Hioki 3290 Instruction Manual

Page 1
INSTRUCTION MANUAL
3290
CLAMP ON
AC/DC HiTESTER
Page 2
Page 3
i
Contents
Inspection ............................................................... 2
Safety Notes ...........................................................3
Usage Notes........................................................... 6
Organization of This Manual .................................10
Chapter 1 Overview 13
1.1 Product Overv iew ........................... ......13
1.2 Features ................................................13
1.3 Parts Names and Functions .................16
1.3.1 3290
CLAMP ON AC/DC HiTESTER ........ 16
1.3.2 9691, 9692, 9693
1.4 Flowchart Of Key Operations ................25
1.4.1 Current Measurements Mode .......... 25
1.4.2 Frequency Mode ..............................28
1.5 Modes ...................................................29
Chapter 2
Basic Measure ment Procedures 35
2.1 DC Measurement (DC A) ......................36
2.2 AC Measurement (AC A) ......................38
2.3 Peak Measurement In AC+DC A Mode 39
CLAMP ON AC/DC SENSOR .......... 24
Page 4
ii
Chapter 3 Measurem en t Proc edu re s 43
3.1 Preparations ..........................................43
3.2 Current Measurement ........................... 45
3.2.1 Measuring DC Current (DC A) ..........46
3.2.2 Measuring AC Current (AC A) ..........47
3.2.3 Measuring AC + DC Current (AC+DC A)
...........................................................49
3.2.4 Bar Graph .........................................50
3.2.5 Peak Measur ement ...........................51
3.2.6 Output Function
(Current Measu rem en t) ............... ..... .53
3.3 Frequency Measurement ......................57
3.3.1 Frequency Measurement ..................57
3.3.2 Output Function
(Frequency Measurement) ................58
3.4 Auto-Zero Adjustment Function .............60
3.5 Data Hold Function ................................61
3.6 Changing the Screen Update Rate .......61
3.6.1 SLOW Mode ......................... ...... ..... .61
3.6.2 FAST Mode .......................................62
3.7 Recording Function ...............................63
3.7.1 Measurement Reading ......................63
3.7.2 Display of Elapsed Time ...................64
3.7.3 Interrupting the Recording Function ..64
3.7.4 Clearing Recorded Data ...................64
3.7.5 Disabling the Recording Function .....65
3.8 Measurement Method
Switching Function ................................66
3.8.1 RMS Measurement ........................... 66
3.8.2 MEAN Measurement .........................66
3.9 Measurement Response Speed
Switching Function ................................67
Page 5
iii
3.9.1 NORMAL Mode ................................67
3.9.2 SLOW Mode ............................... ......67
3.9.3 FAST Mode ...................................... 67
3.10 Filter Function ....................................... 70
3.11 Ke y Loc k Fun cti on ................................71
3.12 Bar Graph Enlargement Function .........71
3.13 Battery Check Function ........................72
3.14 MON Output Filter Function ..................73
3.15 Auto Power Off Function .......................74
3.16 Setting the Save Function .....................75
3.17 Battery-Low Warning ............................76
3.18 Beep Tone ............................................77
Chapter 4 Batteries Replacement 79
Chapter 5 AC Adapter 81
Chapter 6 Attaching the Strap 83
Chapter 7 Specifications 85
7.1 Measurement Specifications .................85
7.2 General Specifications ..........................88
Chapter 8 Clamp Sensor Specifications 93
8.1 9691 CLAMP ON AC/DC SENSOR ......93
8.2 9692 CLAMP ON AC/DC SENSOR ......96
8.3 9693 CLAMP ON AC/DC SENSOR ......99
Page 6
iv
Chapter 9 Carrying Case 103
Chapter 10 Troubleshooting 105
Chapter 11 After - sale Servi ce 111
11.1 Cleaning ..............................................111
11.2 Service ................................................111
Chapter 12 Appe ndix 113
12.1 Combined Accuracy ............................113
12.1.1 Accuracy Tables .............................113
12.1.2 Method of Calculation .....................118
12.2 Use of Intermediate Cable ...................119
Page 7
1
Introduction
Thank you for purchasing the HIOKI "3290 CLAMP ON AC/DC HiTESTER". To obtain maximum performance from the product, please read this manual first, and keep it handy for future refer ence. This manual describes the operation of the 3290 when used with the 9691, 9692, or 9693 CLAMP ON AC/DC SENSOR.
Request
We have tried to bring this manual as close to perfection as we could achieve. If perchance you find any unclear portions, mistakes, omissions, or the like, we would be most obliged if you could please notify us of them via any HIOKI agent, or directly.
Page 8
2
Inspection
• When you receive the product, inspect it carefully to ensure that no damage occurred during ship­ping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.
• Before using the product the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representa­tive.
Accessories
Strap .........................................................................................................1
LR6 Manganese battery............................................................................4
Instruction Manual.....................................................................................1
Options
9400 CARRYING CASE 9094 OUTPUT CORD 9445-02 AC ADAPTER (SA10-0910N, SINO-AMERICAN) 9445-03 AC ADAPTER (EU) (SA10-0910G, SINO-AMERICAN) 9691 CLAMP ON AC/DC SENSOR 9692 CLAMP ON AC/DC SENSOR 9693 CLAMP ON AC/DC SENSOR
Page 9
3
Safety Notes
This manual contains information and warnings essential for safe operation of the product and for maintaining it in safe operating condition. Before using the product, be sure to carefully read the following safety notes.
This product is designed to conform to IEC 61010 Safety Standards, and has been thor­oughly tested for safety prior to shipment. However, mishandling during use could result in injury or death, as well as damage to the product. Be certain that you understand the instructions and precautions in the man­ual before use. We disclaim any responsibil­ity for accidents or injuries not resulting directly from product defects.
Safety Symbols
In the manual, the symbol indicates particu­larly important information that the user should read before using the product.
The symbol printed on the product indicates that the user should refer to a corresponding topic in the manual (marked with the symbol) before using the relevant function.
Indicates a double-insulated device.
Indicates DC (Direct Current). Indicates AC (Alternating Current).
Indicates both DC (Direct Current) and AC (Alter­nating Current).
Indicates that the instrument may be connected to or disconnected from a live circuit.
Page 10
4
The following symbols in this manual indicate the relative importance of cautions and warnings.
Indicates that incorrect operation presents an ex­treme hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a sig­nificant hazard that could result in serious injury or death to the user.
Indicates that incorrect operation presents a pos­sibility of injury to t he user or damage to the prod­uct.
Advisory items related to performance or correct operation of the product.
Other Symbols
Indicates the prohibited action
Accuracy
We define measurement tolerances in terms of f.s. (full scale), rdg. (reading) and dgt. (digit) values, with the following meanings:
(maximum display value or scale length)
f.s.
rdg.
The maximum displayable value or the full length of the scale. This is usually the maximum value of the currently selected range.
(reading or displayed value) The value currently being measured and indicated on the measuring product.
(resolution)
dgt.
The smallest displayable unit on a digital measur­ing product, i.e., the input value that causes the digital display to show a "1".
Page 11
5
Measurement categories (Overvoltage categories)
This product complies with CATIII (600 V), CATII (1000 V) safety requirements. To ensure safe operation of measurement products, I 61010 establishes safety standards for various electrical environments, categorized as CAT I to CAT IV, and called measurement categories. These are defined as follows.
EC
CAT I
CAT II
CAT III
CAT IV
Secondary electrical circuits connected to an AC electri­cal outlet through a transformer or similar device.
Primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.)
Primary electrical circuits of heavy equipment (fixed in­stallations) connected directly to the distribution panel, and feeders from the distribution panel to outlets.
The circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel).
Higher-numbered categories correspond to electrical environments with greater momentary energy. So a measurement device designed for CAT III environments can endure greater momentary energy than a device designed for CAT II. Using a measurement product in an environment designated with a higher-numbered category than that for which the product is rated could result in a severe accident, and must be careful ly avo ided . Never use a CAT I measuring iproduct in CAT II, III, or IV environments. The measurement categories comply with the Overvoltage Categories of the IEC60664 Standards
Page 12
6
Usage Notes
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
To avoid short circuits and potentially life-
threatening hazards, never attach the product to a circuit that operates at more than the 600V, or over bare conductors.
Clamp sensor should only be connected to
the secondary side of a breaker, so the breaker can prevent an accident if a short circuit occurs. Connections should never be made to the primary side of a breaker, because unrestricted curr ent flo w could cause a serious accident if a short circuit occurs.
Use only the specified Model 9445-02 AC
ADAPTER (SA10-0910N, SINO-AMERI­CAN) or 9445-03 (for EU) AC ADAPTER (SA10-0910G, SINO-AMERICAN). AC adapter input voltage range is 100 to 240 VAC (with ±10% stability ) at 50/60 Hz. To avoid electr ica l h az ards an d da ma ge to the product, do not apply voltage outside of this range.
To avoid electric shock, do not allow the
product to get wet, and do not use it when your hands are wet.
Do not use the product where it may be
exposed to corrosive or combustible gases. The product may be damaged or cause an explosion.
Page 13
7
The top part of the clamp sensor above the barrier (including the clamps, but not the lever) is provided with double insulation to ensure safety. Be careful not to drop the sensor or othe rwis e s ub jec t i t to im p act. A damag ed sens or may result in el ectric shock during measurement. In case of sensor damage, contact us immediately for repa ir or disc ard the dam ag ed se nso r to avoid su bse qu ent use .
To avoid damage to the product, do not exceed the maximum input current rating, which depends on the frequency of the current being measured (see Chapter 8 Clamp Sensor Specifications - Frequency­dependent del eti on cha r ac teri sti cs) Be careful about the evolution of heat, when the input frequ enc y is high.
To avoid electri c s ho ck when rep lac in g th e batteries, first disconnect the clamp from the object to be measu red.
After replacing the batteries, replace the cover and screws before usi ng the product.
Do not mix old and new batteries, or different types of batteries. Also, be careful to observe battery polarity during installation. Otherwise, poor performance or damage from battery leakage could result.
To avoid the possibility of explosion, do not short circuit, disassemble or incinerate batteries.
Handle and dispose of batteries in accordance with local regulations.
Page 14
8
Do not use the product when the battery is depleted ( lights on the LCD). When
is on, accuracy and other specifications cannot be guaranteed.
For the inside memory prot ection, make sure the power is turned off before plugging in or unplugging the AC adapter.
Adjust ments and repairs should be made only by technically qualified personnel.
If the protective functions of the product are damaged, either remove it from service or mark it clearly so that others do not use it inadvert­ently.
To avoid corrosion from batter y leakage, remo ve the batteries from the product if it is to be stored for a long time.
Page 15
9
Observe the following to avoid damage to the
-
st
-
n
n
product.
Installation and Operating Environment Between 0°C and 40°C; 80% RH or less; indoors only.
Direct
sunlight
High temper ature, high humidity, du
Electromag
etic radiatio
Do not store or use the product where it could be exposed to direct sunlight, high temperature or humidity, or condensation. Under such condi­tions, the product may be damaged and insula­tion may deteriorate so that it no longer meets specifications.
This product is not designed to be entirely water- or dust-proof. To avoid damage, do not use it in a wet or dusty environment.
Do not use the product near a device that gener­ates a strong electromagnetic field or electro­static charge, as these may cause erroneous measurements.
To avoid damage to the product, protect it from vibration or shock during transport and handling, and be especially careful to avoid dropping.
Impact,
dropping
• Accurate measurement may be impossible in the
• To clean the product, wipe it gently with a soft
presence of strong magnetic fields, such as near transformers and high-current conductors, or in the presence of strong electromagnetic fields such as near radio transmitters.
cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether , ketones, thinners or gasoline, as they can deform and discolor the case.
Page 16
10
Organization of This Manual
Chapter 1 Product Ov erview
Provides an overview of the product and describes the parts and functions.
Chapter 2 Basic Measurement Procedure
Describes the basic measurement procedure for quick reference and the peak measurement procedure in DC current (DC A) mode, AC current (AC A) mode, and AC+DC A mode.
Chapter 3 Measurement Procedure
Describes the measurement procedure for the 3290 CLAMP ON AC/DC HiTESTER when combined with the 9691, 9692, or 9693 CLAMP ON AC/DC SENSOR.
Chapter 4 Battery Replacement
Explains how to replace the battery used for the 3290 Clamp On AC/DC HiTESTER.
Chapter 5 AC Adapter
Explains how to use the 9445-02 AC Adapter (optional).
Chapter 6 Attaching the Strap
Explains how to attach the strap.
Chapter 7 Specifications
Lists the specifications of the 3290 CLAMP ON AC/DC HiTESTER.
Chapter 8 Clamp Sensor Specifications
Lists the specifications of the 9691, 9692, and 9693 CLAMP ON AC/DC SENSORs.
Chapter 9 Carrying Case
Explains how to use the 9400 CARRYING CASE (optional).
Page 17
11
Chapter 10 Troubleshooting
Lists common symptoms that may indicate a malfunction. If you encounter a problem, be sure to read this chapter before requesting repair.
Chapter 11 Service
Describes our customer support service.
Chapter 12 Appendix
• Provides the accuracy table and explains how to
calculate accuracy for the 3290 CLAMP ON AC/ DC HiTESTER when combined with the 9691, 9692, or 9693 CLAMP ON AC/DC SENSOR.
• Describes how to use the intermediate cable.
Page 18
12
Page 19
13
Chapter 1 Overview
Overview Chapter 1
1.1 Product Overview
The 3290 CLAMP ON AC/DC HiTESTER is used with the 9691, 9692, or 9693 CLAMP ON AC/DC SENSOR. These sensors are interchangeable and any model can be used with the 3290. The 3290 automatically detects the sensor connected and sets up the appropriate range. When combined with the 9691, 9692, or 9693 Sensor, the 3290 HiTESTER can perform DC, AC, and AC+DC measurement of a live power line. Various functions are available (using the SHIFT key) to make the 3290 suitable for various current measurement applications. The 3290 is also equipped with two output terminals to enable a variety of advanced measurements, such as simultaneous output of both monitor and rms record. Moreover, the 3290 can operate using battery power or the AC adapter (for long-term measurement use).
1.2 Features
The 3290 has the following features:
Multif unc tion Microcom puter
The 3290 both compact and multifunctional thanks to the built-in microcomputer and use of the SHIFT key.
Large-Screen LCD
The large-screen LCD employed is easy to read.
Clamp Sensor
The 3290 is compatible with three clamp-on AC/DC sensors (9691, 9692, and 9693) having different ratings. Because the 3290 automatically detects the sensor connected, you need not edit the setting each time you change the sensor.
Page 20
Chapter 1 Overview
14
Multif unc tion Microcom puter
The 3290 both compact and multifunctional thanks to the built-in microcomputer and use of the SHIFT key.
Large-Screen LCD
The large-screen LCD employed is easy to read.
Clamp Sensor
The 3290 is compatible with three clamp-on AC/DC sensors (9691, 9692, and 9693) having different ratings. Because the 3290 automatically detects the sensor connected, you need not edit the setting each time you change the sensor.
Two Output Terminals
Two signals are output simultaneously from among the four following: current (REC = record and MON = monitor), current frequency (REC = record), and low battery warning signal.
• Current monitor & low battery warning signal
• Current monitor & current record
• Current record & low battery warning signal
• DC component & AC component (AC/DC sepa­rate output)
• Frequency record & current record
Method of measurement
In AC A mode, you can select either the true rms (RMS) or average rectified rms indication (MEAN).
Measurement Response Speed
For RMS measurement, you can select a response speed from FAST, NORMAL, or SLOW.
Low-frequency Current Measurement
In AC+DC A mode, you can measu re curr en t from 1 Hz when the measurement response speed is set to SLOW.
Filter Function
This function turns on and off a low-pass filter (LPF) of approximately 500 Hz.
Setting Save Function
This function saves the current measurement conditions so that settings are retained even after power is turned OFF.
Key Lock Function
This locks all keys including the POWER key (except the keys used to unlock).
Page 21
15
Chapter 1 Overview
AC+DC Measurement
The 3290 measures AC superimposed on DC, half-wave rectified, and full-wave rectified.
Peak Measurement
The peak hold current can also be measured.
REC Function
You can select whether the maximum or minimum measurement value is displayed.
Dual-power supply
The 3290 run s on either battery power or an AC power source.
Dustproof Cap
The dustproof cap is provided for use in a dusty environment.
Intermediate Cable
In case the 9691, 9692, or 9693 sensor cable (about 2 m long) is too short, an interme diate cable can be used to extend connection length. Insert the intermediate cable between the 3290 and the sensor cable.
Page 22
Chapter 1 Overview
16
5,
1.3 Parts Names and Functions
1.3.1 3290 CLAMP ON AC/DC HiTESTER
21
24
22
23
27
20
14
6, 11
7,15
8,16 9,17
10,18
1
28
19
4, 13
3, 12
2
Page 23
17
26
25
Chapter 1 Overview
Keys for Basic Measurement
1. POWER Key
• Turns power ON/OFF.
• Holding down the CLEAR key and turning power on initializes the setting save function and restores the default settings.
• Holding down the HOLD key and turning power on disables the auto power-off function.
• Holding down the PEAK/Hz key and turning power on turns off the buzzer.
2. SHIFT Key
• Pressing this key causes to light, indicating that the additional function (in blue letters) allo-
cated to each key is available. To turn off, press the SHIFT key again.
• Some keys may be locked when is on.
Page 24
Chapter 1 Overview
18
3. MODE Key
• Selects the current measurement mode.
4. RANGE Key
• Selects auto range or manual range.
• Selects a measurement range when manual range is selected.
• The current ranges for the 9691 and 9692 are 20 A and 200 A; the ranges for the 9693 are 200 A and 2000 A. The frequency ranges are 10 Hz, 100 Hz, and 1000 Hz.
5. OUTPUT Key
• Sets the output (for current and frequency mea­surement). Two outputs are available. The settings avail­able for each current measurement mode are fixed.
AUTO
AC ADC A
L
M (Hz only)
AC+DC A
H
• Disables the auto power-off function.
6. CLEAR Key
• Clears data to “0” during peak measurement.
• Clears data to “0” during activation of the record­ing function.
• Holding down the CLEAR key and turning power on initializes the setting save function and restores the default settings.
7. HOLD Key
• Suspends or deactivates the screen-updating function. (See 15 under “Keys for Advanced Measurement” for details of the key lock func­tion.)
• Holding down the HOLD key and turning power on disables the auto power-off function.
Page 25
19
8. PEAK/Hz Key
• PEAK performs peak measurement of wave­forms (peak hold).
• Hz performs frequency measurement (in AC A mode).
• Holding down the PEAK/Hz key and turning power on turns off the buzzer.
9. MAX/MIN Key
• Activates the recording function and displays the maximum value (MAX), minimum value (MIN), mean value of both maximum & minimum values (AVE), and elapsed time (TIME).
• Pressing this key immediately activates the recording function.
• MAX displays the maximum value of measure­ments taken after the recording function is acti­vated.
Chapter 1 Overview
• MIN displays the minimum value of measure­ments taken after the recording function is acti­vated.
• AVE displays the mean value of the maximum and minimum values measured after the record­ing function is activated.
• TIME displays the time elapsed after the record­ing function is activated.
• The auto power-off function is deactivated.
10. SAMPL Key
• Selects a screen-updating rate from NORMAL (twice/sec.), SLOW (once/3 sec. ), or FAST (4 times/sec.). Each time this key is pressed, the rate changes in the order above.
MINMAX
TIMEAVE
Measurement value
• The rate is usually set to NORMAL. (Note that
• SLOW decelerates the speed of updating on-
• FAST accelerates the speed of updating on-
there is no indication for NORMAL.) screen measurement (with SLOW on). screen measurement (with FAST on).
Page 26
Chapter 1 Overview
20
• During RMS measurement in AC mode or when in AC+DC mode (with RMS on), the measure­ment response speed changes according to the screen-updating rate (SLOW, FAST, or NOR­MAL).
• Y ou can only change the measurement response speed. For details, see 18 under “Keys for Advanced Measurement.”
NORMAL
SLOW
FAST
11. 0 ADJ Key (SHIFT CLEAR Keys)
• Performs auto zero adjustment in DC A or AC+DC A mode.
Keys for Advanced Measurement
12. RMS/MEAN Key (SHIFT MODE Keys)
• Selects the true rms measurement (RMS) or average rectified rms indication (MEAN) in AC A mode. RMS provides accurate measurement of distorted waveforms. MEAN only provides accu­rate measurement of undistorted sine waves.
13. B.GRAPH Key (SHIFT RANGE Keys)
• Enlarges the range of a bar graph.
14. MON.FL Key (SHIFT OUTPUT Keys)
• Turns on a low-pass filter of approx. 1 Hz when MON is selected for OUT1 in DC A mode.
• Turns on a low-pass filter of approx. 1 Hz when MON is selected for OUT1 (REC for OUT2) in AC+DC A mode. This enables separate output of the DC component (OUT1) and AC compo­nent (OUT2). (The value of the AC component is indicated on the LCD.)
Page 27
21
Chapter 1 Overview
15. LOCK Key (SHIFT HOLD Keys)
• Locks all keys including the POWER key.
• To unlock, press the same keys again (SHIFT HOLD keys).
16. FILTER Key (SHIFT PEAK/Hz Keys)
• Turns on a low-pass filter of approx. 500 Hz in AC A or AC+DC A mode .
• To turn the filter off, press the same keys again (SHIFT PEAK/H z key s).
17. B.CHECK Key (SHIFT MAX/MIN Keys)
• Indicates remaining battery power. When power is depleted to 0%, the low battery warning (  )
goes on.
18. RESPONSE Key (SHIFT SAMPL Keys)
Display
19. LCD
• Selects a measurement response speed inde­pendently from the screen-updating rate when RMS is selected in AC A mode or when AC+DC mode is selected.
• Press the SAMPL key to return to the measure­ment response speed that matches the selected screen-updating rate.
The letter of LCD is clearest, when it sees from the front side.
Page 28
Chapter 1 Overview
22
  
Direct Current (DC) Alternating Current (AC) Alternating Current and Direct Current (AC+DC)
Auto zero adjustment is active. Low battery war ning
OUT1 Setting of OUT1 OUT2 Setting of OUT2
Data hold function
MON MON.FL
Monitor is active. 1 Hz filter for monitor is active.
REC Record is active.
Auto power-off is active.
AUTO Auto-range L Current/frequency low range M Frequency mid range H Current/frequency high range FAST Screen-updating 4 times/sec. SLOW Screen updating approx. once/3 sec. FILTER 500 Hz filter is active. MAX Maximum value MIN Mini mum value AVE Average value TIME Elapsed time MEAN Average rectified rms indication RMS True root mean squar e value PEAK Wave peak value Hz Frequency
Key lock is active.P
A Current hour 100 hours/segment (bar graph)

Input over (bar graph) SHIFT
Page 29
23
Chapter 1 Overview
Parts
20. Sensor Connector (SENSOR)
Connect the 9691, 9692, or 9693 CLAMP ON AC/DC SENSOR to this connector.
21. AC Adapter Connection Terminal
Connect the 9445-02 AC ADAPTER (optional) to this terminal when not using the battery or for long-term measurement.
22. Output Terminal 1 (OUT1)
Connect the 9094 OUTPUT CORD (optional) to this terminal to obtain output during current or frequency measurement.
23. Output Terminal 2 (OUT2)
Connect the 9094 OUTPUT CORD (optional) to obtain output during current measurement or at detection of the low battery warning signal.
24. Dustproof Caps (4 pcs.)
Used to protect the terminals in a dusty environment.
25. Strap
Used to hang the 3290 around neck so as not to drop it.
26. Top Case
Cover the 3290 with this case when not in use to protect the LCD, keys and terminals.
27. Top Case Clasp
Insert the clasp into the top case to attach the case. Move the top case to the rear during measurement.
28. Battery Cover
Unfasten the screw, then remove the case to replace the battery.
Page 30
Chapter 1 Overview
24
1.3.2 9691, 9692, 9693 CLAMP ON AC/DC SENSOR
9691
9692
9693
Output connector
Page 31
25
Chapter 1 Overview
R
M
1.4 Flowchart Of Key Operations
1.4.1 Current Measurements Mode
MODE key
AC ADC A
ANGE key
AC+DC A
PEAK/Hz key
AX/MIN key
AUTO range
L range
H range
  DC A
  PEAK
  MAX
AUTO range
L range
H range
  AC A
  PEAK
   Hz
  MAX
AUTO range
L range
H range
AC+DC A
  PEAK
  MAX
   MIN
   AVE
   TIME
Measured
value
   MIN
   AVE
   TIME
Measured
value
   MIN
   AV E
   TIME
Measured
value
Page 32
26
S
R
MODE key
Chapter 1 Overview
DC A
AC A
AC+DC A
OUTPUT key
AMPL key
MS/MEAN key
OUT1 OUT2
OFF OFF
MON B.Lo
NORMAL
  SLOW
  FAST
Disabled
OUT1 OUT2
OFF OFF
REC B.Lo
MON REC
NORMAL
  SLOW
  FAST
OUT1 OUT2
OFF OFF
REC B.Lo
MON REC
NORMAL
  SLOW
  FAST
Disabled
  RMS
  MEAN
Page 33
27
Chapter 1 Overview
MODE key
MON.FL key
FILTER key
RESPONSE キー
  MON
  MON.FL
Disabled
Disabled
AC ADC A
Disabled
  OFF
   ON
NORMAL
(No displayed)
AC+DC A
  MON
  MON.FL
  OFF
   ON
NORMAL
(No displayed)
  SLOW ( )
  FAST ( )
Effective on RMS
  SLOW ( )
  FAST ( )
Page 34
Chapter 1 Overview
28
1.4.2 Frequency Mode
PEAK/Hz key
RANGE key
MAX/MIN key
PEAK
AUTO range
L range
H range
  MAX
   MIN
Hz
(AC A only)
AUTO range
10.00 Hz(L) range
100.0 Hz(M) range
1000 Hz(H) range
  MAX
   MIN
Current Measured value
OUTPUT key
   AV E
   TIME
Measured
value
OUT1 OUT2
OFF OFF
MON B.Lo
   AVE
   TIME
Measured
value
OUT1 OUT2
OFF OFF
REC (Hz) REC(A)
Page 35
29
1.5 Modes
For current, three modes are provided: DC (direct current, ), AC (alternating current, ), and AC+DC (alternating current and direct current, ) modes. Select a proper
mode according to the waveform shown below:
Chapter 1 Overview
0
0
0
0 V
0
0
0
0
0
0
0
0
0
0 V
0
0
0
0
0
0
0
0
0
0
0
0
Page 36
30
Mode
Chapter 1 Overview
Input
Display
waveform
OK
(with polarity)
OUTPUT
(only for current mode)
REC MON
DC
()
AC
No
Not meas urab le
No
Not measurable
No
Not meas urab le
(zero displayed)
OK
No
Not measurable
OK (without polarity)
Setting
impossible
AC+DC
OK
OK
Page 37
31
Chapter 1 Overview
-
-
-
-
-
-
-
-
-
-
Quick Reference
Measurements Operations
DC A
AC A RMS
AC A MEAN
AC+DC A
DC A peak
PEAK
ADJ
RMS
MEAN
ADJ
ADJ
Measure ment
Measure ment
Measure ment
Measure ment
Measure ment
(pa g e)
44
45
46
47
49
AC A RMS peak
AC A MEAN peak
AC+DC A peak
AC A RMS Hz
AC A MEAN Hz
RMS
MEAN
PEAK
RMS
MEAN
PEAK
PEAK
ADJ
Hz
Hz
Measure
49
ment
49
Measure ment
49
Measure ment
55
Measure ment
55
Measure ment
Page 38
Chapter 1 Overview
32
/peak/
H
/peak/
H
W i
/peak/
H
/peak/Hz
/peak/
H
Quick Reference (application)
Operations
(page)
Hold the measured value (Data Hold Function)
//
z/record
Measurement
HOLD
59
"O.L." is displayed (Overflow Warning Function)
//
z/record
hen the measured value exceeds the max-
mum value, "O.L." is displayed.
Change the screen update rate (Changing the Screen Update Rate)
//
(NORMAL/SLOW/FAST) switch
59
z
Measurement speed can be changed (Measurement Response Speed Switching Function)
/
Outputs (Output Function)
//
z
(NORMAL/SLOW/FAST) switch
AC A mode: RMS
AC+DC A mode only
OUT1: OUT2: on
65
51 56
Page 39
33
Chapter 1 Overview
O O
D
/peak
/
/peak/
H
/peak/
H
/peak/
H
Operations
(page)
Obtained through a low-pass filter (LPF) set to approximately 1 Hz (MON Output Filter Function)
/
UT1
MON
UT2
REC
isable:
OUT1:  MON.FL OUT2:  REC
Press the key again.
71
Obtained through a low-pass filter (LPF) to measurement value and output. (Filter Function)
/
FILTER : on
68
500 Hz の Hz
Disable:
Press the key again.
Disabling the auto power off function (Auto Power Off Function)
//
z
+
Power ON
APS : O FF
72
Disabling the beep tone (Beep Tone)
//
z
+
Power ON
75
Hold the maximum measurement, minimum measurement, and mean value. (Recording Function)
//
z/record
Bar graph The elapsed time is displayed.
MAX: maximum value /MIN: minimum value /AVE: average value /TIME: elapsed time
(MAX/MIN/AVE/ TIME/ Instantaneous value switch
61
Disable:
Press the key.
Page 40
Chapter 1 Overview
34
/peak/
H
D
" " display
/Hz
/peak/
H
/peak/
H
on
Operations
(page)
Saved the settings (Setting the Save Function)
//
73
z
(Hold down 2 seconds. )
isable:
+
Power ON
"" ""    display
Enlarged to the Bar graph. (Bar Graph Enlargement Function)
//
Disable:
Bar Graph Enlargement
Press the key again.
69
Battery check
//
z/record
Key lock
//
z
Disable:
70
69
Press the key again.
Page 41
35
Chapter 2 Basic Measurement Procedures
Basic Measurement
Procedures
To avoid short circuits and potentially life­threatening hazards, never attach the product to a circuit that operates at more than the 600V, or over bare conductors.
Clamp sensor should only be connected to the secondary side of a breaker, so the breaker can prevent an accident if a short circuit occurs. Connections should never be made to the primar y side of a breaker, because unre stri ct ed current flow cou ld cause a serious accident if a short circuit occurs.
Use only the specified Model 9445-02 AC
Chapter 2
ADAPTER (SA10-0910N, SINO-AMERI­CAN) or 9445-03 (for EU) AC ADAPTER (SA10-0910G, SINO-AMERICAN). AC adapter input voltage range is 100 to 240 VAC (with ±10% stability ) at 50/6 0 H z. To avoid electri ca l ha za rds and damage t o the product, do not apply voltage outside of this range.
This chapter describes t he basic meas urement procedure. For details of the measurement proce­dure, see Chapter 3, “Measurement Procedure.”
Page 42
Chapter 2 Basic Measurement Procedures
36
2.1 DC Measurement (DC A)
1. Press the MODE key to
display DC A (  ).
2. Press the SHIFT key to
display .
Page 43
37
Chapter 2 Basic Measurement Procedures
.
3. Press the 0 ADJ key
(CLEAR key) to perform zero adjustment.
4. will appear on the
LCD.
9691 CLAMP ON AC/DC SENSOR
Load
Measured conductor
Power source
5. Align the current direction
indicator on the clamp with the current directio n of the conductor to be measured, then clamp the conduct or so that it passes through the center of the clamp core. The figure shows the 9691 CLAMP ON AC/DC SENSOR. When using the 9692 or 9693 CLAMP ON AC/DC SENSOR, note that the current direction indicator is located in a different position.
6. Read the displayed value
Since the measurem en t range is set to AUTO, you need not set the range.
Page 44
Chapter 2 Basic Measurement Procedures
38
.
2.2 AC Measurement (AC A)
1. Press the MODE key to
display AC A ( * When measuring
current up to 10 Hz, use AC+DC A mode and select SLOW.
 ).
9691 CLAMP ON AC/DC SENSOR
Measured conductor
2. Clamp the conductor to
be measured so that it passes through the center of the clamp core. You need not align the direction of the clamp.
3. Read the displayed value
Since measurem ent range is set to AUTO, you need not set the range.
Page 45
39
Chapter 2 Basic Measurement Procedures
2.3 Peak Measurement In AC+DC A Mode
1. Press the MODE key to
display AC+DC A (  ).
2. Press the PEAK/Hz to
display PEAK.
Page 46
Chapter 2 Basic Measurement Procedures
40
3. Press the SHIFT key to
display .
9691 CLAMP ON AC/DC SENSOR
Measured conductor
4. Press the 0 ADJ key
(CLEAR key) to perform zero adjustment.
5. will appear on the
LCD.
6. Clamp the conductor to
be measured so that it passes through the center of the clamp core. You need not align the direction of the clamp.
Page 47
41
Chapter 2 Basic Measurement Procedures
.
7. Press the CLEAR key (0
ADJ key) to clear the
data.
8. Read the displayed value
Since the measurem en t range is set to AUTO, you need not set the range.
Page 48
Chapter 2 Basic Measurement Procedures
42
Page 49
43
Chapter 3 Measurement Procedures
Measurement Procedures
Do not connect or disconnect the output con­nector of the clamp-on sensor while power is supplied to the 3290 or the sensor is clamping a conductor to be measured.
When disconnecting the output connector of the clamp-on sensor from the 3290 sensor connec­tor , be sure to hold it by the metal part and pull it upward. If it is pulled by the cable, the lock will not be released and a broken wire in the cable may result.
The maximum continuous-input limit is obtained from the temperature increase due to self-heat­ing during measurement. To prevent damage to
Chapter 3
the clamp-on sensor, do not input a current exceeding this limit.
The max imum continuous-input limit varies depending on the clamp-on sensor and the fre­quency of the current to be measured. Please see the dilating-characteristics graph by fre­quency in "Chapter 8 Clamp Sensor Specifica­tions".
3.1 Preparations
1. Remove the battery cover and insert the bat-
tery (see Chapter 4 Battery Replacement).
2. Connect the output connector of the clamp-on
sensor to the sensor connector. The type of clamp-on sensor will be detected automati­cally.
Page 50
Chapter 3 Measurement Procedures
44
3. Press the POWER key to turn on the power. Make sure all segments of the LCD light up.
Then, the model No. (3290) is displayed and the charge remaining in the battery is shown by a bar graph.
New battery
50% battery charge The battery charge is 0.
lights up. The beep tone sounds three times.
remaining
4. The 3290 is ready for operation in DC-current measurement mode.
• If the setting save function has been used (if the
HOLD key is held down for approximately 2 sec­onds, the current setting is saved), the 3290 will be in the saved measurement mode.
• To cancel the saved setting, turn off the power
and then turn it on again while holding down the CLEAR key. The 3290 will return to the default setting (DC current).
Low-Battery-Voltage Detection Function
If the battery voltage falls when is on, the power will be turned off. On the LCD, “ ” and “ ” will appear.
Connecting/Disconnecting the Clamp On Sen­sor
When connecting the clamp-on sensor to the 3290, hold it by the area above the metal part of the output connector when inserting it. If it is held by the metal part, it won’t be possible to insert the output connector into the sensor connector. Make sure the wide terminal of the connec tor is at the top when the connector is inserted. When disconnecting the clamp-on sensor from the 3290, hold the metal part of the connector when pulling it out. If i t is held by the area above the metal part, it won’t be possible to disconnect the connector.
Page 51
45
Chapter 3 Measurement Procedures
Wide termina l
Connecting
Disconnecting
3.2 Current Measurement
The ends of the clamp sensor are very delicate.
Be careful not to drop the sensor or subject it to impact. If the sensor is deformed or the contact surfaces of the clamps are damaged, the mea­surement may not be accurate.
• Be sure to read the instruction manual for the
clamp sensor to be used before starting mea­surement.
Output connector
• The hall element is used for the detector of the
• Accurate measurement may be impossible in the
clamp-on sensor. The hall element tends to drift with age or due to the ambient temperature. Keep this fact in mind when performing mea­surement continuously.
presence of strong magnetic fields, such as near transformers and high-current conductors, or in the presence of strong electromagnetic fields such as near radio transmitters.
Page 52
Chapter 3 Measurement Procedures
46
• The reading may show a measurement greater
than the actual value due to magnetic-field inter­ference. The amount of interference varies depending on the sensor. For detail s, see “Exter­nal magnetic-field interference” in Chapter 8 Clamp Sensor Specifications.
• The 9691 has a lock mechanism for the clamps
so that the clamps will not open due to vibration or other similar causes during measurement. Lock the clamps as necessary. (If the clamps should open even slightly during measurement, the accuracy cannot be guaranteed.)
• Attach the clamp around only one conductor.
Single-phase (2-wire) or three-phase (3-wire) cables clamped together will not produce any reading.
OK
3.2.1 Measuring DC Current (DC A)
1. Press the MODE key to display “ .”
2. Select a measurement range suitable for the current to be measured.
3. Before clamping the conductor to be mea­sured, press the 0ADJ key (SHIFT CLEAR key) with the clamp sensor fully closed to per­form the auto zero adjustment (see 3.4 Auto­Zero-Adjustment Function). Upon completion
of the auto zero adjustment, “ ” will light up. (When auto range is selected, the auto zero adjustment is made for the two ranges.)
Page 53
47
Chapter 3 Measurement Procedures
The auto zero adjustment range is ± 450 dgt. for the L range and ± 45 dgt. for the H range. If the sensor’s offset exceeds these ranges, the auto zero adjustment cannot be performed.
4. Open the clamps of the sensor. Orient the cur-
rent direction indicator on the clamps in the current direction of the conductor to be mea­sured, place the conductor in the center of the clamp core, and clamp it.
Load
Power source
Current direction indicator
• In DC A mode, only a DC current that does not
contain an AC component can be measured cor­rectly (see 1.5 Modes).
• When a conductor to be measured is placed in
the center of the clamp core, measurement is performed the most accurately, with no effect of the conductor position.
• The measurement may change if the direction of
the clamp sensor is changed. However, the mea­surements are within the guaranteed accuracy range.
• If the reading fluctuates rapidly and is difficult to
read, press the SAMPL key to select SLOW.
• The L range displays up to 2500 dgt.
3.2.2 Measuring AC Current (AC A)
1. Press the MODE key to display “ .”
2. Select a measurement range suitable for the
current to be measured.
Page 54
Chapter 3 Measurement Procedures
48
3. Open the clamps of the sensor, place the con­ductor to be measured in the center of the clamp core, and clamp it. In addition, make sure the clamp sensor is perpendicular to the conductor.
• When measuring a current of 10 Hz or less, use
AC+DC A mode. Press the SAMPL key to set the screen-updating rate to SLOW (the mea­surement response speed must be SLOW; the unit symbol “A” will blink slowly). If SLOW is selected in AC A mode, the measurements may become smaller than those in AC+DC A mode.
• DC, full-wave rectified waveforms, half-wave
rectified waveforms, and DC+AC waveforms cannot be measured in AC A mode (see 1.5 Modes).
• When a conductor to be measured is placed in
the center of the clamp core, measurement is performed the most accurately without effect of the conductor position. If the clamp sensor is not perpendicular to the conductor, there may be a slight effect on the measurement when the mea­sured current frequency is high.
• We recommend that RMS measurement be per-
formed. By switching over to MEAN (SHIFT MODE key), the MEAN value can be compared with the RMS value and a check can be con­ducted for distortion in the measured current. If the values are nearly the same, the current is close to a sine wave with little distortion (the ratio of the peak to rms is close to 1.4:1).
• When the MEAN and RMS measurements differ ,
the measured current may be distorted.
• During MEAN measurement, the measurement
response speed cannot be changed.
• Select RMS measurement and the FAST
screen-updating rate if the rise in the current waveform is sharp (the measurement response speed must be FA ST; the unit symbol “A” will blink quickly).
• The L range displays up to 2500 dgt.
Page 55
49
Chapter 3 Measurement Procedures
3.2.3 Measuring AC + DC Current (AC+DC A)
1. 1. Press the MODE key to display “  .”
2. 2. Select a measurement range suitable for
the current to be measured.
3. 3. Before clamping the conductor to be mea-
sured, press the 0ADJ key (SHIFT CLEA R key) with the clamp sensor fully closed to per­form the auto zero adjustment. Upon comple-
tion of the auto zero adjustment, “ ” will light up (see 3.4 Auto-Zero-Adjustment Func­tion).
4. 4. Open the clamps of the sensor, place the
conductor to be measured in the center of the clamp core, and clamp it. In addition, make sure the clamp sensor is perpendicular to the conductor.
• When measuring a current of 10 Hz or less,
press the SAMPL key to set the screen updating rate to SLOW (the measurement response speed must be SLOW; the unit symbol “A” will blink slowly).
• When a conductor to be measured is placed in
the center of the clamp core, measurement is performed the most accurately with no effect of the conductor position. If the clamp sensor is not perpendicular to the conductor, there may be a slight effect on the measurement when the mea­sured current frequency is high.
• RMS measurement is performed.
• When DC measurement is performed, the polar-
ity will not be displayed. The measurement may change if the direction of the clamp sensor is changed. However, the measurements are within the guaranteed accuracy range.
• Set the screen-updating rate to FAST if the
increase in the current waveform is sharp (the measurement response speed must be FAST; the unit symbol “A” will blink quickly).
• The L range displays up to 2500 dgt.
Page 56
Chapter 3 Measurement Procedures
50
3.2.4 Bar Graph
The bar graph shows the current measurement range and the measurement. The blinking segment represents the current measurement range. When AUTO range is selected, it represents the range for internal operation. For example, if the segment at “2” is blinking and the number to the right is “x10,” the range is 20 A. In DC A mode, the negative polarity is shown to the left of “0” and the positive polarity is shown to the right. Therefore, two segments will blink. The measurement is represented by the segments that light up. For instance, when the 20-A range is selected in AC A mode, if the segments up to 1 (11 segments) are on, the measurement is 10 A or more but below 11 A.
• Because the segment at the number that repre-
sents the range blinks, if a measurement is 1900 dgt. or more but below 2100 dgt. (1800 dgt. or more but below 2200 dgt. in DC A mode), the bar graph will not show any change.
• If a measurement is 2100 dgt. or more (2200 dgt.
or more in DC A mode), “ ” will appear.
• The numerical reading may differ from the mea-
surement on the bar graph (e.g., “ ” appears even if the numerical reading is 2099 dgt.).
• If the reading shows O.L., “ ” will appear.
• When a measurement is unstable, the segments
that represent the measurement may blink.
Page 57
51
Chapter 3 Measurement Procedures
When the 9691 clamp on the AC/DC sensor is connected and the H range is selected, the indication will differ from that when ot her sensors are used as below.
• Because the sensor rating is 100 A, the segment
at “1” blinks and the number to the right of the bar graph shows “x100,” indicating the 100-A range.
• If a measurement is between 900 dgt. and 1050
dgt., the bar graph will not show any change. (Normally, one segment lights up for every 100 dgt..)
• If a measurement exceeds 1050 dgt., the seg-
ment 12th onward will light up and “” will appear (“ ” will appear in DC A mode).
3.2.5 Peak Measurement
1. Press the MODE key and select a measure-
ment mode for the circuit to be measured.
2. The measurement mode is changed as shown
below each time the PEAK/Hz key is pressed. Sele ct PEAK.
PEAK
3. In DC A or AC+DC A mode, perform the auto
zero adjustment by pressing the 0ADJ key (SHIFT CLEAR key). (See 3.4 Auto-Zero­Adjustment Function.)
4. 4. Select a measurement range. (When AUTO
range is selected, some time may be required for the 3290 to detect an appropriate range and switch over to the range. This may delay the peak measurement. If it is not possible to estimate the peak current value, use the H range.)
 Hz (AC only)
Measurement
Page 58
Chapter 3 Measurement Procedures
52
5. Before conducting the measurement, press the CLEAR key to clear (reset) the previous data. In DC A or AC+DC A mode, when there is no input, pressing the 0ADJ key (SHIFT CLEAR key) performs the auto zero adjust­ment and clears the data at the same time.
6. Open the clamps of the sensor, place the con­ductor to be measured in the center of the clamp core, and clamp it.
• When measuring a current of 10 Hz or less, use
AC+DC A mode (it is not necessary to select SLOW when only peak measurement is per­formed).
• Be careful to prevent the power from being
turned off by the auto-power-off function (see
3.15 Auto-Power-Off Function). If the measure­ment exceeds the auto-power-off time, disable the auto power off.
• When peak measurement is performed, the
polarity will not be displayed. When the direction of the clamp sensor is changed, the measure­ment may change within the guaranteed accu­racy range.
• Clear the data using the CLEAR key, if neces-
sary, after clamping the conductor.
• There is no difference in peak measurement
between RMS and MEAN in AC A mode.
• The output function is not available for peak
measurements. If the OUTPUT key is pressed in peak measurement mode, the current measure­ment will be output.
• To check the transitional peak values, press the
MAX/MIN key to display a measurement (no indicator). (The largest value will be displayed each time the screen is updated.)
• The bar graph is not displayed in peak measure-
ment mode.
• During peak measurement, the peak detection
circuit performs analog processing (no digital sampling is performed).
MAX
MIN
AVE TIME
Measure ment (No indicator)
Page 59
53
Chapter 3 Measurement Procedures
3.2.6 Output Function (Current Measurement)
Fully insert the 9094 output cord (optional) into the output terminal.
To prevent malfunction, be careful not to short-circuit the output terminal or apply a voltage to it.
There are two output terminals, and different signals may be output from these terminals. The current measurement is output at the rate printed on the battery cover of the
3290. The battery-low warning signal is output at approximately
DC 3 V until the warning indicator ( ) appears. After has appeared, it will be reduced to approx. 0 V (see 3.17 Battery-Low Warning) .
• Output 1 (OUT1) is selectable between REC (numerical record output) and MON (waveform output) during current measurement (MON only in DC A mode, see 1.5 Modes). During frequency measurement, REC (numerical record output) can be selected.
• Output 2 (OUT2) is selectable between REC (numerical record output) and B.Lo (battery-low warning signal) during current measurement (B.Lo only in DC A mode). During frequency measure­ment, REC (numerical record output) can be selected.
1. Press the MODE key to select a measurement
mode suitable for the circuit to be measured.
2. Press the RANGE key to fix the measurement
range.
3. Press OUTPUT key , and OUT1 and OUT2 will
light up, indicating that the output function is enabled. This automatically disables the auto
power off (“ ” goes out ).
Page 60
Chapter 3 Measurement Procedures
54
4. 4. Press the OUTPUT key to edit the output setting.
DC A Mode
OUT1
OUT2
OUT1
OUT2
MON
B.Lo ( OFF
OFF
OFF ( OFF)
AC A Mode and AC+DC A Mode
REC
B.Lo ( OFF)
MON
REC ( OFF)
OFF
OFF ( OFF)
5. Determine the range setting from both the measurement range of the 3290 and that of the instrument used, such as a recorder. See the conversion tables below.
Instrument
range /DIV
200 A range 1 A 2 A 5 A 10 A 20 A 50 A 100 A 200 A
20 A range 0.1 A 0.2 A 0.5 A 1 A 2 A 5 A 10 A 20 A
10 mV 20 mV 50 mV 0.1 V 0.2 V 0.5 V 1 V 2V
*The figures above represent the currents per DIV of the measuring instr um ent.
Instrument
range /DIV
2000 A range 10 A 20 A 50 A 100 A 200 A 500 A 1000 A 2000 A
200 A range 1 A 2 A 5 A 10 A 20 A 50 A 100 A 200 A
10 mV 20 mV 50 mV 0.1 V 0.2 V 0.5 V 1 V 2V
*The figures above represent the currents per DIV of the measuring instr um ent.
Page 61
55
Chapter 3 Measurement Procedures
Clamp-On Sensor + 3290 MON Output
Frequency Characteristics
5 0
-5
-10
-15
GAIN (dB)
-20
-25
-30 1
10010
1k
10k
Input frequency (Hz)
100k
• Make sure measuring instruments connected to the 3290 have input impedance over 1 M. If the impedance is low, the readings will be affected.
• When using the output function, be sure to press the OUTPUT key, and make sure OUT1 is on. Data is output even if OUT1 is not shown on the LCD. However, the auto power off is valid, and the power will be turned off approximately 10 minutes later.
• If the auto zero adjustment is not performed in DC A or AC+DC A mode, errors may occur in the reading and output.
• When measuring a current of 10 Hz or less, use AC+DC A mode. Press the SAMPL key to set the screen updating rate to SLOW (the measure­ment response speed must be SLOW; the unit symbol “A” will blink slowly). If SLOW is selected in AC A mode, the measurements may become smaller than those in AC+DC A mode.
Page 62
Chapter 3 Measurement Procedures
56
• The guaranteed output accuracy range is that in which the current measurements are not O.L. in the L range (MON output may saturate if it exceeds 4 V in the L range). In the H range, the accuracy range is determined by the maximum peak current of the clamp-on sensor.
• There is no difference in MON output between RMS and MEAN in AC A mode.
• If the OUTPUT key is pressed in AUTO range, the measurement range is set to the range in effect when the key is pressed (AUTO goes out).
• The REC output is analog. During RMS mea­surement, the measurement response speed changes in accordance with the screen updating rate. The measurement response speed may be changed separately by pressing the RESPONSE
key (SHIFT SAMPL key ). (See 3.9 Measure­ment-Response-Speed Switching Function.)
• During MEAN measurement, the measurement response speed will not change when the screen updating rate is changed.
• While current measurements are being output, peak measurement or frequency measurement can be performed, and the recording function or data hold function can be used (if the screen updating rate, measurement mode, or measure­ment range is changed, or auto zero adjustment is performed, the output will fluctuate).
• If the HOLD key is pressed during output, the output value will not be held.
• If the MODE key is pressed to switch over to the current measurement mode while frequency measurement records are being output, the 3290 continues to output the frequency records. To output current measurements, press the OUTPUT key to cancel the frequency output and edit the output setting.
• Use the 9445-02 or 9445-03 (EU) AC adapter (optional) for long-term recording.
• When the AC adapter is used, if the commercial power line contains a large amount of noise, the display may show several counts or noise may be present in the output. In such a case, connect the ground terminal or the L terminal of the recorder to the ground.
Page 63
57
Chapter 3 Measurement Procedures
3.3 Frequency Measurement
3.3.1 Frequency Measurement
1. Press the MODE key to display “ .”
2. When the current range of the circuit to be measured is known, set the current range to the manual range (if the current range is unknown, perform current measurement and then set the range).
3. Each time the PEAK/Hz key is pressed, the indication will change as shown below.
PEAK
  Hz (AC only)
Measure­ment
4. Select a frequency range suitable for the fre­quency to be measured.
5. Open the clamps of the sensor , and clamp the conductor to be measured.
• When the 3290 is running on the battery, if fre-
quency measurement is performed with the cur­rent showing O.L., the current consumption will be increased and the battery life will be short­ened.
• If the input is too smal l for the range (5% or less),
the measurement may not be accurate. The reading may show “----“ or O.L., or it may fluctu­ate.
• The reading will show “----“ if the input is below
10 Hz in the 100-Hz or 1000-Hz range, and below 1 Hz in the 10-Hz range.
• If the input exceeds 1 kHz, the reading shows
• Some time may be required for the reading to
O.L.. stabilize, depending on the frequency range and
input frequency.
Page 64
Chapter 3 Measurement Procedures
58
• The 10-Hz range or 100-Hz range will display up to 125% of each range.
• Pressing the MAX/MIN key will not affect the out­put.
• The bar graph shows the current measurement.
• Zero cross detection of waveforms is used for frequency measurement. In the case of special waveforms, such as those of inverters, the fre­quencies may not be measured despite the fact that a low-pass filter is used (when the carrier frequencies are as low as several kHz, the car­rier component cannot be removed and zero cross detection therefore cannot be performed correctly, thereby increasing the reading).
• If there is a DC waveform including 0 V, zero cross detection cannot be performed correctly and the reading may be increased or become unstable.
3.3.2 Output Function (Frequency Measurement)
Fully insert the 9094 output cord (optional) into the output terminal.
To prevent malfunction, be careful not to short-circuit the output terminal or apply a voltage to it.
An output of DC 1 V is produced for 1000 counts on the full scale of the frequency range. The output is synchronized with the screen updating rate. (The output waveform will be in a stepped shape due to D/A outputs.)
1. Edit the settings while referring to 3.3.1 Fre­quency Measurement.
2. Press the OUTPUT key; OUT1 and OUT2 will light up, indicating that the output function is enabled. OUT1 outputs the frequency numeri­cal record, and OUT2 outputs the current numerical record.
Page 65
59
Chapter 3 Measurement Procedures
3. The auto-power-off function will be automati-
cally disabled ( goes out).
OUT1
OUT2
REC (Hz)
REC (A) ( OFF)
OFF
OFF ( ON)
4. Determine the range setting from both the
measurement range of the 3290 and that of the instrument used, such as a recorder.
Instrument
range /DIV
1000 Hz range 10 Hz 20 Hz 50 Hz 100 Hz 200 Hz 500 Hz 1000 Hz
100 Hz range 1 Hz 2 Hz 5 Hz 10 Hz 20 Hz 50 Hz 100 Hz
10 Hz range 0.1 Hz 0.2 Hz 0.5 Hz 1 Hz 2 Hz 5 Hz 10 Hz
*The figures above represent the frequencies per DIV of the measuri ng ins trument.
10 mV 20 mV 50 mV 0.1 V 0.2 V 0.5 V 1 V
• Make sure measuring instruments connected to
the 3290 have input impedance over 1 M.
• When using the output function, be sure to press
the OUTPUT key and confirm that OUT1 is on. When OUT1 is off, the output will not reflect the value for frequency measurement.
• When the 3290 is running off battery power, per-
forming frequency measurements with the cur­rent showing O.L. will increase current consumption and reduce battery life.
• If the OUTPUT key is pressed in the AUTO
range, the frequency range is set to the range in effect at the time the key is pressed (AUT O goes out).
• If the HOLD key is pressed, the output value will
also be held.
• If the measurement mode is changed to fre-
quency measurement mode while current mea­surement records are being output, the 3290 will continue to output current records. To output fre­quency measurements, press the OUTPUT key to cancel current output and edit the output set­ting.
Page 66
Chapter 3 Measurement Procedures
60
• When “----“ is displayed, the output is 0 V. When
O.L. is displayed, the output is approx. 1.4 V.
• Use the 9445-02 or 9445-03 (EU) AC adapter
(optional) for long-term recording.
• When the AC adapter is used, if the commercial
power line contains a large amount of noise, the display may show several counts or noise may be present in the output. In such a case, connect the ground terminal or the L terminal of the recorder to the ground.
3.4 Auto-Zero Adjustment Function
The auto-zero adjustment function automatically adjusts the offset in the internal circuit that results from temperature characteristics or clamp-sensor magnetization. This function is used for measurement in DC A mode and AC+DC A mode. The clamp core is magnetized when a large DC current is measured or a powerful magnet is placed close to the clamp core.
1. Fully close the clamps of the sensor and con­firm that the reading has stabilized with no input. Press the 0ADJ key (SHIFT CLEAR
key). “ ” blinks during auto-zero adjust­ment and stops blinking and remains on upon completion.
• Perform auto-zero adjustment even if the read-
ing is zero.
• The adjustable range in current measurement
mode is ±450 dgt in the L range and ±45 dgt. in the H range. If the sensor’s offset exceeds these ranges, auto-zero adjustment cannot be per­formed.
• If auto-zero adjustment is performed during cur-
rent measurement or while the reading fluctu­ates, zero adjustment may not be performed correctly.
• If the 0ADJ key (SHIFT CLEAR key) is pressed
again during auto-zero adjustment ( is blinking), zero adjustment will be canceled.
Page 67
61
Chapter 3 Measurement Procedures
3.5 Data Hold Function
Use this function to hold the current reading. “ ” will light up, and the numerical reading and bar graph will be held. The data hold function is available in any measurement mode.
1. Press the HOLD key; the reading and bar
graph will be held.
2. To cancel, press the HOLD key again.
• If the HOLD key is pressed while current mea-
surements are being output, the output will not be held. While frequency measurements are being output, the output reflects the reading.
3.6 Changing the Screen Update Rate
The screen update rate is set to twice per second when power is turned on. This rate may be changed according to measurement conditions. Every time the SAMPL key is pressed, the indication changes as shown below.
SLOW
3.6.1 SLOW Mode
If the reading fluctuates rapidly, making it hard to read during current or frequency measurement, select a slower screen update rate (approximately once every 3 seconds).
• To meas ure currents of 10 Hz or less, measure-
ment accuracy is guaranteed only when AC+DC A mode is selected and the measurement response speed is set to SLOW.
FAST
NORMAL
• During RMS measurement in AC A mode or in
AC+DC A mode, the measurement response speed (RMS measurement conversion speed) changes according to the screen update rate
(“ ” lights up for approximately 1 second). In SLOW mode, the unit symbol “A” blinks slowly.
Page 68
Chapter 3 Measurement Procedures
62
• During RMS measurement in AC A mode or in
AC+DC A mode, measurement response time is approximately 8 seconds. If there is a transient change in measurement, some time will be required for the reading to stabilize.
3.6.2 FAST Mode
This mo de sets the scree n up da t e ra t e t o ap pro x i mat el y 4 times/second during current or frequency measurement. It is suitable for measurement of abrupt changes such as starting currents. When measuring a starting current, press the MAX/MIN key to use the recording function in order to hold the maximum value (MAX). This will make measurements easier to read.
• When the measurement response speed is set
to FAST during measurement of currents below 45 Hz, measurement accuracy cannot be guar­anteed.
• During RMS measurement in AC A mode or in
AC+DC A mode, the measurement response speed (RMS measurement conversion speed) changes according to the screen update rate
(“ ” lights up for approximately 1 second). In FAST mode, the unit symbol “A” blinks quickly.
• During RMS measurement in AC A mode or in
AC+DC A mode, a continuous waveform of more than 200 ms is required to start current measure­ment. With a shorter waveform, the reading will indicate a value smaller than the actual value.
Page 69
63
Chapter 3 Measurement Procedures
3.7 Recording Function
Use the recording function to hold the maxi mu m measurement, minimum measurement, and mean value of the maximum & minimum.
3.7.1 Measurement Reading
Pressing the MAX/MIN key during current or frequency measurements activates the recording function. The 3290 saves the maximum measurement (MAX), minimum measurement (MIN), and mean value (AVE) in internal memory when the MAX/MIN key is pressed. If the MAX/ MIN key is pressed while the recording function is activated, the indication changes as shown below. When MAX, MIN, AVE, or TIME is not displayed, the reading represents an i nstantaneous value.
The unit for the bar graph is hours.
MAX
Data (MAX, MIN, AVE) will be held while the display is being switched over. However, if the maximum or minimum data is updated in the meantime, the data values will be updated.
When the recording function is activated, the auto power off function is disabled ( off).
The mean value (AVE) is calculated by the following equation: Mean Value = [(Maxim um + Mini mu m)/ 2]
If, after PEAK mode is selected using the PEAK/Hz key, the recording function is activated and an instantaneous value (no indicator) is selected, the transitional peak values can be viewed.
MIN
AVE TIME
Instantan eous value
(No indicator)
Page 70
Chapter 3 Measurement Procedures
64
3.7.2 Display of Elapsed Time
When the MAX/MIN key is pressed to activate the recording function, measurement of elapsed time will begin. The elapsed time is displayed when “TIME” is on. The digital readout can display a maximum of 99 hours 59 minutes. Each block of 100 hours is indicated by the illumination of successive bar graph segments. An elapsed time of up to 2099 hours 59 minutes can be so indicated. When the second segment lights up, 100 hours have elapsed. When the 21st segment lights up, 2000 hours have elapsed.
3.7.3 Interrupting the Recording Function
Press the HOLD key to interrupt the recording function. “ ” will light up, and the elapsed time will stop
incrementing. While the recording function is interrupted, data is not updated even if the clamp sensor is disconnected from the conductor. While the recording function is interrupted, the readings can be changed using the MAX/MIN key. When the HOLD key is pressed again,
” will disappear and recording will resume.
3.7.4 Clearing Recor ded Data
To clear data while the recording function is activated, press the CLEAR key.
Page 71
65
Chapter 3 Measurement Procedures
3.7.5 Disabling the Recording Function
To disable the recording function, press the MODE key during current or frequency measurement. If the recording function is disabled, the auto power off function will be
enabled ( lights up). However, if the output function is on, the auto power off function will not be enabled.
• When minimum and average data is required,
start the recording function during measurement. If the recording function is started when there is no input, the minimum measurements always show zero. To end the recording, press the HOLD key, then end measurement. If the clamp­on sensor is disconnected from the circuit to be measured before the recording function is ended, the minimum measurement will become zero.
• The data will be lost when power is turned off.
• If the recording function is started in the AUTO
range, the measurement range will be set to the range in effect at the time the MAX/MIN key is pressed.
• Use the 9445-02 or 9445-03 (EU) AC adapter
(optional) for long-term recording. If the 3290 is powered by batteries, press the B.CHECK key
(SHIFT MAX/MIN key and check the remain­ing battery power before starting measurement.
Page 72
Chapter 3 Measurement Procedures
66
3.8 Measurement Method Switching Function
In AC A mode, the measurement method is selectable between true rms measurement (RMS) and average rectified rms indication (MEAN). RMS provides accurate measurement of distorted waveforms. MEAN provides accurate measurement of sine waves that are not
distorted. Press the RMS/MEAN key (SHIFT key) to select the measurement method.
3.8.1 RMS Measurement
The true rms measurement uses the rms conversion IC. In addition to distorted waveforms, this method is used for square and triangle waves. In AC+ DC A mode, full-wave rectified and half-wave rectified waveforms are also measured. For this conversion, there is a correlation between frequency characteristics and conversion speed (see 3.9 Measurement Response-Speed Switching Function).
MODE
• In AC+DC A mode, RMS measurement is per­formed.
3.8.2 MEAN Measurem en t
The average rectified rms in dication measurement converts the input waveform to a full-wave rectified waveform using the absolute value detection circuit. The converted waveform is then smoothed by LPF and multiplied by 1.11 for conversion to rms. This method provides accurat e measurement of only sine waves without distortion. If the measurement is nearly the same as the RMS measurement, the waveform will be similar to a sine wave with little distortion. Furthermore, after peak measurement is performed, the ratio of the peak value to
the MEAN measurement is approximately 1:4 ( ), which is even closer to a sine wave.
Page 73
67
Chapter 3 Measurement Procedures
3.9 Measurement Response Speed Switching Function
When RMS is selected in AC A mode or when measurement is performed in AC+DC A mode, the measurement response speed can be changed. Press the
RESPONSE key (SHIFT speed. Pressing the SAMPL key only changes the screen update rate. The measurement response speed will be changed according to the screen update rate.
3.9.1 NORMAL Mode
If the setting is not saved by the setting save function, the 3290 is in NORMAL mode. In NORMAL mode, no indicator is shown on the LCD. The unit symbol “A” remains on without blinking. This mode can be used for almost any measurement.
SAMPL key) to change the
3.9.2 SLOW Mode
When SLOW mode is selected, “ ” lights up for approximately 1 second and the unit “A” blinks slowly. In AC+DC A mode, low frequencies (10 Hz or less) will be measured accurately when SLOW mode is selected. However , due to the slow response speed, the 3290 will be unable to measure sudden changes.
• To meas ure currents of 10 Hz or less, measure­ment accuracy is guaranteed only when AC+DC A mode is selected and measurement response speed is set to SLOW.
3.9.3 FAST Mode
When FAST mode is selected, “ ” lights up for approximately 1 second and the unit “A” blinks quickly. In F AST mode, the 3290 is able to measure sudden changes but cannot measure low frequencies (below 45 Hz) accurately.
• Accuracy is not guaranteed for frequencies
below 45 Hz.
Page 74
Chapter 3 Measurement Procedures
68
8 s
Output Response Waveform (Rise)
FAST
NORMAL
OUT1: MON output waveform
200A
100A
AC A mode 200A range: 0A 180A (45Hz)
0.2 s
OUT1: MON output waveform
200A 100A
OUT2: REC output waveform
OUT2: REC output waveform
SLOW
AC A mode 200A range: 0A 180A (10Hz)
0.8 s
OUT1: MON output waveform
200A
100A
AC+DC A mode 20 0A range: 0A 180A (1Hz)
OUT2: REC output waveform
Page 75
69
Chapter 3 Measurement Procedures
Output Response Waveform (Fall)
8s
FAST
AC A mode 200A range: 200A 20A (45Hz)
NORMAL
OUT1: MON output waveform
OUT2: REC output waveform
0.2 s
OUT1: MON output waveform
OUT2: REC output wavefor m
SLOW
AC A mode 200A range: 200A 20A (10Hz)
0.8 s
OUT1: MON output waveform
OUT2: REC output waveform
AC+DC A mode 20 0A range: 200A20A (1Hz)
Page 76
Chapter 3 Measurement Procedures
70
3.10 Filter Function
The filter function is available in AC A and AC+DC A modes. A low-pass filter (LPF) allows low frequencies to pass through while removing high frequencies. The 3290 sets the cut-off frequency (fc) to approximately 500 Hz. When the filter function is enabled, input current waveforms around 500 Hz can be reduced to –3 dB (attenuation by approx. 30%). (“FILTER” will appear when the filter function is enabled.) Use this function to remove the carrier component of the secondary side of an inverter, or to reduce noi se. Pr ess
the FIL TER key (SHIFT function.
Fundamental: 50 Hz Carrier: 2 kHz
PEAK/Hz key) to enable thi s
Filter OFF
OUT1: MON output
Filter ON
OUT1: MON output
Page 77
71
Chapter 3 Measurement Procedures
3290 Filter Characteristics
5
0
-5
GAIN )dB)
-10
-15 10
100
Input frequency (Hz)
3.11 Key Lock Function
This function locks all keys, including the POWER key, to prevent accidental alteration of the settings. The auto power off function will be disabled.
Press the LOCK key (SHIFT it again to unlock. “ ” will appear when the keys are
locked.
Filter OFF Filter ON
1k
HOLD key) to lock. Press
3.12 Bar Graph Enlargement Function
The bar graph is normally sized according to the measurement range. The bar graph can be enlarged to show 1/10 of the range.
Press the B.GRAPH key (SHIFT enlarge. Press it again to cancel. For example, the L range (20.00-A range when the 9691 clamp-on sensor is used) can be enlarge d to 2-A full scale (f.s .).
RANGE key) to
Page 78
Chapter 3 Measurement Procedures
72
3.13 Battery Check Function
This function uses a bar graph to indicate the remaining battery power. Press the B.CHECK key (SHIFT MAX/MIN key) to display the bar graph. The remaining charge will be indicated as 0%, 25%, 50%, 75%, or 100% for approximately 2 seconds. An indication of 100% means sufficient charge remains. An indication of 0% means battery power is exhausted.
When it falls to 0%, the battery-low warning indicator ( ) will appear. If the unit runs off the 9445-02 or 9445-03 (EU) AC adapter, AC adapter voltage will be indicated instead.
New battery
Reduced to 50%
Reduced to 0. The beep tone sounds 3 times.
lights up.
• After lights up, accuracy cannot be guaran­teed. Replace batteries immediately.
• * If the AC adapter voltage is indicated, the indi­cated value may not reach 100%. However, this does not indicate a problem (when the specified AC adapter 9445-02 or 9445-02 (EU) is used).
Page 79
73
Chapter 3 Measurement Procedures
3.14 MON Output Filter Function
In DC A or AC+DC A mode, when OUT1 is set to MON, output can be obtained through a low-pass filter (LPF) set to approximately 1 Hz. This function will be useful for outputting a DC component only when an AC component or noise is superimposed on the DC component.
Press the MON.FL k ey (SHIFT enable the function. Press it again to disable the function. If this function is enabled in AC+DC A mode, OUT1 outputs the DC component while OUT2 outputs the AC
component (REC). The mode indicator changes from “ to “
“, and the reading shows the AC measurement. If
the function is disabled, the indicator returns to “
• When MON.FL is used, the response speed of OUT1 will be reduced.
OUTPUT key) to
.”
• OUT1 has a 1-Hz filter. If the AC component included in a waveform has is low frequency and large input, the filter may be unable to com­pletely remove the AC component.
• Peak measurement using this function shows the same result as peak measurement in AC A mode (peak measurement of the waveform with the DC component removed).
Page 80
Chapter 3 Measurement Procedures
74
3.15 Auto Power Off Function
When approxima tel y 10 minu tes have ela psed since the last key operation, the power to the 3290 is automatically turned off. This prevents the 3290 from being left on unintentionally and minimizes battery consumption.
When “ ” is displayed, the auto power off function is enabled. “ ” will start blinking and the beep tone will
sound for approximately 30 seconds just before power is turned off. Press any key except the POWER key to extend operation by 10 minutes.
Disabling the Auto Power Off Function
• While holding down the HOLD key, press the POWER key to turn on power (to enable the auto power off function, turn power off, then on again).
• When the setting is saved with the auto power off function disabled, to enable the function, restore the default setting (while holding down the CLEAR key, turn on power).
In the following cases, the power will not be turned off even if the auto power off function is enabled.
• The MAX/MIN key is pressed and the recording function is used.
• The OUTPUT key has been pressed.
• The LOCK key has been pressed.
Page 81
75
Chapter 3 Measurement Procedures
3.16 Setting the Save Function
The 3290 is factory-set as shown below. A setting may be edited (for example) by selecting a frequently used mode and saving it. When the setting is saved, the 3290 will be turned on with the saved setting. The saved setting can be changed. With the power on, set the mode, measurement range and other parameters. Hold down the HOLD key for
approximately 2 seconds (“ ” will appear). To restore the default setting, hold down the CLEAR key and
turn on power (press the POWER key). (“ ” and “ ” will appear.)
Mode Range Output Auto power-off
Screen-
updating
DC AUTO OFF ON NORMAL
The following settings cannot be saved.
• Recording function (Set it after measurement has started.)
• Sensor detection (The 3290 detects the sensor connected when power is turned on.)
• If the setting is saved following auto-zero adjust­ment, the auto-zero adjustment data will be saved. However, be sure to perform auto-zero adjustment before starting measurement in DC A or AC+DC A mode.
• The auto power off setting and the beep tone setting can also be saved. To enable the auto power off function or the beep tone, hold down the CLEAR key and turn on power.
• The Record Function is disabled as soon as the instrument’s power is turnedoff. To enable the Record Function, make the appropriate settings againafter turning on the power.
• If using the 3290 after it has been repaired or
recalibrated, the saved settings may have been deleted . Check that the settings saved prior to repair work are still valid before using the instru­ment; otherwise reset and re-save the settings as appropriate.
Page 82
Chapter 3 Measurement Procedures
76
t
3.17 Battery-Low Warning
If the unit is operated with very low battery power, accuracy cannot be guaranteed. When the battery-low
warning indicator ( fresh batteries.
If the battery-low warning indicator ( long-term current measurement output, it may not be clear which measurements are guaranteed and which are not. If OUT2 is set to B.Lo, around DC 3 V is output if the
battery-low warn i ng indica tor ( output when the indicator is on.
OUT1REC
) appears, replace all batteries with
) is not on, and 0 V is
) appears during
Accuracy is guaranteed.
OUT2: B.Lo (Battery-low warning signal)
If the paper-feed speed is low, the lines of the graph will be thickened.
Accuracy is no guaranteed.
• When the 3290 is running off battery power , per­forming frequency measurements with the cur­rent showing O.L. will increase current consumption and reduce battery life.
• The charge remaining in the battery can be checked using the bar graph immediately after power is turned on or by pressing the B.CHECK
key (SHIFT MAX/MIN key). Please note that the charge indicated is approximate and should only be used as a guide. Exercise caution when performing long-term measurement or when using the recording function.
Page 83
77
Chapter 3 Measurement Procedures
If the power is turned off with the battery-low warning indicator ( ) displayed, when power is turned on again, the battery -low warning indicator () may not light up
immediately. This is because voltage often increases slightly if a battery is not used for some time. The battery-
low warning indicator ( ) will soon light up again. Replace the battery as soon as possible (see Chapter 4 Replacing the Battery).
3.18 Beep Tone
The beep tone sounds once when any key is pressed. It sounds twice when the key operation is not accepted. The key operation is not accepted in the following cases.
• Invalid operation while the recording function is in use
• Invalid operation while is displayed
• Data clear (CLEAR key) when peak measure­ment is not selected
• Filter switching (SHIF T PEAK/Hz key) in DC A mode
• Measurement response speed switching (SHIFT OUTPUT key) in DC A mode
•Auto-zero adjustment (SHIFT CLEAR key) in AC A mode
Disabling the Beep Tone
• While holding down the PEAK/Hz key, press the POWER key (turn power off, then on again to cancel).
• If the setting is saved with the beep tone dis­abled, to enable the beep tone, restore the default setting (hold down the CLEAR key and turn on power).
Page 84
Chapter 3 Measurement Procedures
78
Page 85
79
Chapter 4 Batteries Replacement
Batteries Replacement
To avoid electri c s ho ck when rep lac in g th e
batteries, first disconnect the clamp from the object to be measu red.
After replacing the batteries, replace the
cover and screws before using th e prod­uct.
Do not mix old and new batteries, or differ-
ent types of batteries. Also, be careful to observe b atte ry pol arity during i ns t al lat ion . Otherwise, poor performance or damage from battery leakage could result.
To avoid the possibility of explosion, do not
short circ ui t, d isas sem bl e or i ncine rate bat-
Chapter 4
teries.
Handle and dispose of batteries in accor-
dance with local regulations.
3. Use a Phillip s screwdriver to unfasten the setscrew on the battery cover.
4. Remove the battery cover.
5. Remove the battery.
6. Insert a new batteries while confirm­ing the correct polarity.
7. First fit the tab of the cover, then fit the cover. Secure the cover in place by tightening the setscrew.
Page 86
Chapter 4 Batteries Replacement
80
Page 87
81
Chapter 5 AC Adapter
AC Adapter Chapter 5
Fully insert the optional 9445-02 AC adapter into the AC adapter connection terminal.
AC adapter input voltage range is 100 to 240 VAC (with ±10% stability ) at 50/6 0 H z. To avoid electri ca l ha za rds and damage t o the product, do not apply voltage outside of this range.
For AC adapter operation, use only the Hioki
Model 9445- 02 or 9445-03 (EU ) AC adapter. We cannot accept responsibility for accidents or damage related to the use of any other AC adapter.
The adapter may be used either with or without the batteries. Use of the batteries enables continuous measurement if the AC power source temporarily becomes unavailable, due to a blackout or some other reason. (The duration of measurement backup depends on the remaining battery power.) When using the AC adapter, note that the voltage drops to approximately 6 V. When the battery voltage is higher than the internal voltage (i.e., when using a brand new battery), the battery may be drained. When the battery voltage drops below the internal voltage, the power supply is switched over to the AC adapter.
Page 88
Chapter 5 AC Adapter
82
• When using the AC adapter, note that current of approximately 3 mA is consumed regardless of whether the power switch of the 3290 is on or off. (Battery power is not consumed, however, when power is turned off.)
• When the power supply is switched over from the battery to the AC adapter or vice versa dur­ing peak measurement, an inaccurate measure­ment result (i.e., value larger than the actual value) may be displayed.
• When using the AC adapter, note that the battery indicator (bar graph) indicates the voltage (approx. 6 V) of the AC adapter. Though the voltage may not reach 100%, this poses no prob­lem for measurement operation.
Page 89
83
Chapter 6 Attaching the Strap
Attaching the Strap Chapter 6
Using the 3290 with the strap attached improves operability.
1. Fasten the two straps together into one.
2. 2.Attach each end to the 3290 as shown in the figure above.
Page 90
Chapter 6 Attaching the Strap
84
Page 91
85
Chapter 7 Specifications
Specifications Chapter 7
7.1 Measurem ent Specifications
Measurement Mode
DC A AC A (RMS: True RMS, MEAN: Mean Value) AC+DC A (R MS: True RMS)
Use with Sensor
Connected
sensor
9691
9692
9693
(The specifications vary depending on the maximum continuous input range and maximum peak current of the sensor used.)
• The current measurement ranges of the 3290 are auto range and manual range (L range and H range).
The 3290 automatically detects the sensor connected to it.
Standard Measurement Peak Measurement
Range Resolution
20 A 0.01 A 25.00 A 20 A 0.1 A 50.0 A 100 mV/A
200 A 0.1 A 105.0 A 200 A 0.1 A 150.0 A 10 mV/A
20 A 0.01 A 25.00 A 20 A 0.1 A 50.0 A 100 mV/A 200 A 0.1 A 210.0 A 200 A 0.1 A 300.0 A 10 mV/A 200 A 0.1 A 250.0 A 200 A 1 A 500 A 10 mV/A
2000 A 1 A 2100 A 2000 A 1 A 3000 A 1 mV/A
Maximum displayed
Range Resolution
Maximum displayed
Output
rate
Page 92
Chapter 7 Specifications
86
Measurement Response Speed
The measurement response speed is the time required until the analog output value stabilizes when input is changed from 0% to 90% and from 100% to 10% of the range in true rms measurement (RMS). (1) F AST (0.2 sec) Satisfies the specification at 45 Hz or more. (2) NORMAL (0.8 sec) Satisfies the specification at 10 Hz or more. (3) SLOW (8.0 sec) Satisfies the specification at 1 Hz or more.
Accuracy of the 3290
(23  5  , 73  9 , 80%RH or less, no condensation) Guaranteed accuracy period: 1 year
1. Current Measurement
• In DC or AC+DC mode, the measurement is con­ducted after zero adjustment has been performed.
• FAST, NORMAL, and SLOW indicate the mea­surement response speed.
• When a low-pass filter (500 Hz) is used, the speci­fications shall be satisfied at 60 Hz or less.
(1) DIsplay accuracy (rdg. +dgt.)
DC 1 to 10Hz 10 to 45Hz 45 to 66Hz 66 to 1kHz
AC RMS
AC+DC
DC
FAST
NORMAL
SLOW
(0.3%+5)
(0.3%+7)
(AC +DC
only)
-
(1.0%+3)
(AC +DC
only)
-
(0.3%+3)
--
(0.3%+3) (0.3%+3)
AC MEAN - -
Page 93
87
Chapter 7 Specifications
(2) REC output accuracy (2 V/range) (rdg. +mV)
DC 1 to 10Hz 10 to 45Hz 45 to 66Hz 66 to 1kHz
FAST
AC RMS
AC+DC
NORMAL
SLOW
AC MEAN -
(0.8%+10)
(AC +DC
only)
-
(1.3%+5)
(AC +DC
only)
-
(0.8%+5)  (0.8%+5)
(0.8%+5)
(3) MON output accuracy (2 V/range) (rdg. +mV )
DC 1 to 10Hz 10 to 45Hz 45 to 66Hz 66 to 1kHz
DC
AC+DC
AC - -
(0.3%+5) (0.3%+5)
(0.3%+5) (0.3%+5)  (0.3%+5)
(4) Peak display accuracy (Peak hold function) When sine waves are continuously input after peak reset
L range (rdg. +dgt.)
DC 1 to 10Hz 10 to 45Hz 45 to 66Hz 66 to 1kHz
DC
AC+DC
AC - -
(0.5%+2)  (0.5%+2)
(0.5%+2) (0.5%+2)  (0.5%+2)
H range (rdg. +dgt.)
DC 1 to 10Hz 10 to 45Hz 45 to 66Hz 66 to 1kHz
DC
AC+DC
AC - -
(0.5%+5) (0.5%+5)
(0.5%+5)  (0.5%+5)  (0.5%+5)
Page 94
Chapter 7 Specifications
88
2. Frequency measurement
• When input is 5% or larger of the current range
(1) DIsplay accuracy (rdg. +dgt.)
Range (Accuracy Range) Resolution
10Hz (1.00 to 10.00Hz) 0.01Hz 12.50Hz 100Hz (10.0 to 100.0Hz) 0.1Hz 125.0Hz 1000Hz (100 to 1000Hz) 1Hz 1000Hz
(2) Output accuracy (rdg. +mV)
Range (Accuracy Range) REC
10Hz (1.00 to 10.00Hz) 100Hz (10.0 to 100.0Hz) 1000Hz (100 to 1000Hz)
DC1V/f.s.
Maximum
display
(1.3%+3)
(2.0%+3)
Output response: 4 seconds or less at 1000 Hz and 100 Hz range
(0.3%+1)
(1.0%+1)
6 seconds or less at 10 Hz
7.2 General Specificat ions
Functions
Auto zero adjust­ment
Data hold Press the HOLD key to retain the current
Press the SHIFT CLEAR keys to perform auto zero adjustment (in DC A and AC+DC A modes).
value on the screen.
Page 95
89
Chapter 7 Specifications
Turning the filter on/off
Press the SHIFT PEAK/Hz keys to turn a 500 Hz LPF on and off in AC A and AC+DC A modes. The filter characteristic is -3 dB at 550 Hz ±10%.
Recording Press the MAX/MIN key to select the max-
imum value (MAX), minimum value (MIN), average (AV E), or elapsed time (TIME) to be displayed.
Setting the save function
Hold down the HOLD key (for more than 2 sec.) to save the current settings. To initialize, hold down the CLEAR key and turn on the POWER key.
Switching the method of mea­surement
Expanding the bar graph range
Press the SHIFT MODE keys to select the true rms (RMS) or average rectified rms indication (MEAN) (in AC A mode).
Press the SHIFT RANGE keys to change the bar graph range.
Separate AC/DC output
Press the SHIFT OUTPUT keys to output MON (waveform passed through a 1 Hz LPF) from OUT1 and REC (numeric data coupled with AC) from OUT2. The LCD only shows the AC component (in AC+DC A mode).
Low battery warn­ing output
Press the OUTPUT key (when OUT2 is set to B.Lo) to output 0 V for the low battery warning (normally 3.0 V).
Key lock Press the SHIFT HOLD keys to lock the
keys. Press the same keys again to unlock the keys.
Battery check Press the SHIFT MAX/MIN keys to check
the remaining battery power.
Switching response
Select 0.2 sec. (F AST), 0.8 sec. (NOR­MAL), or 8 sec. (SLOW) when true rms
(RMS) is selected. (Switching mea­surement response speed)
(Interlocked with screen-updating rate in
AC RMS or AC+DC mode. Press the
SHIFT SAMPL keys to set the response
speed independently.)
Page 96
Chapter 7 Specifications
90
Auto power-off Automatic shutdown occurs after 10.5 ±1
minutes. A beep tone warning sounds before shutdown. You can extend the time or disable the function (by holding down the HOLD key and turning power on).
Beep tone ON/OFF (Holding down the PEAK/Hz key
and turning power on switches off the beep
sound.) Display Liquid crystal display (LCD) Digital counter Max. 3000 counts (current)
Max. 1250 counts (frequency) Bar graph display 21 segments Over-range dis-
play
O.L. goes on. When input is outside the
full-scale range, the  bar graph
appears. Low battery warn-
ing
Goes on (during which time accuracy
cannot be guaranteed). Data hold indicator
Auto power-off active indicator
Units A, Hz Zero suppression 5 counts Screen-updating
rate
Digital counter
NORMAL (approx. twice/second)
SLOW (approx. once/3 seconds)
FAST (approx. 4 times/second)
Bar graph display
Approx. 4 times/second Display response
time (The range can be fixed from 0% to 100%.)
Switching the range
Current:
Max. 1 second
Frequency:
Max. 1 second (1000 Hz, 100 Hz range)
Max. 2.5 seconds (10 Hz range)
Selectable from auto range and manual
(fixed) range.
Page 97
91
Chapter 7 Specifications
Circuit dynamic characteristics (crest factor)
Max. 2.5 (The upper limit of H range is the maximum peak current of the sensor con­nected. The maximum output voltage of L
range is 4 V.) Output impedance Max. 100 Ω Temperature char-
acteristics
0 to 40 range: 0.1 x accuracy specifi-
cations/
32 to 104 range: 0.18 x accuracy specifi-
cations/ Location for use Indoor, altitude up to 2000 m (6566 feet) Operating temper-
ature and humid-
0 to 40 , 32 to 104 , 80% RH or less;
no condensation ity range
S torage tempera-
-10 to 50, 14 to 122; no condensation
ture range Power supply Four 1.5 V LR6 batteries or the AC adapter
(optional) Maximum rated
power
500 mVA (running on battery power with a
sensor connected) Battery life Approx. 22 hours (continuous, no load,
with a sensor connected) Applicable stan-
dards
Safety EN61010
EMC EN61326
EN61000-3-2 EN61000-3-3
Effect of con-
Max. 10%f.s. of L range at 3 V ducted radio-fre­quency electromagnetic field
Dimensions Approx . 155W x 98H x 47D mm
Approx. 6.10"W x 3.86"H x 1.85"D
(excluding protruding parts) Mass Approx. 545 g (19.2 oz.) (including battery) Accessories Strap 1
LR6 batteries 4
Instruction manual 1
Page 98
Chapter 7 Specifications
92
Options 9445-02 AC ADAPTER
(SA10-0910N, SINO-AMERICAN) 9445-03 AC ADAPTER (EU) (SA10-0910G, SINO-AMERICAN) 9094 OUTPUT CORD 9400 CARRYING CASE 9691 CLAMP ON AC/DC SENSOR 9692 CLAMP ON AC/DC SENSOR 9693 CLAMP ON AC/DC SENSOR
Page 99
93
Chapter 8 Clamp Sensor Specifications
Clamp Sensor Specifications
Chapter 8
8.1 9691 CLAMP ON AC/DC SENSOR
Rated current AC/DC 100 A Output voltage AC/DC 1 V/100 A Output resistance 1 00 Ω or less Maximum permissible
input Maximum peak current
value Accuracy guarantee for
temperature and humidity
100 Arms continuous (Derating according to frequency)
150 A
23°C±5°C(73°F± 9°F), 80%RH or less, no condensation
Guaranteed accuracy period
(1) Continuous input
Range DC to 66Hz 66 to 500Hz
0 to 80A
1.0%rdg. 0.5mV1.0%rdg. 0.5mV
80 to 100A
(2) Peak input
Range DC to 66Hz 66 to 500Hz
0 to 110A
1.0%rdg. 2mV 1.0%rdg. 2mV
110 to 150A
1 year (Opening and closing of the sensor: 10,000 times)
2.0%rdg.  0.5mV2.5%rdg.  0.5mV
2.0%rdg.  2mV2.5%rdg.  2mV
Page 100
Chapter 8 Clamp Sensor Specifications
94
Frequency band DC to 10 kHz (-3dB) Effect of conductor
Within ±1.0% (at 80 A (55 Hz))
position Effect of external
electromagnetic field Temperature character-
istics
0.5 A equivalent or less (in an exte rnal electromagnetic field of 400 A/m)
0 to 40 range: 0.1 x accuracy specifications/ 32 to 104 range: 0.18 x accuracy specifications/
Operating temperature and humidity range
Storage temperature range
0 to 40 , 32 to 104 , 80% RH or less; no condensation
-10 to 50, 14 to 122, 80% RH or less; no condensation
Location for use Indoor, altitude up to 2000 m (6566
feet)
Maximum permitted
600 V
circuit voltage Dielectric strength 5.55 kV AC for 1 minute (sensor -
case, sensor - circuit)
Power supply ± 5 V ±0.25 V (supply from the out-
side) Maximum rated power 50 mVA Measurable conductor
φ35 mm (1.38") or less diameter
Dimensions Approx. 53W x 129H x 18D mm
Approx. 2.09"W x 5.08"H x 0.71"D
(excluding protruding parts) Mass Approx. 230 g (8.1 oz.) (including bat-
teries) Cable length Approx. 2 m (6.5 feet) Accessory Instruction manual 1
Loading...