURE key is not held down.
2. Read the measured value after it has stabilized.
• End (Shutdown method when is indicated)
1. Press the LOCK (STOP) key or MEASURE key to end measurement. The current measurement is automatically held.
2. When the object to be measured must be discharged, read and follow t he instructions given under "Discharge Function" below.
• If the object to which the test leads are connected is live, even if the MΩ
HiTESTER is not performing measurement, the high-voltage warning lamp and indicator will flash on the display.
• Insulation resistances are unstable by nature. The readings may not stabilize
with some objects; this does not necessarily indicate a malfunction.
• If the object has a capacitive component, the MΩ HiTESTER may indicate a
value smaller than the actual resistance immediately after the start of mea­surement. The reading will gradually increase to indicate the actual resis­tance.
• If the function switch is turned during measurement, the MΩ HiTESTER will
stop measurement.
• If is indicated even if the measurement is performed with the tip of the
test leads short-circuited, a conductor in the test leads may be broken.
• Make sure the high-voltage warning lamp off first, and then select the Resis-
tance Measurement functions or OFF.
• Some objects may require time un til the m easurem ent s are stable . (Appro x. 5 sec.)
Discharge Function
When measuring an insulation resistance that contains a capacitance element, a charge proportional to the measurement voltage accumulates, and if undis­charged could lead to an electric shock accident.
1. End measurement without disconnecting the test leads from the object.
2. The built-in discharge circuit automatically discharges the item.
3. During discharging, the high-voltage warning lamp and indicator flash. They will go out when the voltage falls below approximately 30 V.
4. Discharge time varies with capacity.
AC V oltage Measurement
• Test lead should only be connected to the secondary side of a breaker, so the breaker can prevent an accident if a short circuit occurs. Connections should never be made to the primary side of a breaker, because unrestricted current flow could cause a serious accident if a short circuit occurs .
• The maximum input voltage is 600 V AC. Attempting to measure volt­age in excess of the maximum input could destroy the instrument and result in personal inj ur y or de a th.
• To avoid electrical shock, be careful to avoid shorting live lines with the test lead.
1. Set the function selector to V.
2. Connect the test lead to the instrument's measurement terminal.
3. Connect the test lead to the circuit being measured and re ad the displayed value. Do not use the MEASURE key or LOCK key.
3. Connect the test lead to the object being measured.
Measuring Only While the Key is Pressed
•Start
1. Press the MEASURE key and read the displayed value.
•End
1. Release the MEASURE key to end measurement. The current measurement is automatically held.
Measuring without Holding the Key (Continuous Measurement)
•Start
1. Hold down the LOCK key for more than 2 seconds. The appear on the display.
The MΩ HiTESTER continues with measurement even if the LOCK key or MEASURE key is not held down.
2. Read the measured value.
• End (Shutdown method when is indicated)
1. Press the LOCK (STOP) key or MEASURE key to end measurement. The current measurement is automatically held.
0ΩAdjustment Function
T o indicate the resistance of the object itself, the zero-adjust function stores the resistances of the test leads and fuse and deducts these values from measure­ment values.
1. Set the function selector to Ω.
2. Connect the test lead to the instrument's measurement terminal.
3. Short-circuit the metal tips of the test leads.
4. To start measurement, press the MEASURE key, or hold down the LOCK key for more than 2 seconds.
5. Press the 0ΩADJ key. lights up, and the display indicates "0.00 Ω".
6. Connect the test lead to the object being measured.
7. Read the measured value.
• The indication can be zero-adjusted when the reading is 3 Ω or less. If the
0ΩADJ key is pressed when the reading is over 3 Ω, "Err.1" is displayed.
• If the test leads are short-circuited during resistance measurement, the mea-
surement current will exceed 200 mA, accelerating battery consumption. Perform the zero adjustment as quickly as possible and open the circuit as soon as the adjustment is complete.
• If is indicated even if the measurement is performed with the tip of the test
leads short-circuited, a conductor in the test leads or the fuse may be broken.
Example of measuring the earthing
conductor resistance
If an additional operating circuit is connected in parallel to the circuit under measurement, the measurement error may occur due to the effects of impedance of the circuit connected in parallel or transient currents.
Measure the earthing conductor resistance at Ω range. Please refer to the low resistance measurement for measuring method.
Comparator Function
The comparator function compares the measurement with a set criterion, indi­cates PASS or FAIL, and sounds the buzzer during resistance or insulation­resistance measurement.
Using Comparator
1. Set the function selector to 250 V, 500 V, or 1000 V or Ω.
2. Press the COMP key. The COMP indicator, REF indicator, criterion, and the condition for sounding the buzzer ( or ) appear on the display. The display changes to the criterion-setting screen. The criterion changes each time the COMP key is pressed. Press the key repeatedly until the criterion to be used is displayed. Press the 0ΩADJ key. This switches the display between and , allowing you to select the criterion for sounding the buzzer. For example, if you switch the display to , the buzzer sounds when the comparator result is .
Resistance Measurement
Never apply voltage to test lead when the Resistance Measurement functions is selected. Doing so may damage the instrument and result in personal injury. To avoid electrical accidents, remove power from the circuit before measuring.
Preparing for Measurement
1. Set the function selector to Ω.
2. Connect the test lead to the instrument's measurement terminal.
3. Press the MEASURE key or the LOCK key to start measurement. The REF indicator and the criterion go out. The display returns to the measurement display screen. The MΩ HiTESTER compares measurements and the criterion.
Result Indication
Insulation-resistance measurement: When the measurement is sma ller than the criterion, is indicated. When the measurement is equal to or greater than the criterion, is indicated. Resistance measurement: When the measurement is greater than the crite­rion, is indicated. When the measurement is equal to or smaller than the criterion, is indicated.
Select a criterion from the presets shown in the table below.
Function
250 V/500 V
1000 V
Ω
For example, when the function switch is turned to "250 V," every time the COMP key is pressed, the criterion changes as follows: 0.2 MΩ 0.4
MΩ ... 200 MΩ 0.2 MΩ ....
• When the criterion setting screen is shown (step 2 above), if the MΩ HiT­ESTER remains idle for 2 seconds, the REF indicator, the criterion, and
or are no longer displayed, and the display reverts to the previ­ous screen. However, the COMP indicator remains on, indicating that the comparator function is enabled.
• If power is turned OFF after the comparator function is enabled, the function is automatically enabled when the power is turned ON again. The last crite­rion setting made before the power is turned OFF is also held.
0.2/0.4/0.5/1/2/3/5/10/20/30/50/100/200 unit 1/2/3/5/10/20/30/50/100/200/500/1000/2000 unit
0.5/1/2/3/4/5/6/10/20/50/100/200/1k unit
Established reference value available [MΩ]
[MΩ]
[MΩ]
[Ω]
When Not Using Comparator
To disable the comparator function, hold down the seconds. The
Automatic Pow e r- Sa v i ng Mo d e
The instrument will automatically enter power-saving mode about 10 minutes following the last operation, and all displayed values disappear. When the power is turned ON, the auto power save function is automatically enabled (APS lights up).
COMP
indicator goes out, and the comparator function is disabled.
COMP
key for more than 2
To Switch from Power-Saving Mode
Set the function selector to OFF before returning to the original position.
Disabling the Auto Power Save Function
While holding down LIGHT key, turn the function switch to turn ON the power.
Measurement principles
1.Insulation Resistance Measurement
The insulation resistance of test object Rx is obtained by supplying a voltage V to the test object and measuring the current leaking from the test object and the voltage supplied using the formula (Voltage supplied, V) / (current leakage, l).
2.AC Voltage Measurement
This is obtained from converting the value of the current flowing from the voltage source through the instrument to a voltage value.
3.Low Resistance Measurement
The resistance of test object Rx is obtained by supplying a specific current l to the test object and measuring the voltage occurring between the test ter­minals using the formula (inter-terminal voltage, V) / (supplied current l).
Replacing of Batteries and Fuse
• To avoid electric shock, turn off the function switch and disconnect the test lead before replacing the batteries and fuse.
• After replacing the batteries or fuse, replace the cover and screws before using the instru ment.
• Do not mix old and new batte r ie s , or different types of batteries . Als o, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result.
• To avoid the possibility of explosion, do not short circuit, disassemble or incinerate batteries.
• Handle and dispose of batteries in accordance with local regulations.
• Replace the fuse only with one of the specified characteristics and voltage and current ratings. Using a non-specified fuse or shorting the fuse holder may cause a life-threatening hazard. Fuse type: 7012540 (made by SIBA Inc.), Rating 0.5 A/700 V AC, very fast-Acting.
To avoid corrosion from battery leakage, remove the batteries from the instru­ment if it is to be stored for a long time.