HIOKI 3293 User Manual

Field Measuring Instruments
CLAMP ON HiTESTER
3291
CLAMP ON LEAK HiTESTER
3293
Display
Flips!
Hold
Maximum
Value
Power ON
simply by
opening the jaw
Super­thin jaw for even the narrowest spaces
3293:
Leak to Load
(0.06mA to 1000A)
3291
: Wide Current Range
(0.06A to 1000A)
Measure Leak to Load with One Single Unit
10uA
Resolution
2
Super bright
backlight
Advanced
fi lter cuts off noise components
from harmonic and other sources
Measure for maximum
value simultaneously and
hold max. value on
bar graph
Double-lever jaw
mechanism gets into even the narrowest spaces
Advanced sensor
makes it safe enough to clamp on bare
conductors
Simply open jaws to turn the power on
Flip
the cover to rotate
the display and view measurements from any
angle
Specifi cations
3291
Load
3293
Leak to Load
Measurable conductor size
ø30mm max ø24mm max
Rectifi cation True RMS Maximum rated
voltage to earth
AC600Vrms AC300Vrms
Measurement Range
•Auto Range switching
Zero-suppress at 5 counts
30.00mA
300.0mA
6.000A
60.00A 60.00A
600.0A 600.0A 1000A 1000A
Accuracy
±1.5%rdg. ±5dgt. (50Hz, 60Hz,
fc=180
)
Filter OFF:
±2.0%rdg. ±5dgt. (45 to 66Hz)
±3.0%rdg. ±5dgt. (40 to 45Hz, 66 to 600Hz)
(Add±1%rdg. for readings of 0.270 to 0.540A on the 3293)
Accuracy Guaranteed Measurements
Minimum 5.40A Minimum 1.00mA
Crest Factor 2.0 max (1.5 max when current exceeds 600A)
Response Time
Max. 1.5s
(Refers to time required to achieve 90% of accuracy specifi cations)
Display Refresh Rate
Approx. 2 times/second
(Bar graph: Approx. 4 times/second)
Temperature Coeffi cient
Add (0.05 x accuracy specifi cations)/°C
(at 0°C to 40°C)
Effect of Conductor Position
within ±5.0%
within ±0.1%
(at or below 6A range)
within ±5.0%
(at or below 60A range)
Effect of External Magnetic Field
not defi ned
Max 7.5mA
(within a 400A/m
magnetic fi eld, at or below 6A range)
Other Functions
Display hold, maximum value hold
(hold bar graph position), reversible LCD display, backlight, auto power off, auto power on when jaws are open, battery level indication
Usage Environment
0°C to 40°C/32°F to 104°F, 80%rh max.
(no condensation)
Storage Environment
-10°C to 50°C/14°F to 122°F, 80%rh max.
(no condensation)
Standards
EN61010-2-032
Type A current sensor
CATIII 600V, CATIV 300V, EN61326
EN61010-2-032
Type A current sensor,
CATIII 300V, EN61326
Rated Voltage and maximum rated power
3V DC, 20mVA
Power Supply and max. continuous use
Button type lithium battery (CR2032) x 1,
Approx. 10 hours (continuous, no load)
Dimensions and mass
50W x 136H x 26D mm, 90g
2” x 5” x 1”, 3.2 oz
50W x 130H x 26D mm, 120g
2” x 5” x 1”, 4.2 oz
HEAD O FFICE :
81 Koizumi, Ue da, Nagano, 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail : os-co m@ hiok i.co.jp
HIOKI USA CORPORATION :
6 Corpo rate Dri ve, Cranbury, NJ 08512 USA TEL +1-609-40 9-9109 / FAX +1-609-409-9108 E-mail : hioki@ hiokiusa.com
DISTRIBUTED BY
All in format ion cor rect as of Apr. 30, 2007. All s pecifi cations a re subject to change wit hout notice. 3291E1-74M-0 0P Pri nted in Jap an
Shang hai Representat ive Of fi c e :
1904, Shanghai Tim es Square Of fi ce 93 Huaihai Zhong Road Shanghai, 200 021, P.R.China TEL +86 -21-6391-0090, 0092 FAX +86 -21-6391-0360 E-mail : info@hioki.cn
3291
CAT IV 300V CAT III 600V
3293
CAT III 300V
Loading...